Entry  Thu Jan 24 10:26:09 2019, VP, TD, CA, NH, Start up Merger merger_screenshot.png
1) Click on item "Merger MK2" in top panel
   This kills any existing Merger existence and loads a new copy.
Entry  Thu Apr 25 22:45:30 2024, Tom, Muneerah, Norah, Over night shift AIDA 26 Fri 9x
 

01:02

Aida04 was zero not reading had to call Tom to fix it.
Entry  Thu Mar 12 15:25:08 2020, Thursday, March 12 16:00 to midnight shift, CA, DK, LS 12x
16:30
    All system wide checks passed
    Temps shown in attachment #1 
Entry  Tue Mar 10 08:48:07 2020, TD, OH, NH, LS, CA, DK, PJW, RP, Tuesday 10 March 08.00-16.00 41x

c 09.00 aida03 drops Histogram enable at DAQ GO
        NH determined this was due to Options DB corruption and fixed
Entry  Sun Mar 14 17:35:05 2021, TD, OH, Sunday 14 March - Pulser Walkthrough 52.png51.png40.pngScreenshot_2021-03-14_Spectrum_Browser_aidas-gsi(3).png
18.01 Pulser walkthrough at end of run S452 file S452/R57_0 ... R57_28

      ASIC settings 2019Dec19-16.19.51
Entry  Thu May 6 16:59:20 2021, TD, OH, Wednesday 6 May 13x

18.09 NH has installed jumpers LK2-4 for *all* adaptor PCBs
Entry  Thu May 12 10:10:51 2022, TD, OH, AIDA data throughput 15x

Attachment 1 - top aida06
Entry  Thu Dec 19 13:06:02 2019, TD, NH, AIDA@DESPEC To Do list 
FAIR phase 0 DESPEC experiments are scheduled March-May 2020 see

https://www.gsi.de/en/work/organisation/scientific_boards/user/beam_time.htm
Entry  Thu Feb 20 16:35:55 2020, TD, NH, Thu 20 February 2020 12x
DSSD#3 (most downstream) was replaced

Noted that old DSSD wafer was not sitting properly - detached - in PCB, PCB flexing and/or inadequate epoxy gluing
Entry  Fri Feb 21 09:39:38 2020, TD, NH, Friday 21 February 7x

Attachment 1 - layout 2 - stat spectra
Attachment 2 - detector biases & leakage currents - OK
Entry  Fri Feb 21 12:16:06 2020, TD, NH, Report - high - unusual FEE64 crash during startup ttyUSB4
 
Entry  Mon May 9 08:36:23 2022, TD, NH, Monday 9 May 31x

09.32 Re-start AIDA
Entry  Sat Apr 20 17:52:27 2024, TD, NH, Saturday 20 April contd. Screenshot_from_2024-04-20_18-51-53.pngScreenshot_from_2024-04-20_21-52-08.pngScreenshot_from_2024-04-21_10-52-19.png

Note 40kBq 60Co source placed on bottom surface of snout centred on bPlas detectors and DSSSDs
Entry  Tue Apr 30 09:06:45 2024, TD, NH, Tuesday 30 April 7x

10.00 DSSSD bias & leakasge current - Grafana - attachments 1-2
Entry  Thu May 2 00:16:22 2024, TD, NH, Thursday 2 May Screenshot_from_2024-05-02_13-21-07.pngScreenshot_from_2024-05-02_13-22-16.pngScreenshot_from_2024-05-02_13-22-43.png
01.15 FEE64 temperatures OK
      ADC data item stats
       aida02 & aida04 no data
Entry  Tue Jun 11 10:03:48 2024, TD, NH, 08:00-16:00 Tuesday 10 June 23x
10.57 DSSSD bias & leakage current OK - attachment 1

      FEE64 temps OK - attachment 2
Entry  Wed Dec 11 14:46:21 2024, TD, JB, CC, MP, Mounting of AIDA, bPlast and BB7 for test 20241211_161837.jpg20241211_161833.jpg20241211_161830.jpg20241211_161820.jpg
Snout assembled for detector position and timing test.

The planned test is to use a BGO with a 22-Na source and the implantation stack to characterise the timing and position of back-to-back 511 keV
events (proxy for mock beta decay event).
Entry  Tue Apr 9 13:22:41 2024, TD, JB, CC, Continued of 9th of April Screenshot_from_2024-04-09_14-22-20.pngScreenshot_from_2024-04-09_14-22-16.pngScreenshot_from_2024-04-09_17-18-45.png
Disconnected test - pulser from aida08. Noise conditions improved slightly, probably less pulser events. (Attachments 1 & 2).

We recabled the dew point sensor and flow sensor for the water supply interlock to the FEE64s. When we tried to power up the interlock box, the
PSU (charging block) LED was blinking indicating an internal power instability. We then plugged each of the sensors out and into the interlock box and
Entry  Tue Apr 16 10:49:37 2024, TD, JB, CC, Tuesday 16 April 24x
11.45 DSSSD bias & leakage current OK - attachments 9-10

      FEE64 temps OK - attachment 8
Entry  Mon Apr 8 10:23:16 2024, TD, JB, AIDA Noise optimisation 30x
In the morning we reinstalled FEE64s aida06, aida13, aida07. The rails of the FEE64 boards were found to be incorrect (that of the rails that were used
in CARME).

After replacing the rails on the abovementioned FEE64s we powercycled AIDA and found the following results in the histograms and statistics.