381
|
Tue Nov 29 14:37:56 2022 |
AR TD | Tuesday 29 November |
16:20
DSSD bias and leakage current values:
-99.99 V
-13.2535 uA
Pulsar peak widths (ch FWHM)
aida01 ~ 100
aida02 ~ 103
aida03 ~ 128
aida04 ~ 133
aida05 ~ 22
Pretty similar to what was found in august https://elog.ph.ed.ac.uk/CARME/377
|
Attachment 1: IMG_8240_3.jpg
|
|
Attachment 2: IMG_8241_3.jpg
|
|
Attachment 3: IMG_8242_3.jpg
|
|
Attachment 4: IMG_8243_3.jpg
|
|
Attachment 5: IMG_8246_3.jpg
|
|
Attachment 6: IMG_8247_3.jpg
|
|
Attachment 7: IMG_8248_3.jpg
|
|
Attachment 8: IMG_8249_3.jpg
|
|
Attachment 9: IMG_8250_3.jpg
|
|
Attachment 10: IMG_8252_3.jpg
|
|
Attachment 11: IMG_8253_3.jpg
|
|
Attachment 12: IMG_8254_3.jpg
|
|
Attachment 13: IMG_8255_3.jpg
|
|
Attachment 14: IMG_8257_3.jpg
|
|
Attachment 15: IMG_8258_3.jpg
|
|
Attachment 16: IMG_8259_3.jpg
|
|
Attachment 17: IMG_8264_3.jpg
|
|
Attachment 18: Screenshot_2022-11-29_at_16.40.13.png
|
|
Attachment 19: Screenshot_2022-11-29_at_16.42.37.png
|
|
Attachment 20: Screenshot_2022-11-29_at_16.47.03.png
|
|
Attachment 21: Screenshot_2022-11-29_at_16.49.07.png
|
|
Attachment 22: Screenshot_2022-11-29_at_16.49.44.png
|
|
380
|
Mon Nov 28 13:14:33 2022 |
AR TD | Monday 28 November |
|
Attachment 1: 91F62DFF-B008-4204-91E5-BD041D8E4254.jpeg
|
|
Attachment 2: EA4AFB0D-7960-4C5D-834D-AA8B5932DAC2.jpeg
|
|
Attachment 3: F05ACF42-F97E-4F0F-9174-0B997A86136F.jpeg
|
|
Attachment 4: 02F9137F-A28C-4369-B0DB-EEBCEC2A5F5C.jpeg
|
|
Attachment 5: 9ED571DA-3088-43D6-B186-44ABEA96DB93.jpeg
|
|
Attachment 6: F3526AC5-E4B8-4D8A-9D88-57B4A46B4B0D.jpeg
|
|
Attachment 7: 80088713-1F6E-4AAB-AEA0-77E6FC78DAF9.jpeg
|
|
Attachment 8: 9B5BCDB9-6774-4873-A5BD-FC2960FB3FFF.jpeg
|
|
Attachment 9: 9340303A-4F43-4C31-B420-3A4D6ACF4EA7.jpeg
|
|
Attachment 10: 574D2470-3B5E-499E-A9F7-C561E8535115.jpeg
|
|
Attachment 11: 227ED3B1-7F1E-4529-90C3-CE84CAFA121C.jpeg
|
|
Attachment 12: CB94FBF3-2FFF-4795-B03D-5FD8CFE34242.jpeg
|
|
379
|
Mon Oct 3 16:03:33 2022 |
OH, CB | TTT14 Photos |
Photos of the new TTT14 DSSD delivered to GSI.
Photos of detectors are broken up by photos of SN on box
To big to put on elog. Have uploaded to /Disk/ds-sopa-group/np/GSI/CARME/TTT14_Photos.zip and will change permissions to 777 |
378
|
Sun Aug 14 06:50:58 2022 |
TD, JM | Sunday 14 August |
07.51 DAQ continues files 22Feb/R97
ASIC settings 2022Jan31-11-52-00
slow comparator 0x64
EG&G Ortec 448 Pulser OFF
All system wide checks OK
Attachment 1 - ADC data item stats
Attachment 2 - FEE64 temps OK
Attachments 3 & 4, 8 & 9 - per FEE64 1.8.W spectra 20us FSR and 200us FSR
Attachments 5 & 6 - per FEE64 1.8.L spectra
Attachment 7 - per FEE64 stat spectra
Leakage current remains 8.36 uA, 100V.
Bias power OFF, SIP ion pump power back ON. NIM crate powered off. Motors pi and nnrpi1 still on for remote access.
Compressed air off, water off. Fee relay power off and power cable disconnected. Motors box off. Safety pin for motors reinserted.
