Wed Dec 11 14:46:21 2024, TD, JB, CC, MP, Mounting of AIDA, bPlast and BB7 for test   
|
Snout assembled for detector position and timing test.
The planned test is to use a BGO with a 22-Na source and the implantation stack to characterise the timing and position of back-to-back 511 keV events (proxy for mock beta decay event).
The implantation stack is assembled as follows:
Upstream ----> AIDA test frame (as a spacer) ----> AIDA DSSSD (3208-3/3208-21/3208-22, Upstream AIDA used in S100 & S181, April 2024) ----> bPlast (Downstream bPlast) ----> BB7 (mounted in test frame connected to AIDA adaptor) ----> Downstream
The AIDA DSSSD condition seems to be the same as when it was dismounted after S181 and it, apart from some specks of dust no bond wires appear to be broken to the naked eye.
BGO signals appear on the scope with a 22-Na source but the response is quite messy.
17:36
MSL type BB18(DS)-1000 24cm x 8cm DSSSD upstream position within snout using 8x upstream (23") ribbon cables
MSL type BB(DS)-1000 positioned downstream of downstream bPlast within snout using 1x downstream (29") ribbon cable. For BB7 'isolated' Right Coupler Kapton PCB replaced by unmodified Right Coupler Kapton PCB.
|
Mon Dec 9 12:28:01 2024, NH, CC, MP, Test of BB7 on AIDA Electronics
|
BB7 connected to aida07, sitting in a cardboard box
Update Tcl/Tml files in ASIC, ASIC4 and sys.tml frm PJCS email to allow setting per-ASIC polarity on FEE64
Restart MIDAS@8015
Power on all FEEs
Note nnrpi is not responding to ssh... MP investigates if it ahs been unplugged
No, but a powercycle worked and fixed it
See error: /MIDAS/FEE_ASIC/aida06/ASICs.txt not found, create the folders
We see some data in the L spectra, and changing the settings from one ASIC seems to work properly
Stopped for CC/MP to determine the which ASICs correspond to p+n and n+n
Follow up: Alpha data overnight or source? |
Thu Dec 5 11:18:46 2024, TD, Offline analysis S100 data files R21_0-R21_99 23x
|
Offline analysis of S100 data files R21_0 - R21_99 (162Eu setting)
first WR ts
First timestamp of R21_0 0x17CA09154AE3E636
Epoch converter says ...
GMT: Saturday, April 27, 2024 4:36:35.223 AM
Your time zone: Saturday, April 27, 2024 5:36:35.223 AM GMT+01:00 DST
last WR ts
First timestamp of R21_100 0x17CA16C1904150CE
GMT: Saturday, April 27, 2024 8:47:08.772 AM
Your time zone: Saturday, April 27, 2024 9:47:08.772 AM GMT+01:00 DST
FEE64 configuration
FEE64 a b c
g h
d e f
a b c d e f g h
DSSSD#1 15 3 12 9 1 5 2 4
DSSSD#2 11 7 16 10 14 13 6 8
n+n Ohmic FEE64s 2, 4, 6, 8
Data analysis assumes
- all LEC ADC data channels with valid ADC offset included (1012 of 1024 channels)
LEC calibration ADC offset only
- no clustering
- no multiplex timestamp correction
- no p+n junction side - n+n Ohmic side correlation time gates
- FEE64 *not* DSSSD strip ordering
- hardware - slow comparator setting p+n junction FEE64s 100keV, n+n Ohmic FEE64s 150keV
- LEC energy difference +/-168keV
- HEC energy difference +/- 1.68GeV
- valid LEC events
DSSSD #1
p+n junction side multiplicity = 1 and n+n Ohmic side multiplicity = 1
DSSSD #2
0 < p+n junction side multiplicity < 8
and
0 < n+n Ohmic side multiplicity < 8
151keV < LEC energy < 1008keV
to select candidate beta events and veto higher energy events e.g. light ions
standalone analysis of AIDA data, no downstream veto detector
- valid HEC events
p+n junction side multiplicity > 0 and n+n Ohmic side multiplicity > 0
(x,y) strips corresponding to maximum energy
p+n junction and n+n Ohmic side HEC
- HEC veto
p+n junction side multiplicity > 0 or n+n Ohmic side multiplicity > 0
- per pixel implant-decay correlations
- end of event
difference in WR timestamp between successive ADC data items > 2500
Attachments 1-4
per DSSSD p+n junction - n+n Ohmic strip time difference for HEC and LEC events (2us/channel) linear and log scale
- observe large (> 32us) time differences (on log scale)
- range of time differences increases with multiplicity ( DSSSD#1 cf. DSSSD#2 LEC events)
- distribution of HEC time differences can probably be understood in terms of most/all channels of ASIC being active during HEC event with low LEC thresholds
- AIDA is a triggerless DAQ producing streams of ADC data items *not* events
at high instantaneous rates when events are constructed they may become aggregated in time i.e. > 32us readout time of all channels of one ASIC
