AIDA GELINA BRIKEN nToF CRIB ISOLDE CIRCE nTOFCapture DESPEC DTAS EDI_PSA 179Ta CARME StellarModelling DCF K40
  DESPEC, Page 20 of 35  ELOG logo
  ID Date Author Subject
  320   Mon May 17 15:14:35 2021 LJWMonday 17th May 2021 16:00-20:00

16:25

Informed at shift handover that there has been no beam.

Statistics  - See attachment 1

Temperatures - See attachment 2

Bias & Leakage currents - See attachment 3

System Wide checks - All passed, EXCEPT:

ADC Calibration

   
FEE64 module aida01 failed
FEE64 module aida02 failed
FEE64 module aida03 failed
FEE64 module aida04 failed
FEE64 module aida05 failed
FEE64 module aida06 failed
FEE64 module aida07 failed
FEE64 module aida08 failed
FEE64 module aida09 failed
FEE64 module aida10 failed
FEE64 module aida11 failed
FEE64 module aida12 failed
FEE64 module aida13 failed
FEE64 module aida14 failed
FEE64 module aida15 failed
FEE64 module aida16 failed
Calibration test result: Passed 0, Failed 16

If any modules fail calibration , check the clock status and open the FADC Align and Control browser page to rerun calibration for that module

 

White Rabbit decoder status:

   
         Base         Current     Difference
aida05 fault      0x4da :      0x4dd :      3  
White Rabbit error counter test result: Passed 15, Failed 1

Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR

 

18:00

Usecb scalars now showing DSSD implant rates.

Statistics  - See attachment 4

Temperatures - See attachment 5

Bias & Leakage currents - See attachment 6

System Wide checks - All passed, EXCEPT:

ADC Calibration

   
FEE64 module aida01 failed
FEE64 module aida02 failed
FEE64 module aida03 failed
FEE64 module aida04 failed
FEE64 module aida05 failed
FEE64 module aida06 failed
FEE64 module aida07 failed
FEE64 module aida08 failed
FEE64 module aida09 failed
FEE64 module aida10 failed
FEE64 module aida11 failed
FEE64 module aida12 failed
FEE64 module aida13 failed
FEE64 module aida14 failed
FEE64 module aida15 failed
FEE64 module aida16 failed
Calibration test result: Passed 0, Failed 16

If any modules fail calibration , check the clock status and open the FADC Align and Control browser page to rerun calibration for that module

White rabbit decoder status

   
         Base         Current     Difference
aida05 fault      0x4da :      0x4e1 :      7  
White Rabbit error counter test result: Passed 15, Failed 1

Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WRx

 

19:45

Usecb scalars have continued to show DSSD implant rates since the last recorded elog check.

Statistics  - See attachment 7

Temperatures - See attachment 8

Bias & Leakage currents - See attachment 9

System Wide checks - All passed, EXCEPT:

ADC Calibration:

   
FEE64 module aida01 failed
FEE64 module aida02 failed
FEE64 module aida03 failed
FEE64 module aida04 failed
FEE64 module aida05 failed
FEE64 module aida06 failed
FEE64 module aida07 failed
FEE64 module aida08 failed
FEE64 module aida09 failed
FEE64 module aida10 failed
FEE64 module aida11 failed
FEE64 module aida12 failed
FEE64 module aida13 failed
FEE64 module aida14 failed
FEE64 module aida15 failed
FEE64 module aida16 failed
Calibration test result: Passed 0, Failed 16

If any modules fail calibration , check the clock status and open the FADC Align and Control browser page to rerun calibration for that module


White Rabbit decoder status:

     Base         Current     Difference
aida05 fault      0x4da :      0x4e1 :      7  
White Rabbit error counter test result: Passed 15, Failed 1

Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR

 

19:50 - Just as completing checks, notced usesb rates not showing DSSD implants again for several minutes

19:55 - Beam now back again - DSSD implants showing

 

 

  319   Fri May 14 19:01:09 2021 JSFriday 14th May 20:00-00:00
20:00 Taking over from CB

20:33 Doing full checks.
Temps ok elog:319/8  
Bias ok elog:319/1
Clock check - ok
ADC Calc check - ok
White Rabbit - 
		 Base 		Current 	Difference
aida01 fault 	 0xf932 : 	 0xf934 : 	 2  
aida02 fault 	 0x62ec : 	 0x62ee : 	 2  
aida03 fault 	 0x8679 : 	 0x867b : 	 2  
aida04 fault 	 0xf0e4 : 	 0xf0e6 : 	 2  
aida05 fault 	 0x9db8 : 	 0x9dc5 : 	 13  
aida06 fault 	 0x7f18 : 	 0x7f1a : 	 2  
aida07 fault 	 0xdd2c : 	 0xdd2e : 	 2  
aida08 fault 	 0x1557 : 	 0x1559 : 	 2  
White Rabbit error counter test result: Passed 8, Failed 8

FPGA Timestamp errors -
		 Base 		Current 		Difference
aida05 fault 	 0x0 : 	 0x3 : 	 3  
FPGA Timestamp error counter test result: Passed 15, Failed 1

Memory Info-
Returned 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0  
Mem(KB) :	4	8	16	32	64	128	256	512	1k	2k	4k
aida01 :     35     34     30      8      1      4      1      2      2      3      6   : 35772
aida02 :     37     30     28      5      3      1      2      4      2      3      6   : 36644
aida03 :     41     29     30     16      4      3      1      5      1      3      6   : 36588
aida04 :     13      6     23     12      4      4      4      4      2      2      6   : 35412
aida05 :     20     32     23     10      2      2      2      2      1      4      6   : 36736
aida06 :     14      5     10      5      1      3      2      3      2      3      6   : 35680
aida07 :     44     23     22      9      4      3      1      3      2      3      6   : 36200
aida08 :     39     30     28      9      0      3      2      3      3      2      6   : 35308
aida09 :     41     31     28     10      2      5      1      3      1      3      6   : 35484
aida10 :     43     23     27      7      3      1      1      3      3      3      6   : 36916
aida11 :     39     35     17     10      3      3      2      2      2      3      6   : 35908
aida12 :     37     36     15     10      2      2      2      4      3      2      6   : 35684
aida13 :     47     32     17      7      4      2      3      4      3      2      6   : 36012
aida14 :     23     33     14     10      3      4      3      4      3      2      6   : 36164
aida15 :     44     28     24     12      3      2      4      3      3      2      6   : 35920
aida16 :     22     17     24     17      3      4      2      3      1      3      6   : 35648

Stats
Aida Correlation Info #8 - elog:319/2 same as elog:317/27 
Resume info #3 - elog:319/3 same as elog:317/26 
Pause info #2 - elog:319/4 same as elog:317/32 
AIDA disk info #6 - elog:319/5 same as elog:317/31 
AIDA ADC data items - elog:319/6 same as elog:317/30 
Good Events - elog:319/7 same as elog:317/29
[I had some artifacts in AnyDesk when looking at the stats pages with many zeros, but could see 
clearly once uploaded the screenshots] 

Analysis R5_108 elog:319/9
Pause 149 Resume 150
Highest deadtime FEE10 3%

21:01
ucesb ok - Max 1700 Hz Implant - 1MHz Decay

21:35
Temps ok
Bias ok
Clock check - ok
ADC Calc check - ok
White Rabbit - 
		 Base 		Current 	Difference
aida01 fault 	 0xf932 : 	 0xf934 : 	 2  
aida02 fault 	 0x62ec : 	 0x62ee : 	 2  
aida03 fault 	 0x8679 : 	 0x867b : 	 2  
aida04 fault 	 0xf0e4 : 	 0xf0e6 : 	 2  
aida05 fault 	 0x9db8 : 	 0x9dc5 : 	 13  
aida06 fault 	 0x7f18 : 	 0x7f1a : 	 2  
aida07 fault 	 0xdd2c : 	 0xdd2e : 	 2  
aida08 fault 	 0x1557 : 	 0x1559 : 	 2  
White Rabbit error counter test result: Passed 8, Failed 8

			 Base 		Current 		Difference
aida05 fault 	 0x0 : 	 0x3 : 	 3  
FPGA Timestamp error counter test result: Passed 15, Failed 1

Returned 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0  
Mem(KB) :	4	8	16	32	64	128	256	512	1k	2k	4k
aida01 :     39     34     29      8      1      4      1      2      2      3      6   : 35772
aida02 :     46     29     28      5      3      2      2      3      2      3      6   : 36288
aida03 :     39     25     29     16      4      2      2      5      1      3      6   : 36660
aida04 :     14      6     19     12      4      4      3      4      2      2      6   : 35096
aida05 :     20     30     24     10      2      2      2      2      1      4      6   : 36736
aida06 :     26      4     11      5      2      4      3      2      2      3      6   : 35672
aida07 :     46     22     22      9      4      3      1      3      2      3      6   : 36200
aida08 :     28     32     28      9      0      2      2      4      3      2      6   : 35664
aida09 :     32     30     29     10      3      5      2      3      1      3      6   : 35776
aida10 :     39     23     25      7      3      1      1      3      3      3      6   : 36868
aida11 :     40     31     21     10      3      3      1      4      1      3      6   : 35688
aida12 :     17     31     12     10      2      2      3      4      3      2      6   : 35772
aida13 :     35     28     22      7      4      2      3      4      3      2      6   : 36012
aida14 :     30     33     14     10      3      4      3      4      3      2      6   : 36192
aida15 :     34     26     24     12      3      2      4      3      3      2      6   : 35864
aida16 :     38     24     27     17      3      4      2      3      1      3      6   : 35816

