ID |
Date |
Author |
Subject |
655
|
Wed Jun 12 06:17:23 2024 |
TD | 08:00-16:00 Wednesday 12 June |
07:10 DSSSD bias & leakage current OK - attachment 1
FEE64 temperatures OK - attachment 2
ADC data items stats OK - attachment 3
per FEE64 Rate spectra - attachment 4
Merger, TapeServer etc - attachment 5
per p+n FEE64 1.8.L spectra - attachment 6
aida09 pulser peak width 68 ch FWHM
07.13 peak area 25253 total area 27548 in 221mins => 1.9-2.1Hz => deadtime < 5% or less which is consistent with data file analysis below
analysis data file S181/R6_274 - attachment
max FEE64 deadtime 24% aida08
HEC data rate 4.5kHz, LEC data rate 2.284M
14.00 DAQ crashed by accident
FEE64s power cycled
DAQ restarted
data file S181/R7
DSSSD bias & leakage current OK - attachments 8-9
FEE64 temperatures OK - attachment 10
ADC data item stats - attachment 11
WR timestamps OK - attachment 12
All system wide checks *except* as reported attachments 13-15
per FEE64 Rate spectra - attachment 16
per p+n FEE64 1.8.L spectra - attachment 17
aida09 pulser peak width 66 ch FWHM
per FEE64 1.8.H spectra - attachments 18-19
per FEE64 1.8.W spectra - 20us FSR - attachment 20-21
Merger, TapeServer etc - attachments 22-23
data file S181/R7_16
analysis data file S181/R7_16
max deadtime 7.3% aida08, all other FEE64s < 2%
LEC data rate 1.877M HEC data rate 4.7k |
590
|
Tue Apr 23 07:27:18 2024 |
PP | 08:00-16:00 Tuesday 24 April shift |
Checks 08:00-08:30
Screenshots are attached and everything looks OK. |
650
|
Tue Jun 11 10:03:48 2024 |
TD, NH | 08:00-16:00 Tuesday 10 June |
10.57 DSSSD bias & leakage current OK - attachment 1
FEE64 temps OK - attachment 2
ADC, MBS correlation scaler data item stats - attachments 3-4
ADC 9x < 20k, max 243k aida08
ASIC settings aida01 & aida02 - attachments 5-6
ALL system wide checks OK *except* attachments 7-8
WR timestamps OK - attachment 9
BNC PB-5 settings - attachment 10
per FEE64 Rate spectra - attachment 11
per p+n FEE64 1.8.L spectra - attachment 12
aida09 pulser peak width 57 ch FWHM
per FEE64 1.8.H spectra - attachments 13-14
per FEE64 1.8.W spectra - 20us FSR - attachments 15-16
no p+n FEE64 wavefroms - reason unknown
Merger, TapeServer etc - attachments 17-18
11.25 Lost control of aida-gsi window system
All command terminals lost, could not start new command terminals
aida-gsi restart
FEE64 power cycle
DSSSD bias cycle
Note file manager, calculator apps now working as expected
12.11 DAQ starts
data file S181/R6
ASIC settings 2024Apr19-15.22.49
BNC PB-5 Pulser
Amplitude 10.0V
Attenuation x10
Fall time 1ms
Frequency 2Hz
Polarity + (connected to test + signal daisy chain to all p+n FEE64s)
Sync ASIC clocks OK
All system wide checks OK *except* aida02 FPGA/WR decoder - attachments 19-20
per FEE64 1.8.W spectra - 20us FSR - attachments 21-22
12.29 analysis data file S181/R6_9 - attachment 23
max dead time c. 2% (aida08)
15:34 For some reason the AIDA MBS wasn't connecting to the time sorter (or a check stream) [NH]
Try to reset the AIDA MBS
Due to confusion the AIDA Merger was rebooted too, while the DAQ was running, this broke all the FEEs -> Powecycled
Synced ASIC clocks, fixed all waveforms
Masked all discriminators
All looks ok -> Sending back to MBS and back in DESPEC DAQ |
605
|
Thu Apr 25 08:26:50 2024 |
TD | 08:00-16:00 Thursday 25 April |
09.56 DAQ continues data file S100_alpha/R18_387
FRS switching settings
DSSSD bias & leakage current OK - attachments 1
FEE64 temperatures OK - attachment 2
aida02 ASIC temp - known fault
All system wide checks OK *except* as shown - attachments 3-5
WR timestamps OK - attachment 6
ADC, PAUSE, RESUME & Correlation Scaler data time stats - attachments 7-10
per FEE64 Rate spectra - attachments 11
per p+n FEE64 1.8.L spectra - attachments 12-13
aida09 pulser peak width 57 ch FWHM
common x/y scale => peak height proxy for width => all p+n FEE64s similar pulser peak widths
per n+n FEE64 1.8.L spectra - attachment 14
per FEE64 1.8.H spectra - attachments 15-16
per FEE64 1.8.W spectra - 20us FSR - attachments 17-18
only aida08 & aida09 currently updating - ASIC check load and/or DAQ STOP/GO required?
