ID Date Author Subject
  846   Sun Jun 16 10:06:21 2019 TD, CGFEE64 nnaida14 noise issues
Attachments 1, 3, 5 & 7 *.L spectra log scale

Attachments 2, 4, 6 & 8 *.L spectra linear scale (expanded around pulser peak) 

Attachments 9-12 *.W spectra 20us FSR 

Observe 

2.13.L low amplitude
2.14.L & 3.3.L large number of counts and 'sinusoidal' continuum

... but ... the corresponding waveform spectra look OK

Zero'd all histograms to confirm observations - confirmed - see attachment 16
Pulser peak width spectrum 3.8.L 71 ch FWHM
  845   Sun Jun 16 04:50:03 2019 CATests of calibration on beta data
Analysis of RIBF148/R5_170_190, with 500 < E < 1000 keV energy cut

Note also a 600 keV energy difference cut was applied when sorting the data.

plots in order: Ex - Ey
                Ey vs Ex
                Ex-Ey vs xStrip
                Ex-Ey vs yStrip

DSSD 0 - attachments 1-4

DSSD 1 - attachments 5-8

DSSD 2 - attachments 9-12

DSSD 3 - attachments 13-16

DSSD 4 - attachments 17-20

DSSD 5 - attachments 21-24 (This DSSD has significantly more counts than the rest, so on different scale)

attachment 25 - table comparing peak centroids and sigma values
                calculated by applying gaussian fit to central peak in Ex-Ey plots
  844   Sun Jun 16 04:09:40 2019 CA, CB, ML, TD, CGSunday 16 June noon-midnight
12.10 DAQ continues OK file RIBF147/R2_145

      100Sn setting, no degrader

      ASIC settings 2019Jun13-08.56.35

      BNC PB-4 Pulser 
      Amplitude 90,000
      x5 attenuator IN
      Frequency 25Hz range, variable min. c. 2.5Hz
      t_r 100ns, tau_d 50us
      - polarity
      Delay MIN, tail pulse

      good event statistics - DSSSD#1 noisy - see attachment 3
      10x <20k, 19x <50k, 20x <100k

      p+n FEE64s 1.8.W spectra 20us, 200us, 2ms, 20ms FSR - DSSSD#1 noisy - attachments 4, 6, 7 & 8
      n+n FEE64s 1.8.W spectra 20us FSR - DSSSD#1 noisy - attachment 5

      FEE64 Rate spectra - attachment 1
      Check ASIC Control *all* FEE64s & ASICs x1
      FEE64 Rate spectra - attachment 2

      Observe that nnaida17 is the really noisy FEE64 - nnaida20, 22 & 23 show increased noise but
      not at same level as nnaida17. Is nnaida17 actually the problem? LH kapton PCB is the only
      one we did not replace for example.

12.21 DSSSD#1 stats OK

12.36 AIDA R2_151
      DTAS run 027
      BigRIPS still 2012

13:43 DAQ continues running OK file RIBF147/R2_167
      https://www.reddit.com/r/politics/
      Settings as before

      Good events stats OK - DSSD#1 no longer noisy - see attach 9
      Leakage current OK - see attach 10
      Systems wide checks OK except nnaida7 fails calibration

      Merger ~550k events/second, TapeServer 4435 Kbytes/s

      Will increase thresholds on DSSD#1 and decrease on DSSD#3 on next DTAS stop.

14:20 DSSD#1 was found noisy again (Rate>100kHz & Tapeserver 10Mbytes/s)) but then soon after was back to a
      normal counting rate (Rate back to ~40kHz & Tapeserver back to 5 Mbytes/s)

14:39 DTAS & BigRIPS stops - changing threshold during AIDA R2_174
      DSSD#1 (nnaida17, 20, 22, 23) slow comparator threshold -> 1d (300 keV)
      DSSD#3 (nnaida9, 12, 14, 15) slow comparator threshold -> 8 (80 keV)

      Check ASIC control x1 on all FEE64

      Restarting
      AIDA R2_175
      DTAS 30
      BigRIPS 2013

      New stats look OK - zero'd Counters - see attach  11
      nnaida 17,20,22,23 rates are down
      nnaida 9,12,14,15 rates are up
      It appears the data rate balances out. Data items & write to disk similar to before the changes. 
      See attach 12 & 13

      New sample waveforms in attach 14 & 15. nnaida17 looks significantly better than before.
      Rates spectra also attached - see 16
      
17:12 Data rate still ~600k items/s
      File rolled over to R2_196

17:30 Current F11 rate ~190 cps
      Estimation of implant rate in AIDA of ~270 implants/s (number of HEC events in DSSD1 y strips over duration
of file R2_197)
         > increased AIDA rate likely due to double counting of multi-strip events
      Temps all ok


18.11 DAQ stats zero'd
      *all* histograms zero'd

20:13 DAQ continues to run OK
      Data rate ~620k items/s. FEE64 data rates remain approx. same
      Temps OK
      Just rolled over to file R2_222

20:44 DTAS run 39 started
      BigRIPS run 2015
      AIDA at 1.2GB into R2_226

22:44 DTAS run 42 started
      BigRIPS run 2017
      AIDA at 1.3GB into R2_243

      Carried out ASIC check successfully
      Temps and data rate remain unchanged.

23:16 Been almost 1 day running. 
      DTAS running out of range in some of their timing histograms so reset the correlation scaler
      Reset successfully at start of R2_249

23:19 Reset signal sent again. Still in file R2_249

23:23 DTAS run 43 started
      BigRIPS run 2018
      AIDA run R2_250

23:54 MIDASSORT spectra saved to /homes/npg/spectra/specs-2019Jun16-23.52.55
      MIDASORT online started.
  843   Sat Jun 15 16:53:22 2019 CB, OH, DK, MLSunday 16 June
00.30 AIDA file R2_26
      DTAS run 007
      BigRIPS run 2006

00:55
      Merger status - attachment 1
      TapeSever status - attachment 2

01:40
      Before zeroing stats - attachment 3
      Histos up to now saved in /homes/npg/spectra/specs-2019Jun16-01.46.50
      Before zeroing histos - XY attachment 4

02:22  Up to AIDA R2_41
      Stats - attachment 5
      XY implants - attachment 6

