701
|
Tue Mar 11 11:53:18 2025 |
TD | Monday 10 March |
18.30 AIDA powered up for brief test
|
11x |
700
|
Sat Feb 22 12:35:38 2025 |
TD | Offline analysis data files R9_85-R9_199 (84Mo setting) |
Start 00:43 22.2.25 https://elog.gsi.de/despec/G-24-00302/57
AIDA data file R9_85
|
22x |
699
|
Thu Feb 20 17:06:42 2025 |
JB, CC, TD,MP | WR error for aida 13,14,15,16 | WR timestamp errors resolved after reseating
the HDMI cables to the MACB |
|
698
|
Wed Feb 12 10:46:01 2025 |
TD | Offline analysis data files S181 R4_351-396 | Analysis data files R4_351-396
|
25x |
697
|
Fri Jan 31 17:26:47 2025 |
CC, TD, MP | Friday 31 January contd. |
18.18 bPlas and BB7 installs complete
Restart AIDA DAQ, Merger, Tape Server |
16x |
696
|
Fri Jan 31 13:48:29 2025 |
JB | AIDA analysis from HI-DESPEC meeting 20.11.2024 | AIDA analysis and presentation by J. Bormans
for AIDA made in the HISPEC-DESPEC collaboration
meeting. |
|
695
|
Fri Jan 31 09:37:47 2025 |
CC, TD, MP | Friday 31 January | Implantation stack mounted - BB18(DS)-1000
+ bPlas + bPlas + 3x BB7(DS)-1000
|
26x |
694
|
Wed Jan 15 11:58:08 2025 |
TD | Offline analysis data files S505 R3_9 - R3_128 (207Hg setting) |
Aanlysis data files R3_9 - R3_128 (207Hg
setting)
|
27x |
693
|
Mon Jan 13 17:38:18 2025 |
TD | Offline analysis of S505 data files |
First timestamp of R2_105 0x16FAEC5E54AABF6C
|
10x |
692
|
Mon Jan 13 11:32:30 2025 |
TD | Offline analysis S505 data file R3_150 | Offline analysis of S505 data files R3_150
first WR ts
|
22x |
691
|
Sat Jan 11 14:14:21 2025 |
TD | aida06 crashed and rebooted 10.1.25 | aida06 crashed and rebooted sometime ( 14:48
- 23:25 ) yesterday. System console log appended.
|
|
690
|
Sat Jan 11 14:05:03 2025 |
TD | Saturday 11 January 2024 |
14.37 Detector bias & leakage current OK
- attachment 1
|
17x |
689
|
Fri Jan 10 10:03:43 2025 |
TD | LEC fast comparator threshold too low | > > >
> > > LEC fast comparator has been set to
0x2 (200keV) which is *very* low and means |
|
688
|
Thu Jan 9 17:56:14 2025 |
TD | LEC fast comparator threshold too low | > >
> > LEC fast comparator has been set to 0x2
(200keV) which is *very* low and means the |
|
687
|
Thu Jan 9 17:36:50 2025 |
TD | LEC fast comparator threshold too low | >
> LEC fast comparator has been set to 0x2
(200keV) which is *very* low and means the |
|
686
|
Thu Jan 9 16:25:51 2025 |
TD | LEC fast comparator threshold too low |
LEC fast comparator has been set to 0x2 (200keV)
which is *very* low and means the fast comparator |
|
685
|
Thu Jan 9 15:43:17 2025 |
JB, CC, MP | Timing test platform moved in | 16:30 We moved the platform in. The system
was still biased and the DAQ was still running.
The system seems to be okay, temperatures |
  |
684
|
Thu Jan 9 12:46:43 2025 |
JB, MP, CC | AIDA timing test |
|
   |
683
|
Wed Jan 8 16:47:05 2025 |
JB, MP, CC | AIDA timing test |
|
|
682
|
Wed Jan 8 10:46:50 2025 |
TD | Report - aida06 -boot fail due to network issue | 8:01:25/11:42:48|0x000000d00000-0x000000fe0000
: "user_kernel"
08:01:25/11:42:48|0x000000fe0000-0x000001000000 |
|
681
|
Wed Jan 8 10:08:29 2025 |
JB, GB, SD, MP, CC, JG | AIDA noise test with platform in position | Yesterday we spent time essentially performing
a dry run to get AIDA DAQ and bPlast into
the time sorter, there were some issues with |
   |
680
|
Mon Dec 30 15:43:42 2024 |
TD | [How To] Restart AnyDesk |
See also https://elog.ph.ed.ac.uk/CARME/489
|
|
679
|
Mon Dec 23 10:54:46 2024 |
TD | Monday 23 December 2024 |
11.54 re-established remote access via AnyDesk
- details to follow
|
|
678
|
Tue Dec 17 12:45:36 2024 |
JB, MP, CC | AIDA timing test |
|
  |
677
|
Mon Dec 16 13:07:04 2024 |
JB, MP, CC | AIDA timing test | https://elog.gsi.de/despec/Implantation+Stack/9?suppress=1
- Day 1 ELOG
Day2: |
16x |
676
|
Sat Dec 14 11:52:42 2024 |
TD | Saturday 14 December |
11:00 Visual inspection FEE64 adaptor PCBs
& cabling
|
27x |
675
|
Wed Dec 11 14:46:21 2024 |
TD, JB, CC, MP | Mounting of AIDA, bPlast and BB7 for test | Snout assembled for detector position and
timing test.
