  157   Mon Feb 22 15:19:52 2016 PJCST9 System changed to version 8

I have changed the T9 system to be using Version 8.

Tests using the detector and pulser seem to be ok.

Transmission to the Merge and Merging operating ok

Shaping time 8uS and threshold 64.

Note:-

The Leading Edge Discriminator has changed in Version 8 and is a true implementation of a Leading Edge Discriminator. With hysteresis which requires the input signal to go below the threshold before allowing another trigger.

The Delay setting on the LED web page is removed.

The LED web page is aware of the version being used ( shown at the top of the page ) as the Threshold for the LED is an ADC value ( 0 to 16383 ) rather than a difference from the current baseline.

For a positive waveform where the baseline is expected to be about 8192 a value of 9000 will potentially give a trigger.

For a negative waveform where the baseline is expected to be about 8192 a value of 7000 will potentially give a trigger.

Example window attached.

I would value feedback.

 

  222   Thu May 12 15:50:48 2016 PatrickSystem Checks and firmware update
Added two checks to the System wide checks. Modified the Firmware so the checks work properly. One check reads the status of the FEE64 clocks and checks against expected value. The other check reads the value from all the FEE64 Sync Error Counters, pauses for one second, then reads them again and compares with the first reading. Any change is reported as an error. The Sync Error counter is incremented if the SYNC pulse arrives at a FEE64 when the Timestamp counter in the FEE64 is not at the correct value.
  523   Sun Dec 4 05:46:42 2016 TDSyncCheck RIBF03R1 & RIBF123R1 commissioning tests archived
The SyncCheck etc software used for the RIBF123R1 and RIBF03R1 commissioning
tests on d05 (user aida) has been archived as

aidas1:/homes/npg/SyncCheck/SyncCheck-d05-041216.tar.gz
aidas1:/homes/npg/SyncCheck/DataPackage-d05-041216.tar.gz
aidas1:/homes/npg/SyncCheck/ribfts-d05-041216.tar.gz
  380   Sun Oct 9 11:47:34 2016 CG, TDSunday 9th October
12.30 Following scheduled power outages of October 8

      power up AIDA and Julabo FL11006 recirculating chiller

      Julabo FL11006 set point +20 deg C
      Temperature +25.3 deg C, d.p. +10.6 deg C, RH 39.3%
     
      boot aidas1 
      aidas1 filesystem checks (fsck) run from c. 12.30 to 15.58

      System wide checks OK *except* nnaida21 failed ADC calibration

16.55 DAQ starts (no data transfer)

      ASIC settings 2016Sep02-18.17.01
      slow comparator threshold 0x10 -> 0x0a
      TS_SYNC_PHASE 0

      DSSSD bias & leakage current OK - see attachments 8 & 9
      FEE64 temperatures OK - see attachment 7

17.30 DAQ stops

      DSSSD bias off
      AIDA withdrawn from (CH2)n moderator 
      DSSSD bias on

17.40 DAQ starts

19.15 Noticed DAQ stopped producing SYNC pulses at some point.
      A reload of the run control tab sees them all as going still, but only 20% are reporting any SYNC output in the stats page.
      All FEEs have failed the SYNC errors since baseline check and the values reported by the SYNC pulses received tests vary
      by +- 3 of a central value.
      See attachments 1-6.
      The merger and tape server were not set up before this run and are not currently running.

      TS_SYNC phase changed from 0 -> 1 and ReSYNC

22.20 DAQ OK

      AIDA SYNCs OK
      All system-wide checks OK *except* nnaida21 fails ADC calibration (no change)
  834   Sun Jun 9 03:43:00 2019 CB, TD, CG, OHSunday 9 June
11.43 DAQ continues OK

      ASIC settings 2019May19-17.54.03
      slow comparator 0xa

      BNC PB-4 Pulser 
      Amplitude 90,000
      x5 attenuator IN
      Frequency 25Hz range, variable min. c. 2.5Hz
      t_r 100ns, tau_d 50us
      - polarity
      Delay MIN, tail pulse

      all system wide checks OK *except* nnaida7 fails ADC calibration

      detector biases & leakage currents OK - see attachments 1 & 2

      FEE64 temperatures OK - see attachment 3

      good event statistics - DSSSD#1 noisy - see attachments 4
      10x <20k, 19x <50k, 20x <100k cf. 13x <20k 22x < 50k 24x <100k yesterday afternoon
      Merger OK - see attachment 5
      TapeServer OK - see attachment 6

      p+n FEE64s 1.8.W spectra 20us FSR - DSSSD#1 noisy - attachment 7
      n+n FEE64s 1.8.W spectra 20us FSR - DSSSD#1 noisy - attachment 8
      1.8.L spectra - see attachments 9 & 10
      nnaida1 76 ch FWHM, nnaida3 83 ch FWHM - cf. 104 & 111 ch FWHM respectively yesterday aftenoon