Stats
Good Events - elog:319/10
AIDA ADC data items - elog:319/11
AIDA disk info #6 - elog:319/12 
Pause info #2 - elog:319/13
Resume info #3 - elog:319/14 
Aida Correlation Info #8 - elog:319/15  - rates showed zero, checked again and normal elog:319/17

Analysis R5_128 elog:319/16
Pause 167 Resume 166
Highest deadtime FEE10 5%

22:00
Temps ok
Stats ok 
ucesb ok

22:35 
Temps ok
Stats ok 
ucesb- DSSD 2 is reading consistently 30% lower, not sure if this is dead time issue cropping up 
again elog:319/17

22:54 They are closing the file, we are on R5_155

23:00 
  318   Fri May 14 17:30:56 2021 TDAnalysis of S496 files R5_18, R5_32 & R5_45
S496/R5_18 - attachment 1

LEC fast discriminators - *all* channels disabled

Slow comparator thresholds - p+n FEE64s 0xa *except* aida09 0xf
                             n+n FEE64s 0x20
Sampling ADCs OFF


S496/R5_32 - attachment 2

LEC fast discriminators - *all* channels disabled

Slow comparator thresholds - p+n FEE64s 0xa *except* aida09 & aida15 0xf
                             n+n FEE64s 0x20

Sampling ADCs ON, trigger threshold 15000 


S496/R5_45 - attachment 3

LEC fast discriminators - *all* channels disabled

Slow comparator thresholds - p+n FEE64s 0xc
                             n+n FEE64s 0x20

Sampling ADCs ON, trigger threshold 15000 


* N.B. Ignore 'elapsed idle time(s)' metric which does not work with WR timestamps
 
  317   Fri May 14 14:57:32 2021 CB + OH, TD, NHFriday 14th May 16:00-20:00
16:00 Shift taken over from CA

16:00 
Temperatures OK. Attach 1. 
Upon reloading aida11 first returns "No respose" for all values. 
Upon refreshing, aida11 responds but aida 9 returns "No response". 
Upon refreshing again, attach 1.

Bias OK. Attach 2.

Statistics OK. Attach 3-7

System-wide checks
Clock OK
ADC - all fail
	 Base 		Current 	Difference
aida01 fault 	 0xf932 : 	 0xf933 : 	 1  
aida02 fault 	 0x62ec : 	 0x62ed : 	 1  
aida03 fault 	 0x8679 : 	 0x867a : 	 1  
aida04 fault 	 0xf0e4 : 	 0xf0e5 : 	 1  
aida05 fault 	 0x9db8 : 	 0x9dc3 : 	 11  
aida06 fault 	 0x7f18 : 	 0x7f19 : 	 1  
aida07 fault 	 0xdd2c : 	 0xdd2d : 	 1  
aida08 fault 	 0x1557 : 	 0x1558 : 	 1  
White Rabbit error counter test result: Passed 8, Failed 8

Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR


			 Base 		Current 		Difference
aida05 fault 	 0x0 : 	 0x3 : 	 3  
FPGA Timestamp error counter test result: Passed 15, Failed 1
If any of these counts are reported as in error
The ASIC readout system has detected a timeslip.
That is the timestamp read from the time FIFO is not younger than the last

Returned 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0  
Mem(KB) :	4	8	16	32	64	128	256	512	1k	2k	4k
aida01 :     32     28     25      7      1      4      1      3      2      3      6   : 36112
aida02 :     43     28     29      5      4      2      3      4      1      3      6   : 36092
aida03 :     39     26     28      5      4      3      2      5      1      3      6   : 36428
aida04 :      9      5     21      8      5      4      4      4      2      2      6   : 35292
aida05 :     46     32     26      7      2      2      2      2      2      3      6   : 35768
aida06 :      4      5     16      1      3      4      2      4      1      3      6   : 35352
aida07 :     47     21     25      5      4      3      1      4      1      3      6   : 35604
aida08 :     62     33     32      6      0      2      3      3      3      2      6   : 35520
aida09 :     33     25     28      7      3      5      2      4      1      3      6   : 36140
aida10 :     47     29     22      6      3      2      2      2      2      3      6   : 35716
aida11 :     39     30     29      6      2      3      1      3      2      3      6   : 36124
aida12 :     38     36     20      6      3      2      3      3      3      2      6   : 35448
aida13 :     42     29     24      3      4      2      3      3      3      2      6   : 35440
aida14 :     31     28     18      7      3      3      4      4      2      2      6   : 35228
aida15 :     48     25     30      5      3      2      4      4      2      2      6   : 35272
aida16 :     37     16     28      6      3      3      2      2      2      3      6   : 35796

 Collecting the file size of each FEE64 Options CONTENTS file to check they are all the same

 FEE : aida01 =>   Options file size is 1026 	 Last changed Thu May 13 05:49:42 CEST 2021
 FEE : aida02 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:43:46 CEST 2021
 FEE : aida03 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:43:50 CEST 2021
 FEE : aida04 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:43:53 CEST 2021
 FEE : aida05 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:43:55 CEST 2021
 FEE : aida06 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:43:59 CEST 2021
 FEE : aida07 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:44:02 CEST 2021
 FEE : aida08 =>   Options file size is 1025 	 Last changed Wed May 05 12:15:54 CEST 2021
 FEE : aida09 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:44:08 CEST 2021
 FEE : aida10 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:44:57 CEST 2021
 FEE : aida11 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:44:57 CEST 2021
 FEE : aida12 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:44:57 CEST 2021
 FEE : aida13 =>   Options file size is 1025 	 Last changed Fri May 07 19:40:34 CEST 2021
 FEE : aida14 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:44:57 CEST 2021
 FEE : aida15 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:44:57 CEST 2021
 FEE : aida16 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:44:57 CEST 2021



16:15 Control handed over to OH temporarily.
16:25 AIDA stopped. OH attempts to update Merger. Discriminators off.
16:29 AIDA restart begins.
16:33 AIDA restarted.
16:40 Dead time still very high ~50% in most FEEs.
16:50 Checking effect of threshold on dead time. 
16:53 Checking waveform effect on deadtime. DAQ stopped. Turning waveforms on, threshold increased to max. This
*reduces* deadtime to <10%.
      Sampling ADCs were disabled (switched off) late Wednesday evening.
17:01 DAQ can be restarted.
17:27 Finished checking effect of threshold on dead time. 120 keV front. 320 keV back.
17:33 Analysis of R5_45 - x strips 0xc - y strips 0x20



17:40 OH hands back control.
Temperature OK - attach 10
Bias OK - attach 11
Stats - attach 12-17

System-wide checks
Clock - all pass
ADC - all pass
WR
	 Base 		Current 	Difference
aida01 fault 	 0xf932 : 	 0xf933 : 	 1  
aida02 fault 	 0x62ec : 	 0x62ed : 	 1  
aida03 fault 	 0x8679 : 	 0x867a : 	 1  
aida04 fault 	 0xf0e4 : 	 0xf0e5 : 	 1  
aida05 fault 	 0x9db8 : 	 0x9dc3 : 	 11  
aida06 fault 	 0x7f18 : 	 0x7f19 : 	 1  
aida07 fault 	 0xdd2c : 	 0xdd2d : 	 1  
aida08 fault 	 0x1557 : 	 0x1558 : 	 1  
White Rabbit error counter test result: Passed 8, Failed 8

Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR
	
			 Base 		Current 		Difference
aida05 fault 	 0x0 : 	 0x3 : 	 3  
FPGA Timestamp error counter test result: Passed 15, Failed 1
If any of these counts are reported as in error
The ASIC readout system has detected a timeslip.
That is the timestamp read from the time FIFO is not younger than the last
	
Returned 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0  
Mem(KB) :	4	8	16	32	64	128	256	512	1k	2k	4k
aida01 :     56     35     27      8      1      4      1      2      2      3      6   : 35816
aida02 :     43     31     27      5      3      1      3      3      2      3      6   : 36404
aida03 :     45     28     28     16      4      2      2      4      1      3      6   : 36180
aida04 :      2      2     18     12      4      3      3      4      2      2      6   : 34872
aida05 :     28     28     24     10      2      2      2      2      1      4      6   : 36752
aida06 :     28     11      9      5      2      3      3      3      1      3      6   : 35064
aida07 :     52     24     23      9      4      3      1      2      2      3      6   : 35744
aida08 :     21     26     27      8      0      2      3      3      3      2      6   : 35284
aida09 :     40     32     26     10      3      5      1      3      1      3      6   : 35520
aida10 :     46     23     27      7      3      1      1      3      3      3      6   : 36928
aida11 :     49     31     20     10      2      2      1      2      2      3      6   : 35516
aida12 :     39     35     17      9      2      2      3      4      3      2      6   : 35940
aida13 :     40     32     20      7      4      2      3      4      3      2      6   : 36032
aida14 :     33     33     16     10      3      4      3      4      3      2      6   : 36236
aida15 :     53     27     25     12      3      2      4      4      2      2      6   : 35452
aida16 :     35     22     32     15      3      4      3      3      1      3      6   : 36060

