AIDA GELINA BRIKEN nToF CRIB ISOLDE CIRCE nTOFCapture DESPEC DTAS EDI_PSA 179Ta CARME StellarModelling DCF K40
  DESPEC, Page 6 of 35  ELOG logo
ID Date Author Subject
  601   Wed Apr 24 09:04:55 2024 TD08:00-16:00 Wednesday 24 April
09.56 DAQ continues data file S100_alpha/R18_122
      FRS setting 162Eu

      DSSSD bias & leakage current OK - attachments 1-2

      FEE64 temperatures OK  - attachment 3
       aida02 ASIC temp - known fault

      All system wide checks OK *except* as shown - attachments 4-6

      WR timestamps OK - attachment 7

      ASIC settings aida01 & aida02 - attachments 8-9

      ADC, DISC, PAUSE, RESUME, WR & Correlation Scaler data time stats - attachments 10-15

      per FEE64 Rate spectra - attachments 16-18
       unusually can observe HEC Rates esp. in DSSSD #1 

      per p+n FEE64 1.8.L spectra - attachments 19-21
       aida09 pulser peak width 61 ch FWHM 
       common x/y scale => peak height proxy for width => all p+n FEE64s similar pulser peak widths

      per n+n FEE64 1.8.L spectra - attachment 22

      per FEE64 1.8.H spectra - attachments 23-24

      per FEE64 1.8.W spectra - 20us FSR - attachments 25-26
       only aida08 & aida09 currently updating - ASIC check load and/or DAQ STOP/GO required?

      Merger, TapeServer etc - attachments 27-28

      ucesb - attachments 29-30
       spill cycle reported as 3.5s, extraction 1.5s

10.36 All histograms & stats zero'd

      BNC PB-5 settings - attachment 31

      DSSSD bias & leakage current - Grafana Alerting - attachment 32
       can observe bias & leakage current in greater detail
       DSSSD #1 bias stable, c. 10s nA micro structure observed in leakage current
       DSSSD #2 bias c. 10mV variations, micro structure observed in leakage current but smaller variations cf. DSSSD#1

10.46 analysis of data file R18_131
      max dead time 12% aida04, 6% aida02, all others 0.1% or less
      no timewarps
      HEC data item rates 1.6kHz - correcting for spill cycle => 2.3kHz in spill (online analysis shows no HEC events observed outside spill)
      assume typical HEC event multiplicity = 3 => implant rate ~ 760Hz in spill or 2700 implants per spill
      SC42 L/R currently report c. 2kHz





   

      
  600   Tue Apr 23 22:37:46 2024 TD00:00-08:00 Wednesday 24 April
23.34 23.4.24

      DSSSD bias & leakage current OK - attachment 2

      FEE64 temperatures OK - attachment 2

      ADC data item stats - attachment 3

      per FEE64 rate spectra - attachments 4

      Merger etc - attachment 5

03.41 fDSSSD bias & leakage current OK - attachment 6

      FEE64 temperatures OK - attachment 7

      ADC data item stats - attachment 8

      per FEE64 rate spectra (off and on spill) - attachments 9-10

      Merger etc - attachment 11

03.54 analysis data file R18_67 - attachment 12
      max deadtime 11% (aida04)
      HEC data item rate c. 1.5kHz
      no timewarps

0.35 fDSSSD bias & leakage current OK - attachment 13

      FEE64 temperatures OK - attachment 14

      ADC data item stats - attachment 15

      per FEE64 rate spectra - attachments 16

      Merger etc - attachment 17

03.54 analysis data file R18_96 - attachment 18
      max deadtime 12% (aida04)
      HEC data item rate c. 1.6kHz
      no timewarps
  599   Tue Apr 23 19:23:42 2024 PP20:30 checks

All looks good.

Screenshots attached.

  598   Tue Apr 23 15:11:50 2024 TDTuesday 23 April 16:00-00:00
16.12 Analysis of data file R17_101 - attachment 1

      max deadtime c. 2% (aida02)
      no timewarps observed


16.15 DSSSD bias & leakage current OK - attachment 2

      FEE64 temperatures OK - attachment 7

      ADC data item stats - attachment 3

      per FEE64 rate spectra (on and off spill) - attachments 4-5

      Merger etc - attachment 6


16.53 aida04 stops producing data

      DAQ STOP fails for aida04
      Restart Merger to enable aida04 to complete DAQ STOP

      aida04 rebooted and DAQ restarted OK

      data file S100_alpha/R18
  597   Tue Apr 23 13:15:43 2024 PP14:00 checks

All looks good.