Merger, TapeServer etc - attachments 19-21
BNC PB-5 settings - attachment 22
10.07 data file S100_alpha/R20
new run for new FRS setting
10.08 all histograms, stats, merger stats and data link stats zero'd
10.10 ADC data item stats (beam OFF) - attachment 23
11/16 < 20k, max 129k (aida08)
10.31 analysis of data file R18_349 (closed 06.00 this morning)
max deadtime 12% (aida04)
no timewarps
HEC data item rate 1.7kHz
11.00 tuning 168dy FRS setting
ucesb - attachment 24
spill length 3.5s, extraction time 1.5s
11.30 degraders now set to implant in AIDA DSSSDs
high rates expected
per FEE64 Rate spectra - attachment 25
ADC data item stats - attachment 26
Merger etc - attachment 27
in spill rates to MBS up 30+Mb/s, 2.5M data item/s merged
DSSSD bias & leakage current - attachment 28
c. few hundred nA step in leakage current
analysis of data file R20_33 - attachment 29
max deadtime 18% (aida08)
no timewarps
HEC data item rate 4.1kHz (c. 2.4x previous FRS setting)
Correcting for spill cycle => 5.9kHz in spill (online analysis shows no HEC events observed outside spill)
assume typical HEC event multiplicity = 3 => implant rate ~ 2.0kHz in spill or 6800 implants per spill
SC42 L/R currently report c. 4.5kHz
Note that we currently see a higher fraction of events in DSSSD#2 cf. previous setting so multiplicity assumption may be low
If we assume multipicity = 6 => implant rate ~ 1.0kHz in spill or 3200 implants per spill
we are probably somewhere in between
12.50 DSSSD bias & leakage current - Grafana - attachment 30
leakage current ramping, spill micro structure observable
FEE64 temperatures OK - attachment 31
ADC data item stats - attachment 32
per FEE64 Rate spectra - attachment 33
Merger etc - attachment 34
14.56 analysis of data file R20_106 - attachment 35
max deadtime c. 18% (aida04 & aida08), c. 12% (aida03), c. 11% (aida06), c. 9% (aida02 & aida13), c. 6% (aida05), all others <2%
no timewarps
HEC data item rate 4.2kHz
15.40 DSSSD bias & leakage current - Grafana - attachments 36-37
leakage current ramping, spill micro structure observable
FEE64 temperatures OK - attachment 38
ADC data item stats - attachment 39
per FEE64 Rate spectra - attachment 40
Merger etc - attachment 41
ucesb - attachment 42
15.57 all histograms zero'd
|
615
|
Sun Apr 28 06:57:59 2024 |
Magda Satrazani | 08:00-16:00 Sunday 28 April |
Took over the shift from Marc.
10.51 Added screenshots to illustrate erroneous data from Grafana (AIDA) - attachments 18-19
Compare to the CAEN HV control terminal (attachment 15) which shows the expected values. Have email'd Nic & Helena. TD
10.55 Noticed that the CAEN HV control terminal is showing the corerct data but on the wrong line of the display (attachment 15). Enter commands q (quit) and d (display) to restart display - the correct data now appears on the correct line (attachment 20).
Grafana (AIDA) is now reporting data correctly too (attachment 21).
Summary - CAEN HV terminal display issue ( c. 10.05 - 10.55 )
Note the correct data/values continued to be shown. Leakage current shown correctly by both Grafana and CAEN HV control throughout.
Conclude this was a display issue - not a change in the operating parameters of the CAEN N1419ET.. Data should be OK.