 Collecting the file size of each FEE64 Options CONTENTS file to check they are all the same

 FEE : aida01 =>   Options file size is 1026 	 Last changed Thu May 13 05:49:42 CEST 2021
 FEE : aida02 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:43:46 CEST 2021
 FEE : aida03 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:43:50 CEST 2021
 FEE : aida04 =>   Options file size is 1025 	 Last changed Fri May 14 16:54:56 CEST 2021
 FEE : aida05 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:43:55 CEST 2021
 FEE : aida06 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:43:59 CEST 2021
 FEE : aida07 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:44:02 CEST 2021
 FEE : aida08 =>   Options file size is 1025 	 Last changed Wed May 05 12:15:54 CEST 2021
 FEE : aida09 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:44:08 CEST 2021
 FEE : aida10 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:44:57 CEST 2021
 FEE : aida11 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:44:57 CEST 2021
 FEE : aida12 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:44:57 CEST 2021
 FEE : aida13 =>   Options file size is 1025 	 Last changed Fri May 07 19:40:34 CEST 2021
 FEE : aida14 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:44:57 CEST 2021
 FEE : aida15 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:44:57 CEST 2021
 FEE : aida16 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:44:57 CEST 2021

18:30
Temperatures - OK (attach 18)
Bias OK (attach 19)
Stats - attach 20-24

System-wide checks
Clock - all pass
ADC - all pass
	
		 Base 		Current 	Difference
aida01 fault 	 0xf932 : 	 0xf933 : 	 1  
aida02 fault 	 0x62ec : 	 0x62ed : 	 1  
aida03 fault 	 0x8679 : 	 0x867a : 	 1  
aida04 fault 	 0xf0e4 : 	 0xf0e5 : 	 1  
aida05 fault 	 0x9db8 : 	 0x9dc3 : 	 11  
aida06 fault 	 0x7f18 : 	 0x7f19 : 	 1  
aida07 fault 	 0xdd2c : 	 0xdd2d : 	 1  
aida08 fault 	 0x1557 : 	 0x1558 : 	 1  
White Rabbit error counter test result: Passed 8, Failed 8

Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR
	
			 Base 		Current 		Difference
aida05 fault 	 0x0 : 	 0x3 : 	 3  
FPGA Timestamp error counter test result: Passed 15, Failed 1
If any of these counts are reported as in error
The ASIC readout system has detected a timeslip.
That is the timestamp read from the time FIFO is not younger than the last
	
Returned 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0  
Mem(KB) :	4	8	16	32	64	128	256	512	1k	2k	4k
aida01 :     41     31     31      8      1      4      1      2      2      3      6   : 35788
aida02 :     42     31     29      5      3      2      4      4      1      3      6   : 36304
aida03 :     29     30     29     16      4      3      2      4      1      3      6   : 36276
aida04 :     28      4     20     12      4      3      4      4      2      2      6   : 35280
aida05 :     25     28     23     10      2      2      2      2      1      4      6   : 36724
aida06 :      5      3     10      5      2      4      2      2      2      3      6   : 35308
aida07 :     38     26     22      9      4      3      1      2      2      3      6   : 35688
aida08 :     52     30     28      8      0      2      2      3      3      2      6   : 35200
aida09 :     22     26     21     10      3      4      1      4      1      3      6   : 35704
aida10 :     37     28     25      7      3      1      1      3      3      3      6   : 36900
aida11 :     53     30     20     10      2      3      2      2      2      3      6   : 35908
aida12 :     37     34     17     10      2      2      3      4      3      2      6   : 35956
aida13 :     47     34     16      7      4      2      3      4      3      2      6   : 36012
aida14 :     37     29     17     10      3      4      3      4      3      2      6   : 36236
aida15 :     44     28     25     12      3      2      4      4      2      2      6   : 35424
aida16 :     40     21     33     15      3      4      1      3      1      3      6   : 35576

18:40 Analysed R5_67. Attached. Dead time improved. Max ~1.5%


19:40 Temperatures OK
Bias OK - attach 26
Stats - attach 27-32

System-wide checks

Clock, ADC - all pass
WR

	
		 Base 		Current 	Difference
aida01 fault 	 0xf932 : 	 0xf934 : 	 2  
aida02 fault 	 0x62ec : 	 0x62ee : 	 2  
aida03 fault 	 0x8679 : 	 0x867b : 	 2  
aida04 fault 	 0xf0e4 : 	 0xf0e6 : 	 2  
aida05 fault 	 0x9db8 : 	 0x9dc4 : 	 12  
aida06 fault 	 0x7f18 : 	 0x7f1a : 	 2  
aida07 fault 	 0xdd2c : 	 0xdd2e : 	 2  
aida08 fault 	 0x1557 : 	 0x1559 : 	 2  
White Rabbit error counter test result: Passed 8, Failed 8

Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR
	
			 Base 		Current 		Difference
aida05 fault 	 0x0 : 	 0x3 : 	 3  
FPGA Timestamp error counter test result: Passed 15, Failed 1
If any of these counts are reported as in error
The ASIC readout system has detected a timeslip.
That is the timestamp read from the time FIFO is not younger than the last
	
Returned 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0  
Mem(KB) :	4	8	16	32	64	128	256	512	1k	2k	4k
aida01 :     41     33     30      8      1      4      1      2      2      3      6   : 35788
aida02 :     51     30     30      5      3      2      3      4      1      3      6   : 36092
aida03 :     52     31     28     16      4      2      2      5      1      3      6   : 36744
aida04 :      6      8     20     12      4      4      3      4      2      2      6   : 35096
aida05 :     30     31     22     10      2      2      2      2      1      4      6   : 36752
aida06 :     12      6     10      5      2      3      3      2      2      3      6   : 35488
aida07 :     53     24     21      9      4      3      1      3      2      3      6   : 36228
aida08 :     36     28     28      9      0      2      3      3      3      2      6   : 35408
aida09 :     27     28     25     10      3      4      2      3      1      3      6   : 35548
aida10 :     50     23     26      7      3      1      1      3      3      3      6   : 36928
aida11 :     42     28     22     10      3      3      2      2      2      3      6   : 35944
aida12 :     39     35     15     10      2      2      2      4      3      2      6   : 35684
aida13 :     43     34     17      7      4      2      3      4      3      2      6   : 36012
aida14 :     42     32     16     10      3      4      3      4      3      2      6   : 36264
aida15 :     36     30     25     12      3      2      4      3      3      2      6   : 35920
aida16 :     40     21     25     15      3      4      2      3      1      3      6   : 35704

	
 Collecting the file size of each FEE64 Options CONTENTS file to check they are all the same

 FEE : aida01 =>   Options file size is 1026 	 Last changed Thu May 13 05:49:42 CEST 2021
 FEE : aida02 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:43:46 CEST 2021
 FEE : aida03 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:43:50 CEST 2021
 FEE : aida04 =>   Options file size is 1025 	 Last changed Fri May 14 16:54:56 CEST 2021
 FEE : aida05 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:43:55 CEST 2021
 FEE : aida06 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:43:59 CEST 2021
 FEE : aida07 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:44:02 CEST 2021
 FEE : aida08 =>   Options file size is 1025 	 Last changed Wed May 05 12:15:54 CEST 2021
 FEE : aida09 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:44:08 CEST 2021
 FEE : aida10 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:44:57 CEST 2021
 FEE : aida11 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:44:57 CEST 2021
 FEE : aida12 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:44:57 CEST 2021
 FEE : aida13 =>   Options file size is 1025 	 Last changed Fri May 07 19:40:34 CEST 2021
 FEE : aida14 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:44:57 CEST 2021
 FEE : aida15 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:44:57 CEST 2021
 FEE : aida16 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:44:57 CEST 2021

19:40 Analysed R5_86. Attached. Dead time as before.