Screenshots attached

  595   Tue Apr 23 13:08:28 2024 TDSummary of AIDA FEE64 stops
We have observed a number of AIDA FEE64s stop producing data.

Summary of dates & (approximate) times

aida03	21.4.24	08.20
aida04	22.4.24	09.14             https://elog.ph.ed.ac.uk/DESPEC/586
aida01	22.4.24	16.00
aida02	22.4.24	22.00
aida02	23.4.24	10.21
aida02	23.4.24 11.25
aida04  23.4.24 16.53?            attachment 2
aida04/aida05 25.4.24 22.28       no messages in aida04 system console
aida04  27.8.24 c.23:00
aida04  28.8.24 c. 12-13:00       power cycle
aida04  28.8.24 c. 16.00          reset per Elog 618
aida02  28.4.24 c. 22.20          reset per Elog 618, no error messages in system console
aida04  29.4.24 01.15
aida04  30.4.24 <03.30
aida02  30.4.24 <10.00


aida02 2.5.24 c. 19.30 reset


In future, it would be useful to copy system console reports to the Elog to check whether we can identify common error messages etc.


ssh pi@nnrpi1
<password>

cd logs
cat /ttyUSBx 

The current mapping of USB port # to AIDA FEE64s - attachment 1


aida02 system console c. 19.30 2.5.24

02:05:24/14:40:51|executing generic doStop
02:05:24/19:27:28|Halt (1)
02:05:24/19:27:29|Halt (2)
02:05:24/19:27:29|Halt (3)
02:05:24/19:27:30|Halt (4)
02:05:24/19:27:30|Halt (5)
02:05:24/19:27:31|Halt (6)
02:05:24/19:27:31|Halt (7)
02:05:24/19:27:32|Halt (8)
02:05:24/19:27:32|Halt (9)
02:05:24/19:27:33|Halt (10)
02:05:24/19:27:33|Halt (11)
02:05:24/19:27:34|Halt (12)
02:05:24/19:27:34|Halt (13)
02:05:24/19:27:35|Halt (14)
02:05:24/19:27:35|Halt (15)
02:05:24/19:27:36|Halt (16)
02:05:24/19:27:36|Halt (17)
02:05:24/19:27:37|Halt (18)
02:05:24/19:27:38|Halt (19)
02:05:24/19:27:38|Action has not completed
02:05:24/19:27:39|Giving up waiting
02:05:24/19:27:39|disconnect xfer stream #0 (1)
02:05:24/19:27:39|disconnect xfer stream #0 (2)
02:05:24/19:27:39|disconnect xfer stream #0 (3)
02:05:24/19:27:40|disconnect xfer stream #0 (4)
02:05:24/19:27:40|disconnect xfer stream #0 (5)
02:05:24/19:27:41|disconnect xfer stream #0 (6)
02:05:24/19:27:41|disconnect xfer stream #0 (7)
02:05:24/19:27:42|disconnect xfer stream #0 (8)
02:05:24/19:27:42|disconnect xfer stream #0 (9)
02:05:24/19:27:43|disconnect xfer stream #0 (10)
02:05:24/19:27:43|disconnect xfer stream #0 (11)
02:05:24/19:27:44|disconnect xfer stream #0 (12)
02:05:24/19:27:44|disconnect xfer stream #0 (13)
02:05:24/19:27:45|disconnect xfer stream #0 (14)
02:05:24/19:27:46|disconnect xfer stream #0 (15)
02:05:24/19:27:46|disconnect xfer stream #0 (16)
02:05:24/19:27:47|disconnect xfer stream #0 (17)
02:05:24/19:27:47|disconnect xfer stream #0 (18)
02:05:24/19:27:48|disconnect xfer stream #0 (19)
02:05:24/19:27:48|Action has not completed
02:05:24/19:27:49|Giving up waiting
02:05:24/19:27:49|executing generic Stop_Merger
02:05:24/19:27:49|executing generic Stop_DataStorageServer
02:05:24/19:27:49|completed generic doStop
02:05:24/19:27:49|do_GetState returned z=0 and 1
02:05:24/19:27:49|do_GetState returned z=0 and 1
02:05:24/19:28:01|Transfer Error - : Connection reset by peer
02:05:24/19:28:08|send() failed: 
02:05:24/19:28:08|Aida state now Stopped. AidaExecV10.0_Jul  6 2022: Build 15:10:57 