Checked cable harnesses at cryring. 2 good short sections found and left with boxes in cryring. 2 damaged long sections (bad pin and broken ceramic) taken to repair in
Edinburgh. |
Attachment 1: 60.PNG
|
|
Attachment 2: 59.PNG
|
|
Attachment 3: 58.PNG
|
|
Attachment 4: 57.PNG
|
|
Attachment 5: 56.PNG
|
|
Attachment 6: 55.PNG
|
|
Attachment 7: 54.PNG
|
|
Attachment 8: 61.PNG
|
|
Attachment 9: 62.png
|
|
Attachment 10: 20220814_091209.jpg
|
|
377
|
Sat Aug 13 10:23:59 2022 |
JM, TD | Noise tests pt3 and pulser walk |
Waveforms and discriminator ON. Pulser re-connected to all fees via daisy chain. V=100v, I=8.36uA
Pulser peak widths (ch FWHM)
aida01 ~ 109
aida02 ~ 98
aida05 ~ 24
aida03 ~ 118
aida04 ~ 136
Attachments 2, 3 & 4 1.8.W spectra, 20us FSR
Attachment 5 FEE64 temps OK
Attachment 6 per FEE64 ADC data item rates OK
Attachment 7 per FEE64 rate spectra
Merger reset following procedure from https://elog.ph.ed.ac.uk/DESPEC/36. Keywords for future elog search NetVar, xfer links
Pulser walkthrough R94
EG&G Ortec 448 Pulser settings
Amplitudes 90,000 - 10,000 @ 10,000 step
Attenuators x5 IN
t_r 50ns tau_d 50us
polarity + (- polarity via Cooknell SA1)
Frequency int relay
Motors box turned ON, motors pi ON, dlink network connector ON. Motors box, pi and dlink all in 19 inch rack but power supply is from a uk mains extender daisy chained to vacuum mains extender. 19 inch rack power supply remains on clean line.
Pulser peak widths (ch FWHM)
aida01 ~ 105
aida02 ~ 98
aida05 ~ 23
aida03 ~ 132
aida04 ~ 144
See attachments 8 & 9
Detectors moved in by pnuematic only. Leakage current increased to 9.6 uA (same as leakage current during bias ramping)
Pulser Walk R95
Pulser settings as above
See attachments 10 & 11
Detectors moved back out . Motors box, pi and dlink still turned ON. Leakage current to 8.36 uA
Pulser Walk R96
Pulser settings as above
See attachments 1 & 12
Jumper removed from nn bias fee adaptor board (aida04). No jumpers on any adaptor boards now. V=100V, I=8.36uA
See attachments 14 & 15 - 1.8.W spectra 20us FSR
Pulser on 90,000 x5 attenuate
aida01 ~ 123, centroid ~14691
aida02 ~ 106, centroid ~ 15038
aida05 ~ 23, centroid ~ 14414
aida03 ~ 125, centroid ~ 50838
aida04 ~ 127, centroid ~ 51754
Jumper reinstalled on aida04 nn bias. Note only link 1, link2 inaccesible with fee inserted into ERNI. See attachment 13
aida01 ~ 123, centroid ~14688
aida02 ~ 114, centroid ~ 15034
aida05 ~ 23, centroid ~ 14411
aida03 ~ 118, centroid ~ 50854
aida04 ~ 135, centroid ~ 51697
Pulser off. Background run to be left overnight started 14:24 R97. V=100v, I=8.36 uA. |
Attachment 1: 51.PNG
|
|
Attachment 2: 40.PNG
|
|
Attachment 3: 41.PNG
|
|
Attachment 4: 42.PNG
|
|
Attachment 5: 43.PNG
|
|
Attachment 6: 44.PNG
|
|
Attachment 7: 45.PNG
|
|
Attachment 8: 46.PNG
|
|
Attachment 9: 47.PNG
|
|
Attachment 10: 48.PNG
|
|
Attachment 11: 49.PNG
|
|
Attachment 12: 50.PNG
|
|
Attachment 13: 20220811_172335.jpg
|
|
Attachment 14: 53.PNG
|
|
Attachment 15: 52.PNG
|
|
376
|
Fri Aug 12 15:12:37 2022 |
JM, TD | CARME measurements |
measurement 1 -string
intermediate (smaller) chamber - 2144 (mm) circumference, measured half circumference 1072 mm
outer (larger) chamber - 2160 (mm) circumference, measured 1/4 circumference 540mm
measurement 2 - foil tape
intermediate (smaller) chamber - 2156 (mm) circumference, measured half circumference 1078 mm
outer (larger) chamber - 2172 (mm) circumference, measured 1/4 circumference 543mm
Difference from centre to flat surface on outer chamber is 344 mm on drawing -> say 346 mm radius -> 2174 mm circumference
Part 620FBM600 from pfeiffer intermediate chamber 690mm diameter, 2168mm
Measurements of parts of CARME frame to the target frame as a reference for target frame position.