02:29 AIDA R2_42 
      DTAS run 011
      BigRIPS still 2006

04:08 Up to AIDA R2_55
      Temps - attachment 7
      Biases - attachment 8
      Stats - attachment 9
      XY implants - attachment 10

04:50 Did some major UI updates to merger /data02/2019data/dtas/scripts/parallelscript/runall_xe19_dev.sh
         If no parameters are passed, it asks to select each option one-by-one, showing available options
             For DTAS, if there is only one run for a given header, it automatically selects it
         After that, it checks that all the raw files actually exist
         Finally, it composes the command, asks the user if it looks OK, and then launches the script (or quits)
         (For experts, it can still take the original case where all 7 parameters are passed at the command line)
         There may still be some minor bugs, so for now we don't replace the original one until Alvaro can check

06:09 Stats - attachment 11
      XY implant profiles -- attachment 12

06:30 DTAS 018; AIDA R2_80; BigRIPS 2009

Things getting a bit noisier?
      Merger - attachment 13
      TapeServer - attachment 14
      Stats - attachment 15

   The FEEs corresponding to DSSSD1 are all showing factors 2 to 10 higher than just 30 min. ago

   F11 rates still near 180 cps, primary beam still around 130 pnA.

06:43 Before I can diagnose the high noise problem in DSSSD1, it goes back to normal
      Merger - attachment 16
      TapeServer - attachment 17
      Stats - attachment 18
   Maybe the issue was from looking at the details too often :)

07:46 Shift change.

      DAQ continues running OK.
      Statistics show  noisy DSSD #1 (100-200kHz) - see attach 19
      Tapeserver also shows higher than normal write to disk rate - see attach 20
      After a few seconds write to disk rate goes back to normal - see attach 21
      Data items rate looks OK (500k). This was checked after the rates in DSSD#1 went back to normal - see attach 22
      FEE64 temperatures OK - see attach 23
      Implants look OK - see attach 24

      System-wide checks OK except nnaida7 fails ADC calibration as usual.

      Reset Statistics
      Stats now OK - see attach 25

      Reset HEC implants

      Timewarp events appear to be very frequent, consistently every 3-5 seconds.
      This seems to have been happening ever since beam was put on target.
      Analysed R2_94 using analyser in Desktop 3. No timewarps reported. See attach 26.
      
08:32 AIDA: R2_100,
      DTAS: 21  ,
      BigRips: 2010   

08:41 Another noise spike from DSSD#1
      Observed up to 1,600k data items per second - see attach 27
      Write to disk ~10 MB/s as before.
      See LEC xy plot - attach 28
      HEC xy plot shows no hints of issues - see attach 29
      
      Rate seems manageable - see attach 30 for analysis of noisy run R2_103. 
      Note few seconds deadtime in some FEE64 and unequal number of Pause/Resume in nnaida 22      

09:45 Rate in DSSD#1 back to normal. Data rate 500k, Write to disk 5 Mb/s.

10:00 Started compressing R2_* files to save disk space.
      Used nice gzip to compress to avoid conflicts with the DAQ.
      Status of compression can be checked in ELOG desktop in the terminal window running ls -alth

10:32 DTAS and BigRIPS stop runs
      AIDA on file R2_128

      DTAS:24
      BigRIPS 2011

10:56 Lost beam. BigRIPS and DTAS DAQs continue running. Accelerator operators advise beam should be back in 10 minutes.

11:12 Beam is back. BigRIPS & DTAS continue running and did not stop. AIDA file R2_133.

11:25 System-wide checks OK except nnaida7 fails ADC calibration as usual.
      Stats OK, but nnaida19 is a bit more noisy than befor (80k) - see attach 31
      Temperatures OK - see attach 32
      TapeServer & Merger OK - see attach 33, 34 - rates have been low since 10 am
      HEC xy plot OK - see attach 35

      Reset online spectra

      Leakage current OK - attach 36
     
  842   Sat Jun 15 01:05:41 2019 CA, CB, TD, CG, OH, DK, MLSaturday 15 June
09.05 Julabo FL11006 chiller set point +20 deg C actual +20.0 deg C
      Water level c. 75% M-H markers
      Temperature +24.1 deg C dew point +8.3 deg C RH 36.2%

      DAQ continues OK

      ASIC settings 2019Jun13-08.56.35

      BNC PB-4 Pulser 
      Amplitude 90,000
      x5 attenuator IN
      Frequency 25Hz range
      t_r 100ns, tau_d 50us
      - polarity
      Delay MIN, tail pulse

      all system wide checks OK *except* nnaida7 fails ADC calibration

      detector biases & leakage currents OK - see attachments 1 & 2

      FEE64 temperatures OK - see attachment 3

      good event statistics - OK - see attachment 4
      11x <20k 23x < 50k 24x <100k cf. 9x <20k 23x < 50k 24x <100k 22.39 yesterday
      Merger OK - see attachment 5
      TapeServer OK - see attachment 6

      p+n FEE64s 1.8.W spectra 20us FSR - attachment 7
      n+n FEE64s 1.8.W spectra 20us FSR - attachment 8
      1.8.L spectra - see attachments 9 & 10
      nnaida1 74 ch FWHM, nnaida3 78 ch FWHM - cf. 81 & 77 ch FWHM respectively 23.20 yesterday

      FEE64 Rate spectra - attachment 11
      Check ASIC Control *all* FEE64s & ASICs x2
      Check ASIC Control x2
      FEE64 ASIC
      10    4
      16    4
      17    4
      FEE64 Rate spectra - attachment 12 - some channels recovered

09.19 ADC data synchronised - attachment 13

09.23 zero all histograms
      zero DAQ stats

16:45 DAQ Continues to run
      Rates on 17, 20, 22 amd 23 are up - attachment 14
      Bias and leakage current ok - attachment 15
      System wide checks ok
      ASIC check ok
      
16:58 Statistics back to normal level - attachment 16

17.34 DAQ stop
      replaced kapton pcb on FEE no. 22

      DAQ start -> writing to May19/R36

      system wide checks ok, *except* naida7 fails adc calibration

      good event statistics ok - attachment 17

      TapeService and Merger ok - writing at 4 Mb/s - attachments 18 & 19

17.41 zeroed all histograms

c. 19.30 First beam to AIDA + DTAS - see attachments 20-24

      *no* degraders

      Observe most ions stop in DSSSD#2 with a few stopping in DSSSD#3
              focus looks OK but distribution is to LHS as you look upstream to BigRIPS