- To investigate impose additional end of event criterion
difference in first and last WR timestamp of event < 33us
Attachments 5-6
per DSSSD p+n junction - n+n Ohmic strip time difference for HEC and LEC events (2us/channel) linear and log scale
- blue original end of event criteria, cyan new end of event criteria
- as expected range of time differences is restricted to +/- 32us
- observe somewhat higher fraction of events with low time differences
DSSSD #1 10363098 of 16104322 (64%) events +/-2us
DSSSD #2 860454912 of 1766618199 (49%) events +/-2us
Attachment 7
per DSSSD p+n junction - n+n Ohmic strip time difference for LEC events - x-axis 2us/channel, y-axis 20keV /channel
Attachment 8 per FEE64 LEC data rate (Hz) 268ms/channel
Attachment 9 per FEE64 LEC data rate (Hz) 268ms/channel: 150keV < energy < 1500keV
Attachment 10 per FEE64 LEC data rate (Hz) 268ms/channel: energy > 1500keV
- observe high instantaneous rate on spill
- rate dominated by low energy (<1500keV) events
- rate of higher energy events dominated by on spill events i.e. light ions as expected
- significant deadtime on spill for n+n Ohmic FEE64s, low deadtime off spill
- deadtime low/zero for p+n junction FEE64s on/off spill
Attachment 11 per FEE64 LEC hit pattern: 150keV < energy < 1500keV
Attachment 12 per FEE64 LEC hit pattern: energy > 1500keV
Attachment 13 per FEE64 HEC data rate (Hz) 268ms/channel
Attachment 14 per FEE64 HEC data rate (Hz) 268ms/channel: 100MeV < energy < 1000MeV
Attachment 15 per FEE64 HEC data rate (Hz) 268ms/channel: energy > 1000MeV
- rate dominated by low energy (>1GeV) events
- all HEC events on spill as expected (note FEE64 #7 has a single hot channel which can be disabled in software)
- significant deadtime on spill for n+n Ohmic FEE64s, low deadtime off spill
- deadtime low/zero for p+n junction FEE64s on/off spill
Attachment 16 per DSSSD p+n junction versus n+n Ohmic LEC energy - x-axis & y-axis 20keV/channel
Attachment 17 per DSSSD p+n junction versus n+n Ohmic HEC energy - x-axis & y-axis 20MeV/channel
Attachment 18 per DSSSD p+n junction versus n+n Ohmic HEC strip hit pattern: all HEC events
Attachment 19 per DSSSD p+n junction versus n+n Ohmic HEC strip hit pattern
DSSSD #1 ions stopped in DSSSD #1 i.e. DSSSD #2 HEC multiplicity = 0
DSSSD #1 shows x-y gate used ( 270 < x < 370, 20 < y < 90 ) to identify 166Tb implants
DSSSD #2 ions stopped in DSSSD #2 *and* in transmission (can establish which ions stop in DSSSD#2 from DSSSD#2 HEC energy versus DSSSD#1 HEC energy - see https://elog.ph.ed.ac.uk/DESPEC/672
Attachment 20
DSSSD#1 HEC energy (20MeV/channel) versus HEC-LEC dt (1s/channel)
DSSSD#1 LEC energy (20keV/channel) versus HEC-LEC dt (1s/channel)
DSSSD#1 HEC strip # versus HEC-LEC dt (1s/channel)
- Observe # events in every third channel is lower
- Probably reflects implant-decay correlation livetime
For example (choosing some numbers for illustrative purposes)
on spill: HEC livetime 75%, LEC livetime 75% (FEE64 deadtime common for HEC and LEC data) => implant decay correlation livetime 56%
off spill: HEC live time 75%, LEC livetime 100% => implant decay correlation livetime 75%
- Observe 'hot' x channels 315, 318, 321, 324 - disabled for further analysis
- Do not observe any 'hot' y channels
Attachments 21 & 22
DSSSD#1 per pixel HEC-LEC time (1s/channel): x,y,z gated to select 166Tb events
Naive (parent-daughter decay only, flat background) fit for data t=0-26s ( t1/2 = 27.1(3)s )
Fit ignores data for t=0, 3, 6, 10, 13, 16, 19, 22s to avoid bias of differences in implant-decay correlation deadtime
Suggestion of structure at c. 30s period? Does this reflect spill stucture? 10x spill cycles (30s), 9s spill off, ... etc
Sum of x,y,z gated HEC events (s2112 - see attachment 19) = 670441
Elapsed time of dataset 4h11m = 15060s
# pixels = 100 x 70 = 7000
=> # x,y,z gated HEC events per pixel = 670441/7000/15060 = 0.0064/s or mean time between x,y,z gated HEC event = 157s (estimate needs to be corrected for HEC deadtime)
Sum of implant decay correlations (s2220 - see attachment 21) t=0-150s = 273508 - flat background estimated as 150 x 500 = 75000 = 198508
=> efficiency c. 30% (presumably low due to implant-decay deadtime, LEC multiplicity, per pixel correlations and no clustering)
Summary
$64,000 question - what is the origin of the high instantaneous rate on spill for DSSSD#1 ? On my to do list.
