  65   Fri Aug 16 15:09:05 2019 NHAida Waves (All FEEs)Included waves for all 12 FEEs

FEES 1, 5, 9 all show this HF noise
on it, and are all located at the top of
 12x 
  64   Thu Aug 15 14:09:39 2019 NHAIDA WaveformsNoise seemed less strong today, unsure
of cause. I had re-taped snout as tape was
not sticking well.
 10x 
  63   Wed Aug 14 14:40:18 2019 NHHigh Frequency Noise in AIDAFigures of intense high frequency noise
pickup in AIDA.

Will investigate source in S4.
 HF.pngHFZ.png 
  61   Tue Aug 13 13:55:45 2019 NHNew Merger/MBS Test RunsAIDA Finally working again so can test
VP's added WAVE histogramming mode.

Fig 1. Waveforms for FEE12 ASIC
 6x 
  60   Tue Aug 13 09:17:02 2019 NHAIDA 12/08/2019The other 6 FEEs were powered on one at
a time to check no issues. None found.
All 12 FEEs plugged back into the
 AIDATemps-1208191457.png 
  59   Fri Aug 9 12:33:14 2019 NHAida Test 9/8/19 - FEES 1-6The water issue has been resolved - merely
incorrect temperature guages (replaced).
Water is 20 C as required.
  
  58   Mon Aug 5 12:57:48 2019 NHS4 ConditionsTemp: 26 C
Humidity: 50%
Dew Point: ~15 C
  
  57   Wed Jul 10 10:46:05 2019 NHAIDA/Water 10.07.2019S4 Conditions:

25.2 C, 35.3% RH, Td = 8.8 C

Water temperature: 20 C
  
  56   Mon Jul 8 16:02:52 2019 NH, PJCSAida 28/6/2019 - WATER OFFIt occurs to to wonder if there is now
water inside the modules. If the metal of
the cooling plates is damp on the outside
  
  55   Mon Jul 1 09:42:16 2019 NHAIDA - 01/07/2019Water is still off.

Checked dewpoint sensor still flashing
red light (indicating an error). Pipes warmed.
  
  54   Fri Jun 28 15:05:49 2019 NHAida 28/6/2019 - WATER OFFThe humidity sensor hasn't reported
humdities under 90% yet. In order to try
and help dry the pipes and turn AIDA the
  
  53   Tue Jun 4 09:19:10 2019 NHAIDA InterlockCurrently AIDA's interlock is off (no
lights) and hence AIDA cannot be powered
on. Under investigation - power supply has
  
  52   Mon May 20 13:51:45 2019 NHHowTo Verify WR TimesThe latest version of MIDAS has a new page
to check the WR timestamp of each FEE.

In AIDA Hardware control click:
 HowTo-WR1.pngHowTo-WR2.pngHowTo-WR3.png 
  51   Mon May 20 13:33:06 2019 NHNew Merger/MBS Test RunsNew merger has been worked on by VP which
fixes the timewarp issues and MBS integration.

Currently no WAVE capture is supported,
 8x 
  50   Fri Apr 12 15:15:33 2019 NHReport - FEE Kernel Panics (Update on 48)Update on issue #48 - the "confusing
state" is that the FEE has restarted
and hence is undefined again.
 aida01_log.txt 
  49   Thu Apr 11 15:28:10 2019 CA, OH, CB, TDOffline analysis of 290319 files R13 and R16X projection of high energy channel ADC hits
from R13 (attachment 1)
 6x 
  48   Wed Apr 10 14:53:50 2019 NHReport - FEE stops sending datait seems a FEE somtimes enters a confusing
state and stops sending data
the current merger requires all FEEs
 6x 
  47   Wed Apr 10 12:33:58 2019 NH VPAIDA Firmware UpdateAll AIDA modules have been updated to a
new firmware 0xd330701 which should fix the
merger issues.
 20191004-FeeTemps.png20191004-GSIWR.png 
  46   Thu Apr 4 14:07:45 2019 NHPulser ConfigurationPulser settings during April run:

Rate 2Hz
Delay 250 NS
  
  45   Wed Apr 3 15:24:06 2019 CA, TDreport - low inquire save stateOn Save/Restore Module Settings Tab - "inquire
save state" returns an error. See attachment
1.
 030419_1614_savestate.png 