19.50 Removed upstream PPAC and adjusted ZDS transverse settings to centre - attachments 25-29

      Now observe ions now stop mostly in DSSSD#3 but significant fractions still stop ion DSSSD#2
  
      DSSSD#3 appears well focussed and centred

20.19 all system wide checks OK *except* nnaida7 fails ADC calibration

      detector biases & leakage currents OK - see attachment 30

      FEE64 temperatures OK - see attachment 31

      good event statistics - OK - see attachment 32

20.53 xy hit patters in all DSSDs

      ions mostly in DSSD3, some still depositing in DSSD2

21.06 attachment 33 - beam slightly focused to the left - xy hit pattern

      attachment 34 - beam further focused - xy hit pattern

21.14 DAQ stop

      DAQ start writing to /RIBF147/R1

      Sync pulse sent by DTAS on file R1_0

21.17 DAQ stop/start

      

      writing to file /RIBF147/R2

      sync pulse sent by DTAS - file R2_0

      ASIC settings 2019Jun13-08.56.35

      BNC PB-4 Pulser 
      Amplitude 90,000
      x5 attenuator IN
      Frequency 25Hz range
      t_r 100ns, tau_d 50us
      - polarity
      Delay MIN, tail pulse

      no degrader

21.24 AIDA file on R2_1
      DTAS start 003
      BigRIPS run: 2003

21.31 histograms zeroed

      good event statistics ok - attachment 35

21.34 AIDA file on R2_2
      DTAS start 004
      BigRIPS run: 2004

22.02 good event statistics ok - attachment 36

22.16 AIDA file on R2_8
      DTAS run 004
      BigRIPS run 2004
      
      checked with DTAS
      
22.29 energy x vs y (HEC) - attachment 37 - grouping ~ 3MeV in all detectors
      xy hit pattern (HEC) - attachment 38
      implants centered on DSSD3

22.32 AIDA file R2_9
      DTAS run 005
      BigRIPS run 2005
      beam returns

22.34 HEC multiplicity - most events ~1,2,3 pixels - attachment 39

22.35 checked with DTAS
      AIDA file R2_10
      DTAS run 006
      BigRIPS run 2006

22.38 LEC energy x vs y - clear Ex=Ey locus - attachment 40
      xy hit pattern LEC - attachment 41
      xy multiplicity LEC - attachment 42 - most events 1,2 pixels

22.49 analyzer.f output for file R2_5 - attachment 44
      dead time 0.5% in fee19, zero in all others

22.56 problem with beam, beam stopped on AIDA file R2_8

23.05 all system wide checks OK *except* nnaida7 fails ADC calibration

      detector biases & leakage currents OK - see attachment 45

      FEE64 temperatures OK - see attachment 46

      good event statistics - OK - see attachment 47

23.50 AIDA file R2_21
      DTAS run 006
      BigRIPS run 2006




      
      
      



      

 
      
  841   Fri Jun 14 10:33:44 2019 OH, CA, TD, CG, CB, DKFriday 14th June
Friday 14th June 

Authors: TD, DK, CB

08:40

  elog appears to be down.  Relevant screenshots are in /tmp/june14/ so please upload when we can access

  Julabo chiller 75% M-H markers
  Room Temp +24.5 C deg, dew point +8.1 deg C, Room humidity 35.2%

  Pulser, ASIC settings as yesterday (copy/paste)

      ASIC settings 2019May19-17.54.03
      slow comparator 0xa

      BNC PB-4 Pulser 
      Amplitude 90,000
      x5 attenuator IN
      Frequency 25Hz range
      t_r 100ns, tau_d 50us
      - polarity
      Delay MIN, tail pulse

  DAQ continues OK
  Statistics as before.  nnaida19 counter appears to have overflowed and wrapped around
    See stats.png
  
  System-wide checks OK, except nnaida7 fails ADC calibration as before

  Biases and leak currents reasonable -- see biases.png

  FEE64 temperatures okay, see fee-temps.png

  Merger looks okay, ~600k items / sec; see merger.png

  TapeServer as before, around 4.5 Mbytes/sec; see tapeserver.png


13:52

   Pulser, ASIC settings same as this morning

   DAQ continues OK
   Statistics as before. See attach 190614_1354_stats. Counts RESET.

   System-wide checks OK, except nnaida7 fails ADC calibration as before

   Bias and leak currents OK. See attach 190614_1356_bias

   FEE64 temperatures OK, see attach 190614_1357_temps

   Merger keeps running at ~560k items /s; see 190614_1359_merger
   TapeServer as before. See 190614_tape

17:24 AIDA DAQ crashed by DTAS touching the clock function generator
      Function generator set at 50MHz square high +5V low 0V

18:38 problem loading ASIC settings - mismatch in slow comparator threshold and shaping time
      performed save/restore with 2019Jun13-08.56.35 - all looks ok now
      
      System wide checks ok except nnaida7 fails adc calibration

      FEE64 temperatures ok - attachment 12

      good event statistics ok - attachment 13

      bias/leakage currents ok - attachment 14

     
18:38 for reals

      DAQ Stopped
      R30_0 opened writing to disk
      
      Was to test DTAS sync whilst writing to file
18:43 DAQ stopped on file R30_0

18:45 DAQ restarted on file R31 with no writing to disk

21:50 DAQ found crashed (Valencia group believe it crashed around 20:00)

      Trying to reset FEEs resulted in an error message attachment 15 and 16

      Each FEE was rebooted using telnet to avoid a power cycle
      System reset after reboot

      System wide checks ok
      Check ASIC control 

      FEE temps ok

      DAQ Started again
      On startup FEEs 17, 20, 22 and 23 were all about 150k
      All relate to DSSD0
      Pressed on kapton and noise went down.

      Statistics now good

22.45 Review /var/log/messages and observe nnaida15 rebooted at 19.27 - reason unknown
      This probably explains why DAQ stopped.

