Sat Apr 27 04:59:31 2024, TD, 08:00-16:00 Saturday 27 April 37x
|
06.00 DAQ continues file S100_alpha/R20_740
FRS setting 162Eu
DSSSD bias & leakage current - Grafana - attachments 1-2
leakage currents tracking S4 ambient temperature
FEE64 temperatures OK - attachment 3
All system wide checks OK *except*
as reported - attachments 4-5
*all* ADCs not calibrated (not calibrated post-synchronise ASIC clocks at most recent restart)
WR timestamps OK - attachment 6
ADC, PAUSE, RESUME, Correlation scaler data item stats OK - attachments 7-10
per FEE64 Rate spectra - attachment 11
per p+n FEE64 1.8.L specra - attachments 12-13
aida09 pulser peak width 55 ch FWHM (~38keV FWHM)
per FEE64 1.8.H spectra - attachments 14-15
Merger, TapeServer etc - attachments 16-17
12Mb/s to disk
ucesb - attachment 18
spill length 3.5s, extraction 1.5s
06.30 analysis data file R20_748 - attachment 19
max deadtime 12%
no timewarps
HEC data item rate 2.1kHz
06.35 new run file S100_alpha/R21
09.30 DAQ continues - file S100_alpha/R21_83
DSSSD bias & leakage currents - Grafana - attachment 20
FEE64 temperatures OK - attachment 21
ADC data item stats OK - attachments 22
per FEE64 Rate spectra - attachment 23
Merger etc - attachments 24
ucesb - attachment 25
spill length 3.5s, extraction 1.5s
09.50 per FEE64 1.8.W spectra - 20us FSR - attachments 26-27
no significant change in electronic noise cf. start of run
per p+n FEE64 1.8.L spectra - attachment 28
aida09 pulser peak width 62 ch FWHM
common x/y scale - peak heoght proxy for width
13.57 DAQ continues file R21_141
DSSSD bias & leakage current - Grafana - attachments 29-30
FEE64 temperatures OK - attachment 31
ADC data item stats OK - attachments 32
per p+n FEE64 1.8.L spectra - attachment 33
aida09 pulser peak width 55 ch FWHM (~39keV FWHM)
per FEE64 Rate spectra - attachment 34
Merger etc - attachments 35
ucesb - attachment 36
spill length 3.5s, extraction 1.5s
14.15 analysis data file R21_144 - attachment 35
max deadtime 16% (aida04)
no timewarps
HEC data item rate 2.0kHz |
Fri Apr 26 23:04:51 2024, Marc, 00:00-08:00 Saturday 27 April 12x
|
It's a new day, it's a new shift and I'm feeeeeliiiiing goooood !!!!!
0.15 Checks:
DSSSD bias & leakage current - Grafana - attachments 1-2
leakage current dropped a bit since the evening shift ~22:00.
FEE64 temperatures OK - attachment 3
ADC data item stats - attachment 4
per FEE64 Rate spectra - attachment 5
Merger etc - attachment 6
ucesb - attachment 7
2:20 Nothing new to report. All good. Leakage currents continue to go down slowly. See attachement 8 to 12
3:30 Still running smoothly.
5:20 AIDA 4 stop counting in the last 20 min. I informed the main DAQ and they started a new run without AIDA.
DAQ restarted OK. Main DAQ informed.
|
Fri Apr 26 15:34:00 2024, RDP, 16:00 - 00:00 Friday 26 April 6x
|
This appeared in the system log:
Apr 26 16:30:51 aidas-gsi smartd[1076]: Device: /dev/sda [SAT], 1 Currently unreadable (pending) sectors
Apr 26 16:30:56 aidas-gsi smartd[1076]: Device: /dev/sdd [SAT], 2224 Currently unreadable (pending) sectors
Apr 26 16:30:56 aidas-gsi smartd[1076]: Device: /dev/sdd [SAT], 257 Offline uncorrectable sectors
These messages seem to appear sporadically.
18:10 Link 3 is giving rates of zero (see attachment 3). Is this normal?
18:15 daq error on ucesb (attachment 4). Zero rate through event builder. Emailed Nic & can see he's now on Zoom.
aida07 rebooted itself, but WR out of sequence still
manually reboot aida07, still out of sequence
do full powercycle of aida... now all FEEs are in sequence again. Resume running to disk and alert DESPEC shifters
Thanks Nic!
