Sun Mar 3 16:06:11 2024, TD, Sunday 3 March 24x
|
17.00 FEE64 41:a0:71 ASIC #2 u/s
Replaced ASIC mezzanine with new ASIC mezzanine
Installed as aida04
Unable to boot nnrpi2 - manual/local control of BNC PB-5 and FEE64 PSU - no interlock
Water temperature & pressure as measured outside S4 area - OK
Manually power FEE64s
BNC PB-5 local control/ON
Amplitude 1.0V
Attenuation x10
Polarity -
tau_d 1ms
Frequency 22Hz
FEE64 temps OK - attachment 1
All system wide checks OK
WR timestamp OK - attachment 2
aida04 ASIC settings - attachment 3
ADC data item stats OK - attachment 4
aida04 Rate spectrum - attachment 5
aida04 *.*.L spectra - attachments 6-9
1.8.L pulser peak width 15 ch FWHM
aida04 1.8.W spectra - 20us FSR - attachment 10-11
MIDAS configuration - attachment 12
DHCP config - attachment 13
18.30 FEE64 41:f6:b7 ASIC #1 u/s
Replaced ASIC mezzanine with new ASIC mezzanine
Installed as aida02
All system wide checks OK
FEE64 temps - attachment 14
aida02 ASIC temperature c. 512 deg C !
PSU/Virtex temps OK
Mezzanine ambient to touch - assume faulty sensor/poor connection?
aida02 ASIC settings - attachment 15
WR timestamp OK - attachment 16
ADC data item stats OK - attachment 4
aida04 Rate spectrum - attachment 18
aida02 1.8.W spectra - 20us FSR - attachment 19-20
aida02 *.*.L spectra - attachments 21-24
1.8.L pulser peak width 16 ch FWHM
19.30 Power OFF
|
Mon Feb 26 13:17:51 2024, TD, Offline analysis of DEC23/R9_6 9x
|
First pass analysis of data file /TapeData/DEC23/R9_6
*** TDR format 3.3.0 analyser - TD - May 2021
*** ERROR: READ I/O error: 5002
blocks: 32000
ADC data format: 234532726 ( 99567.1 Hz)
Other data format: 27387282 ( 11626.8 Hz)
Sample trace data format: 0 ( 0.0 Hz)
Undefined format: 0 ( 0.0 Hz)
Other data format type: PAUSE: 53 ( 0.0 Hz)
RESUME: 53 ( 0.0 Hz)
SYNC100: 40775 ( 17.3 Hz)
WR48-63: 40775 ( 17.3 Hz)
FEE64 disc: 2928025 ( 1243.0 Hz)
MBS info: 24377601 ( 10349.1 Hz)
Other info: 0 ( 0.0 Hz)
ADC data range bit set: 287073 ( 121.9 Hz)
Timewarps: ADC: 0 ( 0.0 Hz)
PAUSE: 0 ( 0.0 Hz)
RESUME: 0 ( 0.0 Hz)
SYNC100: 0 ( 0.0 Hz)
WR48-63: 0 ( 0.0 Hz)
FEE64 disc: 0 ( 0.0 Hz)
MBS info: 0 ( 0.0 Hz)
Undefined: 0 ( 0.0 Hz)
Sample trace: 0 ( 0.0 Hz)
*** Timestamp elapsed time: 2355.524 s
*** Statistics
FEE ADC Data Other Data Sample Undefined Pause Resume SYNC100 WR48-63 Disc MBS Other HEC Data
0 59013868 372005 0 0 20 20 9888 9888 281563 70626 0 55467
1 31445079 1098714 0 0 2 2 5017 5017 243009 845667 0 38876
2 22895499 11841669 0 0 1 1 5294 5294 482655 11348424 0 42748
3 18228498 11705354 0 0 0 0 3913 3913 416619 11280909 0 78526
4 25500772 1237178 0 0 3 3 4320 4320 396557 831975 0 28918
5 4 0 0 0 0 0 0 0 0 0 0 0
6 15933921 394210 0 0 0 0 2560 2560 389090 0 0 11489
7 61515085 738152 0 0 27 27 9783 9783 718532 0 0 31049
FEE64 configuration https://elog.ph.ed.ac.uk/AIDA/935
Note
- aida06 not connected to DSSSD ( cabling broken - will be replaced later this week )
- LEC ( 20MeV FSR ) data ADC offset corrected
ADC offsets available for 489 of 512 channels and are included in the analysis - 23 strips for which no ADC offsets could be calculated ( usually because there was
no
pulser data ) are not included in analysis
- LEC ( 20MeV FSR ) front-back strip energy difference cut +/- 50 channels ( c. +/- 280keV )
- HEC ( 2GeV FSR ) front-back strip energy difference cut +/-200 channels ( c. +/- 1120MeV )
- LEC ADC data: 13 < channel < 188 ( c. 73-1053keV )
- HEC ADC data: > 13 channels ( c. 73MeV )
- FEE64 hardware thresholds: LEC c. 100keV, HEC c. 200MeV
Attachments 1 & 2 - per pixel HEC-LEC event time difference spectra - 4.096us/channel
Attachments 3 & 4 - per pixel HEC-LEC event time difference spectra - 65.536us/channel
Attachment 5 - DSSSD x-y hit pattern: HEC-LEC event time difference < 4.3s
Majority of events associated with 'hot' p+n junction strips - few plausible decay candidate events - as expected.
