Entry  Thu Oct 31 15:24:35 2019, TD, waveform spectra issues? 12x
 
Entry  Sat Apr 24 04:07:23 2021, ML, ucesb scalers 24-04-2021_5am_ucesb.png
ucesb is back - see attachment
Entry  Fri Apr 16 05:13:09 2021, ML, system wide checks 16x
system wide checks:

Done around 4am.

#################################
Entry  Sun Apr 18 17:27:44 2021, ML-OH, system wide checks 17x
system wide checks at 6pm CET

##################################################

Difference noted from previous shift:
Entry  Sun Apr 18 21:43:01 2021, ML-OH, system wide checks 20x
New System wide check 10pm CET

Clock status test result: Passed 12, Failed 0

Understand status as follows
Entry  Sat Apr 24 01:44:09 2021, ML, system wide checks 7x
system wide checks:

 

 
Entry  Thu May 13 21:21:25 2021, Philippos and Marc, system wide checks 13x
Wide checks at Time 22:22 CET

Clock status test result: Passed 16, Failed 0

Understand status as follows
Entry  Thu May 20 23:26:11 2021, Marc Labiche (ML), system wide checks 9x
System Wide check at begining of the night shift:

-----

Clock status test result: Passed 16, Failed 0
Entry  Fri May 21 02:17:47 2021, ML, system wide checks 9x
Wide check at 3am -

All fine.

Same output as at the beginning of the shift (see my previous entry) and attachments.
Entry  Fri May 21 06:09:28 2021, ML, system wide checks 8x
System  wide checks at 7am CET

All seems fine, nothing significant to report

Slight increase of dead time noted in FEE 3, 5 and 7.
Entry  Wed Mar 11 16:29:17 2020, CA, LS, TD, slow comparator threshold comparison 100keV_stats.png110keVStats.png120kevstats.png80kevDSSD2.png
17.29 good event statistics for slow comparator threshold at 100 keV (0xa) - attachment 1

17.36 good event statistics for slow comparator threshold at 110 keV (0xb) - attachment 2
Entry  Wed Apr 3 15:24:06 2019, CA, TD, report - low inquire save state 030419_1614_savestate.png
On Save/Restore Module Settings Tab - "inquire save state" returns an error. See attachment 1.
Entry  Wed Apr 3 13:29:20 2019, CA, CB, TD, report - high DAQ stall 030419_1416_aida10stats.png030419_1417_aida09stats.png030419_1417_aida11stats.png030419_1418_aida12stats.png030419_1419_aida10_led.png
aida10 runs very fast for a period, then stops.

Have to restart Merger and TapeServer to get it running again.
Entry  Sat Apr 24 03:32:11 2021, ML, quick checks 24-04-2021_430am_GoodEvent.png
Quick system checks after crash

Temperature ok.

Still cannot connect remotely to /ucesb with my gsi web login and password (?).
Entry  Tue Jul 7 08:50:27 2020, NH, nnrpi2 does not boot 
MIDAS was not working properly on nnrpi2 - power cycled. Now system does not boot properly, it gets stuck trying to login claiming "PAM critical error"
- probable file system corruption?
Will investigate if SD card can be checked and repaired
Entry  Fri Apr 9 16:40:36 2021, NH, nnrpi1 Update 
nnrpi1 is now back in S4 with a new SD card and new raspberry pi.
The freezing originally continued (it was the USB system failing) but was fixed with a change to /boot/cmdline.txt coherent_pool=4M
It seems the kernel update to fix one bug introduced another... 
Entry  Tue Mar 30 13:21:50 2021, NH, nnrpi1 (tty logs) 
We had issues with all TTYs reading out on nnrpi1, I have moved them from an apparently faulty USB hub to a new one.
Trying to reboot the pi to detect the new USB ports and it didn't recover from boot
Trying to repair SD card but I get a lot of errors on my laptop, indications are the SD card is completely dead
Entry  Tue Oct 6 15:48:06 2020, NH, nnrpi kernel update 
FYI:

The SD card on nnrpi2 was full again causing issues, the log files in /var/log (kern.log & messages) were *huge* and filled with the same error many times
Entry  Sun Jun 26 23:04:30 2022, Marc, new shift - Monday 27 June 0:00 to 8:00 12x
0:00 Experiment continues to run smoothly. AIDA DAQ has been rather stable. Leakage current has gone up for the past two days in both DSSDs (combination
of high-Z and high-intensity beam and external high temperature).

Stats & Temperatures (VIRTEX,PSU, ASICs)  all ok.
Entry  Fri May 13 02:28:36 2022, ML, beam dropped 
We see less counts in AIDA implant DSSD1 (~1Hz) than at the beginning of the night shift (~10Hz).

This seems to be beam related as the count rates in DESPEC & FRS Scalers seems to have dropped too.