AIDA GELINA BRIKEN nToF CRIB ISOLDE CIRCE nTOFCapture DESPEC DTAS EDI_PSA 179Ta CARME StellarModelling DCF K40
  ISOLDE, Page 1 of 11  ELOG logo
ID Date Authordown Subject
  3   Sun Feb 19 12:22:16 2017 td, mdVacuum and detector test
------17.2.17 ------------------------------------------------------

closing chamber 
use Leak testing pump
16.50 p = 0.6 mbar --> starting Turbo pump

18.25 Turbo @ 30kRPM, 33W
      Pressure 7.0E-04mbar, backing pressure (leak tester) 0.1mbar
      

------18.2.17 ------------------------------------------------------

12.30 Turbo @ 42kRPM, 25W
      Pressure 6.0E-05mbar, backing pressure (leak tester) 0.1mbar
      FL601 started , temperature 25.8 deg C, set point -5.0 deg C

16.10 Turbo @ 42kRPM, 18W
      Pressure 1.8E-05mbar, backing pressure (leak tester) 9.3E-2mbar
      FL601 set point -5.0 deg C, actual -1.6 deg C

19.2.17

12.20 Turbo @ 42kRPM, 15W
      Pressure 1.5E-05mbar, backing pressure (leak tester) 9.3E-2mbar
      FL601 set point -5.0 deg C, actual -2.0 deg C

------20.2.17 -------------------------------------------------------

14.30 Turbo @ 42kRPM, 14W
      Pressure 1.2E-05mbar, backing pressure (leak tester) 9.2E-2mbar
      FL601 set point -5.0 deg C, actual -2.3 deg C

21.2.17 

10.10 Turbo @ 42kRPM, 14W
      Pressure 1.1E-05mbar, backing pressure (leak tester) 9.2E-2mbar
      FL601 set point -5.0 deg C, actual -2.5 deg C

16.30 Turbo @ 42kRPM, 13W
      Pressure 1.0E-05mbar, backing pressure (leak tester) 9.2E-2mbar
      FL601 set point -5.0 deg C, actual -2.4 deg C

-----22.2.17 -------------------------------------------------------

11.00 Turbo @ 42kRPM, 13W
      Pressure 1.0E-05mbar, backing pressure (leak tester) 9.2E-2mbar
      FL601 set point -5.0 deg C, actual -2.4 deg C

11.02 +/-15V preamp power ON

11.06 Pressure 1.2E-05mbar, FL601 actual temp -1.8 deg C

11.12 Pressure 1.3E-05mbar, FL601 actual temp -1.4 deg C

14.01 Pressure 1.5E-05mbar, FL601 actual temp -1.0 deg C

14.02 +/-15V preamp power OFF

14.15 Pressure 1.1E-05mbar, FL601 actual temp -1.9 deg C

15.27 Pressure 1.0E-05mbar, FL601 actual temp -2.0 deg C

------23.2.17 -----------------------------------------------------

14.40 Pressure 8.6E-06mbar, FL601 actual temp -2.4 deg C

17.10 Pressure 8.1E-06mbar, FL601 actual temp -2.5 deg C

------24.2.17 -----------------------------------------------------

13.08 Turbo @ 42kRPM, 13W
      Pressure 7.0E-06mbar, backing pressure (leak tester) 9.1E-2mbar, leak rate 3.3E-10 mbar.l/s
      FL601 set point -5.0 deg C, actual -2.2 deg C

14.30 Pressure 6.8E-06mbar, FL601 actual temp -2.5 deg C
      +/-15V preamp power ON

14.49 Pressure 9.9E-06mbar, FL601 actual temp -1.0 deg C

17.44 Pressure 1.1E-05mbar, FL601 actual temp -1.0 deg C

19.00 Pressure 1.1E-05mbar, FL601 actual temp -1.0 deg C
      +/-15V preamp power OFF

------26.2.17 ---------------------------------------------------

11.44 Turbo @ 42kRPM, 13W
      Pressure 7.5E-06mbar, backing pressure (leak tester) 9.1E-2mbar, leak rate 1.8E-10 mbar.l/s
      FL601 set point -5.0 deg C, actual -2.0 deg C
      +/-15V preamp power ON

14.02 Pressure 1.2E-05mbar, FL601 actual temp -0.9 deg C


18.19 +/-15V preamp power OFF
      Pressure 1.2E-05mbar, FL601 actual temp -0.9 deg C

------27.2.17 ---------------------------------------------------

11.44 Turbo @ 42kRPM, 13W, 0.7A, 33 degree
      Pressure 6.8E-06mbar, backing pressure (leak tester) 9.1E-2mbar, leak rate 1.3E-10 mbar.l/s
      FL601 set point -5.0 deg C, actual -2.4 deg C

13.15 pressure 6.7E-06mbar;

Turn off Turbo pump and open chamber!
##################################################################
##################################################################


------2.3.17------------------------------------------------------

using leak testing pump as backing/pre pump (start it with open valve --otherwise there will be alarm signal of leak
rate changing dramatically)
showing some time scales for pumping (p_back is the pressure measured in backing pump and p_TC2 is the second gauge
plugged and measured inside the chamber 

16.03 p_back = 39mbar;   p_TC2 = 1.0E+03mbar
16.29 p_back = 1.3mbar;  p_TC2 = 8.4E+02mbar
16.45 p_back = 0.8mbar;  p_TC2 = 6.7E+02mbar
16.59 p_back = 0.6mbar;  p_TC2 = 4.1E+02mbar

18.26 p_back = 0.2mbar;  p_TC2 = 4.0E+00mbar
     --> starting Turbopump (accelerating up with safety check step in between)
18.30 p_back = 0.3mbar;  p_TC2 = 3.5E+00mbar
      (release more gas from the chamber into pipe by TP)
18.34 Turbo @  I=1.8A,   P=28W,         7 kRPM,   27 deg C
      p_TC2 = 1.9E+00mbar 
18.38 Turbo @  I=6.0A,   upto P=90W,   10 kRPM,   29 deg C
18.42 Turbo @  I=2.5A,   P=38W,        11 kRPM,   29 deg C
      p_TC2 =  5.5E-01mbar
18.55 Turbo @  I=1.1A,   P=17W,        15 kRPM,   31 deg C
      p_TC2 =  4.1E-02mbar
18.59 Turbo @  I=4.5A,   P=82W,        19 kRPM,   31 deg C
      p_TC2 =  2.3E-02mbar


------3.3.17-------------------------------------------------------

11.05 Turbo @ 42kRPM, 13W
      Pressure 5.5E-05mbar, FL601 ON
      FL601 set point -5.0 deg C, actual +24 deg C


      +/-15V preamp power ON

14.15 Turbo @ 42kRPM, 0.7A, 14W, 33 deg C
      Pressure 1.2E-05mbar, backing pressure (leak tester) 9.2E-2mbar, leak rate between 2E-12 and 1.4E-11 mbar.l/s
      FL601 set point -5.0 deg C, actual -1.4 deg C

14.30 Adding insulation to water pipe and try to deflect hot air from chiller away from cold pipes

14.40 - 15.40:     +/-15V preamp power ON

15.00 Turbo @ 42kRPM, 0.8A, 16W, 33 deg C
      Pressure 1.8E-05mbar
      FL601 set point -5.0 deg C, actual -0.7 deg C

15.40 Turbo @ 42kRPM, 0.8A, 17W, 33 deg C
      Pressure 2.1E-05mbar
      FL601 set point -5.0 deg C, actual -0.6 deg C

16.10 Turbo @ 42kRPM, 0.7A, 14W, 33 deg C
      Pressure 1.2E-05mbar
      FL601 set point -5.0 deg C, actual -1.6 deg C

17.30 Turbo @ 42kRPM, 0.7A, 14W, 33 deg C
      Pressure 1.1E-05mbar
      FL601 set point -5.0 deg C, actual -1.7 deg C


------06.03.17 ----------------------------------------------

power fluctuations turned off the chiller

13.42 Turbo @ 42kRPM, 1.0A, 19W, 33 deg C
      Pressure 2.2E-05mbar
      FL601 off.
      Turn on FL601, which went from -5.0 up to actual +0.1 deg C and started cooling again.

visible impact on the pressure:

13.46 Turbo @ 42kRPM, 1.0A, 18W, 33 deg C
      Pressure 1.1E-05mbar
      FL601 set point -5.0 deg C, actual -0.1 deg C

14.04 Turbo @ 42kRPM, 0.7A, 14W, 33 deg C
      Pressure 8.7E-06mbar
      FL601 set point -5.0 deg C, actual -1.6 deg C

14.26 Turbo @ 42kRPM, 0.7A, 13W, 33 deg C
      Pressure 8.2E-06mbar
      FL601 set point -5.0 deg C, actual -2.0 deg C

Adding alcohol and water to cooling reservoir up to limit (full reservior), temperature went up due to room
temperature of added liquids

14.44 temp -0.6 deg C
14.49 temp -2.7 deg C
14.52 temp -3.6 deg C
added more alcohol! (one small bottle and a bit of the rest in IPA bottle)

15.00 temp -3.8 deg C

15.01 Turbo @ 42kRPM, 0.7A, 13W, 33 deg C
      Pressure 7.9E-06mbar
      FL601 set point -5.0 deg C, actual -4.1 deg C

15.12 Turbo @ 42kRPM, 0.7A, 13W, 33 deg C
      Pressure 7.4E-06mbar
      FL601 set point -5.0 deg C, actual -5.3 deg C

15.14 set point to -10.0 deg C

15.22 Turbo @ 42kRPM, 0.7A, 13W, 33 deg C
      Pressure 7.3E-06mbar
      FL601 set point -10.0 deg C, actual -6.3 deg C

16.03 Turbo @ 42kRPM, 0.6A, 12W, 33 deg C
      Pressure 6.9E-06mbar, p_back = 9.1E-02mbar, leak rate 6.2E-11 mbar *l/s
      FL601 set point -10.0 deg C, actual -9.9 deg C

17.00 Turbo @ 42kRPM, 0.6A, 12W, 33 deg C
      Pressure 6.7E-06mbar
      FL601 set point -15.0 deg C, actual -10.3 deg C

17.38 Turbo @ 42kRPM, 0.6A, 12W, 33 deg C
      Pressure 6.7E-06mbar, p_back = 9.1E-02mbar
      FL601 set point -15.0 deg C, actual -10.2 deg C
    --> reach minimum temp due to not enough alcohol --> set point to -10.0 deg C

17.41 Turn +/-15V preamp power ON

  17.46 Turbo @ 42kRPM, 0.6A, 12W, 32 deg C
      Pressure 7.2E-06mbar, p_back = 9.1E-02mbar, leak rate 8.3E-11 mbar *l/s
      FL601 set point -10.0 deg C, actual -8.6 deg C

  18.17 Turbo @ 42kRPM, 0.7A, 14W, 32 deg C
      Pressure 1.0E-05mbar, p_back = 9.2E-02mbar, leak rate 2.3E-10 mbar *l/s
      FL601 set point -10.0 deg C, actual -8.0 deg C

  18.44 Turbo @ 42kRPM, 0.8A, 15W, 33 deg C
      Pressure 1.1E-05mbar, p_back = 9.2E-02mbar, leak rate 2.4E-10 mbar *l/s
      FL601 set point -10.0 deg C, actual -7.7 deg C

18.45 Turn +/-15V preamp power OFF

19.35 Turbo @ 42kRPM, 0.6A, 12W, 33 deg C
      Pressure 7.1E-06mbar, p_back = 9.1E-02mbar, leak rate 1.7E-10 mbar *l/s
      FL601 set point -10.0 deg C, actual -9.8 deg C

------07.03.17 ----------------------------------------------

12.13 Turbo @ 42kRPM, 0.7A, 12W, 32 deg C
      Pressure 6.5E-06mbar, p_back = 9.1E-02mbar, leak rate 1.2E-10 mbar *l/s
      FL601 set point -10.0 deg C, actual -10.0 deg C

12.15 Turn +/-15V preamp power ON

  12.31 Turbo @ 42kRPM, 0.7A, 14W, 32 deg C
      Pressure 8.6E-06mbar, p_back = 9.1E-02mbar, leak rate 1.3E-10 mbar *l/s
      FL601 set point -10.0 deg C, actual -8.6 deg C

  13.22 Turbo @ 42kRPM, 0.8A, 15W, 32 deg C
      Pressure 1.0E-05mbar, p_back = 9.2E-02mbar, leak rate 1.8E-10 mbar *l/s
      FL601 set point -10.0 deg C, actual -6.8 deg C

  13.41 Turbo @ 42kRPM, 0.8A, 15W, 32 deg C
      Pressure 1.0E-05mbar, p_back = 9.2E-02mbar, leak rate 1.9E-10 mbar *l/s
      FL601 set point -10.0 deg C, actual -6.9 deg C