z-scale - semi-logarithmic - scattered events are single counts ( blue )
1 count => rate ~ 1/2355s ~ 0.0004Hz, the great majority of pixels have zero counts
Attachment 6 - LEC ( decay ) and HEC ( implant ) events - 262.144us/channel ( 65536 channels = 17.2s )
HEC implants channels 12000-25000 = 216 counts => HEC rate ( in spill ) 63.4Hz
LEC decays channels 12000-25000 = 3865 counts => LEC rate ( in spill ) 1134.1Hz
LEC decays channels 25000-38000 = 2790 counts => LEC rate ( inter spill ) 818.7Hz
Attachment 7 - per pixel HEC-HEC event time difference spectra - 4.096us/channel
Attachment 8 - per pixel LEC-LEC event time difference spectra - 4.096us/channel
Attachment 9 - variables.dat
This is a ( Fortran ) Namelist I/O data file containing of the ADC offsets, FEE64 configuration, LEC and HEC energy difference windows
The ADC offset channel number is calculated as
channel = channel_ident + ( module * 64 ) where module = 0-7 corresponding to AIDA FEE64s aida01-aida08
and is used as follows
ADC data = INT ( RSHIFT( ABS( 32768 - data( i ) ), 3 ) - offset( i ) + 0.5 )
where data(i) is the ADC data item for channel i, offset(i) is the ADC offset for channel i
An ADC offset of -9999 means there was no pulser data for this channel in data files R5 and R11.
|
Wed Jan 31 15:27:07 2024, TD, Photos of water flow & dew point sensor interlock box   
|
Attachment 1 - 4 socket input connectors ( from sensor to interlock box )
Attachment 2 & 2 - 24V dc relays and internal wiring
Attachment 4 - 3 socket input connector ( from interlock box to USB-controlled ac mains relay ) |
Mon Nov 13 11:46:54 2023, TD, Summary of AIDA 24cm x 8cm 'triple' DSSSD tests 
|
|
Thu Oct 12 14:01:43 2023, TD, MSL type BB18 24cm x 8cm DSSSD test - update 
|
DSSSD MSL type BB18 24cm x 8cm 3208-3/3208-21/3208-22
FEE64 configuration see https://elog.ph.ed.ac.uk/AIDA/872 attachment 2
Bias -150V -6.590uA ambient temperature +24.7 deg C d.p. +13.7 deg C RH 50.3%
BNC PB-5
amplitude 10.0V
attenuation x10
decay time 1ms
tail pulse
frequency 25Hz
PB-5 output direct to p+n junction side FEE64 aida01 or aida12, or n+n Ohmic side FEE64 aida02
aida01 1.8.L pulser peak width 61 ch FWHM ~ 46keV FWHM => 5s threshold 98keV
aida12 1.8.L pulser peak width 56 ch FWHM ~ 42keV FWHM => 5s threshold 89keV
aida02 1.8.L pulser peak width 102 ch FWHM ~ 77keV FWHM => 5s threshold 163keV
slow comparator 0xa 100keV ( all p+n junction FEE64s )
0xf 150keV ( all n+n Ohmic FEE64s )
per FEE64 Rate spectra - attachment 1
p+n FEE64s ( aida010, aida01, aida09, aida12, aida03, aida11 ) rates dominated by hot channels, other channels typically <1Hz ( 25Hz pulser to aida12 )
n+n FEE64s ( aida02, aida04 ) rates ~ 10-20Hz/channel
Note aida06 and aida08 are not connected to anything and should be ignored
ADC data item stats - attachment 2
For further information see https://elog.ph.ed.ac.uk/AIDA/906 and https://elog.ph.ed.ac.uk/AIDA/907
To Do
- repair/replace Honeywell HSS-DPS dew point sensor
USB-controlled ac mains relay interlock currently overrriden
do NOT operate AIDA unattended
- aida04 asic #1 u/s - replace ASIC mezzanine
- electrically isolated test signal distribution box req'd
- aida10 asic #4
v. high rates observed and large signal transients
cause unclear ... ASIC/adaptor PCB/cabling/Si wafer ?