The planned test is to use a BGO |
   |
674
|
Mon Dec 9 12:28:01 2024 |
NH, CC, MP | Test of BB7 on AIDA Electronics | BB7 connected to aida07, sitting in a cardboard
box
|
|
673
|
Thu Dec 5 11:18:46 2024 |
TD | Offline analysis S100 data files R21_0-R21_99 | Offline analysis of S100 data files R21_0
- R21_99 (162Eu setting)
|
23x |
672
|
Wed Dec 4 09:57:24 2024 |
TD | HISPEC DESPEC meeting presentation - November 2024 -0 Summary of AIDA performance 2024 | |
|
671
|
Tue Nov 19 11:24:25 2024 |
TD | Anydesk restarted remotely |
Anydesk restarted remotely per https://elog.ph.ed.ac.uk/CARME/489 |
|
670
|
Tue Nov 5 15:19:35 2024 |
TD | S505 offline analysis R5_780 - R5_814 | Offline analysis of data files S505/R5_780
- R5_814 (corresponding to MBS data files
73-74)
|
24x |
669
|
Wed Aug 14 12:40:11 2024 |
JB | Repaired DSSSD delivery 14.09.2024 | Three BB18-1000 triples AIDAs collected
on 14.09.2024
Find attached visual of the wafer |
11x |
668
|
Sun Jul 14 16:52:28 2024 |
TD | S181 R7_38-44 | |
7x |
667
|
Sun Jul 14 14:37:53 2024 |
TD | S181 R6_375-380 | |
6x |
666
|
Mon Jun 17 02:13:21 2024 |
TD | Monday 17 June | 03.11 DSSSD bias & leakage current OK - attachment
1
|
43x |
665
|
Sun Jun 16 03:11:27 2024 |
TD | Sunday 16 June | 04.08 DSSSD bias & leakage current OK - attachment
1
|
26x |
664
|
Sat Jun 15 00:28:01 2024 |
TD | Saturday 15 June | Beam lost due to septum failure c. 18.15 Friday
14 June
|
45x |
663
|
Fri Jun 14 15:00:47 2024 |
Norah , Muneerah | 16:00-20:00 Friday 14 June | Everything appears to be running
smoothly. Attachments 1-4 were taken around
16:00 |
13x |
662
|
Fri Jun 14 06:22:39 2024 |
TD | 08:00-16:00 Friday 14 June |
07.23 All histograms & stats zero'd
|
23x |
661
|
Thu Jun 13 23:04:01 2024 |
Dan Judson, TD | 0.00-8.00 14/6/24 | 0.00 checks - all looks ok
Voltage/currents - attachment 1
Rates attachment 2 + 3 |
40x |
660
|
Thu Jun 13 21:06:44 2024 |
Robert Page | 16:00 to 00:00 shift on 13/6/2024 | Elog went down around the end of the last
shift, so final screenshots were not uploaded
at that point. |
26x |
659
|
Thu Jun 13 06:49:54 2024 |
Magda Satrazani | 08:00 - 16:00 Thursday 13 June 2024 | Took over the shift from Marc.
08:00 FULL CHECK |
40x |
658
|
Wed Jun 12 23:00:16 2024 |
Marc | 00:00-08:00 Thursday 13 June | Starting a new night shift - All good
1am full check:
DSSSD bias & leakage current |
23x |
657
|
Wed Jun 12 15:41:07 2024 |
Betool Alayed | 16:00-00:00 Wed 12 Jun 24 | 5pm full checks:
DSSSD bias & leakage current
ok - attachment # 1-2 |
27x |
656
|
Wed Jun 12 11:06:55 2024 |
PP | Mid-shift checks, 12:00 | All seems normal.
