  AIDA, Page 4 of 46  ELOG logo
Entry  Tue Oct 24 03:16:07 2017, TD, OH, Tuesday 24 October 20x
11.16 DAQ continues
      AIDA file Oct17/R8_107

      BNC PB-4
      Amplitude 90,000
      x5 attenuator IN
      t_r 100ns tau_d 50us
      - polarity
      frequency range 25Hz
      delay MIN, tail pulse, INT ref

      ASIC settings 2017Oct23-14.38.03

11.20 Detector biases & leakage currents OK - see attachments 1 & 2
      FEE64 Temperatures OK - see attachment 3
      good events statistics stable & OK - see attachments 4 & 5
      no merger/tapeserver errors reported since start of run - see attachments 6 & 7
      *all* system wide checks OK *except* nnaida21 fails ADC calibration

11.30 Ambient temperature +23.3 deg C, d.p. +4.1 deg C, RH 28.7%

11.30 analysis file Oct17/R7_108 - zero deadtime, 6x MBS and zero ts timewarps - see attachment 8

11.39 DAQ stop

11.44 DAQ start
      AIDA file Oct17/R9

      Pulser walkthrough

      BNC PB-4
      Amplitude 90,000-10,000 @ 10,000 step
      x5 attenuator IN
      t_r 100ns tau_d 50us
      - polarity
      frequency range 250Hz
      delay MIN, tail pulse, INT ref

11.53 DAQ stop
      See attachment 8

11.44 DAQ start
      AIDA file Oct17/R10

      Pulser walkthrough

      BNC PB-4
      Amplitude 12,000-2,000 @ 2,000 step
      x5 attenuator IN
      t_r 100ns tau_d 50us
      - polarity
      frequency range 250Hz
      delay MIN, tail pulse, INT ref

11.53 DAQ stop
      See attachment 9

12.06 DAQ starts
      AIDA file Oct17/R12

      BNC PB-4
      Amplitude 90,000
      x5 attenuator IN
      t_r 100ns tau_d 50us
      - polarity
      frequency range 25Hz
      delay MIN, tail pulse, INT ref

      ASIC settings 2017Oct23-14.38.03

14.41 online analysis 
      ADC synchronised - see attachment 10


17.12 DAQ starts
      AIDA file Oct17/R13

      Background for LuckyDoll automatic threshold determination

      BNC PB-4 OFF

      ASIC settings 2017Oct23-14.38.03
      Slow comparator nnaida13 0xc, nnaida19-22 0xb, all other FEE64s 0xa

      DAQ statistics - see attachment 11

17.27 DAQ stop
      
18.03 DAQ starts
      AIDA file Oct17/R16

      Background for LuckyDoll automatic threshold determination

      BNC PB-4 OFF

      ASIC settings 2017Oct23-14.38.03
      Slow comparator *all* FEE64s 0xa

      DAQ statistics - see attachment 12

18.22 DAQ stop

18.48 DAQ statistics
      Slow comparator *all* FEE64s 0xa - see attachment 13
      Slow comparator *all* FEE64s 0x9 - see attachment 14
      Slow comparator *all* FEE64s 0x8 - see attachment 15
      Slow comparator *all* FEE64s 0x7 - see attachment 16

18.53 Detector biases & leakage currents OK - see attachments 13 & 14
      FEE64 Temperatures OK - see attachment 15
      *all* system wide checks OK *except* nnaida21 fails ADC calibration
Entry  Wed Oct 25 03:26:55 2017, TD, OH, Wednesday 25 October 7x
11.27 DAQ continues
      Not writing to disk

      BNC PB-4
      Amplitude 90,000
      x5 attenuator IN
      t_r 100ns tau_d 50us
      - polarity
      frequency range 25Hz
      delay MIN, tail pulse, INT ref

      ASIC settings 2017Oct23-14.38.03
      *all* FEE64s slow comparators 0xa

      Detector biases & leakage currents OK - see attachments 1 & 2
      FEE64 Temperatures OK - see attachment 3
      good events statistics stable & OK - see attachments 4
      *all* system wide checks OK *except* nnaida21 fails ADC calibration

11.47 Ambient temperature +23.9 deg C, d.p. +9.0 deg C, RH 39.0%

16:45 System wide checks all ok 
      FEE temps ok - Attachment 5
      Stats - Attachment 6
      Bias and leakage ok - Attachment 7
Entry  Wed Oct 25 18:36:46 2017, TD, OH, Thursday 26 October 28x
02.37 DAQ stalled - USB-controlled ac mains relay fuse blown
      Not writing to disk