- extended background alpha run to check all DSSSD bond wires
pulser OFF
slow comparator 0x64
- bPlas + 2x triple DSSSD + bPlas stack assembly and test
- all up, in beam test |
Sun Oct 1 11:47:07 2023, TD, S4 cooling water 
|
The photographs show the cooling water controls and temperature/pressure gauges outside S4 and the connections used by AIDA within S4. |
Thu Aug 31 15:24:56 2023, NH, New AIDA MBS PC
|
The AIDA MBS FDR will be x86l-119 from now on, not x86l-94
the MBS relay and startup scripts will be changed for this |
Mon Aug 28 12:47:56 2023, NH, Power Failure 24.08.2023
|
There was a power failure in the morning of 24.08.2023 in the Rhein-Main area affecting GSI
The Aida workstation (aida-3) has been restarted, it is unknown if the Pis in S4 rebooted as well (there is a UPS)
29.8.23 TD Both RPi systems rebooted four days ago. |
Sun Jul 16 09:32:57 2023, TD, Firefox browser proxy setting change
|
Firefox browser proxy setting changed to 'Auto-detect proxy settings for this network' ( Firefox -> Edit -> Settings -> Network Settings -> Settings ) |
Wed Jan 18 13:40:33 2023, PJCS TD, MACB settings with either Emulator or VITAR   
|
When using the VETAR connected to the HDMI port of the root MACB the settings for all the MACB in the system 0x3.
This setting takes the Clock and Data line from the input HDMI and outputs it via teh HDMI output ports.
When using the Emulator connected via the SMA connectors on the back of the root MACB then the setting for the root MACB should be 0xD and all others should be 0x3.
Attached is the .jed file for programming the MACB and the .vhd source file to help with understanding of the settings. |
Fri Dec 16 14:02:12 2022, NH, AIDA System off for christmas break
|
The AIDA NIM crate, pis and workstation have been powered off for the Christmas break and will not be accessible |
Thu Oct 6 16:51:49 2022, NH, Oscilloscope analysis 24x
|
Investigating AIDA noise with a TA041 differential probe and oscilloscope
AC Mains (DESPEC platform AC, L-N)
Probe attentuation = 1:100
Fig 1: Main AC waveform [X: 5ms/div, Y: 100 V/div]
Fig 2: Zoomed in at peak (20 V FSR, any less and the waveform clipped) [X: 10us/div, Y: 20V/div]
Fig 3: Longer time base and FFT of 0-5 MHz. No significant frequency harmonics noticed [X: 5ms/div, Y:20 V, FFT X: 500 kHz/div, Y: 10 dBm/div]
No significant noise or distortion present, fully within any AC specification.
Note that at the moment there is almost no load on AC
Equipment on on DESPEC rack: AIDA NIM, AIDA Raspberry Pis, bPlas PC (+ WR) + 2x DESPEC NIM crates
No autofill, VME crate or detectors
All big machines at GSI (SIS, FRS) off (suspect pumps are on)
Ion catcher not on (I think under repair)
-
FEE PSU studies
Probe connected to 5V exposed power pin on FEE64 (+v) and to grounding crimp on FEE64 (-v)
No adapter board connected
Attentuation = 1:10
Fig 4: FFT when FEEs are *off* - essentially probe+scope noise [X: 5ms/div, Y: 100 mV/div, FFT: X: 500 kHz/div, Y: 10 dBm/div]
Fig 5: FFT when FEEs are *on* - notice 1.4 MHz peak in FFT, also seen on ADC waveform readout before (fig 6) [X: 5ms/div, Y: 100 mV/div, FFT: X: 500 kHz/div, Y: 10 dBm/div]
Fig 7: 500 ns/div 5V output on FEE, single FEE on the PSU [X: 500 ns/div, Y: 100 mV/div]
Note average max voltage is 5.31 V (power on) and ~ 70 mV "peak to peak" -might be from probe/scope?