Screenshots attached. |
8x |
655
|
Wed Jun 12 06:17:23 2024 |
TD | 08:00-16:00 Wednesday 12 June | 07:10 DSSSD bias & leakage current OK - attachment
1
|
24x |
654
|
Tue Jun 11 23:13:44 2024 |
TD | 00:00-08:00 Wednesday 11 June | 00.08 DSSSD bias & leakage current OK - attachment
1
|
18x |
653
|
Tue Jun 11 20:20:36 2024 |
TD | ADC data rates from 09.00 11.6.24 | Per NH
"From ELOG (assuming times accurateish)
|
|
652
|
Tue Jun 11 18:45:37 2024 |
PP | Mid-shift checks, 19:45 | All looks good.
Sceenshots attached. |
8x |
651
|
Tue Jun 11 15:01:07 2024 |
PP TD | 16:00-00:00 shift Tuesday 11/06/24 | All seems OK
Screenshots attached.
|
13x |
650
|
Tue Jun 11 10:03:48 2024 |
TD, NH | 08:00-16:00 Tuesday 10 June | 10.57 DSSSD bias & leakage current OK - attachment
1
|
23x |
649
|
Mon Jun 10 23:36:14 2024 |
TD | 0000-08:00 Tuesday 11 June |
00.09 DSSSD bias & leakage current OK - attachment
1
|
20x |
648
|
Mon Jun 10 15:07:24 2024 |
Marc | 16:00-00:00 shift Monday 10/06/24 | Evening shift
5pm full checks:
DSSSD
bias & leakage current ok - attachment |
31x |
647
|
Mon Jun 10 07:15:48 2024 |
Dan | 08.00-16.00 Monday 10th June | 8am checks
temps ok - attachment 2
rates ok - attachment 8 |
50x |
646
|
Sun Jun 9 22:49:56 2024 |
TD | 23:00-07:00 Monday 9 June |
23.47 DSSSD bias & leakage current OK - attachments
1-2
|
27x |
645
|
Sun Jun 9 16:38:30 2024 |
Norah , Muneerah, JB | 16:0-00:00 9 June 2024 | AIDA02 and AIDA06 gave zero attachment
1. After connecting with Tom to fix it, now
it works. |
   |
644
|
Sun Jun 9 13:32:48 2024 |
TD | [How To] Recover DAQ when one FEE64 stops sending data - short(er) version |
1) Select 'No Storage' [workspace 3, display
1, browser tab 'TapeServer']
|
|
643
|
Sat Jun 8 23:00:52 2024 |
TD | Sunday 9 June |
23.58 8.6.24
|
16x |
642
|
Sat Jun 8 13:22:36 2024 |
TD | Saturday 8 June |
14.09 aida02 WR stall
DAQ STOP
|
24x |
641
|
Fri Jun 7 16:08:37 2024 |
TD | Friday 7 June | 05.01 DSSSD bias & leakage current - attachment
1
|
19x |
640
|
Thu Jun 6 11:13:29 2024 |
TD | Thursday 6 June |
12.09 DSSSD bias & leakage current - attachment
1
|
9x |
639
|
Wed Jun 5 08:34:47 2024 |
JB. AM, MP, MY | Detector bias tests | We are trying to bias each system on its
own and in sequence biasing each other detector
system to see the interdependence. |
14x |
638
|
Tue Jun 4 15:40:32 2024 |
NH | Tues 4 Jun | DEGAS Array over Snout again
Situation similar to yesterday, most p+n |
 |
637
|
Mon Jun 3 14:40:34 2024 |
JB, NH | 03 June 2024 | 15:40 reseated FEE64s adaptors of aida01,
aida09, aida13 & aida16.