      FEE64 Rate spectra - attachment 11
      Check ASIC Control *all* FEE64s & ASICs x2
      FEE64 Rate spectra - attachment 12

11.46 ADC data synchronised - attachment 13

      p+n FEE64s 1.8.W spectra 20us, 200us, 2ms & 20ms FSR - DSSSD#1 noisy - attachments 14-17

      Check ASIC Control *all* FEE64s & ASICs x2   - no change DSSSD#1 noise

      DSSSD#1 bias cycle - no change DSSSD#1 noise

      press Kapton Samtec into adaptor PCBs - reduces DSSSD#1 noise to previous levels

12.17 Julabo FL11006 chiller set point +20 deg C actual +20.0 deg C
      Water level c. 60% M-H markers
      Temperature +25.3 deg C dew point +7.6 deg C RH 32.2%

12.23 temporary fix for DSSSD#1 noise? check by
      zero all histograms
      zero DAQ stats

      ...and wait

      good event statistics - DSSSD#1 noisy - see attachments 18
      11x <20k, 21x <50k, 23x <100k

13.30 DAQ stop

      OH replaces LH kapton PCB for DSSSD#1 n+n FEE64s 20 & 23 which we suspect may be damaged due to twist
      placed on kpaton cables by weight of all ribbon cables + cu foil shielding - particularly pronounced
      for nnaida20

      good event statistics - OK - see attachments 19
      7x <20k, 20x <50k, 24x <100k
      p+n FEE64s 1.8.W spectra 20us FSR - OK - attachment 20

      good event statistics - OK - see attachment 21
      7x <20k, 22x <50k, 24x <100k
      p+n FEE64s 1.8.W spectra 20us FSR - OK - attachment 22

17.50 detector biases & leakage currents OK - see attachments 23 & 24

      FEE64 temperatures OK - see attachment 25

      good event statistics - OK - see attachments 26
      9x <20k, 21x < 50k, 24x <100k cf. 10x <20k, 19x <50k, 20x <100k earlier today

      p+n FEE64s 1.8.W spectra 20us FSR - DSSSD#1 noisy - attachment 7
      n+n FEE64s 1.8.W spectra 20us FSR - DSSSD#1 noisy - attachment 8

17.57 all histograms zero'd
      DAQ stats zero'd
  900   Sun Oct 8 14:51:59 2023 TDSunday 8 October contd.
Slow comparator 0xf

DSSSD HV +120V +6.400uA - attachment 3

per FEE64 Rate spectra - attachment 1
ADC data item stats - attachment 2

test + disconnected at start of chain aida10, aida1 ... aida12

ADC data item stats - attachment 4
per FEE64 Rate spectra - attachment 5

No difference observed

DSSSD HV +130V +6.455uA - attachment 7
DSSSD HV +160V +6.605uA - attachment 6

per FEE64 1.8.W spectra - 2ms FSR - attachments 8 & 9
per FEE64 Rate spectra - attachments 10 & 11
ADC data item stats - attachment 12

Depletion+30V bias = 160V - no change to n+n noise, increased noise more generally


17.30 power off, HV off
      test + daisy chain removed from aida10, aida01 ... aida12
      power on
      slow comparator 0x64

      DSSSD bias +120V +6.455uA - attachment 13

      ADC data item stats - attachment 14

      per FEE64 Rate spectra - attachments 15 & 16
  
      1.8.W spectra - 20us & 200us FSR - attachments 17 & 18

      extrinsic noise worse but restricyted to a few hot channels
      
      slow comparator 0xf
      per FEE64 Rate spectra - attachment 19


18.40 DSSD HV to aida04 ( standard coaxial cable ), N1419ET floating output
      HV chain aida12, auda03, aida11 
      aida03 LK1 and LK3 fitted
      slow comparator 0x64

      DSSSD bias +120V +6.435uA - attachment 20
      ADC data item stats - attachment 21

      per FEE64 Rate spectra - attachments 22-24

      1.8.W spectra - 200us FSR - attcahments 25 & 26

       extrinsic noise worse esp. n+n Ohmic side
       p+n rates dominated by hot channels
  899   Sun Oct 8 11:44:35 2023 TDSunday 8 October
12.22 USB-controlled ac mains relay - interlock overidden
      Water flow OK. Water temp 19 deg C OK.
      Power on.

      aida03 LK3 fitted

      test + chaine to all p+n FEE64s
      test - disconnected at EG&G Ortec 534 output and aida04 n+n FEE64 adaptor card