  14.35 Turbo @ 42kRPM, 0.8A, 15W, 33 deg C
      Pressure 1.0E-05mbar, p_back = 9.2E-02mbar, leak rate 1.7E-10 mbar *l/s
      FL601 set point -10.0 deg C, actual -7.3 deg C

  15.24 Turbo @ 42kRPM, 0.8A, 14W, 33 deg C
      Pressure 1.0E-05mbar, p_back = 9.2E-02mbar, leak rate 1.8E-10 mbar *l/s
      FL601 set point -10.0 deg C, actual -7.3 deg C

  16.05 Turbo @ 42kRPM, 0.8A, 14W, 33 deg C
      Pressure 1.0E-05mbar, p_back = 9.2E-02mbar, leak rate 1.8E-10 mbar *l/s
      FL601 set point -10.0 deg C, actual -7.4 deg C

  17.56 Turbo @ 42kRPM, 0.8A, 14W, 33 deg C
      Pressure 1.0E-05mbar, p_back = 9.2E-02mbar, leak rate 1.7E-10 mbar *l/s
      FL601 set point -10.0 deg C, actual -7.6 deg C

18.14 Turn +/-15V preamp power OFF

18.18 Turbo @ 42kRPM, 0.7A, 14W, 33 deg C
      Pressure 9.1E-06mbar, p_back = 9.2E-02mbar, leak rate 1.6E-10 mbar *l/s 
      (decreased after a big outgasing (E-09) just shortly after turning preamp power) 
      FL601 set point -10.0 deg C, actual -7.5 deg C

------08.03.17 --------------------------------------------------------------

12.00 Turbo @ 42kRPM, 1.0A, 18W, 34 deg C
      Pressure 6.2E-06mbar, p_back = 9.1E-02mbar, leak rate 7.9E-11 mbar *l/s
      FL601 set point -10.0 deg C, actual -9.7 deg C

      Power on Turbo pump higher than usual, although pressure is at its lowest since.
      Checked cycle time = 8261 min = 137.68 hours = 5.74 days
              pump life  = 31107 h

Release a bit of coolant to check the concentration in chemistry lab:
density of mixture = 0.965 g/cm3 at temperature of 19.3 degree

concentration of alcohol: x = (density_mixture - density_water)/(density_ethanol - density_water)
                            = (0.965 - 0.998)/(0.79 - 0.998) 
                            = 0.159
around 16% alcohol --> freezing temp should be around -7 to -8 degree C 

13.46 Turbo @ 42kRPM, 1.0A, 18W, 35 deg C
      Pressure 6.1E-06mbar, p_back = 9.1E-02mbar, leak rate 8.6E-11 mbar *l/s
      FL601 set point -10.0 deg C, actual -9.2 deg C

15.30 Turbo @ 42kRPM, 0.8A, 15W, 35 deg C
      Pressure 6.1E-06mbar, p_back = 9.1E-02mbar, leak rate 9.3E-11 mbar *l/s
      FL601 set point -10.0 deg C, actual -9.6 deg C

15.37 Turn +/-15V preamp power ON

  16.38 Turbo @ 42kRPM, 0.8A, 17W, 35 deg C
      Pressure 9.4E-06mbar, p_back = 9.1E-02mbar, leak rate 1.5E-10 mbar *l/s
      FL601 set point -10.0 deg C, actual -7.0 deg C

16.39 Turn +/-15V preamp power OFF

------09.03.17 --------------------------------------------------------------

11.21 Turbo @ 42kRPM, 0.7A, 15W, 35 deg C
      Pressure 6.1E-06mbar, p_back = 9.1E-02mbar, leak rate 8.6E-11 mbar *l/s
      FL601 set point -10.0 deg C, actual -9.6 deg C (fluctuations between -10.0 and -9.3 degree C)

11.22 Turn +/-15V preamp power ON

  14.18 Turbo @ 42kRPM, 0.8A, 17W, 35 deg C
      Pressure 9.3E-06mbar, p_back = 9.1E-02mbar, leak rate 1.7E-10 mbar *l/s
      FL601 set point -10.0 deg C, actual -7.4 deg C

  14.39 Turbo @ 42kRPM, 0.9A, 17W, 35 deg C
      Pressure 9.4E-06mbar, p_back = 9.1E-02mbar, leak rate 1.6E-10 mbar *l/s
      FL601 set point -10.0 deg C, actual -7.4 deg C

14.40 Turn +/-15V preamp power OFF

14.50 Turbo @ 42kRPM, 0.8A, 16W, 35 deg C
      Pressure 7.4E-06mbar, p_back = 9.1E-02mbar, leak rate 1.2E-10 mbar *l/s
      FL601 set point -10.0 deg C, actual -8.3 deg C

15.00 Turbo @ 42kRPM, 0.8A, 16W, 35 deg C
      Pressure 6.9E-06mbar, p_back = 9.1E-02mbar, leak rate 1.3E-10 mbar *l/s
      FL601 set point -10.0 deg C, actual -9.2 deg C

17.14 Turbo @ 42kRPM, 0.8A, 15W, 35 deg C
      Pressure 6.1E-06mbar, p_back = 9.1E-02mbar, leak rate 9.9E-11 mbar *l/s
      FL601 set point -10.0 deg C, actual -10.0 deg C

17.58 Turbo @ 42kRPM, 0.8A, 15W, 34 deg C
      Pressure 6.0E-06mbar, p_back = 9.1E-02mbar, leak rate 4.2E-10 mbar *l/s (did some Helium leak checking)
      FL601 set point -10.0 deg C, actual -9.9 deg C

------10.03.17 --------------------------------------------------------------

10.55 Turbo @ 42kRPM, 0.8A, 15W, 35 deg C
      Pressure 6.0E-06mbar, p_back = 9.1E-02mbar, leak rate 1.0E-10 mbar *l/s
      FL601 set point -10.0 deg C, actual -10.2 deg C

12.01 Turbo @ 42kRPM, 0.7A, 15W, 35 deg C
      Pressure 5.9E-06mbar, p_back = 9.1E-02mbar, leak rate 1.1E-10 mbar *l/s
      FL601 set point -10.0 deg C, actual -10.0 deg C

12.07 leak rate dropped to 3.0E-12 just occasionly

13.07 Turbo @ 42kRPM, 0.7A, 15W, 34 deg C
      Pressure 5.8E-06mbar, p_back = 9.1E-02mbar, leak rate 3.5E-10 mbar *l/s (after leak check - suspious about UP
and DOWN ribbon cable flanges)
      FL601 set point -10.0 deg C, actual -10.1 deg C

13.07 Turn +/-15V preamp power ON

  14.56 Turbo @ 42kRPM, 0.8A, 17W, 34 deg C
      Pressure 9.0E-06mbar, p_back = 9.1E-02mbar, leak rate 6.3E-10 mbar *l/s
      FL601 set point -10.0 deg C, actual -7.2 deg C

14.56 Turn +/-15V preamp power OFF

17.10 Turbo @ 42kRPM, 0.8A, 15W, 34 deg C
      Pressure 5.9E-06mbar, p_back = 9.1E-02mbar, leak rate 2.9E-10 mbar *l/s
      FL601 set point -10.0 deg C, actual -10.1 deg C

17.40 Turbo @ 42kRPM, 0.8A, 15W, 34 deg C
      Pressure 5.7E-06mbar, p_back = 9.1E-02mbar, leak rate 7.4.9E-10 mbar *l/s
      FL601 set point -10.0 deg C, actual -10.2 deg C

------13/03/17 ---------------------------------------------------------------

11.40 Turbo @ 42kRPM, 0.8A, 15W, 34 deg C
      Pressure 6.3E-06mbar, p_back = 9.1E-02mbar, leak rate 9.4E-11 mbar *l/s
      FL601 set point -10.0 deg C, actual -10.1 deg C
      --> pressure went up over the weekend??????????????? (Room temperature can be the reason)

15.53 Turbo @ 42kRPM, 0.8A, 15W, 34 deg C
      Pressure 6.1E-06mbar, p_back = 9.1E-02mbar, leak rate 1.5E-10 mbar *l/s
      FL601 set point -10.0 deg C, actual -10.1 deg C



------14/03/17 ---------------------------------------------------------------

09.38 Turbo @ 42kRPM, 0.7A, 15W, 35 deg C
      Pressure 6.3E-06mbar, p_back = 9.1E-02mbar, leak rate 1.1E-10 mbar *l/s
      FL601 set point -10.0 deg C, actual -9.9 deg C

09.39 Turn +/-15V preamp power ON

  11.14 Turbo @ 42kRPM, 0.8A, 17W, 35 deg C
      Pressure 8.9E-06mbar, p_back = 9.1E-02mbar, leak rate 1.8E-10 mbar *l/s
      FL601 set point -10.0 deg C, actual -7.1 deg C

  11.30 Turbo @ 42kRPM, 0.8A, 17W, 35 deg C
      Pressure 8.7E-06mbar, p_back = 9.1E-02mbar, leak rate 1.8E-10 mbar *l/s
      FL601 set point -10.0 deg C, actual -6.8 deg C

11.30 Turn +/-15V preamp power OFF

Adding same silicone vacuum leak sealant (Aerosol) to the Upstream cable connections (suspious leak)

13.30 Turbo @ 42kRPM, 0.8A, 15W, 34 deg C
      Pressure 6.2E-06mbar, p_back = 9.1E-02mbar, leak rate 1.0E-10 mbar *l/s
      FL601 set point -10.0 deg C, actual -9.8 deg C



------15/03/17 ---------------------------------------------------------------

10.50 Turbo @ 42kRPM, 0.7A, 15W, 35 deg C
      Pressure 6.2E-06mbar, p_back = 9.1E-02mbar, leak rate 1.2E-10 mbar *l/s
      FL601 set point -10.0 deg C, actual -10.0 deg C
      Fluke Thermometer(thermocoupling to preamp board) = 0.3 deg C

13.30 Turbo @ 42kRPM, 0.8A, 15W, 35 deg C
      Pressure 5.9E-06mbar, p_back = 9.1E-02mbar, leak rate 9.9E-11 mbar *l/s
      FL601 set point -10.0 deg C, actual -10.1 deg C

 13.48 Leak test of lower 2/3 of BNC connections, when leak risen slowly from 2E-10 to 3E-09 in like 10minutes.
        At same time p(TC1) raised for some second up to 9E-06mbar, which could indicate a huge outgasing) 

 14.04 Turbo @ 42kRPM, 0.7A, 15W, 35 deg C
      Pressure 6.0E-06mbar, p_back = 9.1E-02mbar, leak rate 3.4E-09 mbar *l/s
      FL601 set point -10.0 deg C, actual -9.7 deg C

16.36 Turbo @ 42kRPM, 0.7A, 15W, 34 deg C
      Pressure 5.9E-06mbar, p_back = 9.1E-02mbar, leak rate 5.9E-10 mbar *l/s
      FL601 set point -10.0 deg C, actual -9.9 deg C



------16/03/17 ---------------------------------------------------------------

11.49 Turbo @ 42kRPM, 0.8A, 15W, 34 deg C
      Pressure 6.1E-06mbar, p_back = 9.1E-02mbar, leak rate 1.8E-10 mbar *l/s
      FL601 set point -10.0 deg C, actual -10.1 deg C

13.11 Turbo @ 42kRPM, 0.8A, 15W, 34 deg C
      Pressure 6.0E-06mbar, p_back = 9.1E-02mbar, leak rate 1.4E-08 mbar *l/s (leak test Toby - found bad o-ring
connection back side downstream end)
      FL601 set point -10.0 deg C, actual -9.7 deg C

----- Switched off TP 15.34 (peter changed Fan of TP controller, new Targetholder did not fit perfectly, so still with
old one, investigation on o-ring leak --> maybe small dust particle)
15.53 chamber close, Roughing pump with bigger pipe ON, 16.02 Turn on TP (Soft start)

16.30 Pressure 1.1E-03mbar (Turbo @ 15kRPM, 0.6A, 10W, 34 deg C)
16.55 Pressure 5.2E-04mbar (Turbo @ 27kRPM, 0.6A, 17W, 35 deg C)
17.32 Pressure 2.8E-04mbar (Turbo @ 42kRPM, 0.6A, 27W, 37 deg C)
18.21 Pressure 2.2E-04mbar (Turbo @ 42kRPM, 0.6A, 20W, 38 deg C)

18.23 pressure = 2.E-04 --> start FL601set point -10.0 deg C, actual -6.5 went up to - 1.5 deg C (start equilibrium)

18.27 Pressure 1.6E-04mbar (Turbo @ 42kRPM, 0.9A, 19W, 38 deg C)
      FL601 set point -10.0 deg C, actual -2.2 deg C

18.40 Pressure 1.3E-04mbar (Turbo @ 42kRPM, 0.9A, 19W, 39 deg C)
      FL601 set point -10.0 deg C, actual -5.5 deg C
      (temperature of TP could have indicated a leak...)