Attachment 23
LEC multiplicity with/without HEC data in event
per DSSSD LEC p+n junction multiplicity versus n+n Ohmic multiplicity
per DSSSD LEC p+n junction multiplicity versus n+n Ohmic multiplicity z_hec=1 and z_hec=2
With HEC data
DSSSD#1 p+n junction multiplicity ~ 17, n+n Ohmic multiplicity ~28
DSSSD#2 p+n junction multiplicity ~ 40, n+n Ohmic multiplicity ~23
Assume 200Hz HEC events => DSSSD#1 LEC rate = 200 x ( 17 + 28 ) => 9k LEC data items cf. >100k LEC data items (attachments 8 & 13)
i.e. not due to HEC events
|
Wed Dec 4 09:57:24 2024, TD, HISPEC DESPEC meeting presentation - November 2024 -0 Summary of AIDA performance 2024
|
|
Tue Nov 19 11:24:25 2024, TD, Anydesk restarted remotely
|
Anydesk restarted remotely per https://elog.ph.ed.ac.uk/CARME/489 |
Tue Nov 5 15:19:35 2024, TD, S505 offline analysis R5_780 - R5_814 24x
|
Offline analysis of data files S505/R5_780 - R5_814 (corresponding to MBS data files 73-74)
Can find an analysis of alpha background runs at RIBF, RIKEN for comparison at https://elog.ph.ed.ac.uk/AIDA/816
Beam off - background runs without sources
MBS 73 27.6.22 09:33-11:33 CEST https://elog.gsi.de/despec/S505/204
MBS 74 27.6.22 11:34-13:34 CEST https://elog.gsi.de/despec/S505/206
MBS 75 27.6.22 13:35-16:29 CEST https://elog.gsi.de/despec/S505/209
ADC offsets per https://elog.ph.ed.ac.uk/DESPEC/556
FEE64 configuration
FEE64 a b c d
DSSSD#1 3 4 1 2
DSSSD#2 7 8 5 6
p+n junction FE64s odd numbered
Data analysis assumes
- all LEC ADC data channels with valid ADC offset included (474 of 512 channels)
- no clustering
- no p+n junction side - n+n Ohmic side correlation time gates
- valid LEC events
0 < p+n junction side multiplicity < 8
and
0 < n+n Ohmic side multiplicity < 8
Attachments 1-2 - per DSSSD p+n versus n+n multiplicity
Attachment 3 - per DSSSD x versus y
Attachments 4-5 - per DSSSD p+n versus n+n energy (20keV/channel nominal)
all combinations of per DSSSD p+n junction and n+n Ohmic energies
with projection of data within window onto x and y axes
too many events for natural (U decay series) background
off leading diagonal correlations anomalous
transverse width of leading diagonal correlation wider than expected - ADC offsets OK?
Attachment 6 - per FEE64 WR timestamp (32.768us/channel)
FEE64 sync test using pulser data - looks OK
Attachments 7-14 - per FEE64 ADC spectra (5.6keV/channel nominal)
note common x/y scale - pulser peak height proxy for peak width
per FEE64 1.8.L pulser peak widths (ch FWHM)
1 11.43
2 16.58
3 12.52
4 17.54
5 9.10
6 12.22
7 17.03
8 16.09
3 of 4 p+n junction FEE64 good (<70keV), 1 of 4 n+n Ohmic FEE64 good - all others < 100keV
Pulser peak indicates noise/gain/offset stable throughout background runs
Attachments 15-22 - per FEE64 ADC spectra (5.6kleV/channel nominal)
observe broad peak-like structures (c. 2.8MeV) in first channel of most ASICs?