14.45 analysis data file R21_500 - attachment 36
max deadtime 22% (aida11), 16% (aida04), 7% (aida02), 6% (aida08), 6% (aida06) , all others < 1%
no timewarps
HEC data item rate 2.7kHz
|
612
|
Sat Apr 27 04:59:31 2024 |
TD | 08:00-16:00 Saturday 27 April |
06.00 DAQ continues file S100_alpha/R20_740
FRS setting 162Eu
DSSSD bias & leakage current - Grafana - attachments 1-2
leakage currents tracking S4 ambient temperature
FEE64 temperatures OK - attachment 3
All system wide checks OK *except*
as reported - attachments 4-5
*all* ADCs not calibrated (not calibrated post-synchronise ASIC clocks at most recent restart)
WR timestamps OK - attachment 6
ADC, PAUSE, RESUME, Correlation scaler data item stats OK - attachments 7-10
per FEE64 Rate spectra - attachment 11
per p+n FEE64 1.8.L specra - attachments 12-13
aida09 pulser peak width 55 ch FWHM (~38keV FWHM)
per FEE64 1.8.H spectra - attachments 14-15
Merger, TapeServer etc - attachments 16-17
12Mb/s to disk
ucesb - attachment 18
spill length 3.5s, extraction 1.5s
06.30 analysis data file R20_748 - attachment 19
max deadtime 12%
no timewarps
HEC data item rate 2.1kHz
06.35 new run file S100_alpha/R21
09.30 DAQ continues - file S100_alpha/R21_83
DSSSD bias & leakage currents - Grafana - attachment 20
FEE64 temperatures OK - attachment 21
ADC data item stats OK - attachments 22
per FEE64 Rate spectra - attachment 23
Merger etc - attachments 24
ucesb - attachment 25
spill length 3.5s, extraction 1.5s
09.50 per FEE64 1.8.W spectra - 20us FSR - attachments 26-27
no significant change in electronic noise cf. start of run
per p+n FEE64 1.8.L spectra - attachment 28
aida09 pulser peak width 62 ch FWHM
common x/y scale - peak heoght proxy for width
13.57 DAQ continues file R21_141
DSSSD bias & leakage current - Grafana - attachments 29-30
FEE64 temperatures OK - attachment 31
ADC data item stats OK - attachments 32
per p+n FEE64 1.8.L spectra - attachment 33
aida09 pulser peak width 55 ch FWHM (~39keV FWHM)
per FEE64 Rate spectra - attachment 34
Merger etc - attachments 35
ucesb - attachment 36
spill length 3.5s, extraction 1.5s
14.15 analysis data file R21_144 - attachment 35
max deadtime 16% (aida04)
no timewarps
HEC data item rate 2.0kHz |
662
|
Fri Jun 14 06:22:39 2024 |
TD | 08:00-16:00 Friday 14 June |
07.23 All histograms & stats zero'd
DAQ continues data file S181/R8_35
analysis data file R8_35 - attachment 1
max deadtime 4.8% (aida08), all other FEE64s < 3%
LEC data rate 1.677M HEC data rate 3.7k
08.46 DSSSD bias & leakage current OK - attachments 2-3
FEE64 temps OK - attachment 4
ADC data item stats - attachment 5
per FEE64 Rate spectra - attachment 6
per p+n FEE64 1.8.L spectra - attachment 8
aida09 pulser peak width 65 ch FWHM
per FEE64 1.8.H spectra - attachment 7 & 10
per FEE64 1.8.W spectra - 20us FSR - attachment 9 & 11
Merger, TapeServer etc - attachments 12-13
data file S181/R8_72
10.51 DSSSD bias & leakage current OK - attachments 14
FEE64 temps OK - attachment 15
ADC data item stats - attachment 16
per FEE64 Rate spectra - attachment 17
10.54 analysis data file R8_122 - attachment 18
max deadtime 7.8% (aida08), all other FEE64s < 3%
ADC data rate 1.88M, HEC data rate 4.7k
14.24 DSSSD bias & leakage current OK - attachments 18
FEE64 temps OK - attachment 19
ADC data item stats - attachment 20
per FEE64 Rate spectra - attachment 21
aida02 no data
14.28 merger restarted and DAQ running OK
14.38 analysis data file R8_213 - attachment 22
max deadtime 6.6% (aida08), all other FEE64s < 3.5%
ADC data rate 1.935M, HEC data rate 3.6k |
659
|
Thu Jun 13 06:49:54 2024 |
Magda Satrazani | 08:00 - 16:00 Thursday 13 June 2024 |
Took over the shift from Marc.