Attachments 3+4 show clearance between CARME frame and target frame just below interaction chamber from inside and outside of the ring. Attachment 5 is clearance in same section but above interaction chamber
Attachment 6 - measurement between outer edge of CARME frame and inner edge of target frame - 165mm
Attachment 7 - measurements between outer edge of CARME frame and inner edge of target frame - 455 mm
Attachment 8 - measurement between lower edge of CARME frame and lower edge of target frame - 1245 mm
Attachment 9 - measurement between top of VAT valve and lower edge of target frame - 233mm (left arrow), 216mm (right arrow)
|
Attachment 1: 20220810_095036.jpg
|
|
Attachment 2: 20220810_111638.jpg
|
|
Attachment 3: 20220813_142817.jpg
|
|
Attachment 4: 20220813_142930.jpg
|
|
Attachment 5: 20220813_142945.jpg
|
|
Attachment 6: measurement1.png
|
|
Attachment 7: measurement2.png
|
|
Attachment 8: measurement3.png
|
|
Attachment 9: measurement4.png
|
|
375
|
Fri Aug 12 09:06:57 2022 |
TD, JM | Noise tests pt2 |
10.07 DSSSD bias -100V leakage current -8.65uA
Waveforms and discriminators off, still on clean power supply
stats are ~10k across all fees
aida03 ~ 220
aida04 ~ 205
aida01 ~ 378
aida02 ~ 259
aida05 ~ 40
Grounding removed from all fees except nn bias fee aida04 and aida05. Grounding goes into both adaptor board and fee body
aida01 - no identifiable peak
aida02 - no identifiable peak
aida05 - double peak separated by 70 chn
aida04 - 400 shouldering
aida03 - no identifiable peak
Grounding removed from all fees except nn and pn bias fees aida04, aida01 and aida05. Grounding goes into both adaptor board and fee body
aida01 - 405
aida02 - no identifiable peak
aida05 - double peak
aida04 - 228
aida03 - 417
Grounding removed from all fees except pn bias fee aida01 and aida05. Grounding goes into both adaptor board and fee body
aida01 - 383
aida02 - no identifiable peak
aida05 - 64
aida04 - no identifiable peak
aida03 - no identifiable peak
Grounding on all fees. Grounding goes into adaptor board only
aida01 - no identifiable peak
aida02 - no identifiable peak
aida05 - 63 (wasn't changed)
aida04 - 700
aida03 - 700
Back to previous grounding on all fees
aida01 - 331
aida02 - 226
aida05 - 37
aida04 - 207
aida03 - 214
power cycle
aida01 - 326
aida02 - 238
aida05 - 46
aida04 - 212
aida03 - 195
Waveforms and discriminator back on
Removed test inputs and terminators from all fees, except aida03 (unable to reach) however the lemo is not touching the chamber
See attached waveforms compared to previous elog 32,33
Inserted test input to aida05 only directly from pulser. Width ~ 28
Inserted test input to aida01 only directly from pulser. Width ~ 329
Waveforms and discriminator off -> no impact on aida01 pulser width
Waveforms and discriminator back on
Connected a grounding cable from one of the water cooling pipes to cryring ground
aida01 ~ 205
aida05 ~ 23
aida02 ~ 105
aida03 ~ 71
aida04 ~ 148
Fees with bias attached to them aida01 (+ve) and aida04 (-ve) both show larger widths than there polarity counterpart
waveform screencap 34,35
Fee probe
7.46V left most bar
-6.26V 2nd bar
5.26V 3rd bar
5.26V 4th bar
5.25V 5th bar
Power supply probe
Top -6.05
2nd -6.05
3rd -6.05
4th +6.49
5th -7.78
new Ortec 448 pulser swapped in
aida04 ~ 166
aida05 ~ 21
Attachment 7-11 oscilloscope picture when probing fee relay power supply. Power ON fees connected. Top is attachment 7, bottom (5th) is attachment 11.
Attachment 12-16 old pulser
Attachment 17-18 new pulser
Attachment 19-23 oscilloscope picture when probing fee relay power supply. Power OFF fees connected. Top is attachment 19, bottom (5th) is attachment 23.
Attachment 24-28 oscilloscope picture when probing fee relay power supply. Power ON, fees disconnected. Top is attachment 24, bottom (5th) is attachment 28.