19:57 Shift handed over to James S
  316   Fri May 14 07:05:08 2021 CAFriday 14th May 08:00 - 16:00 shift
08:00 CA takes over

08:28 Problem at UNILAC - no beam

09:30 all system wide checks ok *except*

      all FEE64 fail ADC calibration (no waveforms)

      WR decoder status:

		 Base 		Current 	Difference
aida01 fault 	 0xf932 : 	 0xf933 : 	 1  
aida02 fault 	 0x62ec : 	 0x62ed : 	 1  
aida03 fault 	 0x8679 : 	 0x867a : 	 1  
aida04 fault 	 0xf0e4 : 	 0xf0e5 : 	 1  
aida05 fault 	 0x9db8 : 	 0x9dc2 : 	 10  
aida06 fault 	 0x7f18 : 	 0x7f19 : 	 1  
aida07 fault 	 0xdd2c : 	 0xdd2d : 	 1  
aida08 fault 	 0x1557 : 	 0x1558 : 	 1  
White Rabbit error counter test result: Passed 8, Failed 8

Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR

    FPGA timestamp errors:

			 Base 		Current 		Difference
aida05 fault 	 0x0 : 	 0x2 : 	 2  
FPGA Timestamp error counter test result: Passed 15, Failed 1
If any of these counts are reported as in error
The ASIC readout system has detected a timeslip.
That is the timestamp read from the time FIFO is not younger than the last

09:36 FEE64 Temperatures ok - attachment 1

      Detector bias / leakage currents ok - attachment 2

      Statistics - attachments 3-8

09:50 beam is back - writing to file R4_265

09:56 FRS adjusting degrader settings (S4) - temporarily remove to check counts in scintillator vs AIDA

10:10 FRS increase degrader thickness

10.28 all histograms, stats and merger stats zero'd

10.50 all system wide checks ok *except*
		 Base 		Current 	Difference
aida01 fault 	 0xf932 : 	 0xf933 : 	 1  
aida02 fault 	 0x62ec : 	 0x62ed : 	 1  
aida03 fault 	 0x8679 : 	 0x867a : 	 1  
aida04 fault 	 0xf0e4 : 	 0xf0e5 : 	 1  
aida05 fault 	 0x9db8 : 	 0x9dc2 : 	 10  
aida06 fault 	 0x7f18 : 	 0x7f19 : 	 1  
aida07 fault 	 0xdd2c : 	 0xdd2d : 	 1  
aida08 fault 	 0x1557 : 	 0x1558 : 	 1  
White Rabbit error counter test result: Passed 8, Failed 8

Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR


			 Base 		Current 		Difference
aida05 fault 	 0x0 : 	 0x3 : 	 3  
FPGA Timestamp error counter test result: Passed 15, Failed 1
If any of these counts are reported as in error
The ASIC readout system has detected a timeslip.
That is the timestamp read from the time FIFO is not younger than the last


Returned 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0  
Mem(KB) :	4	8	16	32	64	128	256	512	1k	2k	4k
aida01 :      3      2      2      2      2      2      2      4      3      3      6   : 36860
aida02 :      7      3      5      1      2      5      2      4      3      3      6   : 37284
aida03 :      7      5      3      1      4      4      2      3      3      3      6   : 36756
aida04 :      3      1     20      4      2      3      3      5      1      3      6   : 36052
aida05 :     19     25     26      6      2      2      3      2      2      3      6   : 35828
aida06 :      1      3     16      1      3      4      1      5      1      3      6   : 35580
aida07 :      6      4      1      2      3      3      2      3      3      3      6   : 36552
aida08 :      8      5      2      3      2      3      2      2      2      4      6   : 37064
aida09 :      9      7      1      2      2      4      2      3      3      3      6   : 36652
aida10 :      2      5      3      1      1      4      2      3      3      3      6   : 36544
aida11 :     16      4      4      2      1      2      2      3      3      3      6   : 36384
aida12 :      2      1      1      3      1      2      2      3      3      3      6   : 36288
aida13 :      0      1      1      2      0      4      3      2      3      3      6   : 36184
aida14 :      6      3      2      0      2      3      2      3      3      3      6   : 36432
aida15 :     25      8      2      1      2      3      1      2      2      4      6   : 36836
aida16 :      3      5      1      2      2      2      3      2      3      3      6   : 36100


 Collecting the file size of each FEE64 Options CONTENTS file to check they are all the same

 FEE : aida01 =>   Options file size is 1026 	 Last changed Thu May 13 05:49:42 CEST 2021
 FEE : aida02 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:43:46 CEST 2021
 FEE : aida03 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:43:50 CEST 2021
 FEE : aida04 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:43:53 CEST 2021
 FEE : aida05 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:43:55 CEST 2021
 FEE : aida06 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:43:59 CEST 2021
 FEE : aida07 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:44:02 CEST 2021
 FEE : aida08 =>   Options file size is 1025 	 Last changed Wed May 05 12:15:54 CEST 2021
 FEE : aida09 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:44:08 CEST 2021
 FEE : aida10 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:44:57 CEST 2021
 FEE : aida11 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:44:57 CEST 2021
 FEE : aida12 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:44:57 CEST 2021
 FEE : aida13 =>   Options file size is 1025 	 Last changed Fri May 07 19:40:34 CEST 2021
 FEE : aida14 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:44:57 CEST 2021
 FEE : aida15 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:44:57 CEST 2021
 FEE : aida16 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:44:57 CEST 2021

10.50 DAQ continues S496/R4_304

11:18 FEE64 Temperatures ok - attachment 9

      Detector bias & leakage currents ok - attachment 10

      statistics - attachments 11-16

15:00 FEE64           avg. CPU usage (%)

      1               52
      2               55
      3               55
      4               60
      5               93
      6               50
      7               53
      8               55
      9               55
      10              51
      11              50
      12              67
      13              56
      14              55
      15              55
      16              58

15:08 all system wide checks ok *except*

      all FEE64 fail ADC calibration (no waveform)

      WR decoder status:

      Base 		Current 	Difference
aida01 fault 	 0xf932 : 	 0xf933 : 	 1  
aida02 fault 	 0x62ec : 	 0x62ed : 	 1  
aida03 fault 	 0x8679 : 	 0x867a : 	 1  
aida04 fault 	 0xf0e4 : 	 0xf0e5 : 	 1  
aida05 fault 	 0x9db8 : 	 0x9dc3 : 	 11  
aida06 fault 	 0x7f18 : 	 0x7f19 : 	 1  
aida07 fault 	 0xdd2c : 	 0xdd2d : 	 1  
aida08 fault 	 0x1557 : 	 0x1558 : 	 1  
White Rabbit error counter test result: Passed 8, Failed 8

Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR

    FPGA errors:

 Base 		Current 		Difference
aida05 fault 	 0x0 : 	 0x3 : 	 3  
FPGA Timestamp error counter test result: Passed 15, Failed 1
If any of these counts are reported as in error
The ASIC readout system has detected a timeslip.
That is the timestamp read from the time FIFO is not younger than the last

15:10 Temperatures ok - attachment 17

      Detector bias / leakage current ok - attachment 18

      Statistics - attachments 18 -24
  315   Thu May 13 23:58:05 2021 OHFriday 14th May 00:00 - 08:00
00:00 OH Takes over.
 PULSER SETTINGS
---------------------------
 Pulse is ON
 Positive Tail Pulse
 Trigger Source is Internal Clock
 Trigger Threshold is 3.5
 Amplitude   : 2.0 Volts
 Rep Rate    : 2.0 hZ
 Delay       : 250.0 ns
 Fall Time   : 1 ms
 Attenuation : 1
 Display is  : Volts
 Equivalent keV is : 200.0
 Ramp Start at 0.01 Volts
 Ramp Stop  at 9.99 Volts
 Ramp Start at 1.0 keV
 Ramp Stop  at 999.0 keV
 Ramp Time  is 60 seconds
 # Ramp Cycles is 1

OK
0x210 

 p+n Thresholds 0xa (100keV)
 n+n Threshold 0x20 (320keV)

01:00 System wide checks
 Clock status ok
 ADC calibration fail as expected (Waveforms off)
 WR Errors
		 Base 		Current 	Difference
aida01 fault 	 0xf932 : 	 0xf933 : 	 1  
aida02 fault 	 0x62ec : 	 0x62ed : 	 1  
aida03 fault 	 0x8679 : 	 0x867a : 	 1  
aida04 fault 	 0xf0e4 : 	 0xf0e5 : 	 1  
aida05 fault 	 0x9db8 : 	 0x9dbf : 	 7  
aida06 fault 	 0x7f18 : 	 0x7f19 : 	 1  
aida07 fault 	 0xdd2c : 	 0xdd2d : 	 1  
aida08 fault 	 0x1557 : 	 0x1558 : 	 1  
White Rabbit error counter test result: Passed 8, Failed 8

Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR

 FPGA Errors
	
                Base Current Difference
aida05 fault 	 0x0 : 	 0x2 : 	 2  
FPGA Timestamp error counter test result: Passed 15, Failed 1
If any of these counts are reported as in error
The ASIC readout system has detected a timeslip.
That is the timestamp read from the time FIFO is not younger than the last

 Statistics ok - attachment 1
 Temp ok - attachment 2
 Bias and leakage current ok - attachment 3

01:42 Comparing the CS of AIDA3 and 4
      Stats have been running for a while (Believe since 5am)
      Looking at the counter ratio - attachment 4
      We still have a deadtime of around 14% in FEE4 (Assuming FEE3 has no dead time)

01:56 Noticed /media/ThirdDrive was at 97% full
      Let DESPEC locals know I would like to use the next run stop to change the directory.
      They said they could do it now so they stopped their run.
      Changed the symbolic link to /media/1e2....
      Now writing to R4_

03:26 MBS DAQ Stopped as beam stopped
03:33 Beam back now and DAQ running again

04:44 System wide checks as before
      Statistics ok - attachment 5
      Temperatures ok - attachment - 7
      Bias and leakage currents ok - attachment 6

05:53 MBS stops run as beam drops
06:00 MBS started a new run as beam is back

07:37 System wide checks clock ok and FPGA as before
      WR additional difference on FEE5
	
		 Base 		Current 	Difference
aida01 fault 	 0xf932 : 	 0xf933 : 	 1  
aida02 fault 	 0x62ec : 	 0x62ed : 	 1  
aida03 fault 	 0x8679 : 	 0x867a : 	 1  
aida04 fault 	 0xf0e4 : 	 0xf0e5 : 	 1  
aida05 fault 	 0x9db8 : 	 0x9dc0 : 	 8  
aida06 fault 	 0x7f18 : 	 0x7f19 : 	 1  
aida07 fault 	 0xdd2c : 	 0xdd2d : 	 1  
aida08 fault 	 0x1557 : 	 0x1558 : 	 1  
White Rabbit error counter test result: Passed 8, Failed 8

Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR

     Statistics ok - attachment 8
     Temp ok - attachment 9
     Bias and leakage ok - attachment 10
 
  314   Thu May 13 21:21:25 2021 Philippos and Marcsystem wide checks

Wide checks at Time 22:22 CET

Clock status test result: Passed 16, Failed 0

Understand status as follows
Status bit 3 : firmware PLL that creates clocks from external clock not locked
Status bit 2 : always logic '1'
Status bit 1 : LMK3200(2) PLL and clock distribution chip not locked to external clock
Status bit 0 : LMK3200(1) PLL and clock distribution chip not locked to external clock
If all these bits are not set then the operation of the firmware is unreliable

FEE64 module aida01 failed
FEE64 module aida02 failed
FEE64 module aida03 failed
FEE64 module aida04 failed
FEE64 module aida05 failed
FEE64 module aida06 failed
FEE64 module aida07 failed
FEE64 module aida08 failed
FEE64 module aida09 failed
FEE64 module aida10 failed
FEE64 module aida11 failed
FEE64 module aida12 failed
FEE64 module aida13 failed
FEE64 module aida14 failed
FEE64 module aida15 failed
FEE64 module aida16 failed
Calibration test result: Passed 0, Failed 16

If any modules fail calibration , check the clock status and open the FADC Align and Control browser page to rerun calibration for that module


         Base         Current     Difference
aida01 fault      0xf932 :      0xf933 :      1  
aida02 fault      0x62ec :      0x62ed :      1  
aida03 fault      0x8679 :      0x867a :      1  
aida04 fault      0xf0e4 :      0xf0e5 :      1  
aida05 fault      0x9db8 :      0x9dbc :      4  
aida06 fault      0x7f18 :      0x7f19 :      1  
aida07 fault      0xdd2c :      0xdd2d :      1  
aida08 fault      0x1557 :      0x1558 :      1  
White Rabbit error counter test result: Passed 8, Failed 8

Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR

             Base         Current         Difference
aida05 fault      0x0 :      0x2 :      2  
FPGA Timestamp error counter test result: Passed 15, Failed 1
If any of these counts are reported as in error
The ASIC readout system has detected a timeslip.
That is the timestamp read from the time FIFO is not younger than the last


Returned 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0  
Mem(KB) :    4    8    16    32    64    128    256    512    1k    2k    4k
aida01 :     35     11      3      5      5      4      1      3      2      4      6   : 37876
aida02 :     20      5      4      3      2      5      2      4      3      3      6   : 37400
aida03 :     20      3      4      3      4      4      1      3      2      4      6   : 37640
aida04 :     25      6      2      3      4      4      3      3      3      3      6   : 37140
aida05 :     30     10      5      7      2      6      2      4      2      4      6   : 38776
aida06 :     21     11      4      4      3      5      2      4      3      3      6   : 37548
aida07 :     23      8      4      4      2      6      1      3      2      4      6   : 37852
aida08 :     27      7      2      3      4      3      2      4      3      3      6   : 37284
aida09 :     31      6      5      6      3      3      2      4      3      3      6   : 37372
aida10 :     22      5      3      4      3      6      2      2      2      4      6   : 37616
aida11 :     23      5      4      2      5      2      1      3      2      4      6   : 37444
aida12 :     20      7      4      7      3      5      2      3      3      3      6   : 37096
aida13 :     28      5      3      3      1      5      2      3      3      3      6   : 36840
aida14 :     38     12      2      4      2      4      1      4      3      3      6   : 37144
aida15 :     27      9      5      4      2      5      3      2      3      3      6   : 36740
aida16 :     21      6      4      4      1      4      2      3      3      3      6   : 36740

 

  313   Thu May 13 19:52:36 2021 Philippos and MarcConnection problem

At 20:15, We lost connection with AnyDesk.

Liliana restarted Anydesk on the AIDA DAQ PC. All seemed fine on AIDA DAQ but there was a more generic issue on the main DAQ, possibly the timesorter.

Helena restarted all after ~30mn and it's all working again now.

 

 

  312   Thu May 13 17:05:42 2021 Muneerah, MArcThu 13 May 04:00-00:00

19:08 System Check

Rates from ucesb below 1 kHz

Atachment 1 Stats

Atachment 2 Temperatures

Atachment 3 Voltages

System wide checks are all ok except

ADC calibration all failed  
FEE64 module aida01 failed
FEE64 module aida02 failed
FEE64 module aida03 failed
FEE64 module aida04 failed
FEE64 module aida05 failed
FEE64 module aida06 failed
FEE64 module aida07 failed
FEE64 module aida08 failed
FEE64 module aida09 failed
FEE64 module aida10 failed
FEE64 module aida11 failed
FEE64 module aida12 failed
FEE64 module aida13 failed
FEE64 module aida14 failed
FEE64 module aida15 failed
FEE64 module aida16 failed
Calibration test result: Passed 0, Failed 16

If any modules fail calibration , check the clock status and open the FADC Align and Control browser page to rerun calibration for that module

 

White rabbit  
         Base         Current     Difference
aida01 fault      0xf932 :      0xf933 :      1  
aida02 fault      0x62ec :      0x62ed :      1  
aida03 fault      0x8679 :      0x867a :      1  
aida04 fault      0xf0e4 :      0xf0e5 :      1  
aida05 fault      0x9db8 :      0x9dba :      2  
aida06 fault      0x7f18 :      0x7f19 :      1  
aida07 fault      0xdd2c :      0xdd2d :      1  
aida08 fault      0x1557 :      0x1558 :      1  
White Rabbit error counter test result: Passed 8, Failed 8

Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR

FPGA all passed

Memory information

   
Returned 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0  
Mem(KB) :    4    8    16    32    64    128    256    512    1k    2k    4k
aida01 :      6      2      2      1      2      4      3      4      2      3      7   : 40424
aida02 :      3      4      2      2      1      2      1      3      3      3      7   : 40140
aida03 :     10     10      0      3      2      4      2      4      2      3      7   : 40280
aida04 :      3      2      0      2      2      3      2      3      1      4      7   : 40540
aida05 :      4      9      2      3      2      4      3      3      1      4      7   : 41048
aida06 :      8      6      1      1      2      3      2      2      3      3      7   : 40064
aida07 :      1      4      0      1      1      3      1      3      3      3      7   : 40196
aida08 :      1      4      2      2      2      4      3      4      2      3      7   : 40452
aida09 :      8      5      2      2      1      2      1      3      3      3      7   : 40168
aida10 :     22      5      4      2      1      3      1      4      2      3      7   : 39872
aida11 :      2      1      2      1      2      2      3      3      2      3      7   : 39632
aida12 :      8      5      2      2      1      3      1      4      2      3      7   : 39784
aida13 :     18      8      2      1      0      2      3      3      2      3      7   : 39624
aida14 :     23     12      2      2      1      2      2      4      2      3      7   : 40028
aida15 :      3      5      3      2      2      3      1      4      2      3      7   : 39844
aida16 :      0      2      1      2      1      4      2      4      2      3      7   : 40096

 

 

 

  311   Thu May 13 07:10:17 2021 CA, TDMay 13th 08:00 - 16:00 shift
08:10 all system wide checks ok *except*

      all FEE64 modules fail ADC calibration

      WR decoder status:

      Base 		Current 	Difference
aida01 fault 	 0xf932 : 	 0xf933 : 	 1  
aida02 fault 	 0x62ec : 	 0x62ed : 	 1  
aida03 fault 	 0x8679 : 	 0x867a : 	 1  
aida04 fault 	 0xf0e4 : 	 0xf0e5 : 	 1  
aida05 fault 	 0x9db8 : 	 0x9db9 : 	 1  
aida06 fault 	 0x7f18 : 	 0x7f19 : 	 1  
aida07 fault 	 0xdd2c : 	 0xdd2d : 	 1  
aida08 fault 	 0x1557 : 	 0x1558 : 	 1  
White Rabbit error counter test result: Passed 8, Failed 8

Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR

08:16 attempt to recalibrate ADCs using FADC align and control - unsuccessful on all

08:18 FEE64 temps ok - attachment 1

      bias/leakage currents ok - attachment 2

      statistics - attachments 3-8

10:13 all system wide checks ok *except*

      all FEE64 modules fail ADC calibration

      WR decoder status:

		 Base 		Current 	Difference
aida01 fault 	 0xf932 : 	 0xf933 : 	 1  
aida02 fault 	 0x62ec : 	 0x62ed : 	 1  
aida03 fault 	 0x8679 : 	 0x867a : 	 1  
aida04 fault 	 0xf0e4 : 	 0xf0e5 : 	 1  
aida05 fault 	 0x9db8 : 	 0x9db9 : 	 1  
aida06 fault 	 0x7f18 : 	 0x7f19 : 	 1  
aida07 fault 	 0xdd2c : 	 0xdd2d : 	 1  
aida08 fault 	 0x1557 : 	 0x1558 : 	 1  
White Rabbit error counter test result: Passed 8, Failed 8

Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR

10:16 FEE64 temps ok - attachment 9

      bias/leakage currents ok - attachment 10

      statistics - attachments 11-16

12:16 all system wide checks ok *except*

      all FEE64 modules fail ADC calibration

      WR decoder status:

		 Base 		Current 	Difference
aida01 fault 	 0xf932 : 	 0xf933 : 	 1  
aida02 fault 	 0x62ec : 	 0x62ed : 	 1  
aida03 fault 	 0x8679 : 	 0x867a : 	 1  
aida04 fault 	 0xf0e4 : 	 0xf0e5 : 	 1  
aida05 fault 	 0x9db8 : 	 0x9dba : 	 2  
aida06 fault 	 0x7f18 : 	 0x7f19 : 	 1  
aida07 fault 	 0xdd2c : 	 0xdd2d : 	 1  
aida08 fault 	 0x1557 : 	 0x1558 : 	 1  
White Rabbit error counter test result: Passed 8, Failed 8

Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR

12:16 FEE64 temps ok - attachment 17

      bias/leakage currents ok - attachment 18

      statistics - attachments 19-25

14:08 all system wide checks ok *except*

      all FEE64 modules fail ADC calibration

      WR decoder status:

		 Base 		Current 	Difference
aida01 fault 	 0xf932 : 	 0xf933 : 	 1  
aida02 fault 	 0x62ec : 	 0x62ed : 	 1  
aida03 fault 	 0x8679 : 	 0x867a : 	 1  
aida04 fault 	 0xf0e4 : 	 0xf0e5 : 	 1  
aida05 fault 	 0x9db8 : 	 0x9dba : 	 2  
aida06 fault 	 0x7f18 : 	 0x7f19 : 	 1  
aida07 fault 	 0xdd2c : 	 0xdd2d : 	 1  
aida08 fault 	 0x1557 : 	 0x1558 : 	 1  
White Rabbit error counter test result: Passed 8, Failed 8

Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR

FEE64 temps ok - attachment 16

      bias/leakage currents ok - attachment 27

      statistics - attachments 28-35
  310   Thu May 13 00:51:40 2021 Muneerah and OHThu 13 May 00:00-08:00
01:50 Beam on
The leakage current and the voltages been monitered every half an hour and logged in the excel sheet, looks decreasing.
Tempetratures OK

Rates OK

Attachment 1 Voltages

Atachment 2 Rates

Atachment 3 Temperatures

Attachment 4 ucesb rates less than 1 kHz


System wide check are ok except:

Base Current Difference
aida05 fault  0x36ca :  0x36ce :  4 
White Rabbit error counter test result: Passed 15, Failed 1

Understand the status reports as follows:
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR

Base Current Difference
aida13 fault 0xa :  0xe9 :  223 
FPGA Timestamp error counter test result: Passed 15, Failed 1
If any of these counts are reported as in error
The ASIC readout system has detected a timeslip.
That is the timestamp read from the time FIFO is not younger than the last


Returned 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 
Mem(KB) : 4 8 16 32 64 128 256 512 1k 2k 4k
aida01 : 0 5 3 0 1 4 3 3 3 3 6 : 36760
aida02 : 7 9 0 2 2 4 2 3 3 3 6 : 36644
aida03 : 1 3 2 1 2 1 2 3 3 3 6 : 36188
aida04 : 3 1 2 0 2 3 2 3 3 3 6 : 36404
aida05 : 20 9 3 2 2 3 2 2 3 3 6 : 36104
aida06 : 4 10 0 2 2 4 2 2 3 3 6 : 36128
aida07 : 3 1 2 1 3 3 2 3 3 3 6 : 36500
aida08 : 2 4 3 2 3 4 2 3 2 3 6 : 35672
aida09 : 1 6 3 3 1 4 2 3 2 3 6 : 35588
aida10 : 5 6 4 2 1 3 2 3 2 3 6 : 35460
aida11 : 23 8 2 2 2 3 3 4 2 3 6 : 36348
aida12 : 8 4 2 2 2 3 2 4 2 3 6 : 36000
aida13 : 2 5 3 1 2 4 3 3 2 3 6 : 35840
aida14 : 3 6 2 2 2 3 2 4 2 3 6 : 35996
aida15 : 2 4 3 3 0 4 1 3 3 3 6 : 36280
aida16 : 7 7 5 1 3 4 2 3 2 3 6 : 35716



<strong>05:00 OH Took over at 04:00</strong>

System wide checks ok except

   Base   Current  Difference
aida05 fault   0x36ca :   0x36d0 :   6 
White Rabbit error counter test result: Passed 15, Failed 1

Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR


    Base   Current   Difference
aida05 fault   0x0 :   0x1 :   1 
aida13 fault   0xa :   0xe9 :   223 
FPGA Timestamp error counter test result: Passed 14, Failed 2
If any of these counts are reported as in error
The ASIC readout system has detected a timeslip.
That is the timestamp read from the time FIFO is not younger than the last



05:41 AIDA 05 has dropped out. Will restart the system

05:58 DAQ recovered from powercycle
 p+n strips at 0xa
 n+n strips at 0x20
 ADC control register at 0xFF
 All waveforms turned off in LED

 Statistics - attachment 5
 Correlation scaler in FEE 3 and 4 are mostly tracking 1:1

06:34 Local users have been unable to recover MBS
      And have been unable to contact local experts

07:22 Bias and leakage currents ok - attachment 6
 Stats attachment -7
  309   Wed May 12 18:53:38 2021 JSWed 12 20:00-00:00
19:51 Shift change

Stats ok
ucesb ok, Max implants ~1500 Hz 

20:30
Stats ok
ucesb ok, Max implants ~1000 Hz
current ok (maybe leveling out)

20:51 
The online people are starting a new sort and are going to change some things with degraders I think. We are on R1_340.

20:59 

Check
Clock status test result: Passed 16, Failed 0
Calibration test result: Passed 16, Failed 0
WR:
		 Base 		Current 	Difference
aida05 fault 	 0x36ca : 	 0x36ce : 	 4  
White Rabbit error counter test result: Passed 15, Failed 1
FPGA time:
			 Base 		Current 		Difference
aida13 fault 	 0xa : 	 0xb6 : 	 172  
FPGA Timestamp error counter test result: Passed 15, Failed 1

--- both errors above same module as earlier shift
	 
Returned 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0  
Mem(KB) :	4	8	16	32	64	128	256	512	1k	2k	4k
aida01 :      1      3      2      0      1      3      3      2      3      3      6   : 36092
aida02 :      6      4      4      2      2      3      1      3      3      3      6   : 36280
aida03 :      6      3      2      0      0      3      3      2      3      3      6   : 36048
aida04 :     10      3      1      1      3      3      1      3      3      3      6   : 36272
aida05 :     19      8      2      0      2      3      1      3      3      3      6   : 36268
aida06 :      2      2      0      2      3      4      2      2      3      3      6   : 36120
aida07 :      6      7      3      0      3      4      3      2      3      3      6   : 36416
aida08 :      7      5      4      0      3      2      2      3      3      3      6   : 36420
aida09 :     21      6      3      2      2      4      2      3      2      3      6   : 35700
aida10 :     13      3      1      1      3      3      1      4      2      3      6   : 35772
aida11 :     25      6      4      2      1      4      2      4      2      3      6   : 36180
aida12 :      0      4      0      3      2      3      3      3      2      3      6   : 35712
aida13 :     18      5      2      3      2      3      2      4      2      3      6   : 36080
aida14 :      2      1      1      1      2      3      2      4      2      3      6   : 35904
aida15 :      0      1      1      3      0      4      2      2      3      3      6   : 35960
aida16 :      4      5      3      2      3      4      1      3      2      3      6   : 35432

ucesb ok elog:309/1
current ok elog:309/2
Stats ok elog:309/3
temps ok elog:309/4

21:36
ucesb ok
Stats ok 

22:01
ucesb ok 1500 Hz peak implant
Stats ok 
temps ok
bias ok - earlier rise seems to have stopped, 0.2uA fluctuations on values.

