  763   Tue Aug 26 14:43:44 2025 JM, CB, EWCARME on, background run

26 August

14:40

Turned CARME on 

Everything started up okay, all detectors on and stable. 

Note aida16 clock and ADC are uncalibrated. Note dssd4 is on MEC

Detector 1 noisy. 

AIDA started - all okay. 

All FEEs set to LEC (0x0) and slow comparator threshold set to 300keV (0x1e). 

Currently writing to data to G22-205/R100, doing background run, ~400 kB/s. 

16:40

Checks OK. 

19:00

Checks OK

22:30

Temps OK, Rates OK

x2go window size weird on different monitor as well.

27th August

Stopped Run 100. We had accidently set the threshold to 140 keV (0xe) not 300 keV (0x1e) as intended. 

Either way we require data on the alphas for electronics testing in October and the data rate at this threshold is too high. Threshold set to 0x64 (1 MeV) to look only at the alphas. 

Run 101 started. See attached leakage currents

20:30

Checks OK

28th August

10:00

Checks OK

CARME daq crashed this afternoon. Aida14 had a disconnection error. Seemed to coincide with dssd4 tripping again. 

Rebooting FEEs. On reboot there are always some FEEs which seem to have some error. The FEE changes each time. i.e aida12 one time, aida09 next time. I will give them a little while to compose themselves before trying again. 

Unable to get the system writing data to disk again. I think the system is unhappy with the merger / tape server not being closed properly. Will have a proper look tomorrow. 

29th August

20:15

Running again R102. 0x64 slow comparator thresholds. DSSD4 off due to leakage current tripping -> MEC

30th August

Daq continues R102

31st August

Daq continues R102

11:30

R102 stopped. CARME shutdown. This week, potentially tomorrow morning, the Kasell group will dismount the VUV spectrometer. CARME off and bias off during these works. 

Attachment 1: bias.png
bias.png
  762   Thu Jul 17 07:07:56 2025 CBCARME off

08:07 CEST

Temperatures ok - attached

DSSDs on. Leakage currents still high. - attached

If ML & co cannot find the reason, will turn off FEEs later today

FEEs turned off, leakage current still high (DSSD off now). 

Motors out. 

Attachment 1: Capture.PNG
Capture.PNG
Attachment 2: 2.PNG
2.PNG
Attachment 3: 3.PNG
3.PNG
Attachment 4: Capture.PNG
Capture.PNG
  761   Wed Jul 9 11:21:14 2025 EW, CBCARME restarted - long background R98

---9 July---

11:00

Turned detectors on - all stable see attachment 1 

AIDA started - all okay. See attachment 2 for FEE temps.  

All FEEs set to LEC (0x0) and slow comparator threshold set to 300keV (0x1e). See attachment 3. 

After some issues tape server and merger all set up, NetVar not working??? have tried to reboot things but still getting issues. See attachment 4 for tape service and attachment 5 for merger. 

Currently writing to data to G22-205/R98 (R97 empty), doing background run.

13:30

All checks okay apart from white rabbit, see attachment 6.

Attachment 7 - stats 

18:30 CEST

More issues with anydesk - now refuses to connect with error message desk_rt_ipc

May be as simple as the monitor being turned off. Will have to be checked in person

Connected to atppc025 via x2go and from there checked

- Temperatures are fine (attached)

- Detector bias / leakage current is fine (attached)

- Data rate from file size being written is around 500 kB/s (fine)

- A number of interesting features in the online background spectrum. Likely all non-physical (attached)

- Stats OK (attached)

20:50 CEST

All checks good.

22:28 CEST

All checks good.

Features in the XY plot rensemble beam hitting the detectors. Attached.

Leakage currents and rates are good - unlikely to be major detector damage. Attached.

Only beam in the ring between now and end of D+D was stable uranium.

Killed CARMEsort via terminal (still no access to anydesk).

Restarted CARMEsort. No obvious structures appear after a few minutes. Was there beam in the ring today?

23:09 CEST

All checks good

Features in the XY plot and Ey plots disappeared.

 

 

---10 July---

03:22 CEST

All checks good.

Features gone. Attached.

05:15 CEST

All checks good.

07:34 CEST

All checks good.

10:40 CEST

All checks good. 

12:35 CEST

All checks good. Some features in plots again appeared in last 2 hours, attached Ey plot (attachment 15). 

14:40 CEST

All checks okay. 

16:40 CEST

All checks okay. 

18:40 CEST

All checks okay. Attachment 16 leakage current stable. 

Attachment 17 Ey plot.

19:35 CEST

All checks OK.

Reset the Root spectra again - highly unlikely to have beam overnight without experiments running.

23:39 CEST

All checks OK.

Still seeing peak at 6 MeV like attachment 17, after resetting the Root spectra. Detector damage? Still seems implausible.

Events are focussed as in attachment 12.

Reset again. No obvious peak or pattern after a few minutes running.

 

From Claude K

"CRYRING was running with the Uranium beam from ESR until Tuesday (8 July) night around 22:00.

From Wednesday noon we started to inject D+ from the local source and had the D+ beam stored at various energies from 14:30 until 17:40. During the night from Wednesday (9 July) to Thursday (10 July) there was no beam.

We then continued measurements with stored D+ On Thursday from ca. 11:00 to 21:00. Since then CRYRING is in shutdown."

 

 

---11 July---

08:07 CEST

All checks OK.

Peak at 6 MeV did not re-appear overnight. Attached x2.

12:00 CEST

All checks good! All plots look good!

16:00 CEST 

All checks okay! 

19:20 CEST

All checks OK.

Background peaks almost visible.

22:53 CEST

All checks OK.

 

---12 July---

07:25 CEST

All checks OK.

10:00 CEST

All checks OK. Temps attached.

11:00 CEST

All checks OK.

19:00 CEST

All checks OK.

23:00 CEST

All checks OK.

 

---13 July---

07:42 CEST

All checks OK.

11:00 CEST

All checks OK.

 

ELOG down during the afternoon

 

--14 July--

11:00 CEST

All checks ok. 

14:40 CEST

Tape service and merger now working on atppc025

All checks ok, briefly went no storage at around 14:38 CEST for 1-2 minutes whilst getting tape service and merger up. 

17:40 CEST

All checks ok. 

21:30 CEST
All checks OK.

23:16 CEST

All checks OK. Several screenshots attached.

Calibration appears to hinders clear alpha peaks.

 

---15 July---

07:47 CEST

All checks OK.

11:55 CEST

All checks ok. 

14:35 CEST

Issue with atppc025, see attachment 29 for error message (desktop 2, detectors etc)

Got error to go, all checks ok. 

17:35 CEST

Same issue with atppc025, otherwise all checks ok. 

20:30 CEST
All checks OK.

23:07 CEST

All checks OK

 

---16 July---

07:41  CEST

All checks OK

12:00 CEST

All check ok. 

15:15 CEST

Some fees have high rates, all other checks ok. 

15:32 CEST

The above behaviour appears to be correlated with an increase in leakage current across all four detectors. Most likely light?

The background is not reliable -> No storage.

Unfortunately the leakage current logger crashed a few days ago on the server, and without anydesk we did not realise. Restarted it on atppc025.

Informed ML about issue. If it cannot be solved we could increase the thresholds and keep going, but the background may not be representative.

16:35 CEST

Issue persists. ML and ZA are not aware of any changes, except power-saving mode enabled for cave AC. No effect on FEE temperatures - not the issue.

aida04 rates shows 0. Checkloaded. Fine for a few seconds. Back to 0. Data event rate 100,000 may be too high.

Increased thresholds to 1 MeV. Data rates lowered, but e.g. on aida04 still 40,000. Goes back to zero. Checkloaded. OK for a short while, then back to zero.

Tape server stopped. DAQ stopped - all FEEs respond, good! Merger still going.

DSSDs off. Will retry tomorrow morning.

23:44 CEST

Temps OK.

Attachment 1: Screenshot_2025-07-09_at_11.19.43.png
Screenshot_2025-07-09_at_11.19.43.png
Attachment 2: Screenshot_2025-07-09_at_11.16.35.png
Screenshot_2025-07-09_at_11.16.35.png
Attachment 3: Screenshot_2025-07-09_at_11.16.07.png
Screenshot_2025-07-09_at_11.16.07.png
Attachment 4: Screenshot_2025-07-09_at_11.15.57.png
Screenshot_2025-07-09_at_11.15.57.png
Attachment 5: Screenshot_2025-07-09_at_11.15.47.png
Screenshot_2025-07-09_at_11.15.47.png
Attachment 6: Screenshot_2025-07-09_at_13.31.55.png
Screenshot_2025-07-09_at_13.31.55.png
Attachment 7: Screenshot_2025-07-09_at_13.39.19.png
Screenshot_2025-07-09_at_13.39.19.png
Attachment 8: 1.png
1.png
Attachment 9: 2.png
2.png
Attachment 10: 3.png
3.png
Attachment 11: 4.png
4.png
Attachment 12: Capture.PNG
Capture.PNG
Attachment 13: Capture2.PNG
Capture2.PNG
Attachment 14: Capture.PNG
Capture.PNG
Attachment 15: Screenshot_2025-07-10_at_11.39.51.png
Screenshot_2025-07-10_at_11.39.51.png
Attachment 16: Screenshot_2025-07-10_at_17.38.52.png
Screenshot_2025-07-10_at_17.38.52.png
Attachment 17: Screenshot_2025-07-10_at_17.38.39.png
Screenshot_2025-07-10_at_17.38.39.png
Attachment 18: Capture.PNG
Capture.PNG
Attachment 19: Capture2.PNG
Capture2.PNG
Attachment 20: Capture.PNG
Capture.PNG
Attachment 21: Capture2.PNG
Capture2.PNG
Attachment 22: 1.PNG
1.PNG
Attachment 23: 2.PNG
2.PNG
Attachment 24: 3.PNG
3.PNG
Attachment 25: 4.PNG
4.PNG
Attachment 26: 5.PNG
5.PNG
Attachment 27: 7.PNG
7.PNG
Attachment 28: 6.PNG
6.PNG
Attachment 29: IMG_4779.jpeg
IMG_4779.jpeg
Attachment 30: 1.PNG
1.PNG
Attachment 31: 2.PNG
2.PNG
Attachment 32: 3.PNG
3.PNG
  760   Sun Jul 6 11:35:28 2025 CBAnydesk version updated

Updated anydesk to 6.3.2 CentOS7 release.

Works now.

  759   Sat Jul 5 08:51:33 2025 EBManydesk connection settings
Attachment 1: anydesk_settings.png
anydesk_settings.png
  758   Wed Jun 18 14:21:39 2025 CBCARME off

15:20

DAQ stopped

Merger & Tape Server stopped

FEEs off

All detectors off

Leakage current plot attached

Leakage current file moved to /TapeData/

 

Will restart for a long background when the next experiment is over.

Attachment 1: Screenshot_from_2025-06-18_15-18-31.png
Screenshot_from_2025-06-18_15-18-31.png
  757   Tue Jun 17 23:51:58 2025 TD, CB, JGRun 96 - continued
00.31 DAQ continues data file G22-205/R96_27
      
      DSSSD bias & leakage currents - attachments 7-8
      FEE64 temps OK - attachment 6
      ADC data item stats - attachment 5
      per FEE64 Rate spectra - attachment 4
      Merger & TapeServer - attachments 2-3

      Online analysis - attachment 1

      Analysis data file R96_29 - attachment 9
       no timewarps, all FEE64 deadtimes <<1%

04.04 FEE64 temps OK - attachment 10
      ADC data item stats - attachment 11
      DSSSD bias & leakage currents - attachments 12 & 15
      per FEE64 Rate spectra - attachment 13
      TapeServer - attachment 14

08.26 DAQ continues data file G22-205/R96_64

      DSSSD bias & leakage currents - attachments 16-17
      FEE64 temps OK - attachment 18
      ADC data item stats - attachment 19
      per FEE64 Rate spectra - attachment 20
      TapeServer - attachment 21

11:03 All checks OK
EyTotal histogram attached.

12:25 Temps OK. Rates OK.

15:03 run stopped - beam time over
all checks ok
EyTotal hist attached
Attachment 1: Screenshot_from_2025-06-18_00-51-35.png
Screenshot_from_2025-06-18_00-51-35.png
Attachment 2: Screenshot_from_2025-06-18_00-50-35.png
Screenshot_from_2025-06-18_00-50-35.png
Attachment 3: Screenshot_from_2025-06-18_00-50-18.png
Screenshot_from_2025-06-18_00-50-18.png
Attachment 4: Screenshot_from_2025-06-18_00-50-00.png
Screenshot_from_2025-06-18_00-50-00.png
Attachment 5: Screenshot_from_2025-06-18_00-48-42.png
Screenshot_from_2025-06-18_00-48-42.png
Attachment 6: Screenshot_from_2025-06-18_00-31-31.png
Screenshot_from_2025-06-18_00-31-31.png
Attachment 7: Screenshot_from_2025-06-18_00-31-06.png
Screenshot_from_2025-06-18_00-31-06.png
Attachment 8: Screenshot_from_2025-06-18_00-30-49.png
Screenshot_from_2025-06-18_00-30-49.png
Attachment 9: R96_29
 *** TDR format 3.3.0 analyser - TD - May 2021
 *** ERROR: READ I/O error:       5002
                   blocks:      32000
          ADC data format:  261850116 (  336634.1 Hz)
        Other data format:      69888 (      89.8 Hz)
 Sample trace data format:          0 (       0.0 Hz)
         Undefined format:          0 (       0.0 Hz)
   Other data format type:      PAUSE:         46 (       0.1 Hz)
                               RESUME:         46 (       0.1 Hz)
                              SYNC100:      34898 (      44.9 Hz)
                              WR48-63:      34898 (      44.9 Hz)
                           FEE64 disc:          0 (       0.0 Hz)
                             MBS info:          0 (       0.0 Hz)
                           Other info:          0 (       0.0 Hz)

   ADC data range bit set:         71 (       0.1 Hz)

                Timewarps:        ADC:          0 (       0.0 Hz)
                                PAUSE:          0 (       0.0 Hz)
                               RESUME:          0 (       0.0 Hz)
                              SYNC100:          0 (       0.0 Hz)
                              WR48-63:          0 (       0.0 Hz)
                           FEE64 disc:          0 (       0.0 Hz)
                             MBS info:          0 (       0.0 Hz)
                            Undefined:          0 (       0.0 Hz)
                         Sample trace:          0 (       0.0 Hz)

 *** Timestamp elapsed time:      777.848 s
 FEE  elapsed dead time(s) elapsed idle time(s)
  0                1.168                0.000
  1                0.000               62.240
  2                0.000               92.323
  3                0.250                0.000
  4                0.066                0.000
  5                0.000              313.625
  6                0.000               26.861
  7                0.000               69.084
  8                0.955                0.000
  9                0.093                0.000
 10                0.016              457.454
 11                0.447                0.000
 12                0.000                0.000
 13                0.000                0.000
 14                0.000                0.000
 15                0.000                0.000
 16                0.000                0.000
 17                0.000                0.000
 18                0.000                0.000
 19                0.000                0.000
 20                0.000                0.000
 21                0.000                0.000
 22                0.000                0.000
 23                0.000                0.000
 24                0.000                0.000
 25                0.000                0.000
 26                0.000                0.000
 27                0.000                0.000
 28                0.000                0.000
 29                0.000                0.000
 30                0.000                0.000
 31                0.000                0.000
 32                0.000                0.000

 *** Statistics
 FEE  ADC Data Other Data     Sample  Undefined      Pause     Resume    SYNC100    WR48-63       Disc        MBS      Other   HEC Data
  0   36906229      10016          0          0         10         10       4998       4998          0          0          0         41
  1   10150880       2708          0          0          0          0       1354       1354          0          0          0         30
  2    8283625       2156          0          0          0          0       1078       1078          0          0          0          0
  3   37453715      10294          0          0          6          6       5141       5141          0          0          0          0
  4   23358879       6376          0          0          4          4       3184       3184          0          0          0          0
  5    4272841       1160          0          0          0          0        580        580          0          0          0          0
  6   11147856       3068          0          0          0          0       1534       1534          0          0          0          0
  7    9321327       2384          0          0          0          0       1192       1192          0          0          0          0
  8   61453512      16278          0          0         14         14       8125       8125          0          0          0          0
  9   19621565       5040          0          0          6          6       2514       2514          0          0          0          0
 10    3914618        986          0          0          1          1        492        492          0          0          0          0
 11   35964891       9422          0          0          5          5       4706       4706          0          0          0          0
 12         36          0          0          0          0          0          0          0          0          0          0          0
 13          5          0          0          0          0          0          0          0          0          0          0          0
 14        137          0          0          0          0          0          0          0          0          0          0          0
 15          0          0          0          0          0          0          0          0          0          0          0          0
 16          0          0          0          0          0          0          0          0          0          0          0          0
 17          0          0          0          0          0          0          0          0          0          0          0          0
 18          0          0          0          0          0          0          0          0          0          0          0          0
 19          0          0          0          0          0          0          0          0          0          0          0          0
 20          0          0          0          0          0          0          0          0          0          0          0          0
 21          0          0          0          0          0          0          0          0          0          0          0          0
 22          0          0          0          0          0          0          0          0          0          0          0          0
 23          0          0          0          0          0          0          0          0          0          0          0          0
 24          0          0          0          0          0          0          0          0          0          0          0          0
 25          0          0          0          0          0          0          0          0          0          0          0          0
 26          0          0          0          0          0          0          0          0          0          0          0          0
 27          0          0          0          0          0          0          0          0          0          0          0          0
 28          0          0          0          0          0          0          0          0          0          0          0          0
 29          0          0          0          0          0          0          0          0          0          0          0          0
 30          0          0          0          0          0          0          0          0          0          0          0          0
 31          0          0          0          0          0          0          0          0          0          0          0          0
 32          0          0          0          0          0          0          0          0          0          0          0          0

 *** Timewarps
 FEE       ADC      Pause     Resume    SYNC100    WR48-63       Disc        MBS  Undefined    Samples
  0          0          0          0          0          0          0          0          0          0
  1          0          0          0          0          0          0          0          0          0
  2          0          0          0          0          0          0          0          0          0
  3          0          0          0          0          0          0          0          0          0
  4          0          0          0          0          0          0          0          0          0
  5          0          0          0          0          0          0          0          0          0
  6          0          0          0          0          0          0          0          0          0
  7          0          0          0          0          0          0          0          0          0
  8          0          0          0          0          0          0          0          0          0
  9          0          0          0          0          0          0          0          0          0
 10          0          0          0          0          0          0          0          0          0
 11          0          0          0          0          0          0          0          0          0
 12          0          0          0          0          0          0          0          0          0
 13          0          0          0          0          0          0          0          0          0
 14          0          0          0          0          0          0          0          0          0
 15          0          0          0          0          0          0          0          0          0
 16          0          0          0          0          0          0          0          0          0
 17          0          0          0          0          0          0          0          0          0
 18          0          0          0          0          0          0          0          0          0
 19          0          0          0          0          0          0          0          0          0
 20          0          0          0          0          0          0          0          0          0
 21          0          0          0          0          0          0          0          0          0
 22          0          0          0          0          0          0          0          0          0
 23          0          0          0          0          0          0          0          0          0
 24          0          0          0          0          0          0          0          0          0
 25          0          0          0          0          0          0          0          0          0
 26          0          0          0          0          0          0          0          0          0
 27          0          0          0          0          0          0          0          0          0
 28          0          0          0          0          0          0          0          0          0
 29          0          0          0          0          0          0          0          0          0
 30          0          0          0          0          0          0          0          0          0
 31          0          0          0          0          0          0          0          0          0
 32          0          0          0          0          0          0          0          0          0

 *** Program elapsed time:   26.919s ( 1188.754 blocks/s,  74.297 Mb/s)
Attachment 10: Screenshot_from_2025-06-18_04-04-16.png
Screenshot_from_2025-06-18_04-04-16.png
Attachment 11: Screenshot_from_2025-06-18_04-04-35.png
Screenshot_from_2025-06-18_04-04-35.png
Attachment 12: Screenshot_from_2025-06-18_04-04-51.png
Screenshot_from_2025-06-18_04-04-51.png
Attachment 13: Screenshot_from_2025-06-18_04-22-08.png
Screenshot_from_2025-06-18_04-22-08.png
Attachment 14: Screenshot_from_2025-06-18_04-22-29.png
Screenshot_from_2025-06-18_04-22-29.png
Attachment 15: Screenshot_from_2025-06-18_04-22-47.png
Screenshot_from_2025-06-18_04-22-47.png
Attachment 16: Screenshot_from_2025-06-18_08-25-28.png
Screenshot_from_2025-06-18_08-25-28.png
Attachment 17: Screenshot_from_2025-06-18_08-25-46.png
Screenshot_from_2025-06-18_08-25-46.png
Attachment 18: Screenshot_from_2025-06-18_08-49-30.png
Screenshot_from_2025-06-18_08-49-30.png
Attachment 19: Screenshot_from_2025-06-18_08-49-45.png
Screenshot_from_2025-06-18_08-49-45.png
Attachment 20: Screenshot_from_2025-06-18_08-50-00.png
Screenshot_from_2025-06-18_08-50-00.png
Attachment 21: Screenshot_from_2025-06-18_08-50-17.png
Screenshot_from_2025-06-18_08-50-17.png
Attachment 22: Screenshot_from_2025-06-18_12-02-48.png
Screenshot_from_2025-06-18_12-02-48.png
Attachment 23: Screenshot_from_2025-06-18_15-00-16.png
Screenshot_from_2025-06-18_15-00-16.png
  756   Tue Jun 17 19:10:07 2025 NH, PMHRun 96
20:00 - Go to YR cave to see what is going on

20:20 - Beam back

20:28 - Set Tape to Storage, rate ca 2500 kB/s
Temperature and rate OK, System wide checks OK

Target density low (4e12/cm2)

21:00 - target density still 4e12, Ey rate histograms attached (attach 4)

21:20 - Target density increased to 6e12, we see more indication of reactions

22:30 - All checks OK, tape rate still 2500 kB/s
Target density ~7e12/cm2, beam intensity ~1e9pps

23:00 - Target density 7.5e12/cm2, beam intensity: 1e9 pps. Particle detector rate: 8 kHz at injection
 Gas bottle pressure still above first tick 
 Target position X=-1000, Y=1400
 All running smoothly. CARME DAQ running OK. (Figs 5-7)
Attachment 1: Screenshot_from_2025-06-17_20-28-22.png
Screenshot_from_2025-06-17_20-28-22.png
Attachment 2: Screenshot_from_2025-06-17_20-28-43.png
Screenshot_from_2025-06-17_20-28-43.png
Attachment 3: Screenshot_from_2025-06-17_20-58-26.png
Screenshot_from_2025-06-17_20-58-26.png
Attachment 4: Screenshot_from_2025-06-17_21-02-19.png
Screenshot_from_2025-06-17_21-02-19.png
Attachment 5: Screenshot_from_2025-06-17_22-59-58.png
Screenshot_from_2025-06-17_22-59-58.png
Attachment 6: Screenshot_from_2025-06-17_22-59-48.png
Screenshot_from_2025-06-17_22-59-48.png
Attachment 7: Screenshot_from_2025-06-17_22-59-38.png
Screenshot_from_2025-06-17_22-59-38.png
  755   Tue Jun 17 12:22:50 2025 CB EW JG AG SNRun 94/95/96

Run 94

Background with beam at 150 keV and no target.

Note energy is 0.15% off vs. beam and target to have a more realistic orbit.

As prior run but thresholds changed to

DSSD1 - 0xc

DSSD2 - 0x8

DSSD3 - 0x8

DSSD4 - MEC

 

Run 95

DSSD 4 worked briefly. Tripped again a few minutes into this run. Ignore

 

Run 96 - as run 94

run stopped at 14:00

Zoran reset to with-target settings for later

The plotted readout for the G35 voltage divider over several injections for this run is attached. This kind of data can be obtained from Konstantin also after the experiment.

The average is something like 88.4808 V.

Attachment 1: G35Voltage.pdf
G35Voltage.pdf
  754   Tue Jun 17 12:14:30 2025 EW, CB, AGRun 93- Background with beam 150keV

Run 93- New run at 150keV, background with beam

13:15 All checks ok

Attachment 1: Screenshot_from_2025-06-17_13-16-52.png
Screenshot_from_2025-06-17_13-16-52.png
Attachment 2: Screenshot_from_2025-06-17_13-16-49.png
Screenshot_from_2025-06-17_13-16-49.png
Attachment 3: Screenshot_from_2025-06-17_13-16-45.png
Screenshot_from_2025-06-17_13-16-45.png
Attachment 4: Screenshot_from_2025-06-17_13-16-43.png
Screenshot_from_2025-06-17_13-16-43.png
  753   Tue Jun 17 08:04:09 2025 EW, AG, CBRun 92- Background with beam 90keV

Run 92- started at 09:00, background with beam

Same settings as run 91

See attachments for rates etc

All checks okay

09:58 Run stopped as now tuning 150keV beam

10:00 All checks ok

 

10:25 D2 bottle pack pressure 30 bar approximately

12:18 All checks ok

Attachment 1: Screenshot_from_2025-06-17_09-03-43.png
Screenshot_from_2025-06-17_09-03-43.png
Attachment 2: Screenshot_from_2025-06-17_09-03-38.png
Screenshot_from_2025-06-17_09-03-38.png
Attachment 3: Screenshot_from_2025-06-17_09-03-34.png
Screenshot_from_2025-06-17_09-03-34.png
Attachment 4: Screenshot_from_2025-06-17_09-03-29.png
Screenshot_from_2025-06-17_09-03-29.png
  752   Mon Jun 16 22:06:00 2025 CB, AR, EWRUN 90-91

23:00 DAQ restored and data now being written okay. Checks all okay, see screenshots below. Histograms reset. 40 bars in bottle @ 23:45 - should last 30 hours.

12:00 Checks all ok. No problems with aidas or writing to disk.

01:10 Checks all ok.

02:05 Checks all ok. See attatchment 6 for peak update after reset histograms 3 hours prior. Target okay, dropped a little to 6e12 but seems to be going up itself. Will monitor and then move position if needed.

03:10 During checks, noticed detector 4 had tripped so turned it off. Writing to disk increased alot, statistics increased to alost 250000. Put aida13-16 in low/medium energy. Started new run. (see attatchment 7-14 for before and afters)

RUN 91

03:15 Checks all ok after turning off detector 4. Moved target around and got abit more density but x is very sensitive and y did not seem to do much. Density 5.5e12.

04:30 Checks all ok. Target decreasing in density - monitoring it.

04:35 Target density back up to 1e13 after moving in y.

05:30 Checks all ok. Target density decreased slightly to 8e12

06:30 Checks all ok, target density stable for past hour

07:30 Checks all ok

08:30 Checks all ok, target density at 1e13 for past two hours

09:00 Run stopped

 

Attachment 1: Screenshot_from_2025-06-16_23-08-33.png
Screenshot_from_2025-06-16_23-08-33.png
Attachment 2: Screenshot_from_2025-06-16_23-08-17.png
Screenshot_from_2025-06-16_23-08-17.png
Attachment 3: Screenshot_from_2025-06-16_23-07-42.png
Screenshot_from_2025-06-16_23-07-42.png
Attachment 4: Screenshot_from_2025-06-16_23-06-56.png
Screenshot_from_2025-06-16_23-06-56.png
Attachment 5: Screenshot_from_2025-06-16_23-06-36.png
Screenshot_from_2025-06-16_23-06-36.png
Attachment 6: Screenshot_from_2025-06-17_02-06-51.png
Screenshot_from_2025-06-17_02-06-51.png
Attachment 7: Screenshot_from_2025-06-17_03-10-21.png
Screenshot_from_2025-06-17_03-10-21.png
Attachment 8: Screenshot_from_2025-06-17_03-09-43.png
Screenshot_from_2025-06-17_03-09-43.png
Attachment 9: Screenshot_from_2025-06-17_02-57-31.png
Screenshot_from_2025-06-17_02-57-31.png
Attachment 10: Screenshot_from_2025-06-17_02-56-20.png
Screenshot_from_2025-06-17_02-56-20.png
Attachment 11: Screenshot_from_2025-06-17_02-55-51.png
Screenshot_from_2025-06-17_02-55-51.png
Attachment 12: Screenshot_from_2025-06-17_02-55-38.png
Screenshot_from_2025-06-17_02-55-38.png
Attachment 13: Screenshot_from_2025-06-17_02-55-25.png
Screenshot_from_2025-06-17_02-55-25.png
Attachment 14: Screenshot_from_2025-06-17_02-06-51.png
Screenshot_from_2025-06-17_02-06-51.png
  751   Mon Jun 16 21:58:00 2025 CB ARRuns 84-89

Found all Merger Links background black - inactive. Nontheless the tapeservice was working, the data rate to disk was normal, and the file was being written and the histograms were being updated.

Decided to stop DAQ, stop Merger, restart merger. Links rest to green background - but no more writing to disk!

 

Killed Merger, killed Tape Service. Restarted.

Netvar read out at 3 - set to 1. When the Output of the Merger is enabled Netvar returns to 3. Tapeservice not writing to disk

Killed Merger, killed Tape Service. Restarted x3. Same behaviour.

Not working. No blocks being written.

With Tape Server status Stopped Ticked No storage, Unticked No storage -> Now writing blocks ?!

 

Tape Server Go. Writing to disk. Merger OK. Histograms updated, observables as normal (see next ELOG).

Lost runs 84-89 in the process.

  750   Mon Jun 16 11:12:22 2025 JG, AK, AR, EHrun 83

12:15 restarted measurement with target - Run 83

same settings as for runs 76, 77 etc

ROOT histograms NOT zeroed

Beam at 8e8, target density at 1.5e13, scintillator @ 0 deg. peak rate 5000

status screenshots attached

15:15: Checks all ok, peaks are peaky (attatchment 8)

16:45 Checks all ok.

17:45 Checks all ok. IMP Froze and has been restarted. Peaks good. Light turned off iin cave.

18:40: Checks all ok. Target great at 1e13.

22:20 Checks done - all okay except aida green-to-red was not working, but still writing data to disk. Carlo trying to figure it out now. All other checks were okay. Daedalus crashed, restarted and target still at 1e13. 

Attachment 1: Screenshot_from_2025-06-16_12-16-08.png
Screenshot_from_2025-06-16_12-16-08.png
Attachment 2: Screenshot_from_2025-06-16_12-15-34.png
Screenshot_from_2025-06-16_12-15-34.png
Attachment 3: Screenshot_from_2025-06-16_12-15-19.png
Screenshot_from_2025-06-16_12-15-19.png
Attachment 4: Screenshot_from_2025-06-16_12-15-07.png
Screenshot_from_2025-06-16_12-15-07.png
Attachment 5: Screenshot_from_2025-06-16_12-14-53.png
Screenshot_from_2025-06-16_12-14-53.png
Attachment 6: Screenshot_from_2025-06-16_12-14-42.png
Screenshot_from_2025-06-16_12-14-42.png
Attachment 7: Screenshot_from_2025-06-16_14-55-13.png
Screenshot_from_2025-06-16_14-55-13.png
Attachment 8: Screenshot_from_2025-06-16_15-19-14.png
Screenshot_from_2025-06-16_15-19-14.png
  749   Mon Jun 16 10:46:11 2025 CB JM JG AKRun82

Background without beam while source is optimised.

  748   Mon Jun 16 08:17:41 2025 JGD2 consumption

The current bottle of D2 was installed on friday 06.06. at around 12:00. I assume we started at 200 bar.

Right now we have 50 bar remaining in this bottle (16.06. at 8:00).

This is an average consumption of ~15 bar/day

 150 bar / 9.83 days

 

 

  747   Sun Jun 15 10:46:35 2025 EW, EM, AR, AB, CB, JG, AKRun 81

11:45 New run started - run 81, same settings as previous runs (75, 76 etc)

ROOT histograms NOT zeroed

Beam at 2.5e8, target density at 3e12

11:50 All checks okay

12:45 All checks okay

13:45 All checks okay

14:45 All checks okay

15:45 All checks okay.

17:00 All Checks okay. Target temp manually dropped from 30.4K to 30.1K to try and improve density as would not get much above 2.5-3ish. Also played with target position for a while - now seemed to have stabilised it at 4.5e12 as of now.

18:00 All checkd okay. Target been stable at 4.5-5.3e12.

18:20 Done checks again (all ok) as the beam is being a little stange. Attatched ss of peaky peaks (some of them very nice) but detector 3 histogram wont cooperate and wont let me see the counts if I zoom further down, but can see proton peak fine. Seeing periodic spikes in the counts plot for the particle detector - monitoring. Also the tail of the beam seems more red than usual.

19:00 More photos of strange spikes in the counts for particle detector uploaded. Sometimes, the baseline has decent sized spikes too. Attatchments 10-14

19:20 Checks ok. Traget stable. Spikes in particle detector seem to be more frequent, uploading more photos below. Also attatched peaks at this stage (attatchemtn 15).

19:36 Very strange spikes in particle det. Two bigish peaks at baseline and then strange beginning at the injection. Attatchment 19.

19:45 Another spike in the baseline for particle det. Attatchment 20. 

20:25 Checks ok. Spikes still in particle det plot.

21:20 Checks ok. Screenshot of peaks in attatchment 21.

22:20 Checks ok. Target been relativly stable all shift at 5e12. Spikes still periodically turning up in particle detector. Histograms are gaining counts, attatchment 22 is the peaks.

22:30 Target suddenly got really dense up to 1e13. Lifetime of the beam is ok. Trending down slowly. Did checks again and all is good.

 

01:45. Checks OK. Target seems thick.

04:30

Temperatures OK - attach

Rates OK - attach

Merger OK - attach

Hisograms look OK - attached

Target looks very thick from ring vacuum worsening.

7:15 all checks okay - historgams attached

beam 2e8, target 1e13, scintillator rate 2000 in peak

10:00 

Run stopped to gain cave access.

Before beam was stopped, target was moved out. We still see some signal on the neutral particle detector, roughly a factor 10 below with target on at 1e13 at/cm2

Will need a beam background run for normalisation procedure. See attachment 29

10:30

Now we have no beam in the ring we still observe some events in the neutrals detector. Some spikes up to 150 counts observed, could explain some of what we have seen so far. See attachment 30

Note we are saving all this data to MBS so it is effectively background for the neutral detector. 

Attachment 1: Screenshot_from_2025-06-15_11-50-19.png
Screenshot_from_2025-06-15_11-50-19.png
Attachment 2: Screenshot_from_2025-06-15_11-50-10.png
Screenshot_from_2025-06-15_11-50-10.png
Attachment 3: Screenshot_from_2025-06-15_11-50-07.png
Screenshot_from_2025-06-15_11-50-07.png
Attachment 4: Screenshot_from_2025-06-15_11-49-32.png
Screenshot_from_2025-06-15_11-49-32.png
Attachment 5: Screenshot_from_2025-06-15_18-23-17.png
Screenshot_from_2025-06-15_18-23-17.png
Attachment 6: Screenshot_from_2025-06-15_18-23-35.png
Screenshot_from_2025-06-15_18-23-35.png
Attachment 7: IMG_7014.JPG
IMG_7014.JPG
Attachment 8: IMG_7013.JPG
IMG_7013.JPG
Attachment 9: IMG_7012.JPG
IMG_7012.JPG
Attachment 10: IMG_7016.JPG
IMG_7016.JPG
Attachment 11: IMG_7017.JPG
IMG_7017.JPG
Attachment 12: IMG_7018.JPG
IMG_7018.JPG
Attachment 13: IMG_7025.JPG
IMG_7025.JPG
Attachment 14: IMG_7027.JPG
IMG_7027.JPG
Attachment 15: Screenshot_from_2025-06-15_19-17-14.png
Screenshot_from_2025-06-15_19-17-14.png
Attachment 16: IMG_7043.JPG
IMG_7043.JPG
Attachment 17: IMG_7036.JPG
IMG_7036.JPG
Attachment 18: IMG_7039.JPG
IMG_7039.JPG
Attachment 19: IMG_7048.JPG
IMG_7048.JPG
Attachment 20: IMG_7050.JPG
IMG_7050.JPG
Attachment 21: Screenshot_from_2025-06-15_21-23-20.png
Screenshot_from_2025-06-15_21-23-20.png
Attachment 22: Screenshot_from_2025-06-15_22-21-21.png
Screenshot_from_2025-06-15_22-21-21.png
Attachment 23: Screenshot_from_2025-06-16_04-29-10.png
Screenshot_from_2025-06-16_04-29-10.png
Attachment 24: Screenshot_from_2025-06-16_04-29-00.png
Screenshot_from_2025-06-16_04-29-00.png
Attachment 25: Screenshot_from_2025-06-16_04-28-51.png
Screenshot_from_2025-06-16_04-28-51.png
Attachment 26: Screenshot_from_2025-06-16_04-28-40.png
Screenshot_from_2025-06-16_04-28-40.png
Attachment 27: Screenshot_from_2025-06-16_04-28-26.png
Screenshot_from_2025-06-16_04-28-26.png
Attachment 28: Screenshot_from_2025-06-16_07-20-26.png
Screenshot_from_2025-06-16_07-20-26.png
Attachment 29: 20250616_100644.jpg
20250616_100644.jpg
Attachment 30: 20250616_104229.jpg
20250616_104229.jpg
  746   Sun Jun 15 07:01:37 2025 EW, EMRun 80

08:00 New run started - run 80

Run whilst they are optimising the beam after losing it, everything else same as previous runs

ROOT histograms NOT zeroed

All checks fine

09:05 All checks okay

Attachment 5 current histograms zoomed into area of interest, detector 3 only proton peak visible

10:00 All checks okay

11:00 All checks okay

11:45 Run stopped as beam is back

Attachment 1: Screenshot_from_2025-06-15_08-09-05.png
Screenshot_from_2025-06-15_08-09-05.png
Attachment 2: Screenshot_from_2025-06-15_08-09-00.png
Screenshot_from_2025-06-15_08-09-00.png
Attachment 3: Screenshot_from_2025-06-15_08-08-05.png
Screenshot_from_2025-06-15_08-08-05.png
Attachment 4: Screenshot_from_2025-06-15_08-07-56.png
Screenshot_from_2025-06-15_08-07-56.png
Attachment 5: Screenshot_from_2025-06-15_09-55-36.png
Screenshot_from_2025-06-15_09-55-36.png
  745   Fri Jun 13 19:41:32 2025 JG, SHB, JM, AR, EW, OFRun 77, 78, 79

New run started - Run 77 after refilling LN2 dewar

Same settings as Run 75 & 76

The root histos have not been reset, i.e. they show combined statistics of run 75, 76 and 77.

20:45 - all checks OK

22:00 - all checks OK

00:10

Observed no data writing to disk. Checked merger no data merging. No ADC data items. Seems daq crashed again.

Appears aida09 had crashed causing system to stop. Tried fix suggested in earlier elog by Tom. No success. All FEEs show error and system became unresponsive. 

Power cycling FEEs. 

00:30

New run started Run 78. Same conditions as previous runs. 

All thresholds remain unchanged ->0x28 (400 keV)

Had to reset monitoring histograms as they were unresponsive too. 

All system checks OK. 

Data rate is ~300 kB/sec

ADC data items rate is similar to previous run (screenshot already attached)

Detector leakage currents look OK. Plots attached. 

WS6 (elog and internet) appeared to freeze before. This could be similar to the issue osserved last weekend. All seems fine now and DAQ continues. Should keep an eye on it.

02:00

Checls OK

05:00

Checks OK

7:10 Checks OK. Target was fluctuating over night, dropped around midnight then back up but had no target for about 2-3 hours overnight. Fixed target thickness this morning and now have 5e12.

8:15 Checks Ok.

9:15 Checks Ok.

10:15 Checks ok. Target good at 5e12. Attatched screenshots of the peaky peaks so far (can't put numbers on excel as lost the ability to see it - see 9 10 and 11)

11:35 All checks ok. Screenshot of peaks attatchment 12 (decently peaky).

12:30 All checks ok. Peaks are beginning to show, especially proton and triton but on det 3, cant distinguish trition from noise peak so no value in spreadhseet for that. Target stable at 4.8e12.

13:30 All Checks ok.

14:30 All Checks ok.

15:40 All checks okay, lost beam at 15:25, let HKR know and they are looking into it, have put to no storage.

16:35 All checks okay, still no beam, taget is steady at 3e12

17:35 All checks okay, beam is back but being optimised

18:33 New run started- run 79, same settings as previous runs, did NOT reset histograms, beam now 2e8 (previously was 7e8)

18:33 All checks okay apart from temp of aida 3 which is a little high (see attachment 13), rates also attached (attachemnt 14), current histogram attached (15)

19:35 All checks okay, aida 3 temp now back within range, target density dropped but have managed to get it back up to 2e12

21:30 All checks okay, target has been very unstable, beam now at 3.8e8

22:30 All checks fine, target at 3e12 for past 45 minutes, beam at 4e8, current histogram attachment 16

23:40 All checks ok. Integral value for range to 4000 seems to be increasing for all histograms - meaning target is probably ok for now. Vacuum plot doesnt show any issues.

02:00 All checks ok. Integral values for proton peak have increased so assuming target is still holding as of now (only peak I can clearly isolate to get value from) - no indication that it is off via vacuum plot. 

5:00 All Checks ok. Integral values increasing. 

7:00 All checks ok, can't see target density as of 6:20

08:00 Run stopped as lost beam, histogram attachment 17

 

Attachment 1: .png
.png
Attachment 2: Screenshot_from_2025-06-13_20-45-21.png
Screenshot_from_2025-06-13_20-45-21.png
Attachment 3: Screenshot_from_2025-06-13_20-44-58.png
Screenshot_from_2025-06-13_20-44-58.png
Attachment 4: Screenshot_from_2025-06-13_20-44-31.png
Screenshot_from_2025-06-13_20-44-31.png
Attachment 5: Screenshot_from_2025-06-13_20-43-48.png
Screenshot_from_2025-06-13_20-43-48.png
Attachment 6: Screenshot_from_2025-06-13_20-43-28.png
Screenshot_from_2025-06-13_20-43-28.png
Attachment 7: Screenshot_from_2025-06-13_20-43-10.png
Screenshot_from_2025-06-13_20-43-10.png
Attachment 8: Screenshot_from_2025-06-14_00-39-34.png
Screenshot_from_2025-06-14_00-39-34.png
Attachment 9: Screenshot_from_2025-06-14_10-11-30.png
Screenshot_from_2025-06-14_10-11-30.png
Attachment 10: Screenshot_from_2025-06-14_10-13-13.png
Screenshot_from_2025-06-14_10-13-13.png
Attachment 11: Screenshot_from_2025-06-14_10-16-10.png
Screenshot_from_2025-06-14_10-16-10.png
Attachment 12: Screenshot_from_2025-06-14_11-35-13.png
Screenshot_from_2025-06-14_11-35-13.png
Attachment 13: Screenshot_from_2025-06-14_18-35-11.png
Screenshot_from_2025-06-14_18-35-11.png
Attachment 14: Screenshot_from_2025-06-14_18-38-46.png
Screenshot_from_2025-06-14_18-38-46.png
Attachment 15: Screenshot_from_2025-06-14_18-36-26.png
Screenshot_from_2025-06-14_18-36-26.png
Attachment 16: Screenshot_from_2025-06-14_22-31-01.png
Screenshot_from_2025-06-14_22-31-01.png
Attachment 17: Screenshot_from_2025-06-15_08-05-50.png
Screenshot_from_2025-06-15_08-05-50.png
  744   Fri Jun 13 16:39:18 2025 JG, SHBRun 76

Same settings as run 75.

Detector 4 is biased at 100V and switched from MEC to LEC.

The root histos have not been reset, i.e. they show combined statistics of run 75 and 76.

 

Attachment 1: Screenshot_from_2025-06-13_17-44-59.png
Screenshot_from_2025-06-13_17-44-59.png
Attachment 2: Screenshot_from_2025-06-13_17-44-42.png
Screenshot_from_2025-06-13_17-44-42.png
Attachment 3: Screenshot_from_2025-06-13_17-43-52.png
Screenshot_from_2025-06-13_17-43-52.png
Attachment 4: Screenshot_from_2025-06-13_17-43-37.png
Screenshot_from_2025-06-13_17-43-37.png
Attachment 5: Screenshot_from_2025-06-13_17-43-18.png
Screenshot_from_2025-06-13_17-43-18.png
Attachment 6: Screenshot_from_2025-06-13_17-42-59.png
Screenshot_from_2025-06-13_17-42-59.png
Attachment 7: Screenshot_from_2025-06-13_16-01-44.png
Screenshot_from_2025-06-13_16-01-44.png
  743   Fri Jun 13 14:45:58 2025 JG, SS, JM, CB, EW, ARRun 75

 

90 keV/u beam on target, Target overlap confirmed with lifetime measurement using bunched beam - @ -12.5mm

We do not expect to see see Rutherford due to energy loss in the dead layer, using other particle detector for normalsiation.

Moved detectors in with servos only. Both arms are IN and stationary.

Left potentiometer = 36.54 mm

Right potentiometer = 37.96 mm

Horizontal beam position = -12.5 mm

Thresholds increased for all to 0x28 (400keV)

data rate is ~0.7 MB/sec

Electron cooler read values; Cooler voltage = 50V (dodgy accuracy need to ask about fc20 on monday), cooler current=0.26 mA.

15:50 All Checks ok. Overlap on beam found. Target at 4e12. Bottle at 90 bars.

 

Attachment 1: Screenshot_from_2025-06-13_16-01-44.png
Screenshot_from_2025-06-13_16-01-44.png
Attachment 2: 2025-06-13_15-59-53_YRT1_CRYRING_2025_D_Bruno_90keV.C1.txt
# Version: 1.3
# Comment: ParamModi Export at 2025-06-13T16:00:01.262304648+02:00[Europe/Berlin]
# Context: YRT1_CRYRING_2025_D_Bruno.C1
# Columns: PARAMETERNAME,VALUE,UNIT,BEAMPROCESSPURPOSE
LOGICAL.SOURCE_HV/U,20000.0,V,TRANSFER_INJECTION
YRT1_TO_YRT1LQ1_BEAM/BEAM_MODE_S,NO_BEAM,,TRANSFER_INJECTION
YRT1_TO_YRT1LQ1_BEAM/PARTICLE_SYMBOL_S,2H,,TRANSFER_INJECTION
YRT1_TO_YRT1LQ1_BEAM/Q_S,1,,TRANSFER_INJECTION
YRT1_TO_YRT1LQ1_TIMEPARAM/TIMING_BI_DELAY_S,2.0E-4,s,TRANSFER_INJECTION
YRT1_TO_YRT1LQ1_TIMEPARAM/TIMING_BI_WINDOW_S,0.002,s,TRANSFER_INJECTION
YRT1_TO_YRT1LQ1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,BEAMOUT_INIT
YRT1_TO_YRT1LQ1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,TRANSFER_INJECTION
YRT1_TO_YRT1LQ1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,BEAMOUT_RESET
YRT1_TO_YRT1LQ1_TIMEPARAM/TIMING_SOURCE_PULSE_LENGTH_S,0.0024,s,TRANSFER_INJECTION
YRT1_TO_YRT1LQ1_TIMEPARAM/T_BP_LENGTH_S,0.12,s,BEAMOUT_INIT
YRT1_TO_YRT1LQ1_TIMEPARAM/T_BP_LENGTH_S,0.11,s,TRANSFER_INJECTION
YRT1_TO_YRT1LQ1_TIMEPARAM/T_BP_LENGTH_S,0.0,s,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/T_BP_LENGTH_S,0.0,s,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/T_BP_LENGTH_S,0.0,s,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/T_BP_LENGTH_S,0.0,s,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/T_BP_LENGTH_S,0.0,s,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/T_BP_LENGTH_S,0.0,s,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/T_BP_LENGTH_S,0.0,s,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/T_BP_LENGTH_S,0.0,s,STORAGE_MANIP_PERFORM
YRT1_TO_YRT1LQ1_TIMEPARAM/T_BP_LENGTH_S,---,s,STORAGE_MANIP_WAIT
YRT1_TO_YRT1LQ1_TIMEPARAM/T_BP_LENGTH_S,---,s,STORAGE_MANIP_RESTART_FGS
YRT1_TO_YRT1LQ1_TIMEPARAM/T_BP_LENGTH_S,---,s,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/T_BP_LENGTH_S,---,s,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/T_BP_LENGTH_S,0.12,s,BEAMOUT_RESET
YRT1LQ1_TO_YRT1LC1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,BEAMOUT_INIT
YRT1LQ1_TO_YRT1LC1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,TRANSFER_INJECTION
YRT1LQ1_TO_YRT1LC1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LQ1_TO_YRT1LC1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LQ1_TO_YRT1LC1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LQ1_TO_YRT1LC1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LQ1_TO_YRT1LC1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LQ1_TO_YRT1LC1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LQ1_TO_YRT1LC1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LQ1_TO_YRT1LC1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LQ1_TO_YRT1LC1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LQ1_TO_YRT1LC1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,BEAMOUT_RESET
LOGICAL.YRT1KH1/HKICK,0.001,rad,TRANSFER_INJECTION
LOGICAL.YRT1KV1/VKICK,0.005,rad,TRANSFER_INJECTION
LOGICAL.YRT1LC1_V/ACTIVE_S,true,,TRANSFER_INJECTION
LOGICAL.YRT1LC1_V/THIGH_S,2.0E-4,s,TRANSFER_INJECTION
LOGICAL.YRT1LC1_V/TOFFSET_S,5.0E-4,s,TRANSFER_INJECTION
LOGICAL.YRT1LC1_V/U,2000.0,V,TRANSFER_INJECTION
LOGICAL.YRT1LD21/ASYMMETRY_KNOB_S,0.0,,TRANSFER_INJECTION
LOGICAL.YRT1LD21/HOR_STEER_KNOB_S,0.0,,TRANSFER_INJECTION
LOGICAL.YRT1LD21/U_QUAD,-350.0,V,TRANSFER_INJECTION
LOGICAL.YRT1LD21/VER_STEER_KNOB_S,0.05,,TRANSFER_INJECTION
LOGICAL.YRT1LD22/ASYMMETRY_KNOB_S,0.0,,TRANSFER_INJECTION
LOGICAL.YRT1LD22/U_QUAD,350.0,V,TRANSFER_INJECTION
LOGICAL.YRT1LT31/ASYMMETRY_KNOB_S,0.0,,TRANSFER_INJECTION
LOGICAL.YRT1LT31/U_QUAD,680.0,V,TRANSFER_INJECTION
LOGICAL.YRT1LT32/ASYMMETRY_KNOB_S,0.0,,TRANSFER_INJECTION
LOGICAL.YRT1LT32/U_QUAD,-680.0,V,TRANSFER_INJECTION
LOGICAL.YRT1LT33/ASYMMETRY_KNOB_S,0.0,,TRANSFER_INJECTION
LOGICAL.YRT1LT33/U_QUAD,680.0,V,TRANSFER_INJECTION
LOGICAL.YRT1LT41/ASYMMETRY_KNOB_S,0.0,,TRANSFER_INJECTION
LOGICAL.YRT1LT41/U_QUAD,-1520.0,V,TRANSFER_INJECTION
LOGICAL.YRT1LT42/ASYMMETRY_KNOB_S,0.0,,TRANSFER_INJECTION
LOGICAL.YRT1LT42/U_QUAD,2080.0,V,TRANSFER_INJECTION
LOGICAL.YRT1LT43/ASYMMETRY_KNOB_S,0.0,,TRANSFER_INJECTION
LOGICAL.YRT1LT43/U_QUAD,-1550.0,V,TRANSFER_INJECTION
LOGICAL.YRT1MH1/HKICK,0.006,rad,TRANSFER_INJECTION
LOGICAL.YRT1MH1/K0L_THEO,-1.57079,rad,TRANSFER_INJECTION
YRT1LC1_TO_YRT1MH2_TIMEPARAM/TIMING_BI_DELAY_S,1.0E-6,s,TRANSFER_INJECTION
YRT1LC1_TO_YRT1MH2_TIMEPARAM/TIMING_BI_WINDOW_S,1.0E-6,s,TRANSFER_INJECTION
YRT1LC1_TO_YRT1MH2_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,BEAMOUT_INIT
YRT1LC1_TO_YRT1MH2_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,TRANSFER_INJECTION
YRT1LC1_TO_YRT1MH2_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LC1_TO_YRT1MH2_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LC1_TO_YRT1MH2_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LC1_TO_YRT1MH2_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LC1_TO_YRT1MH2_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LC1_TO_YRT1MH2_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LC1_TO_YRT1MH2_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LC1_TO_YRT1MH2_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LC1_TO_YRT1MH2_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LC1_TO_YRT1MH2_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,BEAMOUT_RESET
YRT1LC1_TO_YRT1MH2_TIMEPARAM/TIMING_RF_LEAD_S,4.0E-4,s,TRANSFER_INJECTION
LOGICAL.YRT1BR1/ACT_ON_BEAM_S,true,,TRANSFER_INJECTION
LOGICAL.YRT1BR1/THIGH_S,0.0025,s,TRANSFER_INJECTION
LOGICAL.YRT1BR1/URF0_S,0.14,V,TRANSFER_INJECTION
LOGICAL.YRT1BR1/U_PAUSE_PULSE_AMPL_S,4.76,V,TRANSFER_INJECTION
LOGICAL.YRT1KH2/HKICK,-2.0E-4,rad,TRANSFER_INJECTION
LOGICAL.YRT1KV2/VKICK,0.0,rad,TRANSFER_INJECTION
LOGICAL.YRT1LD51/ASYMMETRY_KNOB_S,0.0,,TRANSFER_INJECTION
LOGICAL.YRT1LD51/U_QUAD,12000.0,V,TRANSFER_INJECTION
LOGICAL.YRT1LD52/ASYMMETRY_KNOB_S,0.0,,TRANSFER_INJECTION
LOGICAL.YRT1LD52/U_QUAD,-6500.0,V,TRANSFER_INJECTION
LOGICAL.YRT1QD61/K1L,1.7,1/m,TRANSFER_INJECTION
LOGICAL.YRT1QD62/K1L,-1.7,1/m,TRANSFER_INJECTION
LOGICAL.YRT1QD71/K1L,1.5,1/m,TRANSFER_INJECTION
LOGICAL.YRT1QD72/K1L,-1.5,1/m,TRANSFER_INJECTION
YRT1BR1_TO_YRT1MH2_BEAM/E_S,298000.0,eV/u,TRANSFER_INJECTION
YRT1BR1_TO_YRT1MH2_RF/PHASE_OFFSET_S,0.0,deg,TRANSFER_INJECTION
LOGICAL.GHTYKH4/HKICK,0.0,rad,TRANSFER_INJECTION
LOGICAL.GHTYKV3/VKICK,0.0015,rad,TRANSFER_INJECTION
LOGICAL.GHTYKV4/VKICK,-0.002,rad,TRANSFER_INJECTION
LOGICAL.GHTYQD41/K1L,1.4,1/m,TRANSFER_INJECTION
LOGICAL.GHTYQD42/K1L,-1.2,1/m,TRANSFER_INJECTION
LOGICAL.YR01LP1I/ESEP_HKICK_S,0.02,rad,TRANSFER_INJECTION
LOGICAL.YR01LP1I/ESEP_K0L_THEO_S,-0.05,rad,TRANSFER_INJECTION
LOGICAL.YR01LP2I/ESEP_HKICK_S,-0.03,rad,TRANSFER_INJECTION
LOGICAL.YR01LP2I/ESEP_K0L_THEO_S,0.055,rad,TRANSFER_INJECTION
LOGICAL.YR01MP1I/HKICK,0.0026,rad,TRANSFER_INJECTION
LOGICAL.YR01MP1I/K0L_THEO,0.29322,rad,TRANSFER_INJECTION
LOGICAL.YRT1MH2/HKICK_ESR_S,-0.61087,rad,TRANSFER_INJECTION
LOGICAL.YRT1MH2/HKICK_LOCAL_S,-0.021,rad,TRANSFER_INJECTION
LOGICAL.YRT1MH2/K0L_THEO,-0.61087,rad,TRANSFER_INJECTION
YRT1MH2_TO_CRYRING_TIMEPARAM/TIMING_BI_DELAY_S,1.0E-6,s,TRANSFER_INJECTION
YRT1MH2_TO_CRYRING_TIMEPARAM/TIMING_BI_WINDOW_S,1.0E-6,s,TRANSFER_INJECTION
YRT1MH2_TO_CRYRING_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,BEAMOUT_INIT
YRT1MH2_TO_CRYRING_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,TRANSFER_INJECTION
YRT1MH2_TO_CRYRING_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1MH2_TO_CRYRING_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1MH2_TO_CRYRING_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1MH2_TO_CRYRING_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1MH2_TO_CRYRING_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1MH2_TO_CRYRING_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1MH2_TO_CRYRING_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1MH2_TO_CRYRING_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1MH2_TO_CRYRING_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1MH2_TO_CRYRING_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,BEAMOUT_RESET
YRT1MH2_TO_CRYRING_TIMEPARAM/TIMING_SEPTUM_TRIGGER_OFFSET_S,0.0,s,TRANSFER_INJECTION
CRYRINGB2B/B2B_SID_EXT_S,---,,RING_INJECTION
CRYRINGB2B/B2B_SOURCE_KICKER_DELAY_S,---,s,RING_INJECTION
CRYRINGB2B/B2B_SOURCE_TRIGGER_OFFSET_S,---,deg,RING_INJECTION
CRYRINGB2B/B2B_TRIGGER_OFFSET_S,0.0,deg,RING_INJECTION
CRYRINGB2B/B2B_TRIGGER_OFFSET_TIME_S,0.0,,RING_INJECTION
CRYRINGBEAM/BEAM_MODE_S,NO_BEAM,,RING_INJECTION
CRYRINGBEAM/BRHO_CHIMNEY_S,---,Tm,BEAMOUT_INIT
CRYRINGBEAM/BRHO_CHIMNEY_S,---,Tm,BEAMOUT_RESET
CRYRINGBEAM/BRHO_DISPLAY_S,0.1583,Tm,RING_INJECTION
CRYRINGBEAM/BRHO_DISPLAY_S,0.087,Tm,RING_RF_MANIPULATION
CRYRINGBEAM/BRHO_DISPLAY_S,0.087,Tm,RING_TARGET
CRYRINGBEAM/BRHO_START_END_S,[0.15826692, 0.15826692],Tm,RING_INJECTION
CRYRINGBEAM/BUCKETFILL_RAW_S,[[0.0, 0.0]],,RING_INJECTION
CRYRINGBEAM/BUCKETFILL_RAW_S,[[0.0, 1.2]],,RING_BUNCHING
CRYRINGBEAM/BUCKETFILL_RAW_S,[[1.2, 1.2]],,RING_RAMP
CRYRINGBEAM/BUCKETFILL_RAW_S,[[1.2, 1.2]],,RING_RF_MANIPULATION
CRYRINGBEAM/BUCKETFILL_RAW_S,[[1.2, 1.2]],,RING_RAMP
CRYRINGBEAM/BUCKETFILL_RAW_S,[[1.2, 0.0]],,RING_REBUNCHING
CRYRINGBEAM/BUCKETFILL_RAW_S,[[0.0, 0.0]],,RING_TARGET
CRYRINGBEAM/BUCKETFILL_RAW_S,[[0.0, 0.0]],,RING_INTERMEDIATE
CRYRINGBEAM/BUCKETFILL_SIMPLE_S,1.2,,RING_INJECTION
CRYRINGBEAM/CHIMNEY_B,---,,BEAMOUT_INIT
CRYRINGBEAM/CHIMNEY_B,---,,BEAMOUT_RESET
CRYRINGBEAM/CH_END_S,5.0655,,RING_INJECTION
CRYRINGBEAM/CH_END_S,5.0655,,RING_RAMP
CRYRINGBEAM/CH_END_S,5.0655,,RING_RAMP
CRYRINGBEAM/CH_END_S,---,,STORAGE_MANIP_WAIT
CRYRINGBEAM/COASTING_BEAM_B,true,,RING_REBUNCHING
CRYRINGBEAM/CV_END_S,-5.3921,,RING_INJECTION
CRYRINGBEAM/CV_END_S,-5.3921,,RING_RAMP
CRYRINGBEAM/CV_END_S,-5.3921,,RING_RAMP
CRYRINGBEAM/CV_END_S,---,,STORAGE_MANIP_WAIT
CRYRINGBEAM/DELTAR,0.0,m,RING_INJECTION
CRYRINGBEAM/DELTAR,0.0,m,RING_RF_MANIPULATION
CRYRINGBEAM/DELTAR,0.0,m,STORAGE_MANIP_WAIT
CRYRINGBEAM/DELTAR,0.0,m,RING_TARGET
CRYRINGBEAM/DELTAR,0.0,m,RING_INTERMEDIATE
CRYRINGBEAM/DPBL,0.0,,RING_INJECTION
CRYRINGBEAM/DPBL,0.0,,STORAGE_MANIP_WAIT
CRYRINGBEAM/DPFREV,0.003,,RING_INJECTION
CRYRINGBEAM/DPFREV,0.014,,RING_RF_MANIPULATION
CRYRINGBEAM/DPFREV,0.014,,STORAGE_MANIP_WAIT
CRYRINGBEAM/DPFREV,0.014,,RING_TARGET
CRYRINGBEAM/DPFREV,0.014,,RING_INTERMEDIATE
CRYRINGBEAM/DP_OVER_P_S,0.0013,,RING_INJECTION
CRYRINGBEAM/E_S,298000.0,eV/u,RING_INJECTION
CRYRINGBEAM/E_S,90000.0,eV/u,RING_RAMP
CRYRINGBEAM/E_S,90000.0,eV/u,RING_RAMP
CRYRINGBEAM/FINAL_HARMONIC_S,2,,RING_RF_MANIPULATION
CRYRINGBEAM/FREV,140270.3399,Hz,RING_INJECTION
CRYRINGBEAM/FREV,77786.0577,Hz,RING_RF_MANIPULATION
CRYRINGBEAM/FREV,77786.0577,Hz,RING_TARGET
CRYRINGBEAM/H_INPUT_S,2,,RING_INJECTION
CRYRINGBEAM/INJ_EMIH_S,1.0E-5,m*rad,RING_INJECTION
CRYRINGBEAM/INJ_EMIL_S,0.006189628151069671,eVs/u,RING_INJECTION
CRYRINGBEAM/INJ_EMIV_S,1.0E-5,m*rad,RING_INJECTION
CRYRINGBEAM/KL_AMPL_S,0.0,,RING_INJECTION
CRYRINGBEAM/KL_AMPL_S,0.0,,RING_BUNCHING
CRYRINGBEAM/KL_AMPL_S,0.0,,RING_POST_INJECTION
CRYRINGBEAM/KL_AMPL_S,0.0,,RING_MANI_PRE_RAMP
CRYRINGBEAM/KL_AMPL_S,0.0,,RING_RAMP
CRYRINGBEAM/KL_AMPL_S,0.0,,RING_RF_MANIPULATION
CRYRINGBEAM/KL_AMPL_S,0.0,,RING_RAMP
CRYRINGBEAM/KL_AMPL_S,0.0,,RING_REBUNCHING
CRYRINGBEAM/KL_AMPL_S,0.0,,RING_TARGET
CRYRINGBEAM/KL_AMPL_S,0.0,,RING_INTERMEDIATE
CRYRINGBEAM/KL_HARM_S,1,,RING_INJECTION
CRYRINGBEAM/KL_HARM_S,1,,RING_BUNCHING
CRYRINGBEAM/KL_HARM_S,1,,RING_POST_INJECTION
CRYRINGBEAM/KL_HARM_S,1,,RING_MANI_PRE_RAMP
CRYRINGBEAM/KL_HARM_S,1,,RING_RAMP
CRYRINGBEAM/KL_HARM_S,1,,RING_RF_MANIPULATION
CRYRINGBEAM/KL_HARM_S,1,,RING_RAMP
CRYRINGBEAM/KL_HARM_S,1,,RING_REBUNCHING
CRYRINGBEAM/KL_HARM_S,1,,RING_TARGET
CRYRINGBEAM/KL_HARM_S,1,,RING_INTERMEDIATE
CRYRINGBEAM/KL_OFFSET_S,0.0,,RING_INJECTION
CRYRINGBEAM/KL_OFFSET_S,0.0,,RING_BUNCHING
CRYRINGBEAM/KL_OFFSET_S,0.0,,RING_POST_INJECTION
CRYRINGBEAM/KL_OFFSET_S,0.0,,RING_MANI_PRE_RAMP
CRYRINGBEAM/KL_OFFSET_S,0.0,,RING_RAMP
CRYRINGBEAM/KL_OFFSET_S,0.0,,RING_RF_MANIPULATION
CRYRINGBEAM/KL_OFFSET_S,0.0,,RING_RAMP
CRYRINGBEAM/KL_OFFSET_S,0.0,,RING_REBUNCHING
CRYRINGBEAM/KL_OFFSET_S,0.0,,RING_TARGET
CRYRINGBEAM/KL_OFFSET_S,0.0,,RING_INTERMEDIATE
CRYRINGBEAM/KL_PHASE_S,0.0,deg,RING_INJECTION
CRYRINGBEAM/KL_PHASE_S,0.0,deg,RING_BUNCHING
CRYRINGBEAM/KL_PHASE_S,0.0,deg,RING_POST_INJECTION
CRYRINGBEAM/KL_PHASE_S,0.0,deg,RING_MANI_PRE_RAMP
CRYRINGBEAM/KL_PHASE_S,0.0,deg,RING_RAMP
CRYRINGBEAM/KL_PHASE_S,0.0,deg,RING_RF_MANIPULATION
CRYRINGBEAM/KL_PHASE_S,0.0,deg,RING_RAMP
CRYRINGBEAM/KL_PHASE_S,0.0,deg,RING_REBUNCHING
CRYRINGBEAM/KL_PHASE_S,0.0,deg,RING_TARGET
CRYRINGBEAM/KL_PHASE_S,0.0,deg,RING_INTERMEDIATE
CRYRINGBEAM/NPARTICLES_S,1000.0,,RING_INJECTION
CRYRINGBEAM/PARTICLE_SYMBOL_INPUT_S,2H,,RING_INJECTION
CRYRINGBEAM/QH_END_S,2.35,,RING_INJECTION
CRYRINGBEAM/QH_END_S,---,,RING_POST_INJECTION
CRYRINGBEAM/QH_END_S,2.33,,RING_RAMP
CRYRINGBEAM/QH_END_S,2.33,,RING_RAMP
CRYRINGBEAM/QH_END_S,---,,STORAGE_MANIP_WAIT
CRYRINGBEAM/QV_END_S,2.57,,RING_INJECTION
CRYRINGBEAM/QV_END_S,---,,RING_POST_INJECTION
CRYRINGBEAM/QV_END_S,2.53,,RING_RAMP
CRYRINGBEAM/QV_END_S,2.53,,RING_RAMP
CRYRINGBEAM/QV_END_S,---,,STORAGE_MANIP_WAIT
CRYRINGBEAM/Q_INPUT_S,1,,RING_INJECTION
CRYRINGBEAM/TREV,7.12909E-6,s,RING_INJECTION
CRYRINGBEAM/USE_SIMPLE_BF_B,false,,RING_INJECTION
CRYRINGBUMPER/TFALL_S,1.76E-4,s,RING_INJECTION
CRYRINGIKICKER/KICKER_TOFFSET1_S,0.0,s,RING_INJECTION
CRYRINGOPTICS/OPTIC_NAME_RING_S,CRYRING_STANDARD_OPTICS_TABLE,,RING_INJECTION
CRYRINGOPTICS/TAU_END_S,0.0,,RING_INJECTION
CRYRINGOPTICS/TAU_END_S,---,,RING_POST_INJECTION
CRYRINGOPTICS/TAU_END_S,0.0,,RING_RAMP
CRYRINGOPTICS/TAU_END_S,0.0,,RING_RAMP
CRYRINGORBIT/COH02,0.0,m,RING_INJECTION
CRYRINGORBIT/COH02,0.0,m,RING_RF_MANIPULATION
CRYRINGORBIT/COH02,0.0,m,STORAGE_MANIP_WAIT
CRYRINGORBIT/COH02,0.0,m,RING_TARGET
CRYRINGORBIT/COH03,0.0,m,RING_INJECTION
CRYRINGORBIT/COH03,0.0,m,RING_RF_MANIPULATION
CRYRINGORBIT/COH03,0.0,m,STORAGE_MANIP_WAIT
CRYRINGORBIT/COH03,0.0,m,RING_TARGET
CRYRINGORBIT/COH04,0.0,m,RING_INJECTION
CRYRINGORBIT/COH04,0.0,m,RING_RF_MANIPULATION
CRYRINGORBIT/COH04,0.0,m,STORAGE_MANIP_WAIT
CRYRINGORBIT/COH04,0.0,m,RING_TARGET
CRYRINGORBIT/COH06,0.0,m,RING_INJECTION
CRYRINGORBIT/COH06,0.0,m,RING_RF_MANIPULATION
CRYRINGORBIT/COH06,0.0,m,STORAGE_MANIP_WAIT
CRYRINGORBIT/COH06,0.0,m,RING_TARGET
CRYRINGORBIT/COH07,0.0,m,RING_INJECTION
CRYRINGORBIT/COH07,0.0,m,RING_RF_MANIPULATION
CRYRINGORBIT/COH07,0.0,m,STORAGE_MANIP_WAIT
CRYRINGORBIT/COH07,0.0,m,RING_TARGET
CRYRINGORBIT/COH08,0.0,m,RING_INJECTION
CRYRINGORBIT/COH08,0.0,m,RING_RF_MANIPULATION
CRYRINGORBIT/COH08,0.0,m,STORAGE_MANIP_WAIT
CRYRINGORBIT/COH08,0.0,m,RING_TARGET
CRYRINGORBIT/COH10,0.0,m,RING_INJECTION
CRYRINGORBIT/COH10,0.0,m,RING_RF_MANIPULATION
CRYRINGORBIT/COH10,0.0,m,STORAGE_MANIP_WAIT
CRYRINGORBIT/COH10,0.0,m,RING_TARGET
CRYRINGORBIT/COH11,0.0,m,RING_INJECTION
CRYRINGORBIT/COH11,0.0,m,RING_RF_MANIPULATION
CRYRINGORBIT/COH11,0.0,m,STORAGE_MANIP_WAIT
CRYRINGORBIT/COH11,0.0,m,RING_TARGET
CRYRINGORBIT/COH12,0.0,m,RING_INJECTION
CRYRINGORBIT/COH12,0.0,m,RING_RF_MANIPULATION
CRYRINGORBIT/COH12,0.0,m,STORAGE_MANIP_WAIT
CRYRINGORBIT/COH12,0.0,m,RING_TARGET
CRYRINGORBIT/COV02,0.0,m,RING_INJECTION
CRYRINGORBIT/COV02,0.0,m,RING_RF_MANIPULATION
CRYRINGORBIT/COV02,0.0,m,STORAGE_MANIP_WAIT
CRYRINGORBIT/COV02,0.0,m,RING_TARGET
CRYRINGORBIT/COV03,0.0,m,RING_INJECTION
CRYRINGORBIT/COV03,0.0,m,RING_RF_MANIPULATION
CRYRINGORBIT/COV03,0.0,m,STORAGE_MANIP_WAIT
CRYRINGORBIT/COV03,0.0,m,RING_TARGET
CRYRINGORBIT/COV04,0.0,m,RING_INJECTION
CRYRINGORBIT/COV04,0.0,m,RING_RF_MANIPULATION
... 679 more lines ...
  742   Thu Jun 12 15:33:17 2025 CB, JM, EM, JG, TD, AB, CF, ARRun 74

16:34

Like Run 73, but particle detector in YR10 now active, and recorded in MBS.

Clear effect of target on beam lifetime. Target off -> nearly no counts.

Attachment 22 - IPM (no target)

Attachment 23 - IPM (target)

Attachment 24 - YR10 particle detector (gap in events corresponds to target off period)

16.42 DSSSD bias & leakage current - attachments 1-2

DSSSD #4 currently conditioning at -20V

FEE64 temperatures OK - attachment 3

ADC, DISC, PAUSE & RESUME data item stats - attachments 4-7

System wide checks OK *except* WR deocder status aida04 - attachment 8

WR timestamps OK - attachment 9

per FEE64 Rate spectra - attachments 10-12

per FEE64 1.8.L spectra - attachments 13-14

per FEE64 1.8.W spectra - 20us FSR - attachments 15-16

pulser peak width aida06 69 ch FWHM, aida08 67 ch FWHM

Merger, TapeServer OK - attachments 17-18

 

ASIC settings 2025Apr08-13.45.30

 aida01-aida04 slow comparator 0xc

aida05-aida12 slow comparator 0x8

aida13-aida16 slow comparator 0x64, LEC/MEC 0x1

17.23 CRYRING vacuum and temperature - attachments 19-21

 

17.34 routine checks OK

target density c. 5e+12, beam 3e+6

 

18.40 changed target x,y position to check correlation between target density and detector counts. New position: x:-1500, y:2400.

checks OK.

target density c. 3.7e+12, beam 3.1e+6

20.00 online analysis - per DSSSD energy spectra - attachment 25

 t, h and p energy peaks clearly observed DSSSD #1 and #2 - p energy peak less clear in DSSSD #3

 

20.30 checks ok

target density increased to c. 5e+12 for some minutes then back at c. 3.7e+12, beam 3.1e+6.

 

21.47 checks ok

target density increased to c. 5.7e+12 around 21.35, then back at c. 3.7e+12. beam 3.1e+6.

 

22.30 online analysis - per DSSSD energy spectra - attachment 26

 

22.35 Analysis of data file G22-205/R74_29 - attachment 27

no timewarps, all deadtimes << 1%

 

22.45 DSSSD#4 bias -20V -> -40V, leakage current -2.7uA -> -4.09uA

23.14 DSSS\D#4 leakage current c. -9uA, return to bias -20V

 

23.21 target density slightly unstable in the last hour, changed nozzle position to x=-2000, y=2500.

 

00:04

Temps OK

System wide checks OK

Rates OK

Merger OK (1400 kb/s)

Target oscillating, but looks usable. Need to have another look at temperatures tomorrow morning.

 

01:57

Temps OK

Rates OK

Merger OK

Target getting thicker.

 

05:00

Temps OK

Rates OK

Merger OK

Target fluctuating

7:20 All sysytem checks ok. Target lost density @ 6;55 from 4.6e12 to 2.6e12 - will try moving. Update 7:30 after moving density back up to 4.4e12.

8:45 All Checks ok but target is fluctuating - really wants to be around 2e12. Had it up to 5e12 about 25 mins ago. Naughty target tbh.

9:25 All checks ok, gone to no storage as now tuning next beam energy.

10:30 All system checks OK, still tuning.

11:38 All system checks ok, still tuning. Frank entered cave to adjust magnetic field as the beam was not cooling at 90 keV/A

12:36 All checks okay, cave still open

13:40 All checks okay, tuning beam currently

14:50 All Checks OK. Target beam overlap currently happening - scanning range not showing anything. Target back on but not hitting - thickness 1.2e12.

 

 

Attachment 1: Screenshot_from_2025-06-12_16-42-39.png
Screenshot_from_2025-06-12_16-42-39.png
Attachment 2: Screenshot_from_2025-06-12_16-42-48.png
Screenshot_from_2025-06-12_16-42-48.png
Attachment 3: Screenshot_from_2025-06-12_16-43-19.png
Screenshot_from_2025-06-12_16-43-19.png
Attachment 4: Screenshot_from_2025-06-12_16-43-30.png
Screenshot_from_2025-06-12_16-43-30.png
Attachment 5: Screenshot_from_2025-06-12_16-43-42.png
Screenshot_from_2025-06-12_16-43-42.png
Attachment 6: Screenshot_from_2025-06-12_16-43-56.png
Screenshot_from_2025-06-12_16-43-56.png
Attachment 7: Screenshot_from_2025-06-12_16-44-11.png
Screenshot_from_2025-06-12_16-44-11.png
Attachment 8: Screenshot_from_2025-06-12_16-44-40.png
Screenshot_from_2025-06-12_16-44-40.png
Attachment 9: Screenshot_from_2025-06-12_16-45-21.png
Screenshot_from_2025-06-12_16-45-21.png
Attachment 10: Screenshot_from_2025-06-12_16-46-17.png
Screenshot_from_2025-06-12_16-46-17.png
Attachment 11: Screenshot_from_2025-06-12_16-46-24.png
Screenshot_from_2025-06-12_16-46-24.png
Attachment 12: Screenshot_from_2025-06-12_16-46-35.png
Screenshot_from_2025-06-12_16-46-35.png
Attachment 13: Screenshot_from_2025-06-12_16-47-43.png
Screenshot_from_2025-06-12_16-47-43.png
Attachment 14: Screenshot_from_2025-06-12_16-49-50.png
Screenshot_from_2025-06-12_16-49-50.png
Attachment 15: Screenshot_from_2025-06-12_16-54-34.png
Screenshot_from_2025-06-12_16-54-34.png
Attachment 16: Screenshot_from_2025-06-12_16-55-13.png
Screenshot_from_2025-06-12_16-55-13.png
Attachment 17: Screenshot_from_2025-06-12_16-55-33.png
Screenshot_from_2025-06-12_16-55-33.png
Attachment 18: Screenshot_from_2025-06-12_16-55-41.png
Screenshot_from_2025-06-12_16-55-41.png
Attachment 19: 1.png
1.png
Attachment 20: 3.png
3.png
Attachment 21: 2.png
2.png
Attachment 22: 20250612_163243.jpg
20250612_163243.jpg
Attachment 23: 20250612_161105.jpg
20250612_161105.jpg
Attachment 24: 20250612_163237.jpg
20250612_163237.jpg
Attachment 25: Screenshot_from_2025-06-12_20-00-18.png
Screenshot_from_2025-06-12_20-00-18.png
Attachment 26: Screenshot_from_2025-06-12_22-29-57.png
Screenshot_from_2025-06-12_22-29-57.png
Attachment 27: R74_29
 *** TDR format 3.3.0 analyser - TD - May 2021
 *** ERROR: READ I/O error:       5002
                   blocks:      32000
          ADC data format:  261845231 (  182631.5 Hz)
        Other data format:      74772 (      52.2 Hz)
 Sample trace data format:          0 (       0.0 Hz)
         Undefined format:          0 (       0.0 Hz)
   Other data format type:      PAUSE:         45 (       0.0 Hz)
                               RESUME:         45 (       0.0 Hz)
                              SYNC100:      37341 (      26.0 Hz)
                              WR48-63:      37341 (      26.0 Hz)
                           FEE64 disc:          0 (       0.0 Hz)
                             MBS info:          0 (       0.0 Hz)
                           Other info:          0 (       0.0 Hz)

   ADC data range bit set:        106 (       0.1 Hz)

                Timewarps:        ADC:          0 (       0.0 Hz)
                                PAUSE:          0 (       0.0 Hz)
                               RESUME:          0 (       0.0 Hz)
                              SYNC100:          0 (       0.0 Hz)
                              WR48-63:          0 (       0.0 Hz)
                           FEE64 disc:          0 (       0.0 Hz)
                             MBS info:          0 (       0.0 Hz)
                            Undefined:          0 (       0.0 Hz)
                         Sample trace:          0 (       0.0 Hz)

 *** Timestamp elapsed time:     1433.736 s
 FEE  elapsed dead time(s) elapsed idle time(s)
  0                0.185                0.000
  1                0.000              628.972
  2                0.000              554.120
  3                0.050                2.762
  4                0.005                6.506
  5                0.000             1033.389
  6                0.000              131.775
  7                0.000              759.159
  8                2.740                0.000
  9                0.004               21.901
 10                0.000             1310.256
 11                0.005                9.027
 12                0.000                0.000
 13                0.000                0.000
 14                0.000                0.000
 15                0.000                0.000
 16                0.000                0.000
 17                0.000                0.000
 18                0.000                0.000
 19                0.000                0.000
 20                0.000                0.000
 21                0.000                0.000
 22                0.000                0.000
 23                0.000                0.000
 24                0.000                0.000
 25                0.000                0.000
 26                0.000                0.000
 27                0.000                0.000
 28                0.000                0.000
 29                0.000                0.000
 30                0.000                0.000
 31                0.000                0.000
 32                0.000                0.000

 *** Statistics
 FEE  ADC Data Other Data     Sample  Undefined      Pause     Resume    SYNC100    WR48-63       Disc        MBS      Other   HEC Data
  0   40194292      11718          0          0         11         11       5848       5848          0          0          0         76
  1    7144658       1920          0          0          0          0        960        960          0          0          0         30
  2    7356264       2130          0          0          0          0       1065       1065          0          0          0          0
  3   30372380       8614          0          0          5          5       4302       4302          0          0          0          0
  4   25621460       7438          0          0          3          3       3716       3716          0          0          0          0
  5    4380276       1110          0          0          0          0        555        555          0          0          0          0
  6   13281715       3970          0          0          0          0       1985       1985          0          0          0          0
  7    6038211       1666          0          0          0          0        833        833          0          0          0          0
  8   76901033      21984          0          0         20         20      10972      10972          0          0          0          0
  9   21957648       6166          0          0          1          1       3082       3082          0          0          0          0
 10    1365330        376          0          0          0          0        188        188          0          0          0          0
 11   27231964       7680          0          0          5          5       3835       3835          0          0          0          0
 12          0          0          0          0          0          0          0          0          0          0          0          0
 13          0          0          0          0          0          0          0          0          0          0          0          0
 14          0          0          0          0          0          0          0          0          0          0          0          0
 15          0          0          0          0          0          0          0          0          0          0          0          0
 16          0          0          0          0          0          0          0          0          0          0          0          0
 17          0          0          0          0          0          0          0          0          0          0          0          0
 18          0          0          0          0          0          0          0          0          0          0          0          0
 19          0          0          0          0          0          0          0          0          0          0          0          0
 20          0          0          0          0          0          0          0          0          0          0          0          0
 21          0          0          0          0          0          0          0          0          0          0          0          0
 22          0          0          0          0          0          0          0          0          0          0          0          0
 23          0          0          0          0          0          0          0          0          0          0          0          0
 24          0          0          0          0          0          0          0          0          0          0          0          0
 25          0          0          0          0          0          0          0          0          0          0          0          0
 26          0          0          0          0          0          0          0          0          0          0          0          0
 27          0          0          0          0          0          0          0          0          0          0          0          0
 28          0          0          0          0          0          0          0          0          0          0          0          0
 29          0          0          0          0          0          0          0          0          0          0          0          0
 30          0          0          0          0          0          0          0          0          0          0          0          0
 31          0          0          0          0          0          0          0          0          0          0          0          0
 32          0          0          0          0          0          0          0          0          0          0          0          0

 *** Timewarps
 FEE       ADC      Pause     Resume    SYNC100    WR48-63       Disc        MBS  Undefined    Samples
  0          0          0          0          0          0          0          0          0          0
  1          0          0          0          0          0          0          0          0          0
  2          0          0          0          0          0          0          0          0          0
  3          0          0          0          0          0          0          0          0          0
  4          0          0          0          0          0          0          0          0          0
  5          0          0          0          0          0          0          0          0          0
  6          0          0          0          0          0          0          0          0          0
  7          0          0          0          0          0          0          0          0          0
  8          0          0          0          0          0          0          0          0          0
  9          0          0          0          0          0          0          0          0          0
 10          0          0          0          0          0          0          0          0          0
 11          0          0          0          0          0          0          0          0          0
 12          0          0          0          0          0          0          0          0          0
 13          0          0          0          0          0          0          0          0          0
 14          0          0          0          0          0          0          0          0          0
 15          0          0          0          0          0          0          0          0          0
 16          0          0          0          0          0          0          0          0          0
 17          0          0          0          0          0          0          0          0          0
 18          0          0          0          0          0          0          0          0          0
 19          0          0          0          0          0          0          0          0          0
 20          0          0          0          0          0          0          0          0          0
 21          0          0          0          0          0          0          0          0          0
 22          0          0          0          0          0          0          0          0          0
 23          0          0          0          0          0          0          0          0          0
 24          0          0          0          0          0          0          0          0          0
 25          0          0          0          0          0          0          0          0          0
 26          0          0          0          0          0          0          0          0          0
 27          0          0          0          0          0          0          0          0          0
 28          0          0          0          0          0          0          0          0          0
 29          0          0          0          0          0          0          0          0          0
 30          0          0          0          0          0          0          0          0          0
 31          0          0          0          0          0          0          0          0          0
 32          0          0          0          0          0          0          0          0          0

 *** Program elapsed time:   24.852s ( 1287.645 blocks/s,  80.478 Mb/s)
  741   Thu Jun 12 14:22:27 2025 JM, CB, TD, CF, JG, EMRun 73

15:00

Cave closed, waiting for beam to come back.

Reset all ASIC settings to those in elog 737

DSSD4 remains OFF and is set to the MEC channel.

ADC data items and data writing to disk - Attachments 1+2

While the cave has been closed we warmed the target up to 80K and have now cooled back to 30.5K to increase the density which was dropping into the 10^11 range. We have now recovered densities in the mid 10^12 at/cm2 range.

Detector leakage current plot - Attachment 3,4

All system checks OK

 

Started at 16:20

EM and JG trying to get scintillator signals in MBS

Configuration completed?

Stopped 16:32

Attachment 1: Screenshot_from_2025-06-12_15-23-37.png
Screenshot_from_2025-06-12_15-23-37.png
Attachment 2: Screenshot_from_2025-06-12_15-23-51.png
Screenshot_from_2025-06-12_15-23-51.png
Attachment 3: Screenshot_from_2025-06-12_15-29-14.png
Screenshot_from_2025-06-12_15-29-14.png
Attachment 4: Screenshot_from_2025-06-12_15-31-19.png
Screenshot_from_2025-06-12_15-31-19.png
  740   Thu Jun 12 14:08:42 2025 JM, CBRun 72 - pulser walkthrough

Pulser walkthrough conducted while the cave was open.

Pulser amplitudes 90000, 80000, 70000, 60000, 50000, 40000, 30000, 20000, 10000

DSSD4 was set to MEC and bias off as the detector had previously tripped.

All thresholds set to 0x64 for the pulser walkthrough

Attachment 1: Screenshot_from_2025-06-12_15-33-03.png
Screenshot_from_2025-06-12_15-33-03.png
  739   Thu Jun 12 06:20:50 2025 JMRun 68,69 - background

07:20

Started a background run while the cave is open.

All ASIC settings remain unchanged, we just have no beam.

We see obvious difference in the rate plots for DSSD2 from the monitoring code. Attachment 1 - beam on, Attachment 2- beam off.

Data rate ~1.6 MB.sec

Checks OK.

10:30

Lost DSSD4 again. Writing a lot of junk to disk.

Switched to MEC channels. Still writing lots of data. Stopped the run.

Required ASIC check load on aida13,14,15,16 in order to bring rates back down.

Slowly re-biasing detector. At 35V

Moved detectors fully out for safety. Left potentiometer = 18.86 mm and right potentiometer = 19.8 mm. No effect on leakage currents.

Past 40V we large increases in leakage current. Only went up to 10uA and didn't trip. Set to 40V, current back down to normal levels.

 

Note we have turned the target compressor off to heat up before cooling later to try and get higher densities. This follows a drop in E1 yesterday reducing the target density.

10:45

Started new background run - Run 69.

DSSD4 set to MEC.

Thresholds still as in previous runs (120, 80, 80 keV).

Attachment 1: Screenshot_from_2025-06-12_07-01-00.png
Screenshot_from_2025-06-12_07-01-00.png
Attachment 2: Screenshot_from_2025-06-12_07-20-26.png
Screenshot_from_2025-06-12_07-20-26.png
  738   Wed Jun 11 10:13:42 2025 EW, CB, PJW, AR, JMRun 67

As run 66, optimisation of beam completed

DSSD4 now working and thresholds to 0x30 (480keV), switched to LEC

Tape service at 1.7MB/s

Did NOT reset the root histograms!!

 

11:30 All checks OK

13:48 All checks OK

15:15, fULL CHECK, STATUS - Okay. Target density ~2E12

16:15 All Checks OK. Target 2.9e12

17:30, Full check! status - okay. target 2.3e12

18:30 All Checks Ok

19:30 All Checks OK

21:30: All System Checks OK but targte density started dropping around 8:20. Tried to move the nozel but any x direction made it drop alot and y did nothing. Values hovering steadily between 1.1e12 to 1.4e12. Will try and move nozel again in 1 hour.

22:40: Full check, Status -okay! thickness 1.4E12. For more details look at the attchments

00:55: Full check. All looks okay!

03:55: Full check. All looks good!

06:28 Checks OK. Target density has dropped and is ~9e11 at/cm2. It will require some attention this afternoon

07:00 Run stopped. Final histogram attached. Number of counts somewhat obscured by background level.

Attachment 1: Screenshot_from_2025-06-11_11-15-33.png
Screenshot_from_2025-06-11_11-15-33.png
Attachment 2: Screenshot_from_2025-06-11_11-15-16.png
Screenshot_from_2025-06-11_11-15-16.png
Attachment 3: Screenshot_from_2025-06-11_22-37-25.png
Screenshot_from_2025-06-11_22-37-25.png
Attachment 4: Screenshot_from_2025-06-11_22-37-45.png
Screenshot_from_2025-06-11_22-37-45.png
Attachment 5: Screenshot_from_2025-06-11_22-37-53.png
Screenshot_from_2025-06-11_22-37-53.png
Attachment 6: Screenshot_from_2025-06-11_22-38-01.png
Screenshot_from_2025-06-11_22-38-01.png
Attachment 7: Screenshot_from_2025-06-11_22-38-08.png
Screenshot_from_2025-06-11_22-38-08.png
Attachment 8: Screenshot_from_2025-06-11_22-38-14.png
Screenshot_from_2025-06-11_22-38-14.png
Attachment 9: Screenshot_from_2025-06-11_22-38-31.png
Screenshot_from_2025-06-11_22-38-31.png
Attachment 10: Screenshot_from_2025-06-11_22-38-49.png
Screenshot_from_2025-06-11_22-38-49.png
Attachment 11: Screenshot_from_2025-06-11_22-39-08.png
Screenshot_from_2025-06-11_22-39-08.png
Attachment 12: Screenshot_from_2025-06-11_22-39-13.png
Screenshot_from_2025-06-11_22-39-13.png
Attachment 13: Screenshot_from_2025-06-11_22-40-55.png
Screenshot_from_2025-06-11_22-40-55.png
Attachment 14: Screenshot_from_2025-06-12_07-01-35.png
Screenshot_from_2025-06-12_07-01-35.png
  737   Wed Jun 11 09:16:55 2025 CBASIC Control Settings

11 June 10:18

2025Apr08-13.45.30 settings

DSSD4 not working as of the screenshots (MEC + High Threshold)

See attached.

 

Attachment 1: Screenshot_from_2025-06-11_10-14-06.png
Screenshot_from_2025-06-11_10-14-06.png
Attachment 2: Screenshot_from_2025-06-11_10-14-12.png
Screenshot_from_2025-06-11_10-14-12.png
Attachment 3: Screenshot_from_2025-06-11_10-14-17.png
Screenshot_from_2025-06-11_10-14-17.png
Attachment 4: Screenshot_from_2025-06-11_10-14-22.png
Screenshot_from_2025-06-11_10-14-22.png
Attachment 5: Screenshot_from_2025-06-11_10-14-26.png
Screenshot_from_2025-06-11_10-14-26.png
Attachment 6: Screenshot_from_2025-06-11_10-14-31.png
Screenshot_from_2025-06-11_10-14-31.png
Attachment 7: Screenshot_from_2025-06-11_10-14-37.png
Screenshot_from_2025-06-11_10-14-37.png
Attachment 8: Screenshot_from_2025-06-11_10-14-42.png
Screenshot_from_2025-06-11_10-14-42.png
Attachment 9: Screenshot_from_2025-06-11_10-14-50.png
Screenshot_from_2025-06-11_10-14-50.png
Attachment 10: Screenshot_from_2025-06-11_10-14-56.png
Screenshot_from_2025-06-11_10-14-56.png
Attachment 11: Screenshot_from_2025-06-11_10-15-02.png
Screenshot_from_2025-06-11_10-15-02.png
Attachment 12: Screenshot_from_2025-06-11_10-15-11.png
Screenshot_from_2025-06-11_10-15-11.png
Attachment 13: Screenshot_from_2025-06-11_10-15-20.png
Screenshot_from_2025-06-11_10-15-20.png
Attachment 14: Screenshot_from_2025-06-11_10-15-25.png
Screenshot_from_2025-06-11_10-15-25.png
Attachment 15: Screenshot_from_2025-06-11_10-15-31.png
Screenshot_from_2025-06-11_10-15-31.png
Attachment 16: Screenshot_from_2025-06-11_10-15-36.png
Screenshot_from_2025-06-11_10-15-36.png
Attachment 17: Screenshot_from_2025-06-11_10-15-41.png
Screenshot_from_2025-06-11_10-15-41.png
  736   Wed Jun 11 08:37:48 2025 CB, PJWRun 66

09:38

As Run 65, no changes, CRYRING operators attempting to improve beam intensity.

We should be able to keep running while they do, but there is a chance of some missed injections or other oddities.

Checks OK

Root histograms NOT zeroed.

 

10:20

Checks OK

Attachment 1: Screenshot_from_2025-06-11_10-21-30.png
Screenshot_from_2025-06-11_10-21-30.png
Attachment 2: Screenshot_from_2025-06-11_10-21-44.png
Screenshot_from_2025-06-11_10-21-44.png
  735   Tue Jun 10 22:38:59 2025 JM, CB, PJWRun 65

23:40

Offline anlaysis shows two clear peaks for DSSD1 and 2 at very low energies close to the threshold. One must be elastics and the other a noise peak.

For DSSD3 the noise level is so exceptionally high due to the low threshold that nothing can be discerned from background.

Therefore I decided to stop the previous run and increase the slow comparator threshold for DSSD3 from 50 keV to 80 keV (same as DSSD2).

Run 65 started.

150 keV/u beam on target, Target overlap confirmed with lifetime measurement using bunched beam.

Detectors in with servos only. Both arms are IN and stationary.

Left potentiometer = 36.54 mm

Right potentiometer = 37.96 mm

Data rate has dropped to 1.5 MB/sec and merger is between 600k and 900k.

Stats attached at new thresholds.

Stats zero'd to see number of pause and resume overnight.

02:30

Some reaction counts visible on monitoring histograms (rate v low)

Checks OK, same fee (aida16 fails clock and adc as before)

pause info items after a few hours attached

05:00

We see the peaks much more clearly now. Rate is low but they are there. Target density looks like it has been fluctuating in the low 10^12 region. 

Note for this energy with 1e8 pps, 1e12 target density I estimate ~2 counts/h per detector for the 3He peak.from our spreadsheet calculator. After 6 hours of this run I saw about 10-15 counts in the peak of DSSD2, perhaps the same in DSSD3. DSSD1 has missing strips so is harder to see. This rate is about 48 hours for 5% (if we had 4 dssds).   

Checks OK

Leakage current of dssd4 has been stable at 40V for a while now. Pushed voltage up to 75V with leakage current stable. At 80V it started to go over current

Turned bias to 60V and have left on at this level. It looks stable. Perhaps in the morning it can be pushed to close to 100V.

 

07:15

Temps OK

System wide checks: 16 fails ADC calibration, 14 fails WR decoder

Rates OK, some Pause/resume statementes overnight. See attached x2

Merger OK. 1.5 Mb/s

Peaks clearly visible in two detectors.

 

08:25 All checks OK. DSSD 4 to 75V - OK.

09:10 All checks OK. DSSD 4 to 80V. Trying to improve target (2-4E12).

Attachment 1: Screenshot_from_2025-06-10_23-46-27.png
Screenshot_from_2025-06-10_23-46-27.png
Attachment 2: offline.png
offline.png
Attachment 3: Screenshot_from_2025-06-11_02-29-55.png
Screenshot_from_2025-06-11_02-29-55.png
Attachment 4: Screenshot_from_2025-06-11_05-06-46.png
Screenshot_from_2025-06-11_05-06-46.png
Attachment 5: Screenshot_from_2025-06-11_07-21-12.png
Screenshot_from_2025-06-11_07-21-12.png
Attachment 6: Screenshot_from_2025-06-11_07-21-40.png
Screenshot_from_2025-06-11_07-21-40.png
Attachment 7: Screenshot_from_2025-06-11_07-24-26.png
Screenshot_from_2025-06-11_07-24-26.png
  734   Tue Jun 10 16:46:39 2025 JM, CB, EW, AY, PC150 keV - Run64

Run 64 started.

150 keV/u beam on target, Target overlap confirmed with lifetime measurement using bunched beam.

We do not see Rutherford on the detector - perhaps this is a calibration issue - perhaps too much noise at threshold.

We leave for a few hours to see if we obsrerve any reactions peaks

Moved detectors in with servos only. Both arms are IN and stationary.

Left potentiometer = 36.54 mm

Right potentiometer = 37.96 mm

Horizontal beam position = -3.0 mm

data rate is ~5 MB/sec

Electron cooler read values; Cooler voltage = 88.57V, cooler current=1.28 mA.

19:05

Full check, Status - Okay! Target thickness ~ 2E12. For more details, check the attached screenshots

20:12

Full check, Status - Okay! Target thickness ~ 1.6E12, For more details, check the attached screenshots

21:22

Full check, Status - Okay! For more details, check the attached screenshots

22:30

Full check, Status - Okay! Target thickness ~ 1.8E12, For more details, check the attached screenshots

23:00

Checks OK. I have increased the equal energy cut back to 500 keV and increased bin size on Ey energy histogram to 50 keV to try and see the nuclear peaks better.

Screen shot attached of monitoring code before it was reset. Some small peaks visible.

Attachment 1: Screenshot_from_2025-06-10_18-58-46.png
Screenshot_from_2025-06-10_18-58-46.png
Attachment 2: Screenshot_from_2025-06-10_18-59-09.png
Screenshot_from_2025-06-10_18-59-09.png
Attachment 3: Screenshot_from_2025-06-10_18-59-20.png
Screenshot_from_2025-06-10_18-59-20.png
Attachment 4: Screenshot_from_2025-06-10_18-59-48.png
Screenshot_from_2025-06-10_18-59-48.png
Attachment 5: Screenshot_from_2025-06-10_19-00-45.png
Screenshot_from_2025-06-10_19-00-45.png
Attachment 6: Screenshot_from_2025-06-10_19-01-12.png
Screenshot_from_2025-06-10_19-01-12.png
Attachment 7: Screenshot_from_2025-06-10_19-03-55.png
Screenshot_from_2025-06-10_19-03-55.png
Attachment 8: Screenshot_from_2025-06-10_19-05-08.png
Screenshot_from_2025-06-10_19-05-08.png
Attachment 9: Screenshot_from_2025-06-10_19-05-13.png
Screenshot_from_2025-06-10_19-05-13.png
Attachment 10: Screenshot_from_2025-06-10_19-05-22.png
Screenshot_from_2025-06-10_19-05-22.png
Attachment 11: Screenshot_from_2025-06-10_19-06-15.png
Screenshot_from_2025-06-10_19-06-15.png
Attachment 12: Screenshot_from_2025-06-10_20-13-17.png
Screenshot_from_2025-06-10_20-13-17.png
Attachment 13: Screenshot_from_2025-06-10_20-13-36.png
Screenshot_from_2025-06-10_20-13-36.png
Attachment 14: Screenshot_from_2025-06-10_20-14-36.png
Screenshot_from_2025-06-10_20-14-36.png
Attachment 15: Screenshot_from_2025-06-10_20-15-30.png
Screenshot_from_2025-06-10_20-15-30.png
Attachment 16: Screenshot_from_2025-06-10_20-15-55.png
Screenshot_from_2025-06-10_20-15-55.png
Attachment 17: Screenshot_from_2025-06-10_20-16-07.png
Screenshot_from_2025-06-10_20-16-07.png
Attachment 18: Screenshot_from_2025-06-10_20-16-19.png
Screenshot_from_2025-06-10_20-16-19.png
Attachment 19: Screenshot_from_2025-06-10_21-18-16.png
Screenshot_from_2025-06-10_21-18-16.png
Attachment 20: Screenshot_from_2025-06-10_21-19-10.png
Screenshot_from_2025-06-10_21-19-10.png
Attachment 21: Screenshot_from_2025-06-10_21-19-48.png
Screenshot_from_2025-06-10_21-19-48.png
Attachment 22: Screenshot_from_2025-06-10_21-20-30.png
Screenshot_from_2025-06-10_21-20-30.png
Attachment 23: Screenshot_from_2025-06-10_21-20-55.png
Screenshot_from_2025-06-10_21-20-55.png
Attachment 24: Screenshot_from_2025-06-10_21-22-56.png
Screenshot_from_2025-06-10_21-22-56.png
Attachment 25: Screenshot_from_2025-06-10_21-23-03.png
Screenshot_from_2025-06-10_21-23-03.png
Attachment 26: Screenshot_from_2025-06-10_21-23-10.png
Screenshot_from_2025-06-10_21-23-10.png
Attachment 27: Screenshot_from_2025-06-10_21-56-07.png
Screenshot_from_2025-06-10_21-56-07.png
Attachment 28: Screenshot_from_2025-06-10_22-30-47.png
Screenshot_from_2025-06-10_22-30-47.png
Attachment 29: Screenshot_from_2025-06-10_22-31-13.png
Screenshot_from_2025-06-10_22-31-13.png
Attachment 30: Screenshot_from_2025-06-10_22-31-49.png
Screenshot_from_2025-06-10_22-31-49.png
Attachment 31: Screenshot_from_2025-06-10_22-31-59.png
Screenshot_from_2025-06-10_22-31-59.png
Attachment 32: Screenshot_from_2025-06-10_22-32-06.png
Screenshot_from_2025-06-10_22-32-06.png
Attachment 33: Screenshot_from_2025-06-10_22-32-16.png
Screenshot_from_2025-06-10_22-32-16.png
Attachment 34: Screenshot_from_2025-06-10_22-32-48.png
Screenshot_from_2025-06-10_22-32-48.png
Attachment 35: Screenshot_from_2025-06-10_22-33-04.png
Screenshot_from_2025-06-10_22-33-04.png
Attachment 36: Screenshot_from_2025-06-10_22-33-19.png
Screenshot_from_2025-06-10_22-33-19.png
Attachment 37: Screenshot_from_2025-06-10_22-33-24.png
Screenshot_from_2025-06-10_22-33-24.png
Attachment 38: Screenshot_from_2025-06-10_22-33-38.png
Screenshot_from_2025-06-10_22-33-38.png
Attachment 39: Screenshot_from_2025-06-10_22-33-42.png
Screenshot_from_2025-06-10_22-33-42.png
Attachment 40: Screenshot_from_2025-06-10_22-33-58.png
Screenshot_from_2025-06-10_22-33-58.png
Attachment 41: Screenshot_2025-06-10_22-37-40.png
Screenshot_2025-06-10_22-37-40.png
Attachment 42: Screenshot_from_2025-06-10_23-00-20.png
Screenshot_from_2025-06-10_23-00-20.png
  733   Tue Jun 10 13:29:23 2025 EW, CB, JMThresholds changed

Experimented with slow comparator thresholds as we go down to lower beam energies

Checked deadlayer loses with SRIM, expected ruthford energies are as follows:

150keV/u -> 230keV

100keV/u -> 120keV

80keV/u -> 76keV

New thresholds:

aida1-4 0xc (120keV)

aida5-8 0x8 (80keV)

aida9-12 0x5 (50keV)

aida13-16 0x64(1MeV) set to MEC channel

Statistics and data rate with beam on but no target

Deleted April 24,25 and G22-86 from GSI, backup in Edinburgh

Attachment 1: Screenshot_from_2025-06-10_14-35-07.png
Screenshot_from_2025-06-10_14-35-07.png
Attachment 2: Screenshot_from_2025-06-10_14-34-59.png
Screenshot_from_2025-06-10_14-34-59.png
Attachment 3: Screenshot_from_2025-06-10_14-34-49.png
Screenshot_from_2025-06-10_14-34-49.png
Attachment 4: Screenshot_from_2025-06-10_14-34-17.png
Screenshot_from_2025-06-10_14-34-17.png
  732   Tue Jun 10 09:57:50 2025 CB, JM, PJW, YuL, JGRun 62

10:58

As Run 61, no target for background.

200 keV/A

Saved the ASIC settings for currently running. I made a mistake and overwrote the 2025APR08 settings.

Current settings have aida13-16 set to MEC channel and slow comparator set to 0x64.

All ither channels set to 0xc slow comparator.

 

Energy changed to 150 keV/A a few minutes before the end of this run. No storage. Waiting for the new beam.

  731   Tue Jun 10 09:44:33 2025 JGMBS not running for a week

The MBS daq was not writing data since June 2. 22:39 (run_0035.lmd)

This means, no detector and target timing is available for this period.

The reason is that the rfio server was running on atppc025, which crashed at that time.

We resumed running MBS now at 10.06. ca. 10:35. (run_0036.lmd).

  730   Tue Jun 10 09:44:30 2025 JGMBS not running for a week
The MBS daq was not writing data since June 2. 22:39 (run_0035.lmd) The reason is that the rfio server was running on atppc025 which crashed at that time. We resumed running MBS now at 10.06. ca. 10:35. (run_0036.lmd).
  729   Tue Jun 10 09:44:20 2025 JGMBS not running for a week
The MBS daq was not writing data since June 2. 22:39 (run_0035.lmd) The reason is that the rfio server was running on atppc025 which crashed at that time. We resumed running MBS now at 10.06. ca. 10:35. (run_0036.lmd).
  728   Sun Jun 8 11:55:00 2025 EM, PC, JM, CB, EW, AY, CB, PJW, YuLRun 58,59,60, 61

12:52

Started Run 58 - Beam energy 200 keV/u.

Slow comparator threshold set to 0xc (130 keV) for all channels. Fast discriminators turned OFF for all channels by setting enable mask in discrinator control to 0xfffff. See attachments. 

D+ beam deuterium target. D+ beam energy is now 0.2 MeV/u. (beam calc calculation is 0.2037 MeV/u).

All four detectors positioned fully out and stationary.

Ring cycle time is 5 seconds.

Cooler read value = 117.58 V, 0.00178 mA

Target around 8.9E12 atoms/cm2

All checks ok!

Beam on ionisation profile monitor is lost quite rapidly and looks to increase in size quite a lot due to target interactions. We may have to look at lowering target density

Attachment 25 shows IPM with no target and attachment 26 shows IPM with the target 8e12 at/cm2. Compare also to attachment 12 (IPM with target at 2.6e12).  

14:20

Checks are ok!

15:25

Fees 13-16 switched to MEC and bias on DSSD4 switched off

All other checks ok

17:00

Target dropped to 1-2E12 atoms/cm2

All checks okay- 17:25

17:45

Moved nozzle from x=-1500 to x=-2000, density went from 1E12 to 3E12

Moved nozzle again at 18:00 from x=-2000 to x=-3000 in steps of 500, density at 4E12

18:45

Turned bias ON for dssd4 stable at V=10 but unstable at V=20 so bias turned back OFF

19:30

All checks ok!

IPM attached, target density 2.6 atoms/cm2 (attachment 12)

20:00

Tried moving the nozzle more as target density dropped to 2-3E12 again

Moved in 500 steps either side of x=-2500 but didn't improve, then moved in 100 steps from y=1600 to y=2300 and target density now 6E12 at x=-2500 and y=2300

Tried dssd4 again but only stable to V=20 so bias OFF 

21:50- All checks good, target density stable at 8E12

22:50- All checks fine (see attachments), target density at 6E12

00:10

Checks OK. Have zero'd statistics to see pause and resume statements over an hour.

Rates looks OK. ~140 c/h 3He, and ~80 c/h for 3H and 1H each.

Target has been stable at 7e12 at/cm2 for the last hour. 

02:10

Checks OK. See pause statements after 2 hours - attachment 27, very few overall.

03:07

Checks OK. Target (9-11)E12

 

04:06

Checks OK. Target (9-11)E12

 

07:00

Checks OK. Target 9.8E12

page AIDA/Check/Check.tml.  failed

ptoblem fixed

 

09:00

Checks OK. Target 8.9E12

 

10:32

No target data displayed for 10 minutes, but jet density unchanged. Restored display from remote access by M.L.

 

11:22

Storage data stopped and Faraday cup inserted to check the leakage current of Detector #4. Not good, current increases above 10 uA already at 40 V. 

11:29

HV Detector #4 swithced off, Faraday cup removed, data stored again.

 

11:35

Check OK. Target 8.6E12

 

13:06

Check OK. Target 7.5E12

 

15:00

Temperature check - aida03 returns "No response". No response for rates either. Merger looks fine for 3? Likely just happened.

Attempted to STOP the DAQ. All FEEs still going (refused to stop) and 3 in error. Decided to just turn all FEEs off.

Restored DAQ with no issues. 3 is working. Temperatures fine. Rates fine. Unclear what had happened.

Run59 is EMPTY. New run = run 60.

16:00, Full check, everything working after the restart

17:35  Full check, everything is working smoothly!

18:35: Full check, status -Okay.  thickness 5.7E12,

19:35: full ckeck! Status - OKAY, target theckness ~5E12

20:50 : Full check! status - okay

22:00 : Full check, status -okay. Target thickness 6.4E12

22:55 : Full check, status -okay! ------ target thickness 5.8E12

01:00 : Full check, all okay! 

04:00 : Full check, all okay! 

07:00 : Full check, all okay!

 

10:00

Attempted to change cooling to improve the beam. Changed cooling by 0.5V. Significant improvement in the IPM, but Rutherford goes down. Target missed?

Changed orbit bumps. Increased Rutherford. IPM is back to how it was. Rates are possibly slightly higher? Will attempt more systematically in the afternoon, when we change beam energy again.

Attempted to delay target on by 1 second. No significant changes in beam lifetime or rates. Can try systematically again in the afternoon.

Restored settings to how they were. Started new run 61.

Temps OK

Rates OK

Merger OK

Online histos look OK (did not reset)

 

Checks

Clock status

FEE64 module aida16 global clocks failed, 6
 Clock status test result: Passed 15, Failed 1

ADC calibration

FEE64 module aida16 failed
Calibration test result: Passed 15, Failed 1

White Rabbit decoder

   
         Base         Current     Difference
aida14 fault      0xbf08 :      0xbf08 :      0  
aida14 : WR status 0x10
 White Rabbit error counter test result: Passed 15, Failed 1

Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR

FPGA Timestamp error counter test result: Passed 16, Failed 0

 

 

Attachment 1: Screenshot_from_2025-06-08_12-40-25.png
Screenshot_from_2025-06-08_12-40-25.png
Attachment 2: Screenshot_from_2025-06-08_12-41-01.png
Screenshot_from_2025-06-08_12-41-01.png
Attachment 3: Screenshot_from_2025-06-08_12-41-12.png
Screenshot_from_2025-06-08_12-41-12.png
Attachment 4: Screenshot_from_2025-06-08_12-41-53.png
Screenshot_from_2025-06-08_12-41-53.png
Attachment 5: Screenshot_from_2025-06-08_12-42-01.png
Screenshot_from_2025-06-08_12-42-01.png
Attachment 6: Screenshot_from_2025-06-08_12-42-19.png
Screenshot_from_2025-06-08_12-42-19.png
Attachment 7: Screenshot_from_2025-06-08_12-48-47.png
Screenshot_from_2025-06-08_12-48-47.png
Attachment 8: Screenshot_from_2025-06-08_13-03-41.png
Screenshot_from_2025-06-08_13-03-41.png
Attachment 9: Screenshot_from_2025-06-08_13-03-52.png
Screenshot_from_2025-06-08_13-03-52.png
Attachment 10: Screenshot_from_2025-06-08_13-04-41.png
Screenshot_from_2025-06-08_13-04-41.png
Attachment 11: Screenshot_from_2025-06-08_13-07-41.png
Screenshot_from_2025-06-08_13-07-41.png
Attachment 12: IMG_4311.JPG
IMG_4311.JPG
Attachment 13: Screenshot_from_2025-06-08_14-12-02.png
Screenshot_from_2025-06-08_14-12-02.png
Attachment 14: Screenshot_from_2025-06-08_19-32-22.png
Screenshot_from_2025-06-08_19-32-22.png
Attachment 15: Screenshot_from_2025-06-08_19-32-35.png
Screenshot_from_2025-06-08_19-32-35.png
Attachment 16: Screenshot_from_2025-06-08_19-33-35.png
Screenshot_from_2025-06-08_19-33-35.png
Attachment 17: Screenshot_from_2025-06-08_19-33-39.png
Screenshot_from_2025-06-08_19-33-39.png
Attachment 18: Screenshot_from_2025-06-08_19-33-49.png
Screenshot_from_2025-06-08_19-33-49.png
Attachment 19: Screenshot_from_2025-06-08_19-34-24.png
Screenshot_from_2025-06-08_19-34-24.png
Attachment 20: Screenshot_from_2025-06-08_22-52-14.png
Screenshot_from_2025-06-08_22-52-14.png
Attachment 21: Screenshot_from_2025-06-08_22-47-46.png
Screenshot_from_2025-06-08_22-47-46.png
Attachment 22: Screenshot_from_2025-06-08_22-47-57.png
Screenshot_from_2025-06-08_22-47-57.png
Attachment 23: Screenshot_from_2025-06-08_22-48-27.png
Screenshot_from_2025-06-08_22-48-27.png
Attachment 24: Screenshot_from_2025-06-08_22-48-36.png
Screenshot_from_2025-06-08_22-48-36.png
Attachment 25: 20250608_121153.jpg
20250608_121153.jpg
Attachment 26: 20250608_124137.jpg
20250608_124137.jpg
Attachment 27: Screenshot_from_2025-06-09_02-06-22.png
Screenshot_from_2025-06-09_02-06-22.png
Attachment 28: Screenshot_from_2025-06-09_02-06-22.png
Screenshot_from_2025-06-09_02-06-22.png
Attachment 29: Screenshot_from_2025-06-09_03-05-16.png
Screenshot_from_2025-06-09_03-05-16.png
Attachment 30: Screenshot_from_2025-06-09_03-06-47.png
Screenshot_from_2025-06-09_03-06-47.png
Attachment 31: Screenshot_from_2025-06-09_15-51-46.png
Screenshot_from_2025-06-09_15-51-46.png
Attachment 32: Screenshot_from_2025-06-09_15-52-16.png
Screenshot_from_2025-06-09_15-52-16.png
Attachment 33: Screenshot_from_2025-06-09_15-52-50.png
Screenshot_from_2025-06-09_15-52-50.png
Attachment 34: Screenshot_from_2025-06-09_15-53-06.png
Screenshot_from_2025-06-09_15-53-06.png
Attachment 35: Screenshot_from_2025-06-09_15-53-22.png
Screenshot_from_2025-06-09_15-53-22.png
Attachment 36: Screenshot_from_2025-06-09_15-55-24.png
Screenshot_from_2025-06-09_15-55-24.png
Attachment 37: Screenshot_from_2025-06-09_15-56-00.png
Screenshot_from_2025-06-09_15-56-00.png
Attachment 38: Screenshot_from_2025-06-09_15-56-19.png
Screenshot_from_2025-06-09_15-56-19.png
Attachment 39: Screenshot_from_2025-06-09_15-57-26.png
Screenshot_from_2025-06-09_15-57-26.png
Attachment 40: Screenshot_from_2025-06-09_15-58-04.png
Screenshot_from_2025-06-09_15-58-04.png
Attachment 41: Screenshot_from_2025-06-09_21-54-32.png
Screenshot_from_2025-06-09_21-54-32.png
Attachment 42: Screenshot_from_2025-06-09_21-55-03.png
Screenshot_from_2025-06-09_21-55-03.png
Attachment 43: Screenshot_from_2025-06-09_21-55-21.png
Screenshot_from_2025-06-09_21-55-21.png
Attachment 44: Screenshot_from_2025-06-09_21-55-30.png
Screenshot_from_2025-06-09_21-55-30.png
Attachment 45: Screenshot_from_2025-06-09_21-55-41.png
Screenshot_from_2025-06-09_21-55-41.png
Attachment 46: Screenshot_from_2025-06-09_21-55-49.png
Screenshot_from_2025-06-09_21-55-49.png
Attachment 47: Screenshot_from_2025-06-10_10-05-08.png
Screenshot_from_2025-06-10_10-05-08.png
Attachment 48: Screenshot_from_2025-06-10_10-05-29.png
Screenshot_from_2025-06-10_10-05-29.png
Attachment 49: Screenshot_from_2025-06-10_10-05-39.png
Screenshot_from_2025-06-10_10-05-39.png
Attachment 50: beamCalc-0.2.png
beamCalc-0.2.png
  727   Sun Jun 8 10:05:30 2025 EM, PC, JMRun57

11:06

Run during beam change. We will go down to 200 keV/u

All the checks are ok.

No storage mode

Target off

We played with some of the slow comparator thresholds to judge the statistics and data rate. 

Deemed threshold at 100 keV too low for the current beam energy due to excessive rates observed on all FEEs. If required for lower energies we can go to this level. 

Data rate at 130 keV threshold is about 2 MB/sec compared to 8 MB/sec at 100 keV. Note beam energy is 400 keV so no significant benefit going to the lowest threshold. 

  726   Sun Jun 8 09:07:22 2025 EM, PC, JMRun56

NOTE: RunR55 is missing

Run56: Background run with 250 keV/u. Target off using the usual procedure.

All  checks ok!

We moved to nostorage mode and will change the run during the beam change

Attachment 1: Screenshot_from_2025-06-08_10-10-06.png
Screenshot_from_2025-06-08_10-10-06.png
Attachment 2: Screenshot_from_2025-06-08_10-15-01.png
Screenshot_from_2025-06-08_10-15-01.png
  725   Sat Jun 7 10:23:26 2025 JM, CBcarme-gsi rebooted

10:00

The computer screen controlling the DAQ on carme-gsi appeared to have crashed.

Unable to move throught the different windows etc.

Unable to connect via anydesk. ssh access still available. It appears the x-server had crashed, we noticed earlier in the beam certain functions such as screenshots were glitchy so this was probably expected eventually.

We were able to access the DAQ on another CRYRING machine machine via 192.168.207.143:8015 and 192.168.207.143:8015 in the browser. DAQ is still running. We will try to reboot the x server without disturbing the DAQ.

ssh into carme-gsi and used the command 'systemctl reboot gdml' to attempt to reboot only this server.

This did not have the desired effect, whole of carme-gsi rebooted. We should have turned off the FEEs and stopped the DAQ gracefully but did not.

Restarted carme-gsi and rebooted the FEEs and DAQ system.

All checks OK -> clocks, ADC calibration etc.

ASIC settings loaded are 2025APR08. Only changes made to this settings were reducing the slow comparator threshold from 0x32 (1 MeV) to 0x14 (200 keV). aida03 and aida04 were set to the MEC on these settings, they have now been reset to the LEC channel (0x0). All channels on the LEC.

  724   Sat Jun 7 10:06:27 2025 JM, EM, CB, DB, AY, PC, CBRun 54

11:10

We have started a new run at the same energy as R53 (250 keV/u) due to an issue with the DAQ pc

Server and DAQ restarted

ASIC settings loaded 2025APR08. Only changes made are setting all FEEs to LEC (0x0) on channel 10, and setting the slow comparator thresholds to 0x14 (200 keV).

All beam settings are the same as the previous elog. Detectors remain stationary and in the fully out position.

All checks ok!

ADC data items statistics appear unchanged. See attachment.

Data rate ~ 800 kB/sec

12:40

All checks ok

The target is not super stable.

14:00

 Checks ok!  Target back to >3E12 1/cm2

 

15:00

 Checks ok! 

17:00

Full check, everything looks okay!, Areal density is in beetween 3-5E12. Rate in all the detectors look fine.

19:14

Full check, everything looks great. Average arial density around 5e12 atoms/cm2. For more details please look at the attached screenshots

20:20

Target density dropped to 8e10, We are waiting for 20 minutes before changing the nozzle positions!

20:41 : slowly increased to 2.2e11 but it is not increasing anymore

20:43 Changed the nozzle position X fromm -1000 to -1500

20:56, Target thickness recovered bsck to 5e12

21:25, Full check, everything runnig smoothly. There is nothing unexpected, for details, please have a look at the attached file

22:30 Full check, Stable target thickness of more than 5e12 atoms/cm2. everthig running fine:-) For more details, please look at the attached screenshots!

 

00:52

Temperatures OK

Rates OK

Leakage current stable

System wide checks OK

Merger OK

Beam is on

Target looks fine, 2E-10 mbar in YR09 Etarget

 

02:51

Temperatures OK

Rates OK

Leakage current stable

System wide checks OK

Merger OK

Beam is on

Target looks fine, 3E-10 mbar in YR09 Etarget

 

03:45

Temperatures OK

Rates OK

Leakage current stable

System wide checks OK

Merger OK

Beam is on

Target looks fine, 4E-10 mbar in YR09 Etarget

 

07:05

Temperatures OK

Rates OK

Leakage current stable

System wide checks OK

Merger OK

Beam is on

Target looks great!!! 1.1E13 this morning!!

 

08:05

All checks ok!

 

09:05

All checks ok! Target is perfect to make this Sunday amazing

10:00

All checks ok! We completed the data taking on this energy. Atatched the final stats.

 

Attachment 1: Screenshot_from_2025-06-07_11-10-07.png
Screenshot_from_2025-06-07_11-10-07.png
Attachment 2: Screenshot_from_2025-06-07_11-10-37.png
Screenshot_from_2025-06-07_11-10-37.png
Attachment 3: Screenshot_from_2025-06-07_11-14-12.png
Screenshot_from_2025-06-07_11-14-12.png
Attachment 4: Screenshot_from_2025-06-07_11-15-33.png
Screenshot_from_2025-06-07_11-15-33.png
Attachment 5: Screenshot_from_2025-06-07_11-16-33.png
Screenshot_from_2025-06-07_11-16-33.png
Attachment 6: Screenshot_from_2025-06-07_11-16-55.png
Screenshot_from_2025-06-07_11-16-55.png
Attachment 7: Screenshot_from_2025-06-07_11-10-07.png
Screenshot_from_2025-06-07_11-10-07.png
Attachment 8: Screenshot_from_2025-06-07_12-40-18.png
Screenshot_from_2025-06-07_12-40-18.png
Attachment 9: Screenshot_from_2025-06-07_12-40-53.png
Screenshot_from_2025-06-07_12-40-53.png
Attachment 10: Screenshot_from_2025-06-07_12-41-35.png
Screenshot_from_2025-06-07_12-41-35.png
Attachment 11: Screenshot_from_2025-06-07_12-42-29.png
Screenshot_from_2025-06-07_12-42-29.png
Attachment 12: Screenshot_from_2025-06-07_12-42-37.png
Screenshot_from_2025-06-07_12-42-37.png
Attachment 13: Screenshot_from_2025-06-07_12-42-47.png
Screenshot_from_2025-06-07_12-42-47.png
Attachment 14: Screenshot_from_2025-06-07_12-47-40.png
Screenshot_from_2025-06-07_12-47-40.png
Attachment 15: Screenshot_from_2025-06-07_12-49-36.png
Screenshot_from_2025-06-07_12-49-36.png
Attachment 16: Screenshot_from_2025-06-07_14-04-20.png
Screenshot_from_2025-06-07_14-04-20.png
Attachment 17: Screenshot_from_2025-06-07_14-04-42.png
Screenshot_from_2025-06-07_14-04-42.png
Attachment 18: Screenshot_from_2025-06-07_14-07-18.png
Screenshot_from_2025-06-07_14-07-18.png
Attachment 19: Screenshot_from_2025-06-07_15-01-42.png
Screenshot_from_2025-06-07_15-01-42.png
Attachment 20: Screenshot_from_2025-06-07_15-02-18.png
Screenshot_from_2025-06-07_15-02-18.png
Attachment 21: Screenshot_from_2025-06-07_15-02-43.png
Screenshot_from_2025-06-07_15-02-43.png
Attachment 22: Screenshot_from_2025-06-07_15-03-29.png
Screenshot_from_2025-06-07_15-03-29.png
Attachment 23: Screenshot_from_2025-06-07_15-07-28.png
Screenshot_from_2025-06-07_15-07-28.png
Attachment 24: Screenshot_from_2025-06-07_19-14-13.png
Screenshot_from_2025-06-07_19-14-13.png
Attachment 25: Screenshot_from_2025-06-07_19-14-48.png
Screenshot_from_2025-06-07_19-14-48.png
Attachment 26: Screenshot_from_2025-06-07_19-15-44.png
Screenshot_from_2025-06-07_19-15-44.png
Attachment 27: Screenshot_from_2025-06-07_19-16-50.png
Screenshot_from_2025-06-07_19-16-50.png
Attachment 28: Screenshot_from_2025-06-07_19-19-04.png
Screenshot_from_2025-06-07_19-19-04.png
Attachment 29: Screenshot_from_2025-06-07_19-19-09.png
Screenshot_from_2025-06-07_19-19-09.png
Attachment 30: Screenshot_from_2025-06-07_19-19-19.png
Screenshot_from_2025-06-07_19-19-19.png
Attachment 31: Screenshot_2025-06-07_20-30-44.png
Screenshot_2025-06-07_20-30-44.png
Attachment 32: Screenshot_2025-06-07_20-58-09.png
Screenshot_2025-06-07_20-58-09.png
Attachment 33: Screenshot_from_2025-06-07_21-24-25.png
Screenshot_from_2025-06-07_21-24-25.png
Attachment 34: Screenshot_from_2025-06-07_21-26-10.png
Screenshot_from_2025-06-07_21-26-10.png
Attachment 35: Screenshot_from_2025-06-07_21-26-15.png
Screenshot_from_2025-06-07_21-26-15.png
Attachment 36: Screenshot_from_2025-06-07_21-27-35.png
Screenshot_from_2025-06-07_21-27-35.png
Attachment 37: Screenshot_from_2025-06-07_21-27-42.png
Screenshot_from_2025-06-07_21-27-42.png
Attachment 38: Screenshot_from_2025-06-07_21-28-23.png
Screenshot_from_2025-06-07_21-28-23.png
Attachment 39: Screenshot_from_2025-06-07_22-28-31.png
Screenshot_from_2025-06-07_22-28-31.png
Attachment 40: Screenshot_from_2025-06-07_22-29-31.png
Screenshot_from_2025-06-07_22-29-31.png
Attachment 41: Screenshot_from_2025-06-07_22-29-48.png
Screenshot_from_2025-06-07_22-29-48.png
Attachment 42: Screenshot_from_2025-06-07_22-30-49.png
Screenshot_from_2025-06-07_22-30-49.png
Attachment 43: Screenshot_from_2025-06-07_22-31-25.png
Screenshot_from_2025-06-07_22-31-25.png
Attachment 44: Screenshot_from_2025-06-07_22-31-33.png
Screenshot_from_2025-06-07_22-31-33.png
Attachment 45: Screenshot_from_2025-06-08_07-09-51.png
Screenshot_from_2025-06-08_07-09-51.png
Attachment 46: Screenshot_from_2025-06-08_07-10-07.png
Screenshot_from_2025-06-08_07-10-07.png
Attachment 47: Screenshot_from_2025-06-08_07-11-13.png
Screenshot_from_2025-06-08_07-11-13.png
Attachment 48: Screenshot_from_2025-06-08_07-12-40.png
Screenshot_from_2025-06-08_07-12-40.png
Attachment 49: Screenshot_from_2025-06-08_07-27-03.png
Screenshot_from_2025-06-08_07-27-03.png
Attachment 50: Screenshot_from_2025-06-08_10-06-06.png
Screenshot_from_2025-06-08_10-06-06.png
  723   Fri Jun 6 17:27:03 2025 AY, ST, JM, EM, DBRun 53

Run 53 started at 18:25 (started already earlier but since then data is taken). Everything is running again. We started at 3E12 at/cm2.
Side comment: Gleb told us 1h ago that he will switch to coasted mode and call us back as soon as he is done. So far he didn't call back and he is not in the office... So we decided to start data taking already. But: the software says that coasting mode is already on, but the graph looks like bunched mode... So we started data taking and will update this situation as soon as Gleb called us. -> Update: Okay, we were the entire time in coasting mode

D+ beam deuterium target. D+ beam energy is now 0.25 MeV/u. (beam calc calculation is 0.2570 MeV/u).

All four detectors positioned fully out and stationary.

Left potentiometer = -2.76mm (corresponds to x = -2500, which is the more trustworthy value)

Right potentiometer = 1.77 mm (corresponds to y = 2500, which is the more trustworthy value)

Ring cycle time is 5 seconds.

Cooler set value = 146.25 V, 2 mA. Read value = 146.2 V, 0.00179 mA

All leakage currents look good.

Beam position. Vertical =0.0mm , Horizontal=-2.0 mm

Data rate ~ 812kB/s

 

After starting the run we struggled because we did not know whether we can trust the beam overlap test after going from bunched mode to coasting mode. We therefore stopped data taking for approx. 1h to to this test once again in coasting mode, but did it via the Rutherford measurement. The result showed that the optimum is exactly the same: x=-2mm (see attachment 1).

However, we are strugglign with the areal density... It quickly dropped to 2E10 and slowly recovered in a more or less oscillating way (see attachment 2; the first four columns are the integrals of the four detectors, the next column is the sum of them. Afterwards its the average areal density and the last column is the total integral divided by the average areal density).
 

19:29: Storage turned back on
Left potentiometer now = -2.79mm (corresponds to x = -5000, which is the more trustworthy value)

19:53: Areal density dropped to 2E11, so we moved the left potentiometer once from x=-5000 to x=-4500. Sudden increase in pressure, but slow decrease again.
19:59: Left potentiometer once from x=-4500 to x=-4000. Seems to be the right direction... But we do it slowly...

20:19: Full check. Everything is okay, but areal density is still... aweful... Most of the time its between 1E11 and 1E12 at/cm2
20:20: Left potentiometer once from x=-4000 to x=-3500. Seems again to be the right direction... Sudden increase from 7E11 to 2E12
20:40: Everything stabilized at a higher average areal density. We are now fluctuating between 1E12 to 5E12.

20:55: Left potentiometer once from x=-3500 to x=-3000. Seems again to be the right direction... Sudden increase from 9E11 to 3E12

21:56: Full check. Everything is okay and areal density finally stabilized. Its now since 1h between 3E12 and 5E12.
22:47: Full check. Everything is meraviglioso!

23:20 

Checks OK

Target density remains stable around 5e12 at/cm2. Count rate has improved to around 200 c/h for 3He, 150 c/h for 3H and 80 c/h for 1H.

02:20

All checks OK

Count rates look pretty good, target density has maintained mid-high e12 lavel over last 3 hours.

peaks in spectra very clear

05:00

Checks OK

Now have several thousand counts total. Target level in high e12 level.

07.06.2025

07:00

Full checks! Target level 1-6E12 [1/cm2]

 

10:00

The x server on the CARME-gsi computer appears to be struggling and is not able to be checked. Needs a reboot. 

We still have ssh access and are able to access the DAQ on another CRYRING machine machine via 192.168.207.143:8015. DAQ is still running. We will try to reboot the x server without disturbing the DAQ. 

oopsie whole carme-gsi rebooted not just the x server. Restarting DAQ

Attachment 1: 2025_06_06_19_41_30_Greenshot.png
2025_06_06_19_41_30_Greenshot.png
Attachment 2: Screenshot_2025-06-06_19-46-11.png
Screenshot_2025-06-06_19-46-11.png
Attachment 3: Screenshot_from_2025-06-06_21-58-00.png
Screenshot_from_2025-06-06_21-58-00.png
Attachment 4: Screenshot_from_2025-06-06_21-58-13.png
Screenshot_from_2025-06-06_21-58-13.png
Attachment 5: Screenshot_from_2025-06-06_21-58-43.png
Screenshot_from_2025-06-06_21-58-43.png
Attachment 6: Screenshot_from_2025-06-07_02-26-21.png
Screenshot_from_2025-06-07_02-26-21.png
Attachment 7: Screenshot_from_2025-06-07_07-17-01.png
Screenshot_from_2025-06-07_07-17-01.png
Attachment 8: Screenshot_from_2025-06-07_07-15-30.png
Screenshot_from_2025-06-07_07-15-30.png
Attachment 9: Screenshot_from_2025-06-07_07-14-11.png
Screenshot_from_2025-06-07_07-14-11.png
Attachment 10: Screenshot_from_2025-06-07_07-12-18.png
Screenshot_from_2025-06-07_07-12-18.png
Attachment 11: beamCalc-0.25.png
beamCalc-0.25.png
  722   Fri Jun 6 16:49:59 2025 CB, AY, STBeam overlap results for 200keV and 250keV

For 200keV the optimal horizontal position is -5mm.
For 250keV the optimal horizontal position is -2mm.

Both values are already given to the operators, so tomorrow we can just switch to 200keV and go on :)

Nonetheless, we now went to 250keV and start measuring.

Attachment 1: 2025_06_06_17_46_56_Greenshot.png
2025_06_06_17_46_56_Greenshot.png
  721   Thu Jun 5 16:54:18 2025 EW, ST, CBRun 52 - 0.3 MeV/A

Run 52 started at 17:53. Everything is running again. We started at 3E12 at/cm2

D+ beam deuterium target. D+ beam energy is now 0.3 MeV/u. (beam calc calculation is 0.3067 MeV/u).

All four detectors positioned fully out and stationary and also detector #4 is running again.

Left potentiometer = -3.03mm (corresponds to x = 0, which is the more trustworthy value)

Right potentiometer = 1.82 mm (corresponds to y = -1000, which is the more trustworthy value)

Ring cycle time is 5 seconds.

Cooler set value = 177.9 V, 4 mA. Read value = 177.65 V, 3.53 mA

All leakage currents look good.

Beam position. Vertical =0.0mm , Horizontal=1.0 mm

Data rate ~ 947kB/s

18:09: System check - everything perfect!

Leakage current attachment 1

Rates attachment 2

19:50: System check - everything marvelous!

22:05: System check - everything excellent!

22:50: System check - everything fabulous!

current spectra attachment 3

 

02:05

Temperatures OK

System wide checks OK

Statistics OK

Merger OK

Leakage current OK

Ring vacuum OK

 

04:52

Temps OK

System wide checks OK

Stats OK - attached

Merger OK - attached

Leakage OK - attached

Histos look good - attached

 

07:10

Temps OK

System wide checks OK

Stats OK - attached

Merger OK - attached

Leakage OK - attached

Histos look good - attached

 

11:31

All checks OK

13:54

All checks OK

Rates attached

 

 

 

 

Attachment 1: Screenshot_from_2025-06-05_18-10-19.png
Screenshot_from_2025-06-05_18-10-19.png
Attachment 2: Screenshot_from_2025-06-05_18-10-12.png
Screenshot_from_2025-06-05_18-10-12.png
Attachment 3: Screenshot_from_2025-06-05_22-52-55.png
Screenshot_from_2025-06-05_22-52-55.png
Attachment 4: Screenshot_from_2025-06-06_04-51-26.png
Screenshot_from_2025-06-06_04-51-26.png
Attachment 5: Screenshot_from_2025-06-06_04-51-37.png
Screenshot_from_2025-06-06_04-51-37.png
Attachment 6: Screenshot_from_2025-06-06_04-51-48.png
Screenshot_from_2025-06-06_04-51-48.png
Attachment 7: Screenshot_from_2025-06-06_04-52-07.png
Screenshot_from_2025-06-06_04-52-07.png
Attachment 8: Screenshot_from_2025-06-06_07-18-22.png
Screenshot_from_2025-06-06_07-18-22.png
Attachment 9: Screenshot_from_2025-06-06_07-17-45.png
Screenshot_from_2025-06-06_07-17-45.png
Attachment 10: Screenshot_from_2025-06-06_07-17-21.png
Screenshot_from_2025-06-06_07-17-21.png
Attachment 11: Screenshot_from_2025-06-06_07-16-45.png
Screenshot_from_2025-06-06_07-16-45.png
Attachment 12: Screenshot_from_2025-06-06_13-55-31.png
Screenshot_from_2025-06-06_13-55-31.png
Attachment 13: Screenshot_from_2025-06-06_13-59-12.png
Screenshot_from_2025-06-06_13-59-12.png
Attachment 14: beamCalc-0.3.png
beamCalc-0.3.png
  720   Thu Jun 5 08:50:00 2025 JM, CF, CB, EWRun 51 - background

09:49

DSSD4 has been biased to 100V in steps of 5V from 60V. It was previously stable overnight at 60V - attachment 1

Will keep an eye on it but looks OK at the moment. Current bias and leakage current - attachment 2

I have reset DSSD4 FEES (13,14,15,16) to the LEC and lowered the thresholds to be in line with other FEES.

Current slow comparator threshold for all FEES is 0x14 (200 keV).

Attachment 3 shows current ADC data items statistics.

Started a new run 51 with no beam and no target. Access to the cave is allowed.

Current data rate is 800 kB/sec.

All statistics and histograms in aida zero'd

12:57

Temps OK

Rates OK - attached

Merger OK

System wide checks

 

14:58

Temps - OK

Rates - OK

Merger - OK

 

15:00

Beam tuning to start soon. No storage.

Attachment 1: Screenshot_from_2025-06-05_09-38-05.png
Screenshot_from_2025-06-05_09-38-05.png
Attachment 2: Screenshot_from_2025-06-05_09-46-08.png
Screenshot_from_2025-06-05_09-46-08.png
Attachment 3: Screenshot_from_2025-06-05_09-51-43.png
Screenshot_from_2025-06-05_09-51-43.png
  719   Wed Jun 4 10:44:36 2025 EW, CB, JM, OF, JG, STRun 50 - 0.35 MeV/A

11:30

Run 50 started.

D+ beam deuterium target. D+ beam energy is 0.35 MeV/u. (beam calc calculation is 0.3577 MeV/u)

Target density is about 4e12 at/cm2

All four detectors positioned fully out and stationary.

Left potentiometer = 18.86 mm

Right potentiometer = 19.8 mm

Ring cycle time is 10 seconds.

Target T1=30.1K, T2=31.8K.

Nozzle position unchanged since end of run 48

YR09-Etarget pressure = 2.05e-10 mbar.

Cooler  Read value = 207.1 V, 4.39 mA

 

All leakage currents look good (except dssd4, currently stable at V=60)

Beam position. Vertical =0.0mm , Horizontal=0.5 mm

Data rate ~ 172kB/s

Slow comparator threshold to 0x14 (200 keV) since we decreased beam energy.

13:00

All checks OK

We were seeing some rubbish obscuring the 3He peak. JM remembered setting the EYEX cut very high (3 MeV) on the online plot, Now it is reduced to 500 keV

See attachments of histograms before restarting code.
 

16:32 all checks okay
18:19 all checks okay
20:20 all checks okay
21:45 all checks okay

01:00 all checks okay

04:00 all checks okay

 

7:00

Run stopped - cave to be opened. Beam looks on when run was stopped.

Temps OK - attach

Leakage current stable - attach

Rates OK - attach

Merger OK - attach

Final spectra attached

09:30

Gas target shutter valve closed to save ring vacuum.

All checks OK

Stats at htis threshold with no beam attached.

Biased DSSD4 in steps of 5V from 60V since it has been stable since yesterday,

Was able to bias to 100V with no issues.

Attachment 1: Screenshot_from_2025-06-04_12-07-43.png
Screenshot_from_2025-06-04_12-07-43.png
Attachment 2: Screenshot_from_2025-06-04_12-07-19.png
Screenshot_from_2025-06-04_12-07-19.png
Attachment 3: Screenshot_from_2025-06-04_12-57-24.png
Screenshot_from_2025-06-04_12-57-24.png
Attachment 4: Screenshot_from_2025-06-04_12-57-37.png
Screenshot_from_2025-06-04_12-57-37.png
Attachment 5: Screenshot_from_2025-06-05_07-07-37.png
Screenshot_from_2025-06-05_07-07-37.png
Attachment 6: Screenshot_from_2025-06-05_07-07-30.png
Screenshot_from_2025-06-05_07-07-30.png
Attachment 7: Screenshot_from_2025-06-05_07-02-19.png
Screenshot_from_2025-06-05_07-02-19.png
Attachment 8: Screenshot_from_2025-06-05_07-03-40.png
Screenshot_from_2025-06-05_07-03-40.png
Attachment 9: Screenshot_from_2025-06-05_09-34-41_1.png
Screenshot_from_2025-06-05_09-34-41_1.png
Attachment 10: Screenshot_from_2025-06-05_09-38-05.png
Screenshot_from_2025-06-05_09-38-05.png
Attachment 11: Screenshot_from_2025-06-05_09-46-08.png
Screenshot_from_2025-06-05_09-46-08.png
Attachment 12: Screenshot_from_2025-06-05_13-02-08.png
Screenshot_from_2025-06-05_13-02-08.png
Attachment 13: beamCalc-0.35.png
beamCalc-0.35.png
  718   Wed Jun 4 00:10:39 2025 JMRun 49 - beam only

01:10

Gas target turned off via the shutter valve.

Run 49 started beam only background run.

Slow comparator threshold lowered to 0x14 (200 keV) in preparation for the energy change tomorrow morning.

Other than that no changes to the DAQ or to the experimental setup.

D+ beam remains at 0.5 MeV nominally. All four detectors positioned fully out and stationary,

I tried to re-bias detector #4 however was not able to bias past 70V. DSSD4 remains off and set to the MEC (0x1). (see attchement 1)

Ring cycle is every 10 seconds. Beam position unchanged, Beam position. Vertical =0.0mm , Horizontal=0.0 mm

Ring vacuum is now 3.16e-11 mbar. Before switching off the gas target density was 2e12 at/cm2

Data rate is 89 kB/sec

ADC data items stats remain largely unchanged at new threshold. (see attachment 2)

Aida01 has increased to several thousand. Tried check ASIC control on asic #4 of aida01->no effect on higher rates.

01:20

All checks OK.

After 20 minutes of running see energy histogram and xy histogram (attachments 3,4)

Interesting xy pattern (hydrogen? Injection noise?)

Note put aida13-aida16 slow comparator threshold back to 0x24 as we had some rates coming through (~1kHz) which is now back to zero. If we revive the detector in the morning, the thresholds should be lowered in line with other detectors.

03:00

All checks OK

04:30

All checks ok

7:20

All checks ok

Run stopped, final energy histogram and xy histogram attached.

Attachment 1: Screenshot_from_2025-06-04_01-18-10.png
Screenshot_from_2025-06-04_01-18-10.png
Attachment 2: Screenshot_from_2025-06-04_01-18-59.png
Screenshot_from_2025-06-04_01-18-59.png
Attachment 3: Screenshot_from_2025-06-04_01-20-47.png
Screenshot_from_2025-06-04_01-20-47.png
Attachment 4: Screenshot_from_2025-06-04_01-20-56.png
Screenshot_from_2025-06-04_01-20-56.png
Attachment 5: Screenshot_from_2025-06-04_07-22-01.png
Screenshot_from_2025-06-04_07-22-01.png
Attachment 6: Screenshot_from_2025-06-04_07-21-56.png
Screenshot_from_2025-06-04_07-21-56.png
  717   Tue Jun 3 14:54:29 2025 PMH, STRun 48

Run 48 started. Everything is running again. We started at 3E13 at/cm2

D+ beam deuterium target. D+ beam energy is still 0.5 MeV/u. (beam calc calculation is 0.5143 MeV/u). See previous entry for beamcalc.

All four detectors positioned fully out and stationary, however we switched off detector #4.

Left potentiometer = -3.03mm (corresponds to x = 0, which is the more trustworthy value)

Right potentiometer = 1.74 mm (corresponds to y = -500, which is the more trustworthy value)

Ring cycle time is 10 seconds.

Cooler set value = 297.2 V, 8 mA. Read value = 296.7 V, 7 mA

All leakage currents look good (except dssd4, currently turned off)

Beam position. Vertical =0.0mm , Horizontal=0.0 mm

Data rate ~ 108kB/s

 

16:40  Change Gasjet target Y=-500 to Y=0
17:37  Change Gasjet target Y= 0 to Y= 500

20:30 We now take some time to optimize the areal density
20:30  Change Gasjet target Y= 500 to Y= 1000 (resulting in 2E12)
20:32  Change Gasjet target Y= 1000 to Y= 500 (resulting in 4E12)
20:33  Change Gasjet target Y= 500 to Y= 0 (resulting in 7E12)
20:35  Change Gasjet target Y= 0 to Y= -500 (resulting in 1.1E13)
20:37  Change Gasjet target Y= -500 to Y= -1000 (resulting in 1.8E13, still increasing and amazing for almost 1h)

21:48: Since approximately 20min its just going down systematically from a stable 2E13 to 6E12...) Lets try another round of optimization...
21:48  Change Gasjet target Y= -1000 to Y= -1500 (starting at 7E12 and resulting in 2E12)
21:49  Change Gasjet target Y= -1500 to Y= -1000 (starting at 2E12 and resulting in 6E12)
21:50  Change Gasjet target Y= -1000 to Y= -500 (starting at 6E12 and resulting in 9E12)
21:51  Change Gasjet target Y= -500 to Y= 0 (starting at 9E12 and resulting in 1.3E13 but dropping back to 7E12 within 3min)

21:54  Change Gasjet target. Leaving Y at 0, but moving X from 0 to -500 (starting at 6E12 and resulting in 3E12)
21:55  Change Gasjet target. Leaving Y at 0, but moving X from -500 to 0 (starting at 3E12 and resulting in 4E12)
21:56  Change Gasjet target. Leaving Y at 0, but moving X from 0 to +500 (starting at 4E12 and resulting in 6E12)

21:59  Change Gasjet target. Leaving X at 500, but moving Y from 0 to +500 (starting at 5E12 and resulting in 7E11...)
22:00  Change Gasjet target. Leaving X at 500, but moving Y from +500 to 0 (starting at 7E11 and resulting in 2E12)

Side comment: It seems no matter what we do, that there is a general decrease within the last 30min...

22:00  Change Gasjet target. Leaving X at 500, but moving Y from +500 to 0 (starting at 7E11 and resulting in 2E12)

Okay... No matter what we do... Its still decreasing...

22:03  Change Gasjet target. Leaving Y at 0, but moving X from 500  to 0 (starting at 1E12 and resulting in 2E12)
22:06  Change Gasjet target. Leaving X at 0, but moving Y = 0 from  to -500 (starting at 2E12 and resulting in 5E12
22:09  Change Gasjet target. Leaving X at 0, but moving Y = -500 from  to -1000 (starting at 5E12 and just randomly fluctuating within the next 15min... and more decreasing than increasing...)

22:25  Change Gasjet target. Leaving X at 0, but moving Y = -1000 from  to -1500 (starting at 1E12 and resulting in 2E12... )
22:31  Change Gasjet target. Leaving X at 0, but moving Y = -1500 from  to -2000 (starting at 1E12 and resulting in 8E11... ) Whaaaat????
22:32  Change Gasjet target. Leaving X at 0, but moving Y = -2000 from  to -1000 (starting at 9E11 and resulting in 4E12... ) Okaaaay lets wait and see...

01:00

Run stopped.

Final statistics very high. Over 10,000 counts for 3He and 3H, Over 5000 counts for proton.

Final energy histogram and xy plot on final attachments

Attachment 1: Screenshot_from_2025-06-03_16-57-29.png
Screenshot_from_2025-06-03_16-57-29.png
Attachment 2: Screenshot_from_2025-06-03_16-57-46.png
Screenshot_from_2025-06-03_16-57-46.png
Attachment 3: Screenshot_from_2025-06-03_16-57-55.png
Screenshot_from_2025-06-03_16-57-55.png
Attachment 4: Screenshot_from_2025-06-03_16-58-15.png
Screenshot_from_2025-06-03_16-58-15.png
Attachment 5: Screenshot_from_2025-06-04_00-54-02.png
Screenshot_from_2025-06-04_00-54-02.png
Attachment 6: Screenshot_from_2025-06-04_00-57-33.png
Screenshot_from_2025-06-04_00-57-33.png
  716   Tue Jun 3 11:16:07 2025 EW, EM, JMRun 47, beam only background

Run 47 started. Target cooling down after water cooler failure, beam only background

D+ beam deuterium target. D+ beam energy is 0.5 MeV/u. (beam calc calculation is 0.5143 MeV/u). See previous entry for beamcalc.

All four detectors positioned fully out and stationary.

Left potentiometer = 18.86 mm

Right potentiometer = 19.8 mm

Ring cycle time is 10 seconds.

Cooler set value = 297.2 V, 8 mA. Read value = 296.477 V, 6.97 mA

All leakage currents look good (except dssd4, currently stable at V=40)

Beam position. Vertical =0.0mm , Horizontal=0.0 mm

Data rate ~ 25kB/s

13:00

Run stopped to allow tuning of parameters for 0.35 MeV/u

Attachment 1: Screenshot_from_2025-06-03_12-20-25.png
Screenshot_from_2025-06-03_12-20-25.png
Attachment 2: Screenshot_from_2025-06-03_12-19-36.png
Screenshot_from_2025-06-03_12-19-36.png
  715   Mon Jun 2 20:03:43 2025 JG, AG, CB, EW, EMRun 45/46, 0.5 MeV/A

Run 45 started.

D+ beam deuterium target. D+ beam energy is 0.5 MeV/u. (beam calc calculation is 0.5143 MeV/u). See attachment for beamCalc

Target density is about 2e12 at/cm2

All four detectors positioned fully out and stationary.

Left potentiometer = 18.86 mm

Right potentiometer = 19.8 mm

Ring cycle time is 10 seconds. (optimized with calculation https://web-docs.gsi.de/~stoe_exp/web_programs/optimizer/index.php)

Target T1=29.9K, T2=31.6K.

Nozzle position was not touched since run 40 (not clear how to double check)

YR09-Etarget pressure = 1.33e-10 mbar.

Cooler set value = 297.2 V, 8 mA. Read value = 296.477 V, 6.97 mA

All leakage currents look good

Beam position. Vertical =0.0mm , Horizontal=0.0 mm

Beam intensity ~6e8 on bunched beam, unknown on coasting.

Data rate ~ 422kB/s

Attachements:

- beam cycle time optimzer results (https://web-docs.gsi.de/~stoe_exp/web_programs/optimizer/index.php)

- full paramodi export

- leackage current plot

- beam target overlap measurement

22:55 - all checks okay

 

23:00-00:00 atppc025 crashed, taking down the proxy to the internet and anydesk. Recovered remotely moving proxy to atppc022. Let Michael know, then move back to atppc025 (contact CB or JM). 22 seems to work fine, no rush

Tried to move InteractiveMotion.x screen on carme-gsi since it was on atppc025, which froze. Cannot ssh to carmemotorspi from carme-gsi, except if already ssh in. Must be a proxy issue. Motors are stationary, will fix tomorrow morning.

 

00:00 Found DSSD 4 unbiased. Strong spike in leakage current just before midnight. Correlated with atppc025 freezing? Maybe a power fluctuation? No storage. Attempting to re-bias. Beam and target appear to be fine. Merger appears to be fine.

Moved right arm slightly with servo. No effect on bias.

V = 50 -> 14 uA oscillating

V = 40 -> 7 uA oscillating

V = 30 -> 4uA stable

V = 35 -> 6.5 uA unstable

Cannot bias above 35 V. Will leave DSSD4 condition overnight. LEC-> MEC on DSSD 4.

Run 45 final histograms save. Started new run with same conditions except DSSD 4.

 

04:00

Temps OK

Rates OK (no dssd 4)

Merger OK (38 kb/s)

Histo attached

Could increase bias in 4 to around 70 V before the current got unstable. Ramped down to 30 V again. Need to try significantly slower, 10 V an hour or so -> morning shift.

