Mon May 6 07:10:26 2024, TD, Monday 6 May
|
08.11 DSSSD#1 bias -120V leakage current -17.0uA
DSSSD#2 bias -120V leakage current -10.5uA
|
Mon Jul 5 14:36:15 2021, NH, Monday 5 July
|
AIDA Unbiased and powered down
All FEEs were running before stoppage. None had rebooted.
Aida05 did not respond until merger was restarted, all other DAQs responded fine.
|
Mon Mar 4 11:35:30 2024, TD, Monday 4 March 11x
|
12.30 FEE64 41:d7:cd ASIC mezzanine u/s, no data, ASIC temperature c. 20 deg C ( ambient ) low
Replaced ASIC mezzanine with new ASIC mezzanine
Installed as aida08
|
Mon Apr 4 09:19:49 2022, TD, Monday 4 April 12x
|
AIDA adaptor PCB config - all LKs removed *except*
aida04 & aida08 LK1
|
Mon Jun 3 07:35:56 2024, TD, Monday 3 June 8x
|
08.28 DSSSD bias & leakage current - attachments 1-2
FEE64 temps OK - attachment 3
|
Mon Apr 29 08:11:16 2024, TD, Monday 29 April 41x
|
08.52 Beam off
|
Mon Jun 27 17:03:09 2022, MA, Monday 27th June 16:00-00:00 6x
|
16:00 Took over the shift from OH no beam yet.
18:00 Still no beam yet.
Statistics, Temperature, Current are checked and attached 1-3 |
Mon Jun 27 06:45:22 2022, OH, Monday 27th June 08:00-16:00   
|
07:45 Spoke to David and the beam has been gone since about 05:30
Reason for the loss of beam is a vacuum issue before the FRS
They are waiting for the experts
|
Mon May 27 17:20:49 2024, TD, Monday 27 May 9x
|
17.50 bPlas cabling, grounds etc connected, power OFF
CC reports that bPlas power is actually ON
|
Mon Apr 26 09:41:21 2021, TD, Monday 26 April    
|
10.42 All system wide checks OK *except*
Base Current Difference
|
Mon May 24 09:57:07 2021, TD, Monday 24 May 19x
|
07.30 Hot HEC channels multiple FEE64s
ASIC check fixes
|
Mon May 23 16:41:45 2022, TD, Monday 23 May  
|
17.40 DAQ stopped
Attachment 1 - DSSSD bias & leakage current
|
Mon Dec 23 10:54:46 2024, TD, Monday 23 December 2024
|
11.54 re-established remote access via AnyDesk - details to follow
|
Mon Apr 12 16:49:40 2021, TD, Monday 22 April - pulser walkthrough & alpha background 6x
|
16.54 Pulser walkthrough
ASIC settings 2019Dec19-16.19.51
|
Sun Apr 21 23:20:45 2024, TD, Monday 22 April 38x
|
00.15 Data link to MBS was closed by MBS - reason(s) unknown
NH restarted MBS receiver and data transfer restarted
|
Mon Jun 21 10:21:50 2021, NH, Monday 21 June 8x
|
11.21 - DAQ continues OK (no storage mode)
11.21 - System checks ok *except*
|
Mon Jun 20 09:26:23 2022, OH, TD, Monday 20 June 11x
|
10:26 MACB HDMI for AIDA03 and AIDA04 reseated.
Merger and DAQ restarted - All happily sending data
Has been running for 10 minutes with no WR errors.
|
Mon Apr 19 16:58:29 2021, LJW, Monday 19th April 16:00-20:00 Shift 6x
|
System Checks @ ~18:10 :
|
Mon Apr 19 13:27:08 2021, RDP, Monday 19th April 12 noon shift  
|
12.30 There appear to be fewer HEC events in aida09 cf. aida11 and upstream DSSSDs
Decrease aida09 asic #1-4 HEC fast comparator 0x2 -> 0x1
|
Mon Apr 19 01:08:40 2021, MA, Monday 19th April 00:00-08:00 8x
|
03:09 General check