  433   Sat Aug 12 11:57:51 2023 JMUpdated thermocouple positions and heater channel layoutsHeater channel layouts for the interaction
chamber and carme in attached documents. 

Updated thermocouple positions;
 6x 
  431   Thu Aug 10 07:51:43 2023 JMBaking tent and heater testing09/08/23

Attached is the current pumpdown
plot taken from grafana. 
 8x 
  430   Thu Jul 27 09:43:56 2023 RSSThermocouples connected -2
10:30  IE514 1.80e-08mbar

Most
of the CARME is covered by the baking tent.
The tent needs to be closed and a bit covered
 IMG_9920.jpgIMG_9919.jpg 
  429   Wed Jul 26 08:28:09 2023 RSSBuilding tent
09:26  IE514 1.96e-08mbar
 IMG_9897.jpgIMG_9899.jpgIMG_9901.jpgIMG_9902.jpg 
  428   Tue Jul 25 09:12:15 2023 RSSBuilding tent
10:10  IE514 2.0e-08mbar

Upstream
part of the CARME is covered.
 BD01C0C8-5181-49B8-8751-AD4C99A459CC.jpeg7B3836F1-8A7E-491B-AFDC-46033BF84760.jpeg 
  427   Mon Jul 24 10:40:14 2023 RSS, TDPump down continues11.40 Pirani 1.00e-3 mbar IE514 2.23e-08mbar

15.15 7d/24h vacuum plots
 vacplot.pngvacplot24.png 
  426   Sun Jul 23 10:13:48 2023 RSS, TDBuilding tent11.15 Pirani 1.0e-03mbar IE514 2.72e-08mbar

16.35 Pirani 1.0e-03mbar IE514 2.61e-08mbar
 DBE0DA2E-2F69-469C-B645-3E33E3D82138.jpeg857185AC-3571-41E9-88AE-9F624445135B.jpegE69ED1D1-444E-478A-A3B4-481023B4024D.jpeg610048A3-25F5-459C-9EF1-83EFEFDE51C9.jpeg 
  425   Sat Jul 22 09:19:59 2023 RSS, TDThermocouples connected10.20 Pirani 1e-03mbar IE514 3.26e-08mbar

 10x 
  424   Fri Jul 21 08:21:54 2023 TD, RSSBaking tent manual09.20 Pirani 1.00e-3 mbar IE514 3.90e-8 mbar

      Today - install TCs, connect to Pico
 CARME-baking-guide_(1).pdf 
  423   Thu Jul 20 08:29:26 2023 TDInstalling tent. Two NEG power cable broken.09.28 Pirani 1.00e-3 mbar IE514 5.45e-8 mbar

12.10 Bakeout tent frame installed
 11x 
  422   Wed Jul 19 08:17:26 2023 TDLeak testing09.16 Pirani 1.00e-3 mbar IE514 8.77e-8 mbar
      Leak tester c. 5e-12 mbar.l/s (leak
tester valved off)
  
  421   Tue Jul 18 08:16:28 2023 TDPumping down. Issues found with one detector.09.15 Following overnight pumping of CARME
chamber by Leybold turbo
      Pirani 1.00e-3 mbar IE514 2.83e-7 mbar
  
  420   Mon Jul 17 08:45:11 2023 TDClosing CARME again
Grosmantage are scheduled to arrive 10.00
to inspect COF600 seals and their fit to
 20230717_093106.jpg 
  419   Fri Jul 14 08:27:17 2023 TDCARME open - inspecting seal
09.20 CARME chamber at atmospheric pressure
as expected.
  
  418   Thu Jul 13 08:46:44 2023 TDReplacing wire seal
09.30 CARME chamber status
 20230713_135351.jpg20230713_135416.jpg 
  417   Thu Jul 13 00:12:45 2023 JM, CBCARME closedThe CARME chamber was closed and pushed
back into position on the ring. 

Prior to moving back onto the ring,
 Bias110723.pngfullyin.jpgfullyin-measured.jpg 
  416   Thu Jul 13 00:12:19 2023 JM, CBBellow replacedThe left (inside ring) bellow for the motors
rod was replaced.

Replacing the bellow involved;
 20230705_164523.jpg 
  415   Thu Jul 6 15:26:39 2023 JMLeak updateThe leak on the actuator arm bellows has
been attempted to be sealed using UHV vacuum
seal spray. The spray had the opposite to
 pumping.png 
  414   Wed Jul 5 19:27:13 2023 JMPump down and leak testCARME and the interaction chamber have
been pumping for ~26 hours, P=4.15e-7 mbar,
attached figures show the current pump down
 pumping_1.pngpumping-zoom.pngsignal-2023-07-05-15-13-30-512.jpg 
  413   Tue Jul 4 17:08:08 2023 JMLeak test and ventingThis morning the pressure achieved was
1.44E-7 mbar. Leak testing showed an overall
leak rate of 1.5E-10 mbar l/s, this is below
 20230704_135455.jpg20230704_174524.jpg 