AIDA GELINA BRIKEN nToF CRIB ISOLDE CIRCE nTOFCapture DESPEC DTAS EDI_PSA 179Ta CARME StellarModelling DCF K40
  DESPEC, Page 35 of 36  ELOG logo
ID Date Author Subjectdown
  655   Wed Jun 12 06:17:23 2024 TD08:00-16:00 Wednesday 12 June
07:10 DSSSD bias & leakage current OK - attachment 1

      FEE64 temperatures OK - attachment 2

      ADC data items stats OK - attachment 3
   
      per FEE64 Rate spectra - attachment 4

      Merger, TapeServer etc - attachment 5

      per p+n FEE64 1.8.L spectra - attachment 6
       aida09 pulser peak width 68 ch FWHM 
07.13  peak area 25253 total area 27548 in 221mins => 1.9-2.1Hz => deadtime < 5% or less which is consistent with data file analysis below

      analysis data file S181/R6_274 - attachment 
       max FEE64 deadtime 24% aida08
       HEC data rate 4.5kHz, LEC data rate 2.284M


14.00 DAQ crashed by accident
      FEE64s power cycled 
      DAQ restarted
      data file S181/R7

      DSSSD bias & leakage current OK - attachments 8-9

      FEE64 temperatures OK - attachment 10

      ADC data item stats - attachment 11

      WR timestamps OK - attachment 12

      All system wide checks *except* as reported attachments 13-15

      per FEE64 Rate spectra - attachment 16

      per p+n FEE64 1.8.L spectra - attachment 17
       aida09 pulser peak width 66 ch FWHM

      per FEE64 1.8.H spectra - attachments 18-19

      per FEE64 1.8.W spectra - 20us FSR - attachment 20-21

      Merger, TapeServer etc - attachments 22-23
       data file S181/R7_16

      analysis data file S181/R7_16
       max deadtime 7.3% aida08, all other FEE64s < 2%
       LEC data rate 1.877M HEC data rate 4.7k
  590   Tue Apr 23 07:27:18 2024 PP08:00-16:00 Tuesday 24 April shift

Checks 08:00-08:30

Screenshots are attached and everything looks OK.

  650   Tue Jun 11 10:03:48 2024 TD, NH08:00-16:00 Tuesday 10 June
10.57 DSSSD bias & leakage current OK - attachment 1

      FEE64 temps OK - attachment 2

      ADC, MBS correlation scaler data item stats - attachments 3-4
       ADC 9x < 20k, max 243k aida08

      ASIC settings aida01 & aida02 - attachments 5-6

      ALL system wide checks OK *except* attachments 7-8

      WR timestamps OK - attachment 9

      BNC PB-5 settings - attachment 10

      per FEE64 Rate spectra - attachment 11

      per p+n FEE64 1.8.L spectra - attachment 12
       aida09 pulser peak width 57 ch FWHM

      per FEE64 1.8.H spectra - attachments 13-14

      per FEE64 1.8.W spectra - 20us FSR - attachments 15-16
       no p+n FEE64 wavefroms - reason unknown

      Merger, TapeServer etc - attachments 17-18

11.25 Lost control of aida-gsi window system 
      All command terminals lost, could not start new command terminals

      aida-gsi restart

      FEE64 power cycle
      DSSSD bias cycle

      Note file manager, calculator apps now working as expected

12.11 DAQ starts 
      data file S181/R6

      ASIC settings 2024Apr19-15.22.49

      BNC PB-5 Pulser
      Amplitude 10.0V
      Attenuation x10
      Fall time 1ms
      Frequency 2Hz
      Polarity + (connected to test + signal daisy chain to all p+n FEE64s)

      Sync ASIC clocks OK
      All system wide checks OK *except* aida02 FPGA/WR decoder - attachments 19-20

      per FEE64 1.8.W spectra - 20us FSR - attachments 21-22

12.29 analysis data file S181/R6_9 - attachment 23
      max dead time c. 2% (aida08)

15:34 For some reason the AIDA MBS wasn't connecting to the time sorter (or a check stream) [NH]
      Try to reset the AIDA MBS 
      Due to confusion the AIDA Merger was rebooted too, while the DAQ was running, this broke all the FEEs -> Powecycled
      Synced ASIC clocks, fixed all waveforms
      Masked all discriminators
      All looks ok -> Sending back to MBS and back in DESPEC DAQ
  605   Thu Apr 25 08:26:50 2024 TD08:00-16:00 Thursday 25 April
09.56 DAQ continues data file S100_alpha/R18_387
      FRS switching settings

