ID |
Date |
Author |
Subject |
772
|
Sat Nov 10 03:50:00 2018 |
CA, CB, TD, OH | Saturday 10 November |
12.35 DAQ continues OK - data transfer disabled
12.37 all system wide checks OK *except* nnaida7 fails ADC calibration
detector biases & leakage currents OK - see attachments 1 & 2
Julabo FL11006 chiller set point +20 deg C actual +20 deg C, water level topped up c. 1 litre - 80% (L-H)
ambient tempature + 27.3 deg C, RH 33.0%, d.p. +9.6 deg C
FEE64 temperatures OK - see attachment 3
good event statistics OK - see attachment 4
per FEE64 rate spectra - see attachment 5
1.8.L spectra - see attachments 6 & 7
nnaida1 pulser peak width 71 ch FWHM
1.8.W spectra 20us FSR - see attachments 8 & 9 |
Attachment 1: 1.png
|
|
Attachment 2: 2.png
|
|
Attachment 3: 3.png
|
|
Attachment 4: 4.png
|
|
Attachment 5: 5.png
|
|
Attachment 6: 6.png
|
|
Attachment 7: 7.png
|
|
Attachment 8: 8.png
|
|
Attachment 9: 9.png
|
|
774
|
Sat Nov 10 21:53:29 2018 |
CA, CB, TD, OH | Sunday 11 November |
06.53 DAQ continues OK - data transfer disabled
06.55 all system wide checks OK *except* nnaida7 fails ADC calibration
detector biases & leakage currents OK - see attachments 1 & 2
Julabo FL11006 chiller set point +20 deg C actual +20 deg C, water level c. 75% (L-H)
ambient tempature + 26.9 deg C, RH 21.6%, d.p. +2.9 deg C
FEE64 temperatures OK - see attachment 3
good event statistics OK - see attachment 4
per FEE64 rate spectra - see attachment 5
ASIC check load
per FEE64 rate spectra - see attachment 6
good event statistics OK - see attachment 7
1.8.L spectra - see attachments 8 & 9
nnaida1 pulser peak width 72 ch FWHM
1.8.W spectra 20us FSR - see attachments 10 & 14
1.8.W spectra 200us FSR - see attachments 11 & 15
1.8.W spectra 2ms FSR - see attachments 12 & 16
1.8.W spectra 20ms FSR - see attachments 13 & 17
Note - LED threshold 9000 for *all* FEE64s
07.27 statistics zero'd
some 32bit counters (e.g. nnaida19 & nnaida20) had overflowed/wrapped around
18:19 Computer appears to be running slow though cannot see a reason why.
No programs using to much ram or cpu
FEE temp ok - attachment 18
bias and leakage currents ok - attachment 19
System wide checks ok
21:02 System wide checks ok
DAQ Continues ok - data transfer disabled.
Bias and leakage currents ok - attachment 20
FEE Temperatures ok - attachment 21 |
Attachment 1: 1.png
|
|
Attachment 2: 2.png
|
|
Attachment 3: 3.png
|
|
Attachment 4: 4.png
|
|
Attachment 5: 5.png
|
|
Attachment 6: 6.png
|
|
Attachment 7: 7.png
|
|
Attachment 8: 8.png
|
|
Attachment 9: 9.png
|
|
Attachment 10: 10.png
|
|
Attachment 11: 11.png
|
|
Attachment 12: 12.png
|
|
Attachment 13: 13.png
|
|
Attachment 14: 14.png
|
|
Attachment 15: 15.png
|
|
Attachment 16: 16.png
|
|
Attachment 17: 17.png
|
|
Attachment 18: 181111_1817_Temp.png
|
|
Attachment 19: 181111_1819_Bias.png
|
|
Attachment 20: 181111_2103_Bias.png
|
|
Attachment 21: 181111_2103Temp.png
|
|
850
|
Mon Jun 17 08:03:37 2019 |
CA, CB, TD, OH | Mon Jun 17 16:00-24:00 |
16:04 DAQ continues running OK
All system-wide checks OK except nnaida7 ADC calibration fails
FEE64 temperatures OK - attach 1
Merger and Tape Server OK - see attach 2 & 3
16:30 DTAS and BigRIPS stop
AIDA on file R2_406
16:35 Beam stop.
AIDA plastic gain increased by a factor 2
Moved the lead bricks between AIDA and F11, 1 cm further away to the sides
Chiller water level OK
Humidity 34.6%
Temperature 23.4 C
Dew point 7.1 C
16:40 Beam restart
AIDA run R2_408
DTAS 70 (from DTAS ELOG)
BigRIPS 2030 (from DTAS ELOG)
No obvious changes to the implantation pattern. May need more stats.
Detector leakage current OK - see attach 4
17.31 nnaida14 3.*.H & 4.*.H spectra - attachments 5 & 6
Corresponds to RHS DSSSD#3 - decrease in rate observed but no evidence of systematic variation of HEC energy
Modified implantdecay4.f to select events with x & y energy > ADC ch 1600 (c. 1MeV) to observe x-y hit pattern
of light particles - see attachments 7 & 8
Low stats but light particle distribution appears to be symmetric with no obvious RHS cut
17:37 DTAS & BigRIPS DAQ stop
Beam stop
Lead moved back as before
AIDA plastic gain decreased by a factor 5
17:40 DTAS and BigRIPS stop
AIDA on file R2_416
17.48 zero *all* SortSas spectra
17:50 Beam restarting
AIDA R2_418
DTAS 73 (72 is a mistake - ignore)
BigRIPS 2031
18:43 DAQ continues OK
All system wide checks OK - except nnaida7 fails ADC calibration as usual
FEE64 temperature OK - see attach 9
Stats OK - see attach 10
Merger & Tape Server OK - see attach 11 & 12
Leakage currents OK - see attach 13
HEC & LEC events plots attached - attach 14 & 15 - note LEC in DSSD#3 are > 3 MeV only (light particle-like)
18:50 AIDA R2_427
DTAS 74
BigRIPS 2031
19.08 DTAS & BigRIPS DAQ stop - beam lost
AIDA file R2_429
21:15 Beam is back - briefly
AIDA R2_446
21:17 Beam lost - operators advise "this will take a long time"
22:16 Beam is back!