02:05:24/19:28:08|do_GetState returned z=0 and 0
02:05:24/19:28:13|executing generic doGo
02:05:24/19:28:45|Initialising communication with storage server 192.168.11.99 on port 11002
02:05:24/19:28:45|Setting Transfer Block Size 65536
02:05:24/19:28:45|Setting Transfer Mode 3
02:05:24/19:28:45|Setting Overlap Mode 0
02:05:24/19:28:45|Setting Blocking Mode 0
02:05:24/19:28:45|Setting nice 0
02:05:24/19:28:45|TCP transfer library version 3.12
02:05:24/19:28:45|TCP socket send buffer was 16384 - now 221184
02:05:24/19:28:45|TCP socket receive buffer was 87380 - now 221184
02:05:24/19:28:45|TCP socket created OK - now connecting to 192.168.11.99 port 11002
02:05:24/19:28:45|Connected to 192.168.11.99 port 11002
02:05:24/19:28:45|connect xfer stream #0 (1)
02:05:24/19:28:45|Data Acquisition Statistics counters now cleared
02:05:24/19:28:46|ASIC DMA buffer size 2097152; WAVE DMA buffer size 1048576
02:05:24/19:28:46|Aida state now Going. AidaExecV10.0: Build Jul  6 2022_15:10:57 
02:05:24/19:28:46|Go (1)
02:05:24/19:28:46|Running Go_Electronics for aida02 with TSMaster aida01
02:05:24/19:28:46|ADCs Calibrated. 
02:05:24/19:28:46|Finished Go_Electronics for AIDA
02:05:24/19:28:46|executing generic Go_Merger
02:05:24/19:28:46|executing generic Go_DataStorageServer
02:05:24/19:28:46|completed generic doGo
02:05:24/19:28:46|do_GetState returned z=0 and 1
02:05:24/19:28:46|In RDOGo_Operate: Enabled Correlation, ASIC and discriminator readout
02:05:24/19:28:51|Also in RDOGo_Operate: Enabled waveform readout
02:05:24/19:28:51|get_WAVEBlk (A) which = 0 ---- status = 0048, State machines = 0201
02:05:24/19:28:51|WV:0:Buffer info. Length = 64 : Offset = 0 
02:05:24/19:28:51|WV:1:Buffer info. Length = 62 : Offset = 2 
02:05:24/19:28:51|WV:2:Buffer info. Length = 60 : Offset = 4 
02:05:24/19:28:52|WV:3:Buffer info. Length = 58 : Offset = 6 
02:05:24/19:28:52|WV:4:Buffer info. Length = 56 : Offset = 8 
02:05:24/19:28:52|WV:5:Buffer info. Length = 54 : Offset = 10 
02:05:24/19:28:52|WV:6:Buffer info. Length = 52 : Offset = 12 
02:05:24/19:28:52|WV:7:Buffer info. Length = 50 : Offset = 14 
02:05:24/19:28:52|WV:8:Buffer info. Length = 48 : Offset = 16 
02:05:24/19:28:52|WV:9:Buffer info. Length = 46 : Offset = 18 
02:05:24/19:28:52|get_WAVEBlk (A) which = 1 ---- status = 0088, State machines = 0201
02:05:24/19:28:52|get_WAVEBlk (A) which = 0 ---- status = 0048, State machines = 0908
02:05:24/19:28:52|get_WAVEBlk (A) which = 1 ---- status = 0088, State machines = 0908
02:05:24/19:28:52|get_WAVEBlk (A) which = 0 ---- status = 0048, State machines = 0908
02:05:24/19:28:52|get_WAVEBlk (A) which = 1 ---- status = 0088, State machines = 0908
02:05:24/19:28:52|get_WAVEBlk (A) which = 0 ---- status = 0048, State machines = 0908
02:05:24/19:28:52|get_WAVEBlk (A) which = 1 ---- status = 0088, State machines = 0908
02:05:24/19:28:52|get_WAVEBlk (A) which = 0 ---- status = 1048, State machines = 0201
02:05:24/19:28:52|get_WAVEBlk (A) which = 1 ---- status = 0088, State machines = 0908
02:05:24/19:28:52|
  594   Tue Apr 23 11:09:00 2024 PP12:00 checks