 

07:10

Since 4:35 no electron cooler

7:15 electron cooler back

Temps OK

Rates OK (except dssd4)

Merger OK (38kB/s)

Histo attached

 

8:35

V=40  -> 8uA oscillating 

V=50 -> 8uA oscillating

V=40 -> 7uA oscillating 

V=30 -> 3.9uA oscillating 

 

Performed a V-I curve test of dssd 4

Turned bias OFF, went to 5V and then bias ON again 

V=5 -> 1.245uA stable 

V=10 -> 1.820uA stable 

V=15 -> 2.300uA stable 

V=20 -> 2.720uA stable 

V=25 -> 3.105uA stable 

V=30 -> 3.460uA stable 

V=35 -> 3.795uA stable 

V=40 -> 4.115uA stable 

V=45 -> 4.425uA stable

V=50 -> unstable 

Turned bias OFF again, then back on after few minutes

V=5 -> 1.240uA stable 

V=10 -> 1.820uA stable 

V=15 -> 2.295uA stable 

V=20 -> 2.715uA stable 

V=25 -> 3.100uA stable 

V=30 -> 3.450uA stable 

V=35 -> 3.790uA stable 

V=40 -> 4.110uA stable 

Leaving at 40 for a couple of hours 

 

10:55

Tried 0x1 for medium/low energy selection for aida 13,14,15,16 but rate was roughly 250000 across all 

11:30

Water coolers have gone down, moinitoring the FEE temperatures, stopped the run 

11:48

Water cooling back

 

Attachment 1: Screenshot_2025-06-02_20-58-17.png
Screenshot_2025-06-02_20-58-17.png
Attachment 2: 2025-06-02_20-56-30_YRT1_CRYRING_2025_D_Bruno_500keV.C1.txt
# Version: 1.3
# Comment: ParamModi Export at 2025-06-02T20:57:02.504235715+02:00[Europe/Berlin]
# Context: YRT1_CRYRING_2025_D_Bruno.C1
# Columns: PARAMETERNAME,VALUE,UNIT,BEAMPROCESSPURPOSE
LOGICAL.SOURCE_HV/U,20000.0,V,TRANSFER_INJECTION
YRT1_TO_YRT1LQ1_BEAM/BEAM_MODE_S,NO_BEAM,,TRANSFER_INJECTION
YRT1_TO_YRT1LQ1_BEAM/PARTICLE_SYMBOL_S,2H,,TRANSFER_INJECTION
YRT1_TO_YRT1LQ1_BEAM/Q_S,1,,TRANSFER_INJECTION
YRT1_TO_YRT1LQ1_TIMEPARAM/TIMING_BI_DELAY_S,2.0E-4,s,TRANSFER_INJECTION
YRT1_TO_YRT1LQ1_TIMEPARAM/TIMING_BI_WINDOW_S,0.002,s,TRANSFER_INJECTION
YRT1_TO_YRT1LQ1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,BEAMOUT_INIT
YRT1_TO_YRT1LQ1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,TRANSFER_INJECTION
YRT1_TO_YRT1LQ1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,BEAMOUT_RESET
YRT1_TO_YRT1LQ1_TIMEPARAM/TIMING_SOURCE_PULSE_LENGTH_S,0.0024,s,TRANSFER_INJECTION
YRT1_TO_YRT1LQ1_TIMEPARAM/T_BP_LENGTH_S,0.12,s,BEAMOUT_INIT
YRT1_TO_YRT1LQ1_TIMEPARAM/T_BP_LENGTH_S,0.11,s,TRANSFER_INJECTION
YRT1_TO_YRT1LQ1_TIMEPARAM/T_BP_LENGTH_S,0.0,s,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/T_BP_LENGTH_S,0.0,s,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/T_BP_LENGTH_S,0.0,s,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/T_BP_LENGTH_S,0.0,s,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/T_BP_LENGTH_S,0.0,s,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/T_BP_LENGTH_S,0.0,s,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/T_BP_LENGTH_S,0.0,s,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/T_BP_LENGTH_S,0.0,s,STORAGE_MANIP_PERFORM
YRT1_TO_YRT1LQ1_TIMEPARAM/T_BP_LENGTH_S,---,s,STORAGE_MANIP_WAIT
YRT1_TO_YRT1LQ1_TIMEPARAM/T_BP_LENGTH_S,---,s,STORAGE_MANIP_RESTART_FGS
YRT1_TO_YRT1LQ1_TIMEPARAM/T_BP_LENGTH_S,---,s,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/T_BP_LENGTH_S,---,s,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/T_BP_LENGTH_S,0.12,s,BEAMOUT_RESET
YRT1LQ1_TO_YRT1LC1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,BEAMOUT_INIT
YRT1LQ1_TO_YRT1LC1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,TRANSFER_INJECTION
YRT1LQ1_TO_YRT1LC1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LQ1_TO_YRT1LC1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LQ1_TO_YRT1LC1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LQ1_TO_YRT1LC1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LQ1_TO_YRT1LC1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LQ1_TO_YRT1LC1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LQ1_TO_YRT1LC1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LQ1_TO_YRT1LC1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LQ1_TO_YRT1LC1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LQ1_TO_YRT1LC1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,BEAMOUT_RESET
LOGICAL.YRT1KH1/HKICK,0.001,rad,TRANSFER_INJECTION
LOGICAL.YRT1KV1/VKICK,0.005,rad,TRANSFER_INJECTION
LOGICAL.YRT1LC1_V/ACTIVE_S,true,,TRANSFER_INJECTION
LOGICAL.YRT1LC1_V/THIGH_S,1.0E-4,s,TRANSFER_INJECTION
LOGICAL.YRT1LC1_V/TOFFSET_S,5.0E-4,s,TRANSFER_INJECTION
LOGICAL.YRT1LC1_V/U,2000.0,V,TRANSFER_INJECTION
LOGICAL.YRT1LD21/ASYMMETRY_KNOB_S,0.0,,TRANSFER_INJECTION
LOGICAL.YRT1LD21/HOR_STEER_KNOB_S,0.0,,TRANSFER_INJECTION
LOGICAL.YRT1LD21/U_QUAD,-350.0,V,TRANSFER_INJECTION
LOGICAL.YRT1LD21/VER_STEER_KNOB_S,0.05,,TRANSFER_INJECTION
LOGICAL.YRT1LD22/ASYMMETRY_KNOB_S,0.0,,TRANSFER_INJECTION
LOGICAL.YRT1LD22/U_QUAD,350.0,V,TRANSFER_INJECTION
LOGICAL.YRT1LT31/ASYMMETRY_KNOB_S,0.0,,TRANSFER_INJECTION
LOGICAL.YRT1LT31/U_QUAD,680.0,V,TRANSFER_INJECTION
LOGICAL.YRT1LT32/ASYMMETRY_KNOB_S,0.0,,TRANSFER_INJECTION
LOGICAL.YRT1LT32/U_QUAD,-680.0,V,TRANSFER_INJECTION
LOGICAL.YRT1LT33/ASYMMETRY_KNOB_S,0.0,,TRANSFER_INJECTION
LOGICAL.YRT1LT33/U_QUAD,680.0,V,TRANSFER_INJECTION
LOGICAL.YRT1LT41/ASYMMETRY_KNOB_S,0.0,,TRANSFER_INJECTION
LOGICAL.YRT1LT41/U_QUAD,-1520.0,V,TRANSFER_INJECTION
LOGICAL.YRT1LT42/ASYMMETRY_KNOB_S,0.0,,TRANSFER_INJECTION
LOGICAL.YRT1LT42/U_QUAD,2080.0,V,TRANSFER_INJECTION
LOGICAL.YRT1LT43/ASYMMETRY_KNOB_S,0.0,,TRANSFER_INJECTION
LOGICAL.YRT1LT43/U_QUAD,-1550.0,V,TRANSFER_INJECTION
LOGICAL.YRT1MH1/HKICK,0.006,rad,TRANSFER_INJECTION
LOGICAL.YRT1MH1/K0L_THEO,-1.57079,rad,TRANSFER_INJECTION
YRT1LC1_TO_YRT1MH2_TIMEPARAM/TIMING_BI_DELAY_S,1.0E-6,s,TRANSFER_INJECTION
YRT1LC1_TO_YRT1MH2_TIMEPARAM/TIMING_BI_WINDOW_S,1.0E-6,s,TRANSFER_INJECTION
YRT1LC1_TO_YRT1MH2_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,BEAMOUT_INIT
YRT1LC1_TO_YRT1MH2_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,TRANSFER_INJECTION
YRT1LC1_TO_YRT1MH2_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LC1_TO_YRT1MH2_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LC1_TO_YRT1MH2_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LC1_TO_YRT1MH2_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LC1_TO_YRT1MH2_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LC1_TO_YRT1MH2_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LC1_TO_YRT1MH2_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LC1_TO_YRT1MH2_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LC1_TO_YRT1MH2_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LC1_TO_YRT1MH2_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,BEAMOUT_RESET
YRT1LC1_TO_YRT1MH2_TIMEPARAM/TIMING_RF_LEAD_S,4.0E-4,s,TRANSFER_INJECTION
LOGICAL.YRT1BR1/ACT_ON_BEAM_S,true,,TRANSFER_INJECTION
LOGICAL.YRT1BR1/THIGH_S,0.0025,s,TRANSFER_INJECTION
LOGICAL.YRT1BR1/URF0_S,0.14,V,TRANSFER_INJECTION
LOGICAL.YRT1BR1/U_PAUSE_PULSE_AMPL_S,4.76,V,TRANSFER_INJECTION
LOGICAL.YRT1KH2/HKICK,-2.0E-4,rad,TRANSFER_INJECTION
LOGICAL.YRT1KV2/VKICK,0.0,rad,TRANSFER_INJECTION
LOGICAL.YRT1LD51/ASYMMETRY_KNOB_S,0.0,,TRANSFER_INJECTION
LOGICAL.YRT1LD51/U_QUAD,12000.0,V,TRANSFER_INJECTION
LOGICAL.YRT1LD52/ASYMMETRY_KNOB_S,0.0,,TRANSFER_INJECTION
LOGICAL.YRT1LD52/U_QUAD,-6500.0,V,TRANSFER_INJECTION
LOGICAL.YRT1QD61/K1L,1.7,1/m,TRANSFER_INJECTION
LOGICAL.YRT1QD62/K1L,-1.7,1/m,TRANSFER_INJECTION
LOGICAL.YRT1QD71/K1L,1.5,1/m,TRANSFER_INJECTION
LOGICAL.YRT1QD72/K1L,-1.5,1/m,TRANSFER_INJECTION
YRT1BR1_TO_YRT1MH2_BEAM/E_S,298000.0,eV/u,TRANSFER_INJECTION
YRT1BR1_TO_YRT1MH2_RF/PHASE_OFFSET_S,0.0,deg,TRANSFER_INJECTION
LOGICAL.GHTYKH4/HKICK,0.0,rad,TRANSFER_INJECTION
LOGICAL.GHTYKV3/VKICK,0.0,rad,TRANSFER_INJECTION
LOGICAL.GHTYKV4/VKICK,-0.001,rad,TRANSFER_INJECTION
LOGICAL.GHTYQD41/K1L,1.4,1/m,TRANSFER_INJECTION
LOGICAL.GHTYQD42/K1L,-1.2,1/m,TRANSFER_INJECTION
LOGICAL.YR01LP1I/ESEP_HKICK_S,0.008,rad,TRANSFER_INJECTION
LOGICAL.YR01LP1I/ESEP_K0L_THEO_S,-0.05,rad,TRANSFER_INJECTION
LOGICAL.YR01LP2I/ESEP_HKICK_S,-0.018,rad,TRANSFER_INJECTION
LOGICAL.YR01LP2I/ESEP_K0L_THEO_S,0.055,rad,TRANSFER_INJECTION
LOGICAL.YR01MP1I/HKICK,0.0025,rad,TRANSFER_INJECTION
LOGICAL.YR01MP1I/K0L_THEO,0.29322,rad,TRANSFER_INJECTION
LOGICAL.YRT1MH2/HKICK_ESR_S,-0.61087,rad,TRANSFER_INJECTION
LOGICAL.YRT1MH2/HKICK_LOCAL_S,-0.021,rad,TRANSFER_INJECTION
LOGICAL.YRT1MH2/K0L_THEO,-0.61087,rad,TRANSFER_INJECTION
YRT1MH2_TO_CRYRING_TIMEPARAM/TIMING_BI_DELAY_S,1.0E-6,s,TRANSFER_INJECTION
YRT1MH2_TO_CRYRING_TIMEPARAM/TIMING_BI_WINDOW_S,1.0E-6,s,TRANSFER_INJECTION
YRT1MH2_TO_CRYRING_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,BEAMOUT_INIT
YRT1MH2_TO_CRYRING_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,TRANSFER_INJECTION
YRT1MH2_TO_CRYRING_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1MH2_TO_CRYRING_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1MH2_TO_CRYRING_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1MH2_TO_CRYRING_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1MH2_TO_CRYRING_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1MH2_TO_CRYRING_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1MH2_TO_CRYRING_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1MH2_TO_CRYRING_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1MH2_TO_CRYRING_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1MH2_TO_CRYRING_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,BEAMOUT_RESET
YRT1MH2_TO_CRYRING_TIMEPARAM/TIMING_SEPTUM_TRIGGER_OFFSET_S,0.0,s,TRANSFER_INJECTION
CRYRINGB2B/B2B_SID_EXT_S,---,,RING_INJECTION
CRYRINGB2B/B2B_SOURCE_KICKER_DELAY_S,---,s,RING_INJECTION
CRYRINGB2B/B2B_SOURCE_TRIGGER_OFFSET_S,---,deg,RING_INJECTION
CRYRINGB2B/B2B_TRIGGER_OFFSET_S,0.0,deg,RING_INJECTION
CRYRINGB2B/B2B_TRIGGER_OFFSET_TIME_S,0.0,,RING_INJECTION
CRYRINGBEAM/BEAM_MODE_S,NO_BEAM,,RING_INJECTION
CRYRINGBEAM/BRHO_CHIMNEY_S,---,Tm,BEAMOUT_INIT
CRYRINGBEAM/BRHO_CHIMNEY_S,---,Tm,BEAMOUT_RESET
CRYRINGBEAM/BRHO_DISPLAY_S,0.1583,Tm,RING_INJECTION
CRYRINGBEAM/BRHO_DISPLAY_S,0.205,Tm,RING_RF_MANIPULATION
CRYRINGBEAM/BRHO_DISPLAY_S,0.205,Tm,RING_TARGET
CRYRINGBEAM/BRHO_START_END_S,[0.15826692, 0.15826692],Tm,RING_INJECTION
CRYRINGBEAM/BUCKETFILL_RAW_S,[[0.0, 0.0]],,RING_INJECTION
CRYRINGBEAM/BUCKETFILL_RAW_S,[[0.0, 1.4]],,RING_BUNCHING
CRYRINGBEAM/BUCKETFILL_RAW_S,[[1.4, 1.4]],,RING_RAMP
CRYRINGBEAM/BUCKETFILL_RAW_S,[[1.4, 1.4]],,RING_RF_MANIPULATION
CRYRINGBEAM/BUCKETFILL_RAW_S,[[1.4, 1.4]],,RING_RAMP
CRYRINGBEAM/BUCKETFILL_RAW_S,[[1.4, 0.0]],,RING_REBUNCHING
CRYRINGBEAM/BUCKETFILL_RAW_S,[[0.0, 0.0]],,RING_TARGET
CRYRINGBEAM/BUCKETFILL_RAW_S,[[0.0, 0.0]],,RING_INTERMEDIATE
CRYRINGBEAM/BUCKETFILL_SIMPLE_S,1.4,,RING_INJECTION
CRYRINGBEAM/CHIMNEY_B,---,,BEAMOUT_INIT
CRYRINGBEAM/CHIMNEY_B,---,,BEAMOUT_RESET
CRYRINGBEAM/CH_END_S,5.0655,,RING_INJECTION
CRYRINGBEAM/CH_END_S,5.0655,,RING_RAMP
CRYRINGBEAM/CH_END_S,5.0655,,RING_RAMP
CRYRINGBEAM/CH_END_S,---,,STORAGE_MANIP_WAIT
CRYRINGBEAM/COASTING_BEAM_B,true,,RING_REBUNCHING
CRYRINGBEAM/CV_END_S,-5.3921,,RING_INJECTION
CRYRINGBEAM/CV_END_S,-5.3921,,RING_RAMP
CRYRINGBEAM/CV_END_S,-5.3921,,RING_RAMP
CRYRINGBEAM/CV_END_S,---,,STORAGE_MANIP_WAIT
CRYRINGBEAM/DELTAR,-0.00225,m,RING_INJECTION
CRYRINGBEAM/DELTAR,0.0,m,RING_RF_MANIPULATION
CRYRINGBEAM/DELTAR,-5.0E-4,m,STORAGE_MANIP_WAIT
CRYRINGBEAM/DELTAR,-5.0E-4,m,RING_TARGET
CRYRINGBEAM/DELTAR,-5.0E-4,m,RING_INTERMEDIATE
CRYRINGBEAM/DPBL,0.0,,RING_INJECTION
CRYRINGBEAM/DPBL,0.0,,STORAGE_MANIP_WAIT
CRYRINGBEAM/DPFREV,0.0035,,RING_INJECTION
CRYRINGBEAM/DPFREV,0.004,,RING_RF_MANIPULATION
CRYRINGBEAM/DPFREV,0.00775,,STORAGE_MANIP_WAIT
CRYRINGBEAM/DPFREV,0.00775,,RING_TARGET
CRYRINGBEAM/DPFREV,0.00775,,RING_INTERMEDIATE
CRYRINGBEAM/DP_OVER_P_S,0.0014,,RING_INJECTION
CRYRINGBEAM/E_S,298000.0,eV/u,RING_INJECTION
CRYRINGBEAM/E_S,500000.0,eV/u,RING_RAMP
CRYRINGBEAM/E_S,500000.0,eV/u,RING_RAMP
CRYRINGBEAM/FINAL_HARMONIC_S,2,,RING_RF_MANIPULATION
CRYRINGBEAM/FREV,140363.587,Hz,RING_INJECTION
CRYRINGBEAM/FREV,181811.9976,Hz,RING_RF_MANIPULATION
CRYRINGBEAM/FREV,182373.299,Hz,RING_TARGET
CRYRINGBEAM/H_INPUT_S,2,,RING_INJECTION
CRYRINGBEAM/INJ_EMIH_S,1.0E-5,m*rad,RING_INJECTION
CRYRINGBEAM/INJ_EMIL_S,0.006665753393459646,eVs/u,RING_INJECTION
CRYRINGBEAM/INJ_EMIV_S,1.0E-5,m*rad,RING_INJECTION
CRYRINGBEAM/KL_AMPL_S,0.0,,RING_INJECTION
CRYRINGBEAM/KL_AMPL_S,0.0,,RING_BUNCHING
CRYRINGBEAM/KL_AMPL_S,0.0,,RING_POST_INJECTION
CRYRINGBEAM/KL_AMPL_S,0.0,,RING_MANI_PRE_RAMP
CRYRINGBEAM/KL_AMPL_S,0.0,,RING_RAMP
CRYRINGBEAM/KL_AMPL_S,0.0,,RING_RF_MANIPULATION
CRYRINGBEAM/KL_AMPL_S,0.0,,RING_RAMP
CRYRINGBEAM/KL_AMPL_S,0.0,,RING_REBUNCHING
CRYRINGBEAM/KL_AMPL_S,0.0,,RING_TARGET
CRYRINGBEAM/KL_AMPL_S,0.0,,RING_INTERMEDIATE
CRYRINGBEAM/KL_HARM_S,1,,RING_INJECTION
CRYRINGBEAM/KL_HARM_S,1,,RING_BUNCHING
CRYRINGBEAM/KL_HARM_S,1,,RING_POST_INJECTION
CRYRINGBEAM/KL_HARM_S,1,,RING_MANI_PRE_RAMP
CRYRINGBEAM/KL_HARM_S,1,,RING_RAMP
CRYRINGBEAM/KL_HARM_S,1,,RING_RF_MANIPULATION
CRYRINGBEAM/KL_HARM_S,1,,RING_RAMP
CRYRINGBEAM/KL_HARM_S,1,,RING_REBUNCHING
CRYRINGBEAM/KL_HARM_S,1,,RING_TARGET
CRYRINGBEAM/KL_HARM_S,1,,RING_INTERMEDIATE
CRYRINGBEAM/KL_OFFSET_S,0.0,,RING_INJECTION
CRYRINGBEAM/KL_OFFSET_S,0.0,,RING_BUNCHING
CRYRINGBEAM/KL_OFFSET_S,0.0,,RING_POST_INJECTION
CRYRINGBEAM/KL_OFFSET_S,0.0,,RING_MANI_PRE_RAMP
CRYRINGBEAM/KL_OFFSET_S,0.0,,RING_RAMP
CRYRINGBEAM/KL_OFFSET_S,0.0,,RING_RF_MANIPULATION
CRYRINGBEAM/KL_OFFSET_S,0.0,,RING_RAMP
CRYRINGBEAM/KL_OFFSET_S,0.0,,RING_REBUNCHING
CRYRINGBEAM/KL_OFFSET_S,0.0,,RING_TARGET
CRYRINGBEAM/KL_OFFSET_S,0.0,,RING_INTERMEDIATE
CRYRINGBEAM/KL_PHASE_S,0.0,deg,RING_INJECTION
CRYRINGBEAM/KL_PHASE_S,0.0,deg,RING_BUNCHING
CRYRINGBEAM/KL_PHASE_S,0.0,deg,RING_POST_INJECTION
CRYRINGBEAM/KL_PHASE_S,0.0,deg,RING_MANI_PRE_RAMP
CRYRINGBEAM/KL_PHASE_S,0.0,deg,RING_RAMP
CRYRINGBEAM/KL_PHASE_S,0.0,deg,RING_RF_MANIPULATION
CRYRINGBEAM/KL_PHASE_S,0.0,deg,RING_RAMP
CRYRINGBEAM/KL_PHASE_S,0.0,deg,RING_REBUNCHING
CRYRINGBEAM/KL_PHASE_S,0.0,deg,RING_TARGET
CRYRINGBEAM/KL_PHASE_S,0.0,deg,RING_INTERMEDIATE
CRYRINGBEAM/NPARTICLES_S,1000.0,,RING_INJECTION
CRYRINGBEAM/PARTICLE_SYMBOL_INPUT_S,2H,,RING_INJECTION
CRYRINGBEAM/QH_END_S,2.29,,RING_INJECTION
CRYRINGBEAM/QH_END_S,---,,RING_POST_INJECTION
CRYRINGBEAM/QH_END_S,2.35,,RING_RAMP
CRYRINGBEAM/QH_END_S,2.34,,RING_RAMP
CRYRINGBEAM/QH_END_S,---,,STORAGE_MANIP_WAIT
CRYRINGBEAM/QV_END_S,2.55,,RING_INJECTION
CRYRINGBEAM/QV_END_S,---,,RING_POST_INJECTION
CRYRINGBEAM/QV_END_S,2.55,,RING_RAMP
CRYRINGBEAM/QV_END_S,2.55,,RING_RAMP
CRYRINGBEAM/QV_END_S,---,,STORAGE_MANIP_WAIT
CRYRINGBEAM/Q_INPUT_S,1,,RING_INJECTION
CRYRINGBEAM/TREV,7.12435E-6,s,RING_INJECTION
CRYRINGBEAM/USE_SIMPLE_BF_B,true,,RING_INJECTION
CRYRINGBUMPER/TFALL_S,1.5E-4,s,RING_INJECTION
CRYRINGIKICKER/KICKER_TOFFSET1_S,0.0,s,RING_INJECTION
CRYRINGOPTICS/OPTIC_NAME_RING_S,CRYRING_STANDARD_OPTICS_TABLE,,RING_INJECTION
CRYRINGOPTICS/TAU_END_S,0.0,,RING_INJECTION
CRYRINGOPTICS/TAU_END_S,---,,RING_POST_INJECTION
CRYRINGOPTICS/TAU_END_S,0.0,,RING_RAMP
CRYRINGOPTICS/TAU_END_S,0.0,,RING_RAMP
CRYRINGORBIT/COH02,0.0,m,RING_INJECTION
CRYRINGORBIT/COH02,0.0,m,RING_RF_MANIPULATION
CRYRINGORBIT/COH02,0.0,m,STORAGE_MANIP_WAIT
CRYRINGORBIT/COH02,0.0,m,RING_TARGET
CRYRINGORBIT/COH03,0.0,m,RING_INJECTION
CRYRINGORBIT/COH03,0.0,m,RING_RF_MANIPULATION
CRYRINGORBIT/COH03,0.0,m,STORAGE_MANIP_WAIT
CRYRINGORBIT/COH03,0.0,m,RING_TARGET
CRYRINGORBIT/COH04,0.0,m,RING_INJECTION
CRYRINGORBIT/COH04,0.0,m,RING_RF_MANIPULATION
CRYRINGORBIT/COH04,0.0,m,STORAGE_MANIP_WAIT
CRYRINGORBIT/COH04,0.0,m,RING_TARGET
CRYRINGORBIT/COH06,0.0,m,RING_INJECTION
CRYRINGORBIT/COH06,0.0,m,RING_RF_MANIPULATION
CRYRINGORBIT/COH06,-0.009,m,STORAGE_MANIP_WAIT
CRYRINGORBIT/COH06,-0.009,m,RING_TARGET
CRYRINGORBIT/COH07,0.0,m,RING_INJECTION
CRYRINGORBIT/COH07,0.0,m,RING_RF_MANIPULATION
CRYRINGORBIT/COH07,0.0,m,STORAGE_MANIP_WAIT
CRYRINGORBIT/COH07,0.0,m,RING_TARGET
CRYRINGORBIT/COH08,0.0,m,RING_INJECTION
CRYRINGORBIT/COH08,0.0,m,RING_RF_MANIPULATION
CRYRINGORBIT/COH08,0.0,m,STORAGE_MANIP_WAIT
CRYRINGORBIT/COH08,0.0,m,RING_TARGET
CRYRINGORBIT/COH10,0.0,m,RING_INJECTION
CRYRINGORBIT/COH10,0.0,m,RING_RF_MANIPULATION
CRYRINGORBIT/COH10,0.0,m,STORAGE_MANIP_WAIT
CRYRINGORBIT/COH10,0.0,m,RING_TARGET
CRYRINGORBIT/COH11,0.0,m,RING_INJECTION
CRYRINGORBIT/COH11,0.0,m,RING_RF_MANIPULATION
CRYRINGORBIT/COH11,0.0,m,STORAGE_MANIP_WAIT
CRYRINGORBIT/COH11,0.0,m,RING_TARGET
CRYRINGORBIT/COH12,0.0,m,RING_INJECTION
CRYRINGORBIT/COH12,0.0,m,RING_RF_MANIPULATION
CRYRINGORBIT/COH12,0.0,m,STORAGE_MANIP_WAIT
CRYRINGORBIT/COH12,0.0,m,RING_TARGET
CRYRINGORBIT/COV02,0.0,m,RING_INJECTION
CRYRINGORBIT/COV02,0.0,m,RING_RF_MANIPULATION
CRYRINGORBIT/COV02,0.0,m,STORAGE_MANIP_WAIT
CRYRINGORBIT/COV02,0.0,m,RING_TARGET
CRYRINGORBIT/COV03,0.0,m,RING_INJECTION
CRYRINGORBIT/COV03,0.0,m,RING_RF_MANIPULATION
CRYRINGORBIT/COV03,0.0,m,STORAGE_MANIP_WAIT
CRYRINGORBIT/COV03,0.0,m,RING_TARGET
CRYRINGORBIT/COV04,0.0,m,RING_INJECTION
CRYRINGORBIT/COV04,0.0,m,RING_RF_MANIPULATION
... 679 more lines ...
Attachment 3: Screenshot_from_2025-06-02_21-03-11.png
Screenshot_from_2025-06-02_21-03-11.png
Attachment 4: overlap.png
overlap.png
Attachment 5: Screenshot_from_2025-06-03_00-19-00.png
Screenshot_from_2025-06-03_00-19-00.png
Attachment 6: Screenshot_from_2025-06-03_00-19-13.png
Screenshot_from_2025-06-03_00-19-13.png
Attachment 7: Screenshot_from_2025-06-03_00-19-28.png
Screenshot_from_2025-06-03_00-19-28.png
Attachment 8: Screenshot_from_2025-06-03_00-19-28.png
Screenshot_from_2025-06-03_00-19-28.png
Attachment 9: Screenshot_from_2025-06-03_00-19-52.png
Screenshot_from_2025-06-03_00-19-52.png
Attachment 10: Screenshot_from_2025-06-03_00-20-01.png
Screenshot_from_2025-06-03_00-20-01.png
Attachment 11: Screenshot_from_2025-06-03_04-11-15.png
Screenshot_from_2025-06-03_04-11-15.png
Attachment 12: Screenshot_from_2025-06-03_07-26-29.png
Screenshot_from_2025-06-03_07-26-29.png
Attachment 13: beamCalc-D5.png
beamCalc-D5.png
  714   Mon Jun 2 17:48:07 2025 JG, AGRun 44 scraper tests

Same settings are Run 43 but with coasting beam and scrapers set to:  30 / - 30 mm.

 

all checks ok, but WR timestamp error, see attached

 

19:28 the Run 44 has been stopped.

Attachment 1: Screenshot_from_2025-06-02_18-51-24.png
Screenshot_from_2025-06-02_18-51-24.png
Attachment 2: Screenshot_from_2025-06-02_19-27-31.png
Screenshot_from_2025-06-02_19-27-31.png
  713   Mon Jun 2 17:25:31 2025 JG, AGRun 43 scraper tests

Same settings are Run 42 but bunched beam and scrapers set to:  30 / - 30 mm.

 

  712   Mon Jun 2 16:35:35 2025 EW, AGRun 42 and scraper tests

17:00

Scraper tests conducted. We tried moving in the scrapers both at a fixed position throughout and also moving with the injection cycle.

We were unable to find a good position with the scrapers that did not produce significant scattering in the detectors.

Therefore we have moved the scrapers back to the fully out position.

Run 42 started.

We have increased the storage time to 10 reps (1 minute) in order to investigate the shoulders on the 3He peaks online and whether it is likely from injection. I think it is likely just a calibration issue

D+ beam deuterium target. D+ beam energy is 0.75 MeV/u. (beam calc calculation is 0.7619 MeV/u). See previous entry for beamcalc.

Beam calc - Attachment 1

Target density is about 2e12 at/cm2

All four detectors positioned fully out and stationary.

Left potentiometer = 18.86 mm

Right potentiometer = 19.8 mm

Ring cycle time is 60 seconds. Note we see some hydrogen in in the spectrum, this disapears after a few seconds.

Target T1=30.0K, T2=31.6K.

Nozzle position x=-2.96, z=1.72 mm

YR09-Etarget pressure = 2.0e-10 mbar.

Cooler set value = 439 V, 15 mA. Read value = 439.81 V, 12.88 mA

All leakage currents look good

Beam position. Vertical =0.0mm , Horizontal=-1.5 mm

Beam intensity ~6e8 on bunched beam, 6e8 on coasting.

Data rate ~ 416kB/s

slow comparator threshold to 0x24 (360 keV)

Attachment 1: Screenshot_from_2025-06-02_17-44-07.png
Screenshot_from_2025-06-02_17-44-07.png
Attachment 2: Screenshot_from_2025-06-02_17-44-50.png
Screenshot_from_2025-06-02_17-44-50.png
  711   Mon Jun 2 09:30:38 2025 JM, EW, STRun 41

Run 41 started.

D+ beam deuterium target. D+ beam energy is 0.75 MeV/u. (beam calc calculation is 0.7619 MeV/u). See previous entry for beamcalc.

Beam calc - Attachment 1

Target density is about 3e12 at/cm2

All four detectors positioned fully out and stationary.

Left potentiometer = 18.86 mm

Right potentiometer = 19.8 mm

Ring cycle time is 25 seconds. Note we see some hydrogen in in the spectrum, this disapears after a few seconds.

Target T1=30.0K, T2=31.9K.

Nozzle position x=-2.96, z=1.72 mm

YR09-Etarget pressure = 2.0e-10 mbar.

Cooler set value = 439 V, 15 mA. Read value = 439.81 V, 12.88 mA

All leakage currents look good

Beam position. Vertical =0.0mm , Horizontal=-1.5 mm

Beam intensity ~3e8 on bunched beam, 3e8 on coasting. Gleb has increased the intensity this morning.

Data rate ~ 403kB/s

Lowered slow comparator threshold to 0x24 (360 keV)

13:20 Run stopped

Wolfgang tuning next beam energy

All checks fine

Total stats: 7827,4085,1821

OFFLINE CHECK

I have check Run41 and applied the gain calibration from 2024 in addition to a tigher EyEx energy cut (600 keV from 3 MeV for the online script). We no longer see the larger hump to the side of the 3He peak in DSSD1. See final attachment for comparison.

Note no time cuts have been applied to this spectra.and a more rigourous gain calibration using the igher energy D+ runs will likely improve things further. 

Attachment 1: Screenshot_from_2025-06-02_10-37-39.png
Screenshot_from_2025-06-02_10-37-39.png
Attachment 2: Screenshot_from_2025-06-02_10-38-36.png
Screenshot_from_2025-06-02_10-38-36.png
Attachment 3: Screenshot_from_2025-06-02_13-21-06.png
Screenshot_from_2025-06-02_13-21-06.png
Attachment 4: Screenshot_from_2025-06-02_13-20-36.png
Screenshot_from_2025-06-02_13-20-36.png
Attachment 5: Screenshot_from_2025-06-02_13-20-19.png
Screenshot_from_2025-06-02_13-20-19.png
Attachment 6: DSSD1-Ey.png
DSSD1-Ey.png
  710   Sun Jun 1 15:33:50 2025 EW, JM, CBRun 40

16:30

Run 40 started.

D+ beam deuterium target. D+ beam energy is 0.75 MeV/u. (beam calc calculation is 0.7619 MeV/u)

Beam calc - Attachment 1

Target density is about 1e12 at/cm2

All four detectors positioned fully out and stationary.

Left potentiometer = 18.86 mm

Right potentiometer = 19.8 mm

Ring cycle time is 30 seconds. Note we see some hydrogen in in the spectrum, this disapears after a few seconds.

Target T1=30.1K, T2=31.9K.

Nozzle position x=-2.96, z=1.72 mm

YR09-Etarget pressure = 2.55e-10 mbar.

Cooler set value = 439 V, 15 mA. Read value = 439.81 V, 12.88 mA

All leakage currents look good

Beam position. Vertical =0.0mm , Horizontal=-1.5 mm

Beam intensity ~1e8 on bunched beam, 1e7 on coasting. This is less intensity than higher energies, may require an increase in the morning. 

Data rate ~ 403kB/s

Lowered slow comparator threshold to 0x24 (360 keV) since we decreased beam energy.

18:36

Temps OK - attached

Rates OK - attached

Merger OK - attached.

Spectra appear to have a flat background to the right of Rutherford that was not there at higher energies.

The xy plot also does not seem to be as centered.

21:17

Tems OK

Stats OK

Merger OK

Leakage current stable

Run continues OK

Ring vacuum 2E-10 mbar, likely getting less target.

 

02:00

Temps OK

Stats OK

Merger OK

Leakage current OK

Run continues - attached

Ring vacuum as before

 

04:20

Temps OK

Stats OK

Merger OK

L. current stable

Run continues

Ring vacuum as before

07:00

All checks OK

Leakage currents remain stable.

We have accumulated good statistics. 3He peak does seem a bit off. Cooling could be slightly off. Problem is not so bad looking at Ey compared to Ex. Online code uses pulser calibration only so may be a bit off depending on Ey/Ex.

Target density has been stable in the 1e12-2e12 at/cm2 range overnight. A bit more would be good.

Will ask on call operator to check cooling and try to increase intensity.

Run stopped gas target off.

07:30

Note during the tuning process we observed the same structure at 2.5 MeV in DSSD 1 as we did overnight that may overlap with the 3He peak. Perhaps it can be removed offline. See final attachment

 

Attachment 1: Screenshot_from_2025-06-01_16-38-55.png
Screenshot_from_2025-06-01_16-38-55.png
Attachment 2: beamCalc-D4.png
beamCalc-D4.png
Attachment 3: Screenshot_from_2025-06-01_18-33-17.png
Screenshot_from_2025-06-01_18-33-17.png
Attachment 4: Screenshot_from_2025-06-01_18-33-35.png
Screenshot_from_2025-06-01_18-33-35.png
Attachment 5: Screenshot_from_2025-06-01_18-34-06.png
Screenshot_from_2025-06-01_18-34-06.png
Attachment 6: Screenshot_from_2025-06-01_18-36-19.png
Screenshot_from_2025-06-01_18-36-19.png
Attachment 7: Screenshot_from_2025-06-01_18-36-26.png
Screenshot_from_2025-06-01_18-36-26.png
Attachment 8: Screenshot_from_2025-06-02_02-03-19.png
Screenshot_from_2025-06-02_02-03-19.png
Attachment 9: Screenshot_from_2025-06-02_07-16-07.png
Screenshot_from_2025-06-02_07-16-07.png
Attachment 10: Screenshot_from_2025-06-02_07-37-04.png
Screenshot_from_2025-06-02_07-37-04.png
  709   Sun Jun 1 07:46:34 2025 JM, EM, CBRun 39

08:45

Run 39 started.

D+ beam deuterium target. D+ beam energy is 1.0 MeV/u. (beam calc calculation is 1.0151 MeV/u)

Beam calc - Attachment 1

Target density is about 5e12 at/cm2

All four detectors positioned fully out and stationary.

Left potentiometer = 18.86 mm

Right potentiometer = 19.8 mm

Ring cycle time is 1 minute. Note we see some hydrogen in in the spectrum, this disapears after a few seconds.

Target T1=30.1K, T2=31.9K.

Nozzle position x=-2.96, z=1.72 mm

YR09-Etarget pressure = 2.55e-10 mbar.

Cooler set value = 575 V, 15 mA. Read value = 575.8 V, 12.9 mA

All leakage currents look good

Beam position. Vertical =0.0mm , Horizontal=-1.25 mm

Beam intensity ~1e8. We appear to lose intensity when reducing energy. It is still sufficient to for our purposes.

09.38

CRYRING vacuum and ambient temperature - attachments 2-3

09:58

All checks OK.

Target density is ~7e12 at/cm2

Data rate ~300kB/sec

All leakage currents look OK

After 1 hour we have 3671, 2515 and 1064 in each peak respectively. This is also logged on the google docs form in same tab as elog.

 

10:36

Temps OK

Data rate OK 326 kB/sec

Leakage currents stable

 

12:30

Temps ok - attach

Rates ok - attach

Merger ok - attach

14:20

Run stopped for energy change.

Attachment 1: screenshot-2025-06-01_08-48-132701843244570721227.png
screenshot-2025-06-01_08-48-132701843244570721227.png
Attachment 2: Capture.PNG
Capture.PNG
Attachment 3: Capture_2.PNG
Capture_2.PNG
Attachment 4: Screenshot_from_2025-06-01_12-29-30.png
Screenshot_from_2025-06-01_12-29-30.png
Attachment 5: Screenshot_from_2025-06-01_12-29-48.png
Screenshot_from_2025-06-01_12-29-48.png
Attachment 6: Screenshot_from_2025-06-01_12-30-11.png
Screenshot_from_2025-06-01_12-30-11.png
Attachment 7: Screenshot_from_2025-06-01_14-18-29.png
Screenshot_from_2025-06-01_14-18-29.png
  708   Sat May 31 10:46:42 2025 TDRun 38 - remote monitoring
11.40 DAQ continues
      File G22-205/R38

      ASIC settings
       slow comparator 0x32 all FEE64s

      DSSSD bias & leakage current OK - attachments 1-2

      FEE64 temps OK - attachment 3

      ADC data item stats OK - attachment 4

      WR timestamps OK - attachment 5

      per FEE64 Rate spectra - attachment 6

      Merger & Tape Server - attachments 7-8
       filesystem /mnt/data1 4.7Tb remaining

      Online analysis - per DSSSD energy spectra - elastic peak max count / channel c. 250-300Hz - attachment 9

      CRYRING vacuum & temperature - attachments 10-12
       YR09 e-target gauge indicates vacuum transients from high e-11 to low e-9mbar presumably due to high density D2 target

11.42 All histograms & stats zero'd


17.36 DSSSD bias & leakage current OK - attachments 13-14

      FEE64 temps OK - attachment 15

      ADC data item stats OK - attachment 16

      per FEE64 Rate spectra - attachment 17

      per FEE64 1.8.L spectra - attachments 18-19
       aida11 p+n pulser peak width 63 ch FWHM 
       aida03 n+n pulser peak width 85 ch FWHM

      Merger - attachment 20

      Online analysis spectra - elastic peak max count / channel c. few Hz - attachment 21
       rates much lower than earlier - no beam and/or no target - alert JM
Attachment 1: Screenshot_from_2025-05-31_11-40-15.png
Screenshot_from_2025-05-31_11-40-15.png
Attachment 2: Screenshot_from_2025-05-31_11-40-34.png
Screenshot_from_2025-05-31_11-40-34.png
Attachment 3: Screenshot_from_2025-05-31_11-41-01.png
Screenshot_from_2025-05-31_11-41-01.png
Attachment 4: Screenshot_from_2025-05-31_11-41-18.png
Screenshot_from_2025-05-31_11-41-18.png
Attachment 5: Screenshot_from_2025-05-31_11-42-27.png
Screenshot_from_2025-05-31_11-42-27.png
Attachment 6: Screenshot_from_2025-05-31_11-43-50.png
Screenshot_from_2025-05-31_11-43-50.png
Attachment 7: Screenshot_from_2025-05-31_11-44-54.png
Screenshot_from_2025-05-31_11-44-54.png
Attachment 8: Screenshot_from_2025-05-31_11-45-10.png
Screenshot_from_2025-05-31_11-45-10.png
Attachment 9: Screenshot_from_2025-05-31_11-46-08.png
Screenshot_from_2025-05-31_11-46-08.png
Attachment 10: Capture.PNG
Capture.PNG
Attachment 11: Capture_2.PNG
Capture_2.PNG
Attachment 12: Capture_3.PNG
Capture_3.PNG
Attachment 13: Screenshot_from_2025-05-31_17-26-17.png
Screenshot_from_2025-05-31_17-26-17.png
Attachment 14: Screenshot_from_2025-05-31_17-03-49.png
Screenshot_from_2025-05-31_17-03-49.png
Attachment 15: Screenshot_from_2025-05-31_17-03-18.png
Screenshot_from_2025-05-31_17-03-18.png
Attachment 16: Screenshot_from_2025-05-31_17-03-38.png
Screenshot_from_2025-05-31_17-03-38.png
Attachment 17: Screenshot_from_2025-05-31_17-26-42.png
Screenshot_from_2025-05-31_17-26-42.png
Attachment 18: Screenshot_from_2025-05-31_17-27-39.png
Screenshot_from_2025-05-31_17-27-39.png
Attachment 19: Screenshot_from_2025-05-31_17-27-39.png
Screenshot_from_2025-05-31_17-27-39.png
Attachment 20: Screenshot_from_2025-05-31_17-36-49.png
Screenshot_from_2025-05-31_17-36-49.png
Attachment 21: Screenshot_from_2025-05-31_17-37-07.png
Screenshot_from_2025-05-31_17-37-07.png
  707   Sat May 31 09:58:21 2025 JMRun 38

10:56

Run 38 started.

D+ beam deuterium target. D+ beam energy is 1.5 MeV/u. (beam calc calculation is 1.5160 MeV/u)

Beam calc - Attachment 1

Target density is about 8.5e12 at/cm2 after moving nozzle this morning.

All four detectors positioned fully out and stationary.

Left potentiometer = 18.86 mm

Right potentiometer = 19.8 mm

Ring cycle time is 1 minute. Note we see some hydrogen in in the spectrum, this disapears after a few seconds.

Target T1=30K, T2=31.9K.

Nozzle position x=-2.89, z=1.72 mm

YR09-Etarget pressure = 2.37e-10 mbar.

Cooler set value = 846V, 15 mA. Read value = 847.378V, 13.105mA

Detector 4 looks OK at the moment.

Beam position. Vertical =0.0mm , Horizontal=2.5mm

11:05

All checks OK

Data rate is ~300 kB/sec.

Leakage currents - attachments 2,3

17:56

Rates have dropped significantly. Do not see elastics peak. No beam?

18:23

Check in with HKR. The thunder and lightning storm appears to have knocked out the beam from 17:30 until 18:10.

Beam is now back with an intensity of 2e8 pps.

I have also adjusted the nozzle of the gas target slightly and recovered target pressure of 1e13 from 3e11 at/cm2. This has been stable for the last 15 minutes. I think during shifts we will require some nozzle modifications. It is not vital at higher energies but for energies during the week, we will need.

We again see the very nice elastic scattering pattern on our detector. I am concerned we may be some shadowing on our detector giving rise to the observed circle? Attachment 5

20:30

Checks all OK

Target density in last 30 mins ~1e13 at/cm2

DSSD4 still looking good and  stable

23:00

All checks OK.

Detectors all stable

Target density still in 1e13 range.

We have accumulated lots of stats! This run should provide an excellent run for energy calibration using elastics and reaction peaks.

We run overnight and change energy in the morning

01:00

All checks OK

Detectors stable

Target density stable 1e13.

05:00

All good still

07:05

Run stopped to change energy. Gas target off.

final statistics plot attached.

Leakage currents overnight also. DSSD4 had two transients during the night but recovered.

Attachment 1: beamCalc-D2.png
beamCalc-D2.png
Attachment 2: Screenshot_from_2025-05-31_11-07-00.png
Screenshot_from_2025-05-31_11-07-00.png
Attachment 3: Screenshot_from_2025-05-31_11-06-50.png
Screenshot_from_2025-05-31_11-06-50.png
Attachment 4: Screenshot_from_2025-05-31_17-55-33_1.png
Screenshot_from_2025-05-31_17-55-33_1.png
Attachment 5: Screenshot_from_2025-05-31_18-26-29.png
Screenshot_from_2025-05-31_18-26-29.png
Attachment 6: Screenshot_from_2025-05-31_23-04-18.png
Screenshot_from_2025-05-31_23-04-18.png
Attachment 7: Screenshot_from_2025-06-01_07-10-13.png
Screenshot_from_2025-06-01_07-10-13.png
Attachment 8: Screenshot_from_2025-06-01_07-10-21.png
Screenshot_from_2025-06-01_07-10-21.png
Attachment 9: Screenshot_from_2025-06-01_07-10-30.png
Screenshot_from_2025-06-01_07-10-30.png
Attachment 10: Screenshot_from_2025-06-01_07-30-28.png
Screenshot_from_2025-06-01_07-30-28.png
  706   Fri May 30 21:49:08 2025 JM, NP, CBRun 37

22:49

Run 37 started.

D+ beam deuterium target. D+ beam energy is 2 MeV/u. (beam calc calculation is 1.9211 MeV/u)

Target density is about 1.3e12 at/cm2. FLuctuations between 1e12 and 1e13 at/cm2.

All four detectors positioned fully out and stationary.

Left potentiometer = 18.86 mm

Right potentiometer = 19.8 mm

Ring cycle time is 1 minute. Note we see some hydrogen in in the spectrum, this disapears after a few seconds.

Target T1=30K, T2=31.9K.

Nozzle position x=-3.11, z=1.77 mm

YR09-Etarget pressure = 1.04e-10 mbar.

Cooler set value = 1126V, 15 mA. Read value = 1068V, 13.22 mA

Detector 4 is still a bit unstable in its leakage current but the data coming through looks good. Some crap is periodically written to disk when leakage current spikes. Attachment 1,2 for leakage currents.

23:30

Checks OK - high but managable rates from DSSD4

01:00

Checks OK - DSSD4 has settled down. Leakage current attachment 1

Ex energy histograms - attachment 2

X2go check on gas target shows density has dropped to 5e11 at/cm2 in last half an hour. 

03:56

Checks OK

DSSD had more issues but once more settled

Temps OK

Rates OK

DAQ continuing

06:00

Checks OK

08:00

We can now change to 1.5 MeV/u no storage mode enabled. 

Attachment 1: Screenshot_from_2025-05-31_00-56-32.png
Screenshot_from_2025-05-31_00-56-32.png
Attachment 2: Screenshot_from_2025-05-31_00-58-40.png
Screenshot_from_2025-05-31_00-58-40.png
Attachment 3: Screenshot_from_2025-05-31_06-16-50.png
Screenshot_from_2025-05-31_06-16-50.png
  705   Fri May 30 20:47:47 2025 JM, CB, NPRun 36

21:47

Run 36 started.

D+ beam deuterium target. D+ beam energy is 2 MeV/u.

Target density is about 1.6e12 at/cm2. FLuctuations between 1e12 and 1e13 at/cm2.

Detectors are now IN with the servo motors only. Detectors are stationary.

Left potentiometer = 36.54 mm

Right potentiometer = 37.96 mm

Ring cycle time is 1 minute. Note we see some hydrogen in in the spectrum, this disapears after a few seconds.

Target T1=30K, T2=31.9K.

Nozzle position x=-3.12, z=1.72 mm

YR09-Etarget pressure = 2.29e-10 mbar.

Cooler set value = 1126V, 15 mA. Read value = 1068V, 13.22 mA

Detector 4 is still a bit unstable in its leakage current but the data coming through looks good. Some crap is periodically written to disk when leakage current spikes.

We will run like this for an hour then move detectors back out for overnight run.

after 1 hour of beamtime we stop.

Attachment 1 - xy plot

Attachment 2 - energy histogram NUCLEAR

Attachment 1: Screenshot_from_2025-05-30_22-45-48.png
Screenshot_from_2025-05-30_22-45-48.png
Attachment 2: Screenshot_from_2025-05-30_22-46-48.png
Screenshot_from_2025-05-30_22-46-48.png
  704   Fri May 30 20:06:48 2025 JM, CB, NPRun35

21:00

Run 35 started.

D+ beam deuterium target. D+ beam energy is 2 MeV/u. (beam calc calculation is 1.9211 MeV/u -> may be wrong cooler voltage, need to check grafana)

Target density is about 4e12 at/cm2. FLuctuations between 1e12 and 1e13 at/cm2.

All four detectors positioned fully out and stationary.

Left potentiometer = 18.86 mm

Right potentiometer = 19.8 mm

Ring cycle time is 1 minute. Note we see some hydrogen in in the spectrum, this disapears after a few seconds.

Target T1=30K, T2=31.9K.

Nozzle position x=-3.06, z=1.72 mm

YR09-Etarget pressure = 2.29e-10 mbar.

Cooler set value = 1126V, 15 mA. Read value = 1068V, 13.22 mA

Detector 4 is still a bit unstable in its leakage current but the data coming through looks good. Some crap is periodically written to disk when leakage current spikes. Attachment 1,2 for leakage currents.

Attachment 3,4 data rate to disk and statistics tab for when detector 4 is behaving. When not data rate can spike to 4000 kb/sec. Looks OK at the moment.

Attachment 5 - xy plot. We observe a circular shape on the detector. We are unsure of what this is at the moment. Offline analysis required.

Attachment 6 - energy histogram including Rutherford.

Attachment 7 - energy histogram just nuclear peaks.

Attachment 8 - beam calc with read values. 

Attachment 1: Screenshot_from_2025-05-30_21-22-53.png
Screenshot_from_2025-05-30_21-22-53.png
Attachment 2: Screenshot_from_2025-05-30_21-23-16.png
Screenshot_from_2025-05-30_21-23-16.png
Attachment 3: Screenshot_from_2025-05-30_21-23-30.png
Screenshot_from_2025-05-30_21-23-30.png
Attachment 4: Screenshot_from_2025-05-30_21-23-56.png
Screenshot_from_2025-05-30_21-23-56.png
Attachment 5: Screenshot_from_2025-05-30_21-26-17.png
Screenshot_from_2025-05-30_21-26-17.png
Attachment 6: Screenshot_from_2025-05-30_21-35-00.png
Screenshot_from_2025-05-30_21-35-00.png
Attachment 7: Screenshot_from_2025-05-30_21-36-58.png
Screenshot_from_2025-05-30_21-36-58.png
Attachment 8: beamCalc-D1.png
beamCalc-D1.png
  703   Fri May 30 07:27:41 2025 CB LB JM NP EMWaiting for deuterium

08:27

No run going.

Temperatures OK - attach

 

Attempting to bias channel 3 again. Slowly ramping up.

V = 60, I = 5.3

V = 70, I = 6.1

V = 80, I = 9.0 --> 6.2 (unstable). Decreasing bias.

V = 70, I = 6.0 (stable again)

Leaving at 70 V for a while to condition (?)

 

Resetting the gnuplot graph. Old caen_current.dat moved to G22-205-caen_current.dat and copied in /TapeServer/G22-205

 

10:10

Temperatures OK

Leakage current on channel 3 showed some issues, back to stable. See attached.

 

11:00

Temps OK

Leakage current stable.

Increased bias on channel 3 to 80 V. I = 6.2 uA stable. Leaving for a while.

 

12:45

Temps OK

Leakage current stable

Increased bias on channel 3 to 90 V. I = 6.3 stable. Leaving for a while.

 

13:30

Temps OK

Leakage current stable.

Increased bias on channel 3 to 100V. I = 6.3 stable. Hopefully it holds until the end of the beam time.

 

20:00

Leakage current on channel 3 had issues again and bias was reduced to 50 V. Remained stable. Turned to 60 V now. Still stable. See attachement.

FEEs on detector for -> MEC

Temps OK - attach

Checks as before - attach

Rates OK (four fees not counting as expected) - attach

 

Deuterium beam in the ring at 2 MeV/A approx 2E8 pps, significant H2+ contamination.

Deuterium target working, fluctuating around 5E12-1E13 atoms/cm2

Vacuum in YR09 - 3E-10 mbar

 

Beginning beam-target overlap measurements soon.

Attachment 1: Screenshot_from_2025-05-30_08-28-08.png
Screenshot_from_2025-05-30_08-28-08.png
Attachment 2: Screenshot_from_2025-05-30_08-37-17.png
Screenshot_from_2025-05-30_08-37-17.png
Attachment 3: Screenshot_from_2025-05-30_10-10-12.png
Screenshot_from_2025-05-30_10-10-12.png
Attachment 4: Screenshot_from_2025-05-30_13-29-57.png
Screenshot_from_2025-05-30_13-29-57.png
Attachment 5: Screenshot_from_2025-05-30_20-00-13.png
Screenshot_from_2025-05-30_20-00-13.png
Attachment 6: Screenshot_from_2025-05-30_20-00-17.png
Screenshot_from_2025-05-30_20-00-17.png
Attachment 7: Screenshot_from_2025-05-30_20-00-27.png
Screenshot_from_2025-05-30_20-00-27.png
Attachment 8: Screenshot_from_2025-05-30_20-00-37.png
Screenshot_from_2025-05-30_20-00-37.png
  702   Thu May 29 18:04:47 2025 LB, CB, PMH, JMRun 32,33,34

Oven temperature increased, current up to 3.5E7 pps.

Found beam-target overlap setting horizontal bump in RO9 to -2.5mm.

Target T1 30.5K, density increasing to 2.6E11 at/cm2.

19:08 Run 32 start

Moving left detectors (79.69->34.66mm).

Beam energy 1.85 MeV/u nominal

Electron cooler actual values: voltage 1.051 kV,  current 0.013215 uA, [voltage divider G35 1.0504 kV]

Stopped run, beam was still in bunch mode.

19:15 Run 33 start

All checks Ok.

Target at 3E11 at/cm2

Attachment 1-2 with rates and temperatures

 

20:20

Target density 3.4E11 at/cm2. Beam around 2.2-3.6E7 pps.

Checks ok! Spectra attached in att.3.

21:25

all good. Target density 4E11 at/cm^2

22:24

All checks are OK

Noted some variability in the injection intensity. For 15 minutes about half of shots were below 1e7. Perhaps some variability due to end of lithium in oven or increased temperture. 

Beam calc with read values - Attachment 4

01:30

All checks OK. Injection intensity has stabilised ~2E7

Target density is stable ~4e11 at/cm2

02:44

Suddenly see very high rates in DSSD4. Detector had tripped. Tried to re-bias but kept tripping.

Set to 70V, leakage current is about 17uA but fluctuates quite wildly. Set to 50V and leakage current is about 6uA but again has wild fluctuations. Not sure of the cause, this detector is not moving and is far from the beam.

Decided to turn detector off to try and protect it. We can  re-bias and monitor in the day tomorrow when we do not have beam on target.

This detector has tripped previously but was able to be re-biased at the time. See attachments 5 and 6 for the detector leakage currents.

New Run started R34.

Same conditions as before. DSSD4 turned off. To stop crap writing to disk I have set the corresponding FEEs to the MEC.

In principle we do not need this detector now as it is not moving to see the reaction.

Note I also observed in the system messages we semi regularly get a DHCP request from aida11. It seems to be functioning fine as I see the pulser signal in the spectrum browser. May be worth following up on at a later date. See attachment 7

All other checks OK

Data rate is now ~140 kB/sec

03:15

Tried to bias DSSD4 again. Leakage currents appear fine again.

Left for 5 minutes to gauge stability -> seems fine.

03:22

Have returned aida13, 14, 15, 16 back to LEC.  Maintaing the same run (34)

Scratch that. 03:27 detector tripped again. Will just leave it off tonight. Turned those FEEs back to MEC.

06:00

I have quickly looked at Run 33. We clearly see the 7Be again in both detectors. Locus on detector much smaller than at higher energies. 

All checks OK

07:07

Run stopped, switching to deuterium.

Detectors moved out

Shutter for gas target off.

Moved to next run -> no storage mode.

Attachment 1: Screenshot_from_2025-05-29_19-16-09.png
Screenshot_from_2025-05-29_19-16-09.png
Attachment 2: Screenshot_from_2025-05-29_19-12-46.png
Screenshot_from_2025-05-29_19-12-46.png
Attachment 3: Screenshot_from_2025-05-29_20-23-50.png
Screenshot_from_2025-05-29_20-23-50.png
Attachment 4: beamCalc.png
beamCalc.png
Attachment 5: Screenshot_from_2025-05-30_02-37-55.png
Screenshot_from_2025-05-30_02-37-55.png
Attachment 6: Screenshot_from_2025-05-30_02-51-21.png
Screenshot_from_2025-05-30_02-51-21.png
Attachment 7: Screenshot_from_2025-05-30_02-38-10.png
Screenshot_from_2025-05-30_02-38-10.png
  701   Thu May 29 14:39:52 2025 CB LB PMHRun 31

15:40

Waiting for Serge to come in to change beam energy.

Remaining at the same beam energy and settings as before, except RIGHT arm now moving.

Run 31

37.96 -> 83.69 mm

 

16:53

Serge arrived. Run ended. Motors out. No storage.

Temps OK

Rates OK

 

 

Attachment 1: Screenshot_from_2025-05-29_15-41-46.png
Screenshot_from_2025-05-29_15-41-46.png
Attachment 2: Screenshot_from_2025-05-29_15-42-11.png
Screenshot_from_2025-05-29_15-42-11.png
Attachment 3: Screenshot_from_2025-05-29_15-42-49.png
Screenshot_from_2025-05-29_15-42-49.png
Attachment 4: Screenshot_from_2025-05-29_16-58-12.png
Screenshot_from_2025-05-29_16-58-12.png
Attachment 5: Screenshot_from_2025-05-29_16-58-16.png
Screenshot_from_2025-05-29_16-58-16.png
Attachment 6: Screenshot_from_2025-05-29_16-58-18.png
Screenshot_from_2025-05-29_16-58-18.png
Attachment 7: Screenshot_from_2025-05-29_16-58-21.png
Screenshot_from_2025-05-29_16-58-21.png
Attachment 8: Screenshot_from_2025-05-29_16-58-23.png
Screenshot_from_2025-05-29_16-58-23.png
Attachment 9: Screenshot_from_2025-05-29_16-58-26.png
Screenshot_from_2025-05-29_16-58-26.png
  700   Wed May 28 16:40:48 2025 EH, SN, LB, OF, EBM, CBRun 28/29/30 - 1.88 MeV/A

We start a new run with real data taking condititon: beam is ok, target density is 1e12 cm-2, detectors are moving in-out periodically.

 

Cooler voltage setpoint 1066.6 V. Read value 1068 V.

Current read = 0.01325 uA

Beam energy = 1.88 MeV/u (from HKR)

Attachment 1 - beamcalc with nominal values

Attachment 32 - beamcalc with actual values

 

Right arm at 19.80 mm

Left  arm IN at 77.56 mm (OUT 32.77 mm)

vertical beam position 0.0 mm

horizontal beam position  -3.0mm

Beam intensity ~1.8e+7 particles.

Target density ~1e12 at/cm2

Target temperatures: T1=31.4 K T2=35.1 K

 

17:48

aida01 fault      0x0 :      0x1 :      1  
FPGA Timestamp error counter test result: Passed 15, Failed 1

All checks OK.

The electron current from beamcalc is the setvalue.

 

18:51

aida01 fault      0x0 :      0x1 :      1  
FPGA Timestamp error counter test result: Passed 15, Failed 1

All checks OK.

20:20

aida01 fault      0x0 :      0x1 :      1  
FPGA Timestamp error counter test result: Passed 15, Failed 1

All checks OK.

22:02

         Base         Current         Difference
aida01 fault      0x0 :      0x1 :      1  
FPGA Timestamp error counter test result: Passed 15, Failed 1

All checks OK.

 

23:13

All checks OK. Target stabilized at ~6.7E11 at/cm^2

00:23

All checks good.

01:01

We are observing what seems to be a steady decrease of the target density, currently at around 5e11 at/cm2, whihc also seems to be following a decrease in the temperature T2.

1;40

All checks good. Spectra attached in att. 18. Target density still decreasing (3.7E11 at/vcm2)2

2:54

All checks ok. Attached rates and spectra in att. 19-20. target density decreased as continued, now around 2.5E11 at/cm^2.

We tried to move a bit around the nozzle/target, no effect.

Gone back to original value of x=1000, y=-4200.

Moreover, the screen with the program to move the detector seems to be bugged (gray background, a double window): however, the potentiometer value is chnaging and I see the usual effect of detector moving on the statistics, so i think it is fine

3:38

Checks Ok. Target 2.5E11 at/cm2

4:05

The bug disappeared

4:56

Target density is now 1.9E11 at/cm2. It is very low with respect to before, but our reasoning is that nothing can be done about it at night anyway (apart from moving the nozzle, which we tried without effect). O. pointed out that the beam lifetime is still such that the beam is gone before the next injection.

Checks ok. Attachments 21-22 spectra and xy plot. Some structure in top left?

06:00

All check ok. Target density 1.8E11 at/cm2

6:49

all goo, target 1.7E11 at/cm2

07:30

All checks ok, except timestamp (15 passed, 1 failed, see screenshot)

8:30

Tried moving the nozzle. Minor improvement in target density only. x=-1000 y=-4200

Stopped writing to disk while moving nozzle and stopped the motors. Restarted everything and moved to run 29 - no differences vs. 28.

09:11

Temps OK.

Nautilus does not respond to user commands. Attempint to make it work again without rebooting.

10:05

Stopped to reconsider which arm we are moving. Started to move arms around. Decided to stick with left arm.

Started run 30. Same as 28 and 29.

 

10:30

Temps OK - attach

Rates OK - attach

Stats OK (taken with arm in) - attach

Current histogram attached

 

11:40

Something happened to leakage currents. We see an increase in the top-right detector (which recovered) and a decrease in the bottom-right. Neither is moving.

No changes to left arm detectors, which are moving.

Temperature in the cave stable. No effects on the DAQ.

Temps OK

Rates in aida06 at 10k. Fast discriminator at 100k. Checkloaded. OK. Other rates fine.

Will keep monitoring.

 

11:50

aida06 fast discriminator over 100k. Checkloaded x2. OK now. Other rates fine.

No changes to leakage current.

 

13:20

Temps OK

Rates OK (aida 06 is no longer counting fast)

Leakage current stable

 

14:20

Run continues OK

Temps OK

Stats OK


Attachment 1: Screenshot_2025-05-28_17-51-17.png
Screenshot_2025-05-28_17-51-17.png
Attachment 2: Screenshot_from_2025-05-28_17-47-21.png
Screenshot_from_2025-05-28_17-47-21.png
Attachment 3: Screenshot_from_2025-05-28_17-49-00.png
Screenshot_from_2025-05-28_17-49-00.png
Attachment 4: Screenshot_from_2025-05-28_17-49-11.png
Screenshot_from_2025-05-28_17-49-11.png
Attachment 5: Screenshot_from_2025-05-28_17-49-30.png
Screenshot_from_2025-05-28_17-49-30.png
Attachment 6: Screenshot_from_2025-05-28_17-49-41.png
Screenshot_from_2025-05-28_17-49-41.png
Attachment 7: Screenshot_from_2025-05-28_18-52-35.png
Screenshot_from_2025-05-28_18-52-35.png
Attachment 8: Screenshot_from_2025-05-28_18-52-46.png
Screenshot_from_2025-05-28_18-52-46.png
Attachment 9: Screenshot_from_2025-05-28_18-52-56.png
Screenshot_from_2025-05-28_18-52-56.png
Attachment 10: Screenshot_from_2025-05-28_18-53-04.png
Screenshot_from_2025-05-28_18-53-04.png
Attachment 11: Screenshot_from_2025-05-28_18-53-34.png
Screenshot_from_2025-05-28_18-53-34.png
Attachment 12: Screenshot_from_2025-05-28_20-21-22.png
Screenshot_from_2025-05-28_20-21-22.png
Attachment 13: Screenshot_from_2025-05-28_22-04-57.png
Screenshot_from_2025-05-28_22-04-57.png
Attachment 14: Screenshot_from_2025-05-28_22-04-40.png
Screenshot_from_2025-05-28_22-04-40.png
Attachment 15: Screenshot_from_2025-05-28_22-04-26.png
Screenshot_from_2025-05-28_22-04-26.png
Attachment 16: Screenshot_from_2025-05-28_22-04-36.png
Screenshot_from_2025-05-28_22-04-36.png
Attachment 17: Screenshot_from_2025-05-28_22-04-52.png
Screenshot_from_2025-05-28_22-04-52.png
Attachment 18: Screenshot_from_2025-05-29_01-43-22.png
Screenshot_from_2025-05-29_01-43-22.png
Attachment 19: Screenshot_from_2025-05-29_02-53-41.png
Screenshot_from_2025-05-29_02-53-41.png
Attachment 20: Screenshot_from_2025-05-29_02-54-03.png
Screenshot_from_2025-05-29_02-54-03.png
Attachment 21: Screenshot_from_2025-05-29_05-00-39.png
Screenshot_from_2025-05-29_05-00-39.png
Attachment 22: Screenshot_from_2025-05-29_05-00-13.png
Screenshot_from_2025-05-29_05-00-13.png
Attachment 23: Screenshot_from_2025-05-29_07-31-25.png
Screenshot_from_2025-05-29_07-31-25.png
Attachment 24: Screenshot_from_2025-05-29_07-30-25.png
Screenshot_from_2025-05-29_07-30-25.png
Attachment 25: Screenshot_from_2025-05-29_07-30-10.png
Screenshot_from_2025-05-29_07-30-10.png
Attachment 26: Screenshot_from_2025-05-29_07-29-42.png
Screenshot_from_2025-05-29_07-29-42.png
Attachment 27: Screenshot_from_2025-05-29_08-41-51.png
Screenshot_from_2025-05-29_08-41-51.png
Attachment 28: Screenshot_from_2025-05-29_10-30-00.png
Screenshot_from_2025-05-29_10-30-00.png
Attachment 29: Screenshot_from_2025-05-29_10-30-03.png
Screenshot_from_2025-05-29_10-30-03.png
Attachment 30: Screenshot_from_2025-05-29_10-30-16.png
Screenshot_from_2025-05-29_10-30-16.png
Attachment 31: Screenshot_from_2025-05-29_10-30-19.png
Screenshot_from_2025-05-29_10-30-19.png
Attachment 32: screenshot-2025-05-29_11-13-385986126969656182153.png
screenshot-2025-05-29_11-13-385986126969656182153.png
Attachment 33: Screenshot_from_2025-05-29_11-38-07.png
Screenshot_from_2025-05-29_11-38-07.png
Attachment 34: Screenshot_from_2025-05-29_11-38-10.png
Screenshot_from_2025-05-29_11-38-10.png
Attachment 35: Screenshot_from_2025-05-29_11-44-55.png
Screenshot_from_2025-05-29_11-44-55.png
Attachment 36: Screenshot_from_2025-05-29_11-44-55.png
Screenshot_from_2025-05-29_11-44-55.png
Attachment 37: Screenshot_from_2025-05-29_14-19-52.png
Screenshot_from_2025-05-29_14-19-52.png
  699   Wed May 28 10:54:14 2025 LBRun 27

Run 27 running

28/05/2025

(ignore 26 - in-between run)

We start a new run with real data taking condititon: beam is ok, target density is 9.6e11 cm-2, detectors are moving in-out periodically.

Cooler voltage setpoint 1128 V.

Beam energy = 1.99 MeV/u (from HKR) (same as run 25)

Left arm at 18.86 mm

Right  arm IN at 83.69 mm (OUT 37.96 mm)

vertical beam position 0.0 mm

horizontal beam position  -3.5mm

Beam intensity ~1e+7 particles.

Target density ~9.6e11 at/cm2

Target temperatures: T1=31.5 K T2=35.2 K

Beam cycle set to 10 seconds.

11;57

All checks OK.

Aida10's rates seems low now, but aida07 is instead jumping between 200 and 4000 following beam.

 

15:45

Root histograms refreshed.

 

Temps OK - attach

Stats OK - attach

All checks OK

Leakage OK - attach


             Base         Current         Difference
aida01 fault      0x0 :      0x1 :      1  
FPGA Timestamp error counter test result: Passed 15, Failed 1
If any of these counts are reported as in error
The ASIC readout system has detected a timeslip.
That is the timestamp read from the time FIFO is not younger than the last

 

 

Attachment 1: Screenshot_from_2025-05-28_15-45-13.png
Screenshot_from_2025-05-28_15-45-13.png
Attachment 2: Screenshot_from_2025-05-28_15-46-07.png
Screenshot_from_2025-05-28_15-46-07.png
Attachment 3: Screenshot_from_2025-05-28_15-48-27.png
Screenshot_from_2025-05-28_15-48-27.png
Attachment 4: Screenshot_from_2025-05-28_15-48-38.png
Screenshot_from_2025-05-28_15-48-38.png
Attachment 5: Screenshot_from_2025-05-28_15-49-09.png
Screenshot_from_2025-05-28_15-49-09.png
  698   Tue May 27 22:20:11 2025 CB, LB,PWRun 25 - 1.99/2.02 MeV/A

23:20

JM thinks we cannot see 7Be locus on run 24. Possibly the right arm cannot get close enough.

Stopped moving right arm

Started moving left arm as previously: 32.52 mm -> 77.56 mm

Checked with bunched beam: not clipping. Possibly we could go further in - to be decided.

 

Started run 25

Target - 1.1E12 atoms/cm2

Beam - 1.2E7 pps

 

Root histograms refreshed.

 

Temps OK - attach

Stats OK - attach

All checks OK

Leakage OK - attach

Spectrum browser 1, 7, 8 attached

 

00:00

Run continues OK

Temps OK, Stats OK, Merger OK

 

01:15

Run continues OK

Temps OK - attached

Stats OK - attached

Merger OK

Rates OK - attached

Also attached current histograms

 

03:10

Motors now moving 32.77 -> 77.56 mm. Unclear why, should not affect results.

Target 1.1E12 atoms/cm2

Current 1.4E7 pps

 

Temps OK - attached

Rates OK - attached

All checks OK

Merger OK

 

03:37

7Be locus appears to be clearly visible. Attach 18

 

04:32

Temps OK

Rates OK

Merger OK

 

05:30

Temps OK

Rates - aida10 showes intermittently high rates around 10k. See attached. Seems to be linked to beam, already observed last night.

 

06:10

Temps OK

Rates OK - aida10 still counting 10k with beam, 300 without

Merger OK

Histograms attached

 

7:23- LB,PW

All checks OK, general behaviour consistent with previous observations. Beam current ~ 1E7 pps

8:40

All good, attachment 27 is the spectra. Beam current 1.3E7 pps.

9;50

All good, target very stable at 1E12 at/cm^2. Noticed also Aida01 and aida02 show same behaviour as aida10, but the maximum rate is between 1000-2000 (instead of the 10000-20000 of aida10)

 

Run 25 stopped 10:45, Run 26 started -> laser spectroscopy laser test

Attachment 1: Screenshot_from_2025-05-27_23-21-58.png
Screenshot_from_2025-05-27_23-21-58.png
Attachment 2: Screenshot_from_2025-05-27_23-22-16.png
Screenshot_from_2025-05-27_23-22-16.png
Attachment 3: Screenshot_from_2025-05-27_23-23-12.png
Screenshot_from_2025-05-27_23-23-12.png
Attachment 4: Screenshot_from_2025-05-27_23-23-35.png
Screenshot_from_2025-05-27_23-23-35.png
Attachment 5: Screenshot_from_2025-05-27_23-23-44.png
Screenshot_from_2025-05-27_23-23-44.png
Attachment 6: Screenshot_from_2025-05-27_23-24-18.png
Screenshot_from_2025-05-27_23-24-18.png
Attachment 7: Screenshot_from_2025-05-27_23-24-32.png
Screenshot_from_2025-05-27_23-24-32.png
Attachment 8: Screenshot_from_2025-05-27_23-24-58.png
Screenshot_from_2025-05-27_23-24-58.png
Attachment 9: Screenshot_from_2025-05-28_01-13-04.png
Screenshot_from_2025-05-28_01-13-04.png
Attachment 10: Screenshot_from_2025-05-28_01-13-16.png
Screenshot_from_2025-05-28_01-13-16.png
Attachment 11: Screenshot_from_2025-05-28_01-13-49.png
Screenshot_from_2025-05-28_01-13-49.png
Attachment 12: Screenshot_from_2025-05-28_01-14-01.png
Screenshot_from_2025-05-28_01-14-01.png
Attachment 13: Screenshot_from_2025-05-28_01-14-09.png
Screenshot_from_2025-05-28_01-14-09.png
Attachment 14: Screenshot_from_2025-05-28_01-14-14.png
Screenshot_from_2025-05-28_01-14-14.png
Attachment 15: Screenshot_from_2025-05-28_01-14-20.png
Screenshot_from_2025-05-28_01-14-20.png
Attachment 16: Screenshot_from_2025-05-28_03-09-54.png
Screenshot_from_2025-05-28_03-09-54.png
Attachment 17: Screenshot_from_2025-05-28_03-10-12.png
Screenshot_from_2025-05-28_03-10-12.png
Attachment 18: Screenshot_from_2025-05-28_03-47-08.png
Screenshot_from_2025-05-28_03-47-08.png
Attachment 19: Screenshot_from_2025-05-28_05-30-00.png
Screenshot_from_2025-05-28_05-30-00.png
Attachment 20: Screenshot_from_2025-05-28_05-30-11.png
Screenshot_from_2025-05-28_05-30-11.png
Attachment 21: Screenshot_from_2025-05-28_05-30-20.png
Screenshot_from_2025-05-28_05-30-20.png
Attachment 22: Screenshot_from_2025-05-28_06-09-33.png
Screenshot_from_2025-05-28_06-09-33.png
Attachment 23: Screenshot_from_2025-05-28_06-09-40.png
Screenshot_from_2025-05-28_06-09-40.png
Attachment 24: Screenshot_from_2025-05-28_06-09-43.png
Screenshot_from_2025-05-28_06-09-43.png
Attachment 25: Screenshot_from_2025-05-28_06-09-46.png
Screenshot_from_2025-05-28_06-09-46.png
Attachment 26: Screenshot_from_2025-05-28_06-09-53.png
Screenshot_from_2025-05-28_06-09-53.png
Attachment 27: Screenshot_from_2025-05-28_08-38-27.png
Screenshot_from_2025-05-28_08-38-27.png
  697   Tue May 27 15:18:46 2025 JG, CN, JMrun 24

Run 24 running

27/05/2025

We start a new run with real data taking condititon: beam is ok, target density is 5e11 cm-2, detectors are moving in-out periodically.

Cooler voltage setpoint 1128.55 V.

Beam energy = 1.99 MeV/u (from HKR)

Attachment 1 - beamcalc with set values

Attachment 7 - beamcalc with actual values

 

Left arm at 18.86 mm

Right  arm IN at 83.69 mm (OUT 37.56 mm)

vertical beam position 0.0 mm

horizontal beam position  -3.5mm

Beam intensity ~2e+7 particles.

Target density ~5e11 at/cm2

Target temperatures: T1=31.5 K T2=35.1 K

Beam cycle set to 10 seconds.


17:32 all checks OK

17:35 target density jumped to at 9e11

19:30 all checks OK (spectra attached)

19:45 target density drops to 3.5E11

20:20 target density jumps to 7e11

21:30 all checks OK

22:50 all checks OK

 

Attachment 1: Screenshot_2025-05-27_16-17-15.png
Screenshot_2025-05-27_16-17-15.png
Attachment 2: 2025-05-27_16-08-58_YRT1_CRYRING_2025_Li_Marsh.C1.txt
# Version: 1.3
# Comment: ParamModi Export at 2025-05-27T16:09:10.333896316+02:00[Europe/Berlin]
# Context: YRT1_CRYRING_2025_Li_Marsh.C1
# Columns: PARAMETERNAME,VALUE,UNIT,BEAMPROCESSPURPOSE
LOGICAL.SOURCE_HV/U,35000.0,V,TRANSFER_INJECTION
YRT1_TO_YRT1LQ1_BEAM/BEAM_MODE_S,NO_BEAM,,TRANSFER_INJECTION
YRT1_TO_YRT1LQ1_BEAM/PARTICLE_SYMBOL_S,7Li,,TRANSFER_INJECTION
YRT1_TO_YRT1LQ1_BEAM/Q_S,1,,TRANSFER_INJECTION
YRT1_TO_YRT1LQ1_TIMEPARAM/TIMING_BI_DELAY_S,2.0E-4,s,TRANSFER_INJECTION
YRT1_TO_YRT1LQ1_TIMEPARAM/TIMING_BI_WINDOW_S,0.002,s,TRANSFER_INJECTION
YRT1_TO_YRT1LQ1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,BEAMOUT_INIT
YRT1_TO_YRT1LQ1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,TRANSFER_INJECTION
YRT1_TO_YRT1LQ1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,BEAMOUT_RESET
YRT1_TO_YRT1LQ1_TIMEPARAM/TIMING_SOURCE_PULSE_LENGTH_S,0.0024,s,TRANSFER_INJECTION
YRT1_TO_YRT1LQ1_TIMEPARAM/T_BP_LENGTH_S,0.12,s,BEAMOUT_INIT
YRT1_TO_YRT1LQ1_TIMEPARAM/T_BP_LENGTH_S,0.11,s,TRANSFER_INJECTION
YRT1_TO_YRT1LQ1_TIMEPARAM/T_BP_LENGTH_S,---,s,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/T_BP_LENGTH_S,---,s,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/T_BP_LENGTH_S,---,s,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/T_BP_LENGTH_S,---,s,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/T_BP_LENGTH_S,---,s,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/T_BP_LENGTH_S,0.0,s,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/T_BP_LENGTH_S,0.0,s,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/T_BP_LENGTH_S,0.0,s,STORAGE_MANIP_PERFORM
YRT1_TO_YRT1LQ1_TIMEPARAM/T_BP_LENGTH_S,---,s,STORAGE_MANIP_WAIT
YRT1_TO_YRT1LQ1_TIMEPARAM/T_BP_LENGTH_S,---,s,STORAGE_MANIP_RESTART_FGS
YRT1_TO_YRT1LQ1_TIMEPARAM/T_BP_LENGTH_S,---,s,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/T_BP_LENGTH_S,---,s,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/T_BP_LENGTH_S,0.12,s,BEAMOUT_RESET
YRT1LQ1_TO_YRT1LC1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,BEAMOUT_INIT
YRT1LQ1_TO_YRT1LC1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,TRANSFER_INJECTION
YRT1LQ1_TO_YRT1LC1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LQ1_TO_YRT1LC1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LQ1_TO_YRT1LC1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LQ1_TO_YRT1LC1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LQ1_TO_YRT1LC1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LQ1_TO_YRT1LC1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LQ1_TO_YRT1LC1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LQ1_TO_YRT1LC1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LQ1_TO_YRT1LC1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LQ1_TO_YRT1LC1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,BEAMOUT_RESET
LOGICAL.YRT1KH1/HKICK,-0.018,rad,TRANSFER_INJECTION
LOGICAL.YRT1KV1/VKICK,0.004,rad,TRANSFER_INJECTION
LOGICAL.YRT1LC1_V/ACTIVE_S,true,,TRANSFER_INJECTION
LOGICAL.YRT1LC1_V/THIGH_S,6.0E-4,s,TRANSFER_INJECTION
LOGICAL.YRT1LC1_V/TOFFSET_S,5.0E-4,s,TRANSFER_INJECTION
LOGICAL.YRT1LC1_V/U,2000.0,V,TRANSFER_INJECTION
LOGICAL.YRT1LD21/ASYMMETRY_KNOB_S,0.0,,TRANSFER_INJECTION
LOGICAL.YRT1LD21/HOR_STEER_KNOB_S,0.0,,TRANSFER_INJECTION
LOGICAL.YRT1LD21/U_QUAD,-640.0,V,TRANSFER_INJECTION
LOGICAL.YRT1LD21/VER_STEER_KNOB_S,-0.03,,TRANSFER_INJECTION
LOGICAL.YRT1LD22/ASYMMETRY_KNOB_S,0.0,,TRANSFER_INJECTION
LOGICAL.YRT1LD22/U_QUAD,650.0,V,TRANSFER_INJECTION
LOGICAL.YRT1LT31/ASYMMETRY_KNOB_S,0.0,,TRANSFER_INJECTION
LOGICAL.YRT1LT31/U_QUAD,515.0,V,TRANSFER_INJECTION
LOGICAL.YRT1LT32/ASYMMETRY_KNOB_S,0.0,,TRANSFER_INJECTION
LOGICAL.YRT1LT32/U_QUAD,-945.0,V,TRANSFER_INJECTION
LOGICAL.YRT1LT33/ASYMMETRY_KNOB_S,0.0,,TRANSFER_INJECTION
LOGICAL.YRT1LT33/U_QUAD,1210.0,V,TRANSFER_INJECTION
LOGICAL.YRT1LT41/ASYMMETRY_KNOB_S,0.0,,TRANSFER_INJECTION
LOGICAL.YRT1LT41/U_QUAD,-2100.0,V,TRANSFER_INJECTION
LOGICAL.YRT1LT42/ASYMMETRY_KNOB_S,0.0,,TRANSFER_INJECTION
LOGICAL.YRT1LT42/U_QUAD,2020.0,V,TRANSFER_INJECTION
LOGICAL.YRT1LT43/ASYMMETRY_KNOB_S,0.0,,TRANSFER_INJECTION
LOGICAL.YRT1LT43/U_QUAD,-2100.0,V,TRANSFER_INJECTION
LOGICAL.YRT1MH1/HKICK,0.022,rad,TRANSFER_INJECTION
LOGICAL.YRT1MH1/K0L_THEO,-1.57079,rad,TRANSFER_INJECTION
YRT1LC1_TO_YRT1MH2_TIMEPARAM/TIMING_BI_DELAY_S,1.0E-6,s,TRANSFER_INJECTION
YRT1LC1_TO_YRT1MH2_TIMEPARAM/TIMING_BI_WINDOW_S,1.0E-6,s,TRANSFER_INJECTION
YRT1LC1_TO_YRT1MH2_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,BEAMOUT_INIT
YRT1LC1_TO_YRT1MH2_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,TRANSFER_INJECTION
YRT1LC1_TO_YRT1MH2_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LC1_TO_YRT1MH2_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LC1_TO_YRT1MH2_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LC1_TO_YRT1MH2_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LC1_TO_YRT1MH2_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LC1_TO_YRT1MH2_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LC1_TO_YRT1MH2_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LC1_TO_YRT1MH2_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LC1_TO_YRT1MH2_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LC1_TO_YRT1MH2_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,BEAMOUT_RESET
YRT1LC1_TO_YRT1MH2_TIMEPARAM/TIMING_RF_LEAD_S,4.0E-4,s,TRANSFER_INJECTION
LOGICAL.YRT1BR1/ACT_ON_BEAM_S,true,,TRANSFER_INJECTION
LOGICAL.YRT1BR1/THIGH_S,0.0025,s,TRANSFER_INJECTION
LOGICAL.YRT1BR1/URF0_S,0.1,V,TRANSFER_INJECTION
LOGICAL.YRT1BR1/U_PAUSE_PULSE_AMPL_S,4.76,V,TRANSFER_INJECTION
LOGICAL.YRT1KH2/HKICK,3.0E-4,rad,TRANSFER_INJECTION
LOGICAL.YRT1KV2/VKICK,-2.0E-4,rad,TRANSFER_INJECTION
LOGICAL.YRT1LD51/ASYMMETRY_KNOB_S,0.0,,TRANSFER_INJECTION
LOGICAL.YRT1LD51/U_QUAD,510.0,V,TRANSFER_INJECTION
LOGICAL.YRT1LD52/ASYMMETRY_KNOB_S,0.0,,TRANSFER_INJECTION
LOGICAL.YRT1LD52/U_QUAD,-420.0,V,TRANSFER_INJECTION
LOGICAL.YRT1QD61/K1L,2.05,1/m,TRANSFER_INJECTION
LOGICAL.YRT1QD62/K1L,-1.68,1/m,TRANSFER_INJECTION
LOGICAL.YRT1QD71/K1L,1.85,1/m,TRANSFER_INJECTION
LOGICAL.YRT1QD72/K1L,-2.1,1/m,TRANSFER_INJECTION
YRT1BR1_TO_YRT1MH2_BEAM/E_S,5000.0,eV/u,TRANSFER_INJECTION
YRT1BR1_TO_YRT1MH2_RF/PHASE_OFFSET_S,0.0,deg,TRANSFER_INJECTION
LOGICAL.GHTYKH4/HKICK,-2.0E-4,rad,TRANSFER_INJECTION
LOGICAL.GHTYKV3/VKICK,-0.002,rad,TRANSFER_INJECTION
LOGICAL.GHTYKV4/VKICK,0.0012,rad,TRANSFER_INJECTION
LOGICAL.GHTYQD41/K1L,1.4,1/m,TRANSFER_INJECTION
LOGICAL.GHTYQD42/K1L,-1.25,1/m,TRANSFER_INJECTION
LOGICAL.YR01LP1I/ESEP_HKICK_S,0.017,rad,TRANSFER_INJECTION
LOGICAL.YR01LP1I/ESEP_K0L_THEO_S,-0.05,rad,TRANSFER_INJECTION
LOGICAL.YR01LP2I/ESEP_HKICK_S,-0.035,rad,TRANSFER_INJECTION
LOGICAL.YR01LP2I/ESEP_K0L_THEO_S,0.055,rad,TRANSFER_INJECTION
LOGICAL.YR01MP1I/HKICK,0.0035,rad,TRANSFER_INJECTION
LOGICAL.YR01MP1I/K0L_THEO,0.29322,rad,TRANSFER_INJECTION
LOGICAL.YRT1MH2/HKICK_ESR_S,-0.61087,rad,TRANSFER_INJECTION
LOGICAL.YRT1MH2/HKICK_LOCAL_S,-0.021,rad,TRANSFER_INJECTION
LOGICAL.YRT1MH2/K0L_THEO,-0.61087,rad,TRANSFER_INJECTION
YRT1MH2_TO_CRYRING_TIMEPARAM/TIMING_BI_DELAY_S,1.0E-6,s,TRANSFER_INJECTION
YRT1MH2_TO_CRYRING_TIMEPARAM/TIMING_BI_WINDOW_S,1.0E-6,s,TRANSFER_INJECTION
YRT1MH2_TO_CRYRING_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,BEAMOUT_INIT
YRT1MH2_TO_CRYRING_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,TRANSFER_INJECTION
YRT1MH2_TO_CRYRING_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1MH2_TO_CRYRING_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1MH2_TO_CRYRING_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1MH2_TO_CRYRING_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1MH2_TO_CRYRING_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1MH2_TO_CRYRING_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1MH2_TO_CRYRING_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1MH2_TO_CRYRING_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1MH2_TO_CRYRING_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1MH2_TO_CRYRING_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,BEAMOUT_RESET
YRT1MH2_TO_CRYRING_TIMEPARAM/TIMING_SEPTUM_TRIGGER_OFFSET_S,0.0,s,TRANSFER_INJECTION
CRYRINGB2B/B2B_SID_EXT_S,---,,RING_INJECTION
CRYRINGB2B/B2B_SOURCE_KICKER_DELAY_S,---,s,RING_INJECTION
CRYRINGB2B/B2B_SOURCE_TRIGGER_OFFSET_S,---,deg,RING_INJECTION
CRYRINGB2B/B2B_TRIGGER_OFFSET_S,0.0,deg,RING_INJECTION
CRYRINGB2B/B2B_TRIGGER_OFFSET_TIME_S,0.0,,RING_INJECTION
CRYRINGBEAM/BEAM_MODE_S,NO_BEAM,,RING_INJECTION
CRYRINGBEAM/BRHO_CHIMNEY_S,---,Tm,BEAMOUT_INIT
CRYRINGBEAM/BRHO_CHIMNEY_S,---,Tm,BEAMOUT_RESET
CRYRINGBEAM/BRHO_DISPLAY_S,0.0713,Tm,RING_INJECTION
CRYRINGBEAM/BRHO_DISPLAY_S,1.0103,Tm,RING_RF_MANIPULATION
CRYRINGBEAM/BRHO_DISPLAY_S,1.4256,Tm,RING_TARGET
CRYRINGBEAM/BRHO_START_END_S,[0.07134952, 0.07134952],Tm,RING_INJECTION
CRYRINGBEAM/BUCKETFILL_RAW_S,[[0.0, 0.0]],,RING_INJECTION
CRYRINGBEAM/BUCKETFILL_RAW_S,[[0.0, 1.4]],,RING_BUNCHING
CRYRINGBEAM/BUCKETFILL_RAW_S,[[1.4, 1.4]],,RING_RAMP
CRYRINGBEAM/BUCKETFILL_RAW_S,[[1.4, 1.4]],,RING_RF_MANIPULATION
CRYRINGBEAM/BUCKETFILL_RAW_S,[[1.4, 1.4]],,RING_RAMP
CRYRINGBEAM/BUCKETFILL_RAW_S,[[1.4, 1.4]],,RING_REBUNCHING
CRYRINGBEAM/BUCKETFILL_RAW_S,[[1.4, 1.4]],,RING_TARGET
CRYRINGBEAM/BUCKETFILL_RAW_S,[[1.4, 1.4]],,RING_INTERMEDIATE
CRYRINGBEAM/BUCKETFILL_SIMPLE_S,1.4,,RING_INJECTION
CRYRINGBEAM/CHIMNEY_B,---,,BEAMOUT_INIT
CRYRINGBEAM/CHIMNEY_B,---,,BEAMOUT_RESET
CRYRINGBEAM/CH_END_S,5.0655,,RING_INJECTION
CRYRINGBEAM/CH_END_S,5.0655,,RING_RAMP
CRYRINGBEAM/CH_END_S,5.0655,,RING_RAMP
CRYRINGBEAM/CH_END_S,---,,STORAGE_MANIP_WAIT
CRYRINGBEAM/COASTING_BEAM_B,false,,RING_REBUNCHING
CRYRINGBEAM/CV_END_S,-5.3921,,RING_INJECTION
CRYRINGBEAM/CV_END_S,-5.3921,,RING_RAMP
CRYRINGBEAM/CV_END_S,-5.3921,,RING_RAMP
CRYRINGBEAM/CV_END_S,---,,STORAGE_MANIP_WAIT
CRYRINGBEAM/DELTAR,5.0E-4,m,RING_INJECTION
CRYRINGBEAM/DELTAR,3.0E-4,m,RING_RF_MANIPULATION
CRYRINGBEAM/DELTAR,0.0,m,STORAGE_MANIP_WAIT
CRYRINGBEAM/DELTAR,0.0,m,RING_TARGET
CRYRINGBEAM/DELTAR,0.0,m,RING_INTERMEDIATE
CRYRINGBEAM/DPBL,0.0,,RING_INJECTION
CRYRINGBEAM/DPBL,0.0,,STORAGE_MANIP_WAIT
CRYRINGBEAM/DPFREV,0.0015,,RING_INJECTION
CRYRINGBEAM/DPFREV,0.002,,RING_RF_MANIPULATION
CRYRINGBEAM/DPFREV,0.013,,STORAGE_MANIP_WAIT
CRYRINGBEAM/DPFREV,0.013,,RING_TARGET
CRYRINGBEAM/DPFREV,0.013,,RING_INTERMEDIATE
CRYRINGBEAM/DP_OVER_P_S,0.004,,RING_INJECTION
CRYRINGBEAM/E_S,4990.0,eV/u,RING_INJECTION
CRYRINGBEAM/E_S,1000000.0,eV/u,RING_RAMP
CRYRINGBEAM/E_S,1990000.0,eV/u,RING_RAMP
CRYRINGBEAM/FINAL_HARMONIC_S,6,,RING_RF_MANIPULATION
CRYRINGBEAM/FREV,18132.5287,Hz,RING_INJECTION
CRYRINGBEAM/FREV,256592.872,Hz,RING_RF_MANIPULATION
CRYRINGBEAM/FREV,364898.6686,Hz,RING_TARGET
CRYRINGBEAM/H_INPUT_S,8,,RING_INJECTION
CRYRINGBEAM/INJ_EMIH_S,1.0E-5,m*rad,RING_INJECTION
CRYRINGBEAM/INJ_EMIL_S,6.160687750547807E-4,eVs/u,RING_INJECTION
CRYRINGBEAM/INJ_EMIV_S,1.0E-5,m*rad,RING_INJECTION
CRYRINGBEAM/KL_AMPL_S,0.0,,RING_INJECTION
CRYRINGBEAM/KL_AMPL_S,0.0,,RING_BUNCHING
CRYRINGBEAM/KL_AMPL_S,0.0,,RING_POST_INJECTION
CRYRINGBEAM/KL_AMPL_S,0.0,,RING_MANI_PRE_RAMP
CRYRINGBEAM/KL_AMPL_S,0.0,,RING_RAMP
CRYRINGBEAM/KL_AMPL_S,0.0,,RING_RF_MANIPULATION
CRYRINGBEAM/KL_AMPL_S,0.0,,RING_RAMP
CRYRINGBEAM/KL_AMPL_S,0.0,,RING_REBUNCHING
CRYRINGBEAM/KL_AMPL_S,0.0,,RING_TARGET
CRYRINGBEAM/KL_AMPL_S,0.0,,RING_INTERMEDIATE
CRYRINGBEAM/KL_HARM_S,1,,RING_INJECTION
CRYRINGBEAM/KL_HARM_S,1,,RING_BUNCHING
CRYRINGBEAM/KL_HARM_S,1,,RING_POST_INJECTION
CRYRINGBEAM/KL_HARM_S,1,,RING_MANI_PRE_RAMP
CRYRINGBEAM/KL_HARM_S,1,,RING_RAMP
CRYRINGBEAM/KL_HARM_S,1,,RING_RF_MANIPULATION
CRYRINGBEAM/KL_HARM_S,1,,RING_RAMP
CRYRINGBEAM/KL_HARM_S,1,,RING_REBUNCHING
CRYRINGBEAM/KL_HARM_S,1,,RING_TARGET
CRYRINGBEAM/KL_HARM_S,1,,RING_INTERMEDIATE
CRYRINGBEAM/KL_OFFSET_S,0.0,,RING_INJECTION
CRYRINGBEAM/KL_OFFSET_S,0.0,,RING_BUNCHING
CRYRINGBEAM/KL_OFFSET_S,0.0,,RING_POST_INJECTION
CRYRINGBEAM/KL_OFFSET_S,0.0,,RING_MANI_PRE_RAMP
CRYRINGBEAM/KL_OFFSET_S,0.0,,RING_RAMP
CRYRINGBEAM/KL_OFFSET_S,0.0,,RING_RF_MANIPULATION
CRYRINGBEAM/KL_OFFSET_S,0.0,,RING_RAMP
CRYRINGBEAM/KL_OFFSET_S,0.0,,RING_REBUNCHING
CRYRINGBEAM/KL_OFFSET_S,0.0,,RING_TARGET
CRYRINGBEAM/KL_OFFSET_S,0.0,,RING_INTERMEDIATE
CRYRINGBEAM/KL_PHASE_S,0.0,deg,RING_INJECTION
CRYRINGBEAM/KL_PHASE_S,0.0,deg,RING_BUNCHING
CRYRINGBEAM/KL_PHASE_S,0.0,deg,RING_POST_INJECTION
CRYRINGBEAM/KL_PHASE_S,0.0,deg,RING_MANI_PRE_RAMP
CRYRINGBEAM/KL_PHASE_S,0.0,deg,RING_RAMP
CRYRINGBEAM/KL_PHASE_S,0.0,deg,RING_RF_MANIPULATION
CRYRINGBEAM/KL_PHASE_S,0.0,deg,RING_RAMP
CRYRINGBEAM/KL_PHASE_S,0.0,deg,RING_REBUNCHING
CRYRINGBEAM/KL_PHASE_S,0.0,deg,RING_TARGET
CRYRINGBEAM/KL_PHASE_S,0.0,deg,RING_INTERMEDIATE
CRYRINGBEAM/NPARTICLES_S,1000.0,,RING_INJECTION
CRYRINGBEAM/PARTICLE_SYMBOL_INPUT_S,7Li,,RING_INJECTION
CRYRINGBEAM/QH_END_S,2.31,,RING_INJECTION
CRYRINGBEAM/QH_END_S,---,,RING_POST_INJECTION
CRYRINGBEAM/QH_END_S,2.29,,RING_RAMP
CRYRINGBEAM/QH_END_S,2.29,,RING_RAMP
CRYRINGBEAM/QH_END_S,---,,STORAGE_MANIP_WAIT
CRYRINGBEAM/QV_END_S,2.55,,RING_INJECTION
CRYRINGBEAM/QV_END_S,---,,RING_POST_INJECTION
CRYRINGBEAM/QV_END_S,2.55,,RING_RAMP
CRYRINGBEAM/QV_END_S,2.55,,RING_RAMP
CRYRINGBEAM/QV_END_S,---,,STORAGE_MANIP_WAIT
CRYRINGBEAM/Q_INPUT_S,1,,RING_INJECTION
CRYRINGBEAM/TREV,5.514951E-5,s,RING_INJECTION
CRYRINGBEAM/USE_SIMPLE_BF_B,true,,RING_INJECTION
CRYRINGBUMPER/TFALL_S,5.0E-4,s,RING_INJECTION
CRYRINGIKICKER/KICKER_TOFFSET1_S,0.0,s,RING_INJECTION
CRYRINGOPTICS/OPTIC_NAME_RING_S,CRYRING_STANDARD_OPTICS_TABLE,,RING_INJECTION
CRYRINGOPTICS/TAU_END_S,0.0,,RING_INJECTION
CRYRINGOPTICS/TAU_END_S,---,,RING_POST_INJECTION
CRYRINGOPTICS/TAU_END_S,0.0,,RING_RAMP
CRYRINGOPTICS/TAU_END_S,0.0,,RING_RAMP
CRYRINGORBIT/COH02,0.0,m,RING_INJECTION
CRYRINGORBIT/COH02,0.0,m,RING_RF_MANIPULATION
CRYRINGORBIT/COH02,0.0,m,STORAGE_MANIP_WAIT
CRYRINGORBIT/COH02,0.0,m,RING_TARGET
CRYRINGORBIT/COH03,0.0,m,RING_INJECTION
CRYRINGORBIT/COH03,0.0,m,RING_RF_MANIPULATION
CRYRINGORBIT/COH03,0.0,m,STORAGE_MANIP_WAIT
CRYRINGORBIT/COH03,0.0,m,RING_TARGET
CRYRINGORBIT/COH04,0.0,m,RING_INJECTION
CRYRINGORBIT/COH04,0.0,m,RING_RF_MANIPULATION
CRYRINGORBIT/COH04,0.0,m,STORAGE_MANIP_WAIT
CRYRINGORBIT/COH04,0.0,m,RING_TARGET
CRYRINGORBIT/COH06,0.0,m,RING_INJECTION
CRYRINGORBIT/COH06,0.0,m,RING_RF_MANIPULATION
CRYRINGORBIT/COH06,0.0,m,STORAGE_MANIP_WAIT
CRYRINGORBIT/COH06,0.0,m,RING_TARGET
CRYRINGORBIT/COH07,0.0,m,RING_INJECTION
CRYRINGORBIT/COH07,0.0,m,RING_RF_MANIPULATION
CRYRINGORBIT/COH07,0.0,m,STORAGE_MANIP_WAIT
CRYRINGORBIT/COH07,0.0,m,RING_TARGET
CRYRINGORBIT/COH08,0.0,m,RING_INJECTION
CRYRINGORBIT/COH08,0.0,m,RING_RF_MANIPULATION
CRYRINGORBIT/COH08,0.0,m,STORAGE_MANIP_WAIT
CRYRINGORBIT/COH08,0.0,m,RING_TARGET
CRYRINGORBIT/COH10,0.0,m,RING_INJECTION
CRYRINGORBIT/COH10,0.0,m,RING_RF_MANIPULATION
CRYRINGORBIT/COH10,0.0,m,STORAGE_MANIP_WAIT
CRYRINGORBIT/COH10,0.0,m,RING_TARGET
CRYRINGORBIT/COH11,0.0,m,RING_INJECTION
CRYRINGORBIT/COH11,0.0,m,RING_RF_MANIPULATION
CRYRINGORBIT/COH11,0.0,m,STORAGE_MANIP_WAIT
CRYRINGORBIT/COH11,0.0,m,RING_TARGET
CRYRINGORBIT/COH12,0.0,m,RING_INJECTION
CRYRINGORBIT/COH12,0.0,m,RING_RF_MANIPULATION
CRYRINGORBIT/COH12,0.0,m,STORAGE_MANIP_WAIT
CRYRINGORBIT/COH12,0.0,m,RING_TARGET
CRYRINGORBIT/COV02,0.0,m,RING_INJECTION
CRYRINGORBIT/COV02,0.0,m,RING_RF_MANIPULATION
CRYRINGORBIT/COV02,0.0,m,STORAGE_MANIP_WAIT
CRYRINGORBIT/COV02,0.0,m,RING_TARGET
CRYRINGORBIT/COV03,0.0,m,RING_INJECTION
CRYRINGORBIT/COV03,0.0,m,RING_RF_MANIPULATION
CRYRINGORBIT/COV03,0.0,m,STORAGE_MANIP_WAIT
CRYRINGORBIT/COV03,0.0,m,RING_TARGET
CRYRINGORBIT/COV04,0.0,m,RING_INJECTION
CRYRINGORBIT/COV04,0.0,m,RING_RF_MANIPULATION
... 679 more lines ...
Attachment 3: Screenshot_from_2025-05-27_19-44-25.png
Screenshot_from_2025-05-27_19-44-25.png
Attachment 4: Screenshot_from_2025-05-27_19-44-42.png
Screenshot_from_2025-05-27_19-44-42.png
Attachment 5: Screenshot_from_2025-05-27_19-44-55.png
Screenshot_from_2025-05-27_19-44-55.png
Attachment 6: Screenshot_from_2025-05-27_19-45-06.png
Screenshot_from_2025-05-27_19-45-06.png
Attachment 7: screenshot-2025-05-29_11-13-385986126969656182153.png
screenshot-2025-05-29_11-13-385986126969656182153.png
  696   Tue May 27 12:49:48 2025 MLIon beam / electron beam settings so far.

Screenshots and ParamModi settings export attached.

Attachment 1: 2025-05-27_13-46-30-866.png
2025-05-27_13-46-30-866.png
Attachment 2: 2025-05-27_13-47-04-130.png
2025-05-27_13-47-04-130.png
Attachment 3: 2025-05-27_13-47-23-173.png
2025-05-27_13-47-23-173.png
Attachment 4: 2025-05-27_13-47-52-384.png
2025-05-27_13-47-52-384.png
Attachment 5: 2025-05-27_13-50-26_YRT1_CRYRING_2025_Li_Marsh.C1.txt
# Version: 1.3
# Comment: ParamModi Export at 2025-05-27T13:50:34.912176166+02:00[Europe/Berlin]
# Context: YRT1_CRYRING_2025_Li_Marsh.C1
# Columns: PARAMETERNAME,VALUE,UNIT,BEAMPROCESSPURPOSE
LOGICAL.SOURCE_HV/U,35000.0,V,TRANSFER_INJECTION
YRT1_TO_YRT1LQ1_BEAM/BEAM_MODE_S,NO_BEAM,,TRANSFER_INJECTION
YRT1_TO_YRT1LQ1_BEAM/PARTICLE_SYMBOL_S,7Li,,TRANSFER_INJECTION
YRT1_TO_YRT1LQ1_BEAM/Q_S,1,,TRANSFER_INJECTION
YRT1_TO_YRT1LQ1_TIMEPARAM/TIMING_BI_DELAY_S,2.0E-4,s,TRANSFER_INJECTION
YRT1_TO_YRT1LQ1_TIMEPARAM/TIMING_BI_WINDOW_S,0.002,s,TRANSFER_INJECTION
YRT1_TO_YRT1LQ1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,BEAMOUT_INIT
YRT1_TO_YRT1LQ1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,TRANSFER_INJECTION
YRT1_TO_YRT1LQ1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,BEAMOUT_RESET
YRT1_TO_YRT1LQ1_TIMEPARAM/TIMING_SOURCE_PULSE_LENGTH_S,0.0024,s,TRANSFER_INJECTION
YRT1_TO_YRT1LQ1_TIMEPARAM/T_BP_LENGTH_S,0.12,s,BEAMOUT_INIT
YRT1_TO_YRT1LQ1_TIMEPARAM/T_BP_LENGTH_S,0.11,s,TRANSFER_INJECTION
YRT1_TO_YRT1LQ1_TIMEPARAM/T_BP_LENGTH_S,---,s,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/T_BP_LENGTH_S,---,s,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/T_BP_LENGTH_S,---,s,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/T_BP_LENGTH_S,---,s,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/T_BP_LENGTH_S,---,s,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/T_BP_LENGTH_S,0.0,s,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/T_BP_LENGTH_S,0.0,s,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/T_BP_LENGTH_S,0.0,s,STORAGE_MANIP_PERFORM
YRT1_TO_YRT1LQ1_TIMEPARAM/T_BP_LENGTH_S,---,s,STORAGE_MANIP_WAIT
YRT1_TO_YRT1LQ1_TIMEPARAM/T_BP_LENGTH_S,---,s,STORAGE_MANIP_RESTART_FGS
YRT1_TO_YRT1LQ1_TIMEPARAM/T_BP_LENGTH_S,---,s,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/T_BP_LENGTH_S,---,s,STATIC
YRT1_TO_YRT1LQ1_TIMEPARAM/T_BP_LENGTH_S,0.12,s,BEAMOUT_RESET
YRT1LQ1_TO_YRT1LC1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,BEAMOUT_INIT
YRT1LQ1_TO_YRT1LC1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,TRANSFER_INJECTION
YRT1LQ1_TO_YRT1LC1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LQ1_TO_YRT1LC1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LQ1_TO_YRT1LC1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LQ1_TO_YRT1LC1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LQ1_TO_YRT1LC1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LQ1_TO_YRT1LC1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LQ1_TO_YRT1LC1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LQ1_TO_YRT1LC1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LQ1_TO_YRT1LC1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LQ1_TO_YRT1LC1_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,BEAMOUT_RESET
LOGICAL.YRT1KH1/HKICK,-0.018,rad,TRANSFER_INJECTION
LOGICAL.YRT1KV1/VKICK,0.004,rad,TRANSFER_INJECTION
LOGICAL.YRT1LC1_V/ACTIVE_S,true,,TRANSFER_INJECTION
LOGICAL.YRT1LC1_V/THIGH_S,6.0E-4,s,TRANSFER_INJECTION
LOGICAL.YRT1LC1_V/TOFFSET_S,5.0E-4,s,TRANSFER_INJECTION
LOGICAL.YRT1LC1_V/U,2000.0,V,TRANSFER_INJECTION
LOGICAL.YRT1LD21/ASYMMETRY_KNOB_S,0.0,,TRANSFER_INJECTION
LOGICAL.YRT1LD21/HOR_STEER_KNOB_S,0.0,,TRANSFER_INJECTION
LOGICAL.YRT1LD21/U_QUAD,-640.0,V,TRANSFER_INJECTION
LOGICAL.YRT1LD21/VER_STEER_KNOB_S,-0.03,,TRANSFER_INJECTION
LOGICAL.YRT1LD22/ASYMMETRY_KNOB_S,0.0,,TRANSFER_INJECTION
LOGICAL.YRT1LD22/U_QUAD,650.0,V,TRANSFER_INJECTION
LOGICAL.YRT1LT31/ASYMMETRY_KNOB_S,0.0,,TRANSFER_INJECTION
LOGICAL.YRT1LT31/U_QUAD,515.0,V,TRANSFER_INJECTION
LOGICAL.YRT1LT32/ASYMMETRY_KNOB_S,0.0,,TRANSFER_INJECTION
LOGICAL.YRT1LT32/U_QUAD,-945.0,V,TRANSFER_INJECTION
LOGICAL.YRT1LT33/ASYMMETRY_KNOB_S,0.0,,TRANSFER_INJECTION
LOGICAL.YRT1LT33/U_QUAD,1210.0,V,TRANSFER_INJECTION
LOGICAL.YRT1LT41/ASYMMETRY_KNOB_S,0.0,,TRANSFER_INJECTION
LOGICAL.YRT1LT41/U_QUAD,-2100.0,V,TRANSFER_INJECTION
LOGICAL.YRT1LT42/ASYMMETRY_KNOB_S,0.0,,TRANSFER_INJECTION
LOGICAL.YRT1LT42/U_QUAD,2020.0,V,TRANSFER_INJECTION
LOGICAL.YRT1LT43/ASYMMETRY_KNOB_S,0.0,,TRANSFER_INJECTION
LOGICAL.YRT1LT43/U_QUAD,-2100.0,V,TRANSFER_INJECTION
LOGICAL.YRT1MH1/HKICK,0.022,rad,TRANSFER_INJECTION
LOGICAL.YRT1MH1/K0L_THEO,-1.57079,rad,TRANSFER_INJECTION
YRT1LC1_TO_YRT1MH2_TIMEPARAM/TIMING_BI_DELAY_S,1.0E-6,s,TRANSFER_INJECTION
YRT1LC1_TO_YRT1MH2_TIMEPARAM/TIMING_BI_WINDOW_S,1.0E-6,s,TRANSFER_INJECTION
YRT1LC1_TO_YRT1MH2_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,BEAMOUT_INIT
YRT1LC1_TO_YRT1MH2_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,TRANSFER_INJECTION
YRT1LC1_TO_YRT1MH2_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LC1_TO_YRT1MH2_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LC1_TO_YRT1MH2_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LC1_TO_YRT1MH2_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LC1_TO_YRT1MH2_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LC1_TO_YRT1MH2_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LC1_TO_YRT1MH2_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LC1_TO_YRT1MH2_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LC1_TO_YRT1MH2_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1LC1_TO_YRT1MH2_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,BEAMOUT_RESET
YRT1LC1_TO_YRT1MH2_TIMEPARAM/TIMING_RF_LEAD_S,4.0E-4,s,TRANSFER_INJECTION
LOGICAL.YRT1BR1/ACT_ON_BEAM_S,true,,TRANSFER_INJECTION
LOGICAL.YRT1BR1/THIGH_S,0.0025,s,TRANSFER_INJECTION
LOGICAL.YRT1BR1/URF0_S,0.1,V,TRANSFER_INJECTION
LOGICAL.YRT1BR1/U_PAUSE_PULSE_AMPL_S,4.76,V,TRANSFER_INJECTION
LOGICAL.YRT1KH2/HKICK,3.0E-4,rad,TRANSFER_INJECTION
LOGICAL.YRT1KV2/VKICK,-2.0E-4,rad,TRANSFER_INJECTION
LOGICAL.YRT1LD51/ASYMMETRY_KNOB_S,0.0,,TRANSFER_INJECTION
LOGICAL.YRT1LD51/U_QUAD,510.0,V,TRANSFER_INJECTION
LOGICAL.YRT1LD52/ASYMMETRY_KNOB_S,0.0,,TRANSFER_INJECTION
LOGICAL.YRT1LD52/U_QUAD,-420.0,V,TRANSFER_INJECTION
LOGICAL.YRT1QD61/K1L,2.05,1/m,TRANSFER_INJECTION
LOGICAL.YRT1QD62/K1L,-1.68,1/m,TRANSFER_INJECTION
LOGICAL.YRT1QD71/K1L,1.85,1/m,TRANSFER_INJECTION
LOGICAL.YRT1QD72/K1L,-2.1,1/m,TRANSFER_INJECTION
YRT1BR1_TO_YRT1MH2_BEAM/E_S,5000.0,eV/u,TRANSFER_INJECTION
YRT1BR1_TO_YRT1MH2_RF/PHASE_OFFSET_S,0.0,deg,TRANSFER_INJECTION
LOGICAL.GHTYKH4/HKICK,-5.0E-4,rad,TRANSFER_INJECTION
LOGICAL.GHTYKV3/VKICK,-0.002,rad,TRANSFER_INJECTION
LOGICAL.GHTYKV4/VKICK,0.0015,rad,TRANSFER_INJECTION
LOGICAL.GHTYQD41/K1L,1.4,1/m,TRANSFER_INJECTION
LOGICAL.GHTYQD42/K1L,-1.25,1/m,TRANSFER_INJECTION
LOGICAL.YR01LP1I/ESEP_HKICK_S,0.017,rad,TRANSFER_INJECTION
LOGICAL.YR01LP1I/ESEP_K0L_THEO_S,-0.05,rad,TRANSFER_INJECTION
LOGICAL.YR01LP2I/ESEP_HKICK_S,-0.034,rad,TRANSFER_INJECTION
LOGICAL.YR01LP2I/ESEP_K0L_THEO_S,0.055,rad,TRANSFER_INJECTION
LOGICAL.YR01MP1I/HKICK,0.0035,rad,TRANSFER_INJECTION
LOGICAL.YR01MP1I/K0L_THEO,0.29322,rad,TRANSFER_INJECTION
LOGICAL.YRT1MH2/HKICK_ESR_S,-0.61087,rad,TRANSFER_INJECTION
LOGICAL.YRT1MH2/HKICK_LOCAL_S,-0.021,rad,TRANSFER_INJECTION
LOGICAL.YRT1MH2/K0L_THEO,-0.61087,rad,TRANSFER_INJECTION
YRT1MH2_TO_CRYRING_TIMEPARAM/TIMING_BI_DELAY_S,1.0E-6,s,TRANSFER_INJECTION
YRT1MH2_TO_CRYRING_TIMEPARAM/TIMING_BI_WINDOW_S,1.0E-6,s,TRANSFER_INJECTION
YRT1MH2_TO_CRYRING_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,BEAMOUT_INIT
YRT1MH2_TO_CRYRING_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,TRANSFER_INJECTION
YRT1MH2_TO_CRYRING_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1MH2_TO_CRYRING_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1MH2_TO_CRYRING_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1MH2_TO_CRYRING_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1MH2_TO_CRYRING_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1MH2_TO_CRYRING_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1MH2_TO_CRYRING_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1MH2_TO_CRYRING_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1MH2_TO_CRYRING_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,STATIC
YRT1MH2_TO_CRYRING_TIMEPARAM/TIMING_IS_LENGTH_CHANGED_S,true,,BEAMOUT_RESET
YRT1MH2_TO_CRYRING_TIMEPARAM/TIMING_SEPTUM_TRIGGER_OFFSET_S,0.0,s,TRANSFER_INJECTION
CRYRINGB2B/B2B_SID_EXT_S,---,,RING_INJECTION
CRYRINGB2B/B2B_SOURCE_KICKER_DELAY_S,---,s,RING_INJECTION
CRYRINGB2B/B2B_SOURCE_TRIGGER_OFFSET_S,---,deg,RING_INJECTION
CRYRINGB2B/B2B_TRIGGER_OFFSET_S,0.0,deg,RING_INJECTION
CRYRINGB2B/B2B_TRIGGER_OFFSET_TIME_S,0.0,,RING_INJECTION
CRYRINGBEAM/BEAM_MODE_S,NO_BEAM,,RING_INJECTION
CRYRINGBEAM/BRHO_CHIMNEY_S,---,Tm,BEAMOUT_INIT
CRYRINGBEAM/BRHO_CHIMNEY_S,---,Tm,BEAMOUT_RESET
CRYRINGBEAM/BRHO_DISPLAY_S,0.0713,Tm,RING_INJECTION
CRYRINGBEAM/BRHO_DISPLAY_S,1.0103,Tm,RING_RF_MANIPULATION
CRYRINGBEAM/BRHO_DISPLAY_S,1.4112,Tm,RING_TARGET
CRYRINGBEAM/BRHO_START_END_S,[0.07134952, 0.07134952],Tm,RING_INJECTION
CRYRINGBEAM/BUCKETFILL_RAW_S,[[0.0, 0.0]],,RING_INJECTION
CRYRINGBEAM/BUCKETFILL_RAW_S,[[0.0, 1.4]],,RING_BUNCHING
CRYRINGBEAM/BUCKETFILL_RAW_S,[[1.4, 1.4]],,RING_RAMP
CRYRINGBEAM/BUCKETFILL_RAW_S,[[1.4, 1.4]],,RING_RF_MANIPULATION
CRYRINGBEAM/BUCKETFILL_RAW_S,[[1.4, 1.4]],,RING_RAMP
CRYRINGBEAM/BUCKETFILL_RAW_S,[[1.4, 0.0]],,RING_REBUNCHING
CRYRINGBEAM/BUCKETFILL_RAW_S,[[0.0, 0.0]],,RING_TARGET
CRYRINGBEAM/BUCKETFILL_RAW_S,[[0.0, 0.0]],,RING_INTERMEDIATE
CRYRINGBEAM/BUCKETFILL_SIMPLE_S,1.4,,RING_INJECTION
CRYRINGBEAM/CHIMNEY_B,---,,BEAMOUT_INIT
CRYRINGBEAM/CHIMNEY_B,---,,BEAMOUT_RESET
CRYRINGBEAM/CH_END_S,5.0655,,RING_INJECTION
CRYRINGBEAM/CH_END_S,5.0655,,RING_RAMP
CRYRINGBEAM/CH_END_S,5.0655,,RING_RAMP
CRYRINGBEAM/CH_END_S,---,,STORAGE_MANIP_WAIT
CRYRINGBEAM/COASTING_BEAM_B,true,,RING_REBUNCHING
CRYRINGBEAM/CV_END_S,-5.3921,,RING_INJECTION
CRYRINGBEAM/CV_END_S,-5.3921,,RING_RAMP
CRYRINGBEAM/CV_END_S,-5.3921,,RING_RAMP
CRYRINGBEAM/CV_END_S,---,,STORAGE_MANIP_WAIT
CRYRINGBEAM/DELTAR,5.0E-4,m,RING_INJECTION
CRYRINGBEAM/DELTAR,3.0E-4,m,RING_RF_MANIPULATION
CRYRINGBEAM/DELTAR,0.0,m,STORAGE_MANIP_WAIT
CRYRINGBEAM/DELTAR,0.0,m,RING_TARGET
CRYRINGBEAM/DELTAR,0.0,m,RING_INTERMEDIATE
CRYRINGBEAM/DPBL,0.0,,RING_INJECTION
CRYRINGBEAM/DPBL,0.0,,STORAGE_MANIP_WAIT
CRYRINGBEAM/DPFREV,0.0015,,RING_INJECTION
CRYRINGBEAM/DPFREV,0.002,,RING_RF_MANIPULATION
CRYRINGBEAM/DPFREV,0.013,,STORAGE_MANIP_WAIT
CRYRINGBEAM/DPFREV,0.013,,RING_TARGET
CRYRINGBEAM/DPFREV,0.013,,RING_INTERMEDIATE
CRYRINGBEAM/DP_OVER_P_S,0.004,,RING_INJECTION
CRYRINGBEAM/E_S,4990.0,eV/u,RING_INJECTION
CRYRINGBEAM/E_S,1000000.0,eV/u,RING_RAMP
CRYRINGBEAM/E_S,1950000.0,eV/u,RING_RAMP
CRYRINGBEAM/FINAL_HARMONIC_S,6,,RING_RF_MANIPULATION
CRYRINGBEAM/FREV,18132.5287,Hz,RING_INJECTION
CRYRINGBEAM/FREV,256592.872,Hz,RING_RF_MANIPULATION
CRYRINGBEAM/FREV,361224.731,Hz,RING_TARGET
CRYRINGBEAM/H_INPUT_S,8,,RING_INJECTION
CRYRINGBEAM/INJ_EMIH_S,1.0E-5,m*rad,RING_INJECTION
CRYRINGBEAM/INJ_EMIL_S,6.160687750547807E-4,eVs/u,RING_INJECTION
CRYRINGBEAM/INJ_EMIV_S,1.0E-5,m*rad,RING_INJECTION
CRYRINGBEAM/KL_AMPL_S,0.0,,RING_INJECTION
CRYRINGBEAM/KL_AMPL_S,0.0,,RING_BUNCHING
CRYRINGBEAM/KL_AMPL_S,0.0,,RING_POST_INJECTION
CRYRINGBEAM/KL_AMPL_S,0.0,,RING_MANI_PRE_RAMP
CRYRINGBEAM/KL_AMPL_S,0.0,,RING_RAMP
CRYRINGBEAM/KL_AMPL_S,0.0,,RING_RF_MANIPULATION
CRYRINGBEAM/KL_AMPL_S,0.0,,RING_RAMP
CRYRINGBEAM/KL_AMPL_S,0.0,,RING_REBUNCHING
CRYRINGBEAM/KL_AMPL_S,0.0,,RING_TARGET
CRYRINGBEAM/KL_AMPL_S,0.0,,RING_INTERMEDIATE
CRYRINGBEAM/KL_HARM_S,1,,RING_INJECTION
CRYRINGBEAM/KL_HARM_S,1,,RING_BUNCHING
CRYRINGBEAM/KL_HARM_S,1,,RING_POST_INJECTION
CRYRINGBEAM/KL_HARM_S,1,,RING_MANI_PRE_RAMP
CRYRINGBEAM/KL_HARM_S,1,,RING_RAMP
CRYRINGBEAM/KL_HARM_S,1,,RING_RF_MANIPULATION
CRYRINGBEAM/KL_HARM_S,1,,RING_RAMP
CRYRINGBEAM/KL_HARM_S,1,,RING_REBUNCHING
CRYRINGBEAM/KL_HARM_S,1,,RING_TARGET
CRYRINGBEAM/KL_HARM_S,1,,RING_INTERMEDIATE
CRYRINGBEAM/KL_OFFSET_S,0.0,,RING_INJECTION
CRYRINGBEAM/KL_OFFSET_S,0.0,,RING_BUNCHING
CRYRINGBEAM/KL_OFFSET_S,0.0,,RING_POST_INJECTION
CRYRINGBEAM/KL_OFFSET_S,0.0,,RING_MANI_PRE_RAMP
CRYRINGBEAM/KL_OFFSET_S,0.0,,RING_RAMP
CRYRINGBEAM/KL_OFFSET_S,0.0,,RING_RF_MANIPULATION
CRYRINGBEAM/KL_OFFSET_S,0.0,,RING_RAMP
CRYRINGBEAM/KL_OFFSET_S,0.0,,RING_REBUNCHING
CRYRINGBEAM/KL_OFFSET_S,0.0,,RING_TARGET
CRYRINGBEAM/KL_OFFSET_S,0.0,,RING_INTERMEDIATE
CRYRINGBEAM/KL_PHASE_S,0.0,deg,RING_INJECTION
CRYRINGBEAM/KL_PHASE_S,0.0,deg,RING_BUNCHING
CRYRINGBEAM/KL_PHASE_S,0.0,deg,RING_POST_INJECTION
CRYRINGBEAM/KL_PHASE_S,0.0,deg,RING_MANI_PRE_RAMP
CRYRINGBEAM/KL_PHASE_S,0.0,deg,RING_RAMP
CRYRINGBEAM/KL_PHASE_S,0.0,deg,RING_RF_MANIPULATION
CRYRINGBEAM/KL_PHASE_S,0.0,deg,RING_RAMP
CRYRINGBEAM/KL_PHASE_S,0.0,deg,RING_REBUNCHING
CRYRINGBEAM/KL_PHASE_S,0.0,deg,RING_TARGET
CRYRINGBEAM/KL_PHASE_S,0.0,deg,RING_INTERMEDIATE
CRYRINGBEAM/NPARTICLES_S,1000.0,,RING_INJECTION
CRYRINGBEAM/PARTICLE_SYMBOL_INPUT_S,7Li,,RING_INJECTION
CRYRINGBEAM/QH_END_S,2.31,,RING_INJECTION
CRYRINGBEAM/QH_END_S,---,,RING_POST_INJECTION
CRYRINGBEAM/QH_END_S,2.3,,RING_RAMP
CRYRINGBEAM/QH_END_S,2.3,,RING_RAMP
CRYRINGBEAM/QH_END_S,---,,STORAGE_MANIP_WAIT
CRYRINGBEAM/QV_END_S,2.55,,RING_INJECTION
CRYRINGBEAM/QV_END_S,---,,RING_POST_INJECTION
CRYRINGBEAM/QV_END_S,2.55,,RING_RAMP
CRYRINGBEAM/QV_END_S,2.55,,RING_RAMP
CRYRINGBEAM/QV_END_S,---,,STORAGE_MANIP_WAIT
CRYRINGBEAM/Q_INPUT_S,1,,RING_INJECTION
CRYRINGBEAM/TREV,5.514951E-5,s,RING_INJECTION
CRYRINGBEAM/USE_SIMPLE_BF_B,true,,RING_INJECTION
CRYRINGBUMPER/TFALL_S,5.0E-4,s,RING_INJECTION
CRYRINGIKICKER/KICKER_TOFFSET1_S,0.0,s,RING_INJECTION
CRYRINGOPTICS/OPTIC_NAME_RING_S,CRYRING_STANDARD_OPTICS_TABLE,,RING_INJECTION
CRYRINGOPTICS/TAU_END_S,0.0,,RING_INJECTION
CRYRINGOPTICS/TAU_END_S,---,,RING_POST_INJECTION
CRYRINGOPTICS/TAU_END_S,0.0,,RING_RAMP
CRYRINGOPTICS/TAU_END_S,0.0,,RING_RAMP
CRYRINGORBIT/COH02,0.0,m,RING_INJECTION
CRYRINGORBIT/COH02,0.0,m,RING_RF_MANIPULATION
CRYRINGORBIT/COH02,0.0,m,STORAGE_MANIP_WAIT
CRYRINGORBIT/COH02,0.0,m,RING_TARGET
CRYRINGORBIT/COH03,0.0,m,RING_INJECTION
CRYRINGORBIT/COH03,0.0,m,RING_RF_MANIPULATION
CRYRINGORBIT/COH03,0.0,m,STORAGE_MANIP_WAIT
CRYRINGORBIT/COH03,0.0,m,RING_TARGET
CRYRINGORBIT/COH04,0.0,m,RING_INJECTION
CRYRINGORBIT/COH04,0.0,m,RING_RF_MANIPULATION
CRYRINGORBIT/COH04,0.0,m,STORAGE_MANIP_WAIT
CRYRINGORBIT/COH04,0.0,m,RING_TARGET
CRYRINGORBIT/COH06,0.0,m,RING_INJECTION
CRYRINGORBIT/COH06,0.0,m,RING_RF_MANIPULATION
CRYRINGORBIT/COH06,0.0,m,STORAGE_MANIP_WAIT
CRYRINGORBIT/COH06,0.0,m,RING_TARGET
CRYRINGORBIT/COH07,0.0,m,RING_INJECTION
CRYRINGORBIT/COH07,0.0,m,RING_RF_MANIPULATION
CRYRINGORBIT/COH07,0.0,m,STORAGE_MANIP_WAIT
CRYRINGORBIT/COH07,0.0,m,RING_TARGET
CRYRINGORBIT/COH08,0.0,m,RING_INJECTION
CRYRINGORBIT/COH08,0.0,m,RING_RF_MANIPULATION
CRYRINGORBIT/COH08,0.0,m,STORAGE_MANIP_WAIT
CRYRINGORBIT/COH08,0.0,m,RING_TARGET
CRYRINGORBIT/COH10,0.0,m,RING_INJECTION
CRYRINGORBIT/COH10,0.0,m,RING_RF_MANIPULATION
CRYRINGORBIT/COH10,0.0,m,STORAGE_MANIP_WAIT
CRYRINGORBIT/COH10,0.0,m,RING_TARGET
CRYRINGORBIT/COH11,0.0,m,RING_INJECTION
CRYRINGORBIT/COH11,0.0,m,RING_RF_MANIPULATION
CRYRINGORBIT/COH11,0.0,m,STORAGE_MANIP_WAIT
CRYRINGORBIT/COH11,0.0,m,RING_TARGET
CRYRINGORBIT/COH12,0.0,m,RING_INJECTION
CRYRINGORBIT/COH12,0.0,m,RING_RF_MANIPULATION
CRYRINGORBIT/COH12,0.0,m,STORAGE_MANIP_WAIT
CRYRINGORBIT/COH12,0.0,m,RING_TARGET
CRYRINGORBIT/COV02,0.0,m,RING_INJECTION
CRYRINGORBIT/COV02,0.0,m,RING_RF_MANIPULATION
CRYRINGORBIT/COV02,0.0,m,STORAGE_MANIP_WAIT
CRYRINGORBIT/COV02,0.0,m,RING_TARGET
CRYRINGORBIT/COV03,0.0,m,RING_INJECTION
CRYRINGORBIT/COV03,0.0,m,RING_RF_MANIPULATION
CRYRINGORBIT/COV03,0.0,m,STORAGE_MANIP_WAIT
CRYRINGORBIT/COV03,0.0,m,RING_TARGET
CRYRINGORBIT/COV04,0.0,m,RING_INJECTION
CRYRINGORBIT/COV04,0.0,m,RING_RF_MANIPULATION
... 679 more lines ...
  695   Tue May 27 08:38:29 2025 TD[How To] Recover DAQ when one FEE64 stops sending data - short(er) version
1) Select 'No Storage' [workspace 4, display 1, browser tab 'TapeServer']
    If already in 'No Storage' mode please ignore this step

2) STOP DAQ [workspace 3, display 1, browser tab 'Run Control']
   all FEE64s should change status from 'going' to 'stopped' *except* whichever FEE64(s) have a problem - typically their status is 'going' or 'error'

3) Restart Merger by selecting the Merger icon [workspace 4, display 2]
    terminal window showing the startup will popup

4) Setup Merger by selecting 'SETUP' [workspace 4, display 1, browser tab 'NewMerger Control']

5) Start Merger by selecting 'GO' [workspace 4, display 1, browser tab 'NewMerger Control']
     all data links will be inactive (red)

6) Start DAQ by selecting 'GO' [workspace 3, display 1, browser tab 'Run Control']
     check that DAQ is *actually* going - if not repeat 'GO'

7) Check data is being transferred to Merger and TapeServer

    Select 'Reload' [workspace 4, display 1, browser tab 'NewMerger Control']
     you should see a non-zero 'current merged data rate' and all links should show as green at least some of the time (requires observation over several reloads)

    Select 'Reload' [workspace 4, display 1, browser tab 'TapeServer']
     you should see 'Blocks written' and 'kb written' counters increase

8) Unselect 'No Storage' [workspace 3, display 1, browser tab 'TapeServer']
    If already in 'No Storage' mode at the start of this procedure please ignore this step


   If you continue to have problems, please request expert assistance.
  694   Tue May 27 05:24:05 2025 LB,SN, L. Cs. P. J. W.Run 22, 23 - 1.95 MeV/A

27/05/2025 ca. 6:00

We start a new run: beam is ok, target density ok, detectors are moving in-out periodically.

Cooler voltage setpoint  1105.7 V

Attachment 8 - beamcalc with actual values

Next entry - beamcalc with set values

 

Right arm at 19.80 mm

Left arm IN at 77.56 mm (OUT 32.53 mm)

vertical beam position 0.0 mm

horizontal beam position  -3.5mm

Beam intensity 1.7e+7 particles.

Target density ~6e11 at/cm2

Target temperatures: T1=31.5 K T2=35.2 K

Beam cycle set to 10 seconds.

All check are Ok!

6:30

Noticed a small spike in all leakage currents, see attachment 1

6:50

all checks ok!

 

07:40

All checks are Ok!

Aida 10 shows sometimes higher statistics.

 

09:02

All checks are Ok!

Aida 10 shows sometimes higher statistics.

 

10:00

Checks OK.

Compression works OK.

Moved to run 23. Now moving RIGHT hand side detectors instead of LEFT hand side.

Motion 37.96 mm -> 83.69 mm

Left arm stationary at 18.86 mm

Planning to change energy around lunch.

 

11:35

Parameters are OK! Tragte density dropped down a bit: ~6e11 at/cm2

 

12:35

Parameters are OK! Tragte density dropped down a bit: ~1e12 at/cm2

 

13:10

Starting beam energy change. Detectors OUT.

Attachment 1: Screenshot_from_2025-05-27_06-31-48.png
Screenshot_from_2025-05-27_06-31-48.png
Attachment 2: Screenshot_from_2025-05-27_07-40-10.png
Screenshot_from_2025-05-27_07-40-10.png
Attachment 3: Screenshot_from_2025-05-27_07-40-21.png
Screenshot_from_2025-05-27_07-40-21.png
Attachment 4: Screenshot_from_2025-05-27_09-22-51.png
Screenshot_from_2025-05-27_09-22-51.png
Attachment 5: Screenshot_from_2025-05-27_09-23-06.png
Screenshot_from_2025-05-27_09-23-06.png
Attachment 6: Screenshot_from_2025-05-27_11-34-21.png
Screenshot_from_2025-05-27_11-34-21.png
Attachment 7: Screenshot_from_2025-05-27_11-34-36.png
Screenshot_from_2025-05-27_11-34-36.png
Attachment 8: screenshot-2025-05-29_11-30-30730985236495866200.png
screenshot-2025-05-29_11-30-30730985236495866200.png
  693   Tue May 27 05:03:41 2025 JMDAQ restart

06:00

FEE modules power cycled.

No issues on power cycle restart.

All system check OK - clock status, ADC calibration etc.

ASIC settings loaded are the same as previous startup. Slow comparator thresholds 0x32 (500 keV) and all FEEs set to LEC.

Statistics (ADC data) - attachment 1. Note aida10 fluctuations between 10,000 and 300 which appears to be a physics effect (beam on target)

Spectrum browser - layout 1 - attachment 2

Merger - attachment 3

Tape server - attachment 4 Data rate is about 300 kB/sec.

FEE temperatures OK

Waveforms - attachments 4,5

All Lekagae currents appear normal

Online monitoring scropt started. Beam on target. Run 22 started. See nect elog for details.

Attachment 1: Screenshot_from_2025-05-27_06-09-45.png
Screenshot_from_2025-05-27_06-09-45.png
Attachment 2: Screenshot_from_2025-05-27_06-10-13.png
Screenshot_from_2025-05-27_06-10-13.png
Attachment 3: Screenshot_from_2025-05-27_06-10-29.png
Screenshot_from_2025-05-27_06-10-29.png
Attachment 4: Screenshot_from_2025-05-27_06-11-22.png
Screenshot_from_2025-05-27_06-11-22.png
Attachment 5: Screenshot_from_2025-05-27_06-13-59.png
Screenshot_from_2025-05-27_06-13-59.png
Attachment 6: Screenshot_from_2025-05-27_06-14-43.png
Screenshot_from_2025-05-27_06-14-43.png
Attachment 7: Screenshot_from_2025-05-27_06-16-04.png
Screenshot_from_2025-05-27_06-16-04.png
  692   Mon May 26 18:37:17 2025 LC, AG, LB, SNRun 21

Run 21 running

26/05/2025 19:42 p.m.

We start a new run with real data taking condititon: beam is ok, target density is 1.1e12 cm-2, detectors are moving in-out periodically.

Cooler voltage setpoint 1105.7 V.

Beam energy = 1.95 MeV/u

Right arm at 19.80 mm

Left arm IN at 77.56 mm (OUT 32.53 mm)

vertical beam position 0.0 mm

horizontal beam position  -3.5mm

Beam intensity 2.2e+7 particles.

Target density ~1.1e12 at/cm2

Target temperatures: T1=31.5 K T2=35.2 K

Beam cycle set to 10 seconds.

All check are Ok!

21:44

All check are Ok!

23:10

Data rate to disk ~300 kB/sec

All checks OK

ADC disk Items #6 and ADC data items screenshots - attachments 5 and 6

Target density has dropped slightly to 7e11 at/cm2. This is fine with lithium due to beam lifetime constraints.

00:43

All checks OK

01:52

All checks are OK,

Attached spectra (attachment 7 and 8).

target pressure seems to be back at ca. 1E12 at/cm2

3:27

All checks good.

target pressure went down to 5E11 at/cm2 for a while, then back up to ca. 9E11 at/cm2.

4:49

All checks good

Rates higher on aida10 and aida4 (see attachment 9), monitoring. It is not consistently high but for example aida10 goes 222->20000->300->11000. It seems to follow the detector movement (higher rates when detector are IN)

Data rate to disk ~370kB/s.

Spectra attached in 10-11.

5:11

Still flickering. Attachment 12 with rates.

Data rate to disk around ~400 kB/s

 

05:30

DAQ has crashed. All statistics on aida read zero. No data through the merger or tape server.

Stopped tape server and then Run control STOP. Most FEEs did not respond to the command and continued to read going and some undefined. Reset aida servers and reloaded webpages. Same status. Have to power cycle.

Attachment 1: Screenshot_from_2025-05-26_19-41-13.png
Screenshot_from_2025-05-26_19-41-13.png
Attachment 2: Screenshot_from_2025-05-26_19-41-25.png
Screenshot_from_2025-05-26_19-41-25.png
Attachment 3: Screenshot_from_2025-05-26_21-43-59.png
Screenshot_from_2025-05-26_21-43-59.png
Attachment 4: Screenshot_from_2025-05-26_21-44-11.png
Screenshot_from_2025-05-26_21-44-11.png
Attachment 5: Screenshot_from_2025-05-26_23-13-16.png
Screenshot_from_2025-05-26_23-13-16.png
Attachment 6: Screenshot_from_2025-05-26_23-13-33.png
Screenshot_from_2025-05-26_23-13-33.png
Attachment 7: Screenshot_from_2025-05-27_01-50-56.png
Screenshot_from_2025-05-27_01-50-56.png
Attachment 8: Screenshot_from_2025-05-27_01-50-39.png
Screenshot_from_2025-05-27_01-50-39.png
Attachment 9: Screenshot_from_2025-05-27_04-45-35.png
Screenshot_from_2025-05-27_04-45-35.png
Attachment 10: Screenshot_from_2025-05-27_04-57-15.png
Screenshot_from_2025-05-27_04-57-15.png
Attachment 11: Screenshot_from_2025-05-27_04-57-00.png
Screenshot_from_2025-05-27_04-57-00.png
Attachment 12: Screenshot_from_2025-05-27_05-10-36.png
Screenshot_from_2025-05-27_05-10-36.png
  691   Mon May 26 08:56:56 2025 EM,LBRun20

Run 20 running

We start a new run since the laser guys will do some measurements.

Zoran will increase the beam intensity

Detectors are out position

All checks are OK!

13:30

All checks are ok!

Attachment 1: Screenshot_from_2025-05-26_10-01-01.png
Screenshot_from_2025-05-26_10-01-01.png
Attachment 2: Screenshot_from_2025-05-26_10-01-19.png
Screenshot_from_2025-05-26_10-01-19.png
Attachment 3: Screenshot_from_2025-05-26_10-07-05.png
Screenshot_from_2025-05-26_10-07-05.png
Attachment 4: Screenshot_from_2025-05-26_13-32-57.png
Screenshot_from_2025-05-26_13-32-57.png
Attachment 5: Screenshot_from_2025-05-26_13-33-49.png
Screenshot_from_2025-05-26_13-33-49.png
Attachment 6: Screenshot_from_2025-05-26_13-34-12.png
Screenshot_from_2025-05-26_13-34-12.png
Attachment 7: Screenshot_from_2025-05-26_13-34-50.png
Screenshot_from_2025-05-26_13-34-50.png
Attachment 8: Screenshot_from_2025-05-26_13-35-19.png
Screenshot_from_2025-05-26_13-35-19.png
  690   Sat May 24 18:27:55 2025 EM, LCBackground runs continued R17-R19

Summary: Target not running. Background runs written to disk, some with beam some without.

19:27 Data file R17  running

      DSSSD positions out. Nikos will work on the target.
      Started a new run just after changing the DSSSD position
      No storage enabled
      All the checks are ok!

At 20:30 after 2h there was possible to open the cave again and Nikos started to cool down the target. However, is not clear if they can close the ring again...


22:12  The beam is back again. Cave closed and we are cooling downg the target!

Data file R18 started now.

DSSSD positions out. 

No storage disabled.

All checks ok. See attached files Work on the target.

 

Sunday 25 May 2025

09:45 Data file R18 running

DSSSD positions out. 

All checks ok!

 

12:05 Data file R18 running

DSSSD positions out. 

All checks ok!

 

13:55 Data file R18 running

DSSSD positions out. 

All checks ok!

 

15:50 Data file R19 running

DSSSD positions out. 

Beam is off.

All checks ok!

 

21:50

All checks OK.

 

Monday 26.05.25 09:15

All checks OK.

Attachment 1: Screenshot_from_2025-05-24_20-32-56.png
Screenshot_from_2025-05-24_20-32-56.png
Attachment 2: Screenshot_from_2025-05-24_20-32-43.png
Screenshot_from_2025-05-24_20-32-43.png
Attachment 3: Screenshot_from_2025-05-24_20-32-23.png
Screenshot_from_2025-05-24_20-32-23.png
Attachment 4: Screenshot_from_2025-05-24_20-32-02.png
Screenshot_from_2025-05-24_20-32-02.png
Attachment 5: Screenshot_from_2025-05-24_20-31-28.png
Screenshot_from_2025-05-24_20-31-28.png
Attachment 6: Screenshot_from_2025-05-24_22-19-46.png
Screenshot_from_2025-05-24_22-19-46.png
Attachment 7: Screenshot_from_2025-05-24_22-16-14.png
Screenshot_from_2025-05-24_22-16-14.png
Attachment 8: Screenshot_from_2025-05-24_22-15-07.png
Screenshot_from_2025-05-24_22-15-07.png
Attachment 9: Screenshot_from_2025-05-24_22-14-00.png
Screenshot_from_2025-05-24_22-14-00.png
Attachment 10: Screenshot_from_2025-05-25_09-48-48.png
Screenshot_from_2025-05-25_09-48-48.png
Attachment 11: Screenshot_from_2025-05-25_09-48-28.png
Screenshot_from_2025-05-25_09-48-28.png
Attachment 12: Screenshot_from_2025-05-25_09-47-42.png
Screenshot_from_2025-05-25_09-47-42.png
Attachment 13: Screenshot_from_2025-05-25_09-47-22.png
Screenshot_from_2025-05-25_09-47-22.png
Attachment 14: Screenshot_from_2025-05-25_09-45-08.png
Screenshot_from_2025-05-25_09-45-08.png
Attachment 15: Screenshot_from_2025-05-25_09-45-56.png
Screenshot_from_2025-05-25_09-45-56.png
Attachment 16: Screenshot_from_2025-05-25_12-07-27.png
Screenshot_from_2025-05-25_12-07-27.png
Attachment 17: Screenshot_from_2025-05-25_12-07-53.png
Screenshot_from_2025-05-25_12-07-53.png
Attachment 18: Screenshot_from_2025-05-25_12-08-21.png
Screenshot_from_2025-05-25_12-08-21.png
Attachment 19: Screenshot_from_2025-05-25_12-08-49.png
Screenshot_from_2025-05-25_12-08-49.png
Attachment 20: Screenshot_from_2025-05-25_12-09-06.png
Screenshot_from_2025-05-25_12-09-06.png
Attachment 21: Screenshot_from_2025-05-25_13-55-56.png
Screenshot_from_2025-05-25_13-55-56.png
Attachment 22: Screenshot_from_2025-05-25_13-56-18.png
Screenshot_from_2025-05-25_13-56-18.png
Attachment 23: Screenshot_from_2025-05-25_13-57-12.png
Screenshot_from_2025-05-25_13-57-12.png
Attachment 24: Screenshot_from_2025-05-25_13-57-33.png
Screenshot_from_2025-05-25_13-57-33.png
Attachment 25: Screenshot_from_2025-05-25_13-57-48.png
Screenshot_from_2025-05-25_13-57-48.png
Attachment 26: Screenshot_from_2025-05-25_15-47-50.png
Screenshot_from_2025-05-25_15-47-50.png
Attachment 27: Screenshot_from_2025-05-25_15-48-51.png
Screenshot_from_2025-05-25_15-48-51.png
Attachment 28: Screenshot_from_2025-05-25_15-49-14.png
Screenshot_from_2025-05-25_15-49-14.png
Attachment 29: Screenshot_from_2025-05-25_15-50-03.png
Screenshot_from_2025-05-25_15-50-03.png
Attachment 30: Screenshot_from_2025-05-25_15-50-35.png
Screenshot_from_2025-05-25_15-50-35.png
Attachment 31: Screenshot_from_2025-05-26_09-13-13.png
Screenshot_from_2025-05-26_09-13-13.png
Attachment 32: Screenshot_from_2025-05-26_09-13-47.png
Screenshot_from_2025-05-26_09-13-47.png
Attachment 33: Screenshot_from_2025-05-26_09-14-04.png
Screenshot_from_2025-05-26_09-14-04.png
Attachment 34: Screenshot_from_2025-05-26_09-14-18.png
Screenshot_from_2025-05-26_09-14-18.png
Attachment 35: Screenshot_from_2025-05-26_09-15-52.png
Screenshot_from_2025-05-26_09-15-52.png
Attachment 36: Screenshot_from_2025-05-26_11-19-49.png
Screenshot_from_2025-05-26_11-19-49.png
  689   Sat May 24 08:28:14 2025 TD, EMBackground runs R13-R16
Summary: Target not running. Some background files written to disk, some with beam some without. 

09.27 DAQ going
      Switch from 'no storage' mode - alpha background
      Data file R13 

      All histograms & stats zero'd

      ASIC settings 2025Apr08-13.45.30
       slow comparator 0x32

      Pulser ON

      DSSSD bias & leakage currents OK - attachment 1-2 
      FEE64 temps OK - attachment 3
      ADC data item stats - attachment 4
      per FEE64 Rate spectra - attachment 5
      WR timestamp OK - attachment 6
      System wide checks OK *except* WR decoder/FPGA error - attachments 7-8
      Merger & TapeServer - attachment 9-10
      CRYRING vacuum & temperature - attachment 11-12

10.07 Laser group have completed their work in the CRYRING area - will now request 7Li beam
      Tape Server to 'no storage' mode

10.45 Beam 7Li1+ c. 1e6 
      Switch from 'no storage' mode - alpha background
      Data file R14
      All histograms & stats zero'd; online analysis restarted

11.35 per FEE64 1.8.L spectra - attachments 13-14
       pulser peak width aida03 78 ch FWHM, aida11 63 ch FWHM

      per FEE64 1.8.W spectra - 20us FSR - attachments 15-18

12.55 DSSSD bias & leakage currents OK - attachment 19-20
      FEE64 temps OK - attachment 21
      ADC data item stats - attachment 22

14.33 Switch Tape Server from 'no storage' mode
      Data file R15
      DSSSD positions Left 36.54mm, Right 64.60mm

16:22 Data file R15 still running

      DSSSD positions unchanged (Left 36.54mm, Right 64.60mm)
      All the checks are ok!


18:30 Data file R16  running

      DSSSD positions out. Nikos will work on the target.
      Started a new run just after changing the DSSSD position

      All the checks are ok!
Attachment 1: Screenshot_from_2025-05-24_09-30-00.png
Screenshot_from_2025-05-24_09-30-00.png
Attachment 2: Screenshot_from_2025-05-24_09-30-11.png
Screenshot_from_2025-05-24_09-30-11.png
Attachment 3: Screenshot_from_2025-05-24_09-30-29.png
Screenshot_from_2025-05-24_09-30-29.png
Attachment 4: Screenshot_from_2025-05-24_09-30-39.png
Screenshot_from_2025-05-24_09-30-39.png
Attachment 5: Screenshot_from_2025-05-24_09-31-20.png
Screenshot_from_2025-05-24_09-31-20.png
Attachment 6: Screenshot_from_2025-05-24_09-31-44.png
Screenshot_from_2025-05-24_09-31-44.png
Attachment 7: Screenshot_from_2025-05-24_09-31-59.png
Screenshot_from_2025-05-24_09-31-59.png
Attachment 8: Screenshot_from_2025-05-24_09-32-04.png
Screenshot_from_2025-05-24_09-32-04.png
Attachment 9: Screenshot_from_2025-05-24_09-33-42.png
Screenshot_from_2025-05-24_09-33-42.png
Attachment 10: Screenshot_from_2025-05-24_09-33-47.png
Screenshot_from_2025-05-24_09-33-47.png
Attachment 11: 3.png
3.png
Attachment 12: 4.png
4.png
Attachment 13: Screenshot_from_2025-05-24_11-35-42.png
Screenshot_from_2025-05-24_11-35-42.png
Attachment 14: Screenshot_from_2025-05-24_11-36-51.png
Screenshot_from_2025-05-24_11-36-51.png
Attachment 15: Screenshot_from_2025-05-24_11-41-27.png
Screenshot_from_2025-05-24_11-41-27.png
Attachment 16: Screenshot_from_2025-05-24_11-42-07.png
Screenshot_from_2025-05-24_11-42-07.png
Attachment 17: Screenshot_from_2025-05-24_11-42-24.png
Screenshot_from_2025-05-24_11-42-24.png
Attachment 18: Screenshot_from_2025-05-24_11-42-43.png
Screenshot_from_2025-05-24_11-42-43.png
Attachment 19: Screenshot_from_2025-05-24_12-55-39.png
Screenshot_from_2025-05-24_12-55-39.png
Attachment 20: Screenshot_from_2025-05-24_12-55-48.png
Screenshot_from_2025-05-24_12-55-48.png
Attachment 21: Screenshot_from_2025-05-24_12-56-11.png
Screenshot_from_2025-05-24_12-56-11.png
Attachment 22: Screenshot_from_2025-05-24_12-56-24.png
Screenshot_from_2025-05-24_12-56-24.png
Attachment 23: Screenshot_from_2025-05-24_16-24-01.png
Screenshot_from_2025-05-24_16-24-01.png
Attachment 24: Screenshot_from_2025-05-24_16-25-03.png
Screenshot_from_2025-05-24_16-25-03.png
Attachment 25: Screenshot_from_2025-05-24_16-25-27.png
Screenshot_from_2025-05-24_16-25-27.png
Attachment 26: Screenshot_from_2025-05-24_16-26-50.png
Screenshot_from_2025-05-24_16-26-50.png
Attachment 27: Screenshot_from_2025-05-24_18-33-59.png
Screenshot_from_2025-05-24_18-33-59.png
Attachment 28: Screenshot_from_2025-05-24_18-34-56.png
Screenshot_from_2025-05-24_18-34-56.png
Attachment 29: Screenshot_from_2025-05-24_18-35-51.png
Screenshot_from_2025-05-24_18-35-51.png
  688   Fri May 23 09:19:40 2025 SN, TDTarget wait, switch lithium oven
10.12 DAQ, Merger & Tape Server running OK - no storage mode 

      DSSSD bias leakage current OK - attachments 9-10

      CRYRING vacuum & temperature - attachments 11-12

      FEE64 temps OK - attachment 8

      ADC & DISC data item stats OK - attachments 6-7

      per FEE64 Rate spectra - attachments 1 & 5
       before and after check ASIC load

      WR timestamps OK - attachment 4

      System wide checks OK *except* FPGA & WR status - attachments 2-3

10.33 All histograms & stats zero'd
      Online analysis restarted

13:56 Checks on temperature status, statistics status and power supply : OK 
      
Attachment 1: Screenshot_from_2025-05-23_10-18-44.png
Screenshot_from_2025-05-23_10-18-44.png
Attachment 2: Screenshot_from_2025-05-23_10-15-58.png
Screenshot_from_2025-05-23_10-15-58.png
Attachment 3: Screenshot_from_2025-05-23_10-15-54.png
Screenshot_from_2025-05-23_10-15-54.png
Attachment 4: Screenshot_from_2025-05-23_10-15-36.png
Screenshot_from_2025-05-23_10-15-36.png
Attachment 5: Screenshot_from_2025-05-23_10-14-27.png
Screenshot_from_2025-05-23_10-14-27.png
Attachment 6: Screenshot_from_2025-05-23_10-13-43.png
Screenshot_from_2025-05-23_10-13-43.png
Attachment 7: Screenshot_from_2025-05-23_10-13-32.png
Screenshot_from_2025-05-23_10-13-32.png
Attachment 8: Screenshot_from_2025-05-23_10-13-21.png
Screenshot_from_2025-05-23_10-13-21.png
Attachment 9: Screenshot_from_2025-05-23_10-12-52.png
Screenshot_from_2025-05-23_10-12-52.png
Attachment 10: Screenshot_from_2025-05-23_10-12-40.png
Screenshot_from_2025-05-23_10-12-40.png
Attachment 11: 1.png
1.png
Attachment 12: 2.png
2.png
Attachment 13: Screenshot_from_2025-05-23_12-53-29.png
Screenshot_from_2025-05-23_12-53-29.png
Attachment 14: Screenshot_from_2025-05-23_12-53-24.png
Screenshot_from_2025-05-23_12-53-24.png
Attachment 15: Screenshot_from_2025-05-23_12-50-59.png
Screenshot_from_2025-05-23_12-50-59.png
Attachment 16: Screenshot_from_2025-05-23_13-53-32.png
Screenshot_from_2025-05-23_13-53-32.png
Attachment 17: Screenshot_from_2025-05-23_13-53-16.png
Screenshot_from_2025-05-23_13-53-16.png
Attachment 18: Screenshot_from_2025-05-23_13-52-59.png
Screenshot_from_2025-05-23_13-52-59.png
  687   Thu May 22 06:41:10 2025 JG, TDR6 - R11 scrapers test / pulser walk
07.00 

JG found target off (target data stopped c. 02.30 this morning) and beam off (not re-enabled from previous CRYRNG area entry - target related?) JG re-enables beam 

07.37 

Run R5 stopped Run R6 started - switch to no storage mode 

07.44 

DSSSD bias and leakage currents - attachments 1 
FEE64 Rate spectra OK - attachment 2 
ADC data item stats OK - attachment 3 
FEE64 temps OK - attachment 4 
WR timestamps OK - attachment 5 
Merger and Tape Server - attachments 6 
DSSSD x-y hit patterns - attachment 8 
Per JM instructions will move detectors toward beam axis by 30mm and investigate effect of e-target scrapers 

08.45 

Run R7 JG has setup e-target scrapers (YR09DS1HA and YR09DS1HI) both scrapers are set to move with the beam cycle. They move in with the start of the 'target' period 
(SC3). They move out with the start of the 'intermediate' period (SC4). The movement is rather slow (~1sec) so there is still some unblocked time during 'target' period. 
The positions are YR09DS1HA (outside): 10mm (SC3) and 30mm (SC4) YR09DS1HI (inside): -0.5mm (SC3) and -20mm (SC4) 


11:50 

Jan has been playing with the scraper settings so that we can take a run with the scrapers  and without. 

Since he has been playing. We see the beam halo. This is when scrapers were in and beam was moved to the inside of the ring by 5mm. See attachment 10

Moving the beam to the putside of the ring by 5mm we see halo on the left now. See attachment 11

We believe this is scattering from the edge of the scrapers. See attachment 12 for the energy spectrum which corresponds to attachment 11. It is consistent with 
scattering.

We will now take a run with and without the scrapers in and with the left detector moving in with every cycle via the pneumatic motor.  

12:20 

R8 started. Scrapers in. Left arm moving in synchronous with the ring cycle, potentiometer when IN = 62.71 mm. Right arm stationary and OUT, potentiometer = 19.8 mm

13:00

Run 9 started. Same condition as run 8 except no scrapers.

13:35 

Run 9 stopped, R10 started -> no storage 

13.37 Merger - attachment 13

      per FEE64 n+n Ohmic 1.8.W spectra - 20us, 200us, 2ms, 20ms FSR - attachments 14-18
      per FEE64 p+n junction 1.8.W spectra - 20ms, 2ms, 200us, 20us FSR - attachments 19-22

      per FEE64 Rate spectra - attachment 23

      WR timestamp OK - attachment 24

      ADC, DISC, PAUSE  RESUME data item stats - attachment 25-28

      FEE64 temps OK - attachment 29

      DSSSD bias & leakage current OK - attachment 30


14:00

R11 Pulser walkthrough 
    See attachments 31-32

Following pulser walkthrough R12 started with no beam and no target in the ring. Cave is accessible while we switch to 7Li

20:15

System checks OK
Attachment 1: Screenshot_from_2025-05-22_07-44-10.png
Screenshot_from_2025-05-22_07-44-10.png
Attachment 2: Screenshot_from_2025-05-22_07-44-21.png
Screenshot_from_2025-05-22_07-44-21.png
Attachment 3: Screenshot_from_2025-05-22_07-44-30.png
Screenshot_from_2025-05-22_07-44-30.png
Attachment 4: Screenshot_from_2025-05-22_07-44-44.png
Screenshot_from_2025-05-22_07-44-44.png
Attachment 5: Screenshot_from_2025-05-22_07-45-17.png
Screenshot_from_2025-05-22_07-45-17.png
Attachment 6: Screenshot_from_2025-05-22_07-45-27.png
Screenshot_from_2025-05-22_07-45-27.png
Attachment 7: Screenshot_from_2025-05-22_07-45-36.png
Screenshot_from_2025-05-22_07-45-36.png
Attachment 8: Screenshot_from_2025-05-22_07-46-08.png
Screenshot_from_2025-05-22_07-46-08.png
Attachment 9: Screenshot_from_2025-05-22_07-46-18.png
Screenshot_from_2025-05-22_07-46-18.png
Attachment 10: Screenshot_from_2025-05-22_11-59-16.png
Screenshot_from_2025-05-22_11-59-16.png
Attachment 11: Screenshot_from_2025-05-22_12-09-58.png
Screenshot_from_2025-05-22_12-09-58.png
Attachment 12: Screenshot_from_2025-05-22_12-13-06.png
Screenshot_from_2025-05-22_12-13-06.png
Attachment 13: Screenshot_from_2025-05-22_13-47-20.png
Screenshot_from_2025-05-22_13-47-20.png
Attachment 14: Screenshot_from_2025-05-22_13-47-07.png
Screenshot_from_2025-05-22_13-47-07.png
Attachment 15: Screenshot_from_2025-05-22_13-46-10.png
Screenshot_from_2025-05-22_13-46-10.png
Attachment 16: Screenshot_from_2025-05-22_13-45-40.png
Screenshot_from_2025-05-22_13-45-40.png
Attachment 17: Screenshot_from_2025-05-22_13-45-17.png
Screenshot_from_2025-05-22_13-45-17.png
Attachment 18: Screenshot_from_2025-05-22_13-43-26.png
Screenshot_from_2025-05-22_13-43-26.png
Attachment 19: Screenshot_from_2025-05-22_13-42-53.png
Screenshot_from_2025-05-22_13-42-53.png
Attachment 20: Screenshot_from_2025-05-22_13-42-23.png
Screenshot_from_2025-05-22_13-42-23.png
Attachment 21: Screenshot_from_2025-05-22_13-41-44.png
Screenshot_from_2025-05-22_13-41-44.png
Attachment 22: Screenshot_from_2025-05-22_13-41-14.png
Screenshot_from_2025-05-22_13-41-14.png
Attachment 23: Screenshot_from_2025-05-22_13-39-05.png
Screenshot_from_2025-05-22_13-39-05.png
Attachment 24: Screenshot_from_2025-05-22_13-38-52.png
Screenshot_from_2025-05-22_13-38-52.png
Attachment 25: Screenshot_from_2025-05-22_13-38-07.png
Screenshot_from_2025-05-22_13-38-07.png
Attachment 26: Screenshot_from_2025-05-22_13-37-56.png
Screenshot_from_2025-05-22_13-37-56.png
Attachment 27: Screenshot_from_2025-05-22_13-37-42.png
Screenshot_from_2025-05-22_13-37-42.png
Attachment 28: Screenshot_from_2025-05-22_13-37-25.png
Screenshot_from_2025-05-22_13-37-25.png
Attachment 29: Screenshot_from_2025-05-22_13-37-14.png
Screenshot_from_2025-05-22_13-37-14.png
Attachment 30: Screenshot_from_2025-05-22_13-36-40.png
Screenshot_from_2025-05-22_13-36-40.png
Attachment 31: Screenshot_from_2025-05-22_14-40-46.png
Screenshot_from_2025-05-22_14-40-46.png
Attachment 32: Screenshot_from_2025-05-22_14-40-00.png
Screenshot_from_2025-05-22_14-40-00.png
Attachment 33: Screenshot_from_2025-05-22_15-06-01.png
Screenshot_from_2025-05-22_15-06-01.png
  686   Wed May 21 14:38:22 2025 JM, TDR4 and R5 - background

15:30

Waiting for the target to cool down. ETA is 16:30 for cooling to be finished and then we need to try and make target.