AIDA R2_454
DTAS 78
BigRIPS 2034
Statistics OK - attach 15
FEE64 temperatures OK - attach 16
Merger & Tape Server OK - see attach 17 & 18
System-wide checks as usual
DTAS reports rate in F11 down by 20%, and rate in AIDA plastic sees similar drop. |
Attachment 1: 190617_1607_tape.png
|
|
Attachment 2: 190617_1607_temp.png
|
|
Attachment 3: 190617_1647_bias.png
|
|
Attachment 4: 1.png
|
|
Attachment 5: 2.png
|
|
Attachment 6: 3.png
|
|
Attachment 7: 4.png
|
|
Attachment 8: 190617_1845_stats.png
|
|
Attachment 9: 190617_1845_temps.png
|
|
Attachment 10: 190617_1846_merger.png
|
|
Attachment 11: 190617_1846_tape.png
|
|
Attachment 12: 190617_1847_bias.png
|
|
Attachment 13: 190617_1848_HEC.png
|
|
Attachment 14: 190617_1848_LEC_3MeV.png
|
|
Attachment 15: 190617_2224_merger.png
|
|
Attachment 16: 190617_2224_stats.png
|
|
Attachment 17: 190617_2224_temps.png
|
|
Attachment 18: 190617_2225_tape.png
|
|
Attachment 19: 190617_2230_bias.png
|
|
762
|
Sun Nov 4 01:11:41 2018 |
CA, CB, TD, OH, PJW | Sunday 4 November |
10.03 DAQ continues OK - current file Oct18/R16_9
FEE64 LEC/MEC slow comparator threshold 0x64
beam background
10.05 all system wide checks OK *except* nnaida7 fails ADC calibration
detector biases & leakage currents OK - see attachments 1 & 2
FEE64 temperatures OK - see attachment 3
good event statistics OK *except* nnaida2 - see attachment 4
nnaida2 rate due to one hot channel as before
Merger OK - see attachment 5
TapeServer OK - see attachment 6
no errors or warnings shown by server terminal sessions
1.8.L spectra - see attachments 7 & 8
N.B. evidence of minimally ionising particles (presumably protons) from beam to F11 E detector
~10-15cm Pb bricks were placed within BRIKEN (CH2)n stack to stop heavy particles
1.8.W spectra 20us FSR - see attachments 9 & 10
online analysis with MIDASsort
attachment 11 - per DSSSD m_p versus m_n
12 - per DSSSD x versus y
13 - per DSSSD E_p versus E_n
14 - channel versus E_p
15 - hit pattern (ASIC & DSSSD order)
14.35 DAQ continues OK - current file Oct18/R16_11
FEE64 LEC/MEC slow comparator threshold 0x64
beam background
14.35 all system wide checks OK *except* nnaida7 fails ADC calibration
detector biases & leakage currents OK - see attachments 15 & 16
FEE64 temperatures OK - see attachment 17
good event statistics OK *except* nnaida2 - see attachment 18
nnaida2 rate due to one hot channel as before
14.51 analysis file Oct18/R16_10 - attachment 19 |
Attachment 1: 1.png
|
|
Attachment 2: 2.png
|
|
Attachment 3: 3.png
|
|
Attachment 4: 4.png
|
|
Attachment 5: 5.png
|
|
Attachment 6: 6.png
|
|
Attachment 7: 7.png
|
|
Attachment 8: 8.png
|
|
Attachment 9: 9.png
|
|
Attachment 10: 10.png
|
|
Attachment 11: 11.png
|
|
Attachment 12: 12.png
|
|
Attachment 13: 13.png
|
|
Attachment 14: 14.png
|
|
Attachment 15: 15.png
|
|
Attachment 16: 1.png
|
|
Attachment 17: 2.png
|
|
Attachment 18: 3.png
|
|
Attachment 19: 4.png
|
|
Attachment 20: R16_10
|
*** MIDAS TDR format 3.3.0 AIDA data analyser - TD - November 2018
*** ERROR: READ I/O error: 5002
blocks: 32000
ADC data format: 174098646 ( 19529.0 Hz)
Other data format: 87021354 ( 9761.3 Hz)
Sample trace data format: 0 ( 0.0 Hz)
Undefined format: 0 ( 0.0 Hz)
Other data format type: PAUSE: 0 ( 0.0 Hz)
RESUME: 0 ( 0.0 Hz)
SYNC100: 81618360 ( 9155.3 Hz)
FEE64 disc: 54111 ( 6.1 Hz)
MBS info: 5348883 ( 600.0 Hz)
Other info: 0 ( 0.0 Hz)
ADC data range bit set: 0 ( 0.0 Hz)
Timewarps: ADC: 0 ( 0.0 Hz)
PAUSE: 0 ( 0.0 Hz)
RESUME: 0 ( 0.0 Hz)
SYNC100: 0 ( 0.0 Hz)
FEE64 disc: 0 ( 0.0 Hz)
MBS info: 0 ( 0.0 Hz)
Undefined: 0 ( 0.0 Hz)
Sample trace: 0 ( 0.0 Hz)
*** Timestamp elapsed time: 8914.900 s
FEE elapsed dead time(s) elapsed idle time(s)
1 0.000 0.000
2 0.000 0.000
3 0.000 0.000
4 0.000 0.000
5 0.000 0.000
6 0.000 0.000
7 0.000 0.000
8 0.000 0.000
9 0.000 0.000
10 0.000 0.000
11 0.000 0.000
12 0.000 0.000
13 0.000 0.000
14 0.000 0.000
15 0.000 0.000
16 0.000 0.000
17 0.000 0.000
18 0.000 0.000
19 0.000 0.000
20 0.000 0.000
21 0.000 0.000
22 0.000 0.000
23 0.000 0.000
24 0.000 0.000
25 0.000 0.000
26 0.000 0.000
27 0.000 0.000
28 0.000 0.000
29 0.000 0.000
30 0.000 0.000
31 0.000 0.000
32 0.000 0.000
*** Statistics
FEE ADC Data Other Data Sample Undefined Pause Resume SYNC100 Disc MBS Other HEC Data
0 0 0 0 0 0 0 0 0 0 0 0
1 9170 3403362 0 0 0 0 3400765 2597 0 0 0
2 173967998 3402321 0 0 0 0 3400765 1556 0 0 0
3 3715 3402542 0 0 0 0 3400765 1777 0 0 0
4 3883 3402764 0 0 0 0 3400765 1999 0 0 0
5 3845 3402745 0 0 0 0 3400765 1980 0 0 0
6 32850 8751357 0 0 0 0 3400765 1709 5348883 0 0
7 3935 3403802 0 0 0 0 3400765 3037 0 0 0
8 3765 3402601 0 0 0 0 3400765 1836 0 0 0
9 4441 3404098 0 0 0 0 3400765 3333 0 0 0
10 4229 3403033 0 0 0 0 3400765 2268 0 0 0
11 3938 3403183 0 0 0 0 3400765 2418 0 0 0
12 4105 3403950 0 0 0 0 3400765 3185 0 0 0
13 4122 3402388 0 0 0 0 3400765 1623 0 0 0
14 4251 3403646 0 0 0 0 3400765 2881 0 0 0
15 4324 3402358 0 0 0 0 3400765 1593 0 0 0
16 4135 3402599 0 0 0 0 3400765 1834 0 0 0
17 4548 3402909 0 0 0 0 3400765 2144 0 0 0
18 4259 3402386 0 0 0 0 3400765 1621 0 0 0
19 4304 3403421 0 0 0 0 3400765 2656 0 0 0
20 4575 3403738 0 0 0 0 3400765 2973 0 0 0
21 4369 3403400 0 0 0 0 3400765 2635 0 0 0
22 4607 3403340 0 0 0 0 3400765 2575 0 0 0
23 4818 3402709 0 0 0 0 3400765 1944 0 0 0
24 4460 3402702 0 0 0 0 3400765 1937 0 0 0
25 0 0 0 0 0 0 0 0 0 0 0
26 0 0 0 0 0 0 0 0 0 0 0
27 0 0 0 0 0 0 0 0 0 0 0
28 0 0 0 0 0 0 0 0 0 0 0
29 0 0 0 0 0 0 0 0 0 0 0
30 0 0 0 0 0 0 0 0 0 0 0
31 0 0 0 0 0 0 0 0 0 0 0
32 0 0 0 0 0 0 0 0 0 0 0
*** Timewarps
FEE ADC Pause Resume SYNC100 Disc MBS Undefined Samples
0 0 0 0 0 0 0 0 0
1 0 0 0 0 0 0 0 0
2 0 0 0 0 0 0 0 0
3 0 0 0 0 0 0 0 0
4 0 0 0 0 0 0 0 0
5 0 0 0 0 0 0 0 0
6 0 0 0 0 0 0 0 0
7 0 0 0 0 0 0 0 0
8 0 0 0 0 0 0 0 0
9 0 0 0 0 0 0 0 0
10 0 0 0 0 0 0 0 0
11 0 0 0 0 0 0 0 0
12 0 0 0 0 0 0 0 0
13 0 0 0 0 0 0 0 0
14 0 0 0 0 0 0 0 0
15 0 0 0 0 0 0 0 0
16 0 0 0 0 0 0 0 0
17 0 0 0 0 0 0 0 0
18 0 0 0 0 0 0 0 0
19 0 0 0 0 0 0 0 0
20 0 0 0 0 0 0 0 0
21 0 0 0 0 0 0 0 0
22 0 0 0 0 0 0 0 0
23 0 0 0 0 0 0 0 0
24 0 0 0 0 0 0 0 0
25 0 0 0 0 0 0 0 0
26 0 0 0 0 0 0 0 0
27 0 0 0 0 0 0 0 0
28 0 0 0 0 0 0 0 0
29 0 0 0 0 0 0 0 0
30 0 0 0 0 0 0 0 0
31 0 0 0 0 0 0 0 0
32 0 0 0 0 0 0 0 0
*** MBS Timewarps
FEE
0 0
1 0
2 0
3 0
4 0
5 0
6 0
7 0
8 0
9 0
10 0
11 0
12 0
13 0
14 0
15 0
16 0
17 0
18 0
19 0
20 0
21 0
22 0
23 0
24 0
25 0
26 0
27 0
28 0
29 0
30 0
31 0
32 0
*** Program elapsed time: 8.402s ( 3808.461 blocks/s, 238.029 Mb/s)
|
765
|
Mon Nov 5 01:40:10 2018 |
CA, CB, TD, OH, PJW | Monday 5 November |
10.40 Lowered slow comparator threshold from 100 to 10
Good event statistics ok (attachment 1)
System wide checks ok, except for nnaida7 ADC calibration failure
FEE64 temperatures ok (attachment 2)
Bias/Leakage currents ok (attachment 3)
1.0.L spectra obtained, look ok (attachment 4)
DAQ stopped (R16)
10.49 Slow comparator threshold raised from 10 to 100
DAQ restarted (R17)
Merger and TapeServer ok, no error messages in terminals
Good event statistics ok (attachment 5)
nnaida2 is now at a good rate, after fixing kapton connection.