      BNC PB-4
      Amplitude 90,000
      x5 attenuator IN
      t_r 100ns tau_d 50us
      - polarity
      frequency range 25Hz
      delay MIN, tail pulse, INT ref

      ASIC settings 2017Oct23-14.38.03
      *all* FEE64s slow comparators 0xa

      Detector biases & leakage currents OK - see attachments 1 & 2
      *all* system wide checks OK *except* nnaida21 fails ADC calibration
      Ambient temperature +24.3 deg C, d.p. +8.7 deg C, RH 37.1%

02.54 Replace 250V/16A fuse and power up

      FEE64 Temperatures OK - see attachment 3
      good events statistics stable & OK - see attachments 4

03:30 Stats with nnaida13 slow comparator threshold raised to 140keV - Attachment 5

03:50 Pulser widths done, slow comparator threshold raised to 0xc ~120keV for  nnaida 19-22
      Statistics - Attachment 6

05:40 Beam now going into F11. Wasabi and YSol in front of AIDA
      AIDA is seeing light particles
      Waveforms look good
      Layout 7 - Attachment 7
      layout 8 - Attachment 8
      Statistics with light particles going through - Attachment 9
      Bias and leakage currents ok - Attachment 10

      We are taking the light ion data to file in R17_*

06:50 Light ion data continues
      System wide checks all ok

07:22 Bias and leakage ok - Attachment 11
      FEE Temperatures ok - Attachment 12
      Stats - Attachment 13

08:15 Light ions stop still writing file

08:27 Beam into aida run R17_30
      Zero degrader, just 4mm of WAS3ABI
      WAS3ABI run 120
      BRIKEN started run at 08:27 run 18
      BIGRIPS Run 8
      Timestamp reset at 08:27
      Stats with beam - Attachment 14

08:44 Waveforms with implants in the detector
      Layout 7 - Attachments 15 & 16
      Layout 8 - Attachments 17 & 18
      Layout 1  linear and log attachments 19 & 20 
      Leakage and bias ok - Attachment 21

09:14 Just noticed the stats in nnaida13 has dropped rate. Noise source may have been turned off
      Statistics - Attachment 22

      Spectra from the 0mm degrader run - Attachments 22-27

09:20 BRIKEN and WAS3ABI stop on run 162 aida on R17_35
      1mm degrader moved in
      BigRIPS on run 9
      BRIKEN run 19
      WAS3ABI start on 163
      AIDA on R17_35

      Rates similar to last screenshot of stats 09:14

09:26 Leakage and bias ok - Attachment 28

09:41 Beam stops
      AIDA on file R17_37
      BRIKEN stops running
      WAS3ABI on run 178

AIDA Files written
-rw-rw-rw-. 1 npg npgstaff 2.0G Oct 26 08:22 R17_29
-rw-rw-rw-. 1 npg npgstaff 2.0G Oct 26 08:31 R17_30
-rw-rw-rw-. 1 npg npgstaff 2.0G Oct 26 08:40 R17_31
-rw-rw-rw-. 1 npg npgstaff 2.0G Oct 26 08:50 R17_32
-rw-rw-rw-. 1 npg npgstaff 2.0G Oct 26 09:00 R17_33
-rw-rw-rw-. 1 npg npgstaff 2.0G Oct 26 09:10 R17_34
-rw-rw-rw-. 1 npg npgstaff 2.0G Oct 26 09:21 R17_35
-rw-rw-rw-. 1 npg npgstaff 2.0G Oct 26 09:31 R17_36
-rw-rw-rw-. 1 npg npgstaff 2.0G Oct 26 09:42 R17_37
-rw-rw-rw-. 1 npg npgstaff 329M Oct 26 09:44 R17_38
Entry  Fri Oct 27 01:58:44 2017, TD, OH, Friday 27 October 12x
09.56 DAQ stopped OK yesterday

      BNC PB-4
      Amplitude 90,000
      x5 attenuator IN
      t_r 100ns tau_d 50us
      - polarity
      frequency range 25Hz
      delay MIN, tail pulse, INT ref

      ASIC settings 2017Oct23-14.38.03
      *all* FEE64s slow comparators 0xa *except* nnaida13 0xe, nnaida19-22 0xc