------17/03/17 Friday---------------------------------------------------------------

09.37 Pressure 1.2E-04mbar (Turbo @ 42kRPM, 0.9A, 15W, 40 deg C)
      FL601 set point -10.0 deg C, actual -9.6 deg C
LEAK? (screws on targetholder were not secured enough)

09.43 Pressure 5.4E-05mbar (Turbo @ 42kRPM, 0.6A, 13W, 40 deg C)
      FL601 set point -10.0 deg C, actual -9.6 deg C

10.40 Pressure 5.5E-05mbar (Turbo @ 42kRPM, 0.7A, 13W, 40 deg C)
      FL601 set point -10.0 deg C, actual -9.7 deg C
      Helium test: o-ring (downstream end) front top part 10^-9 to 10^-5 (big leak --> no pressure improvement)

11.06 Turn off FL601
14.35 Turn off TP
###################################################################################
OPEN CHAMBER: Removed pre amp board U1-U4.
over weekend close chamber with roughing pump (~5.0E-01mbar)

------20/03/17 Monday---------------------------------------------------------------

CHANGED/SWITCHED UP and DOWN board inside the chamber.

New UP (1-8) and new DOWN (5-8)

Distance measurements: 




------17/03/17 Friday---------------------------------------------------------------

09.37 Pressure 1.2E-04mbar (Turbo @ 42kRPM, 0.9A, 15W, 40 deg C)
      FL601 set point -10.0 deg C, actual -9.6 deg C
LEAK? (screws on targetholder were not secured enough)

09.43 Pressure 5.4E-05mbar (Turbo @ 42kRPM, 0.6A, 13W, 40 deg C)
      FL601 set point -10.0 deg C, actual -9.6 deg C

10.40 Pressure 5.5E-05mbar (Turbo @ 42kRPM, 0.7A, 13W, 40 deg C)
      FL601 set point -10.0 deg C, actual -9.7 deg C
      Helium test: o-ring (downstream end) front top part 10^-9 to 10^-5 (big leak --> no pressure improvement)

11.06 Turn off FL601
14.35 Turn off TP
###################################################################################
OPEN CHAMBER: Removed pre amp board U1-U4.
over weekend close chamber with roughing pump (~5.0E-01mbar)

------20/03/17 Monday---------------------------------------------------------------

CHANGED/SWITCHED UP and DOWN board inside the chamber.

New UP (1-8) and new DOWN (5-8)

Distance measurements: 

New UPstream: (metal frame stand off = 10mm)
   Position (to end of chamber): 738 - 40 (pos from target) - 32 (plastic standoff) + 10 = 676 (to end of board)

Orientation:

      4
       3 

 6          2
  5        1

       7
       8


New DOWNstream: (metal frame stand off = 20mm) 
   Position (to end of chamber): 738 - 400 (pos from target) - 14 (plastic standoff) + 20 = 344 (to end of board)


Orientation:

      
       

 6    
  5        

       7
       8


***note: 24/03/17: took short standoff for UPstream hold (only ~40mm to targetholder)
                   remeasured short standoffs: 12.5mm


Checked: Cooling (changed circle with one preamp board less), cabling (preamp power signal), 
position, cleaning, test preamp
CABLING changed 16-x19 to D5-8

14.30 close chamber, start roughing pump
15.24 p(TC2) = 4.4E-01 mbar  --> soft start big Turbo Pump
15.39 p = 1.6 E-03mbar (0.7A, 11W, 11kRPM, 27 deg C) 
16.11 p = 5.9 E-04mbar (0.6A, 14W, 23kRPM, 29 deg C)
16.33 p = 3.4 E-04mbar (1.3A, 38W, 34kRPM, 31 deg C)
17.41 p = 1.7 E-04mbar (0.9A, 18W, 42kRPM, 35 deg C)

Turn on FL601 cooling, set point to -5.0 deg C, actual temp starts at 22 deg C

19.12 Pressure 5.9E-05mbar (Turbo @ 42kRPM, 0.6A, 10W, 35 deg C)
      FL601 set point -5.0 deg C, actual -4.6 deg C
   --> changed set point to -10.0 deg C

------21/03/17 Tuesday---------------------------------------------------------------

10.52 p = 1.9 E-05 mbar 
     --> switch on small Turbo pump (TP2)

11.05 Pressure 1.3E-05mbar (Turbo @ 42kRPM, 0.5A, 10W, 35 deg C;  TP2 @ 75kRPM, 0.25A, 13W, 32 deg C)
      FL601 set point -10.0 deg C, actual -5.5 deg C
   --> Turn ア15V preamp power ON
Testing DAQ

11.45 Pressure 2.0E-05mbar (Turbo @ 42kRPM, 0.5A, 10W, 35 deg C;  TP2 @ 75kRPM, 0.26A, 13W, 39 deg C)
      FL601 set point -10.0 deg C, actual -3.9 deg C

12.17 Turn ア15V preamp power OFF 
      Pressure 1.9E-05mbar (Turbo @ 42kRPM, 0.5A, 10W, 35 deg C;  TP2 @ 75kRPM, 0.26A, 13W, 43 deg C)
      FL601 set point -10.0 deg C, actual -4.0 deg C

14.18 Pressure 1.2E-05mbar (Turbo @ 42kRPM, 0.5A, 10W, 35 deg C;  TP2 @ 75kRPM, 0.25A, 13W, 45 deg C)
      FL601 set point -10.0 deg C, actual -5.6 deg C

added some alcohol(mixing takes a short heat up), switch to helium leak tester (DONT MOVE WHILE IT IS ON)
added FAN for small Turbopump

15.40 Pressure 1.1E-05mbar (Turbo @ 42kRPM, 0.5A, 9W, 35 deg C;  TP2 @ 75kRPM, 0.24A, 12W, 33 deg C)
      FL601 set point -10.0 deg C, actual -6.0 deg C

17.05 Pressure 1.0E-05mbar (Turbo @ 42kRPM, 0.5A, 9W, 34 deg C;  TP2 @ 75kRPM, 0.24A, 12W, 34 deg C)
      FL601 set point -10.0 deg C, actual -7.6 deg C

17.05 Turn ア15V preamp power ON

   18.51 Pressure 1.6E-05mbar (Turbo @ 42kRPM, 0.5A, 9W, 34 deg C;  TP2 @ 75kRPM, 0.24A, 12W, 34 deg C)
      FL601 set point -10.0 deg C, actual -5.3 deg C

18.51 Turn ア15V preamp power OFF


------22/03/17 Wednesday---------------------------------------------------------------

10.55 Pressure 7.9E-06mbar (Turbo @ 42kRPM, 0.5A, 8W, 35 deg C;  TP2 @ 75kRPM, 0.22A, 11W, 33 deg C)
      p_back = 9.2E-02mbar, leak rate 10E-11 - 10E-12 mbar*l/s
      FL601 set point -10.0 deg C, actual -7.7 deg C

14.05 pressure 7.7E-06mbar

OPEN CHAMBER FOR ALIGNMENT

------23/03/17 Thursday---------------------------------------------------------------

Si-Detectors: measures- DeltaE-E (65オm - 1000オm) and Downstream E (74オm) 
HANDLE WITH CARE, NO TENSION --> THIN LAYERS. 

Voltage supply: 24V for 74オm. BIAS cable is 4th row and 2nd place

Cabled to U1-4. DONT FLAT THE CHAMBER WITH LOTS OF AIR, use valve at TP.



------24/03/17 Friday---------------------------------------------------------------

Testing Preamps (with Si-"capacitor" - no BIAS) under atmosphere with 5 deg C cooling
Normal resolution with pulser and without Si-Detector = 2-3 channel

Without BIAS:
#1 adc 0-13: FWHM ~10ch
#1 adc 14,15: Broken
#1 adc 16-31: FWHM ~7ch

#2 adc 0-6: FWHM ~27ch
#2 adc 7,8: FWHM ~5-6ch only
#2 adc 9-14: FWHM ~30ch (2peaks each; low peak coming from wrong pulser signal.)
#2 adc 13: FWHM ~90ch
#2 adc 15: FWHM ~27ch

#2 adc 16-31: 8-10ch
except from #2 adc 22,28: ~27ch

Quad BIAS Supply:
0V - 0.20オA
5V - 0.82オA
10V - 0.95オA
15V - 1.07オA
20V - 1.17オA
24V - 1.25オA

With BIAS:
#1 adc 0-31(14,15 broken): FWHM 4-6ch

#2 adc 0-15: FWHM ~ 7ch
except from #2 adc 3,4,7,8,12 with lower tail: FWHM ~7-10ch

#2 adc 16-31: FWHM ~4-6ch
except 22,38: FWHM ~10ch
except 13: FWHM ~25ch


Targetholder (alignment) marker on scale

pos-1:  34.54  (bottom on ladder)

pos-2:  94.05  (middle)

pos-3:  bottom.8 (almost maximum, top on ladder)


New run with alpha source P32 R9344 mixed (only Gold sealing --> wear Cloves)
 (Pu-239,   - 5.155MeV 
  Am-241,   - 5.486MeV
  Cm-244,   - 5.805MeV)

17.15 Close Chamber, roughing pump

17.28 p = 9.9E-01mbar   --> Start TP (big)
17.48 p = 2.0E-03mbar (TP @ 11kRPM, 0.7A, 12W, 28 deg C


------25/03/17 Saturday---------------------------------------------------------------

15.00 Pressure 4.6E-05mbar (Turbo @ 42kRPM, 0.6A, 11W, 34 deg C)
      --> start small TP2 and open valve (quick soft start in 1 minute) - TP2 @ 75kRPM, 0.24A, 12W, 26 deg C)
      start FL601 set point -10.0 deg C, actual 20.7 deg C

15.05 p = 3.8E-05 mbar
15.07 p = 3.4E-05 mbar



------27/03/17 Monday---------------------------------------------------------------

12.17 Pressure 6.2E-06mbar (Turbo @ 42kRPM, 0.8A, 14W, 36 deg C;  TP2 @ 75kRPM, 0.22A, 11W, 32 deg C)
      FL601 set point -10.0 deg C, actual -8.4 deg C

12.21 Turn ア15V preamp power ON

   14.05 Pressure 8.7E-06mbar (Turbo @ 42kRPM, 0.8A, 13W, 36 deg C;  TP2 @ 75kRPM, 0.22A, 11W, 32 deg C)
      FL601 set point -10.0 deg C, actual -5.0 deg C

   14.43 Pressure 8.7E-06mbar (Turbo @ 42kRPM, 0.8A, 13W, 36 deg C;  TP2 @ 75kRPM, 0.22A, 11W, 32 deg C)
      FL601 set point -10.0 deg C, actual -5.1 deg C

14.44 Turn ア15V preamp power OFF

16.25 Pressure 6.1E-06mbar (outgazing still ongoing, saw peak up to 2.3E-05mbar at a glimpse)
      (Turbo @ 42kRPM, 0.8A, 13W, 36 deg C;  TP2 @ 75kRPM, 0.22A, 11W, 32 deg C)
      FL601 set point -10.0 deg C, actual -7.3 deg C

------28/03/17 Tuesday---------------------------------------------------------------

13.06 Pressure 5.7E-06mbar (Turbo @ 42kRPM, 0.8A, 14W, 35 deg C;  TP2 @ 75kRPM, 0.22A, 11W, 32 deg C)
      FL601 set point -10.0 deg C, actual -9.4 deg C

13.09 Turn ア15V preamp power ON

   16.15 Pressure 7.8E-06mbar (Turbo @ 42kRPM, 0.8A, 13W, 35 deg C;  TP2 @ 75kRPM, 0.22A, 11W, 32 deg C)
      FL601 set point -10.0 deg C, actual -5.8 deg C

16.16 Turn ア15V preamp power OFF

18.36 Pressure 6.1E-06mbar (Turbo @ 42kRPM, 0.7A, 13W, 35 deg C;  TP2 @ 75kRPM, 0.22A, 11W, 32 deg C)
      FL601 set point -10.0 deg C, actual -8.5 deg C


------29/03/17 Wednesday---------------------------------------------------------------

13.15 Pressure 5.9E-06mbar (Turbo @ 42kRPM, 0.7A, 13W, 35 deg C;  TP2 @ 75kRPM, 0.22A, 11W, 32 deg C)
      FL601 set point -10.0 deg C, actual -9.5 deg C

13.25 Turn ア15V preamp power ON for alpha measurement

   14.30 Pressure 7.3E-06mbar (Turbo @ 42kRPM, 0.7A, 13W, 35 deg C;  TP2 @ 75kRPM, 0.22A, 11W, 32 deg C)
      FL601 set point -10.0 deg C, actual -7.6 deg C

   15.41 Pressure 7.9E-06mbar (Turbo @ 42kRPM, 0.7A, 13W, 35 deg C;  TP2 @ 75kRPM, 0.22A, 11W, 32 deg C)
      FL601 set point -10.0 deg C, actual -6.6 deg C

   16.31 Pressure 7.8E-06mbar (Turbo @ 42kRPM, 0.7A, 13W, 35 deg C;  TP2 @ 75kRPM, 0.22A, 11W, 32 deg C)
      FL601 set point -10.0 deg C, actual -6.4 deg C
good vacuum condition of ~8E-06mbar with preamps on
16.31 Turn ア15V preamp power OFF
   
17.46 Pressure 6.1E-06mbar (Turbo @ 42kRPM, 0.7A, 12W, 35 deg C;  TP2 @ 75kRPM, 0.22A, 11W, 32 deg C)
      FL601 set point -10.0 deg C, actual -8.7 deg C


------30/03/17 Thursday---------------------------------------------------------------

12.15 Pressure 6.3E-06mbar (Turbo @ 42kRPM, 0.7A, 13W, 35 deg C;  TP2 @ 75kRPM, 0.22A, 11W, 32 deg C)
      FL601 set point -10.0 deg C, actual -9.8 deg C
      shut off TP + TP2 --> spinning down

13.05 p = 3.1E-05mbar, turn off FL601, bringing down to atmosphere takes about 30 minutes.

