AIDA GELINA BRIKEN nToF CRIB ISOLDE CIRCE nTOFCapture DESPEC DTAS EDI_PSA 179Ta CARME StellarModelling DCF K40
  CARME, Page 2 of 38  ELOG logo
Entry  Tue Aug 9 13:44:16 2022, TD, JM, Tuesday 9 August Capture_1.PNG20220809_163450.jpg20220809_135840.jpg20220809_135833.jpg20220809_135801.jpg

14.00 CARME vacuum c. 1.8e-10 mbar - see attachment 1
Entry  Wed Aug 10 10:30:04 2022, TD, JM, Wednesday 10 August 4.PNG3.PNG2.png1.png

09.45 Found CARME electrical power off
Entry  Fri Aug 12 09:06:57 2022, TD, JM, Noise tests pt2 30x

10.07 DSSSD bias -100V leakage current -8.65uA
Entry  Sun Aug 14 06:50:58 2022, TD, JM, Sunday 14 August 10x

07.51 DAQ continues files 22Feb/R97
Entry  Mon Jan 29 15:32:08 2024, TD, JM, Monday 29 January contd. 16x

Test + and test - daisy chain cabling extended from DSSSD #1 to #2 to DSSSD #1 to DSSSD #4
 terminated by 50 Ohm at end of daisy chain
Entry  Wed Apr 23 11:08:34 2025, TD, JM, Background Run 11x

12.08 DSSSD bias & leakage current - attachments 1-2
       CAEN logger program crashed sometime this am - now restarted
Entry  Thu Apr 18 13:34:06 2024, TD, JG, Thursday 18 April 13x

14.30 
Entry  Sat May 24 08:28:14 2025, TD, EM, Background runs R13-R16 29x
Summary: Target not running. Some background files written to disk, some with beam some without. 

09.27 DAQ going
Entry  Wed Feb 23 21:27:29 2022, TD, CN, check  11x
Motor position ok

Data rate: 1030kbytes/s

T ok
Entry  Fri May 10 21:23:35 2024, TD, CB, JM, 20Ne7+ test beam 18x
22.00

Carsten Brandau stopping e-target experiment Unattended beam overnight. No operators watching.

10^5 (nominal) 20Ne7+ at 380keV/u beam available
Entry  Tue Jun 17 23:51:58 2025, TD, CB, JG, Run 96 - continued 23x

00.31 DAQ continues data file G22-205/R96_27
      
Entry  Wed Nov 10 14:41:12 2021, TD RSS + CB (remotely), Wrapping CARME in heating tape 
ML reports that baking of the upstream section (gas jet interaction chamber plus e-target) has commenced - initially to 50 deg C to check thermocouples.
If OK baking 
temperature will be ramped slowly (which I think means c. 10 deg C per h) to limits established for specific components by the e-target.
Entry  Tue Aug 10 12:22:56 2021, TD RSS, Tuesday 10 August  20210810_115555.jpg20210810_115607.jpg20210810_115617.jpg

Attachments 1-3 - 2 of 4 SAES D2000 installed at same ports used for STFC DL XHV tests
Entry  Wed Aug 11 12:30:40 2021, TD RSS, Wednesday 11 August  8x

Attachments 1 & 2 - 2x SAES D-2000 re-fitted to side ports
                    all 4x SAES D-2000 now re-fitted to chamber
Entry  Thu Aug 12 09:16:42 2021, TD RSS, Thursday 12 August 13x

Attachment 1 - IE414/514 feedthrough wiring diagram
Entry  Thu Oct 14 15:54:47 2021, TD RSS, Thursday 14 October IMG_5056.JPG.jpgIMG_5053.JPG
Attachment 1 - what is function of the 4x c/s holes indicated?

Entry  Sun Oct 17 11:48:48 2021, TD RSS, Sunday 17 October 14x
Looking downstream the RHS motor assembly has  loose green/yellow (gnd?) see attachment 1

Install blank base plate to trial install in downstream CARME chamber. The rod is just through all of the mounting holes - see attachments 2 & 3
Entry  Sun Feb 27 12:49:44 2022, TD RSS, Sunday 27 February 07:00-15:00 6x

13.50 Cryostat set point 120K
Entry  Mon Jan 29 08:52:07 2024, TD JM, Monday 29 January 32x

09.50 CARME IE514 5.1e-11 mbar ( ambient temperature 16 deg C )
Entry  Wed Feb 23 18:58:12 2022, TD CN, Wednesday 23 February 15.00-23.00 shift 6x


Assorted screenshots of the CARME, CRYRING and gas jet target control systems in operation
ELOG V3.1.4-unknown