      Detector biases & leakage currents OK - see attachments 1 & 2
      FEE64 Temperatures OK - see attachment 3
      *all* system wide checks OK *except* nnaida21 fails ADC calibration
      Julabo FL11006 recirculating chiller coolant reservoir refilled - set point +20 deg C, actual +20.0 deg C
      Ambient temperature +23.6 deg C, d.p. +6.2 deg C, RH 32.2%

      AIDA withdrawn from (CH2)n matrix

      DAQ started - no data transfer
      good events statistics stable - see attachment 4
      *all* system wide checks OK *except* nnaida21 fails ADC calibration

13:30 System wide checks ok
      Stats - Attachment 6
      FEE temperatures ok - Attachment 7
      leakage and bias ok - Attachment 8

17.15 Detector biases & leakage currents OK - see attachments 9 & 10
      FEE64 Temperatures OK - see attachment 11
      good events statistics stable - see attachment 12
      *all* system wide checks OK *except* nnaida21 fails ADC calibration
      Ambient temperature +25.2 deg C, d.p. +5.7 deg C, RH 28.5%
Entry  Sat Oct 28 03:33:10 2017, TD, OH, Saturday 28 October 13x
11.33 DAQ running OK - date transfer not enabled

      BNC PB-4
      Amplitude 90,000
      x5 attenuator IN
      t_r 100ns tau_d 50us
      - polarity
      frequency range 25Hz
      delay MIN, tail pulse, INT ref

      ASIC settings 2017Oct23-14.38.03
      *all* FEE64s slow comparators 0xa *except* nnaida13 0xe, nnaida19-22 0xc

      Detector biases & leakage currents OK - see attachments 1 & 2
      FEE64 Temperatures OK - see attachment 3
      DAQ statistics OK - see attachment 4
      *all* system wide checks OK *except* nnaida21 fails ADC calibration
      Julabo FL11006 recirculating chiller set point +20 deg C, actual +20.0 deg C
      Ambient temperature +26.2 deg C, d.p. +6.2 deg C, RH 28.3%

14.40 Detector biases & leakage currents OK - see attachments 5 & 6
      FEE64 Temperatures OK - see attachment 7
      DAQ statistics OK - see attachment 8
      *all* system wide checks OK *except* nnaida21 fails ADC calibration

14.50 DAQ starts
      AIDA data file Oct17/R18

14.53 per FEE64 rates (20us/channel) - see attachment 9
      No indication of c. 400us period noise as indicated by DAQ statistics attachment 8

15.46 DAQ stopped

      DAQ start - data tranfser disabled

19.40 Detector biases & leakage currents OK - see attachments 10 & 11
      FEE64 Temperatures OK - see attachment 12
      DAQ statistics OK - see attachment 13
      *all* system wide checks OK *except* nnaida21 fails ADC calibration
Entry  Sun Oct 29 05:16:35 2017, TD, OH, Sunday 29 October 14x
14.16 DAQ running OK - data transfer not enabled

      BNC PB-4
      Amplitude 90,000
      x5 attenuator IN
      t_r 100ns tau_d 50us
      - polarity
      frequency range 25Hz
      delay MIN, tail pulse, INT ref

      ASIC settings 2017Oct23-14.38.03
      *all* FEE64s slow comparators 0xa *except* nnaida13 0xe, nnaida19-22 0xc

      Detector biases & leakage currents OK - see attachments 1 & 2
      FEE64 Temperatures OK - see attachment 3
      DAQ statistics OK - see attachment 4
      *all* system wide checks OK *except* nnaida21 fails ADC calibration
      Julabo FL11006 recirculating chiller set point +20 deg C, actual +20.0 deg C
      Ambient temperature +23.9 deg C, d.p. +10.8 deg C, RH 43.2%

15.05 per FEE64 Rate, 1.8.L and 1.8.W spectra - note common x & y scales
      Peak width nnaida5 63 ch FWHM
      Peak width nnaida8 82 ch FWHM

15.09 All histograms zero'd

18.25 Detector biases & leakage currents OK - see attachments 11 & 12
      FEE64 Temperatures OK - see attachment 13
      DAQ statistics OK - see attachment 14
      *all* system wide checks OK *except* nnaida21 fails ADC calibration
Entry  Mon Oct 30 01:10:28 2017, TD, OH, Monday 30 October 8x
10.10 DAQ running OK - data transfer not enabled

      BNC PB-4
      Amplitude 90,000
      x5 attenuator IN
      t_r 100ns tau_d 50us
      - polarity
      frequency range 25Hz
      delay MIN, tail pulse, INT ref