TEMP OK. HV OK. System wide checks |
10x |
636
|
Mon Jun 3 07:35:56 2024 |
TD | Monday 3 June | 08.28 DSSSD bias & leakage current - attachments
1-2
|
8x |
635
|
Sun Jun 2 01:45:05 2024 |
TD | Sunday 2 June | 02.41 DSSSD bias & leakage current - attachment
1
|
25x |
634
|
Sat Jun 1 12:48:54 2024 |
TD | Saturday 1 June |
13.26 DSSSD bias & leakage current - attachments
1-3
|
31x |
633
|
Mon May 27 17:20:49 2024 |
TD | Monday 27 May | 17.50 bPlas cabling, grounds etc connected,
power OFF
CC reports that bPlas power is actually |
9x |
632
|
Wed May 22 20:10:19 2024 |
NH, CC, HMA, MV, ZC, KZ | AIDA Snout Remounted | AIDA and the Snout have been remounted in
S4
|
9x |
631
|
Thu May 9 09:08:26 2024 |
TD | Current NewMerger install - aida-gsi | [npg@aidas-gsi MIDAS]$ ls -l
total 524
drwxrwxr-x. 2 npg npg 156 Mar 17 2021 |
|
630
|
Tue May 7 13:25:00 2024 |
JB, CC, NH, HMA | Tuesday 6 May - dismount snout | Status before dismount
14:26 DSSSD#1 119.99 V @ 17.968
uA |
 |
629
|
Mon May 6 07:10:26 2024 |
TD | Monday 6 May |
08.11 DSSSD#1 bias -120V leakage current
-17.0uA
|
|
628
|
Sun May 5 10:22:24 2024 |
TD | Sunday 5 May |
11.22 DSSSD#1 bias -120V leakage current
-17.1uA
|
|
627
|
Sat May 4 07:36:07 2024 |
TD | Saturday 4 May |
08.32 DSSSD bias & leakage currents - Grafana
- attachments 1-2
|
11x |
626
|
Thu May 2 23:08:43 2024 |
TD | Friday 3 May | 00.05 FEE64 temperatures OK
ADC data item stats OK
|
|
625
|
Thu May 2 00:16:22 2024 |
TD, NH | Thursday 2 May | 01.15 FEE64 temperatures OK
ADC data item stats
aida02 & aida04 no data
|
  |
624
|
Wed May 1 08:59:15 2024 |
TD | Wednesday 1 May | 04.55 FEE64 temperatures OK
ADC data item stats OK *except* aida04
no data
|
|
623
|
Tue Apr 30 09:06:45 2024 |
TD, NH | Tuesday 30 April |
10.00 DSSSD bias & leakasge current - Grafana
- attachments 1-2
|
7x |
622
|
Mon Apr 29 08:11:16 2024 |
TD | Monday 29 April |
08.52 Beam off
|
41x |
621
|
Sun Apr 28 22:52:06 2024 |
Marc | 0:00-08:00 Monday 29 April | Just another manic magic Monday,
....
Taking over Betool's AIDA shift. |
7x |
620
|
Sun Apr 28 20:35:57 2024 |
Betool Alayed | 16:0-00:00 28 April 2024 | 21:30 screenshots
and
23:24 screenshots |
10x |
619
|
Sun Apr 28 16:26:47 2024 |
Betool Alayed | 16:00-00:00 Sunday 28 April | 19:24 pm screenshots
and
17:18 pm screenshots |
10x |
618
|
Sun Apr 28 16:07:59 2024 |
TD | [How To] Recover DAQ when one FEE64 stops sending data - short(er) version | 0) Please infrom the 'on shift' crew that
you need to restart the AIDA DAQ
|
|
617
|
Sun Apr 28 15:49:58 2024 |
TD | DAQ restart short(er) cut |
16.45 aida04 stopped producing data
|
|
616
|
Sun Apr 28 15:09:34 2024 |
Betool Alayed | 16:0-00:00 28 April 2024 | 16:00 aida04 not producing data
power cycle all FEE64s to recover DAQ |
|
615
|
Sun Apr 28 06:57:59 2024 |
Magda Satrazani | 08:00-16:00 Sunday 28 April | Took over the shift from Marc.
|
36x |
614
|
Sat Apr 27 23:04:42 2024 |
Marc | 00:00-08:00 Sunday 28 April | Encore un matin, un matin pour ..... a
shift !!!ramping up again ! S4 temperature
not really going down either: 25C |
12x |
613
|
Sat Apr 27 15:13:06 2024 |
Norah | 16:00 - 00:00 Saturday 27 April | 14:50 Checks
Leakage current is 20.021
and HV status is Max Voltages ! Is it normal |
15x |
612
|
Sat Apr 27 04:59:31 2024 |
TD | 08:00-16:00 Saturday 27 April |
06.00 DAQ continues file S100_alpha/R20_740
FRS setting 162Eu
|
37x |
611
|
Fri Apr 26 23:04:51 2024 |
Marc | 00:00-08:00 Saturday 27 April | It's a new day, it's a new shift
and I'm feeeeeliiiiing goooood !!!!!