22:20 Nic noted that they were seeing fewer y-strips events (x agreed with FRS rate).
Dead time was suggested as a problem. Oscar investigating, good event rate in midas doesn't show significant dead time.
Looking at the difference in FEEs3 & 4, which should have the same scaler rate but don't, suggestion data lost in the pause and resume.
Oscar is going to raise the threshold on one of the y FEEs as a test
Increased aida4 slow comparator threshold changed
There is some confusion as it seems to be dead time but apart rates shown look low enough, dead time is possibly masking true rate?
Results of test, need thresholds 100keV on p+n and 320keV n+n to get a roughly 1:1 between FEE3 and FEE4 scaler

23:40 FRS magnet issue, possibly power, no beam

00:03 bias ok elog:309/5 has come down because no beam
handing over to muneerah
  308   Wed May 12 15:03:10 2021 CB, OHWed 12 16:00-20:00
16:00 Shift change

Statistics OK (attach 1)
Temps OK (attach 2)
Bias OK (attach 3)

System wide checks
Clock, ADC - OK

WR
		 Base 		Current 	Difference
aida05 fault 	 0x36ca : 	 0x36cb : 	 1  
White Rabbit error counter test result: Passed 15, Failed 1


FPGA

			 Base 		Current 		Difference
aida13 fault 	 0xa : 	 0x4b : 	 65  
FPGA Timestamp error counter test result: Passed 15, Failed 1
If any of these counts are reported as in error
The ASIC readout system has detected a timeslip.
That is the timestamp read from the time FIFO is not younger than the last

Returned 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0  
Mem(KB) :	4	8	16	32	64	128	256	512	1k	2k	4k
aida01 :     15      5      6      1      1      2      2      3      3      3      6   : 36388
aida02 :     21     10      2      2      2      3      2      2      3      3      6   : 36100
aida03 :     25      5      3      1      0      3      3      2      3      3      6   : 36188
aida04 :      5      2      0      3      2      3      2      3      3      3      6   : 36484
aida05 :     16      6      3      1      2      4      1      3      2      4      6   : 37440
aida06 :      2      2      2      2      2      3      1      3      3      3      6   : 36216
aida07 :     16      9      3      0      3      4      3      2      3      3      6   : 36472
aida08 :      0      7      1      1      2      2      3      3      2      3      6   : 35560
aida09 :     15      7      4      2      2      4      2      3      2      3      6   : 35700
aida10 :      3      7      3      4      1      3      1      4      1      4      6   : 36788
aida11 :      5      1      3      1      1      4      3      3      2      3      6   : 35756
aida12 :      5      5      0      3      2      3      2      4      2      3      6   : 35996
aida13 :      2      5      1      2      2      4      3      3      2      3      6   : 35840
aida14 :     19      5      3      1      1      2      1      3      3      3      6   : 36100
aida15 :     20      6      6      2      1      3      2      2      3      3      6   : 36064
aida16 :      7      9      4      1      3      4      2      3      2      3      6   : 35716

File size change
 Collecting the file size of each FEE64 Options CONTENTS file to check they are all the same

 FEE : aida01 =>   Options file size is 1026 	 Last changed Mon May 10 16:47:51 CEST 2021
 FEE : aida02 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:43:46 CEST 2021
 FEE : aida03 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:43:50 CEST 2021
 FEE : aida04 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:43:53 CEST 2021
 FEE : aida05 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:43:55 CEST 2021
 FEE : aida06 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:43:59 CEST 2021
 FEE : aida07 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:44:02 CEST 2021
 FEE : aida08 =>   Options file size is 1025 	 Last changed Wed May 05 12:15:54 CEST 2021
 FEE : aida09 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:44:08 CEST 2021
 FEE : aida10 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:44:57 CEST 2021
 FEE : aida11 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:44:57 CEST 2021
 FEE : aida12 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:44:57 CEST 2021
 FEE : aida13 =>   Options file size is 1025 	 Last changed Fri May 07 19:40:34 CEST 2021
 FEE : aida14 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:44:57 CEST 2021
 FEE : aida15 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:44:57 CEST 2021
 FEE : aida16 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:44:57 CEST 2021


16:20 Beam is back. Currently on File R1_200
16:43 Beam is being optimised.

17:00 UCESB OK
Stats OK (attach 4)
Temps OK
Bias OK (attach 5)
System-wide checks as before

17:30 Beam rate is quite high (up to 2000 Hz peak on ucesb). Stats still OK

17:52 Started compressing R1_1*

18:00 UCESB ok (attach 5)
Stats OK (attach 6)
Temps OK
Bias OK (attach 7)
System wide checks as before

19:00 UCESB ok (as before)
Stats OK (attach 8)
Temps OK
Bias OK (attach 9)
Merger OK (attach 10)

19:32 UCESB ok
Stats OK (attach 11)
Temps OK (attach 12)
Bias OK (attach 13)
System wide checks as before

19:51 Shift handed over to James S
  307   Wed May 12 06:59:55 2021 OH, RDPWed 12 May 08:00-16:00
08:00 While the rates appeared to have quietened around 7am they are back to their previous rate now.

      They still seem to be working on the setting. Implant rate in AIDA about 8 per spill. Thick degrader is still in place
      They have collected 25 minutes of implantation data. They are analysing this data to confirm the PID

08:38 The PID confirmation was apparently run number 4. During this run they were implanting Cd in AIDA DSSD2

09:09 They are now taking data implanting somewhere in the snout. I have turned off no storage.
      Unfortunately we are writing to NOTAPE R3. I was unable to change it as they were already taking data
      Data rate 16.3MB/s
      ucesb shows similar rates in SC42 and AIDA01. We may be implanting Te in AIDA then.

09:35 They stopped the DAQ as their is no beam. Switched the tapeserver over to R1 in directory S496. Setting to no storage

09:47 Beam back turned off no storage. On file R1_10

11:04 Now that we have implants on the detector we get a better look at the strips
      From the 1d pattern it is obvious that there is an adapter card misaligned in DSSD1- attachments 1 and 2
      This adaptor board corresponds to FEE11

12:17 Leakage currents screenshot 512.png

13:30 Beam off. Expected back at 15:00...

14:09 Making use of the beam off time NH and HA entered S4 to investigate FEE1 and FEE11.
      They noticed FEE11 was off and have realigned that. All channels are now observed in that FEE
      They couldn't see anything wrong with the connector for FEE1. We are still missing a block of channels there.
      Rates are as before. No better, no worse.

15:31 Leakage currents screenshot 513.png. They are continuing to rise following the intervention in S4. It could just be because the temperature is rising...
  306   Wed May 12 02:07:59 2021 MA, BAWed 12 May 00:00-08:00

03:12 System Check

attachment 1 : Current

attachment 2 : Temperature

attachment 3 : Rates

 

   
Clock status test result: Passed 16, Failed 0

Understand status as follows
Status bit 3 : firmware PLL that creates clocks from external clock not locked
Status bit 2 : always logic '1'
Status bit 1 : LMK3200(2) PLL and clock distribution chip not locked to external clock
Status bit 0 : LMK3200(1) PLL and clock distribution chip not locked to external clock
If all these bits are not set then the operation of the firmware is unreliable

   
Calibration test result: Passed 16, Failed 0

If any modules fail calibration , check the clock status and open the FADC Align and Control browser page to rerun calibration for that module

   
         Base         Current     Difference
aida05 fault      0x36ca :      0x36cb :      1  
White Rabbit error counter test result: Passed 15, Failed 1

Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR

 

   
             Base         Current         Difference
aida13 fault      0xa :      0xf :      5  
FPGA Timestamp error counter test result: Passed 15, Failed 1
If any of these counts are reported as in error
The ASIC readout system has detected a timeslip.
That is the timestamp read from the time FIFO is not younger than the last

   
Returned 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0  
Mem(KB) :    4    8    16    32    64    128    256    512    1k    2k    4k
aida01 :     22      5      7      2      1      2      2      3      3      3      6   : 36464
aida02 :      9      8      3      3      1      4      1      2      4      3      6   : 36916
aida03 :      5      2      6      2      0      3      2      3      3      3      6   : 36420
aida04 :      6      5      2      3      4      4      2      3      3      3      6   : 36800
aida05 :     17      6      6      2      2      4      1      3      2      4      6   : 37524
aida06 :      7     12      3      4      3      3      1      3      3      3      6   : 36460
aida07 :     17     11      5      1      3      3      3      2      3      3      6   : 36428
aida08 :      3      5      0      1      4      2      2      4      2      3      6   : 35924
aida09 :     27      6      4      2      0      2      2      2      3      3      6   : 35868
aida10 :     16     11      8      0      2      2      1      3      2      4      6   : 37272
aida11 :     15      2      2      3      1      4      3      4      2      3      6   : 36364
aida12 :      1      6      4      3      1      3      2      4      2      3      6   : 35988
aida13 :     22     14     10      2      4      2      2      4      2      3      6   : 36264
aida14 :     26     10      4      3      2      1      1      3      3      3      6   : 36184
aida15 :     14      2      3      2      2      4      1      2      3      3      6   : 35896
aida16 :      7      5      4      0      2      4      2      3      2      3      6   : 35588

06:10 DSSD1 rate high!

