AIDA GELINA BRIKEN nToF CRIB ISOLDE CIRCE nTOFCapture DESPEC DTAS EDI_PSA 179Ta CARME StellarModelling DCF K40
  DESPEC, Page 4 of 36  ELOG logo
Entry  Wed May 19 18:50:01 2021, LS, Wednesday 19th May 20:00 to 24:00 21x
20.00
Stats okay (Fig 1)
Analysed R14_762 deadtimes are okay (Fig 2)
Entry  Wed May 19 06:59:33 2021, OH, Wednesday 19th May 08:00 - 16:00 9x
08:00 OH Takes over for CA
      Thresholds 0xc for p+n and 0x19 for n+n ASICs 1-3 and 0x64 for n+nASIC4
Entry  Wed May 19 14:54:36 2021, CB, Wednesday 19 May 16:00 - 20:00 11x
15:55 Took over from OH

16:00
Entry  Wed May 18 15:29:50 2022, TD, Wednesday 18 May 8x
      file no storage mode, no MBS data transfer

      ADC control register 0x0 
Entry  Wed Dec 18 17:39:27 2019, TD, Wednesday 18 December 23x

18.40 Detector biases & leakage currents OK - attachment 1
      FEE64 temperatures OK - attachment 2
Entry  Wed Mar 17 10:31:09 2021, TD, Wednesday 17 March 11x
11.28 DAQ continues OK - no storage

      ASIC settings 2019Dec19-16.19.51
Entry  Tue Apr 16 23:37:58 2024, TD, Wednesday 17 April 44x

00.40 Observed zero data fron aida03, aida04, aida09
      WR timestamps offset from other FEE64s
Entry  Wed Jun 16 18:05:03 2021, TD, Wednesday 16 June 7x

19.04 All system wide checks OK *except*
Entry  Wed May 11 19:08:41 2022, CB, PP, Wednesday 11th May - evening shift 16.png22.png13.png14.png
20:19 no beam yet.

Statistic check (screenshot attached).

Temperatures OK (screenshot attached).
Entry  Wed May 11 07:04:13 2022, OH, NH, TD, Wednesday 11th May 9x
08:00 Overnight at some point it appears that aida14 has stopped producing any ASIC data
      It is currently sending data pause requests and data push requests but nothing else - attachment 1
      telnet into aida14 and run top to see what the current state of the FEE is. AIDAExec currently using 0% CPU usage.
Entry  Tue Mar 10 23:17:31 2020, DK, LS, PW, TD, CA, Wednesday 11 March 2020 43x
00:15 Shift change near AIDA R6_475

  We continue with spill cycles around 0.2 Hz or one beginning once each ~5 seconds.
Entry  Wed Mar 11 17:21:33 2020, CA, TD, OH, Wednesday 11 March 16.00-00.00 23x
18.25 DAQ currently running No Storage mode

ASIC settings 2019Oct31-13.24.23
Entry  Wed Mar 10 07:16:02 2021, OH, LS, Wednesday 10th March 08:00-24:00 39x
08:16 System wide checks all ok
      Statistics - attachment 1
      Temperatures - attachment 2
Entry  Wed Apr 10 08:37:57 2024, JB, CC, TD, Wednesday 10 April 47x

09.38 CC completed install of Bplast driver PCBs yesterday evening.
      All flat ribbon cables connected - all drain wires grounded
Entry  Wed May 1 08:59:15 2024, TD, Wednesday 1 May Capture.PNG
04.55 FEE64 temperatures OK
      ADC data item stats OK *except* aida04 no data
      TapeServer no storage mode
Entry  Wed Jun 1 07:35:13 2022, TD, Wednesday 1 June 15x

Attachment 1 - DSSSD bias & leakage current - OK
Entry  Wed Mar 20 12:22:27 2024, NH, Wed Mar 20 9x
Turn on AIDA for Dry Run demonstrations and so on

All system wide checks, temp, bias OK

Noise situation is dreadful (but has not been optimised). Deterioriation since first mounted, suspect cabling issues with bPlast and BB7.
Entry  Tue Apr 20 23:08:22 2021, CA, Wed Apr 21 00:00-08:00 23x
00:00 - CA takes over

00:15 - stats/database/ucesb ok - attachment 1
Entry  Wed May 25 14:28:26 2022, NH, Wed 25 May aida_single_alu.jpg
The n+n FEEs were all removed to adjust the platform back to the single configuration

After some confusion we found the orientation of the frame to position the single detectors and moved back
Entry  Wed Apr 21 11:27:55 2021, OH, Wed 21st April 12:00-16:00 6x
12:30 MBS has crashed
      This time AIDA is still running fine.
ELOG V3.1.4-unknown