Sun Jun 26 08:51:20 2022, OH, NH, Sunday 26 June 08:00-24:00 27x
|
09:51 Taken over from Tom following the night shift
Experiment is still running smoothly
Compression of the files is complete up to the start of R5.
|
Tue Jun 28 10:11:35 2022, OH, NH, MIDAS Data Aq V10  
|
11:11 Rebooted FEEs and changed aidacommon in /MIDAS/linux-ppc_4xx/startup to point to the new V10 DataAq that Patrick produced
When using V9 the Merger statistics reported WR items at twice the rate of ADC data items.
i.e for ever data item we were sending and info code 4 and info code 5 item sending 192 bits of data vs 64 for just the data word
|
Fri Apr 16 07:19:00 2021, OH, MA, Friday 16th April 08:00-16:00 14x
|
08:19 System wide checks all ok
Baselines reset on WR and FPGA checks for the start of the day
|
Sat Mar 6 06:52:06 2021, OH, LS, Saturday 6th March 30x
|
07:50 Pulser settings - attachment 1
07:52 System wide checks all ok
|
Wed Mar 10 07:16:02 2021, OH, LS, Wednesday 10th March 08:00-24:00 39x
|
08:16 System wide checks all ok
Statistics - attachment 1
Temperatures - attachment 2
|
Thu Mar 11 07:09:25 2021, OH, LS, Thursday 11th of March 29x
|
08:00 System wide checks all ok *exctept aida07 and 10 on ADC calibration
N.B. wrong date given to the screenshots by accident (It's early)
Statistics - attachment 1
|
Thu Mar 12 08:04:24 2020, OH, CA, TD, Thursday 12th March 08:00-16:00 14x
|
ASIC settings 2019Oct31-13.24.23
slow comparator 0xa
|
Fri May 13 06:56:14 2022, OH, BA, MA, Low Rates
|
Quote:
The rate is very low in all aida, not sure . |
Thu May 12 00:37:52 2022, OH, AM, Thursday 12 May 13x
|
01:37 Over the past hour the degrader settings have been checked by impinging 209Pb onto the detectors.
That is now complete.
Will find optimum thresholds for n+n side and then leave DAQ running with n+n at optimum and p+n at 0xc
|
Tue May 4 12:11:39 2021, OH TD, Bias test of triple    
|
13:11 System wide checks all ok except *aida02 fails ADC calibration*
Collecting the file size of each FEE64 Options CONTENTS file to check they are all the same
|
Sun Mar 14 08:16:00 2021, OH PJCS, Sunday 14th March 2021
|
> 08:07 Realised the correlation scalers hadn't been enabled again. Have enabled them now. R56_151
>
> Looking through the merger messages from the nightshift we see bad timestamps in FEE 11. This FEE does not have any scalers going into it.
|
Tue Jun 21 10:19:05 2022, OH ML, Tuesday 21st June 08:00-00:00 12x
|
11:19 Stats ok 0xa on all FEE - attachment 1
Current pulser settings - attachment 2
FEE Temperatures ok - attachment 3
|
Thu Sep 20 08:57:48 2018, OH, Thursday 20th September   
|
09:57 Alpha run stoppped R2_0 -> R2_2
Thresholds dropped from 0x64 to 0xa
|
Tue Mar 10 10:16:56 2020, OH, AIDA Implant Range Scan
|
A range scan was performed using AIDA.
See the GSI elog: https://elog.gsi.de/despec/S480/19 |
Wed Mar 11 23:44:27 2020, OH, Thursday 12th March 00:00-08:000 10x
|
ASIC settings 2019Oct31-13.24.23
slow comparator 0xa
|
Thu Mar 4 10:13:48 2021, OH, AnyDesk connection instuction
|
| |
Thu Mar 4 11:16:13 2021, OH, Further cases of possible database issues 
|
During testing for the S452 experiment yesterday we were running through the steps of a restart.
The FEEs were not powercycled at this stage but the MIDAS server was restarted.
|
Fri Mar 5 22:53:05 2021, OH, Thursday 5th of March
|
There has been no beam for much of the day. It was lost at 13:00 and not regained until 22:00 German time.
At some point during the day the White Rabbit timestamp cable was unplugged from AIDA. This caused it lose correlation with the rest of the system.
At first a reset and setup was performed to try and regain correlation but that did not prove successful.
|
Sun Mar 7 23:13:50 2021, OH, Monday 8th March 13x
|
00:00 ASIC settings 2019Dec19-16.19.51
DSSSD#1 slow comparator 0xa
DSSSD#2 slow comparator 0xa
|
Mon Mar 8 22:51:47 2021, OH, Tuesday 9th March 10x
|
18:00 ASIC settings 2019Dec19-16.19.51
DSSSD#1 slow comparator 0xa
DSSSD#2 slow comparator 0xa
|