15:09 DAQ still running (Oct18/R17_0)
Merger and TapeServer ok
System wide checks ok, except for nnaida7 ADC calibration failure
FEE64 temperatures ok (attachment 6)
Bias/Leakage currents ok (attachment 7)
Good event statistics ok (attachment 8)
18.48 good event statistics OK *except* nnaida17 c. 400k
ASIC Control -> Check ASIC Control
*all* FEE64 good events statistics now OK
DAQ stop - current file Oct18/R17_20
18.50 DAQ start - file Oct18/R18
BNC PB-4 OFF
LEC/MEC slow comparator 0x64
alpha background run
18.54 all system wide checks OK *except* nnaida7 fails ADC calibration
detector biases & leakage currents OK - see attachments 9 & 10
FEE64 temperatures OK - see attachment 11
good event statistics OK - see attachment 12
Merger - Link16 errors reported - see attachment 13
TapeServer OK - see attachment 14
1.8.W spectra - attachment 15 & 16
|
Attachment 1: 181105_1027_goodeventstats.png
|
|
Attachment 2: 181105_1029_temp.png
|
|
Attachment 3: 181105_1030_bias.png
|
|
Attachment 4: 181105_1026_spectra.png
|
|
Attachment 5: 181105_1051_goodeventstats.png
|
|
Attachment 6: 181105_1513_temp.png
|
|
Attachment 7: 181105_1514_bias.png
|
|
Attachment 8: 181105_1518_goodeventstats.png
|
|
Attachment 9: 100.png
|
|
Attachment 10: 101.png
|
|
Attachment 11: 102.png
|
|
Attachment 12: 104.png
|
|
Attachment 13: 105.png
|
|
Attachment 14: 106.png
|
|
Attachment 15: 107.png
|
|
Attachment 16: 108.png
|
|
768
|
Wed Nov 7 01:00:58 2018 |
CA, CB, TD, OH, PJW | Wednesday 7 November |
10.02 DAQ still running, on file Oct18/R27_9
Merger and TapeServer ok, no error warnings in terminals
10.07 Good event statistics ok, except nnaida6 rate being high (attachment 1)
nnaida17 rate ok, but counts very high. May have run fast at some point overnight.
System wide checks ok, except nnaida7 failure in ADC calibration
FEE64 temperatures ok (attachment 2)
10.13 Bias/leakage currents slightly higher, but ok (attachment 3)
10:30 1.8.W spectra (attachment 4)
11.10 Stopped the DAQ (R27_9)
Checked ASIC control - ok
Started the DAQ again (Oct18/R28_0)
good event statistics ok, aside from nnaida6 running slightly fast (attachment 5)
15.07 DAQ stop - file Oct18/R28_2
Detector bias OFF
Move AIDA downstream to check position of Clover G7
16.05 Move back upstream into BRIKEN (CH2)n moderator assembly
Detector bias ON
16.12 all system wide checks OK *except* nnaida7 fails ADC calibration
sync all ASIC clocks using ReSYNC
detector biases & leakage currents OK - see attachments 6 & 7
FEE64 temperatures OK - see attachment 8
good event statistics OK *except* nnaida2 - see attachment 9
see attachment 10 for comparison to statistics 5.1.18
16.41 1.8.W spectra 20us FSR - see attachments 11 & 12
16.55 slow comparator thresholds raised from 10 to 100
DAQ start (Oct18/R29_0)
Merger and TapeServer ok - no error warnings in terminals
good event statistics ok, except nnaida6 running fast (attachment 13)
|
Attachment 1: 181107_1006_goodeventstats.png
|
|
Attachment 2: 181107_1011_temp.png
|
|
Attachment 3: 181107_1013_bias.png
|
|
Attachment 4: 181107_1029_18W_spectra.png
|
|
Attachment 5: 181107_1114_goodeventstats.png
|
|
Attachment 6: 1.png
|
|
Attachment 7: 2.png
|
|
Attachment 8: 3.png
|
|
Attachment 9: 4.png
|
|
Attachment 10: 181105_1027_goodeventstats.png
|
|
Attachment 11: 5.png
|
|
Attachment 12: 6.png
|
|
Attachment 13: 181107_1658_goodeventstats.png
|
|
751
|
Sat Oct 27 01:22:43 2018 |
CA, OH | Saturday 27 October |
09:22 DAQ found crashed. Appears to have crashed at 20:23 last night whilst writing R3_0 to file.
Run Control is non responsive
Leakage currents okay - attachment 1
Unable to ping any of the nnaida FEEs
Restarted merger in an attempt to clear links still no response.
Have powered down FEE64
Replaced blown fuse
Checked temperatures of FEEs, all okay - attachment 2
10:19 Restarted alpha run, write to file Oct18/R4.
System wide checks - all ok.
Slow comparator threshold changed to 100 from 10
14:12 FEE temperatures ok (attachment 3)
Bias/leakage currents ok (attachment 4)
System wide checks ok
Alpha run on R4_1
17.45 DAQ continues OK, file Oct18/R4_3 |
Attachment 1: 181027_0927_bias.png
|
|
Attachment 2: 181027_1004_temp.png
|
|
Attachment 3: 181027_1410_temp.png
|
|
Attachment 4: 181027_1409_bias.png
|
|
752
|
Sun Oct 28 01:33:34 2018 |
CA, OH | Sunday 28 October |
09:33 DAQ still running, file R4_11.