Attachment 29,30 probe-scope setup |
Attachment 1: 33.PNG
|
|
Attachment 2: 32.PNG
|
|
Attachment 3: 31.PNG
|
|
Attachment 4: 30.PNG
|
|
Attachment 5: 34.PNG
|
|
Attachment 6: 35.PNG
|
|
Attachment 7: 20220812_163204.jpg
|
|
Attachment 8: 20220812_163230.jpg
|
|
Attachment 9: 20220812_163319.jpg
|
|
Attachment 10: 20220812_163339.jpg
|
|
Attachment 11: 20220812_163411.jpg
|
|
Attachment 12: 20220812_163526.jpg
|
|
Attachment 13: 20220812_163601.jpg
|
|
Attachment 14: 20220812_163650.jpg
|
|
Attachment 15: 20220812_163705.jpg
|
|
Attachment 16: 20220812_163744.jpg
|
|
Attachment 17: 20220812_165322.jpg
|
|
Attachment 18: 20220812_165330.jpg
|
|
Attachment 19: 20220812_165619.jpg
|
|
Attachment 20: 20220812_165639.jpg
|
|
Attachment 21: 20220812_165652.jpg
|
|
Attachment 22: 20220812_165704.jpg
|
|
Attachment 23: 20220812_165717.jpg
|
|
Attachment 24: 20220812_165812.jpg
|
|
Attachment 25: 20220812_165839.jpg
|
|
Attachment 26: 20220812_165849.jpg
|
|
Attachment 27: 20220812_165900.jpg
|
|
Attachment 28: 20220812_165911.jpg
|
|
Attachment 29: 20220812_170337.jpg
|
|
Attachment 30: 20220813_113236.jpg
|
|
374
|
Thu Aug 11 09:25:13 2022 |
JM, TD | Noise tests |
Moved pneumatic motor in and out several revolutions
Biased detector up to 100 V, I ~ 12.4 uA ambient temp c. 21 deg C
Moving detector in increased leakge current to ~14.5 uA, moving detector out dropped leakage current back to ~12.4 uA
aida05 - pulser width 21 chn (slightly lower than yesterday)
aida01 - pulser width 124 chn
aida02 - double peak (is not currently grounded)
aida03 - 183 chn
aida04 - 212 chn
Attachment 1 1.8.L spectra p+n FEE64s
Attachment 2 per FEE64 rate spectra
Attachment 3 aida05 1.8.L, 1.8.W and rate spectra
Attachment 4 FEE64 temperatures - OK
Attachment 5 per FEE64 ADC data event stats - rates for aida01-aida04 significabtly lower than yesterday presumably due to detector bias -100V
Attachments 6 & 7 per FEE64 1.8.W spectra
Attachment 8 1.8.L spectra for n+n FEE64s
regrounded aida02 back to cryring ground - required removing star configuration of all fee grounds and making a new
aida05 - pulser width 30 chn
aida01 - pulser width 300 chn
aida02 - 220 -> double peak gone
aida03- 190
aida04 - 200
Perhaps star configuration isn't as tight as before - re tighten star section
Re tightened section -> no impact on pulser widths
Moved detector in using pneumatic -> no impact on pulser widths
Added a grounding cable from the 19 inch rack to the same CRYRING ground that the fees are connected too. Stats around 2-3k, slow comparator = 0x64, bias=100v, I=12.18 uA
aida01 ~ 260
aida02 ~ 177
aida03 ~ 178
aida04 ~ 187
aida05 ~28
Removed CRYRING ground. Fees and rack connected together still.
no change
Bias supply module is floating
ion pumps off, gauge off -> no change in waveforms, leakage current dropped to 8.7 uA
aida01 ~ 255
aida02 ~ 176
aida05 ~ 26
Removed all vacuum systems and motors power from 19 inch rack power. Connected to the rack are now: NIM bin, network switch, 2 inputs to mains relay, raspberrypi, pi dlink usb connector, VME crack
aida01 ~ 220
aida02 ~ 154
aida05 ~ 27
aida03 ~ 181
aida04 ~ 193
Moved power cable for 19 inch rack to CRYRING 'clean' power source
aida03 ~ 211
aida04 ~ 212
aida01 ~ 317
aida02 ~ 215
aida05 ~ 29
Reconnected CRYRING ground
aida01 ~ 301
aida02 ~ 203
aida05 ~ 28
aida03 ~ 181
aida04 ~ 191
Current status 17:00:
All CARME vacuum equipment is off (turbo, SIP power modules, NEG power modules, gauges) - roughing pump is on but is valved off from the turbo which is valved off from chamber
Motors and motors pi are off. Detectors are fully out and the pin has been reinserted.