###########################
13.35 OPEN CHAMBER, longer cable done for detector, Peter add block (15.8mm) the target holder (due to pos 3)
      put stand off on DOWNstream

Test #3,#4 under atmosphere WITHOUT BIAS:
#3 adc 0 lower xmean ~ 1550, FWHM ~20ch
#19 low xmean ~450, FWHM ~3ch
rest of #3: xmean 2400-2600, FWHM ~10ch

#4 xmean more or less between 2000-3000, most ~2400.
#4 adc 4: double peak, FWHM > 30ch
#4 adc 0 - 15: FWHM ~ 30ch, except of
#4 adc 7: FWHM ~8ch.

#4 adc 16-31: FWHM ~12-13ch, right tail
except #4 adc 16: FWHM ~7ch
       #4 adc 18: FWHM 15ch
       #4 adc 25: FWHM ~20ch

BIAS up to 24 V  1.38オA leak current (while ramp <3オA)
alpha source inside chamber on new middle position: 94.05 -15.8 = 78.25
chamber close after 2h 

WITH BIAS: 
#3 0-31: 4-6ch FWHM
except #3 adc 1: ~10ch FWHM
       #3 adc 19: ~3ch FWHM (lower mean)
left tail: adc 0-23      and    right tail: adc 24-31

#4 adc 0-15: broader: adc 0-4: ~7ch FWHM
                      adc 5:   ~3ch FWHM
                      adc 6,13: 11ch FWHM
                      adc 7,15: ~5,6ch FWHM
                      adc 8-12,14: ~7,8ch FWHM

#4 adc 16-31: 5-8ch FWHM  right tail 

##############################################
################################closed chamber

16.10 Roughing pump slowly pump down (open value stepwise)
16.25 p = 1.1 mbar   --> start TP
16.30 p = 1.3E-01 mbar (TP @ 7kRPM, 1.0A, 15W, 28 deg C)
17.21 p = 7.8E-04 mbar (TP @ 27kRPM, 0.6A, 17W, 31 deg C)
18.08 p = 3.3E-04 mbar (TP @ 42kRPM, 1.1A, 22W, 34 deg C)
      --> start TP2 (1min) --> TP2 @ 75kRPM, 0.26A, 13W, 26 deg C)
18.12 p = 2.7E-04 mbar
18.16 start FL601 set to -10.0 deg C, starts at ~ +4.0 deg C
18.18 p = 2.0E-04 mbar
18.54 p = 8.8E-05 mbar (TP @ 42kRPM, 0.8A, 17W, 35 deg C; TP2 @ 75kRPM, 0.25A, 13W, 32 deg C)
      FL601 set point -10.0 deg C, actual -5.5 deg C


------31/03/17 Friday---------------------------------------------------------------

10.35 p = 1.5E-05mbar
12.15 p = 1.4E-05 mbar (TP @ 42kRPM, 0.6A, 12W, 35 deg C; TP2 @ 75kRPM, 0.22A, 11W, 32 deg C)
      FL601 set point -10.0 deg C, actual -6.2 deg C
12.30 Turn ア15V preamp power ON    (for Outgazing and alpha measurement)

   13.40 p = 2.2E-05 mbar;   leak rate: 1.6E-09 mbar l /S
   14.15 p = 2.1E-05 mbar;   T = -4.5 deg C

TURN BIAS on 24V, 1.16オA
14.17 DAQ started
#3 -19: low peak 181 with Amplitude of 478 (14.33), FWHM 4-5ch --> Area ~ 5500counts
rest #3 0-31: xmean ~ 1000, FWHM 8-12 ch

#4 adc 6 and 7: no signal
 --> adc 6 could be preamp problem
 --> adc 7 could be Si-Det issue, because it was observed with #2adc7 before
#4 adc 5: low xmean = 134, FWHM 2.5ch
rest of #4 adc 0-15: (14.40) ~500-800 Amplitude, xmean ~ 1000, FWHM ~12ch
#4 adc 16-31: A~200-250, xmean ~ 950-1050; FWHM ~7-8ch


   14.45 p = 2.1E-05 mbar;   Temp = -4.8 deg C;    leak rate: 8.0E-10 mbar l /S
14.46 Turn ア15V preamp power OFF, switch off Amps and BIAS

16.10 p = 1.3E-05 mbar;   T = -6.2 deg C

18.23 p = 1.2E-05 mbar;   T = -6.2 deg C


------
------03/04/17 Monday ------------------- Peter testing ----------------

15.49 p = 7.6E-06 mbar;

------04/04/17 Tuesday -------------------

14.05 p = 7.1E-06 mbar;

TP Off

------05/04/17 Wednesday -------------------
########################################
##### chamber open, alpha source out ###

10.25 chamber close, start pumping
12.10 large TP ON

13.55 p = 9.4E-05 mbar

15.30 p = 5.7E-05 mbar

------06/04/17 Thursday -------------------

09.50 p = 1.5E-05 mbar

15.30 p = 1.0E-05 mbar


------07/04/17 Friday -------------------

09.25 p = 9.1E-06 mbar (TP @ 42kRPM, 0.5A, 9W, 35 deg C; no TP2)
      FL601 set point -10.0 deg C, actual -9.7 deg C
09.25 Turn ア15V preamp power ON

   12.02 p = 1.2E-05 mbar
         T = -6.9 deg C
   Turn ア15V preamp power OFF

12.13 p = 1.0E-05 mbar
      T = -7.5 deg C

12.54 p = 9.0E-06 mbar
      T = -8.5 deg C

15.04 p = 9.0E-06 mbar
      T = -9.1 deg C

16.47 p = 8.6E-06 mbar
      T = -9.3 deg C

##########################
ACCIDENT Registered on Sunday 09/04/17 ~10 am:
Power went down due to wiring problem
Tom C. closed valves, but chamber was probably already vented to atmosphere. 
During power shut down roughing and turbo pump slowed down and turbo probably deaccelerate against atmosphere.
Possible back flows.


------11/04/17 Tuesday -----------------
OPEN CHAMBER 16.05: readjust Downstream plate to 326mm, cause we changed to longer stand-off
Downstream --> 74オm Si-Det
Upstream   --> 500オm Si-Det

Check DAQ/FWHM of spectra without BIAS on under atmosphere and only PULSER signal:


BIAS up (500オm):
Voltage[V]  Leak current[オA]
  0             0.14
  5             0.6
 10             0.73
 20             0.93
 30             1.11
 40             1.22
 50             1.27
 60             1.32
 70             1.34
 80             1.38
 90             1.42
100             1.44
110             1.48
120             1.49
130             1.52   --> 1.44 (after 5 min)


Check DAQ/FWHM of spectra with BIAS on under atmosphere and only PULSER signal:


CLOSING CHAMBER
18.03 Start roughing pump
18.06 p = 1.0E+03 mbar
18.11 p = 2.0E+01 mbar
18.14 p = 3.8E+00 mbar
18.16 p = 1.9E+00 mbar
18.19 p = 1.1E+00 mbar  --> start TP (and close valve at TP2)
18.22 p = 5.0E-01 mbar
18.25 p = 9.0E-02 mbar (took a while for next stage of acceleration... very slow start)
18.34 p = 2.6E-03 mbar
18.40 p = 2.2E-03 mbar
18.44 p = 1.6E-03 mbar
19.02 p = 1.0E-03 mbar
19.12 p = 8.2E-04 mbar
19.18 p = 7.6E-04 mbar
19.21 p = 7.0E-04 mbar
19.30 p = 5.7E-04 mbar
20.00 p = 3.7E-04 mbar (full speed at TP @ 42 kRPM)
20.04 TURN ON TP2
20.06 p = 3.0E-04 mbar
20.10 p = 2.9E-04 mbar (TP @ 42kRPM, 0.9A, 18W, 34 deg C; TP2 @ 75kRPM, 0.28A, 15W, 28 deg C)

------12/04/17 Wednesday -----------------

9.33  p = 6.1E-05 mbar
10.27 p = 5.9E-05 mbar
      Turn on FL601 chiller set to -5.0 deg C; T_start = 18.7 deg C
10.44 p = 2.9E-05 mbar; T = 4.8 deg C 
      stop chiller to remove 500ml of Coolant and replace it with ~375ml IPA
10.54 p = 2.7E-05 mbar; T = 3.4 deg C

11.08 p = 2.1E-05 mbar; T = -5.0 deg C   (set to -10.0 deg C)
11.15 Turn ア15V preamp power ON

  11.29 p = 2.4E-05 mbar; T = -5.8 deg C
  11.50 p = 2.5E-05 mbar; T = -5.1 deg C

BIAS up for UPSTREAM (500オm - 130 V) and DOWNSTREAM (74オm - 24 V)
11.51 DAQ started for testing Si-det with triple alpha source

  12.10 p = 2.4E-05 mbar; T = -5.1 deg C
  12.17 p = 2.5E-05 mbar; T = -5.2 deg C
  13.12 p = 2.3E-05 mbar; T = -5.2 deg C
  15.08 p = 2.1E-05 mbar; T = -5.1 deg C
  15.50 p = 2.0E-05 mbar; T = -5.1 deg C

15.58 DAQ stopped; ramp down BIAS, switch off amp+preamps
      Turn ア15V preamp power OFF

16.02 p = 2.0E-05 mbar; T = -5.2 deg C
16.35 p = 1.5E-05 mbar; T = -6.0 deg C
17.04 p = 1.5E-05 mbar; T = -6.0 deg C

17.05 Turn ア15V preamp power ON
  17.19 p = 1.7E-05 mbar; T = -1.4 deg C (set to -2.0)
17.19 Turn ア15V preamp power OFF

17.40 p = 1.5E-05 mbar; T = -5.7 deg C (after set to -10.0 deg C)
18.14 p = 1.4E-05 mbar; T = -5.8 deg C

------13/04/17 Thursday -----------------

11.37 p = 1.0E-05 mbar; T = -6.2 deg C
      Turn ア15V preamp power ON
  13.21 p = 1.4E-05 mbar; T = -4.5 deg C
      Turn ア15V preamp power OFF
add ~300ml IPA, set point to +5.0 deg C

13.44 p = 1.3E-05 mbar; T = 4.1 deg C
14.33 p = 1.4E-05 mbar; T = 5.0 deg C

14.38 FL601 turn OFF, peter test cooling pipes with Helium (~4bar He --> leak rate up to 10^{-6} or 10^{-5} mbar*l/s
Test chiller in close circle: 15.05  T = -8.3 deg C

open cooling pipe --> chamber without cooling:
14.50 p = 2.0E-05 mbar
15.05 p = 2.9E-05 mbar
15.35 p = 4.0E-05 mbar (leak rate 5 * 10^{-5} mbar*l/s)

15.50 reconnect cooling to chamber
15.55 p = 1.5E-05 mbar; T = -5.8 deg C;    leak rate 2.0E-05 mbar*l/s
16.55 p = 1.7E-05 mbar; Cooling was off by Daid to test temperature influence

17.00 p = 1.1E-05 mbar; T = -5.7 deg C (chiller back on)
17.50 p = 1.0E-05 mbar; T = -6.6 deg C
18.18 p = 1.0E-05 mbar; T = -6.8 deg C

------14/04/17 Friday -----------------

11.09 p = 8.5E-06 mbar; T = -7.1 deg C
11.15 Turn ア15V preamp power ON

  12.10 p = 1.2E-05 mbar; T = -5.2 deg C    switch to Helium leak pump: rate ~ 4.3E-09 mbar*l/s

  12.29 p = 1.1E-05 mbar; T = -5.2 deg C

  Turn ア15V preamp power OFF

14.20 p = 8.5E-06 mbar; T = -7.2 deg C

14.50 Turn ア15V preamp power ON

  15.05 p = 9.4E-06 mbar;  T = -6.6 deg C;  leak rate 5*10^{-09} mbar*l/s
  15.30 p = 1.0E-05 mbar;  T = -5.7 deg C;  leak rate 6.3*10^{-09} mbar*l/s
  15.50 p = 1.1E-05 mbar;  T = -5.6 deg C;  leak rate 7*10^{-09} mbar*l/s
  16.33 p = 1.1E-05 mbar;  T = -5.5 deg C;  leak rate 7.6*10^{-09} mbar*l/s
  22   Thu Sep 14 14:07:21 2017 pbalignment

Chat with survey team

1. mauf plugs for both ends of chamber with concentric 8mmH7 holes, with nylon tipped grub screws/clamp bars

2. Manuf of target with 8mmH7 hole

3. Manuf detector with 8mmH7 hole

4. Manuf of 4x pad sliders for feet for radial alignment.

 

Report from survey team attached.