22:14 AIDA DSSD 2 HV status = 99 on grafana (attachment 6)
22:43 Noticed that aida02 ASIC temperature reading is now 0.00 instead of 511(!) that it was previously - see attachment 2 in #608. It was similarly high earlier in this shift too.
|
Fri Apr 26 08:15:59 2024, TD, [How To] Recover DAQ when one FEE64 stops sending data
|
Occasionally, one (sometimes more) FEE64(s) will stop producing data whilst the DAQ is running *and* data is being transferred to the Merger and TapeServer.
Typical symptoms are:
- no data in Rate spectrum [workspace 2, display 1, browser tab 'Spectrum Browser']
- zero rate in 'ADC data item' statistics [workspace 2, display 1, browser tab 'Statistics']
- Merger data link inactive [workspace 3, display 1, browser tab 'NewMerger Control']
active data links are green, inactive links are red
N.B. data links are numbered from 0 to n-1, FEE64s are numbered from 1 to n
0) Please infrom the 'on shift' crew that you need to restart the AIDA DAQ
1) STOP DAQ [workspace 2, display 1, browser tab 'Run Control']
all FEE64s should change status from 'going' to 'stopped' *except* whichever FEE64(s) have a problem - typically their status is 'going' or 'error'
2) Restart Merger by selecting the Merger icon [workspace 3, display 2]
terminal window showing the startup will popup
3) Setup Merger by selecting 'SETUP' [workspace 3, display 1, browser tab 'NewMerger Control']
4) Start Merger by selecting 'GO' [workspace 3, display 1, browser tab 'NewMerger Control']
all data links will be inactive (red)
5) Reboot problem FEE64s
Launch a command terminal and telnet to the problem FEE64 for example
telnet aida04
Enter user 'root' and enter password. When logged in reboot the FEE64 using the command
reboot
6) Monitor boot using server system log [workspace 1, display 2, browser tab 'terminal']
you should see two messages showing the FEE64 mounting filesystems
wait c. three minutes for boot to complete
7) Restart DAQ
Select DAQ status 'Update' [workspace 2, display 1, browser tab 'Run Control']
Select DAQ status 'RESET' [workspace 2, display 1, browser tab 'Run Control']
Select DAQ status 'SETUP' [workspace 2, display 1, browser tab 'Run Control']
Disable histogramming by selecting 'Histogramming Disable all' [workspace 2, display 1, browser tab 'Run Control']
Disable waveforms by selecting 'Enable Waveforms Disable all' [workspace 2, display 1, browser tab 'Run Control']
Disable data transfer by selecting 'Data transfer Disable #1 all' [workspace 2, display 1, browser tab 'Run Control']
8) Restore ASIC settings
Select 'Save/Restore Settings' [workspace 2, display 1, browser tab 'Control']
Select the correct 'Data Base Key' [workspace 2, display 1, browser tab 'Module Settings']
Select 'Restore Settings' [workspace 2, display 1, browser tab 'Module Settings']
Wait 10s
9) Check ASIC settings
Select 'Check ASIC Control' [workspace 2, display 1, browser tab 'ASIC Control']
Wait for popup indicating completion of checks and dismiss
10) System wide checks
Select 'Synchronise the ASIC clocks' [workspace 2, display 1, browser tab 'System Wide Checks']
Select 'Collect all WR status error counters ...' [workspace 2, display 1, browser tab 'System Wide Checks']
Select 'Collect all FPGA timestamp error counters ...' [workspace 2, display 1, browser tab 'System Wide Checks']
Select 'Check clock status' [workspace 2, display 1, browser tab 'System Wide Checks']
Select 'Check WR decoder status' [workspace 2, display 1, browser tab 'System Wide Checks']
Select 'Check FPGA timestamp errors' [workspace 2, display 1, browser tab 'System Wide Checks']
make a note or take a screenshot of any error messages observed
11) Start DAQ
Enable histogramming by selecting 'Histogramming Enable All' [workspace 2, display 1, browser tab 'Run Control']
Enable waveforms by selecting 'Enable Waveforms Enable All' [workspace 2, display 1, browser tab 'Run Control']
Start DAQ by selecting 'GO' [workspace 2, display 1, browser tab 'Run Control']
check that DAQ is *actually* going - if not repeat 'GO'
12) Check WR timestamps by selecting 'Collect all timestamps' [workspace 2, display 1, browser tab 'GSI White Rabbit Control']