Bias ON 100V, 8.5 uA. Pulser ON, all equipment connected to 19 inch rack power ON. Jumper is on link 1 for nn bias adaptor card only. |
Attachment 1: 19.PNG
|
|
Attachment 2: 18.PNG
|
|
Attachment 3: 17.PNG
|
|
Attachment 4: 16.PNG
|
|
Attachment 5: 15.PNG
|
|
Attachment 6: 14.PNG
|
|
Attachment 7: 13.PNG
|
|
Attachment 8: 12.PNG
|
|
Attachment 9: 11.PNG
|
|
Attachment 10: 10.PNG
|
|
Attachment 11: 9.PNG
|
|
Attachment 12: 8.PNG
|
|
Attachment 13: 7.PNG
|
|
Attachment 14: 6.PNG
|
|
373
|
Wed Aug 10 10:33:03 2022 |
TD, ML, JM, NP | Meeting at YR09 to discuss options for CARME operations to February 2022 |
Initial discussions of proposals to provide open CARME chamber and provide access to near and far detector chambers
We assume that we need to complete all work installing and testing DSSSDs and re-assembling CARME by Feb 2022 in readiness for bakeout.
Proposal #1
Disconnect far detector chamber section from mid-chamber section
Disconnect DN350CF flange from near detector chamber
Crane to support near detector chamber and mid-chamber sections
Remove feet of near detector chamber and mid-chamber sections
Install cradle(s) beneath near detector chamber and mid-chamber sections
Use for fork lift to remove cradle(s) + near detector chamber and mid-chamber sections - ML confirms that an appropriate fork lift is available
Proposal #2 (suggested by ML)
Design and fabricate table plus rails to extend transverse rails of CARME chamber support assembly
The CARME chamber support assembly comprises the 'feet' plus two-stage welded Al frame
Disconnect far detector chamber section from downstream section
Disconnect DN350CF flange from near detector chamber
Slide CARME chamber plus upper stage of Al support frame transversely onto table
Use crane to move CARME chamber to work area
Other discussions
Discussed workarounds to enable install of FEE64s (outside ring, upper feedthrough) which currently conflicts with longitudinal beam of gas jet target support frame.
To Do
Edinburgh
- investigate feasibility of proposal #2
- investigate how much space is required to install FEE64s (outside ring, upper feedthrough)
e.g. by extension of transverse beams of gas jet target support frame, move current assembly outwards from ring c. 20-25cm
GSI
- investigate feasibility of gas jet target support frame changes outlined above e.g. movement of gas/water services, purchase of longer cross beams etc.
Suggest meeting (online) to discuss results of these investigations and continue discussions w/c August 22 |
Attachment 1: 20220810_111638.jpg
|
|
Attachment 2: 20220810_095036.jpg
|
|
Attachment 3: 20220810_095150.jpg
|
|
Attachment 4: 20220810_095159.jpg
|
|
372
|
Wed Aug 10 10:30:04 2022 |
TD, JM | Wednesday 10 August |
09.45 Found CARME electrical power off
Reset power switch of distribution box to on
Reset fuse of 19" rack extender board
One of the SIP power modules may be faulty - others switch on OK
14.00 Checked coolant water flow OK
Transferred Netgear 24 port Gbit switch from AIDA (S4) to CARME (YR09)
Re-connected FEE64 RJ45 network cables
Re-connected nnrpi1 RJ45 network cable
Re-connected RJ45 network cable (yellow) to bottom left connector of YR09 patch panel
@CRYRING user control rooms
Connected RJ45 netwwork cable (yellow) from bottom left connector of patch panel (labelled "1-2->YR10" and "3-4->YR10") to CARME server carme-gsi p4p1 port
Retarted HttpdTcl servers on nnrpi1 and carme-gsi.
Normal 5x FEE64 power up
All system wide checks OK
FEE64 temperatures OK - attachment 1
ASIC settings
slow comparator 0x64
Pulser running through aida01-05
Detectors not biased -> lot of noise aida01-04
AIDA stats - attachments 3 & 4
aida05 pulser width ~150 channels FWHM, normal settings, shaping time 8us - attachment 4
See attachments 2-4
Tried changing ADC control register to 0xf in local controls tab -> stops waveform producing data -> no effect on pulser width (slight increase to 160 chn)
fast comparator blanked off (enable disable mask in discriminator control) -> no effect on pulser width
aida05 was ungrounded -> grounded aida05 to the same ground as aida01-04 -> pulser width reduced to 46 chns
Set ADC control register back to 0x0 -> waveforms back -> pulser width reduced to 24 chns
Changed shaping time from 8ms to 1ms -> pulser width increased to 67 chns -> changed back to 8ms shaping time -> returns to 24 chn width
Changed shaping time from 8ms to 2ms -> pulser width increased to 52 chns
Changed shaping time from 2ms to 4ms -> pulser width decreased to 33 chns |
Attachment 1: 4.PNG
|
|
Attachment 2: 3.PNG
|
|
Attachment 3: 2.png
|
|
Attachment 4: 1.png
|
|
371
|
Tue Aug 9 16:07:33 2022 |
JM, TD | Anydesk setup command line |
To log into appc218 desktop:
same login and password as to carme@atppc022.gsi.de
To start anydesk
1. make sure a monitor is plugged into the carme computer!
