AIDA GELINA BRIKEN nToF CRIB ISOLDE CIRCE nTOFCapture DESPEC DTAS EDI_PSA 179Ta CARME StellarModelling DCF K40
  DESPEC, Page 26 of 33  ELOG logo
Entry  Sat Mar 6 06:52:06 2021, OH, LS, Saturday 6th March 30x
07:50 Pulser settings - attachment 1

07:52 System wide checks all ok
Entry  Fri Mar 5 23:45:41 2021, TD, S452 information S452__Shifts_checkpoints_v1_(1).pdfS452_Shift_Plan.pdf

aida-gsi Anydesk 897170655
Entry  Fri Mar 5 23:21:59 2021, CA, OH, March 6th 2021 00:00 - 08:00 shift 18x
00:23 206Hg setting - No beam in AIDA yet

      AIDA not writing to file
Entry  Fri Mar 5 22:53:05 2021, OH, Thursday 5th of March 
There has been no beam for much of the day. It was lost at 13:00 and not regained until 22:00 German time.
At some point during the day the White Rabbit timestamp cable was unplugged from AIDA. This caused it lose correlation with the rest of the system.
At first a reset and setup was performed to try and regain correlation but that did not prove successful.
Entry  Thu Mar 4 11:16:13 2021, OH, Further cases of possible database issues ErrorMessage.PNGerrorMessage.txt
During testing for the S452 experiment yesterday we were running through the steps of a restart.

The FEEs were not powercycled at this stage but the MIDAS server was restarted.
Entry  Thu Mar 4 11:03:04 2021, NH, S452 Operating Notes 
Please note the following "quirks" currently present in S452 that are important

[B]a) Strips/Thresholds[/B]
Entry  Thu Mar 4 10:13:48 2021, OH, AnyDesk connection instuction Connecting_to_AnyDesk_at_GSI_for_AIDA.pdf
 
Entry  Wed Mar 3 15:02:19 2021, CA, MBS command line startup 
/MIDAS/Linux/startup/datarelaymbs 
Entry  Mon Feb 22 18:18:47 2021, NH, AIDA Rates/Situation  8x
The ground cable (LEMO00) in aida07 was disconnected, it has been reinserted and the waveforms now look good.

(All data below taken with thresholds 0xA for all)
    Reply  Sat Feb 27 12:47:57 2021, NH, AIDA Rates/Situation  10x
Updated figures as of 27 Feb 2021 - S4 is now closed for experiments but may open on Wednesday.
Snout has been grounded and all FEEs are seemingly grounded properly.
Entry  Fri Feb 12 17:31:04 2021, NH, Background run 
A background run has been started to check the AIDA channels.
It is in /TapeData/BGFEB21
Entry  Mon Feb 1 15:19:07 2021, NH, AIDA Status Feb 2021 
01.02.2021 - Four new sets of PSU, HDMI & Ethernet cables were ran from the DESPEC rack to the AIDA frame, in preparation for 4 new FEE64s.
A new 7 port USB hub has also been placed for the last few serial connections.
    Reply  Sat Feb 6 10:45:55 2021, NH, AIDA Status Feb 2021 pregrd050222_E.pngpregrd050221_O.pngpostgrd050221_O.pngpostgrd050221_E.png
> 01.02.2021 - Four new sets of PSU, HDMI & Ethernet cables were ran from the DESPEC rack to the AIDA frame, in preparation for 4 new FEE64s.
> A new 7 port USB hub has also been placed for the last few serial connections.
Entry  Tue Nov 24 13:10:42 2020, NH, Waveforms with frame grounded waves_odd.pngwaves_even.png20201124_164416.jpg20201124_164413.jpg
AIDA Frame is now grounded to platform (will add photo later)

Waveforms improved from April (c.f. https://elog.ph.ed.ac.uk/DESPEC/125) but comparison not truly fair (other systems may be offline now)
Entry  Tue Oct 6 15:48:06 2020, NH, nnrpi kernel update 
FYI:

The SD card on nnrpi2 was full again causing issues, the log files in /var/log (kern.log & messages) were *huge* and filled with the same error many times
Entry  Tue Jul 7 08:50:27 2020, NH, nnrpi2 does not boot 
MIDAS was not working properly on nnrpi2 - power cycled. Now system does not boot properly, it gets stuck trying to login claiming "PAM critical error"
- probable file system corruption?
Will investigate if SD card can be checked and repaired
Entry  Tue Jul 7 08:04:57 2020, NH, Power failure in Messhuette 
Last week there was a power failure in Messhuette (FRS control room) - AIDA-3 and MBS were affected.
AIDA itself and S4 equipment is on UPS and was not shut down.
Entry  Wed Mar 18 18:02:37 2020, PJCS, New Firmware loaded in all FEE64s 
Loaded new frimware into all 12 FEE64s in the system. IWR_Dec19_4.bin 

Booted ok after power-cycle.  

RESET/SETUP/GO all fine.
Entry  Mon Mar 16 07:25:24 2020, NH, End of Experiment 
08:25 Beam Stopped, end of S480

Tape Server stopped at R12_2097

Started again in no storage mode
Entry  Mon Mar 16 05:08:12 2020, TA, AM, ES, Regular entry 6x

02:14 all system wide checks okay 
      FEE temperatures okay 
      leakage currents okay and recorded to spreadsheet 
      good event stats
okay 
      merger running at around 3M items/sec
      tape service running at 26MB/sec

 

All attached to this entry with screenshots to 1-7
Entry  Mon Mar 16 01:07:29 2020, TA, AM, ES, Regular entry 13x

02:14 all system wide checks okay 
      FEE temperatures okay 
      leakage currents okay and recorded to spreadsheet 
      good event stats
okay 
      merger running at 3M items/sec
      tape service running at 26MB/sec

 

All attached to this entry with screenshots to 1-7
Entry  Sun Mar 15 06:45:00 2020, CA, LS, CG, Sunday 15th March 08:00 - 24:00 35x
ASIC settings 2019Oct31-13.24.23
     slow comparator 0xa