08:00 FULL CHECK
DSSSD bias & leakage current ok - attachment # 1-2
FEE64 temperatures ok - attachment # 3
ADC data item stats - attachment # 4
Merger ok - Attachement # 5
Tape service - attachement # 6
10:00 FULL CHECK
DSSSD bias & leakage current ok - attachment # 7-8
FEE64 temperatures ok - attachment # 9
ADC data item stats - attachment # 10
Merger ok - Attachement # 11
Tape service - attachement # 12
12:00 FULL CHECK
DSSSD bias & leakage current ok - attachment # 13-14
FEE64 temperatures ok - attachment # 15
ADC data item stats - attachment # 16
Merger ok - Attachement # 17
Tape service - attachement # 18
12.34 analysis data file S181/R7_610 - attachment 19
max deadtime 8.7% (aida08), all other FEE64s < 2.7%
LEC data rate 2.000M, HEC data rate 4.7k
12.52 per p+n FEE64 1.8.L spectra - attachment 20
aida09 pulser peak width 63 ch FWHM
per FEE64 1.8.H spectra - attachments 21-22
per FEE64 1.8.W spectra - 20us FSR - attachments 23-24
13.00 all histograms zero'd
14:00 FULL CHECK
DSSSD bias & leakage current ok - attachment # 19-20
FEE64 temperatures ok - attachment # 21
ADC data item stats - attachment # 22
Merger ok - Attachement # 23
Tape service - attachement # 24
13.01 Beam OFF for S4 access
ADC data item stats - attachment 25
per FEE64 Rate spectra - attachment 26
14.12 analysis data file R7_654 - attachment 27
max deadtime 3.9% (aida08), all other FEE64s < 3%
LEC data rate 1.817M, HEC data rate 2.5k
14.00 data file R7_649 - attachment 28
max deadtime 4.5% (aida08), all other FEE64s < 3%
LEC data rate 1.756M, HEC data rate 2.5k
16:00 FULL CHECK
DSSSD bias & leakage current ok - attachment # 25-26
FEE64 temperatures ok - attachment # 27
ADC data item stats - attachment # 28
Merger ok - Attachement # 29
Tape service - attachement # 30 |
608
|
Fri Apr 26 07:01:26 2024 |
Norah | 08:00 - 16:00 Friday 26 April |
10:00 Checks:
All seems smooth.
DSSD bias and leakage current ( Attachment 1)
FEE64 temperatures (Attachment 2)
Per FEE64 Rate spectra (Attachment 3)
ADC data item stats -(Attachment 4)
Ucesb (Attachment 5)
10:21
Aida05 was zero not reading had to call Nic to fix it.
10:34
Now, Aida05 is running well.
.......................................................
12:00 Checks:
Everything appears to be going smoothly, and I took screenshot of them.
12:28, Aida02 showed as not reading , had to email Nic to fix it.
12:30, it resumed reading and returned to normal operation.
.........................................................
14:00 Checks:
Everything is going well , and I have captured screenshots of each one.
......................................................
16:00 Checks:
All seems smooth, and I took screenshot of them.
|
647
|
Mon Jun 10 07:15:48 2024 |
Dan | 08.00-16.00 Monday 10th June |
8am checks
temps ok - attachment 2
rates ok - attachment 8
voltage ok - attachment 5
merger ok - attachment 6
9am checks
temps ok - attachment 10
rates ok - attachment 9
voltage ok - attachment 12
merger ok - attachment 11
(Ignore Grafana plots 13-43 inc - not refreshed)
9.16 beam stopped to optimise beam intensity
10.00 checks - still no beam
temps ok - attachment 17
rates ok - attachment 15+16
voltage ok - attachment 14
merger ok - attachment 18
11.00 checks - still no beam
temps ok - attachment 23
rates ok - attachment 21+22
voltage ok - attachment 20
merger ok - attachment 24
12pm checks
temps ok - attachment 28
rates ok - attachment 26+27
voltage ok - attachment 29
merger ok - attachment 30
12.45 - beam back
temps ok - attachment 34
rates ok - attachment 32+33
voltage ok - attachment 35
merger ok - attachment 36
2pm checks
temps ok - attachment 40
rates ok - attachment 38+39
voltage ok - attachment 41
merger ok - attachment 42
/tmp/R4_276 added - attachment 44
max deadtime 10.5% (aida08)
n+n FEE64s deadtime
3pm checks
temps ok - attachment 48
rates ok - attachment 46+47
voltage ok - attachment 45+50
merger ok - attachment 49 |
637
|
Mon Jun 3 14:40:34 2024 |
JB, NH | 03 June 2024 |
15:40 reseated FEE64s adaptors of aida01, aida09, aida13 & aida16.