      CAEN N1419ET floating output
      Bias chain braid - aida11, aida03 & aida12, core aida04
      Bias +60V +5.28uA stable - attachment 1

      FEE64 temps OK - attachment 2

      1.8.W spectra - 200us FSR - attachments 3 & 4
      per FEE64 p+n 1.8.L spectra - attachment 5
       aida01 pulser peak width 240 ch FWHM

      Slow comparator 0x64

      per FEE64 Rate spectra - attachment 6
       p+n rates dominated by hot channels
      ADC data item stats - attachment 7

12.58 Bias +100V +6.075uA stable - attachment 8

     1.8.W spectra - 200us FSR - attachments 9 & 10
      per FEE64 p+n 1.8.L spectra - attachment 11
       aida01 pulser peak width 116 ch FWHM

      Slow comparator 0x64

      per FEE64 Rate spectra - attachment 12
       n+n rates dominated by hot channels, no p+n hot channels
      ADC data item stats - attachment 13

     1.8.W spectra - 20us FSR - attachments 14 & 15

     1.8.W spectra - 2ms FSR - attachments 16-19

     Observe 100kHz extrinsic noise with 5kHz modulation

13.22 Bias +100V +5.995uA stable - attachment 20

13.36 Bias +100V +6.040uA stable - attachment 25

     per FEE64 Rate spectra - attachment 21
       n+n rates dominated by hot channels, no p+n hot channels
      ADC data item stats - attachment 22

     slow comparator 0x64 -> 0xf

     per FEE64 Rate spectra - attachment 23
     ADC data item stats - attachment 24

13.36 Bias +110V +6.175uA stable - attachment 30

     slow comparator 0xf -> 0x64

     1.8.W spectra - 2ms FSR - attachments 26 & 27

     per FEE64 Rate spectra - attachment 29
      aida02, aida04 & aida10 rates dominated by hot channels
     ADC data item stats - attachment 28

13.36 Bias +120V +6.295uA stable - attachment 36

     slow comparator 0x64

     1.8.W spectra - 2ms FSR - attachments 33 & 34

     per FEE64 p+n 1.8.L spectra - attachment 31
      aida01 pulser peak width 120 ch FWHM

     per FEE64 Rate spectra - attachment 32
      aida02, aida04 & aida10 rates dominated by hot channels

     ADC data item stats - attachment 35


15.07 aida03 LK1 fitted
      Bias +60V +5.600uA stable
      Bias +100V +6.260uA stabl

      slow comparator 0x64

      1.8.W spectra - 2ms FSR - attachments 37 & 38

      per FEE64 p+n 1.8.L spectra - attachment 40
       aida01 pulser peak width 115 ch FWHM

      per FEE64 Rate spectra - attachment 39
       aida02, aida04 & aida10 rates dominated by hot channels

      ADC data item stats - attachment 41

      DSSSD HV - attachment 42




    per FEE64 Rate spectra - attachment 39
       aida02, aida04 & aida10 rates dominated by hot channels

      ADC data item stats - attachment 41



      


      
      
  557   Sun May 7 01:29:50 2017 OH, TDSunday 7 May
09.28 DAQ continues OK 
      Current file May2017/R1_36

      ASIC settings 2017Mar27-16.12.39 *except*
      slow comparator 0x64
      LEC/MEC fast comparator 0xff

      BNC PB-4 pulser OFF

      Detector biases & leakage currents OK - see attachments 1 & 2
      FEE64 Temperatures OK - see attachment 3
      'good events' statistics OK - see attachment 4
      No warnings/errors reported by merger since DAQ GO - see attachment 5

      *all* system wide checks OK

09.40 1.8.L ADC spectra - see attachments 6 & 7 
      1.8.W waveforms - see attachments 8-11

10.00 E17/F11 ambient temperature +24.3 deg C, RH 39.4%, d.p. +9.6 deg C

18:11 System wide checks all ok
      Leakage and bias all ok - see attachment 12
      Fee temperatures all ok - see attachment 13
      LN2 Refill cycle has been running the past few days and we have encountered no issues
      AIDA daq continues to run fine

19:20 Ambient temperature 24.5 Humdity 38.3% D.P +9.2
  698   Sat Nov 4 23:49:14 2017 OHSunday 5 November
08.09 DAQ running OK - data transfer not enabled

      BNC PB-4
      Amplitude 90,000
      x5 attenuator IN
      t_r 100ns tau_d 50us
      - polarity
      frequency range 25Hz
      delay MIN, tail pulse, INT ref

      ASIC settings 2017Oct23-14.38.03
      *all* FEE64s slow comparators 0xa *except* nnaida19-22 0xc

