Entry  Fri Aug 11 09:11:04 2023, Annie, Pulse Settings IMG_2908.JPGIMG_2909.JPGIMG_2910.JPGIMG_2911.JPG
 
Entry  Fri Aug 11 10:12:02 2023, Annie, Pressure, voltage and current check IMG_2915.JPGIMG_2916.JPGIMG_2918.JPG
 
Entry  Fri Jul 18 12:04:36 2025, CLW, Presence at CERN 

Excel sheet for presence at CERN during 41Ca run. Please update if required. 

https://docs.google.com/spreadsheets/d/19OzH7bpnHb-YeOskNBGlrhEkPZmDsPDw9y0tleU0Soo/edit?gid=0#gid=0

Entry  Tue Oct 13 21:04:49 2015, Claudia Lederer, Photos of Test Setup 5-10 October 2015 DSC00042.JPGDSC00048.JPGDSC00056.JPGDSC00065.JPG

some photos in high resolution of the test setup 5-10 October 2015

Entry  Thu Aug 10 09:55:50 2023, CLW, Photos of Chamber, Al26 Run Aug23 13x
 
Entry  Thu Aug 10 10:04:13 2023, CLW, Photos inside chamber Al26 run Aug2023 8x
 
Entry  Thu Aug 10 16:11:57 2023, TD, NS, AR, Noise 20230810_162959.JPGIMG_2906.jpeg20230810_164219.JPG
Observe c. 2mV p to p noise with DSO ( Z_in = 50 Ohm ) c. 60us period with HF structure. DSO connected to junction box via 34-way IDC - 16x Lemo-00 adaptor.

Origin of noise upstream of 4x34-way to 8x16-way Junction Box. Not microphonics from Edwards RV5 Rotary Pump.
No change observed with simple ground connections between NIM bin/+/-15V PSU/Junction Box and MSL type W1 preamplifier units/vacuum chamber/support assembly.


Estimate of electronic noise

Pulser BNC PB-5

Amplitude 0.5V
Attenuation x1
Decay time 1ms
Frequency 50Hz

Preamplifier RAL108
Output impedance 100 Ohm
Sensitivity 20mV/MeV ( into high Z load ), 6.7mV/MeV ( into 50 Ohm load ) 

Amplifier EG&G Ortec 571
Input terminated by 50 Ohm
Gain x1 (internal) x 1.0 (fine gain) x 50 (coarse gain ) = 50
Shaping time 0.5us

MCA Amptek 8000D
Input FSR 10V
12 bit ADC


Nominal gain = 6.7mV/MeV x 50 = 335mV/MeV

12 bit ADC input FSR = 10V / 0.335V/MeV = 29.85MeV FSR or 7.3keV/channel


dE p+n junction strip # 4 ( of 0-15 )

pulser peak centroid = 799.8 ch
pulser peak width = 7.8 ch FWHM = 57 keV FWHM


E p+n junction strip # 4 ( of 0-15 )

pulser peak centroid = 864.8ch
pulser peak width = 3.5 ch FWHM = 26 keV FWHM


E n+n Ohmic strip # 4 ( of 0-15 )

pulser peak centroid = 913.0 ch
pulser peak width = 5.3 ch FWHM = 39 keV FWHM


Noise estimates are probably accurate to c. 10% level.
    Reply  Fri Aug 11 10:02:08 2023, TD, NS, AR, Noise 20230811_104233.jpg20230811_104253.jpg
Check RAL108 +/-15V PSU

Measured output voltages +15.21V -15.18V - OK

Observe output voltages with DSO ( ch #1 AC/1M +15V, ch #2 AC/1M -15V ) - see attachments 1 & 2

What we should observe is c. 1mV rms ( white ) noise but we clearly observe similar noise transients ( c. 60us period with HF structure ) to those observed at RAL108 
outputs. The RAL108 preamplifier units do have RC filters on the +/-15V - typically c. 100 Ohm and 4.7uF. Some additional, inline filtering with a lower rolloff 
frequency may be required.