Everything is OK now.

Screenshots attached.

  593   Tue Apr 23 10:52:39 2024 PPAida02 down and back

We lost aida02 at about 11:25 again. Nick fixed it about 11:50.

It is now back in the DAQ.

  592   Tue Apr 23 10:10:31 2024 PPaida02 is back

Nick rebooted AIDA and aida02 came back at around 10:55.

AIDA was out of the DAQ from about 10:30 until 10:55

  591   Tue Apr 23 09:21:01 2024 PP10:20 checks - aida02 is down

Aida02 is down. Contacted Nick.

The rest looks stable, see screenshots.

10:53 CEST (NH): Rebooted aida02 via telnet
Resync ASIC clocks
All system wide checks look ok
ASIC threshold => 0x14

Go > All 16 writing to merger, looks OK

 

aida02 went down again 11:35 CEST
Rebooted with same procedure and is back again

  590   Tue Apr 23 07:27:18 2024 PP08:00-16:00 Tuesday 24 April shift

Checks 08:00-08:30

Screenshots are attached and everything looks OK.

  589   Mon Apr 22 22:35:59 2024 TD00.00-08.00 Tuesday 22 April
23.36 Analysis of data file R17_34 - attachment 1
      max deadtime c. 1% (aida02)
      no timewarps

23.38 DSSSD bias & leakage current OK - attachment 2

      FEE64 temperatures OK - attachment 3
       aida02 ASIC temp - known fault

      ADC data item stats - attachment 4

      per FEE64 Rate spectra - attachment 5

      Merger etc - attachment 6

01.03 DSSSD bias & leakage current OK - attachment 7

      FEE64 temperatures OK - attachment 8
       aida02 ASIC temp - known fault

      ADC data item stats - attachment 9

      per FEE64 Rate spectra - attachment 10

      Merger etc - attachment 11

03.06 DSSSD bias & leakage current OK - attachment 12

      FEE64 temperatures OK - attachment 13
       aida02 ASIC temp - known fault

      ADC data item stats - attachment 14

      per FEE64 Rate spectra - attachment 15

      Merger etc - attachment 16

05.04 DSSSD bias & leakage current OK - attachment 17

      FEE64 temperatures OK - attachment 18
       aida02 ASIC temp - known fault

      ADC data item stats - attachment 19

      per FEE64 Rate spectra - attachment 20

      Merger etc - attachment 21

07.29 DSSSD bias & leakage current OK - attachment 22

      FEE64 temperatures OK - attachment 23
       aida02 ASIC temp - known fault

      ADC data item stats - attachment 24

      per FEE64 Rate spectra - attachment 25

      Merger etc - attachment 26
  588   Mon Apr 22 15:17:07 2024 DSJ16.00-0.00 22/04/24

First checks of shift. All looks ok.

Screenshot of HV, temps, rates, Merger attached. Attachments 1-4.

Screenshots of all spectrum Layout IDs taken before zeroing at 16.35. (Attachments 5-12)

 

checks at 16.00 - merger has crashed - aida01 dropped out  - Tom tried to stop DAQ but got an error. Restart servers restarted at 16.09 

Reset 01, restarted DAQ, and restarted as R16.

16.15 Data seems to be collected but spectra not being incremented in aida01. Timestamps look to be out of sync. Tom restarting aida01 again.

Did not fix, power cycle all FEEs - all spectra reset. aida3,6,11,12,14,15 wont calibrate adcs so wont have waveforms

RUN17 STARTED 16.51|

 HV, temps, rates, Merger looks ok 17.00. Plots saved as attachments 13-16.