While waiting we havve been running a background (R3). I noticed the rates on aida10 had increased (300k) and aida09 were slowlly creeping up too. Check load and shaking the detectors did not work. These FEEs correspond to the high leakage current detectors and showed similar rates to aida03 and aida04 which were turned to MEC.

Tried some tests. Lowered bias voltage to 50V. Rates increased on aida09 and aida10. Increased voltage to 110V and rates dropped to approx zero. Checked 105V and we get the same effect as at 110V. 

Then decided to try to increase the bias on DSSD1 as well. 110V no ffect but at 115V we see the rates drop to zero.

Data rates of ~300 kB/sec writing to disk.

Some increases in rates on aida12 have also been noticed. We have turned the fast discrimtor off for this FEE which has resolved the issue. We should keep an eye on the rates as we hope this is stable but may not be.

 Attachment 1 - current ADC data items rate with new settings. (still 0x32 slow comparator, all channels LEC)

Attachment 2 - aida12 discriminator settings.

Attachment 3 - current bias/leakage current values.

Started R4 - background we expect beam in the ring soon.

16:00

Beam is in the ring - we see the beam stripe on DSSD1

No change on data rate writing to disk or in ADC statistics rate.

Motors fully out. Left potentiometer = 18.86 mm, Right potentiometer = 19.80 mm

Beam energy = 2 MeV/u - 6Li1+

Cycle time is 25 seconds

Attachment 4 - CARME detectors xy plot

Attachment 5 - Total energy histograms (Ex)

17:56

Nikos requires access to the cave so beam is stopped and background run is going.

R5 started. Same settings as previous run.

21:45

All system checks OK - except white rabbit decoder and FPGA timestamps see attachments 6,7

Data rate had gone high again ~2 MB/sec. No difference on ADC data items. Large increase on aida06 for aida disc info #6.

Checked hit rate for aida06 and see ASIC 2 was misbehaving. ASIC check load on asic 2 -> problem sorted. This is not first time this has happened. We should keep an eye on the data rate.

Attachment 1: Screenshot_from_2025-05-21_15-43-32.png
Screenshot_from_2025-05-21_15-43-32.png
Attachment 2: Screenshot_from_2025-05-21_15-29-16.png
Screenshot_from_2025-05-21_15-29-16.png
Attachment 3: Screenshot_from_2025-05-21_15-52-05.png
Screenshot_from_2025-05-21_15-52-05.png
Attachment 4: Screenshot_from_2025-05-21_16-15-42.png
Screenshot_from_2025-05-21_16-15-42.png
Attachment 5: Screenshot_from_2025-05-21_16-15-59.png
Screenshot_from_2025-05-21_16-15-59.png
Attachment 6: Screenshot_from_2025-05-21_21-54-42.png
Screenshot_from_2025-05-21_21-54-42.png
Attachment 7: Screenshot_from_2025-05-21_21-54-52.png
Screenshot_from_2025-05-21_21-54-52.png
  685   Tue May 20 19:38:36 2025 JMCARME restart

19:00

Arrived at GSI.

We have 6Li1+ beam in the ring at 2 MeV/u. Injection is about 2.5e7 pps and lifetime with no target is about 40 seconds. We have injection every 25 seconds.

DSSD bias situation remains unchanged from previous elogs. See attachment 1.

AIDA started - all looks OK. All FEE temps OK - coolant pipe must still be fine.

Set aida03 and aida04 to MEC (0x1) and slow comparator thresholds for all FEEs to 0x32 (500 keV).

Attachment 2 shows ADC data items rate for 1 MeV threshold. Attachment 3 shows the same but at 500 keV threshold with aida03 and aida04 set to MEC.

DAQ go rates look OK.

Setting up tape server and merger. I notice a strange occurance. NetVar options is set to 3 on loading. I set this equal to 1 so we can write to disk.... but this turns off the enable output to data storage off on the merger. By enabling output to data storage again the neVar options variable returns to 3 and we cannot write data to disk. I have tried reloading the servers for the merger/tape server and rebooting the FEEs -> still have this issue. It is pretty late so have turned FEEs off but left DSSDs biased at 100V.

Note the proxy is back up

09:45

Started up AIDA DAQ.

Started merger and tape server as yesterday. Today I have had no issues (same procedure was used last night?!). We are currently writing data to disk G22-205/R1

All FEEs set to 0x32 thresholds, with aida03 and aida04 set to MEC.

All sytem checks OK - no faults.

Temps OK

Merger running between 100k and 400k. Tape server writing ~300 kb/sec to disk.

Beam is ON. 6Li1+ at 2 MeV/u. Gleb is currently running machine tests.

Motors are not homed yet, waiting until beam is off and access to the cave to double check pins are out, Potentiometers read left=21.45mm and right=19.8mm

Attachment 4 - ADC data rates

Attachment 5 - Merger

Attachment 6 - Tape server

11:00

Safety meeting , beam off and access to the cave

Pins removed from motors and compressed air on. Motors homed.

Started background run (no beam) R3

13:30

We are now waiing for the target before we can begin tests today.

ASIC settings checked. We are using APR2025 settings. Only chnages made are aida03 and aida04 to MEC and slow comparator threshold lowered to 500 keV (0x32)

Fast comparator threshold is 0xff

14:00

Motors movement tested with signals from CRYRING. Dummy test signals sent through number 16 on the control room to CRYRING touch panel. This sent an injection signal followed by a dump signal 5 seconds later. Motors moved according to this pattern.

Command used on CARME motors code for future reference

l t in

l p in

l t out

l p out

l r 20

Attachment 1: Screenshot_from_2025-05-20_20-04-13.png
Screenshot_from_2025-05-20_20-04-13.png
Attachment 2: Screenshot_from_2025-05-20_20-29-54.png
Screenshot_from_2025-05-20_20-29-54.png
Attachment 3: Screenshot_from_2025-05-20_20-38-10.png
Screenshot_from_2025-05-20_20-38-10.png
Attachment 4: Screenshot_from_2025-05-21_09-49-54.png
Screenshot_from_2025-05-21_09-49-54.png
Attachment 5: Screenshot_from_2025-05-21_09-50-08.png
Screenshot_from_2025-05-21_09-50-08.png
Attachment 6: Screenshot_from_2025-05-21_09-50-22.png
Screenshot_from_2025-05-21_09-50-22.png
  684   Sat May 17 11:19:06 2025 JMShifter intructions 2025

EMERGENCY SHUTDOWNS

1. CARME motors: Q key on keyboard while on workspace 1 of appc218 (right of this double-monitor setup)

2. Internal target:

 

ELOG

The shift should put all information relating to the current run in ONE elog entry. If a significant change occurs during the shift i.e beam off, target off, run change, detectors out etc, start a new run and add a new elog entry.

Elog entry should be named by the current run number.

At the start of each new elog entry please state current detector positions, beam energy and run type (e.g. beam on target, background etc).

TO DO FOR THE SHIFTERS - CHECK EVERY HOUR AND POST TO ELOG WITH SCREENSHOTS. ELOG should be updated with current beam energy, injection beam intensity, system checks OK/not, FEE temperature OK/not, Statistics OK/not, Merger OK/not, tape server data rate, Motors OK/not and min/max motors potentiometer value. The time of each update should be noted in the elog. Screenshots of important screens (detailed below) should also be uploaded to the elog regularly (not required every hour).

 

CARME DAQ

CARME computer has 6 workspaces (WS), two monitors for each WS.

WS1 : left monitor - fee power

          right monitor - Detector Bias control, fee power server (just a terminal, do not close). Check detector bias is stable. If any detector current levels become unstable contact Jordan or Carlo and note time in the elog.

 

WS2 : left monitor - DAQ controls

          right monitor - DAQ server (just a terminal, do not close)

          Most important tabs (DAQ controls)

  • Run control : shows daq status, stopped / going
  • Fee temperatures : temperatures of fees labelled aida01, aida02, aida03, aida04, aida05, aida06, aida07, aida08, aida09, aida10, aida11, aida12, aida13, aida14, aida15, aida16. Hit reload, temperatures will update and should all be green. 1-2 C above safe limit (top) is OK. More than that - warn expert shifter immediately.
  • Statistics : shows the statistics for each fee. Make sure ADC Data Items is selected in the left hand menu. Hit update all  (SCREENSHOT once, if no changes record OK in elog)
  • System wide checks : checks for the setup. Click on 'Check Clock Status', 'Check ADC Calibration' and 'Check the White rabbit decoder status' to run checks. (SCREENSHOT once, if no changes record OK in elog)
  • Spectrum browser : loads histograms for fees and fee channels from preset layouts. select layout from 'layout ID' then restore layout from 'Arrange functions' drop down menu. Check layout 1, Select log scale and change xmax as 128 . This displays the data rate per strip. If any strips are a lot higher than the others, or very different from previous entry, make a note in the ELOG. (SCREENSHOT once, if no changes record OK in elog)
  • Asic control : sets asic parameters

 

WS3 : left monitor - Merge control, Tape server

          right monitor - servers (just terminals, do not close) , data being written to file

  • Check NewMerger tab. Hit reload button and see if 'current links with data' and 'current merged data rate' are updating. All links should go green, but not necessarily at the same time. Keep hitting refresh until you see all of them going green at least once.
  • Data rate, hit reload should be around 100,000. Record data rate in elog
  • Check tapeService tab. Hit reload button. Check data rate in Kbytes/sec box. Record data rate to elog. This is where the current run number is displayed
  • On the right-hand monitor, bottom right terminal, check the current run fragment (sub-run) is being written and is slowly increasing in size. It refreshed automatically every 5 seconds.

 

WS4 : Leakage current plot

  • Check leakage current plot is stable. Transients will occur during motors movement but baseline current should remain consistent. Make sure it is updating and time is correct. (SCREENSHOT once, record OK in elog if no changes) Check present values on WS2, CAEN window

 

WS5 : Online data monitoring

  • Check the following plots are updating for each DSSD
    • LowEnergyXY (right click - col - colz). This shows the total number of events, per pixel, since the code was started / re-started.
    • LowEnergyAngleE. This shows the angle vs energy plot of all detected events. Can be used to observed elastic scattering / 7Be from lithium measurement
    • LowEnergyXYRate (right click - col - colz). This shows the rate over the last 20 seconds.
    • LowEnergyExRate (double left click. Right click on y axis - Log scale on y). Rate over the last 20 seconds
    • LowEnergyExTotal (double left click. Right click on y axis - Log scale on y). This shows the total energy deposited on all X strips since code restart.
    • LowEnergyEyTotall (as above). as above, Y strips. (SCREENSHOT of each per DSSD per shift. If significant changes are observed screenshot again)

 

  • During the shift, shifters should try to integrate peaks of interest. Area to integrate will be given in spreadsheet in same workspace as monitoring code. This is particularly important for D+D where peaks from each detectors for reaction products should be integrated every hour to determine how much time before moving to the next datapoint. For lithium beamtime this should also be done in the angle vs energy plot. Note the number of counts in the spreadsheet in the same workspace.

WS6 : Browser, ELOG

-->IMPORTANT NOTE: MIDAS DAQ can respond slowly. DO NOT try to change tab or give new command while the previous one is being executed. Check activity in right-hand part of tab currently in use. Do not double click on buttons.

 

Motors computer (Computer to the right)

  • Check motors cycle is running. The code will cycle through the loaded motors instructions. When an instruction is being done it is highlighted in bold. Check the bold highlighting cycles through entire set of instructions.
  • Check potentiometer is reproducing same max/min values for each cycle of instructions.
  • Detectors are on LEFT and RIGHT arm.
  • No need for screenshots, but record in the ELOG you checked this.

 

  683   Tue May 13 15:07:42 2025 JMOnline monitoring code

CARME_Sort histogramming has been updated to aid our understanding online.

> XY plots have been updated to reflect the true position of the beam relative to each detector. Online we will be able to layout in a grid to reflect looking downstream in YR09 (see attachment 1)

> Angle vs Energy plots have been created (see attachment 2). For this the position of the detectors has to hardcoded in the EventClustering header file of the code. This has been done using the detector positions from the 426 keV/u 15N measurement in 2024 which is the reference position. To reflect the true state during the beamtime a correction to these positions is required in the EventClustering regular file (e.g x_position + 30 mm). Correction can be applied using difference between potentiometer value and reference potentiometer values. 

Ref potentiometer value LEFT = 36.54 mm

Ref potentiometer value RIGHT = 64.60 mm

> An additional condition based on the cluster y position has been added for an event to be added to the histograms. This can be used to remove edge strips affected by the beam stripe (if beam stripe can't be removed via scrapers). This allows angle vs energy and regular energy histograms to be of use online. Currently this option is set to zero, can be enabled if beam stripe persists.

> Histograms not immediately useful (such as high energy XY etc) have been commented out. This makes the histogram browser cleaner and should ease use for online shifters. (see attachment 3)

> Corrections to CARME_Sort found during analysis of 15N data have been added to the code. These include Event timing cutoffs in the EventBuilder and a corrreciton to the BuildTimestamp function in DataItems file which was causing timewarps.

> Previous CARME_Sort versions have been saved and will be uploaded to github in due time.

 

Attachment 1: Screenshot_from_2025-05-13_15-14-12.png
Screenshot_from_2025-05-13_15-14-12.png
Attachment 2: Screenshot_from_2025-05-13_15-14-38.png
Screenshot_from_2025-05-13_15-14-38.png
Attachment 3: Screenshot_from_2025-05-13_15-28-48.png
Screenshot_from_2025-05-13_15-28-48.png
  682   Mon May 12 13:25:09 2025 JGMBS DAQ - overview
An MBS DAQ is running to collect two types of data:

1. WR timestamps for machine events (injection, target ON/OFF, ...)
These are triggering the MBS individually.

2. Scaler rates for machine parameters (target pressures, ring RF, ...)
These are being readout each second by a 1Hz trigger.

DAQ quick manual


MBS node: cry_exp@r4l-70.gsi.de
directory: /esr/usr/cry_exp/2025_JET_CARME/mbs

DAQ running on: cry_exp@atppc028.gsi.de
in a screen session named "jetDAQ"

RFIO server: cry_exp@atppc028.gsi.de
in a screen session named "RFIO"
directory: /u/cry_exp/Beamtimes/202505_Li/lmd

RESTART DAQ on cry_exp@atppc028:
1. cd /u/cry_exp/Beamtimes/202505_Li
2. ./daqstart.sh

ENTER jetDAQ or RFIO screen on cry_exp@atppc028:
- screen -x jetDAQ
- screen -x RFIO

cabling documentation



MBS trigger inputs



trig. type event
1 (PENDING) ring injection
2 (PENDING) detectors IN
3 (PENDING) detectors OUT
4 (PENDING) target ON (shutter)*
5 (PENDING) target OFF (shutter)*
6 (PENDING) beam dumped
11 (PENDING) 1Hz scaler readout


*The GENESYS signal logic for the target shutter was changed several times, this means trigger 4 and 5 could be swapped. The motor signals (trigger 2 and 3), however, have not been touched and should be stable.

scaler inputs



scaler ch. rate/signal
1 gas jet E1 (inlet pressure zone 1)
2 gas jet E2 (inlet pressure zone 2)
3 gas jet E3 (inlet pressure zone 3)
4 gas jet E4 (inlet pressure zone 4)
5 gas jet S1 (outlet pressure zone 1)
6 gas jet S2 (outlet pressure zone 2)
7 gas jet S3 (outlet pressure zone 3)
8 gas jet S4 (outlet pressure zone 4)
9 gas jet T1
10 gas jet T2
11 gas jet VV1
12 gas jet ??
13 Scintillator rate (0 deg. det. YR09) [ONLY AFTER 12-06-2025 16:30]
14 -
15 -
16 -
17 cryring IMP horizontal
18 cryring IMP vertical
19 cryring PCT
20 cryring ICT int
21 cryring BPM int
22 cryring Schottky span 0
23 cry-radio
24 cryring RF
25 -
26 -
27 -
28 -
29 -
30 -
31 -
32 -
  681   Sat Apr 26 22:01:37 2025 CBProxy bridge

Proxy bridge enabling anydesk and generic Internet access dropped due to atppc025 being rebooted. Found it dropped again late today.

Restarted bridge inside a screen session.

  680   Wed Apr 23 11:08:34 2025 TD, JMBackground Run
12.08 DSSSD bias & leakage current - attachments 1-2
       CAEN logger program crashed sometime this am - now restarted
       DSSSD #2 and DSSSD #4 OK - DSSSD #4 stable since c. 03:00 20.4.25
       DSSSD #1 and DSSSD #3 remain high
       All DSSSDS show effect of ambient temperature transient in CRYRING hall

       CRYRING vacuum and temperature - attachments 3-4

14:40

FEEs started up. Background run started. Pulser running in the at 1HZ, amplitude is 9000.
Note aida03 and aida04 set to MEC due to high rate >100k. This resulted in data rates ~3MB/s. Slow comparator thresholds set to 0x64 (1 MeV)
System checks OK. AIDA16 and ADC calibration check clock status fails. 
Run 10 started. 
Attachments 5,6,7,8,9 from start of run. Waveforms, ADC data items statistics, New merger, Tape server respectively. 
Will continue to monitor during the day until tonight when I will turn off the FEEs again. 

Run stopped. 
Noticed spike in leakage current for all detectors. 
This coincides with change in vacuum around the ring. Some operation is ongoing, all FEEs off, detectors off. 
See attachment 10 for leakage currents. 

Vacuum gauge in section 9 is back on to allow for ring interlocks.

29/04/25

Turned the FEEs on to check if aida08 is still connected to the coolant loop. Temperatures in normal range after startup. FEEs turned back off for now. 
Attachment 1: Screenshot_from_2025-04-23_12-07-52.png
Screenshot_from_2025-04-23_12-07-52.png
Attachment 2: Screenshot_from_2025-04-23_12-08-10.png
Screenshot_from_2025-04-23_12-08-10.png
Attachment 3: 1.png
1.png
Attachment 4: 0.PNG
0.PNG
Attachment 5: Screenshot_from_2025-04-23_15-38-33.png
Screenshot_from_2025-04-23_15-38-33.png
Attachment 6: Screenshot_from_2025-04-23_15-40-30.png
Screenshot_from_2025-04-23_15-40-30.png
Attachment 7: Screenshot_from_2025-04-23_15-40-54.png
Screenshot_from_2025-04-23_15-40-54.png
Attachment 8: Screenshot_from_2025-04-23_15-41-13.png
Screenshot_from_2025-04-23_15-41-13.png
Attachment 9: Screenshot_from_2025-04-23_15-41-26.png
Screenshot_from_2025-04-23_15-41-26.png
Attachment 10: Screenshot_from_2025-04-23_16-07-43.png
Screenshot_from_2025-04-23_16-07-43.png
Attachment 11: Screenshot_from_2025-04-29_15-15-52.png
Screenshot_from_2025-04-29_15-15-52.png
  679   Tue Apr 22 06:44:02 2025 TDTuesday 22 April
07.40 DSSSD bias & leakage current - attachments 1-3
       CAEN logger program crashed c. 18.00 April 20 - now restarted
       DSSSD #2 and DSSSD #4 OK - DSSSD #4 stable since c. 03:00 20.4.25
       DSSSD #1 and DSSSD #3 remain high with slow decrease (ambient temperature approx stable c. 16.5 deg C)

       CRYRING vacuum and temperature - attachments 4-5
Attachment 1: Screenshot_from_2025-04-22_07-40-45.png
Screenshot_from_2025-04-22_07-40-45.png
Attachment 2: Screenshot_from_2025-04-22_07-41-10.png
Screenshot_from_2025-04-22_07-41-10.png
Attachment 3: Screenshot_from_2025-04-22_07-47-25.png
Screenshot_from_2025-04-22_07-47-25.png
Attachment 4: 1.png
1.png
Attachment 5: 2.png
2.png
  678   Sun Apr 20 07:30:29 2025 CBDSSD leakage currents over Easter

Just after midnight on 20/04 it appears DSSD #4 leakage current has perfectly stabilised. See attached.

DSSDs remain biased.

 

20 April

19:30 - leakage current still stable but anydesk connection not working.

Machine is connected, but connection fails.

 

21 April

Anydesk connection works again. Leakage current remains stable. What changed?

Attachment 1: Screenshot_2025-04-20_083007.png
Screenshot_2025-04-20_083007.png
Attachment 2: Screenshot_2025-04-20_193226.png
Screenshot_2025-04-20_193226.png
Attachment 3: Screenshot_2025-04-21_085130.png
Screenshot_2025-04-21_085130.png
  677   Fri Apr 18 16:18:33 2025 JMD+D BBN energies test

D+D minimum energy Ecm = 150 keV which corresponds to Ed=150 keV/u or Ed=300 keV. Therefore we need a threshold below this.

Test 1 - Current status

Current slow comparator threshold is 0x64 (1 MeV). ADC data items rate in all FEEs is essentially zero with this threshold. (except aida03 and aida04). Aida15 and 16 also show ~1k rates when DSSD4 current is fluctuating.

Pulser was set to find point it is cut off by the threshold. Attachments 1,2,3 show 1.8L for nn channels with the pulser set to 570,550 and 530. Over this range pulser disappears. Estimate 550 is equal to ~1 MeV.

Test 2 - Noise level at lower threshold 200 keV

Threshold set to 0x14 (200 keV). Pulser kept at 550 for the time being to gauge location of noise compared to pulser. Attachments 4,5 show 1.8L for nn and pn channels respectively. Some noise observable at low energy.

Note zero energy is around 32500 channels. In nn channels (attachment 4) pulser is to the right of the noise peak, wheras in pn channels (attachment 5) pulser is to the left of noise peak.

Attachment 6 shows data rates with pulser off. Rates okay, only aida15 has very high rate (80k), rest of channels <5k.

Test 3 - Noise level at lower threshold 100 keV

Threshold set to 0xa (100 keV). Pulser kept at 550 for the time being. Attachment 7 shows data rates with pulser off. Rates significantly higher all round. Half of channels >50k, some >100k.

Attachments 8,9 show 1.8L for pn and nn channels respectively. A lot more noise observable at low energy. Pulser peaks generally overshadowed by noise. 100 keV probably too low to set threshold.

Test 4 - Noise level at lower threshold 150 keV

Threshold set to 0xf (150 keV). Pulser kept at 550 for the time being. Attachment 10 shows data rates with pulser off. In general rates are pretty low. Most channels <1k. Aida15 and 16 pretty high.

Attachments 11,12 show 1.8L for nn and pn channels respectively. Noise observable at low energy but in general less than that of the pulser peaks.

Test 5 - D+D pulser peak at lower threshold 150 keV

Theshold kept at 0xf (150 keV). Pulser set to 180 (should be about 300 keV based on Test 1). This should be the level of d+d at our lowest beam energy.

Attachments 13,14 show 1.8L for pn and nn channels respectively.

Test 6 - D+D pulser peak at lower threshold 200 keV

Theshold increased to 0x14 (200 keV). Pulser kept at 180 (should be about 300 keV based on Test 1). This should be the level of d+d at our lowest beam energy.

Attachments 15,16 show 1.8L for nn and pn channels respectively.

 

 

 

Attachment 1: Screenshot_from_2025-04-18_15-38-00.png
Screenshot_from_2025-04-18_15-38-00.png
Attachment 2: Screenshot_from_2025-04-18_15-39-21.png
Screenshot_from_2025-04-18_15-39-21.png
Attachment 3: Screenshot_from_2025-04-18_15-40-29.png
Screenshot_from_2025-04-18_15-40-29.png
Attachment 4: Screenshot_from_2025-04-18_15-45-09.png
Screenshot_from_2025-04-18_15-45-09.png
Attachment 5: Screenshot_from_2025-04-18_15-47-36.png
Screenshot_from_2025-04-18_15-47-36.png
Attachment 6: Screenshot_from_2025-04-18_15-48-20.png
Screenshot_from_2025-04-18_15-48-20.png
Attachment 7: Screenshot_from_2025-04-18_15-50-58.png
Screenshot_from_2025-04-18_15-50-58.png
Attachment 8: Screenshot_from_2025-04-18_15-52-24.png
Screenshot_from_2025-04-18_15-52-24.png
Attachment 9: Screenshot_from_2025-04-18_15-54-06.png
Screenshot_from_2025-04-18_15-54-06.png
Attachment 10: Screenshot_from_2025-04-18_15-57-50.png
Screenshot_from_2025-04-18_15-57-50.png
Attachment 11: Screenshot_from_2025-04-18_15-59-38.png
Screenshot_from_2025-04-18_15-59-38.png
Attachment 12: Screenshot_from_2025-04-18_16-00-31.png
Screenshot_from_2025-04-18_16-00-31.png
Attachment 13: Screenshot_from_2025-04-18_16-04-04.png
Screenshot_from_2025-04-18_16-04-04.png
Attachment 14: Screenshot_from_2025-04-18_16-05-24.png
Screenshot_from_2025-04-18_16-05-24.png
Attachment 15: Screenshot_from_2025-04-18_16-08-05.png
Screenshot_from_2025-04-18_16-08-05.png
Attachment 16: Screenshot_from_2025-04-18_16-09-22.png
Screenshot_from_2025-04-18_16-09-22.png
  676   Fri Apr 18 08:17:38 2025 JMMore testing

09:10

Arrived at GSI. Leakage currents status same as before, DSSD4 is still fluctuating. DSSD1 and DSSD3 currents are definitely lower again compared to previous days. I am convinced they are decreasing over time, whether this will be enough for our measurement....

While in the cave I again removed the motors pins so that the detectors can can be shook in an attempt to perform magic. While in the cave I also connected the CRYRING ground cable to the coolant pipe (ground config 11). Will see effect on rates / pulser.

Ground config 11 - see attachments 1-8. Rates in aida03 are massively down by a factor 10! Rates now ~20k. Aida04 still around 300k. On attachment 4 we can see the noise from zero energy on aida03 is massively down and we can now see the pulser, although the peak is double. Similar low energy noise is now appearing on aida15 and aida16 of DSSD4. Is this the early phase of what happended to DSSD1 and DSSD3?

Note the waveforms shown in attachments 7 and 8 change quite alot when updating the spectrum browser. Wave structure very visible in 7 but after updating a few times this disappears before coming back again a bit later.

20 minutes later ...

DAQ continued for about 20 minutes, noticed the leakage current in DSSD4 was no longer fluctuating. Zero'd stats and histograms to look again.

Stats in aida03 have increased back to ~100k, still less than before but back up from ~20k. Check of the pulser widths shows they have decreased across the board from the previous test 20 mins ago.

Leaving DAQ to run for a bit longer, the rates in aida03 continue to increase and are now back ~200k (back where it was).

I am currently trying to cycle motors. No effect on leakage current when cycling pneumatic motor. Note when using single commands such as 'l p in' I have no problems. But when trying a longer command where I ask the code to wait before pulling the pneumatic out again the code crashes. ??? Crash not repeated by Carlo remotely, Perhaps it is my laptop, or a bug relating to windows. Will investigate back in Edinburgh.

Have performed the same bias test for DDSD4 as #1 and #2 in previous elogs. Same behaviour observed. No (0.01uA) leakage current with only nn and current ramps up and falls to zero with only pn. I did notice that the ion pumps were arcing, I have not noticed this in prevous days. They have now been turned off, I think we can leave them off as they do little for the vacuum anyway unless using heavy gas taget gas. VI bias curve looks identical to one performed at the start of the week.

I have moved the PC power to the 19 inch rack. Power cycled the FEEs. I noticed the leakage current on DSSD4 was fine during VI test but after starting the FEEs power cycle they started to fluctuate again. After about 5 minutes after the startup, DSSD4 leakage current returned to a normal stable level. I have seen spikes in leakage current for all detectors during power cycle but perhaps it could also cause the fluctuations in #4. Overnight I have left FEEs off due to uncertainty over the swagelock on aida08 and the fact I was changing ground a lot so a background didn't make sense. Perhaps leaving FEE power on will stabilise DSSD4? I will leave system to stabilise a bit now (after seeing rates and pulser widths change in previosu test) and check pulser widths. Going for an awful GSI coffee.

New configuration: See attachments 9-13. Rates on aida03 and aida04 still remain very high >200k. Pulser widths across the board look pretty good. All have reduced since moving the PC on same power supply as 19 inch rack. DSSD4 leakage current has been stable the last hour since FEE power on. Waveforms look improved too. No longer see such high amplitude oscillations. Updated several times, see no change in waveforms like previous test.

Attachment 14 for pulser widrths from tests this week. Note x means no value. This could be because the width was so very big, or the test input wasn't connected. These values aren't too bad. DSSD2 and 3 have avergae pulser width (1.8L only) of ~60 keV.

Tried ground cable to NIM bin, had no effect / maybe negative effect. Disconnected cable.

D+D tests in next elog

Finally a pulser walkthrough was completed. Run8 has pulser settings unchanged, slow comparator at 0x64, aida03 and aida04 set to MEC. Pulser amplitudes 9000,8000,7000,6000,5000,4000,3000,2000,1000

A second walkthrough (Run9) was done with slow comparator set to 0x14, aida03 and aida04 set to MEC. Pulser amplitudes 9000,8000,7000,6000,5000,4000,3000,2000,1000,700,400 (so that we can check ~BBN energies) - attachment 15

During second walkthrough, DSSD4 starting fluctuating in leakage current again. No-one was in the ring. No observable change. FEEs were ON and were currently acquiring. ?? See noise which in attachment 15 on aida15/16 that only started when current was fluctuating.

FEEs off. Bias on and monitoring. Pins re-inserted into motors. Compressed air off.

Attachment 1: Screenshot_from_2025-04-18_09-54-29.png
Screenshot_from_2025-04-18_09-54-29.png
Attachment 2: Screenshot_from_2025-04-18_09-55-40.png
Screenshot_from_2025-04-18_09-55-40.png
Attachment 3: Screenshot_from_2025-04-18_09-56-51.png
Screenshot_from_2025-04-18_09-56-51.png
Attachment 4: Screenshot_from_2025-04-18_10-02-05.png
Screenshot_from_2025-04-18_10-02-05.png
Attachment 5: Screenshot_from_2025-04-18_10-03-08.png
Screenshot_from_2025-04-18_10-03-08.png
Attachment 6: Screenshot_from_2025-04-18_10-06-45.png
Screenshot_from_2025-04-18_10-06-45.png
Attachment 7: Screenshot_from_2025-04-18_10-07-57.png
Screenshot_from_2025-04-18_10-07-57.png
Attachment 8: Screenshot_from_2025-04-18_10-08-10.png
Screenshot_from_2025-04-18_10-08-10.png
Attachment 9: Screenshot_from_2025-04-18_13-27-39.png
Screenshot_from_2025-04-18_13-27-39.png
Attachment 10: Screenshot_from_2025-04-18_13-36-29.png
Screenshot_from_2025-04-18_13-36-29.png
Attachment 11: Screenshot_from_2025-04-18_13-42-42.png
Screenshot_from_2025-04-18_13-42-42.png
Attachment 12: Screenshot_from_2025-04-18_13-43-22.png
Screenshot_from_2025-04-18_13-43-22.png
Attachment 13: Screenshot_from_2025-04-18_13-44-12.png
Screenshot_from_2025-04-18_13-44-12.png
Attachment 14: pulser-widths.png
pulser-widths.png
Attachment 15: Screenshot_from_2025-04-18_17-03-56.png
Screenshot_from_2025-04-18_17-03-56.png
  675   Thu Apr 17 08:51:29 2025 JMGrounding cont.

09:00

Came in to see the bias for DSSD4 which has been fluctuating for the past few days had tripped during the night. From the gnuplot it looks like it was fluctuating all night and just after midnight it went above 20uA and tripped.

Other leakage currents are around the value they have been recently, day night cycle still observed. Again perhaps the leakage current of DSSD1 and DSSD3 is slowly decreasing and reaching a lower minimum value. Reset the module, DSSD4 does bias back up and is currently around 10uA. See attachments 1+2.

I have also noticed a small spike in current when starting up the FEE power, see the small spikes in attachment 2.

Continuing with different ground configurations from yesterday.

Ground config 5: I have now removed the grounding cables between non-bias FEE modules. So grounding cable goes between pn and nn bias FEEs and then one grounding cable to the coolant pipe. No grounding on any other FEEs. (attachments 3-8). Effect on pulser widths is hard to discern, some widths have decreased i.e 5 and 6 wheras others have increased such as 11. One interesting note is that the rates on aida03 have decreased by just over a factor 2 so now read ~100k. Whether this is the result of different grounding or maybe similar to what was seen for DSSD3 where rates decreased on there own over time? Attachment 8 shows the rates for each ASIC and most have now reached similar level to the other FEEs (note log scale).

Ground config 6: Have kept grounds to only pn or nn bias FEEs. Have now also included a ground cable from the body of the FEE module into the grounding connection. The ADC data items rate increase from 100k to 150k in aida03, no other impact on rates. Pulser widths remain largely unchanged, although some widths saw increases in the pulser widths. (attachments 9-13)

Ground config 7: Wanted to check impact of having no grounds at all. No impact on leakage currents. Rates look pretty similar for all FEEs, aida03 and aida04 200k and 350k respectively. Pulser widths and waveforms significantly worse than grounded case. See attachments 14-18

Ground config 8: Have reattached all ground cables (back to config1). In addition I have added a ground cable from the coolant pipe to effectively the CARME chamber. Rates appear pretty unchanged. Pulser widths better than no grounding but worse than original configuration. Attachments 19-23

Note since after lunch DSSD4 leakage current no longer fluctuating. Nothing has been changed. As I write that currents start fluctuating again. Maybe some ongoing work in the cave is affecting this?

Removed groud cable going to the chamber.

I wanted to rule out any affect from the test inputs. Have removed all test inputs. Only input is directly from pulser to DSSD3 as this previously had best pulser widths. All other cables removed. See attachments 24-26. Still see very high rates in aida03 and aida04 (250k and 350k). Pulser widths in DSSD3 seem worse now for some reason? While pulser detattched decided to lower slow comparator to 0x32 (500keV) and 0x24 (360 keV) to see rates in all FEEs. Rates still look pretty good, other than aida03 and 04. See attachments 27,29. Putting test inputs back.

Test inputs all back. Ground configuration back to configuration 1. Pulser tests, widths are best so far - will upload them later from all tests.Rates good in all except aida03 and aida04 still.

FEEs off, bias on and monitoring. Have put pins back in motors.

20:50

Bias has tripped again on DSSD4. It had remiained stable with noi fluctuations ~20 mins before leaving.

Attachment 1: Screenshot_from_2025-04-17_09-40-03.png
Screenshot_from_2025-04-17_09-40-03.png
Attachment 2: Screenshot_from_2025-04-17_09-40-23.png
Screenshot_from_2025-04-17_09-40-23.png
Attachment 3: Screenshot_from_2025-04-17_09-57-48.png
Screenshot_from_2025-04-17_09-57-48.png
Attachment 4: Screenshot_from_2025-04-17_09-58-31.png
Screenshot_from_2025-04-17_09-58-31.png
Attachment 5: Screenshot_from_2025-04-17_10-02-08.png
Screenshot_from_2025-04-17_10-02-08.png
Attachment 6: Screenshot_from_2025-04-17_10-06-07.png
Screenshot_from_2025-04-17_10-06-07.png
Attachment 7: Screenshot_from_2025-04-17_10-07-29.png
Screenshot_from_2025-04-17_10-07-29.png
Attachment 8: Screenshot_from_2025-04-17_10-11-12.png
Screenshot_from_2025-04-17_10-11-12.png
Attachment 9: Screenshot_from_2025-04-17_11-53-54.png
Screenshot_from_2025-04-17_11-53-54.png
Attachment 10: Screenshot_from_2025-04-17_11-57-21.png
Screenshot_from_2025-04-17_11-57-21.png
Attachment 11: Screenshot_from_2025-04-17_12-02-10.png
Screenshot_from_2025-04-17_12-02-10.png
Attachment 12: Screenshot_from_2025-04-17_12-03-28.png
Screenshot_from_2025-04-17_12-03-28.png
Attachment 13: Screenshot_from_2025-04-17_12-04-34.png
Screenshot_from_2025-04-17_12-04-34.png
Attachment 14: Screenshot_from_2025-04-17_13-20-17.png
Screenshot_from_2025-04-17_13-20-17.png
Attachment 15: Screenshot_from_2025-04-17_13-22-07.png
Screenshot_from_2025-04-17_13-22-07.png
Attachment 16: Screenshot_from_2025-04-17_13-24-07.png
Screenshot_from_2025-04-17_13-24-07.png
Attachment 17: Screenshot_from_2025-04-17_13-25-57.png
Screenshot_from_2025-04-17_13-25-57.png
Attachment 18: Screenshot_from_2025-04-17_13-26-32.png
Screenshot_from_2025-04-17_13-26-32.png
Attachment 19: Screenshot_from_2025-04-17_14-45-50.png
Screenshot_from_2025-04-17_14-45-50.png
Attachment 20: Screenshot_from_2025-04-17_14-47-34.png
Screenshot_from_2025-04-17_14-47-34.png
Attachment 21: Screenshot_from_2025-04-17_14-50-31.png
Screenshot_from_2025-04-17_14-50-31.png
Attachment 22: Screenshot_from_2025-04-17_14-52-29.png
Screenshot_from_2025-04-17_14-52-29.png
Attachment 23: Screenshot_from_2025-04-17_14-53-18.png
Screenshot_from_2025-04-17_14-53-18.png
Attachment 24: Screenshot_from_2025-04-17_15-41-38.png
Screenshot_from_2025-04-17_15-41-38.png
Attachment 25: Screenshot_from_2025-04-17_15-45-34.png
Screenshot_from_2025-04-17_15-45-34.png
Attachment 26: Screenshot_from_2025-04-17_15-46-19.png
Screenshot_from_2025-04-17_15-46-19.png
Attachment 27: Screenshot_from_2025-04-17_15-54-46.png
Screenshot_from_2025-04-17_15-54-46.png
Attachment 28: Screenshot_from_2025-04-17_15-57-26.png
Screenshot_from_2025-04-17_15-57-26.png
Attachment 29: Screenshot_from_2025-04-17_15-58-47.png
Screenshot_from_2025-04-17_15-58-47.png
  674   Wed Apr 16 08:20:05 2025 JMGrounding

09:00

Leakage currents this morning again appear to have lowered during the night. Current plot shows a day/night variation but I think the minimum is decreasing ever so slightly.

This morning DSSD1=15.1uA and DSSD3=14.4uA, this is ~0.3-0.4 uA lower than yesterday. DSSD4 is still fluctuating in its leakage current and looks to have been all night. See attachment 1.

Will now proceed to systematically check all grounds before restarting AIDA.

Several cables have been re-routed this morning to clean up the area, allow us to see clear what cable goes where, and are avoid any unwanted interference. E.g. ion pump cables have been re-routed to not be near the bias cables, one ion pump cable was running between the power and hdmi on one of the FEEs, this was been sorted.

All grounds have been checked, some loose cables have been tightend up. Current ground configuration has all the adaptor cards (nn and pn) on a single detector connected together. One grounding cable then goes from each detector to the right coolant pipe which all the other detectors are also grounded to. This was the configuration for the previous beamtime. Will start pulser tests after lunch. Bias back on, leakage current slowly increasing again, DSSD3 current no longer fluctuating.

13:00

Ground config 1 - attachments 2-7 (described above)

Ground config 2 - attachments 8-10 (grounds now go to the CRYRING ground cable not the coolant pipe and jumper removed form bias module)

pulser widths 1.8L worse pretty much across the board. No sig impact on rates

Ground config 3 - attachments 11-14 (grounds returned to coolant pipe)

pulser widths 1.8L better than config 2 but still worse than config1

Ground configuration returned to first configuration. Pulser widths consistent with that seen before.

Additional bias tests.

Disconnected the pn bias cable (core) from DSSD1 but left nn bias cable connected. Biased. Observed bias ramp up but no leakage current was seen.

Reconnected pn bias cable but now disconnected nn bias cable. Biased. Observed bias ramp up, leakage current was seen to increase during ramp to ~8uA. Once fully ramped, leakage current started to slowly decrease and eventually went to zero.

Repeated same tests with DSSD2. Same behaviour observed as with DSSD1.

Have removed motors pins and tried some movement again. No effect on leakage current. Checked pulser widths before movement and after movement (with pnuematic IN). No change in pulser widths.

End of day status FEEs off, bias on and monitoring. Motor pins out to allow for last movement check, motors code running in screen session.

Attachment 1: Screenshot_from_2025-04-16_09-19-34_1.png
Screenshot_from_2025-04-16_09-19-34_1.png
Attachment 2: Screenshot_from_2025-04-16_14-18-39.png
Screenshot_from_2025-04-16_14-18-39.png
Attachment 3: Screenshot_from_2025-04-16_14-21-10.png
Screenshot_from_2025-04-16_14-21-10.png
Attachment 4: Screenshot_from_2025-04-16_14-25-23.png
Screenshot_from_2025-04-16_14-25-23.png
Attachment 5: Screenshot_from_2025-04-16_14-26-33.png
Screenshot_from_2025-04-16_14-26-33.png
Attachment 6: Screenshot_from_2025-04-16_14-29-51.png
Screenshot_from_2025-04-16_14-29-51.png
Attachment 7: Screenshot_from_2025-04-16_14-30-37.png
Screenshot_from_2025-04-16_14-30-37.png
Attachment 8: Screenshot_from_2025-04-16_15-31-45.png
Screenshot_from_2025-04-16_15-31-45.png
Attachment 9: Screenshot_from_2025-04-16_15-36-24.png
Screenshot_from_2025-04-16_15-36-24.png
Attachment 10: Screenshot_from_2025-04-16_15-41-11.png
Screenshot_from_2025-04-16_15-41-11.png
Attachment 11: Screenshot_from_2025-04-16_16-49-43.png
Screenshot_from_2025-04-16_16-49-43.png
Attachment 12: Screenshot_from_2025-04-16_16-55-09.png
Screenshot_from_2025-04-16_16-55-09.png
Attachment 13: Screenshot_from_2025-04-16_16-55-44.png
Screenshot_from_2025-04-16_16-55-44.png
Attachment 14: Screenshot_from_2025-04-16_17-42-40.png
Screenshot_from_2025-04-16_17-42-40.png
  673   Tue Apr 15 12:23:10 2025 JMLeakage current / CARME issues

14/04/25

Arrived at GSI in the afternoon.

Found that the coolant loop on aida08 had popped out which resulted in the high FEE temperatures observed (see attachment 1). Female swagelock connector on FEE suggested to be the cause as this connector has the locking mechanism. FEE dismounted and female swagelock connector removed. No obvious defect in the connector (see attachment 2).

Turning on detectors we observe ~4uA in DSSD2 and DSSD4 (right hand side detectors) and ~15uA in DSSD1 and DSSD3 (left detectors). Leakage currents in left hand side detectors is down from ~17uA seen at the end of last week and ~19uA at start of last week. Temperatures in the cave do not appear to have changed sufficiently to explain this change. VI curve completed to test diode response (attachment 3+4).

Motors turned on to again see if it changes the leakage currents in any way. Both arms homed. Left arm moved in with pneumatic and servo (10mm). No change in leakage currents during movement or when fully in. Same procedure with the right arm. No change in leakage currents in either. I doubt this is a light affect. If some light was incident on the left hand detectors, moving the detectors should have changed the amount of light on them in some way, reducing the leakage current. Similarly if light can hit the detector when fully in for the left arm, some should be hitting the right arm when it is fully in too, but no change observed on the right arm either. No light sources found on interaction chamber, baking jackets covers all ports.

Movement for several repeptitions with the pneumatic tried to see if this had any affect -> no change

Bias lines for DSSD1 and DSSD2 swapped around on the caen bias module. No change in leakage current for DSSD1 or 2. Channel on bias module not to blame.

Tried turning ion pumps off, no change on leakage currents.

Tried reverse biasing the detectors. Polarity for channels 0 (DSSD1) and 2 (DSSD3) swapped from -ve to +ve on bias module. Core and braid swapped on adaptor card. Completed VI curve when reverse biased for these detectrors. Obtain the same curve as with -ve polarity. Left at 100V for 10-15 minutes. Put polarities back to -ve, no change in leakage currents observed.

Opended up one of the diamond windows with bias off. Shone my phone light through the window for ~30 seconds. Closed window off again, biased, no change observed. Left detectors biased overnight. CH0 - 15.4 uA, CH2-14.7uA.

15/04/25

No change in leakage currents overnight. From gnuplot they look pretty flat overnight. Perhaps slight decrease.

Checked attaching LK2 jumper to aida09 (nn DSSD3). Could not reach other nn adaptor card. No effect on leakage current. Dismounted jumper.

New female connector mounted on aida08 and FEE remounted.

Restarted AIDA system by closing all windows and servers. Error on starup, could not connect to relay16 for FEE power.

Issue related to dew point interlock. Dew point sensor is very sensitive and any movement can cause it to trip. Sensor was not well supported on the coolant pipe. Sensor remounted with plastic zip ties. Issue resolved. Investigate some noise issues in the afternoon

After lunch I have noticed the leakage current have increased slightly. See attachment 5. CH3 is higher and appears to be fluctuating by 1-2uA. Turning off bias and going to cave to investigate.

Bias still higher and fluctating on CH3. Konstantin is in the ring, but he has not started any works yet, some grossmontage are also in the ring and are working on the new source. Could this be affecting us? Either way, will startup daq to check rates/noise etc.

Previous settings applied (i.e all channels back to LEC). Have lost the dew point interlock again!

I have suspended the dew point sensor in the air for the time being -> need to put a dummy in.

Grossmontage working on ion source have left and leakage current on DSSD3 is no longer fluctuating -> coincidence? (attachment 6)

AIDA started again. Rates appear OK except for on 3,4 and 15. (see attachement 7,8,9)  Note the rates on 9,10 are reduced compared to previous elog entry. Note pulser off. Note on aida startup aida16 clock status fail and ADC calibration fail for aida 6,14 and 16

16:00

We again see DSSD3 leakage current fluctuating. I will head back to the cave to check if anyone is working again? Nikos and davide working on the target but can't see how they would impact the leakage current (they were only installing a support for the new turbo)

DSSD3 remains fluctuating even now they have stopped working.

Pulser on. Same settings as elog 664 -> freq = 10 Hz

Overall pulser widths look bad, rates (ADC data items) look okay (~1000) for all except aida03 and aida04 (>250,000). Again rates on aida09 and aida10 are good although the widths are poor. One change I did make was to ground the FEEs to the CRYRING ground wheras previously they were grounded to the coolant pipe. I may try putting them back onto the coolant loop to see if there are any changes. Note grounds all around have not been examined yet.  Attachments 10-17 with pulser on.

Moved grounds back onto the coolant pipe -> no effect on ADC rates.

Also tried lowing slow comparator theshold down to 150 keV (0xf) for most channels (except aida3 and 4). See rates attachment 18.

FEE power off, bias remains on

 

Attachment 1: 20250414_120219.jpg
20250414_120219.jpg
Attachment 2: pic2.jpg
pic2.jpg
Attachment 3: left-VI.png
left-VI.png
Attachment 4: right-VI.png
right-VI.png
Attachment 5: Screenshot_from_2025-04-15_13-53-33.png
Screenshot_from_2025-04-15_13-53-33.png
Attachment 6: Screenshot_from_2025-04-15_15-43-37.png
Screenshot_from_2025-04-15_15-43-37.png
Attachment 7: Screenshot_from_2025-04-15_15-22-44.png
Screenshot_from_2025-04-15_15-22-44.png
Attachment 8: Screenshot_from_2025-04-15_15-28-58.png
Screenshot_from_2025-04-15_15-28-58.png
Attachment 9: Screenshot_from_2025-04-15_15-29-29.png
Screenshot_from_2025-04-15_15-29-29.png
Attachment 10: Screenshot_from_2025-04-15_16-21-16.png
Screenshot_from_2025-04-15_16-21-16.png
Attachment 11: Screenshot_from_2025-04-15_16-23-21.png
Screenshot_from_2025-04-15_16-23-21.png
Attachment 12: Screenshot_from_2025-04-15_16-24-20.png
Screenshot_from_2025-04-15_16-24-20.png
Attachment 13: Screenshot_from_2025-04-15_16-24-54.png
Screenshot_from_2025-04-15_16-24-54.png
Attachment 14: Screenshot_from_2025-04-15_16-25-36.png
Screenshot_from_2025-04-15_16-25-36.png
Attachment 15: Screenshot_from_2025-04-15_16-26-13.png
Screenshot_from_2025-04-15_16-26-13.png
Attachment 16: Screenshot_from_2025-04-15_16-26-48.png
Screenshot_from_2025-04-15_16-26-48.png
Attachment 17: Screenshot_from_2025-04-15_16-28-02.png
Screenshot_from_2025-04-15_16-28-02.png
Attachment 18: Screenshot_from_2025-04-15_17-27-12.png
Screenshot_from_2025-04-15_17-27-12.png
  672   Thu Apr 10 10:53:14 2025 TDThursday 10 April
11.50 DAQ continues
      Data file APR25/R6_13

      DSSSD bias & leakage current - attachment 1
      
      FEE64 temps OK - attachment 2

      ADC data item stats - attachment 3

      Merger & Tape Server - attachment 4-5

      Disk file log - attachment 6
       event rate increased significantly at some point this morning but is currently OK 


19.21 FEE64 power OFF - aida08 high temperature

      DSSSD bias & leakage current - attachment 7

      FEE64 temps - attachment 8
       aida08 high temperature - FEE64 power OFF, detector HV OFF
Attachment 1: Screenshot_from_2025-04-10_11-50-32.png
Screenshot_from_2025-04-10_11-50-32.png
Attachment 2: Screenshot_from_2025-04-10_11-51-02.png
Screenshot_from_2025-04-10_11-51-02.png
Attachment 3: Screenshot_from_2025-04-10_11-51-35.png
Screenshot_from_2025-04-10_11-51-35.png
Attachment 4: Screenshot_from_2025-04-10_11-51-48.png
Screenshot_from_2025-04-10_11-51-48.png
Attachment 5: Screenshot_from_2025-04-10_11-52-02.png
Screenshot_from_2025-04-10_11-52-02.png
Attachment 6: Screenshot_from_2025-04-10_11-53-45.png
Screenshot_from_2025-04-10_11-53-45.png
Attachment 7: Screenshot_from_2025-04-10_19-20-23.png
Screenshot_from_2025-04-10_19-20-23.png
Attachment 8: Screenshot_from_2025-04-10_19-20-51.png
Screenshot_from_2025-04-10_19-20-51.png
  671   Thu Apr 10 07:02:12 2025 CB (remotely)Current logging enabled - sort of

There appear to be issues with Shahab's caenlogging program. It crashes occasionally, with an error indicating the number of columns in the terminal is incorrect. We will have to ask Shahab.

For now

1. Start the caen_screen.sh script on nnrp1 (aka CARMEMotorsPi -  why was it renamed? It should stay as CARMEMotorsPi to avoid duplicate names).

This starts connect to the caen power module via TTY inside a screen called caenhv. Currently on desktop 1

 

2. On nnrpi1, start the caenlogging program --settings caenlogger_settings_gsi.toml

 

3. On carme-gsi, in ~/Programs/caenlogger start the gnuplot script

 

If the caenlogger script crashes, restart it on the second terminal

More investigations required on the crash.

  670   Wed Apr 9 10:02:21 2025 TDCARME equipment in CRYRING area moved
There was some tidying up of the CRYRING hall in advance of this week's EC evaluation meeting at GSI.

CARME equipment stored in central area of the CRYRING hall has been moved to the roof storage area near rooms 2.033/4 and a cupboard near the main entrance to the CRYRING 
hall - attachments 1-2.

Please contact Davide Racano for further information.
Attachment 1: 20250409_102115.jpg
20250409_102115.jpg
Attachment 2: 20250409_101705.jpg
20250409_101705.jpg
  669   Wed Apr 9 02:40:04 2025 TDWednesday 9 April
03.37 DAQ continues
      Data file APR25/R5_8

      DSSSD bias & leakage current - attachment 1
      
      FEE64 temps OK - attachment 2

      ADC data item stats - attachment 3

      Merger & Tape Server - attachment 4-5

10.23 New run 
      Data file APR25/R6
      alpha background - pulser OFF
      all histograms & stats zero'd

      DSSSD bias & leakage current - attachment 6
      
      FEE64 temps OK - attachment 7

      ADC data item stats - attachment 8

      Merger & Tape Server - attachment 9-10

      WR timestamps OK - attachment 11
Attachment 1: Screenshot_from_2025-04-09_03-37-36.png
Screenshot_from_2025-04-09_03-37-36.png
Attachment 2: Screenshot_from_2025-04-09_03-38-05.png
Screenshot_from_2025-04-09_03-38-05.png
Attachment 3: Screenshot_from_2025-04-09_03-38-36.png
Screenshot_from_2025-04-09_03-38-36.png
Attachment 4: Screenshot_from_2025-04-09_03-38-59.png
Screenshot_from_2025-04-09_03-38-59.png
Attachment 5: Screenshot_from_2025-04-09_03-39-18.png
Screenshot_from_2025-04-09_03-39-18.png
Attachment 6: Screenshot_from_2025-04-09_10-24-34.png
Screenshot_from_2025-04-09_10-24-34.png
Attachment 7: Screenshot_from_2025-04-09_10-24-56.png
Screenshot_from_2025-04-09_10-24-56.png
Attachment 8: Screenshot_from_2025-04-09_10-25-49.png
Screenshot_from_2025-04-09_10-25-49.png
Attachment 9: Screenshot_from_2025-04-09_10-26-39.png
Screenshot_from_2025-04-09_10-26-39.png
Attachment 10: Screenshot_from_2025-04-09_10-26-53.png
Screenshot_from_2025-04-09_10-26-53.png
Attachment 11: Screenshot_from_2025-04-09_10-29-58.png
Screenshot_from_2025-04-09_10-29-58.png
  668   Tue Apr 8 10:46:13 2025 TDTuesday 8 April
11.30 CRYRING vacuum and temperature - attachments 1-2

      DSSSD bias & leakage current - attachment 3
       c. 1uA reduction in DSSSD#1 and DSSSD#2 leakage currents cf. yesterday but remain high cf.June 2024

      FEE64 temps OK - attachment 4

      System wide checks - clock fail aida16 - attachments 5-6
       aida16 global clock status 0x6 - non-critical failure, often resolved by a power cycle.

      ADC data item stats - attachment 7
       DSSSD #3 n+n Ohmic side FEE64 rates return to c. normal cf. yesterday


      per FEE64 Rate spectra - attachment 8

      per FEE64 1.8.L spectra - attachments 9-10
       pulser peak width aida11 55 ch FWHM, aida09 75 ch FWHM
       only DSSSD #3 aida11/aida12 have expected peak widths

      per FEE64 1.8.W spectra - 20us FSR - attachments 11-13

      WR timestamps OK - attachment 14

17.29 DSSSD bias & leakage current - attachment 15
      
      FEE64 temps OK - attachment 16

      WR timestamps OK - attachment 17

      ADC data item stats - attachment 18

      per FEE64 Rate spectra - attachment 19

17.35 All histograms & stats zero'd
      Start run
      Data file APR25/R5
      alpha background

      EG&G Ortec 448 Pulser
       amplituide 90,000
       normalise 5.00
       x5 attenuator IN
       INT OSC
       frequency 2Hz
       t_r 100ns
       tau_d 50us (1000us setting does not work)
       polarity +

       ASIC settings 2025Apr08-13.45.30
        slow comparator 0x64
        aida03, aida04, aida09, aida10 LEC/EC range setting -> MEC

      per FEE64 Rate spectra - attachment 20

      Merger & Tape Server - attachment 21-22

20.41 DAQ continues
      Data file APR25/R5_7

      DSSSD bias & leakage current - attachment 23
      
      FEE64 temps OK - attachment 24

      ADC data item stats - attachment 25

      Merger & Tape Server - attachment 26-27

      per FEE64 Rate spectra - attachment 19

        
Attachment 1: Capture_2.PNG
Capture_2.PNG
Attachment 2: Capture.PNG
Capture.PNG
Attachment 3: Screenshot_from_2025-04-08_11-29-48.png
Screenshot_from_2025-04-08_11-29-48.png
Attachment 4: Screenshot_from_2025-04-08_11-30-14.png
Screenshot_from_2025-04-08_11-30-14.png
Attachment 5: Screenshot_from_2025-04-08_11-38-06.png
Screenshot_from_2025-04-08_11-38-06.png
Attachment 6: Screenshot_from_2025-04-08_11-38-29.png
Screenshot_from_2025-04-08_11-38-29.png
Attachment 7: Screenshot_from_2025-04-08_11-38-46.png
Screenshot_from_2025-04-08_11-38-46.png
Attachment 8: Screenshot_from_2025-04-08_11-39-39.png
Screenshot_from_2025-04-08_11-39-39.png
Attachment 9: Screenshot_from_2025-04-08_11-41-09.png
Screenshot_from_2025-04-08_11-41-09.png
Attachment 10: Screenshot_from_2025-04-08_11-43-20.png
Screenshot_from_2025-04-08_11-43-20.png
Attachment 11: Screenshot_from_2025-04-08_11-43-56.png
Screenshot_from_2025-04-08_11-43-56.png
Attachment 12: Screenshot_from_2025-04-08_11-44-24.png
Screenshot_from_2025-04-08_11-44-24.png
Attachment 13: Screenshot_from_2025-04-08_11-45-11.png
Screenshot_from_2025-04-08_11-45-11.png
Attachment 14: Screenshot_from_2025-04-08_13-36-00.png
Screenshot_from_2025-04-08_13-36-00.png
Attachment 15: Screenshot_from_2025-04-08_17-29-11.png
Screenshot_from_2025-04-08_17-29-11.png
Attachment 16: Screenshot_from_2025-04-08_17-29-35.png
Screenshot_from_2025-04-08_17-29-35.png
Attachment 17: Screenshot_from_2025-04-08_17-30-30.png
Screenshot_from_2025-04-08_17-30-30.png
Attachment 18: Screenshot_from_2025-04-08_17-30-56.png
Screenshot_from_2025-04-08_17-30-56.png
Attachment 19: Screenshot_from_2025-04-08_17-28-57.png
Screenshot_from_2025-04-08_17-28-57.png
Attachment 20: Screenshot_from_2025-04-08_17-37-47.png
Screenshot_from_2025-04-08_17-37-47.png
Attachment 21: Screenshot_from_2025-04-08_17-35-42.png
Screenshot_from_2025-04-08_17-35-42.png
Attachment 22: Screenshot_from_2025-04-08_17-35-54.png
Screenshot_from_2025-04-08_17-35-54.png
Attachment 23: Screenshot_from_2025-04-08_20-40-46.png
Screenshot_from_2025-04-08_20-40-46.png
Attachment 24: Screenshot_from_2025-04-08_20-41-39.png
Screenshot_from_2025-04-08_20-41-39.png
Attachment 25: Screenshot_from_2025-04-08_20-41-56.png
Screenshot_from_2025-04-08_20-41-56.png
Attachment 26: Screenshot_from_2025-04-08_20-42-29.png
Screenshot_from_2025-04-08_20-42-29.png
Attachment 27: Screenshot_from_2025-04-08_20-42-48.png
Screenshot_from_2025-04-08_20-42-48.png
  667   Mon Apr 7 08:59:41 2025 TDMonday 7 April
10.00 FEE64 power ON
      CARME IE514 OFF

      Confirm DAQ SETUP restores ASIC settings 2024Apr24-14.14.03 for *all* FEE64s

      DSSSD bias & leakage current - attachment 1
       leakage currents remain high cf. June 2024 - status of vacuum group controlled ion gauge unknown

12.10 FEE64 power ON

      DSSSD bias & leakage current - attachment 2

      FEE64 temps OK - attachment 3

      All system wide checks OK

      WR timestamps OK - attachment 4

      per FEE64 Rate spectra - attachments 5-6

      ADC data item stats - attachment 7

13.45 ML says vacuum group have switched off YR09DD1AS
      DSSSD bias & leakage currents - attachment 8
       DSSSD #2 and DSSSD #4 leakage currents return to 'normal' values cf. June 2024
       DSSSD #1 and DSSSD #3 leakage currents reduce by c. 10-15% only and remain high cf. June 2024 

17.47 DSSSD bias OFF
      FEE6 Power OFF
   
Attachment 1: Screenshot_from_2025-04-07_10-01-15.png
Screenshot_from_2025-04-07_10-01-15.png
Attachment 2: Screenshot_from_2025-04-07_12-14-02.png
Screenshot_from_2025-04-07_12-14-02.png
Attachment 3: Screenshot_from_2025-04-07_12-14-16.png
Screenshot_from_2025-04-07_12-14-16.png
Attachment 4: Screenshot_from_2025-04-07_12-14-57.png
Screenshot_from_2025-04-07_12-14-57.png
Attachment 5: Screenshot_from_2025-04-07_12-16-12.png
Screenshot_from_2025-04-07_12-16-12.png
Attachment 6: Screenshot_from_2025-04-07_12-16-36.png
Screenshot_from_2025-04-07_12-16-36.png
Attachment 7: Screenshot_from_2025-04-07_12-16-45.png
Screenshot_from_2025-04-07_12-16-45.png
Attachment 8: Screenshot_from_2025-04-07_14-06-10.png
Screenshot_from_2025-04-07_14-06-10.png
  666   Sun Apr 6 14:21:53 2025 TD[How To] Startup CARME
- water flow/interlock checks
   check water flow to/from inflow/outflow manifolds
   check all coolant lines to all FEE64s are connected
   check USB-controlled ac mains relay 'water' LED illuminated                                 water flow and dew point sensor interlocks OK

- carme-gsi network cabling
   from patch panel YR10/2
    yellow network cable to carme-gsi operating in 19" rack room (2.033)
    green network cable to carme-gsi operating in counting room (2.034)

- boot carme-gsi
   systemctl start dhcpd                                                                       start DHCP dynamic network configuration 
   ssh -L 8080:proxy.gsi.de:8080 carme@atppc025                                                establish proxy port for AnyDesk access
   ssh -X pi@nnrpi1
    /MIDAS/TclHttpd/linux-arm/TclHttpd-server &                                                start MIDAS server
   ssh -X pi@nnrpi1
    ./caen_screen.sh                                                                           start CAEN N1419ET control interface

- enable current logging
  open a new terminal
  ssh pi@nnrpi1
  cd Programs/caenlogger
  caenlogger --config caenlogger_config_gsi.toml
   
  open a new terminal
  cd Programs/caenlogger
  gnuplot plot_current.gp

- system log
   sudo tail -f /var/log/messages                                                              monitor system log e.g. filesystem mounts for FEE64 embedded processors
   (sudo does not work - use su instead. will have to fix)

- FEE64 power ON
   Start browser on nnrpi1:8015
   Power On

- Start MIDAS
   select server start icon (leftmost)                                                            1x terminal session will appear
   Start browser tab localhost:8015
 
   Data Acquisition Control Services
   Update (nothing should change)

   Act on all Data acquisition server TICKED   
   RESET
   SETUP
   
   Enable histograms and waveforms

  Open ASIC control
  Act on all FEE64 modules
  Act on all ASICS
  System functions -> Check ASIC control

  Check FEE64 temperatures (Reload)

  System wide checks
     Clock
     ADC Calibration
     White Rabbit
     Drop down -> collect FPGA errors
     Check FPGA
     Synchronise ASIC clocks

  FADC Align & Check - all FEE64s
  Drop down -> calibrate ADCs
  System wide checks
    Clock
    ADC calibration
  If any FEE fails, FADC calibrate it again

  DAQ GO

  WR timestamp                                                                  check WR timestamp monotonically increases aida01->aida16          


  To startup Merger and Tape Server

  DAQ STOP                                                 

  8115 server, tape server, merger -> from buttons on top 
         3x terminal sessions will appear

  Start browser tab localhost:8115
  New Merge control
  Tape Service
  NetVar Service

  Merger tab
   (top) setup
   go (ignore error message)
   enable output to data storage                                                Merger State = GOing : merge/running : xfer Links => Merger => Storage
  
  tape server
   Inquire Available Devices
   Allocate Device
   
   Enter volume name (if already defined)
   Mount volume
   Reload   
   Set run number
   Open file

   Open a new terminal
   Navigate to /TapeData/ and the volume name
   watch ls -lht                                                                

   Enable data transfer #1 (in Run Control)
   X should appear next to every FEE


  Check netvar variable NetVar.MERGE.RunOptions = 1

   DAQ GO
    check merger rate and activity of all FEE64s
    check tape server blocks written number is increasing
   Server state -> Go
  665   Sun Apr 6 11:40:48 2025 TDOptions DB reconstructed
Options directory corrupted unable to change Options settings per https://elog.ph.ed.ac.uk/CARME/664

Move old Options directory to Options.BAK-060425

Created new Options directory (777), aida01-aida16 sub directories (777) and populated using CONTENTS file from the old aida01 options (777)
All FEE64s will now use ASIC.settings 2024Apr24-14.14.03

[npg@carme-gsi ~]$ cd /MIDAS/DB/EXPERIMENTS/AIDA
[npg@carme-gsi AIDA]$ ls
2017Nov15-15.21.56  2019Dec19-16.19.51  2021Apr29-13-16-00  2024Feb08-15.59.09  CONTENTS        Options.311019         Options.BAK-060425  Options.PRISTINE         README
2018Aug21-13.17.09  2019Mar31-16.15.57  2022Jan31-11-52-00  2024Jan20-12.02.08  Histograms      Options.BACKUP180920   Options.BAK-130421  Options.PRISTINE_TRIPLE
2018Jun22-11.28.18  2019Oct31-13.24.23  2024Apr24-14.14.03  aida_check_db.sh    Options         Options.BACKUP190425   Options.Current     Options.WORKING
2018Sep06-14.38.28  2020Apr15-20.29.26  2024Feb05-10.22.28  aida_check_db.sh~   Options.220131  Options.BACKUPCorrupt  Options.Dec19       Options.WORKING.Old
[npg@carme-gsi AIDA]$ ls -lR Options
Options:
total 0
drwxrwxrwx. 2 npg npg 22 Apr  6 12:33 aida01
drwxrwxrwx. 2 npg npg 22 Apr  6 12:35 aida02
drwxrwxrwx. 2 npg npg 22 Apr  6 12:35 aida03
drwxrwxrwx. 2 npg npg 22 Apr  6 12:35 aida04
drwxrwxrwx. 2 npg npg 22 Apr  6 12:35 aida05
drwxrwxrwx. 2 npg npg 22 Apr  6 12:35 aida06
drwxrwxrwx. 2 npg npg 22 Apr  6 12:35 aida07
drwxrwxrwx. 2 npg npg 22 Apr  6 12:35 aida08
drwxrwxrwx. 2 npg npg 22 Apr  6 12:35 aida09
drwxrwxrwx. 2 npg npg 22 Apr  6 12:35 aida10
drwxrwxrwx. 2 npg npg 22 Apr  6 12:35 aida11
drwxrwxrwx. 2 npg npg 22 Apr  6 12:35 aida12
drwxrwxrwx. 2 npg npg 22 Apr  6 12:35 aida13
drwxrwxrwx. 2 npg npg 22 Apr  6 12:35 aida14
drwxrwxrwx. 2 npg npg 22 Apr  6 12:35 aida15
drwxrwxrwx. 2 npg npg 22 Apr  6 12:35 aida16

Options/aida01:
total 4
-rwxrwxrwx. 1 npg npg 1027 Apr  6 12:33 CONTENTS

Options/aida02:
total 4
-rwxrwxr-x. 1 npg npg 1027 Apr  6 12:35 CONTENTS

Options/aida03:
total 4
-rwxrwxr-x. 1 npg npg 1027 Apr  6 12:35 CONTENTS

Options/aida04:
total 4
-rwxrwxr-x. 1 npg npg 1027 Apr  6 12:35 CONTENTS

Options/aida05:
total 4
-rwxrwxr-x. 1 npg npg 1027 Apr  6 12:35 CONTENTS

Options/aida06:
total 4
-rwxrwxr-x. 1 npg npg 1027 Apr  6 12:35 CONTENTS

Options/aida07:
total 4
-rwxrwxr-x. 1 npg npg 1027 Apr  6 12:35 CONTENTS

Options/aida08:
total 4
-rwxrwxr-x. 1 npg npg 1027 Apr  6 12:35 CONTENTS

Options/aida09:
total 4
-rwxrwxr-x. 1 npg npg 1027 Apr  6 12:35 CONTENTS

Options/aida10:
total 4
-rwxrwxr-x. 1 npg npg 1027 Apr  6 12:35 CONTENTS

Options/aida11:
total 4
-rwxrwxr-x. 1 npg npg 1027 Apr  6 12:35 CONTENTS

Options/aida12:
total 4
-rwxrwxr-x. 1 npg npg 1027 Apr  6 12:35 CONTENTS

Options/aida13:
total 4
-rwxrwxr-x. 1 npg npg 1027 Apr  6 12:35 CONTENTS

Options/aida14:
total 4
-rwxrwxr-x. 1 npg npg 1027 Apr  6 12:35 CONTENTS

Options/aida15:
total 4
-rwxrwxr-x. 1 npg npg 1027 Apr  6 12:35 CONTENTS

Options/aida16:
total 4
-rwxrwxr-x. 1 npg npg 1027 Apr  6 12:35 CONTENTS
[npg@carme-gsi AIDA]$ ls
2017Nov15-15.21.56  2019Dec19-16.19.51  2021Apr29-13-16-00  2024Feb08-15.59.09  CONTENTS        Options.311019         Options.BAK-060425  Options.PRISTINE         README
2018Aug21-13.17.09  2019Mar31-16.15.57  2022Jan31-11-52-00  2024Jan20-12.02.08  Histograms      Options.BACKUP180920   Options.BAK-130421  Options.PRISTINE_TRIPLE
2018Jun22-11.28.18  2019Oct31-13.24.23  2024Apr24-14.14.03  aida_check_db.sh    Options         Options.BACKUP190425   Options.Current     Options.WORKING
2018Sep06-14.38.28  2020Apr15-20.29.26  2024Feb05-10.22.28  aida_check_db.sh~   Options.220131  Options.BACKUPCorrupt  Options.Dec19       Options.WORKING.Old
[npg@carme-gsi AIDA]$ ls -lR Options.BAK-060425
Options.BAK-060425:
total 0
drwxrwxr-x. 2 npg npg 22 Jan 31  2022 aida01
drwxrwxr-x. 2 npg npg 22 Jan 31  2022 aida02
drwxrwxr-x. 2 npg npg 22 Jan 31  2022 aida03
drwxrwxr-x. 2 npg npg 22 Jan 31  2022 aida04
drwxrwxr-x. 2 npg npg 22 Jan 31  2022 aida05
drwxrwxr-x. 2 npg npg 22 Jan 31  2022 aida06
drwxrwxr-x. 2 npg npg 22 Jan 31  2022 aida07
drwxrwxr-x. 2 npg npg 22 Jan 31  2022 aida08
drwxrwxr-x. 2 npg npg 22 Jan 18  2024 aida09
drwxrwxr-x. 2 npg npg 22 Jan 18  2024 aida10
drwxrwxr-x. 2 npg npg 22 Jan 18  2024 aida11
drwxrwxr-x. 2 npg npg 22 Jan 18  2024 aida12
drwxrwxr-x. 2 npg npg 22 Jan 18  2024 aida13
drwxrwxr-x. 2 npg npg 22 Jan 18  2024 aida14
drwxrwxr-x. 2 npg npg 22 Jan 18  2024 aida15
drwxrwxr-x. 2 npg npg 22 Jan 18  2024 aida16

Options.BAK-060425/aida01:
total 4
-rwxrwxr-x. 1 npg npg 1027 Apr  5 18:05 CONTENTS

Options.BAK-060425/aida02:
total 4
-rwxrwxr-x. 1 npg npg 1026 Feb 19  2022 CONTENTS

Options.BAK-060425/aida03:
total 4
-rwxrwxr-x. 1 npg npg 1026 Jan 20  2024 CONTENTS

Options.BAK-060425/aida04:
total 4
-rwxrwxr-x. 1 npg npg 1026 Dec  1  2022 CONTENTS

Options.BAK-060425/aida05:
total 4
-rwxrwxr-x. 1 npg npg 1027 Nov 30  2022 CONTENTS

Options.BAK-060425/aida06:
total 4
-rwxrwxr-x. 1 npg npg 1026 Jan 20  2024 CONTENTS

Options.BAK-060425/aida07:
total 4
-rwxrwxr-x. 1 npg npg 1026 Jan 31  2022 CONTENTS

Options.BAK-060425/aida08:
total 4
-rwxrwxr-x. 1 npg npg 1026 Feb  4  2022 CONTENTS

Options.BAK-060425/aida09:
total 4
-rwxrwxr-x. 1 npg npg 1026 Jan 18  2024 CONTENTS

Options.BAK-060425/aida10:
total 4
-rwxrwxr-x. 1 npg npg 1026 Feb  1  2024 CONTENTS

Options.BAK-060425/aida11:
total 4
-rwxrwxr-x. 1 npg npg 1026 Feb  3  2024 CONTENTS

Options.BAK-060425/aida12:
total 4
-rwxrwxr-x. 1 npg npg 1026 Jan 18  2024 CONTENTS

Options.BAK-060425/aida13:
total 4
-rwxrwxr-x. 1 npg npg 1026 Jan 18  2024 CONTENTS

Options.BAK-060425/aida14:
total 4
-rwxrwxr-x. 1 npg npg 1026 Feb  8  2024 CONTENTS

Options.BAK-060425/aida15:
total 4
-rwxrwxr-x. 1 npg npg 1027 Feb  3  2024 CONTENTS

Options.BAK-060425/aida16:
total 4
-rwxrwxr-x. 1 npg npg 1027 Jun 19  2024 CONTENTS

[npg@carme-gsi AIDA]$ cat Options/aida01/CONTENTS
Index string Stat.offset&&Aida_Hist_W_Enable&&TS_SYNC_PHASE&&ExtClk&&Aida.shift&&MACB_TRIG_MODE&&Aida.offset&&RunNumber&&Aida_GroupBase&&Rate.channels&&Stat.channels&&Aida.Vchannels&&ASIC.settings&&Aida_Hist_D_Enable&&WAVE_DMA_HWM&&Stat.shift&&Aida.Wchannels&&Aida_Hist_V_Enable&&Include.Aida&&Aida.channels&&DataAcqPgm&&Aida_Hist_H_Enable&&DataFormat&&ASIC_DMA_HWM&&Aida_Hist_L_Enable
Stat.offset string 64
Aida_Hist_W_Enable string 1
Aida.shift string 0
TS_SYNC_PHASE string 0x0
ExtClk string 1
Aida.offset string 0
MACB_TRIG_MODE string 4
RunNumber string 133
Rate.channels string 512
Aida_GroupBase string 1
Stat.channels string 512
Aida_Hist_D_Enable string 1
ASIC.settings string 2024Apr24-14.14.03
Aida.Vchannels string 256
WAVE_DMA_HWM string 0x0007ffff
Aida.Wchannels string 1020
Stat.shift string 6
Include.Aida string 0
Aida_Hist_V_Enable string 0
Aida.channels string 65536
DataFormat string 0x0000
DataAcqPgm string AidaExecV9
Aida_Hist_H_Enable string 1
ASIC_DMA_HWM string 0x000fffff
Aida_Hist_L_Enable string 1
Attachment 1: Screenshot_from_2025-04-06_12-59-39.png
Screenshot_from_2025-04-06_12-59-39.png
  664   Sun Apr 6 10:17:49 2025 TDSunday 6 April
11.15 CARME IE514 ion gauge remains OFF
      ML says there are 2x ion gauges with line of sight to the DSSSDs - CARME IE514 (now off) and another controlled by the vacuum group - he will request that this is 
turned off Monday
      ML says that ion gauges within the upstream e-target should not have line of sight to the DSSSDs
 
      FEE64 power ON

      ASIC settings 
       appear to be using mixture of 2022Jan31-11-52-00, 2024Jan20-12.02.08, 2024Feb08-15.59.09 and 2024Apr24-14.14.03 settings

[npg@carme-gsi Options]$ pwd
/home/npg/MIDAS_Releases/23Jan19/MIDAS_200119/DB/EXPERIMENTS/AIDA/Options
[npg@carme-gsi Options]$ foreach f ( `ls` )
foreach? echo $f
foreach? cat $f/CONTENTS | grep settings
foreach? end
aida01
ASIC.settings string 2024Apr24-14.14.03
aida02
ASIC.settings string 2022Jan31-11-52-00
aida03
ASIC.settings string 2024Jan20-12.02.08
aida04
ASIC.settings string 2022Jan31-11-52-00
aida05
ASIC.settings string 2022Jan31-11-52-00
aida06
ASIC.settings string 2022Jan31-11-52-00
aida07
ASIC.settings string 2022Jan31-11-52-00
aida08
ASIC.settings string 2022Jan31-11-52-00
aida09
ASIC.settings string 2022Jan31-11-52-00
aida10
ASIC.settings string 2022Jan31-11-52-00
aida11
ASIC.settings string 2022Jan31-11-52-00
aida12
ASIC.settings string 2022Jan31-11-52-00
aida13
ASIC.settings string 2022Jan31-11-52-00
aida14
ASIC.settings string 2024Feb08-15.59.09
aida15
ASIC.settings string 2022Jan31-11-52-00
aida16
ASIC.settings string 2024Apr24-14.14.03

      slow comparator 0x64 appears to be set for all FEE64s
      positive input (p+n junction strips) FEE64s aida01, 02, 05, 06, 09, 10, 11, 12, 13, 15 & 16
      negative input (n+n Ohmic strips) FEE64s aida03, 04, 07, 08 & 14

      Check ASIC Load all FEE64s

      All system wide checks OK

      FEE64 temps OK - attachment 1

      DSSSD bias & leakage current - attachment 2
       leakage currents remain high cf. June 2024

      per FEE64 Rate spectra - attachment 3
       high rates observed for DSSSD#1 and DSSSD#3 n+n Ohmic strips
       2x hot p+n junction channels aida15 

      per FEE64 1.8.W spectra - 20us FSR - attachments 4-5

      ADC data item stats - attachment 6
       high rates observed for DSSSD#1 and DSSSD#3 n+n Ohmic strips

      CRYRING vacuum and temperature logs - attachments 7-8

12.51 Options DB re-constructed - see https://elog.ph.ed.ac.uk/CARME/665

      ASIC settings for all FEE64s restored from 2024Apr24-14.14.03
       slow comparator 0x64 appears to be set for all FEE64s
       positive input (p+n junction strips) FEE64s aida01, 02, 05, 06, 11, 12, 15 & 16
       negative input (n+n Ohmic strips) FEE64s aida03, 04, 07, 08, 09, 10, 13 & 14

      WR timestamps OK - attachment 9


14.25 EG&G Ortec 448 Pulser settings

      Amplitude 90,000
      Normalise 5.00
      Attenuator x5 IN
      Frequency Relay Line (50Hz)
      t_r 100ns
      tau_d 50us
      Polarity +

      per FEE64 Rate Spectra - attachments 10-11

      per FEE64 1.8.L spectra - attachments 12-14
       Pulser peak width aida06 87 ch FWHM, aida08 94 ch FWHM
       electronic noise poor consistent with high leakage current and 1.8.W spectra above

      ADC data item stats - attachment 15

15.50 Data transfer to Merger/TapeServer started
       Tape Server -> no storage mode
       Data directory APR25

16.30 DAQ STOP
      FEE64 power OFF
      DSSSD bias OFF
Attachment 1: Screenshot_from_2025-04-06_11-25-51.png
Screenshot_from_2025-04-06_11-25-51.png
Attachment 2: Screenshot_from_2025-04-06_11-28-58.png
Screenshot_from_2025-04-06_11-28-58.png
Attachment 3: Screenshot_from_2025-04-06_11-36-36.png
Screenshot_from_2025-04-06_11-36-36.png
Attachment 4: Screenshot_from_2025-04-06_11-37-19.png
Screenshot_from_2025-04-06_11-37-19.png
Attachment 5: Screenshot_from_2025-04-06_11-37-56.png
Screenshot_from_2025-04-06_11-37-56.png
Attachment 6: Screenshot_from_2025-04-06_11-35-19.png
Screenshot_from_2025-04-06_11-35-19.png
Attachment 7: Capture.PNG
Capture.PNG
Attachment 8: Capture.PNG
Capture.PNG
Attachment 9: Screenshot_from_2025-04-06_12-57-44.png
Screenshot_from_2025-04-06_12-57-44.png
Attachment 10: Screenshot_from_2025-04-06_15-04-47.png
Screenshot_from_2025-04-06_15-04-47.png
Attachment 11: Screenshot_from_2025-04-06_15-09-13.png
Screenshot_from_2025-04-06_15-09-13.png
Attachment 12: Screenshot_from_2025-04-06_15-05-41.png
Screenshot_from_2025-04-06_15-05-41.png
Attachment 13: Screenshot_from_2025-04-06_15-07-03.png
Screenshot_from_2025-04-06_15-07-03.png
Attachment 14: Screenshot_from_2025-04-06_15-08-25.png
Screenshot_from_2025-04-06_15-08-25.png
Attachment 15: Screenshot_from_2025-04-06_15-39-45.png
Screenshot_from_2025-04-06_15-39-45.png
  663   Sat Apr 5 16:15:04 2025 TDFEE64 power interlock installed
Water flow and dew point sensors have been installed on the outlet and inlet cooling water manifolds respectively.

Outputs indicating good water flow and ambient temperature above the dew point are connected via an interlock box (constructed by Peter Black)
to the USB-controlled ac mains relay which provides remote control of ac mains power to the AIDA FEE64 PSUs.

The interlock box (see attachment 1) has 

1x 24V dc power input (ac mains transformer)
4x sensor inputs (1x dew point, 1x flow and 2x unused, dummy cable inputs)
4x outputs (1x connected to USB-controlled ac mains relay, 3x unused)

When all sensor inputs are OK, 4x yellow LEDs (corresponding to each of the 4x sensor inputs) are illuminated and visible through the transparent
lid of the interlock box - see attachment 2
The USB-controlled ac mains relay yellow 'water' LED is illuminated if interlock is OK - see attachment 3


B+B Sensor Dew Point Controller - attachments 4 & 5
 red led (top right) indicates sensor above dew point

Micronics U1000 Mk II Clamp On Ultrasonic Flow Meter - attachments 6-9

 Setup as follows
  Pipe ID: 44mm
  Reading: Flow
  Units: litres
  Flow Units: l/min
  Pipe Material: M/Steel
  Set Separation: C 3

 All other settings default

 Not possible to attach sensor assembly to electronics module for some reason but cabling is attached and unit appears to
 work OK. Typical flow c. 75 l/min with signal strength c. 75% - see video attachment 10

Sunday 6 April
 FEE64 power OFF
 Valve OFF outflow
  Flow rate reported by Micronics U1000 Mk II decreases fro c. 75 l/min to 0 l/min over a period of c. 10-15s
  USB-controlled ac mains relay - yellow 'water' LED off
 Valve ON outflow
  Flow rate recovers c. 10-15s

Monday 7 April
 FEE64 power ON
 Valve OFF outflow
  Flow rate reported by Micronics U1000 Mk II decreases fro c. 75 l/min to 0 l/min over a period of c. 10-15s
  AIDA PSUs OFF as flow -> 0
 Valve ON outflow
  USB-controlled ac mains relay - yellow 'water' LED ON
  AIDA PSUs remain OFF
  Rly16 browser tab continues to display 'power ON' and OK status - attachment 11
   but unable switch ALL off
  To recover - clsoe Rly16 browser tab, reboot nnrpi1 and restart MIDAS server


 Valve ON outflow

  
Attachment 1: 20250405_132338.jpg
20250405_132338.jpg
Attachment 2: 20250405_165907.jpg
20250405_165907.jpg
Attachment 3: 20250405_165916.jpg
20250405_165916.jpg
Attachment 4: 20250405_165850.jpg
20250405_165850.jpg
Attachment 5: Datasheet_dew-point-sensor_0557_0015x.pdf
Datasheet_dew-point-sensor_0557_0015x.pdf Datasheet_dew-point-sensor_0557_0015x.pdf Datasheet_dew-point-sensor_0557_0015x.pdf
Attachment 6: U1000MKII_Issue-3.1.pdf
U1000MKII_Issue-3.1.pdf U1000MKII_Issue-3.1.pdf U1000MKII_Issue-3.1.pdf U1000MKII_Issue-3.1.pdf U1000MKII_Issue-3.1.pdf U1000MKII_Issue-3.1.pdf U1000MKII_Issue-3.1.pdf U1000MKII_Issue-3.1.pdf
Attachment 7: 20250405_165827.jpg
20250405_165827.jpg
Attachment 8: 20250405_165831.jpg
20250405_165831.jpg
Attachment 9: 20250405_165838.jpg
20250405_165838.jpg
Attachment 10: 20250405_165753.mp4
Attachment 11: Screenshot_from_2025-04-07_11-54-48.png
Screenshot_from_2025-04-07_11-54-48.png
  662   Sat Apr 5 16:13:32 2025 TDSaturday 5 April
12.00 CARME IE514 ion gauge 4.2e-11 mbar
      CARME IE514 OFF


      aida08 red coolant line and HDMI cable found disconnected - blue coolant line, network cable and FEE^$ power are still connected
      Now re-connected


13.57 DSSSD bias & leakage current - attachment 1
       leakage currents high (especially DSSSD#1 and DSSSD#3) cf. most recent values from June 2024 values https://elog.ph.ed.ac.uk/CARME/654
       another ion gauge on with line of sight to DSSSDs?

      FEE64 power ON

      FEE64 temps OK - attachment 2

      All system wide checks OK - attachments 3-6

      WR timestamps OK - attachment 7


18.00 DSSSD bias OFF
      FEE64 power OFF
Attachment 1: Screenshot_from_2025-04-05_13-57-06.png
Screenshot_from_2025-04-05_13-57-06.png
Attachment 2: Screenshot_from_2025-04-05_16-00-19.png
Screenshot_from_2025-04-05_16-00-19.png
Attachment 3: Screenshot_from_2025-04-05_16-01-31.png
Screenshot_from_2025-04-05_16-01-31.png
Attachment 4: Screenshot_from_2025-04-05_16-04-22.png
Screenshot_from_2025-04-05_16-04-22.png
Attachment 5: Screenshot_from_2025-04-05_16-04-36.png
Screenshot_from_2025-04-05_16-04-36.png
Attachment 6: Screenshot_from_2025-04-05_16-04-47.png
Screenshot_from_2025-04-05_16-04-47.png
Attachment 7: Screenshot_from_2025-04-05_16-06-41.png
Screenshot_from_2025-04-05_16-06-41.png
  661   Fri Apr 4 10:17:16 2025 TDFriday 4 April
10.43 Reboot carme-gsi - unable to open a working terminal

      Post boot found that it was necessary to manually enable WAN network interface (enp2s0) - perhaps it had been disabled previously?

      Started proxy port for remote access via AnyDesk per https://elog.ph.ed.ac.uk/CARME/489 using the command 

         ssh -L 8080:proxy.gsi.de:8080 carme@atppc02

      AnyDesk 978 207 281 now online/available

13.43 carme-gsi moved from 19" rack room to control/counting room (next door)

      19" rack room
       *yellow* network cable disconnected from YR10/2 patch panel, *green* network cable connected to YR10/2 patch panel
      control counting room
       *green* network cable connected to carme-gsi p4p1 network interface (connection to CARME network switch at YR09)
       network cable connected from carme-gsi enp2s0 network interface (port #2) to Netgear MS510TXM EXPSW3 port (connection to WAN)
      
  660   Mon Mar 10 15:47:27 2025 TDAnalysis data files G22-201 R79_10 & R79_24
Analysis of data files G22-201 R79_10 and R79_24 - attachments 1-2

Observe HIGH deadtime for FEE64 aida11 (attachment 1) and aida16 (attachment 2) associated with *very* high pause/resume/HEC/disc data item rates.

R79_10

deadtime aida11 99.4%
pause/resume ~ 720Hz
HEC ~ 2800Hz
disc ~ 250kHz


R79_24

deadtime aida16 18.7%
pause/resume ~ 140Hz
HEC ~ 540Hz
disc ~ 50kHz
Attachment 1: R79_10.txt
 *** TDR format 3.3.0 analyser - TD - May 2021
 *** ERROR: READ I/O error:       5002
                   blocks:      32000
          ADC data format:   28979939 (   32250.3 Hz)
        Other data format:  232940065 (  259227.5 Hz)
 Sample trace data format:          0 (       0.0 Hz)
         Undefined format:          0 (       0.0 Hz)
   Other data format type:      PAUSE:     647219 (     720.3 Hz)
                               RESUME:     647218 (     720.3 Hz)
                              SYNC100:      35348 (      39.3 Hz)
                              WR48-63:      35348 (      39.3 Hz)
                           FEE64 disc:  231574932 (  257708.4 Hz)
                             MBS info:          0 (       0.0 Hz)
                           Other info:          0 (       0.0 Hz)

   ADC data range bit set:    2508895 (    2792.0 Hz)

                Timewarps:        ADC:          0 (       0.0 Hz)
                                PAUSE:          0 (       0.0 Hz)
                               RESUME:          0 (       0.0 Hz)
                              SYNC100:          0 (       0.0 Hz)
                              WR48-63:          0 (       0.0 Hz)
                           FEE64 disc:          0 (       0.0 Hz)
                             MBS info:          0 (       0.0 Hz)
                            Undefined:          0 (       0.0 Hz)
                         Sample trace:          0 (       0.0 Hz)

 *** Timestamp elapsed time:      898.593 s
 FEE  elapsed dead time(s) elapsed idle time(s)
  0                0.000                0.000
  1                0.000                0.000
  2                0.000                0.000
  3                0.000                0.000
  4                0.000                0.000
  5                0.000                0.000
  6                0.000                0.000
  7                0.000                0.000
  8                0.000                0.000
  9                0.000                0.000
 10              892.873                0.000
 11                0.000                0.000
 12                0.000                0.000
 13                0.000                0.000
 14                0.137                0.000
 15                0.000                0.000
 16                0.000                0.000
 17                0.000                0.000
 18                0.000                0.000
 19                0.000                0.000
 20                0.000                0.000
 21                0.000                0.000
 22                0.000                0.000
 23                0.000                0.000
 24                0.000                0.000
 25                0.000                0.000
 26                0.000                0.000
 27                0.000                0.000
 28                0.000                0.000
 29                0.000                0.000
 30                0.000                0.000
 31                0.000                0.000
 32                0.000                0.000

 *** Statistics
 FEE  ADC Data Other Data     Sample  Undefined      Pause     Resume    SYNC100    WR48-63       Disc        MBS      Other   HEC Data
  0    1104087        558          0          0          0          0        228        228        102          0          0         39
  1       1368         19          0          0          0          0          0          0         19          0          0         11
  2        703          0          0          0          0          0          0          0          0          0          0          0
  3    1210701       1170          0          0          0          0        258        258        654          0          0         86
  4     103506         34          0          0          0          0         17         17          0          0          0          0
  5       3399          2          0          0          0          0          1          1          0          0          0          0
  6       5236        395          0          0          0          0          1          1        393          0          0        218
  7      52921         16          0          0          0          0          8          8          0          0          0          0
  8      43890         53          0          0          0          0         13         13         27          0          0         22
  9       6906       1900          0          0          0          0          2          2       1896          0          0        193
 10    2506650  226682792          0          0     647212     647211      28069      28069  225332231          0          0    2506358
 11      48924         27          0          0          0          0         11         11          5          0          0          0
 12     189100        607          0          0          0          0         40         40        527          0          0         88
 13     163612         80          0          0          0          0         26         26         28          0          0         15
 14   19931577    6246424          0          0          7          7       5816       5816    6234778          0          0        135
 15    3607359       5988          0          0          0          0        858        858       4272          0          0       1730
 16          0          0          0          0          0          0          0          0          0          0          0          0
 17          0          0          0          0          0          0          0          0          0          0          0          0
 18          0          0          0          0          0          0          0          0          0          0          0          0
 19          0          0          0          0          0          0          0          0          0          0          0          0
 20          0          0          0          0          0          0          0          0          0          0          0          0
 21          0          0          0          0          0          0          0          0          0          0          0          0
 22          0          0          0          0          0          0          0          0          0          0          0          0
 23          0          0          0          0          0          0          0          0          0          0          0          0
 24          0          0          0          0          0          0          0          0          0          0          0          0
 25          0          0          0          0          0          0          0          0          0          0          0          0
 26          0          0          0          0          0          0          0          0          0          0          0          0
 27          0          0          0          0          0          0          0          0          0          0          0          0
 28          0          0          0          0          0          0          0          0          0          0          0          0
 29          0          0          0          0          0          0          0          0          0          0          0          0
 30          0          0          0          0          0          0          0          0          0          0          0          0
 31          0          0          0          0          0          0          0          0          0          0          0          0
 32          0          0          0          0          0          0          0          0          0          0          0          0

 *** Timewarps
 FEE       ADC      Pause     Resume    SYNC100    WR48-63       Disc        MBS  Undefined    Samples
  0          0          0          0          0          0          0          0          0          0
  1          0          0          0          0          0          0          0          0          0
  2          0          0          0          0          0          0          0          0          0
  3          0          0          0          0          0          0          0          0          0
  4          0          0          0          0          0          0          0          0          0
  5          0          0          0          0          0          0          0          0          0
  6          0          0          0          0          0          0          0          0          0
  7          0          0          0          0          0          0          0          0          0
  8          0          0          0          0          0          0          0          0          0
  9          0          0          0          0          0          0          0          0          0
 10          0          0          0          0          0          0          0          0          0
 11          0          0          0          0          0          0          0          0          0
 12          0          0          0          0          0          0          0          0          0
 13          0          0          0          0          0          0          0          0          0
 14          0          0          0          0          0          0          0          0          0
 15          0          0          0          0          0          0          0          0          0
 16          0          0          0          0          0          0          0          0          0
 17          0          0          0          0          0          0          0          0          0
 18          0          0          0          0          0          0          0          0          0
 19          0          0          0          0          0          0          0          0          0
 20          0          0          0          0          0          0          0          0          0
 21          0          0          0          0          0          0          0          0          0
 22          0          0          0          0          0          0          0          0          0
 23          0          0          0          0          0          0          0          0          0
 24          0          0          0          0          0          0          0          0          0
 25          0          0          0          0          0          0          0          0          0
 26          0          0          0          0          0          0          0          0          0
 27          0          0          0          0          0          0          0          0          0
 28          0          0          0          0          0          0          0          0          0
 29          0          0          0          0          0          0          0          0          0
 30          0          0          0          0          0          0          0          0          0
 31          0          0          0          0          0          0          0          0          0
 32          0          0          0          0          0          0          0          0          0

 *** Program elapsed time:   24.215s ( 1321.503 blocks/s,  82.594 Mb/s)
Attachment 2: R79_24.txt
 *** TDR format 3.3.0 analyser - TD - May 2021
 *** ERROR: READ I/O error:       5002
                   blocks:      32000
          ADC data format:   83751406 (   26778.5 Hz)
        Other data format:  178168596 (   56967.3 Hz)
 Sample trace data format:          0 (       0.0 Hz)
         Undefined format:          0 (       0.0 Hz)
   Other data format type:      PAUSE:     448350 (     143.4 Hz)
                               RESUME:     448350 (     143.4 Hz)
                              SYNC100:      43650 (      14.0 Hz)
                              WR48-63:      43650 (      14.0 Hz)
                           FEE64 disc:  177184596 (   56652.6 Hz)
                             MBS info:          0 (       0.0 Hz)
                           Other info:          0 (       0.0 Hz)

   ADC data range bit set:    1682572 (     538.0 Hz)

                Timewarps:        ADC:          0 (       0.0 Hz)
                                PAUSE:          0 (       0.0 Hz)
                               RESUME:          0 (       0.0 Hz)
                              SYNC100:          0 (       0.0 Hz)
                              WR48-63:          0 (       0.0 Hz)
                           FEE64 disc:          0 (       0.0 Hz)
                             MBS info:          0 (       0.0 Hz)
                            Undefined:          0 (       0.0 Hz)
                         Sample trace:          0 (       0.0 Hz)

 *** Timestamp elapsed time:     3127.561 s
 FEE  elapsed dead time(s) elapsed idle time(s)
  0                0.000                0.000
  1                0.000                0.000
  2                0.000                0.000
  3                0.000                0.000
  4                0.000                0.000
  5                0.000                0.000
  6                0.000                0.000
  7                0.000                0.000
  8                0.000                0.000
  9                0.000                0.000
 10                0.000                0.000
 11                0.000                0.000
 12                0.000                0.000
 13                0.000                0.000
 14                0.620                0.000
 15              586.109                0.000
 16                0.000                0.000
 17                0.000                0.000
 18                0.000                0.000
 19                0.000                0.000
 20                0.000                0.000
 21                0.000                0.000
 22                0.000                0.000
 23                0.000                0.000
 24                0.000                0.000
 25                0.000                0.000
 26                0.000                0.000
 27                0.000                0.000
 28                0.000                0.000
 29                0.000                0.000
 30                0.000                0.000
 31                0.000                0.000
 32                0.000                0.000

 *** Statistics
 FEE  ADC Data Other Data     Sample  Undefined      Pause     Resume    SYNC100    WR48-63       Disc        MBS      Other   HEC Data
  0    3119762       1529          0          0          0          0        757        757         15          0          0          7
  1       4282          4          0          0          0          0          2          2          0          0          0          0
  2       2544          2          0          0          0          0          1          1          0          0          0          0
  3    3390513       2086          0          0          0          0        934        934        218          0          0         22
  4     360107        196          0          0          0          0         98         98          0          0          0          0
  5      10127          8          0          0          0          0          4          4          0          0          0          0
  6      12521        270          0          0          0          0          5          5        260          0          0        193
  7     202583         64          0          0          0          0         32         32          0          0          0          0
  8     121120         81          0          0          0          0         38         38          5          0          0          2
  9      19612       4370          0          0          0          0          7          7       4356          0          0        411
 10       4704          6          0          0          0          0          3          3          0          0          0          0
 11     134007         68          0          0          0          0         34         34          0          0          0          0
 12     605075       1275          0          0          0          0        139        139        997          0          0        159
 13    1179710        370          0          0          0          0        170        170         30          0          0          7
 14   58524563   21407125          0          0         41         41      18434      18434   21370175          0          0         71
 15   16060176  156751142          0          0     448309     448309      22992      22992  155808540          0          0    1681700
 16          0          0          0          0          0          0          0          0          0          0          0          0
 17          0          0          0          0          0          0          0          0          0          0          0          0
 18          0          0          0          0          0          0          0          0          0          0          0          0
 19          0          0          0          0          0          0          0          0          0          0          0          0
 20          0          0          0          0          0          0          0          0          0          0          0          0
 21          0          0          0          0          0          0          0          0          0          0          0          0
 22          0          0          0          0          0          0          0          0          0          0          0          0
 23          0          0          0          0          0          0          0          0          0          0          0          0
 24          0          0          0          0          0          0          0          0          0          0          0          0
 25          0          0          0          0          0          0          0          0          0          0          0          0
 26          0          0          0          0          0          0          0          0          0          0          0          0
 27          0          0          0          0          0          0          0          0          0          0          0          0
 28          0          0          0          0          0          0          0          0          0          0          0          0
 29          0          0          0          0          0          0          0          0          0          0          0          0
 30          0          0          0          0          0          0          0          0          0          0          0          0
 31          0          0          0          0          0          0          0          0          0          0          0          0
 32          0          0          0          0          0          0          0          0          0          0          0          0

 *** Timewarps
 FEE       ADC      Pause     Resume    SYNC100    WR48-63       Disc        MBS  Undefined    Samples
  0          0          0          0          0          0          0          0          0          0
  1          0          0          0          0          0          0          0          0          0
  2          0          0          0          0          0          0          0          0          0
  3          0          0          0          0          0          0          0          0          0
  4          0          0          0          0          0          0          0          0          0
  5          0          0          0          0          0          0          0          0          0
  6          0          0          0          0          0          0          0          0          0
  7          0          0          0          0          0          0          0          0          0
  8          0          0          0          0          0          0          0          0          0
  9          0          0          0          0          0          0          0          0          0
 10          0          0          0          0          0          0          0          0          0
 11          0          0          0          0          0          0          0          0          0
 12          0          0          0          0          0          0          0          0          0
 13          0          0          0          0          0          0          0          0          0
 14          0          0          0          0          0          0          0          0          0
 15          0          0          0          0          0          0          0          0          0
 16          0          0          0          0          0          0          0          0          0
 17          0          0          0          0          0          0          0          0          0
 18          0          0          0          0          0          0          0          0          0
 19          0          0          0          0          0          0          0          0          0
 20          0          0          0          0          0          0          0          0          0
 21          0          0          0          0          0          0          0          0          0
 22          0          0          0          0          0          0          0          0          0
 23          0          0          0          0          0          0          0          0          0
 24          0          0          0          0          0          0          0          0          0
 25          0          0          0          0          0          0          0          0          0
 26          0          0          0          0          0          0          0          0          0
 27          0          0          0          0          0          0          0          0          0
 28          0          0          0          0          0          0          0          0          0
 29          0          0          0          0          0          0          0          0          0
 30          0          0          0          0          0          0          0          0          0
 31          0          0          0          0          0          0          0          0          0
 32          0          0          0          0          0          0          0          0          0

 *** Program elapsed time:   23.742s ( 1347.812 blocks/s,  84.238 Mb/s)
  659   Tue Jul 2 17:43:32 2024 TDOffline analysis G22-201/R89
Offline analysis data file G22-201/R89_0 to R89_13 (c. 6h40m data)

DSSSD#1 & 3 36.54mm -> ~ 2.3-8.6 deg 
DSSSD#2 & 4 64.60mm -> ~ 4.1-10.4 deg 


Beam ~1.06MeV/u 15N5+


Sort program variables https://elog.ph.ed.ac.uk/CARME/240530_180659/variables.dat

ADC data corrected for ADC offset but not gain (assume 0.7keV/LSB)

per DSSSD m_p verus m_n - attachments 1-2

For spectra below events must satisfy condition 0 < m_p < 4 *and* 0 < m_n < 4
 gated w(1)=50 & w(2)=-50
 20keV/channel nominal

per DSSSD p+n junction strip versus n+n Ohmic strip - attachment 3
 note - x/y channel ordering to be checked

per DSSSD E_p versus E_n - attachments 4 & 6
 check whether vertical structures (correct E_n but low E_p) are associated with specific strips or edge of DSSSD?

per DSSSD E_p versus E_n - attachment 5 & 7
 expanded view high energy events 

DSSSD #1 E_p + DSSSD#4  E_p - attachment 8
DSSSD #2 E_p + DSSSD#3  E_p - attachment 8

DSSSD #1 E_p versus DSSSD#4  E_p - attachment 9
DSSSD #2 E_p versus DSSSD#3  E_p - attachment 9
 note clusters of events at asymmetric energies but only for one combination of detectors - random events?


______________________________________________________________________________________________________________________________

               Reaction: 1H(15N,4He)12C

                      A:           1         15          4         12
                      Z:           1          7          2          6
              Mass(amu):    1.007825  15.000109   4.002603  12.000000
             Error(keV):       0.000      0.001      0.001      0.000
                 Origin:
                 Status:

                  Q(gs):    4.9655 +/-   0.0001 MeV

  Projectile lab energy:   15.900 MeV
              CM energy:    1.001 MeV
      Excitation energy:    0.000 MeV
       Reaction Q-value:    4.965 MeV
     Reaction threshold:    0.000 MeV

                 Particle 3                               Particle 4                           Non-relativistic Jacobians
           ( scattered/ejectile )                         ( recoil )                                           Ejectile Ejectile
                                                                                               Ejectile Recoil dOmega/  dOmega/
  Lab    CM       Lab   Lab     Lab    dOmega      Lab    CM       Lab   Lab     Lab    dOmega  dOmega  dOmega Recoil   Recoil
 Angle  Angle   Energy  Beta    ToF    Lab/CM     Angle  Angle   Energy  Beta    ToF    Lab/CM  Lab/CM  Lab/CM dOmega   dOmega
 (deg)  (deg)    (MeV)        (ns/cm)  ratio      (deg)  (deg)    (MeV)        (ns/cm)  ratio                  Lab/Lab  CM/Lab

  0.00 NaN      16.372 0.0934   0.357 NaN          0.00 NaN       4.494 0.0283   1.177 NaN     NaN     NaN     NaN     NaN
  1.00   1.91   16.367 0.0934   0.357  0.2729      1.10 178.09    4.498 0.0284   1.176 -0.3314  0.2729 -0.3314 -0.8233 -3.0175
  2.00   3.83   16.354 0.0934   0.357  0.2730      2.20 176.17    4.512 0.0284   1.174 -0.3290  0.2730 -0.3290 -0.8296 -3.0393
  3.00   5.74   16.331 0.0933   0.358  0.2732      3.29 174.26    4.535 0.0285   1.171 -0.3251  0.2732 -0.3251 -0.8403 -3.0759
  4.00   7.66   16.299 0.0932   0.358  0.2735      4.37 172.34    4.566 0.0286   1.167 -0.3197  0.2735 -0.3197 -0.8553 -3.1277
  5.00   9.57   16.258 0.0931   0.358  0.2739      5.43 170.43    4.607 0.0287   1.162 -0.3129  0.2738 -0.3130 -0.8750 -3.1954
  6.00  11.49   16.208 0.0929   0.359  0.2743      6.47 168.51    4.657 0.0289   1.156 -0.3049  0.2743 -0.3049 -0.8996 -3.2796
  7.00  13.40   16.150 0.0928   0.360  0.2749      7.49 166.60    4.716 0.0290   1.149 -0.2957  0.2749 -0.2957 -0.9294 -3.3815
  8.00  15.31   16.082 0.0926   0.360  0.2755      8.48 164.69    4.784 0.0292   1.141 -0.2855  0.2755 -0.2855 -0.9648 -3.5023
  9.00  17.23   16.006 0.0924   0.361  0.2762      9.45 162.77    4.860 0.0295   1.132 -0.2745  0.2762 -0.2745 -1.0063 -3.6434
 10.00  19.14   15.921 0.0921   0.362  0.2770     10.38 160.86    4.945 0.0297   1.122 -0.2627  0.2770 -0.2627 -1.0545 -3.8067
 11.00  21.05   15.827 0.0918   0.363  0.2779     11.27 158.95    5.039 0.0300   1.111 -0.2503  0.2779 -0.2503 -1.1100 -3.9944
 12.00  22.96   15.725 0.0916   0.364  0.2789     12.13 157.04    5.141 0.0303   1.100 -0.2376  0.2789 -0.2376 -1.1738 -4.2091
 13.00  24.87   15.614 0.0912   0.366  0.2800     12.96 155.13    5.251 0.0306   1.089 -0.2245  0.2800 -0.2245 -1.2469 -4.4539
 14.00  26.78   15.496 0.0909   0.367  0.2812     13.74 153.22    5.370 0.0310   1.077 -0.2113  0.2811 -0.2113 -1.3305 -4.7326
 15.00  28.69   15.369 0.0905   0.369  0.2824     14.49 151.31    5.496 0.0313   1.064 -0.1980  0.2824 -0.1980 -1.4260 -5.0496

______________________________________________________________________________________________________________________________

 Options available:
    (r)eaction   e(x)citation energy   (e)nergy   (a)ngles   (p)rint   (q)uit

 Enter option: x

 Enter excitation energy (MeV): 4.44

 Enter projectile lab energy (MeV): /

 Enter lab scattering angles [start,stop,step] (deg): /

______________________________________________________________________________________________________________________________

               Reaction: 1H(15N,4He)12C

                      A:           1         15          4         12
                      Z:           1          7          2          6
              Mass(amu):    1.007825  15.000109   4.002603  12.000000
             Error(keV):       0.000      0.001      0.001      0.000
                 Origin:
                 Status:

                  Q(gs):    4.9655 +/-   0.0001 MeV

  Projectile lab energy:   15.900 MeV
              CM energy:    1.001 MeV
      Excitation energy:    4.440 MeV
       Reaction Q-value:    0.525 MeV
     Reaction threshold:    0.000 MeV

                 Particle 3                               Particle 4                           Non-relativistic Jacobians
           ( scattered/ejectile )                         ( recoil )                                           Ejectile Ejectile
                                                                                               Ejectile Recoil dOmega/  dOmega/
  Lab    CM       Lab   Lab     Lab    dOmega      Lab    CM       Lab   Lab     Lab    dOmega  dOmega  dOmega Recoil   Recoil
 Angle  Angle   Energy  Beta    ToF    Lab/CM     Angle  Angle   Energy  Beta    ToF    Lab/CM  Lab/CM  Lab/CM dOmega   dOmega
 (deg)  (deg)    (MeV)        (ns/cm)  ratio      (deg)  (deg)    (MeV)        (ns/cm)  ratio                  Lab/Lab  CM/Lab

  0.00   0.00    9.002 0.0694   0.481  0.0000      0.00 180.00    7.423 0.0364   0.916  0.0000  0.0000  0.0000 NaN     NaN
       180.00    0.740 0.0199   1.675  0.0000      0.00   0.00   15.686 0.0529   0.630  0.0000  0.0000  0.0000 NaN     NaN
  1.00   2.81    8.997 0.0693   0.481  0.1270      0.64 177.19    7.428 0.0364   0.915 -0.0513  0.1270 -0.0513 -2.4762 *******
       179.20    0.740 0.0199   1.674  1.5463      0.13   0.80   15.685 0.0529   0.630  0.0243 -1.5463  0.0243 ******* 41.1048
  2.00   5.61    8.983 0.0693   0.481  0.1271      1.27 174.39    7.443 0.0365   0.915 -0.0509  0.1271 -0.0509 -2.4951 *******
       178.39    0.741 0.0199   1.673  1.5414      0.25   1.61   15.684 0.0529   0.630  0.0243 -1.5414  0.0243 ******* 41.1102
  3.00   8.42    8.958 0.0692   0.482  0.1271      1.90 171.58    7.468 0.0365   0.913 -0.0503  0.1271 -0.0503 -2.5269 *******
       177.58    0.743 0.0200   1.671  1.5334      0.38   2.42   15.682 0.0529   0.630  0.0243 -1.5334  0.0243 ******* 41.1194
  4.00  11.24    8.923 0.0691   0.483  0.1271      2.52 168.76    7.502 0.0366   0.911 -0.0494  0.1271 -0.0494 -2.5722 *******
       176.77    0.746 0.0200   1.668  1.5221      0.50   3.23   15.679 0.0529   0.630  0.0243 -1.5221  0.0243 ******* 41.1323
  5.00  14.05    8.879 0.0689   0.484  0.1272      3.13 165.95    7.547 0.0367   0.908 -0.0483  0.1272 -0.0483 -2.6319 *******
       175.96    0.750 0.0201   1.663  1.5077      0.63   4.04   15.676 0.0529   0.631  0.0243 -1.5077  0.0243 ******* 41.1490
  6.00  16.88    8.824 0.0687   0.486  0.1273      3.73 163.12    7.601 0.0369   0.905 -0.0470  0.1273 -0.0470 -2.7072 *******
       175.13    0.754 0.0201   1.658  1.4901      0.76   4.87   15.671 0.0529   0.631  0.0243 -1.4901  0.0243 ******* 41.1697
  7.00  19.71    8.760 0.0684   0.487  0.1273      4.32 160.29    7.665 0.0370   0.901 -0.0455  0.1273 -0.0455 -2.7995 *******
       174.30    0.760 0.0202   1.652  1.4695      0.89   5.70   15.665 0.0529   0.631  0.0243 -1.4695  0.0243 ******* 41.1945
  8.00  22.55    8.686 0.0681   0.490  0.1274      4.89 157.45    7.739 0.0372   0.897 -0.0438  0.1274 -0.0438 -2.9108 *******
       173.46    0.766 0.0203   1.645  1.4458      1.02   6.54   15.659 0.0529   0.631  0.0243 -1.4458  0.0243 ******* 41.2236
  9.00  25.41    8.603 0.0678   0.492  0.1275      5.44 154.59    7.823 0.0374   0.892 -0.0419  0.1275 -0.0419 -3.0436 *******
       172.61    0.774 0.0204   1.637  1.4191      1.15   7.39   15.652 0.0529   0.631  0.0242 -1.4192  0.0242 ******* 41.2573
 10.00  28.27    8.509 0.0674   0.495  0.1276      5.97 151.73    7.916 0.0376   0.887 -0.0399  0.1276 -0.0399 -3.2011 *******
       171.75    0.782 0.0205   1.628  1.3896      1.28   8.25   15.643 0.0528   0.631  0.0242 -1.3896  0.0242 ******* 41.2958
 11.00  31.15    8.407 0.0670   0.498  0.1277      6.48 148.85    8.019 0.0379   0.881 -0.0377  0.1277 -0.0377 -3.3875 *******
       170.87    0.792 0.0206   1.619  1.3572      1.42   9.13   15.634 0.0528   0.631  0.0242 -1.3572  0.0242 ******* 41.3394
 12.00  34.05    8.294 0.0666   0.501  0.1278      6.97 145.95    8.131 0.0381   0.875 -0.0354  0.1278 -0.0354 -3.6080 *******
       169.97    0.803 0.0207   1.608  1.3221      1.56  10.03   15.623 0.0528   0.632  0.0242 -1.3221  0.0242 ******* 41.3886
 13.00  36.96    8.172 0.0661   0.505  0.1279      7.43 143.04    8.253 0.0384   0.868 -0.0330  0.1279 -0.0330 -3.8695 *******
       169.06    0.815 0.0209   1.596  1.2843      1.70  10.94   15.611 0.0528   0.632  0.0241 -1.2844  0.0241 ******* 41.4437
 14.00  39.90    8.040 0.0656   0.509  0.1280      7.87 140.10    8.385 0.0387   0.862 -0.0306  0.1280 -0.0306 -4.1814 *******
       168.13    0.828 0.0211   1.583  1.2440      1.84  11.87   15.598 0.0528   0.632  0.0241 -1.2441  0.0241 ******* 41.5053
 15.00  42.86    7.899 0.0650   0.513  0.1280      8.28 137.14    8.526 0.0390   0.855 -0.0281  0.1280 -0.0281 -4.5560 *******
       167.17    0.843 0.0213   1.569  1.2012      1.99  12.83   15.583 0.0527   0.632  0.0241 -1.2013  0.0241 ******* 41.5741

______________________________________________________________________________________________________________________________


Assume alphas of coincident events detected at 5 deg lab

1H(15N,4He)12C(gs)    E_alpha + E_12C = 16.528+4.607MeV = 20.865MeV - 0.9MeV 12C energy loss dead layer = 19.965MeV = ch 998


1H(15N,4He)12C(4.44)    forward CM E_alpha + E_12C = 0.750+15.676MeV = 16.426MeV - 0.8MeV 12C energy loss dead layer = 15.626MeV = ch 781
1H(15N,4He)12C(4.44)   backward CM E_alpha + E_12C = 8.879+7.547MeV = 16.426MeV - 0.9MeV 12C energy loss dead layer = 15.526MeV = ch 776
Attachment 1: Screenshot_from_2024-06-06_15-06-55.png
Screenshot_from_2024-06-06_15-06-55.png
Attachment 2: Screenshot_from_2024-06-06_15-07-36.png
Screenshot_from_2024-06-06_15-07-36.png
Attachment 3: Screenshot_from_2024-06-06_15-08-53.png
Screenshot_from_2024-06-06_15-08-53.png
Attachment 4: Screenshot_from_2024-06-06_15-09-42.png
Screenshot_from_2024-06-06_15-09-42.png
Attachment 5: Screenshot_from_2024-06-06_15-11-08.png
Screenshot_from_2024-06-06_15-11-08.png
Attachment 6: Screenshot_from_2024-06-06_15-10-35.png
Screenshot_from_2024-06-06_15-10-35.png
Attachment 7: Screenshot_from_2024-06-06_15-11-53.png
Screenshot_from_2024-06-06_15-11-53.png
Attachment 8: Screenshot_from_2024-06-06_15-13-22.png
Screenshot_from_2024-06-06_15-13-22.png
Attachment 9: Screenshot_from_2024-06-06_15-14-51.png
Screenshot_from_2024-06-06_15-14-51.png
  658   Wed Jun 19 12:11:22 2024 CB, RSSCARME status

DSSD off

FEEs off

Motors out, pin inserted, compressed air closed, power removed from motors box.

Motors and temperature logs are in carme-gsi:logs/G22-200

 

Next experiment will install new setup attached to gas target. Some baking and vacuum operations required. Everything should stay off for safety until they are done.

  657   Wed Jun 19 12:08:34 2024 CB + variousRun 101-105

Run 101: no target. See ELOG entry by TD.

Run 102: ??

Run 103: ??

Run 104: After experiment time is over, Gleb tests beam lifetime at "high energies". Some data acquired for 30 minutes.

Run 105: Quick pulser walkthrough. People are working near CARME. Positive polarity 1000-9000 amplitude.

  656   Wed Jun 19 12:06:50 2024 CB, OF, SSRun 98/99/100 - higher energy

10:21 June 17

New energy: 1.125 MeV/u

Target in. Run 98 started.

 

Run 99 overlap check.

 

Run 100 back to acquisition as run 98.

  655   Wed Jun 19 12:04:02 2024 CB + OF, SSRun 96/97

Run 96 - as run 89, while we wait for beam energy change at higher energies

Run stopped around 08:56 June 17

 

Run 97 - target out, beam optimisation for higher energies.

 

 

  654   Wed Jun 19 12:01:29 2024 TD23:00-07:00 Wednesday 19 June
23.18 DSSSD bias & leakage current OK - attachment 1
      FEE64 temps OK - attachment 2
      ADC data item stats attachment 3
       aida01-aida04 c. 250k
      per FEE64 1.8.W spectra - 20us FSR - attachments 4-5
       large transients aida01-aida04
      per FEE64 Rate spectra - attachment 6
      Merger, TapeServer etc. - attachments 7-8
        data file R101_15
      DSSSD energy spectra (updating) - attachment 9
      DSSSD bias & leakage current log - attachment 10
        transients bottom & top left c. 02:00 18.6.24


      cycle slow comparator 0x32 -> 0x64 -> Check ASIC Control -> 0x32
        aida01 - aida04 rates return to normal

      ADC data item stats - attachment 11
      DSSSD energy spectra - attachment 12

23.40 cycle slow comparator 0x32 -> 0x64 -> Check ASIC Control -> 0x32
        aida01 - aida04 rates return to normal


00.10 cycle slow comparator 0x32 -> 0x64 -> Check ASIC Control -> 0x32
        aida01 - aida04 rates return to normal

01.43 ADC item stats OK - attachment 13
       FEE64 temps OK - attachment 14

04.43 ADC data items stats
       ADC data item stats - attachment 15
       FEE64 temps OK - attachment 16
       cycle slow comparator 0x32 -> 0x64 -> Check ASIC Control -> 0x32
        aida01 - aida04 rates return to normal
       ADC data item stats - attachment 17
       per FEE64 Rate spectra - attachment 18
       Merger, TapeServer etc. - attachment 19
        data file R101_51

07.00 FEE64 temps OK - attachment 21
      ADC data items OK - attachment 20
      Merger, Tape Server OK - attachment 22

      DSSSD bias & leakage current - attachment 23

      DAQ stop, close data file
      DSSSD bias OFF
      FEE64 power OFF
         

Gzip'ing all uncompressed data files
Attachment 1: Screenshot_from_2024-06-18_23-18-30.png
Screenshot_from_2024-06-18_23-18-30.png
Attachment 2: Screenshot_from_2024-06-18_23-19-08.png
Screenshot_from_2024-06-18_23-19-08.png
Attachment 3: Screenshot_from_2024-06-18_23-19-35.png
Screenshot_from_2024-06-18_23-19-35.png
Attachment 4: Screenshot_from_2024-06-18_23-19-58.png
Screenshot_from_2024-06-18_23-19-58.png
Attachment 5: Screenshot_from_2024-06-18_23-20-34.png
Screenshot_from_2024-06-18_23-20-34.png
Attachment 6: Screenshot_from_2024-06-18_23-21-15.png
Screenshot_from_2024-06-18_23-21-15.png
Attachment 7: Screenshot_from_2024-06-18_23-21-39.png
Screenshot_from_2024-06-18_23-21-39.png
Attachment 8: Screenshot_from_2024-06-18_23-21-58.png
Screenshot_from_2024-06-18_23-21-58.png
Attachment 9: Screenshot_from_2024-06-18_23-22-33.png
Screenshot_from_2024-06-18_23-22-33.png
Attachment 10: Screenshot_from_2024-06-18_23-23-38.png
Screenshot_from_2024-06-18_23-23-38.png
Attachment 11: Screenshot_from_2024-06-18_23-27-21.png
Screenshot_from_2024-06-18_23-27-21.png
Attachment 12: Screenshot_from_2024-06-18_23-32-38.png
Screenshot_from_2024-06-18_23-32-38.png
Attachment 13: Screenshot_from_2024-06-19_01-43-09.png
Screenshot_from_2024-06-19_01-43-09.png
Attachment 14: Screenshot_from_2024-06-19_01-43-52.png
Screenshot_from_2024-06-19_01-43-52.png
Attachment 15: Screenshot_from_2024-06-19_04-44-25.png
Screenshot_from_2024-06-19_04-44-25.png
Attachment 16: Screenshot_from_2024-06-19_04-44-51.png
Screenshot_from_2024-06-19_04-44-51.png
Attachment 17: Screenshot_from_2024-06-19_04-46-42.png
Screenshot_from_2024-06-19_04-46-42.png
Attachment 18: Screenshot_from_2024-06-19_04-47-32.png
Screenshot_from_2024-06-19_04-47-32.png
Attachment 19: Screenshot_from_2024-06-19_04-48-46.png
Screenshot_from_2024-06-19_04-48-46.png
Attachment 20: Screenshot_from_2024-06-19_07-01-22.png
Screenshot_from_2024-06-19_07-01-22.png
Attachment 21: Screenshot_from_2024-06-19_07-01-45.png
Screenshot_from_2024-06-19_07-01-45.png
Attachment 22: Screenshot_from_2024-06-19_07-02-17.png
Screenshot_from_2024-06-19_07-02-17.png
Attachment 23: Screenshot_from_2024-06-19_07-03-24.png
Screenshot_from_2024-06-19_07-03-24.png
  653   Wed Jun 19 12:00:56 2024 CBRun 93/94/95 - background

04:00 17 June

Moved the detectors back to the usual configuration (as per run 89).

Moved target away - background run.

 

Some issues with run numbers - see attached run name R93, but run 94 is written. See attached. 93 and 95 contain data. 94 does *not* exist.

 

No obvious differences wrt. target run, except Rutherford.

Attachment 1: Screenshot_from_2024-06-18_05-11-59.png
Screenshot_from_2024-06-18_05-11-59.png
Attachment 2: Screenshot_from_2024-06-18_06-48-46.png
Screenshot_from_2024-06-18_06-48-46.png
Attachment 3: Screenshot_from_2024-06-18_06-48-49.png
Screenshot_from_2024-06-18_06-48-49.png
Attachment 4: Screenshot_from_2024-06-18_06-49-08.png
Screenshot_from_2024-06-18_06-49-08.png
Attachment 5: Screenshot_from_2024-06-18_06-49-16.png
Screenshot_from_2024-06-18_06-49-16.png
Attachment 6: Screenshot_from_2024-06-18_06-49-23.png
Screenshot_from_2024-06-18_06-49-23.png
  652   Wed Jun 19 11:54:26 2024 CBRun 91/92 - moving detectors

~ 2:00 17 June

 

Run 91

Attempted to move the detectors closer to the beam, and move them in sync with the injection.

Found best position was

Right arm: s by 5 mm, r p in (not moving). No benefits from moving. 10 mm + r p in on sync results in beam clipping during injection. 69.55 mm not moving.

Left arm: s by 5 mm, l p in (in sync with motion). Due to the short beam preparation time, only the trigger out was used as per screenshots below. 23.58 -- 67.90 moving.

Saw spike on leakage current for both detectors on the right arm. Unsure if this because they are being moved, or because they were hit with beam. Leakage current recovered.

 

Run 92 - acquired data as above.

Rutherford locus more clear on left arm detectors. No clear sign of nuclear reactions signals.

Attachment 1: Screenshot_from_2024-06-18_03-07-50.png
Screenshot_from_2024-06-18_03-07-50.png
Attachment 2: Screenshot_from_2024-06-18_03-53-51.png
Screenshot_from_2024-06-18_03-53-51.png
Attachment 3: Screenshot_from_2024-06-18_04-20-15.png
Screenshot_from_2024-06-18_04-20-15.png
Attachment 4: Screenshot_from_2024-06-18_04-20-18.png
Screenshot_from_2024-06-18_04-20-18.png
Attachment 5: Screenshot_from_2024-06-18_04-20-21.png
Screenshot_from_2024-06-18_04-20-21.png
  651   Mon Jun 17 23:10:08 2024 CBRun 90

00:10

As run 89, detectors moved fully out.

 

01:20

TL detector noisy. Shook it. Fixed it.

Temps OK. Noise spike reported in Run 89 still there.

Target 6E11. Beam stable.

 

02:00

TL detector noisy. Shook it. Fixed it.

 

ELOG STOPPED WORKING

 

 

Attachment 1: Screenshot_from_2024-06-18_01-21-47.png
Screenshot_from_2024-06-18_01-21-47.png
Attachment 2: Screenshot_from_2024-06-18_01-21-55.png
Screenshot_from_2024-06-18_01-21-55.png
  650   Mon Jun 17 16:26:00 2024 JM, RSS, CBRun 89

17:20

(Run 87 Junk and Run 88 does not exist)

Beam charge state set to 5+. Energy is ~1.06 MeV/u

Cooler voltage setpoint 590.3 V.

FC20 read value 589.07 V.

Right arm at 64.60 mm

Left arm IN at 36.54 mm

FakeMotion script running

horizontal beam position 3.5 mm (new overlap position with new beam)

vertical beam position  0.0 mm

Beam intensity 7.0e+6 particles.

Target density ~5e11 at/cm2

Beam is stored in the ring for ~20 seconds. From beam overlap tests ~ 75% of the beam is destroyed in this time.

Beam halo appears slightly reduced. This may be due to reduced intensity and reduced injection rate (20 s vs 5s previously)

 

18:00

FEE64 temperatures OK

ADC data item stats OK

per FEE64 Rate spectra OK

Beam intensity 1.17e+7 particles

Target density ~5.25E11 atoms/cm2

 

20:28


FEE64 temperatures OK

ADC data item stats OK

per FEE64 Rate spectra OK

Beam intensity 1.08e+7 particles

Target density ~3.84E11 atoms/cm2

 

23:58

Temps OK. Stats OK. Merger OK. Write to disk OK.

Waveforms attached.

Beam keeps going away due to "solenoid" issues.

On-line analysis screenshots attached.

Without beam, we see a clear noise spike in all detectors every 50-200 seconds randomly. This is *not* correlated with beam injection or dump.

Checkloaded - no effect. Beam back - no changes.

Attachment 1: Screenshot_from_2024-06-17_17-59-13.png
Screenshot_from_2024-06-17_17-59-13.png
Attachment 2: Screenshot_from_2024-06-17_17-59-30.png
Screenshot_from_2024-06-17_17-59-30.png
Attachment 3: Screenshot_from_2024-06-17_17-59-38.png
Screenshot_from_2024-06-17_17-59-38.png
Attachment 4: Screenshot_from_2024-06-17_17-59-54.png
Screenshot_from_2024-06-17_17-59-54.png
Attachment 5: Screenshot_from_2024-06-17_18-00-00.png
Screenshot_from_2024-06-17_18-00-00.png
Attachment 6: Screenshot_from_2024-06-17_20-27-54.png
Screenshot_from_2024-06-17_20-27-54.png
Attachment 7: Screenshot_from_2024-06-17_20-28-05.png
Screenshot_from_2024-06-17_20-28-05.png
Attachment 8: Screenshot_from_2024-06-17_20-28-13.png
Screenshot_from_2024-06-17_20-28-13.png
Attachment 9: Screenshot_from_2024-06-17_20-30-10.png
Screenshot_from_2024-06-17_20-30-10.png
Attachment 10: Screenshot_from_2024-06-17_20-30-15.png
Screenshot_from_2024-06-17_20-30-15.png
Attachment 11: Screenshot_from_2024-06-17_23-48-01.png
Screenshot_from_2024-06-17_23-48-01.png
Attachment 12: Screenshot_from_2024-06-17_23-48-08.png
Screenshot_from_2024-06-17_23-48-08.png
Attachment 13: Screenshot_from_2024-06-17_23-48-57.png
Screenshot_from_2024-06-17_23-48-57.png
Attachment 14: Screenshot_from_2024-06-17_23-49-26.png
Screenshot_from_2024-06-17_23-49-26.png
Attachment 15: Screenshot_from_2024-06-17_23-49-34.png
Screenshot_from_2024-06-17_23-49-34.png
Attachment 16: Screenshot_from_2024-06-17_23-49-39.png
Screenshot_from_2024-06-17_23-49-39.png
Attachment 17: Screenshot_from_2024-06-17_23-49-46.png
Screenshot_from_2024-06-17_23-49-46.png
Attachment 18: Screenshot_from_2024-06-17_23-49-52.png
Screenshot_from_2024-06-17_23-49-52.png
Attachment 19: Screenshot_from_2024-06-17_23-50-21.png
Screenshot_from_2024-06-17_23-50-21.png
  649   Mon Jun 17 06:26:45 2024 JM, OF, AK, CBChange to 5+ beam

07:00

We are changing to 5+ beam so that we can access higher energies and possibly observe a higher lying, wider resonance.

Beam energy change to 1.06 MeV/u.

Detectors out

Target nozzle out

Pressure in E1 had reached 2e-1 mbar and turbo was spinning down to 810 Hz. Pressure reduced and E1 has reduced to 8.2e-2 mbar. We may require a pressure increase when we want the target back?

Run changed and out into no storage mode so we do not write rubbish to disk.

 

11:34

Beam tuning in the ring in progress

Merger OK. Not writing to disk.

Temps OK. Rates OK.

Leakage current OK.

Motors out. Target out.

 

13:31

Checks OK. Beam tuning near completion.

Some data acquired during beam tuning attached.

Attachment 1: Screenshot_from_2024-06-17_11-32-46.png
Screenshot_from_2024-06-17_11-32-46.png
Attachment 2: Screenshot_from_2024-06-17_11-33-47.png
Screenshot_from_2024-06-17_11-33-47.png
Attachment 3: Screenshot_from_2024-06-17_11-34-01.png
Screenshot_from_2024-06-17_11-34-01.png
Attachment 4: Screenshot_from_2024-06-17_11-34-11.png
Screenshot_from_2024-06-17_11-34-11.png
Attachment 5: Screenshot_from_2024-06-17_13-30-41.png
Screenshot_from_2024-06-17_13-30-41.png
Attachment 6: Screenshot_from_2024-06-17_13-30-52.png
Screenshot_from_2024-06-17_13-30-52.png
Attachment 7: Screenshot_from_2024-06-17_13-30-56.png
Screenshot_from_2024-06-17_13-30-56.png
  648   Mon Jun 17 05:47:35 2024 JMRun 86

06:50

No sign of the resonance during night shift. New run started but when operators are avaiable we will likely move to higher energies to try and observe the (p,a) reaction.

Cooler voltage setpoint 247.8 V.

FC20 read value 247.25 V.

Right arm at 64.60 mm

Left arm IN at 36.54 mm

FakeMotion script running

horizontal beam position 1.0 mm

vertical beam position  -5mm

Beam intensity 7.0e+7 particles.

Target density ~5e11 at/cm2

  647   Mon Jun 17 04:01:16 2024 JMRun 85

05:00

Cooler voltage setpoint 247.45 V.

FC20 read value 246.9 V.

Right arm at 64.60 mm

Left arm IN at 36.54 mm

FakeMotion script running

horizontal beam position 1.0 mm

vertical beam position  -5mm

Beam intensity 7.0e+7 particles.

Target density ~5e11 at/cm2

Target temperature is 38.6 K. Values of E1 is 819 Hz, 2.7A, 1.85e-1 mbar

All checks OK

  646   Mon Jun 17 02:05:37 2024 JMRun 84

03:00

Cooler voltage setpoint 247.7 V.

FC20 read value 247.15 V.

Right arm at 64.60 mm

Left arm IN at 36.54 mm

FakeMotion script running

horizontal beam position 1.0 mm

vertical beam position  -5mm

Beam intensity 7.0e+7 particles.

Target density ~5e11 at/cm2

Target temperature is 38.6 K. Values of E1 is 820 Hz, 2.7A, 1.8e-1 mbar

All checks OK

03:34

Noise increase in top left dssd. Servo motor moved to resolve issue.

  645   Mon Jun 17 00:05:13 2024 JMRun 83

01:00

Cooler voltage setpoint 247.6 V.

FC20 read value 247.05 V.

Right arm at 64.60 mm

Left arm IN at 36.54 mm

FakeMotion script running

horizontal beam position 1.0 mm

vertical beam position  -5mm

Beam intensity 7.0e+7 particles.

Target density ~5e11 at/cm2

Target temperature is 38.6 K. Values of E1 is 820 Hz, 2.7A, 1.7e-1 mbar

All checks OK

02:50

Noise increase on dssd top left. Moved servo  motor to resolve issue.

  644   Sun Jun 16 22:49:22 2024 JMRun 82

23:38

Cooler voltage setpoint 247.5 V.

FC20 read value 246.95 V.

Right arm at 64.60 mm

Left arm IN at 36.54 mm

FakeMotion script running

horizontal beam position 1.0 mm

vertical beam position  -5mm

Beam intensity 7.0e+7 particles.

Target density ~5e11 at/cm2

Target temperature is 38.6 K. Values of E1 is 820 Hz, 2.7A, 1.7e-1 mbar

All checks OK

  643   Sun Jun 16 22:40:11 2024 JMTarget overlap test - R81

23:10

Beam set to bunched mode to observe beam lifetime. 

Electron cooler voltage setpoint kept at 246.8 V. Note this is 0.5 V lower than the previous test where the setpoint was 247.3 V

This corresponds to ~13.5 keV in the beam energy.

Attachment 1 shows target scan. Maximum still observed at a horizontal beam position of 1.0 mm.

Moving 0.5 V up or down does not impact the orbit enough to shift the beam position off the target.

Beam lifetime off target seems larger than previous, perhaps better vacuum conditions?

Back to coasting beam

Attachment 1: Beam-overlap-updated2.png
Beam-overlap-updated2.png
  642   Sun Jun 16 19:01:30 2024 ML, EMRun 80

20:00

Ecooler setpoint 246.8

Readout 246.25

Collector current 4.3

Target density ~5.3E11

Beam intensity 7E7 ions

  641   Sun Jun 16 13:51:11 2024 CB EM MLRun 79

14:51

Ecooler setpoint 246.9

Readout 246.35

Collector current 4.3

Target density 4.6E11

Beam intensity 8E7 ions

 

Shook left arm to fix it.

 

15:32

Right half on BL detector was not counting. Checkloaded. Fixed.

 

17:20

Moved left arm out/in to fix it.

Reduced temperature to 38.5 K in steps as the target density was declining.

E1: pressure: 1.57E-1 mbar, turbo frequency: 819  Hz

Target density: 1.5E11 atoms/cm2; Ion beam intensity: 6E7 particles (went down slightly, have not been able to get everything back)

DSSSD bias & leakage current OK 
FEE64 temperatures OK 
ADC data item stats OK 
per FEE64 Rate spectra OK 

 

19:50

E1: pressure: 1.61E-1 mbar, turbo frequency: 819  Hz

Target density: 1.5E11 atoms/cm2; Ion beam intensity: 7E7 particles (recovered slightly)

DSSSD bias & leakage current OK 
FEE64 temperatures OK 
ADC data item stats OK 
per FEE64 Rate spectra OK 
Attachment 1: Screenshot_from_2024-06-16_14-53-27.png
Screenshot_from_2024-06-16_14-53-27.png
Attachment 2: Screenshot_from_2024-06-16_14-53-45.png
Screenshot_from_2024-06-16_14-53-45.png
Attachment 3: Screenshot_from_2024-06-16_14-55-19.png
Screenshot_from_2024-06-16_14-55-19.png
Attachment 4: Screenshot_from_2024-06-16_17-21-55.png
Screenshot_from_2024-06-16_17-21-55.png
Attachment 5: Screenshot_from_2024-06-16_17-22-35.png
Screenshot_from_2024-06-16_17-22-35.png
Attachment 6: Screenshot_from_2024-06-16_19-52-34.png
Screenshot_from_2024-06-16_19-52-34.png
Attachment 7: Screenshot_from_2024-06-16_19-53-16.png
Screenshot_from_2024-06-16_19-53-16.png
  640   Sun Jun 16 13:29:05 2024 RSS CBRun 78

Target clogged?

  639   Sun Jun 16 06:02:31 2024 RSSRun 77

Run 77 started at 06:48

Cooler voltage setpoint 247.0 V.

 

07:05

DSSSD bias & leakage current OK - attachments 1 & 2

FEE64 temperatures OK - attachment 3

ADC data item stats OK - attachment 4

per FEE64 Rate spectra OK - attachment 5

Beam intensity 8.45e+7 particles

Target density ~5.34E11 atoms/cm2

 

09:02

DSSSD bias & leakage current OK - attachments 6 & 7

FEE64 temperatures OK - attachment 8

ADC data item stats OK - attachment 9

per FEE64 Rate spectra OK - attachment 10

Beam intensity 7.85e+7 particles

Target density ~5.29E11 atoms/cm2

 

11:00

DSSSD bias & leakage current OK - attachments 11 & 12

FEE64 temperatures OK - attachment 13

ADC data item stats OK - attachment 14

per FEE64 Rate spectra OK - attachment 15

Beam intensity 8.5e+7 particles

Target density ~6.1E11 atoms/cm2

 

13:00

DSSSD bias & leakage current OK - attachments 16 & 17

FEE64 temperatures OK - attachment 18

ADC data item stats OK - attachment 19

per FEE64 Rate spectra OK - attachment 20

Beam intensity 8.2e+7 particles

Target density ~6.1E11 atoms/cm2
Attachment 1: Screenshot_from_2024-06-16_07-05-18.png
Screenshot_from_2024-06-16_07-05-18.png
Attachment 2: Screenshot_from_2024-06-16_07-06-01.png
Screenshot_from_2024-06-16_07-06-01.png
Attachment 3: Screenshot_from_2024-06-16_07-05-33.png
Screenshot_from_2024-06-16_07-05-33.png
Attachment 4: Screenshot_from_2024-06-16_07-05-44.png
Screenshot_from_2024-06-16_07-05-44.png
Attachment 5: Screenshot_from_2024-06-16_07-05-53.png
Screenshot_from_2024-06-16_07-05-53.png
Attachment 6: Screenshot_from_2024-06-16_09-01-29.png
Screenshot_from_2024-06-16_09-01-29.png
Attachment 7: Screenshot_from_2024-06-16_09-02-12.png
Screenshot_from_2024-06-16_09-02-12.png
Attachment 8: Screenshot_from_2024-06-16_09-01-46.png
Screenshot_from_2024-06-16_09-01-46.png
Attachment 9: Screenshot_from_2024-06-16_09-01-55.png
Screenshot_from_2024-06-16_09-01-55.png
Attachment 10: Screenshot_from_2024-06-16_09-02-03.png
Screenshot_from_2024-06-16_09-02-03.png
Attachment 11: Screenshot_from_2024-06-16_11-00-13.png
Screenshot_from_2024-06-16_11-00-13.png
Attachment 12: Screenshot_from_2024-06-16_11-00-18.png
Screenshot_from_2024-06-16_11-00-18.png
Attachment 13: Screenshot_from_2024-06-16_11-00-33.png
Screenshot_from_2024-06-16_11-00-33.png
Attachment 14: Screenshot_from_2024-06-16_11-00-43.png
Screenshot_from_2024-06-16_11-00-43.png
Attachment 15: Screenshot_from_2024-06-16_11-00-52.png
Screenshot_from_2024-06-16_11-00-52.png
Attachment 16: Screenshot_from_2024-06-16_12-59-53.png
Screenshot_from_2024-06-16_12-59-53.png
Attachment 17: Screenshot_from_2024-06-16_13-00-04.png
Screenshot_from_2024-06-16_13-00-04.png
Attachment 18: Screenshot_from_2024-06-16_13-00-24.png
Screenshot_from_2024-06-16_13-00-24.png
Attachment 19: Screenshot_from_2024-06-16_13-00-34.png
Screenshot_from_2024-06-16_13-00-34.png
Attachment 20: Screenshot_from_2024-06-16_13-00-45.png
Screenshot_from_2024-06-16_13-00-45.png
  638   Sat Jun 15 21:14:56 2024 JM, JS, LCRun 76

22:15

New run 76 - beam on target with definite target overlap [coasting]

Cooler voltage setpoint 247.1 V.

FC20 read value 246.55 V.

Right arm at 64.60 mm

Left arm IN at 36.54 mm

FakeMotion script running

horizontal beam position 1.0 mm (new beam target overlap)

vertical beam position  -5mm

Beam intensity 7.5e+7 particles.

Target density ~6.2e11 at/cm2

Target temperature is 38.5 K. Values of E1 is 820 Hz, 2.6A, 1.85e-1 mbar

 

 

00:15

 

System and detector check is OK. E1 pressure is increasing slowly. target 4e+11/cm2

 

02:15

 

System and detector check is OK. E1 pressure is increasing slowly. P=2,2e-1 mbar Pump frequency is 807 Hz, 2.7A. Target 5e+11/cm2

 

03:15

 

System and detector check is OK. E1 pressure reaches 2,3e-1 mbar, thus flow is decrease a bit. Pump frequency is 800 Hz, 2.8A. Target 6e+11/cm2

 

04:15

 

System and detector check is OK. E1 pressure is stabilized. New values: 2.07e-1 mbar, 810 Hz, 2.8 A. Target 5e+11/cm2. Resonance peak not so visible- see attachment 1.

 

05:43

 

System and detector check is OK. E1 pressure is slowly increasing. New values: 2.13e-1 mbar, 807 Hz, 2.8 A. Target 5e+11/cm2. Beam intensity stable at 7e+7 particles.

Attachment 1: Screenshot_from_2024-06-15_22-16-53.png
Screenshot_from_2024-06-15_22-16-53.png
  637   Sat Jun 15 13:30:18 2024 JM, JSRun 75

14:30

New run 75 - beam on target with definite target overlap [coasting]

Cooler voltage setpoint 247.2 V.

FC20 read value 246.65 V.

Right arm at 64.60 mm

Left arm IN at 36.54 mm

FakeMotion script running

horizontal beam position 1.0 mm (new beam target overlap)

vertical beam position  -5mm

Beam intensity 8.5e+7 particles.

Target density ~6e11 at/cm2

Target temperature is 36.6 K. Values of E1 is 820 Hz, 2.6A, 1.4e-1 mbar

 

16:00

Beam intensity 8.5e+7 particles.

Target density ~4e11 at/cm2

Target temperature is 36.6 K. Values of E1 is 820 Hz, 2.6A, 1.68e-1 mbar

DSSSD bias & leakage current OK - attachment 1

FEE64 temperatures OK - attachment 2

ADC data item stats OK - attachment 3

 

16:58

Pressure on E1 has risen to 1.74e-1 mbar. Target density ~4.5e11 at/cm2.

Turbo is decreased to 818-819 Hz. We will increase temperature sligtly to see if this helps.

17:11

Increase in noise on top left DSSD. Correlates with spike in the source seen on lassie. Moved servo motor out then in. Problem resolved. 

System checks OK

18:35

Tempature has been increase in small steps to 38.4 K. E1 appears to have stabalised at 1.71e-1 mbar. Turbo speed 818-819 Hz. 

Target density ~5.5e11 at/cm2. We will wait to see how it evolves over time.

18:47

Increase in noise on top left DSSD. Correlates with spike in the source seen on lassie. Moved servo motor out then in. Problem resolved. 

20:55

All checks OK

Tempature has been increase in small steps to 38.4 K. E1  1.81e-1 mbar. Turbo speed 818-819 Hz. 

21:48

System checks OK

22:14

Run stopped

Attachment 1: Screenshot_from_2024-06-15_15-53-57.png
Screenshot_from_2024-06-15_15-53-57.png
Attachment 2: Screenshot_from_2024-06-15_15-54-30.png
Screenshot_from_2024-06-15_15-54-30.png
Attachment 3: Screenshot_from_2024-06-15_15-54-55.png
Screenshot_from_2024-06-15_15-54-55.png
  636   Sat Jun 15 06:10:21 2024 EM, PMHRun 74

7:00

Cooler voltage setpoint 247.4 V.

 

7:44

E1 Turbo at 808 Hz, 2.75 A

DSSSD bias & leakage current OK 

FEE64 temperatures OK 

ADC data item stats OK 

per FEE64 Rate spectra OK 

Beam intensity 7.5e+7 particles

Target density ~3E10 atoms/cm2

 

8:30

E1 pressure was too high (2.13E-1 mbar) and turbo frequency too low (807 Hz).

We increased T1 in steps of 0.2 K to 39.6 K, which improved in E1 (2E-1 mbar) and turbo freq. (813 Hz).

T1 was decreased again to 39.4 K.

After speaking to Nikos we slightly decreased the gas flow (~15 bar) as desribed in elog entry 631.

Values now: turbo (817 Hz), E1 pressure (1.78E-1 mbar) -> both improved

 

10:06

The gas flow was decreased in three steps, there is now change visible on the gauge (still ~15 bar).

The turbo frequency ist stable at 819 Hz and pressure at E1 is 1.45E-1 mbar.

The temperature is currently at 39.0 K and we will start decreasing it to try and bring the target density up.

DSSSD bias & leakage current

OK FEE64 temperatures OK

ADC data item stats OK

per FEE64 Rate spectra OK

Beam intensity 7.5e+7 particles

Target density ~2E10 atoms/cm2

11:50

All system checks OK

Target density is ~3e11 at/cm2.

Temoerature of the nozzle is curretnly 37.5K with E1=1.6e-1 mbar. Turbo running at 820 Hz

 

14:29

Run stopped

 

 

 

Attachment 1: Screenshot_from_2024-06-15_07-40-18.png
Screenshot_from_2024-06-15_07-40-18.png
Attachment 2: Screenshot_from_2024-06-15_07-40-48.png
Screenshot_from_2024-06-15_07-40-48.png
Attachment 3: Screenshot_from_2024-06-15_07-41-45.png
Screenshot_from_2024-06-15_07-41-45.png
Attachment 4: Screenshot_from_2024-06-15_07-41-51.png
Screenshot_from_2024-06-15_07-41-51.png
Attachment 5: Screenshot_from_2024-06-15_07-41-56.png
Screenshot_from_2024-06-15_07-41-56.png
Attachment 6: Screenshot_from_2024-06-15_07-43-29.png
Screenshot_from_2024-06-15_07-43-29.png
  635   Sat Jun 15 00:46:45 2024 TDCRYRING hall temperature transient
CRYRING hall temperature transient - evening of Friday 14 June

Cooling of recirculating water failed on a number of circuits within the GSI experimental hall including S4 and CRYRING
Subsequently repaired bu on-call maintenance engineer
Attachment 1: Capture.PNG
Capture.PNG
  634   Fri Jun 14 22:18:18 2024 RSS, CB, USRun 73

Beam is back.New run started.

 

Cooler voltage setpoint 247.3 V.

 

23:18

DSSSD bias & leakage current OK 

FEE64 temperatures OK 

ADC data item stats OK 

per FEE64 Rate spectra OK 

Beam intensity 6e+7 particles

Target density ~4E10 atoms/cm2

 

01:01

DSSSD bias & leakage current OK - attachment 1 & 5

FEE64 temperatures OK - attachment 2

ADC data item stats OK - attachment 3

per FEE64 Rate spectra OK - attachment 4

Beam intensity 7e+7 particles

Target density ~5E11 atoms/cm2

 

03:03

DSSSD bias & leakage current OK - attachment 6 & 7

FEE64 temperatures OK - attachment 8

ADC data item stats OK - attachment 9

per FEE64 Rate spectra OK - attachment 10

Beam intensity 7e+7 particles

Target density ~4E11 atoms/cm2

 

05:00

DSSSD bias & leakage current OK - attachment 11 & 15

FEE64 temperatures OK - attachment 12

ADC data item stats OK - attachment 13

per FEE64 Rate spectra OK - attachment 14

Beam intensity 7.4e+7 particles

Target density ~5E11 atoms/cm2

 

05:15

Turbo frequency went below 810 Hz. Set temperature increased from 38.8 K to 39.0 K.

 

06:50

DSSSD bias & leakage current OK - attachment 16 & 17

FEE64 temperatures OK - attachment 18

ADC data item stats OK - attachment 19

per FEE64 Rate spectra OK - attachment 20

Beam intensity 7.5e+7 particles

Target density ~2.3E11 atoms/cm2
Attachment 1: Screenshot_from_2024-06-15_00-59-40.png
Screenshot_from_2024-06-15_00-59-40.png
Attachment 2: Screenshot_from_2024-06-15_00-59-53.png
Screenshot_from_2024-06-15_00-59-53.png
Attachment 3: Screenshot_from_2024-06-15_01-00-03.png
Screenshot_from_2024-06-15_01-00-03.png
Attachment 4: Screenshot_from_2024-06-15_01-00-12.png
Screenshot_from_2024-06-15_01-00-12.png
Attachment 5: Screenshot_from_2024-06-15_01-00-20.png
Screenshot_from_2024-06-15_01-00-20.png
Attachment 6: Screenshot_from_2024-06-15_03-03-21.png
Screenshot_from_2024-06-15_03-03-21.png
Attachment 7: Screenshot_from_2024-06-15_03-03-41.png
Screenshot_from_2024-06-15_03-03-41.png
Attachment 8: Screenshot_from_2024-06-15_03-03-49.png
Screenshot_from_2024-06-15_03-03-49.png
Attachment 9: Screenshot_from_2024-06-15_03-04-00.png
Screenshot_from_2024-06-15_03-04-00.png
Attachment 10: Screenshot_from_2024-06-15_03-04-08.png
Screenshot_from_2024-06-15_03-04-08.png
Attachment 11: Screenshot_from_2024-06-15_05-00-41.png
Screenshot_from_2024-06-15_05-00-41.png
Attachment 12: Screenshot_from_2024-06-15_05-01-00.png
Screenshot_from_2024-06-15_05-01-00.png
Attachment 13: Screenshot_from_2024-06-15_05-01-14.png
Screenshot_from_2024-06-15_05-01-14.png
Attachment 14: Screenshot_from_2024-06-15_05-01-23.png
Screenshot_from_2024-06-15_05-01-23.png
Attachment 15: Screenshot_from_2024-06-15_05-01-34.png
Screenshot_from_2024-06-15_05-01-34.png
Attachment 16: Screenshot_from_2024-06-15_06-49-51.png
Screenshot_from_2024-06-15_06-49-51.png
Attachment 17: Screenshot_from_2024-06-15_06-49-57.png
Screenshot_from_2024-06-15_06-49-57.png
Attachment 18: Screenshot_from_2024-06-15_06-50-13.png
Screenshot_from_2024-06-15_06-50-13.png
Attachment 19: Screenshot_from_2024-06-15_06-50-25.png
Screenshot_from_2024-06-15_06-50-25.png
Attachment 20: Screenshot_from_2024-06-15_06-50-33.png
Screenshot_from_2024-06-15_06-50-33.png
  633   Fri Jun 14 19:59:14 2024 US, LC, CBNew run with name of 72 is started due to target issue

21:00

Target is lost. Background run is being collected.

New run 72 - beam on target with definite target overlap [coasting]

Cooler voltage setpoint 247.3 V.

FC20 read value 246.75 V.

Right arm at 64.60 mm

Left arm IN at 36.54 mm18:00

FakeMotion script running

horizontal beam position 1.0 mm (new beam target overlap)

vertical beam position  -5mm

Beam intensity 6e+7 particles.

Target density ~3e9 at/cm2

Target temperature is 90 K. Values of E1 is 820 Hz, 2.6A, 4.2e-2 mbar

 

Aida03 detector T is stable.SCreenshot is attached.

 

22:27

Cooling fixed. Temps OK.

Attachment 1: Screenshot_from_2024-06-14_21-08-22.png
Screenshot_from_2024-06-14_21-08-22.png
Attachment 2: Screenshot_from_2024-06-14_21-08-22.png
Screenshot_from_2024-06-14_21-08-22.png
  632   Fri Jun 14 14:52:02 2024 JMBeam-Target Overlap Scan

The orbit of the beam in the ring was re-configured so that the beam position is more central.

With the target density at ~6e11 at/cm2 this allowed a target overlap scan using the beam lifetime to be performed. 

Beam put into bunched mode to allow the number of particles to be observed on Grafana.

The number of particles injected at the start of the cycle was then compared to the number of particles at the end of the cycle to gauge the % of the beam lost.

A flat top region was observed where >99% of the particles in the ring were lost. The centre of this flat top corresponds to a horizontal beam position of 1.0 mm

Attachment 1 - Grafana plot of beam intensity

Attachment 2 - Plot of overlap

Attachment 1: 20240614_153532.jpg
20240614_153532.jpg
Attachment 2: Beam-overlap-updated.png
Beam-overlap-updated.png
  631   Fri Jun 14 14:48:02 2024 CB NP ML[How To] Change target pressure. EMERGENCY ONLY!

ONLY FOLLOW THESE INSTRUCTIONS if E1 turbo is below 820 Hz, pressure in E1 is above 2.5E-1 mbar and experts (Nikos, Andreas) cannot be reached.

 

Go to the red cabinet under the stairs.

Open rightmost door. See first photo. Do not press the button on the handle or the door will lock!

Turn down pressure regulator indicated in second photo by approximately 1 degree. Read regulator to know which way to turn to decrease. No immediate change should be obvious on the pressure regulator.

Check E1 pressure for 5 minutes. If no effect, turn pressure down again until an effect appears.

Attachment 1: 20240614_142435.jpg
20240614_142435.jpg
Attachment 2: 20240614_142440.jpg
20240614_142440.jpg
  630   Fri Jun 14 14:46:54 2024 LC, JM, CB, TD, USRun 71 - beam on target

15:45

New run 71 - beam on target with definite target overlap [coasting]

Cooler voltage setpoint 247.3 V.

FC20 read value 246.75 V.

Right arm at 64.60 mm

Left arm IN at 36.54 mm18:00

FakeMotion script running

horizontal beam position 1.0 mm (new beam target overlap)

vertical beam position  -5mm

Beam intensity 7.0e+7 particles.

Target density ~4e11 at/cm2

Target temperature is 36.6 K. Values of E1 is 820 Hz, 2.6A, 1.4e-1 mbar

16:36

No difference observed between bunched and coasting

See attached screenshots to compare to previous run

Saw an increase in the data rate to disk ~ 2000 kB/s. On spectrum browser we saw aida15 had HEC reading hot. ASIC check on aida 15 resolved issue

 

18:00

Beam intensity 6e+7 particles.

Target density ~6e11 at/cm2

Target temperature is 36.6 K. Values of E1 is  820 Hz, 2.7 A, 1.5e-1 mbar

Aida03 detector temperature is 66 degree.

 

20:00

Beam intensity 6e+7 particles.

Target density ~6e11 at/cm2

Target temperature is 36.6 K. Values of E1 is  820 Hz, 2.7 A, 1.6e-1 mbar

Aida03 detector temperature is 63 degree. Screenshot is attached.