> 
> Observe c. 2mV p to p noise with DSO ( Z_in = 50 Ohm ) c. 60us period with HF structure. DSO connected to junction box via 34-way IDC - 16x Lemo-00 adaptor.
> 
> Origin of noise upstream of 4x34-way to 8x16-way Junction Box. Not microphonics from Edwards RV5 Rotary Pump.
> No change observed with simple ground connections between NIM bin/+/-15V PSU/Junction Box and MSL type W1 preamplifier units/vacuum chamber/support assembly.
> 
> 
> Estimate of electronic noise
> 
> Pulser BNC PB-5
> 
> Amplitude 0.5V
> Attenuation x1
> Decay time 1ms
> Frequency 50Hz
> 
> Preamplifier RAL108
> Output impedance 100 Ohm
> Sensitivity 20mV/MeV ( into high Z load ), 6.7mV/MeV ( into 50 Ohm load ) 
> 
> Amplifier EG&G Ortec 571
> Input terminated by 50 Ohm
> Gain x1 (internal) x 1.0 (fine gain) x 50 (coarse gain ) = 50
> Shaping time 0.5us
> 
> MCA Amptek 8000D
> Input FSR 10V
> 12 bit ADC
> 
> 
> Nominal gain = 6.7mV/MeV x 50 = 335mV/MeV
> 
> 12 bit ADC input FSR = 10V / 0.335V/MeV = 29.85MeV FSR or 7.3keV/channel
> 
> 
> dE p+n junction strip # 4 ( of 0-15 )
> 
> pulser peak centroid = 799.8 ch
> pulser peak width = 7.8 ch FWHM = 57 keV FWHM
> 
> 
> E p+n junction strip # 4 ( of 0-15 )
> 
> pulser peak centroid = 864.8ch
> pulser peak width = 3.5 ch FWHM = 26 keV FWHM
> 
> 
> E n+n Ohmic strip # 4 ( of 0-15 )
> 
> pulser peak centroid = 913.0 ch
> pulser peak width = 5.3 ch FWHM = 39 keV FWHM
> 
> 
> Noise estimates are probably accurate to c. 10% level.
Entry  Thu Sep 18 14:11:29 2025, CLW, EO, New processing  Screenshot_2025-09-18_at_12.27.35.pngnewprocessing.pdf

Comparison strip 6 amplitude spectra anew (red) vs (old) processing (Screenshot). Slight shift in amplitude and slightly less counts with new processing in t+alpha peaks (1% level). This could be either due to less noise background, shift in neutron energy calibration (amplitude spectra shown with a cut of 100 keV). pdf plot shows new vs old for neutron energy vs counts. only minor differences at low neutron energy, new processing makes spectra cleaner at higher energy.

Shape of the Li 240 keV resonance is shifted in new processing - either due to different offset in tof-ene calibration, or time stamp for signals in one of the processing versions is wrong. Need a new comparison to simulated spectra. 

Entry  Sat Sep 14 09:30:40 2024, TD, Neutron beam profile  20240914_101751.jpg
Neutron beam profile - film exposure overnight 12/13.9.24
Entry  Tue Aug 15 11:41:27 2023, TD, NS, Monday 14 August - vacuum chamber pressure 20230814_083514.jpg
08.35 Vacuum chamber pressure OK - see attachment 1
Entry  Thu Dec 7 09:48:23 2023, Emmanuel, MWDdetector and User Input UserInput_Silicons.hMWDdetector.cc

MWDdetector and UserInput

Entry  Tue Apr 16 12:24:01 2024, Emmanuel, MWD amplitude (dE and E) 26Al_PDF.pdf
 
Entry  Thu Jul 27 11:53:52 2023, TD, MSL type W1 detector data for 26Al(n,X) experiment EAR2, n_TOF 
MSL type W1(SS)-20

3186-9  20um Depletion voltage 3V Operating voltage 6V
3585-12 20um Depletion voltage 2V Operating voltage 4V

MSL type W1(DS)-150

3458-1 157um Depletion voltage 22V Operating voltage 52V
3458-4 144um Depletion voltage 18V Operating voltage 48V
Entry  Mon Aug 4 16:32:56 2025, Selin, Lithium Consistency Check Li1-8.pngLi9-16.pngLi17-24.pngLi25-32.png

These are the consistency checks done for Lithium sample.

Please note that in the earlier runs, the cables 22 and 23 were switched.

Entry  Thu Jul 28 16:00:43 2016, Sarah, List of to-dos 

Process alpha data
- Compare pulser width and alpha width to calibration spectra (shaping amp with 1us and gain 50x)
- Does photodiode data process OK or is it a problem with the oscillations?

Process LiF data
- Any pile up issues? Problems with gamma flash?

 

Online monitoring:
- Check leakage every few days, check lab temp.
- Check alpha peaks of 26Al contaminants - Gd, B.
 

Entry  Wed Jul 16 13:56:00 2025, CLW, Links to photos and proton counting excel sheet 

Photos setup

Excel Sheet 

Entry  Thu Sep 28 15:32:16 2023, Emmanuel, LiF sample in 20230928_154334.jpg20230928_155923.jpg20230926_115053.jpg20230928_155747.jpg

We switched from 10B target (number 1) to LiF target (number 3). We planned to take this measurement for one day.

Entry  Fri Aug 11 09:17:28 2023, Annie, Li target scans 7Li_foil_side1_post_processed.pdf7Li_foil_side2_post_processed.pdfIMG_2914.JPG

Processed scans of both sides of the Li target showing the beam spot, and an image showing what it looks like to the eye.

Entry  Tue Aug 2 08:14:26 2016, Sarah, Leakage currents spreadsheet (26/07- 01/08) Leakage_currents.xlsx

 Steady for dE, E. Possibly rising on PHDI- monitor closely.