Mon Jul 31 10:21:41 2023, Nikolay Sosnin, n_TOF Data and Filtering
|
|
Wed Feb 12 13:51:02 2025, CLW, expected counts B and Li 6x
|
The rootfiles countsB.root and countsLi.root contain histograms of count spectra that would be expected at n_TOF for a 10B and a 6Li target,
respectively. The units in y are arbitrary. This can be used to check, if we can reproduce the expected trend with our Li and B measurements. So the histograms
called "counts" can be directly compared to the histograms called "energy" in Sili_deed.c (they should have the same binning). You |
Mon Jul 21 10:30:01 2025, Selin, The 2D Energy Spectrum of Ca41
|
The 2D energy spectrum of Ca41 (36 runs). The expected alpha energies are 4.8 MeV (alpha_0) and 2.7 MeV (alpha_1). |
Thu Oct 6 09:34:30 2022, Claudia, Test setup Photos and cabling    
|
Setup 5.10.22
dE 20 um (ID 2837-33) and 494 um double sided (ID 1194-9)
46 channels (2 connectors missing) |
Thu Aug 10 11:30:32 2023, Nikolay Sosnin, Silicons and LiF3+Gafchromic Target 18x
|
|
Fri Jun 20 09:49:11 2025, CLW, Silicon detector setup description
|
|
Sat Oct 8 11:56:49 2022, Nikolay Sosnin, Silicon Strips of Concern
|
DEED 6 and 9 no signal throughout the campaign.
EFED 27 high oscillations in baseline throughout the campaign. |
Fri Jul 11 09:16:18 2025, TD, Silena 7710 Quad Bias Supply manual and internal configuration
|
Manual - see attachment 1
|
Fri Aug 11 13:06:54 2023, Annie, Signal Spreadsheet info  
|
Spreadhseet made for shifters to manually enter that they have checked each signal and that they can see gamma-flashes. It is saved as "26Al_EAR2_Signal_Checks.ods",
it is located on the second monitor on the right of the control room, next to the monitor used to see the beam intensity (same monitor used to see signals).
It will stay open throughout the campeign. Shifters have been told to check signals more than just the once to record that they're okay, |
Fri Sep 2 12:03:40 2016, Sarah, SPD card changed
|
14 bit card replaced for DEED 9-12, to prevent issue with pulser arriving before gamma flash. |
Wed Jul 16 08:55:22 2025, Selin, Run221141 Boron Uncalibrated Data  
|
Plots of Boron #1 run (221141) from the Ca41 campaign with the right voltage applied (11.07.2025).
1- Uncalibrated amplitude spectra of strip 1 (dedicated).
2- Uncalibrated 2d energy plot of Boron dedicated data. |
Wed Jul 16 08:51:06 2025, Selin, Run221136 Lithium Uncalibrated Data  
|
Plots of Lithium #3 run (221136) from the Ca41 campaign with the right voltage applied (11.07.2025).
1- Uncalibrated amplitude spectra of strip 1 (dedicated).
2- Uncalibrated amplitude spectra of strip 1 (parasitic). |
Sat Jul 12 10:30:06 2025, Francisco García Infantes, Report Experimental SetUp Ca41   
|
I'm attaching a report for the experimental setup so far.
There is a main problem which haven't been solved yet and I don't know if that can cause problem in the future analysis.
Channels #23, #24, #31 and #32 are not working well. The signal os these channels look attenuated from the pulser, make impossible to see |
Mon Aug 14 11:00:13 2023, TD, RAL108 +/-15V PSU test in ISOLDE hall 13x
|
This morning 2x RAL108 +/-15V PSUs were borrowed from the Edinburgh equipment in the ISOLDE hall to check whether the same transient noise is observed
at the +/-15V PSU
|
Tue Aug 15 10:17:39 2023, TD, NS, RAL108 +/-15V PSU test at EAR2, n_TOF Monday 14 August 7x
|
On the morning of Monday 14 August 2x RAL108 +/-15V PSUs were borrowed from the Edinburgh equipment in the ISOLDE hall to check whether the same transient
noise is observed at the +/-15V PSU outputs.
|
Mon Aug 21 12:08:59 2023, TD, RAL108 +/-15V PSU test - JCMB 21.8.23 13x
|
PSU Calex
Setup, PSU and ac mains filter - attachments 1-3
|
Sat Aug 19 14:14:26 2023, TD, RAL108 +/-15V PSU test - JCMB 18.8.23 10x
|
PSU Calex
Setup and PSU details - attachments 1-3
|
Wed Aug 6 11:20:52 2025, Selin, Pulser Test   
|
The pulser test was performed with a frequency of 500 Hz and with voltage settings of: 2V, 1.6V, 1.2V, 0.8V, 0.6V, 0.4V, 0.3V, 0.2V, 0.1V, 0.05V, 0.025V.
The run was started with calibration settings and then the voltage was changed every 2.5-3 minutes.
From the attached plots, it can be seen that everything is as expected. This means that the pulser multipolarity is consistent with front and |
Fri Aug 11 09:11:04 2023, Annie, Pulse Settings   
|
|
Fri Aug 11 10:12:02 2023, Annie, Pressure, voltage and current check  
|
|