  84   Fri Nov 1 18:09:03 2019 CA, TD, NHSummary - 29.10-1.11.19> >
> > 
> > - evidence of c. 1MHz extrinsic noise
  
  707   Mon Jul 21 13:01:19 2025 MPStatusStatus before grounding checks  Screenshot_from_2025-07-21_14-00-23.pngScreenshot_from_2025-07-21_14-00-39.png 
  708   Tue Jul 22 17:00:26 2025 MP, CCStatusTest with shielded HV cables for both detectors.

The FWHM for ch1 is now 184 ch.

Yesterday cables for cards 5, 6,
 Screenshot_from_2025-07-22_17-51-52.pngScreenshot_from_2025-07-22_17-59-20.png 
  19   Thu Jan 24 10:26:09 2019 VP, TD, CA, NHStart up Merger1) Click on item "Merger MK2" in top panel
   This kills any existing Merger existence
and loads a new copy.
 merger_screenshot.png 
  274   Sat Apr 24 04:50:01 2021 MLSpectraSome screenshot of spectra attached  8x 
  725   Fri Oct 24 14:33:37 2025 JB, GB, MP, AMSnout grounding, bias and test input layout as of 24.10.2025   AIDA_scheme_241020525(2).pdf 
  30   Thu Mar 21 15:45:48 2019 NH, AS, SA, MPSnout Mounting & DSSD BiasThe AIDA snout (with single DSSD and plastic)
has been mounted in S4 cave again. DSSD was
connected to the outer 4 FEEs again.
 LeakageBias.png 
  5   Wed Sep 19 12:40:41 2018 OH, TDSingle DSSD Installation and TestsOn the 18th of September we installed DSSD
3208-14
 21x 
  417   Mon May 2 09:52:20 2022 OHShifter checklist 2022Every 30 minutes:

- Reload statistics (Web browser tab screen
  
  230   Fri Apr 16 11:42:45 2021 OHShifter checklistEvery 30 minutes:

- Reload statistics (Web browser tab screen
  
  446   Sat May 14 03:14:00 2022 PPShift Checks4:13

AIDA stats OK

Leakage current OK
 8x 
  712   Wed Jul 23 13:59:37 2025 CC, NKShielding of the snoutAttachment1: AIDA05.1.8L when AIDA05 connected,
biased at 100 V with not-shielded cables,
CAEN HT,  after the base of the
 Screenshot_from_2025-07-23_15-07-42.png 
  735   Thu Oct 30 10:06:21 2025 MP, NHSetup and data taking with 207Bi- Initialised the Merger and Tape server

- loading of settings: EXPERIMENTS/AIDA/2025Oct29-16.11.06

- aida16 very noisy (attachments
 Screenshot_from_2025-10-30_11-07-55.pngScreenshot_from_2025-10-30_11-07-39.pngScreenshot_from_2025-10-30_11-07-29.pngScreenshot_from_2025-10-30_11-08-07.pngScreenshot_from_2025-10-30_11-10-06.png 
  736   Thu Oct 30 16:43:36 2025 MP, NHSetup and data taking with 207BiData taking and DAQ stopped now (17.40)




  
  14   Mon Dec 10 11:09:53 2018 CA, TDScintillator pictures/system checks for 07.12.1811.12 System wide checks ok (attachment 1)
      FEE temperatures ok (attachment 2)
      bias/leakage currents ok (attachment
 14x 
  194   Fri Mar 12 23:56:00 2021 OHSaturday March 13th 00:00-08:0024:00 Joined in with TD to debug the dropout
of AIDA05 - See https://elog.ph.ed.ac.uk/DESPEC/193
      After both powercycles and telnet reboots
 11x 
  301   Sat May 8 09:05:04 2021 TDSaturday 8 May
09.54 FEE64 temperatures OK - attachment
1
 27x 
  642   Sat Jun 8 13:22:36 2024 TDSaturday 8 June
14.09 aida02 WR stall
      DAQ STOP
 24x 
  422   Sat May 7 09:55:53 2022 TDSaturday 7 May
Attachments 1 & 2 per FEE64 1.8.L spectra
 15x 
  175   Sat Mar 6 06:52:06 2021 OH, LSSaturday 6th March07:50 Pulser settings - attachment 1

07:52 System wide checks all ok
 30x 
ELOG V3.1.3-7933898