Bias/leakage ok (attachment 1)
Leakage currents whilst still acceptable are up around 400nA on yesterday
Have checked all detectors and rates and pulser peak widths are still good across the board
Will keep an eye on but for now it appears to be ok
BRIKEN is currently doing a Cf source test so could be a result of charging from the source
Humidity: 33.1, Temperature: 24.6, Dew point: 7.1
FEE temperatures ok (attachment 2)
09:38 System wide checks ok
10:25 Pulser peak widths all checked
Widths comparable to 26th apart from 24 which is much wider from ~70 -> ~120
Waveforms, layout 7 and 8 - attachments 3 and 4
10:37 Started new run on file R5 at time 10:37
Threshold changed back to 100 from 10
Pulser on at 2Hz
16:34 Stopped R5 alpha run
Lowered thresholds to 0xa to produce higher rate data to test AIDA<->BRIKEN synchronisation
Started file R6 to do this
DAQ crashed on start ASIC readout with unable to connect to nnaida19 message appearing.
Could not clear despite attempts at restarting the merge if that was the cause.
FEE power cycle
17:03 DAQ recovered, R7 started
AIDA did not crash
BRIKEN started Ge fill cycle
AIDA did not crash during or after
17:17 R7 Stopped sync program is being worked on by Jorge,
17:32 Thresholds raised back to 100 from 10
Pulser switched off
R8 started
leakage current dropping down (attachment 5) |
Attachment 1: 181028_0933_bias.png
|
|
Attachment 2: 181028_0937_temp.png
|
|
Attachment 3: 181028_1023_1.8.W.png
|
|
Attachment 4: 181028_1024_layout8.png
|
|
Attachment 5: 181028_1734_bias.png
|
|
793
|
Fri May 10 05:58:01 2019 |
CA, OH, TD | Friday 10 May 2019 |
11.00 - chiller refilled and switched on
set to 20 degrees celsius
AIDA stand moved and secured in F11
clock connected -> 50 MHz square wave at 5.0 V
AIDA connected to power box
network cables attached
NIM modules switched on
Raspberry Pis switched on
Detector bias switched on for detectors 1-6
13.58 - AIDA power relay 1,2,5,6 sequenced on
bias/leakage currents ok (attachment 1)
DAQ setup
System wide checks ok, except naida07 fails ADC calibration, and clock status check returns error in
nnaida06, 5.
Above issue with clock no longer present
FEE temperatures ok (attachment 2)
good event statistics (attachment 3)
16.30 1.8.W spectra n+n strips
FSR 20us, 200us, 2ms, 20ms attachments 4-7
1.8.W spectra p+n strips
FSR 20us, 200us, 2ms, 20ms attachments 8-11
compare with https://elog.ph.ed.ac.uk/AIDA/746
naida19, naida20, naida23 show low frequency noise structure, worse than in above elog entry
17.22 1.8.L spectra p+n strips
peaks:
aida01 72
aida02 112
aida03 80
aida04 78
aida05 74
aida06 95
aida07 78
aida08 71
aida09 82
aida10 85
aida11 93
aida12 96
aida13 101
aida14 70
aida15 87
aida16 87
aida17 69
aida18 83
aida19 127
aida20 116
aida21 84
aida22 74
aida23 108
aida24 117
|
Attachment 1: 190510_bias_1158.png
|
|
Attachment 2: 190509_1409_temp.png
|
|
Attachment 3: 190510_1418_stats.png
|
|
Attachment 4: 190510_1613_18W.png
|
|
Attachment 5: 190905_1615_18W_10.png
|
|
Attachment 6: 190510_1628_18w_100_back.png
|
|
Attachment 7: 190510_1627_18w_1000_back.png
|
|
Attachment 8: 190510_1622_18w_0_front.png
|
|
Attachment 9: 190510_1620_18w_10_front.png
|
|
Attachment 10: 190510_1624_100_front.png
|
|
Attachment 11: 190510_1625_18w_1000_front.png
|
|
Attachment 12: 190510_1713_back_18L.png
|
|
794
|
Sat May 11 02:50:14 2019 |
CA, OH, TD | Saturday 11th May 2019 |
10:52 system wide checks ok, except naida07 fails ADC calibration
good event statistics, rates in naida19,20,23,24 high (attachment 1)
FEE temperatures ok (attachment 2)
detector bias/leakage currents stable (attachment 3) |
Attachment 1: 190511_1053_stats.png
|
|
Attachment 2: 190511_1053_temp.png
|
|
Attachment 3: 190511_1054_bias.png
|
|
799
|
Tue May 14 02:19:20 2019 |
CA, OH, TD | Tuesday 14th May 2019 |
10.20 System wide checks ok (except naida07 ADC calibration failure)
FEE temperatures ok (attachment 1)
good event statistics ok (attachment 2)
detector bias and leakage currents ok (attachment 3)
DAQ stopped
10.40 pulser switched off
slow comparator threshold raised to 0x64
DAQ, merger, TapeServer started - all ok (attachments 4 and 5)
writing to file R1 in TapeData/May19 at 70 kb/s
good event statistics (attachment 6)
16:11 Tapeserver and merger still ok
DAQ still running
System wide checks all ok *except nnaida7 fails ADC calibration*
Bias and leakage currents ok - attachment 7
FEE Temperatures ok - attachment 8
Statistics still ok
Spectrum are showing alpha events (Layout 4) - attachment 9
18:25 Tektronix clock accidentally unplugged. DAQ has stalled as a result.
Clock reset to 50MHz 5V TTL signal
18:29 System wide check of clock status shows error message in attachment 10
DAQ Reset
With slow comp at oxa statistics are still good - attachment 11
System wide checks all ok *except nnaida7 fails ADC calibration*
Resync to the ASIC clock
All histograms zeroed
18:37 Slow comp threshold raised to 0x64 for alpha run
Writing to file:
May19/R2_0
|
Attachment 1: 190514_1016_temp.png
|
|
Attachment 2: 190514_1018_stats.png
|
|
Attachment 3: 190514_1019_bias.png
|
|
Attachment 4: merger.png
|
|
Attachment 5: tapeserver.png
|
|
Attachment 6: 190514_1054_stats.png
|
|