      Detector biases & leakage currents OK - see attachments 1
      FEE64 Temperatures OK - see attachment 2
      DAQ statistics OK - see attachment 3
      *all* system wide checks OK *except* nnaida21 fails ADC calibration

13:35 DAQ running ok
      Fee temperatures - attachment 4
      Stats - attachment 5
      Leakage currents and bias ok - Attachment 6

13:39 Bias taken off detectors
      Server windows closed and servers halted
      FEE power relays turned off
      Chiller remains on
  277   Sun Jun 5 02:14:10 2016 TD, AE, CG & DKSunday 5 June 2016
10.10 Detector biases & leakage currents - OK - see attachments 1-2 

10.40 Detector biases & leakage currents - OK - see attachments 3-4
      30 minutes ~2000Hz highenergy heavy ions during BigRIPS tuning

17.35 Detector biases & leakage currents - OK - see attachments 5-6
      Rates DSSSD #1 - #6 ~500Hz per DSSSD

19.30 AIDA SYNCs lost
      FEE64s power cycled and rebooted

22.30 Detector biases & leakage currents - OK - see attachments 7-8
      Rates DSSSD #1 - #6 ~1kHz per DSSSD

23.00 2mm plate removed from fixed degrader, 1mm installed => net removal of 1mm
      Variable degrader No. 2 in beam, all others out.
  356   Sun Sep 4 04:26:23 2016 TDSunday 4 September
14.20 R13 continues

      DSSSD bias/leakage current OK - see attachments 1-2
      FEE64 temperatures OK - see attachment 3

      System-wide checks OK *except* nnaida19 fails ASIC timestamp check (no change)

      Merger errors messages for Q 19 - see attachment 4

     AIDA SYNC, PAUSE, RESUME, Correlation L, SYNC time warp, Good Events, AIDA disc, nnaida6, nnaida5 &
nnaida20 statistics - see attachments 6-16

     Analysis of R16_331 - FEE64 data synchronised - see attachment 17

aidas1> ./analyser2 /TapeData/Aug2016/R16_331
 *** GREAT format 3.2.0 analyser - TD - May 2014
 *** ERROR: READ I/O error:       5002
                   blocks:      32000
          ADC data format:  182657196 ( 1048839.0 Hz)
        Other data format:   78430804 (  450358.8 Hz)
 Sample trace data format:          0 (       0.0 Hz)
         Undefined format:          0 (       0.0 Hz)
   Other data format type:      PAUSE:         94 (       0.5 Hz)
                               RESUME:         95 (       0.5 Hz)
                              SYNC100:    1591884 (    9140.8 Hz)
                           FEE64 disc:   76825108 (  441138.8 Hz)
                             MBS info:      13623 (      78.2 Hz)
                           Other info:          0 (       0.0 Hz)

   ADC data range bit set:       4398 (      25.3 Hz)

                Timewarps:        ADC:          0 (       0.0 Hz)
                                PAUSE:          0 (       0.0 Hz)
                               RESUME:          0 (       0.0 Hz)
                              SYNC100:          0 (       0.0 Hz)
                           FEE64 disc:          0 (       0.0 Hz)
                             MBS info:          0 (       0.0 Hz)
                            Undefined:          0 (       0.0 Hz)
                         Sample trace:          0 (       0.0 Hz)

 Timestamp elapsed time:      174.152 s

 FEE module #:   1 elapsed dead time       0.189 s
 FEE module #:   2 elapsed dead time       0.705 s
 FEE module #:   3 elapsed dead time       0.000 s
 FEE module #:   4 elapsed dead time       0.000 s
 FEE module #:   5 elapsed dead time       3.321 s
 FEE module #:   6 elapsed dead time       0.467 s
 FEE module #:   7 elapsed dead time       0.000 s
 FEE module #:   8 elapsed dead time       0.000 s
 FEE module #:   9 elapsed dead time       0.124 s
 FEE module #:  10 elapsed dead time       0.000 s
 FEE module #:  11 elapsed dead time       0.000 s
 FEE module #:  12 elapsed dead time       0.000 s
 FEE module #:  13 elapsed dead time       0.284 s
 FEE module #:  14 elapsed dead time       0.035 s
 FEE module #:  15 elapsed dead time       0.000 s
 FEE module #:  16 elapsed dead time       0.000 s
 FEE module #:  17 elapsed dead time       0.684 s
 FEE module #:  18 elapsed dead time       0.012 s
 FEE module #:  19 elapsed dead time       0.000 s
 FEE module #:  20 elapsed dead time       0.000 s
 FEE module #:  21 elapsed dead time       0.000 s
 FEE module #:  22 elapsed dead time       0.158 s
 FEE module #:  23 elapsed dead time       0.000 s
 FEE module #:  24 elapsed dead time       0.000 s
 FEE module #:  25 elapsed dead time       0.000 s
 FEE module #:  26 elapsed dead time       0.000 s
 FEE module #:  27 elapsed dead time       0.000 s
 FEE module #:  28 elapsed dead time       0.000 s
 FEE module #:  29 elapsed dead time       0.000 s
 FEE module #:  30 elapsed dead time       0.000 s
 FEE module #:  31 elapsed dead time       0.000 s
 FEE module #:  32 elapsed dead time       0.000 s