Attachment 23 per DSSSD p+n versus n+n energy (20keV/channel nominal)
valid LEC events
p+n junction side multiplicity = 1 n+n Ohmic side multiplicity = 1
Attachment 24 p+n FEE64 #0 energy versus each of n+n FEE64s (#1, #3, #5 and #7) energies (20keV/channel nominal)
valid LEC events
p+n junction side multiplicity = 1 n+n Ohmic side multiplicity = 1
Attachments 23 & 24 do not unambiguously identify which FEE64s are attached to which DSSSD
|
Wed Aug 14 12:40:11 2024, JB, Repaired DSSSD delivery 14.09.2024 11x
|
Three BB18-1000 triples AIDAs collected on 14.09.2024
Find attached visual of the wafer and bond wire + factory bias tests accompanying the DSSSDs. elog:669/1 elog:669/2 elog:669/3
Visual inspection carried out showed that bond wires have been fixed + fingerprint on one DSSSD removed and wires repaired. elog:669/4 elog:669/5 elog:669/6 elog:669/7 elog:669/8 elog:669/9 elog:669/10 elog:669/11
DSSSD 1 (defect bias issue 80V): 3208-10 / 3208-18 / 3208-20
DSSSD 2 (3208-6 dysfunctional): 3208-6 / 3208-9 / 3208-16
DSSSD 3 (defect fingerprint): 3131-5 / 3131-10 / 3131-12
|
Sun Jul 14 16:52:28 2024, TD, S181 R7_38-44 7x
|
|
Sun Jul 14 14:37:53 2024, TD, S181 R6_375-380 6x
|
|
Mon Jun 17 02:13:21 2024, TD, Monday 17 June 43x
|
03.11 DSSSD bias & leakage current OK - attachment 1
FEE64 temperatures OK - attachment 2
ADC data item stats OK - attachment 3
per FEE64 Rate spectra - attachment 4
05.33 DSSSD bias & leakage current OK - attachment 5
FEE64 temperatures OK - attachment 6
ADC data item stats OK - attachment 7
per FEE64 Rate spectra - attachment 8
per FEE64 1.8.L spectra - attachments 9-11
aida09 pulser walkthrough 60 ch FWHM
per FEE64 1.8.W spectra - 20us FSR - attachments 12-14
Merger, TapeServer etc - attachments 15-16
05.40 analysis data file R9_1042 - attachment 17
Check ASIC load x2
aida02 zero rate
restart Merger
05.55 Data file S181/R10
Pulser walkthrough
All histograms zero'd
BNC PB-5
Amplitudes 10.0-1.0 @ 1.0 step
Attenuation x10
Polarity +
tau_d 1ms
Frequency 25Hz
per p+n FEE64 1.8.L spectra - attachment 18
per FEE64 Rate spectra - attachment 19
06.43 TapeServer no storage mode
09.14 Data file S181/R11
Pulser walkthrough
All histograms zero'd
BNC PB-5
Amplitudes 10.0-1.0 @ 1.0 step
Attenuation x10
Polarity -
tau_d 1ms
Frequency 25Hz
09.52 TapeServer no storage mode
per n+n FEE64 1.8.L spectra - attachment 20
15.18 DEFGAS/FATIMA arrays moved downstream to permit installation of calibration source at centre of each array - increased distance to AIDA/bPlas => lower rate
15.16 DSSSD bias & leakage current OK - attachments 21-22
FEE64 temperatures OK - attachment 23
ADC data item stats OK - attachment 24
aida02 zero rate
per FEE64 Rate spectra - attachment 25
Merger restart
ADC data item stats OK - attachment 26
7x < 20k, max 207k aida16
15.54 bPlast debiased.
16.29 Data file S181/R12
bPlas OFF but remiains cabled/connected/grounded
16.30 DSSSD bias & leakage current OK - attachment 27
FEE64 temperatures OK - attachment 28
ADC data item stats OK - attachment 29
per FEE64 Rate spectra - attachment 30
per FEE64 1.8.W spectra - 20us, 200us, 2ms, 20ms FSR - see attachments 31-39
note - not all FEE64s updated spectra for all settings
ADC data item stats OK - attachment 40
rates change cf. attachment 29
sources installed/moved/removed, people moving around platform?
per FEE64 Rate spectra - attachment 41
16.57 DSSSD bias OFF
FEE64 power OFF
16.34 analysis data file S181/R12_7
max deadtime 2.2% (aida08)
ADC data rate 1.330M, HEC data rate 2.1k
16.52 analysis data file S181/R12_12
max deadtime 1.7% (aida08)
ADC data rate 945k, HEC data rate 1.8k |
Sun Jun 16 03:11:27 2024, TD, Sunday 16 June 26x
|
04.08 DSSSD bias & leakage current OK - attachment 1
FEE64 temperatures OK - attachment 2
ADC data item stats OK - attachment 3
per FEE64 Rate spectra - attachment 4
07.52 DSSSD bias & leakage current OK - attachment 5
FEE64 temperatures OK - attachment 6
ADC data item stats OK - attachment 7
per FEE64 Rate spectra - attachment 8
per p+n FEE64 1.8.L spectra - attachments 9-10
aida09 pulser peak width 59 ch FWHM
15.01 DSSSD bias & leakage current OK - attachment 11
FEE64 temperatures OK - attachment 12
ADC data item stats OK - attachment 13
per p+n FEE64 1.8.L spectra - attachments 14-15
aida09 pulser peak width 59 ch FWHM
aida05 pulser peak width 68 ch FWHM
aida12 pulser peak width 71 ch FWHM
per FEE64 Rate spectra - attachment 16
Merger, TapeSever etc - attachment 17
no data transfer to MBS - MBS down?
