Thu Apr 25 15:02:50 2024, Marc, 16:00 - 0:00 Thursday 25 April 8x
|
16:20:
Spill length:4s (was 3.5s before) - Extraction time 1.5s (see attachement #1)
16.45
Checks:
DSSSD bias & leakage current - Grafana - attachments 2
leakage current ramping, spill micro structure observable
FEE64
temperatures OK - attachment 3
ADC data item stats - attachment 5
per FEE64 Rate spectra - attachment 4
Merger etc - attachment |
Sat Apr 27 15:13:06 2024, Norah, 16:00 - 00:00 Saturday 27 April 15x
|
14:50 Checks
Leakage current is 20.021 and HV status is Max Voltages ! Is it normal ? - attachments 1
It is quite warm at GSI today and this may be a temperature effect. Hopefully will go down latter in the day (Marc). Any way let's kep an |
Fri Apr 26 15:34:00 2024, RDP, 16:00 - 00:00 Friday 26 April 6x
|
This appeared in the system log:
Apr 26 16:30:51 aidas-gsi smartd[1076]: Device: /dev/sda [SAT], 1 Currently unreadable (pending) sectors
Apr 26 16:30:56 aidas-gsi smartd[1076]: Device: /dev/sdd [SAT], 2224 Currently unreadable (pending) sectors |
Sun Jun 9 16:38:30 2024, Norah , Muneerah, JB, 16:0-00:00 9 June 2024   
|
AIDA02 and AIDA06 gave zero attachment 1. After connecting with Tom to fix it, now it works.
17:00
DSSSD bias & leakage current - attachment 2 |
Sun Apr 28 15:09:34 2024, Betool Alayed, 16:0-00:00 28 April 2024
|
16:00 aida04 not producing data
power cycle all FEE64s to recover DAQ
16.20 analysis data file R21_668 |
Sun Apr 28 20:35:57 2024, Betool Alayed, 16:0-00:00 28 April 2024 10x
|
21:30 screenshots
and
23:24 screenshots |
Mon Apr 22 15:17:07 2024, DSJ, 16.00-0.00 22/04/24 26x
|
First checks of shift. All looks ok.
Screenshot of HV, temps, rates, Merger attached. Attachments 1-4.
Screenshots of all spectrum Layout IDs taken before zeroing at 16.35. (Attachments 5-12) |
Fri Apr 16 15:52:08 2021, DSJ , 16 April 16.00 shift 12x
|
Clock status test result: Passed 12, Failed 0
Understand status as follows
|
Sat Mar 13 23:08:08 2021, CA, 14th March 00:00-08:00 22x
|
00:08 DAQ continues ok - writing to file R46_221
ASIC settings 2019Dec19-16.19.51
|
Tue Apr 23 13:15:43 2024, PP, 14:00 checks 8x
|
All looks good.
Screenshots attached |
Wed Nov 13 09:10:37 2019, CA, NH, OH, 13th November 2019   
|
10:10 moved HV cables from 11 and 12 to 10 and 9 - all detectors biased in same configuration
lowered slow comparator threshold to 0xa
|
Fri May 13 15:00:23 2022, CB, 13 March - 16:00-24:00 shift 10x
|
16:00 Took over remotely from OH, TD who remain in GSI
Stats OK - attach 1
|
Sat Apr 13 14:04:44 2024, JB, 13 April Noise checks 9x
|
15:04 bPlast was left powered over night. Powering up to check noise conditions - also to check shifting of Germanium baseline (no change observed).
TEMP OK.
Noise condition the same as yesterday when platform was moved in. Noise in aida01 and |
Tue Apr 23 11:09:00 2024, PP, 12:00 checks 8x
|
Everything is OK now.
Screenshots attached. |
Thu Apr 11 22:04:50 2024, JB, 12.04.2024 AIDA-bPlast noise optimisation    
|
15:00 Platform in, biasing detector. Temp OK Attachment 1. test - Pulser OFF.
Rates somewhat worse than last night - probable contact on grounding of snout. Attachment
3. |
Thu Apr 11 08:27:21 2024, JB, TD, 11 April noise tests 48x
|
TO-DO for 11.04.2024
- Try bringing bPlast drain wire ground back to the PSU ground for PN 300 and R&SRMP4040.
- Recheck the downstream
detector bias and ground scheme.
-
|
Tue Apr 23 09:21:01 2024, PP, 10:20 checks - aida02 is down 8x
|
Aida02 is down. Contacted Nick.
The rest looks stable, see screenshots.
10:53 CEST (NH): Rebooted aida02 via telnet |
Sun Apr 28 22:52:06 2024, Marc, 0:00-08:00 Monday 29 April 7x
|
Just another manic magic Monday, ....
Taking over Betool's AIDA shift. All is good.
This shift's first hour has been eventless. Temperatures, Rates, bias, all look good. |
Mon Mar 9 13:50:08 2020, LS, CA, DK, 09/03/2020 system checks 13x
|
14:52 Attachment 1 shows low energy spectra. AIDA06 shows strange baseline, possible double-hit?
15:06 Systems checks fine.
|
Wed Apr 24 09:04:55 2024, TD, 08:00-16:00 Wednesday 24 April 33x
|
09.56 DAQ continues data file S100_alpha/R18_122
FRS setting 162Eu
|