 *** Program elapsed time:   27.594s ( 1159.683 blocks/s,  72.480 Mb/s)


14.29 R16 stops


15.30 Noise measurements
      Found incorrect test pulse polarity connected to p+n and n+n FEE64s - corrected - see attachment 18

DSSSD #1 FEE64s, shaping time(us) versus pulser peak widths ch FWHM

      17  22  20  23 spectrum 1.8.L 
0.5  180 202 130 104
1    147 177 121 128
2    110 164 116 128
2.5   97 154 103 131
3.5   82 123  93 110
4.5   75 105  88 104
5.5   71  93  86  99
6     71  90  84  97
6.5   69  91  87  94
7     70  87  80  94
7.5   70  84  86  93
8     71  87  87  92
      
      System-wide checks OK *except* nnaida19 fails ASIC timestamp check (no change)

16.26 R17 starts
      Pulser + 207Bi source

      ASIC settings 2016Sep02-18.17.01
      Discriminator masks -> 0-31 0x0 32-63 0x800000
      Correlation scaler trigger delay 0x180 -> 0x5
      Slow comparator 0x8

      Analysis of R17_0 - FEE64 data synchronised - see attachment 19

      top - attachment 20
      merger 1.9M items/s tapeserver 15Mb/s
  762   Sun Nov 4 01:11:41 2018 CA, CB, TD, OH, PJWSunday 4 November
10.03 DAQ continues OK - current file Oct18/R16_9
      FEE64 LEC/MEC slow comparator threshold 0x64
      beam background

10.05 all system wide checks OK *except* nnaida7 fails ADC calibration

      detector biases & leakage currents OK - see attachments 1 & 2

      FEE64 temperatures OK - see attachment 3
      good event statistics OK *except* nnaida2 - see attachment 4
      nnaida2 rate due to one hot channel as before

      Merger OK - see attachment 5
      TapeServer OK - see attachment 6
      no errors or warnings shown by server terminal sessions 

  
      1.8.L spectra - see attachments 7 & 8

       N.B. evidence of minimally ionising particles (presumably protons) from beam to F11 E detector
            ~10-15cm Pb bricks were placed within BRIKEN (CH2)n stack to stop heavy particles

      1.8.W spectra 20us FSR - see attachments 9 & 10

      online analysis with MIDASsort

      attachment 11 - per DSSSD m_p versus m_n
                 12 - per DSSSD x versus y
                 13 - per DSSSD E_p versus E_n
                 14 - channel versus E_p
                 15 - hit pattern (ASIC & DSSSD order)

14.35 DAQ continues OK - current file Oct18/R16_11
      FEE64 LEC/MEC slow comparator threshold 0x64
      beam background

14.35 all system wide checks OK *except* nnaida7 fails ADC calibration

      detector biases & leakage currents OK - see attachments 15 & 16

      FEE64 temperatures OK - see attachment 17
      good event statistics OK *except* nnaida2 - see attachment 18
      nnaida2 rate due to one hot channel as before

14.51 analysis file Oct18/R16_10 - attachment 19
  629   Sun Jun 4 00:06:18 2017 OH, TD, PJWSunday 4 June 08.00-16.00
08:00 DAQ found stopped at around 04:49 - Attachment 1
      Time obtained from last file written as part of R1
      Tape server still running - Attachment 2
      Merger terminal contents - Attachment 3
      FEE output from monitor for nnaida6 - Attachment 4

      DAQ still passed all system wide checks after the crash

08:22 Merger recovered.
      DAQ start
      AIDA file RIBF148/R2

      BNC PB-4
      Amplitude 90,000
      x5 attenuator IN
      t_r 100ns tau_d 50us
      - polarity
      frequency 25Hz range
      delay MIN, tail pulse, INT ref

      ASIC settings 2017Mar27-16.12.39

08.35 Detector biases & leakage currents OK - see attachments 5 & 6
      FEE64 Temperatures OK - see attachment 7
      good events statistics OK - see attachment 8
      No warnings/errors reported by merger since DAQ GO - see attachment 9

