Tue Feb 1 15:00:28 2022, NH, AIDA CentOS 7 Startup
|
Starting up AIDA on CentOS 7 to confirm things work (based on some CARME experience)
NFS settings:
|
Mon Oct 28 17:02:11 2019, TD, AIDA @ DESPEC cable length estimates
|
AIDA snout length 37cm (from end of AIDA snout support assembly)
downstream scintillator (assembly depth 1cm) 36cm
|
Wed May 4 16:50:04 2022, NH, AIDA 4th May 
|
bPlas has been mostly cabled in now
Reseat LAN cable (AIDA Switch -> AIDA-3) and reachieve 1 gbit/s speed as needed.
|
Wed Apr 27 15:47:46 2022, NH, HA, SA, CJ, AIDA 27 Apr 22 7x
|
Remount snout in S4 (photos to come)
Alignment with DEGAS OK but a bit short... DEGAS platform will be brought 31 mm closer to AIDA by rail adjustment
|
Sat May 1 08:45:19 2021, TD, AIDA 24cm x 8cm DSSSD + triPlast assembly for S496 12x
|
Photos provided by Nic Hubbard and Helena Albers from the assembly of 2x triPlast
plus 2x MSL type BB18(DS)-1000 24cm x 8cm detector assembly - attachments 1-3.
|
Mon Dec 16 14:48:02 2019, NH TD, AIDA 16.12.19    
|
DSSD Snout was reassembled to install the second bPlast (downstream)
(fig 1)
|
Tue Aug 13 09:17:02 2019, NH, AIDA 12/08/2019
|
The other 6 FEEs were powered on one at a time to check no issues. None found.
All 12 FEEs plugged back into the PSU.
Note: The FEE temperatures reported on Friday were before a Reset/Setup of the DAQ, which means the ASICs run very fast and the hence run hotter. |
Mon Jul 1 09:42:16 2019, NH, AIDA - 01/07/2019
|
Water is still off.
Checked dewpoint sensor still flashing red light (indicating an error). Pipes warmed. Not what I expected. Possible fault?
I checked S4's ambient temp/humidity with a home-brought sensor: Reports 27 C/50% RH (Dew Point = 16 C) |
Tue Jun 11 20:20:36 2024, TD, ADC data rates from 09.00 11.6.24
|
Per NH
"From ELOG (assuming times accurateish)
|
Fri Mar 22 14:31:24 2019, NH, A few more setup notes I forgot   
|
Extra FEE information:
HV braid (0 V) goes into FEE10 (beam left) with LK1 jumper
HV core (-160 V) goes into FEE09 (top) |
Sat Dec 7 02:39:02 2019, DK, NH et al., 34Si Fragments PID   
|
See attached
Tentative PID (DK) as attachment #3
|
Wed Jun 1 08:00:18 2022, TD, 2xDSSSDs spectra layouts
|
Spectra layouts for 2x DSSSDs can be found in directory ~/LayOut/2xDSSSDs |
Fri May 13 00:50:16 2022, ML, 2h-Shift Checks  
|
All seems ok (apart for my virgin network connection usually quite reliable but not so tonight !?!).
Statistics seems ok.
Temperatures appears stable |
Fri May 13 02:56:36 2022, ML, 2h-Shift Checks  
|
All seemnormal on AIDA side.
Stats as before
Temperatures stable |
Fri May 13 04:45:23 2022, ML, 2h-Shift Checks  
|
AIDA is running ok. Note that, in the last hour or so, the rate of implant in DSSD1 is back to higher rate ~10 Hzto 15 Hz as at the beginning of this
shift.
Stats seems ok |
Fri May 13 06:47:33 2022, ML, 2h-Shift Checks  
|
It has been a quiet night. AIDA has been stable.
Stats appears stable appear
Temperature also stable |
Fri May 13 23:37:59 2022, ML, 2h-Shift Checks   
|
Took over remotely from CB about an hour ago
At 0h40 (CET):
Stats still ok - attach 1
Leakage still OK - Attach 2
Temps
- attach 3
Grafana OK
UCESB rates ok - attach 5
System-wide checks: aida09 fails clock (1), aida02 06 09 10 13 fail ADC calibration,
all pass WR |
Sat May 14 01:52:03 2022, ML, 2h-Shift Checks    
|
AIDA stats ok
Leakage current ok
Temperatires ok |
Tue Mar 26 14:59:30 2019, NH, 26.03.2019 
|
Fig1: FEE temperatures from 26.03.2019 all OK
AIDA has been on today and merger setting fixed (see previous ELOG entry https://elog.ph.ed.ac.uk/DESPEC/36)
Still diagnosing issues with MBS (figure 2) but local data acquisition should be functional. |
Fri Jan 25 09:54:12 2019, CA, TD, NH, VP, 25th January 2019 8x
|
09.30: AIDA still running, not writing to file
Still running ok
No water leaks
|