Also see voltage changes with FEE power draw:
Power on : 5.45 V (different scale to above)
SETUP ran : 5.51 V
FADCs off : 5.86 V
ACQ Go: : Unchanged; ASIC threshold 0xa: Unchanged
Also check situation on a fully loaded PSU (8 fees connected and powered on)
Power on: 5.29 V (fig 8)
SETUP ran: 5.36 V (fig 9)
FADCs off: 5.64 V (fig 10)
All X: 500 ns/div, Y: 100 mV/div
Both cases observe voltage rises as current draw drops (as expected for voltage drop along a cable)
Noise on 'scope seems to get slightly worse with reduced current (and higher voltage)
No sign of strong 100 kHz noise as seen in ADC traces beforehand
Todo:
- Check -6V and 7V rails
- Check 5V and noise when front-end card is added and pulser/HV connected
- Check between two FEE64 grounds
- Check direct out of PSU vs ground to see if 1.4 MHz appears on PSU side or FEE64 side
-
11.10.22 Updates
Attachement 11 - 5V PSU on upper PSU with no FEEs attached whatsoever. No 1.4 MHz (on FFT) but clear low frequency beats from switching - presumably low/no load behaviour
Attachement 12 - 5V PSU on aida12 with 8 FEEs on PSU. Longer time base to allow lower frequencies in FFT. 1.4 MHz switching spikes visible but nothing around 100 kHz region
Attachments 13-16: 5V PSU on aida12 at 20 mV/div vertical and 1, 0.5, 2, 5 us/div horizontal respectively
12.10.22 Updates
Attachment 17: -6V PSU on aida12 with 8 FEEs on PSU. 2 us/div timebase. 20 mV/div amplitude
Attachment 18: 10 ms timebase and FFT
Average voltage: -6.21 V
Attachment 19: 7V PSU on aida12 with 8 FEEs on PSU. 2 us/div timebase. 20 mV/div amplitude
Attachment 18: 10 ms timebase and FFT
Average voltage: 7.46V
Measurement between AIDA12 ground and Reference ground/copper bar
+ve (red probe) attached to copper bar at ground point (not strong connection at present)
-ve (black probe) attached to ground crimp on aida12 (connected to cooling plate)
aida12 no adapter board connected: connections are PSU, Ethernet, HDMI and TTY only
Attachment 21: 5 us/div 100 mv/div waveform, big oscillations present. Not seen before FEEs turned on (8 FEES, 1-7+12)
Attachment 22: 10 ms/div for FFT, sharp peak at exactly 100 kHz observed...
Attachment 23: Between 5V PSU (+ve) and 19" rack (-ve) with no FEEs connected to PSU
See strong 100 kHz oscillations too, note that voltage isn't 5V as PSU is floating w.r.t. ground
Looks to be common mode noise (on both 5V and Return of PSU)
Attachment 24: Same as 21 but using thick crocodile clips on probe to ground and aida12. Noise is attenuated but still present |
Wed Sep 14 19:07:07 2022, PJCS, INFO : Three Merger Statistics explained
|
There are three Merger statistics that can be used to better understand how the data flow through the Merger system is proceeding.
Two are from the Link task and one is from the Merger.They are all "No data buffers avaliable" with #1, #2, #3 at the end.
#1: This is incremented when the Link task has a data item to put in the queue for the Merge process but there is no room.
#2: This is incremented when the Link task has found no room in the queue for the Merge process ( #1 ) , waited , tried again and failed.