Attachment 7: 190514_1615_bias.png
|
|
Attachment 8: 190514_1615_FEETemp.png
|
|
Attachment 9: 190514_1614_18L.png
|
|
Attachment 10: 190514_1829_CheckClockStatus.png
|
|
Attachment 11: 190514_1834_Stats.png
|
|
800
|
Wed May 15 02:17:23 2019 |
CA, OH, TD | Wednesday 15 May |
10.18 DAQ continues OK to file May19/R2
alpha background
ASIC settings 2018Oct03-13.26.58
*except* slow comparator 0xa -> 0x64
BNC PB-4 Pulser OFF
all system wide checks OK *except* nnaida7 fails ADC calibration
detector biases & leakage currents OK - see attachments 1 & 2
FEE64 temperatures OK - see attachment 3
good event statistics OK - see attachments 4
Merger OK - see attachment 5
TapeServer OK - see attachment 6
1.8.L spectra - attachments 7 & 8
1.8.W spectra 20us FSR - attachments 10 & 11
1.8.W spectra 20ms FSR - attachments 10 & 11
Julabo FL11006 chiller set point +20 deg C actual +20.0 deg C - water topped up
Temperature + deg 24.7C dew point +7.5 deg C RH 33.3%
FEE64 nnaida1-4 temperature logs - see attachments 12-15
all FEE64 temperatures appear to be stable
11.05 check ASIC control *all* FEE64s and ASICs
11.13 analysis of file May19/R2_1 - see attachment 16
ADC rate c. 72Hz dominated by same, single hot channel (Stat spectrum ch# 104) in each of FEE64s nnaida1,
2, 5 & 6 otherwise looks OK
sync pulser rate 10Hz confirmed by JA
17.25 all system wide checks OK *except* nnaida7 fails ADC calibration
FEE64 temperatures OK - see attachment 18
good event statistics OK - see attachments 19
detector biases & leakage currents OK - see attachment 20
Temperature + deg 24.5C dew point +8.2 deg C RH 34.1%
Merger and TapeServer ok |
Attachment 1: 1.png
|
|
Attachment 2: 2.png
|
|
Attachment 3: 3.png
|
|
Attachment 4: 190514_1834_Stats.png
|
|
Attachment 5: 5.png
|
|
Attachment 6: 6.png
|
|
Attachment 7: 7.png
|
|
Attachment 8: 8.png
|
|
Attachment 9: 10.png
|
|
Attachment 10: 11.png
|
|
Attachment 11: 12.png
|
|
Attachment 12: 13.png
|
|
Attachment 13: 1.png
|
|
Attachment 14: 2.png
|
|
Attachment 15: 3.png
|
|
Attachment 16: 4.png
|
|
Attachment 17: R2_1
|
*** MIDAS TDR format 3.3.0 AIDA data analyser - TD - November 2018
*** ERROR: READ I/O error: 5002
blocks: 32000
ADC data format: 2056849 ( 72.9 Hz)
Other data format: 259063151 ( 9185.6 Hz)
Sample trace data format: 0 ( 0.0 Hz)
Undefined format: 0 ( 0.0 Hz)
Other data format type: PAUSE: 0 ( 0.0 Hz)
RESUME: 0 ( 0.0 Hz)
SYNC100: 258208982 ( 9155.3 Hz)
FEE64 disc: 8079 ( 0.3 Hz)
MBS info: 846090 ( 30.0 Hz)
Other info: 0 ( 0.0 Hz)
ADC data range bit set: 0 ( 0.0 Hz)
Timewarps: ADC: 0 ( 0.0 Hz)
PAUSE: 0 ( 0.0 Hz)
RESUME: 0 ( 0.0 Hz)
SYNC100: 0 ( 0.0 Hz)
FEE64 disc: 0 ( 0.0 Hz)
MBS info: 0 ( 0.0 Hz)
Undefined: 0 ( 0.0 Hz)
Sample trace: 0 ( 0.0 Hz)
*** Timestamp elapsed time: 28203.305 s
FEE elapsed dead time(s) elapsed idle time(s)
1 0.000 0.000
2 0.000 0.000
3 0.000 0.000
4 0.000 0.000
5 0.000 0.000
6 0.000 0.000
7 0.000 0.000
8 0.000 0.000
9 0.000 0.000
10 0.000 0.000
11 0.000 0.000
12 0.000 0.000
13 0.000 0.000
14 0.000 0.000
15 0.000 0.000
16 0.000 0.000
17 0.000 0.000
18 0.000 0.000
19 0.000 0.000
20 0.000 0.000
21 0.000 0.000
22 0.000 0.000
23 0.000 0.000
24 0.000 0.000
25 0.000 0.000
26 0.000 0.000
27 0.000 0.000
28 0.000 0.000
29 0.000 0.000
30 0.000 0.000
31 0.000 0.000
32 0.000 0.000
*** Statistics
FEE ADC Data Other Data Sample Undefined Pause Resume SYNC100 Disc MBS Other HEC Data
0 0 0 0 0 0 0 0 0 0 0 0
1 81457 10765234 0 0 0 0 10758707 6527 0 0 0
2 115328 10758862 0 0 0 0 10758707 155 0 0 0
3 83 10758755 0 0 0 0 10758707 48 0 0 0
4 109 10758773 0 0 0 0 10758707 66 0 0 0
5 1558907 10758768 0 0 0 0 10758707 61 0 0 0
6 297802 11604852 0 0 0 0 10758708 54 846090 0 0
7 97 10758755 0 0 0 0 10758708 47 0 0 0
8 85 10758769 0 0 0 0 10758708 61 0 0 0
9 267 10758856 0 0 0 0 10758708 148 0 0 0
10 233 10758894 0 0 0 0 10758708 186 0 0 0
11 84 10758762 0 0 0 0 10758708 54 0 0 0
12 92 10758777 0 0 0 0 10758708 69 0 0 0
13 242 10758744 0 0 0 0 10758708 36 0 0 0
14 270 10758759 0 0 0 0 10758708 51 0 0 0
15 86 10758708 0 0 0 0 10758708 0 0 0 0
16 89 10758739 0 0 0 0 10758708 31 0 0 0
17 281 10758804 0 0 0 0 10758708 96 0 0 0
18 270 10758752 0 0 0 0 10758708 44 0 0 0
19 96 10758741 0 0 0 0 10758708 33 0 0 0
20 155 10758791 0 0 0 0 10758707 84 0 0 0
21 208 10758759 0 0 0 0 10758707 52 0 0 0
22 347 10758792 0 0 0 0 10758707 85 0 0 0
23 157 10758777 0 0 0 0 10758707 70 0 0 0
24 104 10758728 0 0 0 0 10758707 21 0 0 0
25 0 0 0 0 0 0 0 0 0 0 0
26 0 0 0 0 0 0 0 0 0 0 0
27 0 0 0 0 0 0 0 0 0 0 0
28 0 0 0 0 0 0 0 0 0 0 0
29 0 0 0 0 0 0 0 0 0 0 0
30 0 0 0 0 0 0 0 0 0 0 0
31 0 0 0 0 0 0 0 0 0 0 0
32 0 0 0 0 0 0 0 0 0 0 0
*** Timewarps
FEE ADC Pause Resume SYNC100 Disc MBS Undefined Samples
0 0 0 0 0 0 0 0 0
1 0 0 0 0 0 0 0 0
2 0 0 0 0 0 0 0 0
3 0 0 0 0 0 0 0 0
4 0 0 0 0 0 0 0 0
5 0 0 0 0 0 0 0 0
6 0 0 0 0 0 0 0 0
7 0 0 0 0 0 0 0 0
8 0 0 0 0 0 0 0 0
9 0 0 0 0 0 0 0 0
10 0 0 0 0 0 0 0 0
11 0 0 0 0 0 0 0 0
12 0 0 0 0 0 0 0 0
13 0 0 0 0 0 0 0 0
14 0 0 0 0 0 0 0 0
15 0 0 0 0 0 0 0 0
16 0 0 0 0 0 0 0 0
17 0 0 0 0 0 0 0 0
18 0 0 0 0 0 0 0 0
19 0 0 0 0 0 0 0 0
20 0 0 0 0 0 0 0 0
21 0 0 0 0 0 0 0 0
22 0 0 0 0 0 0 0 0
23 0 0 0 0 0 0 0 0
24 0 0 0 0 0 0 0 0
25 0 0 0 0 0 0 0 0
26 0 0 0 0 0 0 0 0
27 0 0 0 0 0 0 0 0
28 0 0 0 0 0 0 0 0