21:00

arget densityis dropped drastically. A temperature is ncreased signifiantly.

Beam intensity 6e+7 particles.

Target density ~2.7e9 at/cm2

Target temperature is 9 K. Values of E1 is  820 Hz, 2.7 A, 4.2e-2mbar

Aida03 detector temperature is 63 degree. Screenshot is attached.

 

Attachment 1: Screenshot_from_2024-06-14_16-35-44.png
Screenshot_from_2024-06-14_16-35-44.png
Attachment 2: Screenshot_from_2024-06-14_16-36-06.png
Screenshot_from_2024-06-14_16-36-06.png
Attachment 3: Screenshot_from_2024-06-14_16-36-28.png
Screenshot_from_2024-06-14_16-36-28.png
Attachment 4: Screenshot_from_2024-06-14_19-49-02.png
Screenshot_from_2024-06-14_19-49-02.png
  629   Fri Jun 14 13:45:21 2024 JM, JS, CB, TD, EHRun 70 - beam on target

14:45

New run 70 - beam on target with definite target overlap

Cooler voltage setpoint 247.3 V.

FC20 read value 246.75 V.

Right arm at 64.60 mm

Left arm IN at 36.54 mm

FakeMotion script running

horizontal beam position 1.0 mm (new beam target overlap)

vertical beam position  -5mm

Beam intensity 7.0e+7 particles.

Target density ~6.3e11 at/cm2

Target temperature is 36.6 K. Values of E1 is 820 Hz, 2.6A, 1.4e-1 mbar

All system checks OK, data rate is ~ 260 kB/s. Statistics attached.

xy and energy histograms following ~20 mins attached.

15:45

I realised we have been running in bunched mode not coasting as we have been previously.

See attached screenshots (4-7) of last 1 hour of running. We will move to coasting.

Attachment 1: Screenshot_from_2024-06-14_14-51-46.png
Screenshot_from_2024-06-14_14-51-46.png
Attachment 2: Screenshot_from_2024-06-14_14-53-57.png
Screenshot_from_2024-06-14_14-53-57.png
Attachment 3: Screenshot_from_2024-06-14_14-55-02.png
Screenshot_from_2024-06-14_14-55-02.png
Attachment 4: Screenshot_from_2024-06-14_15-38-58.png
Screenshot_from_2024-06-14_15-38-58.png
Attachment 5: Screenshot_from_2024-06-14_15-39-16.png
Screenshot_from_2024-06-14_15-39-16.png
Attachment 6: Screenshot_from_2024-06-14_15-39-36.png
Screenshot_from_2024-06-14_15-39-36.png
Attachment 7: Screenshot_from_2024-06-14_15-40-14.png
Screenshot_from_2024-06-14_15-40-14.png
  628   Fri Jun 14 10:21:24 2024 JS, TDRun 68

Frank has been tuning the beam since 08:00

detectors moved fully out

11:20

New run (R68) started at 11:20. Put into no storage mode as rates are high during beam tuning. Cut off the end of last nights run when analysing

System checks OK

 

13.20 DSSSD bias & leakage current OK - attachment 1

FEE64 temperatures OK - attachment 2

ADC data item stats - attachment 3

per FEE64 1.8.W spectra - attachments 4-5

per FEE64 Rate spectra - attachment 6

 aida16 producing HEC data - Chck ASIC load to fix

TapeServer - no storage mode

 

13.35 aida01-aida04 c. 250k ADC data items/s - will await completion of current beam/target scan to fix by cycing Left DSSSD motion

 

14.00 CB cycles Left DSSSD motion to fix aida01-aida04 rate issue

 

CRYRING vacuum & hall temprature - attachments 7-8

 

14.28 New run data file G22-201/R69

Beam/target overlap optimised - details to follow

Attachment 1: Screenshot_from_2024-06-14_13-22-13.png
Screenshot_from_2024-06-14_13-22-13.png
Attachment 2: Screenshot_from_2024-06-14_13-22-18.png
Screenshot_from_2024-06-14_13-22-18.png
Attachment 3: Screenshot_from_2024-06-14_13-22-37.png
Screenshot_from_2024-06-14_13-22-37.png
Attachment 4: Screenshot_from_2024-06-14_13-22-58.png
Screenshot_from_2024-06-14_13-22-58.png
Attachment 5: Screenshot_from_2024-06-14_13-23-30.png
Screenshot_from_2024-06-14_13-23-30.png
Attachment 6: Screenshot_from_2024-06-14_13-24-08.png
Screenshot_from_2024-06-14_13-24-08.png
Attachment 7: Screenshot_from_2024-06-14_13-25-02.png
Screenshot_from_2024-06-14_13-25-02.png
Attachment 8: vac.PNG
vac.PNG
Attachment 9: temp.PNG
temp.PNG
  627   Thu Jun 13 23:01:07 2024 RSS, JS, EHRun 67 - Night shift

Run 67

00:01

      DSSSD bias & leakage current OK - attachment 1

      FEE64 temperatures OK - attachment 2

      ADC data item stats OK - attachment 3

      per FEE64 Rate spectra OK - attachment 4

           Beam intensity 5.7E7 particles

           Target density - 4.98E11 atoms/cm2

 

02:01

      DSSSD bias & leakage current OK - attachment 5

      FEE64 temperatures OK - attachment 6

      ADC data item stats OK - attachment 7

      per FEE64 Rate spectra OK - attachment 8

           Beam intensity 5.8E7 particles

           Target density - 4.39E11 atoms/cm2

 

04:00

      DSSSD bias & leakage current OK - attachment 9

      FEE64 temperatures OK - attachment 10

      ADC data item stats OK - attachment 11

      per FEE64 Rate spectra OK - attachment 12

           Beam intensity 4.1E7 particles

           Target density - 3.57E11 atoms/cm2

 

06:00

      DSSSD bias & leakage current OK - attachment 13

      FEE64 temperatures OK - attachment 14

      ADC data item stats OK - attachment 15

      per FEE64 Rate spectra OK - attachment 16

           Beam intensity 3.7E7 particles

           Target density - 3.89E11 atoms/cm2

08:00

          The orbit adjustments started. The detectors moved out. One of these showed rates of about 20kHz. but after moving it in and out again it returned to normal.

11:00

      DSSSD bias & leakage current OK - attachment 17

      FEE64 temperatures OK - attachment 18

      ADC data item stats OK - attachment 19

      per FEE64 Rate spectra OK - attachment 20

           Beam intensity 4.9E7 particles

           Target density - 5.27E11 atoms/cm2

Attachment 1: Screenshot_from_2024-06-14_00-02-01.png
Screenshot_from_2024-06-14_00-02-01.png
Attachment 2: Screenshot_from_2024-06-14_00-02-20.png
Screenshot_from_2024-06-14_00-02-20.png
Attachment 3: Screenshot_from_2024-06-14_00-02-31.png
Screenshot_from_2024-06-14_00-02-31.png
Attachment 4: Screenshot_from_2024-06-14_00-02-40.png
Screenshot_from_2024-06-14_00-02-40.png
Attachment 5: Screenshot_from_2024-06-14_02-00-42.png
Screenshot_from_2024-06-14_02-00-42.png
Attachment 6: Screenshot_from_2024-06-14_02-00-58.png
Screenshot_from_2024-06-14_02-00-58.png
Attachment 7: Screenshot_from_2024-06-14_02-01-04.png
Screenshot_from_2024-06-14_02-01-04.png
Attachment 8: Screenshot_from_2024-06-14_02-01-13.png
Screenshot_from_2024-06-14_02-01-13.png
Attachment 9: Screenshot_from_2024-06-14_04-00-08.png
Screenshot_from_2024-06-14_04-00-08.png
Attachment 10: Screenshot_from_2024-06-14_04-00-26.png
Screenshot_from_2024-06-14_04-00-26.png
Attachment 11: Screenshot_from_2024-06-14_04-00-38.png
Screenshot_from_2024-06-14_04-00-38.png
Attachment 12: Screenshot_from_2024-06-14_04-00-52.png
Screenshot_from_2024-06-14_04-00-52.png
Attachment 13: Screenshot_from_2024-06-14_06-00-15.png
Screenshot_from_2024-06-14_06-00-15.png
Attachment 14: Screenshot_from_2024-06-14_06-00-31.png
Screenshot_from_2024-06-14_06-00-31.png
Attachment 15: Screenshot_from_2024-06-14_06-00-42.png
Screenshot_from_2024-06-14_06-00-42.png
Attachment 16: Screenshot_from_2024-06-14_06-00-51.png
Screenshot_from_2024-06-14_06-00-51.png
Attachment 17: Screenshot_from_2024-06-14_10-53-10.png
Screenshot_from_2024-06-14_10-53-10.png
Attachment 18: Screenshot_from_2024-06-14_10-53-54.png
Screenshot_from_2024-06-14_10-53-54.png
Attachment 19: Screenshot_from_2024-06-14_10-54-28.png
Screenshot_from_2024-06-14_10-54-28.png
Attachment 20: Screenshot_from_2024-06-14_10-55-05.png
Screenshot_from_2024-06-14_10-55-05.png
  626   Thu Jun 13 21:07:30 2024 LCS, USRun 67- beam on target

22:05

The energy is changed to 247.4 V.

Read back value is 246.849 V.

Right arm at 64.60 mm

Left arm IN at 36.54 mm

horizontal beam position 8.75 mm

vertical beam position  -5mm

Beam intensity 4.0e+7 particles.

Run is R67. Target temperature is 36..6 K.

Values of E1 is 820 Hz, 2.4A, 8.69e-2 mbar,  2.55e+11/cm2. The pressure is slowly increasing since this afternoon. The recent value is E1 is 820 Hz, 2.4A, 9.0e-2 mbar,  3.28e+11/cm2

 

Systen checks all OK

Attachments

13.39 DSSSD bias & leakage current OK

FEE64 temps OK

ADC data item stats OK

per FEE64 Rate spectra

Merger, TapeServer et

 data file R65_11

DSSSD bias & leakge current log

CARME DSSSD motion control

 

 

 

Attachment 1: Screenshot_from_2024-06-13_22-14-46.png
Screenshot_from_2024-06-13_22-14-46.png
Attachment 2: Screenshot_from_2024-06-13_22-15-01.png
Screenshot_from_2024-06-13_22-15-01.png
Attachment 3: Screenshot_from_2024-06-13_22-15-14.png
Screenshot_from_2024-06-13_22-15-14.png
Attachment 4: Screenshot_from_2024-06-13_22-15-38.png
Screenshot_from_2024-06-13_22-15-38.png
Attachment 5: Screenshot_from_2024-06-13_22-15-44.png
Screenshot_from_2024-06-13_22-15-44.png
Attachment 6: Screenshot_from_2024-06-13_22-15-51.png
Screenshot_from_2024-06-13_22-15-51.png
Attachment 7: Screenshot_from_2024-06-13_22-16-07.png
Screenshot_from_2024-06-13_22-16-07.png
  625   Thu Jun 13 14:46:17 2024 TDR66 beam overlap test
15.45 Start data file G22-201/R66 
      Beam/target overlap test
  624   Thu Jun 13 09:48:25 2024 TD, LCSRun 65 - beam on target

10:47

The energy is changed to 247.4 V.

Read back value is 246.849 V.

Right arm at 64.60 mm

Left arm IN at 36.54 mm

MBS working - hopefully

horizontal beam position 8.75 mm

vertical beam position  -5mm

Beam intensity 3.0e+7 particles.

e-cooler export controls before/after energy change - attachments 1 & 2

New run is R65. Target temperature is 37.6 K.

Values of E1 is 820 Hz, 2.12A, 6.01e-2 mbar,  2.8e+11/cm2

12:34

Systen checks all OK

Target density stable at ~1.5e11 at/cm2

E1 = 6e-2 mbar, turbo is fine

 

 

13,38 CB reboots RHS CRYRING control computer (atppc017) - appears to be a now unavoidable system update/reboot

13.39 DSSSD bias & leakage current OK - attachment 3

FEE64 temps OK - attachment 4

ADC data item stats OK - attachment 5

per FEE64 Rate spectra - attachment 6

Merger, TapeServer etc - attachments 7-8

 data file R65_11

DSSSD bias & leakge current log - attachment 12

CARME DSSSD motion control - attachment 13

 

13.42 E1 turbo 820Hz 2.2A 5.2e-2mbar

Target 3.45e+11/cm2 T1 36.6K

Beam c. 3e+7

 

15:20

Stats OK. Temps OK.

Merger OK. Write to disk OK.

 

Attachment 1: 20240613_104403.jpg
20240613_104403.jpg
Attachment 2: 20240613_104840.jpg
20240613_104840.jpg
Attachment 3: Screenshot_from_2024-06-13_13-39-26.png
Screenshot_from_2024-06-13_13-39-26.png
Attachment 4: Screenshot_from_2024-06-13_13-39-40.png
Screenshot_from_2024-06-13_13-39-40.png
Attachment 5: Screenshot_from_2024-06-13_13-40-09.png
Screenshot_from_2024-06-13_13-40-09.png
Attachment 6: Screenshot_from_2024-06-13_13-40-32.png
Screenshot_from_2024-06-13_13-40-32.png
Attachment 7: Screenshot_from_2024-06-13_13-40-52.png
Screenshot_from_2024-06-13_13-40-52.png
Attachment 8: Screenshot_from_2024-06-13_13-41-02.png
Screenshot_from_2024-06-13_13-41-02.png
Attachment 9: Screenshot_from_2024-06-13_13-41-16.png
Screenshot_from_2024-06-13_13-41-16.png
Attachment 10: Screenshot_from_2024-06-13_13-41-23.png
Screenshot_from_2024-06-13_13-41-23.png
Attachment 11: Screenshot_from_2024-06-13_15-12-19.png
Screenshot_from_2024-06-13_15-12-19.png
Attachment 12: Screenshot_from_2024-06-13_15-13-16.png
Screenshot_from_2024-06-13_15-13-16.png
Attachment 13: Screenshot_from_2024-06-13_15-20-28.png
Screenshot_from_2024-06-13_15-20-28.png
Attachment 14: Screenshot_from_2024-06-13_15-20-31.png
Screenshot_from_2024-06-13_15-20-31.png
  623   Thu Jun 13 06:21:53 2024 TD, LCSRun 64 - contd
07.22 E1 pump 799Hz, 2.8A, 2.6e-1 mbar, 4.8e+11/cm2

07.24 DSSSD bias & leakage current OK - attachment 1

      FEE64 temperatures OK - attachment 2

      ADC data item stats OK - attachment 3

      per FEE64 Rate spectra - attachment 4

      per FEE64 1.8.L spectra - attachments 5-6

      per FEE64 1.8.W spectra - 20us FSR - attachments 7-8

      Merger, TapeServer etc - attachments 9-10
       data file G22-201/R64_11

      DSSSD bias & leakage current log - attachment 11

      CRYRING hall temperature log - attachment 12

      CARME DSSSD position controller - attachment 13

07.57 all histograms zero'd


07.58 E1 pump 794Hz, 2.8A, 2.6e-1 mbar, 3.1e+11/cm2
08.55 E1 pump 785Hz, 2.8A, 2.74e-1 mbar, 4.8e+11/cm2. Therefore, the temperature is changed from 38.8 K to 38.4 K. Values of E1 is changed to E1 pump 782Hz, 2.7A, 2.95e-1 mbar, 6e+11/cm2
09.15 The temperature is further changed from 38.4 K to 38.0 K.  
09.25 Values of E1 is changed to E1 pump  780 Hz,  2.79 A,  3.0 e-1 mbar,  5.87 e+11/cm2
09.35 Values of E1 is changed to E1 pump  779 Hz,  2.79 A,  3.1 e-1 mbar,  2.07 e+11/cm2
09.45 Values of E1 is changed to E1 pump  778 Hz,  2.75 A,  3.0 e-1 mbar,  4.6 e+11/cm2
10.00 Values of E1 is changed to E1 pump   776 Hz,   2.78 A,  3.05 e-1 mbar,  5 e+11/cm2

Intervention by Uli. Reduced H2 regulator to c. 12 bar. Briefly opened exhaust to hasten recovery.

10.13 E1 820Hz 2.3A 7.7e-2mbar c. 2e+11/cm2
10.34 E1 820Hz 2.1A 6.2e-2mbar c. 1.2e+11/cm2 38K
      Beam c. 4e+7. Cooler set 247.3V, read 246.751V.
Attachment 1: Screenshot_from_2024-06-13_07-24-49.png
Screenshot_from_2024-06-13_07-24-49.png
Attachment 2: Screenshot_from_2024-06-13_07-28-56.png
Screenshot_from_2024-06-13_07-28-56.png
Attachment 3: Screenshot_from_2024-06-13_07-30-34.png
Screenshot_from_2024-06-13_07-30-34.png
Attachment 4: Screenshot_from_2024-06-13_07-36-35.png
Screenshot_from_2024-06-13_07-36-35.png
Attachment 5: Screenshot_from_2024-06-13_07-40-06.png
Screenshot_from_2024-06-13_07-40-06.png
Attachment 6: Screenshot_from_2024-06-13_07-41-10.png
Screenshot_from_2024-06-13_07-41-10.png
Attachment 7: Screenshot_from_2024-06-13_07-42-32.png
Screenshot_from_2024-06-13_07-42-32.png
Attachment 8: Screenshot_from_2024-06-13_07-43-30.png
Screenshot_from_2024-06-13_07-43-30.png
Attachment 9: Screenshot_from_2024-06-13_08-10-26.png
Screenshot_from_2024-06-13_08-10-26.png
Attachment 10: Screenshot_from_2024-06-13_08-10-33.png
Screenshot_from_2024-06-13_08-10-33.png
Attachment 11: Screenshot_from_2024-06-13_08-10-43.png
Screenshot_from_2024-06-13_08-10-43.png
Attachment 12: Capture.PNG
Capture.PNG
Attachment 13: Screenshot_from_2024-06-13_08-16-07.png
Screenshot_from_2024-06-13_08-16-07.png
  622   Thu Jun 13 00:50:33 2024 JM, JSRun 64 - beam on target

01:50

Right arm at 64.60 mm

Left arm IN at 36.54 mm

MBS working - hopefully

Target ~3E11

Injection ca. 3E7

horizontal beam position 8.75 mm

vertical beam position  -5mm

Changed electron cooler set point to 247.3 V

Read value 246.75 V

Collector current 4.3A

Histograms and statistics zero'd. Code reset

Rates OK. Temps OK.

 

NOTE. Between 23:45 and 12:45 the electron cooler was off? We called HKR and they turned it back on. Are we using the same cooler settings? Is the cooler still definitely working properly? Ask Claude in the morning

02:30

The target density started to oscillate between 2e+10 and 3.5e+10. However, the pump speed seems to be stable.

03:50

The oscillations seems to deepend between 1e+10 and 6e+10. Additionally the E1 pump slowed down to 815. As Nikos suggested during the tuning, I tried to lower the temperature from 41K to 40.8K.

04:00 - 04:30

No major change was observed, thus I increased the temperature to 42.2K in order to try to get the pump to the nominal rotation. After 10 minutes, the pump slowed down to 810 and the pressure in E1 reached 2.2e-1. Since increasing the temperature seemed to have a negative effect, I slowly decresed the temperature with steps of 0.2 K, until reaching 40K. At this point, the E1 pump returned to 819/820, the pressure of E1 lowered to 1.6e-1 and the target density seems to be more stable with a value of 2.8e+11.

05:00

The target density started decreasing to 1.5e+10 and the pump speed slowed to 815. I've tried to lower the temperature to 39.8K, but the target density started to decrease. Thus I've set the temperature to 40.2K. The pump speed raised to 818 but the target density remaind at 1e+11 level with oscillations between 8.5e+10 and 1.5e+11. After some minutes, it started slowly increasing towards the 2e+11 region. The E1 pumping speed is stable at 818 and the pressure of E1 is 2.02e-1.

05:10

Electron cooler off, restarted 10 minutes after by calling HKK.

05:30

Target density at 1.5e+11, however, the E1 raised to 2.3e-1 and the pump speed decreased to 810. Since going down seemed to solve it, I decreased the temperature to 39.6K. After 5 minutes the E1 lowered to 1.9e-1 and the pump speed is going up (currently at 814). The target density is at the same level of 1.5e+11.

06:00

Again the target density started to go down, as well the speed of the E1 pump and the E1 pressure reached 2.3e-1. I lowered the temperature to 39.2K: the target density raised to approximately 3.8e+11 (and keeps going up), the E1 pressure lowered to 1.9e-1 and the E1 pump speed started increasing from 810 to 815.

06:30

The target density starting going down again, the E1 started increasing to 2.3e-1 and the E1 pump speed decreased to 810. I lowered the temperature to 38.8K, the density raised to 3.5e+11, the E1 pressure lowered to 1.9e-1 and the pumping speed starting increasing.

  621   Wed Jun 12 07:16:47 2024 RSS, YL, JM, TD, CBStatus

07:56 Gleb has taken the beam to try settings for the 5+ charge state. As soon as the target is ready, the 1+ charge state will be given back.

Detectors have been moved out for safety

08:10
      Run 63
   
      DSSSD bias & leakage current OK - attachment 1

      FEE64 temperatures OK - attachment 2

      ADC data item stats - attachment 3

      per FEE64 Rate spectra - attachment 4

      
09:55 
      Run 63 

      DSSSD bias & leakage current OK 

      FEE64 temperatures OK  

      ADC data item stats OK

      per FEE64 Rate spectra OK

 

10:30

Data rate appeared very high c. 2000 kB/s instead of the usual 200 kB/s.

Rate on first two ASICs of aida02 was high (~2 kHz) but rest of aida01-04 looked fine.

ASIC check load reduced rate back to normal levels. Data rate reduced back to 200 kB/sec.

 

New ADC offsets loaded into the CARME_Sort code

 

Aida03 appears to be down. Check load does not resolve issue. Moving detector arm does not resolve issue. No spike in leakage current observed. System checks appear OK.  Zeroing ADC data item statistics keeps aida03 at 606023857 items and does not zero.

Issue resolved by resetting the data links for aida03

13:20

DSSSD bias & leakage current OK

FEE64 temperatures OK

ADC data item stats OK

per FEE64 Rate spectra OK

 

21:30

Waiting for the target.

Beam has been running, we have been running without storage.

See attachment showing beam stripe on the detectors.

 

Attachment 1: Screenshot_from_2024-06-12_08-18-27.png
Screenshot_from_2024-06-12_08-18-27.png
Attachment 2: Screenshot_from_2024-06-12_08-18-44.png
Screenshot_from_2024-06-12_08-18-44.png
Attachment 3: Screenshot_from_2024-06-12_08-18-55.png
Screenshot_from_2024-06-12_08-18-55.png
Attachment 4: Screenshot_from_2024-06-12_08-19-08.png
Screenshot_from_2024-06-12_08-19-08.png
Attachment 5: Screenshot_from_2024-06-12_11-10-34.png
Screenshot_from_2024-06-12_11-10-34.png
Attachment 6: Screenshot_from_2024-06-12_21-39-20.png
Screenshot_from_2024-06-12_21-39-20.png
  620   Tue Jun 11 21:16:50 2024 TDRun 63 continued
Run 63 continued

22.06 DSSSD bias & leakage current OK - attachment 1

      FEE64 temperatures OK - attachment 2

      ADC data item stats - attachment 3

      per FEE64 Rate spectra - attachments 4-5
       aida16 producing HEC data, ASIC check load to clear


23.56 DSSSD bias & leakage current OK - attachment 6

      FEE64 temperatures OK - attachment 7

      ADC data item stats - attachment 8

      per FEE64 Rate spectra - attachments 9

       aida01-aida04 high ADC data item rates - left DSSSD motion cycle required to reset


03.28 DSSSD bias & leakage current OK - attachment 10

      FEE64 temperatures OK - attachment 11

      ADC data item stats - attachment 12

      per FEE64 Rate spectra - attachments 13

       aida01-aida04 high ADC data item rates - left DSSSD motion cycle required to reset


07.50 DSSSD bias & leakage current OK - attachment 14

      FEE64 temperatures OK - attachment 15

      ADC data item stats - attachment 16

      per FEE64 Rate spectra - attachments 17

       aida01-aida04 high ADC data item rates (see above) appears to have resolved itself or has been resolved (but not logged)
Attachment 1: Screenshot_from_2024-06-11_22-06-38.png
Screenshot_from_2024-06-11_22-06-38.png
Attachment 2: Screenshot_from_2024-06-11_22-07-38.png
Screenshot_from_2024-06-11_22-07-38.png
Attachment 3: Screenshot_from_2024-06-11_22-08-37.png
Screenshot_from_2024-06-11_22-08-37.png
Attachment 4: Screenshot_from_2024-06-11_22-12-35.png
Screenshot_from_2024-06-11_22-12-35.png
Attachment 5: Screenshot_from_2024-06-11_22-10-38.png
Screenshot_from_2024-06-11_22-10-38.png
Attachment 6: Screenshot_from_2024-06-11_23-56-29.png
Screenshot_from_2024-06-11_23-56-29.png
Attachment 7: Screenshot_from_2024-06-12_00-00-16.png
Screenshot_from_2024-06-12_00-00-16.png
Attachment 8: Screenshot_from_2024-06-12_00-02-06.png
Screenshot_from_2024-06-12_00-02-06.png
Attachment 9: Screenshot_from_2024-06-12_00-02-35.png
Screenshot_from_2024-06-12_00-02-35.png
Attachment 10: Screenshot_from_2024-06-12_03-25-53.png
Screenshot_from_2024-06-12_03-25-53.png
Attachment 11: Screenshot_from_2024-06-12_03-26-33.png
Screenshot_from_2024-06-12_03-26-33.png
Attachment 12: Screenshot_from_2024-06-12_03-27-12.png
Screenshot_from_2024-06-12_03-27-12.png
Attachment 13: Screenshot_from_2024-06-12_03-28-02.png
Screenshot_from_2024-06-12_03-28-02.png
Attachment 14: Screenshot_from_2024-06-12_07-50-16.png
Screenshot_from_2024-06-12_07-50-16.png
Attachment 15: Screenshot_from_2024-06-12_07-52-53.png
Screenshot_from_2024-06-12_07-52-53.png
Attachment 16: Screenshot_from_2024-06-12_07-53-23.png
Screenshot_from_2024-06-12_07-53-23.png
Attachment 17: Screenshot_from_2024-06-12_07-53-53.png
Screenshot_from_2024-06-12_07-53-53.png
  619   Tue Jun 11 20:10:25 2024 JMRun 63 - Beam only background

21:00

Claude has achieved optimal cooling parameters for the beam

No target and internal work inside the cave tomorrow is required

Run 63 going - beam only background

Right arm at 64.60 mm

Left arm IN at 36.54 mm

MBS fakemotion script running

High rates on top left dssd again. Cycle left servo motor two times resolved issue

Rates OK. Temps OK.

  618   Tue Jun 11 19:44:48 2024 OF, SSTarget out and at He

Target has been moved out and H2 switched off. Changed to He.

Nikos will repair nozzle tomorrow morning. Remote checks until then.

  617   Tue Jun 11 12:26:07 2024 RSS, JS, JM, CBRun 62 - Beam optimization + Target change

Beam optimization and target change

  616   Tue Jun 11 08:34:50 2024 JM, RSS, JSRun 61 - beam on target

New run 61

Right arm at 64.60 mm

Left arm IN at 36.54 mm

MBS working - hopefully

Target 4E11

Injection ca. 3E7

horizontal beam position 8.75 mm

vertical beam position  -5mm

Changed electron cooler set point to 244.95 V

Read value 244.4 V

Collector current 4.31A

Histograms and statistics zero'd. Code reset

Rates OK. Temps OK.

10:24

We lost the beam for 5 minutes. No counting spike observed when we got it back.

1L10

We lost the beam again for 5 minutes, but no counting spike obsevered when we got it back.

 

11:50


      DSSSD bias & leakage current OK - attachment 1

      FEE64 temperatures OK - attachment 2

      ADC data item stats - attachment 3

      per FEE64 Rate spectra - attachment 4
      
      Beam intensity - ~ 3.3E7 pps

      Target density - 5.14E11 atoms/cm2

 

12:00 - 12:40

Large fluctuations in the target pressure, going down to 1e-9. We increased the temperature a little but it did not solve the problem.

Increased pressure 0.5 K more. Seems more stable

 

13:09

FEEs on det 2 stopped working. Checkloaded. Fixed.

No beam.

Attachment 1: Screenshot_from_2024-06-11_11-47-57.png
Screenshot_from_2024-06-11_11-47-57.png
Attachment 2: Screenshot_from_2024-06-11_11-48-20.png
Screenshot_from_2024-06-11_11-48-20.png
Attachment 3: Screenshot_from_2024-06-11_11-48-32.png
Screenshot_from_2024-06-11_11-48-32.png
Attachment 4: Screenshot_from_2024-06-11_11-48-44.png
Screenshot_from_2024-06-11_11-48-44.png
  615   Tue Jun 11 04:36:17 2024 CB, RSS, JSRun 60 - as 59, left servo in

05:36

Run started.

Conditions as before, except left servo now really in.

 

06:24

Run continues OK, file R60_9

Temps OK. Stats OK. Merger OK. Write to disk OK.

System wide checks as usual.

 

07:30


      DSSSD bias & leakage current OK - attachment 4

      FEE64 temperatures OK - attachment 5

      ADC data item stats - attachment 6

      per FEE64 Rate spectra - attachment 7
      
      Beam intensity - ~ 4.8E7 pps

      Target density - 7.25E11 atoms/cm2
      

 

9:20


      DSSSD bias & leakage current OK - attachment 8

      FEE64 temperatures OK - attachment 9

      ADC data item stats - attachment 10

      per FEE64 Rate spectra - attachment 11
      
      Beam intensity - ~ 4E7 pps

      Target density - 7E11 atoms/cm2
      

 

Attachment 1: Screenshot_from_2024-06-11_06-23-36.png
Screenshot_from_2024-06-11_06-23-36.png
Attachment 2: Screenshot_from_2024-06-11_06-23-56.png
Screenshot_from_2024-06-11_06-23-56.png
Attachment 3: Screenshot_from_2024-06-11_06-24-11.png
Screenshot_from_2024-06-11_06-24-11.png
Attachment 4: Screenshot_from_2024-06-11_07-27-43.png
Screenshot_from_2024-06-11_07-27-43.png
Attachment 5: Screenshot_from_2024-06-11_07-28-02.png
Screenshot_from_2024-06-11_07-28-02.png
Attachment 6: Screenshot_from_2024-06-11_07-28-14.png
Screenshot_from_2024-06-11_07-28-14.png
Attachment 7: Screenshot_from_2024-06-11_07-28-22.png
Screenshot_from_2024-06-11_07-28-22.png
Attachment 8: Screenshot_from_2024-06-11_09-18-55.png
Screenshot_from_2024-06-11_09-18-55.png
Attachment 9: Screenshot_from_2024-06-11_09-19-10.png
Screenshot_from_2024-06-11_09-19-10.png
Attachment 10: Screenshot_from_2024-06-11_09-19-20.png
Screenshot_from_2024-06-11_09-19-20.png
Attachment 11: Screenshot_from_2024-06-11_09-19-28.png
Screenshot_from_2024-06-11_09-19-28.png
  614   Mon Jun 10 20:31:12 2024 JM, AK, EH, CBRun 59 Beam on target

21:32

New run 59

Right arm at 64.60 mm

Left arm IN at 36.54 mm

MBS working - hopefully

Target 4.7E11

Injection ca. 4E7

horizontal beam position 8.75 mm

vertical beam position  -5mm

Changed electron cooler set point to 244.85 V

Read value 244.3 V

Collector current 4.31A

Histograms and statistics zero'd. Code reset

Rates OK. Temps OK.

22:44

Screenshots of the above

 

23:00

Detector 0 shows usual strong noise. Shook twice. Fixed it. Beam loss in correspondence with this event.

Beam mostly recovered a few minutes later.

23:28

Detector 0 shows noise. Shook twice with servo. No result. Shook with pneumatic. Fixed. Some beam instabilities again.

01:00

Temps OK. Rates OK.

Merger OK. Write to disk OK (file 59_13)

01:36

Detector 0 shows usual strong noise. Shook it. Fixed it.

01:46

Detector 0 shows noise. Shook twice with servo. No result. Shook with pneumatic. Fixed.

The noise increase is definitely correlated with spikes in YRT1DT1MK current reading. Unclear what the reading is.

02:36

Target density has been consistentyl above 1E12 recently, very unstable and trending forward. Increased Lakeshore setpoint by 0.5 K.

Density decreases to high 1E11, and appear significantly more stable. E1 pressure reading also decreasing to 1E-1 mbar.

Temps OK. Stats OK. Merger OK. Write to disk OK.

03:58

Temps OK. Stats OK. Merger OK. Write to disk OK.

Target still unstable and now a bit too low density. Trying to slowly lower the temperature in 0.1 K steps.

04:55

Temps OK. Stats OK. Merger OK. Write to disk OK.

Target around 6E11 atoms/cm2, with Lakeshore temperature at 30.8 K. Waiting for it to stabilise.

05:34

Detector 0 noisy. Shook. Fixed it.

Realised I had not moved the detector back at 01:46 after shaking it. Most of this run has been taken with detector 0 fully out, i.e. 18.86 mm.

Moving in and moving to a new run.

Attachment 1: Screenshot_from_2024-06-10_22-45-11.png
Screenshot_from_2024-06-10_22-45-11.png
Attachment 2: Screenshot_from_2024-06-10_22-45-21.png
Screenshot_from_2024-06-10_22-45-21.png
Attachment 3: Screenshot_from_2024-06-10_22-46-00.png
Screenshot_from_2024-06-10_22-46-00.png
Attachment 4: Screenshot_from_2024-06-10_22-46-17.png
Screenshot_from_2024-06-10_22-46-17.png
Attachment 5: Screenshot_from_2024-06-10_22-46-26.png
Screenshot_from_2024-06-10_22-46-26.png
Attachment 6: Screenshot_from_2024-06-10_22-46-46.png
Screenshot_from_2024-06-10_22-46-46.png
Attachment 7: Screenshot_from_2024-06-10_22-47-47.png
Screenshot_from_2024-06-10_22-47-47.png
Attachment 8: Screenshot_from_2024-06-10_22-47-51.png
Screenshot_from_2024-06-10_22-47-51.png
Attachment 9: Screenshot_from_2024-06-11_00-57-12.png
Screenshot_from_2024-06-11_00-57-12.png
Attachment 10: Screenshot_from_2024-06-11_00-57-20.png
Screenshot_from_2024-06-11_00-57-20.png
Attachment 11: Screenshot_from_2024-06-11_03-58-02.png
Screenshot_from_2024-06-11_03-58-02.png
Attachment 12: Screenshot_from_2024-06-11_04-54-48.png
Screenshot_from_2024-06-11_04-54-48.png
Attachment 13: Screenshot_from_2024-06-11_04-54-59.png
Screenshot_from_2024-06-11_04-54-59.png
Attachment 14: Screenshot_from_2024-06-11_04-55-28.png
Screenshot_from_2024-06-11_04-55-28.png
Attachment 15: Screenshot_from_2024-06-11_04-55-33.png
Screenshot_from_2024-06-11_04-55-33.png
Attachment 16: Screenshot_from_2024-06-11_05-35-24.png
Screenshot_from_2024-06-11_05-35-24.png
Attachment 17: Screenshot_from_2024-06-11_05-35-27.png
Screenshot_from_2024-06-11_05-35-27.png
Attachment 18: Screenshot_from_2024-06-11_05-35-33.png
Screenshot_from_2024-06-11_05-35-33.png
  613   Mon Jun 10 18:22:47 2024 JMElectron cooler calculations

Ion beam energy is calculated using Beamcalc (https://web-docs.gsi.de/~lestinsk/beamcalc/electron.php)

Electron current is set to 4.3 mA and can be taken from Grafana

Alpha factor is 50

Ion energy is then adjusted using the cooler voltage

The cooler voltage inserted into BeamCalc is taken from the FC20 voltage divider. However an offset from the space charge correction is required!

This correction is determined experimentally by changing the electron current and finding the cooler voltage which maintains the optimum cooling of the beam.

Attachment 1 shows this plot for 15N1+ at  ~0.426 MeV/u

Attachment 2 shows this same plot in comparison with BeamCalc

To obtain the correct energy we need to subtract 2V from the voltage used in BeamCalc

An example is shown in attachment 3

Here to obtain an ion energy of 426.6 keV/u requires a cooler voltage of 246.7 V. But, due to the space charge correction this value should be 244.7 V.

The cooler voltage setpoint should therefore be chosen so that the FC20 voltage divider reads 244.7 V.

The setpoint needs to be ~0.55 V higher than the desired value i.e setpoint of 245.25 V to obtain the required 244.7 V (at least this beamtime it does)

 

Attachment 1: space-charge-2.png
space-charge-2.png
Attachment 2: space-charge-1.png
space-charge-1.png
Attachment 3: BeamCalc.png
BeamCalc.png
  612   Mon Jun 10 14:26:52 2024 EH, AKRun 58

15:30

Merger OK, Tape server OK

ADC data item statistics OK

FEE temperatures look OK

System checks OK

20:20

Merger OK, Tape server OK

ADC data item statistics OK

FEE temperatures look OK

System checks OK

Attachment 1: Screenshot_from_2024-06-10_15-24-15.png
Screenshot_from_2024-06-10_15-24-15.png
Attachment 2: Screenshot_from_2024-06-10_15-22-30.png
Screenshot_from_2024-06-10_15-22-30.png
Attachment 3: Screenshot_from_2024-06-10_15-23-28.png
Screenshot_from_2024-06-10_15-23-28.png
Attachment 4: Screenshot_from_2024-06-10_15-23-20.png
Screenshot_from_2024-06-10_15-23-20.png
Attachment 5: Screenshot_from_2024-06-10_20-21-55.png
Screenshot_from_2024-06-10_20-21-55.png
Attachment 6: Screenshot_from_2024-06-10_20-21-48.png
Screenshot_from_2024-06-10_20-21-48.png
Attachment 7: Screenshot_from_2024-06-10_20-20-11.png
Screenshot_from_2024-06-10_20-20-11.png
Attachment 8: Screenshot_from_2024-06-10_20-21-14.png
Screenshot_from_2024-06-10_20-21-14.png
  611   Mon Jun 10 12:49:56 2024 RSS, DB, CB, JM R58 Space Charge Test
13:53

      DSSSD bias & leakage current OK - attachment 1

      FEE64 temperatures OK - attachment 2

      ADC data item stats - attachment 3

      per FEE64 Rate spectra - attachment 4
      
      Beam intensity - ~ 5.0E7 pps

      Target density - off

         

Attachment 1: Screenshot_from_2024-06-10_13-52-45.png
Screenshot_from_2024-06-10_13-52-45.png
Attachment 2: Screenshot_from_2024-06-10_13-53-04.png
Screenshot_from_2024-06-10_13-53-04.png
Attachment 3: Screenshot_from_2024-06-10_13-53-17.png
Screenshot_from_2024-06-10_13-53-17.png
Attachment 4: Screenshot_from_2024-06-10_13-53-28.png
Screenshot_from_2024-06-10_13-53-28.png
  610   Mon Jun 10 07:40:47 2024 RSS, DBBeam

08:26: There is no beam and Gleb is informed.

08:46: Beam is back.

  609   Mon Jun 10 05:24:49 2024 TD, RSS, DBR57
06:45 New data file G22-201/R57

      Right arm at 64.60 mm

      Left arm IN at 36.54 mm

      horizontal beam position 8.75 mm

      vertical beam position -5mm
   
      Changed electron cooler set point to 244.4 V

      Read value 243.851 V

      Collector current 4.31A


      DSSSD bias & leakage current OK - attachment 1

      FEE64 temperatures OK - attachment 2

      ADC data item stats - attachment 3

      per FEE64 Rate spectra - attachment 4
      
      Beam intensity - ~ 5.7E7 pps

      Target density - 2.91E11 atoms/cm2

8:35


      DSSSD bias & leakage current OK - attachment 5

      FEE64 temperatures OK - attachment 6

      ADC data item stats - attachment 7

      per FEE64 Rate spectra - attachment 8
      
      Beam intensity - ~ 4.8E7 pps

      Target density - 2.86E11 atoms/cm2
      
      There is no beam since 8:26 and Gleb is informed. 08:46: Beam is back.

10:49


      DSSSD bias & leakage current OK - attachment 9

      FEE64 temperatures OK - attachment 10

      ADC data item stats - attachment 11

      per FEE64 Rate spectra - attachment 12
      
      Beam intensity - ~ 4.0E7 pps

      Target density - 3.66E11 atoms/cm2


13:18


      DSSSD bias & leakage current OK - attachment 13

      FEE64 temperatures OK - attachment 14

      ADC data item stats - attachment 15

      per FEE64 Rate spectra - attachment 16
      
      Beam intensity - ~ 5.0E7 pps

      Target density - 5.67E11 atoms/cm2
      



     
Attachment 1: Screenshot_from_2024-06-10_06-41-37.png
Screenshot_from_2024-06-10_06-41-37.png
Attachment 2: Screenshot_from_2024-06-10_06-42-40.png
Screenshot_from_2024-06-10_06-42-40.png
Attachment 3: Screenshot_from_2024-06-10_06-43-15.png
Screenshot_from_2024-06-10_06-43-15.png
Attachment 4: Screenshot_from_2024-06-10_06-44-38.png
Screenshot_from_2024-06-10_06-44-38.png
Attachment 5: Screenshot_from_2024-06-10_08-34-30.png
Screenshot_from_2024-06-10_08-34-30.png
Attachment 6: Screenshot_from_2024-06-10_08-34-59.png
Screenshot_from_2024-06-10_08-34-59.png
Attachment 7: Screenshot_from_2024-06-10_08-35-23.png
Screenshot_from_2024-06-10_08-35-23.png
Attachment 8: Screenshot_from_2024-06-10_08-35-42.png
Screenshot_from_2024-06-10_08-35-42.png
Attachment 9: Screenshot_from_2024-06-10_10-46-13.png
Screenshot_from_2024-06-10_10-46-13.png
Attachment 10: Screenshot_from_2024-06-10_10-46-31.png
Screenshot_from_2024-06-10_10-46-31.png
Attachment 11: Screenshot_from_2024-06-10_10-46-42.png
Screenshot_from_2024-06-10_10-46-42.png
Attachment 12: Screenshot_from_2024-06-10_10-46-52.png
Screenshot_from_2024-06-10_10-46-52.png
Attachment 13: Screenshot_from_2024-06-10_13-16-00.png
Screenshot_from_2024-06-10_13-16-00.png
Attachment 14: Screenshot_from_2024-06-10_13-16-19.png
Screenshot_from_2024-06-10_13-16-19.png
Attachment 15: Screenshot_from_2024-06-10_13-16-31.png
Screenshot_from_2024-06-10_13-16-31.png
Attachment 16: Screenshot_from_2024-06-10_13-16-40.png
Screenshot_from_2024-06-10_13-16-40.png
  608   Sun Jun 9 22:14:23 2024 TD, PM, AGRun 56 - Data taking
23:15 New run 56 with

      Right arm at 64.60 mm

      Left arm IN at 36.54 mm

      MBS working - hopefully

      Target 4E11

      Injection ca. 4E7

      horizontal beam position 8.75 mm

      vertical beam position&nbsp; -5mm
   
      Changed electron cooler set point to 244.3 V

      Read value 243.757 V

      Collector current 4.31A

      Rates OK. Temps OK

00.30 DAQ continues file G22-201/R56_9

      DSSSD bias & leakage current OK - attachment 1

      FEE64 temperatures OK - attachment 2

      ADC data item stats - attachment 3

      per FEE64 Rate spectra - attachment 4

      Merger etc - attachment 5

      DSSSD bias & leakage current log, CRYRING hall ambient temperature log - attachments 6-7

01.30 Photographs of current CRYRING controls - attachments 8-14


01.30 DAQ continues file G22-201/R56_14

      DSSSD bias & leakage current OK - attachment 15

      FEE64 temperatures OK - attachment 16

      ADC data item stats - attachment 17

      per FEE64 Rate spectra
       aida16 asics # 1 & # 2 HEC data - attachment 18
       aida16 ASIC load check x2 restores normal operation - attachment 19

      target 5e+11 stable, beam 4e+7 stable 

02.45 beam off - reported to HKR

      DAQ continues file G22-201/R56_16

03.00 beam back

      beam 4e+7 stable, target 5e+11 stable

      DSSSD bias & leakage current OK - attachment 20

      FEE64 temperatures OK - attachment 21

      ADC data item stats - attachment 22

      per FEE64 Rate spectra - attachment 23


04.20 DAQ continues file G22-201/R56_21

      DSSSD bias & leakage current OK - attachment 24

      FEE64 temperatures OK - attachment 25

      ADC data item stats - attachment 26

      per FEE64 Rate spectra - attachments 27

      Cycle L DSSSDs per https://elog.ph.ed.ac.uk/CARME/603


06.04 DAQ continues file G22-201/R56_21

      DSSSD bias & leakage current OK - attachment 28

      FEE64 temperatures OK - attachment 29

      ADC data item stats - attachment 30

      per FEE64 Rate spectra - attachments 31

      beam 4e+7 stable, target 4e+11 stable
Attachment 1: Screenshot_from_2024-06-10_00-23-24.png
Screenshot_from_2024-06-10_00-23-24.png
Attachment 2: Screenshot_from_2024-06-10_00-24-07.png
Screenshot_from_2024-06-10_00-24-07.png
Attachment 3: Screenshot_from_2024-06-10_00-24-20.png
Screenshot_from_2024-06-10_00-24-20.png
Attachment 4: Screenshot_from_2024-06-10_00-25-10.png
Screenshot_from_2024-06-10_00-25-10.png
Attachment 5: Screenshot_from_2024-06-10_00-25-38.png
Screenshot_from_2024-06-10_00-25-38.png
Attachment 6: Screenshot_from_2024-06-10_00-42-25.png
Screenshot_from_2024-06-10_00-42-25.png
Attachment 7: Capture.PNG
Capture.PNG
Attachment 8: 20240610_003428.jpg
20240610_003428.jpg
Attachment 9: 20240610_003436.jpg
20240610_003436.jpg
Attachment 10: 20240610_003446.jpg
20240610_003446.jpg
Attachment 11: 20240610_003458.jpg
20240610_003458.jpg
Attachment 12: 20240610_003510.jpg
20240610_003510.jpg
Attachment 13: 20240610_003518.jpg
20240610_003518.jpg
Attachment 14: 20240610_003529.jpg
20240610_003529.jpg
Attachment 15: Screenshot_from_2024-06-10_01-32-42.png
Screenshot_from_2024-06-10_01-32-42.png
Attachment 16: Screenshot_from_2024-06-10_01-32-58.png
Screenshot_from_2024-06-10_01-32-58.png
Attachment 17: Screenshot_from_2024-06-10_01-33-07.png
Screenshot_from_2024-06-10_01-33-07.png
Attachment 18: Screenshot_from_2024-06-10_01-33-35.png
Screenshot_from_2024-06-10_01-33-35.png
Attachment 19: Screenshot_from_2024-06-10_01-34-12.png
Screenshot_from_2024-06-10_01-34-12.png
Attachment 20: Screenshot_from_2024-06-10_02-55-21.png
Screenshot_from_2024-06-10_02-55-21.png
Attachment 21: Screenshot_from_2024-06-10_02-55-58.png
Screenshot_from_2024-06-10_02-55-58.png
Attachment 22: Screenshot_from_2024-06-10_02-56-13.png
Screenshot_from_2024-06-10_02-56-13.png
Attachment 23: Screenshot_from_2024-06-10_02-56-22.png
Screenshot_from_2024-06-10_02-56-22.png
Attachment 24: Screenshot_from_2024-06-10_04-22-41.png
Screenshot_from_2024-06-10_04-22-41.png
Attachment 25: Screenshot_from_2024-06-10_04-22-55.png
Screenshot_from_2024-06-10_04-22-55.png
Attachment 26: Screenshot_from_2024-06-10_04-23-24.png
Screenshot_from_2024-06-10_04-23-24.png
Attachment 27: Screenshot_from_2024-06-10_04-23-31.png
Screenshot_from_2024-06-10_04-23-31.png
Attachment 28: Screenshot_from_2024-06-10_06-04-17.png
Screenshot_from_2024-06-10_06-04-17.png
Attachment 29: Screenshot_from_2024-06-10_06-04-32.png
Screenshot_from_2024-06-10_06-04-32.png
Attachment 30: Screenshot_from_2024-06-10_06-04-44.png
Screenshot_from_2024-06-10_06-04-44.png
Attachment 31: Screenshot_from_2024-06-10_06-04-52.png
Screenshot_from_2024-06-10_06-04-52.png
  607   Sun Jun 9 13:45:24 2024 CB PMH AGRun 55 - Data taking

14:45

New run 55 with:

Right arm at 64.60 mm

Left arm IN at 36.54 mm

MBS working - hopefully

Target 7E11

Injection ca. 5E7

horizontal beam position 8.75 mm

vertical beam position  -5mm

Changed electron cooler set point to 244.2 V

Read value 243.66 V

Collector current 4.31A

Rates OK. Temps OK.

 

15:00

Detector 0 was noisy. Shook arm. Fixed it.

 

16:15

Detector 0 was noisy. Shook arm. Fixed it.

There is no beam in CRYRING since 16:05.

 

16:25

The beam is back but it is still a bit unstable and the operators are still trying optimize it.

 

17:20

We just did standard detector/electronics checks. Everthing looks OK.

 

17:50

Take screenshots.

 

20:50

Detector 0 was noisy. Shook arm. Fixed it.

But now detector 2 seems to have lost the Rutherford peak.

Screenshots taken and attached.

 

21:05

We did "Check ASIC Control" following Carlo's advise. This fixed the detector 2, but the detector 0 got noisy again. Another movement of the detector fixed the noise again.

 

22:15

Detector 0 was noisy. Shook arm. Fixed it.

 

23:05

This run has finished. The cooler voltage will be increased by 0.1V for the next run.

The latest spectra and detector/DAQ status are in the attachment.

Attachment 1: Screenshot_from_2024-06-09_14-46-33.png
Screenshot_from_2024-06-09_14-46-33.png
Attachment 2: Screenshot_from_2024-06-09_17-48-00.png
Screenshot_from_2024-06-09_17-48-00.png
Attachment 3: Screenshot_from_2024-06-09_17-48-34.png
Screenshot_from_2024-06-09_17-48-34.png
Attachment 4: Screenshot_from_2024-06-09_17-48-51.png
Screenshot_from_2024-06-09_17-48-51.png
Attachment 5: Screenshot_from_2024-06-09_17-51-55.png
Screenshot_from_2024-06-09_17-51-55.png
Attachment 6: Screenshot_from_2024-06-09_17-53-42.png
Screenshot_from_2024-06-09_17-53-42.png
Attachment 7: Screenshot_from_2024-06-09_20-54-38.png
Screenshot_from_2024-06-09_20-54-38.png
Attachment 8: Screenshot_from_2024-06-09_20-54-59.png
Screenshot_from_2024-06-09_20-54-59.png
Attachment 9: Screenshot_from_2024-06-09_20-55-34.png
Screenshot_from_2024-06-09_20-55-34.png
Attachment 10: Screenshot_from_2024-06-09_20-56-21.png
Screenshot_from_2024-06-09_20-56-21.png
Attachment 11: Screenshot_from_2024-06-09_20-56-35.png
Screenshot_from_2024-06-09_20-56-35.png
Attachment 12: Screenshot_from_2024-06-09_20-56-41.png
Screenshot_from_2024-06-09_20-56-41.png
Attachment 13: Screenshot_from_2024-06-09_23-09-38.png
Screenshot_from_2024-06-09_23-09-38.png
Attachment 14: Screenshot_from_2024-06-09_23-09-53.png
Screenshot_from_2024-06-09_23-09-53.png
Attachment 15: Screenshot_from_2024-06-09_23-10-26.png
Screenshot_from_2024-06-09_23-10-26.png
Attachment 16: Screenshot_from_2024-06-09_23-10-37.png
Screenshot_from_2024-06-09_23-10-37.png
Attachment 17: Screenshot_from_2024-06-09_23-10-43.png
Screenshot_from_2024-06-09_23-10-43.png
Attachment 18: Screenshot_from_2024-06-09_23-10-54.png
Screenshot_from_2024-06-09_23-10-54.png
  606   Sun Jun 9 09:45:27 2024 JM, CBBeam-Target Overlap

Target density this morning had stabalised at 7e11 at/cm2

This is sufficient to test the beam/target overlap using the lifetime of the beam in the ring.

Previous attempt at this method was done at 1.5e11 at/cm2 where lifetime from residual gas and target are essentially the same.

Beam set to bunched. Beam position varied by 1mm and the number of particles in the ring taken from Grafana plots. Particles at injection compared with at dump.

Significantly more beam loss at > 8 mm beam position with highest value at 8.75 mm. This is the maximum we can move the beam.

Test repeated with no target. Same beam loss rate for all positions. We see a clear difference with the target only at positions > 8mm.

Test repeated with original electron cooler voltage Gleb tuned with last night. Same outcome -> cooler voltage has no significant impact on overlap (for 0.1V at least)

Overlap using count rate over 7.5 minutes also tested. DSSD 1 (top right) used as gauge as no beam stripe is seen in this detector. 

At horizontal position 4 mm (previous estimation of overlap) we saw a small increase in counts around 4.7 MeV with the target off compared to target on (76 vs 57)

At horizontal position 8.75 mm we saw a large decrease in counts around 4.7 MeV with target off compared to target on (95 vs 290)

It appears clear the overlap is at 8.75 mm. Future runs will use this as horizontal beam position.

Attachment 1: Beam-overlap.png
Beam-overlap.png
  605   Sun Jun 9 08:30:49 2024 JM, CBRun 54 - beam on target

09:30

New run 54 with:

Electron cooler setpoint 244.1 V

Electron cooler read 243.55 V

Collector current 4.3 uA

Right arm at 64.60 mm

Left arm IN at 36.54 mm

MBS working - hopefully

Target 7E11

Injection ca. 5E7

horizontal beam position changed to 8.75 mm

vertical beam position still -5mm

10:30

No beam

ADC stats increased to 250k again same time as beam lost

Detector shake resolved problem

System checks OK

10:35

Beam back

12:17

All system checks OK

12:29

DSSD top left noisy again. Detector shake resolves issuie

 

14:02

Running smoothly. Temps OK. Stats OK, but write to disk rate x10 usual for unclear reasons.

Target mostly stable. Beam OK, missing some injections occasionally.

 

14:25

Detector 0 noisy. Shook. Fixed it.

 

Attachment 1: Screenshot_from_2024-06-09_14-00-21.png
Screenshot_from_2024-06-09_14-00-21.png
Attachment 2: Screenshot_from_2024-06-09_14-00-30.png
Screenshot_from_2024-06-09_14-00-30.png
Attachment 3: Screenshot_from_2024-06-09_14-00-41.png
Screenshot_from_2024-06-09_14-00-41.png
Attachment 4: Screenshot_from_2024-06-09_14-01-26.png
Screenshot_from_2024-06-09_14-01-26.png
Attachment 5: Screenshot_from_2024-06-09_14-01-40.png
Screenshot_from_2024-06-09_14-01-40.png
Attachment 6: Screenshot_from_2024-06-09_14-01-49.png
Screenshot_from_2024-06-09_14-01-49.png
Attachment 7: Screenshot_from_2024-06-09_14-01-56.png
Screenshot_from_2024-06-09_14-01-56.png
  604   Sat Jun 8 20:32:17 2024 LB, JMRun 52 - beam on target

New run 52 with:

Electron cooler setpoint 244.2 V

Electron cooler read 243.66 V

Collector current 4.3 uA

Right arm at 64.60 mm

Left arm IN at 36.54 mm

Target 2E11 fluctuating

Injection ca. 5E7

 

22;00

Detectors moved out while HKR plays with the beam injection to improve its stability

22;56

All system checks good. Detectors still out. Stats in attachment 1

23:08

HKR still has to call, but I have put detectors back in since they are only operating on injection.

Noticed that from time to time the target density still sparks to E12, and this corresponds to when the temperature drops.

23:42

HKR called back, they found the problem and fixed it. Beam seems more stable, with injections at ca. 5E7

23:50

DSSD0 high stats, cycled the motors to clear them.

00:18

All system checks are fine. Attachment 2-3 with stats and a spectrum

00:38

Spikes in the target got more frequent. Increased temperature to 30.5K and monitor if it helps stabilizing it.

01:52

All checks OK. Target behaving, density 5-8E11.

Attachments 4-5

03:10

System checks OK.

Target consistent

Some beam outages which were quickly fixed by HKR

05:03

System checks OK.

Target consistent

07:00

Lost beam. Detectors moved out

Will likely end run once beam is back and change energy. Do not use data past 7 am

All system checks OK

Attachment 1: Screenshot_from_2024-06-08_22-56-26.png
Screenshot_from_2024-06-08_22-56-26.png
Attachment 2: Screenshot_from_2024-06-09_00-17-44.png
Screenshot_from_2024-06-09_00-17-44.png
Attachment 3: Screenshot_from_2024-06-09_00-18-46.png
Screenshot_from_2024-06-09_00-18-46.png
Attachment 4: Screenshot_from_2024-06-09_01-50-54.png
Screenshot_from_2024-06-09_01-50-54.png
Attachment 5: Screenshot_from_2024-06-09_01-51-27.png
Screenshot_from_2024-06-09_01-51-27.png
  603   Sat Jun 8 17:46:12 2024 TD[How To] Cycle Left DSSSDs
Interactive Motion Program [workspace 6, display 2]

c                         stop current command(s)
l s by -20 mm             confirm y at prompt
l s by 20 mm              confirm y at prompt
read FakeMotion.dat       confirm y at prompt
  602   Sat Jun 8 14:49:14 2024 TDR51 - contd
15.40 DSSSD bias & leakage current OK - attachments 1-2
       log shows variations due to ambient temperature changes in CRYRING hall

      FEE64 temperatures OK - attachment 3

      WR timestamps OK - attachment 4

      System wide checks Ok *except* global clock (aida16) - attachments 5-6

      ADC data item stats OK - attachment 7

      per FEE64 Rate spectra - attachment 8

      per FEE64 1.8.L spectra - attachments 9-10
       have not been cleared in some time probably - will zero histograms & stats for later review

     per FEE64 1.8.W spectra - 20us FSR - attachments 11-12

     Merger, TapeServer etc - attachments 13-14
      data file G22-222/R51_15

     Motion control - attachment 15

16.01 all histograms & stats zero'd

16.43 online per DSSSD energy spectra - attachment 16

      photographs of CRYRING beam & target diagnostics - attachments 17-22

18.45 Left DSSSDs cycled to clear high rates (c. 250k/FEE64) aida01-aida04 per https://elog.ph.ed.ac.uk/CARME/603

18.49 DAQ continues data file G22-201/R51_27

      DSSSD bias & leakage current OK - attachment 23

      FEE64 temperatures OK - attachment 24

      ADC data item stats OK - attachment 25

      per FEE64 Rate spectra - attachment 26

      online per DSSSD energy spectra - attachment 27

19.00 Beam less stable but most cycles 5e+7 or higher
      Target peaking c. 1e12 dropping to c. 6e11

19.09 CRYRING hall temperature increasing, DSSSD leakage current increasing - attachments 28-29


20.06 Beam less stable most cycles 1-2e+7 or higher
      Target c. 1-2e12

20.15 JM/CB advise changing Lakeshore 336 Temp Controller set point from 29k to 29.5K - done
      Average pressure decreases but rapid variations still observed - see attachment 30


21.00 contacted HKR to improve beam stability - they were alreday working on it but did not know what was causing the problem - will contact on-call CRYRING expert

      cycled L DSSSD to reset rates
      
21,10 beam 0ff - DSSDs OUT

      Increased target temperature to 30K to solve instability that developed again up to E12 density range.


      


 
Attachment 1: Screenshot_from_2024-06-08_15-40-56.png
Screenshot_from_2024-06-08_15-40-56.png
Attachment 2: Screenshot_from_2024-06-08_15-41-11.png
Screenshot_from_2024-06-08_15-41-11.png
Attachment 3: Screenshot_from_2024-06-08_15-41-44.png
Screenshot_from_2024-06-08_15-41-44.png
Attachment 4: Screenshot_from_2024-06-08_15-42-10.png
Screenshot_from_2024-06-08_15-42-10.png
Attachment 5: Screenshot_from_2024-06-08_15-42-18.png
Screenshot_from_2024-06-08_15-42-18.png
Attachment 6: Screenshot_from_2024-06-08_15-42-26.png
Screenshot_from_2024-06-08_15-42-26.png
Attachment 7: Screenshot_from_2024-06-08_15-42-47.png
Screenshot_from_2024-06-08_15-42-47.png
Attachment 8: Screenshot_from_2024-06-08_15-47-40.png
Screenshot_from_2024-06-08_15-47-40.png
Attachment 9: Screenshot_from_2024-06-08_15-46-08.png
Screenshot_from_2024-06-08_15-46-08.png
Attachment 10: Screenshot_from_2024-06-08_15-46-57.png
Screenshot_from_2024-06-08_15-46-57.png
Attachment 11: Screenshot_from_2024-06-08_15-44-09.png
Screenshot_from_2024-06-08_15-44-09.png
Attachment 12: Screenshot_from_2024-06-08_15-45-20.png
Screenshot_from_2024-06-08_15-45-20.png
Attachment 13: Screenshot_from_2024-06-08_15-48-34.png
Screenshot_from_2024-06-08_15-48-34.png
Attachment 14: Screenshot_from_2024-06-08_15-48-49.png
Screenshot_from_2024-06-08_15-48-49.png
Attachment 15: Screenshot_from_2024-06-08_15-48-57.png
Screenshot_from_2024-06-08_15-48-57.png
Attachment 16: Screenshot_from_2024-06-08_16-42-56.png
Screenshot_from_2024-06-08_16-42-56.png
Attachment 17: 20240608_160331.jpg
20240608_160331.jpg
Attachment 18: 20240608_160545.jpg
20240608_160545.jpg
Attachment 19: 20240608_160526.jpg
20240608_160526.jpg
Attachment 20: 20240608_160516.jpg
20240608_160516.jpg
Attachment 21: 20240608_160508.jpg
20240608_160508.jpg
Attachment 22: 20240608_160536.jpg
20240608_160536.jpg
Attachment 23: Screenshot_from_2024-06-08_18-49-51.png
Screenshot_from_2024-06-08_18-49-51.png
Attachment 24: Screenshot_from_2024-06-08_18-50-10.png
Screenshot_from_2024-06-08_18-50-10.png
Attachment 25: Screenshot_from_2024-06-08_18-50-29.png
Screenshot_from_2024-06-08_18-50-29.png
Attachment 26: Screenshot_from_2024-06-08_18-50-48.png
Screenshot_from_2024-06-08_18-50-48.png
Attachment 27: Screenshot_from_2024-06-08_18-54-00.png
Screenshot_from_2024-06-08_18-54-00.png
Attachment 28: Capture.PNG
Capture.PNG
Attachment 29: Screenshot_from_2024-06-08_19-08-58.png
Screenshot_from_2024-06-08_19-08-58.png
Attachment 30: 20240608_202235.jpg
20240608_202235.jpg
  601   Sat Jun 8 10:13:52 2024 CB, JM, TDRun 50/51 - data taking

11:13

ML enabled ring signals remotely.

Used FakeMotion.dat (read FakeMotion.dat from InteractiveMotion.x).

Started run 50.

 

Cooler voltage moved up by 0.1 V. Did not refocus.

Setpoint = 244.4 V

Read = 243.85 V

Collector current = 4.3 mA

 

Target thickness 3.4E11 atoms/cm2 now stable

Injection 1-4E7 ions

 

Realised motors were not in.

Moved detectors in (36.54 / 64.60 mm)

Started run 51. Ignore run 50.

 

11:46

Realised the left half of DSSD BL was not counting. Checkloaded. Fixed it.

 

12:00

Fire alarm? Stopped later. EM thinks it is a test.

 

13:00

Run continues.

Rates OK. Merger OK. Writing to disk back to 200 kB/s. Nothing changed, reasons unclear.

FakeMotion.dat continues

System wide checks as before

Temperatures OK

 

13:36

Beam lost. Called HKR. Detectors out. Run continues. FakeMotion stopped

14:09

Beam is back. Detectors back in. FakeMotion.dat restarted.

 

15.10 High rates observed aida01-auida04 ( c. 250k/FEE64)

          Cycled Left Servo

         Rates normal

 

Attachment 1: Screenshot_from_2024-06-08_12-58-08.png
Screenshot_from_2024-06-08_12-58-08.png
Attachment 2: Screenshot_from_2024-06-08_12-58-19.png
Screenshot_from_2024-06-08_12-58-19.png
Attachment 3: Screenshot_from_2024-06-08_12-58-36.png
Screenshot_from_2024-06-08_12-58-36.png
Attachment 4: Screenshot_from_2024-06-08_12-58-53.png
Screenshot_from_2024-06-08_12-58-53.png
Attachment 5: Screenshot_from_2024-06-08_12-59-18.png
Screenshot_from_2024-06-08_12-59-18.png
Attachment 6: Screenshot_from_2024-06-08_12-59-26.png
Screenshot_from_2024-06-08_12-59-26.png
Attachment 7: Screenshot_from_2024-06-08_12-59-38.png
Screenshot_from_2024-06-08_12-59-38.png
Attachment 8: Screenshot_from_2024-06-08_12-59-50.png
Screenshot_from_2024-06-08_12-59-50.png
  600   Sat Jun 8 06:13:21 2024 CBRun 47/48/49 - target issues & tests

07:00

After putting the target back, we observe strong periodic (5 s) fluctuations going from 3E11 to 5E12. The reason is unclear. One may warm up the target to solve this, but will ask Nikos later instead.

The target may be a bit too thick for our purposes, and the luminosity will vary greatly depending on injection vs. target fluctuations timing.

Rates OK. Temperatures OK. Merger OK.

System wide checks as before.

 

07:30

Moved detectors fully out to check elastic and signal rate in this configuration.

Half of DSSD BR was missing. Checkloaded. Solved.

Moved to run 48. Target still unstable.

 

08:23

Beam issues. Called HKR. Target out.

 

09:00

Still no beam. Increased target temperature to 29K from 28.5K. It seems the fluctuations disappeared.

Rates as before.

 

10:22

Beam might be back but people are checking the steerer. Target in. Started run 49. Detectors still out.

 

10:13

Some beam. HKR is still working on it.

Data rate up to 2 MB/s. Merger events merged matches. However Stats do not seem to have increased. This is the second time I observe this.

Attachment 1: Screenshot_from_2024-06-08_07-15-27.png
Screenshot_from_2024-06-08_07-15-27.png
Attachment 2: Screenshot_from_2024-06-08_07-15-33.png
Screenshot_from_2024-06-08_07-15-33.png
Attachment 3: Screenshot_from_2024-06-08_10-13-16.png
Screenshot_from_2024-06-08_10-13-16.png
Attachment 4: Screenshot_from_2024-06-08_10-13-23.png
Screenshot_from_2024-06-08_10-13-23.png
Attachment 5: Screenshot_from_2024-06-08_10-13-26.png
Screenshot_from_2024-06-08_10-13-26.png
  599   Sat Jun 8 00:14:39 2024 LBRun 46 - background with beam

RUN 46- started at 00:50

Horizontal beam position = 4.0 mm, vertical beam position = -5.0 mm.

Detectors are Left arm OUT = 36.64, right arm IN =  64.60 mm

Injection at ca. 6E7

Target density 6-7E9

-----

00:51

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

Attachments 1-2

01:19

Attachments 3-4 waveforms. DSSD0 stats fine

01:33

DSSD0 stats fine

Attachments 5-6 spectra and xy, but I am not sure at which time these are cause I am stupid and apparently the AIDAsort code was not running....

01:56

Restarted the AIDAsort code and noted that stats in DSSD0 were bad again. Another shake. This time no change to the beam on my side.

2:43

All checks fine. Attachments 7-10 (also spectra)

4:01

Notice stats on DSSD0 went bad again. Attachment 11. Gave detector a shake.

Now all system checks OK (Attachements 12-14)

4:19

Just noticed target is oscillating a bit, hitting 1E10. I moved it further in X and Y. Seems to not go below 6E9 in any case.

NOTE: target was on at x=-19899 y=500.

5:17

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

Attachments 15-18 (also waves)

5:53

Notice stats on DSSD0 went bad again. Gave detector a shake.

6:10

Stats ok in attachment 19, but

Noticed DSSD3 has started to get only a fate of 140 nstead of previous 5500is (similar to what DSSD2 has done earlier- see previous elog). Attachment 20 for the rates.

I have therefore made the Check ASIC control test from the ASIC control tab- attachment 21 with results.

DSSD3 back to normal operations - attachment 22

Attachment 1: Screenshot_from_2024-06-08_00-51-42.png
Screenshot_from_2024-06-08_00-51-42.png
Attachment 2: Screenshot_from_2024-06-08_00-51-53.png
Screenshot_from_2024-06-08_00-51-53.png
Attachment 3: Screenshot_from_2024-06-08_01-19-58.png
Screenshot_from_2024-06-08_01-19-58.png
Attachment 4: Screenshot_from_2024-06-08_01-21-07.png
Screenshot_from_2024-06-08_01-21-07.png
Attachment 5: Screenshot_from_2024-06-08_01-32-35.png
Screenshot_from_2024-06-08_01-32-35.png
Attachment 6: Screenshot_from_2024-06-08_01-33-28.png
Screenshot_from_2024-06-08_01-33-28.png
Attachment 7: Screenshot_from_2024-06-08_02-42-36.png
Screenshot_from_2024-06-08_02-42-36.png
Attachment 8: Screenshot_from_2024-06-08_02-43-02.png
Screenshot_from_2024-06-08_02-43-02.png
Attachment 9: Screenshot_from_2024-06-08_02-43-12.png
Screenshot_from_2024-06-08_02-43-12.png
Attachment 10: Screenshot_from_2024-06-08_02-43-35.png
Screenshot_from_2024-06-08_02-43-35.png
Attachment 11: Screenshot_from_2024-06-08_03-59-55.png
Screenshot_from_2024-06-08_03-59-55.png
Attachment 12: Screenshot_from_2024-06-08_04-03-02.png
Screenshot_from_2024-06-08_04-03-02.png
Attachment 13: Screenshot_from_2024-06-08_04-03-16.png
Screenshot_from_2024-06-08_04-03-16.png
Attachment 14: Screenshot_from_2024-06-08_04-03-45.png
Screenshot_from_2024-06-08_04-03-45.png
Attachment 15: Screenshot_from_2024-06-08_05-15-37.png
Screenshot_from_2024-06-08_05-15-37.png
Attachment 16: Screenshot_from_2024-06-08_05-15-48.png
Screenshot_from_2024-06-08_05-15-48.png
Attachment 17: Screenshot_from_2024-06-08_05-16-38.png
Screenshot_from_2024-06-08_05-16-38.png
Attachment 18: Screenshot_from_2024-06-08_05-17-07.png
Screenshot_from_2024-06-08_05-17-07.png
Attachment 19: Screenshot_from_2024-06-08_06-11-14.png
Screenshot_from_2024-06-08_06-11-14.png
Attachment 20: Screenshot_from_2024-06-08_06-11-24.png
Screenshot_from_2024-06-08_06-11-24.png
Attachment 21: Screenshot_from_2024-06-08_06-14-01.png
Screenshot_from_2024-06-08_06-14-01.png
Attachment 22: Screenshot_from_2024-06-08_06-14-14.png
Screenshot_from_2024-06-08_06-14-14.png
  598   Fri Jun 7 22:33:13 2024 LBRun 45 - tests on beam-target overlap

Tried to double check beam-target overlap by changing beam horizontal coordinate.

In the tablethe counts in the 4000-5200 keV region in each of the detectors are recorded. Counts were taken for a total time of 5 minutes per beam position. Attached are the xy plots.

I think the tests are inconclusive given that the beam in the first three points were not ideal, with many low intensity injections. Moreover, changing the beam often ends up causing the stats in DSSD0 to rise significantly (consequentely needing to shake the detector to go back to normal operation).

 

horizontal DSSD0 DSSD1 DSSD2 DSSD3 Comments
4.5 mm 2966 60 3826 1880 stats went up at the end of the run (attachment 1 is the xy plot, attachment 1 is the multiplicity showing the noise).
5 mm 3826 68 1945 1719 stats ok. Attachment 3
5.5 mm 2443 64 1 1352 stats ok. beam bad, called 2222, fixed it. Attachment 4
5.5 mm good beam 3587 57 6 2057 repeated measurment with good beam. Suspected something wrong with DSSD2
4.5 mm good beam 5132 64 4 2153 stats bad at the end, confirmed something wrong DSSD2, Jordan did some ASIC check load and made it go back to normal operation. Attachment 5.

After fixing DSSD2, statsin DSSD0 went up again. Stopped the test after shaking again the detector.

----

I have put back the beam at 4mm horizontal. I switched off the target by moving it to x=-10000--->x=-5000 (previosuly X=-19800). Denisity went down to 5E9. However, I decided under Jordan's suggestion to make another test and I put the target back to x=-19800: interestingly the density went up to 5E12 and start oscillating for a while between 5E12 and 5E11.

I did a 10 minute check of counts in the 4000-5200 keV region with this configuration, then switched off the target and did another 10 minutes test. Results:

horizontal 4 mm DSSD0 DSSD1 DSSD2 DSSD3
target on 9719 163 7892 5210
targte off 6579 108 6092 3496

 

Attachment 1: Screenshot_from_2024-06-07_23-11-06.png
Screenshot_from_2024-06-07_23-11-06.png
Attachment 2: Screenshot_from_2024-06-07_23-12-17.png
Screenshot_from_2024-06-07_23-12-17.png
Attachment 3: Screenshot_from_2024-06-07_23-32-22.png
Screenshot_from_2024-06-07_23-32-22.png
Attachment 4: Screenshot_from_2024-06-07_23-43-43.png
Screenshot_from_2024-06-07_23-43-43.png
Attachment 5: Screenshot_from_2024-06-07_23-50-54.png
Screenshot_from_2024-06-07_23-50-54.png
  597   Fri Jun 7 15:53:35 2024 TDDSSD leakage current increase - temperature increase in CRYRING hall

Re-increase in DSSSD leakage current c. 14-15.00 6.6.24 

There was a significant temperature spike in the CRYRING hall on the afternoon of 6.6.24 which has, partly, recovered - see attachment 1

Quote:

Started run 39.

 

We have the 15N+ beam with 426 keV/nucleon. We see a peak in detector top left (mostly) and bottom right, at 4600 keV. We think it is the beam (nominally 6400 keV) after going through silicon dead layer, which however seems to be thicker than expected (expected 0.7 um, but given 1.4 MeV/um energy loss it seems to be > 1 um).

To understand what was going on we first played with detector and beam position.

At 17:28 detectors are fully out, while we are playing with the electron cooler to try to remove this beam halo hitting the detector.

 

Detector DSSD0 top left periodically seems to get noisier (statistics up to >2E5) and we need to move the detector in and out to go back to normal operations.

 

------------------

The beam halo in top left detector, as the detector gets moved in, gets elongated in the y direction instead of the x direction, as expected and previosuly observed.

Run 40 started.

18:36

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

Attachments 1-4

19:43

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

Attachments 5-8

21:15

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

Attachments 9-11

 

00;25

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

-----------------------------------

Movements and observation in xy plot:

- Move right servo in by 20 mm and we see the halo in the top right moving accordingly (attachment 12)

-Move right pneumatic in, saw big increase on bottom right (attachment 13)

-moved beam position down vertical -6 mm and we see reduced beam spot on top left and more on bottom left

-----------------------------------

Beam - target overlap:

We have yet to obtain an overlap

We cannot do this using the beam lifetime as from beam calc the lifetime from the ring is comparable to the lifetime with a target of 1e11 at/cm2

We have tried using Rutherford. However, we will have to move the detectors IN further as the Rutherford peaks at ~3.8 deg.

Possibly we may require other overlap detections from the ring?

I also think the beam requires some optimisation, electron cooler voltage seems to be fluctuating a lot and we see some features on the detectors when moving the beam / detectors as mentioned above.

 

02:25

The beam intensity has decreased signficantly (by about a factor of five) since about an hour. Operators tried different things, also consulted with Frank Herfurth, but they could not solve the problem.

There is no device which indicates malfunction.

So, during the rest of the night shift we will continue searching for the overlap (of the beam with the target) with the available beam intensity.

 

03:30

The beam intensity has improved without any aparent intervention.

We have started a new search for the overlap with 20 minute runs per horiznotal beam position.

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

 

03:40

After changing the horizontal beam position from 0 mm to 1 mm, the bottom right detector lost the Rutherford peak and its position spectrum looks to be half empty.

Moving it OUT and IN again did not change the situation.

There is no reason to think that this behavior has been caused by moving the beam by just 1 mm.

 

04:10

The beam intensity has dropped again without any apparent reason. So, at the moment we can not continue search for the overlap.

The bottom right detector seems to be counting normally again (also without any apparent reason). 

 

06:45

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

 

07:18

Lost beam again. Gleb is looking for it. Moved right detector out temporarily.

System wide checks as before.

Merger OK, Tape Server OK.

Temperatures OK.

Rates currently OK.

Leakage current decreasing.

 

10:10

Beam is back. Target is unusually thick at 3-5E11

Checking beam overlap.

Merger & tape server OK.

Temps OK.

Rates OK, a bit higher in 01 and 04.

 

Attachment 1: Capture.PNG
Capture.PNG
  596   Fri Jun 7 14:09:04 2024 JM, CB, DR, CN, LBRun 44 - 15N Data Taking

15:35

Electron cooler voltage set point = 244.3 V

Electron cooler voltage readout = 243.75 V

Collector current = 4.29 mA

All system checks OK

Cycle time = 6 s

Detector positions: left = 36.54 mm, right = 64.60 mm

Target density = 2.5e11

Injection = 3 - 4 e7 pps

15:50

Injection intensity dropped to 1e7 pps

15:54

Injection intensity back to ~4e7 pps

17:30

All system checks OK

Target at 3.7e11

Beam dropped to 1e7 pps

17:34

Beam back to 3-4 e7 pps

19:10

Beam started varying between 1e7 and 4e7. The operators briefly closed the Faraday cup (~30 seconds), the beam is now back to ~5e7

19:25

All system checks OK

Target at 2.6e11

Beam at 5e7

21:25

All system checks OK

Target at 2.7e11

Beam at 4-5 e7

 

 

  595   Fri Jun 7 10:05:40 2024 CBRun 41/42/43 - 15N data taking start

11:05

Confirmed overlap is at 4 mm under new settings.

For unclear reasons, the bottom-left detector is not seeing the halo anymore since around 10:00. Rutherford still visible and if the left arm is moved in the counts at 5 MeV become apparent.

l s by 20 mm, r p in -> 36.5 mm /  64.6 mm

Target thickness 3E11 atoms/cm2. Beam intensity few E7 at injection. Coasting beam mode.

 

Data rate x10 compared with earlier. Stats as before - the reason is unclear.

Temps & Merger OK. We should be able to tolerate these rates.

Moved detectors in/out to make sure we were not clipping the beam. A few seconds of data may be present in R41_9

 

11:30

Checkloaded. Detector 2 is now counting more.

Moved to run 42. No other changes.

 

13:00

Realised cooler voltage was being read via fault g35 divider. Beam stopped

 

15:00

Switched to F20. Beam back on. Moved to run 43

Electron cooler voltage = 243.155 V

Collector current = 4.29 mA

Beam energy was slightly off, will try again

Attachment 1: Screenshot_from_2024-06-07_11-08-08.png
Screenshot_from_2024-06-07_11-08-08.png
Attachment 2: Screenshot_from_2024-06-07_11-08-21.png
Screenshot_from_2024-06-07_11-08-21.png
Attachment 3: Screenshot_from_2024-06-07_11-08-57.png
Screenshot_from_2024-06-07_11-08-57.png
Attachment 4: Screenshot_from_2024-06-07_11-09-10.png
Screenshot_from_2024-06-07_11-09-10.png
Attachment 5: Screenshot_from_2024-06-07_11-09-56.png
Screenshot_from_2024-06-07_11-09-56.png
Attachment 6: Screenshot_from_2024-06-07_11-10-05.png
Screenshot_from_2024-06-07_11-10-05.png
Attachment 7: Screenshot_from_2024-06-07_11-11-23.png
Screenshot_from_2024-06-07_11-11-23.png
Attachment 8: Screenshot_from_2024-06-07_11-11-26.png
Screenshot_from_2024-06-07_11-11-26.png
  594   Thu Jun 6 16:23:10 2024 LB CB JMRun 39 - Run 40 15N

Started run 39.

 

We have the 15N+ beam with 426 keV/nucleon. We see a peak in detector top left (mostly) and bottom right, at 4600 keV. We think it is the beam (nominally 6400 keV) after going through silicon dead layer, which however seems to be thicker than expected (expected 0.7 um, but given 1.4 MeV/um energy loss it seems to be > 1 um).

To understand what was going on we first played with detector and beam position.

At 17:28 detectors are fully out, while we are playing with the electron cooler to try to remove this beam halo hitting the detector.

 

Detector DSSD0 top left periodically seems to get noisier (statistics up to >2E5) and we need to move the detector in and out to go back to normal operations.

 

------------------

The beam halo in top left detector, as the detector gets moved in, gets elongated in the y direction instead of the x direction, as expected and previosuly observed.

Run 40 started.

18:36

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

Attachments 1-4

19:43

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

Attachments 5-8

21:15

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

Attachments 9-11

 

00;25

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

-----------------------------------

Movements and observation in xy plot:

- Move right servo in by 20 mm and we see the halo in the top right moving accordingly (attachment 12)

-Move right pneumatic in, saw big increase on bottom right (attachment 13)

-moved beam position down vertical -6 mm and we see reduced beam spot on top left and more on bottom left

-----------------------------------

Beam - target overlap:

We have yet to obtain an overlap

We cannot do this using the beam lifetime as from beam calc the lifetime from the ring is comparable to the lifetime with a target of 1e11 at/cm2

We have tried using Rutherford. However, we will have to move the detectors IN further as the Rutherford peaks at ~3.8 deg.

Possibly we may require other overlap detections from the ring?

I also think the beam requires some optimisation, electron cooler voltage seems to be fluctuating a lot and we see some features on the detectors when moving the beam / detectors as mentioned above.

 

02:25

The beam intensity has decreased signficantly (by about a factor of five) since about an hour. Operators tried different things, also consulted with Frank Herfurth, but they could not solve the problem.

There is no device which indicates malfunction.

So, during the rest of the night shift we will continue searching for the overlap (of the beam with the target) with the available beam intensity.

 

03:30

The beam intensity has improved without any aparent intervention.

We have started a new search for the overlap with 20 minute runs per horiznotal beam position.

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

 

03:40

After changing the horizontal beam position from 0 mm to 1 mm, the bottom right detector lost the Rutherford peak and its position spectrum looks to be half empty.

Moving it OUT and IN again did not change the situation.

There is no reason to think that this behavior has been caused by moving the beam by just 1 mm.

 

04:10

The beam intensity has dropped again without any apparent reason. So, at the moment we can not continue search for the overlap.

The bottom right detector seems to be counting normally again (also without any apparent reason). 

 

06:45

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

 

07:18

Lost beam again. Gleb is looking for it. Moved right detector out temporarily.

System wide checks as before.

Merger OK, Tape Server OK.

Temperatures OK.

Rates currently OK.

Leakage current decreasing.

 

10:10

Beam is back. Target is unusually thick at 3-5E11

Checking beam overlap.

Merger & tape server OK.

Temps OK.

Rates OK, a bit higher in 01 and 04.

Attachment 1: Screenshot_from_2024-06-06_18-33-41.png
Screenshot_from_2024-06-06_18-33-41.png
Attachment 2: Screenshot_from_2024-06-06_18-34-15.png
Screenshot_from_2024-06-06_18-34-15.png
Attachment 3: Screenshot_from_2024-06-06_18-34-35.png
Screenshot_from_2024-06-06_18-34-35.png
Attachment 4: Screenshot_from_2024-06-06_18-34-49.png
Screenshot_from_2024-06-06_18-34-49.png
Attachment 5: Screenshot_from_2024-06-06_19-42-49.png
Screenshot_from_2024-06-06_19-42-49.png
Attachment 6: Screenshot_from_2024-06-06_19-42-57.png
Screenshot_from_2024-06-06_19-42-57.png
Attachment 7: Screenshot_from_2024-06-06_19-43-10.png
Screenshot_from_2024-06-06_19-43-10.png
Attachment 8: Screenshot_from_2024-06-06_21-14-02.png
Screenshot_from_2024-06-06_21-14-02.png
Attachment 9: Screenshot_from_2024-06-06_21-14-17.png
Screenshot_from_2024-06-06_21-14-17.png
Attachment 10: Screenshot_from_2024-06-06_21-14-48.png
Screenshot_from_2024-06-06_21-14-48.png
Attachment 11: Screenshot_from_2024-06-06_23-28-29.png
Screenshot_from_2024-06-06_23-28-29.png
Attachment 12: Screenshot_from_2024-06-06_23-42-18.png
Screenshot_from_2024-06-06_23-42-18.png
Attachment 13: Screenshot_from_2024-06-07_07-20-12.png
Screenshot_from_2024-06-07_07-20-12.png
Attachment 14: Screenshot_from_2024-06-07_10-11-17.png
Screenshot_from_2024-06-07_10-11-17.png
Attachment 15: Screenshot_from_2024-06-07_10-11-27.png
Screenshot_from_2024-06-07_10-11-27.png
  593   Thu Jun 6 05:44:45 2024 DR YL DB LB CB JMRun 38 - 15N Tests

06:45

All system checks OK

Target density 3.4e11

11:50

System checks OK.

No beam in the ring yet. Althoug some injection may be taking place at low intensities

Target density appears much more stable than before.

Density currently just above 1e11 at/cm2

Note maintanace is required to appc025 so anydesk is currently down.

 

13:13

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

Attachments 1-5 (also waveforms)

 

15:00

Merger OK, Tape server OK

ADC data item statistics DO NOT looks as before. NOT OK

FEE temperatures look OK

System checks OK

Leakage current stable

Attachments 6-8

15:16

Notice the statistics for aida01-aida04 (DDSD 0, top left) were counting extremely high in the previous check (attachment 6). Previous rate was 0-1 counts.

We tried ASIC check load three times but not effect on the rate.

We then tried 'shaking the detectors' by moving the left pneumatic in and out again.

Rates go back to normal levels (0-1 counts)

 

Attachment 1: Screenshot_from_2024-06-06_13-05-55.png
Screenshot_from_2024-06-06_13-05-55.png
Attachment 2: Screenshot_from_2024-06-06_13-06-24.png
Screenshot_from_2024-06-06_13-06-24.png
Attachment 3: Screenshot_from_2024-06-06_13-07-46.png
Screenshot_from_2024-06-06_13-07-46.png
Attachment 4: Screenshot_from_2024-06-06_13-08-40.png
Screenshot_from_2024-06-06_13-08-40.png
Attachment 5: Screenshot_from_2024-06-06_13-09-03.png
Screenshot_from_2024-06-06_13-09-03.png
Attachment 6: Screenshot_from_2024-06-06_14-59-23.png
Screenshot_from_2024-06-06_14-59-23.png
Attachment 7: Screenshot_from_2024-06-06_14-59-34.png
Screenshot_from_2024-06-06_14-59-34.png
Attachment 8: Screenshot_from_2024-06-06_14-59-49.png
Screenshot_from_2024-06-06_14-59-49.png
  592   Wed Jun 5 13:34:02 2024 JMRun 37 - Change in thresholds

14:20

Target nozzle currently being replaced. No beam

We have decided to switch to 15N beam from now on.

Beam energy for 15N is significantly higher than deuteron beam so we have increased the thresholds (slow comparator) slightly for all detectors to 500 keV (0x32).

ADC data item rates - attachment 1

Run 37 started. Alpha background

Data rate ~ 200 kb/sec

17:44

All system checks OK

20:00

All system checks OK

23:30

All system checks OK

01:25

All system checks OK

03:25

All system checks OK.

Target is stable at 3.5e11, with T1=29 K

05:30

All system checks OK

Target density has slowly drifted down, and is now 1.2e11

Attachment 1: Screenshot_from_2024-06-05_14-36-44.png
Screenshot_from_2024-06-05_14-36-44.png
  591   Wed Jun 5 13:21:45 2024 JM, CBRun 36 - pulser walkthrough

14:00

Access to the cave as Nikos and Andreas have begun work on the target.

Run 36 started - pulser walkthrough.

Pulser amplitudes of 8, 7, 6, 5, 4, 3, 2, 1

Attachment 1 - pn channels

Attachment 2 - nn channels

Attachment 3 - pn channels (zoom)

Attachment 4 - nn channels (zoom)

Best pn channel looks like AIDA09 - pulser width 63

Worst pn channel looks like AIDA14 - pulser width 148

Best nn channel looks like AIDA12 - pulser width 58

Worst nn channel looks like AIDA16 - pulser width 246

This is approximately the same as pulser widths before the beam time.

Attachment 1: Screenshot_from_2024-06-05_14-20-18.png
Screenshot_from_2024-06-05_14-20-18.png
Attachment 2: Screenshot_from_2024-06-05_14-21-26.png
Screenshot_from_2024-06-05_14-21-26.png
Attachment 3: Screenshot_from_2024-06-05_14-27-43.png
Screenshot_from_2024-06-05_14-27-43.png
Attachment 4: Screenshot_from_2024-06-05_14-25-44.png
Screenshot_from_2024-06-05_14-25-44.png
  590   Wed Jun 5 08:34:04 2024 JM. DB, PCTarget nozzle change

09:30

Target density has not improved despite efforts last night. Mid 10 - low 11 density range. We will try to change the nozzle.

Target nozzle moved to the out position.

HKR allowed access to the cave to turn the compressor off. Target nozzle now warming up.

Internal cave work on changing nozzle expected to start at ~11. In the meantime we will let Wolfgang experiment with the beam settings to try to improve injection intensity.

Detectors have been moved out to fully out position

Run 35 started.

  589   Tue Jun 4 19:19:26 2024 TD[How To] Recover DAQ when one FEE64 stops sending data
1) STOP DAQ [workspace 2, display 1, browser tab 'Run Control']
   all FEE64s should change status from 'going' to 'stopped' *except* whichever FEE64(s) have a problem - typically their status is 'going' or 'error'

2) Restart Merger by selecting the Merger icon [workspace 3, display 2]
    terminal window showing the startup will popup

3) Setup Merger by selecting 'SETUP' [workspace 3, display 1, browser tab 'NewMerger Control']

4) Start Merger by selecting 'GO' [workspace 3, display 1, browser tab 'NewMerger Control']
     all data links will be inactive (red)

5)   Disable data transfer by selecting 'Data transfer Disable #1 all' [workspace 2, display 1, browser tab 'Run Control']

6) Start DAQ

    Start DAQ by selecting 'GO' [workspace 2, display 1, browser tab 'Run Control']
     check that DAQ is *actually* going - if not repeat 'GO'

7) Check WR timestamps by selecting 'Collect all timestamps' [workspace 2, display 1, browser tab 'GSI White Rabbit Control']
     readout times will be displayed for each FEE64
     check that the readout times are monotonically increasing as you read down the list of readout times - if not, contact expert support

8) Stop DAQ 

    Start DAQ by selecting 'STOP' [workspace 2, display 1, browser tab 'Run Control']
     check that DAQ has *actually* stopped - if not repeat 'STOP

9) Enable data transfer to Merger and TapeServer and start DAQ

    Enable data transfer by selecting 'Enable data transfer Enable #1 All' [workspace 2, display 1, browser tab 'Run Control']

    Start DAQ by selecting 'GO' [workspace 2, display 1, browser tab 'Run Control']
     check that DAQ is *actually* going - if not repeat 'GO'

10) Check data is being transferred to Merger and TapeServer

    Select 'Reload' [workspace 3, display 1, browser tab 'NewMerger Control']
     you should see a non-zero 'current merged data rate' and all links should show as green at least some of the time (requires observation over several reloads)

    Select 'Reload' [workspace 3, display 1, browser tab 'TapeServer']
     you should see 'Blocks written' and 'kb written' counters increase

11) Check AIDA data is being forwarded from the TapeServer to MBS [workspace 3, display 2, command terminal]

12) Restart should now have been successfully completed. Please inform the 'on shift' crew' that AIDA is forwarding data.

    If not, please request expert assistance.
  588   Tue Jun 4 18:08:04 2024 Boudefla Dounia - CB - LB - SDRun 34

19:08

Beam energy = 3 MeV.

Horizontal beam position = 4.0 mm, vertical beam position = 0.0 mm.

Target density ~ fluctuating

Beam intensity at inj ~ 1.7e7

Detectors are positioned IN. Left arm = 36.54, right arm =  64.60 mm

MBS signals saved from motors code

Histograms and stats zero'd, monitoring code restarted.

Beam in the ring ~ 6 minutes.

Pressure in YR09 E target ~ 4.36e-11 mbar.

 

 

Merger error recovered

White rabbit time stamps

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

 

21:27

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

 

23:00

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

Statistics attachment 5

Merger attachment 6

 

00:42

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

Density is oscillating between E10 and low E11.

attachment 7 is statistics, attachment 8 is temperature, attachment 9 is merger, attachment 10 is spectrum browser.

 

02:20

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

Density is oscillating between E10 and low E11. ( last 15 minutes below 2E10)

attachment 11 -13

 

From 2:00 target density falls below 2E10 and started decreasing to the E9 range. At 2:40 we started playing with x-y to get to higher density. At 3:30 we are at 5/6E10 and waiting to see if it stabilizesto that or if it needs more playing.

04:06

Target density being 1-2E11 since 5 minutes

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

Attachment 14 statistics, attachment 15 merger

 

05;53

Target density 1E10, sometimes e11, before going down to low E10 and we need to play a bit with x-y to recover it

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

Attachment 16-18

 

07:35

Target density 2.5E11, after changiny y position to 1000

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

09:00

Target density has not improved despite efforts last night. Mid 10 - low 11 density range. We have asked HKR to remove beam from the ring and allow access so Nikos can change the nozzle.

Beam off now, access granted. Will begin warming nozzle. Internal cave work on changing nozzle expected to start at 11

Run stopped.

Final monitoring code histogram attached

 

Attachment 1: Screenshot_from_2024-06-04_19-07-40.png
Screenshot_from_2024-06-04_19-07-40.png
Attachment 2: Screenshot_from_2024-06-04_19-11-14.png
Screenshot_from_2024-06-04_19-11-14.png
Attachment 3: Screenshot_from_2024-06-04_21-26-45.png
Screenshot_from_2024-06-04_21-26-45.png
Attachment 4: Screenshot_from_2024-06-04_21-26-50.png
Screenshot_from_2024-06-04_21-26-50.png
Attachment 5: Screenshot_from_2024-06-04_22-40-39.png
Screenshot_from_2024-06-04_22-40-39.png
Attachment 6: Screenshot_from_2024-06-04_22-43-21.png
Screenshot_from_2024-06-04_22-43-21.png
Attachment 7: Screenshot_from_2024-06-05_00-44-49.png
Screenshot_from_2024-06-05_00-44-49.png
Attachment 8: Screenshot_from_2024-06-05_00-45-21.png
Screenshot_from_2024-06-05_00-45-21.png
Attachment 9: Screenshot_from_2024-06-05_00-45-41.png
Screenshot_from_2024-06-05_00-45-41.png
Attachment 10: Screenshot_from_2024-06-05_00-47-33.png
Screenshot_from_2024-06-05_00-47-33.png
Attachment 11: Screenshot_from_2024-06-05_02-21-37.png
Screenshot_from_2024-06-05_02-21-37.png
Attachment 12: Screenshot_from_2024-06-05_02-22-15.png
Screenshot_from_2024-06-05_02-22-15.png
Attachment 13: Screenshot_from_2024-06-05_02-22-44.png
Screenshot_from_2024-06-05_02-22-44.png
Attachment 14: Screenshot_from_2024-06-05_04-06-59.png
Screenshot_from_2024-06-05_04-06-59.png
Attachment 15: Screenshot_from_2024-06-05_04-07-43.png
Screenshot_from_2024-06-05_04-07-43.png
Attachment 16: Screenshot_from_2024-06-05_05-55-15.png
Screenshot_from_2024-06-05_05-55-15.png
Attachment 17: Screenshot_from_2024-06-05_05-55-49.png
Screenshot_from_2024-06-05_05-55-49.png
Attachment 18: Screenshot_from_2024-06-05_05-56-34.png
Screenshot_from_2024-06-05_05-56-34.png
Attachment 19: Screenshot_from_2024-06-05_09-04-45.png
Screenshot_from_2024-06-05_09-04-45.png
  587   Tue Jun 4 15:25:36 2024 BD US CBRun 33

16:26

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

Nicos came in to optimize target density.

 

18:39

aida02 in the MERGER aka aida03 elsewhere has a red background, but still appears to be transferring data. Rates are fine. Temps are fine. System wide checks unchanged.

See attachments. Checkloaded x2 - no difference. Unclear what the issue is.

Attachment 1: Screenshot_from_2024-06-04_18-33-23.png
Screenshot_from_2024-06-04_18-33-23.png
Attachment 2: Screenshot_from_2024-06-04_18-33-38.png
Screenshot_from_2024-06-04_18-33-38.png
Attachment 3: Screenshot_from_2024-06-04_18-35-26.png
Screenshot_from_2024-06-04_18-35-26.png
Attachment 4: Screenshot_from_2024-06-04_18-36-55.png
Screenshot_from_2024-06-04_18-36-55.png
  586   Tue Jun 4 14:13:17 2024 JM, CB, DB, USRun 32- beam only background

15:15

Run 32 started - Beam, no target

Beam energy = 3 MeV.

Horizontal beam position = 4.0 mm, vertical beam position = 0.0 mm.

Beam intensity at inj ~ 2.5e7

Detectors are positioned IN. Left arm = 36.54, right arm =  64.60 mm

Histograms and stats zero'd, monitoring code restarted.

Beam in the ring ~ 5 minutes.

Pressure in YR09 E target ~ 4.5e-11 mbar.

Motors code crashed. Beam removed from ring to re-home detectors.

 

15:20

Beam in the ring again, detectors homed, no target

Detectors have been set up to send signals on injection / dump

15:35

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

 

  585   Tue Jun 4 06:56:02 2024 DFF, PC, LB, JM,NP,ABRun 31 contd

8:13

Since 6:40 the target density dramatically dropped to 10^9 since then it has been unstable and did not recover. Currently it is around 8e8.

09:00

Target density has been increased to the high -10, very low -11 region. We will keep this run going unless the density is increased further.

 

10:21

Beam energy = 3 MeV.

Horizontal beam position = 4.0 mm, vertical beam position = 0.0 mm.

Target density ~ 1e9 at/cm2 (no target, get lost quite quckly)

Beam intensity at inj ~ 1.7e7

Detectors are positioned IN. Left arm = 36.54, right arm =  64.60 mm

Histograms and stats zero'd, monitoring code restarted.

Beam in the ring ~ 6 minutes.

Pressure in YR09 E target ~ 4.36e-11 mbar.

 

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

 

target density has been fluctuating between E9 to E12. We can very clearly see the increase in elastic counts on our detector during high density periods.

Attachment 1 - rate plots at 1E12 at/cm2

Attachment 2 - rate plots at 1E9 at/cm2

 

12:26

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

 

After reaching a density of > 10^12 a bit before 11:00, the valve got closed. When a valve is closed, can be observed a dramatic change in the plot and one of the three green ligths of atps7vaccrz1 (0,2,4) is turned off. Nikos opened it, and now it seems stable again around 1e11.

 

14:07

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

15:05

Run stopped as target density is rubbish. Might as well get beam background

Attachment 1: Screenshot_from_2024-06-04_10-53-58.png
Screenshot_from_2024-06-04_10-53-58.png
Attachment 2: Screenshot_from_2024-06-04_10-58-02.png
Screenshot_from_2024-06-04_10-58-02.png
  584   Tue Jun 4 00:04:39 2024 DR, EHRun 31

Beam energy = 3 MeV

Horizontal beam position = 4.0 mm, vertical beam position = 0.0 mm.

Beam intensity at inj ~ 2.5e7

Detectors are positioned IN. Left arm = 36.54, right arm =  64.60 mm

Cycle time ~ 6 minutes

 

01:05

Target density still very unstable

 

02:45

Target density has dropped to ~1e9 and has been at that level for 30 minutes. We will try to move the target a bit and see if we can get an increase. Target position is currently x=-12400, y=2900

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

 

04:15

We were able to recover some target, but it is still very unstable. The new position is x=-12800, y=2000

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

 

06:15

Target still unstable, fluctuating around the 5e10 region

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

  583   Mon Jun 3 21:37:20 2024 CN, EM, CBRun30

Beam energy = 3 MeV

Horizontal beam position = 7.0 mm, vertical beam position = 0.0 mm.

Beam intensity at inj ~ 2.5e7

Detectors are positioned IN. Left arm = 36.54, right arm =  64.60 mm

22:52 (Attachments 1 -3)

Beam is very unstable

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

 

00:47

Beam is still very unstable.

We checked for beam-target overlap position and found a clear increase in counts at beam position 4.0 mm. We have left the beam at horizontal = 4.0 mm, vertical = 0.0

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Attachment 1: Screenshot_from_2024-06-03_23-15-37.png
Screenshot_from_2024-06-03_23-15-37.png
Attachment 2: Screenshot_from_2024-06-03_23-16-18.png
Screenshot_from_2024-06-03_23-16-18.png
Attachment 3: Screenshot_from_2024-06-03_23-17-16.png
Screenshot_from_2024-06-03_23-17-16.png
  582   Mon Jun 3 15:17:41 2024 CN, EM, CBRun29

16:15

In the previous run while working on the target it was discovered that the valve was closed! Probaly this happened around 10 am, we are not sure.

The gas target inlet valve is open again, working on the target.

 

Beam energy = 3 MeV

Horizontal beam position = 7.0 mm, vertical beam position = 0.0 mm.

Beam intensity at inj ~ 2.5e7

Detectors are positioned IN pneumatic only. Left arm = 62.71, right arm =  64.60 mm

Histograms and stats zero'd, monitoring code restarted.

Beam in the ring ~ 7 minutes.

Pressure in YR09 E target ~ 4.5e-11 mbar.

 

18:03 ( Attchments 1 -4)

Still no target!

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

 

20:23

We have the target back. S1( grafana) 4.8e-8. The target thickness read by the system 5E11 atoms/cm2 but apparently the formula is wrong and the value should be almost e12 atoms/cm2.

. We have now proper beam and target and we

 

Attachment 1: Screenshot_from_2024-06-03_18-04-47.png
Screenshot_from_2024-06-03_18-04-47.png
Attachment 2: Screenshot_from_2024-06-03_18-05-11.png
Screenshot_from_2024-06-03_18-05-11.png
Attachment 3: Screenshot_from_2024-06-03_18-06-09.png
Screenshot_from_2024-06-03_18-06-09.png
Attachment 4: Screenshot_from_2024-06-03_18-06-40.png
Screenshot_from_2024-06-03_18-06-40.png
Attachment 5: Screenshot_from_2024-06-03_20-35-00.png
Screenshot_from_2024-06-03_20-35-00.png
  581   Mon Jun 3 15:10:12 2024 CN, EM, CBRun28

16:10

We start a new run after doing some test with the motor. Injection time is now a 6s, moving of the motors is working fine.

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

No target

 

Attachment 1: Screenshot_from_2024-06-03_16-12-44.png
Screenshot_from_2024-06-03_16-12-44.png
Attachment 2: Screenshot_from_2024-06-03_16-13-11.png
Screenshot_from_2024-06-03_16-13-11.png
Attachment 3: Screenshot_from_2024-06-03_16-14-11.png
Screenshot_from_2024-06-03_16-14-11.png
Attachment 4: Screenshot_from_2024-06-03_16-16-37.png
Screenshot_from_2024-06-03_16-16-37.png
  580   Mon Jun 3 11:46:07 2024 SD, SHB, JM, CN, EMRun 27

12:46 (attachments 1-4)

Run 27 started - Beam no target

Beam energy = 3 MeV

Horizontal beam position = 7.0 mm, vertical beam position = 0.0 mm.

Beam intensity at inj ~ 2.5e7

Detectors are positioned IN pneumatic only. Left arm = 62.71, right arm =  64.60 mm

Histograms and stats zero'd, monitoring code restarted.

Beam in the ring ~ 7 minutes.

Beam on BPM looks to be better than at previous energy.

Pressure in YR09 E target ~ 4.5e-11 mbar.

 

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

 

13:45 (attachments 5-8)

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

 

 

15:15(attachments 9-12)

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

Target density 1e10 atoms/cm2. Still working on the target.

 

 

Attachment 1: Screenshot_from_2024-06-03_12-48-03.png
Screenshot_from_2024-06-03_12-48-03.png
Attachment 2: Screenshot_from_2024-06-03_12-48-37.png
Screenshot_from_2024-06-03_12-48-37.png
Attachment 3: Screenshot_from_2024-06-03_12-49-13.png
Screenshot_from_2024-06-03_12-49-13.png
Attachment 4: Screenshot_from_2024-06-03_12-49-47.png
Screenshot_from_2024-06-03_12-49-47.png
Attachment 5: Screenshot_from_2024-06-03_13-45-42.png
Screenshot_from_2024-06-03_13-45-42.png
Attachment 6: Screenshot_from_2024-06-03_13-46-49.png
Screenshot_from_2024-06-03_13-46-49.png
Attachment 7: Screenshot_from_2024-06-03_13-47-32.png
Screenshot_from_2024-06-03_13-47-32.png
Attachment 8: Screenshot_from_2024-06-03_13-47-51.png
Screenshot_from_2024-06-03_13-47-51.png
Attachment 9: Screenshot_from_2024-06-03_15-16-51.png
Screenshot_from_2024-06-03_15-16-51.png
Attachment 10: Screenshot_from_2024-06-03_15-17-49.png
Screenshot_from_2024-06-03_15-17-49.png
Attachment 11: Screenshot_from_2024-06-03_15-18-19.png
Screenshot_from_2024-06-03_15-18-19.png
Attachment 12: Screenshot_from_2024-06-03_15-22-14.png
Screenshot_from_2024-06-03_15-22-14.png
  579   Mon Jun 3 08:43:22 2024 SD, SHBRun 26

09:44

Run 26 started - Beam no target

Beam energy is being increased from 1.5 to 3 MeV

Horizontal beam position = 7.0 mm, vertical beam position = 0.0 mm.

Detectors are positioned fully OUT

Histograms and stats zero'd, monitoring code restarted.

Pressure in YR09 E target ~ 4.5e-11 mbar.

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

 

10:48 (Attachements 1-4)

 

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

The detectors currently fully out for tuning.

 

 

11:57 (attachments 5-8)

 

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

The detectors currently fully out for tuning.

 

 

 

 

Attachment 1: Screenshot_from_2024-06-03_10-49-19.png
Screenshot_from_2024-06-03_10-49-19.png
Attachment 2: Screenshot_from_2024-06-03_10-50-04.png
Screenshot_from_2024-06-03_10-50-04.png
Attachment 3: Screenshot_from_2024-06-03_10-53-01.png
Screenshot_from_2024-06-03_10-53-01.png
Attachment 4: Screenshot_from_2024-06-03_10-53-44.png
Screenshot_from_2024-06-03_10-53-44.png
Attachment 5: Screenshot_from_2024-06-03_11-55-41.png
Screenshot_from_2024-06-03_11-55-41.png
Attachment 6: Screenshot_from_2024-06-03_11-56-17.png
Screenshot_from_2024-06-03_11-56-17.png
Attachment 7: Screenshot_from_2024-06-03_11-56-36.png
Screenshot_from_2024-06-03_11-56-36.png
Attachment 8: Screenshot_from_2024-06-03_11-57-31.png
Screenshot_from_2024-06-03_11-57-31.png
  578   Sun Jun 2 23:38:17 2024 JM, EMRun 25

00:38

Run 25 started - Beam on target

Beam energy = 1.5 MeV.

Horizontal beam position = 7.0 mm, vertical beam position = 0.0 mm.

Target density ~ 5e11 at/cm2 (Still Fluctuating, after temperature increase)

Beam intensity at inj ~ 2.5e7

Detectors are positioned IN. Left arm = 62.71, right arm =  64.60 mm

Histograms and stats zero'd, monitoring code restarted.

Beam in the ring ~ 5 minutes.

Pressure in YR09 E target ~ 5.2e-11 mbar.

 

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

 

01:30 (Attachments 1 - 4)

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

The target density seems very unstable: fluctuations from 1e11 to 1e12 atoms/cm2 within 5min

 

02:37 (Attachments 5 - 8)

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

The target density is around 5e11  atoms/cm2 with small adjustments

 

03:32 (Attachments 9 - 12)

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

The target density is in average 5e11  atoms/cm2 with some small  adjustments. Max values up to 3e12 atoms/cm2

 

One of the injections at 03:43  was not ok and we lost beam for some min.  Target density 1e12 atoms/cm2.  On 04:00 the beam was back again without any operator intervention.

 

04:35 (Attachments 13 - 16)

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

The target density is in average 1e12  atoms/cm2.

 

05:43 (Attachments 17 - 20)

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

The target density decreased to 2e11 atoms/cm2 in the last 40 min

 

06:43 (Attachments 21 - 24)

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

The target density is fluctuating from 1e11 to 1e12 atoms/cm without any change....

 

07:53 (Attachments 25 - 28)

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

The target density is fluctuating from 1e11 to 1e12 atoms/cm without any change....

 

08:48 (Attachments 29 - 32)

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

The target density is fluctuating from 1e11 to 1e12 atoms/cm without any change....

 

 

09:40 --> Run 25 stopped.

 

 

Attachment 1: Screenshot_from_2024-06-03_01-34-08.png
Screenshot_from_2024-06-03_01-34-08.png
Attachment 2: Screenshot_from_2024-06-03_01-34-32.png
Screenshot_from_2024-06-03_01-34-32.png
Attachment 3: Screenshot_from_2024-06-03_01-34-54.png
Screenshot_from_2024-06-03_01-34-54.png
Attachment 4: Screenshot_from_2024-06-03_01-35-46.png
Screenshot_from_2024-06-03_01-35-46.png
Attachment 5: Screenshot_from_2024-06-03_02-38-30.png
Screenshot_from_2024-06-03_02-38-30.png
Attachment 6: Screenshot_from_2024-06-03_02-38-55.png
Screenshot_from_2024-06-03_02-38-55.png
Attachment 7: Screenshot_from_2024-06-03_02-39-26.png
Screenshot_from_2024-06-03_02-39-26.png
Attachment 8: Screenshot_from_2024-06-03_02-40-07.png
Screenshot_from_2024-06-03_02-40-07.png
Attachment 9: Screenshot_from_2024-06-03_03-33-32.png
Screenshot_from_2024-06-03_03-33-32.png
Attachment 10: Screenshot_from_2024-06-03_03-33-56.png
Screenshot_from_2024-06-03_03-33-56.png
Attachment 11: Screenshot_from_2024-06-03_03-34-20.png
Screenshot_from_2024-06-03_03-34-20.png
Attachment 12: Screenshot_from_2024-06-03_03-35-16.png
Screenshot_from_2024-06-03_03-35-16.png
Attachment 13: Screenshot_from_2024-06-03_04-35-47.png
Screenshot_from_2024-06-03_04-35-47.png
Attachment 14: Screenshot_from_2024-06-03_04-36-07.png
Screenshot_from_2024-06-03_04-36-07.png
Attachment 15: Screenshot_from_2024-06-03_04-36-25.png
Screenshot_from_2024-06-03_04-36-25.png
Attachment 16: Screenshot_from_2024-06-03_04-39-59.png
Screenshot_from_2024-06-03_04-39-59.png
Attachment 17: Screenshot_from_2024-06-03_05-44-44.png
Screenshot_from_2024-06-03_05-44-44.png
Attachment 18: Screenshot_from_2024-06-03_05-45-07.png
Screenshot_from_2024-06-03_05-45-07.png
Attachment 19: Screenshot_from_2024-06-03_05-45-24.png
Screenshot_from_2024-06-03_05-45-24.png
Attachment 20: Screenshot_from_2024-06-03_05-46-22.png
Screenshot_from_2024-06-03_05-46-22.png
Attachment 21: Screenshot_from_2024-06-03_06-42-21.png
Screenshot_from_2024-06-03_06-42-21.png
Attachment 22: Screenshot_from_2024-06-03_06-42-49.png
Screenshot_from_2024-06-03_06-42-49.png
Attachment 23: Screenshot_from_2024-06-03_06-43-48.png
Screenshot_from_2024-06-03_06-43-48.png
Attachment 24: Screenshot_from_2024-06-03_06-44-02.png
Screenshot_from_2024-06-03_06-44-02.png
Attachment 25: Screenshot_from_2024-06-03_07-48-07.png
Screenshot_from_2024-06-03_07-48-07.png
Attachment 26: Screenshot_from_2024-06-03_07-49-47.png
Screenshot_from_2024-06-03_07-49-47.png
Attachment 27: Screenshot_from_2024-06-03_07-50-51.png
Screenshot_from_2024-06-03_07-50-51.png
Attachment 28: Screenshot_from_2024-06-03_07-53-01.png
Screenshot_from_2024-06-03_07-53-01.png
Attachment 29: Screenshot_from_2024-06-03_08-46-03.png
Screenshot_from_2024-06-03_08-46-03.png
Attachment 30: Screenshot_from_2024-06-03_08-46-41.png
Screenshot_from_2024-06-03_08-46-41.png
Attachment 31: Screenshot_from_2024-06-03_08-47-09.png
Screenshot_from_2024-06-03_08-47-09.png
Attachment 32: Screenshot_from_2024-06-03_08-48-17.png
Screenshot_from_2024-06-03_08-48-17.png
  577   Sun Jun 2 22:46:47 2024 JM, EM, NPRun24

23:47

Run 24 started - Beam on target

Beam energy is 1.5 MeV.

Horizontal beam position = 7.0 mm, vertical beam position = -1.0 mm.

Target density ~ 1e11 at/cm2 (Still Fluctuating, after temperature increase)

Beam intensity at inj ~ 2.5e7

Detectors are positioned IN. Left arm = 62.71, right arm =  64.60 mm

Histograms and stats zero'd, monitoring code restarted.

Beam in the ring ~ 10 minutes.

Pressure in YR09 E target ~ 5.2e-11 mbar.

 

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

 

00:23

 

Target density ~4.65e11 atoms/cm2

Beam intensity at inj ~ 2.7e7

Beam intensity dropped to 2.1e7 in 10 minutes. Proportionall this is much greater than before. 

Run 24 stopped.

 

Attachment 1: Screenshot_from_2024-06-02_23-53-00.png
Screenshot_from_2024-06-02_23-53-00.png
Attachment 2: Screenshot_from_2024-06-02_23-53-47.png
Screenshot_from_2024-06-02_23-53-47.png
Attachment 3: Screenshot_from_2024-06-02_23-56-22.png
Screenshot_from_2024-06-02_23-56-22.png
  576   Sun Jun 2 14:30:16 2024 TDOffline analysis G22-201/R22_0 - R22_11
Offline analysis data file G22-201/R22_0 to R22_11

Sort program variables https://elog.ph.ed.ac.uk/CARME/240530_180659/variables.dat

ADC data corrected for ADC offset but not gain

per DSSSD m_p verus m_n - attachment 1

For spectra below events must satisfy condition 0 < m_p < 8 *and* 0 < m_n < 8

per DSSSD p+n junction strip versus n+n Ohmic strip - attachment 2
 note - x/y channel ordering to be checked

per DSSSD E_p versus E_n - attachment 3
 20keV/channel nominal

per DSSSD E_p versus E_n - attachment 4
 gated w(1) & w(2)
 20keV/channel nominal

per DSSSD projection of gated E_p versus E_n spectra - attachments 5-8
 20keV/channel nominal
Attachment 1: Screenshot_from_2024-06-01_23-04-28.png
Screenshot_from_2024-06-01_23-04-28.png
Attachment 2: Screenshot_from_2024-06-01_23-05-20.png
Screenshot_from_2024-06-01_23-05-20.png
Attachment 3: Screenshot_from_2024-06-01_23-05-58.png
Screenshot_from_2024-06-01_23-05-58.png
Attachment 4: Screenshot_from_2024-06-01_22-57-42.png
Screenshot_from_2024-06-01_22-57-42.png
Attachment 5: Screenshot_from_2024-06-01_22-59-30.png
Screenshot_from_2024-06-01_22-59-30.png
Attachment 6: Screenshot_from_2024-06-01_23-00-45.png
Screenshot_from_2024-06-01_23-00-45.png
Attachment 7: Screenshot_from_2024-06-01_23-01-56.png
Screenshot_from_2024-06-01_23-01-56.png
Attachment 8: Screenshot_from_2024-06-01_23-03-22.png
Screenshot_from_2024-06-01_23-03-22.png
  575   Sun Jun 2 13:04:41 2024 JM, AG, CB, DRRun 23 - vertical beam position and beam lifetime tests

14:05

Run 22 stopped. Run 23 started we will now adjust the vertical beam position to see if we observe the beam stripe seen in the last beamtime.

Current vertical beam position 0.0 mm. Horizontal beam position 7.0 mm. Coasting beam.

Beam D+, 1.5 MeV. Storage time ~ 20 minutes. Target ON. Density ~1e11 at/cm2.

XY plot from run 22 with the same settings - attachment 1

Beam stripe looks like it is just on bottom detectors (just )

14:10

New beam settings

Vertical Beam position 5.0mm, Horizontal unchanged. Storage time reduced to ~2 minute

No real change observed. Perhaps some thing more can be seen on top right detector but definitely not like in the Feb experiment.

14:15

New beam settings

Vertical Beam position 9.0mm, Horizontal unchanged.

No real change observed again.

14:20

New beam settings

Vertical Beam position -9.0mm, Horizontal unchanged.

More events observed on bottom left detector but nothing like the Feb measurements. I think we do not have stripe effects like previously.

14:40:50

New Beam settings to check beam lifetime.

Vertical beam position set back to 0.0 mm. Horizontal unchanged

In bunched mode to observed intensity over time. 

Injection into ring 2.2e7

15:15

Beam intensity dropped to 1.1e7. Beam lifetime 35 minutes.

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

 

15:24

We removed the target. Beam intensity at 2.2E7

After 35 minutes, the beam intensity is 1.41e7. So target does have effect on beam lifetime! We would like more.

16:05

We will now bring target back in and try to increase density.

16:10

Some sort of GSI-wide failure, we have lost beam entirely. We don't see any effect on our system. Temperatures and leakage currents are stable

Detectors moved fully out for safety

17:15

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

19:15

Still no beam

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

21:15

Still no beam

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

21:40

Beam has returned!

22:00

Target at around 9e11 and appears stable

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

22:33

Beam back in the ring. Beam settings to be updated.

Rates in right hand detectors look slightly different to before - lower in bottom right, slightly higher in top left. Perhaps beam vertical position needs to be lowered slightly.

ASIC check load just to be sure. Rates in aida15 greatly reduced - only temporarily.

 

Attachment 1: Screenshot_from_2024-06-02_14-04-55.png
Screenshot_from_2024-06-02_14-04-55.png
  574   Sun Jun 2 08:47:01 2024 CB, AGRun 22 - longer rung with 20 min storage

09:47

12 seconds pre acc

200x cycles

Detector pneumatics IN, not moving

Running in this configuration for the rest of the shift

Target is barely 1E11 atoms/cm2 and keeps decreasing

 

Temps OK, checks OK, Stats OK

Writing to disk rate OK

 

11:22

Targets gets thinner and thinner <1E11 atoms/cm2

Temps OK System wide checks OK Merger OK

Rates OK

 

13:02

Temps OK. Target around 1E11 atoms/cm2

Merger OK

Rates OK

Attachment 1: Screenshot_from_2024-06-02_09-47-53.png
Screenshot_from_2024-06-02_09-47-53.png
Attachment 2: Screenshot_from_2024-06-02_09-48-06.png
Screenshot_from_2024-06-02_09-48-06.png
Attachment 3: Screenshot_from_2024-06-02_11-21-59.png
Screenshot_from_2024-06-02_11-21-59.png
Attachment 4: Screenshot_from_2024-06-02_11-22-27.png
Screenshot_from_2024-06-02_11-22-27.png
Attachment 5: Screenshot_from_2024-06-02_13-02-31.png
Screenshot_from_2024-06-02_13-02-31.png
  573   Sun Jun 2 06:41:00 2024 CB, AGRun 21 - tests

07:41

Switching to bunched beam to check D vs H content.

8:24

 

Attach 1&2

ONE spill

Detectors fully out

Switched to bunched beam

12 seconds pre-acceleration

100x cycles (600 seconds) -

Loss over 600 seconds 3.97E7 ions to 3.40E7 ions, i.e. negligible.

 

Attach 3&4

Coasting beam

12 seconds pre-acc

100x cycles

 

Attch 5&6

Coasting beam

12 seconds pre-acc

100x cycles

Detectors pneumatics IN

 

Attach 7& 8

Coasting beam

6 seconds pre-acc

100x cycles

Detector pneumatics IN

 

Attach 8&9

Coasting beam

12 seconds pre acc

200x cycles

Detector pneumatics IN

Running in this configuration for the rest of the shift

Target is barely 1E11 atoms/cm2 and keeps decreasing

Attachment 1: Screenshot_from_2024-06-02_08-23-21.png
Screenshot_from_2024-06-02_08-23-21.png
Attachment 2: Screenshot_from_2024-06-02_08-23-36.png
Screenshot_from_2024-06-02_08-23-36.png
Attachment 3: Screenshot_from_2024-06-02_08-44-38.png
Screenshot_from_2024-06-02_08-44-38.png
Attachment 4: Screenshot_from_2024-06-02_08-44-54.png
Screenshot_from_2024-06-02_08-44-54.png
Attachment 5: Screenshot_from_2024-06-02_08-58-17.png
Screenshot_from_2024-06-02_08-58-17.png
Attachment 6: Screenshot_from_2024-06-02_08-58-25.png
Screenshot_from_2024-06-02_08-58-25.png
Attachment 7: Screenshot_from_2024-06-02_09-18-02.png
Screenshot_from_2024-06-02_09-18-02.png
Attachment 8: Screenshot_from_2024-06-02_09-18-10.png
Screenshot_from_2024-06-02_09-18-10.png
Attachment 9: Screenshot_from_2024-06-02_09-44-16.png
Screenshot_from_2024-06-02_09-44-16.png
Attachment 10: Screenshot_from_2024-06-02_09-44-24.png
Screenshot_from_2024-06-02_09-44-24.png
  572   Sat Jun 1 21:56:17 2024 JM, EM, EM, TY, CB, AGRun 20

22:56

Run 20 started - beam on target

Beam energy is 1.5 MeV. Horizontal beam position = 7.0 mm, vertical beam position = 0.0 mm.

Target density ~ 1e11 at/cm2

Beam intensity at inj ~ 4e7

Detectors are positioned IN. Left arm = 62.71, right arm =  64.60 mm

Histograms and stats zero'd, monitoring code restarted.

Beam in the ring ~ 3minutes.

Pressure in YR09 E target ~ 4.5e-11 mbar.

 

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

 

01:15

Target density has increased slighty: ~ 4e11 at/cm2 (update: this only lasted about 5 min)

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

 

3:10

Target density has decresed slightly but will come back (hopefully) ~7e10 at/cm2.

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

 

05:15

Target density has been somewhat unstable, between ~7E10 and ~3E12 at/cm2.

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

 

07:37

Run continues. Target density still fluctuating around 1E11 atoms/cm2.

Temperatures OK. Merger OK. Rates OK.

System wide checks as before.

Waveforms OK. See attachments.

Some screenshots from the online analysis. We do not see a peak at 1 MeV. We do however clearly see more D than H2+.

Stopping run to make some tests.

 

 

Attachment 1: Screenshot_from_2024-06-01_23-02-05.png
Screenshot_from_2024-06-01_23-02-05.png
Attachment 2: Screenshot_from_2024-06-01_23-02-46.png
Screenshot_from_2024-06-01_23-02-46.png
Attachment 3: Screenshot_from_2024-06-02_01-16-01.png
Screenshot_from_2024-06-02_01-16-01.png
Attachment 4: Screenshot_from_2024-06-02_01-16-22.png
Screenshot_from_2024-06-02_01-16-22.png
Attachment 5: Screenshot_from_2024-06-02_03-11-55.png
Screenshot_from_2024-06-02_03-11-55.png
Attachment 6: Screenshot_from_2024-06-02_03-12-00.png
Screenshot_from_2024-06-02_03-12-00.png
Attachment 7: Screenshot_from_2024-06-02_05-14-40.png
Screenshot_from_2024-06-02_05-14-40.png
Attachment 8: Screenshot_from_2024-06-02_05-15-00.png
Screenshot_from_2024-06-02_05-15-00.png
Attachment 9: Screenshot_from_2024-06-02_07-28-30.png
Screenshot_from_2024-06-02_07-28-30.png
Attachment 10: Screenshot_from_2024-06-02_07-30-22.png
Screenshot_from_2024-06-02_07-30-22.png
Attachment 11: Screenshot_from_2024-06-02_07-31-21.png
Screenshot_from_2024-06-02_07-31-21.png
Attachment 12: Screenshot_from_2024-06-02_07-32-03.png
Screenshot_from_2024-06-02_07-32-03.png
Attachment 13: Screenshot_from_2024-06-02_07-32-21.png
Screenshot_from_2024-06-02_07-32-21.png
Attachment 14: Screenshot_from_2024-06-02_07-32-42.png
Screenshot_from_2024-06-02_07-32-42.png
Attachment 15: Screenshot_from_2024-06-02_07-32-49.png
Screenshot_from_2024-06-02_07-32-49.png
Attachment 16: Screenshot_from_2024-06-02_07-33-00.png
Screenshot_from_2024-06-02_07-33-00.png
Attachment 17: Screenshot_from_2024-06-02_07-33-45.png
Screenshot_from_2024-06-02_07-33-45.png
Attachment 18: Screenshot_from_2024-06-02_07-36-41.png
Screenshot_from_2024-06-02_07-36-41.png
  571   Sat Jun 1 16:38:13 2024 TDOffline analysis G22-201/R17
Offline analysis data file G22-201/R17

Sort program variables https://elog.ph.ed.ac.uk/CARME/240530_180659/variables.dat

ADC data corrected for ADC offset but not gain

per DSSSD m_p verus m_n - attachment 1
 DSSSD #2 spectrum is normal, DSSSD #1 and #3-4 display large multiplicities in p and/or n strips which are, presumably, non-physical events at DSSSD edge

For spectra below events must satisfy condition 0 < m_p < 8 *and* 0 < m_n < 8

per DSSSD p+n junction strip versus n+n Ohmic strip - attachment 2
 note - x/y channel ordering to be checked

per DSSSD E_p versus E_n - attachment 3
 20keV/channel nominal

per DSSSD E_p versus E_n - attachment 4
 gated w(1) & w(2)
 20keV/channel nominal

per DSSSD projection of gated E_p versus E_n spectra - attachments 5-8
 20keV/channel nominal
 log scale
Attachment 1: Screenshot_from_2024-06-01_08-31-53.png
Screenshot_from_2024-06-01_08-31-53.png
Attachment 2: Screenshot_from_2024-06-01_08-33-16.png
Screenshot_from_2024-06-01_08-33-16.png
Attachment 3: Screenshot_from_2024-06-01_08-33-58.png
Screenshot_from_2024-06-01_08-33-58.png
Attachment 4: Screenshot_from_2024-06-01_08-34-44.png
Screenshot_from_2024-06-01_08-34-44.png
Attachment 5: Screenshot_from_2024-06-01_08-36-06.png
Screenshot_from_2024-06-01_08-36-06.png
Attachment 6: Screenshot_from_2024-06-01_08-37-22.png
Screenshot_from_2024-06-01_08-37-22.png
Attachment 7: Screenshot_from_2024-06-01_08-39-06.png
Screenshot_from_2024-06-01_08-39-06.png
Attachment 8: Screenshot_from_2024-06-01_08-40-22.png
Screenshot_from_2024-06-01_08-40-22.png
  570   Sat Jun 1 13:33:50 2024 TDOffline analysis APR24/R3 (alpha background)
Offline analysis data file APR24/R3 (alpha background)

Run started 11:04 26.4.24
Run ended   09:44 29.4.24

Sort program variables https://elog.ph.ed.ac.uk/CARME/240530_180659/variables.dat

ADC data corrected for ADC offset but not gain

per DSSSD m_p verus m_n - attachment 1

For spectra below events must satisfy condition 0 < m_p < 8 *and* 0 < m_n < 8

per DSSSD p+n junction strip versus n+n Ohmic strip - attachment 2
 note - x/y channel ordering to be checked

per DSSSD E_p versus E_n - attachment 3
 20keV/channel nominal

per DSSSD E_p versus E_n - attachment 4
 gated w(1) & w(2)
 20keV/channel nominal

per DSSSD projection of gated E_p versus E_n spectra - attachments 5-8
 20keV/channel nominal

2D spectra z contours/colour map - attachment 9
Attachment 1: Screenshot_from_2024-06-01_05-24-17.png
Screenshot_from_2024-06-01_05-24-17.png
Attachment 2: Screenshot_from_2024-06-01_05-25-05.png
Screenshot_from_2024-06-01_05-25-05.png
Attachment 3: Screenshot_from_2024-06-01_05-26-10.png
Screenshot_from_2024-06-01_05-26-10.png
Attachment 4: Screenshot_from_2024-06-01_05-26-51.png
Screenshot_from_2024-06-01_05-26-51.png
Attachment 5: Screenshot_from_2024-06-01_05-29-43.png
Screenshot_from_2024-06-01_05-29-43.png
Attachment 6: Screenshot_from_2024-06-01_05-32-16.png
Screenshot_from_2024-06-01_05-32-16.png
Attachment 7: Screenshot_from_2024-06-01_05-33-31.png
Screenshot_from_2024-06-01_05-33-31.png
Attachment 8: Screenshot_from_2024-06-01_05-34-39.png
Screenshot_from_2024-06-01_05-34-39.png
Attachment 9: Screenshot_from_2024-06-01_05-34-57.png
Screenshot_from_2024-06-01_05-34-57.png
  569   Sat Jun 1 07:16:15 2024 CB, AB, EM, JMRun 19 - beam target overlap

Morning shift

08:16 - as run 18, target off

9:00 Increased storage time to 5 minutes. Aquired data after initial beam loss. Attach 1 &2. Far less H2+ vs D.

Temps OK

Rates as before

10:00 beam being optimised to reduce 2H+ content

Rates in aida 13, 14 suddenly spike as during overnight. Checkloaded repeatedly. No effect. Increased thresholds in 13, 14 to 0x32 bringing them in line with 15,16.

12:50

We started test with beam

Test1 ( Attached 12-52-55)

Horiz. offset: 8mm

Max vuel on 1.5MeV =1480 counst

Test2

Horiz. offset: 7mm

 

Test3

Horiz. offset: 6mm

target density=2.4E11

 

Test4

Horiz. offset: 5mm

target density=from 1.7E11 to 2.5E11

 

Test5

Horiz. offset: 4mm

target density=from 5.6E11 to 1.1E11 several fluctuations

 

Test6

Horiz. offset: 3mm

target density=1.1E11 several fluctuations. Until now we do not see a clear  signel from the reaction!

 

Test7

Horiz. offset: 2mm

target density=1.1E11 several fluctuations.

 

Test8

 Beam back again. Coasting beam. Detectors IN

Horiz. offset: 8mm

xy and energy histograms - attachment 12 and 13.

 

Test9

Bunched beam Detectors IN

Horiz. offset: 8mm

xy and energy histograms - attachment 14 and 15.

 

Test10

Horiz. offset: 6mm

xy and energy histograms - attachment

target:1.6E11

xy and energy histograms - attachment 16 and 17.

Afternoon shift

From previous overlap tests we can see that moving the beam position does cause a change in the number of events observed in the detectors.

But the change in events appears to be due to beam halo hitting the detectors rather than elastic scattering. New approach required.

We have varied the beam position (horizontal) in 1mm steps from 0mm to 9mm (the maximum allowed). We do this with the target ON and then with the target OFF.

Integral of events ~1.5 MeV taken . Difference between the target and no target plotted to gauge overlap position.

Note we have moved left arm fully out as we were worried about beam hitting dssds on injection. Will check overlap with right dssds only as left dssds have very low rate at this position.

We have determined the overlap position is about 7.0mm horizontal

Attachment 18 shows initial scan.

Attachment 19 shows a second scan in the region we highlighted as possible overlap.

Attachment 1: Screenshot_from_2024-06-01_09-15-54.png
Screenshot_from_2024-06-01_09-15-54.png
Attachment 2: Screenshot_from_2024-06-01_09-16-11.png
Screenshot_from_2024-06-01_09-16-11.png
Attachment 3: Screenshot_from_2024-06-01_10-23-17.png
Screenshot_from_2024-06-01_10-23-17.png
Attachment 4: Screenshot_from_2024-06-01_10-29-48.png
Screenshot_from_2024-06-01_10-29-48.png
Attachment 5: Screenshot_from_2024-06-01_12-52-55.png
Screenshot_from_2024-06-01_12-52-55.png
Attachment 6: Screenshot_from_2024-06-01_12-58-44.png
Screenshot_from_2024-06-01_12-58-44.png
Attachment 7: Screenshot_from_2024-06-01_13-06-01.png
Screenshot_from_2024-06-01_13-06-01.png
Attachment 8: Screenshot_from_2024-06-01_13-12-10.png
Screenshot_from_2024-06-01_13-12-10.png
Attachment 9: Screenshot_from_2024-06-01_13-18-32.png
Screenshot_from_2024-06-01_13-18-32.png
Attachment 10: Screenshot_from_2024-06-01_13-27-12.png
Screenshot_from_2024-06-01_13-27-12.png
Attachment 11: Screenshot_from_2024-06-01_13-34-07.png
Screenshot_from_2024-06-01_13-34-07.png
Attachment 12: Screenshot_from_2024-06-01_14-18-55.png
Screenshot_from_2024-06-01_14-18-55.png
Attachment 13: Screenshot_from_2024-06-01_14-19-10.png
Screenshot_from_2024-06-01_14-19-10.png
Attachment 14: Screenshot_from_2024-06-01_14-28-08.png
Screenshot_from_2024-06-01_14-28-08.png
Attachment 15: Screenshot_from_2024-06-01_14-29-26.png
Screenshot_from_2024-06-01_14-29-26.png
Attachment 16: Screenshot_from_2024-06-01_14-39-31.png
Screenshot_from_2024-06-01_14-39-31.png
Attachment 17: Screenshot_from_2024-06-01_14-39-43.png
Screenshot_from_2024-06-01_14-39-43.png
Attachment 18: Overlap1.png
Overlap1.png
Attachment 19: Overlap2.png
Overlap2.png
  568   Fri May 31 21:53:20 2024 CB EM, JM,EM ABOvernight beam tests

22:53

Run 15 started

Beam tests. We increased the storage time to 2min and switched to bunched beam so that we can see the beam lifetime over the cycle.

D+ Beam energy is set to 1.5 MeV.

Detectors are positioned in with servos only (left arm = 28.06mm, right arm = 29.23 mm)

No target

We observe a large peak at the start of the cycle at 750 keV. This seems to be molecular hydrogen circulating in the ring. We also see a much smaller peak at 1.5 MeV (D+)

After a period of time the 750 keV peak decreases significantly and we are left with the 1,5 MeV peak only. 

Total energy histogram after running for ~ 1 hour - attachment 1

Total xy plot after ~ 1 hour - attachment 2

We observe the large circular structure on the left hand detectors. Is this related to hydrogen or is it the beam type (bunched instead of coasting)?

In the GSI control plots for the number of ions in the ring we observe a large drop in ions circulating at the start of the cycle, and then remains fairly constant throughout the rest of the cycle. - attachment 3

This seems to confirm our hydrogen hypothesis.

23:54

Run 15 stopped, Run 16 started.

All setting the same and still no target, except beam is now coasting. We will compare to Run 15 after one hour of running.

We do not observe the circular distribution on the detector seen in the previous run. It seems this effect comes from the bunched beam. Reason?

We still observe the 750 keV peak at the start of the cycle.

Total energy histogram after running for ~ 1 hour - attachment 4

Total xy plot after ~ 1 hour - attachment 5

00:59

Run16 stopped. Run 17 started.

All settings the same except we have switched back to bunched beam.

Target now on. (X=-13700, Y=0). Target density ~ 1E11 at/cm2.

02:03

Run 17 continues

Target has increased in the average density to ~4e11 at/cm2, no changes made to nozzle position and fluctuates to 1e11 or 1e12 at/cm2.

Energy histograms and the xy plots appear similar to bunched beam with no target. The rates are slightly lower with the target for the left hand dssds, maybe target interactions destroy molecular hydrogen quicker?

Total energy histogram after running for ~ 1 hour - attachment 6

Total xy plot after ~ 1 hour - attachment 7

We will continue to run for another hour or so in this condition

Target density has decreased over the last minutes. We have tried to move the nozzle to maximise density but fluctuations make this difficult.

03:55

Increase in ADC data items statistics for aida 13 and 14 from ~20 kHz to 90 kHz

ASIC checkload has reduced the rates back to ~ 20 kHz

Total energy histogram after running for ~ 3 hour - attachment 8

Total xy plot after ~ 3 hour - attachment 9

Run 17 stopped. Run 18 started

Beam changed back to coasting. All other settings unchanged. Target on. 

04:55

Run 18 continues.

Energy histograms and the xy plots appear similar to bunched beam with no target. The rates are lower with the target than without.

Total energy histogram after running for ~ 1 hour - attachment 10

Total xy plot after ~ 1 hour - attachment 11

Note we observed fluctuations in the T2 temperature for the target. T2 is set to 80 K but rapid fluctuations were observed up to 100K. This occurred for only 5 minutes and I am not sure if it is a real effect.

 

07:20

Temperatures OK

System wide checks OK

Rates OK

Stats OK

Waveforms attached

Leakage current decreasing slightly.

No changes to beam or target.

Attachment 1: Screenshot_from_2024-05-31_23-52-25.png
Screenshot_from_2024-05-31_23-52-25.png
Attachment 2: Screenshot_from_2024-05-31_23-51-56.png
Screenshot_from_2024-05-31_23-51-56.png
Attachment 3: screenshot-2024-05-31_23-53-125456534733007473468.png
screenshot-2024-05-31_23-53-125456534733007473468.png
Attachment 4: Screenshot_from_2024-06-01_00-59-45.png
Screenshot_from_2024-06-01_00-59-45.png
Attachment 5: Screenshot_from_2024-06-01_00-59-55.png
Screenshot_from_2024-06-01_00-59-55.png
Attachment 6: Screenshot_from_2024-06-01_02-01-33.png
Screenshot_from_2024-06-01_02-01-33.png
Attachment 7: Screenshot_from_2024-06-01_02-02-05.png
Screenshot_from_2024-06-01_02-02-05.png
Attachment 8: Screenshot_from_2024-06-01_03-54-39.png
Screenshot_from_2024-06-01_03-54-39.png
Attachment 9: Screenshot_from_2024-06-01_03-54-52.png
Screenshot_from_2024-06-01_03-54-52.png
Attachment 10: Screenshot_from_2024-06-01_04-54-25.png
Screenshot_from_2024-06-01_04-54-25.png
Attachment 11: Screenshot_from_2024-06-01_04-54-43.png
Screenshot_from_2024-06-01_04-54-43.png
Attachment 12: Screenshot_from_2024-06-01_07-17-28.png
Screenshot_from_2024-06-01_07-17-28.png
Attachment 13: Screenshot_from_2024-06-01_07-17-38.png
Screenshot_from_2024-06-01_07-17-38.png
Attachment 14: Screenshot_from_2024-06-01_07-18-00.png
Screenshot_from_2024-06-01_07-18-00.png
Attachment 15: Screenshot_from_2024-06-01_07-18-12.png
Screenshot_from_2024-06-01_07-18-12.png
Attachment 16: Screenshot_from_2024-06-01_07-19-44.png
Screenshot_from_2024-06-01_07-19-44.png
Attachment 17: Screenshot_from_2024-06-01_07-20-07.png
Screenshot_from_2024-06-01_07-20-07.png
  567   Fri May 31 21:03:21 2024 CB, EMRun14

22:03

Before this run we checked quickly  beam with the target on but it does not make sense with the current status of the beam.

We removed target and starte Run14

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

Attachment 1: Screenshot_from_2024-05-31_22-05-57.png
Screenshot_from_2024-05-31_22-05-57.png
Attachment 2: Screenshot_from_2024-05-31_22-07-01.png
Screenshot_from_2024-05-31_22-07-01.png
Attachment 3: Screenshot_from_2024-05-31_22-07-49.png
Screenshot_from_2024-05-31_22-07-49.png
  566   Fri May 31 19:12:01 2024 CB, EMRun 13 - test with beam

20:00

The beam is set at the correct energy, operators cross-checked it again. What we see  at 750 is due to the molecular beam.  The operotors are trying to optimize and get rid of this in the next 30min.

Temps and system checks OK

Leakage current OK

Merger and tape server OK

Potentiometer at 28.06 mm and 29.23 mm

20:30

Keep going with the beam tests. A new peak appears around 950 keV( See attached). Still try to understand what it is. The operators keep going

Attachment 1: Screenshot_from_2024-05-31_20-13-46.png
Screenshot_from_2024-05-31_20-13-46.png
Attachment 2: Screenshot_from_2024-05-31_20-27-58.png
Screenshot_from_2024-05-31_20-27-58.png
Attachment 3: Screenshot_from_2024-05-31_20-32-59.png
Screenshot_from_2024-05-31_20-32-59.png
  565   Fri May 31 06:01:55 2024 JM, TY, PW, EM, CBRun 12 - beam on target

07:00

Run 12 started. D+ beam is incident on the target at 1.5 MeV

Beam intensity in the ring is ~4e7 particles

Target density shows some fluctuations but is ~1e11 atoms/cm2. Target temp 35K

Detectors are positioned fully out (Left arm 18.86mm and right arm 19.8 mm ) and are not moving

Ring pressure looks OK, YR09 E target sensor seems unaffected by the target currently reads ~5e-11 mbar.

Beam is in the ring for ~15s.

Temps and system checks OK

Leakage current OK

Merger and tape server OK

Horizontal beam position 8 mm and vertical beam position 0mm for overlap with target.

Statistics - attachment 1

Energy histogram - attachment 2

XY plot - attachment 3

We see high rate at 1.5 MeV in left dssds but very little in right dssds. XY plots also show counts are concerntrated in a very circular distribution.

07:15

Gleb has just called and requires access to the cave to check on a turbo pump in the source.

Ring vacuum looks good

Beam will be down for 10-15 minutes only hopefully. 

07:45

Beam is back in the ring but injection requires optimisation.

09:00

There are some issues with the source that require a few hours to resolve. Potentially until 7 or 8 pm tonight. We have about an hour before work starts.

We have done a quick test of the motors and moved the left and right arms in using the pnuematic only.

New position left arm = 62.71 mm , right arm = 64.6 mm.

We observed the rates in left detectors went up significantly. In the right detectors we observed little increase.

See attached xy plot - attachment 4

See attached energy histogram - attachment 5

Beam intensity has dropped significantly - below 1e6 particles so even at in positons we see few counts at the moment.

09:25

Source has degraded pretty rapidly so beam will be stopped for now.

Detectors moved back out.

Target nozzle moved from X= -13000, Y=2500 to X= -11000, Y=2500. S1 has dropped to 5e-10 mbar.

Note Gleb mentioned that last night it was likely a large proportion of the ions stored were molecular hydrogen not deuterium.

16:18

Temps and system checks OK

Leakage current OK

Merger and tape server OK

 

18:10

Temps and system checks OK

Leakage current OK

Merger and tape server OK

 

19:38

Operators are trying to optimize some beam. For now the beam is lower than it should be probably because is easy to work with the source like that.

Temps and system checks OK

Leakage current OK

Merger and tape server OK


 

Attachment 1: Screenshot_from_2024-05-31_07-32-33.png
Screenshot_from_2024-05-31_07-32-33.png
Attachment 2: Screenshot_from_2024-05-31_07-33-19.png
Screenshot_from_2024-05-31_07-33-19.png
Attachment 3: Screenshot_from_2024-05-31_07-33-57.png
Screenshot_from_2024-05-31_07-33-57.png
Attachment 4: Screenshot_from_2024-05-31_09-17-08.png
Screenshot_from_2024-05-31_09-17-08.png
Attachment 5: Screenshot_from_2024-05-31_09-21-40.png
Screenshot_from_2024-05-31_09-21-40.png
Attachment 6: Screenshot_from_2024-05-31_16-19-29.png
Screenshot_from_2024-05-31_16-19-29.png
Attachment 7: Screenshot_from_2024-05-31_16-29-22.png
Screenshot_from_2024-05-31_16-29-22.png
Attachment 8: Screenshot_from_2024-05-31_16-31-36.png
Screenshot_from_2024-05-31_16-31-36.png
Attachment 9: Screenshot_from_2024-05-31_18-12-36.png
Screenshot_from_2024-05-31_18-12-36.png
Attachment 10: Screenshot_from_2024-05-31_19-42-42.png
Screenshot_from_2024-05-31_19-42-42.png
Attachment 11: Screenshot_from_2024-05-31_19-45-55.png
Screenshot_from_2024-05-31_19-45-55.png
Attachment 12: Screenshot_from_2024-05-31_20-04-54.png
Screenshot_from_2024-05-31_20-04-54.png
  564   Thu May 30 13:31:01 2024 JM, CB, PW, YLRun 11 - beam on target overlap

14:30

Run 11 started. D+ beam in the ring at 1.5 MeV.

No target at the moment. Valves to the interaction chamber will soon be opened.

Stats and histograms zero'd, monitoring code reset.

All checks and temps OK

Leakage currents OK

ADC data items statistics - attachment 1

Beam in ring ~30s and manipulation ~2s

Beam intensity ~ 5.7E7

 

16:40

System check OK

Temperature OK

Statistics same as before

Merger and server OK

Leakage currents OK

Beam in the ring

Waiting for target , Hydrogen green

 

18:50

System OK

Temperature OK

Statistics same as before

Merger and server OK

Leakage currents OK

Beam ~4x10^7

Target tuning ongoing

 

22:05

System OK

Temperature OK

Statistics same as before

Merger and server OK

Leakage currents OK

Currently no beam -> asked operator to check source

Target tuning ongoing

 

23:22

System OK

Temperature OK

Statistics same as before

Merger and server OK

Leakage currents OK

Currently no beam -> operators are optimizing the intensity of the beam

Optimizing target density as well

 

00:55

The overlaping test started by moving the beam only in the y axis.

 

02:00

Overlap test is over and was found that the optimal position for the horizontal axis is 8mm. Beam on target.

Observation of an increase the 1.5 MeV peak in the CARME monitoring code when overlap with the target is found. Attachment 4.

Drop in the beam lifetime is also observed when overlapped with the target.

System OK

Temperature OK

Statistics same as before

Merger and server OK

Leakage currents OK

 

02:48

CRYRING measurement cycle has been made shorter (from 5 to 3 repetitions) to match it with the decreased beam life time due to optimed overlap with target.

 

04:37

System OK

Temperature OK

Statistics same as before

Merger and server OK

Leakage currents OK

 

 

 

Attachment 1: Screenshot_from_2024-05-30_14-33-19.png
Screenshot_from_2024-05-30_14-33-19.png
Attachment 2: Screenshot_from_2024-05-31_02-09-35.png
Screenshot_from_2024-05-31_02-09-35.png
Attachment 3: Screenshot_from_2024-05-31_06-57-12.png
Screenshot_from_2024-05-31_06-57-12.png
Attachment 4: 20240531_073828.jpg
20240531_073828.jpg
  563   Thu May 30 09:30:33 2024 CB, PW, YL, TDRun 10

10:30

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

 

10.48 FEE64 temperatures OK - attachment 1

WR timestamps OK - attachment 2

ADC data item stats - attachment 3

per FEE64 Rate spectra - attachment 4

 why do we observe HEC data in aida15 asic #1?

per FEE64 1.8.L spectra - attachments 5-6

 ADC to 20MeV FSR - not clear when these spwectra were last zero;d

 currently 750keV/u D + no target ?

10.53 All histograms & stats zero'd

Merger, Taper Server etc - attachments 7-8

  bind error 98 for NewMerger server - not clear when these messahes were generated - added some returns to monitor whether these, or other, messages are repeated

DSSSD bias & leakage current OK - attachment 9

11:30

We noticed some odd peak structure in the rate energy histograms for the right hand side detectors. We took away the beam but this structure persisted. 

This was not observed in the total energy histograms, the reason for this is the applied multiplicity cut to the total histograms. Removing this cut the peaks are clearly visible (attachment 10). Putting the cut back in we see the peaks removed (attachment 11). These events correspond to some noisy strips on the detectors on the right hand side (see attachment 12). Notice the much larger rate on the RHS. Current ADC data items statistics show FEE modules for the RHS also much larger (attachment 13). We did not observe this with no beam (https://elog.ph.ed.ac.uk/CARME/554) but did with beam last night (https://elog.ph.ed.ac.uk/CARME/558) so there nust be some effect of having beam in the ring increasing the number of events on the RHS. Note the detectors are fully out and the events are localised in these strips. 

Note we have moved the detectorsand removed beam from the ring in this time period to investigate these peaks.

12:50

Run 10 continues. Temps OK. System checks OK. Merger OK, Tape server OK.

Leakage current OK.

Notice the 'beam peak' at 1.5 MeV is larger in the left hand side (attachment 14)

14:26

We have started the new_optimeren.py on pi@gas-stepper to try and align the gas target. Valves to the interaction chamber are closed.

Setting the nozzle X=-13000, Y=1000 we observe an increase in E4 from 4E-10 mbar to 2E-9 mbar. Moving either side we observe a drop in E4. We take this as the gas target being aligned.

Anton will now open the valves to the gas target.

Run 10 stopped.         

Attachment 1: Screenshot_from_2024-05-30_10-48-43.png
Screenshot_from_2024-05-30_10-48-43.png
Attachment 2: Screenshot_from_2024-05-30_10-49-43.png
Screenshot_from_2024-05-30_10-49-43.png
Attachment 3: Screenshot_from_2024-05-30_10-50-34.png
Screenshot_from_2024-05-30_10-50-34.png
Attachment 4: Screenshot_from_2024-05-30_10-51-40.png
Screenshot_from_2024-05-30_10-51-40.png
Attachment 5: Screenshot_from_2024-05-30_10-52-51.png
Screenshot_from_2024-05-30_10-52-51.png
Attachment 6: Screenshot_from_2024-05-30_10-53-53.png
Screenshot_from_2024-05-30_10-53-53.png
Attachment 7: Screenshot_from_2024-05-30_10-56-23.png
Screenshot_from_2024-05-30_10-56-23.png
Attachment 8: Screenshot_from_2024-05-30_10-56-42.png
Screenshot_from_2024-05-30_10-56-42.png
Attachment 9: Screenshot_from_2024-05-30_10-57-56.png
Screenshot_from_2024-05-30_10-57-56.png
Attachment 10: Screenshot_from_2024-05-30_11-35-08.png
Screenshot_from_2024-05-30_11-35-08.png
Attachment 11: Screenshot_from_2024-05-30_11-36-07.png
Screenshot_from_2024-05-30_11-36-07.png
Attachment 12: Screenshot_from_2024-05-30_12-58-20.png
Screenshot_from_2024-05-30_12-58-20.png
Attachment 13: Screenshot_from_2024-05-30_13-10-18.png
Screenshot_from_2024-05-30_13-10-18.png
  562   Thu May 30 08:33:21 2024 CB, PW, YLRun 8 & 9- detector motion tests

09:34 - Moved to Run 9 in preparation of moving detectors. No target yet. Beam in the ring at 750 keV/A.

Temperatures OK (attach)

System wide checks OK except 16 fails clock and ADC calib

Rates OK (attach)

Writing to disk OK (attach)

Resetting the sorting script. Moving detectors.

Moved detectors in. See attached. Stripe / shadow quite clear. Some oddities on the top-right DSSD. May be due to ExEy cuts and multiplicity cuts.

Moved on tun run 9 to acquire some stats without motion.

 

10:13 Temperatures still OK. Rates OK. Leakage currents stable.

We were cutting the beam.

Increased repetition to x5.

Motors fully out - run 10 started.

Attachment 1: Screenshot_from_2024-05-30_09-33-55.png
Screenshot_from_2024-05-30_09-33-55.png
Attachment 2: Screenshot_from_2024-05-30_09-35-41.png
Screenshot_from_2024-05-30_09-35-41.png
Attachment 3: Screenshot_from_2024-05-30_09-36-23.png
Screenshot_from_2024-05-30_09-36-23.png
Attachment 4: Screenshot_from_2024-05-30_09-50-27.png
Screenshot_from_2024-05-30_09-50-27.png
Attachment 5: Screenshot_from_2024-05-30_09-52-29.png
Screenshot_from_2024-05-30_09-52-29.png
Attachment 6: Screenshot_from_2024-05-30_09-57-27.png
Screenshot_from_2024-05-30_09-57-27.png
Attachment 7: Screenshot_from_2024-05-30_10-00-25.png
Screenshot_from_2024-05-30_10-00-25.png
  561   Thu May 30 07:25:53 2024 Yuri, Phil, NikosGas target

The cave is closed and beam is back

The d intensity is about half of the intensity before the break

The gas target is prepared but the valvas cannot be opened. Anton is coming in at around 1 PM to open the gate valves.

  560   Thu May 30 06:00:51 2024 Yuri Nikos ...Open the cave

Nikos needs to enter the cave to work on the target cooling.