      ASIC settings 2017Oct23-14.38.03
      *all* FEE64s slow comparators 0xa *except* nnaida13 0xe, nnaida19-22 0xc

      Detector biases & leakage currents OK - see attachments 1 & 2
      FEE64 Temperatures OK - see attachment 3
      DAQ statistics OK - see attachment 4
      *all* system wide checks OK *except* nnaida21 fails ADC calibration

15.56 Detector biases & leakage currents OK - see attachments 5 & 6
      FEE64 Temperatures OK - see attachment 7
      DAQ statistics OK - see attachment 8
      *all* system wide checks OK *except* nnaida21 fails ADC calibration
Entry  Tue Oct 31 00:34:10 2017, TD, OH, Tuesday 31 October 8x
09.34 DAQ running OK - data transfer not enabled

      BNC PB-4
      Amplitude 90,000
      x5 attenuator IN
      t_r 100ns tau_d 50us
      - polarity
      frequency range 25Hz
      delay MIN, tail pulse, INT ref

      ASIC settings 2017Oct23-14.38.03
      *all* FEE64s slow comparators 0xa *except* nnaida13 0xe, nnaida19-22 0xc

      Detector biases & leakage currents OK - see attachments 1 & 2
      FEE64 Temperatures OK - see attachment 3
      DAQ statistics OK - see attachment 4
      *all* system wide checks OK *except* nnaida21 fails ADC calibration

15.02 Detector biases & leakage currents OK - see attachments 5 & 6
      FEE64 Temperatures OK - see attachment 7
      DAQ statistics OK - see attachment 8
      *all* system wide checks OK *except* nnaida21 fails ADC calibration
      Ambient temperature +24.3 deg C, d.p. -0.2 deg C, RH 19.8%
      Julabo FL11006 recirculating chiller set point +20 deg C, actual +20.0 deg C, water level OK c. mid-point
middle-upper marker
     
Entry  Thu Nov 2 01:56:18 2017, TD, OH, Thursday 2 November  16x
10.56 DAQ running OK - data transfer not enabled

      BNC PB-4
      Amplitude 90,000
      x5 attenuator IN
      t_r 100ns tau_d 50us
      - polarity
      frequency range 25Hz
      delay MIN, tail pulse, INT ref

      ASIC settings 2017Oct23-14.38.03
      *all* FEE64s slow comparators 0xa *except* nnaida13 0xe, nnaida19-22 0xc

      Detector biases & leakage currents OK - see attachments 1 & 2
      FEE64 Temperatures OK - see attachment 3
      DAQ statistics OK - see attachment 4
      *all* system wide checks OK *except* nnaida21 fails ADC calibration

      p+n FEE64s 1.8.L spectra - see attachment 5 - nnaida17 FWHM 63 ch
      n+n FEE64s 1.8.L spectra - see attachment 6 - nnaida3 FWHM 78 ch
      per FEE64 rate spectra - see attachments 7 & 8

11.06 all spectra zero'd

17.28 Detector biases & leakage currents OK - see attachments 9 & 10
      FEE64 Temperatures OK - see attachment 11
      DAQ statistics OK - see attachment 12
      *all* system wide checks OK *except* nnaida21 fails ADC calibration

18.46 per PJCS request two screenshots of nnaida21:8015/AIDA/MONLOCK/MONLOCK.tml
      see attachments 13 & 14

23:52 Statistics still good - Attachment 15
      Leakage and bias ok - Attachment 16
Entry  Fri Nov 3 23:09:36 2017, TD, OH, Saturday 4 November 20x
08.09 DAQ running OK - data transfer not enabled

      BNC PB-4
      Amplitude 90,000
      x5 attenuator IN
      t_r 100ns tau_d 50us
      - polarity
      frequency range 25Hz
      delay MIN, tail pulse, INT ref

      ASIC settings 2017Oct23-14.38.03
      *all* FEE64s slow comparators 0xa *except* nnaida19-22 0xc

      Detector biases & leakage currents OK - see attachments 1 & 2
      FEE64 Temperatures OK - see attachment 3
      DAQ statistics OK - see attachment 4
      *all* system wide checks OK *except* nnaida21 fails ADC calibration - see attachment 6
      PLL and DCM Clock Lock Monitor OK - see attachment 5

08.22 per FEE64 Rate, 1.8.L and 1.8.W spectra - see attachments 7-12
      nnaida3 78 ch FWHM
      nnaida17 66 ch FWHM