      DSSSD bias & leakage current OK - attachments 1

      FEE64 temperatures OK  - attachment 2
       aida02 ASIC temp - known fault

      All system wide checks OK *except* as shown - attachments 3-5

      WR timestamps OK - attachment 6

      ADC, PAUSE, RESUME & Correlation Scaler data time stats - attachments 7-10

      per FEE64 Rate spectra - attachments 11

      per p+n FEE64 1.8.L spectra - attachments 12-13
       aida09 pulser peak width 57 ch FWHM 
       common x/y scale => peak height proxy for width => all p+n FEE64s similar pulser peak widths

      per n+n FEE64 1.8.L spectra - attachment 14

      per FEE64 1.8.H spectra - attachments 15-16

      per FEE64 1.8.W spectra - 20us FSR - attachments 17-18
       only aida08 & aida09 currently updating - ASIC check load and/or DAQ STOP/GO required?

      Merger, TapeServer etc - attachments 19-21

      BNC PB-5 settings - attachment 22

10.07 data file S100_alpha/R20
      new run for new FRS setting

10.08 all histograms, stats, merger stats and data link stats zero'd

10.10 ADC data item stats (beam OFF) - attachment 23
      11/16 < 20k, max 129k (aida08)

10.31 analysis of data file R18_349 (closed 06.00 this morning)
      max deadtime 12% (aida04)
      no timewarps
      HEC data item rate 1.7kHz

11.00 tuning 168dy FRS setting
      ucesb - attachment 24
      spill length 3.5s, extraction time 1.5s

11.30 degraders now set to implant in AIDA DSSSDs
      high rates expected 

      per FEE64 Rate spectra - attachment 25

      ADC data item stats - attachment 26

      Merger etc - attachment 27
      in spill rates to MBS up 30+Mb/s, 2.5M data item/s merged

      DSSSD bias & leakage current - attachment 28
       c. few hundred nA step in leakage current 

      analysis of data file R20_33 - attachment 29
      max deadtime 18% (aida08)
      no timewarps
      HEC data item rate 4.1kHz (c. 2.4x previous FRS setting)

      Correcting for spill cycle => 5.9kHz in spill (online analysis shows no HEC events observed outside spill)
      assume typical HEC event multiplicity = 3 => implant rate ~ 2.0kHz in spill or 6800 implants per spill
      SC42 L/R currently report c. 4.5kHz

      Note that we currently see a higher fraction of events in DSSSD#2 cf. previous setting so multiplicity assumption may be low
      If we assume multipicity = 6 => implant rate ~ 1.0kHz in spill or 3200 implants per spill

      we are probably somewhere in between

12.50 DSSSD bias & leakage current - Grafana - attachment 30
       leakage current ramping, spill micro structure observable

      FEE64 temperatures OK - attachment 31

      ADC data item stats - attachment 32

      per FEE64 Rate spectra - attachment 33

      Merger etc - attachment 34

14.56 analysis of data file R20_106 - attachment 35
      max deadtime c. 18% (aida04 & aida08), c. 12% (aida03), c. 11% (aida06), c. 9% (aida02 & aida13), c. 6% (aida05), all others <2%
      no timewarps
      HEC data item rate 4.2kHz


15.40 DSSSD bias & leakage current - Grafana - attachments 36-37
       leakage current ramping, spill micro structure observable

      FEE64 temperatures OK - attachment 38

      ADC data item stats - attachment 39

      per FEE64 Rate spectra - attachment 40

      Merger etc - attachment 41

      ucesb - attachment 42

15.57 all histograms zero'd




      
  615   Sun Apr 28 06:57:59 2024 Magda Satrazani08:00-16:00 Sunday 28 April

Took over the shift from Marc.

 

 

10.51 Added screenshots to illustrate erroneous data from Grafana (AIDA) - attachments 18-19

          Compare to the CAEN HV control terminal (attachment 15) which shows the expected values. Have email'd Nic & Helena. TD

10.55 Noticed that the CAEN HV control terminal is showing the corerct data but on the wrong line of the display (attachment 15). Enter commands q (quit) and d (display) to restart display - the correct data now appears on the correct line (attachment 20).

         Grafana (AIDA) is now reporting data correctly too (attachment 21).

          Summary - CAEN HV terminal display issue ( c. 10.05 - 10.55 )

                             Note the correct data/values continued to be shown. Leakage current shown correctly by both Grafana and CAEN HV control throughout.

