AIDA GELINA BRIKEN nToF CRIB ISOLDE CIRCE nTOFCapture DESPEC DTAS EDI_PSA 179Ta CARME StellarModelling DCF K40
  DESPEC, Page 9 of 36  ELOG logo
Entry  Sun Apr 7 12:02:23 2024, TD, Sunday 7 April 46x

13.00 Cooling water temperature and flow OK - attachment 1
Entry  Sun Apr 7 12:16:27 2024, TD, aida06 system console log 
ISOL Version 1.00 Date 9th January 2017
Flash base address=FC000000
Set Flash to ASync Mode
Entry  Sun Apr 7 13:08:52 2024, TD, FEE64 configuration AIDA_S100_WiringPlan.drawio.png
FEE64 configuration per https://elog.gsi.de/despec/S100/6

Configuration: Wide - 16 FEEs, 2 DSSD
Entry  Sun Apr 7 18:13:56 2024, TD, Sunday 7 April contd. 6x

19.10 CAEN N1419ET LK fitted
      power cycle
Entry  Mon Apr 8 11:18:19 2024, TD, S505 offline analysis data file R3_150 Screenshot_from_2024-03-29_20-21-07.pngScreenshot_from_2024-03-29_20-21-45.pngScreenshot_from_2024-03-29_20-22-29.pngScreenshot_from_2024-03-29_20-23-06.pngvariables.dat

DRG quotes ( from Elog ) S505 FEE64 configuration as
Entry  Tue Apr 9 08:37:47 2024, TD, Tuesday 9 April 16x
09.36 Cooling water temperature and flow OK

      FEE64 power ON
Entry  Tue Apr 9 09:04:07 2024, TD, [How To] Remote control of BNC PB-5 Screenshot_from_2024-04-09_10-03-42.png

BNC PB-5 set to remote control
Entry  Tue Apr 16 23:37:58 2024, TD, Wednesday 17 April 44x

00.40 Observed zero data fron aida03, aida04, aida09
      WR timestamps offset from other FEE64s
Entry  Fri Apr 19 02:35:14 2024, TD, Friday 19 April 38x

03.30
Entry  Sat Apr 20 03:20:23 2024, TD, Saturday 20 April 23x

Note 60Co source (current activity c. 43kBq) was installed yesterday evening - assume it is positioned on the bottom side of the snout as per the other
sources used this week/
Entry  Sat Apr 20 11:40:38 2024, TD, Offline analysis R12 & R13 ( pulser walkthrough ) 17x

S100 ADC offsets using pulser walkthrough data from data files R12 & R13
Entry  Sun Apr 21 00:34:33 2024, TD, Sunday 21 April 37x
01.35 DSSSD bias & leakage current OK - Grafana - attachment 1

      FEE64 temperatures OK - attachment 2
Entry  Sun Apr 21 23:20:45 2024, TD, Monday 22 April 38x
00.15 Data link to MBS was closed by MBS - reason(s) unknown
      NH restarted MBS receiver and data transfer restarted
Entry  Mon Apr 22 09:03:54 2024, TD, Report: aida04 stops producing data Screenshot_from_2024-04-22_09-49-30.png

09.14 aida04 stopped producing data - system console output - attachment 1
Entry  Mon Apr 22 11:43:55 2024, TD, 21-22.4.24 overnight MIDASsort online sort  7x

Attachment 1 - per DSSSD LEC e_p versus e_n (20keV/channel)
Entry  Mon Apr 22 22:35:59 2024, TD, 00.00-08.00 Tuesday 22 April 26x

23.36 Analysis of data file R17_34 - attachment 1
      max deadtime c. 1% (aida02)
Entry  Tue Apr 23 13:08:28 2024, TD, Summary of AIDA FEE64 stops Screenshot_from_2024-04-23_14-22-04.pngaida04

We have observed a number of AIDA FEE64s stop producing data.
Entry  Tue Apr 23 15:11:50 2024, TD, Tuesday 23 April 16:00-00:00 20x

16.12 Analysis of data file R17_101 - attachment 1
Entry  Tue Apr 23 22:37:46 2024, TD, 00:00-08:00 Wednesday 24 April 18x
23.34 23.4.24

      DSSSD bias & leakage current OK - attachment 2
Entry  Wed Apr 24 09:04:55 2024, TD, 08:00-16:00 Wednesday 24 April 33x

09.56 DAQ continues data file S100_alpha/R18_122
      FRS setting 162Eu
ELOG V3.1.4-unknown