15.06 analysis data file R9_747 - attachment 18
max dead time 4.3% aida08, all other FEE64s < 2%
ADC data rate 1.667M, HEC data rate 2.3k
19.12 DSSSD bias & leakage current OK - attachment 19
FEE64 temperatures OK - attachment 20
ADC data item stats OK - attachment 21
per FEE64 Rate spectra - attachment 22
23.34 DSSSD bias & leakage current OK - attachment 23
FEE64 temperatures OK - attachment 24
ADC data item stats OK - attachment 25
per FEE64 Rate spectra - attachment 26 |
Sat Jun 15 00:28:01 2024, TD, Saturday 15 June 45x
|
Beam lost due to septum failure c. 18.15 Friday 14 June
Cooling of recirculating water failed c. 19.00 - water temperature increased from the normal 17-18deg C to 25-26 deg C
FEE64 temperatures increased and FEE64s powered off and on-call engineer requested.
Cooling of recirculating water restored and AIDA restarted
NH started background run 23.37 - data file S181/R9
01.21 DSSSD bias & leakage current OK - attachment 1
FEE64 temperatures OK - attachment 2
ADC data item stats OK - attachment 3
8x < 20k, max 238k aida08
per FEE64 Rate spectra - attachment 4
2x Check ASIC load
Merger, TapeServer etc - attachments 5-6
analysis data file R9_27 - attachment 7
max deadtime 1.9%, all other FEE64s < 1%
ADC data rate 951k, HEC data rate 1.6kHz
04.28 DSSSD bias & leakage current OK - attachment 8
FEE64 temperatures OK - attachment 9
ADC data item stats OK - attachment 10
8x < 20k, max 261k aida08
per FEE64 Rate spectra - attachment 11
08.30 DSSSD bias & leakage current OK - attachments 12-13
FEE64 temperatures OK - attachment 14
ADC data item stats OK - attachment 15
all rates increased - gamma source in use?
per FEE64 Rate spectra - attachments 16-17
note aida08 asic #3 hot HEC channel
per FEE64 1.8.L spectra - attachments 18-20
aida09 pulser peak width 59 ch FWHM
decay alphas aida05, aida12, aida02, aida04
per FEE64 1.8.W spectra - 20us FSR - attachments 21-23
Merger, TapeServer etc - attachments 25-25
analysis data file R9_152 - attachment 26
max deadtime 2.4%, all other FEE64s < 1%
ADC data rate 1.215M, HEC data rate 2.1kHz
12.22 DSSSD bias & leakage current OK - attachment 27
FEE64 temperatures OK - attachment 28
ADC data item stats OK - attachment 29
per FEE64 Rate spectra - attachment 30
15.27 DSSSD bias & leakage current OK - attachment 31
FEE64 temperatures OK - attachment 32
ADC data item stats OK - attachment 33
all rates increased - 370kBq 152Eu in use
per FEE64 Rate spectra - attachments 34
per FEE64 1.8.L spectra - attachments 35-36
aida09 pulser peak width 59 ch FWHM
decay alphas aida05, aida12, aida02, aida04
analysis data file R9_275 - attachment 37
max deadtime 4.4%, all other FEE64s < 2%
ADC data rate 1.520M, HEC data rate 2.3kHz
18.52 DSSSD bias & leakage current OK - attachment 38
FEE64 temperatures OK - attachment 39
ADC data item stats OK - attachment 40
per FEE64 Rate spectra - attachment 41
22.50 DSSSD bias & leakage current OK - attachment 42
FEE64 temperatures OK - attachment 43
ADC data item stats OK - attachment 44
per FEE64 Rate spectra - attachment 45 |
Fri Jun 14 15:00:47 2024, Norah , Muneerah, 16:00-20:00 Friday 14 June 13x
|
Everything appears to be running smoothly. Attachments 1-4 were taken around 16:00
DSSSD Bias & Leakage Current - Check: OK (see attachment 1)
FEE64 Temperature - Check: OK (see attachment 2)
ADC Data Item Stats and Spectrum Browser - Check: OK (see attachments 3 and 4)
Another set of screenshots was taken around 18:00, and everything looks good.