 

20.47, all looks ok - see attachment 17-20. Rates are higher in attachment 20 than in attachment 2 from start of shift

21.59 aida02 (link aida1) stopped taking data and dropped out of the merger (see attachment 21). Came back to life after about 15 miutes. Tom logged in remotely to investigate. Seems to be ok. see attachment 22 for timestamps

22.48 - things apear ok. (Attachments 23-26) 

  587   Mon Apr 22 11:43:55 2024 TD21-22.4.24 overnight MIDASsort online sort
Attachment 1 - per DSSSD LEC e_p versus e_n (20keV/channel)

Attachment 2 - per DSSSD HEC m_p versus m_n 

Attachment 3 - per DSSSD HEC x versus y (I really do need to sort out the mapping!)

Attachment 4 - per DSSSD HEC e_p versus e_n (20MeV/channel)

Attachment 5 - per DSSSD per pixel HEC-HEC time (4.096us/channel)

Attachment 6 - per DSSSD implant/decay event rates (262.144us/channel)
               showing beam spill cycle

Attachment 7 - colours/contour levels for all 2D spectra
  586   Mon Apr 22 09:03:54 2024 TDReport: aida04 stops producing data
09.14 aida04 stopped producing data - system console output - attachment 1

It is not clear (to me at least) whether the cause is radiation-induced single event upsets or
the start/stop of LN2 fills.

Nor is it clear why the FEE64 stops
- WR timestamp error (data link to Merger blocks)
- AIDAExecV10 fails
- ?

rebooted aida04

DAQ RESET/SETUP etc cycle for all FEE64s

10.03 successful restart 
      current data file R15
  585   Sun Apr 21 23:20:45 2024 TDMonday 22 April
00.15 Data link to MBS was closed by MBS - reason(s) unknown
      NH restarted MBS receiver and data transfer restarted

01.34 DSSSD bias & leakage current OK - attachment 1

      FEE64 temperatures OK - attachment 2

      ADC data item stats OK - attachment 3

      Merger, TapeServer etc - attachment 4

04.36 DSSSD bias & leakage current OK - attachment 5

      FEE64 temperatures OK - attachment 6

      ADC data item stats OK - attachment 7

      Merger, TapeServer etc - attachment 8

07.54 data storage to local disk enabled
      file aida-gsi:/TapeData/S100_alpha/R14_

09.36 DAQ continues, current data file R14_46

      168Er setting

      DSSSD bias & leakage current OK - attachments 9 & 10
      leakage current DSSSD#1 ramping and cycling with beam spills

      FEE64 temperatures OK - attachment 11

      System wide checks OK *except* WR/FPGA errors - attachments 12-13

      WR timestamps - attachment 14
       aida04 out of seq

      ADC data item stats - attachment 15
       high rates observed all FEE64s beam on, beam off rates OK

      per FEE64 Rate spectra - attachment 16

      per FEE64 1.8.L spectra - attachments 17-18

      per FEE64 1.8.W spectra - attachments 19-20

      per FEE^4 1.8.H spectra - attachments 21-22
   
      Merger, TapeServer etc - attachments 23-25
       data rates to c. 10Mb/s beam on


09.55 aida04 rebooted - https://elog.ph.ed.ac.uk/DESPEC/586

10.01 DAQ restarted
      data file R15

      ADC data item stats - attachments 26-27
       large variation in rates beam off/on

c. 10.30 beam off (for new FRS setting?)


11.25 analysis of data files R14_19 (beam ON) and R15_24 (beam OFF) - attachments 28-29

      For beam ON max deadtime c. 11% (aida04), beam OFF max deadtime c. 0.1% (aida04)

      Zero timewarps observed

      Notice large # of PAUSE/RESUME data items for aida03 & aida04 presumably due to additional overhead of processing the time machine scalers 

12.35 Beam off

      DSSSD bias & leakage current OK - attachment 30
       Can observe beam spill structure in DSSSD leakage currents - now slowly recovering with beam off

      FEE64 temperatures OK - attachment 31

      ADC data item stats OK - attachment 32

      Merger, TapeServer etc - attachment 33

12.55 New ASIC settings 2024Apr22-12.55.38
       *all* discs disabled to reduce load on aida04
       slow comparator 0x14