Jun 14 19:27:32 aidas1 dhcpd: DHCPDISCOVER from 00:04:a3:2a:f3:ac via p2p2
Jun 14 19:27:32 aidas1 dhcpd: DHCPOFFER on 10.1.2.15 to 00:04:a3:2a:f3:ac via p2p2
Jun 14 19:27:32 aidas1 dhcpd: DHCPREQUEST for 10.1.2.15 (10.1.2.250) from 00:04:a3:2a:f3:ac via p2p2
Jun 14 19:27:32 aidas1 dhcpd: DHCPACK on 10.1.2.15 to 00:04:a3:2a:f3:ac via p2p2
Jun 14 19:27:32 aidas1 rpc.mountd[2868]: authenticated mount request from 10.1.2.15:784 for
/data10/XilinxLinux/ppc_4xx/rfs/nnaida15 (/data10/XilinxLinux/ppc_4xx/rfs)
Jun 14 19:27:59 aidas1 rpc.mountd[2868]: authenticated mount request from 10.1.2.15:864 for /MIDAS@aidas
(/MIDAS@aidas)


22.53 ADC data synchronised - attachment 17

23.05 zero DAQ statistics
      zero all histograms
      

23.20 Check ASIC Control x1 
      FEE64 Rate spectra - attachment 18

      1.8.L spectra - see attachments 19 & 20
      nnaida1 81 ch FWHM, nnaida3 77 ch FWHM

      p+n FEE64s 1.8.W spectra 20us FSR - attachment 21
      n+n FEE64s 1.8.W spectra 20us FSR - attachment 22

23.36 DAQ start file May19/R34_0

      DTAS+AIDA correlation test using gamma ray sources

      Correlation scaler reset issued during R34_0

      ASIC settings 2019Jun13-08.56.35

      BNC PB-4 Pulser 
      Amplitude 90,000
      x5 attenuator IN
      Frequency 25Hz range
      t_r 100ns, tau_d 50us
      - polarity
      Delay MIN, tail pulse

23.53 DAQ stop file May19/R34_2

      nnaida15 did not stop - state unknown - reboot nnaida15

      This did not work - unable to reset nnaida15

      Closed all FireFox broweser tabs, stopped all servers and restarted

00.14 DAQ start

      ASIC settings 2019Jun13-08.56.35

      BNC PB-4 Pulser 
      Amplitude 90,000
      x5 attenuator IN
      Frequency 25Hz range
      t_r 100ns, tau_d 50us
      - polarity
      Delay MIN, tail pulse
  840   Thu Jun 13 05:49:37 2019 CAFurther tests of calibration on Alpha data
Alpha data used: /May19/{R1,R2,R7,R8,R9,R11} (c. 380 hr.)

Absolute calibration calculated with relative calibrated data

Data then resorted with absolute and relative gains applied

Attachment 1 -> Ey vs Ex plot for all detectors, with slope Ey=Ex show in red. Some non-linearity present, but to first order looks good

Attachment 2 -> Alpha peaks pre- absolute calibration in DSSD0
                Ex in blue, Ey in red
                To determine absolute calibration coefficients, assigned leading edge of higher energy peak an energy of 7.4 MeV (Uranium decay series)
                Divide this by the observed channel number (in keV) to get the absolute gain coefficients in keV/ch for Ex and Ey in each detector.
                (Note, data already offset corrected, pulser walkthrough)
                Values typically ~0.72 keV/ch
                Multiply appropriately by relative gains obtained from gainMatch.cpp to get final gain coefficients.

Attachments 3 - 6 -> DSSD0 -> Ex - Ey
                           -> Ey vs Ex
                           -> Ex - Ey vs x strip
                           -> Ex - Ey vs y strip

Attachments 7 - 10 -> DSSD1 -> Ex - Ey
                            -> Ey vs Ex
                            -> Ex - Ey vs x strip
                            -> Ex - Ey vs y strip

Attachments 11 - 14 -> DSSD2 -> Ex - Ey
                             -> Ey vs Ex
                             -> Ex - Ey vs x strip
                             -> Ex - Ey vs y strip

Attachments 15 - 18 -> DSSD3 -> Ex - Ey
                             -> Ey vs Ex
                             -> Ex - Ey vs x strip
                             -> Ex - Ey vs y strip

Attachments 19 - 22 -> DSSD4 -> Ex - Ey
                             -> Ey vs Ex
                             -> Ex - Ey vs x strip
                             -> Ex - Ey vs y strip

Attachments 23 - 26 -> DSSD5 -> Ex - Ey
                             -> Ey vs Ex
                             -> Ex - Ey vs x strip
                             -> Ex - Ey vs y strip

Attachment 27 -> Table comparing sigma and peak centroids of Ex - Ey plots for each DSSD
                 Calculated using simple gaussian fit






                              
  839   Thu Jun 13 00:30:17 2019 CA, CB, TD, CG, OH, DKThursday 13 June
08.32 Julabo FL11006 chiller set point +20 deg C actual +20.0 deg C
      Water level c. 70% M-H markers + 0.5 jugful -> c. 80% M-H markers
      Temperature +24.0 deg C dew point +9.6 deg C RH 39.0%

      DAQ continues OK

      ASIC settings 2019May19-17.54.03
      slow comparator 0xa

      BNC PB-4 Pulser 
      Amplitude 90,000
      x5 attenuator IN
      Frequency 25Hz range
      t_r 100ns, tau_d 50us
      - polarity
      Delay MIN, tail pulse

      all system wide checks OK *except* nnaida7 fails ADC calibration

      detector biases & leakage currents OK - see attachments 1 & 2

      FEE64 temperatures OK - see attachment 3

      good event statistics - OK - see attachment 4
      8x <20k 23x < 50k 24x <100k cf. 7x <20k 23x < 50k 24x <100k yesterday evening
      Merger OK - see attachment 5
      TapeServer OK - see attachment 6

      p+n FEE64s 1.8.W spectra 20us FSR - attachment 7
      n+n FEE64s 1.8.W spectra 20us FSR - attachment 8
      1.8.L spectra - see attachments 9 & 10
      nnaida1 85 ch FWHM, nnaida3 86 ch FWHM - cf. 75 & 79 ch FWHM respectively 17.04 Tuesday 11 June 

      FEE64 Rate spectra - attachment 11
      Check ASIC Control *all* FEE64s & ASICs x2
      Check ASIC Control x3
      FEE64 ASIC
      13    1
      14    4
      16    4
      19    3
      21    2
      FEE64 Rate spectra - attachment 12 - some channels recovered

08.45 ADC data synchronised - attachment 13

08.52 zero all histograms
      zero DAQ stats


08.56 ASIC settings saved to new DB key 2019Jun13-08.56.35
      includes slow comparator 0xa
      DAQ options updated

15:18 
      DAQ continues okay.  Rates seem ok, see attachment 14.
      Biases shown as attachment 15 -- looks fine.