Attached 4

Called OH and he woke up to fix it :)

It was a problem with one of the ASIC according to what he says:

One of the ASICs HEC was running crazy!. Forced the ASICs to check their settings which brings them to back into line see attachment 9. This done by around 06:35

 

07:03 System check

attachment 5 Spectrum rate

attachment 6 Voltages

attachment 7 Rates

attachment 8 Temperature

System wide clock are all ok except

White rabbit

 Base         Current     Difference
aida05 fault      0x36ca :      0x36cb :      1  
White Rabbit error counter test result: Passed 15, Failed 1

Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR

FPGA

 Base         Current         Difference
aida13 fault      0xa :      0x14 :      10  
FPGA Timestamp error counter test result: Passed 15, Failed 1
If any of these counts are reported as in error
The ASIC readout system has detected a timeslip.
That is the timestamp read from the time FIFO is not younger than the last

 

  305   Tue May 11 07:21:24 2021 OHTuesday 11 May
08:20 System wide checks all ok except
		 Base 		Current 	Difference
aida01 fault 	 0xc10c : 	 0xc10e : 	 2  
aida02 fault 	 0xbe2a : 	 0xbe2c : 	 2  
aida03 fault 	 0xf665 : 	 0xf667 : 	 2  
aida04 fault 	 0xeabd : 	 0xeabf : 	 2  
aida05 fault 	 0x36c8 : 	 0x36ca : 	 2  
aida06 fault 	 0x2ea2 : 	 0x2ea4 : 	 2  
aida07 fault 	 0x9186 : 	 0x9188 : 	 2  
aida08 fault 	 0x5443 : 	 0x5445 : 	 2  
White Rabbit error counter test result: Passed 8, Failed 8

Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR

and

	
		 Base 		Current 	Difference
aida01 fault 	 0xc10c : 	 0xc10e : 	 2  
aida02 fault 	 0xbe2a : 	 0xbe2c : 	 2  
aida03 fault 	 0xf665 : 	 0xf667 : 	 2  
aida04 fault 	 0xeabd : 	 0xeabf : 	 2  
aida05 fault 	 0x36c8 : 	 0x36ca : 	 2  
aida06 fault 	 0x2ea2 : 	 0x2ea4 : 	 2  
aida07 fault 	 0x9186 : 	 0x9188 : 	 2  
aida08 fault 	 0x5443 : 	 0x5445 : 	 2  
White Rabbit error counter test result: Passed 8, Failed 8

Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR

    Statistics still high - attachment 1
    Temperatures ok - attachment 2
    Bias and leakage currents ok - attachment 3
  304   Mon May 10 11:45:15 2021 NHConfusing behaviour of rates
Testing the behaviour of powering down the fast ADCs in aida01 (the hot FEE)

Set local control 2 to 0xff in aida01
Wait... see temperature drpos by ~3 degrees (PJCS reports much larger drop in T9)
So far so normal

*but*

aida05 rates go very high when this waveform is disabled!!
Set local control to 0x00 again
They drop back...

Recalibrated FADCs and back to normal
  303   Mon May 10 08:49:37 2021 TDMonday 10 May
09.51 All system wide checks OK *except*


		 Base 		Current 	Difference
aida05 fault 	 0x8441 : 	 0x8442 : 	 1  
aida06 fault 	 0xc5c5 : 	 0xc5c6 : 	 1  
aida07 fault 	 0x6f07 : 	 0x6f08 : 	 1  
aida08 fault 	 0x4814 : 	 0x4815 : 	 1  
White Rabbit error counter test result: Passed 12, Failed 4

Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR

	
Returned 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0  
Mem(KB) :	4	8	16	32	64	128	256	512	1k	2k	4k
aida01 :      1      4      6      3      5      5      2      3      2      3      6   : 36004
aida02 :      4      1      3      6      3      4      3      3      2      3      6   : 36040
aida03 :     12      8      5      2      3      3      3      4      2      3      6   : 36416
aida04 :      1      2      5      4      1      4      3      4      2      3      6   : 36388
aida05 :     22      8      4      2      4      5      1      4      1      4      6   : 37272
aida06 :      5      4      0      6      3      6      3      3      2      3      6   : 36276
aida07 :     38     16      7      7      1      6      1      3      3      3      6   : 37032
aida08 :     16      4      2      3      1      4      1      3      3      3      6   : 36384
aida09 :      2      3      7      6      5      5      2      3      2      3      6   : 36112
aida10 :     42     15      4      6      2      6      2      4      2      3      6   : 36768
aida11 :      1      5     13      7      3      3      3      3      2      3      6   : 36124
aida12 :     23      9      3      4      7      3      2      3      2      3      6   : 35988
aida13 :      2      6      7      3      4      4      3      2      2      3      6   : 35592
aida14 :     17     11      3      4      3      5      2      3      2      3      6   : 35980
aida15 :      4      7      5      2      1      6      2      3      2      3      6   : 35864
aida16 :      4      1      3      3      4      2      3      4      2      3      6   : 36264


 Collecting the file size of each FEE64 Options CONTENTS file to check they are all the same

 FEE : aida01 =>   Options file size is 1026 	 Last changed Sun May 09 15:10:52 CEST 2021
 FEE : aida02 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:43:46 CEST 2021
 FEE : aida03 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:43:50 CEST 2021
 FEE : aida04 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:43:53 CEST 2021
 FEE : aida05 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:43:55 CEST 2021
 FEE : aida06 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:43:59 CEST 2021
 FEE : aida07 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:44:02 CEST 2021
 FEE : aida08 =>   Options file size is 1025 	 Last changed Wed May 05 12:15:54 CEST 2021
 FEE : aida09 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:44:08 CEST 2021
 FEE : aida10 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:44:57 CEST 2021
 FEE : aida11 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:44:57 CEST 2021
 FEE : aida12 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:44:57 CEST 2021
 FEE : aida13 =>   Options file size is 1025 	 Last changed Fri May 07 19:40:34 CEST 2021
 FEE : aida14 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:44:57 CEST 2021
 FEE : aida15 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:44:57 CEST 2021
 FEE : aida16 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:44:57 CEST 2021

09.52 All histograms, stats and merger stats zero'd

      FEE64 temps OK - attachment 1
      ADC data, disc/WR 28-27/pause/correlation stats OK - attachments 2-6
      Rate spectra - attachment 7
       ASIC check load
      Rate spectra - attachment 8
      1.8.W spectra 20us FSR - attachments 9 & 10
      Grafana - most recent 7 days DSSSD bias and leakage current - attachment 11
      Merger stats/tape server/merger - attachment 12
      DSSSD bias and leakage current OK - attachment 13

10.08 1.8.W spectra 20us, 200us, 2ms & 20ms FSR - attachments 14-21
       we are not missing extrinsic noise features at longer time ranges cf. 20us FSR

12.18 Cannot connect to port 8015, MIDAS terminal says "Killed"
      Restart MIDAS 8015 
      Collect FPGA baseline to avoid TCL error
      FEEs unaffected and running as before

13.25 FEEs power cycled to see if any change in weird noise/rates situation
      No change 
      ASICs synchronised and back to normal

14.34 Start a pulser walkthrough in May21/R10
      Attenuation 5x, Rep Rate 25 Hz
      Amplitudes 10..9...1

14.45 Back to 2 Hz @ 2V Pulser, walkthrough complete. No Storage again.

 PULSER SETTINGS
---------------------------
 Pulse is ON
 Positive Tail Pulse
 Trigger Source is Internal Clock
 Trigger Threshold is 3.5
 Amplitude   : 2.0 Volts
 Rep Rate    : 2.0 hZ
 Delay       : 250.0 ns
 Fall Time   : 1 ms
 Attenuation : 1
 Display is  : Volts
 Equivalent keV is : 200.0 (NB. actually it is 12 MeV)
 Ramp Start at 0.01 Volts
 Ramp Stop  at 9.99 Volts
 Ramp Start at 1.0 keV
 Ramp Stop  at 999.0 keV
 Ramp Time  is 60 seconds
 # Ramp Cycles is 1

OK
  302   Sun May 9 19:29:23 2021 TDSunday 9 May
  301   Sat May 8 09:05:04 2021 TDSaturday 8 May
09.54 FEE64 temperatures OK - attachment 1
      ADC data item stats OK - attachment 2
       p+n FEE64s 11x <100k, 1x <200k
      Rate spectra - attachments 3 & 4
       before and after ASIC check load
      1.8.L spectra - attachments 5 & 6
      1.8.W spectra 20us FSR - attachments 7 & 8
      Disc, WR-28-47, pause, resume info data items - attachments 9-12
       note large number pause/resume for aida13 - legacy numbers since last restart/zero?
       attachment 16 for pause stats since stats zero at 10.02 - looks OK
      Grafana DSSSD bias & leakage currents for most recent 48h OK - atatchment 13
      DSSSD bias & leakage currents - attachment 14
      Merger stats/merger/tape server messages - attachment 15

10.02 All histograms and stats zero'd
      
10.06 All system wide checks OK *except*

 Collecting the file size of each FEE64 Options CONTENTS file to check they are all the same

 FEE : aida01 =>   Options file size is 1026 	 Last changed Thu May 06 15:29:51 CEST 2021
 FEE : aida02 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:43:46 CEST 2021
 FEE : aida03 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:43:50 CEST 2021
 FEE : aida04 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:43:53 CEST 2021
 FEE : aida05 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:43:55 CEST 2021
 FEE : aida06 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:43:59 CEST 2021
 FEE : aida07 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:44:02 CEST 2021
 FEE : aida08 =>   Options file size is 1025 	 Last changed Wed May 05 12:15:54 CEST 2021
 FEE : aida09 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:44:08 CEST 2021
 FEE : aida10 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:44:57 CEST 2021
 FEE : aida11 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:44:57 CEST 2021
 FEE : aida12 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:44:57 CEST 2021
 FEE : aida13 =>   Options file size is 1025 	 Last changed Fri May 07 19:40:34 CEST 2021
 FEE : aida14 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:44:57 CEST 2021
 FEE : aida15 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:44:57 CEST 2021
 FEE : aida16 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:44:57 CEST 2021

10.25 Pulser peak width aida01 97 ch FWHM, aida02 350 ch FWHM
       peak widths are increasing cf. earlier this week - see https://elog.ph.ed.ac.uk/DESPEC/291 , https://elog.ph.ed.ac.uk/DESPEC/298
ELOG V3.1.4-unknown