Mon Feb 22 15:19:52 2016, PJCS, T9 System changed to version 8
|
I have changed the T9 system to be using Version 8.
Tests using the detector and pulser seem to be ok.
Transmission to the Merge and Merging operating ok |
Thu May 12 15:50:48 2016, Patrick, System Checks and firmware update 
|
Added two checks to the System wide checks.
Modified the Firmware so the checks work properly.
One check reads the status of the FEE64 clocks and checks
against expected value.
The other check reads the value from all the FEE64 Sync Error Counters, pauses for one second, then reads them again and compares
with the first reading. Any change is reported as an error.
The Sync Error counter is incremented if the SYNC pulse arrives at a FEE64 when the Timestamp |
Sun Dec 4 05:46:42 2016, TD, SyncCheck RIBF03R1 & RIBF123R1 commissioning tests archived
|
The SyncCheck etc software used for the RIBF123R1 and RIBF03R1 commissioning
tests on d05 (user aida) has been archived as
|
Sun Oct 9 11:47:34 2016, CG, TD, Sunday 9th October 9x
|
12.30 Following scheduled power outages of October 8
|
Sun Jun 9 03:43:00 2019, CB, TD, CG, OH, Sunday 9 June 28x
|
11.43 DAQ continues OK
ASIC settings 2019May19-17.54.03
|
Sun Oct 8 14:51:59 2023, TD, Sunday 8 October contd. 28x
|
Slow comparator 0xf
|
Sun Oct 8 11:44:35 2023, TD, Sunday 8 October 50x
|
12.22 USB-controlled ac mains relay - interlock overidden
Water flow OK. Water temp 19 deg C OK.
Power on.
|
Sun May 7 01:29:50 2017, OH, TD, Sunday 7 May 13x
|
09.28 DAQ continues OK
Current file May2017/R1_36
|
Sat Nov 4 23:49:14 2017, OH, Sunday 5 November 6x
|
08.09 DAQ running OK - data transfer not enabled
BNC PB-4
|
Sun Jun 5 02:14:10 2016, TD, AE, CG & DK, Sunday 5 June 2016 8x
|
10.10 Detector biases & leakage currents - OK - see attachments 1-2
10.40 Detector biases & leakage currents - OK - see attachments 3-4
|
Sun Sep 4 04:26:23 2016, TD, Sunday 4 September 20x
|
14.20 R13 continues
DSSSD bias/leakage current OK - see attachments 1-2
|
Sun Nov 4 01:11:41 2018, CA, CB, TD, OH, PJW, Sunday 4 November 20x
|
10.03 DAQ continues OK - current file Oct18/R16_9
FEE64 LEC/MEC slow comparator threshold 0x64
beam background
|
Sun Jun 4 00:06:18 2017, OH, TD, PJW, Sunday 4 June 08.00-16.00 32x
|
08:00 DAQ found stopped at around 04:49 - Attachment 1
Time obtained from last file written as part of R1
Tape server still running - Attachment 2
|
Sun Dec 4 03:44:12 2016, TD, Sunday 4 December 16x
|
12.45 DSSSD bias & leakage currents OK - see attachments 1 & 2
FEE64 temperatures OK - see attachment 3
Ambient temperature +25.1 deg C, d.p. -3.0 deg C, RH 15.6%
|
Sun Sep 30 01:13:47 2018, OH, TD, Sunday 30 September 8x
|
09.10 DAQ continues OK - current file Sep18/R2_6
09.15 all system wide checks OK *except* nnaida7 fails ADC calibration
|
Sun Oct 30 01:53:13 2016, TD, Sunday 30 October 9x
|
10.53 DAQ found stalled
Most recent files R89_102 last accessed 07.17
|
Sun May 3 01:32:18 2015, JA, TD, AE, CG, VP & JLT, Sunday 3 May 2015
|
09.19 Beam off on arrival, now back
R51 running
|
Sun Oct 29 05:16:35 2017, TD, OH, Sunday 29 October 14x
|
14.16 DAQ running OK - data transfer not enabled
BNC PB-4
|
Sun May 29 14:06:31 2016, TD, AE & CG, Sunday 29 May 2016    
|
Throughout day:
DSSD2 (3058-25) would not bias, either short circuit or suspiciously low leakage current.
Carried out some test in situ which were inconclusive.
|
Sun May 29 20:06:15 2016, TD, AE & CG - reply PJCS, Sunday 29 May 2016
|
> Throughout day:
> DSSD2 (3058-25) would not bias, either short circuit or suspiciously low leakage current.
> Carried out some test in situ which were inconclusive.
|