2. ssh into npg@carme-gsi
3. su terminal
4. killall anydesk
5. systemctl start anydesk
returns an anydesk backend, service and tray
Anydesk is now accesible
|
370
|
Tue Aug 9 13:44:16 2022 |
TD, JM | Tuesday 9 August |
14.00 CARME vacuum c. 1.8e-10 mbar - see attachment 1
open water inlet/outlet taps
NIM bin (MACBs, Ortec 480 Pulser, CAEN N1419ET) switched ON
USB-controlled ac mains relay switched ON - water & interlock LEDs active
Require VETAR2 (previous returned to Nik Kurz) and network switch (currently with AIDA@DESPEC)
Turbo pump ON - normal operation
Ion pumps ON - normal operation
16.30 ML informs us that ambient temperature of CRYRING cave has been increased from c. 16 deg C to c. 21 deg C and that this is likely the cause of the change in vacuum
pressure in YR09 and the rest of the storage ring.
obtained VETAR2 and crate
anydesk restarted |
Attachment 1: Capture_1.PNG
|
|
Attachment 2: 20220809_163450.jpg
|
|
Attachment 3: 20220809_135840.jpg
|
|
Attachment 4: 20220809_135833.jpg
|
|
Attachment 5: 20220809_135801.jpg
|
|
369
|
Thu Mar 3 15:14:21 2022 |
JM, OH | CARME packing up |
Pulser walkthrough taken for motor position 23.1mm and motor position 67.43mm.
Points taken at 30,50,70,90,110,130,150,170,190,210,230 for 23.1mm position and 230,210,190,170,150,130,110 for 67.43mm position.
Motors moved to fully OUT position and pin put back in. Current positions before closing program 19.1mm and 19.57mm.
Fee power OFF. Fee power module switched OFF and AC mains cable unplugged.
Bias suppply OFF. NIM crate for bias module switched OFF.
Water supply OFF
CARME ion gauge turned back ON |
368
|
Wed Mar 2 22:29:48 2022 |
CB, SD | Last shift with beam |
22:30 Shift handed over. DAQ OK. Target OK. Beam OK.
00:22 Temperatures OK
Stats OK
Good events OK
Scalers OK
System wide checks as usual - WR fails by 68 ns
Leakage current OK
Target OK
Motors OK
Pressure in ring 5E-10 mbar OK
Beam OK
02:26 - All OK as above
WR now fails by 80
Data rate ~800 KB/s
03:40 All OK as above
Data rate back to ~ 1900 KB/s
04:20 All OK as above
Started compressing files in WS3, monitor 1
Noticed bias is not as stable as it has been so far, but no reason for concern. See attach.
04:30 Saw a spike in counts similar to beam halo. See attachment. Did not see it before. No effects on leakage current. To be checked offline.
06:00 Could not see any more spikes
All other checks OK as before.
Data rate back to ~1000 Kb/s
06:32 Requested beam off for end of beamtime.
Left DAQ on - run R88
Turning off jet target as per Nikos instructions
Final cleanup + pulse walkthrough to be done by JM and OH later today.
Target nozzle removed - pressure in S1 now 1.93E-8 mbar
Valves closed
Entering cave to switch compressor off and change N backing pressure to 10 bar. |
Attachment 1: Stats_2022-03-03_00-23-00.png
|
|
Attachment 2: Scalers_2022-03-03_00-24-03.png
|
|
Attachment 3: GoodEvents_2022-03-03_00-23-38.png
|
|
Attachment 4: Ruth_2022-03-03_00-25-48.png
|
|
Attachment 5: Leakage_022-03-03_00-26-44.png
|
|
Attachment 6: Leakage_2022-03-03_04-22-52.png
|
|
Attachment 7: ExTotal_2022-03-03_04-27-38.png
|
|
Attachment 8: EyTotal_2022-03-03_04-27-10.png
|
|
Attachment 9: Spike_2022-03-03_04-29-57.png
|
|
367
|
Wed Mar 2 21:16:06 2022 |
JM, PM | Current status |
22:16
Temperature setpoint = 113k
Not observed noise burst again
Target density OK, S1 between 2E-8 and 3E-8mbar, E4 between 1E-9 and 1E-8 mbar.
Detector moving well, potentionmer between 23.1 mm and 67.43 mm, move in/move out counters the same. Correlation scalar on aida1
Ruthrford peak varies between 10^3 and 10^2 peak height depending on detector position, noise rate also varies with position.
Beam intensity ~1E8
Merger OK
Data rate ~ 1200 kb/sec
Temperatures OK
System checks OK - wr fails
Base Current Difference
aida01 fault 0xab60 : 0xaba4 : 68
aida02 fault 0x88d3 : 0x8915 : 66
aida03 fault 0x23aa : 0x23ee : 68
aida04 fault 0x15d7 : 0x161b : 68
aida05 fault 0x2f03 : 0x2f20 : 29
White Rabbit error counter test result: Passed 0, Failed 5
Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR
Ring pressure oscillation is 30 minute cycle on all areas of the ring. |
366
|
Wed Mar 2 19:25:25 2022 |
JM, PM | Current status |
20:25
Observed short burst of low energy noise on aida1, lasted <30s (attachment 8).
