Entry  Fri Mar 12 23:56:00 2021, OH, Saturday March 13th 00:00-08:00 11x
24:00 Joined in with TD to debug the dropout of AIDA05 - See https://elog.ph.ed.ac.uk/DESPEC/193
      After both powercycles and telnet reboots we were still seeing 0 in the statistics but did note that the counter was > 0
      The last check we did was a restart of the merger server. This solved the issue.
Entry  Sun Mar 14 07:04:30 2021, OH, Sunday 14th March 2021 20x
08:07 Realised the correlation scalers hadn't been enabled again. Have enabled them now. R56_151

Looking through the merger messages from the nightshift we see bad timestamps in FEE 11. This FEE does not have any scalers going into it.
Entry  Thu Apr 15 22:38:18 2021, OH, Thursday 15th April - 16th April 
23:38 They are testing the finger detector
      It is a lot of U they are putting into it which is passing through to AIDA ~3500Hz
      Most seems to be punching through
Entry  Fri Apr 16 11:42:45 2021, OH, Shifter checklist 
Every 30 minutes:

- Reload statistics (Web browser tab screen 1, workspace 2)
Entry  Mon Apr 19 23:01:02 2021, OH, 20th April 00:00-08:00 17x
23:52 Noticed AIDA07 has dropped out of the merger and is producing no statistics
      Powercycling to recover
Entry  Wed Apr 21 11:27:55 2021, OH, Wed 21st April 12:00-16:00 6x
12:30 MBS has crashed
      This time AIDA is still running fine.
Entry  Thu Apr 29 16:56:42 2021, OH, Config changes for 16FEE 
New set of ASIC settings produced 2021Apr29-13-16-00
- Note no longer odd/even for polarity of signal
Entry  Tue May 4 21:26:43 2021, OH, Alpha run 10x
22:26 Tape server set up to write to directory May21
      DAQ stopped
      Slow comparator set to 0x64 for all FEE64 and an ASIC check performed
Entry  Wed May 5 20:10:16 2021, OH, Wednesday 5 May Alpha run 12x
21:10 After all of the works in S4 today the noise in the system was considerably worse
      It was decided that before starting a new alpha run a powercycle would be performed.
      The powercycle has had no effect on the rates. They are still terrible.
Entry  Tue May 11 07:21:24 2021, OH, Tuesday 11 May 210511_0823_Stats.png210511_0824_Temp.png210511_0825_Bias.png
08:20 System wide checks all ok except
 Base  Current  Difference
aida01 fault   0xc10c :   0xc10e :   2  
Entry  Thu May 13 23:58:05 2021, OH, Friday 14th May 00:00 - 08:00 21x
00:00 OH Takes over.
 PULSER SETTINGS
---------------------------
Entry  Wed May 19 06:59:33 2021, OH, Wednesday 19th May 08:00 - 16:00 9x
08:00 OH Takes over for CA
      Thresholds 0xc for p+n and 0x19 for n+n ASICs 1-3 and 0x64 for n+nASIC4
Entry  Thu May 20 14:56:33 2021, OH, Thursday 20th May 16:00 - 24:00 10x
16:00 p+n threshold 0xc (120keV)
      n+n threshold DSSD2 0x20 (250keV) ASICS 1-3, 0x64 ASIC4
      n+n threshold DSSD2 0x19 (250keV) ASICS 1-3, 0x64 ASIC4
Entry  Fri Sep 3 13:39:42 2021, OH, Additional noise tests performed at GSI DSO_Lab_NoP.jpgDSO_Lab_1.jpgDSO_Lab_50V.jpgDSO_Lab_NoCable.jpg
Contents of an email from NH to OH and TD about tests performed on an additional bias supply:

This email was sent to you by someone outside the University.
Entry  Mon Oct 18 13:01:04 2021, OH, Proxy config proxy.pngproxy2.png
Proxy for firefox and anydesk
Entry  Mon Oct 18 13:03:23 2021, OH, DHCPd.conf backup dhcpd.conf
 
Entry  Mon Oct 18 14:20:32 2021, OH, Top bar launchers 
launcher1
TclHttpd @8015 for AIDA
/MIDAS/Linux/startup/HTTPD
Entry  Tue Oct 19 16:17:49 2021, OH, AIDA-gsi server to centos 7  
Following advice of CU have updated the server to centos 7.

To do this a new 500GB SSD was purchased and installed in the enclosure and a clean install of centos7 was installed to this.
Entry  Thu Oct 21 09:08:41 2021, OH, MIDAS start up messages 8015Startup.txt8115startuptapestart.txtmergerstart.txtdatarelay.txt
 
Entry  Thu Oct 21 09:33:08 2021, OH, Noise tests - 21st Oct 2021 211021_0951_fee8_x8L.png211021_0953_fee14waveform_zoom.png211021_0956_fee14_waveform.png211021_0957_fee8_x8L.png211021_0959_fee8_waveform_zoom.png
Pulser was left on overnight with the bias filters on
First plot is FEE14 not FEE8