AIDA GELINA BRIKEN nToF CRIB ISOLDE CIRCE nTOFCapture DESPEC DTAS EDI_PSA 179Ta CARME StellarModelling DCF K40
  nToF, Page 3 of 4  ELOG logo
ID Date Author Subject Text Attachments
  40   Sun Oct 9 21:07:02 2022 Nikolay SosninBoron Sample 3 Measurement21:15 Changed from detectors in-beam to
boron-10 sample 3 in-beam (in-beam measurement
got 1000 triggers, 500 of which were dedicated).
 14x 
  38   Sun Oct 9 20:57:54 2022 Nikolay SosninDetectors in-beam, no targetAfter long beam-off, changed detector mount
by moving it 2 cm forward into the beam,
after removing LiF 3.
 12x 
  37   Sat Oct 8 11:57:36 2022 Nikolay SosninChange of Detector-Target SeparationIncreased detector-target (LiF 3) separation
by 1.5 cm. Had to replace the support strut
under the detectors due to the previous one
 17x 
  36   Sat Oct 8 11:56:49 2022 Nikolay SosninSilicon Strips of ConcernDEED 6 and 9 no signal throughout the campaign.

EFED 27 high oscillations in baseline
throughout the campaign.
  
  35   Fri Oct 7 20:04:30 2022 Nikolay SosninChange of Pulser SettingThe pulser has been switched from triggering
from beam trigger to running at a 10 Hz rate,
and will now appear in random places in signal
  
  34   Fri Oct 7 17:17:08 2022 Nikolay SosninChange of Pulser DelayStill using pulser triggered from protons.
12 microsecond delay from protons to gamma-flash
+ 11 microsecond delay from gamma flash to
  
  33   Thu Oct 6 14:58:29 2022 Nikolay SosninChange of Pulser DelayStill using pulser triggered from protons.
12 microsecond delay from protons to gamma-flash
+ 1.75 microsecond delay from gamma flash
  
  32   Thu Oct 6 11:45:39 2022 Nikolay SosninAssembly Setup Days 1 and 2Photographs of assembly up to the first
pulser test with LiF3.
 25x 
  31   Thu Oct 6 11:07:47 2022 Nikolay SosninChange of sample/Addition of pulserChanged from LiF 4 to LiF 3 after running
with LiF 4 overnight (run title on n_TOF
DAQ was not updated to refelect the change,
  
  30   Thu Oct 6 09:34:30 2022 ClaudiaTest setup Photos and cablingSetup 5.10.22

dE 20 um (ID 2837-33) and 494 um
double sided (ID 1194-9)
 20221005_174742.jpg20221005_174759.jpg20221005_184752.jpg20221005_184801.jpg20221005_174742.jpg 
  29   Thu Oct 6 09:28:33 2022 ClaudiaItems to be replaced for next runMissing/broken items identified

- long rails for fixing chamber
on support
 20221005_122654.pdf 
  28   Thu May 26 09:20:32 2022 Nick / ClaudiaInventory of Items at n_TOF for (n,cp) experimentsDetectors currently in the boxes:

2x MSL type W1(SS)-50, 2902-4 (53um),
2940-6 (55um)
 rn_image_picker_lib_temp_e85b2099-7b73-4246-843d-6351800c2709.jpgrn_image_picker_lib_temp_3ff5f819-1e0c-4278-bbbc-01038a0a0c94.jpg 
  27   Fri Sep 9 10:01:18 2016 SarahLeakage currents 09.09.16EFED 0.04uA
DEED 0.05uA
PHDI 0.13uA
  
  26   Fri Sep 2 12:03:40 2016 SarahSPD card changed14 bit card replaced for DEED 9-12, to
prevent issue with pulser arriving before
gamma flash.
  
  25   Fri Sep 2 12:01:57 2016 SarahLeakage currents 02.09.16E 0.03
dE 0.05
PHDI 0.12
  
  24   Tue Aug 23 15:34:38 2016 ClaudiaLeakage currents 23.08.16 E = 0.03uA
dE = 0.04uA
PHDI = 0.09uA
  
  23   Wed Aug 17 15:10:05 2016 SarahLeakage currents 12.08.16 E = 0.03uA
dE = 0.04uA
PHDI = 0.06uA
  
  22   Thu Aug 11 08:19:37 2016 SarahLeakage currents 10.8.16 At lunchtime (12.45)
E = 0.03 uA
dE  = 0.04 uA
  
  21   Tue Aug 9 15:57:52 2016 SarahLeakage currents 09.08.16 Morning - 9am
E= 0.03uA
dE= 0.04uA
  
  20   Mon Aug 8 11:02:13 2016 SarahLeakage currents 08.08.16 E 0.03 uA
dE 0.04 uA
PHDI 0.05 uA
  