Attachment 1: 2017.11.02_HISO_XT03_EdinburghChamber.pdf
2017.11.02_HISO_XT03_EdinburghChamber.pdf 2017.11.02_HISO_XT03_EdinburghChamber.pdf 2017.11.02_HISO_XT03_EdinburghChamber.pdf 2017.11.02_HISO_XT03_EdinburghChamber.pdf 2017.11.02_HISO_XT03_EdinburghChamber.pdf
  19   Thu Sep 14 09:47:02 2017 dkdiscussion with beam operators
CERN side
     Faraday cup at the end of SEC, ideally with automated control to insert and remove

Edinburgh side
     Photographs of chamber, diagrams, measurement, nicely organised 
     FC output from SHV to SMA, to input to CERN preamp
     Beam viewer scintillating min beam intensity?
     Irises to protect SSDs are essential since the beams will be tuned for energy case by case and not at once.
     Collimator sizes for the target ladder?
     Window view port to confirm iris settings?
     Photodiodes should be on the opposite side of the target ladder from the collimator.  
  26   Mon Sep 25 14:50:46 2017 dkTo do list before main exp. - update from meeting 28th September
NB: This list may be incomplete!

CERN side: 

     Beam operators hope for a FC which can be automated at the end of SEC: Olof ?

     Gain resistors (and other items) sent to CERN.  Where are they?  KJ has them? Check with Karl

     w26 PC needs to get online.  Contact Olof when we return to ISOLDE to confirm the network port is still free.
Daid. Need to register the device to CERN

     Dry N2 bottle: ask Karl

Edinburgh side:

     Iris automated controls and mounts: in construction. Add additional stop so doesnt close all the way. Peter

     Cooling line chamber feed through: get spare. Peter
 
     Get Al degrader foil for heavy ions (~13 um), mount on dummy frame etc. probably fine from store. Should get 15
um for highest beam energy. With lowest beam energy we might need to change the foil: Daid

     Alpha spectral analysis: to be sorted. Daid, Claudia, Tom, Ruchi

     Sort routines for dE-E plots, etc, for online analysis: Tom

     Photo diodes x2 along with degrader foil: PCB for photodeiodes, PCB for photodiodes. extra holes for ladder. Peter

     Target ladder config finalization: 2 photodiodes, collimator (what size, we have 3 mm and 10mm), Beam viewer,
We need to verify how many positions we have. Need at least 2 CH2 and 1 Carbon. Daid, Claudia

     Extra ribbon cables are needed? need to check with calibration. Also need long ribbon cables

     Beam viewer: see target ladder

     Faraday cup SHV to SMA converter: Peter

     Window port for front of chamber? Peter

     Photographs and diagrams of experimental setup -- organized clearly for beam operators during tuning: add
technical drawing on elog. 

     Alignment -- 2 end caps and one target ladder mount for optics reflector (8 mm diameter?) For survey people. Peter

     Adjustable feet to be produced for radial alignment: Peter / Toby

     Contact CERN about date for beam connection: As soon as possible after 20th October. Talk to Karl/Olof. 

     Collimator at exit of SEC ? needs to be aligned. Ask what the other beamlines do - does olof have collimator
ahead of his chamber

     Calibration of Target ladder positions. Peter / Daid


Email to Karl & Olof about:
- FC after SEC
- Parcel of gain resistors
- date for connection of SEC to our chamber
- collimator in front of chamber ?
- N2 bottle
  Draft   Thu Sep 28 03:34:20 2017 dkheavy ion degrader foil
--text from email--
Using Claudia's code, I added a layer of Al foil before the heavy ion
silicon detector.  The attached plots correspond to the heavy ion
portion of fig 3(a) in the proposal.  Thus it only uses ground-state
transitions and doesn't account for geometry or efficiency etc.  I opt
to show only the highest and lowest beam energies for each simulation.

As I said before, something between 10 and 20 um of Al was likely to
be the correct choice, which is confirmed.  10 um is almost to thin to
guarantee the signals are below 200 MeV, while for 20 um we start
having issues of detecting the lowest energy recoils for the low
energy run.

We can discuss the selection in more detail at the next meeting, but
most probably we can just use some kitchen foil, adopting whatever its
thickness of 13 to 15 um depending what we find.

The attached file starts with no degrader, then adds 10 um Al, and 1
um Al steps up to 20 um.
Attachment 1: degrader.pdf
degrader.pdf degrader.pdf degrader.pdf degrader.pdf degrader.pdf degrader.pdf degrader.pdf degrader.pdf
  28   Thu Sep 28 03:38:18 2017 dkheavy ion degrader foil - updated
--original email text--
Using Claudia's code, I added a layer of Al foil before the heavy ion
silicon detector.  The attached plots correspond to the heavy ion
portion of fig 3(a) in the proposal.  Thus it only uses ground-state
transitions and doesn't account for geometry or efficiency etc.  I opt
to show only the highest and lowest beam energies for each simulation.

As I said before, something between 10 and 20 um of Al was likely to
be the correct choice, which is confirmed.  10 um is almost to thin to
guarantee the signals are below 200 MeV, while for 20 um we start
having issues of detecting the lowest energy recoils for the low
energy run.

We can discuss the selection in more detail at the next meeting, but
most probably we can just use some kitchen foil, adopting whatever its
thickness of 13 to 15 um depending what we find.

The attached file starts with no degrader, then adds 10 um Al, and 1
um Al steps up to 20 um.

--Updated 2 Oct to include 5 um ~ 10 um--

As ~185 MeV is the maximum energy, but the gain resistors may vary by up to ~10%, then the 'maximium safe energy' is about 160 MeV.  This we can get for the highest energy run 
with 15 um Al (we have it in stock); however, then the lowest energy case becomes about 30 MeV, or less than 1 MeV/u, which may be an issue for straggling etc.  So we will 
prepare two foils, as it is seen even for the low energy beam, it comes almost to 190 MeV without a degrader foil.  In fact, any foil thickness between 5 ~ 10 um may be okay 
(we have 8 um in stock).  

Using also all 5 energies (212, 236, 260, 277, 295) then we see that the two lowest energies are okay for 8 um foil, and the three highest energies are suitable for 15 um foil.

NB: 19 Oct I deleted 15 um (previously attachment 2) and now it is attachment 8.  

Note that the beam energy order (right to left, left to right) differs between the first PDF and the second two.

--4 Oct--
Existing 8um Al foils are not the right size.  Check 6 um and 10 um in detail, as we have more stock of these two.  Attached.

--12 Oct--

Attachments 5-7 show the 59Cu energy after the degrader foils for elastic (12C,12C) scattering.  Foil thicknesses are 6, 10, 15 um of Al, respectively.  Note that the angular 
range shows up to 15 degrees (larger than the case of the (p,a)).
Attachment 1: degrader.pdf
degrader.pdf degrader.pdf degrader.pdf degrader.pdf degrader.pdf degrader.pdf degrader.pdf degrader.pdf
Attachment 2: degrader08_full.pdf
degrader08_full.pdf
Attachment 3: degrader06_full.pdf
degrader06_full.pdf
Attachment 4: degrader10_full.pdf
degrader10_full.pdf
Attachment 5: degrader06_12C_full.pdf
degrader06_12C_full.pdf
Attachment 6: degrader10_12C_full.pdf
degrader10_12C_full.pdf
Attachment 7: degrader15_12C_full.pdf
degrader15_12C_full.pdf
Attachment 8: degrader15_full.pdf
degrader15_full.pdf
  29   Wed Oct 4 18:57:16 2017 dkDigitized run notes from tests at CERN
Transcription of paper logbook for runs.  Please check as I cannot read all the writing.

Wednesday 13/9/2017

Detectors installed:
2083-9, 74 um, dE layer
3080-23, 1012 um, E layer
2410-23, 65 um, HI layer

Det.	Bias	Leak
dE	-24	0.46
E	-350	2.61
HI	-14	1.05

Runs 0 - 5 : junk

Run 10
  Header: Alpha calibration
  Date: 13/9/2017
  Start: 18:03 (CERN)
  Stop: 18:12
  Note: Some cables disconnected to test CANNOT READ

Run 11
  Header: Alpha calibration
  Date: 13/9/2017
  Start: 18:16
  Stop: 
  Notes:
  ADC	Ch	Notes
  1	0-15	ok
  1	16-31	
  2	0-15	7+8 only pulser
  2	16-31	ok
  3	0-15	only pulser
  3	16-31	19 low gain (other notes but crossed out)
  4	0-15	6 low (noise?), rest fine
  4	16-31	only pulser
  5	0-15	"probably okay"
  5	16-31	"probably okay"
  6	0-31	"probably okay"

-------
Thursday 14/9/2017

Det.	Bias	Leak
dE	-24	0.44
E	-351	2.4
HI	-14	0.95

Run 12
  Header: 
  Date: 14/9/2017
  Start: 11:50
  Stop: 12:08
  Note: Bad.  Pulser switched on through run

Run 13
  Header: Pulser ON
  Date: 14/9/2017
  Start: 12:07
  Stop: 
  Notes: good
  ADC	Notes
  1	ch 0 ... bad resolution in pulser
  2	ch 7,8 no alphas
  3	ch 0 low gain, ch 0-15 almost only pulser but maybe shieled from alpha source & low noise
  3	ch 24 low resolution
  4	ch 6 no alphas, ch 7 low gain for alphas but not for pulser
  4	ch 16-31 no alpha (maybe shielded & low noise)
  5	okay
  6	okay

 TDC missing #1-32, #1-80,120,123

14:00 before turning off biases:

Det.	Bias	Leak
dE	-24	0.4
E	-351	2.15
HI	-14.5	0.89

15:15 - Vent to remove dE layer

Extra SSD cable arrives

TDC mapping

Det.	Side	TDC ch
dE	p+n	16-63
E	n+n	64-79
HI	p+n	80-127

Bias detectors
Det.	Bias	Leak
E	-350	1.83
HI	-14	0.8

Run 15
  Header:
  Date: 14/9/2017
  Start: 18:22
  Stop: 18:25
  Notes: Bad.  Pulser to 2 Hz

Run 16
  Header: Alpha source in front of E detector 
  Date: 14/9/2017
  Start: 18:31
  Stop: 
  Notes: Bad.  CANNOT READ

Run 17
  Header: 
  Date: 14/9/2017
  Start: 18:32
  Stop:
  Notes: Bad.  ADC #4 ch 6, 7 bad.  
  Switch U7e8 (??) (n+n side p+n) from E detector
  Problem moved -> CANNOT READ

-------
Friday 15/9/2017

10:00 bias detectors

Det.	Bias	Leak
E	-351	2.13
HI	-14.5	0.88

T: -4.9 C
P: 4.9 x 10^-6 mbar

Run 19
  Header: alpha-calib of E layer 3080-23
  Date: 15/9/2017 
  Start:  
  Stop: 
  Notes: 
  No TDC data ...?
  ADC	Ch	Comments
  1	0-31	Pulser (only) in all channels except ch2 which has a few counts across whole spectrum (hair)
  		Resolutions range from 2.0 -> 7.0
		Pulser centroid ~ channel 2800 in all spectra (except ch 2 of course)
  2	0-15	Pulser in all channels, ~channel 2000 (i.e. similar gains)
  		Resolution ~ 2.0 ch, except ch 8 sigma = 4.5 ch
  2	16-31	Pulser in all channels at ~channel 3000
		Resolution ~ 2.0 ch, excehpt ch 24 which is 6.7 ch & ch 16: 4.35 ch.
  3	0-31	Pulser & alphas in all channels
  		(Not 19...see below)
		Resolution is *poor* in ch0
		All gains look similar
		Alpha spectra in similar channel numbers as for other ADCs, but pulser at only 1/2 the channel number
		(A drawing is given to compare ch0 to ch1-15.)
  3	19	Gain seems very low
  		Pulser @ channel 430, alphas @ channel 200
			So about 1/5 of gain of others in ADC 3 channels
  4	0-15	All but ch 6 very similar (good)
  		ch 6: Pulser in similar channel to other spectra, but narrower (~4 ch, c.f. ~8-10 channels for other spectra.)
		      And alphas are much lower gain, appearing ~ ch 70, c.f. 1100 for others
  4	16-31	All okay.  Triple alpha @ ~ channel 1150, pulser @ 2800
  5	0-31	All okay (though low gain - deliberately)
  6	0-31	Ok, see above for ADC 5 notes.