 

  559   Wed May 29 17:42:06 2024 TDOffline analysis APR24/R2 (pulser walkthrough)
ADC offsets using pulser walkthrough data from data file APR24/R2

ch = channel + ( module * 64 ) + ( range * 2048 )

adc_data( ch ) = INT( RSHIFT( ABS( adc_data( ch ) - 32768 ), 3 ) - offset( ch ) + 0.5 )




per FEE64 ADC spectra - attachments 1-16

Note common x & y-scales - peak height proxy for peak width

aida01 1.8.L (s7) pulser peak width 9 ch FWHM (~50keV FWHM)




ADC offset analysis

/home/td/Calibration/offsets /home/td/CARME/R2 50 2048 20 5

First pass sort variables NAMELIST I/O format ( ADC offsets, FEE64 configuration, LEC & HEC equal energy cuts ) - attachment 17
Attachment 1: Screenshot_from_2024-05-28_20-54-05.png
Screenshot_from_2024-05-28_20-54-05.png
Attachment 2: Screenshot_from_2024-05-28_20-54-58.png
Screenshot_from_2024-05-28_20-54-58.png
Attachment 3: Screenshot_from_2024-05-28_20-55-45.png
Screenshot_from_2024-05-28_20-55-45.png
Attachment 4: Screenshot_from_2024-05-28_21-14-54.png
Screenshot_from_2024-05-28_21-14-54.png
Attachment 5: Screenshot_from_2024-05-28_21-15-49.png
Screenshot_from_2024-05-28_21-15-49.png
Attachment 6: Screenshot_from_2024-05-28_21-16-20.png
Screenshot_from_2024-05-28_21-16-20.png
Attachment 7: Screenshot_from_2024-05-28_21-16-46.png
Screenshot_from_2024-05-28_21-16-46.png
Attachment 8: Screenshot_from_2024-05-28_21-18-21.png
Screenshot_from_2024-05-28_21-18-21.png
Attachment 9: Screenshot_from_2024-05-28_21-27-03.png
Screenshot_from_2024-05-28_21-27-03.png
Attachment 10: Screenshot_from_2024-05-28_21-27-42.png
Screenshot_from_2024-05-28_21-27-42.png
Attachment 11: Screenshot_from_2024-05-28_21-28-12.png
Screenshot_from_2024-05-28_21-28-12.png
Attachment 12: Screenshot_from_2024-05-28_21-28-46.png
Screenshot_from_2024-05-28_21-28-46.png
Attachment 13: Screenshot_from_2024-05-28_21-29-16.png
Screenshot_from_2024-05-28_21-29-16.png
Attachment 14: Screenshot_from_2024-05-28_21-29-40.png
Screenshot_from_2024-05-28_21-29-40.png
Attachment 15: Screenshot_from_2024-05-28_21-30-14.png
Screenshot_from_2024-05-28_21-30-14.png
Attachment 16: Screenshot_from_2024-05-28_21-30-48.png
Screenshot_from_2024-05-28_21-30-48.png
Attachment 17: variables.dat
  $variables
  offset(   0) =    31.70
  offset(   1) =     1.56
  offset(   2) =   -20.38
  offset(   3) =   -19.76
  offset(   4) =    13.07
  offset(   5) =     4.85
  offset(   6) =   -21.16
  offset(   7) =    22.27
  offset(   8) =    -5.26
  offset(   9) =    -4.43
  offset(  10) =     1.25
  offset(  11) =    -0.48
  offset(  12) =     9.68
  offset(  13) =     2.46
  offset(  14) =    -5.27
  offset(  15) =     9.41
  offset(  16) =     0.90
  offset(  17) =    -2.60
  offset(  18) =    22.25
  offset(  19) =   -19.17
  offset(  20) =     1.48
  offset(  21) =    -9.92
  offset(  22) =    11.92
  offset(  23) =    12.82
  offset(  24) =    12.26
  offset(  25) =    -7.54
  offset(  26) =    12.66
  offset(  27) =   -23.36
  offset(  28) =     9.50
  offset(  29) =    17.75
  offset(  30) =   -23.58
  offset(  31) =    21.56
  offset(  32) =     7.89
  offset(  33) =   -11.31
  offset(  34) =   -25.22
  offset(  35) =    -5.51
  offset(  36) =   -13.56
  offset(  37) =    -3.33
  offset(  38) =   -15.50
  offset(  39) =     0.09
  offset(  40) =   -12.56
  offset(  41) =   -26.14
  offset(  42) =     3.57
  offset(  43) =   -14.70
  offset(  44) =    12.66
  offset(  45) =   -36.78
  offset(  46) =    24.58
  offset(  47) =    -8.98
  offset(  48) =   -18.05
  offset(  49) =    -7.85
  offset(  50) =   -25.85
  offset(  51) =   -15.22
  offset(  52) =   -10.96
  offset(  53) =   -17.87
  offset(  54) =   -29.36
  offset(  55) =    -9.94
  offset(  56) =    10.20
  offset(  57) =   -23.00
  offset(  58) =   -22.69
  offset(  59) =   -21.56
  offset(  60) =    27.94
  offset(  61) =     3.41
  offset(  62) =     6.03
  offset(  63) =   -10.52
  offset(  64) =     7.65
  offset(  65) =     5.94
  offset(  66) =     0.52
  offset(  67) =     1.01
  offset(  68) =    -7.82
  offset(  69) =   -22.59
  offset(  70) =     9.05
  offset(  71) =    23.35
  offset(  72) =    23.79
  offset(  73) =     3.42
  offset(  74) =     2.83
  offset(  75) =     2.65
  offset(  76) =     3.05
  offset(  77) =     6.05
  offset(  78) =   -12.53
  offset(  79) =   -13.37
  offset(  80) =     5.99
  offset(  81) =    -6.94
  offset(  82) =     0.39
  offset(  83) =   -12.79
  offset(  84) =    -1.48
  offset(  85) =    20.06
  offset(  86) =     1.26
  offset(  87) =   -12.90
  offset(  88) =    18.92
  offset(  89) =    -3.11
  offset(  90) =    -8.23
  offset(  91) =    -6.89
  offset(  92) =    -7.46
  offset(  93) =     0.15
  offset(  94) =   -17.01
  offset(  95) =    -8.25
  offset(  96) =   -31.52
  offset(  97) =    16.13
  offset(  98) =    15.45
  offset(  99) =    24.67
  offset( 100) =   -20.48
  offset( 101) =     2.01
  offset( 102) =    -2.56
  offset( 103) =    42.23
  offset( 104) =     4.75
  offset( 105) =    15.76
  offset( 106) =    11.44
  offset( 107) =   -34.81
  offset( 108) =   -28.62
  offset( 109) =   -21.31
  offset( 110) =   -24.33
  offset( 111) =    -5.05
  offset( 112) =    15.62
  offset( 113) =     5.20
  offset( 114) =    42.22
  offset( 115) =    16.79
  offset( 116) =    -0.78
  offset( 117) =   -12.94
  offset( 118) =    24.25
  offset( 119) =    16.62
  offset( 120) =    -6.90
  offset( 121) =    28.25
  offset( 122) =    23.14
  offset( 123) =    11.89
  offset( 124) =    -6.48
  offset( 125) =    10.15
  offset( 126) =    55.36
  offset( 127) =    21.89
  offset( 128) =   -14.38
  offset( 129) =    -4.43
  offset( 130) =   -17.38
  offset( 131) =    35.78
  offset( 132) =   -42.46
  offset( 133) =    -8.70
  offset( 134) =    15.69
  offset( 135) =   -26.25
  offset( 136) =   -21.07
  offset( 137) =   -35.26
  offset( 138) =   -10.76
  offset( 139) =   -31.95
  offset( 140) =     6.11
  offset( 141) =   -19.00
  offset( 142) =   -13.84
  offset( 143) =   -17.20
  offset( 144) =     7.83
  offset( 145) =    -8.38
  offset( 146) =   -15.68
  offset( 147) =    -5.98
  offset( 148) =     9.85
  offset( 149) =   -21.39
  offset( 150) =   -42.83
  offset( 151) =   -55.99
  offset( 152) =    -8.53
  offset( 153) =     0.97
  offset( 154) =   -29.26
  offset( 155) =   -16.40
  offset( 156) =    -3.32
  offset( 157) =   -42.87
  offset( 158) =   -42.72
  offset( 159) =   -26.85
  offset( 160) =   -36.87
  offset( 161) =   -33.16
  offset( 162) =   -10.69
  offset( 163) =   -23.53
  offset( 164) =    10.43
  offset( 165) =   -10.48
  offset( 166) =   -31.70
  offset( 167) =    21.74
  offset( 168) =     2.27
  offset( 169) =    -7.56
  offset( 170) =   -19.47
  offset( 171) =    25.12
  offset( 172) =   -17.19
  offset( 173) =    -7.30
  offset( 174) =   -44.54
  offset( 175) =   -17.26
  offset( 176) =   -29.31
  offset( 177) =   -12.46
  offset( 178) =   -33.61
  offset( 179) = -9999.99
  offset( 180) =   -31.25
  offset( 181) =    -5.14
  offset( 182) =     5.87
  offset( 183) =   -25.05
  offset( 184) =   -34.61
  offset( 185) =   -19.87
  offset( 186) =    -3.71
  offset( 187) =   -30.59
  offset( 188) =     1.07
  offset( 189) =   -48.50
  offset( 190) =   -34.50
  offset( 191) =   -25.18
  offset( 192) =   -37.17
  offset( 193) =   -40.22
  offset( 194) =   -38.76
  offset( 195) =   -52.65
  offset( 196) =   -37.83
  offset( 197) =    -5.82
  offset( 198) =    -7.01
  offset( 199) =    -4.33
  offset( 200) =    -4.06
  offset( 201) =   -36.15
  offset( 202) =   -14.17
  offset( 203) =   -32.62
  offset( 204) =   -39.21
  offset( 205) =    -7.39
  offset( 206) =   -10.53
  offset( 207) =   -36.56
  offset( 208) =   -34.01
  offset( 209) =   -61.07
  offset( 210) =   -15.10
  offset( 211) =   -35.46
  offset( 212) =    -7.47
  offset( 213) =   -35.09
  offset( 214) =   -78.70
  offset( 215) =   -29.20
  offset( 216) =   -22.09
  offset( 217) =   -22.15
  offset( 218) =   -16.24
  offset( 219) =   -31.62
  offset( 220) =   -32.45
  offset( 221) =   -45.59
  offset( 222) =   -45.27
  offset( 223) =    10.54
  offset( 224) =    10.90
  offset( 225) =   -16.94
  offset( 226) =   -15.20
  offset( 227) =    -2.58
  offset( 228) =   -16.18
  offset( 229) =   -13.20
  offset( 230) =   -54.53
  offset( 231) =   -17.51
  offset( 232) =   -29.77
  offset( 233) =   -30.14
  offset( 234) =   -42.68
  offset( 235) =   -36.03
  offset( 236) =   -17.08
  offset( 237) =   -14.88
  offset( 238) =   -39.70
  offset( 239) =   -64.37
  offset( 240) =   -34.07
  offset( 241) =    16.67
  offset( 242) =     4.33
  offset( 243) =   -11.05
  offset( 244) =    11.48
  offset( 245) =   -30.51
  offset( 246) =    12.23
  offset( 247) =   -12.35
  offset( 248) =    14.03
  offset( 249) =   -27.00
  offset( 250) =     1.20
  offset( 251) =   -24.70
  offset( 252) =   -15.46
  offset( 253) =   -18.55
  offset( 254) =     6.72
  offset( 255) =    -7.39
  offset( 256) = -9999.99
  offset( 257) = -9999.99
  offset( 258) =    29.33
  offset( 259) =    21.68
  offset( 260) = -9999.99
  offset( 261) =    33.37
  offset( 262) =     1.22
  offset( 263) =    30.59
  offset( 264) =     2.99
  offset( 265) =    10.75
  offset( 266) =   -27.39
  offset( 267) = -9999.99
  offset( 268) =    -4.27
  offset( 269) =   -12.65
  offset( 270) =     7.85
  offset( 271) =    -5.75
  offset( 272) =   -12.38
  offset( 273) =    -0.29
  offset( 274) = -9999.99
  offset( 275) =   -39.58
  offset( 276) =   -26.97
  offset( 277) =   -19.65
  offset( 278) =   -28.28
  offset( 279) =   -23.94
  offset( 280) =     5.51
  offset( 281) = -9999.99
  offset( 282) =    -1.87
  offset( 283) =   -34.61
  offset( 284) =     6.36
  offset( 285) =   -21.14
  offset( 286) =   -24.61
  offset( 287) =   -25.32
  offset( 288) = -9999.99
  offset( 289) =    -4.38
  offset( 290) =   -13.00
  offset( 291) =    -1.66
  offset( 292) =    34.08
  offset( 293) =    -2.70
  offset( 294) =     2.43
  offset( 295) = -9999.99
  offset( 296) =     1.82
  offset( 297) =     7.79
  offset( 298) =   -27.47
... 747 more lines ...
  558   Wed May 29 16:46:21 2024 NP, AB, PMRun7 - Measurements only with ion beam.

18:19 Start or run 7: In beam has been optimzed up to now by Zoran and Claude. ~7E7 particles at injection. From now on, only the injection beamline is optimized by Operators.

Detectors IN (left arm pot =18.86 mm, right arm pot = 19.80 mm)

No Target

Beam is in the ring for ~6 seconds and total cycle time between injections is ~8 seconds

Injection beamline is still in process of optimization by operators

21:30

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

23:30

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

01:30

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

Beam and Electron Coller failure (~00:35); Beam back (~01:08); Cooler back (~01:20)

03:00

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

Beam and Electron Coller failure (~01:35); Beam back (~01:55); Cooler back (~01:55)

04:40

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

06:00

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

07:00

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

 

 

Attachment 1: Screenshot_from_2024-05-29_21-49-57.png
Screenshot_from_2024-05-29_21-49-57.png
Attachment 2: Screenshot_from_2024-05-29_21-50-14.png
Screenshot_from_2024-05-29_21-50-14.png
  557   Wed May 29 09:49:35 2024 JM, CBThreshold changes / D+ beam

09:00

R4 alpha background continues, currently R4_12

Temps OK, system checks unchanged from previous status (OK, aida16 clock and ADC issue)

10:30

R4 stopped and moved to the next run R5.

Lithium 3+ beam is not feasible at sufficient densities so we are switching to D+ at a beam energy of 0.75 MeV/u. A measurement of d+p at this energy would produce 3He at energies of ~900 keV. 

The current threshold (slow comparator) has been 1 MeV (0x64). We lowered the thresholds in all FEE modules to 200 keV (0x14). ADC rates look fine in all FEEs (~tens of Hz) except in aida 15 and aida 16 which was ~several hundred kHz. Data rate to disk was 1.2 Mb/s

We increased the threshold in aida 15 and 16 only to 500 keV (0x32) and rates in aida 15 and 16 have dropped to <20 kHz. Data rate to disk now 220 kB/s.

11:00

R5 stopped. R6 started - background with reduced thresholds. 

Attachment 1 - ADC data items statistics

Attachment 2 - Histogram browser statistics

Attachment 3 - Tape server

Attachment 4 - New Merger

13:12

R6 continues R6_9

Edits to the CARME sort code have been made. EyEx energy cut for online monitoring reduced to 100 keV. Multiplicity cut applies to lowEnergyEy, lowEnergyEx and lowEnergyXY histograms. Multiplicity cut considers only single strip events.

15:00

D+ beam is now in the ring and is being optimised by the beam operators. The CARME detectors are currently fullt out, but we can see some effects of the beam on the online monitoring code.

Attachment 1: Screenshot_from_2024-05-29_11-00-20.png
Screenshot_from_2024-05-29_11-00-20.png
Attachment 2: Screenshot_from_2024-05-29_11-01-30.png
Screenshot_from_2024-05-29_11-01-30.png
Attachment 3: Screenshot_from_2024-05-29_11-01-47.png
Screenshot_from_2024-05-29_11-01-47.png
Attachment 4: Screenshot_from_2024-05-29_11-02-03.png
Screenshot_from_2024-05-29_11-02-03.png
  556   Wed May 29 08:10:32 2024 JMRun4 - alpha background

09:00

R4 alpha background continues, currently R4_12

Temps OK, system checks unchanged from previous status (OK, aida16 clock and ADC issue)

  555   Tue May 28 15:25:21 2024 TD, JM, CBPulser tests

R3 started

Pulser ON. Approx 2 kHZ.

1.8*w waveforms attachments 1+2

ADC data items statistics - attchment 3

1.8*L spectra - attachments 4+5

Pulser widths are larger than during testing prior to the Feb beamtime. Widths in Channels;

aida01 = 76

aida02 = 82

aida03 = will upate all later tonight..

R3 ended. Pulser OFF

17:15

R4 Started. Alpha background run.

 

Attachment 1: Screenshot_from_2024-05-28_16-14-56.png
Screenshot_from_2024-05-28_16-14-56.png
Attachment 2: Screenshot_from_2024-05-28_16-15-43.png
Screenshot_from_2024-05-28_16-15-43.png
Attachment 3: Screenshot_from_2024-05-28_16-16-18.png
Screenshot_from_2024-05-28_16-16-18.png
Attachment 4: Screenshot_from_2024-05-28_16-20-05.png
Screenshot_from_2024-05-28_16-20-05.png
Attachment 5: Screenshot_from_2024-05-28_16-28-46.png
Screenshot_from_2024-05-28_16-28-46.png
  554   Tue May 28 12:38:31 2024 TD, JM, CBRun 2 - Alpha background run
13.35 FEE64 temperatures OK - attachment 1

      ADC data item rates OK - attachment 2
       overall rate dominated by 2x hot channels aida15

     per FEE64 Stat spectra - attachment 3
      spectra consistent with April alpha background test 
      note high HEC data stats for aida12 asic#2 - can probably be fixed by ASIC control check

     per 1.8.W spectra - 20us FSR - attachments 4-5
      note 100kHz ripple observed for some FEE64s cf. https://elog.ph.ed.ac.uk/CARME/544

Run stopped 15:15 
Attachment 1: Screenshot_from_2024-05-28_13-35-03.png
Screenshot_from_2024-05-28_13-35-03.png
Attachment 2: Screenshot_from_2024-05-28_13-34-30.png
Screenshot_from_2024-05-28_13-34-30.png
Attachment 3: Screenshot_from_2024-05-28_13-33-19.png
Screenshot_from_2024-05-28_13-33-19.png
Attachment 4: Screenshot_from_2024-05-28_13-36-30.png
Screenshot_from_2024-05-28_13-36-30.png
Attachment 5: Screenshot_from_2024-05-28_13-37-16.png
Screenshot_from_2024-05-28_13-37-16.png
  553   Mon May 27 15:25:08 2024 JMStartup

27/05/24

The carme-gsi computer has been moved from the server room into the shifter control room.

Startup of the DAQ looks OK. Settings loaded 2024Apr24

System wide checks OK. Aida16 clock status fails. Aida16 ADC calibration fails. Rest OK.

All FEEs look OK, we again see high rates in aida15 ~ 12k data items/sec, rate in other FEEs is 0.

Testing of the pulser widths has been done. Pulser widths will be uploaded soon.

Background run started. No beam, no target. File G22-201 and R2. Data being written is ~128 kb/s. Run will be left overnight, no night shift.

28/05/24

DAQ continues with alpha background run.

System wide checks OK. Aida16 clock status fails. Aida16 ADC calibration fails. Rest OK. Temps OK

Data can be seen in the spectrum browser on MIDAS. CARME monitoring code was started incorrectly yesterday. It has now be restarted and a small number of events are seen

  552   Fri May 10 21:23:35 2024 TD, CB, JM20Ne7+ test beam

22.00

Carsten Brandau stopping e-target experiment Unattended beam overnight. No operators watching.

10^5 (nominal) 20Ne7+ at 380keV/u beam available

Carsten reports beam is unstable, often disappearing, and in bunched mode.

4x MSL type TTT14 DSSSDs not moving, locked in out position

Starting CARME MIDAS DAQ ASIC settings 2024Apr24-14.14.03

slow comparator 0x64 E

G&G Ortec 448 Pulser OFF

DSSSD bias & leakage currents - attachment 1

FEE64 temperatures OK - attachment 2

All system wide checks OK, WR timestamps OK - attachment 3

ADC data item stats - attachment 4

Rates dominated by two hot channels aida15 per FEE64 Stat spectra - attachment 5

per FEE64 1.8.L spectra - attachments 6-7 no obvious beam energy peak, continuum of energies from below beam energy to threshold

per FEE64 1.8.W spectra - attachments 8-9

 

22.55 Start data file APR24/R5 Merger, TapeServer etc - attachments 10-12

270k data items/s

 

23:29 Activated CARMESort script. See attachments 13-15.

No obvious beam or signal of any kind. Temperatures OK.

 

00:13 DAQ continues OK. System-wide check OK. Temps OK. Still no clear sign of beam.

 

01:03 Daq continues OK. System wide checks OK, Temps OK.

Only low energies observed in the energy spectrum. Looks a bit like the low energy noise observed in our previous runs with no/very few counts above what would be beam energy (~7.6 MeV)

Note the xy plots are not all orientated correctly (strips are correct in the sort code but root axis need to be flipped - I can;t remember how atm). The part of the plots with counts are those closest to the beam. 

02:00 DAQ continues OK. System wide checks OK. Temps OK.

Spectra still looks the same. I have noticed the right hand side rate is higher than the left. EyEx cut spectrum also looks like much more counts outside of the cut.

03:40

DAQ continues. All checks and temps OK.

I can now see a small peak at 5 MeV in the bottom left dssd. This dssd also has the least number of counts in the low energy continuum.

I have logged back into grafana on x2go to get the number ions in the ring. Attachment 17

04:23 (05:23 EU)

DAQ stopped, merger stopped, tape server stopped. Bias OFF. FEEs turned OFF. Screen shots taken of the monitoring code plots - not yet uploaded.

Windows and servers remain up in case we want to do some more tests in the near future.

Attachment 18 shows beam intensity for total run time from Grafana

Attachment 1: Screenshot_from_2024-05-10_22-31-48.png
Screenshot_from_2024-05-10_22-31-48.png
Attachment 2: Screenshot_from_2024-05-10_22-32-52.png
Screenshot_from_2024-05-10_22-32-52.png
Attachment 3: Screenshot_from_2024-05-10_22-33-45.png
Screenshot_from_2024-05-10_22-33-45.png
Attachment 4: Screenshot_from_2024-05-10_22-34-59.png
Screenshot_from_2024-05-10_22-34-59.png
Attachment 5: Screenshot_from_2024-05-10_22-37-12.png
Screenshot_from_2024-05-10_22-37-12.png
Attachment 6: Screenshot_from_2024-05-10_22-38-32.png
Screenshot_from_2024-05-10_22-38-32.png
Attachment 7: Screenshot_from_2024-05-10_22-39-48.png
Screenshot_from_2024-05-10_22-39-48.png
Attachment 8: Screenshot_from_2024-05-10_22-40-38.png
Screenshot_from_2024-05-10_22-40-38.png
Attachment 9: Screenshot_from_2024-05-10_22-41-24.png
Screenshot_from_2024-05-10_22-41-24.png
Attachment 10: Screenshot_from_2024-05-10_22-56-53.png
Screenshot_from_2024-05-10_22-56-53.png
Attachment 11: Screenshot_from_2024-05-10_22-56-20.png
Screenshot_from_2024-05-10_22-56-20.png
Attachment 12: Screenshot_from_2024-05-10_22-56-35.png
Screenshot_from_2024-05-10_22-56-35.png
Attachment 13: Screenshot_from_2024-05-10_23-28-37.png
Screenshot_from_2024-05-10_23-28-37.png
Attachment 14: Screenshot_from_2024-05-10_23-28-45.png
Screenshot_from_2024-05-10_23-28-45.png
Attachment 15: Screenshot_from_2024-05-10_23-28-51.png
Screenshot_from_2024-05-10_23-28-51.png
Attachment 16: Screenshot_from_2024-05-11_00-12-43.png
Screenshot_from_2024-05-11_00-12-43.png
Attachment 17: 20240511_025056.jpg
20240511_025056.jpg
Attachment 18: 20240511_042236.jpg
20240511_042236.jpg
  550   Sun Apr 28 23:12:11 2024 TDMonday 28 April
00.09 DAQ continues data file R3_21

      DSSSD bias & leakage current OK - attachment 1

      FEE64 temps OK - attachment 2

      ADC data item stats OK - attachment 3

      per FEE64 Rate spectra - attachment 4

      Merger, TapeServer etc - attachments 5-7

04.50 DAQ continues data file R3_22

      DSSSD bias & leakage current OK - attachment 8

      FEE64 temps OK - attachment 9

      ADC data item stats OK - attachment 10

      per FEE64 Rate spectra - attachment 11

      Merger, TapeServer etc - attachments 12-14


09.41 DAQ continues data file R3_23

      DSSSD bias & leakage current OK - attachment 15

      FEE64 temps OK - attachment 16

      ADC data item stats OK - attachment 17

      per FEE64 Rate spectra - attachment 18

      Merger, TapeServer etc - attachments 19-21


09.43 DAQ stopped
      DSSSD bias OFF
      FEE64 power OFF
Attachment 1: Screenshot_from_2024-04-29_00-09-18.png
Screenshot_from_2024-04-29_00-09-18.png
Attachment 2: Screenshot_from_2024-04-29_00-09-45.png
Screenshot_from_2024-04-29_00-09-45.png
Attachment 3: Screenshot_from_2024-04-29_00-10-33.png
Screenshot_from_2024-04-29_00-10-33.png
Attachment 4: Screenshot_from_2024-04-29_00-10-52.png
Screenshot_from_2024-04-29_00-10-52.png
Attachment 5: Screenshot_from_2024-04-29_00-11-04.png
Screenshot_from_2024-04-29_00-11-04.png
Attachment 6: Screenshot_from_2024-04-29_00-11-29.png
Screenshot_from_2024-04-29_00-11-29.png
Attachment 7: Screenshot_from_2024-04-29_00-11-47.png
Screenshot_from_2024-04-29_00-11-47.png
Attachment 8: Screenshot_from_2024-04-29_04-50-25.png
Screenshot_from_2024-04-29_04-50-25.png
Attachment 9: Screenshot_from_2024-04-29_04-50-52.png
Screenshot_from_2024-04-29_04-50-52.png
Attachment 10: Screenshot_from_2024-04-29_04-51-38.png
Screenshot_from_2024-04-29_04-51-38.png
Attachment 11: Screenshot_from_2024-04-29_04-52-01.png
Screenshot_from_2024-04-29_04-52-01.png
Attachment 12: Screenshot_from_2024-04-29_04-52-43.png
Screenshot_from_2024-04-29_04-52-43.png
Attachment 13: Screenshot_from_2024-04-29_04-53-05.png
Screenshot_from_2024-04-29_04-53-05.png
Attachment 14: Screenshot_from_2024-04-29_04-53-22.png
Screenshot_from_2024-04-29_04-53-22.png
Attachment 15: Screenshot_from_2024-04-29_09-41-28.png
Screenshot_from_2024-04-29_09-41-28.png
Attachment 16: Screenshot_from_2024-04-29_09-42-01.png
Screenshot_from_2024-04-29_09-42-01.png
Attachment 17: Screenshot_from_2024-04-29_09-42-43.png
Screenshot_from_2024-04-29_09-42-43.png
Attachment 18: Screenshot_from_2024-04-29_09-42-58.png
Screenshot_from_2024-04-29_09-42-58.png
Attachment 19: Screenshot_from_2024-04-29_09-43-08.png
Screenshot_from_2024-04-29_09-43-08.png
Attachment 20: Screenshot_from_2024-04-29_09-43-28.png
Screenshot_from_2024-04-29_09-43-28.png
Attachment 21: Screenshot_from_2024-04-29_09-43-41.png
Screenshot_from_2024-04-29_09-43-41.png
  549   Sat Apr 27 23:16:02 2024 TDSunday 28 April
00.12 DAQ continues data file R3_15

      DSSSD bias & leakage current OK - attachment 1

      FEE64 temps OK - attachment 2

      ADC data item stats OK - attachment 3

      per FEE64 Rate spectra - attachment 4

      Merger, TapeServer etc - attachments 5-7


05.32 DAQ continues data file R3_16

      DSSSD bias & leakage current OK - attachment 8

      FEE64 temps OK - attachment 9

      ADC data item stats OK - attachment 10

      per FEE64 Rate spectra - attachment 11

      Merger, TapeServer etc - attachments 12-14

11.27 DAQ continues data file R3_17

      DSSSD bias & leakage current OK - attachment 15

      FEE64 temps OK - attachment 16

      ADC data item stats OK - attachment 17

      per FEE64 Rate spectra - attachment 21

      Merger, TapeServer etc - attachments 18-20

11.33 aida16 asic #2 2x Check ASIC control
      aida16 asic #4 2x Check ASIC control


18.53 DAQ continues data file R3_19

      DSSSD bias & leakage current OK - attachment 22

      FEE64 temps OK - attachment 23

      ADC data item stats OK - attachment 24

      per FEE64 Rate spectra - attachment 25

      Merger, TapeServer etc - attachments 26-28

      per FEE64 1.8.H - attachments 29-30

      per FEE64 1.8.L - attachments 31-32

      per FEE64 1.8.W - 20us FSR - attachments 33-36

19.10 CRYRING ambient temperature per https://web-docs.gsi.de/~lestinsk/tempplot.php - attachment 37
Attachment 1: Screenshot_from_2024-04-28_00-11-59.png
Screenshot_from_2024-04-28_00-11-59.png
Attachment 2: Screenshot_from_2024-04-28_00-12-25.png
Screenshot_from_2024-04-28_00-12-25.png
Attachment 3: Screenshot_from_2024-04-28_00-13-20.png
Screenshot_from_2024-04-28_00-13-20.png
Attachment 4: Screenshot_from_2024-04-28_00-14-35.png
Screenshot_from_2024-04-28_00-14-35.png
Attachment 5: Screenshot_from_2024-04-28_00-15-03.png
Screenshot_from_2024-04-28_00-15-03.png
Attachment 6: Screenshot_from_2024-04-28_00-15-40.png
Screenshot_from_2024-04-28_00-15-40.png
Attachment 7: Screenshot_from_2024-04-28_00-15-20.png
Screenshot_from_2024-04-28_00-15-20.png
Attachment 8: Screenshot_from_2024-04-28_05-32-48.png
Screenshot_from_2024-04-28_05-32-48.png
Attachment 9: Screenshot_from_2024-04-28_05-33-11.png
Screenshot_from_2024-04-28_05-33-11.png
Attachment 10: Screenshot_from_2024-04-28_05-34-03.png
Screenshot_from_2024-04-28_05-34-03.png
Attachment 11: Screenshot_from_2024-04-28_05-34-36.png
Screenshot_from_2024-04-28_05-34-36.png
Attachment 12: Screenshot_from_2024-04-28_05-34-50.png
Screenshot_from_2024-04-28_05-34-50.png
Attachment 13: Screenshot_from_2024-04-28_05-35-08.png
Screenshot_from_2024-04-28_05-35-08.png
Attachment 14: Screenshot_from_2024-04-28_05-35-22.png
Screenshot_from_2024-04-28_05-35-22.png
Attachment 15: Screenshot_from_2024-04-28_11-27-15.png
Screenshot_from_2024-04-28_11-27-15.png
Attachment 16: Screenshot_from_2024-04-28_11-27-35.png
Screenshot_from_2024-04-28_11-27-35.png
Attachment 17: Screenshot_from_2024-04-28_11-27-52.png
Screenshot_from_2024-04-28_11-27-52.png
Attachment 18: Screenshot_from_2024-04-28_11-28-48.png
Screenshot_from_2024-04-28_11-28-48.png
Attachment 19: Screenshot_from_2024-04-28_11-29-02.png
Screenshot_from_2024-04-28_11-29-02.png
Attachment 20: Screenshot_from_2024-04-28_11-29-13.png
Screenshot_from_2024-04-28_11-29-13.png
Attachment 21: Screenshot_from_2024-04-28_11-31-05.png
Screenshot_from_2024-04-28_11-31-05.png
Attachment 22: Screenshot_from_2024-04-28_18-53-09.png
Screenshot_from_2024-04-28_18-53-09.png
Attachment 23: Screenshot_from_2024-04-28_18-53-32.png
Screenshot_from_2024-04-28_18-53-32.png
Attachment 24: Screenshot_from_2024-04-28_18-53-54.png
Screenshot_from_2024-04-28_18-53-54.png
Attachment 25: Screenshot_from_2024-04-28_18-54-40.png
Screenshot_from_2024-04-28_18-54-40.png
Attachment 26: Screenshot_from_2024-04-28_18-54-50.png
Screenshot_from_2024-04-28_18-54-50.png
Attachment 27: Screenshot_from_2024-04-28_18-55-06.png
Screenshot_from_2024-04-28_18-55-06.png
Attachment 28: Screenshot_from_2024-04-28_18-55-17.png
Screenshot_from_2024-04-28_18-55-17.png
Attachment 29: Screenshot_from_2024-04-28_19-00-46.png
Screenshot_from_2024-04-28_19-00-46.png
Attachment 30: Screenshot_from_2024-04-28_19-01-27.png
Screenshot_from_2024-04-28_19-01-27.png
Attachment 31: Screenshot_from_2024-04-28_18-59-12.png
Screenshot_from_2024-04-28_18-59-12.png
Attachment 32: Screenshot_from_2024-04-28_19-00-08.png
Screenshot_from_2024-04-28_19-00-08.png
Attachment 33: Screenshot_from_2024-04-28_19-01-57.png
Screenshot_from_2024-04-28_19-01-57.png
Attachment 34: Screenshot_from_2024-04-28_19-02-21.png
Screenshot_from_2024-04-28_19-02-21.png
Attachment 35: Screenshot_from_2024-04-28_19-02-53.png
Screenshot_from_2024-04-28_19-02-53.png
Attachment 36: Screenshot_from_2024-04-28_19-03-18.png
Screenshot_from_2024-04-28_19-03-18.png
Attachment 37: CRY_temps.PNG
CRY_temps.PNG
  548   Fri Apr 26 23:41:16 2024 TDSaturday 27 April
00.34 DAQ continues data file R3_9

      DSSSD bias & leakage current OK - attachment 1

      FEE64 temps OK - attachment 2

      ADC data item stats OK - attachment 3

      per FEE64 Rate spectra - attachment 4

      Merger, TapeServer etc - attachments 5-7

06.41 DAQ continues data file R3_10

      DSSSD bias & leakage current OK - attachment 8

      FEE64 temps OK - attachment 9

      ADC data item stats OK - attachment 10

      per FEE64 Rate spectra - attachment 11

      Merger, TapeServer etc - attachments 12-14


10.56 DAQ continues data file R3_11

      DSSSD bias & leakage current OK - attachment 15

      FEE64 temps OK - attachment 16

      ADC data item stats OK - attachment 17

      per FEE64 Rate spectra - attachment 18

      Merger, TapeServer etc - attachments 19-21

14.33 DAQ continues data file R3_12

      DSSSD bias & leakage current OK - attachment 22

      FEE64 temps OK - attachment 23

      ADC data item stats OK - attachment 24

      per FEE64 Rate spectra - attachment 25

      Merger, TapeServer etc - attachments 26-28

18.31 DAQ continues data file R3_13

      DSSSD bias & leakage current OK - attachment 29

      FEE64 temps OK - attachment 30

      ADC data item stats OK - attachment 31

      per FEE64 Rate spectra - attachment 32

      per FEE64 1.8.L spectra - attachments 33-34

      Merger, TapeServer etc - attachments 35-37
Attachment 1: Screenshot_from_2024-04-27_00-34-52.png
Screenshot_from_2024-04-27_00-34-52.png
Attachment 2: Screenshot_from_2024-04-27_00-35-25.png
Screenshot_from_2024-04-27_00-35-25.png
Attachment 3: Screenshot_from_2024-04-27_00-36-07.png
Screenshot_from_2024-04-27_00-36-07.png
Attachment 4: Screenshot_from_2024-04-27_00-36-34.png
Screenshot_from_2024-04-27_00-36-34.png
Attachment 5: Screenshot_from_2024-04-27_00-36-49.png
Screenshot_from_2024-04-27_00-36-49.png
Attachment 6: Screenshot_from_2024-04-27_00-37-14.png
Screenshot_from_2024-04-27_00-37-14.png
Attachment 7: Screenshot_from_2024-04-27_00-37-37.png
Screenshot_from_2024-04-27_00-37-37.png
Attachment 8: Screenshot_from_2024-04-27_06-41-34.png
Screenshot_from_2024-04-27_06-41-34.png
Attachment 9: Screenshot_from_2024-04-27_06-42-06.png
Screenshot_from_2024-04-27_06-42-06.png
Attachment 10: Screenshot_from_2024-04-27_06-42-48.png
Screenshot_from_2024-04-27_06-42-48.png
Attachment 11: Screenshot_from_2024-04-27_06-43-08.png
Screenshot_from_2024-04-27_06-43-08.png
Attachment 12: Screenshot_from_2024-04-27_06-43-19.png
Screenshot_from_2024-04-27_06-43-19.png
Attachment 13: Screenshot_from_2024-04-27_06-43-36.png
Screenshot_from_2024-04-27_06-43-36.png
Attachment 14: Screenshot_from_2024-04-27_06-43-50.png
Screenshot_from_2024-04-27_06-43-50.png
Attachment 15: Screenshot_from_2024-04-27_10-55-54.png
Screenshot_from_2024-04-27_10-55-54.png
Attachment 16: Screenshot_from_2024-04-27_10-56-21.png
Screenshot_from_2024-04-27_10-56-21.png
Attachment 17: Screenshot_from_2024-04-27_10-56-38.png
Screenshot_from_2024-04-27_10-56-38.png
Attachment 18: Screenshot_from_2024-04-27_10-57-26.png
Screenshot_from_2024-04-27_10-57-26.png
Attachment 19: Screenshot_from_2024-04-27_10-57-45.png
Screenshot_from_2024-04-27_10-57-45.png
Attachment 20: Screenshot_from_2024-04-27_10-58-04.png
Screenshot_from_2024-04-27_10-58-04.png
Attachment 21: Screenshot_from_2024-04-27_10-58-16.png
Screenshot_from_2024-04-27_10-58-16.png
Attachment 22: Screenshot_from_2024-04-27_14-33-33.png
Screenshot_from_2024-04-27_14-33-33.png
Attachment 23: Screenshot_from_2024-04-27_14-34-04.png
Screenshot_from_2024-04-27_14-34-04.png
Attachment 24: Screenshot_from_2024-04-27_14-34-52.png
Screenshot_from_2024-04-27_14-34-52.png
Attachment 25: Screenshot_from_2024-04-27_14-35-08.png
Screenshot_from_2024-04-27_14-35-08.png
Attachment 26: Screenshot_from_2024-04-27_14-35-19.png
Screenshot_from_2024-04-27_14-35-19.png
Attachment 27: Screenshot_from_2024-04-27_14-35-50.png
Screenshot_from_2024-04-27_14-35-50.png
Attachment 28: Screenshot_from_2024-04-27_14-35-34.png
Screenshot_from_2024-04-27_14-35-34.png
Attachment 29: Screenshot_from_2024-04-27_18-31-01.png
Screenshot_from_2024-04-27_18-31-01.png
Attachment 30: Screenshot_from_2024-04-27_18-31-22.png
Screenshot_from_2024-04-27_18-31-22.png
Attachment 31: Screenshot_from_2024-04-27_18-31-40.png
Screenshot_from_2024-04-27_18-31-40.png
Attachment 32: Screenshot_from_2024-04-27_18-32-27.png
Screenshot_from_2024-04-27_18-32-27.png
Attachment 33: Screenshot_from_2024-04-27_18-33-04.png
Screenshot_from_2024-04-27_18-33-04.png
Attachment 34: Screenshot_from_2024-04-27_18-33-47.png
Screenshot_from_2024-04-27_18-33-47.png
Attachment 35: Screenshot_from_2024-04-27_18-34-04.png
Screenshot_from_2024-04-27_18-34-04.png
Attachment 36: Screenshot_from_2024-04-27_18-34-22.png
Screenshot_from_2024-04-27_18-34-22.png
Attachment 37: Screenshot_from_2024-04-27_18-34-44.png
Screenshot_from_2024-04-27_18-34-44.png
  547   Fri Apr 26 07:05:01 2024 TDFriday 26 April
08.02 DAQ continues data file R1_14

      DSSSD bias & leakage current OK - attachment 1

      FEE64 temps OK - attachment 2

      ADC data item stats OK - attachment 3

      per FEE64 Rate spectra - attachment 4 

      Merger, TapeServer etc - attachments 5-7

10.40 Close file R1 and open file R2

      Switch on pulser (relay line 50Hz)

      per FEE64 Rate spectra - attachment 8

      Check ASIC load x2 for aida09, aida10, aida12, aida15  aida16

      per FEE64 Rate spectra - attachment 12
       status quo ante

      Pulser walkthrough 

      Pulser settings per https://elog.ph.ed.ac.uk/CARME/544
      Amplituides 70,000-10,000 @ 10,000 step

      per FEE64 1.8.L spectra - attachments 9-10
 
      per FEE64 Rate spectra - attachment 11

11.05 Pulser OFF
      Close file R2, open file R3

      zero all histograms  stats

      background alpha run continues


20.33 DAQ continues data file R3_8

      DSSSD bias & leakage current OK - attachment 13

      FEE64 temps OK - attachment 14

      ADC data item stats OK - attachment 15

      per FEE64 Rate spectra - attachment 16

      Merger, TapeServer etc - attachments 1-19


      CAEN data log stopped 00:21 Friday 26 April 

pi@nnrpi1:~/Programs/caenlogger $ caenlogger --config caenlogger_config_gsi.toml
2024-04-25 15:25:16.194 | INFO     | caenlogger.__main__:main:117 - Config file provided.
Traceback (most recent call last):
  File "/home/pi/.local/bin/caenlogger", line 10, in <module>
    sys.exit(main())
  File "/home/pi/.local/lib/python3.7/site-packages/caenlogger/__main__.py", line 135, in main
    curses.wrapper(process, config_dic)
  File "/usr/lib/python3.7/curses/__init__.py", line 94, in wrapper
    return func(stdscr, *args, **kwds)
  File "/home/pi/.local/lib/python3.7/site-packages/caenlogger/__main__.py", line 57, in process
    Vmon = [ f for f in HV[5].strip().split("  ") if f ]
IndexError: list index out of range

     Old log gzip'd and moved to directory ~/logs

[npg@carme-gsi ~]$ mv caen_current.dat.gz logs/0031260424-caen_current.dat.gz

20.45 Restarted logger program
Attachment 1: Screenshot_from_2024-04-26_08-02-37.png
Screenshot_from_2024-04-26_08-02-37.png
Attachment 2: Screenshot_from_2024-04-26_08-02-57.png
Screenshot_from_2024-04-26_08-02-57.png
Attachment 3: Screenshot_from_2024-04-26_08-03-17.png
Screenshot_from_2024-04-26_08-03-17.png
Attachment 4: Screenshot_from_2024-04-26_08-03-56.png
Screenshot_from_2024-04-26_08-03-56.png
Attachment 5: Screenshot_from_2024-04-26_08-04-06.png
Screenshot_from_2024-04-26_08-04-06.png
Attachment 6: Screenshot_from_2024-04-26_08-04-25.png
Screenshot_from_2024-04-26_08-04-25.png
Attachment 7: Screenshot_from_2024-04-26_08-04-41.png
Screenshot_from_2024-04-26_08-04-41.png
Attachment 8: Screenshot_from_2024-04-26_10-38-00.png
Screenshot_from_2024-04-26_10-38-00.png
Attachment 9: Screenshot_from_2024-04-26_11-01-53.png
Screenshot_from_2024-04-26_11-01-53.png
Attachment 10: Screenshot_from_2024-04-26_11-02-51.png
Screenshot_from_2024-04-26_11-02-51.png
Attachment 11: Screenshot_from_2024-04-26_11-03-23.png
Screenshot_from_2024-04-26_11-03-23.png
Attachment 12: Screenshot_from_2024-04-26_10-43-19.png
Screenshot_from_2024-04-26_10-43-19.png
Attachment 13: Screenshot_from_2024-04-26_20-33-24.png
Screenshot_from_2024-04-26_20-33-24.png
Attachment 14: Screenshot_from_2024-04-26_20-33-49.png
Screenshot_from_2024-04-26_20-33-49.png
Attachment 15: Screenshot_from_2024-04-26_20-34-30.png
Screenshot_from_2024-04-26_20-34-30.png
Attachment 16: Screenshot_from_2024-04-26_20-36-04.png
Screenshot_from_2024-04-26_20-36-04.png
Attachment 17: Screenshot_from_2024-04-26_20-36-17.png
Screenshot_from_2024-04-26_20-36-17.png
Attachment 18: Screenshot_from_2024-04-26_20-36-37.png
Screenshot_from_2024-04-26_20-36-37.png
Attachment 19: Screenshot_from_2024-04-26_20-36-49.png
Screenshot_from_2024-04-26_20-36-49.png
  546   Thu Apr 25 02:31:25 2024 TDThursday 25 April
03.27 DAQ continues - file APR24/R1_7

      DSSSD bias & leakage current OK - attachment 1
      ambient temperature 16.1 deg C

      FEE64 temperatures OK - attachment 2

      ADC data item stats OK - attachment 3

      per n+n FEE64 1.8.L spectra - attachment 4

      per FEE64 Stat spectra - attachment 5

      Merger, TapeServer etc - attachments 6-8


10.19 DAQ continues - file APR24/R1_8

      DSSSD bias & leakage current OK - attachment 9
      ambient temperature 16.2 deg C

      FEE64 temperatures OK - attachment 10

      ADC data item stats OK - attachment 11

      per FEE64 Stat spectra - attachment 12

      Merger, TapeServer etc - attachments 13-15

      NetVar merge variables - attachment 16


14.44 DAQ continues - file APR24/R1_10

      DSSSD bias & leakage current OK - attachment 17
      ambient temperature 16.5 deg C

      FEE64 temperatures OK - attachment 18

      ADC data item stats OK - attachment 19

      per FEE64 Stat spectra - attachment 20
       some HEC data has appeared

      Merger, TapeServer etc - attachments 21-23

19.12 DAQ continues - file APR24/R1_11

      DSSSD bias & leakage current OK - attachment 24
      ambient temperature 16.5 deg C

      FEE64 temperatures OK - attachment 25

      ADC data item stats OK - attachment 26

      per FEE64 Stat spectra - attachment 27

      per FEE64 1.8.L spectra - attachments 28-29

      Merger, TapeServer etc - attachments 30-32

23.14 DAQ continues - file APR24/R1_12

      DSSSD bias & leakage current OK - attachment 33
      ambient temperature 16.1 deg C

      FEE64 temperatures OK - attachment 34

      ADC data item stats OK - attachment 35

      per FEE64 Stat spectra - attachment 36

      Merger, TapeServer etc - attachments 37-39
Attachment 1: Screenshot_from_2024-04-25_03-26-58.png
Screenshot_from_2024-04-25_03-26-58.png
Attachment 2: Screenshot_from_2024-04-25_03-27-48.png
Screenshot_from_2024-04-25_03-27-48.png
Attachment 3: Screenshot_from_2024-04-25_03-28-19.png
Screenshot_from_2024-04-25_03-28-19.png
Attachment 4: Screenshot_from_2024-04-25_03-28-41.png
Screenshot_from_2024-04-25_03-28-41.png
Attachment 5: Screenshot_from_2024-04-25_03-32-43.png
Screenshot_from_2024-04-25_03-32-43.png
Attachment 6: Screenshot_from_2024-04-25_03-30-13.png
Screenshot_from_2024-04-25_03-30-13.png
Attachment 7: Screenshot_from_2024-04-25_03-30-37.png
Screenshot_from_2024-04-25_03-30-37.png
Attachment 8: Screenshot_from_2024-04-25_03-30-55.png
Screenshot_from_2024-04-25_03-30-55.png
Attachment 9: Screenshot_from_2024-04-25_10-19-09.png
Screenshot_from_2024-04-25_10-19-09.png
Attachment 10: Screenshot_from_2024-04-25_10-19-31.png
Screenshot_from_2024-04-25_10-19-31.png
Attachment 11: Screenshot_from_2024-04-25_10-19-54.png
Screenshot_from_2024-04-25_10-19-54.png
Attachment 12: Screenshot_from_2024-04-25_10-20-14.png
Screenshot_from_2024-04-25_10-20-14.png
Attachment 13: Screenshot_from_2024-04-25_10-20-26.png
Screenshot_from_2024-04-25_10-20-26.png
Attachment 14: Screenshot_from_2024-04-25_10-20-53.png
Screenshot_from_2024-04-25_10-20-53.png
Attachment 15: Screenshot_from_2024-04-25_10-21-22.png
Screenshot_from_2024-04-25_10-21-22.png
Attachment 16: Screenshot_from_2024-04-25_10-21-45.png
Screenshot_from_2024-04-25_10-21-45.png
Attachment 17: Screenshot_from_2024-04-25_14-44-05.png
Screenshot_from_2024-04-25_14-44-05.png
Attachment 18: Screenshot_from_2024-04-25_14-44-29.png
Screenshot_from_2024-04-25_14-44-29.png
Attachment 19: Screenshot_from_2024-04-25_14-44-49.png
Screenshot_from_2024-04-25_14-44-49.png
Attachment 20: Screenshot_from_2024-04-25_14-45-51.png
Screenshot_from_2024-04-25_14-45-51.png
Attachment 21: Screenshot_from_2024-04-25_14-46-35.png
Screenshot_from_2024-04-25_14-46-35.png
Attachment 22: Screenshot_from_2024-04-25_14-46-53.png
Screenshot_from_2024-04-25_14-46-53.png
Attachment 23: Screenshot_from_2024-04-25_14-47-05.png
Screenshot_from_2024-04-25_14-47-05.png
Attachment 24: Screenshot_from_2024-04-25_19-12-30.png
Screenshot_from_2024-04-25_19-12-30.png
Attachment 25: Screenshot_from_2024-04-25_19-13-01.png
Screenshot_from_2024-04-25_19-13-01.png
Attachment 26: Screenshot_from_2024-04-25_19-13-39.png
Screenshot_from_2024-04-25_19-13-39.png
Attachment 27: Screenshot_from_2024-04-25_19-14-02.png
Screenshot_from_2024-04-25_19-14-02.png
Attachment 28: Screenshot_from_2024-04-25_19-14-40.png
Screenshot_from_2024-04-25_19-14-40.png
Attachment 29: Screenshot_from_2024-04-25_19-15-24.png
Screenshot_from_2024-04-25_19-15-24.png
Attachment 30: Screenshot_from_2024-04-25_19-15-58.png
Screenshot_from_2024-04-25_19-15-58.png
Attachment 31: Screenshot_from_2024-04-25_19-16-12.png
Screenshot_from_2024-04-25_19-16-12.png
Attachment 32: Screenshot_from_2024-04-25_19-16-30.png
Screenshot_from_2024-04-25_19-16-30.png
Attachment 33: Screenshot_from_2024-04-25_23-14-01.png
Screenshot_from_2024-04-25_23-14-01.png
Attachment 34: Screenshot_from_2024-04-25_23-14-23.png
Screenshot_from_2024-04-25_23-14-23.png
Attachment 35: Screenshot_from_2024-04-25_23-14-41.png
Screenshot_from_2024-04-25_23-14-41.png
Attachment 36: Screenshot_from_2024-04-25_23-16-04.png
Screenshot_from_2024-04-25_23-16-04.png
Attachment 37: Screenshot_from_2024-04-25_23-16-12.png
Screenshot_from_2024-04-25_23-16-12.png
Attachment 38: Screenshot_from_2024-04-25_23-16-25.png
Screenshot_from_2024-04-25_23-16-25.png
Attachment 39: Screenshot_from_2024-04-25_23-16-42.png
Screenshot_from_2024-04-25_23-16-42.png
  545   Wed Apr 24 13:18:09 2024 TDExamples of poor CARME cable management/routing
Have observed some instances of poor CARME cable management/routing
Attachment 1 - SAES D-2000 ion pump HV cable (netal braid) transiting FEE64 cabling
Attachment 2 - CARME motor controller cable (grey plastic) transiting FEE64 cabling

General check and tidying up required.
Attachment 1: 20240424_125858.jpg
20240424_125858.jpg
Attachment 2: 20240424_125929.jpg
20240424_125929.jpg
  544   Wed Apr 24 12:48:35 2024 TDWednesday 24 April
      Replace AIDA ASIC mezzanine of aida01

13.43 DSSSD bias & leakage current OK - attachment 1
      ambient temperature 16.5 deg C

      FEE64 temperatures OK - attachment 2

      All system wide checks OK 

      WR timestamps OK - attachment 3

      ADC data item stats OK - attachment 4

      per FEE64 Rate spectra - attachment 5
       c. 10 strips (of 1024) not producing data
       most missing strips are from aida05 - misaligned or damaged adaptor PCB?

      per p+n FEE64 1.8.L spectra - attachment 6
       aida01 pulser peak width 75 ch FWHM
       common x/y scale => peak height proxy for peak width

      per n+n FEE64 1.8.L spectra - attachment 7
       aida01 pulser peak width 70 ch FWHM

      per FEE64 1.8.W spectra - 20us FSR - attachments 8-11

Summary

All FEE64s appear to be working (except aida05 which has c. 8 missing channels) with pulser - which 
validates the FEE64, ASICs and adaptor PCB.

Electronic noise c. 50keV FWHM for aida01 & aida03 - other FEE64s better/worse.

Preamplifier noise appears to be similar for all FEE64s which may suggest an issue with the test signal chain

Note
1) e-target is starting up
2) have identified some instances of poor CARME cable management/routing see https://elog.ph.ed.ac.uk/CARME/545



EG&G Ortec 448 Pulser
Amplitude 50,000
Normalise 5.00
x5 attenuator IN
t_r 100ns
tau_d 50us
Polarity +
Frequency relay line (50Hz)

Cooknell SA1 Summing Amplifier
inverted output to test - daisy chain

Current ASIC settings saved to 2024Apr24-14.14.03


14.48 Pulser OFF
 
      All histograms & stats zero'd

      LEC slow comparator 0x64 -> 0xff
      HEC fast comparator 0x2 -> 0x5

      background alpha run to validate signal path from DSSSD to FEE64 adaptor PCB

22.18 Start Merger & TapeServer - attachments 12-14

      DSSSD bias & leakage current OK - attachment 21
      ambient temperature 16.0 deg C


      FEE64 temperatures OK - attachment 15

      WR timestamps OK - attachment 16

      ADC data item stats OK - attachment 17

      per FEE64 Rate spectra - attachment 18

      per p+n FEE64 1.8.L spectra - attachments 19-20
Attachment 1: Screenshot_from_2024-04-24_13-43-29.png
Screenshot_from_2024-04-24_13-43-29.png
Attachment 2: Screenshot_from_2024-04-24_13-44-02.png
Screenshot_from_2024-04-24_13-44-02.png
Attachment 3: Screenshot_from_2024-04-24_13-45-01.png
Screenshot_from_2024-04-24_13-45-01.png
Attachment 4: Screenshot_from_2024-04-24_13-45-28.png
Screenshot_from_2024-04-24_13-45-28.png
Attachment 5: Screenshot_from_2024-04-24_13-45-46.png
Screenshot_from_2024-04-24_13-45-46.png
Attachment 6: Screenshot_from_2024-04-24_13-47-59.png
Screenshot_from_2024-04-24_13-47-59.png
Attachment 7: Screenshot_from_2024-04-24_13-52-50.png
Screenshot_from_2024-04-24_13-52-50.png
Attachment 8: Screenshot_from_2024-04-24_13-55-02.png
Screenshot_from_2024-04-24_13-55-02.png
Attachment 9: Screenshot_from_2024-04-24_13-55-32.png
Screenshot_from_2024-04-24_13-55-32.png
Attachment 10: Screenshot_from_2024-04-24_13-56-48.png
Screenshot_from_2024-04-24_13-56-48.png
Attachment 11: Screenshot_from_2024-04-24_13-55-59.png
Screenshot_from_2024-04-24_13-55-59.png
Attachment 12: Screenshot_from_2024-04-24_22-19-55.png
Screenshot_from_2024-04-24_22-19-55.png
Attachment 13: Screenshot_from_2024-04-24_22-20-16.png
Screenshot_from_2024-04-24_22-20-16.png
Attachment 14: Screenshot_from_2024-04-24_22-20-30.png
Screenshot_from_2024-04-24_22-20-30.png
Attachment 15: Screenshot_from_2024-04-24_22-21-58.png
Screenshot_from_2024-04-24_22-21-58.png
Attachment 16: Screenshot_from_2024-04-24_22-22-22.png
Screenshot_from_2024-04-24_22-22-22.png
Attachment 17: Screenshot_from_2024-04-24_22-22-57.png
Screenshot_from_2024-04-24_22-22-57.png
Attachment 18: Screenshot_from_2024-04-24_22-23-27.png
Screenshot_from_2024-04-24_22-23-27.png
Attachment 19: Screenshot_from_2024-04-24_22-24-37.png
Screenshot_from_2024-04-24_22-24-37.png
Attachment 20: Screenshot_from_2024-04-24_22-25-30.png
Screenshot_from_2024-04-24_22-25-30.png
Attachment 21: Screenshot_from_2024-04-24_22-27-58.png
Screenshot_from_2024-04-24_22-27-58.png
  543   Tue Apr 23 17:44:41 2024 TDTuesday 23 April
18.43 ASIC settings 2024Feb08-15.59.09

aida06 pulser peak width 66 ch FWHM
Attachment 1: Screenshot_from_2024-04-23_18-53-39.png
Screenshot_from_2024-04-23_18-53-39.png
Attachment 2: Screenshot_from_2024-04-23_18-56-31.png
Screenshot_from_2024-04-23_18-56-31.png
Attachment 3: Screenshot_from_2024-04-23_18-58-49.png
Screenshot_from_2024-04-23_18-58-49.png
Attachment 4: Screenshot_from_2024-04-23_18-59-18.png
Screenshot_from_2024-04-23_18-59-18.png
Attachment 5: Screenshot_from_2024-04-23_19-02-02.png
Screenshot_from_2024-04-23_19-02-02.png
Attachment 6: Screenshot_from_2024-04-23_19-05-12.png
Screenshot_from_2024-04-23_19-05-12.png
Attachment 7: Screenshot_from_2024-04-23_19-05-42.png
Screenshot_from_2024-04-23_19-05-42.png
Attachment 8: Screenshot_from_2024-04-23_19-06-15.png
Screenshot_from_2024-04-23_19-06-15.png
Attachment 9: Screenshot_from_2024-04-23_19-10-40.png
Screenshot_from_2024-04-23_19-10-40.png
Attachment 10: Screenshot_from_2024-04-23_19-14-23.png
Screenshot_from_2024-04-23_19-14-23.png
Attachment 11: Screenshot_from_2024-04-23_19-16-12.png
Screenshot_from_2024-04-23_19-16-12.png
Attachment 12: Screenshot_from_2024-04-23_19-16-46.png
Screenshot_from_2024-04-23_19-16-46.png
  542   Fri Apr 19 02:47:59 2024 TDFriday 19 April
03.45




10.01 




10.05 DSSSD bias OFF
      FEE64 power OFF
Attachment 1: Screenshot_from_2024-04-19_03-45-09.png
Screenshot_from_2024-04-19_03-45-09.png
Attachment 2: Screenshot_from_2024-04-19_03-46-43.png
Screenshot_from_2024-04-19_03-46-43.png
Attachment 3: Screenshot_from_2024-04-19_03-47-13.png
Screenshot_from_2024-04-19_03-47-13.png
Attachment 4: Screenshot_from_2024-04-19_03-48-48.png
Screenshot_from_2024-04-19_03-48-48.png
Attachment 5: Screenshot_from_2024-04-19_10-53-30.png
Screenshot_from_2024-04-19_10-53-30.png
Attachment 6: Screenshot_from_2024-04-19_10-53-54.png
Screenshot_from_2024-04-19_10-53-54.png
Attachment 7: Screenshot_from_2024-04-19_10-54-37.png
Screenshot_from_2024-04-19_10-54-37.png
Attachment 8: Screenshot_from_2024-04-19_10-57-32.png
Screenshot_from_2024-04-19_10-57-32.png
Attachment 9: Screenshot_from_2024-04-19_10-58-15.png
Screenshot_from_2024-04-19_10-58-15.png
Attachment 10: Screenshot_from_2024-04-19_10-59-15.png
Screenshot_from_2024-04-19_10-59-15.png
Attachment 11: Screenshot_from_2024-04-19_11-00-45.png
Screenshot_from_2024-04-19_11-00-45.png
Attachment 12: Screenshot_from_2024-04-19_11-04-21.png
Screenshot_from_2024-04-19_11-04-21.png
  541   Thu Apr 18 13:34:06 2024 TD, JGThursday 18 April
14.30 


       per FEE64 1.8.L spectra
        aida11 pulser peak width 52 ch FWHM



14.45 EG&G Ortec 448 Pulser OFF

      all stats & histograms zero'd
 
      alpha background


22.00 DSSSD bias and leakage current OK - attachment 10

      FEE64 temperatures OK - attachment 11

      ADC data item stats - attachment 12

      per FEE64 Stat spectra - attachment 13
       Comparison of attachment 9 (per FEE64 Rate spectra with pulser) indicates significant issues (downstream of adaptor PCB) for DSSSD #1
       Possible issues asics #3 aida01 and possible adaptor PCB misalignment aida05
Attachment 1: Screenshot_from_2024-04-18_14-22-41.png
Screenshot_from_2024-04-18_14-22-41.png
Attachment 2: Screenshot_from_2024-04-18_14-23-10.png
Screenshot_from_2024-04-18_14-23-10.png
Attachment 3: Screenshot_from_2024-04-18_14-23-51.png
Screenshot_from_2024-04-18_14-23-51.png
Attachment 4: Screenshot_from_2024-04-18_14-24-18.png
Screenshot_from_2024-04-18_14-24-18.png
Attachment 5: Screenshot_from_2024-04-18_14-27-33.png
Screenshot_from_2024-04-18_14-27-33.png
Attachment 6: Screenshot_from_2024-04-18_14-28-26.png
Screenshot_from_2024-04-18_14-28-26.png
Attachment 7: Screenshot_from_2024-04-18_14-31-14.png
Screenshot_from_2024-04-18_14-31-14.png
Attachment 8: Screenshot_from_2024-04-18_14-32-31.png
Screenshot_from_2024-04-18_14-32-31.png
Attachment 9: Screenshot_from_2024-04-18_14-43-27.png
Screenshot_from_2024-04-18_14-43-27.png
Attachment 10: Screenshot_from_2024-04-18_21-57-51.png
Screenshot_from_2024-04-18_21-57-51.png
Attachment 11: Screenshot_from_2024-04-18_21-58-31.png
Screenshot_from_2024-04-18_21-58-31.png
Attachment 12: Screenshot_from_2024-04-18_21-59-11.png
Screenshot_from_2024-04-18_21-59-11.png
Attachment 13: Screenshot_from_2024-04-18_21-59-35.png
Screenshot_from_2024-04-18_21-59-35.png
  540   Wed Mar 6 10:47:11 2024 JG, TDWednesday 6 March - PEXARIA5 WR test
Courtesy Nik Kurz, JG has installed a PEXARIA5 PCIe card adjacent to CARME - BuTIS single-mode fibre input
and HDMI output to root MACB input for distribution to the other MACBs and FEE64s.

The PEXARIA5 card provides equivalent functionality to the VETAR2 VME module that we have used to date. 

Further information https://www.gsi.de/fileadmin/EE/Module/EXPLODER/pexaria5_14.pdf 

Cooling water ON 

Via Anydesk startup MIDAS and FEE64s

Power ON
DSSSD HV OFF ( not required for WR test )

FEE64 temperatures OK - attachment 1

All system wide checks OK

Info code 4 & 5 stats OK - attachments 2 & 3

10:48:46 WR timestamps OK - attachment 4

         Timestamp (ns) converted to (s) corresponds to 09.49.22 GMT 6.3.24 which appears to be correct

11.00 Power OFF
Attachment 1: temps.PNG
temps.PNG
Attachment 2: stats_1.PNG
stats_1.PNG
Attachment 3: stats_2.PNG
stats_2.PNG
Attachment 4: wr_0949.PNG
wr_0949.PNG
  539   Wed Feb 21 07:38:32 2024 TDWedneday 21 February
10.54 23.2.24 Bias OFF
Attachment 1: Screenshot_from_2024-02-20_18-40-08.png
Screenshot_from_2024-02-20_18-40-08.png
Attachment 2: Screenshot_from_2024-02-20_21-30-08.png
Screenshot_from_2024-02-20_21-30-08.png
Attachment 3: Screenshot_from_2024-02-21_01-00-13.png
Screenshot_from_2024-02-21_01-00-13.png
Attachment 4: Screenshot_from_2024-02-21_05-10-07.png
Screenshot_from_2024-02-21_05-10-07.png
Attachment 5: Screenshot_from_2024-02-21_08-30-06.png
Screenshot_from_2024-02-21_08-30-06.png
Attachment 6: Screenshot_from_2024-02-21_20-45-58.png
Screenshot_from_2024-02-21_20-45-58.png
Attachment 7: Screenshot_from_2024-02-21_20-46-12.png
Screenshot_from_2024-02-21_20-46-12.png
Attachment 8: Screenshot_from_2024-02-22_00-51-21.png
Screenshot_from_2024-02-22_00-51-21.png
Attachment 9: Screenshot_from_2024-02-22_14-02-29.png
Screenshot_from_2024-02-22_14-02-29.png
Attachment 10: Screenshot_from_2024-02-22_14-02-44.png
Screenshot_from_2024-02-22_14-02-44.png
Attachment 11: Screenshot_from_2024-02-22_19-00-08.png
Screenshot_from_2024-02-22_19-00-08.png
Attachment 12: Screenshot_from_2024-02-22_19-00-31.png
Screenshot_from_2024-02-22_19-00-31.png
Attachment 13: Screenshot_from_2024-02-23_11-53-32.png
Screenshot_from_2024-02-23_11-53-32.png
Attachment 14: Screenshot_from_2024-02-23_11-53-45.png
Screenshot_from_2024-02-23_11-53-45.png
Attachment 15: Capture_10.55.PNG
Capture_10.55.PNG
Attachment 16: Capture_18.00.PNG
Capture_18.00.PNG
Attachment 17: Capture_13.04.PNG
Capture_13.04.PNG
Attachment 18: Capture_23.52.PNG
Capture_23.52.PNG
Attachment 19: Capture_19.46.PNG
Capture_19.46.PNG
Attachment 20: Capture_07.30.PNG
Capture_07.30.PNG
Attachment 21: Capture_04.10.PNG
Capture_04.10.PNG
Attachment 22: Capture_00.00.PNG
Capture_00.00.PNG
Attachment 23: Capture_20.30.PNG
Capture_20.30.PNG
Attachment 24: Capture_17.40.PNG
Capture_17.40.PNG
  538   Tue Feb 20 14:12:32 2024 JMShutdown

CARME DAQ shutdown. FEEs turned OFF and all MIDAS server windows closed.

Detector bias turned OFF.

Final leakage current plot - attachment 1

Attachment 1: Screenshot_from_2024-02-20_15-11-40.png
Screenshot_from_2024-02-20_15-11-40.png
  537   Tue Feb 20 13:20:44 2024 OF, AG, JG, MLR48 status check

Doing afternoon check. Checked the leakage currents, all ok. Attachment 1

Temperatures ok, attchment 2

System wide check, all correct.

Statistics ok (as far as we can tell) attachment 3

Spectrum browser ok, attachment 4

Checked merger, spinner spinning...

 

Beam current is around 2.2 E7 and the target dencity is 6.3E11 cm2.

Attachment 1: Screenshot_from_2024-02-20_14-22-05.png
Screenshot_from_2024-02-20_14-22-05.png
Attachment 2: Screenshot_from_2024-02-20_14-22-49.png
Screenshot_from_2024-02-20_14-22-49.png
Attachment 3: Screenshot_from_2024-02-20_14-29-39.png
Screenshot_from_2024-02-20_14-29-39.png
Attachment 4: Screenshot_from_2024-02-20_14-34-55.png
Screenshot_from_2024-02-20_14-34-55.png
  536   Tue Feb 20 09:57:57 2024 OF, AG, MLRun 48 status check

Doing morning check. Checked the leakage currents, all ok. Attachment 1

Temperatures ok, attchment 2

System wide check, all correct.

Statistics ok (as far as we can tell) attachment 3

Spectrum browser ok, attachment 4

Checked merger, spinner spinning...

Attachment 5 shows the BPM scope (i.e. induced signals from the ion buchnes) for the duration of injection, bunching, acceleration, debunching.

 

Beam current is around 2.5E7 and the target dencity is 6.3E11 cm2.

Attachment 1: Screenshot_from_2024-02-20_11-06-48.png
Screenshot_from_2024-02-20_11-06-48.png
Attachment 2: Screenshot_from_2024-02-20_11-08-05.png
Screenshot_from_2024-02-20_11-08-05.png
Attachment 3: Screenshot_from_2024-02-20_11-12-26.png
Screenshot_from_2024-02-20_11-12-26.png
Attachment 4: Screenshot_from_2024-02-20_11-13-58.png
Screenshot_from_2024-02-20_11-13-58.png
Attachment 5: 2024-02-20_11-37-46-304.png
2024-02-20_11-37-46-304.png
  535   Tue Feb 20 09:51:37 2024 JMRun 48 beam on target

Run 48 started

  534   Tue Feb 20 08:41:08 2024 OF, AGstill no beam CRYRING

We have mosty empty injections into CRYRING. Only sporadically the beam with good intensity is injected.

Michael Lestinsky is currently working on the problem.

  533   Tue Feb 20 07:50:10 2024 JMDAQ restart

DAQ restarted. Run46. All appears OK, aida01 shows high rates across ASIC 4 even with no beam which could not be resolved via ASIC control. May suggest some damge.

Merger is running and tape server is writing data to disk. Data rate is about 1300 kB/sec from 800 kb/sec due to increase in aida01 rate.

FEE temps OK

All system checks OK

Detectors moved to IN position (left arm pot = 36.5 mm, right arm =70.49 mm). Target ON.

Lost beam again.

Run 47 started.

 

  532   Mon Feb 19 21:46:32 2024 EH, EM, CBRun 45 + end of late shift
22:54
We have moved the gas jet target out (y=4000), the density has dropped to zero.
Started a new run (Run45) for background measurement.
Informed HKR that the experiment will be monitored remotely without on-site personnel. They will check on the beam intensity.

02:08
System wide checks OK
Temperatures OK
Leakage current going up - OK
Stats OK
Merger continues OK

Reset the online spectra (had not been done yet for this run).
20s rate looks fine, around 1000 counts/s on top left detector. Maybe we are hitting it with beam? But then we should be see a reduction in beam intensity, we do not.

04:00
System wide checks OK
Temps OK
Stats OK
Merger OK

06:20
Beam intensity quite unstable, but HKR is correcting - attach 7
Systems wide checks OK
Temps OK
Merger OK
Stats OK
Some interesting features can be seen in the online histograms, e.g. no peak at 16 MeV and a clear straight line in the Ex vs x plot.

07:20
DAQ crashed due to multiple merger / tape server / HTTPD server instances being opened on workspace 5
Cannot soft reboot. FEEs off.
Attachment 1: Screenshot_from_2024-02-20_02-02-14.png
Screenshot_from_2024-02-20_02-02-14.png
Attachment 2: Screenshot_from_2024-02-20_02-02-47.png
Screenshot_from_2024-02-20_02-02-47.png
Attachment 3: Screenshot_from_2024-02-20_02-03-07.png
Screenshot_from_2024-02-20_02-03-07.png
Attachment 4: Screenshot_from_2024-02-20_02-03-31.png
Screenshot_from_2024-02-20_02-03-31.png
Attachment 5: Screenshot_from_2024-02-20_02-04-40.png
Screenshot_from_2024-02-20_02-04-40.png
Attachment 6: Screenshot_from_2024-02-20_02-06-37.png
Screenshot_from_2024-02-20_02-06-37.png
Attachment 7: Beam_intensity.png
Beam_intensity.png
Attachment 8: Screenshot_from_2024-02-20_06-15-54.png
Screenshot_from_2024-02-20_06-15-54.png
Attachment 9: Screenshot_from_2024-02-20_06-16-09.png
Screenshot_from_2024-02-20_06-16-09.png
Attachment 10: Screenshot_from_2024-02-20_06-16-39.png
Screenshot_from_2024-02-20_06-16-39.png
Attachment 11: Screenshot_from_2024-02-20_06-17-11.png
Screenshot_from_2024-02-20_06-17-11.png
Attachment 12: Screenshot_from_2024-02-20_06-17-35.png
Screenshot_from_2024-02-20_06-17-35.png
Attachment 13: Screenshot_from_2024-02-20_06-18-29.png
Screenshot_from_2024-02-20_06-18-29.png
Attachment 14: Screenshot_from_2024-02-20_06-18-58.png
Screenshot_from_2024-02-20_06-18-58.png
  531   Mon Feb 19 08:45:22 2024 JM, JG, SHB, EM, EHRun 44 - cooled beam on target

09:43

Run 44 started cooled beam on target

Detectors IN (left arm pot =36.5 mm, right arm pot = 70.49 mm). Stationary. Same as previous runs

Electron cooler voltage 751.4 V this corresponds to 1.3436 MeV/u. This should be back on the peak of the resonance.

Target density is 5.2e11 at/cm2, ring pressure looks OK YR09 1.72e-10 mbar. Target density is not very stable and requires small corrections in the x position to maintain a target

Horizontal beam position is -7.5 mm, vertical beam position is 0.0 mm. Same as previous run

Beam is in the ring for 6 seconds and total cycle time between injections is 10 seconds. Notice this is different than previous runs. Aim is to have a few different cycle time configurations to see effect on halo

Statistics, merger statistics, histograms all zero'd

10:05

FEE temps OK

System checks OK

Merger OK, tape server OK

Leakage current plot attachment 1

ADC data items stats attachment 2

Monitoring code rates attachment 3

11:08

FEE temps OK

System checks OK

Merger OK, tape server OK

Leakage current OK

ADC data items stats OK

Online spectra OK

12:30

FEE temps OK

System checks OK

Merger OK, tape server OK

Leakage current OK

ADC data items stats OK

Online spectra OK

14:03

FEE temps OK

System checks OK

Merger OK, tape server OK

Leakage current OK

ADC data items stats OK

Online spectra OK

15:36

FEE temps OK, system checks OK, merger OK,

Screenshot 4: temperatures

Screenshot 5: statistics

Screenhot 6 - 9: spectra (there is a persistent peak in aida13 in Layout8)

Screenshot 10: leakage current (current 1 seems to be going up slowly but steadily)

Screenshot 11: spectrum

 

17:18

FEE temps OK, system checks OK, merger OK,

Screenshot 12: temperatures

Screenshot 13: statistics

Screenhot 14 - 17: spectra (we still see a peak in aida13 in Layout8, but not always)

Screenshot 18: leakage current (current 1 seems to be going up slowly but steadily)

Screenshot 19: spectrum

 

19:27

FEE temps OK, system checks OK, merger OK,

Screenshot 20: temperatures

Screenshot 21: statistics

Screenhot 22 - 25: spectra (we still see a peak in aida13 in Layout8, but not always)

Screenshot 26: leakage current (current 1 seems to be going up slowly but steadily)

Screenshot 27: spectrum

 

21:57

FEE temps OK, system checks OK, merger OK,

Screenshot 28: temperatures

Screenshot 29: statistics

Screenhot 30 - 33: spectra (we still see a peak in aida13 in Layout8, but not always)

Screenshot 34: leakage current

Screenshot 35: spectrum

 

 

Attachment 1: Screenshot_from_2024-02-19_10-07-26.png
Screenshot_from_2024-02-19_10-07-26.png
Attachment 2: Screenshot_from_2024-02-19_10-10-24.png
Screenshot_from_2024-02-19_10-10-24.png
Attachment 3: Screenshot_from_2024-02-19_10-10-50.png
Screenshot_from_2024-02-19_10-10-50.png
Attachment 4: Screenshot_from_2024-02-19_15-31-30.png
Screenshot_from_2024-02-19_15-31-30.png
Attachment 5: Screenshot_from_2024-02-19_15-31-47.png
Screenshot_from_2024-02-19_15-31-47.png
Attachment 6: Screenshot_from_2024-02-19_15-32-57.png
Screenshot_from_2024-02-19_15-32-57.png
Attachment 7: Screenshot_from_2024-02-19_15-33-22.png
Screenshot_from_2024-02-19_15-33-22.png
Attachment 8: Screenshot_from_2024-02-19_15-33-46.png
Screenshot_from_2024-02-19_15-33-46.png
Attachment 9: Screenshot_from_2024-02-19_15-34-51.png
Screenshot_from_2024-02-19_15-34-51.png
Attachment 10: Screenshot_from_2024-02-19_15-35-23.png
Screenshot_from_2024-02-19_15-35-23.png
Attachment 11: Screenshot_from_2024-02-19_15-35-29.png
Screenshot_from_2024-02-19_15-35-29.png
Attachment 12: Screenshot_from_2024-02-19_17-13-45.png
Screenshot_from_2024-02-19_17-13-45.png
Attachment 13: Screenshot_from_2024-02-19_17-14-05.png
Screenshot_from_2024-02-19_17-14-05.png
Attachment 14: Screenshot_from_2024-02-19_17-15-12.png
Screenshot_from_2024-02-19_17-15-12.png
Attachment 15: Screenshot_from_2024-02-19_17-15-52.png
Screenshot_from_2024-02-19_17-15-52.png
Attachment 16: Screenshot_from_2024-02-19_17-16-24.png
Screenshot_from_2024-02-19_17-16-24.png
Attachment 17: Screenshot_from_2024-02-19_17-17-25.png
Screenshot_from_2024-02-19_17-17-25.png
Attachment 18: Screenshot_from_2024-02-19_17-17-44.png
Screenshot_from_2024-02-19_17-17-44.png
Attachment 19: Screenshot_from_2024-02-19_17-17-51.png
Screenshot_from_2024-02-19_17-17-51.png
Attachment 20: Screenshot_from_2024-02-19_19-24-13.png
Screenshot_from_2024-02-19_19-24-13.png
Attachment 21: Screenshot_from_2024-02-19_19-26-55.png
Screenshot_from_2024-02-19_19-26-55.png
Attachment 22: Screenshot_from_2024-02-19_19-24-59.png
Screenshot_from_2024-02-19_19-24-59.png
Attachment 23: Screenshot_from_2024-02-19_19-25-21.png
Screenshot_from_2024-02-19_19-25-21.png
Attachment 24: Screenshot_from_2024-02-19_19-25-51.png
Screenshot_from_2024-02-19_19-25-51.png
Attachment 25: Screenshot_from_2024-02-19_19-26-41.png
Screenshot_from_2024-02-19_19-26-41.png
Attachment 26: Screenshot_from_2024-02-19_19-27-22.png
Screenshot_from_2024-02-19_19-27-22.png
Attachment 27: Screenshot_from_2024-02-19_19-27-29.png
Screenshot_from_2024-02-19_19-27-29.png
Attachment 28: Screenshot_from_2024-02-19_21-54-34.png
Screenshot_from_2024-02-19_21-54-34.png
Attachment 29: Screenshot_from_2024-02-19_21-54-46.png
Screenshot_from_2024-02-19_21-54-46.png
Attachment 30: Screenshot_from_2024-02-19_21-55-17.png
Screenshot_from_2024-02-19_21-55-17.png
Attachment 31: Screenshot_from_2024-02-19_21-55-41.png
Screenshot_from_2024-02-19_21-55-41.png
Attachment 32: Screenshot_from_2024-02-19_21-56-02.png
Screenshot_from_2024-02-19_21-56-02.png
Attachment 33: Screenshot_from_2024-02-19_21-56-50.png
Screenshot_from_2024-02-19_21-56-50.png
Attachment 34: Screenshot_from_2024-02-19_21-57-23.png
Screenshot_from_2024-02-19_21-57-23.png
Attachment 35: Screenshot_from_2024-02-19_21-57-30.png
Screenshot_from_2024-02-19_21-57-30.png
  530   Mon Feb 19 06:01:17 2024 SHB, JM, JGRun 42 morning shift

07:00

Observed reduction in YR09 presure over last few hours in ML online plot. Beam intensity looks OK but target density was fluctuating quite a lot. Current value was 1.5e11.

Changed the nozzle position slightly which increased target density to 5.5e11. Pressure in YR09 looks OK.

I accidently scrolled slightly with the target code (only by 100 but am not sure). Target looks stable but I will not touch remotely again for fear of losing total target position.

Rates in ADC data items statistics and in the monitoring code look consistent with beam on target. 

08:56

Target density reduced again to 1.5e11. Changed the nozzle position slightly in X and recovered 5.5e11 at/cm2

Beam intensity looks OK

FEE temps OK

System checks OK

Merger and tape server OK

Statistics ADC data items and spectrum browser rates OK

Merger stats attachement 5

leakage current plot attachment 6

Online monitoring code rates OK

09:33

Run stopped to change beam energy

 

Attachment 1: Screenshot_from_2024-02-19_08-11-42.png
Screenshot_from_2024-02-19_08-11-42.png
Attachment 2: Screenshot_from_2024-02-19_08-12-02.png
Screenshot_from_2024-02-19_08-12-02.png
Attachment 3: Screenshot_from_2024-02-19_08-12-27.png
Screenshot_from_2024-02-19_08-12-27.png
Attachment 4: Screenshot_from_2024-02-19_08-12-47.png
Screenshot_from_2024-02-19_08-12-47.png
Attachment 5: Screenshot_from_2024-02-19_08-57-23.png
Screenshot_from_2024-02-19_08-57-23.png
Attachment 6: Screenshot_from_2024-02-19_08-59-08.png
Screenshot_from_2024-02-19_08-59-08.png
  529   Mon Feb 19 05:10:48 2024 TDRun 42 night shift contd.
06.03 DAQ continues file G22-86/R42_32

      DSSSD bias & leakage current - attachments 1-2
       DSSSD #0 trending upwards , DSSSD #1-3 trending downwards with decreasing ambient temperature

      FEE64 temperatures OK - attachments 3-4

      ADC data item stats - attachment 5

      per FEE64 Rate spectra - attachments 6

      target density 5.27e+11/cm2
      beam stable for last 3 hours ( cannot read scale with X2Go )
Attachment 1: Screenshot_from_2024-02-19_06-03-26.png
Screenshot_from_2024-02-19_06-03-26.png
Attachment 2: Screenshot_from_2024-02-19_06-03-41.png
Screenshot_from_2024-02-19_06-03-41.png
Attachment 3: Screenshot_from_2024-02-19_06-04-07.png
Screenshot_from_2024-02-19_06-04-07.png
Attachment 4: Screenshot_from_2024-02-19_06-05-27.png
Screenshot_from_2024-02-19_06-05-27.png
Attachment 5: Screenshot_from_2024-02-19_06-05-51.png
Screenshot_from_2024-02-19_06-05-51.png
Attachment 6: Screenshot_from_2024-02-19_06-06-15.png
Screenshot_from_2024-02-19_06-06-15.png
  528   Sun Feb 18 22:03:56 2024 TDRun 42 night shift
23.03 DAQ continues file G22-86/R42_21

      ML CRYRING ambient temperature and vacuum - attachments 1 & 2

      DSSSD bias & leakage current - attachments 3 & 24
       DSSSD #0 trending upwards , DSSSD #1-3 trending downwards with decreasing ambient temperature

      FEE64 temperatures OK - attachments 4-5

23.07 All system wide checks OK

      WR timestamps OK - attachment 6

      ADC, DISC, PAUSE and correlation stats OK - attachments 7-10

      per FEE64 Rate spectra - attachments 11-12
      per FEE64 Stat spectra - attachments 13-14

      per FEE64 1.8.L, 1.8.H and 1.8.W 20us FSR spectra - attachments 15-20

      Merger, Merger stats and TapeServer OK - attachments 21-23

      online analysis spectra - attachments 25-30

23.28 all histograms, stats, Merger stats zero'd
      online analysis restarted

23.35 target density 5.83e+11/cm2
      beam no change in last 3 hours ( cannot read scale with X2Go )
      see attachments 31-33


00.58 DAQ continues file G22-86/R42_24

      DSSSD bias & leakage current - attachments 34-35
       DSSSD #0 trending upwards , DSSSD #1-3 trending downwards with decreasing ambient temperature

      FEE64 temperatures OK - attachments 36-37

      ADC data item stats - attachment 38

      per FEE64 Rate spectra - attachments 39

      target density 5.75e+11/cm2
      beam no change in last 3 hours ( cannot read scale with X2Go )


03.14 DAQ continues file G22-86/R42_28

      DSSSD bias & leakage current - attachments 40-41
       DSSSD #0 trending upwards , DSSSD #1-3 trending downwards with decreasing ambient temperature

      FEE64 temperatures OK - attachments 42-43

      ADC data item stats - attachment 44

      per FEE64 Rate spectra - attachments 45

      target density 5.91e+11/cm2
      beam decreased in last 3 hours ( cannot read scale with X2Go )
      see attachments 46-47

03:30 beam recovered
      online analysis rates DSSSD#0 recovered to values observed at start of shift
Attachment 1: Capture.PNG
Capture.PNG
Attachment 2: Capture_2.PNG
Capture_2.PNG
Attachment 3: Screenshot_from_2024-02-18_23-05-47.png
Screenshot_from_2024-02-18_23-05-47.png
Attachment 4: Screenshot_from_2024-02-18_23-06-14.png
Screenshot_from_2024-02-18_23-06-14.png
Attachment 5: Screenshot_from_2024-02-18_23-06-56.png
Screenshot_from_2024-02-18_23-06-56.png
Attachment 6: Screenshot_from_2024-02-18_23-07-42.png
Screenshot_from_2024-02-18_23-07-42.png
Attachment 7: Screenshot_from_2024-02-18_23-08-07.png
Screenshot_from_2024-02-18_23-08-07.png
Attachment 8: Screenshot_from_2024-02-18_23-08-29.png
Screenshot_from_2024-02-18_23-08-29.png
Attachment 9: Screenshot_from_2024-02-18_23-08-53.png
Screenshot_from_2024-02-18_23-08-53.png
Attachment 10: Screenshot_from_2024-02-18_23-09-16.png
Screenshot_from_2024-02-18_23-09-16.png
Attachment 11: Screenshot_from_2024-02-18_23-10-18.png
Screenshot_from_2024-02-18_23-10-18.png
Attachment 12: Screenshot_from_2024-02-18_23-10-41.png
Screenshot_from_2024-02-18_23-10-41.png
Attachment 13: Screenshot_from_2024-02-18_23-11-31.png
Screenshot_from_2024-02-18_23-11-31.png
Attachment 14: Screenshot_from_2024-02-18_23-11-48.png
Screenshot_from_2024-02-18_23-11-48.png
Attachment 15: Screenshot_from_2024-02-18_23-12-38.png
Screenshot_from_2024-02-18_23-12-38.png
Attachment 16: Screenshot_from_2024-02-18_23-13-20.png
Screenshot_from_2024-02-18_23-13-20.png
Attachment 17: Screenshot_from_2024-02-18_23-14-03.png
Screenshot_from_2024-02-18_23-14-03.png
Attachment 18: Screenshot_from_2024-02-18_23-14-45.png
Screenshot_from_2024-02-18_23-14-45.png
Attachment 19: Screenshot_from_2024-02-18_23-15-21.png
Screenshot_from_2024-02-18_23-15-21.png
Attachment 20: Screenshot_from_2024-02-18_23-15-58.png
Screenshot_from_2024-02-18_23-15-58.png
Attachment 21: Screenshot_from_2024-02-18_23-16-33.png
Screenshot_from_2024-02-18_23-16-33.png
Attachment 22: Screenshot_from_2024-02-18_23-16-49.png
Screenshot_from_2024-02-18_23-16-49.png
Attachment 23: Screenshot_from_2024-02-18_23-17-01.png
Screenshot_from_2024-02-18_23-17-01.png
Attachment 24: Screenshot_from_2024-02-18_23-18-10.png
Screenshot_from_2024-02-18_23-18-10.png
Attachment 25: Screenshot_from_2024-02-18_23-18-26.png
Screenshot_from_2024-02-18_23-18-26.png
Attachment 26: Screenshot_from_2024-02-18_23-18-38.png
Screenshot_from_2024-02-18_23-18-38.png
Attachment 27: Screenshot_from_2024-02-18_23-18-48.png
Screenshot_from_2024-02-18_23-18-48.png
Attachment 28: Screenshot_from_2024-02-18_23-18-59.png
Screenshot_from_2024-02-18_23-18-59.png
Attachment 29: Screenshot_from_2024-02-18_23-19-18.png
Screenshot_from_2024-02-18_23-19-18.png
Attachment 30: Screenshot_from_2024-02-18_23-19-39.png
Screenshot_from_2024-02-18_23-19-39.png
Attachment 31: 20240218_223433.jpg
20240218_223433.jpg
Attachment 32: 20240218_223341.jpg
20240218_223341.jpg
Attachment 33: 20240218_223509.jpg
20240218_223509.jpg
Attachment 34: Screenshot_from_2024-02-19_00-54-56.png
Screenshot_from_2024-02-19_00-54-56.png
Attachment 35: Screenshot_from_2024-02-19_00-55-14.png
Screenshot_from_2024-02-19_00-55-14.png
Attachment 36: Screenshot_from_2024-02-19_00-55-44.png
Screenshot_from_2024-02-19_00-55-44.png
Attachment 37: Screenshot_from_2024-02-19_00-56-14.png
Screenshot_from_2024-02-19_00-56-14.png
Attachment 38: Screenshot_from_2024-02-19_00-56-42.png
Screenshot_from_2024-02-19_00-56-42.png
Attachment 39: Screenshot_from_2024-02-19_00-57-31.png
Screenshot_from_2024-02-19_00-57-31.png
Attachment 40: Screenshot_from_2024-02-19_03-14-25.png
Screenshot_from_2024-02-19_03-14-25.png
Attachment 41: Screenshot_from_2024-02-19_03-14-48.png
Screenshot_from_2024-02-19_03-14-48.png
Attachment 42: Screenshot_from_2024-02-19_03-15-18.png
Screenshot_from_2024-02-19_03-15-18.png
Attachment 43: Screenshot_from_2024-02-19_03-15-43.png
Screenshot_from_2024-02-19_03-15-43.png
Attachment 44: Screenshot_from_2024-02-19_03-16-04.png
Screenshot_from_2024-02-19_03-16-04.png
Attachment 45: Screenshot_from_2024-02-19_03-16-50.png
Screenshot_from_2024-02-19_03-16-50.png
Attachment 46: 20240219_022309.jpg
20240219_022309.jpg
Attachment 47: 20240219_022316.jpg
20240219_022316.jpg
  527   Sun Feb 18 13:58:33 2024 AR, OF, JMRun 42 - cooled beam on target

Run 42 - 15:00

Detectors IN (left arm pot =36.54 mm, right arm pot = 70.49 mm). Stationary. Same as run 40

Electron cooler voltage 751.89 V this corresponds to 1.3444 MeV/u. This should be on the right of the peak of the resonance.

Target density is 5.03e11 at/cm2, ring pressure looks OK YR09 2.36e-10 mbar.

Horizontal beam position is -7.5 mm, vertical beam position is 0.0 mm. Same as run 40.

Beam is in the ring for 4 seconds and total cycle time between injections is 8 seconds. Notice this is the samet as run 40, different to run 35.

Statistics, merger statistics, histograms all zero'd

System wide checks ok. Temperatures ok. Leakage current ok. Merger ok.

Attatchment 1: Stats

Attatchment 2: Temperatures

Attatchments 3-6: Waveforms

Attatchment 7:Leakage current

Attatchment 8: Monitoring code rates

 

 

15.35 DSSSD #0 leakage current trending somewhat upwards cf. somewhat downwards for DSSSD #1-3

          ML CRYRING hall ambient temprature c. 17.7. deg C and on downward trend.

         Probably indicates radiation damage/surface charging of DSSSD #0 - leave DSSSD #0 biased for 24-48h following run and monitor leakage current

15:46

Seeing some beam intensity drops occasionally. Small drop that instantamnously went back up around 15:20 and then stbale beam unitl around 15:44 and then two small-time beam drops within 2 minutes. See attatchment 9. Continuing to monitor. Target density remains stable at 5e11.

16:45

Beam dropping again and continuing to drop  (began at 16:35). Target density 4.7e11. Called HKR and they will try to increase the intensity. See attatchment 10.

17:00

System checks ok. Temperatures ok. Leakage current plot okay. Merger ok. Beam intensity still reduced to same amount as in attatchment 10). Target density approx 5e11.

Attatchment 1:1 Stats

Attatchment 12: Temperatures

Attatchments 13-16: Waveforms

Attatchment 17:Leakage current

Attatchment 18: Monitoring code rates

17:10

HKR cryring asked for target to be removed to check intensity. Target was removed.

17:17

Target back in, intensity back to around 2.5e7. Stable. Target desity 5e11.

18:42

All system checks and tempeatures ok. Beam stable. Target 2.3e11.

Leakage current: red line still trending upwards but slowly. See image 24.

10:30

Beam went down about 21:45. Called HKR and they were able to almost immediately restore the beam. Beam intensity ~2.5e7.

Target density 5.6e11 at/cm2, pressure in YR09 ~2.5e-10 mbar.

FEE temps and system check OK

Merger and tape server OK

Merger stats (attachment 25)

All rates in spectrum browser look OK

Monitoring code rates indicate beam on target.

 

 

 

 

 

Attachment 1: Screenshot_from_2024-02-18_15-06-55.png
Screenshot_from_2024-02-18_15-06-55.png
Attachment 2: Screenshot_from_2024-02-18_15-07-13.png
Screenshot_from_2024-02-18_15-07-13.png
Attachment 3: Screenshot_from_2024-02-18_15-07-56.png
Screenshot_from_2024-02-18_15-07-56.png
Attachment 4: Screenshot_from_2024-02-18_15-08-32.png
Screenshot_from_2024-02-18_15-08-32.png
Attachment 5: Screenshot_from_2024-02-18_15-09-10.png
Screenshot_from_2024-02-18_15-09-10.png
Attachment 6: Screenshot_from_2024-02-18_15-10-02.png
Screenshot_from_2024-02-18_15-10-02.png
Attachment 7: Screenshot_from_2024-02-18_15-10-44.png
Screenshot_from_2024-02-18_15-10-44.png
Attachment 8: Screenshot_from_2024-02-18_15-11-03.png
Screenshot_from_2024-02-18_15-11-03.png
Attachment 9: IMG_7922.jpg
IMG_7922.jpg
Attachment 10: IMG_7932.jpg
IMG_7932.jpg
Attachment 11: Screenshot_from_2024-02-18_16-58-04.png
Screenshot_from_2024-02-18_16-58-04.png
Attachment 12: Screenshot_from_2024-02-18_16-58-17.png
Screenshot_from_2024-02-18_16-58-17.png
Attachment 13: Screenshot_from_2024-02-18_16-59-30.png
Screenshot_from_2024-02-18_16-59-30.png
Attachment 14: Screenshot_from_2024-02-18_17-00-01.png
Screenshot_from_2024-02-18_17-00-01.png
Attachment 15: Screenshot_from_2024-02-18_17-00-32.png
Screenshot_from_2024-02-18_17-00-32.png
Attachment 16: Screenshot_from_2024-02-18_17-00-53.png
Screenshot_from_2024-02-18_17-00-53.png
Attachment 17: Screenshot_from_2024-02-18_17-01-32.png
Screenshot_from_2024-02-18_17-01-32.png
Attachment 18: Screenshot_from_2024-02-18_17-01-45.png
Screenshot_from_2024-02-18_17-01-45.png
Attachment 19: IMG_7934.jpg
IMG_7934.jpg
Attachment 20: Screenshot_from_2024-02-18_18-42-41.png
Screenshot_from_2024-02-18_18-42-41.png
Attachment 21: Screenshot_from_2024-02-18_18-43-16.png
Screenshot_from_2024-02-18_18-43-16.png
Attachment 22: Screenshot_from_2024-02-18_18-43-35.png
Screenshot_from_2024-02-18_18-43-35.png
Attachment 23: Screenshot_from_2024-02-18_18-43-58.png
Screenshot_from_2024-02-18_18-43-58.png
Attachment 24: Screenshot_from_2024-02-18_18-44-24.png
Screenshot_from_2024-02-18_18-44-24.png
Attachment 25: Screenshot_from_2024-02-18_18-44-44.png
Screenshot_from_2024-02-18_18-44-44.png
Attachment 26: Screenshot_from_2024-02-18_18-44-52.png
Screenshot_from_2024-02-18_18-44-52.png
Attachment 27: Screenshot_from_2024-02-18_22-42-10.png
Screenshot_from_2024-02-18_22-42-10.png
  526   Sun Feb 18 13:10:18 2024 TDSunday 18 February - DSSSD 0 aida01 and aida04 *.*.L spectra
DSSSD 0 aida01 1*L spectra, aida04 **L spectra

Double peak structure feature of aida04 ( n+n Ohmic side ) strips. Single ( albeit sometimes broad peak ) observed in aida01 ( p+n junction side ) strips.
Not currently observed by DSSSD 1-3. DSSSD 0 high energy peak rate far higher than DSSSD 1-3.

Data from last week did not display this double peak but DSSSDs and beam have been moved since.

Unclear whether this reflects radiation damage and/or wafer/active area edge effects on the n+n side.
Might also reflect beam characteristics throughout cycle.

Requires calibration and dE cut to investigate.

JM comment -> early time cut analysis indicates this is a beam effect. DSSD#0 has majority of halo on it so double peak is very pronounced. 
We observe double peak all DSSDs just to a much lower level. 
Attachment 1: Screenshot_from_2024-02-18_13-58-29.png
Screenshot_from_2024-02-18_13-58-29.png
Attachment 2: Screenshot_from_2024-02-18_13-59-55.png
Screenshot_from_2024-02-18_13-59-55.png
Attachment 3: Screenshot_from_2024-02-18_14-01-19.png
Screenshot_from_2024-02-18_14-01-19.png
Attachment 4: Screenshot_from_2024-02-18_14-08-36.png
Screenshot_from_2024-02-18_14-08-36.png
Attachment 5: Screenshot_from_2024-02-18_14-09-44.png
Screenshot_from_2024-02-18_14-09-44.png
  525   Sun Feb 18 07:53:51 2024 HW AGRun 40 Morning Shift

08:57 Doing morning check. Checked the leakage currents, all ok. Attachment 1

Tempritures of, attchment 2

System wide check, all correct.

Statistics ok, aida 4 was a little high, carlo oked it attachment 3 and 4 (hist)

Spectrum brozer, ok, attachment 5 and 6

Waveforem atachment 7 and 8

Checked merger, spinner spinning...

Attached screenshot of merger  9

 

11:15 Beam current is around 2.5E7 and the target dencity is 5.3E11 cm2, there was small isntability (dips) from the beam, reason unknown, all detectors stable.

Check leakage current, is stable,  attachment 1

Tempreture ok, attachment 2

System wide check, all pass

Statistics checked, attachment 3, aida 4 is still high, histograms attachment 4

Spectrum browser ok, attachment 5 and 6

Waveforms attachment 7 and 8

Checked merger, ok, spinner is spinning,

Attachment 9 is the merger info

12:21 Targer intencity started dropping quite dramatically. The beam position X value was adjusted to -11000, we got back above 4E11 /cm2

13:08 Beam current is around 2.5E7 pretty stable, target is working, 61E11 /cm2, stable

Checked leakage current, stable, attachment 1

Temp. stable, atatchment 2

System wide chack, all pass

Statistics checked, attachemnt 3, aida 4 should be checked, histogram of rates 4 attached

Spectrum browser, attached 5 and 6

Waveform attached 7 and 8

Merger ok, spinner is pinning,

Attached 9 merger info

 

 

Attachment 1: Screenshot_from_2024-02-18_08-58-55.png
Screenshot_from_2024-02-18_08-58-55.png
Attachment 2: Screenshot_from_2024-02-18_09-02-22.png
Screenshot_from_2024-02-18_09-02-22.png
Attachment 3: Screenshot_from_2024-02-18_09-09-25.png
Screenshot_from_2024-02-18_09-09-25.png
Attachment 4: Screenshot_from_2024-02-18_09-11-15.png
Screenshot_from_2024-02-18_09-11-15.png
Attachment 5: Screenshot_from_2024-02-18_09-14-10.png
Screenshot_from_2024-02-18_09-14-10.png
Attachment 6: Screenshot_from_2024-02-18_09-14-41.png
Screenshot_from_2024-02-18_09-14-41.png
Attachment 7: Screenshot_from_2024-02-18_09-15-52.png
Screenshot_from_2024-02-18_09-15-52.png
Attachment 8: Screenshot_from_2024-02-18_09-16-34.png
Screenshot_from_2024-02-18_09-16-34.png
Attachment 9: Screenshot_from_2024-02-18_09-20-35.png
Screenshot_from_2024-02-18_09-20-35.png
Attachment 10: Screenshot_from_2024-02-18_11-17-59.png
Screenshot_from_2024-02-18_11-17-59.png
Attachment 11: Screenshot_from_2024-02-18_11-18-31.png
Screenshot_from_2024-02-18_11-18-31.png
Attachment 12: Screenshot_from_2024-02-18_11-20-03.png
Screenshot_from_2024-02-18_11-20-03.png
Attachment 13: Screenshot_from_2024-02-18_11-21-37.png
Screenshot_from_2024-02-18_11-21-37.png
Attachment 14: Screenshot_from_2024-02-18_11-23-10.png
Screenshot_from_2024-02-18_11-23-10.png
Attachment 15: Screenshot_from_2024-02-18_11-24-12.png
Screenshot_from_2024-02-18_11-24-12.png
Attachment 16: Screenshot_from_2024-02-18_11-24-59.png
Screenshot_from_2024-02-18_11-24-59.png
Attachment 17: Screenshot_from_2024-02-18_11-25-49.png
Screenshot_from_2024-02-18_11-25-49.png
Attachment 18: Screenshot_from_2024-02-18_11-28-06.png
Screenshot_from_2024-02-18_11-28-06.png
Attachment 19: Screenshot_from_2024-02-18_13-09-43.png
Screenshot_from_2024-02-18_13-09-43.png
Attachment 20: Screenshot_from_2024-02-18_13-11-14.png
Screenshot_from_2024-02-18_13-11-14.png
Attachment 21: Screenshot_from_2024-02-18_13-14-48.png
Screenshot_from_2024-02-18_13-14-48.png
Attachment 22: Screenshot_from_2024-02-18_13-18-19.png
Screenshot_from_2024-02-18_13-18-19.png
Attachment 23: Screenshot_from_2024-02-18_13-20-35.png
Screenshot_from_2024-02-18_13-20-35.png
Attachment 24: Screenshot_from_2024-02-18_13-21-01.png
Screenshot_from_2024-02-18_13-21-01.png
Attachment 25: Screenshot_from_2024-02-18_13-21-26.png
Screenshot_from_2024-02-18_13-21-26.png
Attachment 26: Screenshot_from_2024-02-18_13-21-54.png
Screenshot_from_2024-02-18_13-21-54.png
Attachment 27: Screenshot_from_2024-02-18_13-23-21.png
Screenshot_from_2024-02-18_13-23-21.png
  524   Sun Feb 18 00:16:55 2024 JM CBRun 40 - night shift

01:17

Run 40 continues.

Beam intensity looks OK. Oscillates between just below 2.5e7 and just below 3e7. (attachment 1)

Target density 4.3e11, YR09 pressure 1.54e-10 mbar.

Rates in monitoring code show beam on target.

FEE temperatures all OK - plot over time (attachment 2)

System check OK

Leakage current OK -trending down (attachment 3)

Tape server and merger OK. Merger stats (attachment 4)

ADC data items looks OK, spectrum browser rates OK

Spectrum browser stat (attachment 7 LOG scale) - looks like some HEC were firing previously. Spec browser histograms zero'd to monitor

pn + nn waveforms (attachments 5+6)

Monitoring code energy histograms (attachment 8)

02:40

X2go screen size has decreased further -> cannot see target density but can view beam intensity. Will use ML vac plot to help gauge beam on target

Beam intensity has decreased over the last hour but has now stabilised and is now ~2e7 at injection. Pressure in YR09 has decreased slightly accordingly from ML online but not significantly (still >1e-10 mbar)

Monitoring code rate plots indicate beam on target

All daq checks OK

Will continue to monitor beam intensity

04:14

Ring intensity has increased and appears stable back at 2.5e7.

Target density has slightly decreased to 3e11, ring pressure YR09 stable at 1.5e-10 mbar. I am reluctant to try and move the nozzle while my x2go screen is not 100%.

FEE temps OK

System checks OK

Merger and tape server OK

Rates in all FEEs look OK

Spectrum browser stats (attchment 9) no HEC firing

Monitoring code rates look OK -> beam on target (attachment 10)

05:14

Beam intensity looks OK, Target density 3.11e11. Ring pressure YR09 2e-10 mbar.

All checks OK

Monitoring code rate looks OK -> beam on target

06:20

Beam intensity OK, Target density 3.1e11. Pressure YR09 1.9e-10 mbar.

FEE temps OK

System checks OK

Statistics and rates of all FEEs OK

Merger and Tape server OK

Monitoring code rates look OK -> beam on target

Merger statistics (attachment 11)

7:15

Beam intensity 2E7, target density getting bad 2eE11. WIll fix it from the lab.

FEE temps OK

System checks OK

Merger OK

Stats OK

Attachment 1: 20240218_011615.jpg
20240218_011615.jpg
Attachment 2: Screenshot_from_2024-02-18_01-23-04.png
Screenshot_from_2024-02-18_01-23-04.png
Attachment 3: Screenshot_from_2024-02-18_01-25-06.png
Screenshot_from_2024-02-18_01-25-06.png
Attachment 4: Screenshot_from_2024-02-18_01-26-35.png
Screenshot_from_2024-02-18_01-26-35.png
Attachment 5: Screenshot_from_2024-02-18_01-36-02.png
Screenshot_from_2024-02-18_01-36-02.png
Attachment 6: Screenshot_from_2024-02-18_01-36-38.png
Screenshot_from_2024-02-18_01-36-38.png
Attachment 7: Screenshot_from_2024-02-18_01-32-12.png
Screenshot_from_2024-02-18_01-32-12.png
Attachment 8: Screenshot_from_2024-02-18_01-39-29.png
Screenshot_from_2024-02-18_01-39-29.png
Attachment 9: Screenshot_from_2024-02-18_04-18-35.png
Screenshot_from_2024-02-18_04-18-35.png
Attachment 10: Screenshot_from_2024-02-18_04-21-02.png
Screenshot_from_2024-02-18_04-21-02.png
Attachment 11: Screenshot_from_2024-02-18_06-35-40.png
Screenshot_from_2024-02-18_06-35-40.png
Attachment 12: Screenshot_from_2024-02-18_07-12-05.png
Screenshot_from_2024-02-18_07-12-05.png
Attachment 13: Screenshot_from_2024-02-18_07-12-44.png
Screenshot_from_2024-02-18_07-12-44.png
Attachment 14: Screenshot_from_2024-02-18_07-13-08.png
Screenshot_from_2024-02-18_07-13-08.png
Attachment 15: Screenshot_from_2024-02-18_07-13-25.png
Screenshot_from_2024-02-18_07-13-25.png
  523   Sat Feb 17 14:29:03 2024 JM, CB, AR, ABRun 40 - cooled beam on target

15:27

Run 40 started cooled beam on target

Detectors IN (left arm pot =36.5 mm, right arm pot = 70.49 mm). Stationary. Same as run 35

Electron cooler voltage 751.095 this corresponds to 1.338 MeV/u. This should be on the left of the peak of the resonance.

Target density is 4e11 at/cm2, ring pressure looks OK YR09 1.43e-10 mbar.

Horizontal beam position is -7.5 mm, vertical beam position is 0.0 mm. Same as run 35.

Beam is in the ring for 4 seconds and total cycle time between injections is 8 seconds. Notice this is different than run 35.

Statistics, merger statistics, histograms all zero'd

15:39

System wide checks ok. Merger ok. Temperatures ok.

Attatchment 1: statistics

Attatchment 2: Temperatures

Attatchment 3-6: Waveforms. Layout 2 - (attach 4) had a strnageish waveform for aida11, stays after reload.It was the HEC firing so did an ASIC control and Zero'd histograms and all is okay.

Attatchment 7: Leakage current

17:04

System wide checks ok. Merger ok. Temperatures ok. The beam is on target and everything runs smoothly.

Attatchment 8: statistics

Attatchment 9: Temperatures

Attatchment 10-13: Waveforms. Everything looks ok.

Attatchment 14: Leakage current

18:55

System wide checks ok. Merger ok. Temperatures ok. The beam is on target and everything runs smoothly.

Beam transformer shows a small decrease in beam intensity, not drastic but small slow decrease, see attachment 22. Target density 4e11. Vacuum pressue ok.

Attatchment 15: Statistics

Attatchment 16: Temperatures

Attatchment 17-20: Waveforms

Attatchment 21: Leakage current

19:28

Run40 - Beam dropped intesntiy at 19:17, moved detectors out, called gleb. Beam intensity started increasing again - not sure whats happenig at present (see image 23). 

20:26

Around 19:50 beam decreased again and stayed down for a while. Beam not stable. Gleb was called, target turned off, beam restabilied - around 2.5e7 ish. Unsure if it will remain stable during the night (image 24).

20:35

detectors moved back in. Target back in, stable at 4e11.

Potentiometer Left: 36.54 mm

Potentiometer Right: 70.49 mm

Beam stable at the moment (image 25).

 

System checks 20:40

   
             Base         Current         Difference
aida03 fault      0x0 :      0x1 :      1  
aida04 fault      0x0 :      0x1 :      1  
aida05 fault      0x0 :      0x1 :      1  
aida06 fault      0x0 :      0x1 :      1  
aida07 fault      0x0 :      0x1 :      1  
aida08 fault      0xe :      0xf :      1  
aida09 fault      0xb :      0xc :      1  
aida10 fault      0x0 :      0x1 :      1  
aida13 fault      0x0 :      0x1 :      1  
aida14 fault      0x0 :      0x1 :      1  
FPGA Timestamp error counter test result: Passed 6, Failed 10
If any of these counts are reported as in error
The ASIC readout system has detected a timeslip.
That is the timestamp read from the time FIFO is not younger than the last

System wide checks ok except the FPGA timestamps - see error above - collected them and it went away. Merger ok. Temperatures ok. The beam is on target and everything runs smoothly for now.

Attatchment 26: Temperatures

Attatchment 27-30: Waveforms

Attatchment 31: Leakage current

 

21:50

FEE temperatures OK

System wide checks OK

Statistics and spectrum browser rate look OK

Merger and tape server OK

Monitoring code rates are consistent with beam on target

X2go seems to not be loading full screen for me - may be my laptop misbehaving?

Can see ring intensity has been stable for the last hour, target density is 4.33e11 and YR09 pressure is 1-2e-10 mbar.

 

23:22

Target density 4.3e11, beam intensity stable.

Monitoring code rates are consistent with beam on target

Fee temps all OK

System checks OK

Statistics and FEErates OK.

Merger and tape server OK

 

 

 

Attachment 1: Screenshot_from_2024-02-17_15-39-50.png
Screenshot_from_2024-02-17_15-39-50.png
Attachment 2: Screenshot_from_2024-02-17_15-40-05.png
Screenshot_from_2024-02-17_15-40-05.png
Attachment 3: Screenshot_from_2024-02-17_15-40-49.png
Screenshot_from_2024-02-17_15-40-49.png
Attachment 4: Screenshot_from_2024-02-17_15-43-04.png
Screenshot_from_2024-02-17_15-43-04.png
Attachment 5: Screenshot_from_2024-02-17_15-44-12.png
Screenshot_from_2024-02-17_15-44-12.png
Attachment 6: Screenshot_from_2024-02-17_15-44-36.png
Screenshot_from_2024-02-17_15-44-36.png
Attachment 7: Screenshot_from_2024-02-17_15-45-02.png
Screenshot_from_2024-02-17_15-45-02.png
Attachment 8: Screenshot_from_2024-02-17_16-58-47.png
Screenshot_from_2024-02-17_16-58-47.png
Attachment 9: Screenshot_from_2024-02-17_16-59-18.png
Screenshot_from_2024-02-17_16-59-18.png
Attachment 10: Screenshot_from_2024-02-17_17-00-48.png
Screenshot_from_2024-02-17_17-00-48.png
Attachment 11: Screenshot_from_2024-02-17_17-01-41.png
Screenshot_from_2024-02-17_17-01-41.png
Attachment 12: Screenshot_from_2024-02-17_17-02-26.png
Screenshot_from_2024-02-17_17-02-26.png
Attachment 13: Screenshot_from_2024-02-17_17-03-01.png
Screenshot_from_2024-02-17_17-03-01.png
Attachment 14: Screenshot_from_2024-02-17_17-03-51.png
Screenshot_from_2024-02-17_17-03-51.png
Attachment 15: Screenshot_from_2024-02-17_18-55-13.png
Screenshot_from_2024-02-17_18-55-13.png
Attachment 16: Screenshot_from_2024-02-17_18-55-48.png
Screenshot_from_2024-02-17_18-55-48.png
Attachment 17: Screenshot_from_2024-02-17_18-56-55.png
Screenshot_from_2024-02-17_18-56-55.png
Attachment 18: Screenshot_from_2024-02-17_18-57-26.png
Screenshot_from_2024-02-17_18-57-26.png
Attachment 19: Screenshot_from_2024-02-17_18-57-54.png
Screenshot_from_2024-02-17_18-57-54.png
Attachment 20: Screenshot_from_2024-02-17_18-58-27.png
Screenshot_from_2024-02-17_18-58-27.png
Attachment 21: Screenshot_from_2024-02-17_18-58-44.png
Screenshot_from_2024-02-17_18-58-44.png
Attachment 22: IMG_7902.jpeg
IMG_7902.jpeg
Attachment 23: IMG_7903.jpeg
IMG_7903.jpeg
Attachment 24: IMG_7906.jpeg
IMG_7906.jpeg
Attachment 25: IMG_7907.jpeg
IMG_7907.jpeg
Attachment 26: Screenshot_from_2024-02-17_20-38-37.png
Screenshot_from_2024-02-17_20-38-37.png
Attachment 27: Screenshot_from_2024-02-17_20-42-41.png
Screenshot_from_2024-02-17_20-42-41.png
Attachment 28: Screenshot_from_2024-02-17_20-43-17.png
Screenshot_from_2024-02-17_20-43-17.png
Attachment 29: Screenshot_from_2024-02-17_20-43-50.png
Screenshot_from_2024-02-17_20-43-50.png
Attachment 30: Screenshot_from_2024-02-17_20-55-50.png
Screenshot_from_2024-02-17_20-55-50.png
Attachment 31: Screenshot_from_2024-02-17_20-45-30.png
Screenshot_from_2024-02-17_20-45-30.png
  522   Sat Feb 17 13:21:44 2024 JM, CBBeam on target - motors tests

14:01

Run 36 started.

We put both detectors to the fully out position. Detectors then driven IN using the pnuematic motor only once the beam is ready and driven fully out before beam dumping.

We notice the right arm is significantly slower than the left arm and does not manage to go fully out before being told to go back in. This is not ideal.

Run stopped.

14:10

Run 37 started

Right arm is positioned at pot = 70.49 and remains stationary. The left arm is moved with the ring cycle. When IN left arm pot = 62.71 mm. When out left arm pot = 18.89 mm

Movement signals are being saved to jan MBS system and can be used to seperate out injection periods on other runs where both detectors will be stationary.

All beam and target settings are the same as for Run 35.

Spectrum browser rates for each FEE when left arm IN shown in (attachment 1). Rates dependent on beam cycle, when detectors move out, rates in left arm drop significantly as expected.

XY and energy histogram plots from monitoring code (attachments 2+3)

We see the Rutherford much more well defined on the bottom left detector now (it is further in and halo effect is reduced). Still some low energy counts (may be due to detector movement -> need to test without beam)

Top left detector energy histogram looks a lot different, still much halo effect.

we will collect some stats for 30 mins or so

14:57

Run 37 Stopped. Run 38 started.

Faraday cup to the ring closed. No beam in the ring. Left arm of the detectors still moving IN and out of the beam with the ring cycle.

Very little counts observed in energy spectra.

15:02

Run 38 stopped so Gleb can change the beam energy. Run 39 started.

 

Attachment 1: Screenshot_from_2024-02-17_14-33-51.png
Screenshot_from_2024-02-17_14-33-51.png
Attachment 2: Screenshot_from_2024-02-17_14-36-27.png
Screenshot_from_2024-02-17_14-36-27.png
Attachment 3: Screenshot_from_2024-02-17_14-39-38.png
Screenshot_from_2024-02-17_14-39-38.png
  521   Sat Feb 17 06:19:27 2024 JM, CBRun 35 - contd

07:10

FEE temperatures OK

System checks OK
Merger and tape server OK
Statistics look OK,
Rates in the detectors from the mointoring code indicate we have beam on target.

Detector leakge current appears to be following same pattern as yesterday -> temperature (attachment 1)

Merger stats (attachment 2) - quite a few more pause / resume statements than normal - noise last night?

Target density appears stable over the last 3 hours ~3.8e11 at/cm2, ring pressure stable YR09 ~1.5e-10 mbar

Beam injection looks stable over the last few hours.

09:08

FEE temperatures OK -aida03 and aida09 0.5 C over 65C warning.

system checks OK

ADC data item stats  OK

Spectrum browser rate (attachment 3)

Merger and Tape server OK

Monitoring  xy and energy histograms since zero'd last night (attachments 4+5)

Target density slight drop to ~3.5e11 at/cm2. Pressure in the ring stable. CB attempting to optimise target density and ring pressure.

Beam injection into the ring still good

11:04

FEE temperatures OK

System checks OK

ADC data items and spectrum browser rates OK

Target density ~4.6e11 at/cm2, ring pressure looks OK ~1e-10 mbar

Beam injection still good.

Merger and tape server OK

12:45

Beam injection over last 2 hours has been stable

Target density 4.75e11 and stable. ring pressures are stable.

FEE temperatures OK

System checks OK

ADC data items stats and spectrum browser rates look OK

Merger and tape server OK

14:01

We wish to test the beam ahlo on injection so are ending run

Run stopped.

Attachment 1: Screenshot_from_2024-02-17_07-20-01.png
Screenshot_from_2024-02-17_07-20-01.png
Attachment 2: Screenshot_from_2024-02-17_07-21-27.png
Screenshot_from_2024-02-17_07-21-27.png
Attachment 3: Screenshot_from_2024-02-17_09-09-35.png
Screenshot_from_2024-02-17_09-09-35.png
Attachment 4: Screenshot_from_2024-02-17_09-13-44.png
Screenshot_from_2024-02-17_09-13-44.png
Attachment 5: Screenshot_from_2024-02-17_09-14-35.png
Screenshot_from_2024-02-17_09-14-35.png
  520   Fri Feb 16 22:03:27 2024 TDRun 35 - night shift
22.51 DAQ continues file G22-86/R35_36

      DSSSD bias OK - attachments 1-2

      FEE64 temperatures - attachments 3-4
      CRYRING hall ambient temperature 20
      FEE64 temps tracking ambient temperature - aida01, aida03, aida09 and aida15 - all ( just ) over 65 deg C

      All system wide checks OK

      WR timestamps OK - attachment 5

      ADC, DISC and PAUSE data item stats - attachments 6-8
       hot channels aida01 ( 91k ), aida04 ( 12k ), aida15 ( 11k ) 

      per FEE64 Rate spectra - attachments 9-10
       aida04 & aida13 only FEE64s with significant rates and systematic rate variation with channel #

      per FEE64 1.8.L, 1.8.H and 1.8.W 20us FSR spectra - attachments 11-16

      Merger, Merger stats and TapeServer OK - attachments 17-19

23.16 All histograms, stats and merger stats zero'd

      Online analysis spectra - attachments 21-26
       only the inner quadrants of the 4x DSSSDs are correctly mapped - attachment 21
       observe double peak structure in DSSSD0 rate spectra - radiation damage most forward lab angle strips? - attachment 24

23.23 restart online analysis - zero spectra

23.30 YR12DTIC 2.5e7/s
      YR09 vacuum 7e-11 - 1.2e-10 mbar
      density 4e11/cm2
      See attachments 27-34



01.30 DAQ continues file G22-86/R35_39

      FEE64 temperatures - attachments 35-36

      ADC data item stats - attachment 37

      DSSSD leakage currents - attachment 38

      YR12DTIC 2.5e7/s
      YR09 vacuum 7e-11 - 1.2e-10 mbar
      density 4.1e11/cm2


03.05 YR12DTIC 2.2e7/s
      YR09 vacuum 7e-11 - 1.2e-10 mbar
      density 3.5e11/cm2

      DAQ continues file G22-86/R35_42

      DSSSD leakage currents - attachment 39

      FEE64 temperatures - attachments 40-41

      ADC data item stats - attachment 42

05.05 DAQ continues file G22-86/R35_47

      YR12DTIC 2.5e7/s
      density 3.8e11/cm2

      DAQ continues file G22-86/R35_42

      DSSSD leakage currents - attachment 43

      FEE64 temperatures - attachments 44-45

      ADC data item stats - attachment 46

      per FEE64 Rate spectra - attachments 47-48
       aida12 asic #1 and # 2 HEC activity 
       and following Check ASIC load aida12

      ADC data item stats - attachment 49


Shift Summary

Beam, target density and vacuum appear to have been stable overnight

FEE64 temperatures continue to cycle

Evidence of some radiation damage DSSSDD 0 presumably for most forward lab angle strips?

aida12 asic #1 and #2 high rate HEC activity - fixed by ASIC check load
Attachment 1: Screenshot_from_2024-02-16_22-51-27.png
Screenshot_from_2024-02-16_22-51-27.png
Attachment 2: Screenshot_from_2024-02-16_22-52-15.png
Screenshot_from_2024-02-16_22-52-15.png
Attachment 3: Screenshot_from_2024-02-16_22-53-02.png
Screenshot_from_2024-02-16_22-53-02.png
Attachment 4: Screenshot_from_2024-02-16_22-54-10.png
Screenshot_from_2024-02-16_22-54-10.png
Attachment 5: Screenshot_from_2024-02-16_22-55-11.png
Screenshot_from_2024-02-16_22-55-11.png
Attachment 6: Screenshot_from_2024-02-16_22-55-38.png
Screenshot_from_2024-02-16_22-55-38.png
Attachment 7: Screenshot_from_2024-02-16_22-55-55.png
Screenshot_from_2024-02-16_22-55-55.png
Attachment 8: Screenshot_from_2024-02-16_22-56-24.png
Screenshot_from_2024-02-16_22-56-24.png
Attachment 9: Screenshot_from_2024-02-16_22-57-14.png
Screenshot_from_2024-02-16_22-57-14.png
Attachment 10: Screenshot_from_2024-02-16_22-57-37.png
Screenshot_from_2024-02-16_22-57-37.png
Attachment 11: Screenshot_from_2024-02-16_22-58-23.png
Screenshot_from_2024-02-16_22-58-23.png
Attachment 12: Screenshot_from_2024-02-16_22-59-08.png
Screenshot_from_2024-02-16_22-59-08.png
Attachment 13: Screenshot_from_2024-02-16_22-59-46.png
Screenshot_from_2024-02-16_22-59-46.png
Attachment 14: Screenshot_from_2024-02-16_23-00-33.png
Screenshot_from_2024-02-16_23-00-33.png
Attachment 15: Screenshot_from_2024-02-16_23-01-18.png
Screenshot_from_2024-02-16_23-01-18.png
Attachment 16: Screenshot_from_2024-02-16_23-01-59.png
Screenshot_from_2024-02-16_23-01-59.png
Attachment 17: Screenshot_from_2024-02-16_23-02-21.png
Screenshot_from_2024-02-16_23-02-21.png
Attachment 18: Screenshot_from_2024-02-16_23-02-36.png
Screenshot_from_2024-02-16_23-02-36.png
Attachment 19: Screenshot_from_2024-02-16_23-02-48.png
Screenshot_from_2024-02-16_23-02-48.png
Attachment 20: Capture.PNG
Capture.PNG
Attachment 21: Screenshot_from_2024-02-16_23-18-08.png
Screenshot_from_2024-02-16_23-18-08.png
Attachment 22: Screenshot_from_2024-02-16_23-18-22.png
Screenshot_from_2024-02-16_23-18-22.png
Attachment 23: Screenshot_from_2024-02-16_23-18-34.png
Screenshot_from_2024-02-16_23-18-34.png
Attachment 24: Screenshot_from_2024-02-16_23-19-00.png
Screenshot_from_2024-02-16_23-19-00.png
Attachment 25: Screenshot_from_2024-02-16_23-19-13.png
Screenshot_from_2024-02-16_23-19-13.png
Attachment 26: Screenshot_from_2024-02-16_23-19-26.png
Screenshot_from_2024-02-16_23-19-26.png
Attachment 27: Capture.PNG
Capture.PNG
Attachment 28: 20240216_222850.jpg
20240216_222850.jpg
Attachment 29: 20240216_222907.jpg
20240216_222907.jpg
Attachment 30: 20240216_222736.jpg
20240216_222736.jpg
Attachment 31: 20240216_222950.jpg
20240216_222950.jpg
Attachment 32: 20240216_222747.jpg
20240216_222747.jpg
Attachment 33: 20240216_222928.jpg
20240216_222928.jpg
Attachment 34: 20240216_222941.jpg
20240216_222941.jpg
Attachment 35: Screenshot_from_2024-02-17_01-24-57.png
Screenshot_from_2024-02-17_01-24-57.png
Attachment 36: Screenshot_from_2024-02-17_01-26-57.png
Screenshot_from_2024-02-17_01-26-57.png
Attachment 37: Screenshot_from_2024-02-17_01-27-34.png
Screenshot_from_2024-02-17_01-27-34.png
Attachment 38: Screenshot_from_2024-02-17_01-29-05.png
Screenshot_from_2024-02-17_01-29-05.png
Attachment 39: Screenshot_from_2024-02-17_03-07-38.png
Screenshot_from_2024-02-17_03-07-38.png
Attachment 40: Screenshot_from_2024-02-17_03-08-05.png
Screenshot_from_2024-02-17_03-08-05.png
Attachment 41: Screenshot_from_2024-02-17_03-08-57.png
Screenshot_from_2024-02-17_03-08-57.png
Attachment 42: Screenshot_from_2024-02-17_03-09-26.png
Screenshot_from_2024-02-17_03-09-26.png
Attachment 43: Screenshot_from_2024-02-17_05-06-46.png
Screenshot_from_2024-02-17_05-06-46.png
Attachment 44: Screenshot_from_2024-02-17_05-07-11.png
Screenshot_from_2024-02-17_05-07-11.png
Attachment 45: Screenshot_from_2024-02-17_05-08-19.png
Screenshot_from_2024-02-17_05-08-19.png
Attachment 46: Screenshot_from_2024-02-17_05-09-42.png
Screenshot_from_2024-02-17_05-09-42.png
Attachment 47: Screenshot_from_2024-02-17_05-11-03.png
Screenshot_from_2024-02-17_05-11-03.png
Attachment 48: Screenshot_from_2024-02-17_05-11-27.png
Screenshot_from_2024-02-17_05-11-27.png
Attachment 49: Capture.PNG
Capture.PNG
  519   Fri Feb 16 10:02:41 2024 JM, OF, CB, AR, EMRun 35 - cooled beam on target

11:01

Run 35 started cooled beam on target

Detectors IN (left arm pot =36.5 mm, right arm pot = 70.49 mm)

Electron cooler voltage 751.4 this corresponds to 1.3436 MeV/u. This should be on the peak of the resonance.

Target density is 3.75e11 at/cm2, ring pressure looks OK YR09 1.43e-10 mbar.

Horizontal beam position is -7.5 mm, vertical beam position is 0.0 mm. Same as run 33.

Beam is in the ring for 8 seconds and total cycle time between injections is 12 seconds. Same as run 33

Beam injection looks stable. ~2e7 injected and no missed injections so far.

Merger OK, Tape server OK

Merger statistics zero'd

All histograms and statistics zero'd, monitoring code reset.

ADC data item statistics looks as before, very few counts until injection into the ring. OK

FEE temperatures look OK

System checks OK

Leakage current trending upwards, same as previous elog.

13:33

FEE temperatures OK - all FEEs at least 1.5C below vertex warning

ADC data items statistics OK

Merger and Tape server OK

Target density 3.8e11 at/cm2, ring pressure stable.

Ring injections look very stable

Leakage current now decreasing (attachment 1)

Monitoring code xy and energy histograms (attachment 2+3)

15:03

System wide checks ok. Merger ok. Temperatures ok.

Attatchment 4: Stats

Attatchment 5: Temperatures

Attatchments 6-9: Waveforms.

Attatchment 10: Leakage current

Attatchment 11: lowenergyEytotal plots

17:09

System wide check ok. Merger ok. Temperature ok.

Attachment 12: Stats

Attachment 13: Temperatures

Attachment 14-18: Waveforms (Layout 7: strange behaviour in aida01, went to ASIC Control and did "Check ASIC control" three times, which resolved the problem)

Attachment 19: Leakage current

Attachment 20: lowenergyEytotal plots

Gas pressure was going up a little bit: changed Y position of the nozzle from 1000 to 950.

17:55

The gas jet target pressure started to fall sharply at 17:34 and went down to zero within 2 minutes.

It came back while it we were changing the nozzle position but at 17:49 fell to zero again.

Currently waiting for Nikos to came and have look at it.

18:10

Target back in place and has a density of 4.04e11. Beam transformer looks similar to pre plummeting target density.

 

 

18.55 Check ASIC load FEE64 aida16 asic#3 HEC channels active

 Following check load merger rate decreased significantly.

19:14

Gas density remains stable around 4e11 after the incident at 17:34 and the beam transformer singal looks ok too. Gas pressure ok.

Merger: All aidas turn green but the merger has decreased significantly (too about 306 000 items/sec) as mentioned above.

System wide check ok. Temperature ok.

Attachment 21: Stats

Attachment 22: Temperatures

Attachment 23-26: Waveforms

Attachment 27: Leakage current

 

 

 

 

Attachment 1: Screenshot_from_2024-02-16_13-44-25.png
Screenshot_from_2024-02-16_13-44-25.png
Attachment 2: Screenshot_from_2024-02-16_13-46-26.png
Screenshot_from_2024-02-16_13-46-26.png
Attachment 3: Screenshot_from_2024-02-16_13-46-52.png
Screenshot_from_2024-02-16_13-46-52.png
Attachment 4: Screenshot_from_2024-02-16_14-56-20.png
Screenshot_from_2024-02-16_14-56-20.png
Attachment 5: Screenshot_from_2024-02-16_14-56-38.png
Screenshot_from_2024-02-16_14-56-38.png
Attachment 6: Screenshot_from_2024-02-16_14-58-01.png
Screenshot_from_2024-02-16_14-58-01.png
Attachment 7: Screenshot_from_2024-02-16_14-58-31.png
Screenshot_from_2024-02-16_14-58-31.png
Attachment 8: Screenshot_from_2024-02-16_14-59-59.png
Screenshot_from_2024-02-16_14-59-59.png
Attachment 9: Screenshot_from_2024-02-16_15-00-11.png
Screenshot_from_2024-02-16_15-00-11.png
Attachment 10: Screenshot_from_2024-02-16_15-00-54.png
Screenshot_from_2024-02-16_15-00-54.png
Attachment 11: Screenshot_from_2024-02-16_15-02-24.png
Screenshot_from_2024-02-16_15-02-24.png
Attachment 12: Screenshot_from_2024-02-16_16-54-55.png
Screenshot_from_2024-02-16_16-54-55.png
Attachment 13: Screenshot_from_2024-02-16_16-55-17.png
Screenshot_from_2024-02-16_16-55-17.png
Attachment 14: Screenshot_from_2024-02-16_16-56-23.png
Screenshot_from_2024-02-16_16-56-23.png
Attachment 15: Screenshot_from_2024-02-16_16-57-08.png
Screenshot_from_2024-02-16_16-57-08.png
Attachment 16: Screenshot_from_2024-02-16_16-58-01.png
Screenshot_from_2024-02-16_16-58-01.png
Attachment 17: Screenshot_from_2024-02-16_17-04-41.png
Screenshot_from_2024-02-16_17-04-41.png
Attachment 18: Screenshot_from_2024-02-16_17-06-07.png
Screenshot_from_2024-02-16_17-06-07.png
Attachment 19: Screenshot_from_2024-02-16_17-08-45.png
Screenshot_from_2024-02-16_17-08-45.png
Attachment 20: Screenshot_from_2024-02-16_17-20-51.png
Screenshot_from_2024-02-16_17-20-51.png
Attachment 21: Screenshot_from_2024-02-16_19-10-26.png
Screenshot_from_2024-02-16_19-10-26.png
Attachment 22: Screenshot_from_2024-02-16_19-10-39.png
Screenshot_from_2024-02-16_19-10-39.png
Attachment 23: Screenshot_from_2024-02-16_19-11-56.png
Screenshot_from_2024-02-16_19-11-56.png
Attachment 24: Screenshot_from_2024-02-16_19-12-21.png
Screenshot_from_2024-02-16_19-12-21.png
Attachment 25: Screenshot_from_2024-02-16_19-12-50.png
Screenshot_from_2024-02-16_19-12-50.png
Attachment 26: Screenshot_from_2024-02-16_19-13-32.png
Screenshot_from_2024-02-16_19-13-32.png
Attachment 27: Screenshot_from_2024-02-16_19-14-12.png
Screenshot_from_2024-02-16_19-14-12.png
  518   Thu Feb 15 21:07:00 2024 JM, AB, AR, OFRun33 - cooled beam on target

22:27

Run 33 started cooled beam on target.

Detectors IN (left arm pot =36.5 mm, right arm pot = 70.49 mm)

Electron cooler voltage 754.945 this corresponds to 1.3499 MeV/u. This is above the resonance of interest but cooler settings cannot be altered without major beam tuning.

Target density is 4.1e11 at/cm2, ring pressure looks OK YR09 1.2e-10 mbar.

Horizontal beam position is -7.5 mm, vertical beam position is now 0.0 mm. This results in no beam halo in bottom left detector which should be good for seeing Rutherford.

Beam is in the ring for 8 seconds and total cycle time between injections is 12 seconds

FEE temperatures are OK. 3 FEEs are slightly too. (attachment 1)

System checks are OK.

Statistics look OK. High in the top left detector which is expected as majority of halo is there. (attachment 2+3)

Merger OK Tape server OK - 851 kB/s

Merger statistics (attachment 4)

Leakage current is increasing but is likely temperature related. ML temp plot has same pattern and it is a WARM day. (attachment 5)

Monitoring code - xy plot and Ey energy histograms (attachment 6+7). Rutherford rate looks higher -> looking peaky

00:14

System checks all OK

FEE temperatures unchanged. Tom has now set up plotting of aida09 fee temperatures in the second workspace.

Merger OK, tape server OK.

Target density remains about 4.1e11 at/cm2. Ring pressure ~1e-10 mbar. YR08 read close to 1e-9 but beam is able to circulate in the ring so will not change for now.

Rate statistics in all FEEs remain unchanged.

02:10

System checks OK

FEE temperatures unchanged

Merger OK

Target density OK. Ring pressure not great but cannot change remotely due to script not in screen and poor coding.

Beam stil on.

Leakage current trending up.

 

04.00 FEE64 temperatures - attachments 10-11

          aida09 ( and other FEE64s ) on downward trend

          CRYRING hall temperature - attachment 12

 

05:00

Ring pressure worsening and target getting thinner (4E11). Tweak nozzle when in the lab physically.

Beam still on around 3E7

 

Temperatures & Leakage currents OK. Same trends

Merger continues OK

Stats OK

System wide checks OK

 

07:00

Fee temperatures have reduced - all green

System check OK

ADC data items statistics unchanged. 

Merger and tape server OK

Leakage current plot (attachment 15)
monitoring  plots - xy plot and EY energy histogram (attachment 16+17 )

Pressure has dropped slightly over the last hour, target density remains stable 4.2e11. Beam transformer has dropped off in the last hour -> less beam target interactions -> lower pressure?

We still see count rates in the detectors.

08:18

Inconsistent beam into the ring continues. HKR are aware.

Temperatures OK

System checks OK

10:00

Beam is much more consistent now. We observe vast majority of shots injected into the ring and corresponding rate increase in the detectors via the monitoring code.

FEE temperatures OK

System wide checks OK

Merger and tape server OK

New merger stats (attachment 18)

ADC data item statistics OK

Leakage current plot (attachment 19)

Attachments 20+21: Cooler settings before and after changing to the new beam energy of 1.3435 MeV/u (copy from GSI olog)

10:24

Run33 stopped. Run 34 started. Changing beam energy to be on resonance.

Detectors moved fully out. Target moved out

Final spectrum browser stat for all FEEs before run ended (attachment 22)

Attachment 1: Screenshot_from_2024-02-15_22-15-04.png
Screenshot_from_2024-02-15_22-15-04.png
Attachment 2: Screenshot_from_2024-02-15_22-17-29.png
Screenshot_from_2024-02-15_22-17-29.png
Attachment 3: Screenshot_from_2024-02-15_22-18-21.png
Screenshot_from_2024-02-15_22-18-21.png
Attachment 4: Screenshot_from_2024-02-15_22-20-52.png
Screenshot_from_2024-02-15_22-20-52.png
Attachment 5: Screenshot_from_2024-02-15_22-21-43.png
Screenshot_from_2024-02-15_22-21-43.png
Attachment 6: Screenshot_from_2024-02-15_22-24-12.png
Screenshot_from_2024-02-15_22-24-12.png
Attachment 7: Screenshot_from_2024-02-15_22-24-21.png
Screenshot_from_2024-02-15_22-24-21.png
Attachment 8: Screenshot_from_2024-02-16_02-07-38.png
Screenshot_from_2024-02-16_02-07-38.png
Attachment 9: Screenshot_from_2024-02-16_02-10-16.png
Screenshot_from_2024-02-16_02-10-16.png
Attachment 10: Screenshot_from_2024-02-16_04-03-13.png
Screenshot_from_2024-02-16_04-03-13.png
Attachment 11: Screenshot_from_2024-02-16_04-04-08.png
Screenshot_from_2024-02-16_04-04-08.png
Attachment 12: Capture.PNG
Capture.PNG
Attachment 13: Screenshot_from_2024-02-16_05-04-19.png
Screenshot_from_2024-02-16_05-04-19.png
Attachment 14: Screenshot_from_2024-02-16_05-04-47.png
Screenshot_from_2024-02-16_05-04-47.png
Attachment 15: Screenshot_from_2024-02-16_07-06-22.png
Screenshot_from_2024-02-16_07-06-22.png
Attachment 16: Screenshot_from_2024-02-16_07-06-40.png
Screenshot_from_2024-02-16_07-06-40.png
Attachment 17: Screenshot_from_2024-02-16_07-07-40.png
Screenshot_from_2024-02-16_07-07-40.png
Attachment 18: Screenshot_from_2024-02-16_10-07-21.png
Screenshot_from_2024-02-16_10-07-21.png
Attachment 19: Screenshot_from_2024-02-16_10-08-02.png
Screenshot_from_2024-02-16_10-08-02.png
Attachment 20: olog-before.pdf
olog-before.pdf
Attachment 21: olog-after.pdf
olog-after.pdf
Attachment 22: Screenshot_from_2024-02-16_10-13-24.png
Screenshot_from_2024-02-16_10-13-24.png
  517   Thu Feb 15 13:18:30 2024 JGMBS DAQ: target & ring values

We are running an MBS DAQ to record the attached list of slow control values of the ring and the target. In most cases this is achieved by converting analog monitor voltages to frequencies and recording them with a VME scaler (VULOM4B with scaler-firmware). The complete list of signals is attached including a channel mapping.

!! Target values are currently not provided, we can only record ring/cooler values !!

We do a scaler readout with a 1Hz trigger rate right now. This can be changed. This generates events with trigger number 1.

Additionally, we receive triggers from motor movement (trigger 2 & trigger 3) and from target ON and target OFF (trigger 12 & trigger 13).

Each event of any trigger type is WR timestamped.

There  is also a Go4 configured for online monitoring.

Details below. Or contact Jan Glorius.

----------------------------------------

Basic info and how-to for MBS

----------------------------------------

linux-user: cry_exp

VME CPU: r4l-58

DAQ PC: atppc025

data storage:  lxg1275:/data.local2/2024_CRYRING_16O_jet

The DAQ is running in a screen session "target_DAQ" on atppc025. Enter by typing

> screen -x "target_DAQ"

SSH connections to r4l-58 are active in tabs 0 and 1. Type [ctrl]+[a] then [x] to switch to tab x. 

If the SSH connection is lost type

> ssh cry_exp@r4l-58

> cd /esr/usr/cry_exp/2024_CRYRING_jet/mbs

MBS is running in tab 1. If it has an error, type [ctrl]+[z] to exit MBS, then

> resl

> mbs

mbs> @startup

mbs> @connect

to start a new file recording type

mbs> @runstart

If an error occures during @runstart, try @connect again.

Inside MBS you monitoring options:

mbs> show acq   (will show you trigger and data rates and status)

mbs> show file    (will show you the file status)

mbs> type ev -v 10    (will display 10 subsequent events from the data stream)

The trigger rate is currently 1Hz.

The trigger logic and setup in configured in /esr/usr/cry_exp/2024_CRYRING_jet/mbs/jet.trlo

-----------------------

Go4 monitoring

-----------------------

Go4 is running on atppc025 in the directory /u/cry_exp/Beamtimes/202402_O2+/go4_target

From this directory you can run it by typing

> go4 monitoring.hotstart

 

 

Attachment 1: 2024_jet_daq.txt
ch 1: jet_E1
ch 2: jet_E2
ch 3: jet_E3
ch 4: jet_E4
ch 5: jet_S1
ch 6: jet_S2
ch 7: jet_S3
ch 8: jet_S4
ch 9: jet_WWK
ch 10: jet_VV1
ch 11: jet_VV2
ch 12: jet_P0
ch 13: jet_T1
ch 14: jet_T2
ch 15: ring_RF
ch 16: Schottky_span0
ch 17: HV_cooler
ch 18: collector_current
ch 19: inhibit
  516   Thu Feb 15 12:40:52 2024 JM. CB, AR, ABRun30 - uncooled beam on target

Run started 13:40

Target density 4.3e11, Beam injection is ~1e7, Pressure in the ring is ~1e-10 mbar.

Detectors have been moved IN with servo motors only (left arm pot = 36.5 mm, right arm pot =37.7 mm

Time between injections is 12 seconds. Beam is in the ring for 3 seconds of the cycle.

System wide checks OK

Temperatures OK

Leakage current plot attached -> temperature fluctuations (attachment 1)

ADC data items stats (attachment 2)

Merger and tape server OK

All histograms and statistics zero'd

14:17

Cave access required. No beam

15:24

No beam. System wide checks ok. Merger ok.

Attatchment 3: stats

Attatchment 4: Temperatures generally okay, 3 were red before reload, one red after reload.

Attatchment 5-8: Waveforms. Layout 7 and 8 look different from yesterday, consistnely different after several reloads.

Attatchment 9: Leakage current

Attatchment 10: LOWENERGYEXEY PLOTS

17:12

Tunning beam. System wide checks ok. Merger ok.

Attatchment 11: stats

Attatchment 12: Temperatures generally okay, 3 were red reloaed and stayed red but not so high.

Attatchment 13-16: Waveforms. Were the same as previous check.

Attatchment 17: Leakage current. They are tunning the beam so its a little bit changed.

19:28

Still hvae unstable noncooled beam, currently being worked on.

Tunning beam. System wide checks ok. Merger ok.

Attatchment 18: Temperatures generally okay, 3 were red reloaed and stayed red but not so high and were slighlty dicrease.

Attatchment 19-22: Waveforms. Were the same as previous check.

Attatchment 23: Leakage current. Slowly increasing from last check in, highest its been but conditions are not great at the moment.

21:10

We did not have any update about the beam. They still work on it.

Tunning beam. System wide checks ok. Merger ok. Stats were ok.

Attatchment 24: Temperatures generally okay, Same 3 were red and they increased a little bit. Nothing too concerning.

Attatchment 25-28: Waveforms. Were the same as previous check.

Attatchment 29: Leakage current. Contining to increase by small amount from last check.

 

 

 

 

Attachment 1: Screenshot_from_2024-02-15_13-45-42.png
Screenshot_from_2024-02-15_13-45-42.png
Attachment 2: Screenshot_from_2024-02-15_13-48-57.png
Screenshot_from_2024-02-15_13-48-57.png
Attachment 3: Screenshot_from_2024-02-15_15-19-31.png
Screenshot_from_2024-02-15_15-19-31.png
Attachment 4: Screenshot_from_2024-02-15_15-19-55.png
Screenshot_from_2024-02-15_15-19-55.png
Attachment 5: Screenshot_from_2024-02-15_15-21-25.png
Screenshot_from_2024-02-15_15-21-25.png
Attachment 6: Screenshot_from_2024-02-15_15-21-53.png
Screenshot_from_2024-02-15_15-21-53.png
Attachment 7: Screenshot_from_2024-02-15_15-22-32.png
Screenshot_from_2024-02-15_15-22-32.png
Attachment 8: Screenshot_from_2024-02-15_15-23-13.png
Screenshot_from_2024-02-15_15-23-13.png
Attachment 9: Screenshot_from_2024-02-15_15-23-38.png
Screenshot_from_2024-02-15_15-23-38.png
Attachment 10: Screenshot_from_2024-02-15_15-24-02.png
Screenshot_from_2024-02-15_15-24-02.png
Attachment 11: Screenshot_from_2024-02-15_17-00-51.png
Screenshot_from_2024-02-15_17-00-51.png
Attachment 12: Screenshot_from_2024-02-15_17-01-28.png
Screenshot_from_2024-02-15_17-01-28.png
Attachment 13: Screenshot_from_2024-02-15_17-03-31.png
Screenshot_from_2024-02-15_17-03-31.png
Attachment 14: Screenshot_from_2024-02-15_17-05-14.png
Screenshot_from_2024-02-15_17-05-14.png
Attachment 15: Screenshot_from_2024-02-15_17-07-20.png
Screenshot_from_2024-02-15_17-07-20.png
Attachment 16: Screenshot_from_2024-02-15_17-09-17.png
Screenshot_from_2024-02-15_17-09-17.png
Attachment 17: Screenshot_from_2024-02-15_17-11-12.png
Screenshot_from_2024-02-15_17-11-12.png
Attachment 18: Screenshot_from_2024-02-15_19-22-45.png
Screenshot_from_2024-02-15_19-22-45.png
Attachment 19: Screenshot_from_2024-02-15_19-23-53.png
Screenshot_from_2024-02-15_19-23-53.png
Attachment 20: Screenshot_from_2024-02-15_19-24-43.png
Screenshot_from_2024-02-15_19-24-43.png
Attachment 21: Screenshot_from_2024-02-15_19-25-59.png
Screenshot_from_2024-02-15_19-25-59.png
Attachment 22: Screenshot_from_2024-02-15_19-26-38.png
Screenshot_from_2024-02-15_19-26-38.png
Attachment 23: Screenshot_from_2024-02-15_19-28-03.png
Screenshot_from_2024-02-15_19-28-03.png
Attachment 24: Screenshot_from_2024-02-15_21-05-57.png
Screenshot_from_2024-02-15_21-05-57.png
Attachment 25: Screenshot_from_2024-02-15_21-07-39.png
Screenshot_from_2024-02-15_21-07-39.png
Attachment 26: Screenshot_from_2024-02-15_21-08-10.png
Screenshot_from_2024-02-15_21-08-10.png
Attachment 27: Screenshot_from_2024-02-15_21-08-48.png
Screenshot_from_2024-02-15_21-08-48.png
Attachment 28: Screenshot_from_2024-02-15_21-09-29.png
Screenshot_from_2024-02-15_21-09-29.png
Attachment 29: Screenshot_from_2024-02-15_21-10-01.png
Screenshot_from_2024-02-15_21-10-01.png
  515   Thu Feb 15 10:39:50 2024 JM, JG, ML, AG, CBRun29 - beam optimisation

11:27

Run 29 started. Beam energy is 1.3435 MeV/u with a cooler voltage of 751.3 V.

We observe significantly greater count rate for top right detector (DSSD#1) now that the beam energy has been lowered. ADC over range? Beam target overlap change?

Plan to check optimisation of beam-target overlap and beam settings again.

12:40

Beam lost. Detectors out. R29_11

13:37

Electron cooler is not working again. We will have un-cooled beam again.

Will start a new run 30 for un-cooled beam on target.

  514   Wed Feb 14 16:52:46 2024 JM, AR, CB, JGRun28 - beam on target

 

Beam energy - 1.361 MeV/u

17:12

Run 28 Started - beam on target

Detectors IN (left arm pot = 36.5 mm, right arm pot = 70.5 mm)

Beam position (vertical offset -9mm, horizontal offset -7.5mm)

Ions circulate in the ring for 3 seconds before they are dumped. Total time between injections is 12 seconds.

Target density apporx 4.2e11 atoms/cm2, pressures in the ring are below 1e-10 mbar. (Attachment 1+2)

Leakage current plot (attachment 3) - change in leakage current is consistent with temperature variation from ML online plot.

18:18

FEE temperatures OK - aida09 is slightly hot - 65.06 C so slightly above vertex temp value of 65C and stays this way over several reloads (attachment 4)

System wide checks OK

Statistics (attachments 5+6) - aida09 and aida12 slightly high stats but still <1K. Check load does not resolve.

Tape server OK - writing ~730 kB/sec

Merger OK

Monitoring code XY plots and Ey energy histograms (attachments 7+8)

18:53 All checks OK

21:58

All system checks ok. Temperatures and leakage current ok. Merger ok.

Attatchment 9: Rates and stats - all below 1k

Attatchment 10: Temperatures

Attatchment 11-14: Waveforms layouts 1,2,7,8 respectively look okay

Attatchment 15: Leakge current

Attatchment 16: lowenergyXY plots

Vacuum pressure is ok.

23:01

system wide checks ok. Temperatures ok. Leakage current ok. Merger ok all went green. Waveforms same as 10pm check.

Statistics - aida09 and aida12 went over 1k (attach 17), but reloaded and settled down again seconds appart (attatch 18). Reloaded again at 23:14 and stats were aove 1k again (attatch 19).

Attatchment 20 is the leakage current, still ok. 

Vacuum pressure ok (attatchment 21) - rises after small bump down from the regular value.

01:00

System wide checks ok. Temperatures ok. Leakage current ok (see attatchment 22). Merger ok all went green. Waveforms the same as 10pm check.

Statistics - aida09 and aida12 occasionally just over 1k when reloading a few times, but goes below mostly.

Vacuum pressure ok, stablised after small bump around 10ish (attatchment 23)

03:00

Ring vacuum suffered from long-term creep, especially in YR08. Tweaked nozzle position. Pressure improves, target density slightly affected. Attachment 24-25

Leakage current and temperature indicate we have an increase in temperature in the ring. This is confirmed by vacplot. Attachment 26-27

Merger continues OK 28

Rates and Stats show we have no beam. Going to the lab 29-30.

 

R28_22 NO BEAM

 

03:45

No beam in the ring, we are missing all the spills. Called HKR.

Detectors out in case beam manipulation is required.

Stat plot looked OK. 31

04:35

DAQ continues OK on run R28_24, still no beam

Temperatures slightly lower - OK

System wide checks - OK

Stats - OK

Merger - OK

05:44

Beam is back. It had been away since 1:30 am due to a fault in the new EC power supply. Fault reason unclear.

Detectors back in (left 36.54, right 70.49 mm)

 

R28_26 BEAM ON

(Beam likely off from R28_ 21-25 inclusive)

 

Temps as before

Stats now OK - att 32

Merger OK - att 33, 34

Stat spectra OK - att 35. Note slope in ASICS (Rutherford?)

Leakage current trending up - likely temperature related.

Ring pressure OK

Gas target 4.14E11

Beam at injection around 7.5E6. Could be higher. Asked for optimisation later today assuming current configuration holds.

06:59

Temperatures OK

System checks OK

Stats looking good in all FEEs except aida01 and aida15

Leakage current has levelled off - looks OK

Ring pressure looks OK and is trending down - target density 4.1e11

Merger and tape server OK

08:30

System checks all OK

Temperatures OK

Merger / tape server OK - data being written R28_29

Target density 4.1e11, ring pressure stable

09:00

Attachment 38 Spectrum browser stats for all FEEs (log scale)

Attachment 39-42 Spectrum browser *L histograms for select channels for each detector. MIDAS data is consistent with inline monitoring plots.

System checks OK

Beam still on - target density and ring pressure remain stable

11:28 We have access to the voltage dividers for the electron cooler. The voltage is -763.625 V. This corresponds to an electron energy of 746.795 eV and a beam energy of 1.361 MeV/u  using ML ion calc.

This is above the resonance energy of 1.3435 MeV/u. We will now end the run and set the cooler voltage to this ion energy.

Run ended. Run 29 started while setting beam energy.

 

Attachment 1: 20240214_180331.jpg
20240214_180331.jpg
Attachment 2: 20240214_180351.jpg
20240214_180351.jpg
Attachment 3: Screenshot_from_2024-02-14_17-26-37.png
Screenshot_from_2024-02-14_17-26-37.png
Attachment 4: Screenshot_from_2024-02-14_18-11-58.png
Screenshot_from_2024-02-14_18-11-58.png
Attachment 5: Screenshot_from_2024-02-14_18-14-06.png
Screenshot_from_2024-02-14_18-14-06.png
Attachment 6: Screenshot_from_2024-02-14_18-14-15.png
Screenshot_from_2024-02-14_18-14-15.png
Attachment 7: Screenshot_from_2024-02-14_18-17-32.png
Screenshot_from_2024-02-14_18-17-32.png
Attachment 8: Screenshot_from_2024-02-14_18-17-46.png
Screenshot_from_2024-02-14_18-17-46.png
Attachment 9: Screenshot_2024-02-14_at_21.51.31.png
Screenshot_2024-02-14_at_21.51.31.png
Attachment 10: Screenshot_2024-02-14_at_21.52.19.png
Screenshot_2024-02-14_at_21.52.19.png
Attachment 11: Screenshot_2024-02-14_at_21.53.59.png
Screenshot_2024-02-14_at_21.53.59.png
Attachment 12: Screenshot_2024-02-14_at_21.54.41.png
Screenshot_2024-02-14_at_21.54.41.png
Attachment 13: Screenshot_2024-02-14_at_21.55.47.png
Screenshot_2024-02-14_at_21.55.47.png
Attachment 14: Screenshot_2024-02-14_at_21.56.24.png
Screenshot_2024-02-14_at_21.56.24.png
Attachment 15: Screenshot_2024-02-14_at_21.57.05.png
Screenshot_2024-02-14_at_21.57.05.png
Attachment 16: Screenshot_2024-02-14_at_21.58.02.png
Screenshot_2024-02-14_at_21.58.02.png
Attachment 17: Screenshot_2024-02-14_at_23.02.54.png
Screenshot_2024-02-14_at_23.02.54.png
Attachment 18: Screenshot_2024-02-14_at_23.03.31.png
Screenshot_2024-02-14_at_23.03.31.png
Attachment 19: Screenshot_2024-02-14_at_23.14.11.png
Screenshot_2024-02-14_at_23.14.11.png
Attachment 20: Screenshot_2024-02-14_at_23.08.34.png
Screenshot_2024-02-14_at_23.08.34.png
Attachment 21: Screenshot_2024-02-14_at_23.17.27.png
Screenshot_2024-02-14_at_23.17.27.png
Attachment 22: Screenshot_2024-02-15_at_01.08.36.png
Screenshot_2024-02-15_at_01.08.36.png
Attachment 23: Screenshot_2024-02-15_at_01.19.22.png
Screenshot_2024-02-15_at_01.19.22.png
Attachment 24: Gasjet.png
Gasjet.png
Attachment 25: Pressure.png
Pressure.png
Attachment 26: Screenshot_from_2024-02-15_03-07-02.png
Screenshot_from_2024-02-15_03-07-02.png
Attachment 27: Screenshot_from_2024-02-15_03-07-44.png
Screenshot_from_2024-02-15_03-07-44.png
Attachment 28: Screenshot_from_2024-02-15_03-20-37.png
Screenshot_from_2024-02-15_03-20-37.png
Attachment 29: Screenshot_from_2024-02-15_03-14-58.png
Screenshot_from_2024-02-15_03-14-58.png
Attachment 30: Screenshot_from_2024-02-15_03-20-18.png
Screenshot_from_2024-02-15_03-20-18.png
Attachment 31: Screenshot_from_2024-02-15_03-59-38.png
Screenshot_from_2024-02-15_03-59-38.png
Attachment 32: Screenshot_from_2024-02-15_05-47-18.png
Screenshot_from_2024-02-15_05-47-18.png
Attachment 33: Screenshot_from_2024-02-15_05-47-31.png
Screenshot_from_2024-02-15_05-47-31.png
Attachment 34: Screenshot_from_2024-02-15_05-47-45.png
Screenshot_from_2024-02-15_05-47-45.png
Attachment 35: Screenshot_from_2024-02-15_05-48-27.png
Screenshot_from_2024-02-15_05-48-27.png
Attachment 36: Screenshot_from_2024-02-15_05-48-41.png
Screenshot_from_2024-02-15_05-48-41.png
Attachment 37: Screenshot_from_2024-02-15_07-01-48.png
Screenshot_from_2024-02-15_07-01-48.png
Attachment 38: Screenshot_from_2024-02-15_08-48-58.png
Screenshot_from_2024-02-15_08-48-58.png
Attachment 39: Screenshot_from_2024-02-15_08-52-55.png
Screenshot_from_2024-02-15_08-52-55.png
Attachment 40: Screenshot_from_2024-02-15_08-54-09.png
Screenshot_from_2024-02-15_08-54-09.png
Attachment 41: Screenshot_from_2024-02-15_08-56-27.png
Screenshot_from_2024-02-15_08-56-27.png
Attachment 42: Screenshot_from_2024-02-15_08-58-01.png
Screenshot_from_2024-02-15_08-58-01.png
  513   Wed Feb 14 14:59:13 2024 JM, CB, NP, ML, AG, ABRun 27 -Beam on target optimisation

Run 27 Started to check optimsation of the beam - target overlap.

Beam intensity at injection is roughly 1e7 particles

Target density apporx 4.2e11 atoms/cm2 and stable.