29 0 0 0 0 0 0 0 0
30 0 0 0 0 0 0 0 0
31 0 0 0 0 0 0 0 0
32 0 0 0 0 0 0 0 0
*** MBS Timewarps
FEE
0 0
1 0
2 0
3 0
4 0
5 0
6 0
7 0
8 0
9 0
10 0
11 0
12 0
13 0
14 0
15 0
16 0
17 0
18 0
19 0
20 0
21 0
22 0
23 0
24 0
25 0
26 0
27 0
28 0
29 0
30 0
31 0
32 0
*** Program elapsed time: 13.602s ( 2352.671 blocks/s, 147.042 Mb/s)
|
Attachment 18: 150519_1707_temp.png
|
|
Attachment 19: 150519_1708_stats.png
|
|
Attachment 20: 150519_1709_bias.png
|
|
801
|
Thu May 16 02:13:10 2019 |
CA, OH, TD | Thursday 16th May |
10:13 all system wide checks OK *except* nnaida7 fails ADC calibration
FEE64 temperatures OK - see attachment 1
good event statistics OK - see attachments 2
detector biases & leakage currents OK - see attachment 3
Temperature + deg 24.7C dew point +8.6 deg C RH 35.0%
Merger and TapeServer ok
10:30 DAQ, merger, TapeServer stop
Alpha run stopped at AIDA file TapeData/May19/R2_5
slow comparator threshold lowered to 10 keV (0xa)
DAQ start, not writing to file
good event statistics after lowering threshold, ok (attachment 4)
10:49 1.8.W spectra 20ms FSR - attachments 5 & 6
11:25 started writing to file R3 (sync test)
low threshold run at 5mb/s
11:33 DAQ stop (R3)
slow comparator threshold raised to 100 kev (0x64)
DAQ start, alpha run
Writing to file TapeData/May19/R4
18.49 alpha background
ASIC settings 2018Oct03-13.26.58
*except* slow comparator 0xa -> 0x64
BNC PB-4 Pulser OFF
FEE64 temperatures OK - see attachment 7
good event statistics - see attachments 8 - nnaida1 405k & nnaida17 368k, all other FEE64s c. 380Hz as
expected
ASIC check for *all* ASICs & FEE64s restored all FEE64 rates to expected c. 380Hz - see attachment 13
detector biases & leakage currents OK - see attachments 9 & 10
Merger OK - no error messages - see attachment 11
TapeServer OK - see attachment 12
all system wide checks OK *except* nnaida7 fails ADC calibration *and* Check Sync errors not counted Since
Baseline - see attachment 14
DAQ stopped May19/R4_3
aidas1> ls -l /TapeData/May19
total 19808048
-rw-rw-rw-. 1 npg npgstaff 1908736000 May 14 18:28 R1_0
-rw-rw-rw-. 1 npg npgstaff 2097152000 May 15 02:30 R2_0
-rw-rw-rw-. 1 npg npgstaff 2097152000 May 15 10:20 R2_1
-rw-rw-rw-. 1 npg npgstaff 2097152000 May 15 18:08 R2_2
-rw-rw-rw-. 1 npg npgstaff 2097152000 May 16 01:57 R2_3
-rw-rw-rw-. 1 npg npgstaff 2097152000 May 16 09:45 R2_4
-rw-rw-rw-. 1 npg npgstaff 190447616 May 16 10:28 R2_5
-rw-rw-rw-. 1 npg npgstaff 1109327872 May 16 11:28 R3_0
-rw-rw-rw-. 1 npg npgstaff 2097152000 May 16 18:39 R4_0
-rw-rw-rw-. 1 npg npgstaff 2097152000 May 16 18:45 R4_1
-rw-rw-rw-. 1 npg npgstaff 2097152000 May 16 18:50 R4_2
-rw-rw-rw-. 1 npg npgstaff 225312768 May 16 18:54 R4_3
-rw-rw-rw-. 1 npg npgstaff 52494336 May 16 19:10 R5_0
Wahtever the issue was it probably occurred before 18.39
DAQ RESET
DAQ SETUP
ASIC check for *all* ASICs & FEE64s
all system wide checks OK *except* nnaida7 fails ADC calibration
Sync all ASIC clocks using Re-SYNC
DAQ started May19/R5
19.23 1.8.W spectra - see attachments 15 & 16 |
Attachment 1: 190516_1016_temp.png
|
|
Attachment 2: 190516_1017_stats.png
|
|
Attachment 3: 190516_1018_bias.png
|
|
Attachment 4: 190516_10.png
|
|
Attachment 5: 190516_1046_18wfront.png
|
|
Attachment 6: 190516_1048_18wback.png
|
|
Attachment 7: 1.png
|
|
Attachment 8: 2.png
|
|
Attachment 9: 3.png
|
|
Attachment 10: 4.png
|
|
Attachment 11: 5.png
|
|
Attachment 12: 6.png
|
|
Attachment 13: 7.png
|
|
Attachment 14: 8.png
|
|
Attachment 15: 11.png
|
|
Attachment 16: 12.png
|
|
807
|
Mon May 20 02:16:58 2019 |
CA, TD | Monday 20 May 2019 |
10:18 DAQ Continues ok
Alpha run continues - Currently on file R8_2
System wide checks all ok
Detector bias/leakage currents ok - attachment 1
Temp +25.6 deg C, Humidity 34.5% AND D.P +8.8 deg C
FEE Temperatures ok - attachment 2
Good event statistics ok - attachment 3
Merger & TapeServer ok - attachments 4 & 5
17.15 DAQ continues OK to file May19/R8_2
alpha background
ASIC settings 2019May19-17.54.03
slow comparator 0x64
BNC PB-4 Pulser OFF
all system wide checks OK *except* nnaida7 fails ADC calibration
detector biases & leakage currents OK - see attachments 6 & 7
FEE64 temperatures OK - see attachment 8
good event statistics OK - see attachments 9
Merger OK - see attachment 10
TapeServer OK - see attachment 11
1.8.L spectra - see attachments 12 & 13
p+n FEE64s 1.8.W spectra 20us FSR - attachment 14
n+n FEE64s 1.8.W spectra 20us FSR - attachment 15
17.25 analysis of May19/R8_2 - see attachment 16
deadtime OK
ADC data for FEE64s nnaida5 & nnaida6 hot channels (Stat spectrum channel # 104)
Also observed at a much lower rate for nnaida1 & nnaida2
17.27 Check ASIC Control *all* ASICs & FEE64s
13.27 Julabo FL11006 chiller set point +20 deg C actual +20.0 deg C
Water level c. 60% M-H markers - water topped up (1x jugful) to H level marker
Temperature +25.4 deg C dew point +9.6 deg C RH 36.3% |
Attachment 1: 200519_bias.png
|
|
Attachment 2: 190520_1022_temp.png
|
|
Attachment 3: 190520_1023_stats.png
|
|
Attachment 4: 190519_merger.png
|
|
Attachment 5: 190519_tape.png
|
|
Attachment 6: 1.png
|
|
Attachment 7: 2.png
|
|
Attachment 8: 3.png
|
|
Attachment 9: 4.png
|
|
Attachment 10: 5.png
|
|
Attachment 11: 6.png
|
|