      (These are all as before we assume)
      ASIC settings 2019May19-17.54.03
      slow comparator 0xa

      BNC PB-4 Pulser 
      Amplitude 90,000
      x5 attenuator IN
      Frequency 25Hz range
      t_r 100ns, tau_d 50us
      - polarity
      Delay MIN, tail pulse

      all system wide checks OK *except* nnaida7 fails ADC calibration

      FEE64 temperatures OK - see attachment 16

      Merge ok, 570k items per second, see attachment 17
      TapeServer not storing but sitting around 4 MBytes/sec, see attachment 18
  838   Wed Jun 12 01:54:04 2019 OH, TD, CG, CB, CAWednesday 12 June
09:54 DAQ Continues ok with merger running at ~ 560k data items per second

      ASIC settings 2019May19-17.54.03
      slow comparator 0xa

      BNC PB-4 Pulser 
      Amplitude 90,000
      x5 attenuator IN
      Frequency 25Hz range, variable min. c. 2.5Hz
      t_r 100ns, tau_d 50us
      - polarity
      Delay MIN, tail pulse

      System wide checks all ok - except nnaida7 fails adc calibration

      Bias and leakage currents ok - attachment 1

      FEE Temps ok - attachment 2

      Good event statistics good - attachment 3

      Merger and tape service good - attachment 4 and 5


16:33  DAQ continues OK with merger running at ~ 600k data items per second

     ASIC settings 2019May19-17.54.03
      slow comparator 0xa

      BNC PB-4 Pulser 
      Amplitude 90,000
      x5 attenuator IN
      Frequency 25Hz range, variable min. c. 2.5Hz
      t_r 100ns, tau_d 50us
      - polarity
      Delay MIN, tail pulse

      All system-wide checks OK except nnaida7 fails ADC calibration

      Bias and leakage currents OK - attachment 6

      FEE temperatures OK - attach 7

      Good event statistics OK - attach 8
      check nnaida10 & nnaida20 counts vs. rate

      Merger and tape service OK - attachment 9 & 10

17.00 Average FEE64 rates since last DAQ stats zero @ 17.05 yesterday - data from attachment 11

      DSSSD#1   count      rate  
                           24h   most recent
      nnaida17 	1914894752 22.1k 23.2k
      nnaida20  3061905664 35.4k 35.6k
      nnaida22  1872820000 21.7k 21.8k
      nnaida23  2217883584 25.7k 24.8k  

      DSSSD#2

      nnaida18  1760374528 20.3k 21.1k
      nnaida19  1402771232 16.2k 66.6k => 32 bit overflow => 4.3E+09 + 1.4E+09 = 5.6E+09 => 65.9k
      nnaida21  2038432092 23.6k 23.5k
      nnaida24  2844103808 32.9k 31.3k

      Conclude rates have bee stable overnight - no evidence of noise transients for DSSSD#1 to 150-250k good
events/s

17.15 DAQ stats zero'd
      all spectra zero'd
  837   Tue Jun 11 01:45:55 2019 CA, CB, TD, CG, OH, DKTuesday 11 June
09:46 DAQ continues ok

      all system wide checks OK *except* nnaida7 fails ADC calibration
      
      detector biases & leakage currents OK - see attachment 1

      FEE64 temperatures OK - see attachment 2

      good event statistics ok - attachment 3

10:00 Merger & TapeServer ok - attachments 4 & 5

15.13 DAQ continues OK

      ASIC settings 2019May19-17.54.03
      slow comparator 0xa

      BNC PB-4 Pulser 
      Amplitude 90,000
      x5 attenuator IN
      Frequency 25Hz range, variable min. c. 2.5Hz
      t_r 100ns, tau_d 50us
      - polarity
      Delay MIN, tail pulse

      all system wide checks OK *except* nnaida7 fails ADC calibration

      detector biases & leakage currents OK - see attachments 6 & 7

      FEE64 temperatures OK - see attachment 8

      good event statistics - DSSSD#1 noisy - see attachments 9
      7x <20k 15x < 50k 20 < 100k cf. 8x <20k, 23x <50k, 24x <100k earlier today

      Check ASIC Control x2

      good event statistics - DSSSD#1 noisy - see attachments 10
      8x <20k 5x < 50k 20 < 100k cf. 8x <20k, 23x <50k, 24x <100k earlier today

      A, B. C & D Samtec connectors pressed into adaptor PCBs

15.41 good event statistics - OK - see attachments 11
      8x <20k 19x < 50k 20x < 100k

16.53 good event statistics - OK - see attachments 11
      8x <20k 19x < 50k 20x < 100k

17.04 p+n FEE64s 1.8.W spectra 20us FSR - DSSSD#1 noisy - attachment 14
      n+n FEE64s 1.8.W spectra 20us FSR - DSSSD#1 noisy - attachment 15
      1.8.L spectra - see attachments 12 & 13 
      nnaida1 75 ch FWHM, nnaida3 79 ch FWHM - cf. 76 & 83 ch FWHM respectively yesterday

17.05 all spectra zero'd
      DAQ stats zero'd

17.07 Julabo FL11006 chiller set point +20 deg C actual +20.0 deg C
      Water level c. 80% M-H markers 
      Temperature +23.4 deg C dew point +9.6 deg C RH 42.0%

17.10 leakage currents/detector bias ok - attachment 17
  836   Mon Jun 10 07:39:35 2019 CA, CB, TD, CG, OH, DKVarying per DSSSD slow comparator thresholds
DSSSDs #1-6 slow comparator 0xa = 100keV - attachments 1-4

DSSSD2: attachments 5 to 8 (0xa to 0x8)

DSSSD3: attachments 9 to 12 (0xa to 0x8)

DSSSD4: attachments 13 to 16 (0xa to 0x8)

DSSSD5: attachments 17 to 20 (0xa to 0x8)

Global rates are >500 to <1000k items per second.  