TEMP OK. HV OK. System wide checks all passed excl. aida07 ADC calibration, see attached.
Histograms and rates look same as previous, reseating aida01 seems to have worked. aida16 still unhappy.
re-reseated aida16. 16:19 - appears to have fixed the problem in the channel.
Pulser test + connected.
bPlast ground removed from frame.
To-do:
- Need clarification about links connected on AIDA n+n FEE64 adaptor boards. Currently, the situation is given by the attachment 8 & 9.
- Fix BB7 grounding, continuity between snout and frame observed.
- Pulser walkthrough for n+n, can do p+n when convienient.
18.30 Noted FEE64 ADC stats now significantly higher ( attachments 9-10 ) cf. earlier this afternoon (attachments 1-2)
DSSSD bias OFF. FEE64 power OFF.
|
600
|
Tue Apr 23 22:37:46 2024 |
TD | 00:00-08:00 Wednesday 24 April |
23.34 23.4.24
DSSSD bias & leakage current OK - attachment 2
FEE64 temperatures OK - attachment 2
ADC data item stats - attachment 3
per FEE64 rate spectra - attachments 4
Merger etc - attachment 5
03.41 fDSSSD bias & leakage current OK - attachment 6
FEE64 temperatures OK - attachment 7
ADC data item stats - attachment 8
per FEE64 rate spectra (off and on spill) - attachments 9-10
Merger etc - attachment 11
03.54 analysis data file R18_67 - attachment 12
max deadtime 11% (aida04)
HEC data item rate c. 1.5kHz
no timewarps
0.35 fDSSSD bias & leakage current OK - attachment 13
FEE64 temperatures OK - attachment 14
ADC data item stats - attachment 15
per FEE64 rate spectra - attachments 16
Merger etc - attachment 17
03.54 analysis data file R18_96 - attachment 18
max deadtime 12% (aida04)
HEC data item rate c. 1.6kHz
no timewarps |
654
|
Tue Jun 11 23:13:44 2024 |
TD | 00:00-08:00 Wednesday 11 June |
00.08 DSSSD bias & leakage current OK - attachment 1
FEE64 temperatures OK - attachment 3
ADC data items stats OK - attachment 2
per FEE64 Rate spectra - attachment 7
per p+n FEE64 1.8.L spectra - attachments 6
aida09 pulser peak width 63 ch FWHM
Merger, TapeServer etc - attachments 4-5
analysis data file S181/R6_274 - attachment 8
max FEE64 deadtime 19% aida08
HEC data rate 3.8kHz, LEC data rate 2.258M
per FEE64 1.8.W spectra - 20us FSR - attachments 9-10
03.12 Merger, TapeServer etc - attachments 11-12
before & after Merger restart
per FEE64 Rate spectra - attachment 13
ADC data items stats OK - attachment 14
aida02 zero rate
FEE64 temperatures OK - attachment 15
DSSSD bias & leakage current OK - attachment 16
analysis data file S181/R6_372 - attachment 17
max FEE64 deadtime 22% aida08
HEC data rate 4.4kHz, LEC data rate 2.295M
03.32 all histograms zero'd
03.57 per p+n FEE64 1.8.L spectra - attachment 18
25 mins = 1500 @ 2Hz = 3000 cf. peak area c. 3000 => deadtime low |
604
|
Wed Apr 24 21:39:51 2024 |
TD | 00:00-08:00 Thursday 25 April |
22.43 24.4.24
DSSSD bias & leakage current OK - attachment 1
FEE64 temperatures OK - attachment 2
ADC data item stats - attachment 3
per FEE64 rate spectra - attachments 4
Merger etc - attachment 5
analysis of data file R18_266 - attachment 6
max dead time 12% (aida04)
no timewarps
HEC data item rate 1.7kHz
03.37 DSSSD bias & leakage current OK - attachments 7-8
FEE64 temperatures OK - attachment 9
ADC data item stats - attachment 10
per FEE64 rate spectra - attachments 11
Merger etc - attachment 12 |
658
|
Wed Jun 12 23:00:16 2024 |
Marc | 00:00-08:00 Thursday 13 June |
Starting a new night shift - All good
1am full check:
DSSSD bias & leakage current ok - attachment # 1-2
FEE64 temperatures ok - attachment # 3
ADC data item stats - attachment # 4
Merger ok - Attachement # 5
Tape service - attachement # 6
2:30am: AIDA02 FEE stopped working. DESPEC main DAQ was alerted and remobe AIDA from timestitcher while I was restarting AIDA DAQ.