                             Conclude this was a display issue - not a change in the operating parameters of the CAEN N1419ET.. Data should be OK.

 

14.45 analysis data file R21_500 - attachment 36

         max deadtime 22% (aida11), 16% (aida04), 7% (aida02), 6% (aida08), 6% (aida06) , all others < 1%

         no timewarps

         HEC data item rate 2.7kHz

 

  612   Sat Apr 27 04:59:31 2024 TD08:00-16:00 Saturday 27 April
06.00 DAQ continues file S100_alpha/R20_740
      FRS setting 162Eu

      DSSSD bias & leakage current - Grafana - attachments 1-2
       leakage currents tracking S4 ambient temperature

      FEE64 temperatures OK - attachment 3

      All system wide checks OK *except* 
       as reported - attachments 4-5
       *all* ADCs not calibrated (not calibrated post-synchronise ASIC clocks at most recent restart)

      WR timestamps OK - attachment 6

      ADC, PAUSE, RESUME, Correlation scaler data item stats OK - attachments 7-10

      per FEE64 Rate spectra - attachment 11

      per p+n FEE64 1.8.L specra - attachments 12-13
       aida09 pulser peak width 55 ch FWHM (~38keV FWHM)

      per FEE64 1.8.H spectra - attachments 14-15

      Merger, TapeServer etc - attachments 16-17
       12Mb/s to disk

      ucesb - attachment 18
       spill length 3.5s, extraction 1.5s

06.30 analysis data file R20_748 - attachment 19
      max deadtime 12%
      no timewarps
      HEC data item rate 2.1kHz

06.35 new run file S100_alpha/R21

09.30 DAQ continues - file S100_alpha/R21_83

      DSSSD bias & leakage currents - Grafana - attachment 20

      FEE64 temperatures OK - attachment 21

      ADC data item stats OK - attachments 22

      per FEE64 Rate spectra - attachment 23

      Merger etc - attachments 24

      ucesb - attachment 25
       spill length 3.5s, extraction 1.5s

09.50 per FEE64 1.8.W spectra - 20us FSR - attachments 26-27
       no significant change in electronic noise cf. start of run

      per p+n FEE64 1.8.L spectra - attachment 28
       aida09 pulser peak width 62 ch FWHM
       common x/y scale - peak heoght proxy for width


13.57 DAQ continues file R21_141

      DSSSD bias & leakage current - Grafana - attachments 29-30

      FEE64 temperatures OK - attachment 31

      ADC data item stats OK - attachments 32

      per p+n FEE64 1.8.L spectra - attachment 33
       aida09 pulser peak width 55 ch FWHM (~39keV FWHM)

      per FEE64 Rate spectra - attachment 34

      Merger etc - attachments 35

      ucesb - attachment 36
       spill length 3.5s, extraction 1.5s

14.15 analysis data file R21_144 - attachment 35
      max deadtime 16% (aida04)
      no timewarps
      HEC data item rate 2.0kHz
  662   Fri Jun 14 06:22:39 2024 TD08:00-16:00 Friday 14 June
07.23 All histograms & stats zero'd

      DAQ continues data file S181/R8_35

      analysis data file R8_35 - attachment 1
      max deadtime 4.8% (aida08), all other FEE64s < 3%
      LEC data rate 1.677M HEC data rate 3.7k

08.46 DSSSD bias & leakage current OK - attachments 2-3

      FEE64 temps OK - attachment 4

      ADC data item stats - attachment 5

      per FEE64 Rate spectra - attachment 6

      per p+n FEE64 1.8.L spectra - attachment 8
       aida09 pulser peak width 65 ch FWHM

      per FEE64 1.8.H spectra - attachment 7 & 10

      per FEE64 1.8.W spectra - 20us FSR - attachment 9 & 11

      Merger, TapeServer etc - attachments 12-13
       data file S181/R8_72


10.51 DSSSD bias & leakage current OK - attachments 14

      FEE64 temps OK - attachment 15

      ADC data item stats - attachment 16

      per FEE64 Rate spectra - attachment 17

10.54 analysis data file R8_122 - attachment 18
       max deadtime 7.8% (aida08), all other FEE64s < 3%
       ADC data rate 1.88M, HEC data rate 4.7k


14.24 DSSSD bias & leakage current OK - attachments 18

      FEE64 temps OK - attachment 19

      ADC data item stats - attachment 20

      per FEE64 Rate spectra - attachment 21

      aida02 no data

14.28 merger restarted and DAQ running OK 


14.38 analysis data file R8_213 - attachment 22
       max deadtime 6.6% (aida08), all other FEE64s < 3.5%
       ADC data rate 1.935M, HEC data rate 3.6k
  659   Thu Jun 13 06:49:54 2024 Magda Satrazani08:00 - 16:00 Thursday 13 June 2024

Took over the shift from Marc.