16.43 Detector biases & leakage currents OK - see attachments 13 & 14
      FEE64 Temperatures OK - see attachment 15
      DAQ statistics OK - see attachment 16
      *all* system wide checks OK *except* nnaida21 fails ADC calibration - see attachment 6
      PLL and DCM Clock Lock Monitor OK - see attachment 5

      per FEE64 Rate, 1.8.L pectra - see attachments 17-20
      nnaida3 78 ch FWHM
      nnaida17 68 ch FWHM
Entry  Thu Jun 21 02:31:49 2018, TD, OH, Thursday 21 June 30.png31.png32.png33.png34.png
10.20 DAQ continues OK

10.29 DAQ stop June18/R1

      ASIC settings 2017Nov17-21.24.47
      *all* FEE64 slow comparators 0x64

      Detector biases & leakage currents OK - see attachments 1 & 2
      FEE64 Temperatures OK - see attachment 3
      DAQ statistics OK - see attachment 4
      *all* system wide checks OK
      PLL and DCM Clock Lock Monitor OK - see attachment 5
Entry  Thu Jun 21 05:23:16 2018, TD, OH, Offline analysis R1 - alpha background 8x
Analysis of overnight alpha background run (June18/R1_0 ... _2) indicates that all FEE64s and DSSSDs are
operating OK.

Spectra for DSSSD #1-6 - see attachments 1-5

Sort program, variables and spectra titles - see attachments 6-7

Sort program statistics

 *** ENTRY finish

 *** data items:  671431805 ( 363493.31 Hz)
 *** ADC events:     743540 (    402.53 Hz)
 *** time warps:          0 (      0.00 Hz)
 *** DSSSD # 1 count:       266 old count:       266 dt:  73278.38 s  LEC rate:      0.00 Hz
 *** DSSSD # 2 count:       119 old count:       119 dt:  73278.38 s  LEC rate:      0.00 Hz
 *** DSSSD # 3 count:       131 old count:       131 dt:  73278.38 s  LEC rate:      0.00 Hz
 *** DSSSD # 4 count:        97 old count:        97 dt:  73278.38 s  LEC rate:      0.00 Hz
 *** DSSSD # 5 count:       103 old count:       103 dt:  73278.38 s  LEC rate:      0.00 Hz
 *** DSSSD # 6 count:       122 old count:       122 dt:  73278.38 s  LEC rate:      0.00 Hz
 *** DSSSD # 1 count:         0 old count:         0 dt:  73278.38 s  HEC rate:      0.00 Hz
 *** DSSSD # 2 count:         0 old count:         0 dt:  73278.38 s  HEC rate:      0.00 Hz
 *** DSSSD # 3 count:         0 old count:         0 dt:  73278.38 s  HEC rate:      0.00 Hz
 *** DSSSD # 4 count:         0 old count:         0 dt:  73278.38 s  HEC rate:      0.00 Hz
 *** DSSSD # 5 count:         0 old count:         0 dt:  73278.38 s  HEC rate:      0.00 Hz
 *** DSSSD # 6 count:         0 old count:         0 dt:  73278.38 s  HEC rate:      0.00 Hz

 *** ENTRY finish ends
Entry  Thu Jun 21 06:48:22 2018, TD, OH, DSSSD HV cabling 50.png51.png
The DSSSD HV cabling has been changed in preparation for configuration change
from 6x DSSSDs to 8x DSSSDs as follows:

FROM                 TO

DSSSD  N1419  HV     DSSSD  N1419  HV
#      Addr   #      #      Addr   #

                     1      0      1
                     2      0      2
1      0      1      3      0      3
2      0      7      4      0      4
3      0      3      5      1      5
4      0      4      6      1      6
5      1      5      7      1      7
6      1      6      8      1      8

See attachments 1 & 2 - DSSSDs # 1-2 not installed yet
Entry  Sat Jun 23 02:28:13 2018, TD, OH, New fuses fitted to USB-controlled ac mains relay 2290195.pdf
The input fuses of the USB-controlled ac mains relay have been changed from 2x 5x20mm 250V/16A to 2x 5x20mm
250V/20A cartridge, time-lag Littelfuse 0215020.MXP
Entry  Sat Nov 18 02:28:26 2023, TD, NH, JB, Saturday 18 November 10x
03.20 c. 3k per spill 13N beam to S4
      Observe scatter of events at all energies for LEC