Set all back to 0xa.

Now we try going to 60 keV (0x6)

DSSSD2: attachments 21 to 24 (0xa to 0x6) global ~1100k

DSSSD3: attachments 25 to 28 (0xa to 0x6) global ~1000k

DSSSD4: attachments 29 to 32 (0xa to 0x6) global ~1100k

Set all back to 0xa again.
  835   Mon Jun 10 01:43:16 2019 CA, CB, TD, CG, OH, DKMonday 10 June
09:45 DAQ continues ok 

      all system wide checks OK *except* nnaida7 fails ADC calibration
      
      detector biases & leakage currents OK - see attachment 1

      FEE64 temperatures OK - see attachment 2

      Good event statistics prior to reload - glitch with display? - attachment 3

      Reload done - good event statistics ok - attachment 4

10.00 Merger & TapeServer ok - attachments 5 & 6

15.14 DAQ continues OK

      ASIC settings 2019May19-17.54.03
      slow comparator 0xa

      BNC PB-4 Pulser 
      Amplitude 90,000
      x5 attenuator IN
      Frequency 25Hz range, variable min. c. 2.5Hz
      t_r 100ns, tau_d 50us
      - polarity
      Delay MIN, tail pulse

      all system wide checks OK *except* nnaida7 fails ADC calibration

      detector biases & leakage currents OK - see attachments 7 & 8

      FEE64 temperatures OK - see attachment 9

      good event statistics - DSSSD#1 noisy - see attachments 10
      8x <20k, 23x <50k, 24x <100k cf. same as earlier today
      Merger OK - see attachment 11
      TapeServer OK - see attachment 12

      p+n FEE64s 1.8.W spectra 20us FSR - DSSSD#1 noisy - attachment 13
      n+n FEE64s 1.8.W spectra 20us FSR - DSSSD#1 noisy - attachment 14
      1.8.L spectra - see attachments 9 & 10
      nnaida1 73 ch FWHM, nnaida3 78 ch FWHM - cf. 76 & 83 ch FWHM respectively yesterday morning

      FEE64 Rate spectra - attachment 15
      Check ASIC Control *all* FEE64s & ASICs x2
      FEE64 Rate spectra - attachment 16

15.26 ADC data synchronised - attachment 17

16.55 Julabo FL11006 chiller set point +20 deg C actual +20.0 deg C
      Water level c. 80% M-H markers - topup to 100% with 0.5x jugful water
      Temperature +24.6 deg C dew point +8.1 deg C RH 34.5%
  834   Sun Jun 9 03:43:00 2019 CB, TD, CG, OHSunday 9 June
11.43 DAQ continues OK

      ASIC settings 2019May19-17.54.03
      slow comparator 0xa

      BNC PB-4 Pulser 
      Amplitude 90,000
      x5 attenuator IN
      Frequency 25Hz range, variable min. c. 2.5Hz
      t_r 100ns, tau_d 50us
      - polarity
      Delay MIN, tail pulse

      all system wide checks OK *except* nnaida7 fails ADC calibration

      detector biases & leakage currents OK - see attachments 1 & 2

      FEE64 temperatures OK - see attachment 3

      good event statistics - DSSSD#1 noisy - see attachments 4
      10x <20k, 19x <50k, 20x <100k cf. 13x <20k 22x < 50k 24x <100k yesterday afternoon
      Merger OK - see attachment 5
      TapeServer OK - see attachment 6

      p+n FEE64s 1.8.W spectra 20us FSR - DSSSD#1 noisy - attachment 7
      n+n FEE64s 1.8.W spectra 20us FSR - DSSSD#1 noisy - attachment 8
      1.8.L spectra - see attachments 9 & 10
      nnaida1 76 ch FWHM, nnaida3 83 ch FWHM - cf. 104 & 111 ch FWHM respectively yesterday aftenoon

      FEE64 Rate spectra - attachment 11
      Check ASIC Control *all* FEE64s & ASICs x2
      FEE64 Rate spectra - attachment 12

11.46 ADC data synchronised - attachment 13

      p+n FEE64s 1.8.W spectra 20us, 200us, 2ms & 20ms FSR - DSSSD#1 noisy - attachments 14-17

      Check ASIC Control *all* FEE64s & ASICs x2   - no change DSSSD#1 noise

      DSSSD#1 bias cycle - no change DSSSD#1 noise

      press Kapton Samtec into adaptor PCBs - reduces DSSSD#1 noise to previous levels

12.17 Julabo FL11006 chiller set point +20 deg C actual +20.0 deg C
      Water level c. 60% M-H markers
      Temperature +25.3 deg C dew point +7.6 deg C RH 32.2%

12.23 temporary fix for DSSSD#1 noise? check by
      zero all histograms
      zero DAQ stats

      ...and wait

      good event statistics - DSSSD#1 noisy - see attachments 18
      11x <20k, 21x <50k, 23x <100k

13.30 DAQ stop

      OH replaces LH kapton PCB for DSSSD#1 n+n FEE64s 20 & 23 which we suspect may be damaged due to twist
      placed on kpaton cables by weight of all ribbon cables + cu foil shielding - particularly pronounced
      for nnaida20

      good event statistics - OK - see attachments 19
      7x <20k, 20x <50k, 24x <100k
      p+n FEE64s 1.8.W spectra 20us FSR - OK - attachment 20

      good event statistics - OK - see attachment 21
      7x <20k, 22x <50k, 24x <100k
      p+n FEE64s 1.8.W spectra 20us FSR - OK - attachment 22

17.50 detector biases & leakage currents OK - see attachments 23 & 24

      FEE64 temperatures OK - see attachment 25

      good event statistics - OK - see attachments 26
      9x <20k, 21x < 50k, 24x <100k cf. 10x <20k, 19x <50k, 20x <100k earlier today

      p+n FEE64s 1.8.W spectra 20us FSR - DSSSD#1 noisy - attachment 7
      n+n FEE64s 1.8.W spectra 20us FSR - DSSSD#1 noisy - attachment 8