2:45am: AIDA02 FEE is back and AIDA has been added to the timestitcher again in the main DAQ.
The AIDA runs at the time of the crash to look at are R7_348, 349.
All spectra were zeroed at this time and checked.
4:00 am full check:
DSSSD bias & leakage current ok - attachment # 7-8
FEE64 temperatures ok - attachment # 9
ADC data item stats - attachment # 10
Merger ok - attachement # 11
Tape service - attachement # 12
Spectra - attachement # 13
ucesb - attachement # 14
6:00am full check:
DSSSD bias & leakage current ok - attachment # 15-16
FEE64 temperatures ok - attachment # 17
ADC data item stats - attachment # 18
Merger ok - attachement # 19
Tape service - attachement # 20
Spectra - attachement # 21
ucesb - attachement # 22
6:30 the nearline/offline analysis team reported possible noise issue in DSSD#2 and asked if the threshold were changed at any point. I can't find any mention of changes in the elog.
I've not done a setup of the all DAQ when AIDA02 fee crashes earlier this morning. I just did a setup of the merger. Thus I don't think the threshold have changed when I stopeed and restarted the DAQ.
I can see that currently slow comparator threshod are set to 0xa for most of the DSSD2 fees except for AIDA06 which is set to 0xf. I'm not sure this is the thershold we would need to increase. Something to check during the next shift.
In the UCESB window I can see that the number of decay in DSSD2 is double the number of decay in DSSD1 while the number of implant is similar. Looking at other shift it looks that this was alsready the case in prvious shifts.
07.06 analysis of data file S181/R7_457 - attachment 23
max deadtime 5.8% aida08, all other FEE64s < 1.4%
LEC data rate 1.788M, HEC data rate 4.3k
|
614
|
Sat Apr 27 23:04:42 2024 |
Marc | 00:00-08:00 Sunday 28 April |
Encore un matin, un matin pour ..... a shift !!!ramping up again ! S4 temperature not really going down either: 25C
0:40 AIDA04 is down. I informed the DAQ shift crew and attempting to restart AIDA04.
After rebooting AIDA04 all seemed going until I reached the merger check. No data was being merged.
I realise then that AIDA01 data transfer was not active. I did not spot it when I enabled the data transfer for all.
I stopped the daq and rebooted AIDA01 following all the instructions in the ELOG and this worked.
We are now back in business with AIDA back in main DAQ at 01:40
01:45 full checks:
Bias: AIDA-Graphana - Attachment 1
Leakage current ramping up again ! S4 temperaturenot really going down either: 25C
FEE64 temperatures OK - attachment 2
ADC data item stats - attachment 3
per FEE64 Rate spectra - attachment 4
Merger etc - attachment 5
ucesb - attachment 6
03:00 full checks
No change to be highlighted. All pretty much the same as during previous full checks.
05:00 full checks
Expt still running very smoothly.
Bias: AIDA-Graphana - Attachment 7-8
Leakage current going down slightly
FEE64 temperatures OK - attachment 9
ADC data item stats - attachment 10
ramping up again ! S4 temperaturenot really going down either: 25C
per FEE64 Rate spectra - attachment 11
Merger etc - attachment 12
ucesb - attachment 13
07:15 all running smoothly.
07:20 Spoke too fast AIDA DAQ is down. DAQ shift crew informed.
07:40: AIDA10 was rebooted successfully and AIDA is now back in the main DAQ.
|
611
|
Fri Apr 26 23:04:51 2024 |
Marc | 00:00-08:00 Saturday 27 April |
It's a new day, it's a new shift and I'm feeeeeliiiiing goooood !!!!!
