Sun Mar 3 16:06:11 2024, TD, Sunday 3 March 24x
|
17.00 FEE64 41:a0:71 ASIC #2 u/s
Replaced ASIC mezzanine with new ASIC mezzanine
|
Mon Feb 26 13:17:51 2024, TD, Offline analysis of DEC23/R9_6 9x
|
First pass analysis of data file /TapeData/DEC23/R9_6
|
Wed Jan 31 15:27:07 2024, TD, Photos of water flow & dew point sensor interlock box   
|
Attachment 1 - 4 socket input connectors ( from sensor to interlock box )
Attachment 2 & 2 - 24V dc relays and internal wiring
|
Mon Nov 13 11:46:54 2023, TD, Summary of AIDA 24cm x 8cm 'triple' DSSSD tests 
|
|
Thu Oct 12 14:01:43 2023, TD, MSL type BB18 24cm x 8cm DSSSD test - update 
|
DSSSD MSL type BB18 24cm x 8cm 3208-3/3208-21/3208-22
|
Sun Oct 1 11:47:07 2023, TD, S4 cooling water 
|
The photographs show the cooling water controls and temperature/pressure gauges outside S4 and the connections used by AIDA within S4. |
Thu Aug 31 15:24:56 2023, NH, New AIDA MBS PC
|
The AIDA MBS FDR will be x86l-119 from now on, not x86l-94
the MBS relay and startup scripts will be changed for this |
Mon Aug 28 12:47:56 2023, NH, Power Failure 24.08.2023
|
There was a power failure in the morning of 24.08.2023 in the Rhein-Main area affecting GSI
The Aida workstation (aida-3) has been restarted, it is unknown if the Pis in S4 rebooted as well (there is a UPS)
|
Sun Jul 16 09:32:57 2023, TD, Firefox browser proxy setting change
|
Firefox browser proxy setting changed to 'Auto-detect proxy settings for this network' ( Firefox -> Edit -> Settings -> Network Settings -> Settings ) |
Wed Jan 18 13:40:33 2023, PJCS TD, MACB settings with either Emulator or VITAR   
|
When using the VETAR connected to the HDMI port of the root MACB the settings for all the MACB in the system 0x3.
This setting takes the Clock and Data line from the input HDMI and outputs it via teh HDMI output ports.
When using the Emulator connected via the SMA connectors on the back of the root MACB then the setting for the root MACB |
Fri Dec 16 14:02:12 2022, NH, AIDA System off for christmas break
|
The AIDA NIM crate, pis and workstation have been powered off for the Christmas break and will not be accessible |
Thu Oct 6 16:51:49 2022, NH, Oscilloscope analysis 24x
|
Investigating AIDA noise with a TA041 differential probe and oscilloscope
AC Mains (DESPEC platform AC, L-N)
|
Wed Sep 14 19:07:07 2022, PJCS, INFO : Three Merger Statistics explained
|
There are three Merger statistics that can be used to better understand how the data flow through the Merger system is proceeding.
Two are from the Link task and one is from the Merger.They are all "No data buffers avaliable" with #1, #2, #3 at the end.
#1: This is incremented when the Link task has a data item to put in the queue for the Merge process but there is no room. |
Thu Sep 8 12:37:18 2022, NH, Proxy Port Changed
|
The proxy in Firefox, Yum and AnyDesk has been changed as the old wasn't working
proxy.gsi.de port 3128 is now in use |
Thu Sep 8 12:31:25 2022, NH, Retrying AIDA DataAcq v10  
|
Startup AIDA with ribbon cable connected to aida03 and aida07 for noise
Setup and run with waveforms enabled. Discriminators ADC power etc as default
|
Tue Aug 30 13:48:32 2022, NH, AIDA Single Switch Configuration
|
The second switch was moved back to CARME so AIDA has been configured back to using a single switch
aida02/aida04/aida06/aida08 updated back to first switch as per https://elog.ph.ed.ac.uk/DESPEC/433
|
Tue Jul 5 08:53:20 2022, TD, To Do
|
In no particular order
|
Wed Jun 29 10:48:26 2022, NH, OH, AIDA Dismounted
|
All detectors removed from single and triple AIDA snouts
Empty snouts *and* DSSDs (in boxes) stored in NH office |
Tue Jun 28 10:11:35 2022, OH, NH, MIDAS Data Aq V10  
|
11:11 Rebooted FEEs and changed aidacommon in /MIDAS/linux-ppc_4xx/startup to point to the new V10 DataAq that Patrick produced
When using V9 the Merger statistics reported WR items at twice the rate of ADC data items.
i.e for ever data item we were sending and info code 4 and info code 5 item sending 192 bits of data vs 64 for just the data word
|
Tue Jun 28 09:37:19 2022, NH, Tues 28 June 08:00-
|
Experiment over
10:37 - Stop DAQ & Tape
|