readout times will be displayed for each FEE64
check that the readout times are monotonically increasing as you read down the list of readout times - if not, contact expert support
13) Stop DAQ
Start DAQ by selecting 'STOP' [workspace 2, display 1, browser tab 'Run Control']
check that DAQ has *actually* stopped - if not repeat 'STOP
14) Enable data transfer to Merger and TapeServer and start DAQ
Enable data transfer by selecting 'Enable data transfer Enable #1 All' [workspace 2, display 1, browser tab 'Run Control']
Start DAQ by selecting 'GO' [workspace 2, display 1, browser tab 'Run Control']
check that DAQ is *actually* going - if not repeat 'GO'
15) Check data is being transferred to Merger and TapeServer
Select 'Reload' [workspace 3, display 1, browser tab 'NewMerger Control']
you should see a non-zero 'current merged data rate' and all links should show as green at least some of the time (requires observation over several reloads)
Select 'Reload' [workspace 3, display 1, browser tab 'TapeServer']
you should see 'Blocks written' and 'kb written' counters increase
16) Check AIDA data is being forwarded from the TapeServer to MBS [workspace 3, display 2, command terminal]
17) Restart should now have been successfully completed. Please inform the 'on shift' crew' that AIDA is forwarding data.
If not, please request expert assistance.
|
Fri Apr 26 07:01:26 2024, Norah, 08:00 - 16:00 Friday 26 April    
|
10:00 Checks:
All seems smooth.
DSSD bias and leakage current ( Attachment 1)
FEE64 temperatures (Attachment 2)
Per FEE64 Rate spectra (Attachment 3)
ADC data item stats -(Attachment 4)
Ucesb (Attachment 5)
10:21
Aida05 was zero not reading had to call Nic to fix it.
10:34
Now, Aida05 is running well.
.......................................................
12:00 Checks:
Everything appears to be going smoothly, and I took screenshot of them.
12:28, Aida02 showed as not reading , had to email Nic to fix it.
12:30, it resumed reading and returned to normal operation.
.........................................................
14:00 Checks:
Everything is going well , and I have captured screenshots of each one.
......................................................
16:00 Checks:
All seems smooth, and I took screenshot of them.
|
Thu Apr 25 22:45:30 2024, Tom, Muneerah, Norah, Over night shift AIDA 26 Fri 9x
|
01:02
Aida04 was zero not reading had to call Tom to fix it.
02:00 checked all good and attached.
Its been running good after that till 7:30 again the same problem happened and Tom fixed it.
|
Thu Apr 25 15:02:50 2024, Marc, 16:00 - 0:00 Thursday 25 April 8x
|
16:20:
Spill length:4s (was 3.5s before) - Extraction time 1.5s (see attachement #1)
16.45 Checks:
DSSSD bias & leakage current - Grafana - attachments 2
leakage current ramping, spill micro structure observable
FEE64 temperatures OK - attachment 3
ADC data item stats - attachment 5
per FEE64 Rate spectra - attachment 4
Merger etc - attachment 6
ucesb - attachment 7
XX.XX Checks
DSSSD bias & leakage current - Grafana - attachments X
FEE64 temperatures OK - attachment X
ADC data item stats - attachment X
per FEE64 Rate spectra - attachment X
Merger etc - attachment X
ucesb - attachment X |
Thu Apr 25 08:26:50 2024, TD, 08:00-16:00 Thursday 25 April 42x
|
09.56 DAQ continues data file S100_alpha/R18_387
FRS switching settings
DSSSD bias & leakage current OK - attachments 1
FEE64 temperatures OK - attachment 2
aida02 ASIC temp - known fault
All system wide checks OK *except* as shown - attachments 3-5
WR timestamps OK - attachment 6
ADC, PAUSE, RESUME & Correlation Scaler data time stats - attachments 7-10
per FEE64 Rate spectra - attachments 11
per p+n FEE64 1.8.L spectra - attachments 12-13
aida09 pulser peak width 57 ch FWHM
common x/y scale => peak height proxy for width => all p+n FEE64s similar pulser peak widths
per n+n FEE64 1.8.L spectra - attachment 14
per FEE64 1.8.H spectra - attachments 15-16
per FEE64 1.8.W spectra - 20us FSR - attachments 17-18
only aida08 & aida09 currently updating - ASIC check load and/or DAQ STOP/GO required?