0.15 Checks:
DSSSD bias & leakage current - Grafana - attachments 1-2
leakage current dropped a bit since the evening shift ~22:00.
FEE64 temperatures OK - attachment 3
ADC data item stats - attachment 4
per FEE64 Rate spectra - attachment 5
Merger etc - attachment 6
ucesb - attachment 7
2:20 Nothing new to report. All good. Leakage currents continue to go down slowly. See attachement 8 to 12
3:30 Still running smoothly.
5:20 AIDA 4 stop counting in the last 20 min. I informed the main DAQ and they started a new run without AIDA.
DAQ restarted OK. Main DAQ informed.
|
649
|
Mon Jun 10 23:36:14 2024 |
TD | 0000-08:00 Tuesday 11 June |
00.09 DSSSD bias & leakage current OK - attachment 1
FEE64 temperatures OK - attachment 2
ADC data items stats OK - attachment 3
per FEE64 Rate spectra - attachment 4
per p+n FEE64 1.8.L spectra - attachments 5-6
aida09 pulser peak width 53 ch FWHM
per FEE64 1.8.H spectra - attachments 7-8
per n+n FEE64 1.8.W spectra - 20us FSR - attachment 9
p+n 1.8.W not running currently
Merger, TapeServer etc - attachments 10-11
analysis data file S181/R4_427 - attachment 12
max FEE64 deadtime 1.7%
HEC data rate 3.1kHz
00.28 all histograms zero'd
02.39 DAQ continues data file S181/R4_460
Beam off since 02.08
DSSSD bias & leakage current OK - attachment 13
FEE64 temperatures OK - attachment 14
ADC data items stats OK - attachment 15
9x < 20k, max 239k aida08
per FEE64 Rate spectra - attachment 16
07.15 DAQ continues data file S181/R4_517
Beam off since 02.08
HKR report c 3-4 h until UNILAC fixed
DSSSD bias & leakage current OK - attachment 17
FEE64 temperatures OK - attachment 18
ADC data items stats OK - attachment 19
8x < 20k, max 238k aida08
per FEE64 Rate spectra - attachment 20 |
589
|
Mon Apr 22 22:35:59 2024 |
TD | 00.00-08.00 Tuesday 22 April |
23.36 Analysis of data file R17_34 - attachment 1
max deadtime c. 1% (aida02)
no timewarps
23.38 DSSSD bias & leakage current OK - attachment 2
FEE64 temperatures OK - attachment 3
aida02 ASIC temp - known fault
ADC data item stats - attachment 4
per FEE64 Rate spectra - attachment 5
Merger etc - attachment 6
01.03 DSSSD bias & leakage current OK - attachment 7
FEE64 temperatures OK - attachment 8
aida02 ASIC temp - known fault
ADC data item stats - attachment 9
per FEE64 Rate spectra - attachment 10
Merger etc - attachment 11
03.06 DSSSD bias & leakage current OK - attachment 12
FEE64 temperatures OK - attachment 13
aida02 ASIC temp - known fault
ADC data item stats - attachment 14
per FEE64 Rate spectra - attachment 15
Merger etc - attachment 16
05.04 DSSSD bias & leakage current OK - attachment 17
FEE64 temperatures OK - attachment 18
aida02 ASIC temp - known fault
ADC data item stats - attachment 19
per FEE64 Rate spectra - attachment 20
Merger etc - attachment 21
07.29 DSSSD bias & leakage current OK - attachment 22
FEE64 temperatures OK - attachment 23
aida02 ASIC temp - known fault
ADC data item stats - attachment 24
per FEE64 Rate spectra - attachment 25
Merger etc - attachment 26 |
661
|
Thu Jun 13 23:04:01 2024 |
Dan Judson, TD | 0.00-8.00 14/6/24 |
0.00 checks - all looks ok
Voltage/currents - attachment 1
Rates attachment 2 + 3
Temps - attachment 4
Merger - attachment 5
ucesb - attachment 6
Grafana - attachment 7
1.30 -aida1 in merger Links, aida02 in rate spectra, has stopped - resetting DAQ.
Datafile R7_961 being written when stopped
1.40 - seems ok
1.50 checks - seems ok
Voltage/currents - attachment 8
Rates attachment 9 + 10
Temps - attachment 11
Merger - attachment 12
ucesb - attachment 13
Grafana - attachment 14
3.30 -aida1 in merger Links, aida02 in rate spectra, has stopped again - resetting DAQ.