08:00 FULL CHECK
DSSSD bias & leakage current  ok - attachment # 1-2
FEE64 temperatures  ok - attachment # 3
ADC data item stats - attachment # 4
Merger ok - Attachement # 5
Tape service - attachement # 6

10:00 FULL CHECK
DSSSD bias & leakage current  ok - attachment # 7-8
FEE64 temperatures  ok - attachment # 9
ADC data item stats - attachment # 10
Merger ok - Attachement # 11
Tape service - attachement # 12

12:00 FULL CHECK
DSSSD bias & leakage current  ok - attachment # 13-14
FEE64 temperatures  ok - attachment # 15
ADC data item stats - attachment # 16
Merger ok - Attachement # 17
Tape service - attachement # 18

 

 

12.34 analysis data file S181/R7_610 - attachment 19

max deadtime 8.7% (aida08), all other FEE64s < 2.7%

LEC data rate 2.000M, HEC data rate 4.7k

 

12.52 per p+n FEE64 1.8.L spectra - attachment 20

 aida09 pulser peak width 63 ch FWHM

per FEE64 1.8.H spectra - attachments 21-22

per FEE64 1.8.W spectra - 20us FSR - attachments 23-24

 

13.00 all histograms zero'd

 

14:00 FULL CHECK
DSSSD bias & leakage current  ok - attachment # 19-20
FEE64 temperatures  ok - attachment # 21
ADC data item stats - attachment # 22
Merger ok - Attachement # 23
Tape service - attachement # 24

 

13.01 Beam OFF for S4 access

ADC data item stats - attachment 25

per FEE64 Rate spectra - attachment 26

14.12 analysis data file R7_654 - attachment 27

max deadtime 3.9% (aida08), all other FEE64s < 3%

LEC data rate 1.817M, HEC data rate 2.5k

14.00 data file R7_649 - attachment 28

max deadtime 4.5% (aida08), all other FEE64s < 3%

LEC data rate 1.756M, HEC data rate 2.5k

 

 

16:00 FULL CHECK
DSSSD bias & leakage current  ok - attachment # 25-26
FEE64 temperatures  ok - attachment # 27
ADC data item stats - attachment # 28
Merger ok - Attachement # 29
Tape service - attachement # 30

  608   Fri Apr 26 07:01:26 2024 Norah08:00 - 16:00 Friday 26 April

10:00 Checks:

All seems smooth.

DSSD bias and  leakage current ( Attachment 1)

FEE64 temperatures (Attachment 2)

Per FEE64 Rate spectra (Attachment 3)

ADC data item stats -(Attachment 4)

Ucesb (Attachment 5)

 

10:21

Aida05 was zero not reading had to call Nic to fix it.

10:34

Now, Aida05 is running well.

.......................................................

12:00 Checks:

Everything appears to be going smoothly, and I took screenshot of them.

 

12:28, Aida02 showed as not reading , had to email Nic to fix it.

12:30, it resumed reading and returned to normal operation.

.........................................................

14:00 Checks:

Everything is going well , and I have captured screenshots of each one.

......................................................

16:00 Checks:

All seems smooth, and I took screenshot of them.

 

 

 

 

 

  647   Mon Jun 10 07:15:48 2024 Dan08.00-16.00 Monday 10th June

8am checks

temps ok - attachment 2

rates ok - attachment 8

voltage ok - attachment 5

merger ok - attachment  6

 

9am checks

temps ok - attachment 10

rates ok - attachment 9

voltage ok - attachment 12

merger ok - attachment  11

 

(Ignore  Grafana plots 13-43 inc - not refreshed)

 

9.16 beam stopped to optimise beam intensity

 

10.00 checks - still no beam

temps ok - attachment 17

rates ok - attachment 15+16

voltage ok - attachment 14

merger ok - attachment  18

 

11.00 checks - still no beam

temps ok - attachment 23

rates ok - attachment 21+22

voltage ok - attachment 20

merger ok - attachment  24

 

12pm checks

temps ok - attachment 28

rates ok - attachment 26+27

voltage ok - attachment 29

merger ok - attachment  30

 