Merger, TapeServer etc - attachments 19-21
BNC PB-5 settings - attachment 22
10.07 data file S100_alpha/R20
new run for new FRS setting
10.08 all histograms, stats, merger stats and data link stats zero'd
10.10 ADC data item stats (beam OFF) - attachment 23
11/16 < 20k, max 129k (aida08)
10.31 analysis of data file R18_349 (closed 06.00 this morning)
max deadtime 12% (aida04)
no timewarps
HEC data item rate 1.7kHz
11.00 tuning 168dy FRS setting
ucesb - attachment 24
spill length 3.5s, extraction time 1.5s
11.30 degraders now set to implant in AIDA DSSSDs
high rates expected
per FEE64 Rate spectra - attachment 25
ADC data item stats - attachment 26
Merger etc - attachment 27
in spill rates to MBS up 30+Mb/s, 2.5M data item/s merged
DSSSD bias & leakage current - attachment 28
c. few hundred nA step in leakage current
analysis of data file R20_33 - attachment 29
max deadtime 18% (aida08)
no timewarps
HEC data item rate 4.1kHz (c. 2.4x previous FRS setting)
Correcting for spill cycle => 5.9kHz in spill (online analysis shows no HEC events observed outside spill)
assume typical HEC event multiplicity = 3 => implant rate ~ 2.0kHz in spill or 6800 implants per spill
SC42 L/R currently report c. 4.5kHz
Note that we currently see a higher fraction of events in DSSSD#2 cf. previous setting so multiplicity assumption may be low
If we assume multipicity = 6 => implant rate ~ 1.0kHz in spill or 3200 implants per spill
we are probably somewhere in between
12.50 DSSSD bias & leakage current - Grafana - attachment 30
leakage current ramping, spill micro structure observable
FEE64 temperatures OK - attachment 31
ADC data item stats - attachment 32
per FEE64 Rate spectra - attachment 33
Merger etc - attachment 34
14.56 analysis of data file R20_106 - attachment 35
max deadtime c. 18% (aida04 & aida08), c. 12% (aida03), c. 11% (aida06), c. 9% (aida02 & aida13), c. 6% (aida05), all others <2%
no timewarps
HEC data item rate 4.2kHz
15.40 DSSSD bias & leakage current - Grafana - attachments 36-37
leakage current ramping, spill micro structure observable
FEE64 temperatures OK - attachment 38
ADC data item stats - attachment 39
per FEE64 Rate spectra - attachment 40
Merger etc - attachment 41
ucesb - attachment 42
15.57 all histograms zero'd
|
Wed Apr 24 21:39:51 2024, TD, 00:00-08:00 Thursday 25 April 12x
|
22.43 24.4.24
DSSSD bias & leakage current OK - attachment 1
FEE64 temperatures OK - attachment 2
ADC data item stats - attachment 3
per FEE64 rate spectra - attachments 4
Merger etc - attachment 5
analysis of data file R18_266 - attachment 6
max dead time 12% (aida04)
no timewarps
HEC data item rate 1.7kHz
03.37 DSSSD bias & leakage current OK - attachments 7-8
FEE64 temperatures OK - attachment 9
ADC data item stats - attachment 10
per FEE64 rate spectra - attachments 11
Merger etc - attachment 12 |
Wed Apr 24 18:44:05 2024, PP, 19:30 checks 8x
|
All seems smooth.
Screenshots attached. |
Wed Apr 24 15:37:59 2024, PP, 16:00-00:00 shift Wednesday 24 April 8x
|
All seems OK.