0.15 Checks:
DSSSD bias |
12x |
610
|
Fri Apr 26 15:34:00 2024 |
RDP | 16:00 - 00:00 Friday 26 April | This appeared in the system log:
Apr 26 16:30:51 aidas-gsi smartd[1076]:
Device: /dev/sda [SAT], 1 Currently unreadable |
6x |
609
|
Fri Apr 26 08:15:59 2024 |
TD | [How To] Recover DAQ when one FEE64 stops sending data |
Occasionally, one (sometimes more) FEE64(s)
will stop producing data whilst the DAQ is |
|
608
|
Fri Apr 26 07:01:26 2024 |
Norah | 08:00 - 16:00 Friday 26 April | 10:00 Checks:
All seems smooth.
DSSD bias and leakage current |
    |
607
|
Thu Apr 25 22:45:30 2024 |
Tom, Muneerah, Norah | Over night shift AIDA 26 Fri |
01:02
Aida04 was zero not reading had |
9x |
606
|
Thu Apr 25 15:02:50 2024 |
Marc | 16:00 - 0:00 Thursday 25 April |
16:20:
Spill length:4s (was 3.5s
before) - Extraction time 1.5s (see attachement
#1)
|
8x |
605
|
Thu Apr 25 08:26:50 2024 |
TD | 08:00-16:00 Thursday 25 April | 09.56 DAQ continues data file S100_alpha/R18_387
FRS switching settings
|
42x |
604
|
Wed Apr 24 21:39:51 2024 |
TD | 00:00-08:00 Thursday 25 April | 22.43 24.4.24
DSSSD bias & leakage current OK - attachment |
12x |
603
|
Wed Apr 24 18:44:05 2024 |
PP | 19:30 checks | All seems smooth.
Screenshots attached. |
8x |
602
|
Wed Apr 24 15:37:59 2024 |
PP | 16:00-00:00 shift Wednesday 24 April | All seems OK.
Screenshots attached. |
8x |
601
|
Wed Apr 24 09:04:55 2024 |
TD | 08:00-16:00 Wednesday 24 April |
09.56 DAQ continues data file S100_alpha/R18_122
FRS setting 162Eu
|
33x |
600
|
Tue Apr 23 22:37:46 2024 |
TD | 00:00-08:00 Wednesday 24 April | 23.34 23.4.24
DSSSD bias & leakage current OK - attachment |
18x |
599
|
Tue Apr 23 19:23:42 2024 |
PP | 20:30 checks | All looks good.
Screenshots attached. |
8x |
598
|
Tue Apr 23 15:11:50 2024 |
TD | Tuesday 23 April 16:00-00:00 |
16.12 Analysis of data file R17_101 - attachment
1
|
20x |
597
|
Tue Apr 23 13:15:43 2024 |
PP | 14:00 checks | All looks good.
Screenshots attached |
8x |
595
|
Tue Apr 23 13:08:28 2024 |
TD | Summary of AIDA FEE64 stops |
We have observed a number of AIDA FEE64s
stop producing data.
|
 |
594
|
Tue Apr 23 11:09:00 2024 |
PP | 12:00 checks | Everything is OK now.
Screenshots attached. |
8x |
593
|
Tue Apr 23 10:52:39 2024 |
PP | Aida02 down and back | We lost aida02 at about 11:25 again. Nick
fixed it about 11:50.
It is now back in the DAQ. |
|
592
|
Tue Apr 23 10:10:31 2024 |
PP | aida02 is back | Nick rebooted AIDA and aida02 came back
at around 10:55.
AIDA was out of the DAQ from about |
|
591
|
Tue Apr 23 09:21:01 2024 |
PP | 10:20 checks - aida02 is down | Aida02 is down. Contacted Nick.
The rest looks stable, see screenshots.
10:53 CEST (NH): Rebooted aida02 |
8x |
590
|
Tue Apr 23 07:27:18 2024 |
PP | 08:00-16:00 Tuesday 24 April shift | Checks 08:00-08:30
Screenshots are attached and everything
looks OK. |
8x |
589
|
Mon Apr 22 22:35:59 2024 |
TD | 00.00-08.00 Tuesday 22 April |
23.36 Analysis of data file R17_34 - attachment
1
|
26x |
588
|
Mon Apr 22 15:17:07 2024 |
DSJ | 16.00-0.00 22/04/24 | First checks of shift. All looks ok.