      BNC PB-5 settings (since yesterday evening)

      Amplitude 10.0V
      Attenuation x10
      tau_d 1ms
      Frequency 2Hz
      Tail pulse
      Polarity +


13.46 all histograms & stats zero'd

13.48 Beam off
      per FEE64 Rate spectra - attachment 34
       2Hz pulser to p+n FEE64s

      per FEE64 1.8.W spectra - attachments 35-36
       p+n FEE64 threshold = 6000 to capture pulser events
       n+n FEE64 threshold = 8000 to capture noise

      per p+n FEE64 1.8.L spectra - attachments 37-38
       common x/y-scale - peak height proxy for peak width
       aida09 pulser peak width 57 ch FWHM ~40keV FWHM ( 5 x sigma ~ 85keV )

14.28 168Eu setting (check only - setup still in progress)
  584   Sun Apr 21 00:34:33 2024 TDSunday 21 April
01.35 DSSSD bias & leakage current OK - Grafana - attachment 1

      FEE64 temperatures OK - attachment 2
 
      ADC data item stats OK - attachment 3

      Merger OK - attachment 4

04.05 DSSSD bias & leakage current OK - Grafana - attachment 5

      FEE64 temperatures OK - attachment 6
 
      ADC data item stats OK - attachment 7

      Merger OK - attachment 8

08.20 DSSSD bias & leakage current OK - Grafana - attachment 9

      FEE64 temperatures OK - attachment 10
 
      ADC data item stats - attachment 11
       increased rates observed esp. DSSSD#2 n+n

      Merger etc - attachment 12
      rate increased x2 cf. overnight

      WR timestamp - attachment 13
       aida03 out sequence

09.25 MBS data relay restarted

      Merger data links - attachment 14
       aida03 rate zero

      per FEE64 1.8.W spectra - 20us FSR - attachments 15-16

      per FEE64 Rate spectra - attachment 17
      increased rate for DSSSD#2 n+n FEE64s in particular

      MIDASsort online analysis - attachments 18-19
      per DSSSD e_front versus e_back for LEC (20keV/channel) and HEC (20MeV/channel)
      observe light ions (LEC) c. 2MeV
      observe low rate of high energy ions (HEC) c. 600MeV
      similar energies in both DSSSDs for LEC and HEC implying both types of event have range >> 2mm Si


10.09 DAQ STOP
      aida03 does not respond - reboot aida03

      DAQ RESET/SETUP etc     

      Synchronise ASIC clocks
      All system wide checks OK *except* FPGA errors

      Data transfer disabled
      Histograms & stats zero'd
      DAQ GO

      Merger restarted 

      Data transfer enabled
      DAQ GO

      WR timestamp OK - attachments 20-21

      ADC data item stats OK - attachment 22
       rates status quo ante
  
      Merger etc OK - attachment 23


14.58 DSSSD bias & leakage current OK - Grafana - attachment 24

      FEE64 temperatures OK - attachment 25

      ADC data item stats OK - attachment 26

      Merger etc OK - attachment 27

18.13 DSSSD bias & leakage current OK - Grafana - attachment 28

      FEE64 temperatures OK - attachment 29

      ADC data item stats OK - attachment 30

      Merger etc OK - attachment 31


20.30 Primary c, few kHz 170Er beam to S4 - upstream s/steel flange removed
      Start of stopping range calibration in the implantation stack - currently ion appear to be stopping in DSSSD#1 (upstream)


      Decay and implant events as function of time showing spill structure - attachment 32

      per DSSSD e_front versus e_back (HEC, 20MeV/channel) - attachment 33

      per DSSSD x versus y hit pattern (HEC) - attachment 34
       mapping to be configured but clear ions are mostly implanted in centre of DSSSD#1 

      2D colour/contour levels - attachment 35

      per DSSSD m_front versus m_back (HEC) - attachment 36

      per DSSSD per pixel HEC-HEC time (4.096us/channel) - attachment 37

23.44 aida02 has stopped running
      DAQ STOP
      aida02 rebooted

      Data transfer disabled

      DAQ RESET/SETUP 

      Synchronise ASIC clocks
      All system wide checks OK *except* FPGA errors

      Histograms & stats zero'd
      DAQ GO

      Merger restarted 

      Data transfer enabled
      DAQ GO







        