DSSSD bias & leakage current ok - attachment # 5-6
FEE64 temperatures ok - attachment # 7
ADC data item stats - attachment # 8
Merger ok - Attachement #9
Tape service - attachement # 10
ucesb - attachment #11
At 19:17, AIDA03 temperatures gave no response (see attachment 12), but after two minutes, they returned to normal.
I noted that the bplse in ucesb has no data (see attachment 13). It may need to change the disc thresholds again
it does not affect AIDA
At 19:30 most aida temperature are high and I spoke to people on site via zoom one of them she is working with AIDA so she was able to help. She called Nick and it is found that there is a problem in the cooling system so she stopped AIDA
At 22 the cooling system fixed and Nic and the other person on site restarted AIDA
|
Fri Jun 14 06:22:39 2024, TD, 08:00-16:00 Friday 14 June 23x
|
07.23 All histograms & stats zero'd
DAQ continues data file S181/R8_35
analysis data file R8_35 - attachment 1
max deadtime 4.8% (aida08), all other FEE64s < 3%
LEC data rate 1.677M HEC data rate 3.7k
08.46 DSSSD bias & leakage current OK - attachments 2-3
FEE64 temps OK - attachment 4
ADC data item stats - attachment 5
per FEE64 Rate spectra - attachment 6
per p+n FEE64 1.8.L spectra - attachment 8
aida09 pulser peak width 65 ch FWHM
per FEE64 1.8.H spectra - attachment 7 & 10
per FEE64 1.8.W spectra - 20us FSR - attachment 9 & 11
Merger, TapeServer etc - attachments 12-13
data file S181/R8_72
10.51 DSSSD bias & leakage current OK - attachments 14
FEE64 temps OK - attachment 15
ADC data item stats - attachment 16
per FEE64 Rate spectra - attachment 17
10.54 analysis data file R8_122 - attachment 18
max deadtime 7.8% (aida08), all other FEE64s < 3%
ADC data rate 1.88M, HEC data rate 4.7k
14.24 DSSSD bias & leakage current OK - attachments 18
FEE64 temps OK - attachment 19
ADC data item stats - attachment 20
per FEE64 Rate spectra - attachment 21
aida02 no data
14.28 merger restarted and DAQ running OK
14.38 analysis data file R8_213 - attachment 22
max deadtime 6.6% (aida08), all other FEE64s < 3.5%
ADC data rate 1.935M, HEC data rate 3.6k |
Thu Jun 13 23:04:01 2024, Dan Judson, TD, 0.00-8.00 14/6/24 40x
|
0.00 checks - all looks ok
Voltage/currents - attachment 1
Rates attachment 2 + 3
Temps - attachment 4
Merger - attachment 5
ucesb - attachment 6
Grafana - attachment 7
1.30 -aida1 in merger Links, aida02 in rate spectra, has stopped - resetting DAQ.
Datafile R7_961 being written when stopped
1.40 - seems ok
1.50 checks - seems ok
Voltage/currents - attachment 8
Rates attachment 9 + 10
Temps - attachment 11
Merger - attachment 12
ucesb - attachment 13
Grafana - attachment 14
3.30 -aida1 in merger Links, aida02 in rate spectra, has stopped again - resetting DAQ.
Datafile R7_1010 being written when stopped
3.38 - seems ok
3.40 checks - all looks ok
Voltage/currents - attachment 15
Rates attachment 16 + 17
Temps - attachment 18
Merger - attachment 19
ucesb - attachment 20
Grafana - attachment 21
Zeroed spectra - attachments 22-28
5.39 -aida1 in merger Links, aida02 in rate spectra, has stopped again - resetting DAQ.