Screenshot of HV, temps,
rates, Merger attached. Attachments 1-4. |
26x |
587
|
Mon Apr 22 11:43:55 2024 |
TD | 21-22.4.24 overnight MIDASsort online sort |
Attachment 1 - per DSSSD LEC e_p versus e_n
(20keV/channel)
|
7x |
586
|
Mon Apr 22 09:03:54 2024 |
TD | Report: aida04 stops producing data |
09.14 aida04 stopped producing data - system
console output - attachment 1
|
|
585
|
Sun Apr 21 23:20:45 2024 |
TD | Monday 22 April | 00.15 Data link to MBS was closed by MBS -
reason(s) unknown
NH restarted MBS receiver and data |
38x |
584
|
Sun Apr 21 00:34:33 2024 |
TD | Sunday 21 April | 01.35 DSSSD bias & leakage current OK - Grafana
- attachment 1
|
37x |
583
|
Sat Apr 20 17:52:27 2024 |
TD, NH | Saturday 20 April contd. |
Note 40kBq 60Co source placed on bottom surface
of snout centred on bPlas detectors and DSSSDs
|
  |
582
|
Sat Apr 20 11:40:38 2024 |
TD | Offline analysis R12 & R13 ( pulser walkthrough ) |
S100 ADC offsets using pulser walkthrough
data from data files R12 & R13
|
17x |
581
|
Sat Apr 20 03:20:23 2024 |
TD | Saturday 20 April |
Note 60Co source (current activity c. 43kBq)
was installed yesterday evening - assume |
23x |
580
|
Fri Apr 19 13:52:34 2024 |
JB, CC, TD | Friday 19 April contd. |
12.00 Replaced AIDA ASIC mezzanine of aida01
to fix issue with asic #3
|
23x |
579
|
Fri Apr 19 02:35:14 2024 |
TD | Friday 19 April |
03.30
|
38x |
578
|
Thu Apr 18 13:59:55 2024 |
TD, JB | WR timestamp OK, readout time not OK |
WR timestamps appear to be OK but readout
times fort aida03, aida04 and aida09 appear |
    |
577
|
Thu Apr 18 08:57:48 2024 |
JB, TD | Thursday 18 April | 09.52 DAQ continues file S100_alpha/R5
note merger not functioning correctly
due to aida03, aida04 & aida09 out of WR |
36x |
576
|
Tue Apr 16 23:37:58 2024 |
TD | Wednesday 17 April |
00.40 Observed zero data fron aida03, aida04,
aida09
|
44x |
575
|
Tue Apr 16 10:49:37 2024 |
TD, JB, CC | Tuesday 16 April | 11.45 DSSSD bias & leakage current OK - attachments
9-10
|
24x |
574
|
Mon Apr 15 09:23:43 2024 |
CC, JB, HMA, TD | Noise in AIDA when connecting BB7 | 9:40 started up AIDA. TEMP OK. Noise condition
same as Saturday. Attachments 1-3.
9:56 BB7 premaps connected but |
13x |
573
|
Sat Apr 13 14:04:44 2024 |
JB | 13 April Noise checks | 15:04 bPlast was left powered over night.
Powering up to check noise conditions - also
to check shifting of Germanium baseline (no |
9x |
572
|
Thu Apr 11 22:04:50 2024 |
JB | 12.04.2024 AIDA-bPlast noise optimisation | 15:00 Platform in, biasing detector. Temp
OK Attachment 1. test - Pulser OFF.
|
    |
571
|
Thu Apr 11 08:27:21 2024 |
JB, TD | 11 April noise tests |
TO-DO for 11.04.2024
- Try bringing
bPlast drain wire ground back to the PSU
ground for PN 300 and R&SRMP4040.
- |
48x |
570
|
Wed Apr 10 08:37:57 2024 |
JB, CC, TD | Wednesday 10 April |
09.38 CC completed install of Bplast driver
PCBs yesterday evening.
|
47x |
569
|
Tue Apr 9 13:22:41 2024 |
TD, JB, CC | Continued of 9th of April | Disconnected test - pulser from aida08.