Screenshots attached. |
Wed Apr 24 09:04:55 2024, TD, 08:00-16:00 Wednesday 24 April 33x
|
09.56 DAQ continues data file S100_alpha/R18_122
FRS setting 162Eu
DSSSD bias & leakage current OK - attachments 1-2
FEE64 temperatures OK - attachment 3
aida02 ASIC temp - known fault
All system wide checks OK *except* as shown - attachments 4-6
WR timestamps OK - attachment 7
ASIC settings aida01 & aida02 - attachments 8-9
ADC, DISC, PAUSE, RESUME, WR & Correlation Scaler data time stats - attachments 10-15
per FEE64 Rate spectra - attachments 16-18
unusually can observe HEC Rates esp. in DSSSD #1
per p+n FEE64 1.8.L spectra - attachments 19-21
aida09 pulser peak width 61 ch FWHM
common x/y scale => peak height proxy for width => all p+n FEE64s similar pulser peak widths
per n+n FEE64 1.8.L spectra - attachment 22
per FEE64 1.8.H spectra - attachments 23-24
per FEE64 1.8.W spectra - 20us FSR - attachments 25-26
only aida08 & aida09 currently updating - ASIC check load and/or DAQ STOP/GO required?
Merger, TapeServer etc - attachments 27-28
ucesb - attachments 29-30
spill cycle reported as 3.5s, extraction 1.5s
10.36 All histograms & stats zero'd
BNC PB-5 settings - attachment 31
DSSSD bias & leakage current - Grafana Alerting - attachment 32
can observe bias & leakage current in greater detail
DSSSD #1 bias stable, c. 10s nA micro structure observed in leakage current
DSSSD #2 bias c. 10mV variations, micro structure observed in leakage current but smaller variations cf. DSSSD#1
10.46 analysis of data file R18_131
max dead time 12% aida04, 6% aida02, all others 0.1% or less
no timewarps
HEC data item rates 1.6kHz - correcting for spill cycle => 2.3kHz in spill (online analysis shows no HEC events observed outside spill)
assume typical HEC event multiplicity = 3 => implant rate ~ 760Hz in spill or 2700 implants per spill
SC42 L/R currently report c. 2kHz
|
Tue Apr 23 22:37:46 2024, TD, 00:00-08:00 Wednesday 24 April 18x
|
23.34 23.4.24
DSSSD bias & leakage current OK - attachment 2
FEE64 temperatures OK - attachment 2
ADC data item stats - attachment 3
per FEE64 rate spectra - attachments 4
Merger etc - attachment 5
03.41 fDSSSD bias & leakage current OK - attachment 6
FEE64 temperatures OK - attachment 7
ADC data item stats - attachment 8
per FEE64 rate spectra (off and on spill) - attachments 9-10
Merger etc - attachment 11
03.54 analysis data file R18_67 - attachment 12
max deadtime 11% (aida04)
HEC data item rate c. 1.5kHz
no timewarps
0.35 fDSSSD bias & leakage current OK - attachment 13
FEE64 temperatures OK - attachment 14
ADC data item stats - attachment 15
per FEE64 rate spectra - attachments 16
Merger etc - attachment 17
03.54 analysis data file R18_96 - attachment 18
max deadtime 12% (aida04)
HEC data item rate c. 1.6kHz
no timewarps |
Tue Apr 23 19:23:42 2024, PP, 20:30 checks 8x
|
All looks good.
Screenshots attached. |
Tue Apr 23 15:11:50 2024, TD, Tuesday 23 April 16:00-00:00 20x
|
16.12 Analysis of data file R17_101 - attachment 1
max deadtime c. 2% (aida02)
no timewarps observed
16.15 DSSSD bias & leakage current OK - attachment 2
FEE64 temperatures OK - attachment 7
ADC data item stats - attachment 3
per FEE64 rate spectra (on and off spill) - attachments 4-5
Merger etc - attachment 6
16.53 aida04 stops producing data
DAQ STOP fails for aida04
Restart Merger to enable aida04 to complete DAQ STOP
aida04 rebooted and DAQ restarted OK
data file S100_alpha/R18 |
Tue Apr 23 13:15:43 2024, PP, 14:00 checks 8x
|
All looks good.
Screenshots attached |
Tue Apr 23 13:08:28 2024, TD, Summary of AIDA FEE64 stops 
|
We have observed a number of AIDA FEE64s stop producing data.