Ring pressure is below 1e-10 mbar for all ring sections and stable

First optimisation procedure was to vary the horizontal beam position and see the effect on the count rate in detector #0 (top left) (see elog 509). Beam position was scanned from -9 and -4. Attachment 1 shows the maximum in count rate around the -7.5 position. 

Second optimisation procedure was to change the time ions are allowed to circulate in the ring before being dumped. The count rate over 5 minutes in detector #0 (top left) at ~20 MeV was used as a gauge. This is the only part of the cycle we can control.

Time (s)    Counts

2               148

3               200

4               178

Attachment 1: Target-overlap-scan-optimized.pdf
Target-overlap-scan-optimized.pdf Target-overlap-scan-optimized.pdf Target-overlap-scan-optimized.pdf
  512   Wed Feb 14 14:06:28 2024 CB AR JMPulser Walkthrough
15:06
Run25

Pulser walkthrough

EGG Ortec 448 Pulser

Amplitide 90,000-10,000 @ 10,000 step
Normalise 5.0
Attenuators x5 IN
t_r 100ns
Tau_d 50us
Frequency Int Osc 50Hz
Polarity + ( - polarity via Cooknell SA1 Sum & Invert Amp )
  511   Wed Feb 14 08:44:32 2024 CB AR JMMotors test and turbo ON

Run number changed 09:44 - Now run 23

Moving motors with the injection

Ring signal very weird, moving in for short time and staying out for 15+ seconds.

 

R24

Right arm FULLY in (70.49 mm)

Left arm SERVOS ONLY in (36.54)

No appreciable effect on beam (not clipping beam at injection / during circulation)

 

ML fixed signal cycle. It's now 8 seconds in, 5 seconds out. Motors still not moving. Acquiring more stats in the current conditions.

 

11:07 - Temps & System wide checks good

Stats as before.

Leakage currents & pressure good.

11:44

Leakage current ok. pressure good. Temperatures ok. Merger ok. System wide checks ok.

In with the left 36.5

In with the right 70.5

Attatchment 1: Rates

Attatchment 2: Temperatures

Attatchment 3: Waveforms layout 7

Attatchment 4: Waveforms layout 8

Attatchment 5: Waveforms layout 1

Attatchment 6: Waveforms layout 2

Attatchment 7: Leakage current

Attatchment 8: lowenergyXYtotal all 4 dets

Attatchment 9: lowenergyExEy all dets

Attatchment 10: loveenergyvExEy

Attachment 11: screenshot from the Shottky

 

14:00 Run 24 stopped to allow access to the cave.

Helium and liquid nitrogen re-filled in the target.

Pulser walkthrough performed R25.

Turned CARME turbo ON, pressure in the ring with target on is reduced by a factor 2.

Tried to find electron cooler voltages to determine the beam energy. We were unable to do this. Hopefully we find out tomorrow. 

 

Attachment 1: Screenshot_from_2024-02-14_11-31-38.png
Screenshot_from_2024-02-14_11-31-38.png
Attachment 2: Screenshot_from_2024-02-14_11-31-52.png
Screenshot_from_2024-02-14_11-31-52.png
Attachment 3: Screenshot_from_2024-02-14_11-32-34.png
Screenshot_from_2024-02-14_11-32-34.png
Attachment 4: Screenshot_from_2024-02-14_11-33-52.png
Screenshot_from_2024-02-14_11-33-52.png
Attachment 5: Screenshot_from_2024-02-14_11-34-51.png
Screenshot_from_2024-02-14_11-34-51.png
Attachment 6: Screenshot_from_2024-02-14_11-35-17.png
Screenshot_from_2024-02-14_11-35-17.png
Attachment 7: Screenshot_from_2024-02-14_11-38-43.png
Screenshot_from_2024-02-14_11-38-43.png
Attachment 8: Screenshot_from_2024-02-14_11-44-19.png
Screenshot_from_2024-02-14_11-44-19.png
Attachment 9: Screenshot_from_2024-02-14_11-44-26.png
Screenshot_from_2024-02-14_11-44-26.png
Attachment 10: Screenshot_from_2024-02-14_11-44-32.png
Screenshot_from_2024-02-14_11-44-32.png
Attachment 11: screenshot-2024-02-14_11-31-236868864019057183143.png
screenshot-2024-02-14_11-31-236868864019057183143.png
  510   Tue Feb 13 17:19:43 2024 JM, AR, NP, AB, RSSRun22 - cooled beam on target (definitely maybe)

18:05 We believe we have an overlap between the beam and the target. Run 22 started.

          CARME detectors in with the servo ONLY. Left arm pot 36.5 mm, right arm pot 37.7 mm

          Target density apporx 2e11 atoms/cm2.

          Shots into the rin partially stable (sometimes consecutive injections at high intensity, sometime apporx 1 in 4 fail rate).

          We plan beam run overnight, remote shifters will check on the progress every 2 hours.

18:32

Everything ok. Temperatures and leakage current are okay.

Attachment 1: Rates

Attachment 2: Temperatures

Attachment 3: Waveforms layout 7

Attachment 4: Waveforms layout 8

Attachment 5: Waveforms layout 1

Attachment 6: Waveforms layout 2

Attachment 7: Leakage current

Attachment 8: xy plot for all four detectors.

 

21:00 Checks made, everything okay!

23:00 Checks made, everything okay!

Attachment 9: Rates

Attachment 10: Temperatures

Attachment 11: Waveforms layout 1

Attachment 12: Waveforms layout 2

Attachment 13: Waveforms layout 7

Attachment 14: Waveforms layout 8

Attachment 15: Leakage current

 

14.02.2023 01:00

Checks made, everything okay!

Attachment 16: Waveforms layout 1

Attachment 17: Waveforms layout 2

Attachment 18: Waveforms layout 7

Attachment 19: Waveforms layout 8

Attachment 20: Rates

Attachment 21: Temperatures

Attachment 22: Leakage current

02:27

FEE temperatures OK

System checks OK

ADC data items appears ok for all FEEs, except aoda09 shows rates of ~1kHz. Spectrum borwser shows this is from a single strip. Checl load aida09 several times but no change to the rate.

Leakage current appears fine. (Attachment 23)

Merger and tape server OK

Ring pressure appeared to have dropped and the target density had dropped to approx 5e10 at/cm2 range. Target nozzle moved 200 steps. Density back to 2.5e11 range and pressure is high -11 low -10 mbar range. <----- Operator mistake

XY plot all detector (attachment 24)

EyTotal all detectors (attachment 25)

 

05:08

Target density appears to have been very stable over the last few hours ~2e11 at/cm2. Ring pressure also appears stable.

FEE temperatures OK

System checks OK

ADC data items appears ok for all FEEs,Aida09 rate appears to be reduced again.

Leakage current appears fine.

Spectrum browser rates (attachment 26)

Spectrum browser stats (attachment 27)

Spectrum browser 1.8*L pn (attachment 28)

Spectrum browser 1.8*L nn (attachment 29)

New merger stats (attachment 30)

I am not confident we are seeing significant Rutherford scattering accumalated so far.

 

07:18

FEE temperatures OK

System checks OK

Stats OK

Leakage current appears fine.

Spectrum browser rates OK

Spectrum browser stats OK

Spectrum browser 1.8*L OK

Spectrum browser 1.8*L nn OK

New merger stats OK

 

7:40

Noticed operator mistake at 3 am. Ring pressure was very bad in YR08. Unlikely we had much beam since then. Moved nozzle to improve pressure.

See attach 38 & 39.

08:51

Checks all ok. Temperatures and leakage current ok. Merger ok. 

Attatchment 40: Rates

Attatchment 41: Temperatures

Attatchment 42: Waveforms layout 7

Attatchment 43: Waveforms layout 8

Attahcment 44: Waveforms layout 1

Attatchment 45: Waveforms layout 2

Attatchment 46: Leakage current

Attatchment 47: Gasjet board - not much change from last check

Attatchment 48: Vacuum pressures 

 

Attachment 1: Screenshot_from_2024-02-13_18-33-30.png
Screenshot_from_2024-02-13_18-33-30.png
Attachment 2: Screenshot_from_2024-02-13_18-33-52.png
Screenshot_from_2024-02-13_18-33-52.png
Attachment 3: Screenshot_from_2024-02-13_18-34-33.png
Screenshot_from_2024-02-13_18-34-33.png
Attachment 4: Screenshot_from_2024-02-13_18-34-53.png
Screenshot_from_2024-02-13_18-34-53.png
Attachment 5: Screenshot_from_2024-02-13_18-35-12.png
Screenshot_from_2024-02-13_18-35-12.png
Attachment 6: Screenshot_from_2024-02-13_18-37-04.png
Screenshot_from_2024-02-13_18-37-04.png
Attachment 7: Screenshot_from_2024-02-13_18-37-52.png
Screenshot_from_2024-02-13_18-37-52.png
Attachment 8: Screenshot_from_2024-02-13_18-48-19.png
Screenshot_from_2024-02-13_18-48-19.png
Attachment 9: Screenshot_from_2024-02-13_22-01-02.png
Screenshot_from_2024-02-13_22-01-02.png
Attachment 10: Screenshot_from_2024-02-13_22-01-31.png
Screenshot_from_2024-02-13_22-01-31.png
Attachment 11: Screenshot_from_2024-02-13_22-03-33.png
Screenshot_from_2024-02-13_22-03-33.png
Attachment 12: Screenshot_from_2024-02-13_22-04-11.png
Screenshot_from_2024-02-13_22-04-11.png
Attachment 13: Screenshot_from_2024-02-13_22-05-01.png
Screenshot_from_2024-02-13_22-05-01.png
Attachment 14: Screenshot_from_2024-02-13_22-05-40.png
Screenshot_from_2024-02-13_22-05-40.png
Attachment 15: Screenshot_from_2024-02-13_22-06-45.png
Screenshot_from_2024-02-13_22-06-45.png
Attachment 16: Screenshot_from_2024-02-14_00-04-10.png
Screenshot_from_2024-02-14_00-04-10.png
Attachment 17: Screenshot_from_2024-02-14_00-04-46.png
Screenshot_from_2024-02-14_00-04-46.png
Attachment 18: Screenshot_from_2024-02-14_00-05-20.png
Screenshot_from_2024-02-14_00-05-20.png
Attachment 19: Screenshot_from_2024-02-14_00-06-03.png
Screenshot_from_2024-02-14_00-06-03.png
Attachment 20: Screenshot_from_2024-02-14_00-06-23.png
Screenshot_from_2024-02-14_00-06-23.png
Attachment 21: Screenshot_from_2024-02-14_00-06-39.png
Screenshot_from_2024-02-14_00-06-39.png
Attachment 22: Screenshot_from_2024-02-14_00-07-30.png
Screenshot_from_2024-02-14_00-07-30.png
Attachment 23: Screenshot_from_2024-02-14_02-30-22.png
Screenshot_from_2024-02-14_02-30-22.png
Attachment 24: Screenshot_from_2024-02-14_02-44-18.png
Screenshot_from_2024-02-14_02-44-18.png
Attachment 25: Screenshot_from_2024-02-14_02-46-22.png
Screenshot_from_2024-02-14_02-46-22.png
Attachment 26: 20240214_023337.jpg
20240214_023337.jpg
Attachment 27: 20240214_023311.jpg
20240214_023311.jpg
Attachment 28: Screenshot_from_2024-02-14_04-58-12.png
Screenshot_from_2024-02-14_04-58-12.png
Attachment 29: Screenshot_from_2024-02-14_04-59-33.png
Screenshot_from_2024-02-14_04-59-33.png
Attachment 30: Screenshot_from_2024-02-14_05-00-33.png
Screenshot_from_2024-02-14_05-00-33.png
Attachment 31: Screenshot_from_2024-02-14_05-01-23.png
Screenshot_from_2024-02-14_05-01-23.png
Attachment 32: Screenshot_from_2024-02-14_05-03-25.png
Screenshot_from_2024-02-14_05-03-25.png
Attachment 33: Screenshot_from_2024-02-14_07-13-36.png
Screenshot_from_2024-02-14_07-13-36.png
Attachment 34: Screenshot_from_2024-02-14_07-14-32.png
Screenshot_from_2024-02-14_07-14-32.png
Attachment 35: Screenshot_from_2024-02-14_07-15-29.png
Screenshot_from_2024-02-14_07-15-29.png
Attachment 36: Screenshot_from_2024-02-14_07-16-11.png
Screenshot_from_2024-02-14_07-16-11.png
Attachment 37: Screenshot_from_2024-02-14_07-16-30.png
Screenshot_from_2024-02-14_07-16-30.png
Attachment 38: Pressures.png
Pressures.png
Attachment 39: GasTarget.png
GasTarget.png
Attachment 40: Screenshot_from_2024-02-14_08-52-02.png
Screenshot_from_2024-02-14_08-52-02.png
Attachment 41: Screenshot_from_2024-02-14_08-52-15.png
Screenshot_from_2024-02-14_08-52-15.png
Attachment 42: Screenshot_from_2024-02-14_08-54-01.png
Screenshot_from_2024-02-14_08-54-01.png
Attachment 43: Screenshot_from_2024-02-14_08-54-22.png
Screenshot_from_2024-02-14_08-54-22.png
Attachment 44: Screenshot_from_2024-02-14_08-55-45.png
Screenshot_from_2024-02-14_08-55-45.png
Attachment 45: Screenshot_from_2024-02-14_08-56-15.png
Screenshot_from_2024-02-14_08-56-15.png
Attachment 46: Screenshot_from_2024-02-14_08-57-29.png
Screenshot_from_2024-02-14_08-57-29.png
Attachment 47: IMG_7850.HEIC
Attachment 48: IMG_7851.HEIC
  509   Tue Feb 13 08:34:44 2024 Hw, JM, AR, CB, NP, AB, RSSStatus of DAQ - beam overlap

Check the status of the DAQ system, with Tom. All correct. See attachments:

1 Leackage currents

2 ADC data item stats

3 White rabbit time stamps

Note: At 9:36 we thing they are working on cooling the beam, there is no target and the detectors are fully out.

9:16 cooled beam, carlo is starting new run r21

Target dencity = 1.47E11 atom/cm2 (10:30 am)

Horizon Y09 from 9 mm to 8 mm, checking different beam position (10:55 am)

Right arm is fully in (11:02 am)

By moving in the detector we are reducing the intencity (11:04 am)

Right numatic out, right servo in, potentiiometer at 38.38 mm (11:05 am)

Losing every second shot, tuning the numatic to get better parameters, beam tuning  (11:13 am)

Leaving the detectors out while going for lunch, still beam (11:42 am)

 

 

12.23 Both detectors 37mm in ( DSSSD wafer edge c. 5cm from beam axis )

System Wide Checks Update - Run 21

Temperatures ok, leakage current ok

AIDA16 said that there was an FPGA timestamp error, collected again and now all passed - timestamps on merger statistics page all good when checked

Attatchment 4: Temperatures

Attatchment 5: Waveforms layout 7

Attatchment 6: Waveforms layout 8

Attatchment 7: Wavforms layout 2

Attatchment 8: Waveforms layout 1

Attatchment 9: Leakage current ok

16:11 Both detectors 37mm in ( DSSSD wafer edge c. 5cm from beam axis ), target off, beam being tuned, what ever, still run 21

System Wide Checks Update - Run 21

Temperatures ok, leakage current ok

 

18:00 Run 21 contd. We have spent the afternoon trying to confirm whether we observe an overlap between the target and the beam. We varied the horizontal beam position and used the top left detector around 19-20 MeV region as a gauge for the count rate.

          This should be reasonably safe as the beam halo in this detector is primarily at low energies. We took the count rate over 10 shots into the ring as our gauge. Attached is a plot of the results of varying the beam position. Target density and gas pressure was apporximately constant throughout.

          We estimate -7 mm is the correct horiziontal beam position. We also had a call off Zoran while at this setting as he observed a decrease in the beam lifetime which was observed in HKR. See attached plot of the scan (attachment 10)

 

 

Attachment 1: Screenshot_from_2024-02-13_09-21-05.png
Screenshot_from_2024-02-13_09-21-05.png
Attachment 2: Screenshot_from_2024-02-13_09-33-31.png
Screenshot_from_2024-02-13_09-33-31.png
Attachment 3: Screenshot_from_2024-02-13_09-33-47.png
Screenshot_from_2024-02-13_09-33-47.png
Attachment 4: Screenshot_from_2024-02-13_13-04-54.png
Screenshot_from_2024-02-13_13-04-54.png
Attachment 5: Screenshot_from_2024-02-13_13-05-40.png
Screenshot_from_2024-02-13_13-05-40.png
Attachment 6: Screenshot_from_2024-02-13_13-06-04.png
Screenshot_from_2024-02-13_13-06-04.png
Attachment 7: Screenshot_from_2024-02-13_13-06-33.png
Screenshot_from_2024-02-13_13-06-33.png
Attachment 8: Screenshot_from_2024-02-13_13-06-53.png
Screenshot_from_2024-02-13_13-06-53.png
Attachment 9: Screenshot_from_2024-02-13_13-11-18.png
Screenshot_from_2024-02-13_13-11-18.png
Attachment 10: Target-overlap-scan.pdf
Target-overlap-scan.pdf
  508   Mon Feb 12 01:40:48 2024 TD, JM, CB, AR, RSSRun17 - Run20
02.35 DAQ continues file G22-88/R17_30

02.38 Check ASIC load
      aida01 asics # 1 and #4 & aida12 asic #2 HEC events removed



09.05 DAQ continues file G22-88/R17_53
      DSSSD bias & leakage current OK - attachments 10-11

      FEE64 temperatures OK - attachment 12

      ADC data item stats OK - attachment 13
       hot channels aida01 ( 67k ) and aida15 ( 7k )

      WR timestamps OK - attachment 14

      per FEE64 Rate and Stat spectra - attachments 15-16

      Merger, Merger stats and TapeServer - attachments 17-19

09.19 New run G22-88/R18
      all histograms & stats zero'd

      CB starting up 4He target preparatory to returning 16O2+ beam to CRYRING

      L & R arms all out

09.26 4He target c. 2e11c.2
      16O2+ beam c. 3e5/s

09.58 4He target c. 3e11/cm2
      16O2+ beam c. 5e5/s at start of cycle
      See attachments 20-22

10.40 New run G22-88/R19
      all histograms & stats zero'd

      CB will attempt to steer the beam 
      all steering options set to 0mm except YR09 target offset (vertical = -9mm) and (horizontal = 9mm) 


10.52 R arm fully in, L arm fully out
      beam steering hard right

14:24 Frank has informed us that the CRYRING team will be examining the electron cooler for th enext few hours. Run 19 stopped, Run 20 started. 
      Detectors moved fully out, target OFF. 
      All histograms and stats zero'd
      Access to the cave is required, beam OFF

15:45 Channel 1 back to 150V.


17.40 DAQ continues file G22-88/R20_13
      DSSSD bias & leakage current OK - attachments 23-24

      FEE64 temperatures OK - attachment 25

      All system wide checks OK except aida08 FPGA error - attachment 26

      WR timestamps OK - attachment 27

      ADC data item stats OK - attachment 28
       hot channels aida01 ( 80k ) and aida15 ( 7k )

      per FEE64 Rate and Stat spectra - attachments 29-30

      Merger, Merger stats and TapeServer - attachments 31-32

23.12 DAQ continues file G22-88/R20_20
      DSSSD bias & leakage current OK - attachments 33-34

      FEE64 temperatures OK

      All system wide checks OK except aida08 FPGA error - attachment 35

      WR timestamps OK - attachment 36

      ADC data item stats OK - attachment 37
       hot channels aida01 ( 80k ) and aida15 ( 7k )

      per FEE64 Rate and Stat spectra - attachments 38

      Merger, Merger stats and TapeServer - attachments 39-40



03.47 DAQ continues file G22-88/R20_28

      DSSSD bias & leakage current OK - attachments 41-42

      FEE64 temperatures OK - attachment 43

      ADC data item stats - attachment 44
       aida12 rate - HEC events 

      per FEE64 Rate spectra - attachment 45

      Check ASIC load 

      ADC data item stats - attachment 46

      per FEE64 Rate spectra - attachment 47

      Merger, Merger stats and TapeServer - attachments 48-49





      
Attachment 1: Screenshot_from_2024-02-12_02-34-50.png
Screenshot_from_2024-02-12_02-34-50.png
Attachment 2: Screenshot_from_2024-02-12_02-35-12.png
Screenshot_from_2024-02-12_02-35-12.png
Attachment 3: Screenshot_from_2024-02-12_02-35-44.png
Screenshot_from_2024-02-12_02-35-44.png
Attachment 4: Screenshot_from_2024-02-12_02-36-21.png
Screenshot_from_2024-02-12_02-36-21.png
Attachment 5: Screenshot_from_2024-02-12_02-37-20.png
Screenshot_from_2024-02-12_02-37-20.png
Attachment 6: Screenshot_from_2024-02-12_02-38-35.png
Screenshot_from_2024-02-12_02-38-35.png
Attachment 7: Screenshot_from_2024-02-12_02-39-32.png
Screenshot_from_2024-02-12_02-39-32.png
Attachment 8: Screenshot_from_2024-02-12_02-39-52.png
Screenshot_from_2024-02-12_02-39-52.png
Attachment 9: Screenshot_from_2024-02-12_02-40-05.png
Screenshot_from_2024-02-12_02-40-05.png
Attachment 10: Screenshot_from_2024-02-12_09-05-53.png
Screenshot_from_2024-02-12_09-05-53.png
Attachment 11: Screenshot_from_2024-02-12_09-06-03.png
Screenshot_from_2024-02-12_09-06-03.png
Attachment 12: Screenshot_from_2024-02-12_09-06-42.png
Screenshot_from_2024-02-12_09-06-42.png
Attachment 13: Screenshot_from_2024-02-12_09-07-02.png
Screenshot_from_2024-02-12_09-07-02.png
Attachment 14: Screenshot_from_2024-02-12_09-07-31.png
Screenshot_from_2024-02-12_09-07-31.png
Attachment 15: Screenshot_from_2024-02-12_09-07-42.png
Screenshot_from_2024-02-12_09-07-42.png
Attachment 16: Screenshot_from_2024-02-12_09-08-23.png
Screenshot_from_2024-02-12_09-08-23.png
Attachment 17: Screenshot_from_2024-02-12_09-08-48.png
Screenshot_from_2024-02-12_09-08-48.png
Attachment 18: Screenshot_from_2024-02-12_09-08-54.png
Screenshot_from_2024-02-12_09-08-54.png
Attachment 19: Screenshot_from_2024-02-12_09-09-03.png
Screenshot_from_2024-02-12_09-09-03.png
Attachment 20: 20240212_095432.jpg
20240212_095432.jpg
Attachment 21: 20240212_095444.jpg
20240212_095444.jpg
Attachment 22: 20240212_095449.jpg
20240212_095449.jpg
Attachment 23: Screenshot_from_2024-02-12_17-40-16.png
Screenshot_from_2024-02-12_17-40-16.png
Attachment 24: Screenshot_from_2024-02-12_17-40-25.png
Screenshot_from_2024-02-12_17-40-25.png
Attachment 25: Screenshot_from_2024-02-12_17-40-43.png
Screenshot_from_2024-02-12_17-40-43.png
Attachment 26: Screenshot_from_2024-02-12_17-41-01.png
Screenshot_from_2024-02-12_17-41-01.png
Attachment 27: Screenshot_from_2024-02-12_17-41-25.png
Screenshot_from_2024-02-12_17-41-25.png
Attachment 28: Screenshot_from_2024-02-12_17-41-39.png
Screenshot_from_2024-02-12_17-41-39.png
Attachment 29: Screenshot_from_2024-02-12_17-42-14.png
Screenshot_from_2024-02-12_17-42-14.png
Attachment 30: Screenshot_from_2024-02-12_17-42-43.png
Screenshot_from_2024-02-12_17-42-43.png
Attachment 31: Screenshot_from_2024-02-12_17-43-00.png
Screenshot_from_2024-02-12_17-43-00.png
Attachment 32: Screenshot_from_2024-02-12_17-43-12.png
Screenshot_from_2024-02-12_17-43-12.png
Attachment 33: Screenshot_from_2024-02-12_23-12-12.png
Screenshot_from_2024-02-12_23-12-12.png
Attachment 34: Screenshot_from_2024-02-12_23-12-35.png
Screenshot_from_2024-02-12_23-12-35.png
Attachment 35: Screenshot_from_2024-02-12_23-12-58.png
Screenshot_from_2024-02-12_23-12-58.png
Attachment 36: Screenshot_from_2024-02-12_23-13-38.png
Screenshot_from_2024-02-12_23-13-38.png
Attachment 37: Screenshot_from_2024-02-12_23-13-57.png
Screenshot_from_2024-02-12_23-13-57.png
Attachment 38: Screenshot_from_2024-02-12_23-15-02.png
Screenshot_from_2024-02-12_23-15-02.png
Attachment 39: Screenshot_from_2024-02-12_23-15-21.png
Screenshot_from_2024-02-12_23-15-21.png
Attachment 40: Screenshot_from_2024-02-12_23-15-39.png
Screenshot_from_2024-02-12_23-15-39.png
Attachment 41: Screenshot_from_2024-02-13_03-47-40.png
Screenshot_from_2024-02-13_03-47-40.png
Attachment 42: Screenshot_from_2024-02-13_03-48-14.png
Screenshot_from_2024-02-13_03-48-14.png
Attachment 43: Screenshot_from_2024-02-13_03-50-53.png
Screenshot_from_2024-02-13_03-50-53.png
Attachment 44: Screenshot_from_2024-02-13_03-51-24.png
Screenshot_from_2024-02-13_03-51-24.png
Attachment 45: Screenshot_from_2024-02-13_03-51-50.png
Screenshot_from_2024-02-13_03-51-50.png
Attachment 46: Screenshot_from_2024-02-13_03-53-21.png
Screenshot_from_2024-02-13_03-53-21.png
Attachment 47: Screenshot_from_2024-02-13_03-53-41.png
Screenshot_from_2024-02-13_03-53-41.png
Attachment 48: Screenshot_from_2024-02-13_03-54-00.png
Screenshot_from_2024-02-13_03-54-00.png
Attachment 49: Screenshot_from_2024-02-13_03-54-15.png
Screenshot_from_2024-02-13_03-54-15.png
  507   Sun Feb 11 13:28:46 2024 JM, CB, TDRun16 - beam on target

13:37 Run started - no beam in the ring (One of the ring magnets was still on RED when it should have been set to green by HKR)

14:28 Beam ON - still observe beam halo on DSSD0 (Top left)

          YR09 Vertical and horizontal offsets set to -9.0 mm

14:36 Target ON, Beam ON

         We can see the beam halo on the detectors again in the same position as previous tests. The energy histograms for the detectors show only low energy events, except bottom left detector where we see a peak around 20 MeV. With the target on we see additional events but we only see events at ~20MeV for the    bottom detector still in the bottom left detector. We are moving the detectors IN to try and see events from scattering of the beam on the target. We may not be hitting the target with the beam.

 

15.29  We observe data in DSSD # 0, 2, 3 (top left, bottom left, bottom right) which is consistent with a line beam halo on the detector. The top right detector is positioned higher by ~10 mm than the top left detector so it is not unphysical to not see any events in this detector from the halo.

           The energy of events in each detector is not consistent however. In DSSD #0 and 3 we see mainly low energy events (some small number of 20 MeV events is seen) but in DSSD #2 we see mainly higher energy (20 MeV) events. See attached energy histograms, xy and Ex vs x plots (attachments 1-3).

          This is consistent with the data from channels seen in MIDAS for aida04 (DSSD #0) and aida12 (DSSD#2).

          Switched FEE64 range bit frpom 20MeV FSR to 1GeV FSR to see if any events are over the 20 MeV range of the low energy channel . Slow comparator from 0x64 to 0x5. We see no events just above 20 MeV  in aida04 but do in aida12. See attachments 4 & 5

          No data ( > 20MeV nominal ) out of range in DSSSD #1

         FEE range set back to FSR to 20 MeV and slow comparator set back to 0x64

15:57  Beam on target. Left arm fully out (pot 18.8 mm), right arm fully in (pot 83.69 mm). Beam injection ~1e6, target density 2e11. Cycle time between 4-13 s.

        Leaving for more stats

 

 

17.30 Routine system checks

         DAQ continues file G22-88/R16_14

        DSSSD bias and leakage current OK - attachments 8 & 28

         DSSSD #2 bias -100V, all other DSSSDs -150V

         ambient temperature +17.0 deg C

 

      FEE64 temperatures OK - attachment 9

      All system wide checks OK

      WR timestamp OK - attachment 10

      aida01 & aida03 ASIC settings - attachments 11-12

      AIDA ADC, PAUSE, DISC, correlation - attachments 13-16

        ADC data item rates OK

     per FEE64 Rate spectra - attachments 17-18

      note change in rates for aida12 & aida13 - presumably due to beam cycle?

    per FEE64 1.8.L, 1.8.H and 1.8.W 20us FSR spectra - attachments 19-24

    Merger, Merger stats and TapeServer - attachments 25-27

18.10 ML CRYRING vacuum - attachment 29

FEE spectrum browser statistics attachment 30 - not many missing channels - aida04 asic 2 perhaps?

18:40 End of run. Target x position moved out by 3000. Faraday cup to the ring closed. Detectors moved fully OUT.

Run 17 started - alpha background

 

 

 

Attachment 1: Screenshot_from_2024-02-11_15-32-52.png
Screenshot_from_2024-02-11_15-32-52.png
Attachment 2: Screenshot_from_2024-02-11_15-33-06.png
Screenshot_from_2024-02-11_15-33-06.png
Attachment 3: Screenshot_from_2024-02-11_15-33-35.png
Screenshot_from_2024-02-11_15-33-35.png
Attachment 4: Screenshot_from_2024-02-11_15-24-58.png
Screenshot_from_2024-02-11_15-24-58.png
Attachment 5: Screenshot_from_2024-02-11_15-27-54.png
Screenshot_from_2024-02-11_15-27-54.png
Attachment 6: 20230524_103830.jpg
20230524_103830.jpg
Attachment 7: Screenshot_from_2024-02-11_15-57-48.png
Screenshot_from_2024-02-11_15-57-48.png
Attachment 8: Screenshot_from_2024-02-11_17-29-16.png
Screenshot_from_2024-02-11_17-29-16.png
Attachment 9: Screenshot_from_2024-02-11_17-29-35.png
Screenshot_from_2024-02-11_17-29-35.png
Attachment 10: Screenshot_from_2024-02-11_17-30-08.png
Screenshot_from_2024-02-11_17-30-08.png
Attachment 11: Screenshot_from_2024-02-11_17-30-18.png
Screenshot_from_2024-02-11_17-30-18.png
Attachment 12: Screenshot_from_2024-02-11_17-30-24.png
Screenshot_from_2024-02-11_17-30-24.png
Attachment 13: Screenshot_from_2024-02-11_17-30-54.png
Screenshot_from_2024-02-11_17-30-54.png
Attachment 14: Screenshot_from_2024-02-11_17-36-57.png
Screenshot_from_2024-02-11_17-36-57.png
Attachment 15: Screenshot_from_2024-02-11_17-37-25.png
Screenshot_from_2024-02-11_17-37-25.png
Attachment 16: Screenshot_from_2024-02-11_17-37-52.png
Screenshot_from_2024-02-11_17-37-52.png
Attachment 17: Screenshot_from_2024-02-11_17-31-54.png
Screenshot_from_2024-02-11_17-31-54.png
Attachment 18: Screenshot_from_2024-02-11_17-32-01.png
Screenshot_from_2024-02-11_17-32-01.png
Attachment 19: Screenshot_from_2024-02-11_17-33-15.png
Screenshot_from_2024-02-11_17-33-15.png
Attachment 20: Screenshot_from_2024-02-11_17-33-51.png
Screenshot_from_2024-02-11_17-33-51.png
Attachment 21: Screenshot_from_2024-02-11_17-34-25.png
Screenshot_from_2024-02-11_17-34-25.png
Attachment 22: Screenshot_from_2024-02-11_17-35-03.png
Screenshot_from_2024-02-11_17-35-03.png
Attachment 23: Screenshot_from_2024-02-11_17-35-56.png
Screenshot_from_2024-02-11_17-35-56.png
Attachment 24: Screenshot_from_2024-02-11_17-36-35.png
Screenshot_from_2024-02-11_17-36-35.png
Attachment 25: Screenshot_from_2024-02-11_17-38-32.png
Screenshot_from_2024-02-11_17-38-32.png
Attachment 26: Screenshot_from_2024-02-11_17-38-49.png
Screenshot_from_2024-02-11_17-38-49.png
Attachment 27: Screenshot_from_2024-02-11_17-38-58.png
Screenshot_from_2024-02-11_17-38-58.png
Attachment 28: Screenshot_from_2024-02-11_17-41-37.png
Screenshot_from_2024-02-11_17-41-37.png
Attachment 29: Capture.PNG
Capture.PNG
Attachment 30: Screenshot_from_2024-02-11_18-27-56.png
Screenshot_from_2024-02-11_18-27-56.png
  506   Sun Feb 11 07:03:34 2024 TDSunday 11 February
02.02 DAQ continues G22-86/R15_21
      alpha background

      DSSSD leakage current = attachment 1
       DSSSD #2 transient c. few hundred nA continue

      FEE64 temperatures OK - attachment 2

      ADC data item stats - attachment 3
       rates dominated by 'hot' channels aida01 ( 70k ) and aida15 ( 11k ) - all other FEE64s c. 0Hz

      per FEE64 Rate and Stat spectra - attachments 4-5

02.07 All stats zero'd 


05.52 DAQ continues G22-86/R15_28
      alpha background

      FEE64 temperatures OK - attachment 6

      ADC data item stats - attachment 7
       rates dominated by 'hot' channels aida01 ( 72k ), aida15 ( 11k ), aida12 ( 3k ) and aida05 ( 600 )
       all other FEE64s c. 0Hz

      per FEE64 Stat spectra - attachment 8
       aida12 asic #2 HEC channels high rate

      Merger and TapeServer - attachments 9-10

      DSSSD leakage current = attachment 11
       DSSSD #2 transient c. few hundred nA continue

      online per DSSSD 2D x-y spectra - attachnent 12

05.56 All histograms & stats zero'd
      2x Check ASIC load

      ADC data item stats - attachment 7
       rates dominated by 'hot' channels aida01 ( 72k ), aida15 ( 11k )
       all other FEE64s c. 0Hz

10:28 DAQ continues G22-88/R15

      FEE temperatures OK

      ADC data items stats dominated by aida01 (86k) and aida15 (9k) rest of stats approx OHz. 
      Rates dominated by 2 hot channels per FEE 
      Leakage current plot (attachment 15)
      System wide checks OK
      Merger OK (500,000 items per sec), Tape server OK (710 kb/sec being written)
      DSSD XY plot (attachment 16) - some noise coming through in certain strips on DSSD 3
      DSSD Ey energy histograms (attachment 17)
      DSSD Ey vs Ex plots (attachment 18)     -  statistics for DSSD0,1,3 are consistent (DSSD 2 stats appear higher - unknown reason as yet)

11:10 Leakage current for DSSD1 (I2 on leakage current plot) remained 'turbulent'. We dropped the voltage from 150V to 100V. Leakage current dropped from ~4.5uA to ~4.2 uA. We will see if it remains more stable. 
      Rates in DSSD1 increased during change in voltage so all histograms were zero'd and monitoring code reset. Transient also observed for I3 in the leake current plot (attachment 19)

12:10 No progress on the target so far ...
      Noticed in aida03 and aida04 several missing channels in the stats spectrum browser (attachment 20)
      ASIC check load several times on aida03 and aida04. All stats and histograms zero'd 

13.37 Target now available - will attempt beam + target
      New data file G22-88/R16

13.39 per FEE64 Stat spectra

13.43 All histograms and stats zero'd

      multiple channels missing aida03 and aida04 ( and possibly other FEE64s too )
Attachment 1: Screenshot_from_2024-02-11_02-02-18.png
Screenshot_from_2024-02-11_02-02-18.png
Attachment 2: Screenshot_from_2024-02-11_02-03-32.png
Screenshot_from_2024-02-11_02-03-32.png
Attachment 3: Screenshot_from_2024-02-11_02-04-09.png
Screenshot_from_2024-02-11_02-04-09.png
Attachment 4: Screenshot_from_2024-02-11_02-04-43.png
Screenshot_from_2024-02-11_02-04-43.png
Attachment 5: Screenshot_from_2024-02-11_02-07-08.png
Screenshot_from_2024-02-11_02-07-08.png
Attachment 6: Screenshot_from_2024-02-11_05-52-02.png
Screenshot_from_2024-02-11_05-52-02.png
Attachment 7: Screenshot_from_2024-02-11_05-52-21.png
Screenshot_from_2024-02-11_05-52-21.png
Attachment 8: Screenshot_from_2024-02-11_05-52-59.png
Screenshot_from_2024-02-11_05-52-59.png
Attachment 9: Screenshot_from_2024-02-11_05-53-41.png
Screenshot_from_2024-02-11_05-53-41.png
Attachment 10: Screenshot_from_2024-02-11_05-54-04.png
Screenshot_from_2024-02-11_05-54-04.png
Attachment 11: Screenshot_from_2024-02-11_05-54-29.png
Screenshot_from_2024-02-11_05-54-29.png
Attachment 12: Screenshot_from_2024-02-11_05-54-53.png
Screenshot_from_2024-02-11_05-54-53.png
Attachment 13: Screenshot_from_2024-02-11_05-57-44.png
Screenshot_from_2024-02-11_05-57-44.png
Attachment 14: Screenshot_from_2024-02-11_10-33-04.png
Screenshot_from_2024-02-11_10-33-04.png
Attachment 15: Screenshot_from_2024-02-11_10-59-13.png
Screenshot_from_2024-02-11_10-59-13.png
Attachment 16: Screenshot_from_2024-02-11_11-01-26.png
Screenshot_from_2024-02-11_11-01-26.png
Attachment 17: Screenshot_from_2024-02-11_11-03-29.png
Screenshot_from_2024-02-11_11-03-29.png
Attachment 18: Screenshot_from_2024-02-11_11-16-26.png
Screenshot_from_2024-02-11_11-16-26.png
Attachment 19: Screenshot_from_2024-02-11_12-08-20.png
Screenshot_from_2024-02-11_12-08-20.png
Attachment 20: Screenshot_from_2024-02-11_13-39-20.png
Screenshot_from_2024-02-11_13-39-20.png
  505   Sat Feb 10 17:05:13 2024 TDRun 15 - background
17.55 DAQ continues G22-86/R15_11
      alpha background

      DSSSD bias OK - attachment 1
       ambient temperature 17.0 deg C

      FEE64 temperatures OK - attachment 2

      WR timestamps OK - attachment 3

      ADC data item stats - attachment 4
       rates dominated by 'hot' channels aida01 ( 85k ) and aida15 ( 11k ) - all other FEE64s c. 0Hz
       stats zero'd 17.57

      per FEE64 Rate & Stat spectra - attachments 5-6
       Stat spectra not zero'd recently - evidence of scattering ( rates varies with strip # )?

      aida01 ( p+n ) and aida03 ( n+n ) ASIC settings - attachments 7-8
       slow comparator 0x64

      Merger and Merger stats OK - attachments 9-10

      DSSSD leakage current = attachment 11
       shows data for 28-30.1.24 ?

      online sort per DSSSD x-y hit patterns - attachment 12

      ML vacuum pressure & cave temperature - attachments 13-14

23.01 DAQ continues G22-86/R15_17
      alpha background

      DSSSD bias OK - attachment 15
       ambient temperature 17.0 deg C

      FEE64 temperatures OK - attachment 16

      ADC data item stats - attachment 17
       rates dominated by 'hot' channels aida01 ( 72k ) and aida15 ( 11k ) - all other FEE64s c. 0Hz

      TapeServer - attachment 18

      DSSSD leakage current = attachment 19
       DSSSD #2 transient c. few hundred nA

      online sort per DSSSD x-y hit patterns - attachment 20
Attachment 1: Screenshot_from_2024-02-10_17-54-47.png
Screenshot_from_2024-02-10_17-54-47.png
Attachment 2: Screenshot_from_2024-02-10_17-55-42.png
Screenshot_from_2024-02-10_17-55-42.png
Attachment 3: Screenshot_from_2024-02-10_17-56-11.png
Screenshot_from_2024-02-10_17-56-11.png
Attachment 4: Screenshot_from_2024-02-10_17-56-47.png
Screenshot_from_2024-02-10_17-56-47.png
Attachment 5: Screenshot_from_2024-02-10_17-57-29.png
Screenshot_from_2024-02-10_17-57-29.png
Attachment 6: Screenshot_from_2024-02-10_17-58-40.png
Screenshot_from_2024-02-10_17-58-40.png
Attachment 7: Screenshot_from_2024-02-10_17-59-30.png
Screenshot_from_2024-02-10_17-59-30.png
Attachment 8: Screenshot_from_2024-02-10_17-59-50.png
Screenshot_from_2024-02-10_17-59-50.png
Attachment 9: Screenshot_from_2024-02-10_18-00-09.png
Screenshot_from_2024-02-10_18-00-09.png
Attachment 10: Screenshot_from_2024-02-10_18-00-44.png
Screenshot_from_2024-02-10_18-00-44.png
Attachment 11: Screenshot_from_2024-02-10_18-03-01.png
Screenshot_from_2024-02-10_18-03-01.png
Attachment 12: Screenshot_from_2024-02-10_18-03-19.png
Screenshot_from_2024-02-10_18-03-19.png
Attachment 13: tep.PNG
tep.PNG
Attachment 14: vac.PNG
vac.PNG
Attachment 15: Screenshot_from_2024-02-10_23-01-27.png
Screenshot_from_2024-02-10_23-01-27.png
Attachment 16: Screenshot_from_2024-02-10_23-00-50.png
Screenshot_from_2024-02-10_23-00-50.png
Attachment 17: Screenshot_from_2024-02-10_23-01-08.png
Screenshot_from_2024-02-10_23-01-08.png
Attachment 18: Screenshot_from_2024-02-10_23-02-34.png
Screenshot_from_2024-02-10_23-02-34.png
Attachment 19: Screenshot_from_2024-02-10_23-02-53.png
Screenshot_from_2024-02-10_23-02-53.png
Attachment 20: Screenshot_from_2024-02-10_23-03-09.png
Screenshot_from_2024-02-10_23-03-09.png
  504   Sat Feb 10 10:00:44 2024 AR RSS JM CBRun 13, 14

System checks: all ok

Attatchment 1: Rates ok

Attatchment 2: Temperatures ok

Attatchment 3: Waveform layout 7

Attatchment 4: Waveform layout 8

Attatchment 5: Waveform layout 1

Attatchment 6: Waveform layout 2

Attatchment 7: Leakage current ok

 

Moved pneumatic / servo for right / left arms in /out

Attachment 8: left servo fully in

9: left pneumatic in (beam is clipped)

10: left pneumatic & servo in (beam is clipped)

11: right servo in

12: right pneumatic in

13: right pneumatic & serve in

 

Some clear issues with strip mapping persist.

 

Strip map

#map strip: 01:00 left/top side, 02:00 right/bottom side (from beam view)            
stripMap    1    2                                            
stripMap    2    1                                            
stripMap    3    2                                            
stripMap    4    1                                            
stripMap    5    2
stripMap    6    1
stripMap    7    2
stripMap    8    1
stripMap    9    1
stripMap    10    2
stripMap    11    1
stripMap    12    2
stripMap    13    1
stripMap    14    2
stripMap    15    1
stripMap    16    2  

 

/////// UPDATE /////////

 

System checks: all ok - except waveform for AIDA 15 in layout 2, it has a low rate compared to others

Attatchment 14: Rates ok

Attatchment 15: Temperatures ok

Attatchment 16: Waveform layout 7

Attatchment 17: Waveform layout 8

Attatchment 18: Waveform layout 2

Attatchment 19: Waveform layout 1

Attatchment 20: Leakage current ok

/////// UPDATE //////

After beam stiring - layout of xy histograms are updated

DSSD0 and DSSD3 have the correct orientation,

DSSD1 needs to be flipped mirrored on the yaxis (i.e. right needs to be left, and left needs to be right but up and down are correct)

DSSD2 needs to be flipped on the xaxis (i.e. up needs to be down and down needs to be up but left and right are correct)

Horizontal offset in YR3 target = 9mm

Horizontal offset in YR7 target = 9mm

Horizontan offset in YR9 target = -9mm

Vertial offset in YR9 target = -8mm

Attatchment 21: Beam profile before stiring

Attatchment 22: Beam profile after stiritng - much better, hitting detectors less.

 

/// UPDATE ///

Mapping for the sort code has been fixed so that detectors are orientated correctly as if looking at the detectors downstream.We was missing dssdPNisX command which lets the sort code know if the PN FEEs correspond to X or Y. Correct mapping is now;

       
stripMap    1    1                                            
stripMap    2    2                                            
stripMap    3    1                                            
stripMap    4    2                                            
stripMap    5    1
stripMap    6    2
stripMap    7    2
stripMap    8    1
stripMap    9    2
stripMap    10    1
stripMap    11    1
stripMap    12    2
stripMap    13    1
stripMap    14    2
stripMap    15    1
stripMap    16    2   

 

 

 

 

Attachment 1: Screenshot_from_2024-02-10_11-01-27.png
Screenshot_from_2024-02-10_11-01-27.png
Attachment 2: Screenshot_from_2024-02-10_11-01-42.png
Screenshot_from_2024-02-10_11-01-42.png
Attachment 3: Screenshot_from_2024-02-10_11-02-37.png
Screenshot_from_2024-02-10_11-02-37.png
Attachment 4: Screenshot_from_2024-02-10_11-02-58.png
Screenshot_from_2024-02-10_11-02-58.png
Attachment 5: Screenshot_from_2024-02-10_11-03-25.png
Screenshot_from_2024-02-10_11-03-25.png
Attachment 6: Screenshot_from_2024-02-10_11-03-47.png
Screenshot_from_2024-02-10_11-03-47.png
Attachment 7: Screenshot_from_2024-02-10_11-05-40.png
Screenshot_from_2024-02-10_11-05-40.png
Attachment 8: l_s_in.png
l_s_in.png
Attachment 9: l_p_in.png
l_p_in.png
Attachment 10: l_p&s_in.png
l_p&s_in.png
Attachment 11: r_s_in.png
r_s_in.png
Attachment 12: r_p_in.png
r_p_in.png
Attachment 13: r_s&p_in.png
r_s&p_in.png
Attachment 14: Screenshot_from_2024-02-10_13-41-54.png
Screenshot_from_2024-02-10_13-41-54.png
Attachment 15: Screenshot_from_2024-02-10_13-42-11.png
Screenshot_from_2024-02-10_13-42-11.png
Attachment 16: Screenshot_from_2024-02-10_13-43-10.png
Screenshot_from_2024-02-10_13-43-10.png
Attachment 17: Screenshot_from_2024-02-10_13-47-04.png
Screenshot_from_2024-02-10_13-47-04.png
Attachment 18: Screenshot_from_2024-02-10_13-48-38.png
Screenshot_from_2024-02-10_13-48-38.png
Attachment 19: Screenshot_from_2024-02-10_13-49-39.png
Screenshot_from_2024-02-10_13-49-39.png
Attachment 20: Screenshot_from_2024-02-10_13-50-13.png
Screenshot_from_2024-02-10_13-50-13.png
Attachment 21: Screenshot_from_2024-02-10_14-12-54.png
Screenshot_from_2024-02-10_14-12-54.png
Attachment 22: Screenshot_from_2024-02-10_14-56-15.png
Screenshot_from_2024-02-10_14-56-15.png
Attachment 23: Screenshot_from_2024-02-10_16-15-27.png
Screenshot_from_2024-02-10_16-15-27.png
  503   Fri Feb 9 17:04:39 2024 JM, AR, CB, NHRun12 - Background
  502   Fri Feb 9 11:58:04 2024 TD + CB, JM, AR, NH, ABRun10 contd, Run11
12.46 DAQ continues G22-88/R10_14

      c. 20MeV 16O2+ + 4He 

      DSSSD bias OK - attachment 1
      ambient temperature 17.7 deg C ( c. 1deg C increase cf. earlier this week )

      FEE64 temperatures OK - attachment 2

      All system wide checks OK
      WR timestamp control OK - attachment 3

      ADC data item stats - attachment 4
      FEE64 rates c. 0Hz except 'hot' channels aida01 ( 71k ) and aida15 ( 8k )

      per FEE64 Rate and Stat spectra - attachments 5-8
       events ( beam? ) aida01 asic #4 and aida04 asic #4 ( c. 200k per ASIC channel )
       
      per FEE64 1.8.L spectra - attachments 9-10
       most spectra showing background and, possibly, a small amount of scattered beam?

      per FEE64 1.8.W spectra - 20us FSR - attachments 11-12

      aida04 3*L and 4*L spectra - attachments 13-15
       distribution by strip # - uniform across asic# #4, continuing across asic #3 with abrupt fall off c. channel 9
       not Rutherford - more consistent halo/beam scattering?
       what are the current lab angles subtended by the DSSSDs?
   
      Rutherford cross section for 20MeV 16O+4He



               Reaction: 4He(16O,16O)4He

                      A:           4         16         16          4
                      Z:           2          8          8          2
              Mass(amu):    4.002603  15.994915  15.994915   4.002603
             Error(keV):       0.001      0.002      0.002      0.001
                 Origin:
                 Status:

  Projectile lab energy:   20.000 MeV
              CM energy:    4.003 MeV

                     Particle 3                                 Particle 4
                   ( scattered )                                ( recoil )

  Lab         Cross Section         CM    Lab/CM    Lab        Lab      Lab
 Angle      Lab           CM       Angle   Ratio   Energy     Angle    Energy
 (deg)           (mb/sr)           (deg)           (MeV)      (deg)    (MeV)

  0.00 NaN          Inf            0.000 NaN       20.000     90.000    0.000
  1.00 0.143028E+09 0.572301E+07   4.999   0.040   19.976     87.500    0.024
  2.00 0.894079E+07  356425.      10.017   0.040   19.902     84.992    0.098
  3.00 0.176669E+07  69982.6      15.072   0.040   19.780     82.464    0.220
  4.00  559329.      21951.4      20.186   0.039   19.607     79.907    0.393
  5.00  229335.      8887.05      25.382   0.039   19.382     77.309    0.618
  6.00  110782.      4221.76      30.690   0.038   19.103     74.655    0.897
  7.00  59959.1      2235.75      36.144   0.037   18.768     71.928    1.232
  8.00  35299.6      1279.52      41.790   0.036   18.371     69.105    1.629
  9.00  22192.2      775.052      47.692   0.035   17.907     66.154    2.093
 10.00  14729.2      489.347      53.941   0.033   17.366     63.029    2.634
 11.00  10261.1      317.919      60.685   0.031   16.732     59.658    3.268
 12.00  7514.20      209.725      68.185   0.028   15.976     55.907    4.024
 13.00  5900.47      137.756      77.018   0.023   15.035     51.491    4.965
 14.00  5693.43      85.2271      89.184   0.015   13.688     45.408    6.312

 Maximum lab scattering angle (  14.49 deg ) exceeded

________________________________________________________________________________



      aida01 4*L spectra - attachments 19-23
       channels 0-8 - high stats structure(s) at low energy
       channels 9-16 - high energy peak cf. aida04 asic #4 spectra

      Merger, Merger stats, TapeServer - attachments 16-18

Started Run 11 with beam ON. 

We observe the same pattern of counts in DSSD0 (Top left detector) but not in DSSD1,2,3 (attachment 25). Moving in the left arm with the pneumatic we see a large drop in 
the rate on DSSD0 to essentially zero. The beam on the CRYRING monitor appears to disappear. 
When we move the right pneumatic IN, we observe an increase in the number of counts in the bottom-left part of the detector. The position is likely to be a mistake in the 
strip assignment.  
No increase observed in the leakage current. Will try systematically tomorrow.

Beam position monitoring from the CRYRING are shown in attachments 25-27
Attachment 1: Screenshot_from_2024-02-09_12-45-57.png
Screenshot_from_2024-02-09_12-45-57.png
Attachment 2: Screenshot_from_2024-02-09_12-46-32.png
Screenshot_from_2024-02-09_12-46-32.png
Attachment 3: Screenshot_from_2024-02-09_12-46-59.png
Screenshot_from_2024-02-09_12-46-59.png
Attachment 4: Screenshot_from_2024-02-09_12-47-37.png
Screenshot_from_2024-02-09_12-47-37.png
Attachment 5: Screenshot_from_2024-02-09_12-47-59.png
Screenshot_from_2024-02-09_12-47-59.png
Attachment 6: Screenshot_from_2024-02-09_12-48-59.png
Screenshot_from_2024-02-09_12-48-59.png
Attachment 7: Screenshot_from_2024-02-09_12-49-17.png
Screenshot_from_2024-02-09_12-49-17.png
Attachment 8: Screenshot_from_2024-02-09_12-49-39.png
Screenshot_from_2024-02-09_12-49-39.png
Attachment 9: Screenshot_from_2024-02-09_12-50-18.png
Screenshot_from_2024-02-09_12-50-18.png
Attachment 10: Screenshot_from_2024-02-09_12-51-04.png
Screenshot_from_2024-02-09_12-51-04.png
Attachment 11: Screenshot_from_2024-02-09_12-51-48.png
Screenshot_from_2024-02-09_12-51-48.png
Attachment 12: Screenshot_from_2024-02-09_12-52-25.png
Screenshot_from_2024-02-09_12-52-25.png
Attachment 13: Screenshot_from_2024-02-09_12-54-05.png
Screenshot_from_2024-02-09_12-54-05.png
Attachment 14: Screenshot_from_2024-02-09_12-56-04.png
Screenshot_from_2024-02-09_12-56-04.png
Attachment 15: Screenshot_from_2024-02-09_12-56-20.png
Screenshot_from_2024-02-09_12-56-20.png
Attachment 16: Screenshot_from_2024-02-09_12-56-52.png
Screenshot_from_2024-02-09_12-56-52.png
Attachment 17: Screenshot_from_2024-02-09_12-57-06.png
Screenshot_from_2024-02-09_12-57-06.png
Attachment 18: Screenshot_from_2024-02-09_12-57-36.png
Screenshot_from_2024-02-09_12-57-36.png
Attachment 19: Screenshot_from_2024-02-09_13-03-20.png
Screenshot_from_2024-02-09_13-03-20.png
Attachment 20: Screenshot_from_2024-02-09_13-03-38.png
Screenshot_from_2024-02-09_13-03-38.png
Attachment 21: Screenshot_from_2024-02-09_13-03-54.png
Screenshot_from_2024-02-09_13-03-54.png
Attachment 22: Screenshot_from_2024-02-09_13-04-22.png
Screenshot_from_2024-02-09_13-04-22.png
Attachment 23: Screenshot_from_2024-02-09_13-04-59.png
Screenshot_from_2024-02-09_13-04-59.png
Attachment 24: Screenshot_from_2024-02-10_10-07-57.png
Screenshot_from_2024-02-10_10-07-57.png
Attachment 25: 20240209_174604.jpg
20240209_174604.jpg
Attachment 26: 20240209_174705.jpg
20240209_174705.jpg
Attachment 27: 20240209_174718.jpg
20240209_174718.jpg
  501   Fri Feb 9 07:57:38 2024 AR, CB, AB, JMRun10 - First cool beam on target

8:58

System wide checks okay, except

   
             Base         Current         Difference
aida08 fault      0x0 :      0xc :      12  
aida09 fault      0x0 :      0xb :      11  
FPGA Timestamp error counter test result: Passed 14, Failed 2
If any of these counts are reported as in error
The ASIC readout system has detected a timeslip.
That is the timestamp read from the time FIFO is not younger than the last

Temperatures okay

Attatchment 1: Layout 7 Waveforms

Attatchment 2: Temperatures

Attatchment 3: Layout 8 Waveforms

Attatchment 4: Layout 2 counter okay

Attatchment 5: Layout 4 spectrum okay

Attatchment 6: Leakage current okay

Switched to run 10 (run 9 is overnight background)

 

10:00 Target pressure on. Attachment 7

Beam on target

Observe low counts. Only aida4 showing anything (attachment 8)

See a peak around 19 MeV in LowEnergyExratedssd0 (can only see in the dssd0) that is assumed to be Rutherford scattering - Attachment 12

Temperatures ok

Wavefroms on attatchment 9 show only signals from aida04 - attachment 9

/////// UPDATE ////////

Peak observed is not from Rutherford, we think it is beam (or part of it) hitting the detector

Attachment 10 shows waveforms after the beam was turned off -> counts in aida 4 are gone.

Beam off = no peak at 19 MeV for LowEnergyExdssd0.

Leakage current ok (attachment 11)

//// BEAM ON AGAIN UPDATE ////

Still see peak at 20 MeV, see attachment 13

Peak is not in dssds 1,2,3 only in dssd 0 (Top left detector) (attatchments 14,15,16 respectively)

Leakage currents OK (attatchment 17)

Attachment 18: LowEnergyXYtotalDSSD0 plot showing all hits in one spot (this is with BEAM AND TARGET on)

TEST: Turning off target while beam is running, can still see hotspot in DSSD0 but no others, so definitely beam related. (attatchment 18)

Tom's estimate: 10^6 particles in 10 minutes - not as many as can be handled

/////// UPDATE ////////

No beam, no target (target off)

All checks ok (see attachment 20 and 21 for rates and temperatures)

Attachment 22: Layout 7 waveforms

Attatchment 23: Layout 8 waveforms

Attachment 24: Layout 2 waveforms 

Attachment 25: Layout 1 waveforms (rate)

Atatchment 26: Leakage current

Attachment 27: LowEnergyExRatedssd 0,1,2,3 respectively

Note: Data similar to that seen in DSSD0 (19 MeV peak) is not observed in MIDAS for the other DSSDs -> Issue is not related to the monitoring code 

 

Attachment 1: Screenshot_from_2024-02-09_09-03-01.png
Screenshot_from_2024-02-09_09-03-01.png
Attachment 2: Screenshot_from_2024-02-09_09-03-28.png
Screenshot_from_2024-02-09_09-03-28.png
Attachment 3: Screenshot_from_2024-02-09_09-04-12.png
Screenshot_from_2024-02-09_09-04-12.png
Attachment 4: Screenshot_from_2024-02-09_09-08-45.png
Screenshot_from_2024-02-09_09-08-45.png
Attachment 5: Screenshot_from_2024-02-09_09-11-08.png
Screenshot_from_2024-02-09_09-11-08.png
Attachment 6: Screenshot_from_2024-02-09_09-13-29.png
Screenshot_from_2024-02-09_09-13-29.png
Attachment 7: 20240209_100016.jpg
20240209_100016.jpg
Attachment 8: Screenshot_from_2024-02-09_11-19-45.png
Screenshot_from_2024-02-09_11-19-45.png
Attachment 9: Screenshot_from_2024-02-09_11-21-36.png
Screenshot_from_2024-02-09_11-21-36.png
Attachment 10: Screenshot_from_2024-02-09_11-40-36.png
Screenshot_from_2024-02-09_11-40-36.png
Attachment 11: Screenshot_from_2024-02-09_11-39-09.png
Screenshot_from_2024-02-09_11-39-09.png
Attachment 12: dfcbe734-8730-4e0f-acfd-5e778e2b3f9f.JPG
dfcbe734-8730-4e0f-acfd-5e778e2b3f9f.JPG
Attachment 13: Screenshot_from_2024-02-09_13-17-12.png
Screenshot_from_2024-02-09_13-17-12.png
Attachment 14: Screenshot_from_2024-02-09_13-17-27.png
Screenshot_from_2024-02-09_13-17-27.png
Attachment 15: Screenshot_from_2024-02-09_13-17-31.png
Screenshot_from_2024-02-09_13-17-31.png
Attachment 16: Screenshot_from_2024-02-09_13-17-40.png
Screenshot_from_2024-02-09_13-17-40.png
Attachment 17: Screenshot_from_2024-02-09_13-18-46.png
Screenshot_from_2024-02-09_13-18-46.png
Attachment 18: Screenshot_from_2024-02-09_13-41-38.png
Screenshot_from_2024-02-09_13-41-38.png
Attachment 19: Screenshot_from_2024-02-09_13-55-21.png
Screenshot_from_2024-02-09_13-55-21.png
Attachment 20: Screenshot_from_2024-02-09_16-25-22.png
Screenshot_from_2024-02-09_16-25-22.png
Attachment 21: Screenshot_from_2024-02-09_16-25-38.png
Screenshot_from_2024-02-09_16-25-38.png
Attachment 22: Screenshot_from_2024-02-09_16-27-06.png
Screenshot_from_2024-02-09_16-27-06.png
Attachment 23: Screenshot_from_2024-02-09_16-27-40.png
Screenshot_from_2024-02-09_16-27-40.png
Attachment 24: Screenshot_from_2024-02-09_16-28-11.png
Screenshot_from_2024-02-09_16-28-11.png
Attachment 25: Screenshot_from_2024-02-09_16-28-39.png
Screenshot_from_2024-02-09_16-28-39.png
Attachment 26: Screenshot_from_2024-02-09_16-29-55.png
Screenshot_from_2024-02-09_16-29-55.png
Attachment 27: Screenshot_from_2024-02-09_16-30-11.png
Screenshot_from_2024-02-09_16-30-11.png
  500   Thu Feb 8 17:11:50 2024 CB, EHPulser walkthrough
17:30 File G22-88/R8

Pulser walkthrough

EGG Ortec 448 Pulser

Amplitide 90,000-10,000 @ 10,000 step
Normalise 5.0
Attenuators x5 IN
t_r 100ns
Tau_d 50us
Frequency Int Osc 50Hz
Polarity + ( - polarity via Cooknell SA1 Sum & Invert Amp )

18:10 Moved to new run R9

Pulser off. Overnight background.

Attachment 1: 4.png
4.png
Attachment 2: 3.png
3.png
  499   Thu Feb 8 14:59:59 2024 JM, TDChange p+n FEE64 settings
ASIC settings for the pn FEEs has been reset to the default aida settings. This was done to combat the observed dropping of several  channels in pn only FEE modules 
following motors movement. This effect was also observed in 2022 in several pulser walkthroughs conducted pre and post detector movement. No dropped channels were 
observed in nn FEEs. 

ASIC seetings changed;

IBias LF feedback from 0xf to 0x8
diode link threshold from 0xbf to 0xca

We also set the slow comparatort for all FEE modules back to 0x64 (1 MeV)

Settings tested using only 2 pn FEE modules (aida05 and aida16). Movement left and right. Dropped channels observed in all pn FEEs except aida05 and aida16. All pn ASIC 
settings then changed. Movement for 10 repeptitions. No dropped channels observed in pn FEEs. 

New settings saved in 2024Feb08-15.59.09
  498   Thu Feb 8 11:22:28 2024 JMShifters instruction 2024

EMERGENCY SHUTDOWNS

1. CARME motors: Q key on keyboard while on workspace 1 of appc218 (right of this double-monitor setup)

2. Internal target:

 

ELOG

Each shift should put all information relating to the shift in ONE elog entry unless a significant change occurs during the shift i.e beam off, target off, run change, detectors out etc.

Elog entry should be named by the shift - morning/afternoon/night followed by the current run number

TO DO FOR THE SHIFTERS - CHECK EVERY HOUR AND POST TO ELOG WITH SCREENSHOTS. ELOG should be updated with current beam energy, injection beam intensity, system checks OK/not, FEE temperature OK/not, Statistics OK/not, Merger OK/not, tape server data rate, Motors OK/not and min/max motors potentiometer value. The time of each update should be noted in the elog. Screenshots of important screens (detailed below) should also be uploaded to the elog.

 

CARME DAQ

CARME computer has 6 workspaces (WS), two monitors for each WS.

WS1 : left monitor - fee power

          right monitor - Detector Bias control, fee power server (just a terminal, do not close)

 

WS2 : left monitor - DAQ controls

          right monitor - DAQ server (just a terminal, do not close)

          Most important tabs (DAQ controls)

  • Run control : shows daq status, stopped / going
  • Fee temperatures : temperatures of fees labelled aida01, aida02, aida03, aida04, aida05, aida06, aida07, aida08, aida09, aida10, aida11, aida12, aida13, aida14, aida15, aida16. Hit reload, temperatures will update and should all be green. 1-2 C above safe limit (top) is OK. More than that - warn expert shifter immediately.
  • Statistics : shows the statistics for each fee. Make sure ADC Data Items is selected in the left hand menu. Hit update all  (SCREENSHOT once, if no changes record OK in elog)
  • System wide checks : checks for the setup. Click on 'Check Clock Status', 'Check ADC Calibration' and 'Check the White rabbit decoder status' to run checks. (SCREENSHOT once, if no changes record OK in elog)
  • Spectrum browser : loads histograms for fees and fee channels from preset layouts. select layout from 'layout ID' then restore layout from 'Arrange functions' drop down menu. Check layout 1, Select log scale and change xmax as 128 . This displays the data rate per strip. If any strips are a lot higher than the others, or very different from previous entry, make a note in the ELOG. (SCREENSHOT once, if no changes record OK in elog)
  • Asic control : sets asic parameters

 

WS3 : left monitor - Merge control, Tape server

          right monitor - servers (just terminals, do not close) , data being written to file

  • Check NewMerger tab. Hit reload button and see if 'current links with data' and 'current merged data rate' are updating. All links should go green, but not necessarily at the same time. Keep hitting refresh until you see all of them going green at least once.
  • Data rate, hit reload should be around 100,000. Record data rate in elog
  • Check tapeService tab. Hit reload button. Check data rate in Kbytes/sec box. Record data rate to elog. This is where the current run number is displayed
  • On the right-hand monitor, bottom right terminal, check the current run fragment (sub-run) is being written and is slowly increasing in size. It refreshed automatically every 5 seconds.

 

WS4 : Leakage current plot

  • Check leakage current plot is stable. Transients will occur during motors movement but baseline current should remain consistent. Make sure it is updating and time is correct. (SCREENSHOT once, record OK in elog if no changes) Check present values on WS2, CAEN window

 

WS5 : Online data monitoring

  • Check the following plots are updating for each DSSD
    • LowEnergyXYTotal (right click - col - colz). This shows the total number of events, per pixel, since the code was started / re-started.
    • LowEnergyXYRate (right click - col - colz). This shows the rate over the last 20 seconds.
    • LowEnergyExRate (double left click. Right click on y axis - Log scale on y). Rate over the last 20 seconds
    • LowEnergyExTotal (double left click. Right click on y axis - Log scale on y). This shows the total energy deposited on all X strips since code restart.
    • LowEnergyEyTotal (as above). as above, Y strips. (SCREENSHOT of each per DSSD per shift. If significant changes are observed screenshot again)

 

WS6 : Browser, ELOG

-->IMPORTANT NOTE: MIDAS DAQ can respond slowly. DO NOT try to change tab or give new command while the previous one is being executed. Check activity in right-hand part of tab currently in use. Do not double click on buttons.

 

Motors computer (Computer to the right)

  • Check motors cycle is running. The code will cycle through the loaded motors instructions. When an instruction is being done it is highlighted in bold. Check the bold highlighting cycles through entire set of instructions.
  • Check potentiometer is reproducing same max/min values for each cycle of instructions.
  • Detectors are on LEFT and RIGHT arm.
  • No need for screenshots, but record in the ELOG you checked this.

 

  497   Wed Feb 7 13:50:20 2024 Elena Hanu, Carlo Bruno, Michael LestinskyAfternoon shift

14:50

 

Temperature, Statistics fine (see screenshots).

System wide checks fine (see below).

All FEEs pass clock check.

All FEEs pass White Rabbit check.

ADC Calibration check:

FEE64 module aida02 failed
FEE64 module aida16 failed
Calibration test result: Passed 14, Failed 2

If any modules fail calibration , check the clock status and open the FADC Align and Control browser page to rerun calibration for that module

FPGA timestamp error check:

         Base         Current         Difference
aida09 fault      0x1 :      0xa :      9  
FPGA Timestamp error counter test result: Passed 15, Failed 1
If any of these counts are reported as in error
The ASIC readout system has detected a timeslip.
That is the timestamp read from the time FIFO is not younger than the last

 

16:00 Slow comp thresh lowered to 100 keV in preparation for detector moving. Will have to test systematically what is the highest threshold we can accept

 

17:00 Temps fine. No overnight run.

 

 

 

Attachment 1: Screenshot_from_2024-02-07_14-53-36.png
Screenshot_from_2024-02-07_14-53-36.png
Attachment 2: Screenshot_from_2024-02-07_14-56-37.png
Screenshot_from_2024-02-07_14-56-37.png
Attachment 3: Screenshot_from_2024-02-07_14-51-21.png
Screenshot_from_2024-02-07_14-51-21.png
  496   Wed Feb 7 10:09:14 2024 CB, JG, EMMorning Shift

11:00

Changed slow comp threshold to 0x32 (500 keV). Checkloaded.

 

Temperature, Statistics fine (see screenshots).

System wide checks fine (see below).

All FEEs pass clock check.

All FEEs pass White Rabbit check.

ADC Calibration check:

FEE64 module aida02 failed
FEE64 module aida16 failed
Calibration test result: Passed 14, Failed 2

If any modules fail calibration , check the clock status and open the FADC Align and Control browser page to rerun calibration for that module

FPGA timestamp error check:

         Base         Current         Difference
aida09 fault      0x1 :      0xa :      9  
FPGA Timestamp error counter test result: Passed 15, Failed 1
If any of these counts are reported as in error
The ASIC readout system has detected a timeslip.
That is the timestamp read from the time FIFO is not younger than the last

 

 

Attachment 1: Screenshot_from_2024-02-07_11-10-08.png
Screenshot_from_2024-02-07_11-10-08.png
Attachment 2: Screenshot_from_2024-02-07_11-10-59.png
Screenshot_from_2024-02-07_11-10-59.png
Attachment 3: Screenshot_from_2024-02-07_11-15-53.png
Screenshot_from_2024-02-07_11-15-53.png
Attachment 4: Screenshot_from_2024-02-07_11-20-38.png
Screenshot_from_2024-02-07_11-20-38.png
  495   Tue Feb 6 09:39:54 2024 CBDetector bias tests

10:30 Found ADC#4 on aida 1 and 4 are suddenly noisy. See attach 1-2. Checkloaded multiple times - no effect.

Leakage current overnight was stable, but a clear transient can be observed yesterday around the times of the DSSD motion test. We did not observe this in the cave looking at the CAEN module - it might have been too fast.

Now reducing bias to 50 V and will check differences later. FEE temperatures OK.

 

10:50 Started DAQ. Rates are much higher in n+n strips, but p+n strips appear unaffected. See attach 4-5.

Stopped DAQ. Will leave 50V until after lunch.

 

11:27 Checked motor control status after yesterday's setup (see previous ELOG entry). See attach 6.

IN counters are respectively: read, increases, latch (left), latch (right)

Read is the number of pulses sent with the instruction to move. Increases is the number of times the instruction was considered. Latch on/off indicates the 1 s time interval during which commands sent to a motor arm (left or right) are considered double-firing or mistake, and ignored.

The number of increases is the same for move in/out - good!

The number of increases for move out is 1 unit less than move in - one double pulse / double command must have been received overnight and ignored.

Note torque on right arm is 2 (arbitrary units) which is unusually high and may indicate some tension in the kapton wires. No apparent effect on DSSD2,4 on that arm. Will not move for now.

 

13:24 FEE temperatures OK, leakage currents decreasing slighly.

13:41 Slowly increased bias to 100V. Interestingly it is now DSSD#4 (Ch#3) that displays the highest leakage current. Attach 7

13:45 Stats & Rates look rather similar to earlier this morning at 150V, but aida01 and 04 are quieter. Attach 8-9. Will leave at 100V for a while longer before moving to 150V.

 

15:20 Set bias to 150V

15:50 Nikos testing target. Started short run. Stats & Rates attached (10-11). Same as yesterday, aida01 and 04 ADC#4 no longer noisy.

V-I on attach 12-13. Legends unreliable!

 

17:00 Nikos completed target test. YR09 pressure around 3E-10 mbar with around 5E11 atoms/cm2 He target. Target not quite stable (10-20%), some optimisation required, but pressure holds at 3E-10 mbar for several minutes.

Beam on target will increase ring pressure. Can't test for now. Target tests without beam complete. Need to log all the data remotely. No logging available at present.

Leaving bias at 150V overnight. No run.

 

 

Attachment 1: Screenshot_from_2024-02-06_10-39-06.png
Screenshot_from_2024-02-06_10-39-06.png
Attachment 2: Screenshot_from_2024-02-06_10-39-15.png
Screenshot_from_2024-02-06_10-39-15.png
Attachment 3: Screenshot_from_2024-02-06_10-39-27.png
Screenshot_from_2024-02-06_10-39-27.png
Attachment 4: Screenshot_from_2024-02-06_10-51-42.png
Screenshot_from_2024-02-06_10-51-42.png
Attachment 5: Screenshot_from_2024-02-06_10-51-50.png
Screenshot_from_2024-02-06_10-51-50.png
Attachment 6: Screenshot_from_2024-02-06_11-26-53.png
Screenshot_from_2024-02-06_11-26-53.png
Attachment 7: Screenshot_from_2024-02-06_13-41-40.png
Screenshot_from_2024-02-06_13-41-40.png
Attachment 8: Screenshot_from_2024-02-06_13-44-26.png
Screenshot_from_2024-02-06_13-44-26.png
Attachment 9: Screenshot_from_2024-02-06_13-44-38.png
Screenshot_from_2024-02-06_13-44-38.png
Attachment 10: Screenshot_from_2024-02-06_15-49-10.png
Screenshot_from_2024-02-06_15-49-10.png
Attachment 11: Screenshot_from_2024-02-06_15-49-16.png
Screenshot_from_2024-02-06_15-49-16.png
Attachment 12: Screenshot_from_2024-02-06_15-53-22.png
Screenshot_from_2024-02-06_15-53-22.png
Attachment 13: Screenshot_from_2024-02-06_15-54-40.png
Screenshot_from_2024-02-06_15-54-40.png
  494   Mon Feb 5 20:15:31 2024 CB, TDDetector move in/out signals
17.43 setup signals from detector movement controller.

See attachment 1
                                                         NIM               TTL                 TTL             +12V
Signal from CRYRING ------> CAEN N89 ------> LF4000 -------> CAEN N89 -------> LeCroy 222 -------> PCB amp -------> start move in
( active high )         TTL              NIM  FIFO  -------> CAEN N89 -------> LeCroy 222 -------> PCB amp -------> start move out
                                                    not(NIM)               TTL                 TTL    ^        +12V
                                                                                                      |
                                                                                                      |
CAEN N89 mode switch to OUT                                                                                                  PSU +12V 

Widths of signals from LeCroy 222
 move in = 75ms
 move out = 25ms

Note - the outputs of EG&G/ESN LF4000 FIFO could be connected directly to the LeCroy 222 as the inputs accept NIM and TTL
Attachment 1: 20240205_174307.jpg
20240205_174307.jpg
  493   Mon Feb 5 19:43:52 2024 CB, TD, MLBuTIS reference for VETAR2
16.40 CB/ML install fibre optic cable from patch panel beneath dipole downstream of YR09/CARME chamber to VETAR2 front panel fibre connector - see attachments 1 & 2.

VETAR2 LCD display changes from magenta to cyan.

DAQ stop. DAQ reset. DAQ setup.

All system wide checks OK.

WR timestamp control - attachment 3 is timed 16.43.11 UTC+1

WR timestamp 0x17B101E7 7FCD1ACD -> 1707147827543349965ns -> 1707147827s -> Monday, February 5, 2024 3:43:47 PM UTC ( https://www.epochconverter.com/ )
Attachment 1: 20240205_173029.jpg
20240205_173029.jpg
Attachment 2: 20240205_172957.jpg
20240205_172957.jpg
Attachment 3: Screenshot_from_2024-02-05_16-43-18.png
Screenshot_from_2024-02-05_16-43-18.png
  492   Mon Feb 5 12:48:50 2024 CB TDPulser walkthough - Run4
13.41 DAQ start - file G22-88/R4

Pulser walkthrough

EGG Ortec 448 Pulser

Amplitide 90,000-10,000 @ 10,000 step
Normalise 5.0
Attenuators x5 IN
t_r 100ns
Tau_d 50us
Frequency Int Osc 50Hz
Polarity + ( - polarity via Cooknell SA1 Sum & Invert Amp )

File G22-88/R4

14.08 DAQ stop file G22-88/R4_8
Attachment 1: Screenshot_from_2024-02-05_14-04-33.png
Screenshot_from_2024-02-05_14-04-33.png
Attachment 2: Screenshot_from_2024-02-05_14-05-16.png
Screenshot_from_2024-02-05_14-05-16.png
  491   Mon Feb 5 05:45:26 2024 CB TDRun3 - end
06.40 DAQ continues file G22-88/R3_42

      DSSSD bias OK - attachment 1
      ambient temperature 16.8 deg C

      FEE64 temperatures OK - attachment 2

      ADC data item stats - attachments 3
       aida15 rate c. 34k ( 2x 'hot' channels, 1x channel missing, all other channels OK )
       aida01 rate c. 75k ( 2x 'hot' channels )

      per FEE64 Stat spectra - attachments 4-5

      Merger and TapeServer - attachments 6-7

      All system wide checks OK except
       aida02 & aida08 FPGA timestamp error - attachment 8

      WR timestamps OK - attachment 9

09.47 DAQ continues file G22-88/R3_46

      DSSSD bias OK - attachment 10
      ambient temperature 16.8 deg C

      FEE64 temperatures OK - attachment 11

      ADC data item stats - attachments 12
       aida15 rate c. 34k ( 2x 'hot' channels, 1x channel missing, all other channels OK )
       aida01 rate c. 75k ( 2x 'hot' channels )

      per FEE64 Stat spectra - attachments 13-14

      Merger and TapeServer - attachments 15-16

09.52 per FEE64 Rate spectra - attachments 17-18

      per FEE64 1.8.L spectra - attachments 19-20

      per FEE64 1.8.H spectra - attachments 21-22

      per p+n FEE64 1.8.W spectra - 20us, 200us, 2ms & 20ms FSR - attachments 23-26

      per n+n FEE64 1.8.W spectra - 20us, 200us, 2ms & 20ms FSR - attachments 27-30


10.11 Merger stats blocks 0-4, 14-16 - attachments 31-38

10.14 aida01-aida04, aida14-aida16 stats- attachments 39-45

10.21 Overnight V-I attach 46

10.23 Save current ASIC settings 2024Feb05-10.22.28
       changes to disc mask and LED waveform threshold for some FEE64s

10.32 Analysis of data file G22-88/R3_45 - attachments 47
       deadtime aida01 c. 0.5%, all other FEE64s deadtime c. 0%


13.07 DAQ stop file G22-88/R3_50
      Stop merger
      Stop tapeserver

Afternoon: Tested detector motion. Channels on p+n strips drop out with high threshold (1 MeV), but not with low threshold (100 keV). No issues on n+n strips. Reasons unclear.
Attachment 1: Screenshot_from_2024-02-05_06-40-57.png
Screenshot_from_2024-02-05_06-40-57.png
Attachment 2: Screenshot_from_2024-02-05_06-41-22.png
Screenshot_from_2024-02-05_06-41-22.png
Attachment 3: Screenshot_from_2024-02-05_06-41-47.png
Screenshot_from_2024-02-05_06-41-47.png
Attachment 4: Screenshot_from_2024-02-05_06-42-04.png
Screenshot_from_2024-02-05_06-42-04.png
Attachment 5: Screenshot_from_2024-02-05_06-42-29.png
Screenshot_from_2024-02-05_06-42-29.png
Attachment 6: Screenshot_from_2024-02-05_06-42-55.png
Screenshot_from_2024-02-05_06-42-55.png
Attachment 7: Screenshot_from_2024-02-05_06-43-16.png
Screenshot_from_2024-02-05_06-43-16.png
Attachment 8: Screenshot_from_2024-02-05_06-44-10.png
Screenshot_from_2024-02-05_06-44-10.png
Attachment 9: Screenshot_from_2024-02-05_06-44-55.png
Screenshot_from_2024-02-05_06-44-55.png
Attachment 10: Screenshot_from_2024-02-05_09-46-59.png
Screenshot_from_2024-02-05_09-46-59.png
Attachment 11: Screenshot_from_2024-02-05_09-47-39.png
Screenshot_from_2024-02-05_09-47-39.png
Attachment 12: Screenshot_from_2024-02-05_09-47-52.png
Screenshot_from_2024-02-05_09-47-52.png
Attachment 13: Screenshot_from_2024-02-05_09-48-22.png
Screenshot_from_2024-02-05_09-48-22.png
Attachment 14: Screenshot_from_2024-02-05_09-48-29.png
Screenshot_from_2024-02-05_09-48-29.png
Attachment 15: Screenshot_from_2024-02-05_09-48-51.png
Screenshot_from_2024-02-05_09-48-51.png
Attachment 16: Screenshot_from_2024-02-05_09-49-05.png
Screenshot_from_2024-02-05_09-49-05.png
Attachment 17: Screenshot_from_2024-02-05_09-52-04.png
Screenshot_from_2024-02-05_09-52-04.png
Attachment 18: Screenshot_from_2024-02-05_09-52-14.png
Screenshot_from_2024-02-05_09-52-14.png
Attachment 19: Screenshot_from_2024-02-05_09-52-57.png
Screenshot_from_2024-02-05_09-52-57.png
Attachment 20: Screenshot_from_2024-02-05_09-53-33.png
Screenshot_from_2024-02-05_09-53-33.png
Attachment 21: Screenshot_from_2024-02-05_09-55-33.png
Screenshot_from_2024-02-05_09-55-33.png
Attachment 22: Screenshot_from_2024-02-05_09-56-09.png
Screenshot_from_2024-02-05_09-56-09.png
Attachment 23: Screenshot_from_2024-02-05_09-57-05.png
Screenshot_from_2024-02-05_09-57-05.png
Attachment 24: Screenshot_from_2024-02-05_09-58-51.png
Screenshot_from_2024-02-05_09-58-51.png
Attachment 25: Screenshot_from_2024-02-05_09-59-23.png
Screenshot_from_2024-02-05_09-59-23.png
Attachment 26: Screenshot_from_2024-02-05_09-59-56.png
Screenshot_from_2024-02-05_09-59-56.png
Attachment 27: Screenshot_from_2024-02-05_10-03-26.png
Screenshot_from_2024-02-05_10-03-26.png
Attachment 28: Screenshot_from_2024-02-05_10-02-25.png
Screenshot_from_2024-02-05_10-02-25.png
Attachment 29: Screenshot_from_2024-02-05_10-01-57.png
Screenshot_from_2024-02-05_10-01-57.png
Attachment 30: Screenshot_from_2024-02-05_10-01-25.png
Screenshot_from_2024-02-05_10-01-25.png
Attachment 31: Screenshot_from_2024-02-05_10-11-41.png
Screenshot_from_2024-02-05_10-11-41.png
Attachment 32: Screenshot_from_2024-02-05_10-11-46.png
Screenshot_from_2024-02-05_10-11-46.png
Attachment 33: Screenshot_from_2024-02-05_10-11-51.png
Screenshot_from_2024-02-05_10-11-51.png
Attachment 34: Screenshot_from_2024-02-05_10-11-57.png
Screenshot_from_2024-02-05_10-11-57.png
Attachment 35: Screenshot_from_2024-02-05_10-12-04.png
Screenshot_from_2024-02-05_10-12-04.png
Attachment 36: Screenshot_from_2024-02-05_10-12-39.png
Screenshot_from_2024-02-05_10-12-39.png
Attachment 37: Screenshot_from_2024-02-05_10-12-44.png
Screenshot_from_2024-02-05_10-12-44.png
Attachment 38: Screenshot_from_2024-02-05_10-13-11.png
Screenshot_from_2024-02-05_10-13-11.png
Attachment 39: Screenshot_from_2024-02-05_10-14-43.png
Screenshot_from_2024-02-05_10-14-43.png
Attachment 40: Screenshot_from_2024-02-05_10-14-50.png
Screenshot_from_2024-02-05_10-14-50.png
Attachment 41: Screenshot_from_2024-02-05_10-15-43.png
Screenshot_from_2024-02-05_10-15-43.png
Attachment 42: Screenshot_from_2024-02-05_10-15-48.png
Screenshot_from_2024-02-05_10-15-48.png
Attachment 43: Screenshot_from_2024-02-05_10-15-54.png
Screenshot_from_2024-02-05_10-15-54.png
Attachment 44: Screenshot_from_2024-02-05_10-16-00.png
Screenshot_from_2024-02-05_10-16-00.png
Attachment 45: Screenshot_from_2024-02-05_10-16-06.png
Screenshot_from_2024-02-05_10-16-06.png
Attachment 46: Screenshot_from_2024-02-05_10-21-03.png
Screenshot_from_2024-02-05_10-21-03.png
Attachment 47: R3_45
 *** TDR format 3.3.0 analyser - TD - May 2021
 *** ERROR: READ I/O error:       5002
                   blocks:      32000
          ADC data format:  261748120 (   83699.2 Hz)
        Other data format:     171882 (      55.0 Hz)
 Sample trace data format:          0 (       0.0 Hz)
         Undefined format:          0 (       0.0 Hz)
   Other data format type:      PAUSE:        194 (       0.1 Hz)
                               RESUME:        194 (       0.1 Hz)
                              SYNC100:      43648 (      14.0 Hz)
                              WR48-63:      43648 (      14.0 Hz)
                           FEE64 disc:      84198 (      26.9 Hz)
                             MBS info:          0 (       0.0 Hz)
                           Other info:          0 (       0.0 Hz)

   ADC data range bit set:          3 (       0.0 Hz)

                Timewarps:        ADC:          0 (       0.0 Hz)
                                PAUSE:          0 (       0.0 Hz)
                               RESUME:          0 (       0.0 Hz)
                              SYNC100:          0 (       0.0 Hz)
                              WR48-63:          0 (       0.0 Hz)
                           FEE64 disc:          0 (       0.0 Hz)
                             MBS info:          0 (       0.0 Hz)
                            Undefined:          0 (       0.0 Hz)
                         Sample trace:          0 (       0.0 Hz)

 *** Timestamp elapsed time:     3127.246 s
 FEE  elapsed dead time(s) elapsed idle time(s)
  0               17.539                0.000
  1                0.000                0.000
  2                0.000                0.000
  3                0.000              991.338
  4                0.000                0.000
  5                0.000                0.000
  6                0.000                0.000
  7                0.000                0.000
  8                0.000                0.000
  9                0.000                0.000
 10                0.000                0.000
 11                0.000                0.000
 12                0.000                0.000
 13                0.000                0.000
 14                0.500                0.000
 15                0.000                0.000
 16                0.000                0.000
 17                0.000                0.000
 18                0.000                0.000
 19                0.000                0.000
 20                0.000                0.000
 21                0.000                0.000
 22                0.000                0.000
 23                0.000                0.000
 24                0.000                0.000
 25                0.000                0.000
 26                0.000                0.000
 27                0.000                0.000
 28                0.000                0.000
 29                0.000                0.000
 30                0.000                0.000
 31                0.000                0.000
 32                0.000                0.000

 *** Statistics
 FEE  ADC Data Other Data     Sample  Undefined      Pause     Resume    SYNC100    WR48-63       Disc        MBS      Other   HEC Data
  0  172462915     139302          0          0        170        170      27567      27567      83828          0          0          0
  1         45          2          0          0          0          0          0          0          2          0          0          1
  2         58          0          0          0          0          0          0          0          0          0          0          0
  3     167035         79          0          0          0          0         24         24         31          0          0          0
  4        332          0          0          0          0          0          0          0          0          0          0          0
  5        257          0          0          0          0          0          0          0          0          0          0          0
  6        254          0          0          0          0          0          0          0          0          0          0          0
  7        370          0          0          0          0          0          0          0          0          0          0          0
  8        336          0          0          0          0          0          0          0          0          0          0          0
  9         75          0          0          0          0          0          0          0          0          0          0          0
 10         99          0          0          0          0          0          0          0          0          0          0          0
 11        240        164          0          0          0          0          0          0        164          0          0          0
 12       1266          0          0          0          0          0          0          0          0          0          0          0
 13        218          0          0          0          0          0          0          0          0          0          0          0
 14   89113402      32165          0          0         24         24      16057      16057          3          0          0          2
 15       1218        170          0          0          0          0          0          0        170          0          0          0
 16          0          0          0          0          0          0          0          0          0          0          0          0
 17          0          0          0          0          0          0          0          0          0          0          0          0
 18          0          0          0          0          0          0          0          0          0          0          0          0
 19          0          0          0          0          0          0          0          0          0          0          0          0
 20          0          0          0          0          0          0          0          0          0          0          0          0
 21          0          0          0          0          0          0          0          0          0          0          0          0
 22          0          0          0          0          0          0          0          0          0          0          0          0
 23          0          0          0          0          0          0          0          0          0          0          0          0
 24          0          0          0          0          0          0          0          0          0          0          0          0
 25          0          0          0          0          0          0          0          0          0          0          0          0
 26          0          0          0          0          0          0          0          0          0          0          0          0
 27          0          0          0          0          0          0          0          0          0          0          0          0
 28          0          0          0          0          0          0          0          0          0          0          0          0
 29          0          0          0          0          0          0          0          0          0          0          0          0
 30          0          0          0          0          0          0          0          0          0          0          0          0
 31          0          0          0          0          0          0          0          0          0          0          0          0
 32          0          0          0          0          0          0          0          0          0          0          0          0

 *** Timewarps
 FEE       ADC      Pause     Resume    SYNC100    WR48-63       Disc        MBS  Undefined    Samples
  0          0          0          0          0          0          0          0          0          0
  1          0          0          0          0          0          0          0          0          0
  2          0          0          0          0          0          0          0          0          0
  3          0          0          0          0          0          0          0          0          0
  4          0          0          0          0          0          0          0          0          0
  5          0          0          0          0          0          0          0          0          0
  6          0          0          0          0          0          0          0          0          0
  7          0          0          0          0          0          0          0          0          0
  8          0          0          0          0          0          0          0          0          0
  9          0          0          0          0          0          0          0          0          0
 10          0          0          0          0          0          0          0          0          0
 11          0          0          0          0          0          0          0          0          0
 12          0          0          0          0          0          0          0          0          0
 13          0          0          0          0          0          0          0          0          0
 14          0          0          0          0          0          0          0          0          0
 15          0          0          0          0          0          0          0          0          0
 16          0          0          0          0          0          0          0          0          0
 17          0          0          0          0          0          0          0          0          0
 18          0          0          0          0          0          0          0          0          0
 19          0          0          0          0          0          0          0          0          0
 20          0          0          0          0          0          0          0          0          0
 21          0          0          0          0          0          0          0          0          0
 22          0          0          0          0          0          0          0          0          0
 23          0          0          0          0          0          0          0          0          0
 24          0          0          0          0          0          0          0          0          0
 25          0          0          0          0          0          0          0          0          0
 26          0          0          0          0          0          0          0          0          0
 27          0          0          0          0          0          0          0          0          0
 28          0          0          0          0          0          0          0          0          0
 29          0          0          0          0          0          0          0          0          0
 30          0          0          0          0          0          0          0          0          0
 31          0          0          0          0          0          0          0          0          0
 32          0          0          0          0          0          0          0          0          0

 *** Program elapsed time:   24.477s ( 1307.373 blocks/s,  81.711 Mb/s)
  490   Sun Feb 4 08:40:04 2024 CB TDRun3 - contd
09.30 DAQ continues file G22-88/R3_17

      DSSSD bias OK - attachment 1
      ambient temperature 16.8 deg C

      FEE64 temperatures OK - attachment 2

      ADC data item stats - attachment 3
       aida15 rate c. 25k ( 2x 'hot' channels, 1x channel missing, all other channels OK )
       aida01 rate c. 45k ( 2x 'hot' channels )

      per FEE64 Stat spectra - attachments 4-5

      per FEE64 1.8.L spectra - attachments 6-7

      per FEE64 1.8.W spectra - 20us FSR - attachments 8-9

      Merger and TapeServer - attachments 10-11

09.36 Check ASIC load
      No change to aida01 ADC data item rate

      Last access times of data files indicate that data rate changed c. 05.00 - 06.00 this am.
      No obvious vacuum/temperature transients in last six hours.

09.55 All system wide checks OK except
       FPGA timestamp errors - attachment 12
      WR timestamps OK - attachment 13

10:06 V-I plot overnight - attach 14.

13.35 DAQ continues file G22-88/R3_22

      DSSSD bias OK - attachment 15
      ambient temperature 16.8 deg C

      FEE64 temperatures OK - attachment 16

      ADC data item stats - attachment 17
       aida15 rate c. 35k ( 2x 'hot' channels, 1x channel missing, all other channels OK )
       aida01 rate c. 51k ( 2x 'hot' channels )

      per FEE64 Rate spectra - attachments 18

      per FEE64 Stat spectra - attachments 19-20

      Merger and TapeServer - attachments 21-22
Attachment 1: Screenshot_from_2024-02-04_09-30-41.png
Screenshot_from_2024-02-04_09-30-41.png
Attachment 2: Screenshot_from_2024-02-04_09-31-07.png
Screenshot_from_2024-02-04_09-31-07.png
Attachment 3: Screenshot_from_2024-02-04_09-31-40.png
Screenshot_from_2024-02-04_09-31-40.png
Attachment 4: Screenshot_from_2024-02-04_09-32-26.png
Screenshot_from_2024-02-04_09-32-26.png
Attachment 5: Screenshot_from_2024-02-04_09-32-42.png
Screenshot_from_2024-02-04_09-32-42.png
Attachment 6: Screenshot_from_2024-02-04_09-33-23.png
Screenshot_from_2024-02-04_09-33-23.png
Attachment 7: Screenshot_from_2024-02-04_09-34-00.png
Screenshot_from_2024-02-04_09-34-00.png
Attachment 8: Screenshot_from_2024-02-04_09-34-37.png
Screenshot_from_2024-02-04_09-34-37.png
Attachment 9: Screenshot_from_2024-02-04_09-35-21.png
Screenshot_from_2024-02-04_09-35-21.png
Attachment 10: Screenshot_from_2024-02-04_09-37-23.png
Screenshot_from_2024-02-04_09-37-23.png
Attachment 11: Screenshot_from_2024-02-04_09-38-28.png
Screenshot_from_2024-02-04_09-38-28.png
Attachment 12: Screenshot_from_2024-02-04_09-55-02.png
Screenshot_from_2024-02-04_09-55-02.png
Attachment 13: Screenshot_from_2024-02-04_09-55-18.png
Screenshot_from_2024-02-04_09-55-18.png
Attachment 14: Screenshot_from_2024-02-04_10-05-07.png
Screenshot_from_2024-02-04_10-05-07.png
Attachment 15: Screenshot_from_2024-02-04_13-35-51.png
Screenshot_from_2024-02-04_13-35-51.png
Attachment 16: Screenshot_from_2024-02-04_13-36-21.png
Screenshot_from_2024-02-04_13-36-21.png
Attachment 17: Screenshot_from_2024-02-04_13-36-41.png
Screenshot_from_2024-02-04_13-36-41.png
Attachment 18: Screenshot_from_2024-02-04_13-37-35.png
Screenshot_from_2024-02-04_13-37-35.png
Attachment 19: Screenshot_from_2024-02-04_13-38-27.png
Screenshot_from_2024-02-04_13-38-27.png
Attachment 20: Screenshot_from_2024-02-04_13-38-52.png
Screenshot_from_2024-02-04_13-38-52.png
Attachment 21: Screenshot_from_2024-02-04_13-40-02.png
Screenshot_from_2024-02-04_13-40-02.png
Attachment 22: Screenshot_from_2024-02-04_13-40-20.png
Screenshot_from_2024-02-04_13-40-20.png
  489   Sat Feb 3 15:05:10 2024 CB, TD[How To] Restart Anydesk

1. Check Proxy server is correct

localhost : 8080

incoming connections 7070

 

Establish port

ssh -L 8080:proxy.gsi.de:8080 carme@atppc025

 

2. Check a password for remote access has been set up

 

3. Close anydesk and check it is *really* closed using

ps -o pid= -C anydesk

kill -9 any process (as root)

 

4 Restart AnyDesk

systemctl restart anydesk (as npg)

  488   Sat Feb 3 00:13:03 2024 CB TDRun1 end - R3 start
01.07 DAQ continues file G22-88/R1_46
      background alpha

      DSSSD bias OK - attachment 1
      FEE64 temps OK - attachment 2
      ADC data item stats - attachment 3
       aida15 asic #2 u/s - rate c. 266k 

      aida11 1*L spectra - attachment 4

      ADC data items/s Merger, Tape Server & Merger stats - attachments 5-7
       data rate to disk c. 2Mb/s dominated by aida15 asic 2


10.05 DAQ continues file G22-88/R1_79
      background alpha

      DSSSD bias OK - attachment 8
       ambient temperature 16.8 deg C
      FEE64 temps OK - attachment 9
      ADC data item stats - attachment 10
       aida15 asic #2 u/s - rate c. 266k 

      aida11 1*L spectra - attachment 11

      ADC data items/s Merger, Tape Server & Merger stats - attachments 12-14
       data rate to disk c. 2Mb/s dominated by aida15 asic 2

10.12 per FEE64 Stat spectra - attachment 15
      missing channels aida03, aida04, aida05 & aida15
      Check ASIC load aida03, aida04, aida05 & aida15

13:25 Overnight leakage current plot - att 16


13.38 DAQ continues file G22-88/R1_92
      background alpha

      DSSSD bias OK - attachment 17
       ambient temperature 16.8 deg C
      FEE64 temps OK - attachment 18
      ADC data item stats - attachment 19
       aida15 asic #2 u/s - rate c. 266k 

      aida11 1*L spectra - attachment 20
       ASIC check load at 10.12 should result in 1-2 hits per channel in the Stat spectra aida03, aida04, aida05 and aida15
       Channels which continue to show zero counts likely indicate a FEE64-adaptor PCB connection issue
       Channels which now show 1, or more, counts cf. attachment 15 likely correspond to channels now working following the check ASIC load

      ADC data items/s Merger, Tape Server & Merger stats - attachments 21-23
       data rate to disk c. 2Mb/s dominated by aida15 asic 2

13.48 Stopped DAQ file G22-88/R1_93

      DSSSD bias OFF
      FEE64 power OFF


16.00 Installed FEE64 power cable to aida16
      Replaced aida15 with new FEE64 MAC 41:f6:5a

      Update dhcpd.conf
      Update MIDAS config to include aida16

      FEE64 power ON
      
      All system wide checks OK

      DSSSD bias OK - attachment 24

      FEE64 temperatures OK - attachment 25

      ADC data item stats - attachment 26
       aida15 rate changes from c. 260k to c. 25k ( 2x 'hot' channels, 1x channel missing, all other channels OK )

      per FEE64 Rate spectra - attachment 27

      per FEE64 1.8.L spectra - attachments 28-29
       pulser peak width aida11 44 ch FWHM, aida14 57 ch FWHM

      per FEE64 1.8.W spectra - 20us FSR - attachments 30-31

      
     Update NewMerger startup script and Options for 16x FEE64s - see attachments 32-33


16.20 file G22-88/R2
      Pulser 50Hz

16.28 DAQ stop

      Pulser OFF

      All histograms zero'd

16.33 DAQ starts file G22-88/R3
      alpha background

      ADC data item stats OK - attachment 34

      NewMerger, TapeServer and Merger stats - attachments 35-38

16.50 Pulser peak width spreadsheet for file R2 - attachment 39

20.32 DAQ continues file G22-88/R3_9

      DSSSD bias OK - attachment 40

      FEE64 temperatures OK - attachment 41

      ADC data item stats - attachment 42
       aida15 rate c. 25k ( 2x 'hot' channels, 1x channel missing, all other channels OK )

      per FEE64 Stat spectra - attachment 43

      Merger, Merger stats and TapServer - attachments 44-46

      Disc data items - attachment 47
       aida15 disc rate c. 8k
      Disable aida15 disc ( fast comparator ) outputs for asic#3 - attachment 48 
      Disc data items- attachment 49
       all FEE64s rate << 1Hz
    
     Update NewMerger startup script and Options for 16x FEE64s - see attachments 32-33
Attachment 1: Screenshot_from_2024-02-03_01-07-28.png
Screenshot_from_2024-02-03_01-07-28.png
Attachment 2: Screenshot_from_2024-02-03_01-08-15.png
Screenshot_from_2024-02-03_01-08-15.png
Attachment 3: Screenshot_from_2024-02-03_01-09-00.png
Screenshot_from_2024-02-03_01-09-00.png
Attachment 4: Screenshot_from_2024-02-03_01-09-39.png
Screenshot_from_2024-02-03_01-09-39.png
Attachment 5: Screenshot_from_2024-02-03_01-10-20.png
Screenshot_from_2024-02-03_01-10-20.png
Attachment 6: Screenshot_from_2024-02-03_01-11-45.png
Screenshot_from_2024-02-03_01-11-45.png
Attachment 7: Screenshot_from_2024-02-03_01-12-23.png
Screenshot_from_2024-02-03_01-12-23.png
Attachment 8: Screenshot_from_2024-02-03_10-02-22.png
Screenshot_from_2024-02-03_10-02-22.png
Attachment 9: Screenshot_from_2024-02-03_10-02-50.png
Screenshot_from_2024-02-03_10-02-50.png
Attachment 10: Screenshot_from_2024-02-03_10-03-29.png
Screenshot_from_2024-02-03_10-03-29.png
Attachment 11: Screenshot_from_2024-02-03_10-04-15.png
Screenshot_from_2024-02-03_10-04-15.png
Attachment 12: Screenshot_from_2024-02-03_10-04-46.png
Screenshot_from_2024-02-03_10-04-46.png
Attachment 13: Screenshot_from_2024-02-03_10-05-09.png
Screenshot_from_2024-02-03_10-05-09.png
Attachment 14: Screenshot_from_2024-02-03_10-05-48.png
Screenshot_from_2024-02-03_10-05-48.png
Attachment 15: Screenshot_from_2024-02-03_10-12-32.png
Screenshot_from_2024-02-03_10-12-32.png
Attachment 16: Screenshot_from_2024-02-03_13-24-46.png
Screenshot_from_2024-02-03_13-24-46.png
Attachment 17: Screenshot_from_2024-02-03_13-38-40.png
Screenshot_from_2024-02-03_13-38-40.png
Attachment 18: Screenshot_from_2024-02-03_13-38-47.png
Screenshot_from_2024-02-03_13-38-47.png
Attachment 19: Screenshot_from_2024-02-03_13-39-04.png
Screenshot_from_2024-02-03_13-39-04.png
Attachment 20: Screenshot_from_2024-02-03_13-40-06.png
Screenshot_from_2024-02-03_13-40-06.png
Attachment 21: Screenshot_from_2024-02-03_13-40-43.png
Screenshot_from_2024-02-03_13-40-43.png
Attachment 22: Screenshot_from_2024-02-03_13-41-00.png
Screenshot_from_2024-02-03_13-41-00.png
Attachment 23: Screenshot_from_2024-02-03_13-41-06.png
Screenshot_from_2024-02-03_13-41-06.png
Attachment 24: Screenshot_from_2024-02-03_16-01-18.png
Screenshot_from_2024-02-03_16-01-18.png
Attachment 25: Screenshot_from_2024-02-03_16-01-33.png
Screenshot_from_2024-02-03_16-01-33.png
Attachment 26: Screenshot_from_2024-02-03_16-01-42.png
Screenshot_from_2024-02-03_16-01-42.png
Attachment 27: Screenshot_from_2024-02-03_16-02-27.png
Screenshot_from_2024-02-03_16-02-27.png
Attachment 28: Screenshot_from_2024-02-03_16-03-16.png
Screenshot_from_2024-02-03_16-03-16.png
Attachment 29: Screenshot_from_2024-02-03_16-06-11.png
Screenshot_from_2024-02-03_16-06-11.png
Attachment 30: Screenshot_from_2024-02-03_16-07-05.png
Screenshot_from_2024-02-03_16-07-05.png
Attachment 31: Screenshot_from_2024-02-03_16-07-35.png
Screenshot_from_2024-02-03_16-07-35.png
Attachment 32: CONTENTS
0x0006dead string 0x0000
MERGE.LinksAvailable string 16
Index string MERGE.LinksAvailable&&0x0006dead&&MERGE.RunOptions&&MERGE.LinksInUse
Merge.LinksInUse string 1%1%1%1%1%1%1%1%1%1%1%1%1%1%1%1%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%
MERGE.LinksInUse string 1%1%1%1%1%1%1%1%1%1%1%1%1%1%1%1%
MERGE.RunOptions string 3
Attachment 33: NewMerger
#!/bin/csh -f 

echo "Tidy up"

killall -s9 MERGE
killall -s9 merge64.AD

killall -s9 master64
killall -s9 link64
killall -s9 merge64
killall -s9 statrate64

echo "Starting New Merger"

/bin/echo -e "\033];New Merger\007" 

  if (! $?MIDASBASE) setenv MIDASBASE /MIDAS

#
#

   setenv PATH ${MIDASBASE}/Linux/bin64:${PATH}
   setenv LD_LIBRARY_PATH ${MIDASBASE}/TclHttpd/Linux64:${MIDASBASE}/Linux/lib64:${MIDASBASE}/lib_Linux64

netint MERGE_SyncStep 0x0000000

cd ${MIDASBASE}/NewMerger/MergeServer/bin64
./master64 -i 16 -l 16 -p 11001  &

sleep 100000000000

#
echo startup complete
#




Attachment 34: Screenshot_from_2024-02-03_16-35-07.png
Screenshot_from_2024-02-03_16-35-07.png
Attachment 35: Screenshot_from_2024-02-03_16-35-20.png
Screenshot_from_2024-02-03_16-35-20.png
Attachment 36: Screenshot_from_2024-02-03_16-35-24.png
Screenshot_from_2024-02-03_16-35-24.png
Attachment 37: Screenshot_from_2024-02-03_16-35-31.png
Screenshot_from_2024-02-03_16-35-31.png
Attachment 38: Screenshot_from_2024-02-03_16-35-40.png
Screenshot_from_2024-02-03_16-35-40.png
Attachment 39: Pulser-widths-carme-2024.xlsx
Attachment 40: Screenshot_from_2024-02-03_20-32-07.png
Screenshot_from_2024-02-03_20-32-07.png
Attachment 41: Screenshot_from_2024-02-03_20-32-40.png
Screenshot_from_2024-02-03_20-32-40.png
Attachment 42: Screenshot_from_2024-02-03_20-33-01.png
Screenshot_from_2024-02-03_20-33-01.png
Attachment 43: Screenshot_from_2024-02-03_20-33-24.png
Screenshot_from_2024-02-03_20-33-24.png
Attachment 44: Screenshot_from_2024-02-03_20-34-17.png
Screenshot_from_2024-02-03_20-34-17.png
Attachment 45: Screenshot_from_2024-02-03_20-34-40.png
Screenshot_from_2024-02-03_20-34-40.png
Attachment 46: Screenshot_from_2024-02-03_20-35-02.png
Screenshot_from_2024-02-03_20-35-02.png
Attachment 47: Screenshot_from_2024-02-03_20-35-52.png
Screenshot_from_2024-02-03_20-35-52.png
Attachment 48: Screenshot_from_2024-02-03_20-36-55.png
Screenshot_from_2024-02-03_20-36-55.png
Attachment 49: Screenshot_from_2024-02-03_20-37-14.png
Screenshot_from_2024-02-03_20-37-14.png
Attachment 50: Screenshot_from_2024-02-04_09-55-02.png
Screenshot_from_2024-02-04_09-55-02.png
  487   Fri Feb 2 09:58:58 2024 TDTapeServer not writng data to disk - plus fix
DAQ and NewMerger appear to be cofigured correctly.

Can allocate device, create volume and open file with TapeServer but cannot write data.

See attachmnents 1-3




Noted aida09 WR timestamp issue in attachment 3 - FEE64 Reboot fixed aida09 WR timestamp issue

Restarted DAQ/Merger/TapeServer - still cannot write to disk


14.30 CU provides the following list of NetVar variables used by the Merger

"MERGE_PID"
"MERGE_ID"
"MERGE_Action"
"MERGE_Signal"
   "MERGE_Parameter"
   "MERGE_Response"
   "MERGE_State"
   "MERGE_Trace"
   "MERGE_Paused"
   "MERGE_RunOptions"
   "MERGE_HardwareVersion"
   "MERGE_LinksAvailable"
   "MERGE_TestMode"
   "MERGE_PollTicks"
   "MERGE_LinksInUse"
   "MERGE_LinksAlive"
   "MERGE_StatsTime"
   "MERGE_Command"
   "MERGE_ZeroStats"
   "MERGE_SyncStep"
   "MERGE_InitialTimeStamp"
"MERGE_LinkState"
"MERGE_LinkStatus"
"MERGE_LinkAlive"
"MERGE_LinkInUse"

15.30 Changed Merger startup and NetVars files ( attachments 6 & 7 )

      Change server command line from

./master64 -i 16 -l 16 -p 11001  &

to

./master64 -i 15 -l 15 -p 11001  &


i.e. *exactly* matches number of FEE64s currently in use ( 15x FEE64s, aida01-aida15 )

      Change NetVars from 

MERGE.LinksInUse string 1%1%1%1%1%1%1%1%1%1%1%1%1%1%1%0%

to

MERGE.LinksInUse string 1%1%1%1%1%1%1%1%1%1%1%1%1%1%1%

i.e. string has 15x 1% strings *not* 15x 1% *plus 1x 0% string 

With *both* changes we are now able to write to disk.

N.B. We did not test the above changes individually - it is possible/probable that just one of the changes above fixes the problem.
Attachment 1: Screenshot_from_2024-02-02_10-57-18.png
Screenshot_from_2024-02-02_10-57-18.png
Attachment 2: Screenshot_from_2024-02-02_10-57-51.png
Screenshot_from_2024-02-02_10-57-51.png
Attachment 3: Screenshot_from_2024-02-02_10-58-23.png
Screenshot_from_2024-02-02_10-58-23.png
Attachment 4: Screenshot_from_2024-02-02_11-01-22.png
Screenshot_from_2024-02-02_11-01-22.png
Attachment 5: Screenshot_from_2024-02-02_11-18-45.png
Screenshot_from_2024-02-02_11-18-45.png
Attachment 6: NewMerger
#!/bin/csh -f 

echo "Tidy up"

killall -s9 MERGE
killall -s9 merge64.AD

killall -s9 master64
killall -s9 link64
killall -s9 merge64
killall -s9 statrate64

echo "Starting New Merger"

/bin/echo -e "\033];New Merger\007" 

  if (! $?MIDASBASE) setenv MIDASBASE /MIDAS

#
#

   setenv PATH ${MIDASBASE}/Linux/bin64:${PATH}
   setenv LD_LIBRARY_PATH ${MIDASBASE}/TclHttpd/Linux64:${MIDASBASE}/Linux/lib64:${MIDASBASE}/lib_Linux64

netint MERGE_SyncStep 0x0000000

cd ${MIDASBASE}/NewMerger/MergeServer/bin64
./master64 -i 15 -l 15 -p 11001  &

sleep 100000000000

#
echo startup complete
#




Attachment 7: CONTENTS
0x0006dead string 0x0000
MERGE.LinksAvailable string 15
Index string MERGE.LinksAvailable&&0x0006dead&&MERGE.RunOptions&&MERGE.LinksInUse
Merge.LinksInUse string 1%1%1%1%1%1%1%1%1%1%1%1%1%1%1%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%
MERGE.LinksInUse string 1%1%1%1%1%1%1%1%1%1%1%1%1%1%1%
MERGE.RunOptions string 3
  486   Fri Feb 2 08:59:36 2024 JM, TD, CBRun1 start - pulser widths spreadsheet
Current pulser widths excel spreadsheet attached

09.58 DSSSD bias OK - attachment 1 

Ambient temperature +16.1 deg C FEE64 temps OK - attachment 2 

ADC data item stats OK - attachment 3 
 aida15 asic #2 high rate - asic u/s

system wide tests OK *except* aida09 - attachments 4-5 
 aida09 WR status 0x20 aida09 WR timestamp incorrect cf. all other FEE64s

per FEE64 Rate spectra - attachment 6

per FEE64 1.8.L spectra - attachments 7-8 
 aida14 pulse peak witdh 58 ch FWHM
 aida11 pulse peak witdh 48 ch FWHM

per FEE64 1.8.W spetra - 20us FSR - attachments 9-10
 

12.00 

aida09 WR timestamp issue resolved by FEE64 reboot 


14.10 

Safety pins removed from detector motion system. CB performs one, complete DSSSD movement cycle. Zero all histogram and stats per FEE64 Stat spectra ( log scale ) and 
ADC 
data item stats for the complete DSSSD movement cycle - attachments 11-12 
No change observed in DSSSD leakage currents during DSSSD movement cycle. 

15.35 

Resolved issue from previous elog. Begin background run writing data to disk.
ASIC check load 
Start DAQ G22-88/R1 background alpha 
EG&G Ortec 448 Pulser OFF 
ASIC settings 2024Jan20-12.02.08 
slow comparator 0x64 ( all FEE64s *except aida15 asic #2 0xff ) 
All histograms set to zero

DSSSD bias OK - attachment 13 
FEE64 temps OK - attachment 14 

WR timestamps OK - attachment 15 

ADC data item stats - attachment 16 
aida15 asic #2 u/s - rate c. 250k 

ADC data items/s Merger, Tape Server & Merger stats - attachments 17-19 
 data rate to disk c. 2Mb/s dominated by aida15 asic 2
Attachment 1: Screenshot_from_2024-02-02_09-59-10.png
Screenshot_from_2024-02-02_09-59-10.png
Attachment 2: Screenshot_from_2024-02-02_10-10-46.png
Screenshot_from_2024-02-02_10-10-46.png
Attachment 3: Screenshot_from_2024-02-02_10-10-59.png
Screenshot_from_2024-02-02_10-10-59.png
Attachment 4: Screenshot_from_2024-02-02_10-11-18.png
Screenshot_from_2024-02-02_10-11-18.png
Attachment 5: Screenshot_from_2024-02-02_10-11-37.png
Screenshot_from_2024-02-02_10-11-37.png
Attachment 6: Screenshot_from_2024-02-02_10-12-32.png
Screenshot_from_2024-02-02_10-12-32.png
Attachment 7: Screenshot_from_2024-02-02_10-13-17.png
Screenshot_from_2024-02-02_10-13-17.png
Attachment 8: Screenshot_from_2024-02-02_10-15-11.png
Screenshot_from_2024-02-02_10-15-11.png
Attachment 9: Screenshot_from_2024-02-02_10-16-33.png
Screenshot_from_2024-02-02_10-16-33.png
Attachment 10: Screenshot_from_2024-02-02_10-18-01.png
Screenshot_from_2024-02-02_10-18-01.png
Attachment 11: Screenshot_from_2024-02-02_14-09-42.png
Screenshot_from_2024-02-02_14-09-42.png
Attachment 12: Screenshot_from_2024-02-02_14-09-46.png
Screenshot_from_2024-02-02_14-09-46.png
Attachment 13: Screenshot_from_2024-02-02_15-41-14.png
Screenshot_from_2024-02-02_15-41-14.png
Attachment 14: Screenshot_from_2024-02-02_15-41-35.png
Screenshot_from_2024-02-02_15-41-35.png
Attachment 15: Screenshot_from_2024-02-02_15-42-14.png
Screenshot_from_2024-02-02_15-42-14.png
Attachment 16: Screenshot_from_2024-02-02_15-42-31.png
Screenshot_from_2024-02-02_15-42-31.png
Attachment 17: Screenshot_from_2024-02-02_15-43-02.png
Screenshot_from_2024-02-02_15-43-02.png
Attachment 18: Screenshot_from_2024-02-02_15-43-11.png
Screenshot_from_2024-02-02_15-43-11.png
Attachment 19: Screenshot_from_2024-02-02_15-43-25.png
Screenshot_from_2024-02-02_15-43-25.png
Attachment 20: Pulser-widths-carme-2024.xlsx
  485   Thu Feb 1 14:45:09 2024 JM, CBNewMerger setup

To include alll 16 FEE modules in the NewMerger we edited the file /MIDAS/DB/EXPERIMENTS/MERGE/Options/carme-gsi from MERGE.LinksAvailable string 5 to MERGE.LinksAvailable string 15

We also edited the file /MIDAS/Linux/startup/NewMerger from ./master64 -i 8 -l 16 -p 11001  & to ./master64 -i 16 -l 16 -p 11001  &

We have started the Merger and tape server. The merger settings were 'corrupted' and did not display xfer Links => Merger. We implemented a fix from https://elog.ph.ed.ac.uk/DESPEC/36 and this resolved the issue.

We started the tape server, merger and go on the run control. We observe the merger working and data links to the tape server. A new file is created each time but we observe no data being written to that file. After 5 minutes of no data being written the merger crashed and the fees crashed soon after. We will leave until tomorrow to resolve.

Issue subsequently resolved see elog https://elog.ph.ed.ac.uk/CARME/487

  484   Thu Feb 1 10:04:11 2024 JM, CBShaping time testing

Arrived at the control room to find we had lost connection to the power relay. We rebooted the nnrpi1 and re-connected according to the steps laid out in elog 471 -> all okay

11:00 Restarted daq with no changes to the system from yesterday. Bias consistent with previous days. All pulser peaks are slightly improved (even more for the pn pulser, one goes from 91 chn to 78). No pulser observed in nn FEE modules 9, 10, 13, 14 again.

ADC data items check OK - all ~3k

Rates - all OK  - attachment 1

1.8L spectra. - attachment 2,3

1.8W waveforms - attachment 4,5

11:20 Changed shaping time from 0xf to 0x3 (2us)

ADC data items check OK - all ~3k attachment 6

Rates - all OK  - attachment 7

1.8L pn spectra. - attachment 8

1.8W waveforms - attachment 9,10

11:35 Changed shaping time from 0x3 to 0x7 (4us)

ADC data items check OK - all ~3k attachment 11

Rates - all OK  - attachment 12

1.8L pn spectra. - attachment 13

1.8W waveforms - attachment 14,15

11:45 Changed shaping time from 0x7 to 0xb (6us)

ADC data items check OK - all ~3k - attachment 16

Rates - all OK  - attachment 17

1.8L spectra. - attachment 18,19

1.8W waveforms - attachment 20,21

11:59 Changed shaping time from 0xb to 0xd (7us)

ADC data items check OK - all ~3k - attchment 22

Rates - all OK  - attachment 23

1.8L pn spectra. - attachment 24

1.8W waveforms - attachment 25, 26

12:09 Changed shaping time back to 0xf (8us)

ADC data items check OK - all ~3k - attchment 27

Rates - all OK  - attachment 28

1.8L pn spectra. - attachment 29

1.8W waveforms - attachment 30,31

13:10 Power cycled FEE modules. nn pulser for FEEs 9,10,13,14 are now active again (no changes made in the cave) Pulser widths for nn and pn  both consistent with before lunch

 

Attachment 1: Screenshot_from_2024-02-01_10-55-39.png
Screenshot_from_2024-02-01_10-55-39.png
Attachment 2: Screenshot_from_2024-02-01_11-01-55.png
Screenshot_from_2024-02-01_11-01-55.png
Attachment 3: Screenshot_from_2024-02-01_11-04-08.png
Screenshot_from_2024-02-01_11-04-08.png
Attachment 4: Screenshot_from_2024-02-01_11-06-41.png
Screenshot_from_2024-02-01_11-06-41.png
Attachment 5: Screenshot_from_2024-02-01_11-07-25.png
Screenshot_from_2024-02-01_11-07-25.png
Attachment 6: Screenshot_from_2024-02-01_11-21-46.png
Screenshot_from_2024-02-01_11-21-46.png
Attachment 7: Screenshot_from_2024-02-01_11-24-21.png
Screenshot_from_2024-02-01_11-24-21.png
Attachment 8: Screenshot_from_2024-02-01_11-25-17.png
Screenshot_from_2024-02-01_11-25-17.png
Attachment 9: Screenshot_from_2024-02-01_11-29-19.png
Screenshot_from_2024-02-01_11-29-19.png
Attachment 10: Screenshot_from_2024-02-01_11-29-58.png
Screenshot_from_2024-02-01_11-29-58.png
Attachment 11: Screenshot_from_2024-02-01_11-34-19.png
Screenshot_from_2024-02-01_11-34-19.png
Attachment 12: Screenshot_from_2024-02-01_11-36-16.png
Screenshot_from_2024-02-01_11-36-16.png
Attachment 13: Screenshot_from_2024-02-01_11-37-15.png
Screenshot_from_2024-02-01_11-37-15.png
Attachment 14: Screenshot_from_2024-02-01_11-41-14.png
Screenshot_from_2024-02-01_11-41-14.png
Attachment 15: Screenshot_from_2024-02-01_11-41-50.png
Screenshot_from_2024-02-01_11-41-50.png
Attachment 16: Screenshot_from_2024-02-01_11-45-20.png
Screenshot_from_2024-02-01_11-45-20.png
Attachment 17: Screenshot_from_2024-02-01_11-46-03.png
Screenshot_from_2024-02-01_11-46-03.png
Attachment 18: Screenshot_from_2024-02-01_11-46-56.png
Screenshot_from_2024-02-01_11-46-56.png
Attachment 19: Screenshot_from_2024-02-01_11-50-59.png
Screenshot_from_2024-02-01_11-50-59.png
Attachment 20: Screenshot_from_2024-02-01_11-52-32.png
Screenshot_from_2024-02-01_11-52-32.png
Attachment 21: Screenshot_from_2024-02-01_11-53-17.png
Screenshot_from_2024-02-01_11-53-17.png
Attachment 22: Screenshot_from_2024-02-01_11-59-19.png
Screenshot_from_2024-02-01_11-59-19.png
Attachment 23: Screenshot_from_2024-02-01_12-00-18.png
Screenshot_from_2024-02-01_12-00-18.png
Attachment 24: Screenshot_from_2024-02-01_12-01-00.png
Screenshot_from_2024-02-01_12-01-00.png
Attachment 25: Screenshot_from_2024-02-01_12-07-16.png
Screenshot_from_2024-02-01_12-07-16.png
Attachment 26: Screenshot_from_2024-02-01_12-07-48.png
Screenshot_from_2024-02-01_12-07-48.png
Attachment 27: Screenshot_from_2024-02-01_12-10-03.png
Screenshot_from_2024-02-01_12-10-03.png
Attachment 28: Screenshot_from_2024-02-01_12-10-47.png
Screenshot_from_2024-02-01_12-10-47.png
Attachment 29: Screenshot_from_2024-02-01_12-12-15.png
Screenshot_from_2024-02-01_12-12-15.png
Attachment 30: Screenshot_from_2024-02-01_12-17-28.png
Screenshot_from_2024-02-01_12-17-28.png
Attachment 31: Screenshot_from_2024-02-01_12-18-20.png
Screenshot_from_2024-02-01_12-18-20.png
  483   Wed Jan 31 09:49:38 2024 JM, TDOptimising - Wednesday 31

Removed ground from DSSD #1 to the chamber and set MACBs so that timestamp is from the NIM WR emulator not the VME VETAR2.

Root MACB mode 0x3 -> 0xd - all other MACB modes 0x3 ( unchanged ).

No timestamps observed from the emulator. Re-set to take timestamps from the VETAR" unit -> all timestamps observed.

10:30

No pulser observed in nn FEE modules 9, 10, 13, 14.

ADC data items check OK - all ~3k - attachment 2

Rates - all OK  - attachment 3

1.8L spectra. - attachment 4,5

1.8W waveforms - attachment 6,7

11:00

Checked daisy chain which goes from 8->14->13->10->9. Replaced the lemo cable from 8->14 and pulsers re-appear. Rates all OK except aida10 which has now disappeared, tried asic check and reset and startup again no change in aida10.

Now majority of pulsers are observable the widths will be noted in a spreadsheet

After replacing lemo for nn pulser chain, all nn pulser widths decreased slightly ~10% for all nn FEEs, little change in pn pulser widths but n pulser for DSSD#1,2 appears more gaussian

11:44

FEE64s power cycled. aida10 now prioducing data.

Removed internal jumper from CAEN N1419ET 

ADC data items check OK - all ~3k - attachment 8

Rates - all OK  - attachment 9

1.8L spectra - attachment 10,11 little change in pn pulser widths but nn pulser widths are slightly broader.

1.8W waveforms - attachment 12,13

15:50 Jumper re-mounted onto caen bias module

ADC data items check OK - all ~3k - attachment 14

Rates - all OK  - attachment 15

1.8L spectra - attachment 16,17

1.8W waveforms - attachment 18,19

16:45 Ground cables attached from water pipe (chamber ground) to lemo cable on aida08 (nn), aida13 (nn), aida09 (nn), aida01 (pn) so that all DSSDs are connected to chamber ground.

ADC data items check OK - all ~3k - attachment 20

Rates - all OK  - attachment 21

1.8L spectra - attachment 22,23

1.8W waveforms - attachment 24,25

 

Attachment 1: Screenshot_from_2024-01-31_10-30-51.png
Screenshot_from_2024-01-31_10-30-51.png
Attachment 2: Screenshot_from_2024-01-31_10-30-32.png
Screenshot_from_2024-01-31_10-30-32.png
Attachment 3: Screenshot_from_2024-01-31_10-35-29.png
Screenshot_from_2024-01-31_10-35-29.png
Attachment 4: Screenshot_from_2024-01-31_10-36-25.png
Screenshot_from_2024-01-31_10-36-25.png
Attachment 5: Screenshot_from_2024-01-31_10-40-38.png
Screenshot_from_2024-01-31_10-40-38.png
Attachment 6: Screenshot_from_2024-01-31_10-50-09.png
Screenshot_from_2024-01-31_10-50-09.png
Attachment 7: Screenshot_from_2024-01-31_10-51-15.png
Screenshot_from_2024-01-31_10-51-15.png
Attachment 8: Screenshot_from_2024-01-31_12-13-56.png
Screenshot_from_2024-01-31_12-13-56.png
Attachment 9: Screenshot_from_2024-01-31_12-15-41.png
Screenshot_from_2024-01-31_12-15-41.png
Attachment 10: Screenshot_from_2024-01-31_12-16-54.png
Screenshot_from_2024-01-31_12-16-54.png
Attachment 11: Screenshot_from_2024-01-31_12-21-42.png
Screenshot_from_2024-01-31_12-21-42.png
Attachment 12: Screenshot_from_2024-01-31_12-26-28.png
Screenshot_from_2024-01-31_12-26-28.png
Attachment 13: Screenshot_from_2024-01-31_12-27-33.png
Screenshot_from_2024-01-31_12-27-33.png
Attachment 14: Screenshot_from_2024-01-31_16-12-15.png
Screenshot_from_2024-01-31_16-12-15.png
Attachment 15: Screenshot_from_2024-01-31_16-14-25.png
Screenshot_from_2024-01-31_16-14-25.png
Attachment 16: Screenshot_from_2024-01-31_16-15-28.png
Screenshot_from_2024-01-31_16-15-28.png
Attachment 17: Screenshot_from_2024-01-31_16-22-14.png
Screenshot_from_2024-01-31_16-22-14.png
Attachment 18: Screenshot_from_2024-01-31_16-27-15.png
Screenshot_from_2024-01-31_16-27-15.png
Attachment 19: Screenshot_from_2024-01-31_16-28-06.png
Screenshot_from_2024-01-31_16-28-06.png
Attachment 20: Screenshot_from_2024-01-31_17-03-14.png
Screenshot_from_2024-01-31_17-03-14.png
Attachment 21: Screenshot_from_2024-01-31_17-05-33.png
Screenshot_from_2024-01-31_17-05-33.png
Attachment 22: Screenshot_from_2024-01-31_17-06-37.png
Screenshot_from_2024-01-31_17-06-37.png
Attachment 23: Screenshot_from_2024-01-31_17-09-40.png
Screenshot_from_2024-01-31_17-09-40.png
Attachment 24: Screenshot_from_2024-01-31_17-13-54.png
Screenshot_from_2024-01-31_17-13-54.png
Attachment 25: Screenshot_from_2024-01-31_17-14-32.png
Screenshot_from_2024-01-31_17-14-32.png
  482   Tue Jan 30 14:04:30 2024 NH[HowTo] Update MACB Firmware
This is how to update the MACB firmware using the opensource tool xc3sprog


Setup (only has to be done first time)
--------------------
Install xc3sprog and fxload (ubuntu: sudo apt install xc3sprog fxload)

To configure the USB programmer:
Download xilinx.tar.gz and put the hex files in /usr/share

Download the udev rules and put it in /etc/udev/rules.d

Run:
sudo udevadm control --reload-rules

[Taken from https://github.com/timvideos/litex-buildenv/wiki/Xilinx-Platform-Cable-USB-under-Linux]

Plug in programmer and should work

----------------------------------
Programming a MACB

Open MACB side panel and connect JTAG programmer to port 

Plug MACB into NIM bin and power NIM bin on

Run
xc3sprog -c xpc -m -j

Expected output:
XC3SPROG (c) 2004-2011 xc3sprog project $Rev$ OS: Linux
Free software: If you contribute nothing, expect nothing!
Feedback on success/failure/enhancement requests:
        http://sourceforge.net/mail/?group_id=170565 
Check Sourceforge for updates:
        http://sourceforge.net/projects/xc3sprog/develop

JTAG loc.:   0  IDCODE: 0x16d8c093  Desc:                  XC2C128_TQ144 Rev: A  IR length:  8

This confirms the JTAG is working

Then to program  run

xc3sprog -c xpc -m . macb_apr20.jed

The directory should contain the macb_apr20.jed and the xc2c128.map files

Expected output:
XC3SPROG (c) 2004-2011 xc3sprog project $Rev$ OS: Linux
Free software: If you contribute nothing, expect nothing!
Feedback on success/failure/enhancement requests:
        http://sourceforge.net/mail/?group_id=170565 
Check Sourceforge for updates:
        http://sourceforge.net/projects/xc3sprog/develop