      Switch LE -> MEC
      aida02 & aida04 slow comparator 0xf -> 0x3 ( c.1.5MeV )
      All other ( p+n ) FEE64s slow comparator 0xa ( c. 5MeV )
      For c. 700MeV/u 14N + 6cm s.steel -> c. 350MeV/u 14N -> c. 30MeV energy loss in 1mm Silicon, c. 45MeV energy loss in 3mm plastic

      per n+n FEE64 1.8L spectra - attachment 1
      observe 3x peaks above threshold - different Z fragments?
      per FEE64 Stat spectra


04.10 DSSSD bias -150V leakage current -5.380uA - attachment 3
      FEE64 temps OK - attachment 4
      ADC data item stats - attachment 5
      per FEE64 Stat spectra - attachment 6
      per FEE64 1.8.L spectra - attachments 7-8
      per FEE64 1.8.W spectra - 20us FSR - attachments 9-10



05.03 aida02 & aida04 slow comparators 0x3 -> 0xa
      All spectra zero'd


c. 06.30 Beam tests end
         DAQ stopped. DSSSD HV OFF. AIDA power OFF.
      
Entry  Fri Dec 8 12:55:38 2023, TD, NH, FEE64 configuration 
FEE64 configuration for 1x MSL type BB18(DS)-1000 24cm x 8cm DSSSD

*Beam into page*

         
                  aida09                           BB7 FEE64

         aida05   aida01   aida06                  p+n junction FEE64s

aida02                              aida04         n+n Ohmic FEE64s
           *
         aida07   aida03   aida08                  p+n junction FEE64s

* approximate position of c. 1.5cm diamond-shaped hole in black cloth optical screen positioned between 'upstream' bPlast and BB18 DSSSD.
Entry  Tue Apr 1 15:34:16 2025, TD, MR, CU, Tests of new AIDA ASIC mezzanines - 2024 production batch 11x
Tests of new AIDA ASIC mezzanines - 2024 production batch

Using nnlycca4/MezzX PCB test bench T9, STFC DL - attachments 1 & 2

nnlycca4 193.62.115.228

BNC PB-4 Pulser
Amplitude ?9000
Attenuator x5 IN
Frequency 25Hz
Polarity +
t_r 50ns
tau_d 200us
Delay MIN
Ref INT

Output amplitude (DSO hi Z input) c. 500mV


MACB
Program switch = 1


Power on sequence - 1 2 3 Mezzanine - attachment 3
1 - FEE64 and ASIC mezzanine cooling fans
2 - FEE64 +6.2V/+6.8A
3 - FEE64 -6.5V/-0.69A, +8V/+0.88A
4 - Mezzanine +5V/+1.7A +7V/+0.03A

At power up ASIC mezzanine +5V/+2.1A which decreases to +5V/+1.7A following DAQ SETUP

Pulser to *all* inputs via test adaptor PCB (PJCS design)

Examples of test signals as follows:

Data Ready DRDY (blue), Address bits A0-A3 (magenta), pulser (yellow) - attachments 4-7 

MUX Out (magenta) - attachment 8

LEC fast comparator (magenta) - attachment 9

Preamp output (magenta) - attachment 10


Following power up and the boot of the embedded PowerPC processor

DAQ RESET
DAQ SETUP
DAQ GO
Start ASIC Readout
*multiple* Check ASIC Load for ASIC #2 (note this check loads all ASICs)

Use DSO scope probe at MezzX PCB test points - attachment 11



All ASICs of the following ASIC mezzanines tested OK

Batch 0216240101
-0009

Batch 0213110101 
-0002

Batch 0324390101 (*new* batch 2024/Q4)

-0010
-0011
-0019
Entry  Mon Jun 6 11:17:30 2016, TD, ML & YS, Re-configure merger to ignore nnaida19 11.png12.png
19.00 Power-cycled and rebooted AIDA FEE64s

      Temperatures are lower - see attachment 1

      Re-configured merge to ignore problem child nnaida19 which continues to ReSYNC but fails
      to pass SYNC errors counting and ASIC clock timestamp system-wide checks - see attachment 2
Entry  Mon Jun 6 13:46:45 2016, TD, ML & YS, /var/log/messages messages
aidas1 log
Entry  Mon Jun 6 13:52:10 2016, TD, ML & YS, FEE64 power-cycle, FEE64 & aidas1 reboot 
21.40 toggled merge input pause state  - merge pause stopped at link 18 (nnaida19 - this link is supposed
      to be disabled 
      
      obtained 'unable to connect' to nnaida1 for DAQ stop

      reboot aidas1, power cycle and reboot FEE64s 