AIDA GELINA BRIKEN nToF CRIB ISOLDE CIRCE nTOFCapture DESPEC DTAS EDI_PSA 179Ta CARME StellarModelling DCF K40
  DESPEC, Page 18 of 36  ELOG logo
New entries since:Thu Jan 1 01:00:00 1970
ID Date Authordown Subject Text Attachments
  456   Sun May 15 07:40:52 2022 OHSunday 15th May 08:00-16:0008:40 FEE Temperatures ok - attachment 1
      FEE statistics - Attachment 2
      Bias and leakage currents ok - attachment
 10x 
  464   Tue May 17 06:58:51 2022 OHTuesday 17th May08:00 Still have beam for the time being.
      Currently on file R4_2049
 6x 
  484   Sun Jun 19 12:36:10 2022 OHSunday 19th JuneYesterday the sum/inverter was found to be
a considerable source of noise in the system.
 1.jpg2.jpg3.jpg4.jpg220618_1753_stats_no_waveform.png 
  493   Thu Jun 23 06:56:02 2022 OHThursday 23 June 08:00-16:0007:56 Taken over from AM
      Current White Rabbit Error status:
        Base  Current  Difference
 11x 
  496   Fri Jun 24 06:56:06 2022 OHFriday 24 June 008:00-16:0007:56 Took over from AM. No issues reported
overnight
      Stats ok - attachment 1
 16x 
  504   Mon Jun 27 06:45:22 2022 OHMonday 27th June 08:00-16:0007:45 Spoke to David and the beam has been
gone since about 05:30
      Reason for the loss of beam is a vacuum
 220626_0830_Stats.png220626_0830_Temp.png220626_0831_Bias.pngR5_770_analysis.txt 
  663   Fri Jun 14 15:00:47 2024 Norah , Muneerah 16:00-20:00 Friday 14 June Everything appears to be running
smoothly. Attachments 1-4 were taken around
16:00
 13x 
  608   Fri Apr 26 07:01:26 2024 Norah08:00 - 16:00 Friday 26 April10:00 Checks:

All seems smooth.

DSSD bias and  leakage current
 Screenshot_from_2024-04-26_10-00-39.pngScreenshot_from_2024-04-26_10-02-35.pngScreenshot_from_2024-04-26_10-04-40.pngScreenshot_from_2024-04-26_10-06-18.pngScreenshot_from_2024-04-26_10-09-52.png 
  613   Sat Apr 27 15:13:06 2024 Norah16:00 - 00:00 Saturday 27 April 14:50 Checks

Leakage current  is 20.021
and HV status is Max Voltages ! Is it normal
 15x 
  81   Fri Nov 1 14:17:15 2019 Nic, PatrickReply: WR Timestamps> > All 12 FEEs have valid WR Timestamps
> > Had to powercycle aida09 once as before
raw readout was displaying upper 12 bits
  
  90   Thu Nov 7 10:22:26 2019 Nic, PatrickReply: WR Timestamps> > > All 12 FEEs have valid WR Timestamps
> > > Had to powercycle aida09 once as before
raw readout was displaying upper 12 bits
  
  36   Tue Mar 26 14:58:09 2019 NH, VPHow to reset merger If the merger doesn't show "xfer
Links => Merger" at the top, it is
probably the merger setting has been corrupted
  
  534   Tue Mar 12 16:16:49 2024 NH, TD, JB, HASummary of DSSSD Biasing 12.03DSSD#1 undergoes a breakdown at 90V, two of
the three wafers show this
- The adapter PCBs themselves have no breakdown
  
  119   Fri Dec 20 12:04:03 2019 NH, TDFriday 20 December 2019Fig 1: DSSD Leakage currents ok

Fig 2: FEE Temps ok
 7x 
  277   Sat Apr 24 13:25:23 2021 NH, TDAIDA FEE64 crash log for S460
https://docs.google.com/spreadsheets/d/1CyL5L-WLsh9fsS7CN989GC3ktqiPQOkG8YIDJrrGxYA/edit#gid=0
  
  375   Tue Aug 17 12:39:25 2021 NH, TDAIDA NoiseInvestigate noise situation of AIDA (no DSSSDs
connected)
 16x 
  404   Mon Mar 7 12:20:26 2022 NH, TDAIDA Grounding Checks Mon 7 MarRemove all DSSD ribbon cable braid ground
from FEE64s to check n+n ground loops
 14x 
  474   Tue May 31 09:05:00 2022 NH, TDTuesday 31 MayDSSD stack will be as follows

1      3208-7  1019       125           
  
  56   Mon Jul 8 16:02:52 2019 NH, PJCSAida 28/6/2019 - WATER OFFIt occurs to to wonder if there is now
water inside the modules. If the metal of
the cooling plates is damp on the outside
  
  127   Wed Feb 19 10:58:29 2020 NH, PJCSReport: aida09 Kernel Panic & Lost WR > aida09 crashed over the weekend and automatically
rebooted.
> After the reboot the WR timestamp sent
  
ELOG V3.1.4-unknown