12.45 - beam back

 

temps ok - attachment 34

rates ok - attachment 32+33

voltage ok - attachment 35

merger ok - attachment  36

 

2pm checks

temps ok - attachment 40

rates ok - attachment 38+39

voltage ok - attachment 41

merger ok - attachment  42

 

/tmp/R4_276 added - attachment 44

max deadtime 10.5% (aida08)

n+n FEE64s deadtime 

 

3pm checks

temps ok - attachment 48

rates ok - attachment 46+47

voltage ok - attachment 45+50

merger ok - attachment  49

  637   Mon Jun 3 14:40:34 2024 JB, NH03 June 2024

15:40 reseated FEE64s adaptors of aida01, aida09, aida13 & aida16.

TEMP OK. HV OK. System wide checks all passed excl. aida07 ADC calibration, see attached.

Histograms and rates look same as previous, reseating aida01 seems to have worked. aida16 still unhappy.

re-reseated aida16. 16:19 - appears to have fixed the problem in the channel.

Pulser test + connected.

bPlast ground removed from frame.

To-do:

- Need clarification about links connected on AIDA n+n FEE64 adaptor boards. Currently, the situation is given by the attachment 8 & 9.

- Fix BB7 grounding, continuity between snout and frame observed.

- Pulser walkthrough for n+n, can do p+n when convienient.

 

 

18.30 Noted FEE64 ADC stats now significantly higher ( attachments 9-10 ) cf. earlier this afternoon (attachments 1-2)

DSSSD bias OFF. FEE64 power OFF.

 

 

  600   Tue Apr 23 22:37:46 2024 TD00:00-08:00 Wednesday 24 April
23.34 23.4.24

      DSSSD bias & leakage current OK - attachment 2

      FEE64 temperatures OK - attachment 2

      ADC data item stats - attachment 3

      per FEE64 rate spectra - attachments 4

      Merger etc - attachment 5

03.41 fDSSSD bias & leakage current OK - attachment 6

      FEE64 temperatures OK - attachment 7

      ADC data item stats - attachment 8

      per FEE64 rate spectra (off and on spill) - attachments 9-10

      Merger etc - attachment 11

03.54 analysis data file R18_67 - attachment 12
      max deadtime 11% (aida04)
      HEC data item rate c. 1.5kHz
      no timewarps

0.35 fDSSSD bias & leakage current OK - attachment 13

      FEE64 temperatures OK - attachment 14

      ADC data item stats - attachment 15

      per FEE64 rate spectra - attachments 16

      Merger etc - attachment 17

03.54 analysis data file R18_96 - attachment 18
      max deadtime 12% (aida04)
      HEC data item rate c. 1.6kHz
      no timewarps
  654   Tue Jun 11 23:13:44 2024 TD00:00-08:00 Wednesday 11 June
00.08 DSSSD bias & leakage current OK - attachment 1

      FEE64 temperatures OK - attachment 3

      ADC data items stats OK - attachment 2
   
      per FEE64 Rate spectra - attachment 7

      per p+n FEE64 1.8.L spectra - attachments 6
       aida09 pulser peak width 63 ch FWHM 

      Merger, TapeServer etc - attachments 4-5

      analysis data file S181/R6_274 - attachment 8
       max FEE64 deadtime 19% aida08
       HEC data rate 3.8kHz, LEC data rate 2.258M

      per FEE64 1.8.W spectra - 20us FSR - attachments 9-10

03.12 Merger, TapeServer etc - attachments 11-12
       before & after Merger restart

      per FEE64 Rate spectra - attachment 13

      ADC data items stats OK - attachment 14
       aida02 zero rate

      FEE64 temperatures OK - attachment 15

      DSSSD bias & leakage current OK - attachment 16

      analysis data file S181/R6_372 - attachment 17
       max FEE64 deadtime 22% aida08
       HEC data rate 4.4kHz, LEC data rate 2.295M

03.32 all histograms zero'd

03.57 per p+n FEE64 1.8.L spectra - attachment 18
       25 mins = 1500 @ 2Hz = 3000 cf. peak area c. 3000 => deadtime low
  604   Wed Apr 24 21:39:51 2024 TD00:00-08:00 Thursday 25 April
22.43 24.4.24

      DSSSD bias & leakage current OK - attachment 1

      FEE64 temperatures OK - attachment 2

      ADC data item stats - attachment 3

      per FEE64 rate spectra - attachments 4

      Merger etc - attachment 5

      analysis of data file R18_266 - attachment 6
       max dead time 12% (aida04)
       no timewarps
       HEC data item rate 1.7kHz