      *all* system wide checks OK *except* ADC calibration of nnaida1, 21 and nnaida1 clock status (status=6)
      ADC calibration reports for nnaida1 & nnaida21 - see attachments 10 & 11

08.50 E17/F11 ambient temperature +24.1 deg C, RH 25.1%, d.p. +2.9 deg C
      Julabo FL11006 set point +20 deg C, actual +20.0 deg C, water level OK (c. 75%)

09.05 analysis of file RIBF148/R2_7 - deadtime <0.5%, no timewarps - see attachment 12

09.12 offline analysis RIBF148/R2_10 - ADC & disc data synchronised - see attachment 13

09.40 System stats showing aidas1 was 100% busy from c. 04.50 - see attachment 14

11.55 RIBF148 DAQ starts
      BigRIPS setting 167Sm

      AIDA file RIBF148/R2_48
      BigRIPS file 2002
      BRIKEN file 3

      Degrader setting  F11 3mm Al variable

      BNC PB-4
      Amplitude 90,000
      x5 attenuator IN
      t_r 100ns tau_d 50us
      - polarity
      frequency 25Hz range
      delay MIN, tail pulse, INT ref

      ASIC settings 2017Mar27-16.12.39

12.04 Detector biases & leakage currents OK - see attachments 15 & 16

12.29 DAQ stop
      AIDA file RIBF148/R2_57

      per DSSSD HEC x versus y - see attachment 17

12.32 RIBF148 DAQ starts

      AIDA file RIBF148/R2_57
      BigRIPS file 2003
      BRIKEN file 4

13:00 Bias and leakage currents - attachment 18

13:29 BRIKEN and BigRIPS stop daqs
      AIDA on file R2_73

13:31 BRIKEN stars run 5
      BigRIPS starts run 2004
      AIDA on file R2_274

      No beam at start of run

14.00 Detector biases & leakage currents OK but rate of increase significantly increased due to higher A & Z -
see attachments 19 & 20

14.10 DAQ stop
      AIDA file RIBF148/R2_84

     per DSSSD HEC x versus y - see attachment 21

     All SortSas spectra zero'd


14.12 RIBF148 DAQ starts

      Degrader setting  F11 1mm Al variable

      AIDA file RIBF148/R2_85
      BigRIPS file 2005
      BRIKEN file 6

      F11 rate ~200Hz

14.32 analysis file RIBF148/R2_89 - nnaida2 deadtime 0.6%, all other deatimes <0.5%, zero timewarps, three MBS
timewarps - see attachment 22

      offline analysis RIBF148/R2_91

 *** DSSSD # 1 count:    219696 old count:    166600 dt:      1.01 s  LEC rate:  52719.37 Hz
 *** DSSSD # 2 count:    346281 old count:    260940 dt:      1.01 s  LEC rate:  84735.65 Hz
 *** DSSSD # 3 count:    127888 old count:     95704 dt:      1.01 s  LEC rate:  31955.71 Hz
 *** DSSSD # 4 count:     89409 old count:     67186 dt:      1.01 s  LEC rate:  22065.37 Hz
 *** DSSSD # 5 count:     53702 old count:     40701 dt:      1.01 s  LEC rate:  12908.78 Hz
 *** DSSSD # 6 count:     83402 old count:     62481 dt:      1.01 s  LEC rate:  20772.60 Hz
 *** DSSSD # 1 count:         1 old count:         0 dt:      1.01 s  HEC rate:      0.99 Hz
 *** DSSSD # 2 count:        11 old count:         9 dt:      1.01 s  HEC rate:      1.99 Hz
 *** DSSSD # 3 count:         2 old count:         2 dt:      1.01 s  HEC rate:      0.00 Hz
 *** DSSSD # 4 count:         8 old count:         6 dt:      1.01 s  HEC rate:      1.99 Hz
 *** DSSSD # 5 count:        22 old count:        13 dt:      1.01 s  HEC rate:      8.94 Hz
 *** DSSSD # 6 count:       447 old count:       333 dt:      1.01 s  HEC rate:    113.19 Hz

     ADC & disc data synchronised - see attachment 24

     online analysis - per DSSSD HEC x versus y - see attachment 23

14.43 all SortSas spectra zero'd

15.08 Detector biases & leakage currents OK - see attachments 25 & 26
      FEE64 Temperatures OK - see attachment 27
      good events statistics OK - see attachment 28
      No warnings/errors reported by merger since DAQ GO - see attachment 29

      *all* system wide checks OK *except* ADC calibration of nnaida1, 21 and nnaida1 clock status (status=6)