Programming row  81                                        
Verify: Success   

Now power off the NIM bin and remove the programmer and put the cover back on
Attachment 1: xusbdfwu.rules
# version 0003
ATTRS{idVendor}=="03fd", ATTRS{idProduct}=="0008", MODE="666"
SUBSYSTEMS=="usb", ACTION=="add", ATTRS{idVendor}=="03fd", ATTRS{idProduct}=="0007", RUN+="/sbin/fxload -v -t fx2 -I /usr/share/xusbdfwu.hex -D $tempnode"
SUBSYSTEMS=="usb", ACTION=="add", ATTRS{idVendor}=="03fd", ATTRS{idProduct}=="0009", RUN+="/sbin/fxload -v -t fx2 -I /usr/share/xusb_xup.hex -D $tempnode"
SUBSYSTEMS=="usb", ACTION=="add", ATTRS{idVendor}=="03fd", ATTRS{idProduct}=="000d", RUN+="/sbin/fxload -v -t fx2 -I /usr/share/xusb_emb.hex -D $tempnode"
SUBSYSTEMS=="usb", ACTION=="add", ATTRS{idVendor}=="03fd", ATTRS{idProduct}=="000f", RUN+="/sbin/fxload -v -t fx2 -I /usr/share/xusb_xlp.hex -D $tempnode"
SUBSYSTEMS=="usb", ACTION=="add", ATTRS{idVendor}=="03fd", ATTRS{idProduct}=="0013", RUN+="/sbin/fxload -v -t fx2 -I /usr/share/xusb_xp2.hex -D $tempnode"
SUBSYSTEMS=="usb", ACTION=="add", ATTRS{idVendor}=="03fd", ATTRS{idProduct}=="0015", RUN+="/sbin/fxload -v -t fx2 -I /usr/share/xusb_xse.hex -D $tempnode"
Attachment 2: xilinx.tar.gz
Attachment 3: xc2c128.map
																																																																																	user
6507	6523		6536	6552	6565	6581		6594	6610	6623	6639		6652	6668		6684			6700		6716			6732	6745	6761		6774	6790	6803	6819		6832	6848	6861	6877		6890	6906	20323	20339		20352	20368	20381	20397		20410	20426	20439	20455		20468	20484	20497	20513			20529		20545			20561	20574	20590	55324	20603	20619	20632	20648		20661	20677	20690	20706		20719	20735	sec_0	
6508	6524		6537	6553	6566	6582		6595	6611	6624	6640		6653	6669		6685			6701		6717			6733	6746	6762		6775	6791	6804	6820		6833	6849	6862	6878		6891	6907	20324	20340		20353	20369	20382	20398		20411	20427	20440	20456		20469	20485	20498	20514			20530		20546			20562	20575	20591	55325	20604	20620	20633	20649		20662	20678	20691	20707		20720	20736	sec_1	
6502	6509		6531	6538	6560	6567		6589	6596	6618	6625		6647	6654	6675			6691		6707			6723		6740	6747		6769	6776	6798	6805		6827	6834	6856	6863		6885	6892	20318	20325		20347	20354	20376	20383		20405	20412	20434	20441		20463	20470	20492	20499		20520		20536			20552		20569	20576		20598	20605	20627	20634		20656	20663	20685	20692		20714	20721	sec_2	
6503	6510		6532	6539	6561	6568		6590	6597	6619	6626		6648	6655	6676			6692		6708			6724		6741	6748	55326	6770	6777	6799	6806		6828	6835	6857	6864		6886	6893	20319	20326		20348	20355	20377	20384		20406	20413	20435	20442		20464	20471	20493	20500		20521		20537			20553		20570	20577	55330	20599	20606	20628	20635		20657	20664	20686	20693		20715	20722	sec_3	
6501	6511		6530	6540	6559	6569		6588	6598	6617	6627		6646	6656											6739	6749	55327	6768	6778	6797	6807		6826	6836	6855	6865		6884	6894	20317	20327		20346	20356	20375	20385		20404	20414	20433	20443		20462	20472	20491	20501									20568	20578	55331	20597	20607	20626	20636		20655	20665	20684	20694		20713	20723	sec_4	
6504	6512		6533	6541	6562	6570		6591	6599	6620	6628		6649	6657											6742	6750	55328	6771	6779	6800	6808		6829	6837	6858	6866		6887	6895	20320	20328		20349	20357	20378	20386		20407	20415	20436	20444		20465	20473	20494	20502									20571	20579	55332	20600	20608	20629	20637		20658	20666	20687	20695		20716	20724	sec_5	
6505	6521		6534	6550	6563	6579		6592	6608	6621	6637		6650	6666											6743	6759	55329	6772	6788	6801	6817		6830	6846	6859	6875		6888	6904	20321	20337		20350	20366	20379	20395		20408	20424	20437	20453		20466	20482	20495	20511									20572	20588	55333	20601	20617	20630	20646		20659	20675	20688	20704		20717	20733	sec_6	
6522			6551		6580			6609		6638			6667												6760			6789		6818			6847		6876			6905		20338			20367		20396			20425		20454			20483		20512										20589		55336	20618		20647			20676		20705			20734		done_0	
6496	6516		6525	6545	6554	6574		6583	6603	6612	6632		6641	6661	6670			6686		6702			6718		6734	6754		6763	6783	6792	6812		6821	6841	6850	6870		6879	6899	20312	20332		20341	20361	20370	20390		20399	20419	20428	20448		20457	20477	20486	20506		20515		20531			20547		20563	20583	55338	20592	20612	20621	20641		20650	20670	20679	20699		20708	20728	done_1	
6497	6517		6526	6546	6555	6575		6584	6604	6613	6633		6642	6662	6671	6680		6687	6696	6703	6712		6719	6728	6735	6755		6764	6784	6793	6813		6822	6842	6851	6871		6880	6900	20313	20333		20342	20362	20371	20391		20400	20420	20429	20449		20458	20478	20487	20507		20516	20525	20532	20541		20548	20557	20564	20584	55340	20593	20613	20622	20642		20651	20671	20680	20700		20709	20729		
6498	6518		6527	6547	6556	6576		6585	6605	6614	6634		6643	6663	6672	6681		6688	6697	6704	6713		6720	6729	6736	6756		6765	6785	6794	6814		6823	6843	6852	6872		6881	6901	20314	20334		20343	20363	20372	20392		20401	20421	20430	20450		20459	20479	20488	20508		20517	20526	20533	20542		20549	20558	20565	20585		20594	20614	20623	20643		20652	20672	20681	20701		20710	20730		
6499	6519		6528	6548	6557	6577		6586	6606	6615	6635		6644	6664	6673	6682		6689	6698	6705	6714		6721	6730	6737	6757		6766	6786	6795	6815		6824	6844	6853	6873		6882	6902	20315	20335		20344	20364	20373	20393		20402	20422	20431	20451		20460	20480	20489	20509		20518	20527	20534	20543		20550	20559	20566	20586		20595	20615	20624	20644		20653	20673	20682	20702		20711	20731		
6500	6520		6529	6549	6558	6578		6587	6607	6616	6636		6645	6665	6674	6683		6690	6699	6706	6715		6722	6731	6738	6758		6767	6787	6796	6816		6825	6845	6854	6874		6883	6903	20316	20336		20345	20365	20374	20394		20403	20423	20432	20452		20461	20481	20490	20510		20519	20528	20535	20544		20551	20560	20567	20587		20596	20616	20625	20645		20654	20674	20683	20703		20712	20732		
6506	6513		6535	6542	6564	6571		6593	6600	6622	6629		6651	6658		6677			6693		6709			6725	6744	6751		6773	6780	6802	6809		6831	6838	6860	6867		6889	6896	20322	20329		20351	20358	20380	20387		20409	20416	20438	20445		20467	20474	20496	20503			20522		20538			20554	20573	20580		20602	20609	20631	20638		20660	20667	20689	20696		20718	20725		
6515	6514		6544	6543	6573	6572		6602	6601	6631	6630		6660	6659	6679	6678		6695	6694	6711	6710		6727	6726	6753	6752		6782	6781	6811	6810		6840	6839	6869	6868		6898	6897	20331	20330		20360	20359	20389	20388		20418	20417	20447	20446		20476	20475	20505	20504		20524	20523	20540	20539		20556	20555	20582	20581		20611	20610	20640	20639		20669	20668	20698	20697		20727	20726		
5728	5760		5792	5824	5856		5888	5920			5952	5984	6016		6048	6080	5600		5632	5696		5664	6128	6160		6192	6224	6256			6288	6320		6352	6384	6416		6448	6480	19544	19576		19608	19640	19672		19704	19736			19768	19800	19832		19864	19896	19416		19448	19512		19480	19944	19976		20008	20040	20072			20104	20136		20168	20200	20232		20264	20296		
5744	5776		5808	5840	5872		5904	5936			5968	6000	6032		6064	6096	5616		5648	5712		5680	6112	6144		6176	6208	6240			6272	6304		6336	6368	6400		6432	6464	19560	19592		19624	19656	19688		19720	19752			19784	19816	19848		19880	19912	19432		19464	19528		19496	19928	19960		19992	20024	20056			20088	20120		20152	20184	20216		20248	20280		
5729	5761		5793	5825	5857		5889	5921			5953	5985	6017		6049	6081	5601		5633	5697		5665	6129	6161		6193	6225	6257			6289	6321		6353	6385	6417		6449	6481	19545	19577		19609	19641	19673		19705	19737			19769	19801	19833		19865	19897	19417		19449	19513		19481	19945	19977		20009	20041	20073			20105	20137		20169	20201	20233		20265	20297		
5745	5777		5809	5841	5873		5905	5937			5969	6001	6033		6065	6097	5617		5649	5713		5681	6113	6145		6177	6209	6241			6273	6305		6337	6369	6401		6433	6465	19561	19593		19625	19657	19689		19721	19753			19785	19817	19849		19881	19913	19433		19465	19529		19497	19929	19961		19993	20025	20057			20089	20121		20153	20185	20217		20249	20281		
5730	5762		5794	5826	5858		5890	5922			5954	5986	6018		6050	6082	5602		5634	5698		5666	6130	6162		6194	6226	6258			6290	6322		6354	6386	6418		6450	6482	19546	19578		19610	19642	19674		19706	19738			19770	19802	19834		19866	19898	19418		19450	19514		19482	19946	19978		20010	20042	20074			20106	20138		20170	20202	20234		20266	20298		
5746	5778		5810	5842	5874		5906	5938			5970	6002	6034		6066	6098	5618		5650	5714		5682	6114	6146		6178	6210	6242			6274	6306		6338	6370	6402		6434	6466	19562	19594		19626	19658	19690		19722	19754			19786	19818	19850		19882	19914	19434		19466	19530		19498	19930	19962		19994	20026	20058			20090	20122		20154	20186	20218		20250	20282		
5731	5763		5795	5827	5859		5891	5923			5955	5987	6019		6051	6083	5603		5635	5699		5667	6131	6163		6195	6227	6259			6291	6323		6355	6387	6419		6451	6483	19547	19579		19611	19643	19675		19707	19739			19771	19803	19835		19867	19899	19419		19451	19515		19483	19947	19979		20011	20043	20075			20107	20139		20171	20203	20235		20267	20299		
5747	5779		5811	5843	5875		5907	5939			5971	6003	6035		6067	6099	5619		5651	5715		5683	6115	6147		6179	6211	6243			6275	6307		6339	6371	6403		6435	6467	19563	19595		19627	19659	19691		19723	19755			19787	19819	19851		19883	19915	19435		19467	19531		19499	19931	19963		19995	20027	20059			20091	20123		20155	20187	20219		20251	20283		
5732	5764		5796	5828	5860		5892	5924			5956	5988	6020		6052	6084	5604		5636	5700		5668	6132	6164		6196	6228	6260			6292	6324		6356	6388	6420		6452	6484	19548	19580		19612	19644	19676		19708	19740			19772	19804	19836		19868	19900	19420		19452	19516		19484	19948	19980		20012	20044	20076			20108	20140		20172	20204	20236		20268	20300		
5748	5780		5812	5844	5876		5908	5940			5972	6004	6036		6068	6100	5620		5652	5716		5684	6116	6148		6180	6212	6244			6276	6308		6340	6372	6404		6436	6468	19564	19596		19628	19660	19692		19724	19756			19788	19820	19852		19884	19916	19436		19468	19532		19500	19932	19964		19996	20028	20060			20092	20124		20156	20188	20220		20252	20284		
5733	5765		5797	5829	5861		5893	5925			5957	5989	6021		6053	6085	5605		5637	5701		5669	6133	6165		6197	6229	6261			6293	6325		6357	6389	6421		6453	6485	19549	19581		19613	19645	19677		19709	19741			19773	19805	19837		19869	19901	19421		19453	19517		19485	19949	19981		20013	20045	20077			20109	20141		20173	20205	20237		20269	20301		
5749	5781		5813	5845	5877		5909	5941			5973	6005	6037		6069	6101	5621		5653	5717		5685	6117	6149		6181	6213	6245			6277	6309		6341	6373	6405		6437	6469	19565	19597		19629	19661	19693		19725	19757			19789	19821	19853		19885	19917	19437		19469	19533		19501	19933	19965		19997	20029	20061			20093	20125		20157	20189	20221		20253	20285		
5734	5766		5798	5830	5862		5894	5926			5958	5990	6022		6054	6086	5606		5638	5702		5670	6134	6166		6198	6230	6262			6294	6326		6358	6390	6422		6454	6486	19550	19582		19614	19646	19678		19710	19742			19774	19806	19838		19870	19902	19422		19454	19518		19486	19950	19982		20014	20046	20078			20110	20142		20174	20206	20238		20270	20302		
5750	5782		5814	5846	5878		5910	5942			5974	6006	6038		6070	6102	5622		5654	5718		5686	6118	6150		6182	6214	6246			6278	6310		6342	6374	6406		6438	6470	19566	19598		19630	19662	19694		19726	19758			19790	19822	19854		19886	19918	19438		19470	19534		19502	19934	19966		19998	20030	20062			20094	20126		20158	20190	20222		20254	20286		
5735	5767		5799	5831	5863		5895	5927			5959	5991	6023		6055	6087	5607		5639	5703		5671	6135	6167		6199	6231	6263			6295	6327		6359	6391	6423		6455	6487	19551	19583		19615	19647	19679		19711	19743			19775	19807	19839		19871	19903	19423		19455	19519		19487	19951	19983		20015	20047	20079			20111	20143		20175	20207	20239		20271	20303		
5751	5783		5815	5847	5879		5911	5943			5975	6007	6039		6071	6103	5623		5655	5719		5687	6119	6151		6183	6215	6247			6279	6311		6343	6375	6407		6439	6471	19567	19599		19631	19663	19695		19727	19759			19791	19823	19855		19887	19919	19439		19471	19535		19503	19935	19967		19999	20031	20063			20095	20127		20159	20191	20223		20255	20287		
5736	5768		5800	5832	5864		5896	5928			5960	5992	6024		6056	6088	5608		5640	5704		5672	6136	6168		6200	6232	6264			6296	6328		6360	6392	6424		6456	6488	19552	19584		19616	19648	19680		19712	19744			19776	19808	19840		19872	19904	19424		19456	19520		19488	19952	19984		20016	20048	20080			20112	20144		20176	20208	20240		20272	20304		
5752	5784		5816	5848	5880		5912	5944			5976	6008	6040		6072	6104	5624		5656	5720		5688	6120	6152		6184	6216	6248			6280	6312		6344	6376	6408		6440	6472	19568	19600		19632	19664	19696		19728	19760			19792	19824	19856		19888	19920	19440		19472	19536		19504	19936	19968		20000	20032	20064			20096	20128		20160	20192	20224		20256	20288		
5737	5769		5801	5833	5865		5897	5929			5961	5993	6025		6057	6089	5609		5641	5705		5673	6137	6169		6201	6233	6265			6297	6329		6361	6393	6425		6457	6489	19553	19585		19617	19649	19681		19713	19745			19777	19809	19841		19873	19905	19425		19457	19521		19489	19953	19985		20017	20049	20081			20113	20145		20177	20209	20241		20273	20305		
5753	5785		5817	5849	5881		5913	5945			5977	6009	6041		6073	6105	5625		5657	5721		5689	6121	6153		6185	6217	6249			6281	6313		6345	6377	6409		6441	6473	19569	19601		19633	19665	19697		19729	19761			19793	19825	19857		19889	19921	19441		19473	19537		19505	19937	19969		20001	20033	20065			20097	20129		20161	20193	20225		20257	20289		
5738	5770		5802	5834	5866		5898	5930			5962	5994	6026		6058	6090	5610		5642	5706		5674	6138	6170		6202	6234	6266			6298	6330		6362	6394	6426		6458	6490	19554	19586		19618	19650	19682		19714	19746			19778	19810	19842		19874	19906	19426		19458	19522		19490	19954	19986		20018	20050	20082			20114	20146		20178	20210	20242		20274	20306		
5754	5786		5818	5850	5882		5914	5946			5978	6010	6042		6074	6106	5626		5658	5722		5690	6122	6154		6186	6218	6250			6282	6314		6346	6378	6410		6442	6474	19570	19602		19634	19666	19698		19730	19762			19794	19826	19858		19890	19922	19442		19474	19538		19506	19938	19970		20002	20034	20066			20098	20130		20162	20194	20226		20258	20290		
5739	5771		5803	5835	5867		5899	5931			5963	5995	6027		6059	6091	5611		5643	5707		5675	6139	6171		6203	6235	6267			6299	6331		6363	6395	6427		6459	6491	19555	19587		19619	19651	19683		19715	19747			19779	19811	19843		19875	19907	19427		19459	19523		19491	19955	19987		20019	20051	20083			20115	20147		20179	20211	20243		20275	20307		
5755	5787		5819	5851	5883		5915	5947			5979	6011	6043		6075	6107	5627		5659	5723		5691	6123	6155		6187	6219	6251			6283	6315		6347	6379	6411		6443	6475	19571	19603		19635	19667	19699		19731	19763			19795	19827	19859		19891	19923	19443		19475	19539		19507	19939	19971		20003	20035	20067			20099	20131		20163	20195	20227		20259	20291		
5740	5772		5804	5836	5868		5900	5932			5964	5996	6028		6060	6092	5612		5644	5708		5676	6140	6172		6204	6236	6268			6300	6332		6364	6396	6428		6460	6492	19556	19588		19620	19652	19684		19716	19748			19780	19812	19844		19876	19908	19428		19460	19524		19492	19956	19988		20020	20052	20084			20116	20148		20180	20212	20244		20276	20308		
5756	5788		5820	5852	5884		5916	5948			5980	6012	6044		6076	6108	5628		5660	5724		5692	6124	6156		6188	6220	6252			6284	6316		6348	6380	6412		6444	6476	19572	19604		19636	19668	19700		19732	19764			19796	19828	19860		19892	19924	19444		19476	19540		19508	19940	19972		20004	20036	20068			20100	20132		20164	20196	20228		20260	20292		
5741	5773		5805	5837	5869		5901	5933			5965	5997	6029		6061	6093	5613		5645	5709		5677	6141	6173		6205	6237	6269			6301	6333		6365	6397	6429		6461	6493	19557	19589		19621	19653	19685		19717	19749			19781	19813	19845		19877	19909	19429		19461	19525		19493	19957	19989		20021	20053	20085			20117	20149		20181	20213	20245		20277	20309		
5757	5789		5821	5853	5885		5917	5949			5981	6013	6045		6077	6109	5629		5661	5725		5693	6125	6157		6189	6221	6253			6285	6317		6349	6381	6413		6445	6477	19573	19605		19637	19669	19701		19733	19765			19797	19829	19861		19893	19925	19445		19477	19541		19509	19941	19973		20005	20037	20069			20101	20133		20165	20197	20229		20261	20293		
5742	5774		5806	5838	5870		5902	5934			5966	5998	6030		6062	6094	5614		5646	5710		5678	6142	6174		6206	6238	6270			6302	6334		6366	6398	6430		6462	6494	19558	19590		19622	19654	19686		19718	19750			19782	19814	19846		19878	19910	19430		19462	19526		19494	19958	19990		20022	20054	20086			20118	20150		20182	20214	20246		20278	20310		
5758	5790		5822	5854	5886		5918	5950			5982	6014	6046		6078	6110	5630		5662	5726		5694	6126	6158		6190	6222	6254			6286	6318		6350	6382	6414		6446	6478	19574	19606		19638	19670	19702		19734	19766			19798	19830	19862		19894	19926	19446		19478	19542		19510	19942	19974		20006	20038	20070			20102	20134		20166	20198	20230		20262	20294		
5743	5775		5807	5839	5871		5903	5935			5967	5999	6031		6063	6095	5615		5647	5711		5679	6143	6175		6207	6239	6271			6303	6335		6367	6399	6431		6463	6495	19559	19591		19623	19655	19687		19719	19751			19783	19815	19847		19879	19911	19431		19463	19527		19495	19959	19991		20023	20055	20087			20119	20151		20183	20215	20247		20279	20311		
5759	5791		5823	5855	5887		5919	5951			5983	6015	6047		6079	6111	5631		5663	5727		5695	6127	6159		6191	6223	6255			6287	6319		6351	6383	6415		6447	6479	19575	19607		19639	19671	19703		19735	19767			19799	19831	19863		19895	19927	19447		19479	19543		19511	19943	19975		20007	20039	20071			20103	20135		20167	20199	20231		20263	20295		
1121	1123	1125	1127	1129	1131	1133	1135	1137	1139	1141	1143	1145	1147	1149	1151	1153	1155	1157	1159	1161	1163	1165	1167	1169	1171	1173	1175	1177	1179	1181	1183	1185	1187	1189	1191	1193	1195	1197	1199	14937	14939	14941	14943	14945	14947	14949	14951	14953	14955	14957	14959	14961	14963	14965	14967	14969	14971	14973	14975	14977	14979	14981	14983	14985	14987	14989	14991	14993	14995	14997	14999	15001	15003	15005	15007	15009	15011	15013	15015		
1120	1122	1124	1126	1128	1130	1132	1134	1136	1138	1140	1142	1144	1146	1148	1150	1152	1154	1156	1158	1160	1162	1164	1166	1168	1170	1172	1174	1176	1178	1180	1182	1184	1186	1188	1190	1192	1194	1196	1198	14936	14938	14940	14942	14944	14946	14948	14950	14952	14954	14956	14958	14960	14962	14964	14966	14968	14970	14972	14974	14976	14978	14980	14982	14984	14986	14988	14990	14992	14994	14996	14998	15000	15002	15004	15006	15008	15010	15012	15014		
1201	1203	1205	1207	1209	1211	1213	1215	1217	1219	1221	1223	1225	1227	1229	1231	1233	1235	1237	1239	1241	1243	1245	1247	1249	1251	1253	1255	1257	1259	1261	1263	1265	1267	1269	1271	1273	1275	1277	1279	15017	15019	15021	15023	15025	15027	15029	15031	15033	15035	15037	15039	15041	15043	15045	15047	15049	15051	15053	15055	15057	15059	15061	15063	15065	15067	15069	15071	15073	15075	15077	15079	15081	15083	15085	15087	15089	15091	15093	15095		
1200	1202	1204	1206	1208	1210	1212	1214	1216	1218	1220	1222	1224	1226	1228	1230	1232	1234	1236	1238	1240	1242	1244	1246	1248	1250	1252	1254	1256	1258	1260	1262	1264	1266	1268	1270	1272	1274	1276	1278	15016	15018	15020	15022	15024	15026	15028	15030	15032	15034	15036	15038	15040	15042	15044	15046	15048	15050	15052	15054	15056	15058	15060	15062	15064	15066	15068	15070	15072	15074	15076	15078	15080	15082	15084	15086	15088	15090	15092	15094		
1281	1283	1285	1287	1289	1291	1293	1295	1297	1299	1301	1303	1305	1307	1309	1311	1313	1315	1317	1319	1321	1323	1325	1327	1329	1331	1333	1335	1337	1339	1341	1343	1345	1347	1349	1351	1353	1355	1357	1359	15097	15099	15101	15103	15105	15107	15109	15111	15113	15115	15117	15119	15121	15123	15125	15127	15129	15131	15133	15135	15137	15139	15141	15143	15145	15147	15149	15151	15153	15155	15157	15159	15161	15163	15165	15167	15169	15171	15173	15175		
1280	1282	1284	1286	1288	1290	1292	1294	1296	1298	1300	1302	1304	1306	1308	1310	1312	1314	1316	1318	1320	1322	1324	1326	1328	1330	1332	1334	1336	1338	1340	1342	1344	1346	1348	1350	1352	1354	1356	1358	15096	15098	15100	15102	15104	15106	15108	15110	15112	15114	15116	15118	15120	15122	15124	15126	15128	15130	15132	15134	15136	15138	15140	15142	15144	15146	15148	15150	15152	15154	15156	15158	15160	15162	15164	15166	15168	15170	15172	15174		
1361	1363	1365	1367	1369	1371	1373	1375	1377	1379	1381	1383	1385	1387	1389	1391	1393	1395	1397	1399	1401	1403	1405	1407	1409	1411	1413	1415	1417	1419	1421	1423	1425	1427	1429	1431	1433	1435	1437	1439	15177	15179	15181	15183	15185	15187	15189	15191	15193	15195	15197	15199	15201	15203	15205	15207	15209	15211	15213	15215	15217	15219	15221	15223	15225	15227	15229	15231	15233	15235	15237	15239	15241	15243	15245	15247	15249	15251	15253	15255		
1360	1362	1364	1366	1368	1370	1372	1374	1376	1378	1380	1382	1384	1386	1388	1390	1392	1394	1396	1398	1400	1402	1404	1406	1408	1410	1412	1414	1416	1418	1420	1422	1424	1426	1428	1430	1432	1434	1436	1438	15176	15178	15180	15182	15184	15186	15188	15190	15192	15194	15196	15198	15200	15202	15204	15206	15208	15210	15212	15214	15216	15218	15220	15222	15224	15226	15228	15230	15232	15234	15236	15238	15240	15242	15244	15246	15248	15250	15252	15254		
1441	1443	1445	1447	1449	1451	1453	1455	1457	1459	1461	1463	1465	1467	1469	1471	1473	1475	1477	1479	1481	1483	1485	1487	1489	1491	1493	1495	1497	1499	1501	1503	1505	1507	1509	1511	1513	1515	1517	1519	15257	15259	15261	15263	15265	15267	15269	15271	15273	15275	15277	15279	15281	15283	15285	15287	15289	15291	15293	15295	15297	15299	15301	15303	15305	15307	15309	15311	15313	15315	15317	15319	15321	15323	15325	15327	15329	15331	15333	15335		
1440	1442	1444	1446	1448	1450	1452	1454	1456	1458	1460	1462	1464	1466	1468	1470	1472	1474	1476	1478	1480	1482	1484	1486	1488	1490	1492	1494	1496	1498	1500	1502	1504	1506	1508	1510	1512	1514	1516	1518	15256	15258	15260	15262	15264	15266	15268	15270	15272	15274	15276	15278	15280	15282	15284	15286	15288	15290	15292	15294	15296	15298	15300	15302	15304	15306	15308	15310	15312	15314	15316	15318	15320	15322	15324	15326	15328	15330	15332	15334		
1521	1523	1525	1527	1529	1531	1533	1535	1537	1539	1541	1543	1545	1547	1549	1551	1553	1555	1557	1559	1561	1563	1565	1567	1569	1571	1573	1575	1577	1579	1581	1583	1585	1587	1589	1591	1593	1595	1597	1599	15337	15339	15341	15343	15345	15347	15349	15351	15353	15355	15357	15359	15361	15363	15365	15367	15369	15371	15373	15375	15377	15379	15381	15383	15385	15387	15389	15391	15393	15395	15397	15399	15401	15403	15405	15407	15409	15411	15413	15415		
1520	1522	1524	1526	1528	1530	1532	1534	1536	1538	1540	1542	1544	1546	1548	1550	1552	1554	1556	1558	1560	1562	1564	1566	1568	1570	1572	1574	1576	1578	1580	1582	1584	1586	1588	1590	1592	1594	1596	1598	15336	15338	15340	15342	15344	15346	15348	15350	15352	15354	15356	15358	15360	15362	15364	15366	15368	15370	15372	15374	15376	15378	15380	15382	15384	15386	15388	15390	15392	15394	15396	15398	15400	15402	15404	15406	15408	15410	15412	15414		
1601	1603	1605	1607	1609	1611	1613	1615	1617	1619	1621	1623	1625	1627	1629	1631	1633	1635	1637	1639	1641	1643	1645	1647	1649	1651	1653	1655	1657	1659	1661	1663	1665	1667	1669	1671	1673	1675	1677	1679	15417	15419	15421	15423	15425	15427	15429	15431	15433	15435	15437	15439	15441	15443	15445	15447	15449	15451	15453	15455	15457	15459	15461	15463	15465	15467	15469	15471	15473	15475	15477	15479	15481	15483	15485	15487	15489	15491	15493	15495		
1600	1602	1604	1606	1608	1610	1612	1614	1616	1618	1620	1622	1624	1626	1628	1630	1632	1634	1636	1638	1640	1642	1644	1646	1648	1650	1652	1654	1656	1658	1660	1662	1664	1666	1668	1670	1672	1674	1676	1678	15416	15418	15420	15422	15424	15426	15428	15430	15432	15434	15436	15438	15440	15442	15444	15446	15448	15450	15452	15454	15456	15458	15460	15462	15464	15466	15468	15470	15472	15474	15476	15478	15480	15482	15484	15486	15488	15490	15492	15494		
1681	1683	1685	1687	1689	1691	1693	1695	1697	1699	1701	1703	1705	1707	1709	1711	1713	1715	1717	1719	1721	1723	1725	1727	1729	1731	1733	1735	1737	1739	1741	1743	1745	1747	1749	1751	1753	1755	1757	1759	15497	15499	15501	15503	15505	15507	15509	15511	15513	15515	15517	15519	15521	15523	15525	15527	15529	15531	15533	15535	15537	15539	15541	15543	15545	15547	15549	15551	15553	15555	15557	15559	15561	15563	15565	15567	15569	15571	15573	15575		
1680	1682	1684	1686	1688	1690	1692	1694	1696	1698	1700	1702	1704	1706	1708	1710	1712	1714	1716	1718	1720	1722	1724	1726	1728	1730	1732	1734	1736	1738	1740	1742	1744	1746	1748	1750	1752	1754	1756	1758	15496	15498	15500	15502	15504	15506	15508	15510	15512	15514	15516	15518	15520	15522	15524	15526	15528	15530	15532	15534	15536	15538	15540	15542	15544	15546	15548	15550	15552	15554	15556	15558	15560	15562	15564	15566	15568	15570	15572	15574		
1761	1763	1765	1767	1769	1771	1773	1775	1777	1779	1781	1783	1785	1787	1789	1791	1793	1795	1797	1799	1801	1803	1805	1807	1809	1811	1813	1815	1817	1819	1821	1823	1825	1827	1829	1831	1833	1835	1837	1839	15577	15579	15581	15583	15585	15587	15589	15591	15593	15595	15597	15599	15601	15603	15605	15607	15609	15611	15613	15615	15617	15619	15621	15623	15625	15627	15629	15631	15633	15635	15637	15639	15641	15643	15645	15647	15649	15651	15653	15655		
1760	1762	1764	1766	1768	1770	1772	1774	1776	1778	1780	1782	1784	1786	1788	1790	1792	1794	1796	1798	1800	1802	1804	1806	1808	1810	1812	1814	1816	1818	1820	1822	1824	1826	1828	1830	1832	1834	1836	1838	15576	15578	15580	15582	15584	15586	15588	15590	15592	15594	15596	15598	15600	15602	15604	15606	15608	15610	15612	15614	15616	15618	15620	15622	15624	15626	15628	15630	15632	15634	15636	15638	15640	15642	15644	15646	15648	15650	15652	15654		
1841	1843	1845	1847	1849	1851	1853	1855	1857	1859	1861	1863	1865	1867	1869	1871	1873	1875	1877	1879	1881	1883	1885	1887	1889	1891	1893	1895	1897	1899	1901	1903	1905	1907	1909	1911	1913	1915	1917	1919	15657	15659	15661	15663	15665	15667	15669	15671	15673	15675	15677	15679	15681	15683	15685	15687	15689	15691	15693	15695	15697	15699	15701	15703	15705	15707	15709	15711	15713	15715	15717	15719	15721	15723	15725	15727	15729	15731	15733	15735		
1840	1842	1844	1846	1848	1850	1852	1854	1856	1858	1860	1862	1864	1866	1868	1870	1872	1874	1876	1878	1880	1882	1884	1886	1888	1890	1892	1894	1896	1898	1900	1902	1904	1906	1908	1910	1912	1914	1916	1918	15656	15658	15660	15662	15664	15666	15668	15670	15672	15674	15676	15678	15680	15682	15684	15686	15688	15690	15692	15694	15696	15698	15700	15702	15704	15706	15708	15710	15712	15714	15716	15718	15720	15722	15724	15726	15728	15730	15732	15734		
1921	1923	1925	1927	1929	1931	1933	1935	1937	1939	1941	1943	1945	1947	1949	1951	1953	1955	1957	1959	1961	1963	1965	1967	1969	1971	1973	1975	1977	1979	1981	1983	1985	1987	1989	1991	1993	1995	1997	1999	15737	15739	15741	15743	15745	15747	15749	15751	15753	15755	15757	15759	15761	15763	15765	15767	15769	15771	15773	15775	15777	15779	15781	15783	15785	15787	15789	15791	15793	15795	15797	15799	15801	15803	15805	15807	15809	15811	15813	15815		
1920	1922	1924	1926	1928	1930	1932	1934	1936	1938	1940	1942	1944	1946	1948	1950	1952	1954	1956	1958	1960	1962	1964	1966	1968	1970	1972	1974	1976	1978	1980	1982	1984	1986	1988	1990	1992	1994	1996	1998	15736	15738	15740	15742	15744	15746	15748	15750	15752	15754	15756	15758	15760	15762	15764	15766	15768	15770	15772	15774	15776	15778	15780	15782	15784	15786	15788	15790	15792	15794	15796	15798	15800	15802	15804	15806	15808	15810	15812	15814		
5521	5523	5525	5527	5529	5531	5533	5535	5537	5539	5541	5543	5545	5547	5549	5551	5553	5555	5557	5559	5561	5563	5565	5567	5569	5571	5573	5575	5577	5579	5581	5583	5585	5587	5589	5591	5593	5595	5597	5599	19337	19339	19341	19343	19345	19347	19349	19351	19353	19355	19357	19359	19361	19363	19365	19367	19369	19371	19373	19375	19377	19379	19381	19383	19385	19387	19389	19391	19393	19395	19397	19399	19401	19403	19405	19407	19409	19411	19413	19415		
5520	5522	5524	5526	5528	5530	5532	5534	5536	5538	5540	5542	5544	5546	5548	5550	5552	5554	5556	5558	5560	5562	5564	5566	5568	5570	5572	5574	5576	5578	5580	5582	5584	5586	5588	5590	5592	5594	5596	5598	19336	19338	19340	19342	19344	19346	19348	19350	19352	19354	19356	19358	19360	19362	19364	19366	19368	19370	19372	19374	19376	19378	19380	19382	19384	19386	19388	19390	19392	19394	19396	19398	19400	19402	19404	19406	19408	19410	19412	19414		
5441	5443	5445	5447	5449	5451	5453	5455	5457	5459	5461	5463	5465	5467	5469	5471	5473	5475	5477	5479	5481	5483	5485	5487	5489	5491	5493	5495	5497	5499	5501	5503	5505	5507	5509	5511	5513	5515	5517	5519	19257	19259	19261	19263	19265	19267	19269	19271	19273	19275	19277	19279	19281	19283	19285	19287	19289	19291	19293	19295	19297	19299	19301	19303	19305	19307	19309	19311	19313	19315	19317	19319	19321	19323	19325	19327	19329	19331	19333	19335		
5440	5442	5444	5446	5448	5450	5452	5454	5456	5458	5460	5462	5464	5466	5468	5470	5472	5474	5476	5478	5480	5482	5484	5486	5488	5490	5492	5494	5496	5498	5500	5502	5504	5506	5508	5510	5512	5514	5516	5518	19256	19258	19260	19262	19264	19266	19268	19270	19272	19274	19276	19278	19280	19282	19284	19286	19288	19290	19292	19294	19296	19298	19300	19302	19304	19306	19308	19310	19312	19314	19316	19318	19320	19322	19324	19326	19328	19330	19332	19334		
5361	5363	5365	5367	5369	5371	5373	5375	5377	5379	5381	5383	5385	5387	5389	5391	5393	5395	5397	5399	5401	5403	5405	5407	5409	5411	5413	5415	5417	5419	5421	5423	5425	5427	5429	5431	5433	5435	5437	5439	19177	19179	19181	19183	19185	19187	19189	19191	19193	19195	19197	19199	19201	19203	19205	19207	19209	19211	19213	19215	19217	19219	19221	19223	19225	19227	19229	19231	19233	19235	19237	19239	19241	19243	19245	19247	19249	19251	19253	19255		
5360	5362	5364	5366	5368	5370	5372	5374	5376	5378	5380	5382	5384	5386	5388	5390	5392	5394	5396	5398	5400	5402	5404	5406	5408	5410	5412	5414	5416	5418	5420	5422	5424	5426	5428	5430	5432	5434	5436	5438	19176	19178	19180	19182	19184	19186	19188	19190	19192	19194	19196	19198	19200	19202	19204	19206	19208	19210	19212	19214	19216	19218	19220	19222	19224	19226	19228	19230	19232	19234	19236	19238	19240	19242	19244	19246	19248	19250	19252	19254		
2001	2003	2005	2007	2009	2011	2013	2015	2017	2019	2021	2023	2025	2027	2029	2031	2033	2035	2037	2039	2041	2043	2045	2047	2049	2051	2053	2055	2057	2059	2061	2063	2065	2067	2069	2071	2073	2075	2077	2079	15817	15819	15821	15823	15825	15827	15829	15831	15833	15835	15837	15839	15841	15843	15845	15847	15849	15851	15853	15855	15857	15859	15861	15863	15865	15867	15869	15871	15873	15875	15877	15879	15881	15883	15885	15887	15889	15891	15893	15895		
2000	2002	2004	2006	2008	2010	2012	2014	2016	2018	2020	2022	2024	2026	2028	2030	2032	2034	2036	2038	2040	2042	2044	2046	2048	2050	2052	2054	2056	2058	2060	2062	2064	2066	2068	2070	2072	2074	2076	2078	15816	15818	15820	15822	15824	15826	15828	15830	15832	15834	15836	15838	15840	15842	15844	15846	15848	15850	15852	15854	15856	15858	15860	15862	15864	15866	15868	15870	15872	15874	15876	15878	15880	15882	15884	15886	15888	15890	15892	15894		
2081	2083	2085	2087	2089	2091	2093	2095	2097	2099	2101	2103	2105	2107	2109	2111	2113	2115	2117	2119	2121	2123	2125	2127	2129	2131	2133	2135	2137	2139	2141	2143	2145	2147	2149	2151	2153	2155	2157	2159	15897	15899	15901	15903	15905	15907	15909	15911	15913	15915	15917	15919	15921	15923	15925	15927	15929	15931	15933	15935	15937	15939	15941	15943	15945	15947	15949	15951	15953	15955	15957	15959	15961	15963	15965	15967	15969	15971	15973	15975		
2080	2082	2084	2086	2088	2090	2092	2094	2096	2098	2100	2102	2104	2106	2108	2110	2112	2114	2116	2118	2120	2122	2124	2126	2128	2130	2132	2134	2136	2138	2140	2142	2144	2146	2148	2150	2152	2154	2156	2158	15896	15898	15900	15902	15904	15906	15908	15910	15912	15914	15916	15918	15920	15922	15924	15926	15928	15930	15932	15934	15936	15938	15940	15942	15944	15946	15948	15950	15952	15954	15956	15958	15960	15962	15964	15966	15968	15970	15972	15974		
2161	2163	2165	2167	2169	2171	2173	2175	2177	2179	2181	2183	2185	2187	2189	2191	2193	2195	2197	2199	2201	2203	2205	2207	2209	2211	2213	2215	2217	2219	2221	2223	2225	2227	2229	2231	2233	2235	2237	2239	15977	15979	15981	15983	15985	15987	15989	15991	15993	15995	15997	15999	16001	16003	16005	16007	16009	16011	16013	16015	16017	16019	16021	16023	16025	16027	16029	16031	16033	16035	16037	16039	16041	16043	16045	16047	16049	16051	16053	16055		
2160	2162	2164	2166	2168	2170	2172	2174	2176	2178	2180	2182	2184	2186	2188	2190	2192	2194	2196	2198	2200	2202	2204	2206	2208	2210	2212	2214	2216	2218	2220	2222	2224	2226	2228	2230	2232	2234	2236	2238	15976	15978	15980	15982	15984	15986	15988	15990	15992	15994	15996	15998	16000	16002	16004	16006	16008	16010	16012	16014	16016	16018	16020	16022	16024	16026	16028	16030	16032	16034	16036	16038	16040	16042	16044	16046	16048	16050	16052	16054		
5281	5283	5285	5287	5289	5291	5293	5295	5297	5299	5301	5303	5305	5307	5309	5311	5313	5315	5317	5319	5321	5323	5325	5327	5329	5331	5333	5335	5337	5339	5341	5343	5345	5347	5349	5351	5353	5355	5357	5359	19097	19099	19101	19103	19105	19107	19109	19111	19113	19115	19117	19119	19121	19123	19125	19127	19129	19131	19133	19135	19137	19139	19141	19143	19145	19147	19149	19151	19153	19155	19157	19159	19161	19163	19165	19167	19169	19171	19173	19175		
5280	5282	5284	5286	5288	5290	5292	5294	5296	5298	5300	5302	5304	5306	5308	5310	5312	5314	5316	5318	5320	5322	5324	5326	5328	5330	5332	5334	5336	5338	5340	5342	5344	5346	5348	5350	5352	5354	5356	5358	19096	19098	19100	19102	19104	19106	19108	19110	19112	19114	19116	19118	19120	19122	19124	19126	19128	19130	19132	19134	19136	19138	19140	19142	19144	19146	19148	19150	19152	19154	19156	19158	19160	19162	19164	19166	19168	19170	19172	19174		
5201	5203	5205	5207	5209	5211	5213	5215	5217	5219	5221	5223	5225	5227	5229	5231	5233	5235	5237	5239	5241	5243	5245	5247	5249	5251	5253	5255	5257	5259	5261	5263	5265	5267	5269	5271	5273	5275	5277	5279	19017	19019	19021	19023	19025	19027	19029	19031	19033	19035	19037	19039	19041	19043	19045	19047	19049	19051	19053	19055	19057	19059	19061	19063	19065	19067	19069	19071	19073	19075	19077	19079	19081	19083	19085	19087	19089	19091	19093	19095		
5200	5202	5204	5206	5208	5210	5212	5214	5216	5218	5220	5222	5224	5226	5228	5230	5232	5234	5236	5238	5240	5242	5244	5246	5248	5250	5252	5254	5256	5258	5260	5262	5264	5266	5268	5270	5272	5274	5276	5278	19016	19018	19020	19022	19024	19026	19028	19030	19032	19034	19036	19038	19040	19042	19044	19046	19048	19050	19052	19054	19056	19058	19060	19062	19064	19066	19068	19070	19072	19074	19076	19078	19080	19082	19084	19086	19088	19090	19092	19094		
5121	5123	5125	5127	5129	5131	5133	5135	5137	5139	5141	5143	5145	5147	5149	5151	5153	5155	5157	5159	5161	5163	5165	5167	5169	5171	5173	5175	5177	5179	5181	5183	5185	5187	5189	5191	5193	5195	5197	5199	18937	18939	18941	18943	18945	18947	18949	18951	18953	18955	18957	18959	18961	18963	18965	18967	18969	18971	18973	18975	18977	18979	18981	18983	18985	18987	18989	18991	18993	18995	18997	18999	19001	19003	19005	19007	19009	19011	19013	19015		
5120	5122	5124	5126	5128	5130	5132	5134	5136	5138	5140	5142	5144	5146	5148	5150	5152	5154	5156	5158	5160	5162	5164	5166	5168	5170	5172	5174	5176	5178	5180	5182	5184	5186	5188	5190	5192	5194	5196	5198	18936	18938	18940	18942	18944	18946	18948	18950	18952	18954	18956	18958	18960	18962	18964	18966	18968	18970	18972	18974	18976	18978	18980	18982	18984	18986	18988	18990	18992	18994	18996	18998	19000	19002	19004	19006	19008	19010	19012	19014		
2241	2243	2245	2247	2249	2251	2253	2255	2257	2259	2261	2263	2265	2267	2269	2271	2273	2275	2277	2279	2281	2283	2285	2287	2289	2291	2293	2295	2297	2299	2301	2303	2305	2307	2309	2311	2313	2315	2317	2319	16057	16059	16061	16063	16065	16067	16069	16071	16073	16075	16077	16079	16081	16083	16085	16087	16089	16091	16093	16095	16097	16099	16101	16103	16105	16107	16109	16111	16113	16115	16117	16119	16121	16123	16125	16127	16129	16131	16133	16135		
2240	2242	2244	2246	2248	2250	2252	2254	2256	2258	2260	2262	2264	2266	2268	2270	2272	2274	2276	2278	2280	2282	2284	2286	2288	2290	2292	2294	2296	2298	2300	2302	2304	2306	2308	2310	2312	2314	2316	2318	16056	16058	16060	16062	16064	16066	16068	16070	16072	16074	16076	16078	16080	16082	16084	16086	16088	16090	16092	16094	16096	16098	16100	16102	16104	16106	16108	16110	16112	16114	16116	16118	16120	16122	16124	16126	16128	16130	16132	16134		
2321	2323	2325	2327	2329	2331	2333	2335	2337	2339	2341	2343	2345	2347	2349	2351	2353	2355	2357	2359	2361	2363	2365	2367	2369	2371	2373	2375	2377	2379	2381	2383	2385	2387	2389	2391	2393	2395	2397	2399	16137	16139	16141	16143	16145	16147	16149	16151	16153	16155	16157	16159	16161	16163	16165	16167	16169	16171	16173	16175	16177	16179	16181	16183	16185	16187	16189	16191	16193	16195	16197	16199	16201	16203	16205	16207	16209	16211	16213	16215		
2320	2322	2324	2326	2328	2330	2332	2334	2336	2338	2340	2342	2344	2346	2348	2350	2352	2354	2356	2358	2360	2362	2364	2366	2368	2370	2372	2374	2376	2378	2380	2382	2384	2386	2388	2390	2392	2394	2396	2398	16136	16138	16140	16142	16144	16146	16148	16150	16152	16154	16156	16158	16160	16162	16164	16166	16168	16170	16172	16174	16176	16178	16180	16182	16184	16186	16188	16190	16192	16194	16196	16198	16200	16202	16204	16206	16208	16210	16212	16214		
2401	2403	2405	2407	2409	2411	2413	2415	2417	2419	2421	2423	2425	2427	2429	2431	2433	2435	2437	2439	2441	2443	2445	2447	2449	2451	2453	2455	2457	2459	2461	2463	2465	2467	2469	2471	2473	2475	2477	2479	16217	16219	16221	16223	16225	16227	16229	16231	16233	16235	16237	16239	16241	16243	16245	16247	16249	16251	16253	16255	16257	16259	16261	16263	16265	16267	16269	16271	16273	16275	16277	16279	16281	16283	16285	16287	16289	16291	16293	16295		
2400	2402	2404	2406	2408	2410	2412	2414	2416	2418	2420	2422	2424	2426	2428	2430	2432	2434	2436	2438	2440	2442	2444	2446	2448	2450	2452	2454	2456	2458	2460	2462	2464	2466	2468	2470	2472	2474	2476	2478	16216	16218	16220	16222	16224	16226	16228	16230	16232	16234	16236	16238	16240	16242	16244	16246	16248	16250	16252	16254	16256	16258	16260	16262	16264	16266	16268	16270	16272	16274	16276	16278	16280	16282	16284	16286	16288	16290	16292	16294		
5041	5043	5045	5047	5049	5051	5053	5055	5057	5059	5061	5063	5065	5067	5069	5071	5073	5075	5077	5079	5081	5083	5085	5087	5089	5091	5093	5095	5097	5099	5101	5103	5105	5107	5109	5111	5113	5115	5117	5119	18857	18859	18861	18863	18865	18867	18869	18871	18873	18875	18877	18879	18881	18883	18885	18887	18889	18891	18893	18895	18897	18899	18901	18903	18905	18907	18909	18911	18913	18915	18917	18919	18921	18923	18925	18927	18929	18931	18933	18935		
5040	5042	5044	5046	5048	5050	5052	5054	5056	5058	5060	5062	5064	5066	5068	5070	5072	5074	5076	5078	5080	5082	5084	5086	5088	5090	5092	5094	5096	5098	5100	5102	5104	5106	5108	5110	5112	5114	5116	5118	18856	18858	18860	18862	18864	18866	18868	18870	18872	18874	18876	18878	18880	18882	18884	18886	18888	18890	18892	18894	18896	18898	18900	18902	18904	18906	18908	18910	18912	18914	18916	18918	18920	18922	18924	18926	18928	18930	18932	18934		
4961	4963	4965	4967	4969	4971	4973	4975	4977	4979	4981	4983	4985	4987	4989	4991	4993	4995	4997	4999	5001	5003	5005	5007	5009	5011	5013	5015	5017	5019	5021	5023	5025	5027	5029	5031	5033	5035	5037	5039	18777	18779	18781	18783	18785	18787	18789	18791	18793	18795	18797	18799	18801	18803	18805	18807	18809	18811	18813	18815	18817	18819	18821	18823	18825	18827	18829	18831	18833	18835	18837	18839	18841	18843	18845	18847	18849	18851	18853	18855		
4960	4962	4964	4966	4968	4970	4972	4974	4976	4978	4980	4982	4984	4986	4988	4990	4992	4994	4996	4998	5000	5002	5004	5006	5008	5010	5012	5014	5016	5018	5020	5022	5024	5026	5028	5030	5032	5034	5036	5038	18776	18778	18780	18782	18784	18786	18788	18790	18792	18794	18796	18798	18800	18802	18804	18806	18808	18810	18812	18814	18816	18818	18820	18822	18824	18826	18828	18830	18832	18834	18836	18838	18840	18842	18844	18846	18848	18850	18852	18854		
4881	4883	4885	4887	4889	4891	4893	4895	4897	4899	4901	4903	4905	4907	4909	4911	4913	4915	4917	4919	4921	4923	4925	4927	4929	4931	4933	4935	4937	4939	4941	4943	4945	4947	4949	4951	4953	4955	4957	4959	18697	18699	18701	18703	18705	18707	18709	18711	18713	18715	18717	18719	18721	18723	18725	18727	18729	18731	18733	18735	18737	18739	18741	18743	18745	18747	18749	18751	18753	18755	18757	18759	18761	18763	18765	18767	18769	18771	18773	18775		
4880	4882	4884	4886	4888	4890	4892	4894	4896	4898	4900	4902	4904	4906	4908	4910	4912	4914	4916	4918	4920	4922	4924	4926	4928	4930	4932	4934	4936	4938	4940	4942	4944	4946	4948	4950	4952	4954	4956	4958	18696	18698	18700	18702	18704	18706	18708	18710	18712	18714	18716	18718	18720	18722	18724	18726	18728	18730	18732	18734	18736	18738	18740	18742	18744	18746	18748	18750	18752	18754	18756	18758	18760	18762	18764	18766	18768	18770	18772	18774		
2481	2483	2485	2487	2489	2491	2493	2495	2497	2499	2501	2503	2505	2507	2509	2511	2513	2515	2517	2519	2521	2523	2525	2527	2529	2531	2533	2535	2537	2539	2541	2543	2545	2547	2549	2551	2553	2555	2557	2559	16297	16299	16301	16303	16305	16307	16309	16311	16313	16315	16317	16319	16321	16323	16325	16327	16329	16331	16333	16335	16337	16339	16341	16343	16345	16347	16349	16351	16353	16355	16357	16359	16361	16363	16365	16367	16369	16371	16373	16375		
2480	2482	2484	2486	2488	2490	2492	2494	2496	2498	2500	2502	2504	2506	2508	2510	2512	2514	2516	2518	2520	2522	2524	2526	2528	2530	2532	2534	2536	2538	2540	2542	2544	2546	2548	2550	2552	2554	2556	2558	16296	16298	16300	16302	16304	16306	16308	16310	16312	16314	16316	16318	16320	16322	16324	16326	16328	16330	16332	16334	16336	16338	16340	16342	16344	16346	16348	16350	16352	16354	16356	16358	16360	16362	16364	16366	16368	16370	16372	16374		
2561	2563	2565	2567	2569	2571	2573	2575	2577	2579	2581	2583	2585	2587	2589	2591	2593	2595	2597	2599	2601	2603	2605	2607	2609	2611	2613	2615	2617	2619	2621	2623	2625	2627	2629	2631	2633	2635	2637	2639	16377	16379	16381	16383	16385	16387	16389	16391	16393	16395	16397	16399	16401	16403	16405	16407	16409	16411	16413	16415	16417	16419	16421	16423	16425	16427	16429	16431	16433	16435	16437	16439	16441	16443	16445	16447	16449	16451	16453	16455		
2560	2562	2564	2566	2568	2570	2572	2574	2576	2578	2580	2582	2584	2586	2588	2590	2592	2594	2596	2598	2600	2602	2604	2606	2608	2610	2612	2614	2616	2618	2620	2622	2624	2626	2628	2630	2632	2634	2636	2638	16376	16378	16380	16382	16384	16386	16388	16390	16392	16394	16396	16398	16400	16402	16404	16406	16408	16410	16412	16414	16416	16418	16420	16422	16424	16426	16428	16430	16432	16434	16436	16438	16440	16442	16444	16446	16448	16450	16452	16454		
2641	2643	2645	2647	2649	2651	2653	2655	2657	2659	2661	2663	2665	2667	2669	2671	2673	2675	2677	2679	2681	2683	2685	2687	2689	2691	2693	2695	2697	2699	2701	2703	2705	2707	2709	2711	2713	2715	2717	2719	16457	16459	16461	16463	16465	16467	16469	16471	16473	16475	16477	16479	16481	16483	16485	16487	16489	16491	16493	16495	16497	16499	16501	16503	16505	16507	16509	16511	16513	16515	16517	16519	16521	16523	16525	16527	16529	16531	16533	16535		
2640	2642	2644	2646	2648	2650	2652	2654	2656	2658	2660	2662	2664	2666	2668	2670	2672	2674	2676	2678	2680	2682	2684	2686	2688	2690	2692	2694	2696	2698	2700	2702	2704	2706	2708	2710	2712	2714	2716	2718	16456	16458	16460	16462	16464	16466	16468	16470	16472	16474	16476	16478	16480	16482	16484	16486	16488	16490	16492	16494	16496	16498	16500	16502	16504	16506	16508	16510	16512	16514	16516	16518	16520	16522	16524	16526	16528	16530	16532	16534		
4801	4803	4805	4807	4809	4811	4813	4815	4817	4819	4821	4823	4825	4827	4829	4831	4833	4835	4837	4839	4841	4843	4845	4847	4849	4851	4853	4855	4857	4859	4861	4863	4865	4867	4869	4871	4873	4875	4877	4879	18617	18619	18621	18623	18625	18627	18629	18631	18633	18635	18637	18639	18641	18643	18645	18647	18649	18651	18653	18655	18657	18659	18661	18663	18665	18667	18669	18671	18673	18675	18677	18679	18681	18683	18685	18687	18689	18691	18693	18695		
4800	4802	4804	4806	4808	4810	4812	4814	4816	4818	4820	4822	4824	4826	4828	4830	4832	4834	4836	4838	4840	4842	4844	4846	4848	4850	4852	4854	4856	4858	4860	4862	4864	4866	4868	4870	4872	4874	4876	4878	18616	18618	18620	18622	18624	18626	18628	18630	18632	18634	18636	18638	18640	18642	18644	18646	18648	18650	18652	18654	18656	18658	18660	18662	18664	18666	18668	18670	18672	18674	18676	18678	18680	18682	18684	18686	18688	18690	18692	18694		
4721	4723	4725	4727	4729	4731	4733	4735	4737	4739	4741	4743	4745	4747	4749	4751	4753	4755	4757	4759	4761	4763	4765	4767	4769	4771	4773	4775	4777	4779	4781	4783	4785	4787	4789	4791	4793	4795	4797	4799	18537	18539	18541	18543	18545	18547	18549	18551	18553	18555	18557	18559	18561	18563	18565	18567	18569	18571	18573	18575	18577	18579	18581	18583	18585	18587	18589	18591	18593	18595	18597	18599	18601	18603	18605	18607	18609	18611	18613	18615		
4720	4722	4724	4726	4728	4730	4732	4734	4736	4738	4740	4742	4744	4746	4748	4750	4752	4754	4756	4758	4760	4762	4764	4766	4768	4770	4772	4774	4776	4778	4780	4782	4784	4786	4788	4790	4792	4794	4796	4798	18536	18538	18540	18542	18544	18546	18548	18550	18552	18554	18556	18558	18560	18562	18564	18566	18568	18570	18572	18574	18576	18578	18580	18582	18584	18586	18588	18590	18592	18594	18596	18598	18600	18602	18604	18606	18608	18610	18612	18614		
4641	4643	4645	4647	4649	4651	4653	4655	4657	4659	4661	4663	4665	4667	4669	4671	4673	4675	4677	4679	4681	4683	4685	4687	4689	4691	4693	4695	4697	4699	4701	4703	4705	4707	4709	4711	4713	4715	4717	4719	18457	18459	18461	18463	18465	18467	18469	18471	18473	18475	18477	18479	18481	18483	18485	18487	18489	18491	18493	18495	18497	18499	18501	18503	18505	18507	18509	18511	18513	18515	18517	18519	18521	18523	18525	18527	18529	18531	18533	18535		
4640	4642	4644	4646	4648	4650	4652	4654	4656	4658	4660	4662	4664	4666	4668	4670	4672	4674	4676	4678	4680	4682	4684	4686	4688	4690	4692	4694	4696	4698	4700	4702	4704	4706	4708	4710	4712	4714	4716	4718	18456	18458	18460	18462	18464	18466	18468	18470	18472	18474	18476	18478	18480	18482	18484	18486	18488	18490	18492	18494	18496	18498	18500	18502	18504	18506	18508	18510	18512	18514	18516	18518	18520	18522	18524	18526	18528	18530	18532	18534		
2721	2723	2725	2727	2729	2731	2733	2735	2737	2739	2741	2743	2745	2747	2749	2751	2753	2755	2757	2759	2761	2763	2765	2767	2769	2771	2773	2775	2777	2779	2781	2783	2785	2787	2789	2791	2793	2795	2797	2799	16537	16539	16541	16543	16545	16547	16549	16551	16553	16555	16557	16559	16561	16563	16565	16567	16569	16571	16573	16575	16577	16579	16581	16583	16585	16587	16589	16591	16593	16595	16597	16599	16601	16603	16605	16607	16609	16611	16613	16615		
2720	2722	2724	2726	2728	2730	2732	2734	2736	2738	2740	2742	2744	2746	2748	2750	2752	2754	2756	2758	2760	2762	2764	2766	2768	2770	2772	2774	2776	2778	2780	2782	2784	2786	2788	2790	2792	2794	2796	2798	16536	16538	16540	16542	16544	16546	16548	16550	16552	16554	16556	16558	16560	16562	16564	16566	16568	16570	16572	16574	16576	16578	16580	16582	16584	16586	16588	16590	16592	16594	16596	16598	16600	16602	16604	16606	16608	16610	16612	16614		
2801	2803	2805	2807	2809	2811	2813	2815	2817	2819	2821	2823	2825	2827	2829	2831	2833	2835	2837	2839	2841	2843	2845	2847	2849	2851	2853	2855	2857	2859	2861	2863	2865	2867	2869	2871	2873	2875	2877	2879	16617	16619	16621	16623	16625	16627	16629	16631	16633	16635	16637	16639	16641	16643	16645	16647	16649	16651	16653	16655	16657	16659	16661	16663	16665	16667	16669	16671	16673	16675	16677	16679	16681	16683	16685	16687	16689	16691	16693	16695		
2800	2802	2804	2806	2808	2810	2812	2814	2816	2818	2820	2822	2824	2826	2828	2830	2832	2834	2836	2838	2840	2842	2844	2846	2848	2850	2852	2854	2856	2858	2860	2862	2864	2866	2868	2870	2872	2874	2876	2878	16616	16618	16620	16622	16624	16626	16628	16630	16632	16634	16636	16638	16640	16642	16644	16646	16648	16650	16652	16654	16656	16658	16660	16662	16664	16666	16668	16670	16672	16674	16676	16678	16680	16682	16684	16686	16688	16690	16692	16694		
2881	2883	2885	2887	2889	2891	2893	2895	2897	2899	2901	2903	2905	2907	2909	2911	2913	2915	2917	2919	2921	2923	2925	2927	2929	2931	2933	2935	2937	2939	2941	2943	2945	2947	2949	2951	2953	2955	2957	2959	16697	16699	16701	16703	16705	16707	16709	16711	16713	16715	16717	16719	16721	16723	16725	16727	16729	16731	16733	16735	16737	16739	16741	16743	16745	16747	16749	16751	16753	16755	16757	16759	16761	16763	16765	16767	16769	16771	16773	16775		
2880	2882	2884	2886	2888	2890	2892	2894	2896	2898	2900	2902	2904	2906	2908	2910	2912	2914	2916	2918	2920	2922	2924	2926	2928	2930	2932	2934	2936	2938	2940	2942	2944	2946	2948	2950	2952	2954	2956	2958	16696	16698	16700	16702	16704	16706	16708	16710	16712	16714	16716	16718	16720	16722	16724	16726	16728	16730	16732	16734	16736	16738	16740	16742	16744	16746	16748	16750	16752	16754	16756	16758	16760	16762	16764	16766	16768	16770	16772	16774		
4561	4563	4565	4567	4569	4571	4573	4575	4577	4579	4581	4583	4585	4587	4589	4591	4593	4595	4597	4599	4601	4603	4605	4607	4609	4611	4613	4615	4617	4619	4621	4623	4625	4627	4629	4631	4633	4635	4637	4639	18377	18379	18381	18383	18385	18387	18389	18391	18393	18395	18397	18399	18401	18403	18405	18407	18409	18411	18413	18415	18417	18419	18421	18423	18425	18427	18429	18431	18433	18435	18437	18439	18441	18443	18445	18447	18449	18451	18453	18455		
4560	4562	4564	4566	4568	4570	4572	4574	4576	4578	4580	4582	4584	4586	4588	4590	4592	4594	4596	4598	4600	4602	4604	4606	4608	4610	4612	4614	4616	4618	4620	4622	4624	4626	4628	4630	4632	4634	4636	4638	18376	18378	18380	18382	18384	18386	18388	18390	18392	18394	18396	18398	18400	18402	18404	18406	18408	18410	18412	18414	18416	18418	18420	18422	18424	18426	18428	18430	18432	18434	18436	18438	18440	18442	18444	18446	18448	18450	18452	18454		
4481	4483	4485	4487	4489	4491	4493	4495	4497	4499	4501	4503	4505	4507	4509	4511	4513	4515	4517	4519	4521	4523	4525	4527	4529	4531	4533	4535	4537	4539	4541	4543	4545	4547	4549	4551	4553	4555	4557	4559	18297	18299	18301	18303	18305	18307	18309	18311	18313	18315	18317	18319	18321	18323	18325	18327	18329	18331	18333	18335	18337	18339	18341	18343	18345	18347	18349	18351	18353	18355	18357	18359	18361	18363	18365	18367	18369	18371	18373	18375		
4480	4482	4484	4486	4488	4490	4492	4494	4496	4498	4500	4502	4504	4506	4508	4510	4512	4514	4516	4518	4520	4522	4524	4526	4528	4530	4532	4534	4536	4538	4540	4542	4544	4546	4548	4550	4552	4554	4556	4558	18296	18298	18300	18302	18304	18306	18308	18310	18312	18314	18316	18318	18320	18322	18324	18326	18328	18330	18332	18334	18336	18338	18340	18342	18344	18346	18348	18350	18352	18354	18356	18358	18360	18362	18364	18366	18368	18370	18372	18374		
4401	4403	4405	4407	4409	4411	4413	4415	4417	4419	4421	4423	4425	4427	4429	4431	4433	4435	4437	4439	4441	4443	4445	4447	4449	4451	4453	4455	4457	4459	4461	4463	4465	4467	4469	4471	4473	4475	4477	4479	18217	18219	18221	18223	18225	18227	18229	18231	18233	18235	18237	18239	18241	18243	18245	18247	18249	18251	18253	18255	18257	18259	18261	18263	18265	18267	18269	18271	18273	18275	18277	18279	18281	18283	18285	18287	18289	18291	18293	18295		
4400	4402	4404	4406	4408	4410	4412	4414	4416	4418	4420	4422	4424	4426	4428	4430	4432	4434	4436	4438	4440	4442	4444	4446	4448	4450	4452	4454	4456	4458	4460	4462	4464	4466	4468	4470	4472	4474	4476	4478	18216	18218	18220	18222	18224	18226	18228	18230	18232	18234	18236	18238	18240	18242	18244	18246	18248	18250	18252	18254	18256	18258	18260	18262	18264	18266	18268	18270	18272	18274	18276	18278	18280	18282	18284	18286	18288	18290	18292	18294		
2961	2963	2965	2967	2969	2971	2973	2975	2977	2979	2981	2983	2985	2987	2989	2991	2993	2995	2997	2999	3001	3003	3005	3007	3009	3011	3013	3015	3017	3019	3021	3023	3025	3027	3029	3031	3033	3035	3037	3039	16777	16779	16781	16783	16785	16787	16789	16791	16793	16795	16797	16799	16801	16803	16805	16807	16809	16811	16813	16815	16817	16819	16821	16823	16825	16827	16829	16831	16833	16835	16837	16839	16841	16843	16845	16847	16849	16851	16853	16855		
2960	2962	2964	2966	2968	2970	2972	2974	2976	2978	2980	2982	2984	2986	2988	2990	2992	2994	2996	2998	3000	3002	3004	3006	3008	3010	3012	3014	3016	3018	3020	3022	3024	3026	3028	3030	3032	3034	3036	3038	16776	16778	16780	16782	16784	16786	16788	16790	16792	16794	16796	16798	16800	16802	16804	16806	16808	16810	16812	16814	16816	16818	16820	16822	16824	16826	16828	16830	16832	16834	16836	16838	16840	16842	16844	16846	16848	16850	16852	16854		
3041	3043	3045	3047	3049	3051	3053	3055	3057	3059	3061	3063	3065	3067	3069	3071	3073	3075	3077	3079	3081	3083	3085	3087	3089	3091	3093	3095	3097	3099	3101	3103	3105	3107	3109	3111	3113	3115	3117	3119	16857	16859	16861	16863	16865	16867	16869	16871	16873	16875	16877	16879	16881	16883	16885	16887	16889	16891	16893	16895	16897	16899	16901	16903	16905	16907	16909	16911	16913	16915	16917	16919	16921	16923	16925	16927	16929	16931	16933	16935		
3040	3042	3044	3046	3048	3050	3052	3054	3056	3058	3060	3062	3064	3066	3068	3070	3072	3074	3076	3078	3080	3082	3084	3086	3088	3090	3092	3094	3096	3098	3100	3102	3104	3106	3108	3110	3112	3114	3116	3118	16856	16858	16860	16862	16864	16866	16868	16870	16872	16874	16876	16878	16880	16882	16884	16886	16888	16890	16892	16894	16896	16898	16900	16902	16904	16906	16908	16910	16912	16914	16916	16918	16920	16922	16924	16926	16928	16930	16932	16934		
3121	3123	3125	3127	3129	3131	3133	3135	3137	3139	3141	3143	3145	3147	3149	3151	3153	3155	3157	3159	3161	3163	3165	3167	3169	3171	3173	3175	3177	3179	3181	3183	3185	3187	3189	3191	3193	3195	3197	3199	16937	16939	16941	16943	16945	16947	16949	16951	16953	16955	16957	16959	16961	16963	16965	16967	16969	16971	16973	16975	16977	16979	16981	16983	16985	16987	16989	16991	16993	16995	16997	16999	17001	17003	17005	17007	17009	17011	17013	17015		
3120	3122	3124	3126	3128	3130	3132	3134	3136	3138	3140	3142	3144	3146	3148	3150	3152	3154	3156	3158	3160	3162	3164	3166	3168	3170	3172	3174	3176	3178	3180	3182	3184	3186	3188	3190	3192	3194	3196	3198	16936	16938	16940	16942	16944	16946	16948	16950	16952	16954	16956	16958	16960	16962	16964	16966	16968	16970	16972	16974	16976	16978	16980	16982	16984	16986	16988	16990	16992	16994	16996	16998	17000	17002	17004	17006	17008	17010	17012	17014		
4321	4323	4325	4327	4329	4331	4333	4335	4337	4339	4341	4343	4345	4347	4349	4351	4353	4355	4357	4359	4361	4363	4365	4367	4369	4371	4373	4375	4377	4379	4381	4383	4385	4387	4389	4391	4393	4395	4397	4399	18137	18139	18141	18143	18145	18147	18149	18151	18153	18155	18157	18159	18161	18163	18165	18167	18169	18171	18173	18175	18177	18179	18181	18183	18185	18187	18189	18191	18193	18195	18197	18199	18201	18203	18205	18207	18209	18211	18213	18215		
4320	4322	4324	4326	4328	4330	4332	4334	4336	4338	4340	4342	4344	4346	4348	4350	4352	4354	4356	4358	4360	4362	4364	4366	4368	4370	4372	4374	4376	4378	4380	4382	4384	4386	4388	4390	4392	4394	4396	4398	18136	18138	18140	18142	18144	18146	18148	18150	18152	18154	18156	18158	18160	18162	18164	18166	18168	18170	18172	18174	18176	18178	18180	18182	18184	18186	18188	18190	18192	18194	18196	18198	18200	18202	18204	18206	18208	18210	18212	18214		
4241	4243	4245	4247	4249	4251	4253	4255	4257	4259	4261	4263	4265	4267	4269	4271	4273	4275	4277	4279	4281	4283	4285	4287	4289	4291	4293	4295	4297	4299	4301	4303	4305	4307	4309	4311	4313	4315	4317	4319	18057	18059	18061	18063	18065	18067	18069	18071	18073	18075	18077	18079	18081	18083	18085	18087	18089	18091	18093	18095	18097	18099	18101	18103	18105	18107	18109	18111	18113	18115	18117	18119	18121	18123	18125	18127	18129	18131	18133	18135		
4240	4242	4244	4246	4248	4250	4252	4254	4256	4258	4260	4262	4264	4266	4268	4270	4272	4274	4276	4278	4280	4282	4284	4286	4288	4290	4292	4294	4296	4298	4300	4302	4304	4306	4308	4310	4312	4314	4316	4318	18056	18058	18060	18062	18064	18066	18068	18070	18072	18074	18076	18078	18080	18082	18084	18086	18088	18090	18092	18094	18096	18098	18100	18102	18104	18106	18108	18110	18112	18114	18116	18118	18120	18122	18124	18126	18128	18130	18132	18134		
4161	4163	4165	4167	4169	4171	4173	4175	4177	4179	4181	4183	4185	4187	4189	4191	4193	4195	4197	4199	4201	4203	4205	4207	4209	4211	4213	4215	4217	4219	4221	4223	4225	4227	4229	4231	4233	4235	4237	4239	17977	17979	17981	17983	17985	17987	17989	17991	17993	17995	17997	17999	18001	18003	18005	18007	18009	18011	18013	18015	18017	18019	18021	18023	18025	18027	18029	18031	18033	18035	18037	18039	18041	18043	18045	18047	18049	18051	18053	18055		
4160	4162	4164	4166	4168	4170	4172	4174	4176	4178	4180	4182	4184	4186	4188	4190	4192	4194	4196	4198	4200	4202	4204	4206	4208	4210	4212	4214	4216	4218	4220	4222	4224	4226	4228	4230	4232	4234	4236	4238	17976	17978	17980	17982	17984	17986	17988	17990	17992	17994	17996	17998	18000	18002	18004	18006	18008	18010	18012	18014	18016	18018	18020	18022	18024	18026	18028	18030	18032	18034	18036	18038	18040	18042	18044	18046	18048	18050	18052	18054		
3201	3203	3205	3207	3209	3211	3213	3215	3217	3219	3221	3223	3225	3227	3229	3231	3233	3235	3237	3239	3241	3243	3245	3247	3249	3251	3253	3255	3257	3259	3261	3263	3265	3267	3269	3271	3273	3275	3277	3279	17017	17019	17021	17023	17025	17027	17029	17031	17033	17035	17037	17039	17041	17043	17045	17047	17049	17051	17053	17055	17057	17059	17061	17063	17065	17067	17069	17071	17073	17075	17077	17079	17081	17083	17085	17087	17089	17091	17093	17095		
3200	3202	3204	3206	3208	3210	3212	3214	3216	3218	3220	3222	3224	3226	3228	3230	3232	3234	3236	3238	3240	3242	3244	3246	3248	3250	3252	3254	3256	3258	3260	3262	3264	3266	3268	3270	3272	3274	3276	3278	17016	17018	17020	17022	17024	17026	17028	17030	17032	17034	17036	17038	17040	17042	17044	17046	17048	17050	17052	17054	17056	17058	17060	17062	17064	17066	17068	17070	17072	17074	17076	17078	17080	17082	17084	17086	17088	17090	17092	17094		
3281	3283	3285	3287	3289	3291	3293	3295	3297	3299	3301	3303	3305	3307	3309	3311	3313	3315	3317	3319	3321	3323	3325	3327	3329	3331	3333	3335	3337	3339	3341	3343	3345	3347	3349	3351	3353	3355	3357	3359	17097	17099	17101	17103	17105	17107	17109	17111	17113	17115	17117	17119	17121	17123	17125	17127	17129	17131	17133	17135	17137	17139	17141	17143	17145	17147	17149	17151	17153	17155	17157	17159	17161	17163	17165	17167	17169	17171	17173	17175		
3280	3282	3284	3286	3288	3290	3292	3294	3296	3298	3300	3302	3304	3306	3308	3310	3312	3314	3316	3318	3320	3322	3324	3326	3328	3330	3332	3334	3336	3338	3340	3342	3344	3346	3348	3350	3352	3354	3356	3358	17096	17098	17100	17102	17104	17106	17108	17110	17112	17114	17116	17118	17120	17122	17124	17126	17128	17130	17132	17134	17136	17138	17140	17142	17144	17146	17148	17150	17152	17154	17156	17158	17160	17162	17164	17166	17168	17170	17172	17174		
3361	3363	3365	3367	3369	3371	3373	3375	3377	3379	3381	3383	3385	3387	3389	3391	3393	3395	3397	3399	3401	3403	3405	3407	3409	3411	3413	3415	3417	3419	3421	3423	3425	3427	3429	3431	3433	3435	3437	3439	17177	17179	17181	17183	17185	17187	17189	17191	17193	17195	17197	17199	17201	17203	17205	17207	17209	17211	17213	17215	17217	17219	17221	17223	17225	17227	17229	17231	17233	17235	17237	17239	17241	17243	17245	17247	17249	17251	17253	17255		
3360	3362	3364	3366	3368	3370	3372	3374	3376	3378	3380	3382	3384	3386	3388	3390	3392	3394	3396	3398	3400	3402	3404	3406	3408	3410	3412	3414	3416	3418	3420	3422	3424	3426	3428	3430	3432	3434	3436	3438	17176	17178	17180	17182	17184	17186	17188	17190	17192	17194	17196	17198	17200	17202	17204	17206	17208	17210	17212	17214	17216	17218	17220	17222	17224	17226	17228	17230	17232	17234	17236	17238	17240	17242	17244	17246	17248	17250	17252	17254		
4081	4083	4085	4087	4089	4091	4093	4095	4097	4099	4101	4103	4105	4107	4109	4111	4113	4115	4117	4119	4121	4123	4125	4127	4129	4131	4133	4135	4137	4139	4141	4143	4145	4147	4149	4151	4153	4155	4157	4159	17897	17899	17901	17903	17905	17907	17909	17911	17913	17915	17917	17919	17921	17923	17925	17927	17929	17931	17933	17935	17937	17939	17941	17943	17945	17947	17949	17951	17953	17955	17957	17959	17961	17963	17965	17967	17969	17971	17973	17975		
4080	4082	4084	4086	4088	4090	4092	4094	4096	4098	4100	4102	4104	4106	4108	4110	4112	4114	4116	4118	4120	4122	4124	4126	4128	4130	4132	4134	4136	4138	4140	4142	4144	4146	4148	4150	4152	4154	4156	4158	17896	17898	17900	17902	17904	17906	17908	17910	17912	17914	17916	17918	17920	17922	17924	17926	17928	17930	17932	17934	17936	17938	17940	17942	17944	17946	17948	17950	17952	17954	17956	17958	17960	17962	17964	17966	17968	17970	17972	17974		
4001	4003	4005	4007	4009	4011	4013	4015	4017	4019	4021	4023	4025	4027	4029	4031	4033	4035	4037	4039	4041	4043	4045	4047	4049	4051	4053	4055	4057	4059	4061	4063	4065	4067	4069	4071	4073	4075	4077	4079	17817	17819	17821	17823	17825	17827	17829	17831	17833	17835	17837	17839	17841	17843	17845	17847	17849	17851	17853	17855	17857	17859	17861	17863	17865	17867	17869	17871	17873	17875	17877	17879	17881	17883	17885	17887	17889	17891	17893	17895		
4000	4002	4004	4006	4008	4010	4012	4014	4016	4018	4020	4022	4024	4026	4028	4030	4032	4034	4036	4038	4040	4042	4044	4046	4048	4050	4052	4054	4056	4058	4060	4062	4064	4066	4068	4070	4072	4074	4076	4078	17816	17818	17820	17822	17824	17826	17828	17830	17832	17834	17836	17838	17840	17842	17844	17846	17848	17850	17852	17854	17856	17858	17860	17862	17864	17866	17868	17870	17872	17874	17876	17878	17880	17882	17884	17886	17888	17890	17892	17894		
3921	3923	3925	3927	3929	3931	3933	3935	3937	3939	3941	3943	3945	3947	3949	3951	3953	3955	3957	3959	3961	3963	3965	3967	3969	3971	3973	3975	3977	3979	3981	3983	3985	3987	3989	3991	3993	3995	3997	3999	17737	17739	17741	17743	17745	17747	17749	17751	17753	17755	17757	17759	17761	17763	17765	17767	17769	17771	17773	17775	17777	17779	17781	17783	17785	17787	17789	17791	17793	17795	17797	17799	17801	17803	17805	17807	17809	17811	17813	17815		
3920	3922	3924	3926	3928	3930	3932	3934	3936	3938	3940	3942	3944	3946	3948	3950	3952	3954	3956	3958	3960	3962	3964	3966	3968	3970	3972	3974	3976	3978	3980	3982	3984	3986	3988	3990	3992	3994	3996	3998	17736	17738	17740	17742	17744	17746	17748	17750	17752	17754	17756	17758	17760	17762	17764	17766	17768	17770	17772	17774	17776	17778	17780	17782	17784	17786	17788	17790	17792	17794	17796	17798	17800	17802	17804	17806	17808	17810	17812	17814		
3441	3443	3445	3447	3449	3451	3453	3455	3457	3459	3461	3463	3465	3467	3469	3471	3473	3475	3477	3479	3481	3483	3485	3487	3489	3491	3493	3495	3497	3499	3501	3503	3505	3507	3509	3511	3513	3515	3517	3519	17257	17259	17261	17263	17265	17267	17269	17271	17273	17275	17277	17279	17281	17283	17285	17287	17289	17291	17293	17295	17297	17299	17301	17303	17305	17307	17309	17311	17313	17315	17317	17319	17321	17323	17325	17327	17329	17331	17333	17335		
3440	3442	3444	3446	3448	3450	3452	3454	3456	3458	3460	3462	3464	3466	3468	3470	3472	3474	3476	3478	3480	3482	3484	3486	3488	3490	3492	3494	3496	3498	3500	3502	3504	3506	3508	3510	3512	3514	3516	3518	17256	17258	17260	17262	17264	17266	17268	17270	17272	17274	17276	17278	17280	17282	17284	17286	17288	17290	17292	17294	17296	17298	17300	17302	17304	17306	17308	17310	17312	17314	17316	17318	17320	17322	17324	17326	17328	17330	17332	17334		
3521	3523	3525	3527	3529	3531	3533	3535	3537	3539	3541	3543	3545	3547	3549	3551	3553	3555	3557	3559	3561	3563	3565	3567	3569	3571	3573	3575	3577	3579	3581	3583	3585	3587	3589	3591	3593	3595	3597	3599	17337	17339	17341	17343	17345	17347	17349	17351	17353	17355	17357	17359	17361	17363	17365	17367	17369	17371	17373	17375	17377	17379	17381	17383	17385	17387	17389	17391	17393	17395	17397	17399	17401	17403	17405	17407	17409	17411	17413	17415		
3520	3522	3524	3526	3528	3530	3532	3534	3536	3538	3540	3542	3544	3546	3548	3550	3552	3554	3556	3558	3560	3562	3564	3566	3568	3570	3572	3574	3576	3578	3580	3582	3584	3586	3588	3590	3592	3594	3596	3598	17336	17338	17340	17342	17344	17346	17348	17350	17352	17354	17356	17358	17360	17362	17364	17366	17368	17370	17372	17374	17376	17378	17380	17382	17384	17386	17388	17390	17392	17394	17396	17398	17400	17402	17404	17406	17408	17410	17412	17414		
3601	3603	3605	3607	3609	3611	3613	3615	3617	3619	3621	3623	3625	3627	3629	3631	3633	3635	3637	3639	3641	3643	3645	3647	3649	3651	3653	3655	3657	3659	3661	3663	3665	3667	3669	3671	3673	3675	3677	3679	17417	17419	17421	17423	17425	17427	17429	17431	17433	17435	17437	17439	17441	17443	17445	17447	17449	17451	17453	17455	17457	17459	17461	17463	17465	17467	17469	17471	17473	17475	17477	17479	17481	17483	17485	17487	17489	17491	17493	17495		
3600	3602	3604	3606	3608	3610	3612	3614	3616	3618	3620	3622	3624	3626	3628	3630	3632	3634	3636	3638	3640	3642	3644	3646	3648	3650	3652	3654	3656	3658	3660	3662	3664	3666	3668	3670	3672	3674	3676	3678	17416	17418	17420	17422	17424	17426	17428	17430	17432	17434	17436	17438	17440	17442	17444	17446	17448	17450	17452	17454	17456	17458	17460	17462	17464	17466	17468	17470	17472	17474	17476	17478	17480	17482	17484	17486	17488	17490	17492	17494		
3841	3843	3845	3847	3849	3851	3853	3855	3857	3859	3861	3863	3865	3867	3869	3871	3873	3875	3877	3879	3881	3883	3885	3887	3889	3891	3893	3895	3897	3899	3901	3903	3905	3907	3909	3911	3913	3915	3917	3919	17657	17659	17661	17663	17665	17667	17669	17671	17673	17675	17677	17679	17681	17683	17685	17687	17689	17691	17693	17695	17697	17699	17701	17703	17705	17707	17709	17711	17713	17715	17717	17719	17721	17723	17725	17727	17729	17731	17733	17735		
3840	3842	3844	3846	3848	3850	3852	3854	3856	3858	3860	3862	3864	3866	3868	3870	3872	3874	3876	3878	3880	3882	3884	3886	3888	3890	3892	3894	3896	3898	3900	3902	3904	3906	3908	3910	3912	3914	3916	3918	17656	17658	17660	17662	17664	17666	17668	17670	17672	17674	17676	17678	17680	17682	17684	17686	17688	17690	17692	17694	17696	17698	17700	17702	17704	17706	17708	17710	17712	17714	17716	17718	17720	17722	17724	17726	17728	17730	17732	17734		
3761	3763	3765	3767	3769	3771	3773	3775	3777	3779	3781	3783	3785	3787	3789	3791	3793	3795	3797	3799	3801	3803	3805	3807	3809	3811	3813	3815	3817	3819	3821	3823	3825	3827	3829	3831	3833	3835	3837	3839	17577	17579	17581	17583	17585	17587	17589	17591	17593	17595	17597	17599	17601	17603	17605	17607	17609	17611	17613	17615	17617	17619	17621	17623	17625	17627	17629	17631	17633	17635	17637	17639	17641	17643	17645	17647	17649	17651	17653	17655		
3760	3762	3764	3766	3768	3770	3772	3774	3776	3778	3780	3782	3784	3786	3788	3790	3792	3794	3796	3798	3800	3802	3804	3806	3808	3810	3812	3814	3816	3818	3820	3822	3824	3826	3828	3830	3832	3834	3836	3838	17576	17578	17580	17582	17584	17586	17588	17590	17592	17594	17596	17598	17600	17602	17604	17606	17608	17610	17612	17614	17616	17618	17620	17622	17624	17626	17628	17630	17632	17634	17636	17638	17640	17642	17644	17646	17648	17650	17652	17654		
3681	3683	3685	3687	3689	3691	3693	3695	3697	3699	3701	3703	3705	3707	3709	3711	3713	3715	3717	3719	3721	3723	3725	3727	3729	3731	3733	3735	3737	3739	3741	3743	3745	3747	3749	3751	3753	3755	3757	3759	17497	17499	17501	17503	17505	17507	17509	17511	17513	17515	17517	17519	17521	17523	17525	17527	17529	17531	17533	17535	17537	17539	17541	17543	17545	17547	17549	17551	17553	17555	17557	17559	17561	17563	17565	17567	17569	17571	17573	17575		
3680	3682	3684	3686	3688	3690	3692	3694	3696	3698	3700	3702	3704	3706	3708	3710	3712	3714	3716	3718	3720	3722	3724	3726	3728	3730	3732	3734	3736	3738	3740	3742	3744	3746	3748	3750	3752	3754	3756	3758	17496	17498	17500	17502	17504	17506	17508	17510	17512	17514	17516	17518	17520	17522	17524	17526	17528	17530	17532	17534	17536	17538	17540	17542	17544	17546	17548	17550	17552	17554	17556	17558	17560	17562	17564	17566	17568	17570	17572	17574		
27	55	83	111	139	167	195	223	251	279	307	335	363	391	419	447	475	503	531	559	587	615	643	671	699	727	755	783	811	839	867	895	923	951	979	1007	1035	1063	1091	1119	13843	13871	13899	13927	13955	13983	14011	14039	14067	14095	14123	14151	14179	14207	14235	14263	14291	14319	14347	14375	14403	14431	14459	14487	14515	14543	14571	14599	14627	14655	14683	14711	14739	14767	14795	14823	14851	14879	14907	14935		
6935	6963	6991	7019	7047	7075	7103	7131	7159	7187	7215	7243	7271	7299	7327	7355	7383	7411	7439	7467	7495	7523	7551	7579	7607	7635	7663	7691	7719	7747	7775	7803	7831	7859	7887	7915	7943	7971	7999	8027	20764	20792	20820	20848	20876	20904	20932	20960	20988	21016	21044	21072	21100	21128	21156	21184	21212	21240	21268	21296	21324	21352	21380	21408	21436	21464	21492	21520	21548	21576	21604	21632	21660	21688	21716	21744	21772	21800	21828	21856		
26	54	82	110	138	166	194	222	250	278	306	334	362	390	418	446	474	502	530	558	586	614	642	670	698	726	754	782	810	838	866	894	922	950	978	1006	1034	1062	1090	1118	13842	13870	13898	13926	13954	13982	14010	14038	14066	14094	14122	14150	14178	14206	14234	14262	14290	14318	14346	14374	14402	14430	14458	14486	14514	14542	14570	14598	14626	14654	14682	14710	14738	14766	14794	14822	14850	14878	14906	14934		
6934	6962	6990	7018	7046	7074	7102	7130	7158	7186	7214	7242	7270	7298	7326	7354	7382	7410	7438	7466	7494	7522	7550	7578	7606	7634	7662	7690	7718	7746	7774	7802	7830	7858	7886	7914	7942	7970	7998	8026	20763	20791	20819	20847	20875	20903	20931	20959	20987	21015	21043	21071	21099	21127	21155	21183	21211	21239	21267	21295	21323	21351	21379	21407	21435	21463	21491	21519	21547	21575	21603	21631	21659	21687	21715	21743	21771	21799	21827	21855		
25	53	81	109	137	165	193	221	249	277	305	333	361	389	417	445	473	501	529	557	585	613	641	669	697	725	753	781	809	837	865	893	921	949	977	1005	1033	1061	1089	1117	13841	13869	13897	13925	13953	13981	14009	14037	14065	14093	14121	14149	14177	14205	14233	14261	14289	14317	14345	14373	14401	14429	14457	14485	14513	14541	14569	14597	14625	14653	14681	14709	14737	14765	14793	14821	14849	14877	14905	14933		
6933	6961	6989	7017	7045	7073	7101	7129	7157	7185	7213	7241	7269	7297	7325	7353	7381	7409	7437	7465	7493	7521	7549	7577	7605	7633	7661	7689	7717	7745	7773	7801	7829	7857	7885	7913	7941	7969	7997	8025	20762	20790	20818	20846	20874	20902	20930	20958	20986	21014	21042	21070	21098	21126	21154	21182	21210	21238	21266	21294	21322	21350	21378	21406	21434	21462	21490	21518	21546	21574	21602	21630	21658	21686	21714	21742	21770	21798	21826	21854		
24	52	80	108	136	164	192	220	248	276	304	332	360	388	416	444	472	500	528	556	584	612	640	668	696	724	752	780	808	836	864	892	920	948	976	1004	1032	1060	1088	1116	13840	13868	13896	13924	13952	13980	14008	14036	14064	14092	14120	14148	14176	14204	14232	14260	14288	14316	14344	14372	14400	14428	14456	14484	14512	14540	14568	14596	14624	14652	14680	14708	14736	14764	14792	14820	14848	14876	14904	14932		
6932	6960	6988	7016	7044	7072	7100	7128	7156	7184	7212	7240	7268	7296	7324	7352	7380	7408	7436	7464	7492	7520	7548	7576	7604	7632	7660	7688	7716	7744	7772	7800	7828	7856	7884	7912	7940	7968	7996	8024	20761	20789	20817	20845	20873	20901	20929	20957	20985	21013	21041	21069	21097	21125	21153	21181	21209	21237	21265	21293	21321	21349	21377	21405	21433	21461	21489	21517	21545	21573	21601	21629	21657	21685	21713	21741	21769	21797	21825	21853		
23	51	79	107	135	163	191	219	247	275	303	331	359	387	415	443	471	499	527	555	583	611	639	667	695	723	751	779	807	835	863	891	919	947	975	1003	1031	1059	1087	1115	13839	13867	13895	13923	13951	13979	14007	14035	14063	14091	14119	14147	14175	14203	14231	14259	14287	14315	14343	14371	14399	14427	14455	14483	14511	14539	14567	14595	14623	14651	14679	14707	14735	14763	14791	14819	14847	14875	14903	14931		
6931	6959	6987	7015	7043	7071	7099	7127	7155	7183	7211	7239	7267	7295	7323	7351	7379	7407	7435	7463	7491	7519	7547	7575	7603	7631	7659	7687	7715	7743	7771	7799	7827	7855	7883	7911	7939	7967	7995	8023	20760	20788	20816	20844	20872	20900	20928	20956	20984	21012	21040	21068	21096	21124	21152	21180	21208	21236	21264	21292	21320	21348	21376	21404	21432	21460	21488	21516	21544	21572	21600	21628	21656	21684	21712	21740	21768	21796	21824	21852		
22	50	78	106	134	162	190	218	246	274	302	330	358	386	414	442	470	498	526	554	582	610	638	666	694	722	750	778	806	834	862	890	918	946	974	1002	1030	1058	1086	1114	13838	13866	13894	13922	13950	13978	14006	14034	14062	14090	14118	14146	14174	14202	14230	14258	14286	14314	14342	14370	14398	14426	14454	14482	14510	14538	14566	14594	14622	14650	14678	14706	14734	14762	14790	14818	14846	14874	14902	14930		
6930	6958	6986	7014	7042	7070	7098	7126	7154	7182	7210	7238	7266	7294	7322	7350	7378	7406	7434	7462	7490	7518	7546	7574	7602	7630	7658	7686	7714	7742	7770	7798	7826	7854	7882	7910	7938	7966	7994	8022	20759	20787	20815	20843	20871	20899	20927	20955	20983	21011	21039	21067	21095	21123	21151	21179	21207	21235	21263	21291	21319	21347	21375	21403	21431	21459	21487	21515	21543	21571	21599	21627	21655	21683	21711	21739	21767	21795	21823	21851		
21	49	77	105	133	161	189	217	245	273	301	329	357	385	413	441	469	497	525	553	581	609	637	665	693	721	749	777	805	833	861	889	917	945	973	1001	1029	1057	1085	1113	13837	13865	13893	13921	13949	13977	14005	14033	14061	14089	14117	14145	14173	14201	14229	14257	14285	14313	14341	14369	14397	14425	14453	14481	14509	14537	14565	14593	14621	14649	14677	14705	14733	14761	14789	14817	14845	14873	14901	14929		
6929	6957	6985	7013	7041	7069	7097	7125	7153	7181	7209	7237	7265	7293	7321	7349	7377	7405	7433	7461	7489	7517	7545	7573	7601	7629	7657	7685	7713	7741	7769	7797	7825	7853	7881	7909	7937	7965	7993	8021	20758	20786	20814	20842	20870	20898	20926	20954	20982	21010	21038	21066	21094	21122	21150	21178	21206	21234	21262	21290	21318	21346	21374	21402	21430	21458	21486	21514	21542	21570	21598	21626	21654	21682	21710	21738	21766	21794	21822	21850		
20	48	76	104	132	160	188	216	244	272	300	328	356	384	412	440	468	496	524	552	580	608	636	664	692	720	748	776	804	832	860	888	916	944	972	1000	1028	1056	1084	1112	13836	13864	13892	13920	13948	13976	14004	14032	14060	14088	14116	14144	14172	14200	14228	14256	14284	14312	14340	14368	14396	14424	14452	14480	14508	14536	14564	14592	14620	14648	14676	14704	14732	14760	14788	14816	14844	14872	14900	14928		
6928	6956	6984	7012	7040	7068	7096	7124	7152	7180	7208	7236	7264	7292	7320	7348	7376	7404	7432	7460	7488	7516	7544	7572	7600	7628	7656	7684	7712	7740	7768	7796	7824	7852	7880	7908	7936	7964	7992	8020	20757	20785	20813	20841	20869	20897	20925	20953	20981	21009	21037	21065	21093	21121	21149	21177	21205	21233	21261	21289	21317	21345	21373	21401	21429	21457	21485	21513	21541	21569	21597	21625	21653	21681	21709	21737	21765	21793	21821	21849		
19	47	75	103	131	159	187	215	243	271	299	327	355	383	411	439	467	495	523	551	579	607	635	663	691	719	747	775	803	831	859	887	915	943	971	999	1027	1055	1083	1111	13835	13863	13891	13919	13947	13975	14003	14031	14059	14087	14115	14143	14171	14199	14227	14255	14283	14311	14339	14367	14395	14423	14451	14479	14507	14535	14563	14591	14619	14647	14675	14703	14731	14759	14787	14815	14843	14871	14899	14927		
6927	6955	6983	7011	7039	7067	7095	7123	7151	7179	7207	7235	7263	7291	7319	7347	7375	7403	7431	7459	7487	7515	7543	7571	7599	7627	7655	7683	7711	7739	7767	7795	7823	7851	7879	7907	7935	7963	7991	8019	20756	20784	20812	20840	20868	20896	20924	20952	20980	21008	21036	21064	21092	21120	21148	21176	21204	21232	21260	21288	21316	21344	21372	21400	21428	21456	21484	21512	21540	21568	21596	21624	21652	21680	21708	21736	21764	21792	21820	21848		
18	46	74	102	130	158	186	214	242	270	298	326	354	382	410	438	466	494	522	550	578	606	634	662	690	718	746	774	802	830	858	886	914	942	970	998	1026	1054	1082	1110	13834	13862	13890	13918	13946	13974	14002	14030	14058	14086	14114	14142	14170	14198	14226	14254	14282	14310	14338	14366	14394	14422	14450	14478	14506	14534	14562	14590	14618	14646	14674	14702	14730	14758	14786	14814	14842	14870	14898	14926		
6926	6954	6982	7010	7038	7066	7094	7122	7150	7178	7206	7234	7262	7290	7318	7346	7374	7402	7430	7458	7486	7514	7542	7570	7598	7626	7654	7682	7710	7738	7766	7794	7822	7850	7878	7906	7934	7962	7990	8018	20755	20783	20811	20839	20867	20895	20923	20951	20979	21007	21035	21063	21091	21119	21147	21175	21203	21231	21259	21287	21315	21343	21371	21399	21427	21455	21483	21511	21539	21567	21595	21623	21651	21679	21707	21735	21763	21791	21819	21847		
17	45	73	101	129	157	185	213	241	269	297	325	353	381	409	437	465	493	521	549	577	605	633	661	689	717	745	773	801	829	857	885	913	941	969	997	1025	1053	1081	1109	13833	13861	13889	13917	13945	13973	14001	14029	14057	14085	14113	14141	14169	14197	14225	14253	14281	14309	14337	14365	14393	14421	14449	14477	14505	14533	14561	14589	14617	14645	14673	14701	14729	14757	14785	14813	14841	14869	14897	14925		
6925	6953	6981	7009	7037	7065	7093	7121	7149	7177	7205	7233	7261	7289	7317	7345	7373	7401	7429	7457	7485	7513	7541	7569	7597	7625	7653	7681	7709	7737	7765	7793	7821	7849	7877	7905	7933	7961	7989	8017	20754	20782	20810	20838	20866	20894	20922	20950	20978	21006	21034	21062	21090	21118	21146	21174	21202	21230	21258	21286	21314	21342	21370	21398	21426	21454	21482	21510	21538	21566	21594	21622	21650	21678	21706	21734	21762	21790	21818	21846		
16	44	72	100	128	156	184	212	240	268	296	324	352	380	408	436	464	492	520	548	576	604	632	660	688	716	744	772	800	828	856	884	912	940	968	996	1024	1052	1080	1108	13832	13860	13888	13916	13944	13972	14000	14028	14056	14084	14112	14140	14168	14196	14224	14252	14280	14308	14336	14364	14392	14420	14448	14476	14504	14532	14560	14588	14616	14644	14672	14700	14728	14756	14784	14812	14840	14868	14896	14924		
6924	6952	6980	7008	7036	7064	7092	7120	7148	7176	7204	7232	7260	7288	7316	7344	7372	7400	7428	7456	7484	7512	7540	7568	7596	7624	7652	7680	7708	7736	7764	7792	7820	7848	7876	7904	7932	7960	7988	8016	20753	20781	20809	20837	20865	20893	20921	20949	20977	21005	21033	21061	21089	21117	21145	21173	21201	21229	21257	21285	21313	21341	21369	21397	21425	21453	21481	21509	21537	21565	21593	21621	21649	21677	21705	21733	21761	21789	21817	21845		
15	43	71	99	127	155	183	211	239	267	295	323	351	379	407	435	463	491	519	547	575	603	631	659	687	715	743	771	799	827	855	883	911	939	967	995	1023	1051	1079	1107	13831	13859	13887	13915	13943	13971	13999	14027	14055	14083	14111	14139	14167	14195	14223	14251	14279	14307	14335	14363	14391	14419	14447	14475	14503	14531	14559	14587	14615	14643	14671	14699	14727	14755	14783	14811	14839	14867	14895	14923		
6923	6951	6979	7007	7035	7063	7091	7119	7147	7175	7203	7231	7259	7287	7315	7343	7371	7399	7427	7455	7483	7511	7539	7567	7595	7623	7651	7679	7707	7735	7763	7791	7819	7847	7875	7903	7931	7959	7987	8015	20752	20780	20808	20836	20864	20892	20920	20948	20976	21004	21032	21060	21088	21116	21144	21172	21200	21228	21256	21284	21312	21340	21368	21396	21424	21452	21480	21508	21536	21564	21592	21620	21648	21676	21704	21732	21760	21788	21816	21844		
14	42	70	98	126	154	182	210	238	266	294	322	350	378	406	434	462	490	518	546	574	602	630	658	686	714	742	770	798	826	854	882	910	938	966	994	1022	1050	1078	1106	13830	13858	13886	13914	13942	13970	13998	14026	14054	14082	14110	14138	14166	14194	14222	14250	14278	14306	14334	14362	14390	14418	14446	14474	14502	14530	14558	14586	14614	14642	14670	14698	14726	14754	14782	14810	14838	14866	14894	14922		
6922	6950	6978	7006	7034	7062	7090	7118	7146	7174	7202	7230	7258	7286	7314	7342	7370	7398	7426	7454	7482	7510	7538	7566	7594	7622	7650	7678	7706	7734	7762	7790	7818	7846	7874	7902	7930	7958	7986	8014	20751	20779	20807	20835	20863	20891	20919	20947	20975	21003	21031	21059	21087	21115	21143	21171	21199	21227	21255	21283	21311	21339	21367	21395	21423	21451	21479	21507	21535	21563	21591	21619	21647	21675	21703	21731	21759	21787	21815	21843		
13	41	69	97	125	153	181	209	237	265	293	321	349	377	405	433	461	489	517	545	573	601	629	657	685	713	741	769	797	825	853	881	909	937	965	993	1021	1049	1077	1105	13829	13857	13885	13913	13941	13969	13997	14025	14053	14081	14109	14137	14165	14193	14221	14249	14277	14305	14333	14361	14389	14417	14445	14473	14501	14529	14557	14585	14613	14641	14669	14697	14725	14753	14781	14809	14837	14865	14893	14921		
6921	6949	6977	7005	7033	7061	7089	7117	7145	7173	7201	7229	7257	7285	7313	7341	7369	7397	7425	7453	7481	7509	7537	7565	7593	7621	7649	7677	7705	7733	7761	7789	7817	7845	7873	7901	7929	7957	7985	8013	20750	20778	20806	20834	20862	20890	20918	20946	20974	21002	21030	21058	21086	21114	21142	21170	21198	21226	21254	21282	21310	21338	21366	21394	21422	21450	21478	21506	21534	21562	21590	21618	21646	21674	21702	21730	21758	21786	21814	21842		
12	40	68	96	124	152	180	208	236	264	292	320	348	376	404	432	460	488	516	544	572	600	628	656	684	712	740	768	796	824	852	880	908	936	964	992	1020	1048	1076	1104	13828	13856	13884	13912	13940	13968	13996	14024	14052	14080	14108	14136	14164	14192	14220	14248	14276	14304	14332	14360	14388	14416	14444	14472	14500	14528	14556	14584	14612	14640	14668	14696	14724	14752	14780	14808	14836	14864	14892	14920		
6920	6948	6976	7004	7032	7060	7088	7116	7144	7172	7200	7228	7256	7284	7312	7340	7368	7396	7424	7452	7480	7508	7536	7564	7592	7620	7648	7676	7704	7732	7760	7788	7816	7844	7872	7900	7928	7956	7984	8012	20749	20777	20805	20833	20861	20889	20917	20945	20973	21001	21029	21057	21085	21113	21141	21169	21197	21225	21253	21281	21309	21337	21365	21393	21421	21449	21477	21505	21533	21561	21589	21617	21645	21673	21701	21729	21757	21785	21813	21841		
11	39	67	95	123	151	179	207	235	263	291	319	347	375	403	431	459	487	515	543	571	599	627	655	683	711	739	767	795	823	851	879	907	935	963	991	1019	1047	1075	1103	13827	13855	13883	13911	13939	13967	13995	14023	14051	14079	14107	14135	14163	14191	14219	14247	14275	14303	14331	14359	14387	14415	14443	14471	14499	14527	14555	14583	14611	14639	14667	14695	14723	14751	14779	14807	14835	14863	14891	14919		
6919	6947	6975	7003	7031	7059	7087	7115	7143	7171	7199	7227	7255	7283	7311	7339	7367	7395	7423	7451	7479	7507	7535	7563	7591	7619	7647	7675	7703	7731	7759	7787	7815	7843	7871	7899	7927	7955	7983	8011	20748	20776	20804	20832	20860	20888	20916	20944	20972	21000	21028	21056	21084	21112	21140	21168	21196	21224	21252	21280	21308	21336	21364	21392	21420	21448	21476	21504	21532	21560	21588	21616	21644	21672	21700	21728	21756	21784	21812	21840		
10	38	66	94	122	150	178	206	234	262	290	318	346	374	402	430	458	486	514	542	570	598	626	654	682	710	738	766	794	822	850	878	906	934	962	990	1018	1046	1074	1102	13826	13854	13882	13910	13938	13966	13994	14022	14050	14078	14106	14134	14162	14190	14218	14246	14274	14302	14330	14358	14386	14414	14442	14470	14498	14526	14554	14582	14610	14638	14666	14694	14722	14750	14778	14806	14834	14862	14890	14918		
6918	6946	6974	7002	7030	7058	7086	7114	7142	7170	7198	7226	7254	7282	7310	7338	7366	7394	7422	7450	7478	7506	7534	7562	7590	7618	7646	7674	7702	7730	7758	7786	7814	7842	7870	7898	7926	7954	7982	8010	20747	20775	20803	20831	20859	20887	20915	20943	20971	20999	21027	21055	21083	21111	21139	21167	21195	21223	21251	21279	21307	21335	21363	21391	21419	21447	21475	21503	21531	21559	21587	21615	21643	21671	21699	21727	21755	21783	21811	21839		
9	37	65	93	121	149	177	205	233	261	289	317	345	373	401	429	457	485	513	541	569	597	625	653	681	709	737	765	793	821	849	877	905	933	961	989	1017	1045	1073	1101	13825	13853	13881	13909	13937	13965	13993	14021	14049	14077	14105	14133	14161	14189	14217	14245	14273	14301	14329	14357	14385	14413	14441	14469	14497	14525	14553	14581	14609	14637	14665	14693	14721	14749	14777	14805	14833	14861	14889	14917		
6917	6945	6973	7001	7029	7057	7085	7113	7141	7169	7197	7225	7253	7281	7309	7337	7365	7393	7421	7449	7477	7505	7533	7561	7589	7617	7645	7673	7701	7729	7757	7785	7813	7841	7869	7897	7925	7953	7981	8009	20746	20774	20802	20830	20858	20886	20914	20942	20970	20998	21026	21054	21082	21110	21138	21166	21194	21222	21250	21278	21306	21334	21362	21390	21418	21446	21474	21502	21530	21558	21586	21614	21642	21670	21698	21726	21754	21782	21810	21838		
8	36	64	92	120	148	176	204	232	260	288	316	344	372	400	428	456	484	512	540	568	596	624	652	680	708	736	764	792	820	848	876	904	932	960	988	1016	1044	1072	1100	13824	13852	13880	13908	13936	13964	13992	14020	14048	14076	14104	14132	14160	14188	14216	14244	14272	14300	14328	14356	14384	14412	14440	14468	14496	14524	14552	14580	14608	14636	14664	14692	14720	14748	14776	14804	14832	14860	14888	14916		
6916	6944	6972	7000	7028	7056	7084	7112	7140	7168	7196	7224	7252	7280	7308	7336	7364	7392	7420	7448	7476	7504	7532	7560	7588	7616	7644	7672	7700	7728	7756	7784	7812	7840	7868	7896	7924	7952	7980	8008	20745	20773	20801	20829	20857	20885	20913	20941	20969	20997	21025	21053	21081	21109	21137	21165	21193	21221	21249	21277	21305	21333	21361	21389	21417	21445	21473	21501	21529	21557	21585	21613	21641	21669	21697	21725	21753	21781	21809	21837		
7	35	63	91	119	147	175	203	231	259	287	315	343	371	399	427	455	483	511	539	567	595	623	651	679	707	735	763	791	819	847	875	903	931	959	987	1015	1043	1071	1099	13823	13851	13879	13907	13935	13963	13991	14019	14047	14075	14103	14131	14159	14187	14215	14243	14271	14299	14327	14355	14383	14411	14439	14467	14495	14523	14551	14579	14607	14635	14663	14691	14719	14747	14775	14803	14831	14859	14887	14915		
6915	6943	6971	6999	7027	7055	7083	7111	7139	7167	7195	7223	7251	7279	7307	7335	7363	7391	7419	7447	7475	7503	7531	7559	7587	7615	7643	7671	7699	7727	7755	7783	7811	7839	7867	7895	7923	7951	7979	8007	20744	20772	20800	20828	20856	20884	20912	20940	20968	20996	21024	21052	21080	21108	21136	21164	21192	21220	21248	21276	21304	21332	21360	21388	21416	21444	21472	21500	21528	21556	21584	21612	21640	21668	21696	21724	21752	21780	21808	21836		
6	34	62	90	118	146	174	202	230	258	286	314	342	370	398	426	454	482	510	538	566	594	622	650	678	706	734	762	790	818	846	874	902	930	958	986	1014	1042	1070	1098	13822	13850	13878	13906	13934	13962	13990	14018	14046	14074	14102	14130	14158	14186	14214	14242	14270	14298	14326	14354	14382	14410	14438	14466	14494	14522	14550	14578	14606	14634	14662	14690	14718	14746	14774	14802	14830	14858	14886	14914		
6914	6942	6970	6998	7026	7054	7082	7110	7138	7166	7194	7222	7250	7278	7306	7334	7362	7390	7418	7446	7474	7502	7530	7558	7586	7614	7642	7670	7698	7726	7754	7782	7810	7838	7866	7894	7922	7950	7978	8006	20743	20771	20799	20827	20855	20883	20911	20939	20967	20995	21023	21051	21079	21107	21135	21163	21191	21219	21247	21275	21303	21331	21359	21387	21415	21443	21471	21499	21527	21555	21583	21611	21639	21667	21695	21723	21751	21779	21807	21835		
5	33	61	89	117	145	173	201	229	257	285	313	341	369	397	425	453	481	509	537	565	593	621	649	677	705	733	761	789	817	845	873	901	929	957	985	1013	1041	1069	1097	13821	13849	13877	13905	13933	13961	13989	14017	14045	14073	14101	14129	14157	14185	14213	14241	14269	14297	14325	14353	14381	14409	14437	14465	14493	14521	14549	14577	14605	14633	14661	14689	14717	14745	14773	14801	14829	14857	14885	14913		
6913	6941	6969	6997	7025	7053	7081	7109	7137	7165	7193	7221	7249	7277	7305	7333	7361	7389	7417	7445	7473	7501	7529	7557	7585	7613	7641	7669	7697	7725	7753	7781	7809	7837	7865	7893	7921	7949	7977	8005	20742	20770	20798	20826	20854	20882	20910	20938	20966	20994	21022	21050	21078	21106	21134	21162	21190	21218	21246	21274	21302	21330	21358	21386	21414	21442	21470	21498	21526	21554	21582	21610	21638	21666	21694	21722	21750	21778	21806	21834		
4	32	60	88	116	144	172	200	228	256	284	312	340	368	396	424	452	480	508	536	564	592	620	648	676	704	732	760	788	816	844	872	900	928	956	984	1012	1040	1068	1096	13820	13848	13876	13904	13932	13960	13988	14016	14044	14072	14100	14128	14156	14184	14212	14240	14268	14296	14324	14352	14380	14408	14436	14464	14492	14520	14548	14576	14604	14632	14660	14688	14716	14744	14772	14800	14828	14856	14884	14912		
6912	6940	6968	6996	7024	7052	7080	7108	7136	7164	7192	7220	7248	7276	7304	7332	7360	7388	7416	7444	7472	7500	7528	7556	7584	7612	7640	7668	7696	7724	7752	7780	7808	7836	7864	7892	7920	7948	7976	8004	20741	20769	20797	20825	20853	20881	20909	20937	20965	20993	21021	21049	21077	21105	21133	21161	21189	21217	21245	21273	21301	21329	21357	21385	21413	21441	21469	21497	21525	21553	21581	21609	21637	21665	21693	21721	21749	21777	21805	21833		
3	31	59	87	115	143	171	199	227	255	283	311	339	367	395	423	451	479	507	535	563	591	619	647	675	703	731	759	787	815	843	871	899	927	955	983	1011	1039	1067	1095	13819	13847	13875	13903	13931	13959	13987	14015	14043	14071	14099	14127	14155	14183	14211	14239	14267	14295	14323	14351	14379	14407	14435	14463	14491	14519	14547	14575	14603	14631	14659	14687	14715	14743	14771	14799	14827	14855	14883	14911		
6911	6939	6967	6995	7023	7051	7079	7107	7135	7163	7191	7219	7247	7275	7303	7331	7359	7387	7415	7443	7471	7499	7527	7555	7583	7611	7639	7667	7695	7723	7751	7779	7807	7835	7863	7891	7919	7947	7975	8003	20740	20768	20796	20824	20852	20880	20908	20936	20964	20992	21020	21048	21076	21104	21132	21160	21188	21216	21244	21272	21300	21328	21356	21384	21412	21440	21468	21496	21524	21552	21580	21608	21636	21664	21692	21720	21748	21776	21804	21832		
2	30	58	86	114	142	170	198	226	254	282	310	338	366	394	422	450	478	506	534	562	590	618	646	674	702	730	758	786	814	842	870	898	926	954	982	1010	1038	1066	1094	13818	13846	13874	13902	13930	13958	13986	14014	14042	14070	14098	14126	14154	14182	14210	14238	14266	14294	14322	14350	14378	14406	14434	14462	14490	14518	14546	14574	14602	14630	14658	14686	14714	14742	14770	14798	14826	14854	14882	14910		
6910	6938	6966	6994	7022	7050	7078	7106	7134	7162	7190	7218	7246	7274	7302	7330	7358	7386	7414	7442	7470	7498	7526	7554	7582	7610	7638	7666	7694	7722	7750	7778	7806	7834	7862	7890	7918	7946	7974	8002	20739	20767	20795	20823	20851	20879	20907	20935	20963	20991	21019	21047	21075	21103	21131	21159	21187	21215	21243	21271	21299	21327	21355	21383	21411	21439	21467	21495	21523	21551	21579	21607	21635	21663	21691	21719	21747	21775	21803	21831		
1	29	57	85	113	141	169	197	225	253	281	309	337	365	393	421	449	477	505	533	561	589	617	645	673	701	729	757	785	813	841	869	897	925	953	981	1009	1037	1065	1093	13817	13845	13873	13901	13929	13957	13985	14013	14041	14069	14097	14125	14153	14181	14209	14237	14265	14293	14321	14349	14377	14405	14433	14461	14489	14517	14545	14573	14601	14629	14657	14685	14713	14741	14769	14797	14825	14853	14881	14909		
6909	6937	6965	6993	7021	7049	7077	7105	7133	7161	7189	7217	7245	7273	7301	7329	7357	7385	7413	7441	7469	7497	7525	7553	7581	7609	7637	7665	7693	7721	7749	7777	7805	7833	7861	7889	7917	7945	7973	8001	20738	20766	20794	20822	20850	20878	20906	20934	20962	20990	21018	21046	21074	21102	21130	21158	21186	21214	21242	21270	21298	21326	21354	21382	21410	21438	21466	21494	21522	21550	21578	21606	21634	21662	21690	21718	21746	21774	21802	21830		
0	28	56	84	112	140	168	196	224	252	280	308	336	364	392	420	448	476	504	532	560	588	616	644	672	700	728	756	784	812	840	868	896	924	952	980	1008	1036	1064	1092	13816	13844	13872	13900	13928	13956	13984	14012	14040	14068	14096	14124	14152	14180	14208	14236	14264	14292	14320	14348	14376	14404	14432	14460	14488	14516	14544	14572	14600	14628	14656	14684	14712	14740	14768	14796	14824	14852	14880	14908		
6908	6936	6964	6992	7020	7048	7076	7104	7132	7160	7188	7216	7244	7272	7300	7328	7356	7384	7412	7440	7468	7496	7524	7552	7580	7608	7636	7664	7692	7720	7748	7776	7804	7832	7860	7888	7916	7944	7972	8000	20737	20765	20793	20821	20849	20877	20905	20933	20961	20989	21017	21045	21073	21101	21129	21157	21185	21213	21241	21269	21297	21325	21353	21381	21409	21437	21465	21493	21521	21549	21577	21605	21633	21661	21689	21717	21745	21773	21801	21829		
10588	10590	10592	10594	10596	10598	10600	10602	10604	10606	10608	10610	10612	10614	10616	10618	10620	10622	10624	10626	10628	10630	10632	10634	10636	10638	10640	10642	10644	10646	10648	10650	10652	10654	10656	10658	10660	10662	10664	10666	24417	24419	24421	24423	24425	24427	24429	24431	24433	24435	24437	24439	24441	24443	24445	24447	24449	24451	24453	24455	24457	24459	24461	24463	24465	24467	24469	24471	24473	24475	24477	24479	24481	24483	24485	24487	24489	24491	24493	24495		
10589	10591	10593	10595	10597	10599	10601	10603	10605	10607	10609	10611	10613	10615	10617	10619	10621	10623	10625	10627	10629	10631	10633	10635	10637	10639	10641	10643	10645	10647	10649	10651	10653	10655	10657	10659	10661	10663	10665	10667	24418	24420	24422	24424	24426	24428	24430	24432	24434	24436	24438	24440	24442	24444	24446	24448	24450	24452	24454	24456	24458	24460	24462	24464	24466	24468	24470	24472	24474	24476	24478	24480	24482	24484	24486	24488	24490	24492	24494	24496		
10668	10670	10672	10674	10676	10678	10680	10682	10684	10686	10688	10690	10692	10694	10696	10698	10700	10702	10704	10706	10708	10710	10712	10714	10716	10718	10720	10722	10724	10726	10728	10730	10732	10734	10736	10738	10740	10742	10744	10746	24497	24499	24501	24503	24505	24507	24509	24511	24513	24515	24517	24519	24521	24523	24525	24527	24529	24531	24533	24535	24537	24539	24541	24543	24545	24547	24549	24551	24553	24555	24557	24559	24561	24563	24565	24567	24569	24571	24573	24575		
10669	10671	10673	10675	10677	10679	10681	10683	10685	10687	10689	10691	10693	10695	10697	10699	10701	10703	10705	10707	10709	10711	10713	10715	10717	10719	10721	10723	10725	10727	10729	10731	10733	10735	10737	10739	10741	10743	10745	10747	24498	24500	24502	24504	24506	24508	24510	24512	24514	24516	24518	24520	24522	24524	24526	24528	24530	24532	24534	24536	24538	24540	24542	24544	24546	24548	24550	24552	24554	24556	24558	24560	24562	24564	24566	24568	24570	24572	24574	24576		
10748	10750	10752	10754	10756	10758	10760	10762	10764	10766	10768	10770	10772	10774	10776	10778	10780	10782	10784	10786	10788	10790	10792	10794	10796	10798	10800	10802	10804	10806	10808	10810	10812	10814	10816	10818	10820	10822	10824	10826	24577	24579	24581	24583	24585	24587	24589	24591	24593	24595	24597	24599	24601	24603	24605	24607	24609	24611	24613	24615	24617	24619	24621	24623	24625	24627	24629	24631	24633	24635	24637	24639	24641	24643	24645	24647	24649	24651	24653	24655		
10749	10751	10753	10755	10757	10759	10761	10763	10765	10767	10769	10771	10773	10775	10777	10779	10781	10783	10785	10787	10789	10791	10793	10795	10797	10799	10801	10803	10805	10807	10809	10811	10813	10815	10817	10819	10821	10823	10825	10827	24578	24580	24582	24584	24586	24588	24590	24592	24594	24596	24598	24600	24602	24604	24606	24608	24610	24612	24614	24616	24618	24620	24622	24624	24626	24628	24630	24632	24634	24636	24638	24640	24642	24644	24646	24648	24650	24652	24654	24656		
10508	10510	10512	10514	10516	10518	10520	10522	10524	10526	10528	10530	10532	10534	10536	10538	10540	10542	10544	10546	10548	10550	10552	10554	10556	10558	10560	10562	10564	10566	10568	10570	10572	10574	10576	10578	10580	10582	10584	10586	24337	24339	24341	24343	24345	24347	24349	24351	24353	24355	24357	24359	24361	24363	24365	24367	24369	24371	24373	24375	24377	24379	24381	24383	24385	24387	24389	24391	24393	24395	24397	24399	24401	24403	24405	24407	24409	24411	24413	24415		
10509	10511	10513	10515	10517	10519	10521	10523	10525	10527	10529	10531	10533	10535	10537	10539	10541	10543	10545	10547	10549	10551	10553	10555	10557	10559	10561	10563	10565	10567	10569	10571	10573	10575	10577	10579	10581	10583	10585	10587	24338	24340	24342	24344	24346	24348	24350	24352	24354	24356	24358	24360	24362	24364	24366	24368	24370	24372	24374	24376	24378	24380	24382	24384	24386	24388	24390	24392	24394	24396	24398	24400	24402	24404	24406	24408	24410	24412	24414	24416		
10428	10430	10432	10434	10436	10438	10440	10442	10444	10446	10448	10450	10452	10454	10456	10458	10460	10462	10464	10466	10468	10470	10472	10474	10476	10478	10480	10482	10484	10486	10488	10490	10492	10494	10496	10498	10500	10502	10504	10506	24257	24259	24261	24263	24265	24267	24269	24271	24273	24275	24277	24279	24281	24283	24285	24287	24289	24291	24293	24295	24297	24299	24301	24303	24305	24307	24309	24311	24313	24315	24317	24319	24321	24323	24325	24327	24329	24331	24333	24335		
10429	10431	10433	10435	10437	10439	10441	10443	10445	10447	10449	10451	10453	10455	10457	10459	10461	10463	10465	10467	10469	10471	10473	10475	10477	10479	10481	10483	10485	10487	10489	10491	10493	10495	10497	10499	10501	10503	10505	10507	24258	24260	24262	24264	24266	24268	24270	24272	24274	24276	24278	24280	24282	24284	24286	24288	24290	24292	24294	24296	24298	24300	24302	24304	24306	24308	24310	24312	24314	24316	24318	24320	24322	24324	24326	24328	24330	24332	24334	24336		
10348	10350	10352	10354	10356	10358	10360	10362	10364	10366	10368	10370	10372	10374	10376	10378	10380	10382	10384	10386	10388	10390	10392	10394	10396	10398	10400	10402	10404	10406	10408	10410	10412	10414	10416	10418	10420	10422	10424	10426	24177	24179	24181	24183	24185	24187	24189	24191	24193	24195	24197	24199	24201	24203	24205	24207	24209	24211	24213	24215	24217	24219	24221	24223	24225	24227	24229	24231	24233	24235	24237	24239	24241	24243	24245	24247	24249	24251	24253	24255		
10349	10351	10353	10355	10357	10359	10361	10363	10365	10367	10369	10371	10373	10375	10377	10379	10381	10383	10385	10387	10389	10391	10393	10395	10397	10399	10401	10403	10405	10407	10409	10411	10413	10415	10417	10419	10421	10423	10425	10427	24178	24180	24182	24184	24186	24188	24190	24192	24194	24196	24198	24200	24202	24204	24206	24208	24210	24212	24214	24216	24218	24220	24222	24224	24226	24228	24230	24232	24234	24236	24238	24240	24242	24244	24246	24248	24250	24252	24254	24256		
10828	10830	10832	10834	10836	10838	10840	10842	10844	10846	10848	10850	10852	10854	10856	10858	10860	10862	10864	10866	10868	10870	10872	10874	10876	10878	10880	10882	10884	10886	10888	10890	10892	10894	10896	10898	10900	10902	10904	10906	24657	24659	24661	24663	24665	24667	24669	24671	24673	24675	24677	24679	24681	24683	24685	24687	24689	24691	24693	24695	24697	24699	24701	24703	24705	24707	24709	24711	24713	24715	24717	24719	24721	24723	24725	24727	24729	24731	24733	24735		
10829	10831	10833	10835	10837	10839	10841	10843	10845	10847	10849	10851	10853	10855	10857	10859	10861	10863	10865	10867	10869	10871	10873	10875	10877	10879	10881	10883	10885	10887	10889	10891	10893	10895	10897	10899	10901	10903	10905	10907	24658	24660	24662	24664	24666	24668	24670	24672	24674	24676	24678	24680	24682	24684	24686	24688	24690	24692	24694	24696	24698	24700	24702	24704	24706	24708	24710	24712	24714	24716	24718	24720	24722	24724	24726	24728	24730	24732	24734	24736		
10908	10910	10912	10914	10916	10918	10920	10922	10924	10926	10928	10930	10932	10934	10936	10938	10940	10942	10944	10946	10948	10950	10952	10954	10956	10958	10960	10962	10964	10966	10968	10970	10972	10974	10976	10978	10980	10982	10984	10986	24737	24739	24741	24743	24745	24747	24749	24751	24753	24755	24757	24759	24761	24763	24765	24767	24769	24771	24773	24775	24777	24779	24781	24783	24785	24787	24789	24791	24793	24795	24797	24799	24801	24803	24805	24807	24809	24811	24813	24815		
10909	10911	10913	10915	10917	10919	10921	10923	10925	10927	10929	10931	10933	10935	10937	10939	10941	10943	10945	10947	10949	10951	10953	10955	10957	10959	10961	10963	10965	10967	10969	10971	10973	10975	10977	10979	10981	10983	10985	10987	24738	24740	24742	24744	24746	24748	24750	24752	24754	24756	24758	24760	24762	24764	24766	24768	24770	24772	24774	24776	24778	24780	24782	24784	24786	24788	24790	24792	24794	24796	24798	24800	24802	24804	24806	24808	24810	24812	24814	24816		
10988	10990	10992	10994	10996	10998	11000	11002	11004	11006	11008	11010	11012	11014	11016	11018	11020	11022	11024	11026	11028	11030	11032	11034	11036	11038	11040	11042	11044	11046	11048	11050	11052	11054	11056	11058	11060	11062	11064	11066	24817	24819	24821	24823	24825	24827	24829	24831	24833	24835	24837	24839	24841	24843	24845	24847	24849	24851	24853	24855	24857	24859	24861	24863	24865	24867	24869	24871	24873	24875	24877	24879	24881	24883	24885	24887	24889	24891	24893	24895		
10989	10991	10993	10995	10997	10999	11001	11003	11005	11007	11009	11011	11013	11015	11017	11019	11021	11023	11025	11027	11029	11031	11033	11035	11037	11039	11041	11043	11045	11047	11049	11051	11053	11055	11057	11059	11061	11063	11065	11067	24818	24820	24822	24824	24826	24828	24830	24832	24834	24836	24838	24840	24842	24844	24846	24848	24850	24852	24854	24856	24858	24860	24862	24864	24866	24868	24870	24872	24874	24876	24878	24880	24882	24884	24886	24888	24890	24892	24894	24896		
10268	10270	10272	10274	10276	10278	10280	10282	10284	10286	10288	10290	10292	10294	10296	10298	10300	10302	10304	10306	10308	10310	10312	10314	10316	10318	10320	10322	10324	10326	10328	10330	10332	10334	10336	10338	10340	10342	10344	10346	24097	24099	24101	24103	24105	24107	24109	24111	24113	24115	24117	24119	24121	24123	24125	24127	24129	24131	24133	24135	24137	24139	24141	24143	24145	24147	24149	24151	24153	24155	24157	24159	24161	24163	24165	24167	24169	24171	24173	24175		
10269	10271	10273	10275	10277	10279	10281	10283	10285	10287	10289	10291	10293	10295	10297	10299	10301	10303	10305	10307	10309	10311	10313	10315	10317	10319	10321	10323	10325	10327	10329	10331	10333	10335	10337	10339	10341	10343	10345	10347	24098	24100	24102	24104	24106	24108	24110	24112	24114	24116	24118	24120	24122	24124	24126	24128	24130	24132	24134	24136	24138	24140	24142	24144	24146	24148	24150	24152	24154	24156	24158	24160	24162	24164	24166	24168	24170	24172	24174	24176		
10188	10190	10192	10194	10196	10198	10200	10202	10204	10206	10208	10210	10212	10214	10216	10218	10220	10222	10224	10226	10228	10230	10232	10234	10236	10238	10240	10242	10244	10246	10248	10250	10252	10254	10256	10258	10260	10262	10264	10266	24017	24019	24021	24023	24025	24027	24029	24031	24033	24035	24037	24039	24041	24043	24045	24047	24049	24051	24053	24055	24057	24059	24061	24063	24065	24067	24069	24071	24073	24075	24077	24079	24081	24083	24085	24087	24089	24091	24093	24095		
10189	10191	10193	10195	10197	10199	10201	10203	10205	10207	10209	10211	10213	10215	10217	10219	10221	10223	10225	10227	10229	10231	10233	10235	10237	10239	10241	10243	10245	10247	10249	10251	10253	10255	10257	10259	10261	10263	10265	10267	24018	24020	24022	24024	24026	24028	24030	24032	24034	24036	24038	24040	24042	24044	24046	24048	24050	24052	24054	24056	24058	24060	24062	24064	24066	24068	24070	24072	24074	24076	24078	24080	24082	24084	24086	24088	24090	24092	24094	24096		
10108	10110	10112	10114	10116	10118	10120	10122	10124	10126	10128	10130	10132	10134	10136	10138	10140	10142	10144	10146	10148	10150	10152	10154	10156	10158	10160	10162	10164	10166	10168	10170	10172	10174	10176	10178	10180	10182	10184	10186	23937	23939	23941	23943	23945	23947	23949	23951	23953	23955	23957	23959	23961	23963	23965	23967	23969	23971	23973	23975	23977	23979	23981	23983	23985	23987	23989	23991	23993	23995	23997	23999	24001	24003	24005	24007	24009	24011	24013	24015		
10109	10111	10113	10115	10117	10119	10121	10123	10125	10127	10129	10131	10133	10135	10137	10139	10141	10143	10145	10147	10149	10151	10153	10155	10157	10159	10161	10163	10165	10167	10169	10171	10173	10175	10177	10179	10181	10183	10185	10187	23938	23940	23942	23944	23946	23948	23950	23952	23954	23956	23958	23960	23962	23964	23966	23968	23970	23972	23974	23976	23978	23980	23982	23984	23986	23988	23990	23992	23994	23996	23998	24000	24002	24004	24006	24008	24010	24012	24014	24016		
11068	11070	11072	11074	11076	11078	11080	11082	11084	11086	11088	11090	11092	11094	11096	11098	11100	11102	11104	11106	11108	11110	11112	11114	11116	11118	11120	11122	11124	11126	11128	11130	11132	11134	11136	11138	11140	11142	11144	11146	24897	24899	24901	24903	24905	24907	24909	24911	24913	24915	24917	24919	24921	24923	24925	24927	24929	24931	24933	24935	24937	24939	24941	24943	24945	24947	24949	24951	24953	24955	24957	24959	24961	24963	24965	24967	24969	24971	24973	24975		
11069	11071	11073	11075	11077	11079	11081	11083	11085	11087	11089	11091	11093	11095	11097	11099	11101	11103	11105	11107	11109	11111	11113	11115	11117	11119	11121	11123	11125	11127	11129	11131	11133	11135	11137	11139	11141	11143	11145	11147	24898	24900	24902	24904	24906	24908	24910	24912	24914	24916	24918	24920	24922	24924	24926	24928	24930	24932	24934	24936	24938	24940	24942	24944	24946	24948	24950	24952	24954	24956	24958	24960	24962	24964	24966	24968	24970	24972	24974	24976		
11148	11150	11152	11154	11156	11158	11160	11162	11164	11166	11168	11170	11172	11174	11176	11178	11180	11182	11184	11186	11188	11190	11192	11194	11196	11198	11200	11202	11204	11206	11208	11210	11212	11214	11216	11218	11220	11222	11224	11226	24977	24979	24981	24983	24985	24987	24989	24991	24993	24995	24997	24999	25001	25003	25005	25007	25009	25011	25013	25015	25017	25019	25021	25023	25025	25027	25029	25031	25033	25035	25037	25039	25041	25043	25045	25047	25049	25051	25053	25055		
11149	11151	11153	11155	11157	11159	11161	11163	11165	11167	11169	11171	11173	11175	11177	11179	11181	11183	11185	11187	11189	11191	11193	11195	11197	11199	11201	11203	11205	11207	11209	11211	11213	11215	11217	11219	11221	11223	11225	11227	24978	24980	24982	24984	24986	24988	24990	24992	24994	24996	24998	25000	25002	25004	25006	25008	25010	25012	25014	25016	25018	25020	25022	25024	25026	25028	25030	25032	25034	25036	25038	25040	25042	25044	25046	25048	25050	25052	25054	25056		
11228	11230	11232	11234	11236	11238	11240	11242	11244	11246	11248	11250	11252	11254	11256	11258	11260	11262	11264	11266	11268	11270	11272	11274	11276	11278	11280	11282	11284	11286	11288	11290	11292	11294	11296	11298	11300	11302	11304	11306	25057	25059	25061	25063	25065	25067	25069	25071	25073	25075	25077	25079	25081	25083	25085	25087	25089	25091	25093	25095	25097	25099	25101	25103	25105	25107	25109	25111	25113	25115	25117	25119	25121	25123	25125	25127	25129	25131	25133	25135		
11229	11231	11233	11235	11237	11239	11241	11243	11245	11247	11249	11251	11253	11255	11257	11259	11261	11263	11265	11267	11269	11271	11273	11275	11277	11279	11281	11283	11285	11287	11289	11291	11293	11295	11297	11299	11301	11303	11305	11307	25058	25060	25062	25064	25066	25068	25070	25072	25074	25076	25078	25080	25082	25084	25086	25088	25090	25092	25094	25096	25098	25100	25102	25104	25106	25108	25110	25112	25114	25116	25118	25120	25122	25124	25126	25128	25130	25132	25134	25136		
10028	10030	10032	10034	10036	10038	10040	10042	10044	10046	10048	10050	10052	10054	10056	10058	10060	10062	10064	10066	10068	10070	10072	10074	10076	10078	10080	10082	10084	10086	10088	10090	10092	10094	10096	10098	10100	10102	10104	10106	23857	23859	23861	23863	23865	23867	23869	23871	23873	23875	23877	23879	23881	23883	23885	23887	23889	23891	23893	23895	23897	23899	23901	23903	23905	23907	23909	23911	23913	23915	23917	23919	23921	23923	23925	23927	23929	23931	23933	23935		
10029	10031	10033	10035	10037	10039	10041	10043	10045	10047	10049	10051	10053	10055	10057	10059	10061	10063	10065	10067	10069	10071	10073	10075	10077	10079	10081	10083	10085	10087	10089	10091	10093	10095	10097	10099	10101	10103	10105	10107	23858	23860	23862	23864	23866	23868	23870	23872	23874	23876	23878	23880	23882	23884	23886	23888	23890	23892	23894	23896	23898	23900	23902	23904	23906	23908	23910	23912	23914	23916	23918	23920	23922	23924	23926	23928	23930	23932	23934	23936		
9948	9950	9952	9954	9956	9958	9960	9962	9964	9966	9968	9970	9972	9974	9976	9978	9980	9982	9984	9986	9988	9990	9992	9994	9996	9998	10000	10002	10004	10006	10008	10010	10012	10014	10016	10018	10020	10022	10024	10026	23777	23779	23781	23783	23785	23787	23789	23791	23793	23795	23797	23799	23801	23803	23805	23807	23809	23811	23813	23815	23817	23819	23821	23823	23825	23827	23829	23831	23833	23835	23837	23839	23841	23843	23845	23847	23849	23851	23853	23855		
9949	9951	9953	9955	9957	9959	9961	9963	9965	9967	9969	9971	9973	9975	9977	9979	9981	9983	9985	9987	9989	9991	9993	9995	9997	9999	10001	10003	10005	10007	10009	10011	10013	10015	10017	10019	10021	10023	10025	10027	23778	23780	23782	23784	23786	23788	23790	23792	23794	23796	23798	23800	23802	23804	23806	23808	23810	23812	23814	23816	23818	23820	23822	23824	23826	23828	23830	23832	23834	23836	23838	23840	23842	23844	23846	23848	23850	23852	23854	23856		
9868	9870	9872	9874	9876	9878	9880	9882	9884	9886	9888	9890	9892	9894	9896	9898	9900	9902	9904	9906	9908	9910	9912	9914	9916	9918	9920	9922	9924	9926	9928	9930	9932	9934	9936	9938	9940	9942	9944	9946	23697	23699	23701	23703	23705	23707	23709	23711	23713	23715	23717	23719	23721	23723	23725	23727	23729	23731	23733	23735	23737	23739	23741	23743	23745	23747	23749	23751	23753	23755	23757	23759	23761	23763	23765	23767	23769	23771	23773	23775		
9869	9871	9873	9875	9877	9879	9881	9883	9885	9887	9889	9891	9893	9895	9897	9899	9901	9903	9905	9907	9909	9911	9913	9915	9917	9919	9921	9923	9925	9927	9929	9931	9933	9935	9937	9939	9941	9943	9945	9947	23698	23700	23702	23704	23706	23708	23710	23712	23714	23716	23718	23720	23722	23724	23726	23728	23730	23732	23734	23736	23738	23740	23742	23744	23746	23748	23750	23752	23754	23756	23758	23760	23762	23764	23766	23768	23770	23772	23774	23776		
11308	11310	11312	11314	11316	11318	11320	11322	11324	11326	11328	11330	11332	11334	11336	11338	11340	11342	11344	11346	11348	11350	11352	11354	11356	11358	11360	11362	11364	11366	11368	11370	11372	11374	11376	11378	11380	11382	11384	11386	25137	25139	25141	25143	25145	25147	25149	25151	25153	25155	25157	25159	25161	25163	25165	25167	25169	25171	25173	25175	25177	25179	25181	25183	25185	25187	25189	25191	25193	25195	25197	25199	25201	25203	25205	25207	25209	25211	25213	25215		
11309	11311	11313	11315	11317	11319	11321	11323	11325	11327	11329	11331	11333	11335	11337	11339	11341	11343	11345	11347	11349	11351	11353	11355	11357	11359	11361	11363	11365	11367	11369	11371	11373	11375	11377	11379	11381	11383	11385	11387	25138	25140	25142	25144	25146	25148	25150	25152	25154	25156	25158	25160	25162	25164	25166	25168	25170	25172	25174	25176	25178	25180	25182	25184	25186	25188	25190	25192	25194	25196	25198	25200	25202	25204	25206	25208	25210	25212	25214	25216		
11388	11390	11392	11394	11396	11398	11400	11402	11404	11406	11408	11410	11412	11414	11416	11418	11420	11422	11424	11426	11428	11430	11432	11434	11436	11438	11440	11442	11444	11446	11448	11450	11452	11454	11456	11458	11460	11462	11464	11466	25217	25219	25221	25223	25225	25227	25229	25231	25233	25235	25237	25239	25241	25243	25245	25247	25249	25251	25253	25255	25257	25259	25261	25263	25265	25267	25269	25271	25273	25275	25277	25279	25281	25283	25285	25287	25289	25291	25293	25295		
11389	11391	11393	11395	11397	11399	11401	11403	11405	11407	11409	11411	11413	11415	11417	11419	11421	11423	11425	11427	11429	11431	11433	11435	11437	11439	11441	11443	11445	11447	11449	11451	11453	11455	11457	11459	11461	11463	11465	11467	25218	25220	25222	25224	25226	25228	25230	25232	25234	25236	25238	25240	25242	25244	25246	25248	25250	25252	25254	25256	25258	25260	25262	25264	25266	25268	25270	25272	25274	25276	25278	25280	25282	25284	25286	25288	25290	25292	25294	25296		
11468	11470	11472	11474	11476	11478	11480	11482	11484	11486	11488	11490	11492	11494	11496	11498	11500	11502	11504	11506	11508	11510	11512	11514	11516	11518	11520	11522	11524	11526	11528	11530	11532	11534	11536	11538	11540	11542	11544	11546	25297	25299	25301	25303	25305	25307	25309	25311	25313	25315	25317	25319	25321	25323	25325	25327	25329	25331	25333	25335	25337	25339	25341	25343	25345	25347	25349	25351	25353	25355	25357	25359	25361	25363	25365	25367	25369	25371	25373	25375		
11469	11471	11473	11475	11477	11479	11481	11483	11485	11487	11489	11491	11493	11495	11497	11499	11501	11503	11505	11507	11509	11511	11513	11515	11517	11519	11521	11523	11525	11527	11529	11531	11533	11535	11537	11539	11541	11543	11545	11547	25298	25300	25302	25304	25306	25308	25310	25312	25314	25316	25318	25320	25322	25324	25326	25328	25330	25332	25334	25336	25338	25340	25342	25344	25346	25348	25350	25352	25354	25356	25358	25360	25362	25364	25366	25368	25370	25372	25374	25376		
9788	9790	9792	9794	9796	9798	9800	9802	9804	9806	9808	9810	9812	9814	9816	9818	9820	9822	9824	9826	9828	9830	9832	9834	9836	9838	9840	9842	9844	9846	9848	9850	9852	9854	9856	9858	9860	9862	9864	9866	23617	23619	23621	23623	23625	23627	23629	23631	23633	23635	23637	23639	23641	23643	23645	23647	23649	23651	23653	23655	23657	23659	23661	23663	23665	23667	23669	23671	23673	23675	23677	23679	23681	23683	23685	23687	23689	23691	23693	23695		
9789	9791	9793	9795	9797	9799	9801	9803	9805	9807	9809	9811	9813	9815	9817	9819	9821	9823	9825	9827	9829	9831	9833	9835	9837	9839	9841	9843	9845	9847	9849	9851	9853	9855	9857	9859	9861	9863	9865	9867	23618	23620	23622	23624	23626	23628	23630	23632	23634	23636	23638	23640	23642	23644	23646	23648	23650	23652	23654	23656	23658	23660	23662	23664	23666	23668	23670	23672	23674	23676	23678	23680	23682	23684	23686	23688	23690	23692	23694	23696		
9708	9710	9712	9714	9716	9718	9720	9722	9724	9726	9728	9730	9732	9734	9736	9738	9740	9742	9744	9746	9748	9750	9752	9754	9756	9758	9760	9762	9764	9766	9768	9770	9772	9774	9776	9778	9780	9782	9784	9786	23537	23539	23541	23543	23545	23547	23549	23551	23553	23555	23557	23559	23561	23563	23565	23567	23569	23571	23573	23575	23577	23579	23581	23583	23585	23587	23589	23591	23593	23595	23597	23599	23601	23603	23605	23607	23609	23611	23613	23615		
9709	9711	9713	9715	9717	9719	9721	9723	9725	9727	9729	9731	9733	9735	9737	9739	9741	9743	9745	9747	9749	9751	9753	9755	9757	9759	9761	9763	9765	9767	9769	9771	9773	9775	9777	9779	9781	9783	9785	9787	23538	23540	23542	23544	23546	23548	23550	23552	23554	23556	23558	23560	23562	23564	23566	23568	23570	23572	23574	23576	23578	23580	23582	23584	23586	23588	23590	23592	23594	23596	23598	23600	23602	23604	23606	23608	23610	23612	23614	23616		
9628	9630	9632	9634	9636	9638	9640	9642	9644	9646	9648	9650	9652	9654	9656	9658	9660	9662	9664	9666	9668	9670	9672	9674	9676	9678	9680	9682	9684	9686	9688	9690	9692	9694	9696	9698	9700	9702	9704	9706	23457	23459	23461	23463	23465	23467	23469	23471	23473	23475	23477	23479	23481	23483	23485	23487	23489	23491	23493	23495	23497	23499	23501	23503	23505	23507	23509	23511	23513	23515	23517	23519	23521	23523	23525	23527	23529	23531	23533	23535		
9629	9631	9633	9635	9637	9639	9641	9643	9645	9647	9649	9651	9653	9655	9657	9659	9661	9663	9665	9667	9669	9671	9673	9675	9677	9679	9681	9683	9685	9687	9689	9691	9693	9695	9697	9699	9701	9703	9705	9707	23458	23460	23462	23464	23466	23468	23470	23472	23474	23476	23478	23480	23482	23484	23486	23488	23490	23492	23494	23496	23498	23500	23502	23504	23506	23508	23510	23512	23514	23516	23518	23520	23522	23524	23526	23528	23530	23532	23534	23536		
11548	11550	11552	11554	11556	11558	11560	11562	11564	11566	11568	11570	11572	11574	11576	11578	11580	11582	11584	11586	11588	11590	11592	11594	11596	11598	11600	11602	11604	11606	11608	11610	11612	11614	11616	11618	11620	11622	11624	11626	25377	25379	25381	25383	25385	25387	25389	25391	25393	25395	25397	25399	25401	25403	25405	25407	25409	25411	25413	25415	25417	25419	25421	25423	25425	25427	25429	25431	25433	25435	25437	25439	25441	25443	25445	25447	25449	25451	25453	25455		
11549	11551	11553	11555	11557	11559	11561	11563	11565	11567	11569	11571	11573	11575	11577	11579	11581	11583	11585	11587	11589	11591	11593	11595	11597	11599	11601	11603	11605	11607	11609	11611	11613	11615	11617	11619	11621	11623	11625	11627	25378	25380	25382	25384	25386	25388	25390	25392	25394	25396	25398	25400	25402	25404	25406	25408	25410	25412	25414	25416	25418	25420	25422	25424	25426	25428	25430	25432	25434	25436	25438	25440	25442	25444	25446	25448	25450	25452	25454	25456		
11628	11630	11632	11634	11636	11638	11640	11642	11644	11646	11648	11650	11652	11654	11656	11658	11660	11662	11664	11666	11668	11670	11672	11674	11676	11678	11680	11682	11684	11686	11688	11690	11692	11694	11696	11698	11700	11702	11704	11706	25457	25459	25461	25463	25465	25467	25469	25471	25473	25475	25477	25479	25481	25483	25485	25487	25489	25491	25493	25495	25497	25499	25501	25503	25505	25507	25509	25511	25513	25515	25517	25519	25521	25523	25525	25527	25529	25531	25533	25535		
11629	11631	11633	11635	11637	11639	11641	11643	11645	11647	11649	11651	11653	11655	11657	11659	11661	11663	11665	11667	11669	11671	11673	11675	11677	11679	11681	11683	11685	11687	11689	11691	11693	11695	11697	11699	11701	11703	11705	11707	25458	25460	25462	25464	25466	25468	25470	25472	25474	25476	25478	25480	25482	25484	25486	25488	25490	25492	25494	25496	25498	25500	25502	25504	25506	25508	25510	25512	25514	25516	25518	25520	25522	25524	25526	25528	25530	25532	25534	25536		
11708	11710	11712	11714	11716	11718	11720	11722	11724	11726	11728	11730	11732	11734	11736	11738	11740	11742	11744	11746	11748	11750	11752	11754	11756	11758	11760	11762	11764	11766	11768	11770	11772	11774	11776	11778	11780	11782	11784	11786	25537	25539	25541	25543	25545	25547	25549	25551	25553	25555	25557	25559	25561	25563	25565	25567	25569	25571	25573	25575	25577	25579	25581	25583	25585	25587	25589	25591	25593	25595	25597	25599	25601	25603	25605	25607	25609	25611	25613	25615		
11709	11711	11713	11715	11717	11719	11721	11723	11725	11727	11729	11731	11733	11735	11737	11739	11741	11743	11745	11747	11749	11751	11753	11755	11757	11759	11761	11763	11765	11767	11769	11771	11773	11775	11777	11779	11781	11783	11785	11787	25538	25540	25542	25544	25546	25548	25550	25552	25554	25556	25558	25560	25562	25564	25566	25568	25570	25572	25574	25576	25578	25580	25582	25584	25586	25588	25590	25592	25594	25596	25598	25600	25602	25604	25606	25608	25610	25612	25614	25616		
9548	9550	9552	9554	9556	9558	9560	9562	9564	9566	9568	9570	9572	9574	9576	9578	9580	9582	9584	9586	9588	9590	9592	9594	9596	9598	9600	9602	9604	9606	9608	9610	9612	9614	9616	9618	9620	9622	9624	9626	23377	23379	23381	23383	23385	23387	23389	23391	23393	23395	23397	23399	23401	23403	23405	23407	23409	23411	23413	23415	23417	23419	23421	23423	23425	23427	23429	23431	23433	23435	23437	23439	23441	23443	23445	23447	23449	23451	23453	23455		
9549	9551	9553	9555	9557	9559	9561	9563	9565	9567	9569	9571	9573	9575	9577	9579	9581	9583	9585	9587	9589	9591	9593	9595	9597	9599	9601	9603	9605	9607	9609	9611	9613	9615	9617	9619	9621	9623	9625	9627	23378	23380	23382	23384	23386	23388	23390	23392	23394	23396	23398	23400	23402	23404	23406	23408	23410	23412	23414	23416	23418	23420	23422	23424	23426	23428	23430	23432	23434	23436	23438	23440	23442	23444	23446	23448	23450	23452	23454	23456		
9468	9470	9472	9474	9476	9478	9480	9482	9484	9486	9488	9490	9492	9494	9496	9498	9500	9502	9504	9506	9508	9510	9512	9514	9516	9518	9520	9522	9524	9526	9528	9530	9532	9534	9536	9538	9540	9542	9544	9546	23297	23299	23301	23303	23305	23307	23309	23311	23313	23315	23317	23319	23321	23323	23325	23327	23329	23331	23333	23335	23337	23339	23341	23343	23345	23347	23349	23351	23353	23355	23357	23359	23361	23363	23365	23367	23369	23371	23373	23375		
9469	9471	9473	9475	9477	9479	9481	9483	9485	9487	9489	9491	9493	9495	9497	9499	9501	9503	9505	9507	9509	9511	9513	9515	9517	9519	9521	9523	9525	9527	9529	9531	9533	9535	9537	9539	9541	9543	9545	9547	23298	23300	23302	23304	23306	23308	23310	23312	23314	23316	23318	23320	23322	23324	23326	23328	23330	23332	23334	23336	23338	23340	23342	23344	23346	23348	23350	23352	23354	23356	23358	23360	23362	23364	23366	23368	23370	23372	23374	23376		
9388	9390	9392	9394	9396	9398	9400	9402	9404	9406	9408	9410	9412	9414	9416	9418	9420	9422	9424	9426	9428	9430	9432	9434	9436	9438	9440	9442	9444	9446	9448	9450	9452	9454	9456	9458	9460	9462	9464	9466	23217	23219	23221	23223	23225	23227	23229	23231	23233	23235	23237	23239	23241	23243	23245	23247	23249	23251	23253	23255	23257	23259	23261	23263	23265	23267	23269	23271	23273	23275	23277	23279	23281	23283	23285	23287	23289	23291	23293	23295		
9389	9391	9393	9395	9397	9399	9401	9403	9405	9407	9409	9411	9413	9415	9417	9419	9421	9423	9425	9427	9429	9431	9433	9435	9437	9439	9441	9443	9445	9447	9449	9451	9453	9455	9457	9459	9461	9463	9465	9467	23218	23220	23222	23224	23226	23228	23230	23232	23234	23236	23238	23240	23242	23244	23246	23248	23250	23252	23254	23256	23258	23260	23262	23264	23266	23268	23270	23272	23274	23276	23278	23280	23282	23284	23286	23288	23290	23292	23294	23296		
11788	11790	11792	11794	11796	11798	11800	11802	11804	11806	11808	11810	11812	11814	11816	11818	11820	11822	11824	11826	11828	11830	11832	11834	11836	11838	11840	11842	11844	11846	11848	11850	11852	11854	11856	11858	11860	11862	11864	11866	25617	25619	25621	25623	25625	25627	25629	25631	25633	25635	25637	25639	25641	25643	25645	25647	25649	25651	25653	25655	25657	25659	25661	25663	25665	25667	25669	25671	25673	25675	25677	25679	25681	25683	25685	25687	25689	25691	25693	25695		
11789	11791	11793	11795	11797	11799	11801	11803	11805	11807	11809	11811	11813	11815	11817	11819	11821	11823	11825	11827	11829	11831	11833	11835	11837	11839	11841	11843	11845	11847	11849	11851	11853	11855	11857	11859	11861	11863	11865	11867	25618	25620	25622	25624	25626	25628	25630	25632	25634	25636	25638	25640	25642	25644	25646	25648	25650	25652	25654	25656	25658	25660	25662	25664	25666	25668	25670	25672	25674	25676	25678	25680	25682	25684	25686	25688	25690	25692	25694	25696		
11868	11870	11872	11874	11876	11878	11880	11882	11884	11886	11888	11890	11892	11894	11896	11898	11900	11902	11904	11906	11908	11910	11912	11914	11916	11918	11920	11922	11924	11926	11928	11930	11932	11934	11936	11938	11940	11942	11944	11946	25697	25699	25701	25703	25705	25707	25709	25711	25713	25715	25717	25719	25721	25723	25725	25727	25729	25731	25733	25735	25737	25739	25741	25743	25745	25747	25749	25751	25753	25755	25757	25759	25761	25763	25765	25767	25769	25771	25773	25775		
11869	11871	11873	11875	11877	11879	11881	11883	11885	11887	11889	11891	11893	11895	11897	11899	11901	11903	11905	11907	11909	11911	11913	11915	11917	11919	11921	11923	11925	11927	11929	11931	11933	11935	11937	11939	11941	11943	11945	11947	25698	25700	25702	25704	25706	25708	25710	25712	25714	25716	25718	25720	25722	25724	25726	25728	25730	25732	25734	25736	25738	25740	25742	25744	25746	25748	25750	25752	25754	25756	25758	25760	25762	25764	25766	25768	25770	25772	25774	25776		
11948	11950	11952	11954	11956	11958	11960	11962	11964	11966	11968	11970	11972	11974	11976	11978	11980	11982	11984	11986	11988	11990	11992	11994	11996	11998	12000	12002	12004	12006	12008	12010	12012	12014	12016	12018	12020	12022	12024	12026	25777	25779	25781	25783	25785	25787	25789	25791	25793	25795	25797	25799	25801	25803	25805	25807	25809	25811	25813	25815	25817	25819	25821	25823	25825	25827	25829	25831	25833	25835	25837	25839	25841	25843	25845	25847	25849	25851	25853	25855		
11949	11951	11953	11955	11957	11959	11961	11963	11965	11967	11969	11971	11973	11975	11977	11979	11981	11983	11985	11987	11989	11991	11993	11995	11997	11999	12001	12003	12005	12007	12009	12011	12013	12015	12017	12019	12021	12023	12025	12027	25778	25780	25782	25784	25786	25788	25790	25792	25794	25796	25798	25800	25802	25804	25806	25808	25810	25812	25814	25816	25818	25820	25822	25824	25826	25828	25830	25832	25834	25836	25838	25840	25842	25844	25846	25848	25850	25852	25854	25856		
9308	9310	9312	9314	9316	9318	9320	9322	9324	9326	9328	9330	9332	9334	9336	9338	9340	9342	9344	9346	9348	9350	9352	9354	9356	9358	9360	9362	9364	9366	9368	9370	9372	9374	9376	9378	9380	9382	9384	9386	23137	23139	23141	23143	23145	23147	23149	23151	23153	23155	23157	23159	23161	23163	23165	23167	23169	23171	23173	23175	23177	23179	23181	23183	23185	23187	23189	23191	23193	23195	23197	23199	23201	23203	23205	23207	23209	23211	23213	23215		
9309	9311	9313	9315	9317	9319	9321	9323	9325	9327	9329	9331	9333	9335	9337	9339	9341	9343	9345	9347	9349	9351	9353	9355	9357	9359	9361	9363	9365	9367	9369	9371	9373	9375	9377	9379	9381	9383	9385	9387	23138	23140	23142	23144	23146	23148	23150	23152	23154	23156	23158	23160	23162	23164	23166	23168	23170	23172	23174	23176	23178	23180	23182	23184	23186	23188	23190	23192	23194	23196	23198	23200	23202	23204	23206	23208	23210	23212	23214	23216		
9228	9230	9232	9234	9236	9238	9240	9242	9244	9246	9248	9250	9252	9254	9256	9258	9260	9262	9264	9266	9268	9270	9272	9274	9276	9278	9280	9282	9284	9286	9288	9290	9292	9294	9296	9298	9300	9302	9304	9306	23057	23059	23061	23063	23065	23067	23069	23071	23073	23075	23077	23079	23081	23083	23085	23087	23089	23091	23093	23095	23097	23099	23101	23103	23105	23107	23109	23111	23113	23115	23117	23119	23121	23123	23125	23127	23129	23131	23133	23135		
9229	9231	9233	9235	9237	9239	9241	9243	9245	9247	9249	9251	9253	9255	9257	9259	9261	9263	9265	9267	9269	9271	9273	9275	9277	9279	9281	9283	9285	9287	9289	9291	9293	9295	9297	9299	9301	9303	9305	9307	23058	23060	23062	23064	23066	23068	23070	23072	23074	23076	23078	23080	23082	23084	23086	23088	23090	23092	23094	23096	23098	23100	23102	23104	23106	23108	23110	23112	23114	23116	23118	23120	23122	23124	23126	23128	23130	23132	23134	23136		
9148	9150	9152	9154	9156	9158	9160	9162	9164	9166	9168	9170	9172	9174	9176	9178	9180	9182	9184	9186	9188	9190	9192	9194	9196	9198	9200	9202	9204	9206	9208	9210	9212	9214	9216	9218	9220	9222	9224	9226	22977	22979	22981	22983	22985	22987	22989	22991	22993	22995	22997	22999	23001	23003	23005	23007	23009	23011	23013	23015	23017	23019	23021	23023	23025	23027	23029	23031	23033	23035	23037	23039	23041	23043	23045	23047	23049	23051	23053	23055		
9149	9151	9153	9155	9157	9159	9161	9163	9165	9167	9169	9171	9173	9175	9177	9179	9181	9183	9185	9187	9189	9191	9193	9195	9197	9199	9201	9203	9205	9207	9209	9211	9213	9215	9217	9219	9221	9223	9225	9227	22978	22980	22982	22984	22986	22988	22990	22992	22994	22996	22998	23000	23002	23004	23006	23008	23010	23012	23014	23016	23018	23020	23022	23024	23026	23028	23030	23032	23034	23036	23038	23040	23042	23044	23046	23048	23050	23052	23054	23056		
12028	12030	12032	12034	12036	12038	12040	12042	12044	12046	12048	12050	12052	12054	12056	12058	12060	12062	12064	12066	12068	12070	12072	12074	12076	12078	12080	12082	12084	12086	12088	12090	12092	12094	12096	12098	12100	12102	12104	12106	25857	25859	25861	25863	25865	25867	25869	25871	25873	25875	25877	25879	25881	25883	25885	25887	25889	25891	25893	25895	25897	25899	25901	25903	25905	25907	25909	25911	25913	25915	25917	25919	25921	25923	25925	25927	25929	25931	25933	25935		
12029	12031	12033	12035	12037	12039	12041	12043	12045	12047	12049	12051	12053	12055	12057	12059	12061	12063	12065	12067	12069	12071	12073	12075	12077	12079	12081	12083	12085	12087	12089	12091	12093	12095	12097	12099	12101	12103	12105	12107	25858	25860	25862	25864	25866	25868	25870	25872	25874	25876	25878	25880	25882	25884	25886	25888	25890	25892	25894	25896	25898	25900	25902	25904	25906	25908	25910	25912	25914	25916	25918	25920	25922	25924	25926	25928	25930	25932	25934	25936		
12108	12110	12112	12114	12116	12118	12120	12122	12124	12126	12128	12130	12132	12134	12136	12138	12140	12142	12144	12146	12148	12150	12152	12154	12156	12158	12160	12162	12164	12166	12168	12170	12172	12174	12176	12178	12180	12182	12184	12186	25937	25939	25941	25943	25945	25947	25949	25951	25953	25955	25957	25959	25961	25963	25965	25967	25969	25971	25973	25975	25977	25979	25981	25983	25985	25987	25989	25991	25993	25995	25997	25999	26001	26003	26005	26007	26009	26011	26013	26015		
12109	12111	12113	12115	12117	12119	12121	12123	12125	12127	12129	12131	12133	12135	12137	12139	12141	12143	12145	12147	12149	12151	12153	12155	12157	12159	12161	12163	12165	12167	12169	12171	12173	12175	12177	12179	12181	12183	12185	12187	25938	25940	25942	25944	25946	25948	25950	25952	25954	25956	25958	25960	25962	25964	25966	25968	25970	25972	25974	25976	25978	25980	25982	25984	25986	25988	25990	25992	25994	25996	25998	26000	26002	26004	26006	26008	26010	26012	26014	26016		
12188	12190	12192	12194	12196	12198	12200	12202	12204	12206	12208	12210	12212	12214	12216	12218	12220	12222	12224	12226	12228	12230	12232	12234	12236	12238	12240	12242	12244	12246	12248	12250	12252	12254	12256	12258	12260	12262	12264	12266	26017	26019	26021	26023	26025	26027	26029	26031	26033	26035	26037	26039	26041	26043	26045	26047	26049	26051	26053	26055	26057	26059	26061	26063	26065	26067	26069	26071	26073	26075	26077	26079	26081	26083	26085	26087	26089	26091	26093	26095		
12189	12191	12193	12195	12197	12199	12201	12203	12205	12207	12209	12211	12213	12215	12217	12219	12221	12223	12225	12227	12229	12231	12233	12235	12237	12239	12241	12243	12245	12247	12249	12251	12253	12255	12257	12259	12261	12263	12265	12267	26018	26020	26022	26024	26026	26028	26030	26032	26034	26036	26038	26040	26042	26044	26046	26048	26050	26052	26054	26056	26058	26060	26062	26064	26066	26068	26070	26072	26074	26076	26078	26080	26082	26084	26086	26088	26090	26092	26094	26096		
9068	9070	9072	9074	9076	9078	9080	9082	9084	9086	9088	9090	9092	9094	9096	9098	9100	9102	9104	9106	9108	9110	9112	9114	9116	9118	9120	9122	9124	9126	9128	9130	9132	9134	9136	9138	9140	9142	9144	9146	22897	22899	22901	22903	22905	22907	22909	22911	22913	22915	22917	22919	22921	22923	22925	22927	22929	22931	22933	22935	22937	22939	22941	22943	22945	22947	22949	22951	22953	22955	22957	22959	22961	22963	22965	22967	22969	22971	22973	22975		spare
9069	9071	9073	9075	9077	9079	9081	9083	9085	9087	9089	9091	9093	9095	9097	9099	9101	9103	9105	9107	9109	9111	9113	9115	9117	9119	9121	9123	9125	9127	9129	9131	9133	9135	9137	9139	9141	9143	9145	9147	22898	22900	22902	22904	22906	22908	22910	22912	22914	22916	22918	22920	22922	22924	22926	22928	22930	22932	22934	22936	22938	22940	22942	22944	22946	22948	22950	22952	22954	22956	22958	22960	22962	22964	22966	22968	22970	22972	22974	22976		spare
8988	8990	8992	8994	8996	8998	9000	9002	9004	9006	9008	9010	9012	9014	9016	9018	9020	9022	9024	9026	9028	9030	9032	9034	9036	9038	9040	9042	9044	9046	9048	9050	9052	9054	9056	9058	9060	9062	9064	9066	22817	22819	22821	22823	22825	22827	22829	22831	22833	22835	22837	22839	22841	22843	22845	22847	22849	22851	22853	22855	22857	22859	22861	22863	22865	22867	22869	22871	22873	22875	22877	22879	22881	22883	22885	22887	22889	22891	22893	22895		spare
8989	8991	8993	8995	8997	8999	9001	9003	9005	9007	9009	9011	9013	9015	9017	9019	9021	9023	9025	9027	9029	9031	9033	9035	9037	9039	9041	9043	9045	9047	9049	9051	9053	9055	9057	9059	9061	9063	9065	9067	22818	22820	22822	22824	22826	22828	22830	22832	22834	22836	22838	22840	22842	22844	22846	22848	22850	22852	22854	22856	22858	22860	22862	22864	22866	22868	22870	22872	22874	22876	22878	22880	22882	22884	22886	22888	22890	22892	22894	22896		spare
8908	8910	8912	8914	8916	8918	8920	8922	8924	8926	8928	8930	8932	8934	8936	8938	8940	8942	8944	8946	8948	8950	8952	8954	8956	8958	8960	8962	8964	8966	8968	8970	8972	8974	8976	8978	8980	8982	8984	8986	22737	22739	22741	22743	22745	22747	22749	22751	22753	22755	22757	22759	22761	22763	22765	22767	22769	22771	22773	22775	22777	22779	22781	22783	22785	22787	22789	22791	22793	22795	22797	22799	22801	22803	22805	22807	22809	22811	22813	22815		user_0
8909	8911	8913	8915	8917	8919	8921	8923	8925	8927	8929	8931	8933	8935	8937	8939	8941	8943	8945	8947	8949	8951	8953	8955	8957	8959	8961	8963	8965	8967	8969	8971	8973	8975	8977	8979	8981	8983	8985	8987	22738	22740	22742	22744	22746	22748	22750	22752	22754	22756	22758	22760	22762	22764	22766	22768	22770	22772	22774	22776	22778	22780	22782	22784	22786	22788	22790	22792	22794	22796	22798	22800	22802	22804	22806	22808	22810	22812	22814	22816		user_1
... 453 more lines ...
Attachment 4: macb_apr20.jed
  481   Tue Jan 30 09:14:50 2024 JM, TDTuesday 30 January
09.00 CARME IE514 5.1e-11mbar

09.15 IE514 not working - one of cables disconnected?

      DSSSD #1 bias cable changed from SHV - Lemo 00.250 to SHV - 2x Lemo 00.250

      DSSSD #1 & #2 LK1 removed from n+n Ohmic side bias FEE64s

      aida11 HDMI cabling re-seated at MACB

10.26 DSSSD bias OK - attachment 1
      Ambient temperature +16.2 deg C

10.30 ADC data item stats OK - attachment 2
      aida06 & aida15 > 100k, all others c. 3k consistent with 50Hz pulser

10.34 per FEE64 Rate spectra - attachment 3
       OK except for known issues aida05, aida07 and aida16 - all others OK

      per p+n FEE64s 1.8.L spectra - attachment 4
       pulser peak width 52 ch FWHM aida11- double peaking DSSSD #1 & #2

      per n+n FEE64s 1.8.L spectra - attachment 7
       pulser peak width 77 ch FWHM aida04 - more uniform noise across all DSSSDs

10.42 per FEE64 1.8.W spectra - 20us FSR - attachments 5-6

11:00 LK2 removed from aida01-04. Configuration for all adaptor cards/FEE modules is now the same. No other change to the setup.

12:00 FEEs power on, bias to 150 V. No change in bias observed. 

12.02 ADC data item stats OK 
      aida06 & aida15 > 100k, all others c. 3k consistent with 50Hz pulser

12.07 per FEE64 Rate spectra 
       OK except for known issues aida05, aida07 and aida16 - all others OK
      
12:09  per p+n FEE64s 1.8.L spectra 
       We now observe a large peak around chn 30,000 on all FEEs. Perhaps an effect of some change in the ring?
       pulser peak width 53 ch FWHM aida11- still observe double peaking DSSSD #1 & #2

      per n+n FEE64s 1.8.L spectra 
      Lots of events at ~chn 65,000 - double peaking now observed on aida03 and aida04
      pulser peak width 93 ch FWHM aida14 

12.20 per FEE64 1.8.W spectra - 20us FSR 

      Perhaps bias was off while acquiring and spectra was not cleared properly. Re-do tests

13:40 ADC data item stats OK - attachment 7 
      aida06 & aida15 > 100k, all others c. 3k consistent with 50Hz pulser

       per FEE64 Rate spectra - attachment 8
       OK except for known issues aida05, aida07 and aida16 - all others OK

      per p+n FEE64s 1.8.L spectra - attachment 9 
       pulser peak width 55 ch FWHM aida11- double peaking DSSSD #1 & #2

      per n+n FEE64s 1.8.L spectra - attachment 10
       pulser peak width 65 ch FWHM aida14 - more uniform noise across all DSSSDs

       per FEE64 1.8.W spectra - 20us FSR - attachments 11-12 

      With help from NH updated firmware on all MACB modules to version 4/20 - see https://elog.ph.ed.ac.uk/CARME/482

      Timestamps for all FEE modules now correct - attachment 13

      MAC address of aida07 was found to be 41-b4-16 not 41-d4-16 as was labelled on the FEE module
       re-labelled FEE64
       dhcpd.conf and MIDAS startup.tcl updated, dhcpd and MIDAS 8015 server restarted - aida07 is now working OK
 
      CAEN N1419ET internal jumper fitted - outputs are no longer floating

15:20 ADC data item stats OK 
      aida06 & aida15 > 100k, all others c. 3k consistent with 50Hz pulser - attachment 14

       per FEE64 Rate spectra
       OK except for known issues aida05, and aida16 - all others OK  - attachment 15

      per p+n FEE64s 1.8.L spectra -
       pulser peak width 50 ch FWHM aida11- double peaking DSSSD #1 & #2 - attachment 16

      per n+n FEE64s 1.8.L spectra 
       pulser peak width 67 ch FWHM aida14 - more uniform noise across all DSSSDs

       per FEE64 1.8.W spectra - 20us FSR - aida14 missing from waveforms - attachments 17,18
 
   Swapped the test inputs so that +ve is now from the inverter and -ve direct from pulser.  - attachments 19,20. Observed reduced noise for when plugged directly into the pulser than when connected from the inverter. 



   Lowered threshold from 0x64 (1 MeV) to 0x20 (320 keV) - attachment 21, 22 - Observed increase in rates, in particular for the DSSD #1,2 pn side. Conclusion -> increased pulser width for DSSD #1,2 is real and not from the pulser
 


   aida01 lemo to water pipe (chamber ground) connection installed 

   ADC data item stats OK 
      aida06 & aida15 > 100k, all others c. 3k consistent with 50Hz pulser 

       per FEE64 Rate spectra
       OK except for known issues aida16 - all others OK 

      per p+n FEE64s 1.8.L spectra -
       pulser peak width 48 ch FWHM aida11- double peaking DSSSD #1 & #2 - attachment 23

      per n+n FEE64s 1.8.L spectra 
       pulser peak width 67 ch FWHM aida14 - attachment 24

       per FEE64 1.8.W spectra - 20us FSR - slightly reduced 100 kHz on pn waveforms - attachments 25,26

   Disconnected aida06 and replaced with the spare FEE module (mac address: 42:0d:16). Power cycled and updated config files. 

   ADC data item stats OK 
      aida15 > 100k, all others including aida06 c. 3k consistent with 50Hz pulser - attachment 27

       per FEE64 Rate spectra
       OK except for known issues aida16 - all others OK - attachment 28

      per p+n FEE64s 1.8.L spectra -
       pulser peak width 48 ch FWHM aida11- no longer observe double peaking DSSSD #1 & #2 but the pulser peaks are still broader compared to DSSD #3,4 - attachment 29

      per n+n FEE64s 1.8.L spectra 
       pulser peak width 55 ch FWHM aida14 - attachment 30

       per FEE64 1.8.W spectra - 20us FSR - greatly reduced 100 kHz on pn waveforms, dodgy FEE previously aida06 must@ve had some impact on DSSD#1,2 - attachments 31,32
      
Attachment 1: Screenshot_from_2024-01-30_10-32-41.png
Screenshot_from_2024-01-30_10-32-41.png
Attachment 2: Screenshot_from_2024-01-30_10-36-20.png
Screenshot_from_2024-01-30_10-36-20.png
Attachment 3: Screenshot_from_2024-01-30_10-40-21.png
Screenshot_from_2024-01-30_10-40-21.png
Attachment 4: Screenshot_from_2024-01-30_10-46-39.png
Screenshot_from_2024-01-30_10-46-39.png
Attachment 5: Screenshot_from_2024-01-30_10-48-17.png
Screenshot_from_2024-01-30_10-48-17.png
Attachment 6: Screenshot_from_2024-01-30_10-48-59.png
Screenshot_from_2024-01-30_10-48-59.png
Attachment 7: Screenshot_from_2024-01-30_13-44-18.png
Screenshot_from_2024-01-30_13-44-18.png
Attachment 8: Screenshot_from_2024-01-30_13-45-18.png
Screenshot_from_2024-01-30_13-45-18.png
Attachment 9: Screenshot_from_2024-01-30_13-46-20.png
Screenshot_from_2024-01-30_13-46-20.png
Attachment 10: Screenshot_from_2024-01-30_13-47-54.png
Screenshot_from_2024-01-30_13-47-54.png
Attachment 11: Screenshot_from_2024-01-30_13-52-14.png
Screenshot_from_2024-01-30_13-52-14.png
Attachment 12: Screenshot_from_2024-01-30_13-53-09.png
Screenshot_from_2024-01-30_13-53-09.png
Attachment 13: Screenshot_from_2024-01-30_15-25-38.png
Screenshot_from_2024-01-30_15-25-38.png
Attachment 14: Screenshot_from_2024-01-30_15-33-37.png
Screenshot_from_2024-01-30_15-33-37.png
Attachment 15: Screenshot_from_2024-01-30_15-38-08.png
Screenshot_from_2024-01-30_15-38-08.png
Attachment 16: Screenshot_from_2024-01-30_15-39-28.png
Screenshot_from_2024-01-30_15-39-28.png
Attachment 17: Screenshot_from_2024-01-30_15-43-00.png
Screenshot_from_2024-01-30_15-43-00.png
Attachment 18: Screenshot_from_2024-01-30_15-46-32.png
Screenshot_from_2024-01-30_15-46-32.png
Attachment 19: Screenshot_from_2024-01-30_16-01-08.png
Screenshot_from_2024-01-30_16-01-08.png
Attachment 20: Screenshot_from_2024-01-30_16-03-27.png
Screenshot_from_2024-01-30_16-03-27.png
Attachment 21: Screenshot_from_2024-01-30_16-08-50.png
Screenshot_from_2024-01-30_16-08-50.png
Attachment 22: Screenshot_from_2024-01-30_16-09-27.png
Screenshot_from_2024-01-30_16-09-27.png
Attachment 23: Screenshot_from_2024-01-30_16-48-21.png
Screenshot_from_2024-01-30_16-48-21.png
Attachment 24: Screenshot_from_2024-01-30_16-49-38.png
Screenshot_from_2024-01-30_16-49-38.png
Attachment 25: Screenshot_from_2024-01-30_16-50-41.png
Screenshot_from_2024-01-30_16-50-41.png
Attachment 26: Screenshot_from_2024-01-30_16-51-51.png
Screenshot_from_2024-01-30_16-51-51.png
Attachment 27: Screenshot_from_2024-01-30_17-35-01.png
Screenshot_from_2024-01-30_17-35-01.png
Attachment 28: Screenshot_from_2024-01-30_17-36-01.png
Screenshot_from_2024-01-30_17-36-01.png
Attachment 29: Screenshot_from_2024-01-30_17-36-57.png
Screenshot_from_2024-01-30_17-36-57.png
Attachment 30: Screenshot_from_2024-01-30_17-39-02.png
Screenshot_from_2024-01-30_17-39-02.png
Attachment 31: Screenshot_from_2024-01-30_17-40-05.png
Screenshot_from_2024-01-30_17-40-05.png
Attachment 32: Screenshot_from_2024-01-30_17-41-19.png
Screenshot_from_2024-01-30_17-41-19.png
  480   Mon Jan 29 15:32:08 2024 TD, JMMonday 29 January contd.
Test + and test - daisy chain cabling extended from DSSSD #1 to #2 to DSSSD #1 to DSSSD #4
 terminated by 50 Ohm at end of daisy chain

Grounding cabling extended from DSSSD #1 ( only ) to DSSSD #1 to DSSSD #4
 grounding cables daisy-chained between 4x adaptor PCBs of each DSSSD - no inter-DSSSD grounding cables

DSSSD#1 bias cable SHV-Lemo 00.250 ( to p+n bias adaptor PCB )
DSSSD#2 - DSSSD #4 bias cables SHV - 2x Lemo 00.250 ( to p+n/n+n bias adaptor PCBs )

LK1 fitted for DSSSD #1 & #2 n+n Ohmic side bias adaptor PCBs
LK2 not fitted on any adaptor PCBs? to be checked

CAEN N1419ET internal jumper fitted ( non-floating outputs )? to be checked

16.54 DSSSD bias OK - see attachment 1

      FEE64 temperatures - attachment 2
       aida06 ASIC temp low

      System wide checks and WR - attachments 3-6
       global clock status aid04 0x6, aida06 0x4
       multiple FPGA timestamp and WR decoder errors for MACBs 4 and 5 ( of 5 )

       WR timestamps OK for aida01-aida08 ( w/ MACB firmware version 04/20 )

     ADC data item stats - attachment 7
      all rates consistent with 50Hz pulser *except* 
       aida11 global clock fail
       aida07 no boot
       aida16 no power
       aida06 ? 
       aida15 faulty asic #2

     per FEE64 Rate spectra - attachment 8

     per p+n and n+n FEE64 1.8.L spectra - attachments 9-12
      Pulser peak width p+n side 52 ch FWHM ( aida12 ), n+n side 70 ch FWHM ( aida14 )

     per p+n and n+n 1.8.W spectra - 20us FSR - attachments 13-16
Attachment 1: Screenshot_from_2024-01-29_16-54-37.png
Screenshot_from_2024-01-29_16-54-37.png
Attachment 2: Screenshot_from_2024-01-29_16-56-12.png
Screenshot_from_2024-01-29_16-56-12.png
Attachment 3: Screenshot_from_2024-01-29_16-55-07.png
Screenshot_from_2024-01-29_16-55-07.png
Attachment 4: Screenshot_from_2024-01-29_16-55-24.png
Screenshot_from_2024-01-29_16-55-24.png
Attachment 5: Screenshot_from_2024-01-29_16-55-39.png
Screenshot_from_2024-01-29_16-55-39.png
Attachment 6: Screenshot_from_2024-01-29_16-55-53.png
Screenshot_from_2024-01-29_16-55-53.png
Attachment 7: Screenshot_from_2024-01-29_16-59-13.png
Screenshot_from_2024-01-29_16-59-13.png
Attachment 8: Screenshot_from_2024-01-29_17-03-10.png
Screenshot_from_2024-01-29_17-03-10.png
Attachment 9: Screenshot_from_2024-01-29_17-05-44.png
Screenshot_from_2024-01-29_17-05-44.png
Attachment 10: Screenshot_from_2024-01-29_17-06-15.png
Screenshot_from_2024-01-29_17-06-15.png
Attachment 11: Screenshot_from_2024-01-29_17-08-53.png
Screenshot_from_2024-01-29_17-08-53.png
Attachment 12: Screenshot_from_2024-01-29_17-09-23.png
Screenshot_from_2024-01-29_17-09-23.png
Attachment 13: Screenshot_from_2024-01-29_17-11-01.png
Screenshot_from_2024-01-29_17-11-01.png
Attachment 14: Screenshot_from_2024-01-29_17-11-39.png
Screenshot_from_2024-01-29_17-11-39.png
Attachment 15: Screenshot_from_2024-01-29_17-12-26.png
Screenshot_from_2024-01-29_17-12-26.png
Attachment 16: Screenshot_from_2024-01-29_17-13-06.png
Screenshot_from_2024-01-29_17-13-06.png
  479   Mon Jan 29 08:52:07 2024 TD JMMonday 29 January
09.50 CARME IE514 5.1e-11 mbar ( ambient temperature 16 deg C )

10.24 DSSD bias OK - attachment 1

10.40 MACB configuration

      1              2      3      4      5

mode  0x3            0x3    0x3    0x3    0x3

      from VETAR2    from 1 from 1 from 1 from 1
      to 2           aida01 aida05 aida09 aida13
      to 3           :      :      :      :
      to 4           :      :      :      :
      to 5           aida04 aida08 aida12 aida16

WR emulator cables connected to MACB #1 rear panel

DAQ reset and DAQ setup between each test below - the FEE64s were NOT power cycled between tests


      Per https://elog.ph.ed.ac.uk/CARME/478 test WR

      connecting/disconnecting HDMI cable from VETAR2 to MACB level 0 input was tested 20.1.24 per https://elog.ph.ed.ac.uk/CARME/477

      With VETAR2 abd all MACB mode sttings 0x3 observe same WR issues as previously

      Noted clock & sync cables from WR emulator were swapped
      ( pay attention to SMA cables cf. Zybo board *not* NIM module as shown in https://elog.ph.ed.ac.uk/DESPEC/517 )

      Change MACB level 0 mode from 0x3 to 0xd - all other MACBs mode 0x3
      HDMI cabkle from VETAR2 to MACB level 0 input disconnected

      System wide checks - global clocks all fail - see attachments 2-5


11.14 Reduce MAC config to MACB level 0 plus one MACB connected to FEE64s aida01-aida4
      All MACB modes 0x3

      System wide checks aida01-aida04 OK. Same problem observed with WR timestamp for aida01-aida04

      See attachments 6-9

11.22 Reduce MAC config to one MACB connected to FEE64s aida01-aida4
      MACB mode 0x3

      System wide checks aida01-aida04 OK. Same problem observed with WR timestamp for aida01-aida04

      See attachments 10-13

11.26 One MACB config using MACB 2 of 5 in NIM bin connected to FEE64s aida01-aida4
      MACB mode 0x3

      System wide checks aida01-aida04 OK. Same problem observed with WR timestamp for aida01-aida04

      See attachments 14-17

11.36 Connect VETAR2 output directly to aida09

      System wide checks for aida09 OK.Valid WR timestamp appears to be generated.

      See attachments 18-20

      Check NIM bin with MACBs etc
        +/-6V, 12V and 24V lines OK

11.42 Connect VETAR2 to MACB 1 of 5 ( mode 0x3 )
      Connect aida09 to MACB 1 of 5
      Global clock OK. WR decoder status error 0xd reported for aida09. Valid WR timestamp for aida09.

      See attachments 21-24

11.53 Connect VETAR2 to MACB 1 of 5 ( mode 0x3 )
      Connect aida09-aida12 to MACB 1 of 5
      Global clock OK. WR decoder status error 0xd reported for aida09-aida12. WR timestamps for aida09-aida12 very different - aida10 might be correct timestamp?

      See attachments 25-28

11.53 Connect VETAR2 to MACB 2 of 5 ( mode 0x3 )
      Connect aida09-aida12 to MACB 2 of 5
      Global clock OK. WR decoder status error 0xd reported for aida09-aida12. WR timestamps for aida09-aida12 very different - aida10 & 11 approx correct but 11 has 
lower timestamp than 10 so probably not quite right

      See attachments 29-32


15.00 Per https://elog.ph.ed.ac.uk/CARME/478 systematically test 1x MACB configurations to find 'working' MACBs

      Identify 3x MACB modules which produce sensible WR timestamps - all labelled ' firmware version 04/20 '.
      The other 5x MACBs are either unlabelled or labelled unknown.