03.37 DSSSD bias & leakage current OK - attachments 7-8

      FEE64 temperatures OK - attachment 9

      ADC data item stats - attachment 10

      per FEE64 rate spectra - attachments 11

      Merger etc - attachment 12
  658   Wed Jun 12 23:00:16 2024 Marc00:00-08:00 Thursday 13 June

Starting a new night shift - All good

1am full check:
DSSSD bias & leakage current  ok - attachment # 1-2
FEE64 temperatures  ok - attachment # 3
ADC data item stats - attachment # 4
Merger ok - Attachement # 5
Tape service - attachement # 6

2:30am: AIDA02 FEE stopped working. DESPEC main DAQ was alerted and remobe AIDA from timestitcher while I was restarting AIDA DAQ.

2:45am: AIDA02  FEE is back and AIDA has been added to the timestitcher again in the main DAQ.

The AIDA runs at the time of the crash to look at are R7_348, 349.

All spectra were zeroed at this time and checked.

4:00 am full check:
DSSSD bias & leakage current  ok - attachment # 7-8
FEE64 temperatures  ok - attachment # 9
ADC data item stats - attachment # 10
Merger ok - attachement # 11
Tape service - attachement # 12
Spectra - attachement # 13
ucesb - attachement # 14

6:00am full check:
DSSSD bias & leakage current  ok - attachment # 15-16
FEE64 temperatures  ok - attachment # 17
ADC data item stats - attachment # 18
Merger ok - attachement # 19
Tape service - attachement # 20
Spectra - attachement # 21
ucesb - attachement # 22

6:30 the nearline/offline analysis team reported possible noise issue in DSSD#2 and asked if the threshold were changed at any point. I can't find any mention of changes in the elog. 
I've not done a setup of the all DAQ when AIDA02 fee crashes earlier this morning. I just did a setup of the merger. Thus I don't think the threshold have changed when I stopeed and restarted the DAQ.
I can see that currently slow comparator threshod are set to 0xa for most of the DSSD2 fees except for AIDA06 which is set to 0xf. I'm not sure this is the thershold we would need to increase. Something to check during the next shift.

In the UCESB window I can see that the number of decay in DSSD2 is double the number of decay in DSSD1 while the number of implant is similar. Looking at other shift it looks that this was alsready the case in prvious shifts.

 

07.06 analysis of data file S181/R7_457 - attachment 23

 max deadtime 5.8% aida08, all other FEE64s < 1.4%

LEC data rate 1.788M, HEC data rate 4.3k

 

  614   Sat Apr 27 23:04:42 2024 Marc00:00-08:00 Sunday 28 April

Encore un matin, un matin pour ..... a shift !!!ramping up again ! S4 temperature not really going down either: 25C

0:40 AIDA04 is down. I informed the DAQ shift crew and attempting to restart AIDA04.

After rebooting AIDA04 all seemed going until I reached the merger check. No data was being merged.

I realise then that AIDA01 data transfer was not active. I did not spot it when I enabled the data transfer for all.

I stopped the daq and rebooted AIDA01 following all the instructions in the ELOG and this worked.

We are now back in business with AIDA back in main DAQ at  01:40

01:45 full checks:

           Bias: AIDA-Graphana - Attachment 1

               Leakage current ramping up again ! S4 temperaturenot really going down either: 25C

      FEE64 temperatures OK - attachment 2

      ADC data item stats - attachment 3

      per FEE64 Rate spectra - attachment 4

      Merger etc - attachment 5

      ucesb - attachment 6

03:00 full checks

          No change to be highlighted. All pretty much the same as during previous full checks.

05:00 full checks

Expt still running very smoothly.

           Bias: AIDA-Graphana - Attachment 7-8

               Leakage current going down slightly

      FEE64 temperatures OK - attachment 9

      ADC data item stats - attachment 10
ramping up again ! S4 temperaturenot really going down either: 25C
      per FEE64 Rate spectra - attachment 11

      Merger etc - attachment 12

      ucesb - attachment 13

07:15 all running smoothly.

07:20 Spoke too fast AIDA DAQ is down. DAQ shift crew informed.

07:40: AIDA10 was rebooted successfully and AIDA is now back in the main DAQ.

 

  611   Fri Apr 26 23:04:51 2024 Marc00:00-08:00 Saturday 27 April

It's a new day, it's a new shift and I'm feeeeeliiiiing goooood !!!!!