Datafile R7_1010 being written when stopped
3.38 - seems ok
3.40 checks - all looks ok
Voltage/currents - attachment 15
Rates attachment 16 + 17
Temps - attachment 18
Merger - attachment 19
ucesb - attachment 20
Grafana - attachment 21
Zeroed spectra - attachments 22-28
5.39 -aida1 in merger Links, aida02 in rate spectra, has stopped again - resetting DAQ.
Datafile R7_1062 being written when stopped
AIDA out of the main data
Could not get it to restart folowing the notes on the elog - SOAP errors as shown in attachment 29. called Tom who''s logged in
6.10 - seems to be working again. Not sure what happened. Tom thinks multiple restarts fixed it
Started data file R8
AIDA back in the main data
FRS shifters report a problem with the AIDA white rabbit time stamp - attachment 30
The following are timestamp values from each of the FEEs taken in sequence
If time does not increase in a reasonable manner run the system wide checks
aida01 : White Rabbit=> 17D8C4ED 8F46CF7B , WR/10=> 2627A17C18714BF, Readout Time => 2627A17C2A74000
aida02 : White Rabbit=> 17D8C4ED A0A1B623 , WR/10=> 2627A17C3435F03, Readout Time => 2627A17C4410000
aida03 : White Rabbit=> 17D8C4ED AFDACE68 , WR/10=> 2627A17C4C914A4, Readout Time => 2627A17C5D3C000
aida04 : White Rabbit=> 17D8C4ED C0516FC2 , WR/10=> 2627A17C66E8B2D, Readout Time => 2627A17C7AC4000
aida05 : White Rabbit=> 17D8C4ED D3B93DAB , WR/10=> 2627A17C85F52F7, Readout Time => 2627A17C9AEC000
aida06 : White Rabbit=> 17D8C4ED E8C9B243 , WR/10=> 2627A17CA7A91D3, Readout Time => 2627A17CB704000
aida07 : White Rabbit=> 17D8C4ED FFE475FD , WR/10=> 2627A17CCCA0BCC, Readout Time => 2627A17CDD94000
aida08 : White Rabbit=> 17D8C4EE 103232A6 , WR/10=> 2627A17CE6B6B77, Readout Time => 2627A17CF7BC000
aida09 : White Rabbit=> 17D8C4EE 20A1F612 , WR/10=> 2627A17D0103235, Readout Time => 2627A17D112C000
aida10 : White Rabbit=> 17D8C4EE 307D8FE7 , WR/10=> 2627A17D1A627FD, Readout Time => 2627A17D2958000
aida11 : White Rabbit=> 17D8C4EE 4402829D , WR/10=> 2627A17D399D9DC, Readout Time => 2627A17D4570000
aida12 : White Rabbit=> 17D8C4EE 6288B04D , WR/10=> 2627A17D6A744D4, Readout Time => 2627A17D7D94000
aida13 : White Rabbit=> 17D8C4EE 751E1306 , WR/10=> 2627A17D88301E7, Readout Time => 2627A17D9C34000
aida14 : White Rabbit=> 17D8C4EE 894E8806 , WR/10=> 2627A17DA87DA67, Readout Time => 2627A17DBFF4000
aida15 : White Rabbit=> 17D8C4EE 9EEEB997 , WR/10=> 2627A17DCB178F5, Readout Time => 2627A17DDC18000
aida16 : White Rabbit=> 17D8C4EE AF0BC923 , WR/10=> 2627A17DE4DFA83, Readout Time => 2627A17DF5C8000
Tom thinks everything appears ok with AIDA
Have converted WR timestamp to data/time - see attachment 40
At the ~second timescale AIDA WR timestamp looks OK, i.e. no gross errors. Will need to check correlations with other detector sub-systems to confirm WR on the ~us time timescale..
However, ~10us offset reported by DESPEC online crew is correct (AIDA timestamps data at a later point in the signal processing cycle than other detector sub-systems). Have suggested online crew contact Nic or Calum to check what the AIDA WR timediff spectra should look like.
7.00 checks - all looks ok
Voltage/currents - attachment 31
Rates attachment 32 + 33
Temps - attachment 36
Merger - attachment 37
ucesb - attachment 39
Grafana - attachment 38
Handed over to Tom
|