Det.	Bias	Leak
E	-351	2.08
HI	-14.5	0.88

Restart MIDAS; VME crate
  => TDC signals return

12:02 Removed E layer #3080-23 (HI layer #2410-23 remains unchanged)

New 1035 um Si installed #2951-22.  It uses a different bias voltage than the previous E layer (-160 vs. -351 V)

12:50 Stop DAQ (run 22 larger than necessary)

Run 23 (?? no note)
  Header: 
  Date: 
  Start: 
  Stop: 
  Notes: 
  ADC	ch	Comments
  1	0-15	Pulser only, similar gains
  		ch 2 no pulser (hair)
		No alphas (unsurprising!)
  1 	16-31	Pulser only, similar gains
  		All okay.
		(No alphas, unsurprising!)
  2	0-31	Pulser only, no alphas, similar gains
  		All okay
  3	0-15	Pulser + alphas
  		alphas in all strips
		alpha similar gains and resolutions
		pulser in ch 0 at ~1/2 amplitude
  3	16-31	Pulser + alphas.
  		ch 19 pulser and alphas at much reduced channel number ~1/5 gain
  4	0-15	Pulser + alphas
  		ch 6 no alphas - Pulser narrow (no detector capacitence?)
		ch 7 alphas at reduced gain (~1/5), pulser narrow
  4 	16-31 	All good
  5	0-31	All good (low gains - by design)
  6	0-31	All good (low gains - by design)

  v767 #1 TDC
  	ch	Comments
	0	1 peak
	1-15	No data
	16-31	1 peak
	32	No peaoks
	33	1 peak + 1 small peak
	34-44	2 peaks
	45-47	1 peak
	48	1 peak + hair
	49-52	1 peak
	53	1 peak + hair
	54-56	2 peaks
	57-63	1 peak
	64-79	broad region of data (alphas?)
	80-86	1 peak
	87	1 peak + 1 smaller peak
	88	Mainly 1 or 2 peaks + lots of hair
	89	1 peak
	90,91	2 peaks
	92-95	1 peak
	96	1 peak
	97-127	Mainly 1 peak

Cable for 1 mm E detector must be faulty

Before opening the chamber:
Det.	Bias	Leak
E	-160	1.12
HI	-14.5	0.92

(When trying to switch the cable to the SSD, apparently it was not fully connected.  Some time is lost to pump, vent, check, etc)

After new cabling is confirmed and evacuation is complete:
Det.	Bias	Leak
E	-160	1.17
HI	14.5	0.96

Run 24
  Header: Check new cable
  Date: 15/9/2017
  Start: 16:18 (UK)
  Stop: 16:35 (UK)
  Notes:
  ADC	ch	Comments
  3	0-15	okay
  3	16-31	ch 19 low gain
  4	0-15	ch 5, ch 7 low gain
  4	16-31	okay
  208   Sat Nov 10 18:48:04 2018 dkTravel schedule spreadsheet
https://docs.google.com/spreadsheets/d/1kvCWF13RMH83fssuN2_JLZXJwNL4rUGSEeKOWAIHUi0/edit#gid=0

or

https://tinyurl.com/isolde-44ti
  210   Tue Nov 20 14:07:37 2018 dkBlue Loan Book Scan
Scan is attached, including all items denoted in the loan book as being taken to ISOLDE.

This will be useful for a complete compilation of our inventory.
Attachment 1: loan-book-scan.pdf
loan-book-scan.pdf loan-book-scan.pdf
  211   Tue Nov 20 16:06:09 2018 dkProjected Inventory of ISOLDE
(To be confirmed / finalized)

==Rack Electronics==

Two (2) 19" electronics racks
  =Rack #1=
  2 (Two) RAL Amp crate (1 full with modules, 1 with five (5) modules)
     21 RAL 109 modules in total.  (NB: Extras are borrowed from Miniball)
  Two (2) NIM Fans
  Four (4) Cable patch boxes
    Ribbon cables for patch

  Two (2) Preamp power supplies
     +/-15 Volt type, up to several Amperes

  =Rack #2=
  VME crate, with power cable
    Seven (7) Silena ADCs
    One (1) S9418 ACQ controller
    Two (2) CAEN V767 TDC
    One (1) CAEN 560
    One (1) MVME 24XX Motorola crate Controller
  26 ribbon cables
    Twelve (12) for analog
    Twelve (12) for discrim
    Two (2) extra

  Camac crate
    Five (5) Logic Shaped Outputs (CAMAC)
    Two (2) CAMAC / VME converter
    One (1) LeCroy 622 Quad Logic (NIM)
    One (1) Ortec 566 TAC (NIM)
    One (1) Lecroy 222 dual GDG
    One (1) Ortec 459 bias supply
  
  NIM crate
    One (1) BNC PB-5 digital pulse generator
    One (1) BNC PB-4 analog pulse generator -- returned to Edinburgh 2018.12
    One (1) Analog ratemeter 4678
    One (1) ECL-NIM-ECL EC 1601
    One (1) Silena Quad Bias Supply 7710 -- swapped with functional unit 2018.11 from problems in 59Cu
    Two (2) LeCroy 419A Logic FI/FO
    One (1) Ortec 572 Amplifier
    One (1) LeCroy Quad Discriminator 821
    One (1) NIM Cooknell homebrew PA Power Supply

One (1) Cooknell Preamplifier
4 (Four) additional Long ribbon cables
2x spare HVBNC cables
BNC to Lemo 
Lemo to Ribbon box
lemo cables

==Preamp setup==
2 Preamp feedthrough
  Bolts and washers
  2 O-Rings + 1 spare
16way BNC multicore
o-ring for ribbon feedthroughs

Julabo Chiller & Insulated External Chiller Lines
Swagelok nylon and brasss ferrels for gas line / coolant piping
spare straight and right angle swagelok fittings
thin green polymer tubing (gas lines, cooling system)

thermometer
thermocouple cable
1 thermocouple vacuum gauge

==Generic Power Connections==
3 Power Bars
20x IEC leads
2x Swiss to UK leads

==Vacuum systems==

Chamber, sliding mount

Scrolling pump
Backflow interlock

Large TMP from TUDA
  Controller
Small TMP
  Controller
TMP bolts (large and small)

Two (2) vacuum hoses
Valve for TP 1 & 2
vacuum clamps

Two (2) inverse magnetron ion gauges
Multi gauge reader
Old inv magnetron vac gauge meter

Upstream beamline pipe
    Gate valve, nipple, adaptor flange to SEC
N2 pipe with valve
Mechanical pressure gauge and "plus" type connector

Blanking kindey plate
bellows
bellow cage

TMP/gauge mounting pipe

==Silicon Detectors and Components==

Two (2) Micron S2-type 1000 um SSDs
Three (3) Micron S2-type 75 um SSDs

dummy detectors
20 ~1 cm spacers for SSD mounting
9 ~2 cm spacers for SSD mounting

Degrader Foils

Irises and controller box

Al heavy ion degraders of various thicknesses.

3 Hamamatsu Photodiodes
1 PCB for Hamamatsu Photodiodes
3 degrader foils (10, 15 20 um Al)

==Beam Diagnostics==

Faraday cup
current integrator for FC
cctv camera power lead
ZnS (or similar) scint screen

==Computer==

Sun Microsystems Tower
Dell monitor
Sun keyboard
Sun mouse
Ethernet hub


==Misc==

alignment plugs for chamber

2 ch oscilloscope 
2 (two) basic voltmeters

middle foot sections of blue frame
6 feet for chamber stand
All bolts for chamber stand
Spacers
3 large O-Rings for chamber (spare)

Bolts, nuts and spacers for chamber end flanges and center

short bits of blue frame
braces off blue frame

Wheels for frame
Tools

Safety boots and helmets-
    -Daid
    -Toby
    -Mirco
    -Claudia
    -Sarah
    -Conor

==In Edinburgh==
  Target ladder and components
  33   Thu Oct 19 14:41:19 2017 daidheavy ion degrader foil - full / new
See entry https://elog.ph.ed.ac.uk/ISOLDE/28 for a full description, but editing the same elog is altering attachment numbers etc.

Here I've added 20 um foil as well as 5.3 MeV/u (~313 MeV) 59Cu beam.

The first set is 59Cu(p,alpha)56Ni showing 56Ni in the heavy ion detector, with different beam energies and Al degrader foil thicknesses.

The second set is for 59Cu(12C,12C)59Cu scattering.
Attachment 1: degrader06_full.pdf
degrader06_full.pdf
Attachment 2: degrader10_full.pdf
degrader10_full.pdf
Attachment 3: degrader15_full.pdf
degrader15_full.pdf
Attachment 4: degrader20_full.pdf
degrader20_full.pdf
Attachment 5: degrader06_12C_full.pdf
degrader06_12C_full.pdf
Attachment 6: degrader10_12C_full.pdf
degrader10_12C_full.pdf
Attachment 7: degrader15_12C_full.pdf
degrader15_12C_full.pdf
Attachment 8: degrader20_12C_full.pdf
degrader20_12C_full.pdf
  37   Sat Nov 4 14:10:20 2017 daidJulabo chiller operation
This log entry contains a basic instruction on how to operate the Julabo liquid flow chiller.

There are 5 buttons (left to right): Up/Down, Enter, 1/0.  

To toggle power to the device, use 1/0 (on/off).  NB: This only turns the electronics on, and does not initiate circulation / cooling.

When switched on, it will display sequentially some basic information sequentially.  One may generally ignore these, and wait until it displays the current temperature.

To adjust or check the set point, tap either of Up/Down.  To change the set point, press and hold either up or down.  Release the button when the desired set point is reached, 
and tap up/down to do the fine tuning.

Once the desired set point is entered, the display will return to the current temperature reading.

To turn on circulation / chilling, press and hold Enter (the down/left facing arrow).  One needs to hold this button several seconds.  It is obvious when it begins to operate, as 
it will make more sounds, and additionally the Snowflake LED indicator should turn on (unless the current temperature is lower than the set point, it may simply flash briefly).

The set point can be adjusted while the circulation / chilling is on, as above.

To turn the circulation / chilling off, hold down Enter again for several seconds.

Best to raise the set point to ~20 degrees and wait for it to reach the value, before turning off circulation, and finally the device (0 button).
  49   Wed Nov 8 22:00:00 2017 daidScalars
Present scalars:
0: Ungated
1: Gated
2: dE
3: E
4: HI
5: Pulser
6: 1 kHz Clock