0.15 Checks:
      DSSSD bias & leakage current - Grafana - attachments 1-2
       leakage current dropped a bit since the evening shift ~22:00.

      FEE64 temperatures OK - attachment 3

      ADC data item stats - attachment 4

      per FEE64 Rate spectra - attachment 5

      Merger etc - attachment 6

      ucesb - attachment 7

2:20 Nothing new to report. All good. Leakage currents continue to go down slowly. See attachement 8 to 12

3:30 Still running smoothly. 

5:20 AIDA 4 stop counting in the last 20 min. I informed the main DAQ and they started a new run without AIDA.

        DAQ restarted OK. Main DAQ informed.

 

  649   Mon Jun 10 23:36:14 2024 TD0000-08:00 Tuesday 11 June
00.09 DSSSD bias & leakage current OK - attachment 1

      FEE64 temperatures OK - attachment 2

      ADC data items stats OK - attachment 3
   
      per FEE64 Rate spectra - attachment 4

      per p+n FEE64 1.8.L spectra - attachments 5-6
       aida09 pulser peak width 53 ch FWHM 

      per FEE64 1.8.H spectra - attachments 7-8

      per n+n FEE64 1.8.W spectra - 20us FSR - attachment 9
       p+n 1.8.W not running currently

      Merger, TapeServer etc - attachments 10-11

      analysis data file S181/R4_427 - attachment 12
       max FEE64 deadtime 1.7%
       HEC data rate 3.1kHz

00.28 all histograms zero'd


02.39 DAQ continues data file S181/R4_460
      Beam off since 02.08

      DSSSD bias & leakage current OK - attachment 13

      FEE64 temperatures OK - attachment 14

      ADC data items stats OK - attachment 15
       9x < 20k, max 239k aida08
   
      per FEE64 Rate spectra - attachment 16


07.15 DAQ continues data file S181/R4_517
      Beam off since 02.08
      HKR report c 3-4 h until UNILAC fixed

      DSSSD bias & leakage current OK - attachment 17

      FEE64 temperatures OK - attachment 18

      ADC data items stats OK - attachment 19
       8x < 20k, max 238k aida08
   
      per FEE64 Rate spectra - attachment 20
  589   Mon Apr 22 22:35:59 2024 TD00.00-08.00 Tuesday 22 April
23.36 Analysis of data file R17_34 - attachment 1
      max deadtime c. 1% (aida02)
      no timewarps

23.38 DSSSD bias & leakage current OK - attachment 2

      FEE64 temperatures OK - attachment 3
       aida02 ASIC temp - known fault

      ADC data item stats - attachment 4

      per FEE64 Rate spectra - attachment 5

      Merger etc - attachment 6

01.03 DSSSD bias & leakage current OK - attachment 7

      FEE64 temperatures OK - attachment 8
       aida02 ASIC temp - known fault

      ADC data item stats - attachment 9

      per FEE64 Rate spectra - attachment 10

      Merger etc - attachment 11

03.06 DSSSD bias & leakage current OK - attachment 12

      FEE64 temperatures OK - attachment 13
       aida02 ASIC temp - known fault

      ADC data item stats - attachment 14

      per FEE64 Rate spectra - attachment 15

      Merger etc - attachment 16

05.04 DSSSD bias & leakage current OK - attachment 17

      FEE64 temperatures OK - attachment 18
       aida02 ASIC temp - known fault

      ADC data item stats - attachment 19

      per FEE64 Rate spectra - attachment 20

      Merger etc - attachment 21

07.29 DSSSD bias & leakage current OK - attachment 22

      FEE64 temperatures OK - attachment 23
       aida02 ASIC temp - known fault

      ADC data item stats - attachment 24

      per FEE64 Rate spectra - attachment 25

      Merger etc - attachment 26
  661   Thu Jun 13 23:04:01 2024 Dan Judson, TD0.00-8.00 14/6/24

0.00 checks - all looks ok

Voltage/currents - attachment 1

Rates attachment 2 + 3

Temps - attachment 4

Merger - attachment 5

ucesb - attachment 6

Grafana - attachment 7

 

1.30 -aida1 in merger Links, aida02 in rate spectra, has stopped - resetting DAQ.

Datafile R7_961 being written when stopped

1.40 - seems ok

 

1.50 checks - seems ok

Voltage/currents - attachment 8

Rates attachment 9 + 10

Temps - attachment 11

Merger - attachment 12

ucesb - attachment 13

Grafana - attachment 14

 

3.30 -aida1 in merger Links, aida02 in rate spectra, has stopped again - resetting DAQ.