Summary of dates & (approximate) times
aida03 21.4.24 08.20
aida04 22.4.24 09.14 https://elog.ph.ed.ac.uk/DESPEC/586
aida01 22.4.24 16.00
aida02 22.4.24 22.00
aida02 23.4.24 10.21
aida02 23.4.24 11.25
aida04 23.4.24 16.53? attachment 2
aida04/aida05 25.4.24 22.28 no messages in aida04 system console
aida04 27.8.24 c.23:00
aida04 28.8.24 c. 12-13:00 power cycle
aida04 28.8.24 c. 16.00 reset per Elog 618
aida02 28.4.24 c. 22.20 reset per Elog 618, no error messages in system console
aida04 29.4.24 01.15
aida04 30.4.24 <03.30
aida02 30.4.24 <10.00
aida02 2.5.24 c. 19.30 reset
In future, it would be useful to copy system console reports to the Elog to check whether we can identify common error messages etc.
ssh pi@nnrpi1
<password>
cd logs
cat /ttyUSBx
The current mapping of USB port # to AIDA FEE64s - attachment 1
aida02 system console c. 19.30 2.5.24
02:05:24/14:40:51|executing generic doStop
02:05:24/19:27:28|Halt (1)
02:05:24/19:27:29|Halt (2)
02:05:24/19:27:29|Halt (3)
02:05:24/19:27:30|Halt (4)
02:05:24/19:27:30|Halt (5)
02:05:24/19:27:31|Halt (6)
02:05:24/19:27:31|Halt (7)
02:05:24/19:27:32|Halt (8)
02:05:24/19:27:32|Halt (9)
02:05:24/19:27:33|Halt (10)
02:05:24/19:27:33|Halt (11)
02:05:24/19:27:34|Halt (12)
02:05:24/19:27:34|Halt (13)
02:05:24/19:27:35|Halt (14)
02:05:24/19:27:35|Halt (15)
02:05:24/19:27:36|Halt (16)
02:05:24/19:27:36|Halt (17)
02:05:24/19:27:37|Halt (18)
02:05:24/19:27:38|Halt (19)
02:05:24/19:27:38|Action has not completed
02:05:24/19:27:39|Giving up waiting
02:05:24/19:27:39|disconnect xfer stream #0 (1)
02:05:24/19:27:39|disconnect xfer stream #0 (2)
02:05:24/19:27:39|disconnect xfer stream #0 (3)
02:05:24/19:27:40|disconnect xfer stream #0 (4)
02:05:24/19:27:40|disconnect xfer stream #0 (5)
02:05:24/19:27:41|disconnect xfer stream #0 (6)
02:05:24/19:27:41|disconnect xfer stream #0 (7)
02:05:24/19:27:42|disconnect xfer stream #0 (8)
02:05:24/19:27:42|disconnect xfer stream #0 (9)
02:05:24/19:27:43|disconnect xfer stream #0 (10)
02:05:24/19:27:43|disconnect xfer stream #0 (11)
02:05:24/19:27:44|disconnect xfer stream #0 (12)
02:05:24/19:27:44|disconnect xfer stream #0 (13)
02:05:24/19:27:45|disconnect xfer stream #0 (14)
02:05:24/19:27:46|disconnect xfer stream #0 (15)
02:05:24/19:27:46|disconnect xfer stream #0 (16)
02:05:24/19:27:47|disconnect xfer stream #0 (17)
02:05:24/19:27:47|disconnect xfer stream #0 (18)
02:05:24/19:27:48|disconnect xfer stream #0 (19)
02:05:24/19:27:48|Action has not completed
02:05:24/19:27:49|Giving up waiting
02:05:24/19:27:49|executing generic Stop_Merger
02:05:24/19:27:49|executing generic Stop_DataStorageServer
02:05:24/19:27:49|completed generic doStop
02:05:24/19:27:49|do_GetState returned z=0 and 1
02:05:24/19:27:49|do_GetState returned z=0 and 1
02:05:24/19:28:01|Transfer Error - : Connection reset by peer
02:05:24/19:28:08|send() failed:
02:05:24/19:28:08|Aida state now Stopped. AidaExecV10.0_Jul 6 2022: Build 15:10:57
02:05:24/19:28:08|do_GetState returned z=0 and 0
02:05:24/19:28:13|executing generic doGo
02:05:24/19:28:45|Initialising communication with storage server 192.168.11.99 on port 11002
02:05:24/19:28:45|Setting Transfer Block Size 65536
02:05:24/19:28:45|Setting Transfer Mode 3
02:05:24/19:28:45|Setting Overlap Mode 0
02:05:24/19:28:45|Setting Blocking Mode 0
02:05:24/19:28:45|Setting nice 0
02:05:24/19:28:45|TCP transfer library version 3.12
02:05:24/19:28:45|TCP socket send buffer was 16384 - now 221184
02:05:24/19:28:45|TCP socket receive buffer was 87380 - now 221184
02:05:24/19:28:45|TCP socket created OK - now connecting to 192.168.11.99 port 11002
02:05:24/19:28:45|Connected to 192.168.11.99 port 11002
02:05:24/19:28:45|connect xfer stream #0 (1)
02:05:24/19:28:45|Data Acquisition Statistics counters now cleared
02:05:24/19:28:46|ASIC DMA buffer size 2097152; WAVE DMA buffer size 1048576
02:05:24/19:28:46|Aida state now Going. AidaExecV10.0: Build Jul 6 2022_15:10:57
02:05:24/19:28:46|Go (1)
02:05:24/19:28:46|Running Go_Electronics for aida02 with TSMaster aida01
02:05:24/19:28:46|ADCs Calibrated.
