Entry  Tue Aug 17 12:39:25 2021, NH, TD, AIDA Noise 16x
Investigate noise situation of AIDA (no DSSSDs connected)

Figs 1-6: Waves, Pulser Peaks + Widths *before* any changes
Entry  Mon Mar 7 12:20:26 2022, NH, TD, AIDA Grounding Checks Mon 7 Mar 14x
Remove all DSSD ribbon cable braid ground from FEE64s to check n+n ground loops

Removing all on n+n side only and test resistance to ground: 4, 6 and 8 show continuity to ground, 2 shows no connection (braid ring terminal to copper
Entry  Tue May 31 09:05:00 2022, NH, TD, Tuesday 31 May 
DSSD stack will be as follows

1      3208-7  1019       125              upstream
    Reply  Mon Jul 8 16:02:52 2019, NH, PJCS, Aida 28/6/2019 - WATER OFF 
It occurs to to wonder if there is now water inside the modules. If the metal of the cooling plates is damp on the outside then it is safe to assume
there is water on the inside too . Can I suggest that one  of the modules is dismounted, disassembled and checked?

Patrick
    Reply  Wed Feb 19 10:58:29 2020, NH, PJCS, Report: aida09 Kernel Panic & Lost WR  
> aida09 crashed over the weekend and automatically rebooted.
> After the reboot the WR timestamp sent to the merger is in the future and hence incorrect
Entry  Thu May 12 16:30:47 2022, NH, OH, TD, Notes for mounting a second switch 
Update DHCPD as expected
Make sure new ethernet device is trusted in firewall
Make sure hosts is updated
Entry  Tue Nov 12 13:59:07 2019, NH, OH, CA, Awfully noisy waveforms 8x
Rates in all FEEs much higher than before - waveform shows very noisy in all systems. - attachment 1

Good event statistics - attachment 2
Entry  Tue Nov 12 14:10:18 2019, NH, OH, CA, Aida10 Unusual Pulser Peak aida10-pulser-121119.pngpulsers121119.pngpulsersodd121119.png
aida10 (no alpha rate recorded) has a strange pulser peak - looks like double peak?

Pulser terminator checked and OK...

All others "OK" but noise quite bad (see previous entry)
Entry  Wed Oct 20 13:26:08 2021, NH, OH, AIDA Noise + Filtering 6x
Turn on AIDA (SL6) to investigate noise situation and impact of new RC filters on bias input

Current setup:
Entry  Wed Jun 29 10:48:26 2022, NH, OH, AIDA Dismounted 
All detectors removed from single and triple AIDA snouts

Empty snouts *and* DSSDs (in boxes) stored in NH office 
Entry  Fri Mar 22 07:34:54 2024, NH, JB, Au Beam 
 
Entry  Tue Mar 26 10:32:05 2024, NH, JB, Tue 26 March 9x
Taken 4 FEE64s from CRYRING (the 4 easiest to access)
  41:d8:2b
  41:f6:5a
Entry  Fri Apr 9 16:45:29 2021, NH, HA. GA, GZ, OH, TD, AIDA Noise 6x
AIDA snout was reconnected and we saw originally a very good reduction in noise from S452 (fig 1)
Occasional bursts of increased rates were noticed, perhaps related to works in S4.
Entry  Wed Apr 27 15:47:46 2022, NH, HA, SA, CJ, AIDA 27 Apr 22 7x
Remount snout in S4 (photos to come) 
Alignment with DEGAS OK but a bit short...  DEGAS platform will be brought 31 mm closer to AIDA by rail adjustment
Entry  Mon Dec 9 12:28:01 2024, NH, CC, MP, Test of BB7 on AIDA Electronics 
BB7 connected to aida07, sitting in a cardboard box

Update Tcl/Tml files in ASIC, ASIC4 and sys.tml frm PJCS email to allow setting per-ASIC polarity on FEE64
Entry  Wed May 22 20:10:19 2024, NH, CC, HMA, MV, ZC, KZ, AIDA Snout Remounted 9x
AIDA and the Snout have been remounted in S4

I had a check of the AIDA connectors and all looked good, the kapton PCB connectors were sitting comfortably above the interstage and I saw no stressed/loose
Entry  Thu Mar 21 15:45:48 2019, NH, AS, SA, MP, Snout Mounting & DSSD Bias LeakageBias.png
The AIDA snout (with single DSSD and plastic) has been mounted in S4 cave again. DSSD was connected to the outer 4 FEEs again.

-160 V Bias applied and DSSD holds voltage with leakage current of around 1.3 uA (was dropping as only biased briefly)

Bias off and FEEs offline again now as day nearly done. MBS testing will be performed shortly, as well as resolution checks.
Entry  Thu Mar 21 15:41:49 2024, NH, AM, MP, CC, Thu Mar 21 6x
Fig 1-3: Noise situation at real thresholds (0xa p+n, 0xf n+n)

Fig 4-6: After AM and MP turn off Mesytec Preamps
    Reply  Fri Mar 22 08:22:43 2024, NH, AM, MP, CC, Thu Mar 21 
> Fig 1-3: Noise situation at real thresholds (0xa p+n, 0xf n+n)

> Fig 4-6: After AM and MP turn off Mesytec Preamps
Entry  Wed Apr 10 12:33:58 2019, NH VP, AIDA Firmware Update 20191004-FeeTemps.png20191004-GSIWR.png
All AIDA modules have been updated to a new firmware 0xd330701 which should fix the merger issues.
A new GSI WR timestamp page has been added to MIDAS - the SYNC rollover is suggested to be 0xe for now.
(VP is working on lowering the rate of WR SYNC infowords)