Datafile R7_1010 being written when stopped

3.38 - seems ok

 

3.40 checks - all looks ok

Voltage/currents - attachment 15

Rates attachment 16 + 17

Temps - attachment 18

Merger - attachment 19

ucesb - attachment 20

Grafana - attachment 21

 

Zeroed spectra - attachments 22-28

 

 

5.39 -aida1 in merger Links, aida02 in rate spectra, has stopped again - resetting DAQ.

Datafile R7_1062 being written when stopped

AIDA out of the main data

Could not get it to restart folowing the notes on the elog - SOAP errors as shown in attachment 29. called Tom who''s logged in 

6.10 - seems to be working again. Not sure what happened. Tom thinks multiple restarts fixed it

Started data file R8

AIDA back in the main data

 

FRS shifters report a problem with the AIDA white rabbit time stamp - attachment 30

The following are timestamp values from each of the FEEs taken in sequence
If time does not increase in a reasonable manner run the system wide checks

aida01 : White Rabbit=>  17D8C4ED 8F46CF7B , WR/10=>   2627A17C18714BF, Readout Time =>   2627A17C2A74000
aida02 : White Rabbit=>  17D8C4ED A0A1B623 , WR/10=>   2627A17C3435F03, Readout Time =>   2627A17C4410000
aida03 : White Rabbit=>  17D8C4ED AFDACE68 , WR/10=>   2627A17C4C914A4, Readout Time =>   2627A17C5D3C000
aida04 : White Rabbit=>  17D8C4ED C0516FC2 , WR/10=>   2627A17C66E8B2D, Readout Time =>   2627A17C7AC4000
aida05 : White Rabbit=>  17D8C4ED D3B93DAB , WR/10=>   2627A17C85F52F7, Readout Time =>   2627A17C9AEC000
aida06 : White Rabbit=>  17D8C4ED E8C9B243 , WR/10=>   2627A17CA7A91D3, Readout Time =>   2627A17CB704000
aida07 : White Rabbit=>  17D8C4ED FFE475FD , WR/10=>   2627A17CCCA0BCC, Readout Time =>   2627A17CDD94000
aida08 : White Rabbit=>  17D8C4EE 103232A6 , WR/10=>   2627A17CE6B6B77, Readout Time =>   2627A17CF7BC000
aida09 : White Rabbit=>  17D8C4EE 20A1F612 , WR/10=>   2627A17D0103235, Readout Time =>   2627A17D112C000
aida10 : White Rabbit=>  17D8C4EE 307D8FE7 , WR/10=>   2627A17D1A627FD, Readout Time =>   2627A17D2958000
aida11 : White Rabbit=>  17D8C4EE 4402829D , WR/10=>   2627A17D399D9DC, Readout Time =>   2627A17D4570000
aida12 : White Rabbit=>  17D8C4EE 6288B04D , WR/10=>   2627A17D6A744D4, Readout Time =>   2627A17D7D94000
aida13 : White Rabbit=>  17D8C4EE 751E1306 , WR/10=>   2627A17D88301E7, Readout Time =>   2627A17D9C34000
aida14 : White Rabbit=>  17D8C4EE 894E8806 , WR/10=>   2627A17DA87DA67, Readout Time =>   2627A17DBFF4000
aida15 : White Rabbit=>  17D8C4EE 9EEEB997 , WR/10=>   2627A17DCB178F5, Readout Time =>   2627A17DDC18000
aida16 : White Rabbit=>  17D8C4EE AF0BC923 , WR/10=>   2627A17DE4DFA83, Readout Time =>   2627A17DF5C8000

Tom thinks everything appears ok with AIDA

Have converted WR timestamp to data/time - see attachment 40

At the ~second timescale AIDA WR timestamp looks OK, i.e. no gross errors. Will need to check correlations with other detector sub-systems to confirm WR on the ~us time timescale.. 

However, ~10us offset reported by DESPEC online crew is correct (AIDA timestamps data at a later point in the signal processing cycle than other detector sub-systems). Have suggested online crew contact Nic or Calum to check what the AIDA WR timediff spectra should look like.

 

 

7.00 checks - all looks ok

Voltage/currents - attachment 31

Rates attachment 32 + 33

Temps - attachment 36

Merger - attachment 37

ucesb - attachment 39

Grafana - attachment 38

 

Handed over to Tom

 

ELOG V3.1.4-unknown