Temperature setpoint = 113 K
Target density OK S1 vaies between 2E-8 and 3E-8 mbar, E4 between 1E-9 and 1E-8 mbar.
Detector moving well between 23.1 mm and 67.43mm, move in and move out counters the same. Correlation scalar on aida1.
Rutherford peak varies between 10^3 and 10^2 depending on detector position, noise peak varies with detector position -> lower further in.
Beam intensity ~1E8
Merger OK
Data rate ~ 1900 Kb/sec
Leakage current = 3.208 uA
Temperatures OK
System checks - wr fails
Base Current Difference
aida01 fault 0xab60 : 0xaba4 : 68
aida02 fault 0x88d3 : 0x8915 : 66
aida03 fault 0x23aa : 0x23ee : 68
aida04 fault 0x15d7 : 0x161b : 68
aida05 fault 0x2f03 : 0x2f20 : 29
White Rabbit error counter test result: Passed 0, Failed 5
Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR
Avg ring pressure = 5.5 E-10 mbar. Ring pressure tends to oosciilate between ~4 and 5.5 E-10 mbar
21:21
Burst of noise not observed again
Temperature setpoint = 113 K
Target density OK S1 vaies between 2E-8 and 3E-8 mbar, E4 between 1E-9 and 1E-8 mbar.
Detector moving well between 23.1 mm and 67.43mm, move in and move out counters the same. Correlation scalar on aida1.
Rutherford peak varies between 10^3 and 10^2 depending on detector position, noise peak varies with detector position -> lower further in.
Beam intensity ~1E8
Merger OK
Data rate ~ 1900 Kb/sec
Leakage current = 3.206 uA
Temperatures OK
System checks - wr fails
Base Current Difference
aida01 fault 0xab60 : 0xaba4 : 68
aida02 fault 0x88d3 : 0x8915 : 66
aida03 fault 0x23aa : 0x23ee : 68
aida04 fault 0x15d7 : 0x161b : 68
aida05 fault 0x2f03 : 0x2f20 : 29
White Rabbit error counter test result: Passed 0, Failed 5
Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR
Avg ring pressure 5.4E-10 mbar |
Attachment 1: 2022-03-02_20-37-20-rateIN.png
|
|
Attachment 2: 2022-03-02_20-36-42-rateOUT.png
|
|
Attachment 3: 2022-03-02_20-38-16-XYtotal.png
|
|
Attachment 4: 2022-03-02_20-39-02-Extotal.png
|
|
Attachment 5: 2022-03-02_20-42-28-ADCitems.png
|
|
Attachment 6: 2022-03-02_20-42-02-goodEvents.png
|
|
Attachment 7: 2022-03-02_20-41-36-correlation8.png
|
|
Attachment 8: 2022-03-02_20-48-27-noise.png
|
|
Attachment 9: 2022-03-02_21-24-40-ExTotalZoom.png
|
|
Attachment 10: 2022-03-02_21-24-07-EYTotalZoom.png
|
|
365
|
Wed Mar 2 17:21:35 2022 |
JM, PM | Current status |
18:21
Temperature setpoint = 113k
Target density OK S1 varies between 2E-8 and 2.5E-8 mbar, E4 between 1E-9 and 5E-9 mbar. (slightly lower than previous check)
Detectors moving well between 23.1mm and 67.43 mm, Move in and move out counters the same. Correlation scalar is on aida1
Rutherford peak varies between 10^3 and 10^2 depending on detector position. No halo peak in rate. Noise peak varies from ~100 peak height when fully in to ~400 peak height when fully out.
Beam intensity ~ 1E8
Merger OK
Data rate ~ 1800 kb/sec
Leakage current ~3.21 uA
Temperature OK
System checks OK - wr fails
Base Current Difference
aida01 fault 0xab60 : 0xaba3 : 67
aida02 fault 0x88d3 : 0x8914 : 65
aida03 fault 0x23aa : 0x23ed : 67
aida04 fault 0x15d7 : 0x161a : 67
aida05 fault 0x2f03 : 0x2f20 : 29
White Rabbit error counter test result: Passed 0, Failed 5
Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR
Avg ring pressure = 4.4 E-10 mbar
19:20
Temperature setpoint = 113k
Target density OK S1 varies between 2E-8 and 3E-8 mbar, E4 between 1E-9 and 1E-8 mbar. (slightly higher than previous check)
Detectors moving well between 23.1mm and 67.43 mm, Move in and move out counters the same. Correlation scalar is on aida1
Rutherford peak varies between 10^3 and 10^2 depending on detector position. No halo peak in rate.