02:05:24/19:28:46|Finished Go_Electronics for AIDA
02:05:24/19:28:46|executing generic Go_Merger
02:05:24/19:28:46|executing generic Go_DataStorageServer
02:05:24/19:28:46|completed generic doGo
02:05:24/19:28:46|do_GetState returned z=0 and 1
02:05:24/19:28:46|In RDOGo_Operate: Enabled Correlation, ASIC and discriminator readout
02:05:24/19:28:51|Also in RDOGo_Operate: Enabled waveform readout
02:05:24/19:28:51|get_WAVEBlk (A) which = 0 ---- status = 0048, State machines = 0201
02:05:24/19:28:51|WV:0:Buffer info. Length = 64 : Offset = 0
02:05:24/19:28:51|WV:1:Buffer info. Length = 62 : Offset = 2
02:05:24/19:28:51|WV:2:Buffer info. Length = 60 : Offset = 4
02:05:24/19:28:52|WV:3:Buffer info. Length = 58 : Offset = 6
02:05:24/19:28:52|WV:4:Buffer info. Length = 56 : Offset = 8
02:05:24/19:28:52|WV:5:Buffer info. Length = 54 : Offset = 10
02:05:24/19:28:52|WV:6:Buffer info. Length = 52 : Offset = 12
02:05:24/19:28:52|WV:7:Buffer info. Length = 50 : Offset = 14
02:05:24/19:28:52|WV:8:Buffer info. Length = 48 : Offset = 16
02:05:24/19:28:52|WV:9:Buffer info. Length = 46 : Offset = 18
02:05:24/19:28:52|get_WAVEBlk (A) which = 1 ---- status = 0088, State machines = 0201
02:05:24/19:28:52|get_WAVEBlk (A) which = 0 ---- status = 0048, State machines = 0908
02:05:24/19:28:52|get_WAVEBlk (A) which = 1 ---- status = 0088, State machines = 0908
02:05:24/19:28:52|get_WAVEBlk (A) which = 0 ---- status = 0048, State machines = 0908
02:05:24/19:28:52|get_WAVEBlk (A) which = 1 ---- status = 0088, State machines = 0908
02:05:24/19:28:52|get_WAVEBlk (A) which = 0 ---- status = 0048, State machines = 0908
02:05:24/19:28:52|get_WAVEBlk (A) which = 1 ---- status = 0088, State machines = 0908
02:05:24/19:28:52|get_WAVEBlk (A) which = 0 ---- status = 1048, State machines = 0201
02:05:24/19:28:52|get_WAVEBlk (A) which = 1 ---- status = 0088, State machines = 0908
02:05:24/19:28:52| |
Tue Apr 23 11:09:00 2024, PP, 12:00 checks 8x
|
Everything is OK now.
Screenshots attached. |
Tue Apr 23 10:52:39 2024, PP, Aida02 down and back
|
We lost aida02 at about 11:25 again. Nick fixed it about 11:50.
It is now back in the DAQ. |
Tue Apr 23 10:10:31 2024, PP, aida02 is back
|
Nick rebooted AIDA and aida02 came back at around 10:55.