To add:
  113   Sat Nov 11 20:27:28 2017 TD, DK, RG, MBRun 100
Header: Continue CH2 at 4.0 MeV/u
Ender: 
====================================================
Run number			: 100
Date				: 11 / 11 / 2017
Start time			: 21:28
Stop time			: 00:27 
Trigger rate (MIDAS)	        : 6
Blocks written		        : 784
覧覧覧覧覧覧覧覧覧覧覧覧覧覧覧
Run Scalars (approx rates)
0- Ungated                      : 4
1- Gated                        : 4
2- dE                           : 150k
3- E                            : 20
4- HI                           : 556
5- Pulser                       : 1
6- Clock			: 968
覧覧覧覧覧覧覧覧覧覧覧覧覧覧覧
Beam energy (MeV/u)		: 4.0 MeV/u (nominal)
Beam intensity (epA)	        : 1.75
Target				: CH2 30+25
Ladder position	(mm)	        : 67.26
HI Degrader foil (um)		: 10 um Al
覧覧覧覧覧覧覧覧覧覧覧覧覧覧覧
del_E voltage (V)		: -40
del_E leakage (uA)		: 0.68
E voltage (V)			: -350
E leakage (uA)		        : 2.10
E_hi voltage (V)		: -14
E_hi leakage (uA)		: 2.59
覧覧覧覧覧覧覧覧覧覧覧覧覧覧覧
PA power (Up, positive)	(A)	: 1.55
PA power (Up, negative) (A)     : 0.55
PA power (Down, positive) (A)	: 0.8
PA power (Down, negative) (A)   : 0.25
覧覧覧覧覧覧覧覧覧覧覧覧覧覧覧
Pressure (IMG1)	(mbar)	        : 6.8 x 10^-6
Pressure (old gauge) (mbar)     : 1.9 x 10^-5
Pressure (SEC)	(mbar)         	: 3.8 x 10^-6
覧覧覧覧覧覧覧覧覧覧覧覧覧覧覧
Temp (Chiller)	deg	        : -9.9
Temp (chamber)	deg	        : 6.0
覧覧覧覧覧覧覧覧覧覧覧覧覧覧覧
Comments: At 21:57, we are notified by RILIS team that they would like to do some cleaning and laser maintenance before leaving for the night.  It should take 15 minutes.  We 
will watch the scalars but continue the run.  22:14 beam is off.  22:18 the laser tuning is finished.  
====================================================
  110   Sat Nov 11 18:06:42 2017 TD, DKRun 98
Header: CH2 target in, Laser #2 off, 4.0 MeV/u
Ender: ends OK
====================================================
Run number			: 98
Date				: 11 / 11 / 2017
Start time			: 18:55
Stop time			: 19:23
Trigger rate (MIDAS)	        : 24
Blocks written		        : 1089
覧覧覧覧覧覧覧覧覧覧覧覧覧覧覧
Run Scalars (approx rates)
0- Ungated                      : 24
1- Gated                        : 24
2- dE                           : 142k
3- E                            : 62
4- HI                           : 42
5- Pulser                       : 24
6- Clock			: 969
覧覧覧覧覧覧覧覧覧覧覧覧覧覧覧
Beam energy (MeV/u)		: 4.0 MeV/u nominal
Beam intensity (epA)	        : 0.05 to 0.06
Target				: CH_2 30+25
Ladder position	(mm)	        : 67.26
HI Degrader foil (um)		: 10 um Al
覧覧覧覧覧覧覧覧覧覧覧覧覧覧覧
del_E voltage (V)		: -40
del_E leakage (uA)		: -0.67
E voltage (V)			: -350
E leakage (uA)		        : -2.10
E_hi voltage (V)		: -14
E_hi leakage (uA)		: -2.67
覧覧覧覧覧覧覧覧覧覧覧覧覧覧覧
PA power (Up, positive)	(A)	: 1.5
PA power (Up, negative) (A)     : 0.55
PA power (Down, positive) (A)	: 0.8
PA power (Down, negative) (A)   : 0.25
覧覧覧覧覧覧覧覧覧覧覧覧覧覧覧
Pressure (IMG1)	(mbar)	        : 6.9E-6
Pressure (old gauge) (mbar)     : 1.95E-5
Pressure (SEC)	(mbar)         	: 3.9E-6
覧覧覧覧覧覧覧覧覧覧覧覧覧覧覧
Temp (Chiller)	deg	        : -10.1
Temp (chamber)	deg	        : +5.8
覧覧覧覧覧覧覧覧覧覧覧覧覧覧覧
Comments: DAQ trigger changed to OR( E p+n strips) .OR. PRE-SCALED-OR( E_HI p+n strips ) .OR. Pulser
====================================================
  111   Sat Nov 11 18:31:10 2017 TD, DKRun 99
Header: 4.0MeV/u 59Cu20+ + (CH2)n, Laser #2 *on*
Ender: Normal run stop
====================================================
Run number			: 99
Date				: 11.11.17
Start time			: 19.31
Stop time			: 21:27
Trigger rate (MIDAS)	        : 
Blocks written		        : 501
覧覧覧覧覧覧覧覧覧覧覧覧覧覧覧
Run Scalars (approx rates)
0- Ungated                      : 4
1- Gated                        : 4
2- dE                           : 142k
3- E                            : 27
4- HI                           : 596
5- Pulser                       : 1
6- Clock			: 968
覧覧覧覧覧覧覧覧覧覧覧覧覧覧覧
Beam energy (MeV/u)		: 4.0 MeV/u nominal
Beam intensity (epA)	        : 1.7-2.5
Target				: CH_2 30+25
Ladder position	(mm)	        : 67.26
HI Degrader foil (um)		: 10 um Al
覧覧覧覧覧覧覧覧覧覧覧覧覧覧覧
del_E voltage (V)		: -40
del_E leakage (uA)		: -0.67
E voltage (V)			: -350
E leakage (uA)		        : -2.10
E_hi voltage (V)		: -14
E_hi leakage (uA)		: -2.57
覧覧覧覧覧覧覧覧覧覧覧覧覧覧覧
PA power (Up, positive)	(A)	: 1.5
PA power (Up, negative) (A)     : 0.55
PA power (Down, positive) (A)	: 0.8
PA power (Down, negative) (A)   : 0.25
覧覧覧覧覧覧覧覧覧覧覧覧覧覧覧
Pressure (IMG1)	(mbar)	        : 6.9E-6
Pressure (old gauge) (mbar)     : 1.9E-5
Pressure (SEC)	(mbar)         	: 3.8E-6
覧覧覧覧覧覧覧覧覧覧覧覧覧覧覧
Temp (Chiller)	deg	        : -10.1
Temp (chamber)	deg	        : +5.8
覧覧覧覧覧覧覧覧覧覧覧覧覧覧覧
Comments: BNC PB-4 Frequency to minimum (c. 1Hz)
          *All* SortSas and hist spectra zero'd at start of run
          Online sort program sort4.f

End of run had 2 alphas.  In ~12 hours (w/o subtracting down time) for runs 85~93 we had 22.  Stats seem about right.
====================================================
  14   Fri Sep 8 19:03:23 2017 TDw26 network configuration
Network configuration files for workstation w26 (Sun Ultra 45)

serial # 0743TFC0DQ
host ID # 840e6006
MAC address 0:14:4f:e:60:6

The MIDAS DAQ configured to use ppc1 (MVME 2431 CPU) connected
to w26 via port 1 (/dev/bge1)
Attachment 1: hosts
#
# Internet host table
#
#
# Merged entries from ipnodes into hosts on <Wednesday April  2 18:57:41 BST 2008>
# Backup files saved in /etc/inet/ directory: hosts.premerge, ipnodes.premerge
#
127.0.0.1	localhost	
129.215.73.26	w26	loghost	egmsg
192.168.1.25    w26daq
192.168.1.1     ppc0            # TUDA Motorola MVME 2431 CPU
192.168.1.2     ppc1            #
192.168.1.3     ppc2            #
192.168.1.4     ppc3
192.168.1.5     ppc4
192.168.1.6     ppc5
192.168.1.7     ppc6
Attachment 2: ethers
08:00:3e:2b:59:76       ppc0    # TRIUMF, TUDA
00:01:af:0b:11:22       ppc1    # HIE-ISOLDE, CERN
00:01:af:0b:10:7a       ppc2    # u/s
00:01:af:09:86:88       ppc3    # GELINA, JRC IRMM 
00:01:af:08:2c:38       ppc4
00:01:af:09:df:db       ppc5    # Edinburgh
00:01:af:09:86:83       ppc6    # u/s
Attachment 3: netmasks
#
# The netmasks file associates Internet Protocol (IP) address
# masks with IP network numbers.
# 
# 	network-number	netmask
#
# The term network-number refers to a number obtained from the Internet Network
# Information Center.
#
# Both the network-number and the netmasks are specified in
# "decimal dot" notation, e.g:
#
# 		128.32.0.0 255.255.255.0
#
129.215.72.0	255.255.252.0
192.168.1.0     255.255.255.0
Attachment 4: nsswitch.conf
#
# Copyright 2006 Sun Microsystems, Inc.  All rights reserved.
# Use is subject to license terms.
#
# ident	"@(#)nsswitch.files	1.14	06/05/03 SMI"

#
# /etc/nsswitch.files:
#
# An example file that could be copied over to /etc/nsswitch.conf; it
# does not use any naming service.
#
# "hosts:" and "services:" in this file are used only if the
# /etc/netconfig file has a "-" for nametoaddr_libs of "inet" transports.

passwd:     files
group:      files
hosts:      files
ipnodes:    files
networks:   files
protocols:  files
rpc:        files
ethers:     files
netmasks:   files
bootparams: files
publickey:  files
# At present there isn't a 'files' backend for netgroup;  the system will 
#   figure it out pretty quickly, and won't use netgroups at all.
netgroup:   files
automount:  files
aliases:    files
services:   files
printers:	user files

auth_attr:  files
prof_attr:  files
project:    files

tnrhtp:     files
tnrhdb:     files
Attachment 5: hostname.bge0
w26
Attachment 6: hostname6.bge0

Attachment 7: hostname.bge1
w26daq
Attachment 8: hostname6.bge1

Attachment 9: defaultrouter
129.215.75.254
Attachment 10: dfstab
#	Place share(1M) commands here for automatic execution
#	on entering init state 3.
#
#	Issue the command '/etc/init.d/nfs.server start' to run the NFS
#	daemon processes and the share commands, after adding the very
#	first entry to this file.
#
#	share [-F fstype] [ -o options] [-d "<text>"] <pathname> [resource]
#	.e.g,
#	share  -F nfs  -o rw=engineering  -d "home dirs"  /export/home2