17.57 all histograms zero'd
      DAQ stats zero'd
  833   Sat Jun 8 06:08:52 2019 CB, TDSaturday 8 June
14.14 Julabo FL11006 chiller set point +20 deg C actual +20.0 deg C
      Water level c. 70% M-H markers
      Temperature +25.3 deg C dew point +8.5 deg C RH 34.7%

      DAQ continues OK

      ASIC settings 2019May19-17.54.03
      slow comparator 0xa

      BNC PB-4 Pulser 
      Amplitude 90,000
      x5 attenuator IN
      Frequency 25Hz range
      t_r 100ns, tau_d 50us
      - polarity
      Delay MIN, tail pulse

      all system wide checks OK *except* nnaida7 fails ADC calibration

      detector biases & leakage currents OK - see attachments 1 & 2

      FEE64 temperatures OK - see attachment 3

      good event statistics OK - see attachments 4
      13x <20k 22x < 50k 24x <100k cf. 13x <20k 21x < 50k 24x <100k yesterday afternoon
      Merger OK - see attachment 5
      TapeServer OK - see attachment 6

      p+n FEE64s 1.8.W spectra 20us FSR - attachment 7
      n+n FEE64s 1.8.W spectra 20us FSR - attachment 8
      1.8.L spectra - see attachments 9 & 10
      nnaida1 104 ch FWHM, nnaida3 111 ch FWHM - cf. 80 & 87 ch FWHM respectively yesterday aftenoon

      FEE64 Rate spectra - attachment 11
      Check ASIC Control *all* FEE64s & ASICs x2
      FEE64 Rate spectra - attachment 12 - some channels recovered

14.29 ADC data synchronised - attachment 13

14.31 zero all histograms
      zero DAQ stats
  832   Fri Jun 7 05:58:11 2019 TDper FEE64 rate/20.48us versus slow comparator threshold
per FEE64 rate/20.48us & 'good event' DAQ statistics 

slow comparator threshold

100keV - attachments 1 & 2
110keV - attachments 3 & 4
120keV - attachments 5 & 6
150keV - attachments 7 & 8
200keV - attachments 9 & 10
  831   Fri Jun 7 05:30:23 2019 TDFriday 7 June
14.05 Julabo FL11006 chiller set point +20 deg C actual +20.0 deg C
      Water level c. 70% M-H markers
      Temperature +25.6 deg C dew point +10.3 deg C RH 37.7%

13.30 DAQ continues OK

      ASIC settings 2019May19-17.54.03
      slow comparator 0xa

      BNC PB-4 Pulser 
      Amplitude 90,000
      x5 attenuator IN
      Frequency 25Hz range
      t_r 100ns, tau_d 50us
      - polarity
      Delay MIN, tail pulse

      all system wide checks OK *except* nnaida7 fails ADC calibration

      detector biases & leakage currents OK - see attachments 1 & 2

      FEE64 temperatures OK - see attachment 3

      good event statistics OK - see attachments 4
      13x <20k 21x < 50k 24x <100k cf. same yesterday afternoon
      Merger OK - see attachment 5
      TapeServer OK - see attachment 6

      p+n FEE64s 1.8.W spectra 20us FSR - attachment 7
      n+n FEE64s 1.8.W spectra 20us FSR - attachment 8
      1.8.L spectra - see attachments 9 & 10
      nnaida1 80 ch FWHM, nnaida3 87 ch FWHM - cf. 96 & 103 ch FWHM respectively yesterday aftenoon

      FEE64 Rate spectra - attachment 11
      Check ASIC Control *all* FEE64s & ASICs x2
      FEE64 Rate spectra - attachment 12 - some channels recovered

14.20 BNC PB-4 pulser
      Frequency changed from c. 4Hz to c. 2.5Hz (minimum)
      25Hz range, variable fully anti-clockwise

14.33 zero *all* spectra
      zero DAQ stats

      ADC data synchronised - attachment 13

16.50 DAQ start file May19/R18

      AIDA+DTAS merger test - 12kBq 60Co positioned on top of AIDA snout

      DAQ sync issued by DTAS during R18_0

      ASIC settings 2019May19-17.54.03
      slow comparator 0xa

      BNC PB-4 Pulser 
      Amplitude 90,000
      x5 attenuator IN
      Frequency 25Hz range
      t_r 100ns, tau_d 50us
      - polarity
      Delay MIN, tail pulse

17.28 DAQ stop file May19/R18_4

17.40 DAQ start file May19/R19

      AIDA+DTAS merger test - c. 30kBq? 137Cs positioned on top of AIDA snout

      DAQ sync issued by DTAS during R18_0

      ASIC settings 2019May19-17.54.03
      slow comparator 0xa

      BNC PB-4 Pulser 
      Amplitude 90,000
      x5 attenuator IN
      Frequency 25Hz range
      t_r 100ns, tau_d 50us
      - polarity
      Delay MIN, tail pulse

18.18 DAQ stop file May19/R19_5

19.36 all system wide checks OK *except* nnaida7 fails ADC calibration

      detector biases & leakage currents OK - see attachments 14 & 15

      FEE64 temperatures OK - see attachment 16

      good event statistics OK - see attachments 17
      12x <20k 21x < 50k 24x <100k cf. 13x <20k 21x < 50k 24x <100k earlier this afternoon

19.45 Julabo FL11006 chiller set point +20 deg C actual +20.0 deg C
      Water level c. 70% M-H markers
      Temperature +25.1 deg C dew point +7.9 deg C RH 33.6%
      
  830   Thu Jun 6 08:55:15 2019 TDThursday 6 June
16.40 DAQ continues OK

      ASIC settings 2019May19-17.54.03
      slow comparator 0xa

      BNC PB-4 Pulser 
      Amplitude 90,000
      x5 attenuator IN
      Frequency 25Hz range
      t_r 100ns, tau_d 50us
      - polarity
      Delay MIN, tail pulse

      all system wide checks OK *except* nnaida7 fails ADC calibration

      detector biases & leakage currents OK - see attachments 1 & 2

      FEE64 temperatures OK - see attachment 3

      good event statistics OK - see attachments 4
      13x <20k 21x < 50k 24x <100k cf. same yesterday afternoon
      Merger OK - see attachment 5
      TapeServer OK - see attachment 6

      p+n FEE64s 1.8.W spectra 20us FSR - attachment 7
      n+n FEE64s 1.8.W spectra 20us FSR - attachment 8
      1.8.L spectra - see attachments 9 & 10
      nnaida1 96 ch FWHM, nnaida3 103 ch FWHM - cf. 88 & 92 ch FWHM respectively yesterday afternoon - now
observe high amplitude tailing on pulser peaks