Datafile R7_1062 being written when stopped
AIDA out of the main data
Could not get it to restart folowing the notes on the elog - SOAP errors as shown in attachment 29. called Tom who''s logged in
6.10 - seems to be working again. Not sure what happened. Tom thinks multiple restarts fixed it
Started data file R8
AIDA back in the main data
FRS shifters report a problem with the AIDA white rabbit time stamp - attachment 30
The following are timestamp values from each of the FEEs taken in sequence
If time does not increase in a reasonable manner run the system wide checks
aida01 : White Rabbit=> 17D8C4ED 8F46CF7B , WR/10=> 2627A17C18714BF, Readout Time => 2627A17C2A74000
aida02 : White Rabbit=> 17D8C4ED A0A1B623 , WR/10=> 2627A17C3435F03, Readout Time => 2627A17C4410000
aida03 : White Rabbit=> 17D8C4ED AFDACE68 , WR/10=> 2627A17C4C914A4, Readout Time => 2627A17C5D3C000
aida04 : White Rabbit=> 17D8C4ED C0516FC2 , WR/10=> 2627A17C66E8B2D, Readout Time => 2627A17C7AC4000
aida05 : White Rabbit=> 17D8C4ED D3B93DAB , WR/10=> 2627A17C85F52F7, Readout Time => 2627A17C9AEC000
aida06 : White Rabbit=> 17D8C4ED E8C9B243 , WR/10=> 2627A17CA7A91D3, Readout Time => 2627A17CB704000
aida07 : White Rabbit=> 17D8C4ED FFE475FD , WR/10=> 2627A17CCCA0BCC, Readout Time => 2627A17CDD94000
aida08 : White Rabbit=> 17D8C4EE 103232A6 , WR/10=> 2627A17CE6B6B77, Readout Time => 2627A17CF7BC000
aida09 : White Rabbit=> 17D8C4EE 20A1F612 , WR/10=> 2627A17D0103235, Readout Time => 2627A17D112C000
aida10 : White Rabbit=> 17D8C4EE 307D8FE7 , WR/10=> 2627A17D1A627FD, Readout Time => 2627A17D2958000
aida11 : White Rabbit=> 17D8C4EE 4402829D , WR/10=> 2627A17D399D9DC, Readout Time => 2627A17D4570000
aida12 : White Rabbit=> 17D8C4EE 6288B04D , WR/10=> 2627A17D6A744D4, Readout Time => 2627A17D7D94000
aida13 : White Rabbit=> 17D8C4EE 751E1306 , WR/10=> 2627A17D88301E7, Readout Time => 2627A17D9C34000
aida14 : White Rabbit=> 17D8C4EE 894E8806 , WR/10=> 2627A17DA87DA67, Readout Time => 2627A17DBFF4000
aida15 : White Rabbit=> 17D8C4EE 9EEEB997 , WR/10=> 2627A17DCB178F5, Readout Time => 2627A17DDC18000
aida16 : White Rabbit=> 17D8C4EE AF0BC923 , WR/10=> 2627A17DE4DFA83, Readout Time => 2627A17DF5C8000
Tom thinks everything appears ok with AIDA
Have converted WR timestamp to data/time - see attachment 40
At the ~second timescale AIDA WR timestamp looks OK, i.e. no gross errors. Will need to check correlations with other detector sub-systems to confirm WR on the ~us time timescale..
However, ~10us offset reported by DESPEC online crew is correct (AIDA timestamps data at a later point in the signal processing cycle than other detector sub-systems). Have suggested online crew contact Nic or Calum to check what the AIDA WR timediff spectra should look like.
7.00 checks - all looks ok
Voltage/currents - attachment 31
Rates attachment 32 + 33
Temps - attachment 36
Merger - attachment 37
ucesb - attachment 39
Grafana - attachment 38
Handed over to Tom
|
Thu Jun 13 21:06:44 2024, Robert Page, 16:00 to 00:00 shift on 13/6/2024 26x
|
Elog went down around the end of the last shift, so final screenshots were not uploaded at that point.
However, AIDA appears to have contuned running smoothly in the interim (at least until 22:00).
Attachments 1 - 6 were taken around 16:00, 7 - 13 at around 18:00 , 14 - 18 at around 21:00.
Another set of sceenshots taken at around 23:00 - attachments 19 - 26.
|
Thu Jun 13 06:49:54 2024, Magda Satrazani, 08:00 - 16:00 Thursday 13 June 2024 40x
|
Took over the shift from Marc.