Attachment 12: 7.png
|
|
Attachment 13: 8.png
|
|
Attachment 14: 9.png
|
|
Attachment 15: 10.png
|
|
Attachment 16: R8_2
|
*** MIDAS TDR format 3.3.0 AIDA data analyser - TD - November 2018
*** ERROR: READ I/O error: 5002
blocks: 29623
ADC data format: 2975396 ( 114.5 Hz)
Other data format: 238748284 ( 9187.4 Hz)
Sample trace data format: 0 ( 0.0 Hz)
Undefined format: 0 ( 0.0 Hz)
Other data format type: PAUSE: 0 ( 0.0 Hz)
RESUME: 0 ( 0.0 Hz)
SYNC100: 237913266 ( 9155.3 Hz)
FEE64 disc: 55432 ( 2.1 Hz)
MBS info: 779586 ( 30.0 Hz)
Other info: 0 ( 0.0 Hz)
ADC data range bit set: 2 ( 0.0 Hz)
Timewarps: ADC: 0 ( 0.0 Hz)
PAUSE: 0 ( 0.0 Hz)
RESUME: 0 ( 0.0 Hz)
SYNC100: 0 ( 0.0 Hz)
FEE64 disc: 0 ( 0.0 Hz)
MBS info: 0 ( 0.0 Hz)
Undefined: 0 ( 0.0 Hz)
Sample trace: 0 ( 0.0 Hz)
*** Timestamp elapsed time: 25986.471 s
FEE elapsed dead time(s) elapsed idle time(s)
1 0.000 0.000
2 0.000 0.000
3 0.000 0.000
4 0.000 0.000
5 0.000 0.000
6 0.000 0.000
7 0.000 0.000
8 0.000 0.000
9 0.000 0.000
10 0.000 0.000
11 0.000 0.000
12 0.000 0.000
13 0.000 0.000
14 0.000 0.000
15 0.000 0.000
16 0.000 0.000
17 0.000 0.000
18 0.000 0.000
19 0.000 0.000
20 0.000 0.000
21 0.000 0.000
22 0.000 0.000
23 0.000 0.000
24 0.000 0.000
25 0.000 0.000
26 0.000 0.000
27 0.000 0.000
28 0.000 0.000
29 0.000 0.000
30 0.000 0.000
31 0.000 0.000
32 0.000 0.000
*** Statistics
FEE ADC Data Other Data Sample Undefined Pause Resume SYNC100 Disc MBS Other HEC Data
0 0 0 0 0 0 0 0 0 0 0 0
1 134034 9966618 0 0 0 0 9913052 53566 0 0 2
2 216203 9913277 0 0 0 0 9913052 225 0 0 0
3 84 9913112 0 0 0 0 9913052 60 0 0 0
4 114 9913181 0 0 0 0 9913052 129 0 0 0
5 2118344 9913102 0 0 0 0 9913052 50 0 0 0
6 503593 10692674 0 0 0 0 9913052 36 779586 0 0
7 85 9913102 0 0 0 0 9913053 49 0 0 0
8 93 9913117 0 0 0 0 9913053 64 0 0 0
9 218 9913217 0 0 0 0 9913053 164 0 0 0
10 230 9913294 0 0 0 0 9913053 241 0 0 0
11 79 9913162 0 0 0 0 9913053 109 0 0 0
12 95 9913091 0 0 0 0 9913053 38 0 0 0
13 248 9913085 0 0 0 0 9913053 32 0 0 0
14 257 9913120 0 0 0 0 9913053 67 0 0 0
15 81 9913053 0 0 0 0 9913053 0 0 0 0
16 89 9913098 0 0 0 0 9913053 45 0 0 0
17 280 9913108 0 0 0 0 9913053 55 0 0 0
18 257 9913106 0 0 0 0 9913053 53 0 0 0
19 100 9913123 0 0 0 0 9913053 70 0 0 0
20 137 9913137 0 0 0 0 9913053 84 0 0 0
21 266 9913131 0 0 0 0 9913053 78 0 0 0
22 268 9913170 0 0 0 0 9913053 117 0 0 0
23 152 9913124 0 0 0 0 9913053 71 0 0 0
24 89 9913082 0 0 0 0 9913053 29 0 0 0
25 0 0 0 0 0 0 0 0 0 0 0
26 0 0 0 0 0 0 0 0 0 0 0
27 0 0 0 0 0 0 0 0 0 0 0
28 0 0 0 0 0 0 0 0 0 0 0
29 0 0 0 0 0 0 0 0 0 0 0
30 0 0 0 0 0 0 0 0 0 0 0
31 0 0 0 0 0 0 0 0 0 0 0
32 0 0 0 0 0 0 0 0 0 0 0
*** Timewarps
FEE ADC Pause Resume SYNC100 Disc MBS Undefined Samples
0 0 0 0 0 0 0 0 0
1 0 0 0 0 0 0 0 0
2 0 0 0 0 0 0 0 0
3 0 0 0 0 0 0 0 0
4 0 0 0 0 0 0 0 0
5 0 0 0 0 0 0 0 0
6 0 0 0 0 0 0 0 0
7 0 0 0 0 0 0 0 0
8 0 0 0 0 0 0 0 0
9 0 0 0 0 0 0 0 0
10 0 0 0 0 0 0 0 0
11 0 0 0 0 0 0 0 0
12 0 0 0 0 0 0 0 0
13 0 0 0 0 0 0 0 0
14 0 0 0 0 0 0 0 0
15 0 0 0 0 0 0 0 0
16 0 0 0 0 0 0 0 0
17 0 0 0 0 0 0 0 0
18 0 0 0 0 0 0 0 0
19 0 0 0 0 0 0 0 0
20 0 0 0 0 0 0 0 0
21 0 0 0 0 0 0 0 0
22 0 0 0 0 0 0 0 0
23 0 0 0 0 0 0 0 0
24 0 0 0 0 0 0 0 0
25 0 0 0 0 0 0 0 0
26 0 0 0 0 0 0 0 0
27 0 0 0 0 0 0 0 0
28 0 0 0 0 0 0 0 0
29 0 0 0 0 0 0 0 0
30 0 0 0 0 0 0 0 0
31 0 0 0 0 0 0 0 0
32 0 0 0 0 0 0 0 0
*** MBS Timewarps
FEE
0 0
1 0
2 0
3 0
4 0
5 0
6 0
7 0
8 0
9 0
10 0
11 0
12 0
13 0
14 0
15 0
16 0
17 0
18 0
19 0
20 0
21 0
22 0
23 0
24 0
25 0
26 0
27 0
28 0
29 0
30 0
31 0
32 0
*** Program elapsed time: 18.426s ( 1607.693 blocks/s, 100.481 Mb/s)
|
857
|
Thu Jun 20 07:35:46 2019 |
CA, TD | Bias & leakage current spreadsheet |
https://docs.google.com/spreadsheets/d/12hgbrywB10hGsKt5uc2HnLfZymh8_uvM6cEASbbqnBE/edit#gid=1560332343 |
Attachment 1: AIDA_leakage_current(1).xlsx
|
858
|
Thu Jun 20 07:42:54 2019 |
CA, TD | AIDA shutdown checklist |
stop the daq
pause merger
stop TapeServer
housekeeping checks
bias off
individually turn power relays off
switch off 2x NIN bins, 1x fan tray
do not turn off clock!!
turn off aida circuits at power box (careful!)
turn chiller off - press and hold arrow button
front panel switch on chiller - off
wall 3 phase power - off
withdraw aida clear from DTAS - slowly!
return power tool, temperature probe and water container to corral
*if cables*
network cables - blue, disconnect at switch, reel up at stairwell
disconnect clock
aida plastic high voltage and signal
cables from root macb trigger and reset to DTAS
mains cables from mains box
then move aida! |
780
|
Tue Nov 13 07:35:37 2018 |
CA, TD, OH | Tuesday 13 November 16:00 - 00:00 |
16.36 DAQ continues ok
AIDA file RIBF148/R5_198
BRIKEN file 37
BigRIPS file 1022
16.38 system wide checks ok, except nnaida7 ADC calibration failure
16.40 FEE64 temperatures ok (attachment 1)
16.42 good event statistics ok (attachment 2)
16.43 detector bias/leakage currents ok (attachment 3)
16.44 Merger and TapeServer ok, no error messages in terminals (attachments 4 and 5)
16.51 BRIKEN stop
AIDA file RIBF148/R5_200
17.00 Al Degrader thickness changed from 2.3mm to 2.8mm
The 2.3mm plate was previously assumed to be 2.5mm, but this has been corrected by measuring individual Al
plates with digital micrometer.