      FEE64 Rate spectra - attachment 11
      Check ASIC Control *all* FEE64s & ASICs x2
      FEE64 Rate spectra - attachment 12 - some channels recovered

16.59 zero *all* spectra
      zero DAQ stats

      ADC data synchronised - attachment 11
  829   Wed Jun 5 06:57:38 2019 TDWednesday 5 June
14.50 DAQ continues OK

      ASIC settings 2019May19-17.54.03
      slow comparator 0xa

      BNC PB-4 Pulser 
      Amplitude 90,000
      x5 attenuator IN
      Frequency 25Hz range
      t_r 100ns, tau_d 50us
      - polarity
      Delay MIN, tail pulse

      all system wide checks OK *except* nnaida7 fails ADC calibration

      detector biases & leakage currents OK - see attachments 1 & 2

      FEE64 temperatures OK - see attachment 3

      good event statistics OK - see attachments 4
      13x <20k 21x < 50k 24x <100k cf. same yesterday afternoon
      Merger OK - see attachment 5
      TapeServer OK - see attachment 6

      p+n FEE64s 1.8.W spectra 20us FSR - attachment 7
      n+n FEE64s 1.8.W spectra 20us FSR - attachment 8
      1.8.L spectra - see attachments 9 & 10
      nnaida1 88 ch FWHM, nnaida3 92 ch FWHM - cf. 75 & 82 ch FWHM respectively yesterday afternoon

      FEE64 Rate spectra - attachment 11
      Check ASIC Control *all* FEE64s & ASICs x2
      FEE64 Rate spectra - attachment 12 - some channels recovered

15.09 zero *all* spectra
      zero DAQ stats

      ADC data synchronised - attachment 11
  828   Tue Jun 4 07:10:27 2019 TDTuesday 4 June
15.01 Julabo FL11006 chiller set point +20 deg C actual +20.0 deg C
      Water level c. 70% M-H markers - topup to 100% with 1x jugful water
      Temperature +25.8 deg C dew point +8.0 deg C RH 32.4%

      Access to F11 from 14.00 whilst F3 & F7 plastic scintillator and PPAC are replaced due to rad damage
during online experiment

      DAQ continues OK

      ASIC settings 2019May19-17.54.03
      slow comparator 0xa

      BNC PB-4 Pulser 
      Amplitude 90,000
      x5 attenuator IN
      Frequency 25Hz range
      t_r 100ns, tau_d 50us
      - polarity
      Delay MIN, tail pulse

      all system wide checks OK *except* nnaida7 fails ADC calibration

      detector biases & leakage currents OK - see attachments 1 & 2

      FEE64 temperatures OK - see attachment 3

      good event statistics OK - see attachments 4
      13x <20k 21x < 50k 24x <100k cf. 15x <20k, 21x < 50k, 24x < 100k yesterday evening

      Merger OK - see attachment 5
      TapeServer OK - see attachment 6

      p+n FEE64s 1.8.W spectra 20us FSR - attachment 7
      n+n FEE64s 1.8.W spectra 20us FSR - attachment 8
      1.8.L spectra - see attachments 9 & 10
      nnaida1 75 ch FWHM, nnaida3 82 ch FWHM - cf. 95 & 99 ch FWHM respectively yesterday afternoon

      FEE64 Rate spectra - attachment 11
      Check ASIC Control *all* FEE64s & ASICs x2
      FEE64 Rate spectra - attachment 12 - some channels recovered

      ADC data synchronised - attachment 11
  827   Mon Jun 3 05:40:19 2019 TDMonday 3 June
13.40 DAQ continues OK

      ASIC settings 2019May19-17.54.03
      slow comparator 0xa

      BNC PB-4 Pulser 
      Amplitude 90,000
      x5 attenuator IN
      Frequency 25Hz range
      t_r 100ns, tau_d 50us
      - polarity
      Delay MIN, tail pulse

      all system wide checks OK *except* nnaida7 fails ADC calibration

      detector biases & leakage currents OK - see attachments 1 & 2
      FEE64 temperatures OK - see attachment 3
      good event statistics OK - see attachments 4
      14x <20k, 21x <50k, 24x <100k - cf. 8x <20k, 15x <50k, 21x <100k yesterday evening

      Merger OK - see attachment 5
      TapeServer OK - see attachment 6

      1.8.L spectra - see attachments 7 & 8
      nnaida1 95 ch FWHM, nnaida3 99 ch FWHM - cf. 75 & 76 ch FWHM respectively yesterday afternoon
      p+n FEE64s 1.8.W spectra 20us FSR - attachment 9
      n+n FEE64s 1.8.W spectra 20us, 200us, 2ms & 20ms FSR - attachments 10-13

      Check ASIC Control *all* FEE64s & ASICs
      zero all spectra

14.11 MIDASsort running online (sampling TapeServer buffers, 5% say) overnight (last time spectra
      were zero'd) observed high energy heavy ions in DSSSD #1-3 - see attachments 14-17

      Currently high energy, heavy ion rate is zero.

19.58 DAQ continues OK

      ASIC settings 2019May19-17.54.03
      slow comparator 0xa

      BNC PB-4 Pulser 
      Amplitude 90,000
      x5 attenuator IN
      Frequency 25Hz range
      t_r 100ns, tau_d 50us
      - polarity
      Delay MIN, tail pulse

      all system wide checks OK *except* nnaida7 fails ADC calibration

      detector biases & leakage currents OK - see attachments 18 & 19
      FEE64 temperatures OK - see attachment 20
      good event statistics OK - see attachments 21
      15x <20k, 21x < 50k, 24x < 100k - cf. 14x <20k, 21x <50k, 24x <100k at 13.40 today

      Merger OK - see attachment 22
      TapeServer OK - see attachment 23

      p+n FEE64s 1.8.W spectra 20us FSR - attachment 24
      n+n FEE64s 1.8.W spectra 20us FSR - attachment 25