        
  583   Sat Apr 20 17:52:27 2024 TD, NHSaturday 20 April contd.
Note 40kBq 60Co source placed on bottom surface of snout centred on bPlas detectors and DSSSDs

18.50 Primary beam 750MeV/u Er to S4
      Primary beam stopped by flange (5cm s/steel in position upstream of implantation stack)

      Observe c. 2MeV dE peak in AIDA DSSSD#1 and DSSSD#2 using MIDASsort online analysis - attachment 1



21.50 Primary beam 1000MeV/u Er to S4 - attachment 2

      per DSSSD e_front versus e_back, 20keV/channel
      DSSSD #1 x & y -axis projections

10.50 Overnight stats - attachment 3
      Primary beam ?
      per DSSSD e_front versus e_back, 20keV/channel
      DSSSD #1 x & y -axis projections
      note multiple dE peaks observed in both DSSSDs
  582   Sat Apr 20 11:40:38 2024 TDOffline analysis R12 & R13 ( pulser walkthrough )
S100 ADC offsets using pulser walkthrough data from data files R12 & R13

ch = channel + ( module * 64 ) + ( range * 2048 )

adc_data( ch ) = INT( RSHIFT( ABS( adc_data( ch ) - 32768 ), 3 ) - offset( ch ) + 0.5 )




per p+n FEE64 ADC spectra (aida01-aida16 *not* aida02, aida04, aida06 & aida08) - attachments 1-12

Note common x & y-scales - peak height proxy for peak width

aida09 1.8.L (s520) pulser peak width 7 ch FWHM (~40keV FWHM)




per n+n FEE64 ADC spectra (aida02, aida04, aida06 & aida08) - attachments 13-16

Note common x & y-scales - peak height proxy for peak width

aida02 1.8.L (s72) pulser peak width 16 ch FWHM (~88keV FWHM)




ADC offset analysis

/home/npg/td/Calibration/offsets /home/npg/td/S100/R12-R13 50 1024 20 5


First pass sort variables ( ADC offsets, FEE64 configuration, LEC & HEC equal energy cuts ) - attachment 17
  581   Sat Apr 20 03:20:23 2024 TDSaturday 20 April
Note 60Co source (current activity c. 43kBq) was installed yesterday evening - assume it is positioned on the bottom side of the snout as per the other sources used this week/

04.17 DSSSD bias & leakage current OK - Grafana - attachment 1

      FEE64 temperatures OK - attachment 2
       aida02 ASIC temp known to be u/s

      ADC data item stats - attachment 3
       all FEE64 slow comparators 0x14 from yesterday's pulser walkthroughs

      per FEE64 Rate spectra - attachment 4

      Merger etc - attachment 5


11.24 DSSSD bias & leakage current OK - Grafana - attachment 6

      FEE64 temperatures OK - attachment 7
       aida02 ASIC temp known to be u/s

      ADC data item stats - attachment 8
       all FEE64 slow comparators 0x14 from yesterday's pulser walkthroughs

      per FEE64 Rate spectra - attachment 9

      Merger etc - attachment 10


17.50 DSSSD bias & leakage current OK - Grafana - attachment 11

      FEE64 temperatures OK - attachment 12
       aida02 ASIC temp known to be u/s

      ALl syhstem wide checks OK *except* WR/FPGA errors - attachments 13-14

      WR timestamps OK - attachment 15

      per FEE64 Rate spectra - attachment 16

      per FEE64 1.8.W spectra - 20us FSR - attachments 17-18

      Merger, TapeServer, Data Links - attachments 19-20
       no data storage

      ADC data item stats - attachment 21
       all FEE64 slow comparators 0x14 from yesterday's pulser walkthroughs


18.01 Grafana alerting shows detail of DSSSD leakage current and S4 temperature variations over time - attachment  22
       DEGAS LN2 fill cycle every 6h 10.00 - 16.00 - 22.00 ... etc

https://despec-vm-01.gsi.de/grafana/d/tHRVJYfGz/aida-alerting?orgId=1
ELOG V3.1.4-unknown