ID |
Date |
Author |
Subject |
676
|
Sat Dec 14 11:52:42 2024 |
TD | Saturday 14 December |
11:00 Visual inspection FEE64 adaptor PCBs & cabling
aida02 ground from/to other FEE64s disconnected - re-connected
aida01, aida09, aida12 - ground cabling screws to Lemo 00.250 housings loose - tightemed
12:50 Cooling water pressure & temperature OK - attachments 1 2
12:59 relay #1 power ON
13:01 relay #2 power ON
13:07 aida06 starts - panic during startup, automatic restart following 3 minute timeout
DAQ reset, setup
Check ASIC Control - browser tab timeout
AIDA MIDAS HTTPD server console log - attachment 3
Appears to have restored ASIC settings 2024Dec13-17.02.45 saved yesterday
aida10 ASICs #1 & #2 positive input, ASICs #3 & #4 negative input
slow comparator 0xa (all p+n junction FEE64s and aida10), 0xf (n+n Ohmic FEE64s)
13:28 tar ASIC settings - attachment 4
[npg@aidas-gsi]$ cd /MIDAS/FEE_ASIC
[npg@aidas-gsi FEE_ASIC]$ tar cvf /tmp/FEE_ASIC.tar .
System wide checks
Sync ASIC clocks - attachment 5
Clock, ADC calibration, WR decoder, FPGA timestamp, PLL checks - attachments 6-11
all OK *except* aida02 WR decoder error
WR timestamps OK - attachment 12
FEE64 temps OK - attachment 13
13:45 Detector bias ON - attachment 14
BNC PB-5 pulser settings - attachment 15
Pulser connected to all p+n junction FEE64s *except* aida10
ADC, DISC, PAUSE and MBS correlation scaler stats - attachments 16-19
aida02 rate significantly lower than yesterday - https://elog.gsi.de/despec/Implantation+Stack/8
high rates observed for aida08, aida11 and aida14 - which are not connected to a DSSSD!
per FEE64 Rate spectra - attachment 20
per p+n junction FEE64 1.8.L spectra - attachment 21
aida09 pulser peak width 56 ch FWHM = 39 keV FWHM
consistent electronic noise for all p+n junction FEE64s (cabling+DSSSD)
electronic noise of p+n junction FEE64s (cabling *only*) higher and more variable cf. https://elog.gsi.de/despec/Implantation+Stack/8 attachment 5
per p+n junction FEE64 1.8.W spectra 20us FSR - attachments 22-23
per n+n Ohmic FEE64 1.8.W spectra 20us FSR - attachment 24
WR timestamps OK - attachment 25
14:43 DAQ STOP
Data transfer enabled
Select Tape Server -> Next Run
DAQ GO data file R3
Merger, Tape Server - attachments 26-27
data transfer rate c. 900k data items/s cf c. 300k data items/s yesterday https://elog.gsi.de/despec/Implantation+Stack/8
16:45 DAQ STOP
Data transfer disabled
Detector bias OFF
FEE64 power OFF |
645
|
Sun Jun 9 16:38:30 2024 |
Norah , Muneerah, JB | 16:0-00:00 9 June 2024 | AIDA02 and AIDA06 gave zero attachment 1. After connecting with Tom to fix it, now it works.
17:00
DSSSD bias & leakage current - attachment 2
FEE64 temperatures OK - attachment 3
Statistics attachment 4
17:39
Most of AIDA0 gave zero. I followed the instructions that Tom gave me to fix it, and now they work.
23:19 Flange removed. Starting to take beam.
|
458
|
Sun May 15 23:00:08 2022 |
& TD | Monday 16 May 00:00-08:00 |
00.01 DAQ continues
file S450/R4_1319
ADC control register 0xff
all disc outputs disabled
ASIC settings 2021Apr29-13-16-00
slow comparator aida02 & aida04 0x16, aida06 & aida08 0x23, aida12 0xd, all others 0xc
BNC PB-5 settings
amplitude 1.000V
attenuation x1
tau_d 1ms
frequency 2Hz
polarity +
Disk space OK - /media/SecondDrive
[npg@aidas-gsi S450]$ df -h
Filesystem Size Used Avail Use% Mounted on
devtmpfs 7.8G 0 7.8G 0% /dev
tmpfs 7.8G 389M 7.4G 5% /dev/shm
tmpfs 7.8G 19M 7.7G 1% /run
tmpfs 7.8G 0 7.8G 0% /sys/fs/cgroup
/dev/mapper/centos-root 50G 16G 35G 31% /
/dev/sda2 1014M 226M 789M 23% /boot
/dev/sda1 200M 12M 189M 6% /boot/efi
/dev/sde1 7.2T 4.1T 2.8T 61% /media/SecondDrive
/dev/mapper/centos-home 407G 91G 316G 23% /home
tmpfs 1.6G 52K 1.6G 1% /run/user/1000
/dev/sdd1 7.2T 6.5T 310G 96% /run/media/npg/ThirdDrive
00.03 all histograms zero'd
system wide checks counter baseline
00.08 check ASIC control - all FEE64s all ASICs
Attachments 1 & 2 - DSSSD bias & Leakage current - OK
grafana DSSSD bias, leakage current & temp - OK
Attachment 3 - FEE64 temps OK
Attachments 4-9 - adc, pause, resume & correlation scaler data items, push, flush stats
Attachments 10-16 - TapeServer, NewMerger, NewMerger stats
Attachment 17 - ucesb
04.08
Attachment 18 - DSSSD bias & Leakage current - OK
Attachment 19 - FEE64 temps OK
Attachment 20 - adc data item stats
Attachments 21 - ucesb
system wide checks - all OK *except* aida09 clock fail status 6
06.36
Attachment 22 - DSSSD bias & Leakage current - OK
Attachment 23 - FEE64 temps OK
Attachment 24 - adc data item stats
Attachments 25 - ucesb
system wide checks - all OK *except* aida09 clock fail status 6 |
|