share -F nfs -o rw=np-suns /scratch
share -F nfs -o rw=ppc1,root=ppc1 /MIDAS
Attachment 11: hosts.allow
rpcbind: 192.168.1.0/255.255.255.0
in.tftpd: 192.168.1.0/255.255.255.0
rpcbind: 129.215.72.0/255.255.252.0 
rpcbind: 255.255.255.255 0.0.0.0
sshd: .gsi.de .ed.ac.uk .dl.ac.uk .cern.ch @np-suns
Attachment 12: hosts.deny
ALL: ALL: spawn (/usr/sfw/sbin/safe_finger -l @%h | /bin/mailx -s %d-%h td) &: DENY
  15   Fri Sep 8 19:12:19 2017 TDMIDAS DAQ configuration
MIDAS DAQ configuration files from directories w26:/MIDAS/experiments/daq
and w26:/MIDAS/TapeServer/SunOS
Attachment 1: latest
SAC sac#1 vme 0x0400:0x000000:0x00000000:0:0 1 0 0 0 0 0 1 {} 1
{Silena S9418} s9418#1 vme 0x0000:0x000000:0x02000000:0:0 1 0 0 0 0 0 1 {} 1
{Silena S9418} s9418#2 vme 0x0000:0x000000:0x02010000:0:0 2 0 0 0 0 0 1 {} 2
{Silena S9418} s9418#3 vme 0x0000:0x000000:0x02020000:0:0 3 0 0 0 0 0 1 {} 3
{Silena S9418} s9418#4 vme 0x0000:0x000000:0x02030000:0:0 4 0 0 0 0 0 1 {} 4
{Silena S9418} s9418#5 vme 0x0000:0x000000:0x02040000:0:0 5 0 0 0 0 0 1 {} 5
{Silena S9418} s9418#7 vme 0x0000:0x000000:0x02060000:0:0 7 0 0 0 0 0 1 {} 7
{CAEN V560} v560#1 vme 0x0000:0x000000:0x00660000:0:0 1 0 0 0 0 0 1 {} 1
{Silena S9418} s9418#6 vme 0x0000:0x000000:0x02050000:0:0 6 0 0 0 0 0 1 {} 6
{CAEN V767} v767#1 vme 0x0000:0x000000:0x01800000:0:0 1 0 0 0 0 0 1 {} 1
{CAEN V767} v767#2 vme 0x0000:0x000000:0x01810000:0:0 2 0 0 0 0 0 1 {} 2
Attachment 2: latest
{"s9418#1.RTP" "0x02"}
{"s9418#2.RTP" "0x02"}
{"s9418#3.RTP" "0x02"}
{"s9418#4.RTP" "0x02"}
{"s9418#5.RTP" "0x02"}
{"s9418#6.RTP" "0x02"}
{"s9418#7.RTP" "0x02"}
{"s9418#1.Threshold0" "16"}
{"s9418#1.Threshold1" "16"}
{"s9418#1.Threshold10" "16"}
{"s9418#1.Threshold11" "16"}
{"s9418#1.Threshold12" "16"}
{"s9418#1.Threshold13" "16"}
{"s9418#1.Threshold14" "16"}
{"s9418#1.Threshold15" "16"}
{"s9418#1.Threshold16" "16"}
{"s9418#1.Threshold17" "16"}
{"s9418#1.Threshold18" "16"}
{"s9418#1.Threshold19" "16"}
{"s9418#1.Threshold2" "16"}
{"s9418#1.Threshold20" "16"}
{"s9418#1.Threshold21" "16"}
{"s9418#1.Threshold22" "16"}
{"s9418#1.Threshold23" "16"}
{"s9418#1.Threshold24" "16"}
{"s9418#1.Threshold25" "16"}
{"s9418#1.Threshold26" "16"}
{"s9418#1.Threshold27" "16"}
{"s9418#1.Threshold28" "16"}
{"s9418#1.Threshold29" "16"}
{"s9418#1.Threshold3" "16"}
{"s9418#1.Threshold30" "16"}
{"s9418#1.Threshold31" "16"}
{"s9418#1.Threshold4" "16"}
{"s9418#1.Threshold5" "16"}
{"s9418#1.Threshold6" "16"}
{"s9418#1.Threshold7" "16"}
{"s9418#1.Threshold8" "16"}
{"s9418#1.Threshold9" "16"}
{"s9418#2.Threshold0" "16"}
{"s9418#2.Threshold1" "16"}
{"s9418#2.Threshold10" "16"}
{"s9418#2.Threshold11" "16"}
{"s9418#2.Threshold12" "16"}
{"s9418#2.Threshold13" "16"}
{"s9418#2.Threshold14" "16"}
{"s9418#2.Threshold15" "16"}
{"s9418#2.Threshold16" "16"}
{"s9418#2.Threshold17" "16"}
{"s9418#2.Threshold18" "16"}
{"s9418#2.Threshold19" "16"}
{"s9418#2.Threshold2" "16"}
{"s9418#2.Threshold20" "16"}
{"s9418#2.Threshold21" "16"}
{"s9418#2.Threshold22" "16"}
{"s9418#2.Threshold23" "16"}
{"s9418#2.Threshold24" "16"}
{"s9418#2.Threshold25" "16"}
{"s9418#2.Threshold26" "16"}
{"s9418#2.Threshold27" "16"}
{"s9418#2.Threshold28" "16"}
{"s9418#2.Threshold29" "16"}
{"s9418#2.Threshold3" "16"}
{"s9418#2.Threshold30" "16"}
{"s9418#2.Threshold31" "16"}
{"s9418#2.Threshold4" "16"}
{"s9418#2.Threshold5" "16"}
{"s9418#2.Threshold6" "16"}
{"s9418#2.Threshold7" "16"}
{"s9418#2.Threshold8" "16"}
{"s9418#2.Threshold9" "16"}
{"s9418#3.Threshold0" "16"}
{"s9418#3.Threshold1" "16"}
{"s9418#3.Threshold10" "16"}
{"s9418#3.Threshold11" "16"}
{"s9418#3.Threshold12" "16"}
{"s9418#3.Threshold13" "16"}
{"s9418#3.Threshold14" "16"}
{"s9418#3.Threshold15" "16"}
{"s9418#3.Threshold16" "16"}
{"s9418#3.Threshold17" "16"}
{"s9418#3.Threshold18" "16"}
{"s9418#3.Threshold19" "16"}
{"s9418#3.Threshold2" "16"}
{"s9418#3.Threshold20" "16"}
{"s9418#3.Threshold21" "16"}
{"s9418#3.Threshold22" "16"}
{"s9418#3.Threshold23" "16"}
{"s9418#3.Threshold24" "16"}
{"s9418#3.Threshold25" "16"}
{"s9418#3.Threshold26" "16"}
{"s9418#3.Threshold27" "16"}
{"s9418#3.Threshold28" "16"}
{"s9418#3.Threshold29" "16"}
{"s9418#3.Threshold3" "16"}
{"s9418#3.Threshold30" "16"}
{"s9418#3.Threshold31" "16"}
{"s9418#3.Threshold4" "16"}
{"s9418#3.Threshold5" "16"}
{"s9418#3.Threshold6" "16"}
{"s9418#3.Threshold7" "16"}
{"s9418#3.Threshold8" "16"}
{"s9418#3.Threshold9" "16"}
{"s9418#4.Threshold0" "16"}
{"s9418#4.Threshold1" "16"}
{"s9418#4.Threshold10" "16"}
{"s9418#4.Threshold11" "16"}
{"s9418#4.Threshold12" "16"}
{"s9418#4.Threshold13" "16"}
{"s9418#4.Threshold14" "16"}
{"s9418#4.Threshold15" "16"}
{"s9418#4.Threshold16" "16"}
{"s9418#4.Threshold17" "16"}
{"s9418#4.Threshold18" "16"}
{"s9418#4.Threshold19" "16"}
{"s9418#4.Threshold2" "16"}
{"s9418#4.Threshold20" "16"}
{"s9418#4.Threshold21" "16"}
{"s9418#4.Threshold22" "16"}
{"s9418#4.Threshold23" "16"}
{"s9418#4.Threshold24" "16"}
{"s9418#4.Threshold25" "16"}
{"s9418#4.Threshold26" "16"}
{"s9418#4.Threshold27" "16"}
{"s9418#4.Threshold28" "16"}
{"s9418#4.Threshold29" "16"}
{"s9418#4.Threshold3" "16"}
{"s9418#4.Threshold30" "16"}
{"s9418#4.Threshold31" "16"}
{"s9418#4.Threshold4" "16"}
{"s9418#4.Threshold5" "16"}
{"s9418#4.Threshold6" "16"}
{"s9418#4.Threshold7" "16"}
{"s9418#4.Threshold8" "16"}
{"s9418#4.Threshold9" "16"}
{"s9418#5.Threshold0" "16"}
{"s9418#5.Threshold1" "16"}
{"s9418#5.Threshold10" "16"}
{"s9418#5.Threshold11" "16"}
{"s9418#5.Threshold12" "16"}
{"s9418#5.Threshold13" "16"}
{"s9418#5.Threshold14" "16"}
{"s9418#5.Threshold15" "16"}
{"s9418#5.Threshold16" "16"}
{"s9418#5.Threshold17" "16"}
{"s9418#5.Threshold18" "16"}
{"s9418#5.Threshold19" "16"}
{"s9418#5.Threshold2" "16"}
{"s9418#5.Threshold20" "16"}
{"s9418#5.Threshold21" "16"}
{"s9418#5.Threshold22" "16"}
{"s9418#5.Threshold23" "16"}
{"s9418#5.Threshold24" "16"}
{"s9418#5.Threshold25" "16"}
{"s9418#5.Threshold26" "16"}
{"s9418#5.Threshold27" "16"}
{"s9418#5.Threshold28" "16"}
{"s9418#5.Threshold29" "16"}
{"s9418#5.Threshold3" "16"}
{"s9418#5.Threshold30" "16"}
{"s9418#5.Threshold31" "16"}
{"s9418#5.Threshold4" "16"}
{"s9418#5.Threshold5" "16"}
{"s9418#5.Threshold6" "16"}
{"s9418#5.Threshold7" "16"}
{"s9418#5.Threshold8" "16"}
{"s9418#5.Threshold9" "16"}
{"s9418#6.Threshold0" "16"}
{"s9418#6.Threshold1" "16"}
{"s9418#6.Threshold10" "16"}
{"s9418#6.Threshold11" "16"}
{"s9418#6.Threshold12" "16"}
{"s9418#6.Threshold13" "16"}
{"s9418#6.Threshold14" "16"}
{"s9418#6.Threshold15" "16"}
{"s9418#6.Threshold16" "16"}
{"s9418#6.Threshold17" "16"}
{"s9418#6.Threshold18" "16"}
{"s9418#6.Threshold19" "16"}
{"s9418#6.Threshold2" "16"}
{"s9418#6.Threshold20" "16"}
{"s9418#6.Threshold21" "16"}
{"s9418#6.Threshold22" "16"}
{"s9418#6.Threshold23" "16"}
{"s9418#6.Threshold24" "16"}
{"s9418#6.Threshold25" "16"}
{"s9418#6.Threshold26" "16"}
{"s9418#6.Threshold27" "16"}
{"s9418#6.Threshold28" "16"}
{"s9418#6.Threshold29" "16"}
{"s9418#6.Threshold3" "16"}
{"s9418#6.Threshold30" "16"}
{"s9418#6.Threshold31" "16"}
{"s9418#6.Threshold4" "16"}
{"s9418#6.Threshold5" "16"}
{"s9418#6.Threshold6" "16"}
{"s9418#6.Threshold7" "16"}
{"s9418#6.Threshold8" "16"}
{"s9418#6.Threshold9" "16"}
{"s9418#7.Threshold0" "16"}
{"s9418#7.Threshold1" "16"}
{"s9418#7.Threshold10" "16"}
{"s9418#7.Threshold11" "16"}
{"s9418#7.Threshold12" "16"}
{"s9418#7.Threshold13" "16"}
{"s9418#7.Threshold14" "16"}
{"s9418#7.Threshold15" "16"}
{"s9418#7.Threshold16" "16"}
{"s9418#7.Threshold17" "16"}
{"s9418#7.Threshold18" "16"}
{"s9418#7.Threshold19" "16"}
{"s9418#7.Threshold2" "16"}
{"s9418#7.Threshold20" "16"}
{"s9418#7.Threshold21" "16"}
{"s9418#7.Threshold22" "16"}
{"s9418#7.Threshold23" "16"}
{"s9418#7.Threshold24" "16"}
{"s9418#7.Threshold25" "16"}
{"s9418#7.Threshold26" "16"}
{"s9418#7.Threshold27" "16"}
{"s9418#7.Threshold28" "16"}
{"s9418#7.Threshold29" "16"}
{"s9418#7.Threshold3" "16"}
{"s9418#7.Threshold30" "16"}
{"s9418#7.Threshold31" "16"}
{"s9418#7.Threshold4" "16"}
{"s9418#7.Threshold5" "16"}
{"s9418#7.Threshold6" "16"}
{"s9418#7.Threshold7" "16"}
{"s9418#7.Threshold8" "16"}
{"s9418#7.Threshold9" "16"}
{"sac#1.Mode" "0"}
{"sac#1.DELR" "6"}
Attachment 3: v767#1
AcqMde14 0 
Enble26  -1 -1 -1 -1 -1 -1 -1 -1 
Trig31 100 
Trig33 -50 
Trig35 0 
Trig3a 3 
Strt47  9 
Edge67  1 1 1 
DRdy73  2 
Attachment 4: v767#2
AcqMde14 0 
Enble26  -1 -1 -1 -1 -1 -1 -1 -1 
Trig31 100 
Trig33 -50 
Trig35 0 
Trig3a 3 
Strt47  9 
Edge67  1 1 1 
DRdy73  2 
Attachment 5: MidasOptions
## sys
v1190_GroupBase 20
Include.VME 1
Include.V1190 1
v1190.channels 4096
v879_GroupBase 32
v767.shift 0
v1290.offset 0
v879.offset 0
Include.V767 1
Histogram.V767 1
v767.offset 0
v878.offset 0
Include.V775 1
Include.V1290 1
s9418.shift 0
Histogram.V775 1
Histogram.V1190 1
s9418.offset 0
Rate.channels 4096
Stat.channels 4096
v785.shift 0
v878.channels 4096
s9418_GroupBase 1
vme.channels 8192
s9418.channels 4096
Include.S9418 1
Include.V785 1
v1290.channels 4096
v775.shift 0
Histogram.V785 1
v775.offset 0
v785.offset 0
v792_GroupBase 32
v775_GroupBase 48
v879.channels 4096
v1290_GroupBase 24
Include.V792 1
v879.shift 0
v775.channels 4096
v767.channels 4096
Histogram.V792 1
vme.offset 0
Histogram.V1290 1
Include.V878 1
Include.V879 1
Histogram.V878 1
v767_GroupBase 20
v792.channels 4096
Histogram.V879 1
vme.shift 3
Histogram.S9418 1
v820_GroupBase 60
v560_GroupBase 56
v878.shift 0
Histogram.VME 1
v785.channels 4096
v878_GroupBase 32
v792.offset 0
v785_GroupBase 32
v792.shift 0
v1190.shift 0
v1290.shift 4
v1190.offset 0
##
Attachment 6: TapeServerParameters
IS607 1 R
Attachment 7: TS_configuration
#device configuration information
#    any line starting with a # is a comment and is ignored
stats  ./stats
#    devices available   -   list ends with a null line or a line starting with a &
#    format       class_name     device_name    driver_task_path_name
#       class_name MUST be one of dlt, exabyte, scsitape, sink
dlt           /dev/rmt/1     ./driver
exabyte       /dev/rmt/0     ./driver
file          /dev/file/0    ./driver
sink          /dev/null/0    ./driver
&
#data link configuration information  - list end with a null line
#    format   link_task_path_name  
./linkTCP
&
#program options - list end with a null line
msg_reporting_level 0x0180fff8
#   use 0x0080fff8 to enable msg logging
msg_logging_level 0xfff8
tapeserver_options 2
#    default for following is 16Kbytes - both MUST be the same at present
data_buffer_size 32
tape_block_size 32
disc_file_size 2000
#
file_path_base  /scratch/data
&
#end of information


  23   Thu Sep 14 15:40:42 2017 TDMSL type S2 bias data
Attachment 1: MSLtypeS2.txt
MSL S2		Depletion	Operating 	Wafer		Location/Comment
Serial #	Voltage (V)	Voltage (V)	Thickness (mm)


2083-9		12		24		0.074		Edinburgh
2410-23		7		14		0.065		Edinburgh
2948-6		14		40		0.065		Edinburgh


2155-13		70		130		0.493		TRIUMF
2182-5		70		130		0.471		TRIUMF
2323-18				130		0.5		Edinburgh
2623-11				130		0.5		Edinburgh	I_L > 20uA @ 130V
2623-18		60		130		0.496		Edinburgh
2623-26				130		0.5		Edinburgh

2951-22		100		160             1.035		Edinburgh
3080-23		110		350		1.012		Edinburgh

Note:

1) Detectors on this list are not necessarily currently operational.
2) Operating voltages are typically 2-3x depletion voltage.
3) Thicknesses are from test results supplied by Micron Semiconductor Ltd.

T.Davinson

Revised - 10.5.17
  36   Sat Nov 4 06:29:03 2017 TDMIDASsort online sort programs, spectrum titles, calibration s/w & READMEs
Attachment 1: IS607.tar.gz
Attachment 2: Calibration.tar.gz
Attachment 3: READMEs.tar.gz
ELOG V3.1.4-unknown