Beam intensity ~ 1E8
Merger OK
Data rate ~ 1800 kb/sec
Leakage current ~3.20 uA
Temperature OK
System checks OK - wr fails
Base Current Difference
aida01 fault 0xab60 : 0xaba4 : 68
aida02 fault 0x88d3 : 0x8915 : 66
aida03 fault 0x23aa : 0x23ee : 68
aida04 fault 0x15d7 : 0x161b : 68
aida05 fault 0x2f03 : 0x2f20 : 29
White Rabbit error counter test result: Passed 0, Failed 5
Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR
Avg ring pressure = 4.3 E-10 mbar
Statistics similar to previous check. |
Attachment 1: 2022-03-02_18-28-37-Exrate.png
|
|
Attachment 2: 2022-03-02_18-29-38-Xytotal.png
|
|
Attachment 3: 2022-03-02_18-30-33-ExTotal.png
|
|
Attachment 4: 2022-03-02_18-32-51-correlation8.png
|
|
Attachment 5: 2022-03-02_18-31-50-GoodEvents.png
|
|
Attachment 6: 2022-03-02_18-31-24-ADCitems.png
|
|
Attachment 7: 2022-03-02_19-48-56-rateIN.png
|
|
Attachment 8: 2022-03-02_19-50-51-rateOUT.png
|
|
364
|
Wed Mar 2 14:15:16 2022 |
JM, PM | Current status |
15:15
Temperature setpoint = 113 K
Target density is OK for now. S1 vaies between 2E-8 and 3.5E-8 mbar. E4 between 1E-9 and 1E-8 mbar.
Detector motors functioning well. Moving between 23.1 mm and 67.43mm. Move in and move out counters the same. Correlation scalar 8 for detectors moving is seen on aida1
Rutherford rate peak varies between 10^3 peak height when fully in and ~10^2 when fully out. No peak from halo seen in rate.
Beam intensity ~ 1E8
Merger OK
Data rate ~ 1700 kb/sec
Run is Run 87 - started 10:34
Data spy restarted at 15:00 so only include Run87 data.
Leakage current = 3.205 uA
Fee Temperatures OK
System checks OK - wr fails
Base Current Difference
aida01 fault 0xab60 : 0xaba3 : 67
aida02 fault 0x88d3 : 0x8914 : 65
aida03 fault 0x23aa : 0x23ed : 67
aida04 fault 0x15d7 : 0x161a : 67
aida05 fault 0x2f03 : 0x2f20 : 29
White Rabbit error counter test result: Passed 0, Failed 5
Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR
Avg ring Pressure = 5.5E-10 mbar
16:45
Temperature setpoint = 113 K
Target density is OK for now. S1 vaies between 2E-8 and 3E-8 mbar. E4 between 1E-9 and 1E-8 mbar.
Detector motors functioning well. Moving between 23.1 mm and 67.43mm. Move in and move out counters the same. Correlation scalar 8 for detectors moving is seen on aida1
Rutherford rate peak varies between 10^3 peak height when fully in and ~10^2 when fully out. No peak from halo seen in rate.
Beam intensity ~ 1E8
Merger OK
Data rate ~ 1300 kb/sec
Temperatures OK
Leakage current ~3.205 uA
System chacks OK - wr fail
Base Current Difference
aida01 fault 0xab60 : 0xaba3 : 67
aida02 fault 0x88d3 : 0x8914 : 65
aida03 fault 0x23aa : 0x23ed : 67
aida04 fault 0x15d7 : 0x161a : 67
aida05 fault 0x2f03 : 0x2f20 : 29
White Rabbit error counter test result: Passed 0, Failed 5
Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR
Avg ring pressure ~ 4.4 E-10 mbar
Statistics similar to previous check.
|
Attachment 1: 2022-03-02_15-23-02-ExRate.png
|
|
Attachment 2: 2022-03-02_15-24-14-XYtotal.png
|
|
Attachment 3: 2022-03-02_15-28-04-correlation8.png
|
|
Attachment 4: 2022-03-02_15-27-13-ADCitems.png
|
|
Attachment 5: 2022-03-02_15-26-34-GoodEvents.png
|
|
Attachment 6: 2022-03-02_16-51-19-aida1-4-stats.png
|
|
363
|
Wed Mar 2 09:38:50 2022 |
Spillmann, Uwe | ion source restarted |
detector position
out: 23.1mm
in: 67.4mm
tape: next file
gas jet S1: 2.0-2.2 ** -8
|
362
|
Wed Mar 2 06:00:49 2022 |
Spillmann, Uwe | waiting |
07:00
Target density is hovering between 2.0e-8 and 2.3e-8 |