Noise conditions improved slightly, probably
less pulser events. (Attachments 1 & |
  |
568
|
Tue Apr 9 09:04:07 2024 |
TD | [How To] Remote control of BNC PB-5 |
BNC PB-5 set to remote control
|
|
567
|
Tue Apr 9 08:37:47 2024 |
TD | Tuesday 9 April | 09.36 Cooling water temperature and flow OK
FEE64 power ON
|
16x |
566
|
Mon Apr 8 16:39:00 2024 |
JB, CC, TD | Monday 8 April |
17.32 Power and detector bias cycle
|
11x |
565
|
Mon Apr 8 11:18:19 2024 |
TD | S505 offline analysis data file R3_150 |
DRG quotes ( from Elog ) S505 FEE64 configuration
as
|
    |
564
|
Mon Apr 8 10:23:16 2024 |
TD, JB | AIDA Noise optimisation | In the morning we reinstalled FEE64s aida06,
aida13, aida07. The rails of the FEE64 boards
were found to be incorrect (that of the rails |
30x |
563
|
Sun Apr 7 18:13:56 2024 |
TD | Sunday 7 April contd. |
19.10 CAEN N1419ET LK fitted
power cycle
|
6x |
562
|
Sun Apr 7 13:08:52 2024 |
TD | FEE64 configuration | FEE64 configuration per https://elog.gsi.de/despec/S100/6
Configuration: Wide - 16 FEEs, 2 DSSD
|
|
561
|
Sun Apr 7 12:16:27 2024 |
TD | aida06 system console log | ISOL Version 1.00 Date 9th January 2017
Flash base address=FC000000
Set Flash to ASync Mode
|
|
560
|
Sun Apr 7 12:02:23 2024 |
TD | Sunday 7 April |
13.00 Cooling water temperature and flow
OK - attachment 1
|
46x |
559
|
Wed Apr 3 13:02:19 2024 |
NH | Merger for 16 FEEs | Changed /MIDAS/Linux/startup/NewMerger
Change parameters -i and -l in master64 to |
    |
558
|
Wed Apr 3 12:42:57 2024 |
NH | Report aida02 WR errors | The WR error counter for aida02 seems to
consantly rise
Tried reseating cable on both ends, |
    |
557
|
Wed Apr 3 12:09:47 2024 |
NH | Report - aida06 frequently fails to boot first time (PHY error) | When booting up AIDA aida06 usually crashes
the first time, it fails to get IP from DHCP
After 180 seconds it reboots and seems |
|
556
|
Tue Apr 2 18:37:21 2024 |
TD | S505 ADC offsets |
S505 ADC offsets using pulser walkthrough
data from data file R1
|
|
555
|
Tue Apr 2 12:36:25 2024 |
JB, CC, NH | Installing FEE64s of DSSSD2 cont. | Some additional checks
- check 'top
hat' electrical isolators are correctly |
|
553
|
Thu Mar 28 09:18:53 2024 |
TD | Installing FEE64s of DSSSD2 | Some additional checks
- check 'top hat' electrical
isolators are correctly installed at each |
|
552
|
Wed Mar 27 14:22:35 2024 |
JB, NH | Installing FEE64s of DSSSD2 | Mounted on frame:
DSSD 1 (Upstream) :
3208-2/3208-5/3208-8 |
14x |
551
|
Tue Mar 26 18:56:03 2024 |
TD | USB-controlled ac mains relay interlock box - wiring |
Sensor ( 4 pins )
|
|
550
|
Tue Mar 26 10:32:05 2024 |
NH, JB | Tue 26 March | Taken 4 FEE64s from CRYRING (the 4 easiest
to access)
41:d8:2b
|
9x |
549
|
Fri Mar 22 08:31:39 2024 |
TD | Friday 22 March |
09.30 Systems check
|
39x |
547
|
Fri Mar 22 08:29:55 2024 |
TD | Anydesk restarted remotely | Anydesk restarted remotely per https://elog.ph.ed.ac.uk/CARME/489
Anydesk address now restored to 832827869 |
|
546
|
Fri Mar 22 08:22:43 2024 |
NH, AM, MP, CC | Thu Mar 21 | > Fig 1-3: Noise situation at real thresholds
(0xa p+n, 0xf n+n)
>
|
|
545
|
Fri Mar 22 07:34:54 2024 |
NH, JB | Au Beam | |
|
544
|
Thu Mar 21 15:41:49 2024 |
NH, AM, MP, CC | Thu Mar 21 | Fig 1-3: Noise situation at real thresholds
(0xa p+n, 0xf n+n)
|
6x |
543
|
Wed Mar 20 17:02:53 2024 |
NH | /dev/sdd | The following messages are in the system log
very often:
|
|
542
|
Wed Mar 20 12:22:27 2024 |
NH | Wed Mar 20 | Turn on AIDA for Dry Run demonstrations
and so on
All system wide checks, temp, bias |
9x |
541
|