17.02 BRIKEN start
BRIKEN file 38
BigRIPS file 1023
AIDA file RIBF148/R5_202
18.01 BRIKEN stop
AIDA file RIBF148/R5_211
18.03 BRIKEN start
AIDA file RIBF148/R5_212
BRIKEN file 39
BigRIPS file 1024
18.08 system wide checks ok, except nnaida7 ADC calibration failure
FEE64 temperatures ok (attachment 6)
good event statistics ok (attachment 8)
detector bias/leakage currents ok (attachment 7)
18.11 Merger and TapeServer ok, no error messages in terminal
19.00 BRIKEN stop
AIDA file RIBF148/R5_221
19.02 BRIKEN start
AIDA file RIBF148/R5_221
BRIKEN file 40
BigRIPS file 1025
19.57 system wide checks ok, except nnaida7 ADC calibration failure
FEE64 temperatures ok (attachment 9)
good event statistics ok (attachment 10)
detector bias/leakage current ok (attachment 11)
20.01 BRIKEN stop
AIDA file RIBF148/R5_231
20.03 BRIKEN start
AIDA file RIBF148/R5_231
BRIKEN file 41
BigRIPS file 1026
21.01 BRIKEN stop
AIDA file RIBF148/R5_240
21.03 BRIKEN start
AIDA file RIBF148/R5_240
BRIKEN file 42
BigRIPS file 1027
21:18 system wide checks ok - except nnaida7 ADC calibration failure
FEE64 temperatures ok (attachment 12)
good event statistics ok (attachment 13)
detector bias/leakage current ok (attachment 14)
22:01 BRIKEN stop
AIDA file RIBF148/R5_250
22:03 BRIKEN start
AIDA file RIBF148/R5_250
BRIKEN file 43
BigRIPS file 1028
22:10 system wide checks ok - except nnaida7 ADC calibration failure
FEE64 temperatures ok (attachment 15)
good event statistics ok (attachment 16)
detector bias/leakage current ok (attachment 17)
Merger and TapeServer ok, no error messages in terminal
23:01 BRIKEN stop
AIDA file RIBF148/R5_260
23:03 BRIKEN start
AIDA file RIBF148/R5_260
BRIKEN file 44
BigRIPS file 1029
23.05 system wide checks ok - except nnaida7 ADC calibration failure
FEE64 temperatures ok (attachment 18)
good event statistics ok (attachment 20)
detector bias/leakage current ok (attachment 19)
Merger and TapeServer ok, no error messages in terminal
|
Attachment 1: 181113_1641_temp.png
|
|
Attachment 2: 181113_1642_stats.png
|
|
Attachment 3: 181113_1643_bias.png
|
|
Attachment 4: 181113_1644_merger.png
|
|
Attachment 5: 181113_1645_tapeserver.png
|
|
Attachment 6: 181113_1805_temp.png
|
|
Attachment 7: 181113_1807_bias.png
|
|
Attachment 8: 181113_1807_stats.png
|
|
Attachment 9: 181113_1954_temp.png
|
|
Attachment 10: 181113_1955_stats.png
|
|
Attachment 11: 181113_1956_bias.png
|
|
Attachment 12: 181113_2111_temp.png
|
|
Attachment 13: 181113_2114_stats.png
|
|
Attachment 14: 181113_2116_bias.png
|
|
Attachment 15: 181113_2207_temp.png
|
|
Attachment 16: 181113_2208_stats.png
|
|
Attachment 17: 181113_2209_bias.png
|
|
Attachment 18: 181113_2306_temp.png
|
|
Attachment 19: 181113_2307_bias.png
|
|
Attachment 20: 181113_2307_stats.png
|
|
789
|
Fri Nov 16 04:47:39 2018 |
CA, TD, OH | Correlation analysis at different degrader Al thickness |
13.53 Correlation analysis plot of all detectors with 3.3mm Al degrader thickness (attachment 1)
13.54 Correlation analysis plot of detectors 1 and 2 with 3mm, 2.3mm and 2.8mm degrader thickness (attachment 2)
Correlation analysis plot of detectors 3 and 4 with 3mm, 2.3mm and 2.8mm degrader thickness (attachment 3)
Correlation analysis plot of detectors 5 and 6 with 3mm, 2.3mm and 2.8mm degrader thickness (attachment 4)
14.02 Plots made over first 200ms of data |
Attachment 1: 3mmcorr.png
|
|
Attachment 2: 181116_1350_det12.png
|
|
Attachment 3: 181116_1351_det34.png
|
|
Attachment 4: 181116_1352_det56.png
|
|
795
|
Mon May 13 03:23:58 2019 |
CA, TD, OH | Monday 13 May |
09.55 all system wide checks OK *except* nnaida7 fails ADC calibration
FEE64 temperatures OK - see attachment 1
good event statistics OK - see attachment 2
all FEE64s <100k, 11x < 20k
1.8.L spectra - see attachments 3 & 4
nnaida1 70ch FWHM, nnaida3 80 ch FWHM
detector biases & leakage currents OK - see attachments 5 & 6
1.8.W spectra 20us FSR - see attachments 7 & 8
nnaida1 Options, Local Controls, Data Redaout, Discriminator Controls - see attachment 9-13
Lost Activity Monitor -see attachment 14
18.01 DAQ found stopped
This was caused by a chnage to the AIDA external clock frequency - see https://elog.ph.ed.ac.uk/AIDA/798
Following change of AIDA external clock frequency to 50MHz
DAQ RESET
DAQ SETUP
Sync all ASIC clocks using Resync
all system wide checks OK *except* nnaida7 fails ADC calibration
zero all histograms
DAQ GO
18.31 FEE64 temperatures OK - see attachment 15
good event statistics OK - see attachment 16
all FEE64s <100k, 10x < 20k
detector biases & leakage currents OK - see attachments 17 & 18 |
Attachment 1: 10.png
|
|
Attachment 2: 21.png
|
|
Attachment 3: 26.png
|
|
Attachment 4: 27.png
|
|
Attachment 5: 22.png
|
|
Attachment 6: 23.png
|
|
Attachment 7: 70.png
|
|
Attachment 8: 71.png
|
|
Attachment 9: 60.png
|
|
Attachment 10: 61.png
|
|
Attachment 11: 62.png
|
|
Attachment 12: 63.png
|
|
Attachment 13: 64.png
|
|
Attachment 14: 65.png
|
|
Attachment 15: 3.png
|
|
Attachment 16: 4.png
|
|
Attachment 17: 5.png
|
|
Attachment 18: 6.png
|
|
796
|
Mon May 13 03:43:08 2019 |
CA, TD, OH | Sunday 12 May |
19.12 all system wide checks OK *except* nnaida7 fails ADC calibration
FEE64 temperatures OK - see attachment 9
detector biases & leakage currents OK - see attachments 10 & 11
good event statistics OK - see attachment 12
all FEE64s <100k, 10x < 20k
1.8.W spectra 20us FSR - see attachments 13 & 14 |
Attachment 1: 190512_1130_temp.png
|
|
Attachment 2: 190512_1131_bias.png
|
|
Attachment 3: 190512_1138_hit.png
|
|
Attachment 4: 190512_1139_stats.png
|
|
Attachment 5: 190512_1143_18w_back.png
|
|
Attachment 6: 190512_1144_18w_front.png
|
|
Attachment 7: 190512_1147_18l_front.png
|
|
Attachment 8: 190512_1150_18l_back.png
|
|
Attachment 9: 1.png
|
|
Attachment 10: 2.png
|
|
Attachment 11: 3.png
|
|
Attachment 12: 4.png
|
|
Attachment 13: 5.png
|
|
Attachment 14: 6.png
|
|