Mon Jun 10 15:07:24 2024, Marc, 16:00-00:00 shift Monday 10/06/24 31x
|
Evening shift
5pm full checks:
DSSSD bias & leakage current ok - attachment # 1-2
FEE64 temperatures ok - attachment # 3
ADC
data item stats - attachment # 4
per FEE64 Rate spectra - attachment # 5
Merger ok - Attachement # 6
Tape service - attachement # 7
7pm |
Mon Jun 10 07:15:48 2024, Dan, 08.00-16.00 Monday 10th June 50x
|
8am checks
temps ok - attachment 2
rates ok - attachment 8 |
Sun Jun 9 22:49:56 2024, TD, 23:00-07:00 Monday 9 June 27x
|
23.47 DSSSD bias & leakage current OK - attachments 1-2
|
Sun Jun 9 16:38:30 2024, Norah , Muneerah, JB, 16:0-00:00 9 June 2024   
|
AIDA02 and AIDA06 gave zero attachment 1. After connecting with Tom to fix it, now it works.
17:00
DSSSD bias & leakage current - attachment 2 |
Sun Jun 9 13:32:48 2024, TD, [How To] Recover DAQ when one FEE64 stops sending data - short(er) version
|
1) Select 'No Storage' [workspace 3, display 1, browser tab 'TapeServer']
If already in 'No Storage' mode please ignore this step
|
Sat Jun 8 23:00:52 2024, TD, Sunday 9 June 16x
|
23.58 8.6.24
|
Sat Jun 8 13:22:36 2024, TD, Saturday 8 June 24x
|
14.09 aida02 WR stall
DAQ STOP
|
Fri Jun 7 16:08:37 2024, TD, Friday 7 June 19x
|
05.01 DSSSD bias & leakage current - attachment 1
FEE64 temperatures OK - attachment 2
|
Thu Jun 6 11:13:29 2024, TD, Thursday 6 June 9x
|
12.09 DSSSD bias & leakage current - attachment 1
|
Wed Jun 5 08:34:47 2024, JB. AM, MP, MY, Detector bias tests 14x
|
We are trying to bias each system on its own and in sequence biasing each other detector system to see the interdependence.
HV OK. TEMP OK. Attachment 1 &2. |
Tue Jun 4 15:40:32 2024, NH, Tues 4 Jun 
|
DEGAS Array over Snout again
Situation similar to yesterday, most p+n rates good |
Mon Jun 3 14:40:34 2024, JB, NH, 03 June 2024 10x
|
15:40 reseated FEE64s adaptors of aida01, aida09, aida13 & aida16.
TEMP OK. HV OK. System wide checks all passed excl. aida07 ADC calibration, see attached.
Histograms and rates look same as previous, reseating aida01 seems to have worked. aida16 still unhappy. |
Mon Jun 3 07:35:56 2024, TD, Monday 3 June 8x
|
08.28 DSSSD bias & leakage current - attachments 1-2
FEE64 temps OK - attachment 3
|
Sun Jun 2 01:45:05 2024, TD, Sunday 2 June 25x
|
02.41 DSSSD bias & leakage current - attachment 1
FEE64 temps OK - attachment 2
|
Sat Jun 1 12:48:54 2024, TD, Saturday 1 June 31x
|
13.26 DSSSD bias & leakage current - attachments 1-3
|
Mon May 27 17:20:49 2024, TD, Monday 27 May 9x
|
17.50 bPlas cabling, grounds etc connected, power OFF
CC reports that bPlas power is actually ON
|
Wed May 22 20:10:19 2024, NH, CC, HMA, MV, ZC, KZ, AIDA Snout Remounted 9x
|
AIDA and the Snout have been remounted in S4
I had a check of the AIDA connectors and all looked good, the kapton PCB connectors were sitting comfortably above the interstage and I saw no stressed/loose |
Thu May 9 09:08:26 2024, TD, Current NewMerger install - aida-gsi
|
[npg@aidas-gsi MIDAS]$ ls -l
total 524
drwxrwxr-x. 2 npg npg 156 Mar 17 2021 Aida
|
Tue May 7 13:25:00 2024, JB, CC, NH, HMA, Tuesday 6 May - dismount snout 
|
Status before dismount
14:26 DSSSD#1 119.99 V @ 17.968 uA
DSSSD#2 119.95 V @ 11.212 uA |
Mon May 6 07:10:26 2024, TD, Monday 6 May
|
08.11 DSSSD#1 bias -120V leakage current -17.0uA
DSSSD#2 bias -120V leakage current -10.5uA
|