Tue May 17 06:58:51 2022, OH, Tuesday 17th May 6x
|
08:00 Still have beam for the time being.
Currently on file R4_2049
|
Tue May 17 16:53:20 2022, TD, Tuesday 17 May 16.00-00.00 24x
|
17.35 zero all histograms
zero statistics
|
Tue Mar 16 09:26:47 2021, TD, Tuesday 16 March 6x
|
10.25 DAQ continues OK - no storage
ASIC settings 2019Dec19-16.19.51
|
Tue Apr 16 10:49:37 2024, TD, JB, CC, Tuesday 16 April 24x
|
11.45 DSSSD bias & leakage current OK - attachments 9-10
FEE64 temps OK - attachment 8
|
Tue Jun 15 12:35:48 2021, TD, Tuesday 15 June 8x
|
13.35 DAQ continues OK - no storage mode
All system wide checks OK *except*
|
Tue Apr 13 16:49:41 2021, TD, Tuesday 13 April 8x
|
Rename Options directory
|
Wed Apr 14 09:55:12 2021, TD, Tuesday 13 April
|
Two further comments
|
Tue May 11 07:21:24 2021, OH, Tuesday 11 May  
|
08:20 System wide checks all ok except
Base Current Difference
aida01 fault 0xc10c : 0xc10e : 2
|
Tue May 10 07:41:28 2022, NH TD, Tuesday 10 May 29x
|
8:15am S4 CONTROLLED ACCESS to remove beam plug and check bPlas/AIDA response to beam
Reseat aida05 ribbon cable on adapter board while in...
|
Tue Mar 10 08:48:07 2020, TD, OH, NH, LS, CA, DK, PJW, RP, Tuesday 10 March 08.00-16.00 41x
|
c 09.00 aida03 drops Histogram enable at DAQ GO
NH determined this was due to Options DB corruption and fixed
|
Tue Jun 4 15:40:32 2024, NH, Tues 4 Jun 
|
DEGAS Array over Snout again
Situation similar to yesterday, most p+n rates good |
Tue Jun 28 09:37:19 2022, NH, Tues 28 June 08:00-
|
Experiment over
10:37 - Stop DAQ & Tape
|
Tue Mar 26 10:32:05 2024, NH, JB, Tue 26 March 9x
|
Taken 4 FEE64s from CRYRING (the 4 easiest to access)
41:d8:2b
41:f6:5a
|
Tue May 24 08:35:00 2022, NH, Tue 24 May
|
09:00 - Remove bias of DSSDs and pwoer of FEEs
Remove snout from S4 so DEGAS frame can be mechnically removed
|
Mon Oct 18 14:20:32 2021, OH, Top bar launchers
|
launcher1
TclHttpd @8015 for AIDA
/MIDAS/Linux/startup/HTTPD
|
Mon Mar 11 15:19:44 2024, JB, NH, Priyanka, Michael Armstrong, Helena Albers, To Do: AIDA PCB tests
|
To do:
1) Disconnect ribbon cables from p+n junction FEE64 adaptor PCBs of upstream DSSSD, apply c. 100V bias and check leakage current is zero
i.e. eliminate shorts in PCBs |
Tue Jul 5 08:53:20 2022, TD, To Do
|
In no particular order
|
Thu Jan 9 15:43:17 2025, JB, CC, MP, Timing test platform moved in  
|
16:30 We moved the platform in. The system was still biased and the DAQ was still running.
The system seems to be okay, temperatures OK. Bias voltage OK 685/1 and 685/2. Two FEEs - aida07 and aida12 are receiving a lot of trigger data
items from the fast discriminator. This is already a bit worrying. 685/3 |
Thu Dec 19 09:22:52 2019, TD, Thusrday 19 December 21x
|
10.15 Found DAQ running at c. 1M data items/s
Rate spectra show aida09 1.*.H and one 1x LEC channel of aida08 asic running hot
|
Thu Mar 11 00:38:33 2021, TD, Thusrday 11 March 00.00-08.00 11x
|
01.38 DAQ continues OK - file R22_388
ASIC settings 2019Dec19-16.19.51
|