      MACB configuration now uses 04/20 MACBs for MACBs 1-3 and unlabelled 2xMACBs fpr MACBs 4-5
      aida01-aida08 should produce WR timestamps OK

      Will attempt to update firmwre of 5x MACBs tomorrow pm with NH
Attachment 1: Screenshot_from_2024-01-29_10-24-10.png
Screenshot_from_2024-01-29_10-24-10.png
Attachment 2: Screenshot_from_2024-01-29_11-05-18.png
Screenshot_from_2024-01-29_11-05-18.png
Attachment 3: Screenshot_from_2024-01-29_11-05-49.png
Screenshot_from_2024-01-29_11-05-49.png
Attachment 4: Screenshot_from_2024-01-29_11-08-15.png
Screenshot_from_2024-01-29_11-08-15.png
Attachment 5: Screenshot_from_2024-01-29_11-08-41.png
Screenshot_from_2024-01-29_11-08-41.png
Attachment 6: Screenshot_from_2024-01-29_11-20-44.png
Screenshot_from_2024-01-29_11-20-44.png
Attachment 7: Screenshot_from_2024-01-29_11-21-07.png
Screenshot_from_2024-01-29_11-21-07.png
Attachment 8: Screenshot_from_2024-01-29_11-21-29.png
Screenshot_from_2024-01-29_11-21-29.png
Attachment 9: Screenshot_from_2024-01-29_11-22-00.png
Screenshot_from_2024-01-29_11-22-00.png
Attachment 10: Screenshot_from_2024-01-29_11-28-20.png
Screenshot_from_2024-01-29_11-28-20.png
Attachment 11: Screenshot_from_2024-01-29_11-28-40.png
Screenshot_from_2024-01-29_11-28-40.png
Attachment 12: Screenshot_from_2024-01-29_11-28-52.png
Screenshot_from_2024-01-29_11-28-52.png
Attachment 13: Screenshot_from_2024-01-29_11-29-26.png
Screenshot_from_2024-01-29_11-29-26.png
Attachment 14: Screenshot_from_2024-01-29_11-33-16.png
Screenshot_from_2024-01-29_11-33-16.png
Attachment 15: Screenshot_from_2024-01-29_11-33-30.png
Screenshot_from_2024-01-29_11-33-30.png
Attachment 16: Screenshot_from_2024-01-29_11-33-43.png
Screenshot_from_2024-01-29_11-33-43.png
Attachment 17: Screenshot_from_2024-01-29_11-34-15.png
Screenshot_from_2024-01-29_11-34-15.png
Attachment 18: Screenshot_from_2024-01-29_11-40-51.png
Screenshot_from_2024-01-29_11-40-51.png
Attachment 19: Screenshot_from_2024-01-29_11-41-12.png
Screenshot_from_2024-01-29_11-41-12.png
Attachment 20: Screenshot_from_2024-01-29_11-41-59.png
Screenshot_from_2024-01-29_11-41-59.png
Attachment 21: Screenshot_from_2024-01-29_11-48-20.png
Screenshot_from_2024-01-29_11-48-20.png
Attachment 22: Screenshot_from_2024-01-29_11-48-41.png
Screenshot_from_2024-01-29_11-48-41.png
Attachment 23: Screenshot_from_2024-01-29_11-48-58.png
Screenshot_from_2024-01-29_11-48-58.png
Attachment 24: Screenshot_from_2024-01-29_11-49-22.png
Screenshot_from_2024-01-29_11-49-22.png
Attachment 25: Screenshot_from_2024-01-29_11-57-41.png
Screenshot_from_2024-01-29_11-57-41.png
Attachment 26: Screenshot_from_2024-01-29_11-58-11.png
Screenshot_from_2024-01-29_11-58-11.png
Attachment 27: Screenshot_from_2024-01-29_11-58-25.png
Screenshot_from_2024-01-29_11-58-25.png
Attachment 28: Screenshot_from_2024-01-29_11-59-11.png
Screenshot_from_2024-01-29_11-59-11.png
Attachment 29: Screenshot_from_2024-01-29_12-05-28.png
Screenshot_from_2024-01-29_12-05-28.png
Attachment 30: Screenshot_from_2024-01-29_12-05-59.png
Screenshot_from_2024-01-29_12-05-59.png
Attachment 31: Screenshot_from_2024-01-29_12-06-11.png
Screenshot_from_2024-01-29_12-06-11.png
Attachment 32: Screenshot_from_2024-01-29_12-07-03.png
Screenshot_from_2024-01-29_12-07-03.png
  478   Sat Jan 27 20:38:15 2024 PJCS[INFO] Steps to diagnose White Rabbit problems

This is a suggested sequence of actions to help with the current problem with the White Rabbit timestamp distribution. Test the system between each intervention. If a test provides an answer to the problem then proceed accordingly rather than continue with the suggestions.

  • Please sketch the current MACB wiring.
  • Unplug and replug the connectors of the VETAR HDMI connecting cable. If possible clean the pcb and cable HDMI connectors with contact cleaner.
  • Disconnect the VETAR and connect the Emulator. If the clocks are not correct then at the Local and Waveform control page change the entry for LMK03200 control register at offset 5 to 0xB. If clocks are now ok then repeat for all the FEEs. Be aware that this value will be set back to 0x1B during SETUP. If the Emulator fails then continue with VETAR and setting 0x1B.
  • Connect one or more FEE64 to the base MACB.
  • Change the base MACB and repeat the test above.
  • Connect a FEE to the VETAR direct. 

I am at present unable to advise further. However should there be more then I will update this entry.

  477   Mon Jan 22 16:29:31 2024 TDSaturday 20 January
11.00 No WLAN/WiFi
      came-gsi private LAN appears to be OK

      AIDA FEE64 power ON

11.09 FEE64 temps - attachment 1
      aida06 low, otherwise OK

11:10 CARME IE514 4.0e-11 mbar, temp unknown
      IE514 OFF

11.22 Detector bias ON - attachment 2
      found ch 2 up ramp set to 63V/s! changed to 1V/s

11.30 system wide checks OK except 
      aida02 & aida06 fail WR decoder status

      WR ts control - attachments 3-4


12.02 ASIC settings updated for aida09-aida16
      ASIC settings DB key 2024Jan20-12:02:08

      Unable to save DB key to Options - use ASIC settings restore as workaround

12.30 Layouts for 16x FEE64s created - directory ~/LayOut/16xFEE64
      Note layouts include spectra from aida07, aida08 and aida16 so an error message will be generated but it can be ignored


12.40 per FEE64 Rate spectra - attachments 5-6
      per FEE64 p+n & nn 1.8.W spectra - attachments 7-8

12.57 per FEE64 p+n & n+n 1.8.W spectra 20us FSR - attachments 9-10

Test signals 

EG&G Ortec 448 Pulser
Amplitude 90,000
attenuation x10
Normalise 5.00
t_r 100ns
tau_d 50us
Frequency 50Hz ( relay line freq )
Polarity +

test + direct from 448, test - from Cooknell SA1 Sum & Invert 

Test + daisy chain -> aida01 - aida02 - aida05 - aida06 - 50 Ohm
Test - daisy chain -> aida03 - aida04 - aida07 - aida08 - 50 Ohm

aida08 ( DSSSD #2 n+n Ohmic bias ) LK2 removed, LK1 added
aida04 ( DSSSD #1 n+n Ohmic bias ) no LK2 fitted, LK1 added

DSSSD#1 SHV - 2x Lemo 00.250 cable replaced by standard SHV - BNC/BNC Lemo-00.250 coaxial cable

Adaptor PCB ground daisy chain aida01 - aida02 - aida03 - aida04

DSSSD#2 - DSSSD#4 HV cabling unchanged

15.30 Power ON
      System wide checks OK except
          aida02 fail WR decoder status - attachment 11

      FEE64 temperatures - attachment 12
       aida06 ASIC temp low

      DSSSD bias - attachment 13
     
      ADC data item stats - attachment 14
       slow comparator 0x64

      per FEE64 Rate and 1.8.W 20us FSR spectra - attachments 15-17

16.00 per FEE64 rate spectra - attachment 18

16.15 per FEE64 Rate and 1.8.W 20us FSR spectra - attachments 19-21

17.36 Switch from VETAR2 ( all MACB modules mode 0x3 ) to NIM WR emulator ( root MACB module 0xd, all others 0x3 )
      All FEE64s fail Global clock and WR decoder - attachments 22-24
      WR timetsamp control - attachment 25


18.10 Noted MACB #4 mode actually set to 0x2 *not* 0x3 - changed to 0x3

      Re-test with VETAR2 WR - same behaviour as observed this morning - see attachments 26-30




Summary

n+n sides appear to be OK but noisy

no data observed from aida06 & aida11
 aida11 known to be disconnected from adaptor PCB

aida15 asic #2 may be faulty

aida01, aida05, aida12 - data suggest FEE64/adaptor PCB alignment issue ?

aida02, aida03, aida09, aida14 - data suggest adaptor PCB damaged pins ?


aida07 - network config issue ?

aida08 - not completing boot

need to attach sys console to aida07 & aida08 to diagnose problem.
Attachment 1: Screenshot_from_2024-01-20_11-10-03.png
Screenshot_from_2024-01-20_11-10-03.png
Attachment 2: Screenshot_from_2024-01-20_11-22-20.png
Screenshot_from_2024-01-20_11-22-20.png
Attachment 3: Screenshot_from_2024-01-20_11-31-31.png
Screenshot_from_2024-01-20_11-31-31.png
Attachment 4: Screenshot_from_2024-01-20_11-31-39.png
Screenshot_from_2024-01-20_11-31-39.png
Attachment 5: Screenshot_from_2024-01-20_12-18-27.png
Screenshot_from_2024-01-20_12-18-27.png
Attachment 6: Screenshot_from_2024-01-20_12-18-37.png
Screenshot_from_2024-01-20_12-18-37.png
Attachment 7: Screenshot_from_2024-01-20_12-44-07.png
Screenshot_from_2024-01-20_12-44-07.png
Attachment 8: Screenshot_from_2024-01-20_12-42-58.png
Screenshot_from_2024-01-20_12-42-58.png
Attachment 9: Screenshot_from_2024-01-20_12-59-12.png
Screenshot_from_2024-01-20_12-59-12.png
Attachment 10: Screenshot_from_2024-01-20_13-00-18.png
Screenshot_from_2024-01-20_13-00-18.png
Attachment 11: Screenshot_from_2024-01-20_15-30-48.png
Screenshot_from_2024-01-20_15-30-48.png
Attachment 12: Screenshot_from_2024-01-20_15-40-44.png
Screenshot_from_2024-01-20_15-40-44.png
Attachment 13: Screenshot_from_2024-01-20_15-40-58.png
Screenshot_from_2024-01-20_15-40-58.png
Attachment 14: Screenshot_from_2024-01-20_15-43-06.png
Screenshot_from_2024-01-20_15-43-06.png
Attachment 15: Screenshot_from_2024-01-20_15-44-18.png
Screenshot_from_2024-01-20_15-44-18.png
Attachment 16: Screenshot_from_2024-01-20_15-45-14.png
Screenshot_from_2024-01-20_15-45-14.png
Attachment 17: Screenshot_from_2024-01-20_15-45-47.png
Screenshot_from_2024-01-20_15-45-47.png
Attachment 18: Screenshot_from_2024-01-20_16-03-18.png
Screenshot_from_2024-01-20_16-03-18.png
Attachment 19: Screenshot_from_2024-01-20_16-15-24.png
Screenshot_from_2024-01-20_16-15-24.png
Attachment 20: Screenshot_from_2024-01-20_16-15-52.png
Screenshot_from_2024-01-20_16-15-52.png
Attachment 21: Screenshot_from_2024-01-20_16-16-37.png
Screenshot_from_2024-01-20_16-16-37.png
Attachment 22: Screenshot_from_2024-01-20_17-36-22.png
Screenshot_from_2024-01-20_17-36-22.png
Attachment 23: Screenshot_from_2024-01-20_17-36-28.png
Screenshot_from_2024-01-20_17-36-28.png
Attachment 24: Screenshot_from_2024-01-20_17-36-43.png
Screenshot_from_2024-01-20_17-36-43.png
Attachment 25: Screenshot_from_2024-01-20_17-38-22.png
Screenshot_from_2024-01-20_17-38-22.png
Attachment 26: 1000009752.jpg
1000009752.jpg
Attachment 27: 1000009753.jpg
1000009753.jpg
Attachment 28: 1000009754.jpg
1000009754.jpg
Attachment 29: 1000009755.jpg
1000009755.jpg
Attachment 30: 1000009756.jpg
1000009756.jpg
  476   Fri Jan 19 11:17:50 2024 JG, TDFriday 19 January
12.20 CARME IE514 4.1e-11mbar, ambient temperature 16.7 deg C

      FEE64 cabling complete

      WR source - VETAR2 borrowed from Nik Kurz and opwerating standalone ( no BuTIS reference )
      All MACBs mode 0x3


15.20 Power ON FEE64s #1 - #15

      aida02 did not boot - /etc/dhcp/dhcpd.conf typo - fixed

      aida04 did not boot - AIDA FEE64 PSU output #4 fuses open circuit - fuses replaced from AIDA FEE64 PSU spare channel ( aida16 not connected yet ) - fixed

      aida07 did not boot - power OK - FEE64 comfort LEDs OK - network switch link LED OK - network config?

      aida08 boots - power OK - comfort LEDs OK - network switch link LED OK - requests RFS / but does not request RFS MIDAS - can ping aida08 OK, cannot telnet aida08 'connection refused'

      aida09 did not boot - FEE64 comfort LEDS OFF - swap FEE64 #9 & #10 power cables - aida09 boots OK, aida10 does not boot
      Restore #9 & #10 power cables - swap AIDA FEE64 outputs #1 & #2 ( aida09 & aida10 ) - aida09 boots OK
      Conclude AIDA FEE64 PSU ( no label ) output #1 is faulty. Replace AIDA FEE64 PSU ( label ISOL3 ). aida09 & aida10 boot OK

      aida15/aida16 MAC addresses were swapped in https://elog.ph.ed.ac.uk/CARME/467 - fixed

      aida01-aida06, aida09-aida15 boot OK - aida07 does not boot, aida08 does not complete boot

16.25 Configure MIDAS DEE64 servers aida01 - aida06, aida10 - aida15 

      FEE64 temperatures - attachment 1
      aida06 looks low ( c. 25 deg C cf. c. 40 deg C )

      System wide checks - attachments 2-8
       aida02 fails WR decoder

      WR timestamps ( attachments 7 & 8 ) look odd/low - values always c. 0x2???


17.30 CARME IE514 OFF

      WR NIM emulator SMA cables to root MACB ( rear panel ) disconnected at front panel

      Power cycle FEE64s

      System wide checks & WR timestamp control - attachments 9-14
       no change

17.45 Detector bias -20, -100 & 150V - attachments 15-17
      CAEN N1419ET non-floating outputs ( to be checked ), HV cable SHV - 2x Lemo 00.250, LK1 not fitted 

18.40 Detector bias -150V attachment 18




To Do
Check ASIC settings
Configure HV
Check LKs
Ground cabling
      
Attachment 1: Screenshot_from_2024-01-19_16-30-27.png
Screenshot_from_2024-01-19_16-30-27.png
Attachment 2: Screenshot_from_2024-01-19_16-23-24.png
Screenshot_from_2024-01-19_16-23-24.png
Attachment 3: Screenshot_from_2024-01-19_16-23-33.png
Screenshot_from_2024-01-19_16-23-33.png
Attachment 4: Screenshot_from_2024-01-19_16-23-51.png
Screenshot_from_2024-01-19_16-23-51.png
Attachment 5: Screenshot_from_2024-01-19_16-24-02.png
Screenshot_from_2024-01-19_16-24-02.png
Attachment 6: Screenshot_from_2024-01-19_16-24-07.png
Screenshot_from_2024-01-19_16-24-07.png
Attachment 7: Screenshot_from_2024-01-19_16-24-44.png
Screenshot_from_2024-01-19_16-24-44.png
Attachment 8: Screenshot_from_2024-01-19_16-31-32.png
Screenshot_from_2024-01-19_16-31-32.png
Attachment 9: Screenshot_from_2024-01-19_17-48-30.png
Screenshot_from_2024-01-19_17-48-30.png
Attachment 10: Screenshot_from_2024-01-19_17-48-38.png
Screenshot_from_2024-01-19_17-48-38.png
Attachment 11: Screenshot_from_2024-01-19_17-48-46.png
Screenshot_from_2024-01-19_17-48-46.png
Attachment 12: Screenshot_from_2024-01-19_17-48-51.png
Screenshot_from_2024-01-19_17-48-51.png
Attachment 13: Screenshot_from_2024-01-19_17-49-11.png
Screenshot_from_2024-01-19_17-49-11.png
Attachment 14: Screenshot_from_2024-01-19_17-49-23.png
Screenshot_from_2024-01-19_17-49-23.png
Attachment 15: Screenshot_from_2024-01-19_17-43-55.png
Screenshot_from_2024-01-19_17-43-55.png
Attachment 16: Screenshot_from_2024-01-19_17-54-33.png
Screenshot_from_2024-01-19_17-54-33.png
Attachment 17: Screenshot_from_2024-01-19_17-56-01.png
Screenshot_from_2024-01-19_17-56-01.png
Attachment 18: Screenshot_from_2024-01-19_18-41-45.png
Screenshot_from_2024-01-19_18-41-45.png
  475   Thu Jan 18 17:21:00 2024 TDThursday 18 January
Network ( dhcpd.conf ) and MIDAS startup.tcl/Options/ASIC settings/rfs filesystems configured


17.20 CARME IE514 4.1e-11mbar, ambient temperature 16.7 deg C
      Pressure appears to be drifting upwards - see attachment 1
Attachment 1: Capture.PNG
Capture.PNG
  474   Wed Jan 17 19:21:20 2024 TDAIDA spares
Spare AIDA hardware 

1x AIDA FEE64 PSU
1x USB-controlled ac mains relay 

Transferred from EX 2.039 store ( ESR roof ) to CARME storage bins adjacent to CRYRING counting room ( CRYRING roof ).

CRYRING Hall

12 USB-Jtag cables ( sys console ) + 1x to Edinburgh Jan 2024 to fabricate more with spare Jtag PCBs
7x HDMI FEE64 - MACB cables
0x 0.3m HDMI-HDMI cables - require 8x 0.3m HDMI - HDMI cables + 2x 3m HDMI-HDMI cables
2x MACB
1x NIM WR emulator
1x Dlink 7 port USB extender
1x Netgear 24-port Gbit network switch
1x CAEN N1419ET ( ch #0 u/s ? )
0x 10m RJ45 Cat 6 network cables yellow/green/purple/brown - require 2x of each
  473   Wed Jan 17 19:15:47 2024 TDWednesday 17 January
CAEN N1419ET remote control via carme-gsi and nnrpi1 - see https://elog.ph.ed.ac.uk/CARME/472

15x ( of 16x ) AIDA power cables available, labelled #1 - #15, installed

16x ( of 16x ) RJ45 Cat 6 network cables available, 15x labelled #1 - #15

16x ( of 16x ) HDMI cables available, labelled #1 - #16
  472   Wed Jan 17 10:31:05 2024 TDRemote control of CAEN N1419ET from nnrpi1
Login to nnrpi1 from carme-gsi using ssh

Execute the caen_screen.sh script

pi@nnrpi1:~ $ cat ./caen_screen.sh
if ! screen -list | grep -q "caenhv"; then
	screen -f -S caenhv /dev/ttyACM0 9600
else
	screen -x caenhv
fi
pi@nnrpi1:~ $ ./caen_screen.sh

Enter the text string 'caen'

If the CAEN display program is already running you will only see page updates/changes.
- enter q ( quit ) and then d ( display ) to re-display the entire screen.

See attachment 1
Attachment 1: Capture.PNG
Capture.PNG
  471   Tue Jan 16 16:01:48 2024 TDChange CARMEraspberrypi to nnrpi1
Change RPi from CARMEraspberrypi to nnrpi1 to provide carme-gsi access to Rly16 to control USB-controlled ac mains relay.
nnrpi1 is already defined in carme-gsi:/etc/dhcp/dhcpd.conf

Transfer RPi network connection from CRYRING local network to AIDA 24-port network Gbit switch

Change /etc/hostname from

CARMEraspberrypi

to

nnrpi1

Change /etc/hosts from

127.0.0.1	localhost
::1		localhost ip6-localhost ip6-loopback
ff02::1		ip6-allnodes
ff02::2		ip6-allrouters

127.0.1.1	raspberrypi	

to

127.0.0.1	localhost
::1		localhost ip6-localhost ip6-loopback
ff02::1		ip6-allnodes
ff02::2		ip6-allrouters

127.0.1.1	nnrpi1

Reboot RPi

Login to RPi and start MIDAS 8015 service

/MIDAS/TclHttpd/linux-arm/TclHttpd-server &

Open new browser tab on carme-gsi

http://nnrpi1:8015/AIDA/Rly16/Rly16.tml
Attachment 1: Capture.PNG
Capture.PNG
  470   Tue Jan 16 14:36:30 2024 TDCARME configuration
FEE64

#1 - #16 -> aida01 - aida16

Network configuration - /etc/dhcpd/dhcpd.conf - attachment 1

MIDAS Options

Added Options sub-directories for aida09-aida16 - attachment 2

MIDAS ASIC settings

Added sub-directories for aida09-aida16 - attachment 3

MIDAS configuration - /MIDAS/config/TclHttpd/carme-gsi"8015 - attachment 4
Currently only aida01-aida08 defined

MIDAS embedded rfs filesystems for aida01-aida24 alreday exist - attachment 5

All FEE64s connected to rev B CARME adaptor PCBs
LK1 fitted to n+n bias FEE64 ( aidaxx, aidaxx, aidaxx, aidaxx )
No other LKs fitted


MACB

WR source NIM WR emulator or VETAR2

See https://elog.ph.ed.ac.uk/DESPEC/517 for MACB mode configuration


CAEN N1419ET


EG&G Ortec 448 Pulser


ASIC settings


Options
Attachment 1: dhcpd.conf
#
# DHCP Server Configuration file.
#   see /usr/share/doc/dhcp*/dhcpd.conf.example
#   see dhcpd.conf(5) man page
#
# 2 days
#default-lease-time 172800;
# 4 days
default-lease-time 345600;
# 8 days
max-lease-time 691200;

option domain-search code 119 = string;

option domain-name "dl.ac.uk";
option domain-name-servers 193.62.115.16, 148.79.80.78;
option netbios-name-servers 148.79.160.89;
option netbios-node-type 8;
option nis-domain "nuclear.physics";
option nis-servers 193.62.115.77;


subnet 192.168.11.0 netmask 255.255.255.0 {
   option subnet-mask 255.255.255.0;
   option broadcast-address 192.168.11.255;
   pool {
 range 192.168.11.118 192.168.11.199;
   }
}

group {
   use-host-decl-names true;
   default-lease-time 3600;
   max-lease-time 14400;

   server-name "192.168.11.99";
   next-server 192.168.11.99;

   host nnrpi1 {
     hardware ethernet dc:a6:32:1d:21:db;
     fixed-address 192.168.11.251;
   }

   host nnrpi2 {
     hardware ethernet b8:27:eb:40:53:e8;
     fixed-address 192.168.11.117;
   }

   host aida01 {
     hardware ethernet d8:80:39:41:d8:20;
     fixed-address 192.168.11.1;
    option root-path "/home/Embedded/XilinxLinux/ppc_4xx/rfs/aida01";
   }

   host aida02 {
     hardware ethernet d8:80:39:41:d8:2a;
     fixed-address 192.168.12.2;
    option root-path "/home/Embedded/XilinxLinux/ppc_4xx/rfs/aida02";
   }

   host aida03 {
     hardware ethernet d8:80:39:41:d0:05;
     fixed-address 192.168.11.3;
    option root-path "/home/Embedded/XilinxLinux/ppc_4xx/rfs/aida03";
   }

   host aida04 {
     hardware ethernet d8:80:39:41:b4:0b;
     fixed-address 192.168.11.4;
    option root-path "/home/Embedded/XilinxLinux/ppc_4xx/rfs/aida04";
   }

   host aida05 {
     hardware ethernet d8:80:39:42:0d:0c;
     fixed-address 192.168.11.5;
    option root-path "/home/Embedded/XilinxLinux/ppc_4xx/rfs/aida05";
   }

   host aida06 {
     hardware ethernet d8:80:39:41:d7:cd;
     fixed-address 192.168.11.6;
    option root-path "/home/Embedded/XilinxLinux/ppc_4xx/rfs/aida06";
   }

   host aida07 {
     hardware ethernet d8:80:39:41:d4:16;
     fixed-address 192.168.11.7;
    option root-path "/home/Embedded/XilinxLinux/ppc_4xx/rfs/aida07";
   }

   host aida08 {
     hardware ethernet d8:80:39:41:ba:2b;
     fixed-address 192.168.11.8;
    option root-path "/home/Embedded/XilinxLinux/ppc_4xx/rfs/aida08";
   }

   host aida09 {
     hardware ethernet d8:80:39:41:d0:0e;
     fixed-address 192.168.11.9;
    option root-path "/home/Embedded/XilinxLinux/ppc_4xx/rfs/aida09";
   }

   host aida10 {
     hardware ethernet d8:80:39:41:d8:2b;
     fixed-address 192.168.11.10;
    option root-path "/home/Embedded/XilinxLinux/ppc_4xx/rfs/aida10";
   }

   host aida11 {
     hardware ethernet d8:80:39:41:d7:cc;
     fixed-address 192.168.11.11;
    option root-path "/home/Embedded/XilinxLinux/ppc_4xx/rfs/aida11";
   }

   host aida12 {
     hardware ethernet d8:80:39:41:d7:c2;
     fixed-address 192.168.11.12;
    option root-path "/home/Embedded/XilinxLinux/ppc_4xx/rfs/aida12";
   }

   host aida13 {
     hardware ethernet d8:80:39:41:ba:22;
     fixed-address 192.168.11.13;
    option root-path "/home/Embedded/XilinxLinux/ppc_4xx/rfs/aida13";
   }

   host aida14 {
     hardware ethernet d8:80:39:41:ee:72;
     fixed-address 192.168.11.14;
    option root-path "/home/Embedded/XilinxLinux/ppc_4xx/rfs/aida14";
   }

   host aida15 {
     hardware ethernet d8:80:39:41:ee:71;
     fixed-address 192.168.11.15;
    option root-path "/home/Embedded/XilinxLinux/ppc_4xx/rfs/aida15";
   }

   host aida16 {
     hardware ethernet d8:80:39:41:a0:71;
     fixed-address 192.168.11.16;
    option root-path "/home/Embedded/XilinxLinux/ppc_4xx/rfs/aida16";
   }

}


subnet 192.168.12.0 netmask 255.255.255.0 {
   option subnet-mask 255.255.255.0;
   option broadcast-address 192.168.12.255;
   pool {
 range 192.168.12.100 192.168.12.199;
   }
}

group {
   use-host-decl-names true;
   default-lease-time 3600;
   max-lease-time 14400;

   server-name "192.168.12.99";
   next-server 192.168.12.99

}




Attachment 2: Options
[npg@carme-gsi Options]$ pwd
/MIDAS/DB/EXPERIMENTS/AIDA/Options
[npg@carme-gsi Options]$ ls -lR
.:
total 0
drwxrwxr-x. 2 npg npg 22 Jan 31  2022 aida01
drwxrwxr-x. 2 npg npg 22 Jan 31  2022 aida02
drwxrwxr-x. 2 npg npg 22 Jan 31  2022 aida03
drwxrwxr-x. 2 npg npg 22 Jan 31  2022 aida04
drwxrwxr-x. 2 npg npg 22 Jan 31  2022 aida05
drwxrwxr-x. 2 npg npg 22 Jan 31  2022 aida06
drwxrwxr-x. 2 npg npg 22 Jan 31  2022 aida07
drwxrwxr-x. 2 npg npg 22 Jan 31  2022 aida08
drwxrwxr-x. 2 npg npg 22 Jan 18 15:15 aida09
drwxrwxr-x. 2 npg npg 22 Jan 18 15:15 aida10
drwxrwxr-x. 2 npg npg 22 Jan 18 15:15 aida11
drwxrwxr-x. 2 npg npg 22 Jan 18 15:15 aida12
drwxrwxr-x. 2 npg npg 22 Jan 18 15:15 aida13
drwxrwxr-x. 2 npg npg 22 Jan 18 15:15 aida14
drwxrwxr-x. 2 npg npg 22 Jan 18 15:15 aida15
drwxrwxr-x. 2 npg npg 22 Jan 18 15:15 aida16

./aida01:
total 4
-rwxrwxr-x. 1 npg npg 1027 Nov 28  2022 CONTENTS

./aida02:
total 4
-rwxrwxr-x. 1 npg npg 1026 Feb 19  2022 CONTENTS

./aida03:
total 4
-rwxrwxr-x. 1 npg npg 1026 Feb 13  2022 CONTENTS

./aida04:
total 4
-rwxrwxr-x. 1 npg npg 1026 Dec  1  2022 CONTENTS

./aida05:
total 4
-rwxrwxr-x. 1 npg npg 1027 Nov 30  2022 CONTENTS

./aida06:
total 4
-rwxrwxr-x. 1 npg npg 1026 Feb  4  2022 CONTENTS

./aida07:
total 4
-rwxrwxr-x. 1 npg npg 1026 Jan 31  2022 CONTENTS

./aida08:
total 4
-rwxrwxr-x. 1 npg npg 1026 Feb  4  2022 CONTENTS

./aida09:
total 4
-rwxrwxr-x. 1 npg npg 1026 Jan 18 15:15 CONTENTS

./aida10:
total 4
-rwxrwxr-x. 1 npg npg 1026 Jan 18 15:15 CONTENTS

./aida11:
total 4
-rwxrwxr-x. 1 npg npg 1026 Jan 18 15:15 CONTENTS

./aida12:
total 4
-rwxrwxr-x. 1 npg npg 1026 Jan 18 15:15 CONTENTS

./aida13:
total 4
-rwxrwxr-x. 1 npg npg 1026 Jan 18 15:15 CONTENTS

./aida14:
total 4
-rwxrwxr-x. 1 npg npg 1026 Jan 18 15:15 CONTENTS

./aida15:
total 4
-rwxrwxr-x. 1 npg npg 1026 Jan 18 15:15 CONTENTS

./aida16:
total 4
-rwxrwxr-x. 1 npg npg 1026 Jan 18 15:15 CONTENTS

[npg@carme-gsi Options]$ pwd
/MIDAS/DB/EXPERIMENTS/AIDA/Options
[npg@carme-gsi Options]$ ls -lR
.:
total 0
drwxrwxr-x. 2 npg npg 22 Jan 31  2022 aida01
drwxrwxr-x. 2 npg npg 22 Jan 31  2022 aida02
drwxrwxr-x. 2 npg npg 22 Jan 31  2022 aida03
drwxrwxr-x. 2 npg npg 22 Jan 31  2022 aida04
drwxrwxr-x. 2 npg npg 22 Jan 31  2022 aida05
drwxrwxr-x. 2 npg npg 22 Jan 31  2022 aida06
drwxrwxr-x. 2 npg npg 22 Jan 31  2022 aida07
drwxrwxr-x. 2 npg npg 22 Jan 31  2022 aida08
drwxrwxr-x. 2 npg npg 22 Jan 18 15:15 aida09
drwxrwxr-x. 2 npg npg 22 Jan 18 15:15 aida10
drwxrwxr-x. 2 npg npg 22 Jan 18 15:15 aida11
drwxrwxr-x. 2 npg npg 22 Jan 18 15:15 aida12
drwxrwxr-x. 2 npg npg 22 Jan 18 15:15 aida13
drwxrwxr-x. 2 npg npg 22 Jan 18 15:15 aida14
drwxrwxr-x. 2 npg npg 22 Jan 18 15:15 aida15
drwxrwxr-x. 2 npg npg 22 Jan 18 15:15 aida16

./aida01:
total 4
-rwxrwxr-x. 1 npg npg 1027 Nov 28  2022 CONTENTS

./aida02:
total 4
-rwxrwxr-x. 1 npg npg 1026 Feb 19  2022 CONTENTS

./aida03:
total 4
-rwxrwxr-x. 1 npg npg 1026 Feb 13  2022 CONTENTS

./aida04:
total 4
-rwxrwxr-x. 1 npg npg 1026 Dec  1  2022 CONTENTS

./aida05:
total 4
-rwxrwxr-x. 1 npg npg 1027 Nov 30  2022 CONTENTS

./aida06:
total 4
-rwxrwxr-x. 1 npg npg 1026 Feb  4  2022 CONTENTS

./aida07:
total 4
-rwxrwxr-x. 1 npg npg 1026 Jan 31  2022 CONTENTS

./aida08:
total 4
-rwxrwxr-x. 1 npg npg 1026 Feb  4  2022 CONTENTS

./aida09:
total 4
-rwxrwxr-x. 1 npg npg 1026 Jan 18 15:15 CONTENTS

./aida10:
total 4
-rwxrwxr-x. 1 npg npg 1026 Jan 18 15:15 CONTENTS

./aida11:
total 4
-rwxrwxr-x. 1 npg npg 1026 Jan 18 15:15 CONTENTS

./aida12:
total 4
-rwxrwxr-x. 1 npg npg 1026 Jan 18 15:15 CONTENTS

./aida13:
total 4
-rwxrwxr-x. 1 npg npg 1026 Jan 18 15:15 CONTENTS

./aida14:
total 4
-rwxrwxr-x. 1 npg npg 1026 Jan 18 15:15 CONTENTS

./aida15:
total 4
-rwxrwxr-x. 1 npg npg 1026 Jan 18 15:15 CONTENTS

./aida16:
total 4
-rwxrwxr-x. 1 npg npg 1026 Jan 18 15:15 CONTENTS


[npg@carme-gsi Options]$ cat aida02/CONTENTS
Index string Stat.offset&&Aida_Hist_W_Enable&&TS_SYNC_PHASE&&ExtClk&&Aida.shift&&MACB_TRIG_MODE&&Aida.offset&&RunNumber&&Aida_GroupBase&&Rate.channels&&Stat.channels&&Aida.Vchannels&&ASIC.settings&&Aida_Hist_D_Enable&&WAVE_DMA_HWM&&Stat.shift&&Aida.Wchannels&&Aida_Hist_V_Enable&&Include.Aida&&Aida.channels&&DataAcqPgm&&Aida_Hist_H_Enable&&DataFormat&&ASIC_DMA_HWM&&Aida_Hist_L_Enable
Stat.offset string 64
Aida_Hist_W_Enable string 1
Aida.shift string 0
TS_SYNC_PHASE string 0x0
ExtClk string 1
Aida.offset string 0
MACB_TRIG_MODE string 4
RunNumber string 63
Rate.channels string 512
Aida_GroupBase string 1
Stat.channels string 512
Aida_Hist_D_Enable string 1
ASIC.settings string 2022Jan31-11-52-00
Aida.Vchannels string 256
WAVE_DMA_HWM string 0x0007ffff
Aida.Wchannels string 1020
Stat.shift string 6
Include.Aida string 0
Aida_Hist_V_Enable string 0
Aida.channels string 65536
DataFormat string 0x0000
DataAcqPgm string AidaExecV9
Aida_Hist_H_Enable string 1
ASIC_DMA_HWM string 0x000fffff
Aida_Hist_L_Enable string 1

Attachment 3: Settings
[npg@carme-gsi 2022Jan31-11-52-00]$ pwd
/MIDAS/DB/EXPERIMENTS/AIDA/2022Jan31-11-52-00
[npg@carme-gsi 2022Jan31-11-52-00]$ ls -lR
.:
total 0
drwxrwxr-x. 2 npg npg 22 Jan 31  2022 aida01
drwxrwxr-x. 2 npg npg 22 Jan 31  2022 aida02
drwxrwxr-x. 2 npg npg 22 Jan 31  2022 aida03
drwxrwxr-x. 2 npg npg 22 Jan 31  2022 aida04
drwxrwxr-x. 2 npg npg 22 Jan 31  2022 aida05
drwxrwxr-x. 2 npg npg 22 Jan 31  2022 aida06
drwxrwxr-x. 2 npg npg 22 Jan 31  2022 aida07
drwxrwxr-x. 2 npg npg 22 Jan 31  2022 aida08
drwxrwxr-x. 2 npg npg 22 Jan 18 15:21 aida09
drwxrwxr-x. 2 npg npg 22 Jan 18 15:21 aida10
drwxrwxr-x. 2 npg npg 22 Jan 18 15:21 aida11
drwxrwxr-x. 2 npg npg 22 Jan 18 15:21 aida12
drwxrwxr-x. 2 npg npg 22 Jan 18 15:21 aida13
drwxrwxr-x. 2 npg npg 22 Jan 18 15:21 aida14
drwxrwxr-x. 2 npg npg 22 Jan 18 15:21 aida15
drwxrwxr-x. 2 npg npg 22 Jan 18 15:21 aida16

./aida01:
total 4
-rwxrwxr-x. 1 npg npg 2726 Jan 31  2022 CONTENTS

./aida02:
total 4
-rwxrwxr-x. 1 npg npg 2726 Jan 31  2022 CONTENTS

./aida03:
total 4
-rwxrwxr-x. 1 npg npg 2726 Jan 31  2022 CONTENTS

./aida04:
total 4
-rwxrwxr-x. 1 npg npg 2726 Jan 31  2022 CONTENTS

./aida05:
total 4
-rwxrwxr-x. 1 npg npg 2726 Jan 31  2022 CONTENTS

./aida06:
total 4
-rwxrwxr-x. 1 npg npg 2726 Jan 31  2022 CONTENTS

./aida07:
total 4
-rwxrwxr-x. 1 npg npg 2726 Jan 31  2022 CONTENTS

./aida08:
total 4
-rwxrwxr-x. 1 npg npg 2726 Jan 31  2022 CONTENTS

./aida09:
total 4
-rwxrwxr-x. 1 npg npg 2726 Jan 18 15:21 CONTENTS

./aida10:
total 4
-rwxrwxr-x. 1 npg npg 2726 Jan 18 15:21 CONTENTS

./aida11:
total 4
-rwxrwxr-x. 1 npg npg 2726 Jan 18 15:21 CONTENTS

./aida12:
total 4
-rwxrwxr-x. 1 npg npg 2726 Jan 18 15:21 CONTENTS

./aida13:
total 4
-rwxrwxr-x. 1 npg npg 2726 Jan 18 15:21 CONTENTS

./aida14:
total 4
-rwxrwxr-x. 1 npg npg 2726 Jan 18 15:21 CONTENTS

./aida15:
total 4
-rwxrwxr-x. 1 npg npg 2726 Jan 18 15:21 CONTENTS

./aida16:
total 4
-rwxrwxr-x. 1 npg npg 2726 Jan 18 15:21 CONTENTS

[npg@carme-gsi 2022Jan31-11-52-00]$ cat aida02/CONTENTS
ASIC.1.3 string 0x02
ASIC.2.19 string 0x5c
ASIC.2.20 string 0x68
ASIC.1.4 string 0x07
ASIC.2.21 string 0x0f
ASIC.2.22 string 0x5c
ASIC.1.5 string 0x08
ASIC.1.6 string 0x09
ASIC.2.23 string 0x08
ASIC.2.24 string 0x68
ASIC.1.7 string 0x05
ASIC.2.25 string 0x08
ASIC.1.8 string 0x00
ASIC.2.26 string 0xbf
ASIC.1.9 string 0x0f
ASIC.2.0 string 0x04
ASIC.2.1 string 0x05
ASIC.3.10 string 0x00
ASIC.2.2 string 0x06
ASIC.3.11 string 0x0d
ASIC.2.3 string 0x02
ASIC.3.12 string 0x64
ASIC.2.4 string 0x07
ASIC.3.13 string 0x34
ASIC.2.5 string 0x08
ASIC.3.14 string 0x02
ASIC.2.6 string 0x09
ASIC.3.15 string 0xff
ASIC.2.7 string 0x05
ASIC.3.16 string 0xd2
ASIC.2.8 string 0x00
ASIC.3.17 string 0x80
ASIC.2.9 string 0x0f
ASIC.3.18 string 0xb2
WAVE.0 string 0x00000000
ASIC.3.20 string 0x68
ASIC.3.19 string 0x5c
WAVE.1 string 0x00000000
ASIC.3.21 string 0x0f
WAVE.2 string 0x00000000
LED.0 string 0x00011d4c
ASIC.3.22 string 0x5c
WAVE.3 string 0x00000000
LED.1 string 0x0000ffff
ASIC.3.23 string 0x08
LED.2 string 0x00011b58
ASIC.3.24 string 0x68
ASIC.1.10 string 0x00
LED.3 string 0x00000000
ASIC.3.25 string 0x08
ASIC.3.0 string 0x04
ASIC.3.26 string 0xbf
ASIC.1.11 string 0x0d
LED.4 string 0x00011b58
LED.5 string 0x00000000
ASIC.3.1 string 0x05
ASIC.1.12 string 0x64
ASIC.1.13 string 0x34
ASIC.3.2 string 0x06
LED.6 string 0x00011b58
LED.7 string 0x00000000
ASIC.1.14 string 0x02
ASIC.3.3 string 0x02
ASIC.1.15 string 0xff
ASIC.3.4 string 0x07
LED.8 string 0x000003fc
LED.9 string 0x000000c8
ASIC.1.16 string 0xd2
ASIC.3.5 string 0x08
ASIC.1.17 string 0x80
ASIC.3.6 string 0x09
ASIC.1.18 string 0xb2
ASIC.3.7 string 0x05
ASIC.1.20 string 0x68
ASIC.3.8 string 0x00
ASIC.1.19 string 0x5c
ASIC.3.9 string 0x0f
ASIC.1.21 string 0x0f
ASIC.4.10 string 0x00
ASIC.1.22 string 0x5c
ASIC.1.23 string 0x08
ASIC.4.11 string 0x0d
ASIC.4.12 string 0x64
ASIC.1.24 string 0x68
ASIC.1.25 string 0x08
ASIC.4.13 string 0x34
ASIC.4.14 string 0x02
ASIC.1.26 string 0xbf
ASIC.4.15 string 0xff
ASIC.4.16 string 0xd2
DISC.6 string 0x00000002
DISC.7 string 0x24800000
ASIC.4.0 string 0x04
ASIC.4.17 string 0x80
ASIC.4.18 string 0xb2
ASIC.4.1 string 0x05
ASIC.4.20 string 0x68
ASIC.4.2 string 0x06
ASIC.4.19 string 0x5c
ASIC.4.3 string 0x02
ASIC.4.21 string 0x0f
ASIC.4.22 string 0x5c
ASIC.4.4 string 0x07
ASIC.4.5 string 0x08
ASIC.4.23 string 0x08
ASIC.4.24 string 0x68
ASIC.4.6 string 0x09
ASIC.2.10 string 0x00
ASIC.4.7 string 0x05
ASIC.4.25 string 0x08
ASIC.4.26 string 0xbf
ASIC.4.8 string 0x00
ASIC.2.11 string 0x0d
ASIC.2.12 string 0x64
ASIC.4.9 string 0x0f
ASIC.2.13 string 0x34
ASIC.2.14 string 0x02
ASIC.2.15 string 0xff
ASIC.2.16 string 0xd2
ASIC.1.0 string 0x04
ASIC.1.1 string 0x05
ASIC.2.17 string 0x80
ASIC.2.18 string 0xb2
ASIC.1.2 string 0x06
Attachment 4: startup.tcl

   Setting BGColor "BlanchedAlmond"
   Setting TextColor "red"

      Setting ProjectName  AIDA
      Setting ProjectImage AIDA.gif
      Setting ProjectURL   "http://npg.dl.ac.uk/NPG/Projects/Fair.html"


  Setting Option1Name STFC
  Setting Option1URL  "stfc.ac.uk"
  Setting Option1Image stfc.jpg


# global MessageLoggerServerHost
#  set  MessageLoggerServerHost nndhcp077.dl.ac.uk

#    package require  MessageLoggerClient 

#  Setting Option3Name "Elog Server"
#  Setting Option3URL  "http://nndhcp077.dl.ac.uk:8080/DataAcq/"

#    Install required services

    if {![namespace exists ::AIDA]} {namespace eval ::AIDA {puts "created namespace ::AIDA"}}
#    set ::AIDA::TestMode 1

    global SpectrumServerHost
    set SpectrumServerHost aidas

   package require XAIDAAccessClient  ;# for Client end - ie AIDA server workstation  sets MIDAS_XAIDAAccessClient_Exists (SOAP)

   package require XAD9252AccessClient ;# for Client end - ie AIDA server workstation  sets MIDAS_XAD9252AccessClient_Exists (SOAP)

namespace eval DataAcquisition {
     variable PROJECT AIDA
     variable ACQSERVERS
     set ACQSERVERS [list aida01 aida02 aida03 aida04 aida05 aida06 aida07 aida08]
     variable ACQSERVER [first $ACQSERVERS]
    variable DataTransferStreams; set DataTransferStreams 2
#     set ::DataAcquisition::EnableMerger 1
#     set ::DataAcquisition::EnableTape 1
    variable RDOGo 0 
}

namespace eval AIDA {
      variable ModuleMenu; set ModuleMenu $::DataAcquisition::ACQSERVERS
      variable TSMaster; set TSMaster aida01
      variable FEE64 [first $ModuleMenu]
      variable ConfigBase; set ConfigBase [file join $env(MIDASBASE) config TclHttpd]
}

    Setting DataBaseRoot "/MIDAS/DB"
    package require DataBaseAccessServer

    package require NetVarsServer 1.1
    package require SigTaskServer

#    package require MemSasServer

#    Install services for Run Control

namespace eval $::DataAcquisition::PROJECT {
    variable DBRootName [file join EXPERIMENTS AIDA]
    variable DBNodeBase 2012
}

  package require AIDARunControl
#  package require TapeServer
#  package require MergerControl
#  package require MERGERunControl


  Setting Option7Name "Project Home Page"
  Setting Option7URL  "http://npg.dl.ac.uk/NPG/Projects/Fair.html"

  Setting Option8Name "Server Software Home Page"
  Setting Option8URL  "http://npg.dl.ac.uk/MIDAS/download/WebServices.html"


Attachment 5: rfs
[npg@carme-gsi rfs]$ pwd
/MIDAS/Embedded/XilinxLinux/ppc_4xx/rfs
[npg@carme-gsi rfs]$ ls -l
total 1491780
drwxrwxrwx. 19 10101  500        227 Dec  1  2022 aida01
drwxrwxrwx. 19 10101  500        227 Dec  1  2022 aida02
drwxrwxrwx. 19 10101  500        227 Dec  1  2022 aida03
drwxrwxrwx. 19 10101  500        227 Dec  1  2022 aida04
drwxrwxrwx. 19 10101  500        227 Dec  1  2022 aida05
drwxrwxrwx. 19 10101  500        227 Feb 10  2022 aida06
drwxrwxrwx. 19 10101  500        227 Feb 10  2022 aida07
drwxrwxrwx. 19 10101  500        227 Feb 10  2022 aida08
drwxrwxrwx. 19 10101  500        227 Oct 29  2021 aida09
drwxrwxrwx. 19 10101  500        227 Oct 29  2021 aida10
drwxrwxrwx. 19 10101  500        227 Oct 29  2021 aida11
drwxrwxrwx. 19 10101  500        227 Oct 29  2021 aida12
drwxrwxrwx. 19 10101  500        227 Oct 29  2021 aida13
drwxrwxrwx. 19 10101  500        227 Oct 29  2021 aida14
drwxrwxrwx. 19 10101  500        227 Oct 29  2021 aida15
drwxrwxrwx. 19 10101  500        227 Oct 29  2021 aida16
drwxrwxrwx. 19 10101  500        227 Apr 28  2021 aida17
drwxrwxrwx. 19 10101  500        227 Apr 28  2021 aida18
drwxrwxrwx. 19 10101  500        227 Apr 28  2021 aida19
drwxrwxrwx. 19 10101  500        227 Apr 28  2021 aida20
drwxrwxrwx. 19 10101  500        227 Apr 28  2021 aida21
drwxrwxrwx. 19 10101  500        227 Apr 28  2021 aida22
drwxrwxrwx. 19 10101  500        227 Apr 28  2021 aida23
drwxrwxrwx. 19 10101  500        227 Apr 28  2021 aida24
-rw-r--r--.  1 root  root 1527582720 Apr 28  2021 rfs.tar

  469   Tue Jan 16 08:57:44 2024 JG, DR, TDTuesday 16 January
10.00 Outflow manifold returned from workshop with leaking attachment point welding re-worked.

11.30 Outflow hoses re-attached to outflow manifold. Faulty male Swagelok replaced.
      Outflow manifold tested on floor - no leaks observed.

      Mounted outflow manifold. All inflow/outflow hoses attached to FEE64s - see attachments 1-4.
      FEE64 #11 detached adaptor PCB - temporarily secured by cable tie.

      Inflow/outflow valves opened. Pin hole leak observed at 3/8" connector gasket - indicated by arrow attachment 
3.
      Leak fixed by tightening fitting.

      1x spare inflow/outflow connected together - attachment 4.

11.40 CARME IE514 3.8e-11mbar, ambient temperature 16.5 deg C

15.05 8x AIDA FEE64 power cables ( #1 - #8 ) installed FEE64s #1 - #8
      Note - Samtec lock clips broken on 3 of 8 cables due to careless handling/poor storage
      AIDA FEE64 power cables tied to one of the FEE64 copper coolant pipes.

      See attachments 5-7

15.30 USB-controlled ac mains relay - USB connected to RPi CARMEbakingpi currently used for CAEN N1419ET 
      Rear panel key switched to override - no flow/d.p. sensors currently installed - front panel water ( yellow ) and override ( red ) LEDs active.

16.15 Transfer CARMEbakingpi from CRYRING network to AIDA 24-port Gbit network switch connected to carme-gsi
      CARMEbakingpi becomes nnrpi1 p- can connect OK via ssh from carme-gsi

      Start server ( port 8015 ) manually

      ssh pi@nnrpi1
      
      /MIDAS/Tcl/Httpd/linux-arm/Tcl-Httpd-server &

      See attachment 8
Attachment 1: 1000009654.jpg
1000009654.jpg
Attachment 2: 1000009651.jpg
1000009651.jpg
Attachment 3: 1000009653.jpg
1000009653.jpg
Attachment 4: 1000009652.jpg
1000009652.jpg
Attachment 5: 1000009663.jpg
1000009663.jpg
Attachment 6: 1000009662.jpg
1000009662.jpg
Attachment 7: 1000009664.jpg
1000009664.jpg
Attachment 8: Capture.PNG
Capture.PNG
  468   Mon Jan 15 18:51:28 2024 JG, DR, TDMonday 15 January
Cooling water manifolds


The manifolds were returned this afternoon.

Inflow manifold - blue hose, female Swagelok connectors

Attached to inflow hose and bled using bleed/drain valve.
Small leak observed at the 3/8" connection between the manifold and the bleed/drain valve - attachment tightened and leak fixed.
Inflow manifold installed.

Outflow manifold - red hose, male Swagelok connectors

Attached to outflow hose and bled using bleed/drain valve.
Small leak observed at the weld immediately *below* the weld that had been re-worked - this weld had not been re-worked.
Significant leak observed at one of the male Swagelok connectors in the lower half of the manifold - it does not seal correctly.


DR will return this manifold to the workshop for welding re-work Tuesday am.
The faulty male Swagelok will be replaced.


AIDA cabling


16x AIDA PSU, HDMI and RJ45 Cat 6 cables and 4x SHV-BNC cables required.

Available

8x AIDA PSU cables 
1x AIDA PSU cable ( in one of two white plastic storage boxs with 7x FEE64s returned from STFC DL ) now appears to be missing

16x AIDA HDMI cables 

8x RJ45 Cat 6 network cables
1x RJ45 Cat 6 network cable ( network switch to CRYRING network patch panel - to be replaced )
3x RJ45 Cat 6 network cables found in AIDA plastic box ( S4 Messehutte )

3x SHV-BNC 

Will check DESPEC store above ESR with NH tomorrow afternoon.
  467   Sun Jan 14 12:31:32 2024 JM, CBFEE positions mac addresses

*Updated following FEE changes. Correct as of 31/01/24

FEE modules are labelled according to the diagram in attachment. Note attachment is looking downstream. Each FEE module is also labelled with its assigned number and whether it is a bias connection on the FEE holders. MAC addresses and FEE descriptions are as follows;

1 -  41:d8:20  - top left detector -  pn bias

2 -  41:d8:2a  - top left detector - 

3 -  41:d0:05  - top left detector - 

4 -  41:b4:0b  - top left detector -  nn bias

5 -  42:0d:0c  - top right detector -  pn bias

6 -  41:0d:16  - top rightdetector - 

7 -  41:b4:16  - top right detector - 

8 -  41:ba:2b  - top right detector -  nn bias

9 -  41:d0:0e  - bottom left detector -  nn bias

10 -  41:d8:2b  - bottom left detector - 

11 -  41:d7:cc  - bottom left detector - 

12 -  41:d7:c2  - bottom left detector -  pn bias

13 -  41:ba:22  - bottom right detector -  nn bias

14 -  41:ee:72  - bottom right detector - 

15 -  41:a0:71  - bottom right detector -

16 -  41:ee:71  - bottom right detector -  pn bias

Attachment 1: CARME-FEEs.jpg
CARME-FEEs.jpg
  466   Sat Jan 13 14:32:21 2024 CB, JMLeaky manifold and cable management

Found two leaks in the welding of the manifolds, marked them in tape. Manifold to be repaired on Monday. Remember to re-install missing hoses, which are  in a bucket in the cave.

See attached how it is supposed to look like.

Left hand hose (inlet) was too short and was cut to dismount the manifold. Davide will obtain a new tube of the same type and mount it for us. NOTE THAT THE INLET connector (BLUE manifold) was leaking. Unclear if this was the hose or poor plumbing.

Installed cable management system that was never used during the last beamtime. Started to thread cables through it. There should be sufficient space to re-install the manifolds with the cable tray in place. If this is difficult, the tray is installed with zip ties only so altering position should be relativly straightforward. The mounting points of the outlet manifold on the CARME chamber can also be rotated around to give a bit more space.

Cleaned up some boxes and moved heaters and some empty boxes to the CRYRING roof.

Attachment 1: 20240112_111838.jpg
20240112_111838.jpg
Attachment 2: 20240112_105936.jpg
20240112_105936.jpg
Attachment 3: 20240113_183019.jpg
20240113_183019.jpg
  465   Sat Jan 13 14:25:05 2024 CB, JMBias tests with all FEEs connected

Connected all FEEs.

The two top fees corresponding to the bottom right detector are connected to damaged flexi rigid PCBs. The top one is missing an pin that should not be connected to anything, while the bottom one has a slightly cracked connector and is missing the top left pin. Spares should be in Edinburgh.

Found that the flexi rigid PCB connected to the bottom most FEE of the bottom-left detector has a damaged ground lemo connector. Possibly pulled during mounting? The bias cables were off most of the time. Seems to bias fine using the other bias lemo, but not ideal.

Attached, a V-I plot of the four detectors being biased. Note bottom-left was only biased to 120V vs. 150 since it shows relatively high leakage current. Time is wrong - bias was given over ~2 minutes. Unclear why the x-axis is incorrect.

Leakage currents for all detectors, and especially bottom-left and bottom-right drop significantly waiting a sufficently long amount of time, e.g. down to ~5.5 uA after 15 minutes.

 

 

To create this plots in case it is needed.

ssh to CARMEbakingpi

cd Programs/caenlogger

python3 caenlogger.py

This will capture the data and pass it to the server (carme-gsi) in ~/caen_current.dat where they will be APPENDED. Delete / save / move the file from the CARME server if required. Backups are in Programs/caenlogger/bias_logs

 

THEN Bias on

 

THEN

ssh carme-gsi

cd Programs/caenlogger

gnuplot plot_current.gp

 

 

Attachment 1: Screenshot_2024-01-13_15-24-40.png
Screenshot_2024-01-13_15-24-40.png
Attachment 2: 20240113_150158.jpg
20240113_150158.jpg
  464   Thu Jan 11 19:57:20 2024 CB, JMAll FEE holders mounted

After discussions with Davide, changed the holes in the FEE holders for the bottom-right detector to slots in order to slightly change the angle that the FEE inserts into the flexi rigid PCB. This was sufficient, with some difficulties, to insert both FEEs into the flexi rigid.

Attempted to finally mount the last two fees, the top two for the bottom-right detector, which should have been very easy. Somehow managed to break one pin on each PCB in similar spots. One is in the third column (not in use), and one is in the fourth column (in use). This latter PCB connector was unfortunately dropped while trying to dismount it, and the ERNI connector cracked. That connector can be used anymore? Cannot find spares in GSI, must be in Edinburgh?

 

Dismounted all the hoses from the manifold, and starting to remove teflon from the plugs. Will use instead gaskets + anti-sieze compound. Quite a number of plug thread are damaged, likely in welding. Unclear if this was the reason behind the leaks. Should be done with the manifolds tomorrow.

  463   Wed Jan 10 19:27:08 2024 CB, JMManifold and more bias tests

Mounted water manifold on CARME with help from GSI plumber. Connected all the hoses with Swagelok self-locking connectors. Tested for water leaks - it leaks from nearly every port. The issue appear to be the connections between the manifold and the plugs, as opposed to the plugs and the hoses which do not leak.

Will try to find a solution tomorrow.

 

FEEs connected from bottom-left, top-left and top-right detector.

Top right detector showed same behaviour described in previous ELOG entry. Removed all jumpers from PCBs. Beahaviour is gone.

Removed all jumpers from flexi-rigid PCBs for bottom-left detector. Behaviour is improved, but not gone - still 13 uA at 150V. Around 3.5 uA at 120V with no conditioning - that may be the working point.

 

Missing

- Power cables and ethernet cables for half the fees

- Fix leaks on water manifold

- Fix / change / re-arrange fee holder for bottom-right detector to avoid it fouling on CARME frame.

  462   Wed Jan 10 10:22:41 2024 CB, JMSome FEEs mounted - issues with Det#2 (bottom left)

Started mounting some FEEs. We ran into some mechanical issues due to the new ERNI connectors on the new flexi-rigid adaptor PCBs. Davide skimmed 6.5 mm off the edge with the groove (the groove is gone). This is suffiicent to support the flexi-rigid and in spite of our earlier concerns, the flexi-rigid can be mated/un-mated with the FEEs without major alignment issues.

We still have issues for the bottom-right detector, where the FEE holder will just not fit - will probably figure something out with Davide.

 

We observe a peculiar issue with the bottom-left detector. At 150V, the leakage current for this detector as tested with just the adaptor cards on the feedthroughs that carry the bias (no FEEs) is of the order of 3 uA.

We mounted all four flexi-rigid PCBs - still 3 uA.

We mounted the FEE that slots into the bottom-most feedthrough (which corresponds to the n-n bias). Still 3 uA.

We mounted the FEE that slots in to the second bottom feedthrough (not corresponding to bias). 15 uA! Upon lowerint the bias, the leakage current goes to zero around 30V as opposed to 0V like all the other detectors, and like this detector when this particular FEE is not mounted.

Tests were performed exchanging all PCBs, and all FEEs. The same behaviour is observed. It is not an issue with the FEE or the PCB. We suspect some cross-contact between detector bias and some strip that feeds into the second FEE? In any case, it is most likely a vacuum side issue - nothing we can do now.

 

The detector can be biased to 80V or so without major issue, and it appears this behaviour improves over time, i.e. leaving the bias on reduces the leakage current dramatically so it may even be possible to reach 150V.

See plots attached. Note in 2nd plot the plotting script has some issues - detector names are correct until the bias in Ch#2 is brough to 150V, then the colours are incorrect. Only Ch#2 was being changed. Note the drop in leakage current by a factor 2 just increasing the bias to 150V for a few seconds and returning down.

Attachment 1: Screenshot_2024-01-10_11-59-39.png
Screenshot_2024-01-10_11-59-39.png
Attachment 2: Screenshot_2024-01-10_12-56-36.png
Screenshot_2024-01-10_12-56-36.png
  461   Mon Jan 8 19:29:57 2024 CB, JMManifold repaired + fee assembly in bench

Davide extracted the stuck blind plugs and re-threaded all missing ports. Completed manifold assembly, except for main inlet/outlet - left for plumber.

Cut 8x2 2m hoses and assembled with Swagelok quikfit and teflon ferrules.

We're missing back ferrules in brass for the repaired fees. It should be possible to buy them from GSI.

 

Attempted assembly of fee, Flexi rigid and black delrin support on bench. Davide shimmed the black delrin supports by 6.5 mm from ridge. They now fit length wise on new adaptor cards. Unfortunately, the decision to flip the connector means the Flexi rigid is no longer at the correct height for the erni connector in the fee. See attached.

Inserting the Flexi rigid without the black support is too dangerous, and defeats the point of having rails. Will probably mount Flexi rigids tomorrow on black support, not screw into the fee holder, and align by eye. Tried once and it works - see last two attachments - but it will have unclear impact on the ease to unmount the fee without damaging the Flexi rigid and will make mounting fee holder on flanges harder. No alternative solutions in current timeframe.

Attachment 1: 20240108_164756.jpg
20240108_164756.jpg
Attachment 2: 20240108_163218.jpg
20240108_163218.jpg
Attachment 3: 20240108_163916.jpg
20240108_163916.jpg
Attachment 4: 20240108_164010.jpg
20240108_164010.jpg
  460   Mon Jan 8 19:27:36 2024 CB, JMLeak current tests overnight

Left bias in overnight. Leakage current decreased and stabilised. See attached. Bias off as we prepare to mount the fees.

Attachment 1: Screenshot_2024-01-08_09-56-20.png
Screenshot_2024-01-08_09-56-20.png
  459   Sun Jan 7 15:38:57 2024 CB, JMManifolds mostly assembled + bias overnight check

Assembled most of the connectors on the new (green) manifolds. Two blanking pulgs stuck when mounting - cannot remove them or tighten them. Swarf? Will ask Davide's help tomorrow. Cannot use the manifold until this is fixed.

Checked FEE assembly in cassette. Need to shorten black Delrin supports by ~6 mm. Holes should end up in the correct position thanks to rails being effectively longer (same length, different hole position on FEE cassette). Will ask Davide to skim one if possible. If not, will have to mount without supports.

Finally managed to get the CAEN program working to control N1419 remotely from Baking Pi. Plotting bias and leakage current overnight to test stability. See attached. IE514 sensor off for these tests.

Attachment 1: Screenshot_2024-01-07_16-38-26.png
Screenshot_2024-01-07_16-38-26.png
  458   Fri Jan 5 12:29:10 2024 JM, CBFurther bias tests

Biased detectors this morning to 150V, the leakage current for all detectors appears stable and is ~3.5 uA. We have also replaced the CAEN bias module so that all channels are now functional. Detectors are connected as follows;

CHN0 -> Top left

CHN1 -> Top Right

CHN2 -> Bottom Left

CHN3 -> Bottom Right

Reconnected the motors and moved the detectors while biased. No observable change in the leakage current during movement at 150V. An increase in the pressure of the electron target was observed during movement (up to 1e-9 mbar). This increase was only momentarily, and the pressure recovered after a couple of seconds. We expect the pressure increase from movement to decrease over time.

 

  457   Thu Jan 4 13:15:10 2024 CB, JMTent removed - bias test

Removed tent, octagon, Neg cables.

Disconnected and removed Baking Pi, TC08, thermocouples.

Tested new rails, plates, fees - it fits. Need to check on the flange next.

Will soon turn off CARME IE514 gauge to bias test the DSSDs.

All four detectors were bias tested using the new flexi-rigid adaptor cards. Note the LK-1 jumper was removed from the adaptor cards so that 'open line' was observed between the bias lemo and the ground. Attachment 1 shows VI curves for all detectors. The maximum leakage current for all detectors appears lower than previous bias tests eg elog 400 -> temperature was ~17C today compared to ~20C for elog 400. We also performed a bias test using the old adaptor cards to see if there was any difference between the old and new cards. No significant difference was observed. For the top left detector we initally observed a significantly large leakage current (4.2 uA at 20V, 8.7 uA at 30V) which had ~1 uA fluctuations. We left for 5 minutes to see if it stabilised and the leakage current suddenly dropped to what is shown in the plot. This leakage current then appeared stable for the rest of the test. Perhaps this effect was observed by Tom but for the top right detector in elog 421?

We plan to connect the bias as we did in the 14N+d measurement so that we can monitor and plot the leakage current over the weekend.

Attachment 1: Bias.png
Bias.png
  456   Wed Oct 18 16:47:36 2023 JG, CBPower outage

Local power outage at the CRYRING today, while JG was trying to activate the ion pumps. Unlcear what caused it.

Valve on scroll worked as intended, but CARME reacehd 1E-7 mbar from turbo (?) venting.

Power restored. Pressure back in the 1E-11 mbar range, decreasing.

Temperature still well above room.

  455   Tue Oct 17 08:43:47 2023 CBNEG activation -2

09:40 - Found temperature and pressure stablised, but too high to comfortably activate NEGs closest to detectors. Detector temperature around 90C. This was expected. Ramped down UHV1400 #1,2,3,4 to P=50W (200C?). Ramped down heaters to 100, 80, 100C. Ramped down heating jacket to 60C.

10:00 - Activated set #5 (x2) at 80W tramp=30, thold=infinity; set#6 (x3) at 120W rest as #5. This is different from previous procedure because we have 4x DSSD but sensor only one one of them. It seems less dangerous to heat up everything at once so that the one TC we have is hopefully representative of the situation.

10:15 - CARME cooling. No visible effects on temperature. Tried P#5 100, P#6 150 W. (500C roughly?) Will probably have to decrease.

10:25 - Detectors getting hotter quickly, but still below 100C. No changes for now.

10:35 - As expected, detectors are getting too hot, too quickly. Decreased power to 60, 90 W

10:43 - Ramped down #5,6 to 40, 60 W. Should be around 200C at equilibrium.

10:45 - Also decreased #1,2,3,4 to 40W, and D2000 to 150C.

10:52 - Further decreased #5,6 to 10, 15W.

11:04 - Temperature back under control

11:10 - Detectors actually cooling now. Increased #5,6 to 20,30 W.

11:18 - Detectors still cooling. Increased #5.6 to 30,45W. If stable plan to leave like this until later this afternoon.

13:23 - Temperatures stable. Increased #5,6 to 40, 60 W. Decreased heaters to 80,80,80 C. Decreased jacket to 40C. Opened the tent partly to connect the vacuum division non-bakeable ion pump cable.

13:35 - Ion pump cable connected and tent closed as best I can. Interestingly, the pressure is going up probably as a result of the detector/kapton cables temperature going up due to the slight increase in #5,6 power.

14:15? - Graziano S. flashed the vacuum divison ion pump.

14:25 - Pressure keeps increasing slightly. I am taking this as a sign the kapton is outgassing, which is being made to out gass which is good. Decided to leave temperatures as they are until later today and start cooling down in the late afternoon / evening. Decreased slightly #5,6 to 34, 51W. Increased D2000 to 200C. Will leave like this if stable.

15:00 - Looks stable. See attachment 3. Leaving for a few hours, before proceeding to gradual cool down overnight. Will ask Jan for help in turning everything really off.

18:00 - All NEGs off. All apadev 10&11 channels to 40C. Ion pumps remain off. Asked JG to come tomorrow late morning and turn the apadev channels off and the ions pumps on.

Attachment 1: Screenshot_2023-10-17_09-45-23.png
Screenshot_2023-10-17_09-45-23.png
Attachment 2: Screenshot_2023-10-17_11-04-09.png
Screenshot_2023-10-17_11-04-09.png
Attachment 3: Screenshot_2023-10-17_14-56-55.png
Screenshot_2023-10-17_14-56-55.png
Attachment 4: Screenshot_2023-10-17_17-58-24.png
Screenshot_2023-10-17_17-58-24.png
Attachment 5: Screenshot_2023-10-17_17-58-51.png
Screenshot_2023-10-17_17-58-51.png
  454   Mon Oct 16 08:28:02 2023 CBNEG activation -1

9:00 - Michael's online vacuum pressure plot is no longer working - an old image is being displayed. CARME temperatures around 80-85C. Lowered heater temperatures to 150, 100, 150 C for safety. Lowered jacket temperature to 80C

9:15 - Flashed ion pumps several times. Small pressure increase, up to ~7E-9 but not macroscopic

9:30 - Started D-2000 #1,2. T=400, tramp 30, thold 90 (this was a mistake 1,3 are behind the detectors, 2,4 are to the side. They should ideally be activated together. Got confused. Not a big deal.)

9:50 - Online vacuum plot is working again. Pressure changing, internals around 80C

10:00 - Internal temperature still stable at 80C. Started D-2000 #3,4. T=400, tramp 30, thold 90

11:30 - For unclerar reasons the pressure has been increasing over the last half hour, after the four NEGs were activated. Nikos and the alignment group were adjusting the feet of the interaction chamber to improve the alignment. Hopefully this is not a leak. Section 10 was apparently vented overnight, and may be in air at the moment.

11:45 - D2000 activation complete. Set temperature to 200C for all four of them. Altough the temperature is still >375C, the pressure drops very quickly. This is good, but I have not observed this behaviour before.

13:10 - Pressure and temperature mostly stable. Turning on UHV1400 groups #1, 2 with at P=250W, tramp 30s, thold 90s.

13:35 - No effect on DSSD temperature. Increasing P->300W as per previous activation procedure.

13:43 - Still no sizeable effect. Increased P->350 W

15:00 - Completed activation of UHV1400 sets #1,2. Set power to 100W. Activated sets #3,4 at P=250W, same settings as other two.

15:30 - Increased UHV1400 #3,4 to P=300 W. Temperatures OK.

15:37 - Increased UHV1400 #3,4 to P=350 W. Temperatures increasing but still acceptable.

16:06 - Temperatures are starting to get a bit too high. Decreased P->300W. Also decreased heaters to 125, 80, 125C and baking jacket to 70 C.

16:26 - Decreased UHV1400 #3,4 to P=250W.

17:05 - Completed activation of UHV1400 sets #3,4. Set power to 100W. Planning to leave stabilising overnight.

Attachment 1: Screenshot_2023-10-16_17-17-30.png
Screenshot_2023-10-16_17-17-30.png
  453   Sun Oct 15 09:35:03 2023 CBStarting ramp down

10:40 - Status as attached. Heaters setpoint/read at 250 C, 150 C, 225 C. Power stable at 53%, 41%, 63%

10:45 - Informed HKR ramp down / pump activation is starting for CARME only. Setpoints to 200, 125, 200 C. Monitoring slow ramp down.

14:10 - Heater temperatures reached the set points. Power stabilising. CARME internals temperature dropped to around 100 C and still falling. Waiting until later today to further reduce heater setpoint. Will probably aim for 80 C before beginning activation.

15:45 - Heater setpoints to 175, 100, 175 C

15:50 - Interaction chamber jacket setpoints to 100C

18:00 -  CARME temperature decreasing, not stable yet. Will check later tonight.

20:43 - CARME internals around 85C, not quite stable still. Waiting overnight.

Attachment 1: Screenshot_2023-10-15_10-35-47.png
Screenshot_2023-10-15_10-35-47.png
Attachment 2: Screenshot_2023-10-15_17-56-00.png
Screenshot_2023-10-15_17-56-00.png
  452   Tue Oct 3 08:48:36 2023 TDTuesday 3 October
09.40 CARME IE514 1.79e-8 mbar, ambient temperature 20.7 deg C

      CARME bakeout monitor, ELOTECH R4000 temperature controller front panel and Grafana bakeout dashboard- attachments 1-3

      No change to ELOTECH R4000 set point
Attachment 1: 1000007432.jpg
1000007432.jpg
Attachment 2: 1000007431.jpg
1000007431.jpg
Attachment 3: 1000007434.jpg
1000007434.jpg
  451   Mon Oct 2 08:41:43 2023 TDMonday 2 October
09.35 CARME IE514 1.95e-8 mbar, ambient temperature 20.8 deg C

      CARME bakeout monitor, ELOTECH R4000 temperature controller front panel and Grafana bakeout dashboard- attachments 1-3

      No change to ELOTECH R4000 set point

17.45 CARME IE514 1.91e-8 mbar, ambient temperature 20.8 deg C

      CARME bakeout monitor, ELOTECH R4000 temperature controller front panel and Grafana bakeout dashboard- attachments 4-6

      No change to ELOTECH R4000 set point
Attachment 1: 1000007409.jpg
1000007409.jpg
Attachment 2: 1000007408.jpg
1000007408.jpg
Attachment 3: 1000007407.jpg
1000007407.jpg
Attachment 4: 1000007416.jpg
1000007416.jpg
Attachment 5: 1000007415.jpg
1000007415.jpg
Attachment 6: 1000007413.jpg
1000007413.jpg
  450   Sun Oct 1 10:24:50 2023 TDSunday 1 October
11.20 CARME IE514 3.53e-9 mbar, ambient temperature 20.6 deg C

      CARME bakeout monitor, ELOTECH R4000 temperature controller front panel and Grafana bakeout dashboard- attachments 1-3

      Change ELOTECH R4000 set points for CARME heaters #1 and #3 to 200 deg C, heater #2 to 150 deg C

14.10 CARME IE514 6.12e-9 mbar, ambient temperature 20.7 deg C

      CARME bakeout monitor, ELOTECH R4000 temperature controller front panel and Grafana bakeout dashboard- attachments 4-6

      Change ELOTECH R4000 set points for CARME heaters #1, #2 and #3 to 225, 175 and 225 deg C respectively.


18.10 CARME IE514 1.92e-8 mbar, ambient temperature 20.8 deg C

      CARME bakeout monitor, ELOTECH R4000 temperature controller front panel and Grafana bakeout dashboard- attachments 7-9

      Change ELOTECH R4000 set points for CARME heaters #1, #2 and #3 to 250, 150 and 225 deg C respectively.
      All other ( active ) ELOTECH R4000 set points ( interaction chamber etc ) to 130 deg C.

20.25 CARME IE514 2.14e-8 mbar, ambient temperature 20.8 deg C

      CARME bakeout monitor, ELOTECH R4000 temperature controller front panel and Grafana bakeout dashboard- attachments 10-12

      No change to ELOTECH R4000 set points
Attachment 1: 1000007352.jpg
1000007352.jpg
Attachment 2: 1000007354.jpg
1000007354.jpg
Attachment 3: 1000007353.jpg
1000007353.jpg
Attachment 4: 1000007360.jpg
1000007360.jpg
Attachment 5: 1000007358.jpg
1000007358.jpg
Attachment 6: 1000007361.jpg
1000007361.jpg
Attachment 7: 1000007368.jpg
1000007368.jpg
Attachment 8: 1000007369.jpg
1000007369.jpg
Attachment 9: 1000007370.jpg
1000007370.jpg
Attachment 10: 1000007377.jpg
1000007377.jpg
Attachment 11: 1000007375.jpg
1000007375.jpg
Attachment 12: 1000007374.jpg
1000007374.jpg
  449   Sat Sep 30 10:58:14 2023 TDSaturday 30 September
10.50 CARME IE514 7.14e-10 mbar, ambient temperature 20.8 deg C

      CARME bakeout monitor, ELOTECH R4000 temperature controller front panel and Grafana bakeout dashboard- attachments 1-3

      Change ELOTECH R4000 set points for CARME heaters #1 and #3 to 125 deg C

14.20 CARME IE514 8.25e-10 mbar, ambient temperature 20.8 deg C

      CARME bakeout monitor, ELOTECH R4000 temperature controller front panel and Grafana bakeout dashboard- attachments 4-6

      Change ELOTECH R4000 set points for CARME heaters #1 and #3 to 150 deg C

18.30 CARME IE514 1.29e-9 mbar, ambient temperature 20.7 deg C

      CARME bakeout monitor, ELOTECH R4000 temperature controller front panel and Grafana bakeout dashboard- attachments 7-9

      Change ELOTECH R4000 set points for CARME heaters #1 and #3 to 175 deg C and heater #2 to 125 deg C
Attachment 1: 1000007236.jpg
1000007236.jpg
Attachment 2: 1000007239.jpg
1000007239.jpg
Attachment 3: 1000007237.jpg
1000007237.jpg
Attachment 4: 1000007251.jpg
1000007251.jpg
Attachment 5: 1000007250.jpg
1000007250.jpg
Attachment 6: 1000007252.jpg
1000007252.jpg
Attachment 7: 1000007256.jpg
1000007256.jpg
Attachment 8: 1000007257.jpg
1000007257.jpg
Attachment 9: 1000007255.jpg
1000007255.jpg
  448   Fri Sep 29 08:54:29 2023 TDFriday 29 September
09.50 CARME IE514 9.0e-11 mbar, ambient temperature 21.3 deg C

      CARME bakeout monitor and ELOTECH temperature controller - attachments 1 & 2


10.15 Commence bakeout ramp
      Set all ELOTECH PID temperature controller zones ( channels ) to 50 deg C
      Expect all non-CARME zones to reach temperature set point quickly. There will be a delay for the TCs corresponding to the CARME heaters but the internal CAME TCs 
will respond more slowly- will observe response of CARME TCs over time.

10.35 ELOTECH temperature controller - attachment 3

11.10 CARME TCs change from c. 25-28 deg C to 28-35 deg C

11.15 Set all ELOTECH PID temperature controller zones ( channels ) to 100 deg C

12.15 CARME IE514 1.39e-10 mbar, ambient temperature 20.8 deg C

      CARME bakeout monitor and ELOTECH temperature controller - attachments 4 & 5


14.20 CARME IE514 3.62e-10 mbar, ambient temperature 20.8 deg C

      CARME bakeout monitor, ELOTECH temperature controller front panel and Grafana bakeout dashboard- attachments 6-8

15.20 Whilst all of the ELOTECH temperature controller TCs are at their current set point ( 100 deg C ), the CARME TCs are not yet at equilibrium. Plan to leave system 
in current state until tomorrow morning. Assuming CARME is at equilibrium will then increment ELOTECH temperature controller TC set points.

17.50 CARME IE514 6.59e-10 mbar, ambient temperature 20.6 deg C

      CARME bakeout monitor, ELOTECH temperature controller front panel and Grafana bakeout dashboard- attachments 9-11
Attachment 1: 1000007175.jpg
1000007175.jpg
Attachment 2: 1000007174.jpg
1000007174.jpg
Attachment 3: 1000007176.jpg
1000007176.jpg
Attachment 4: 1000007178.jpg
1000007178.jpg
Attachment 5: 1000007177.jpg
1000007177.jpg
Attachment 6: 1000007189.jpg
1000007189.jpg
Attachment 7: 1000007190.jpg
1000007190.jpg
Attachment 8: 1000007188.jpg
1000007188.jpg
Attachment 9: 1000007227.jpg
1000007227.jpg
Attachment 10: 1000007228.jpg
1000007228.jpg
Attachment 11: 1000007229.jpg
1000007229.jpg
  447   Thu Sep 28 09:19:50 2023 TDThursday 28 September
10.20 CARME IE514 8.6e-11 mbar

      CARME bakeout monitor and ELOTECH bakeout controller status - see attachments 1 & 2

      The CARME bakeout monitor shows recovery from test yesterday evening. The ELOTECH controller was switched to manual mode for the 3x CARME zones and 5% power selected 
for 45 minutes. CARME TCs increased from mid-20s to high 20s/low 30s over a period of 45 minutes. At the conclusion of the test power was set to 0% for all 3x CARME zones 
and the mode changed from manual to controlled.
Attachment 1: 1000007142.jpg
1000007142.jpg
Attachment 2: 1000007143.jpg
1000007143.jpg
  446   Wed Sep 27 15:56:41 2023 TDWednesday 27 September
16.55 CARME IE514 8.2e-11 mbar

      Vacuum pressure increased slowly overnight from c. 6e-11 mbar yesterday to c. 8e-11 mbar today following low power startup tests of the bakeout controller heating 
the CARME interaction and detector chambers yesterday and today.


Attachment 1 - bakeout controller config courtesy Anton Kalinin

Attachments 2 & 3 - bakeout controller summary displays and CARME bakeout monitor display c. 17.00 today

The 3x CARME heaters are connected to controller #4 zones # 1-3, i.e. lower right display, zones 1-3, mains fuses F9-F11

17.30 3x ( of 4x ) SAES D-2000 ion pump controllers ( SIP modules ) switched off. Fourth controller was already switched off and SHV cable disconnected from rear panel of 
SIP module.
Attachment 1: 1000007129.jpg
1000007129.jpg
Attachment 2: 1000007127.jpg
1000007127.jpg
Attachment 3: 1000007128.jpg
1000007128.jpg
  445   Tue Sep 26 10:13:14 2023 TDELOTECH PID Temperature Controller R4000 manual ( German and English )
Attachment 1: BA-Isoheat_R4000-XX-X-000-X-X_DE_V1.03.pdf
BA-Isoheat_R4000-XX-X-000-X-X_DE_V1.03.pdf BA-Isoheat_R4000-XX-X-000-X-X_DE_V1.03.pdf BA-Isoheat_R4000-XX-X-000-X-X_DE_V1.03.pdf BA-Isoheat_R4000-XX-X-000-X-X_DE_V1.03.pdf BA-Isoheat_R4000-XX-X-000-X-X_DE_V1.03.pdf BA-Isoheat_R4000-XX-X-000-X-X_DE_V1.03.pdf BA-Isoheat_R4000-XX-X-000-X-X_DE_V1.03.pdf BA-Isoheat_R4000-XX-X-000-X-X_DE_V1.03.pdf
Attachment 2: R4000-XX-X-000-X-X_EN_V1.30.pdf
R4000-XX-X-000-X-X_EN_V1.30.pdf R4000-XX-X-000-X-X_EN_V1.30.pdf R4000-XX-X-000-X-X_EN_V1.30.pdf R4000-XX-X-000-X-X_EN_V1.30.pdf R4000-XX-X-000-X-X_EN_V1.30.pdf R4000-XX-X-000-X-X_EN_V1.30.pdf R4000-XX-X-000-X-X_EN_V1.30.pdf R4000-XX-X-000-X-X_EN_V1.30.pdf
  444   Tue Sep 26 08:08:42 2023 TDTuesday 26 September
09.07 IE514 pressure 6.4e-11 mbar, ambient temperature +20.3 deg C

      Using appc218 ( account carme ) workstation by the CARME 19" rack 
     
      ssh pi@carmebakingpi

      Re-connect to bakeout monitor app 

      screen -dR

      See attcahment 1


12.02 Vacuum group ion pump disconnected and RGA removed from CARME chamber -  outside of CARME bakeout tent 'closed' - see attachments 2-4

      Vacuum group IE514 to be disconnected


14.45 Vacuum group disconnect their IE514

      Inside of CARME bakeout tent closed

      Photos of the CARME bakeout tent inside and outside CRYRING - see attachments 5-7  
Attachment 1: 1000007087.jpg
1000007087.jpg
Attachment 2: 1000007090.jpg
1000007090.jpg
Attachment 3: 1000007091.jpg
1000007091.jpg
Attachment 4: 1000007089.jpg
1000007089.jpg
Attachment 5: 1000007096.jpg
1000007096.jpg
Attachment 6: 1000007095.jpg
1000007095.jpg
Attachment 7: 1000007094.jpg
1000007094.jpg
  443   Mon Sep 25 16:33:32 2023 TDRemoved tape from heaters TC
17.33 IE514 pressure 6.4e-11 mbar, ambient temperature +20.8 deg C

      The thermocouples on the 3x groups of heater elements have been detached from direct contact with the heating element and now float abpove the heating element by 2-
3cm - see attachments 1 & 2

      The vacuum group IE514 and Ametek RGA are still attached to CARME - see attachment 3. Have requested vacuum group remove them.

      The 'air side' of the Be window on the interaction chamber is under vacuum provided by small Agilent scroll pump.

      
Attachment 1: 1000007072.jpg
1000007072.jpg
Attachment 2: 1000007067.jpg
1000007067.jpg
Attachment 3: 1000007070.jpg
1000007070.jpg
  442   Fri Sep 22 12:35:46 2023 JMBakeout plots

Attached are plots for the previous bakeout of CARME which may aid in the bakeout to come in the following week.  

Plots relate the heater temperatures and heater powers with the chamber temperature (internal floating thermocouple) and pressure inside CARME. 

Attachment 1: Heater-chamber-temps.png
Heater-chamber-temps.png
Attachment 2: pressure-chamber-temp.png
pressure-chamber-temp.png
Attachment 3: Heater-power-chamber-temp.png
Heater-power-chamber-temp.png
Attachment 4: Heater-power-temp.png
Heater-power-temp.png
  441   Wed Sep 13 15:00:39 2023 MLVacuum and Heater temperature plots from Grafana

Attached are the plotted curves from vacuum and heater temperature (actual and set values from ISOHEAT controllers) during the bakeout process. Channel assignment as per post 433.

Attachment 1: Screenshot_2023-09-13_15-56-34.png
Screenshot_2023-09-13_15-56-34.png
Attachment 2: Screenshot_2023-09-13_15-58-06.png
Screenshot_2023-09-13_15-58-06.png
Attachment 3: Screenshot_2023-09-13_15-59-37.png
Screenshot_2023-09-13_15-59-37.png
Attachment 4: Screenshot_2023-09-13_15-59-01.png
Screenshot_2023-09-13_15-59-01.png
Attachment 5: Screenshot_2023-09-13_15-58-31.png
Screenshot_2023-09-13_15-58-31.png
  440   Sat Sep 2 10:04:55 2023 CBTurbo GV closed

Found CARME in XHV at 1.1E-11 mbar. Chamber temperature around 30C cooling slowly.

Closed GV to top turbo. Pressure goes up but appears to recover around 2.5E-11 mbar, on line with several previous observations that sectioning off the turbo results in a higher reading in the IE514.

Left turbo on. Remember to not turn it off when removing the tent.

  439   Fri Sep 1 08:11:40 2023 CBNEG activation - 2

9:10 Set heaters temperature 300 -> 250 C in preparation for cooling down and to help keep DSSD temperature down. Started activating set 6 (3 X DSSD) at 120 W as per previous procedure.

9:15 Realised previous choice is a mistake in the current configuration as the DSSD TC is closer to set 5. Set 6 off. Set 5 to 60W. No effects in temperature. Ramped slowly to 200W. No immediate effect in temperature, while vacuum goes up.

9:30 Started activating set 6 at 300W (=100 per module). This should give us the best feedback on DSSD temperature from the only TC we have since they should all be at similar temperatures. 

9:40 DSSD TC shows temperature decreasing rapidly (-17C/h). That's nonsense. The TC cannot be working properly, which means activation is now quite dangerous. Looking at strain relief TC instead from now, which is still at 85C. Set 5 down to 30W for safety while set 6 finishes ramping up.

9:45 DSSD TC still reading nonsense at 35C cooling by 20C per hour. Strain relief TC at 85C heating up by 10C per hour is more realistic. Vacuum worsening as set 6 reached 300W. Immediately ramped down to 120W. Leaving for 30 minutes, or less if strain relief temperature increases too much.

9:57 Strain relief heating quickly at 90C. Decided to turn off ALL negs for safety. Set heaters to 200C.

10:18 Strain relief temperature finally slowing down heating rate. DSSD TC beings displaying increasing temperature. The TC wire on the TC-08 module is clearly damaged.

 

After lunch, turned on the four ion elements of the D-2000. Pump#4 is not working, and apparently module #2 also has issues. Used module #4 on pump#2 to get three working pumps. Others will have to be investigated when the tent is removed and we can have some access to the pumps.

Tried closing gate valve to turbo. Vacuum does not seem to really improve. Opened. Will try again tomorrow morning.

Vacuum division ion pump on as well.

  438   Thu Aug 31 08:36:34 2023 CBNEG activation

Found CARME at 50C, a bit cooler than expected, probably due to invasive tent changes yesterday to mount missing wires. Increased heaters temperature to 300C

 

9:35 Started D-2000 #1,3 (behind the DSSD). T=400, tramp 30, thold 90

11:20 Set D-2000 1,3 to 200C. Turned on D-2000 2,4 as above 

13:15 Set D-2000 2,4 to 200C. Turned on UHV1400 set 1 at 250W, tramp 30, thold infinity (will ramp down manually as required)

13:20 No appreciable effects on DSSD temperature. Turned on set 2 as well, which should be mounted close nearby.

13:45 Both sets reached 250W. No effects on detector temperature. Increased to 300W. Vacuum gets slightly worse, but recovers quickly. Not worth further increasing power. Leaving for 90 minutes.

15:20 Strain relief TC at 75C quite hotter than detector TC at 35C. Might be normal. Vacuum division IE514 must have been degassed remotely? Decreased set 1,2 to 75W. Turned on sets 3,4 to 250W as above. 

15:45 Sets 3, 4 now at 250W. Detector TC heating quickly at 13C/h but nowhere near dangerous temperatures yet. Decreased sets 1,2 to 50W. Leaving for 90 minutes.

17:15 Sets 3, 4 to 50W. Slightly worrying temperature difference between detectors and strain relief around 65C. Different materials and different positions could explain this.

17:40 Oddly, detector temperature now going up sharply (temperature diff. down to 40C) since the uhv1400 power was decreased. Cannot think of s logical reason,  it may just be a coincidence and heat transfer is catching up.

 

Leaving overnight. Will turn on the last two groups of UHV1400 tomorrow morning.

  437   Wed Aug 30 10:28:36 2023 CBPreparing for pumps on

We need to connect the non-bakeable (max 80C) wire of the ion pump belonging to the vacuum division, and we also need to connect the two newly-repaired NEG cables to the D2000 modules. That requires partly opening the tent to gain access.

Set heaters temperature from 450 C -> 400 C - CARME floating TC temperature goes from ~98 C to ~88C.

Set heaters temperature to 350 C. Left to cool over lunch.

 

 Connected GSI vacuum division ion pump and IE514

Connected D-2000 upstream left/right with repaired cables

 

Flashed all ion pumps. Ion pump #4 was flashed a couple of times, but has now started arcing. The issue is not with the module - either the wire or the ion element are damaged.

Set heaters to 300 C. Left to cool a bit more overnight. Planning on NEG on tomorrow.

  436   Wed Aug 30 09:33:13 2023 CB, OGNEG cables repaired

Repaired Neg wires for D2000 with Oemer's help.

Cable #2 is fine. 

In Cable #4, wires 1 and 2 are swapped but it's impossible this happened when the connector came apart. The pins are crimped and secured with a screw. Most likely it was sold like this and there's no difference between wire 1 and 2 (same voltage).

Edit: SAES confirmed 1&2 are interchangeable.

Attachment 1: 20230830_102519.jpg
20230830_102519.jpg
Attachment 2: 20230830_102514.jpg
20230830_102514.jpg
  435   Mon Aug 14 09:22:56 2023 JMBakeout

14/08/23

Temperature in the baking volume is ~ 42 C and the pressure is 4.6e-8 mbar. Current heaters set point is 135 C. This will now be increased to 200 C to start increasing the baking temperature more significantly.

Some issues with the interaction chamber heaters were noticed this morning. This included very large increases in temperature despite only small increases in the setpoint. When this occurred the heater channels were simply turned off. Channel 3 on the interaction chamber (main jacket on the centre of the chamber was also observed to smoke slightly). This smoking was seen when the temperature for channel 3 was ~ 70 C. 

Changing the settings of these heaters appears to have resolved most of the issues. The heaters now ramp in a much more controlled manner and remain roughly constant at the setpoint value.  The interaction chamber heaters were all increased to 70 C with no issues observed. Channel 3 also appears OK.

The CARME heaters settings have also been changed and now the heaters temperature and power output do not oscillate as observed previously. Power output per heater is approximately 15 % of total at heater temperatures of 220 C, and this value remains consistent over time. The temperatures inside the baking volume are now ~ 50 C with the pressure equal to 7e-8 mbar. Heating has not been the priority while investigating the issues observed with the heaters.

As a precaution CARME will now be maintained at the 50C level. The interaction chamber and e-target temperatures have been set to 50 C also. If the system remains stable overnight we will again begin increasing the temperature in the morning.

15/08/23

Temperatures across the whole of YR09 have remained very stable overnight. Temperature of CARME has increased slightly and is now ~ 55 C. Pressure is ~1e-7 mbar. A brief lincrease in pressure was observed yesterday due to the de-gassing of the GSI ion pump on CARME. Plan is to now increase the temperatures of YR09 to the 100 C level 

Temperature is ~ 70 C. Heating is slower than expected as I am aiming to minimise thermal gradients across the chamber, which was previously done automatically by our baking code. Current heater temperature values are ~ 320 C. I have increased the temperatures across YR09 to 70 C to match carme . Pressure in CARME is 2.1e-7 mbar.

16/08/23 

The temperatures of the CARME heaters were further increased to ~ 420 C to further increase the chamber temperature. Average chamber temperatures of ~93C achieved. Heater temperatures can be pushed to 450C which should allow us to achieve 100 C. Will allow chamber to equillibriate as temperature gradient between top and bottom is quite large.

17/08/23  

Baking will be unsupervised over the weekend. To remain safe, the berylium window and the interaction chamber will be kept at a temperature of 50C. Carme will be reduced to ~60C during this period.

Attached are plots for the bakeout period concerning this week.  

Attachment 1: 20230814_150430.jpg
20230814_150430.jpg
Attachment 2: 20230815_094400.jpg
20230815_094400.jpg
Attachment 3: Heater-power.png
Heater-power.png
Attachment 4: Heater-ITC-temp.png
Heater-ITC-temp.png
Attachment 5: Pressure-ITC-temp.png
Pressure-ITC-temp.png
  434   Sat Aug 12 15:06:51 2023 JMInitial bakeout tests

12/08/23

Baking testing. Aim to find relationship between heater temperature setpoint and the temperature of the chamber and floating temperature in the baking volume. Behaviour of the heater tempertures and ability to control the heating is also required. Ambient chamber temperature ~22 C. Interaction chamber temperatures will be raised to the level of the CARME chamber throughout, starting temperature of the interaction chamber with heating off is 28 C. P = 7.5e-9 mbar

CARME thermocouple temperatures can be read from the monitoring code which is currently running in a screen session on the baking pi. Heater temperatures and pressure in YR09 can be read from Grafana.

15:40 CARME heaters turned on, set point 50 C. Ramping is ~ 1C per minute for the heaters. 

16:00 Thermocouples inside baking volume show small increases. Setpoint increased to 60 C.

16:10 Setpoint increased to 80 C.

16:30 Setpoint increased to 100 C.

16:45 Setpoint increased to 120 C

17:00 Setpoint raised to 150 C. Average chamber temp ~ 25.2 C, bottom of the chamber is ~28C. 

17:45 Setpoint was maintained at 150 C, temperatures of the heaters continued to rise past this setpoint and were floating about 200-230 C with large variations in the temperature. At lower setpoints the heater temperature appeared to rise linearly with the 1 C per minute. I do not believe I currently have control of the heating. Average chamber temp 28.8 C with bottom of the chamber ~35-37 C. Heating has been turned off while I grab some food. P = 1e-8 mbar

Next test will be to try and maintain a consistent heater tempreture for a longer period of time, probabably at a lower set temperture. 

13/08/23

10:00 Heater setpoint to 50 C. Will maintain at this level for a while to look at increase in chamber temps and see if heater temp remains stable. P = 8.52e-9 mbar

11:47 Heater temperatures have been constant at ~ 55 C for the last hour. Little heating observed in the chamber, requires much larger temperature to heat the baking volume. Setpoint increased to 100 C.

13:10 Heater temperatures have been constant at ~103 C for the last  40 mins or so and appears stable. Temperatures are rising at a steady rate but temps are still low. Interaction chamber heating jackets activated to be a same temperature as CARME. Will require higher temperatures to heat the baking volume.

16:15 Heater temperaturtes have remained constant at ~103 C. Temps inside the baking volume remain low, Increasing heater temperature to 135 C.

18:30 Heaters temperatures remain stable. Larger fluctations are apparent than at a setpoint of 100C with temps between 140 and 150 C. Chamber temperatures increasing avg is ~ 35C, was aiming for 50C over the weekend. Increasing setpoints to 170 C.

20:50 Heater tempretures are stable. Average chamber temp is ~ 40 C. Interaction chamber heaters set to 40 C to keep constant with CARME. Pressure has increased to 2.9e-8 mbar. Heater power outputs are visible in last image, heaters currently turn on at ~80% max power about twice every 5 mins. I expect temps of 250-300 C are required to obtain significant temp increase up to 100 C on the chamber. Chamber temps appear to be fairly stable at this point, will leave for now and check back in a few hours.

23:00 Heater setpoint temp reduced to 135 C to maintain temperatures inside the baking volume overnight with no increase. Heater power has now reduced to < 50 % full power when the eaters are on. Pressure in the chamber is 4.4e-8 mbar.

I believe I now understand the heating system and plan to increase the chamber temperature throughout the day tomorrow. Aim is to get close to full bakeout temperatures.

Attachment 1: 20230813_114600.jpg
20230813_114600.jpg
Attachment 2: 20230813_132550.jpg
20230813_132550.jpg
Attachment 3: 20230813_170829.jpg
20230813_170829.jpg
Attachment 4: 20230813_205613.jpg
20230813_205613.jpg
Attachment 5: 20230813_210238.jpg
20230813_210238.jpg
  433   Sat Aug 12 11:57:51 2023 JMUpdated thermocouple positions and heater channel layouts

Heater channel layouts for the interaction chamber and carme in attached documents. 

Updated thermocouple positions; (front section is closest to gas-target, rear section is moving detector end)

  • 1-1 Internal TC 1
  • 1-2 Internal TC 2
  • 1-3 Internal TC 3
  • 1-4 CARME - top of the front section
  • 1-5 CARME - bottom of the front section
  • 1-6 CARME - top of the mid section
  • 1-7 CARME - middle of the mid section
  • 1-8 CARME - bottom of the mid section
  • 2-1 Float - bottom of mid section 
  • 2-2 CARME - middle of rear section 
  • 2-3 Float - middle of the front section
  • 2-4 CARME - bottom VAT valve
  • 2-5 Float - middle of the rear section 
  • 2-6 CARME - Top elbow
  • 2-7 Float - top of the mid section
  • 2-8 CARME - rear beam pipe

Thermocouple posiitons have been updated in the monitoring code, including the averages for the CARME chamber. 

Attachment 1: Cryring-YR09-CARME-Layout.pdf
Cryring-YR09-CARME-Layout.pdf
Attachment 2: Cryring-YR09-CARME-Layout2.pdf
Cryring-YR09-CARME-Layout2.pdf
Attachment 3: Cryring-YR09-CARME-Layout3.pdf
Cryring-YR09-CARME-Layout3.pdf
Attachment 4: Cryring-YR09-eTarget-Layout1.pdf
Cryring-YR09-eTarget-Layout1.pdf
Attachment 5: Cryring-YR09-eTarget-Layout2.pdf
Cryring-YR09-eTarget-Layout2.pdf
Attachment 6: Cryring-YR09-eTarget-Layout3.pdf
Cryring-YR09-eTarget-Layout3.pdf
  431   Thu Aug 10 07:51:43 2023 JMBaking tent and heater testing

09/08/23

Attached is the current pumpdown plot taken from grafana. 

I have started to re-work the baking tent in order to close gaps from its initial installation. The heaters work and were tested to a temperature of 50C. Testing is required to gauge the heater temperature required to heat the baking volume, we suspect the heaters will be several hundred degrees. 

11/08/23

Baking tent has been fully constructed again, see photos attached.  Some thermocouples have been re-positioned, new thermocouples positions are given in elog 433. GSI ion pump cable has been connected. Roughing pump on the outside of the berylium window has been connected and berylium window is at < 10 mbar. 

Attachment 1: current-pump-down.png
current-pump-down.png
Attachment 2: 20230812_120951.jpg
20230812_120951.jpg
Attachment 3: 20230812_120925.jpg
20230812_120925.jpg
Attachment 4: 20230811_152942.jpg
20230811_152942.jpg
Attachment 5: 20230811_152926.jpg
20230811_152926.jpg
Attachment 6: 20230811_152923.jpg
20230811_152923.jpg
Attachment 7: 20230811_152918.jpg
20230811_152918.jpg
Attachment 8: 20230811_152913.jpg
20230811_152913.jpg
  430   Thu Jul 27 09:43:56 2023 RSSThermocouples connected -2
10:30  IE514 1.80e-08mbar

Most of the CARME is covered by the baking tent. The tent needs to be closed and a bit covered from the front (upstream), end (downstream), and bottom side. TCs are labeled with a marker in the sequence 1-1 to 2-8.

1-1 Internal TC 1
1-2 Internal TC 2
1-3 Internal TC 3
1-4 CARME front top
1-5 CARME front bottom
1-6 Top hanging middle
1-7 CARME middle middle
1-8 CARME middle bottom

2-1 CARME middle top
2-2 CARME back middle
2-3 Elbow
2-4 Valve
2-5 Magnet pipe
2-6 CARME back top
2-7 CARME back float
2-8 CARME back bottom

Also, TC extensions would be needed to connect to TC-8 modules.

Attachment 1: IMG_9920.jpg
IMG_9920.jpg
Attachment 2: IMG_9919.jpg
IMG_9919.jpg
  429   Wed Jul 26 08:28:09 2023 RSSBuilding tent
09:26  IE514 1.96e-08mbar
Attachment 1: IMG_9897.jpg
IMG_9897.jpg
Attachment 2: IMG_9899.jpg
IMG_9899.jpg
Attachment 3: IMG_9901.jpg
IMG_9901.jpg
Attachment 4: IMG_9902.jpg
IMG_9902.jpg
  428   Tue Jul 25 09:12:15 2023 RSSBuilding tent
10:10  IE514 2.0e-08mbar

Upstream part of the CARME is covered.

Attachment 1: BD01C0C8-5181-49B8-8751-AD4C99A459CC.jpeg
BD01C0C8-5181-49B8-8751-AD4C99A459CC.jpeg
Attachment 2: 7B3836F1-8A7E-491B-AFDC-46033BF84760.jpeg
7B3836F1-8A7E-491B-AFDC-46033BF84760.jpeg
  427   Mon Jul 24 10:40:14 2023 RSS, TDPump down continues
11.40 Pirani 1.00e-3 mbar IE514 2.23e-08mbar

15.15 7d/24h vacuum plots
Attachment 1: vacplot.png
vacplot.png
Attachment 2: vacplot24.png
vacplot24.png
  426   Sun Jul 23 10:13:48 2023 RSS, TDBuilding tent
11.15 Pirani 1.0e-03mbar IE514 2.72e-08mbar

16.35 Pirani 1.0e-03mbar IE514 2.61e-08mbar
Attachment 1: DBE0DA2E-2F69-469C-B645-3E33E3D82138.jpeg
DBE0DA2E-2F69-469C-B645-3E33E3D82138.jpeg
Attachment 2: 857185AC-3571-41E9-88AE-9F624445135B.jpeg
857185AC-3571-41E9-88AE-9F624445135B.jpeg
Attachment 3: E69ED1D1-444E-478A-A3B4-481023B4024D.jpeg
E69ED1D1-444E-478A-A3B4-481023B4024D.jpeg
Attachment 4: 610048A3-25F5-459C-9EF1-83EFEFDE51C9.jpeg
610048A3-25F5-459C-9EF1-83EFEFDE51C9.jpeg
  425   Sat Jul 22 09:19:59 2023 RSS, TDThermocouples connected
10.20 Pirani 1e-03mbar IE514 3.26e-08mbar


12.53 CARME Leybold IE514 gauge OFF


Top right DSSSD bias -20V leakage current -10.51uA


Three internal ThermoCouples (TCs) -labeled 1, 2, and 3 in attachment 1 - were attached and tested on the TC-08 boards. Only 2 showed the ambient temperature, see attachment 3. 1 and 3 showed some strange numbers, see attachments 2 and 4 respectively.


15.28 CARME Leybold IE514 gauge OFF. Top right DSSSD bias -20V leakage current -10.44A


All the TCs are attached to TC-08 modules. See attachments 5-7. Once all TCs were plugged into TC-08 modules, they were fine.


TC-08 A0067

1. Internal TC 1
2. Internal TC 2
3. Internal TC 3
4. CARME front top
5. CARME front bottom
6. Top hanging middle
7. CARME middle middle
8. CARME middle bottom



TC-08 A0066

1. CARME middle top
2. CARME back middle
3. Elbow
4. Valve
5. Magnet pipe
6. CARME back top
7. CARME back float
8. CARME back bottom

Heating Jackets placed beneath the heaters. See attachments 8-10.
Attachment 1: 8FE76383-1CCE-45AE-A589-ABD38F1BB304.jpeg
8FE76383-1CCE-45AE-A589-ABD38F1BB304.jpeg
Attachment 2: 482F59E4-AC5E-4DEC-8D99-3DE1E9CD77C2.jpeg
482F59E4-AC5E-4DEC-8D99-3DE1E9CD77C2.jpeg
Attachment 3: 6FAE2064-8F7B-4CF7-ADA2-B89B7DD39FB7.jpeg
6FAE2064-8F7B-4CF7-ADA2-B89B7DD39FB7.jpeg
Attachment 4: F6628AD2-D2B2-4DF3-A2E9-2E56BEFC6D98.jpeg
F6628AD2-D2B2-4DF3-A2E9-2E56BEFC6D98.jpeg
Attachment 5: 8D296DD9-0285-4254-8691-7E2CB42A0623.jpeg
8D296DD9-0285-4254-8691-7E2CB42A0623.jpeg
Attachment 6: 72C0BF3E-A0C3-496A-A37D-5893C7302462.jpeg
72C0BF3E-A0C3-496A-A37D-5893C7302462.jpeg
Attachment 7: FDF5D49E-FAA2-4CC9-A416-727FDA95E8C3.jpeg
FDF5D49E-FAA2-4CC9-A416-727FDA95E8C3.jpeg
Attachment 8: B852C067-86CC-4B9A-9B19-2786C6DF96E6.jpeg
B852C067-86CC-4B9A-9B19-2786C6DF96E6.jpeg
Attachment 9: 4DD9700B-B59C-4BE5-99D7-00364C44D590.jpeg
4DD9700B-B59C-4BE5-99D7-00364C44D590.jpeg
Attachment 10: E99CCFF8-3084-426A-AA1D-F992780E7E29.jpeg
E99CCFF8-3084-426A-AA1D-F992780E7E29.jpeg
  424   Fri Jul 21 08:21:54 2023 TD, RSSBaking tent manual
09.20 Pirani 1.00e-3 mbar IE514 3.90e-8 mbar

      Today - install TCs, connect to Pico controller and test

10.00 Bakeout tent setup guide (draft) - attachment 1

14:30 Tested all TCs and are fine

05.25 Pirani 1.00e-3 mbar IE514 3.71e-8 mbar
Attachment 1: CARME-baking-guide_(1).pdf
CARME-baking-guide_(1).pdf CARME-baking-guide_(1).pdf CARME-baking-guide_(1).pdf CARME-baking-guide_(1).pdf CARME-baking-guide_(1).pdf CARME-baking-guide_(1).pdf
  423   Thu Jul 20 08:29:26 2023 TDInstalling tent. Two NEG power cable broken.
09.28 Pirani 1.00e-3 mbar IE514 5.45e-8 mbar

12.10 Bakeout tent frame installed
      SAES UHV1400 NEG pump power cables #1-6 installed
      See attachments 1 & 2

16.35 Attachments 3-6
      upstream baseplate
      3x sets of heaters beneath CARME chamber connected to CRYRING heater/TC manifold - attachments 7-8

      
16.45 Pirani 1.00e-3 mbar IE514 4.80e-8 mbar


17.10 SAES D-2000 #1 downstream, outside   NEG power + HV ( attachment 9 - OK cable connector )
      SAES D-2000 #3 downstream, inside   NEG power + HV
      SAES D-2000 #4 upstream, inside   HV
      SAES D-2000 #2 upstream, outside   HV

2x ( of 4x ) SAES D-2000 neg power cables were broken - forward-shell detached (unscrewed ) from back-shell remaining attached to D-2000 neg power connector. 
Re-assembly is straightforward but it is unclear how the 6x female sockets are aligned with the connector keys of the forward-shell (attachments 10-11 cf. attachment 9).
Until this is understood these 2x SAES D-2000 neg power cables should not be used.
Attachment 1: 20230720_120957.jpg
20230720_120957.jpg
Attachment 2: 20230720_121016.jpg
20230720_121016.jpg
Attachment 3: 20230720_162938.jpg
20230720_162938.jpg
Attachment 4: 20230720_163700.jpg
20230720_163700.jpg
Attachment 5: 20230720_163651.jpg
20230720_163651.jpg
Attachment 6: 20230720_163655.jpg
20230720_163655.jpg
Attachment 7: 20230721_093447.jpg
20230721_093447.jpg
Attachment 8: 20230721_093452.jpg
20230721_093452.jpg
Attachment 9: 20230721_092923.jpg
20230721_092923.jpg
Attachment 10: 20230721_094608.jpg
20230721_094608.jpg
Attachment 11: 20230721_094642.jpg
20230721_094642.jpg
  422   Wed Jul 19 08:17:26 2023 TDLeak testing
09.16 Pirani 1.00e-3 mbar IE514 8.77e-8 mbar
      Leak tester c. 5e-12 mbar.l/s (leak tester valved off)

09.48 Pirani 1.00e-3 mbar IE514 8.67e-8 mbar
      Leak tester c. 3.5e-12 mbar.l/s (leak tester valved off)

10.33 Pirani 1.00e-3 mbar IE514 8.55e-8 mbar
      Leak tester c. 4e-10 mbar.l/s (leak tester connected to CARME roughing line)

10.55 Leak tester remains at few e-10 mbar.l/s - we do not understand why
      With leak tester in differential mode the highest reading around the COF600 seal is c. 3.5e-12 mbar.l/s
      Following checks elsewhere on the chamber we observe c. 6e-12 mbar.l/s which presumably reflects accumulation of He around the chamber 

      Dispersing He with compressed air reduces diff mode reading to c. 2e-12 mbar.l/s

11.00 Pirani 9.80e-4 mbar IE514 8.47e-8 mbar

14.49 Pirani 1.00e-3 mbar IE514 7.71e-8 mbar
      Leak tester isolated from CARME backing line - CARME Edwards scroll pump now backing turbo.
  421   Tue Jul 18 08:16:28 2023 TDPumping down. Issues found with one detector.
09.15 Following overnight pumping of CARME chamber by Leybold turbo
      Pirani 1.00e-3 mbar IE514 2.83e-7 mbar

     The Leybold IE514 gauge currently in use is the CARME unit ( serial # 1054 ). The calibration data can be found at https://elog.ph.ed.ac.uk/CARME/56


09.44 Pirani 1.00e-3 mbar IE514 2.71e-7 mbar

10.10 Leak tester attached to backing line

      Valves to leak tester and turbo (connection to Edwards scroll pump in between) closed.

      Stop Edwards scroll pump with magnetic valve 'red' button - magnetic valves closes and Edwards scroll pump is switched off.

      Open valves to leak tester and then (slowly) turbo.

      Wait for leak tester to settle in e-12 mbar.l/s range. Leak tester pressure 3.6e-3 mbar. IE514 2.69e-7 mbar.

11.00 Pirani 1.00e-3 mbar IE514 2.42e-7 mbar
      Leak tester 1.8e-10 mbar.l/s

13.04 Pirani 1.00e-3 mbar IE514 2.06e-7 mbar

      Leak testing.
      Downstream DN100 flanges from CARME to dipole OK
      Lower NEG power connector feedhtrough (downstream chamber) OK
      Left hand side (looking downstream) bellows OK

      c. few e-9 mbar.l/s leak bottom left (lookimg downstream) of COF600 wire seal

      Additional torque to M16 bolts bottom left - c. 10 bolts -> 220Nm+

      c. 0.1-0.3e-11 mbar.l/s

      Further checks around circumference of COF600 seal reveal some small leaks c. e-11-12 mbar.l/s.? However background rate
      in leak tester too high for sensitive test - will await recovery to resume checking.

14.00 Pirani 9.8e-4 mbar IE514 1.96e-7 mbar
      Leak tester 4.9e-10 mbar.l/s


14.35 DSSSD test
      IE514 OFF

Ambient temperature 21.5 deg C

Top right
-10V -4uA
-20V 10uA
c. -30V > -20uA trip

Top left
-150V -5.15uA

Bottom left
-150V -4.97uA

Bottom right
-150V -6.33uA

Top right (repeat)
-20V -10uA
-25V -14uA

15.28 IE514 ON
      Pirani 9.8e-4 mbar IE514 1.87e-7 mbar

15.40 c. few e-9 mbar.l/s leak around magnetic valve - probably not an issue in terms of backing performance but may interfere with/confuse leak testing

16.05 Pirani 9.8e-4 mbar IE514 1.74e-7 mbar
      Leak tester 1.2e-9 mbar.l/s

Will leave CARME pumping with leak tester on roughing line and see whether leak rate recovers. If so, will continue leak tests tomorrow. 

18.50 Pirani 1.00e-3 mbar IE514 1.48e-7 mbar
      Leak tester < 1e-12 mbar.l/s (leak tester valved off)
  420   Mon Jul 17 08:45:11 2023 TDClosing CARME again
Grosmantage are scheduled to arrive 10.00 to inspect COF600 seals and their fit to the COF600 flange. They may decide to adjust the seals so that they fit the flange 
under 
tension.

Once Grosmontage are satisfied with the seal and its fit we can close the CARME chamber and start pumping down. 

Attachment 1 - photo of bakeout tent parts including those adapted/modified by Davide.


18.40 Grosmontage were re-tasked this morning.

      Davide, Omer, Cosmas checked alignment of COF600 male and female flanges.

      4x COF600 seals were checked - for some seals we observed wave over a ~ 10cm section of the circumference and/or surface scratches/roughness.
      Also observed that COF600 seals appeared to be too large - consequently the wire seal would be seated well ( i.e. in contact with the flange surface )
      for the majority of the circumference but would lift from flange surface for part(s) of the circumference.
      The fourth COF600 seal seated well over all of the circumference and we decided to proceed with closing the chamber.

      COF600 flange bolts tightened sequentially 5, 10, 20, 30, 40, 60, 80, 100, 120, 140, 160Nm.

      ML roughed CARME chamber slowly ( < 1mbar/s ) to protect Be window. Turbo started at 1.3mbar. Pirani gauge indicated e-3 mbar range when turbo at speed.


18.53 IE514 gauge ON pressure 4.05e-5mbar.

19.10 Pirani 1.42e-3 mbar IE514 3.02e-5 mbar
Attachment 1: 20230717_093106.jpg
20230717_093106.jpg
  419   Fri Jul 14 08:27:17 2023 TDCARME open - inspecting seal
09.20 CARME chamber at atmospheric pressure as expected.

      Will open CARME chamber and inspect COF600 wire seal for similar (or other) issues to the seal removed yesterday. One possibility is that the COF600 flange 
surfaces/structures are misaligned and part of seal is 'escaping' the seal zone. Given the flanges close easily the misalignment would have to be small.

13.30 Observations

      The bolts on the upstream feet of the far detector chamber are all loose and have probably been so for some time

      Based on measurements with calipers and the insertion of shims into the *radial* gap between COF600 flange surfaces the COF600 flanges may be misaligned and rotated 
transversely by a fraction of a millimeter. The challenge is that the chamber is not designed for fine transverse adjustments and it is not easy to measure the changes.

      The Pfeiffer recommended procedure for closing COF seals is to start tightening bolts *diametrically* and then to tighten sequentially

https://www.pfeiffer-vacuum.com/en/know-how/mechanical-components-in-vacuum/connections/detachable-flange-connections/cof-flanges/

     The plan

     Adjust position of far detector chamber and secure upstream feet bolts

     Close COF600 flange using Pfeiffer recommended procedure

     Today is Friday so it will probably be Monday before we can pump the CARME chamber

13.50 Another view

     Grosmontage also inspected the COF600 seal this afternoon. They said that it was important that the seal should be installed under tension to ensure that it remained 
in position as the flanges were closed and the bolts begin to be tightened. They think the seals are a little too large in circumference but that they can shorten one on 
Monday morning. 
  418   Thu Jul 13 08:46:44 2023 TDReplacing wire seal
09.30 CARME chamber status

      Edwards scroll pump ON
      Edwards Speedivac valves #1 & #2 OPEN
      VAT gate valves OPEN

      Re-connect Leybold Pirani gauge to local display/control
      CARME chamber pressure 2.4mbar

      Scroll pump should be able to achieve < 0.1mbar - indicates significant leak

      Davide will first check bolt torques for all flanges opened during most recent vacuum cycle

12.05 Additional torque did not resolve problems
      ML test COF600 seal with isopropanol and observed increased chamber pressure

      Will open COF600 seal after lunch

13.55 COF600 seal removed - one section marked by Kapton tape shows some distortion and a burr on the inner diameter - see attachments 1 & 2. 


      ML speculated that perhaps the seal was lightly too long and a section of circumference was forced outwards?
      Vacuum surfaces cleaned with isopropanol and tissues - red/brown discoloration probably indicating copper observed

14.30 New COF600 seal installed, flanges closed, bolts finger tight, bolts then systematically torqued in small increments 

      COF600 torques finger tight, 10, 20, 30, 40, 60, 80, 100, 120, 140, 160Nm x2
      DN100 flanges doenstream finger tight, 8, 16, 24Nm

16.40 Start scroll pump and slowly rough chamber

18.00 Roughing CARME chamber pressure flatlines c. 30mbar indicating a large leak.
      ML ran some isopropanol around COF600 flange and observed response in roughing pressure of c. 0.2-0.3mbar at c. 30mbar which appears to confirm COF600 is leaking.


      
Attachment 1: 20230713_135351.jpg
20230713_135351.jpg
Attachment 2: 20230713_135416.jpg
20230713_135416.jpg
  417   Thu Jul 13 00:12:45 2023 JM, CBCARME closed

The CARME chamber was closed and pushed back into position on the ring. 

Prior to moving back onto the ring, the detectors were bias tested and the distance between the detectors was also checked using the motors. Bias test was performed before the detector movement tests and before moving the chamber back onto the ring.  

Attached images shows the separation between the detectors when moved in togther. Separation is < 10 mm. Motors logs will be checked to place safety limits on the movement.

The chamber was pumped down. At 2 mbar the turbo was turned on, but the pressure would not decrease below 1e-2 mbar. The turbo was switched off and the pressure increased to 2 mbar. Isopropanol was sprayed on the wire seal causing an increase in the pressure. Leak on the wire seal. 

Attachment 1: Bias110723.png
Bias110723.png
Attachment 2: fullyin.jpg
fullyin.jpg
Attachment 3: fullyin-measured.jpg
fullyin-measured.jpg
  416   Thu Jul 13 00:12:19 2023 JM, CBBellow replaced

The left (inside ring) bellow for the motors rod was replaced.

Replacing the bellow involved;

  • One person holding the metal plate the detectors are mounted on (cabling remains plugged in) while the operation is performed. 
  • The motors box is supported by the cryring crane by its mounting points
  • Bolts on the flange connecting to the chamber can then be undone. The whole motors section (rod included) is then moved out of the chamber using the crane and someone to stabilise the rod as it slides off the detector plate. 
  • Once off the chamber, the steel cover can be removed by disconnecting from the flange and from the motors box. Best to perform on a table. 
  • The bellow is then removed by removing the nuts on the small flange (kf 25) connecting the bellow to the box. The bolts for these bolts are very long and go through the motors box from the moving metal plate seen in the attached image. 
  • Follow these steps in reverse to install the new bellow. 

While replacing the bellow, it was noticed the long bolts for the small flange were too short. It was therefore very difficult to re-attach the nuts with any torque. The thickness of the head of the bolts was shortend to effectively lenghten the bolt which was sufficient to tighten the nuts. New bolts should be bought in case of replacing the bellows again. 

 

Attachment 1: 20230705_164523.jpg
20230705_164523.jpg
  415   Thu Jul 6 15:26:39 2023 JMLeak update

The leak on the actuator arm bellows has been attempted to be sealed using UHV vacuum seal spray. The spray had the opposite to the desired effect and the leak rate increased. When spraying with helium the leak rate had increased to 1e-9 mbar l/s. After lunch a second application of the vacuum seal spray was attempted, however the leak was not sealed. The base leak rate had also increased to the 1e-10 level up from in the morning and yesterday. With the leak increasing in size and risk of a full failure, we are no longer able to move forward with the bakeout. 

The chamber has been vented and we will move forward with plans for replacing the bellow tomorrow. 

Attached plot shows the most current pumpdown.  

Attachment 1: pumping.png
pumping.png
  414   Wed Jul 5 19:27:13 2023 JMPump down and leak test

CARME and the interaction chamber have been pumping for ~26 hours, P=4.15e-7 mbar, attached figures show the current pump down plot in comparison to the two previous pumpdowns in the last 2 weeks and the pump down with the bare chamber in 2020. The gradient of the pressure reduction appears to be improving for each pump down, either due to conditioning of the chamber or the reduction in leaks.  

CARME and the interaction chamber were leaked tested again today. The base leak rate has reduced significantly from 1.5e-10 mbar l/s to <1e-12 mbar l/s. All flanges on the interaction chamber were tested. A leak is still visible on the gate valve, however the magnitude of the leak has decreased significantly from 2.2e-10 to 2e-11. No other leaks observed on the interaction chamber. 

All flanges on CARME checked again, no leaks observed on flanges or wire seal. Leak was observed on the bellows of the left hand side (inner ring) actuator arm. Leak is not on either the small flange box side or the large flange on the chamber, but on the bellows itself. Image attached. Leak rate is 2e-10 mbar l/s. The actuator was moved with the pneumatic motor once to observed any effects, no effect on leak rate or the chamber pressure.

 Replacing the bellows is very challenging and will likely require removing the detectors installed on the rod. Current plan is to continue with the baking and re-evaluate post bakeout. The bellows will be baked but should not experience the same temperatures as the flanges on the chamber. Future experiments do not require extremely forward angles so that the actuator arm can remain fixed in place to not cause a n increase in the leak, the leak rate can be compensated by the large neg pumping power. 

 

Attachment 1: pumping_1.png
pumping_1.png
Attachment 2: pumping-zoom.png
pumping-zoom.png
Attachment 3: signal-2023-07-05-15-13-30-512.jpg
signal-2023-07-05-15-13-30-512.jpg
  413   Tue Jul 4 17:08:08 2023 JMLeak test and venting

This morning the pressure achieved was 1.44E-7 mbar. Leak testing showed an overall leak rate of 1.5E-10 mbar l/s, this is below the previous leak rate but still significantly above the leak rate from elog 141 of the CARME chamber on its own. Spraying helium, a leak is visible on the VAT gate valve on the interaction chamber (inside of the ring). The leak rate increases to ~2.3E-10 mbar l/s. No increase in the leak rate was observed elsewhere on CARME or the interaction chamber. It was decided to vent again and replace the gasket and inspect the knife edges for the valve. 

The valve was removed and the gasket was replaced. Several scratches are visible on the gasket which may be the cause of the leak (or damage may have been done during dismounting). While vented, the backing line was re-organised. The scroll pump is now on the outside of the ring for easier access (see attached image), the turbo pump was rotated to move all power cables and the backing line to the outside of the ring. 

CARME and the interaction chamber were pumped down again and the ion gauge was activated. Vacuum can be followed via the vac plot or by grafana on the carme-gsi computer (anydesk required). 

Attachment 1: 20230704_135455.jpg
20230704_135455.jpg
Attachment 2: 20230704_174524.jpg
20230704_174524.jpg
  412   Mon Jun 26 14:49:05 2023 JM, TDCARME vented

26/06/23

This morning the pressure was 4e-7 mbar. The bolts on the windows and the large flange on the bottom of the interaction chamber were tightend. No leak was present on the windows following tightening. However, the leak on the rotatable flange, although decreased in magnitude, was still in the 1e-9 mbar l/s range when helium was sprayed. The torque on these bolts was 36 Nm, which is already past the 33 Nm recommendation for M8 bolts. It was decided to vent the section and examine the flange. 

Tomorrow morning the gasket will be replaced and we will begin pumping again in the afternoon. 

27/06/23

Flange was removed and examined. Several scratches on the surface of the flange including on the knife edge were visible. These scratches were sanded away by the grossmontage and the flange was remounted. CARME was pumped down in the afternoon. Pressure was in the 1.5E-3 mbar range when the scroll pump stalled causing the pressure to increase and the turbo to increase in current to keep up. Error symbol was observed on the scroll, unknown to why it stalled. VAT gate valve to the chamber was closed and all pumps turned off to be safe, P~ 1mbar. 

28/06/23

Jan started the scroll pump on  just the backing line using the gas balast to see if too much water was pumped. Scroll running fine so the gate valve to the chamber was opened and the turbo started. 

 

  411   Sun Jun 25 14:30:06 2023 JM, TDMotors bias test

The motors were moved using the pneumatic only, in and out for several cycles to gauge the effect on the vacuum. For the left arm no change in pressure was observed. For the right arm a small increase of 0.04e-7 mbar was observed which was quickly reduced by the turbo. 

The detectors were then biased to see if movement had any effect on the leakage current (large effects were observed in 2021 beamtime). Movement of the pneumatics saw no observable change in the leakage current above normal fluctuations. 

Both motors arms had been sucked into the chamber by 5-10mm due to the vacuum before starting the movement cycle. After one movement cycle, the motors returned to their normal fully out positions and remained fully out. 

  410   Sun Jun 25 10:08:30 2023 JM, TDPump-down

23/06/23

Pumpdown started with the scroll pump 17:15 . The turbo-pump was started 17:55. Gate valve was closed and the turbo turned of at 19:20 due to no working fan on the turbo. 

24/06/23

Ion gauge was activated at 11:56, P= 3.7e-5 mbar. Attached pump-down curves begin from when the ion gauge was activated.

The first plot is a comparison was a calculation for water outgassing in the chamber (which has many assumptions) but is a good guide on the expected 1/t decrease in pressure over time. The second plot is a comparison to previous turbo pump-downs in 2020 and 2021. In 2020 the chamber was bare with no instrumentation. In 2021, kapton and detectors were installed, but the pump down was conducted in several stages. Pressure readings are taken from the gauge reader in the cave, the vac plot viewer online appears to show a slightly different reading. 

 25/06/23

10:30 P= 7.11e-7 mbar. Opened valve on leak cart and saw a large increase in the pressure (above ion gauge level). Closed valve to leak cart, pressure recovered quickly (1.28e-6 mbar after ~ 30 mins). Pressure in the leak cart ~3e-3 mbar. Possibly small trapped volume between valves on the leak cart. Opened valve to leak cart again, no pressure increase observed in the chamber.  

11:40 Leak testing. Base leak rate ~5-6e-10 mbar.l/s.  Spraying helium on the seal between the bottom of the interaction chamber and the newly installed valve saw a large increase to  >1e-8 mbar.l/s. Closer attention suggests the leak may be from the rotatable flange on the bottom of the interaction chamber. An increased leak rate to 1e-9 mbar.l/s was also observed on the diamond window on the inside of the ring. Spraying around wire seal and feed-through flanges on chamber on the inside of the ring gave a leak rate of 6e-10 mbar.l/s. This may be a small leak, but is most likely noise on the leak detector or drifting helium to the large leak on the interaction chamber. Cannot determine prescence of small leaks while the large leaks are active. Closing valve to leak cart saw the leak level drop to <1e-12 mbar.l/s. 

Attached images show sites of the major leaks on the interaction chamber. 

Attachment 1: Figure_1.png
Figure_1.png
Attachment 2: 20230625_120640.jpg
20230625_120640.jpg
Attachment 3: 20230625_120648.jpg
20230625_120648.jpg
Attachment 4: pump-down.png
pump-down.png
  409   Sun Jun 25 09:41:06 2023 JM, TDNormal turbo-pump operation

Attached are images showing the messages for normal turbo-pump operation

Attachment 1: 20230623_191510.jpg
20230623_191510.jpg
Attachment 2: 20230623_191459.jpg
20230623_191459.jpg
Attachment 3: 20230623_191450.jpg
20230623_191450.jpg
Attachment 4: 20230623_191443.jpg
20230623_191443.jpg
Attachment 5: 20230623_191437.jpg
20230623_191437.jpg
Attachment 6: 20230623_191432.jpg
20230623_191432.jpg
  408   Sat Jun 24 10:38:42 2023 JM, TDSection 9 closed, begin pumping

23/06/23

The CARME chamber and interaction chamber were connected and section 9 has been fully closed. New valves have been installed on the interaction chamber, in addition to new diamond windows. A leak cart has been connected to the backing line of CARME. Before begining the pumpdown, the manual valve between the interaction chamber and the electron target was closed. 

The pumpdown was started at 17:15,  the valve on the scroll pump was very slowly opened to not damage the berylium window. By 18:00 a pressure of 1 mbar was achieved and the turbopump was activated. By 18:20 a pressure of 2E-3 mbar was achieved, the ion gauge was activated and briefly flashed 5E-4 mbar but this is over its limit so the gauge shutdown. At 19:20, 1.3E-3 mbar was achieved, the ion gauge was activated but the reading was 1.7E-4 mbar which is still over its limit. The turbo fan was not working and temperatures on the turbo were ~50C and increasing, so it was decided to close the main gate valve and turno off the turbo overnight in order to fix the fan.

A bias test of the detectors was also performed while in vacuum, T=20.5 C. All detectors show expected V-I curve and the leakage current is ~1-2 uA lower than the previous test in air, probably due to reduced light leaks. The top right detector did initially produce >10 uA at 20 V but the adaptor card was not properly seated on the p-n feed-through. Seating the adaptor card correctly yielded the expected V-I curve. Further inspection of the pins on this feed-through will be performed tomorrow.  

 

24/06/23

Overnight the pressure had increased to 0.8 mbar. Opening the gate valve decreased the pressure by an order of magnitude. The turbo fan was fixed and the turbo was re-activated, the pressure dropped to 1E-3 mbar level within 20 minutes.  Ion gauge was turned on after 30 minutes with a pressure of 3.7e-5 mbar. Pump down plot and comparison to previous pump downs will be uploaded in a seperate elog. Turbo pump maximum temperature reads 46C and has been consistent for several hours. 

V-I tests completed for the right hand side detectors. Leakage currents ~0.4 uA higher than yesterday which is consistent with the increase in temperature of ~ 0.5-0.7C. 

Attachment 1: 23-06-23-VI.png
23-06-23-VI.png
Attachment 2: 24-06-23-VI.png
24-06-23-VI.png
  407   Thu Jun 22 18:58:02 2023 JM, TDSlide back onto ring

The wire seal has been tightend further to 160 Nm. Open neg feed-through flange has also been fully tightend. 

The CARME chamber was then slid back onto the aluminium frame on the ring. Two carriages were damaged during the sliding process, which have now been replaced. Carriages were replaced by lifting one end of the upper frame section using the crane by only ~1mm. This was sufficient to slide out the broken carriage and replace it (see images to be uploaded)..  It appears there is some deformation in the rails on the temporary frame as CARME is slid across, which resulted in the damaged carriages -> further refinements are required for future slidings. The height of the chamber has been checked using the laser allignment.  

The turbo pump has also been mounted to the top of the chamber . Current plan is to close to the interaction chamber tomorrow and begin pumping.  

Attachment 1: 20230622_112534.jpg
20230622_112534.jpg
Attachment 2: 20230622_112827.jpg
20230622_112827.jpg
Attachment 3: 20230622_120307.jpg
20230622_120307.jpg
  406   Wed Jun 21 18:47:36 2023 JM, TDClosing CARME

Internal CARME installations complete. 

  • New strain reliefs (flange side) installed on each cable harness.
  • 2x sets of SAES UHV1400 NEG pumps installed and connected to 2x power feed-through flanges -> continuity checked OK -> 3x UHV1400 modules gives a resistance of 7.4 Ohms, 2x UHV1400 modules gives resistance of 4.8 Ohms. 
  • Pins B & D open circuit wrt ground ( DMM OL > 20M ? )
  • Pin C connection to ground OK 
  • Internal thermocouples connected. Middle thermocouple connected to strain reliefs, right thermocouple connected to a standoff on the detector, left thermocouple floating (see attached images). Resistance on thermocouples checked for middle and right thermocouples ~ 4 Ohms per thermocouple. 

V-I tests performed prior to closing CARME. All detectors show expected curve, but light leaks are present. Pressing my hand against the aluminium shield yielded 0.5-1.5 uA drop in leakage current. Could not turn off lights for the test due to other people working in the cave. Will repeat test when closed. Temperature in the cave ~ 21.5 C. 

Both chamber sections have now been closed. Wire seal appeared correctly positioned immediately prior to final close. Bolts closing the wire seal have been tightend to 100 Nm, plan to increase to 140 Nm tomorrow. DN100 feed-through flanges have been tightend with a torque of 24 Nm. 

Plan is to move chamber back onto the ring tomorrow and get ready for leak testing. 

Attachment 1: 20230621_104911.jpg
20230621_104911.jpg
Attachment 2: 20230621_104918.jpg
20230621_104918.jpg
Attachment 3: 20230621_110010.jpg
20230621_110010.jpg
Attachment 4: 21-06-23-VI.png
21-06-23-VI.png
  405   Fri May 26 09:08:06 2023 JM, CBDetector movement

Hopefully we will not be opening the chamber for quite some time so find attached images of all four installed detectors and a video of them both moving into the beam axis for future reference

Attachment 1: 20230525_135756.jpg
20230525_135756.jpg
Attachment 2: detector-movement-in.mp4
  404   Fri May 26 08:53:54 2023 JM, CBStrain reliefs (flange side) installed

Strain reliefs on the flange side of the cable harness have been installed for two detectors (top right and bottom right detectors). 

We require the strain relief parts from edinburgh in order to install strain reliefs for the left two detectors. 

Find attached images for the strain relief installed and a video of the clamped cables during a movement cycle.

A V-I test was performed for the two detectors which have all strain reliefs installed (see attachment 4).  V-I curve appears similar to the previous test (elog 398). Aluminium foil shield is now a bit battered so may have some small light leaks. Detectors were moved using the motors while bias was applied at 150V. Small change in leakage current observed (~0.01-0.02 uA)

Attachment 1: strain-relief-flange-side.jpg
strain-relief-flange-side.jpg
Attachment 2: strain-relief1_(1).mp4
Attachment 3: 20230526_140923.jpg
20230526_140923.jpg
Attachment 4: V-I-plot.png
V-I-plot.png
  403   Wed May 24 09:54:56 2023 JM, CBMaximum detector positions

The detectors were moved into the beam axis using the motors to check the maximum distance they can be moved in without collision with one another. 

Movement of both pneumatic motors results in a separation of ~ 4 cm between the detectors.

The servo motors for left and right were each moved in by 1 cm -> results in separation of 2 cm between the detectors

The servo motors for left and right were then each moved in by 5 mm -> results in separation of 1 cm between the detectors

The separation between the jacking plate of one of the detectors and the side of the adjacent detectors is closer than the separation of the detector silicons. The detectors could not be moved in any further than this to avoid collisions from the jacking plate

Attached are photos after movement of the detectors. Parallax may affect the distance on the ruler in the photos. 

During beam time -> Maximum distance that can be moved on the servo motors is 30 mm combined for left and right. (15 mm left and right or 10 mm left, 20mm right etc)  

 

Attachment 1: pneumatic-in.jpg
pneumatic-in.jpg
Attachment 2: pneumatic-in2.jpg
pneumatic-in2.jpg
Attachment 3: pneumatic-in-servo-10mm.jpg
pneumatic-in-servo-10mm.jpg
Attachment 4: pneumatic-in-servo-15mm.jpg
pneumatic-in-servo-15mm.jpg
Attachment 5: max-in.jpg
max-in.jpg
Attachment 6: max-in2.jpg
max-in2.jpg
Attachment 7: max-in-zoom.jpg
max-in-zoom.jpg
Attachment 8: max-in-jacking-plate.jpg
max-in-jacking-plate.jpg
Attachment 9: max-in-jacking-plate2.jpg
max-in-jacking-plate2.jpg
  402   Wed May 24 08:35:12 2023 JM, CBStrain reliefs (detector side) installed

The new MACOR strain reliefs on the (detector side) have been installed. See attached photos

The two pieces  of MACOR are not fully closed but we are confident that the kapton wires are sufficiently clamped.

In order for the strain reliefs to be installed in the correct poisition, two nuts on the jacking plate of the cable harness had to be removed. This is as the groove in the strain relief was not deep enough to accomadate the nut. 

Attached is also a video of the wires (not) moving with the strain reliefs mounted

Attachment 1: 20230524_092010.jpg
20230524_092010.jpg
Attachment 2: 20230524_092017.jpg
20230524_092017.jpg
Attachment 3: 20230524_092101.jpg
20230524_092101.jpg
Attachment 4: 20230524_093000.jpg
20230524_093000.jpg
Attachment 5: 20230524_092159.mp4
  401   Mon May 15 07:34:22 2023 TDMonday 15 May
08.35 DSSSD#3 bias -149.95V leakage current -6.418uA temperature +19.6 deg C ( https://web-docs.gsi.de/~lestinsk/tempplot.php )

DSSSD#3 bias OFF
  400   Sun May 14 11:55:12 2023 TDSunday 14 May
12.44 DSSSD#3 bias -149.95V leakage current -6.448uA temperature +19.7 deg C ( https://web-docs.gsi.de/~lestinsk/tempplot.php )

CRYRING cave lights ON - leakage current measurements @ 60s intervals

6.448
6.449
6.448
6.448
6.449
6.448
6.448
6.448
6.449
6.448

18.24 DSSSD#3 bias -149.95V leakage current -6.442uA temperature + 21.0 deg C ( https://web-docs.gsi.de/~lestinsk/tempplot.php )

CRYRING cave lights ON - leakage current measurements @ 60s intervals

6.441
6.441
6.441
6.441
6.441
6.442
6.442
6.442
6.440
6.441
  399   Sat May 13 11:34:25 2023 TD, RSSSaturday 13 May
12.34 DSSSD#3 bias -149.95V leakage current -6.763uA temperature +21.2 deg C ( https://web-docs.gsi.de/~lestinsk/tempplot.php )

CRYRING cave lights ON - leakage current measurements @ 60s intervals

6.763
6.764
6.764
6.765
6.767
6.767
6.767
6.767
6.768
6.769

12.48 CRYRING cave lights OFF

13.03 leakage current -6.771uA - no obvious change due to the CRYRING cave lighting

17.31 DSSSD#3 bias -149.95V leakage current -6.658uA temperature +21.0 deg C ( https://web-docs.gsi.de/~lestinsk/tempplot.php )

leakage current measurements @ 60s intervals

6.656
6.655
6.655
6.655
6.654
6.652
6.652
6.651
6.650
6.649

CRYRING cave temperature variations/cycles ~0.1 deg C at +21 deg C ambient temperature can produce leakage current variations ~60nA for a leakage current of ~6.6uA

td@winder:~/f77$ ./jgen
 *** Enter reference temperature (K)
294
 *** Enter minimum temperature, maximum temperature
 *** and temperature increment (K)
290 298 0.1

 Temperature (K)   Ratio (J(T)/J(294.0))

      290.0                0.702
      290.1                0.708
      290.2                0.714
      290.3                0.721
      290.4                0.727
      290.5                0.734
      290.6                0.740
      290.7                0.747
      290.8                0.754
      290.9                0.761
      291.0                0.767
      291.1                0.774
      291.2                0.781
      291.3                0.788
      291.4                0.795
      291.5                0.802
      291.6                0.809
      291.7                0.817
      291.8                0.824
      291.9                0.831
      292.0                0.839
      292.1                0.846
      292.2                0.854
      292.3                0.861
      292.4                0.869
      292.5                0.877
      292.6                0.884
      292.7                0.892
      292.8                0.900
      292.9                0.908
      293.0                0.916
      293.1                0.924
      293.2                0.932
      293.3                0.941
      293.4                0.949
      293.5                0.957
      293.6                0.966
      293.7                0.974
      293.8                0.983
      293.9                0.991
      294.0                1.000
      294.1                1.009
      294.2                1.018
      294.3                1.027
      294.4                1.036
      294.5                1.045
      294.6                1.054
      294.7                1.063
      294.8                1.072
      294.9                1.082
      295.0                1.091
      295.1                1.101
      295.2                1.110
      295.3                1.120
      295.4                1.130
      295.5                1.139
      295.6                1.149
      295.7                1.159
      295.8                1.169
      295.9                1.179
      296.0                1.190
      296.1                1.200
      296.2                1.210
      296.3                1.221
      296.4                1.231
      296.5                1.242
      296.6                1.253
      296.7                1.263
      296.8                1.274
      296.9                1.285
      297.0                1.296
      297.1                1.308
      297.2                1.319
      297.3                1.330
      297.4                1.342
      297.5                1.353
      297.6                1.365
      297.7                1.376
      297.8                1.388
      297.9                1.400

STOP *** Program ends
  398   Fri May 12 16:24:20 2023 RSS, TDV-I test

Since yesterday evening, the bottom right detector was at -150 V. In the morning, the current was quite stable ~ 6.77 uA. The bias was switched off, the foils opened, both of the detector arms were moved in and out three times, and the chamber was tightly covered again. A negative bias voltage was applied from 0 to 150 V in steps of 10 V one by one to all four detectors, and the V-I results can be seen in Attachments 1-4. All detectors seem to be fine. The variation of the current with time at a constant bias voltage of -150 V was checked for all detectors and the results are in Attachments 5-8. Also, with an applied bias voltage of -150 V to the bottom right and left detector, the detector arms were moved in and out three times which showed a variation of 1-2 nA and 1 nA in the leakage current. For further conditioning of the bottom right detector, it is again left with a negative bias voltage of 150 V over the weekend.

Attachment 1: detector_1.png
detector_1.png
Attachment 2: detector_2.png
detector_2.png
Attachment 3: detector_3.png
detector_3.png
Attachment 4: detector_4.png
detector_4.png
Attachment 5: detector_1_variation.png
detector_1_variation.png
Attachment 6: detector_2_variation.png
detector_2_variation.png
Attachment 7: detector_3_variation.png
detector_3_variation.png
Attachment 8: detector_4_variation.png
detector_4_variation.png
  397   Thu May 11 21:22:54 2023 RSS, TDV-I test

Today, the foil was opened, the pins on the bottom right detector were checked and looked fine, the right-hand arm was moved a bit in and then out, and the foil was tightly closed again. The negative bias voltage was applied from 0 to 50 V in steps of 10 V. During the measurement, the lights in the cave were switched off to ensure no leakage current from the light*. The results are shown in Attachment 1. To condition the detector, it was left at a negative bias voltage of 50 V for ~ 3 hours with lights on. No increase in current was seen (see Attachment 2), indicating the chamber was nicely covered with foil. Then the negative voltage was increased from 60 V to 150 V in steps of 10 V. No large current readings, like yesterday, were seen today (see Attachment 3). The detector was then left at -150 V for ~ 30 minutes with the current to be nearly stable with fluctuations in the range of 70-100 nA. The foil was then taken off to check if there are any dust particles on the detector surface. As can be seen in Attachment 6, some dust particle was found at the bias/multi-guard ring (MGR) boundary structures of the DSSSD. The area was then nicely cleaned with isopropanol + lab tissue wipe (see Attachment 7) and the chamber was covered with aluminum foil again. The negative bias voltage was applied from 0 to 150 V in steps of 10 V again and the results can be seen in attachment 5, indicating the response of the detector to be normal, unlike yesterday. Before calling the day off, the detector was left at -150 V overnight.

 

* Every time, before switching off the lights, it was checked that no one else was in the cave.

Attachment 1: test1.png
test1.png
Attachment 2: test2.png
test2.png
Attachment 3: test3.png
test3.png
Attachment 4: test4.png
test4.png
Attachment 5: test5.png
test5.png
Attachment 6: IMG_8800.jpg
IMG_8800.jpg
Attachment 7: IMG_8804.jpg
IMG_8804.jpg
  396   Wed May 10 19:06:18 2023 RSS, TDV-I test

To test the detectors, adaptors were installed on the feed through flanges. A negative bias voltage was applied in steps of 10 V from 0 to 150 V. The V-I plots are attached for the four detectors. Three detectors - top right, top left, and bottom left (in the direction of the beam) - are fine. For the bottom right detector (in the direction of the beam), the current was fluctuating and was considerably high (1st test). The adaptors connecting to this detector were removed and checked by applying the bias voltage which showed zero current for voltages ranging from 0 to - 150 V, confirming the adaptors to be fine. They were then exchanged by other adaptors and the V-I test was repeated again (2nd test) with fluctuating and high current values. One of the flanges connecting to this detector was found to be light-tight. These tests were done at an ambient temperature of 21 degrees. Also, we couldn't find the rest of the three SHV to 2x Lemo 00 cables for the detector HV.

Attachment 1: detector_1.png
detector_1.png
Attachment 2: detector_2.png
detector_2.png
Attachment 3: detector_3.png
detector_3.png
Attachment 4: detector_4.png
detector_4.png
  395   Tue May 2 15:02:09 2023 CB, JMAll detectors and harnesses mounted

Completed mounting harnesses and detectors.

The strain relief support for the bottom most two detectors on the feedthrough side was not mounted because we could not find enough MACOR parts and mounting flanges. We cannot find them in GSI, possibly they are somewhere not obvious in Edinburgh.

The strain relief on the detectors' side are still being manufactured.

 

We plan to test the detectors electrically from next week without moving the dolly. DR and OG to find a temporary mounting point for the manifold so that FEEs can be cooled in current configuration.

Attachment 1: 20230502_153410.jpg
20230502_153410.jpg
Attachment 2: 20230502_153503.jpg
20230502_153503.jpg
Attachment 3: 20230502_153509.jpg
20230502_153509.jpg
Attachment 4: 20230502_153511.jpg
20230502_153511.jpg
  394   Mon May 1 18:21:18 2023 CB, JMAll detectors mounted

Mounted four detectors on Carme ad well as four cable harnesses.

 

Cables for new, shorter harnesses appear still too long. Likely the model is not correct somehow.

 

One feedthrough flange was found to be damaged by an adaptor card. Tried to fix in vain. One pin is missing. Replaced with another.

Attachment 1: 20230501_191533.jpg
20230501_191533.jpg
Attachment 2: 20230501_191511.jpg
20230501_191511.jpg
Attachment 3: 20230501_145740.jpg
20230501_145740.jpg
  393   Mon May 1 08:26:09 2023 CB, JMLeft hand detectors remounted

(31 April)

Remounted 3335-12 on top and new 3335-1 on bottom of left hand plate. Used new dog legs and old standoffs.

 

Dismounted all harnesses and feedthroughs from CARME. Dismounted all blind flanges on right hand side. Also had to dismount bottom SAES feedthrough due to a bolt falling inside.

Remounted feedthough. Remounted left hand detector plate.

 

Prepared new 3335-11 and 3335-13 on top and bottom of right hand plate. Planning to mount tomorrow.

Also planning to start jacking in detectors to new harnesses.

 

Not possible to test any detector on the bench due to lack of appropriate multimeter probe. Will have to test with harnesses.

  392   Sun Apr 30 10:01:39 2023 CB, JMLeft hand detectors unmounted

(29 April)

Inspected CARME after opening. See attached. Some key takeaways.

- No visible damage to the detectors or the Kapton cables. No visible deposits or dust on the detectors either. 

- No damage even to the fragile thermocouple wires

- The strain relief on the top and bottom did nothing. Those on the sides may have helped.

- The split funnels on top fell and fortunately missed the detectors. We should not keep them. Unclear if they feel when moving CARME or during the beamtime but either is too dangerous. We need another solution or we'll have to do without.

- Obvious traces of white dust particles over the chamber. Most likely MACOR.

 

Jacked out harness for bottom left detector that was not accepting bias. MACOR shavings fell off while jacking out. MACOR connector on side with no pins is clearly damaged. See attached. Should not affect performance.

Tested bias pins - they work. No visible damage to the detector except one bias bond wire has disconnected. Possibly one more nearby. TD thinks this is not enough to explain observed behaviour.

 

Mounting a DSSD on the plate while it hangs in Carme appears possible but very risky. Dismounted whole left support plate, and right support plate. 

 

Unless a clear fault can be identified for either the bottom left detector or its harnesses, both the detector and the harnesses will be replaced. Plan to remount top left detector with the same harnesses.

Plan to mount two more detectors on the right hand plate.

Attachment 1: 20230428_181020.jpg
20230428_181020.jpg
Attachment 2: 20230428_181123.jpg
20230428_181123.jpg
Attachment 3: 20230428_181126.jpg
20230428_181126.jpg
Attachment 4: 20230428_181129.jpg
20230428_181129.jpg
Attachment 5: 20230428_181131.jpg
20230428_181131.jpg
Attachment 6: 20230428_181133.jpg
20230428_181133.jpg
Attachment 7: 20230428_181139.jpg
20230428_181139.jpg
Attachment 8: 20230428_181141.jpg
20230428_181141.jpg
Attachment 9: 20230429_150436.jpg
20230429_150436.jpg
Attachment 10: 20230429_150717.jpg
20230429_150717.jpg
Attachment 11: 20230429_151421.jpg
20230429_151421.jpg
Attachment 12: 20230429_151425.jpg
20230429_151425.jpg
Attachment 13: 20230429_151428.jpg
20230429_151428.jpg
Attachment 14: 20230429_151431.jpg
20230429_151431.jpg
Attachment 15: 20230429_151432.jpg
20230429_151432.jpg
Attachment 16: 20230429_151435.jpg
20230429_151435.jpg
Attachment 17: 20230429_151436.jpg
20230429_151436.jpg
Attachment 18: 20230429_151444.jpg
20230429_151444.jpg
Attachment 19: 20230429_151449.jpg
20230429_151449.jpg
Attachment 20: 20230429_151453.jpg
20230429_151453.jpg
Attachment 21: 20230429_152026.jpg
20230429_152026.jpg
Attachment 22: 20230429_152045.jpg
20230429_152045.jpg
Attachment 23: 20230429_160614.jpg
20230429_160614.jpg
Attachment 24: 20230429_162251.jpg
20230429_162251.jpg
Attachment 25: 20230429_162257.jpg
20230429_162257.jpg
Attachment 26: 20230429_172209.jpg
20230429_172209.jpg
  391   Sat Apr 29 10:12:58 2023 CB, JMCARME opened

(28 April)

Mounted rails on dolly. The rails don't smoothly connect to those on the Aluminium frame and some thin metal shims were required to get the correct height. A metal support had to be placed above the Delrin pads to support the rails above the gap between frames.

 

All side feet were mounted but extensive modifications were required to slotted feet due to them not being manufactured to drawing, and to Carme middle section flange width being incorrect (see previous entry).

 

Removed brakes and started sliding CARME out. The dolly and the aluminium frame came partly separated as the weight of the side feet was shifted on the latter. Strapped them together. No damage.

 

Kept sliding. Sliding when the carriages passed from one rail to the other was quite hard. The back carriage on the left looking downstream lost a single ball bearing when crossing over. A more sophisticated solution is probably required.

 

Once Carme was on the dolly, opened it using M8 jacking bolts and lifted away front and middle section. 

 

No obvious damage inside but strain relief did not work as intended. Photos to be uploaded after more detailed inspection on Saturday.

 

Plan to install right hand side detectors over the weekend. Unclear if bottom left detector is damaged and needs to be replaced.

Attachment 1: 20230428_094931.jpg
20230428_094931.jpg
Attachment 2: 20230428_155649.jpg
20230428_155649.jpg
Attachment 3: 20230428_164926.jpg
20230428_164926.jpg
Attachment 4: 20230428_155748.jpg
20230428_155748.jpg
Attachment 5: 20230428_165448.jpg
20230428_165448.jpg
Attachment 6: 20230428_174021.jpg
20230428_174021.jpg
Attachment 7: 20230428_181020.jpg
20230428_181020.jpg
Attachment 8: 20230428_181919.jpg
20230428_181919.jpg
  390   Thu Apr 27 19:43:23 2023 CB, JMPreparing to open

Removed turbo pump and set aside in a box. Considering switching to the other turbo to avoid degradation due to lack of use.

 

Installed bellows support and OG devised support to avoid bellows sagging when Carme is split open.

 

Attempted to install new CARME feet, but there were issues. Feet with slots produced do not match design drawings by PB. Slot is in the wrong side. Furthermore Carme interaction chamber flange thickness is 36 mm vs. 36.5 mm in drawing. DR can modify feet to make them fit tomorrow.

 

Now just waiting for rails to open CARME. Will arrive tomorrow or on Tuesday, after weekend and Mayday.

 

Attachment 1: 20230427_123456.jpg
20230427_123456.jpg
Attachment 2: 20230427_161602.jpg
20230427_161602.jpg
  389   Wed Apr 26 16:49:28 2023 CB, JMCARME dolly mostly built

Disconnected cables from CARME motors and turbo. The turbo has to go to allow Carme to slide out. Crane may be required to lift because it's in a very hard spot to access.

 

Dolly mostly built. See attached. Left side doesn't quite fit right by 1-2 mm. Might be okay anyway but will put some shims to avoid any shaking during sliding.

 

Waiting for the rails to arrive due to GSI procurement issues.

 

Plan to mount new feet and bellow support tomorrow.

Also plan to split CARME at the smaller flange towards the interaction chamber, after the bellows and before the bellows towards the magnet on the downstream side.

Attachment 1: 20230426_152121.jpg
20230426_152121.jpg
Attachment 2: 20230426_162619.jpg
20230426_162619.jpg
Attachment 3: 20230426_162627.jpg
20230426_162627.jpg
  388   Tue Apr 25 13:22:19 2023 JMCARME venting

CARME venting using nitrogen gas bottle. 

Coolant pipes dismounted from the chamber

Feed-through flange for n+n bias flange for bottom detector has a bent pin image attached. Clearer image of bent pin for ERNI connector used for aida03 also attached. All other feed-throughs and adaptor cards appear undamaged. 

Attachment 1: 20230425_134119.jpg
20230425_134119.jpg
Attachment 2: 20230425_132944.jpg
20230425_132944.jpg
  387   Mon Apr 24 21:24:26 2023 JMPreparation for CARME dismount

CARME currently in vacuum, pressure ~3E-10 mbar. Chamber to be vented tomorrow afternoon by vacuum division. We have active both pirani and hot cathode gauge to aid. 

Equipment removed from the chamber in preparation for dismount of CARME from the ring this week. Equipment removed includes:

  • Neg pump cabling
  • Heating wire
  • FEE electronic units and associated cabling (power, hdmi etc)
  • Stainless steel FEE holders
  • Red and blue cooling hoses

Attached image shows chamber following equipment removal. Adaptor cards still attached to feed-through flanges. ERNI connnectors examined, all fine except for FEE aida03 which has a bent pin (see attached image). Will dismount adaptor cards to examine further tomorrow. Feed-through flanges will also be examined to check for bent pins following adaptor card removal. 

New CARME frame is part constructed and is in the CRYRING area. Requires rails and castors which should arrive this week. The height of the frame will require shortening, to be level with the current frame. This will be looked into further tomorrow. 

Attachment 1: 20230424_155635.jpg
20230424_155635.jpg
Attachment 2: 20230424_155607.jpg
20230424_155607.jpg
Attachment 3: 20230424_150526.jpg
20230424_150526.jpg
  386   Wed Mar 22 16:33:32 2023 TDNIM WR source
Test of NIM WR source at Edinburgh

Attachment 1 - NIM WR source

NIM module upper SMA - DSO ch 1
NIM module lower SMA - DSO ch 2

Attachment 2

NO SYNC or CLK signals observed.

Noise observed with NIM bin power ON *and* OFF

Attachment 3

NIM WR source in EG&G Ortec NIM bin



... and with the Digilent Zybo board switched ON!


Attacments 4-7
Trigger ch 2
4, 40, 400 & 4000ns/div

Attachment 8
Trigger ch 1
4ns/div
Attachment 1: 20230322_155735.jpg
20230322_155735.jpg
Attachment 2: 20230322_160448.jpg
20230322_160448.jpg
Attachment 3: 20230322_160453.jpg
20230322_160453.jpg
Attachment 4: 20230322_181015.jpg
20230322_181015.jpg
Attachment 5: 20230322_181027.jpg
20230322_181027.jpg
Attachment 6: 20230322_181045.jpg
20230322_181045.jpg
Attachment 7: 20230322_181054.jpg
20230322_181054.jpg
Attachment 8: 20230322_180945.jpg
20230322_180945.jpg
ELOG V3.1.3-7933898