15.11 DAQ stop
      AIDA file RIBF148/R2_101


15.13 RIBF148 DAQ starts

      Degrader setting  F11 1mm Al variable

      AIDA file RIBF148/R2_102
      BigRIPS file 2006
      BRIKEN file 7

15.37 DAQ stop
      AIDA file RIBF148/R2_109

      online analysis - per DSSSD HEC x versus y
                        per DSSSD HEC E_p versus E_n - see attachments 30 & 31
                        heavy-ions stopped:transmitted ~50:50

15.42 RIBF148 DAQ starts

      Degrader setting F11 *all out*

      AIDA file RIBF148/R2_110
      BigRIPS file 2007
      BRIKEN file 8

      *all* SortSas spectra and online histograms zero'd
  521   Sun Dec 4 03:44:12 2016 TDSunday 4 December
12.45 DSSSD bias & leakage currents OK - see attachments 1 & 2
      FEE64 temperatures OK - see attachment 3
      Ambient temperature +25.1 deg C, d.p. -3.0 deg C, RH 15.6%

14.05 per FEE64 Rate, Hitrate, 1*L and 1*W spectra, good events stats - see attachments 4-10
      Note - good events stats indicate increased extrinsic noise (c. 100kHz period) during these screenshots

14.15 1*W and 4*W spectra for nnida24, 17 & 20 - see attachments 11-16
      Note - nnida24 ASIC#1 probably has cable/connector fault at DSSSD end
  743   Sun Sep 30 01:13:47 2018 OH, TDSunday 30 September
09.10 DAQ continues OK - current file Sep18/R2_6

09.15 all system wide checks OK *except* nnaida7 fails ADC calibration

      detector biased & leakage currents OK - see attachments 1 & 2


      FEE64 temperatures OK - see attachment 3
      good event statistics OK - see attachment 4
      nnaida2 rate due to one hot channel

      Merger OK - see attachment 5
      TapeServer OK - see attachment 6
      no errors or warnings shown by server terminal sessions 

      1.8.W - 20us FSR - see attachments 7 & 8

      all spectra zeroed
      
16.05 temperature +24.2 deg C, RH 45.4%, d.p. +11.6deg C
      Julabo FL11006 water level c. 75-80% (i.e. approx mid-way between middle and upper level indicators)
  408   Sun Oct 30 01:53:13 2016 TDSunday 30 October
10.53 DAQ found stalled 
      Most recent files R89_102 last accessed 07.17

      DSSSD bias & leakage current OK - see attachments 1 & 2
      FEE64 temperatures OK - see attachment 3

      All system wide checks OK *except* Clock status, SYNC errors & SYNC pulses - see attachments 4-8 - unusual
messages/reports

      One merger error messages - link 18 - see attachment 9

      Restart merger
      DAQ STOP OK
      DAQ RESET & SETUP OK
      ReSYNC OK
      All system-wide checks OK
      Disable H & data transfer #1
      Restart merger & tapeserver
      Enable H & data transfer #1

      Standard soft reset did not work - one or more FEE64s dropped data transfer #1 link at DAQ GO

      Retry - one, more FEE64s indicate error state - run control tab unresponsive 

      Power cycle FEE64s 
  94   Sun May 3 01:32:18 2015 JA, TD, AE, CG, VP & JLTSunday 3 May 2015
09.19 Beam off on arrival, now back
      R51 running

      BigRIPS 88Ge setting

      Degraders 4mm Al + 1.3mm Al variable, MINOS target filled 

      FEE64 temperatures OK
      Julabo FL11006 recirculating chiller OK, operating temp +20 deg C
      Detector HV OK
      AIDA SYNC/PAUSE/RESUME rates OK

10.05 R51 ends

10.51 Change BigRIPS setting to 94Se

      OR MACB triggers from nnaida1, 2, 6, 10, 13 & 14 input EURICA trigger module input #7
      All nnaida1, 2, 6, 10, 13 & 14 fast discriminators currently disabled.

11.33 R53

      BigRIPS 94Se setting

      c. 15.54 EURICA DAQ LUPO timestamp reset

14.00 Degrader changed t0 4mm fixed + 0.7mm variable (Al)
  690   Sun Oct 29 05:16:35 2017 TD, OHSunday 29 October
14.16 DAQ running OK - data transfer not enabled

      BNC PB-4
      Amplitude 90,000
      x5 attenuator IN
      t_r 100ns tau_d 50us
      - polarity
      frequency range 25Hz
      delay MIN, tail pulse, INT ref

      ASIC settings 2017Oct23-14.38.03
      *all* FEE64s slow comparators 0xa *except* nnaida13 0xe, nnaida19-22 0xc