#3: This the other end of the queue. When the Merge task requests a data item from a Link task queue and there is nothing available.
|
Thu Sep 8 12:37:18 2022, NH, Proxy Port Changed
|
The proxy in Firefox, Yum and AnyDesk has been changed as the old wasn't working
proxy.gsi.de port 3128 is now in use |
Thu Sep 8 12:31:25 2022, NH, Retrying AIDA DataAcq v10  
|
Startup AIDA with ribbon cable connected to aida03 and aida07 for noise
Setup and run with waveforms enabled. Discriminators ADC power etc as default
Try to push above 200k as this is where we saw issues before... lowering threshold to 0x3 pushes rates to
aida03 - 320k
aida07 - 254k
Startup merger and observe rates
aida03 - 224k
aida07 - 213k
Rate drop observed as before.
Now update aidacommon to point to AidaExecV10 and powercycle FEEs
Rates again with 0x3
aida03 - 315k
aida07 - 252k
Restart with data transfer ON
aida03 - 317k
aida07 - 262k
No errors in merger terminal or "Merge time errors" statistic
Will keep running |
Tue Aug 30 13:48:32 2022, NH, AIDA Single Switch Configuration
|
The second switch was moved back to CARME so AIDA has been configured back to using a single switch
aida02/aida04/aida06/aida08 updated back to first switch as per https://elog.ph.ed.ac.uk/DESPEC/433
Additionally a ribbon cable is attached to aida01 and aida05 to introduce some noise into the system |
Tue Jul 5 08:53:20 2022, TD, To Do
|
In no particular order
1) CAEN 83xx series NIM bin (Ortec 533A output noise issue)
observe +/- 6V, 12V, 24V lines with/without load
try new CAEN NIM bin and/or NIM bin of different type
2) Measure actual voltages at FEE64 power connector input
OH suggests fab of power adaptor for safe observation - contact EW
3) rev B adaptor PCB
invert 125 way ERNI - check for mech conflicts
paired HV input (avoid Lemo-00 T pieces)
consider isolating test/HV Lemo-00 shells from PCB ground (loop elimination)
straight jumpers
shrouded Samtec headers - consider mech issues/consequences of using eject clips too
re-visit HV filtering & separate trace ground
4) isolation transformer
as practical matter may be necessary to operate all platform from isolation transformer
consider hire of appropriate unit
need method to measure isolation - will require permit to work or equiv
5) investigate S4 area ac mains
NH discussing with GSI electricians
6) Systematic measurement of AIDA PSU noise
Spec linear AIDA FEE64 PSU
7) Redesign of snout
Return to 1mm welded box Al for lower stage of snout for added rigidity
8) Revisit calculation of cable lengths. Particularly for the triple
9)
|
Wed Jun 29 10:48:26 2022, NH, OH, AIDA Dismounted
|
All detectors removed from single and triple AIDA snouts
Empty snouts *and* DSSDs (in boxes) stored in NH office |
Tue Jun 28 10:11:35 2022, OH, NH, MIDAS Data Aq V10  
|
11:11 Rebooted FEEs and changed aidacommon in /MIDAS/linux-ppc_4xx/startup to point to the new V10 DataAq that Patrick produced
When using V9 the Merger statistics reported WR items at twice the rate of ADC data items.
i.e for ever data item we were sending and info code 4 and info code 5 item sending 192 bits of data vs 64 for just the data word
This was causing significant deadtime when FEEs were running in the range of around 200kHz. These WR items were not reported by the MIDAS Acquisition server but were in the Merger statistics
Patrick has produced V10 which removes these.
When running V10 we can confirm in the Merger statistics that this rate is no longer determined by the ADC data rate and instead controlled via Sync Rollover Target in GSI WhiteRabbit Control.
WR items for 0xE - attachment 1
WR items for 0x7 - attachment 2
However we see in the NewMerger terminal the message shown in attachment 3 frequently.
Also we note that the merger time error counter is also going up.
Our thoughts for this are we have a rollover issue (Is the merger expecting the rollover of the LSB to be one value when the MSB is updated but MIDAS is happening on another?)
Are we having dead time issues which is causing time warps?
Does each buffer from the MIDAS Data Acq start with a full WR timestamp?
aidacommon has been changed back to point to V9 to not cause issues when we run the DAQ and forget we changed it to be this way? |
Tue Jun 28 09:37:19 2022, NH, Tues 28 June 08:00-
|
Experiment over
10:37 - Stop DAQ & Tape
S4 enters controlled access and they uncable bPlas
Will dismount AIDA snout after |