08:00 FULL CHECK
DSSSD bias & leakage current ok - attachment # 1-2
FEE64 temperatures ok - attachment # 3
ADC data item stats - attachment # 4
Merger ok - Attachement # 5
Tape service - attachement # 6
10:00 FULL CHECK
DSSSD bias & leakage current ok - attachment # 7-8
FEE64 temperatures ok - attachment # 9
ADC data item stats - attachment # 10
Merger ok - Attachement # 11
Tape service - attachement # 12
12:00 FULL CHECK
DSSSD bias & leakage current ok - attachment # 13-14
FEE64 temperatures ok - attachment # 15
ADC data item stats - attachment # 16
Merger ok - Attachement # 17
Tape service - attachement # 18
12.34 analysis data file S181/R7_610 - attachment 19
max deadtime 8.7% (aida08), all other FEE64s < 2.7%
LEC data rate 2.000M, HEC data rate 4.7k
12.52 per p+n FEE64 1.8.L spectra - attachment 20
aida09 pulser peak width 63 ch FWHM
per FEE64 1.8.H spectra - attachments 21-22
per FEE64 1.8.W spectra - 20us FSR - attachments 23-24
13.00 all histograms zero'd
14:00 FULL CHECK
DSSSD bias & leakage current ok - attachment # 19-20
FEE64 temperatures ok - attachment # 21
ADC data item stats - attachment # 22
Merger ok - Attachement # 23
Tape service - attachement # 24
13.01 Beam OFF for S4 access
ADC data item stats - attachment 25
per FEE64 Rate spectra - attachment 26
14.12 analysis data file R7_654 - attachment 27
max deadtime 3.9% (aida08), all other FEE64s < 3%
LEC data rate 1.817M, HEC data rate 2.5k
14.00 data file R7_649 - attachment 28
max deadtime 4.5% (aida08), all other FEE64s < 3%
LEC data rate 1.756M, HEC data rate 2.5k
16:00 FULL CHECK
DSSSD bias & leakage current ok - attachment # 25-26
FEE64 temperatures ok - attachment # 27
ADC data item stats - attachment # 28
Merger ok - Attachement # 29
Tape service - attachement # 30 |
Wed Jun 12 23:00:16 2024, Marc, 00:00-08:00 Thursday 13 June 23x
|
Starting a new night shift - All good
1am full check:
DSSSD bias & leakage current ok - attachment # 1-2
FEE64 temperatures ok - attachment # 3
ADC data item stats - attachment # 4
Merger ok - Attachement # 5
Tape service - attachement # 6
2:30am: AIDA02 FEE stopped working. DESPEC main DAQ was alerted and remobe AIDA from timestitcher while I was restarting AIDA DAQ.
2:45am: AIDA02 FEE is back and AIDA has been added to the timestitcher again in the main DAQ.
The AIDA runs at the time of the crash to look at are R7_348, 349.
All spectra were zeroed at this time and checked.
4:00 am full check:
DSSSD bias & leakage current ok - attachment # 7-8
FEE64 temperatures ok - attachment # 9
ADC data item stats - attachment # 10
Merger ok - attachement # 11
Tape service - attachement # 12
Spectra - attachement # 13
ucesb - attachement # 14
6:00am full check:
DSSSD bias & leakage current ok - attachment # 15-16
FEE64 temperatures ok - attachment # 17
ADC data item stats - attachment # 18
Merger ok - attachement # 19
Tape service - attachement # 20
Spectra - attachement # 21
ucesb - attachement # 22
6:30 the nearline/offline analysis team reported possible noise issue in DSSD#2 and asked if the threshold were changed at any point. I can't find any mention of changes in the elog.
I've not done a setup of the all DAQ when AIDA02 fee crashes earlier this morning. I just did a setup of the merger. Thus I don't think the threshold have changed when I stopeed and restarted the DAQ.
I can see that currently slow comparator threshod are set to 0xa for most of the DSSD2 fees except for AIDA06 which is set to 0xf. I'm not sure this is the thershold we would need to increase. Something to check during the next shift.
In the UCESB window I can see that the number of decay in DSSD2 is double the number of decay in DSSD1 while the number of implant is similar. Looking at other shift it looks that this was alsready the case in prvious shifts.
07.06 analysis of data file S181/R7_457 - attachment 23
max deadtime 5.8% aida08, all other FEE64s < 1.4%
LEC data rate 1.788M, HEC data rate 4.3k
|
Wed Jun 12 15:41:07 2024, Betool Alayed, 16:00-00:00 Wed 12 Jun 24 27x
|
5pm full checks:
DSSSD bias & leakage current ok - attachment # 1-2
FEE64 temperatures ok - attachment # 3
ADC data item stats - attachment # 4
Merger ok - Attachement # 5
Tape service - attachement # 6
7pm full checks:
DSSSD bias & leakage current ok - attachment # 7-8
FEE64 temperatures ok - attachment # 9
ADC data item stats - attachment # 10
Merger ok - Attachement # 11
Tape service - attachement # 12
ucesb - attachment # 13
9pm full checks:
DSSSD bias & leakage current ok - attachment# 14-15
FEE64 temperatures ok - attachment # 16
ADC data item stats - attachment # 17
Merger ok - Attachement # 18
Tape service - attachement # 19
ucesb - attachment # 20
11pm full checks:
DSSSD bias & leakage current ok - attachment# 21-22
FEE64 temperatures ok - attachment # 23
ADC data item stats - attachment # 24
Merger ok - Attachement # 25
Tape service - attachement # 26
ucesb - attachment # 27 |
Wed Jun 12 11:06:55 2024, PP, Mid-shift checks, 12:00 8x
|
All seems normal.
Screenshots attached. |
|