      Detector biases & leakage currents OK - see attachments 1 & 2
      FEE64 Temperatures OK - see attachment 3
      DAQ statistics OK - see attachment 4
      *all* system wide checks OK *except* nnaida21 fails ADC calibration
      Julabo FL11006 recirculating chiller set point +20 deg C, actual +20.0 deg C
      Ambient temperature +23.9 deg C, d.p. +10.8 deg C, RH 43.2%

15.05 per FEE64 Rate, 1.8.L and 1.8.W spectra - note common x & y scales
      Peak width nnaida5 63 ch FWHM
      Peak width nnaida8 82 ch FWHM

15.09 All histograms zero'd

18.25 Detector biases & leakage currents OK - see attachments 11 & 12
      FEE64 Temperatures OK - see attachment 13
      DAQ statistics OK - see attachment 14
      *all* system wide checks OK *except* nnaida21 fails ADC calibration
  247   Sun May 29 14:06:31 2016 TD, AE & CGSunday 29 May 2016
Throughout day:
      DSSD2 (3058-25) would not bias, either short circuit or suspiciously low leakage current.
      Carried out some test in situ which were inconclusive.
      Removed snout and took DSSD2 out.
      Measuring the resistance across the two pins providing the n+n bias yielded 20 Ohms.
      For the p+n the same measurement saw an open circuit => at least one of the bond wires broken.

      DSSD1 (3131-9) was moved into the DSSD2 position and DSSD1 is now BB18-1000 3058-4.

      All detectors were biased downstairs and all looked fine - leakage currents of 3.1 - 3.8 uA.

      First detector sits 54.5cm from Al base plate and 10cm from the upstream end of the snout.
      9mm between each of the DSSDs.

      Stack now:
           |   3058-4
        b  |   3131-9
        e  |   2998-23
        a  |   3058-5
        m  |   3058-6
           V   3058-7

      207Bi source installed at end of snout.

      Snout moved upstairs and mounted.
      DSSD2 would not bias. This was due to the HV cable providing bias for this DSSD having a short to ground.
      Replaced with HV cable 7 and biased successfully.
      All DSSDs now bias. Screenshots of leakage currents attached (attachments 4+5) at room temp 25.3oC.
      Caen module Addr 00 provides bias to DSSDs 1, 2, 3 and 4 via channels 0, 1, 2 and 3 respectively.
      Module Addr 01 provides bias to DSSDs 5 and 6 on channels 0 and 1, respectively.
 
      Snout pushed in to EURICA.

22.05 nnaida17 ASIC parameters not loading
      power cycle & reboot

      nnaida5 (master) clock not locked - status 0x6 - see attachments 1-3
  248   Sun May 29 20:06:15 2016 TD, AE & CG - reply PJCSSunday 29 May 2016
> Throughout day:
>       DSSD2 (3058-25) would not bias, either short circuit or suspiciously low leakage current.
>       Carried out some test in situ which were inconclusive.
>       Removed snout and took DSSD2 out.
>       Measuring the resistance across the two pins providing the n+n bias yielded 20 Ohms.
>       For the p+n the same measurement saw an open circuit => at least one of the bond wires broken.
> 
>       DSSD1 (3131-9) was moved into the DSSD2 position and DSSD1 is now BB18-1000 3058-4.
> 
>       All detectors were biased downstairs and all looked fine - leakage currents of 3.1 - 3.8 uA.
> 
>       First detector sits 54.5cm from Al base plate and 10cm from the upstream end of the snout.
>       9mm between each of the DSSDs.
> 
>       Stack now:
>            |   3058-4
>         b  |   3131-9
>         e  |   2998-23
>         a  |   3058-5
>         m  |   3058-6
>            V   3058-7
> 
>       207Bi source installed at end of snout.
> 
>       Snout moved upstairs and mounted.
>       DSSD2 would not bias. This was due to the HV cable providing bias for this DSSD having a short to ground.
>       Replaced with HV cable 7 and biased successfully.
>       All DSSDs now bias. Screenshots of leakage currents attached (attachments 4+5) at room temp 25.3oC.
>       Caen module Addr 00 provides bias to DSSDs 1, 2, 3 and 4 via channels 0, 1, 2 and 3 respectively.
>       Module Addr 01 provides bias to DSSDs 5 and 6 on channels 0 and 1, respectively.
>  
>       Snout pushed in to EURICA.
> 
> 22.05 nnaida17 ASIC parameters not loading
>       power cycle & reboot
> 
>       nnaida5 (master) clock not locked - status 0x6 - see attachments 1-3

I suggest you check the hdmi hasn't been disturbed in the Macb or the FEE64.. Since the others are ok. I suggest it is the cable from the 
Macb 
to the FEE64 #5. If this fails to correct then please send a screen shot of the Local Controls status for nnaida5. 