Tue Mar 19 10:17:30 2024 |
NH | Dry Run 2024 - 19th March 24 | AIDA has 8 FEEs and 1 DSSSD
Aida08 (HDMI#12) had no WR again, I moved |
    |
540
|
Mon Mar 18 18:04:43 2024 |
NH | Preparation for pre-s100 dry run (and test beam???) | In preparation for the dry run the following
*temporary* changes to the FEE numbering
have been prepared
|
|
539
|
Fri Mar 15 16:29:57 2024 |
NH | Leakage currents | The behaviour of the DSSSD leakage current
at low voltages and during biases is unusual
and varies depending on how the adapter boards |
|
537
|
Thu Mar 14 13:00:23 2024 |
JB, NH, MA, AM, GA | Mounting and biasing DSSD 2 | new Downstream DSSD2: 3208-2/3208-5/3208-8
Covered with black cloth. |
|
536
|
Thu Mar 14 12:13:17 2024 |
JB, NH, MA | Mounting and biasing DSSD 1 | Upstream bPlast mounted
new Upstream DSSD: 3208-2/3208-5/3208-8
|
|
535
|
Wed Mar 13 13:02:19 2024 |
JB, NH, HA, MA | DSSD 1 biasing tests | Dismounted Snout and biased DSSSD1 channels
leftmost waifer working
middle waifer reaches 90V then |
|
534
|
Tue Mar 12 16:16:49 2024 |
NH, TD, JB, HA | Summary of DSSSD Biasing 12.03 | DSSD#1 undergoes a breakdown at 90V, two of
the three wafers show this
- The adapter PCBs themselves have no breakdown |
|
533
|
Mon Mar 11 15:19:44 2024 |
JB, NH, Priyanka, Michael Armstrong, Helena Albers | To Do: AIDA PCB tests | To do:
1) Disconnect ribbon cables
from p+n junction FEE64 adaptor PCBs of upstream |
|
532
|
Sun Mar 10 17:08:12 2024 |
NH | AIDA FEE Layout + Cabling Plan for S100 | Proposed FEE numbering and wiring plan for
upcoming experiment S100 (2x Wide DSSSDs)
|
   |
531
|
Fri Mar 8 16:05:57 2024 |
JB, CC, TD, NH | Friday 8 March | Bias tests of AIDA on individual wafers
and in parallel. Spreadsheet can be found
in attachment 1, and graphic results can |
  |
530
|
Thu Mar 7 15:15:29 2024 |
JB, NH, CC | AIDA HV Bias Test IV Curve |
AIDA I-V
Test
Voltage |
|
529
|
Thu Mar 7 11:09:05 2024 |
HA, JB, CC, TD, MG | Thursday 7 March |
Snout assembly
|
|
528
|
Tue Mar 5 09:10:01 2024 |
MA, JB, TD | Tuesday 5 March |
Old ribbon cable assemblies - attachments
1 2
|
  |
527
|
Mon Mar 4 11:35:30 2024 |
TD | Monday 4 March | 12.30 FEE64 41:d7:cd ASIC mezzanine u/s, no
data, ASIC temperature c. 20 deg C ( ambient
) low
|
11x |
526
|
Sun Mar 3 16:06:11 2024 |
TD | Sunday 3 March |
17.00 FEE64 41:a0:71 ASIC #2 u/s
Replaced ASIC mezzanine with new ASIC |
24x |
525
|
Mon Feb 26 13:17:51 2024 |
TD | Offline analysis of DEC23/R9_6 |
First pass analysis of data file /TapeData/DEC23/R9_6
|
9x |
524
|
Wed Jan 31 15:27:07 2024 |
TD | Photos of water flow & dew point sensor interlock box | Attachment 1 - 4 socket input connectors (
from sensor to interlock box )
|
   |
523
|
Mon Nov 13 11:46:54 2023 |
TD | Summary of AIDA 24cm x 8cm 'triple' DSSSD tests | |
 |
522
|
Thu Oct 12 14:01:43 2023 |
TD | MSL type BB18 24cm x 8cm DSSSD test - update |
DSSSD MSL type BB18 24cm x 8cm 3208-3/3208-21/3208-22
|
 |
521
|
Sun Oct 1 11:47:07 2023 |
TD | S4 cooling water |
The photographs show the cooling water controls
and temperature/pressure gauges outside S4 |
 |
520
|
Thu Aug 31 15:24:56 2023 |
NH | New AIDA MBS PC | The AIDA MBS FDR will be x86l-119 from now
on, not x86l-94
|
|
519
|
Mon Aug 28 12:47:56 2023 |
NH | Power Failure 24.08.2023 | There was a power failure in the morning of
24.08.2023 in the Rhein-Main area affecting
GSI
|
|
518
|
Sun Jul 16 09:32:57 2023 |
TD | Firefox browser proxy setting change |
Firefox browser proxy setting changed to
'Auto-detect proxy settings for this network' |
|
517
|
Wed Jan 18 13:40:33 2023 |
PJCS TD | MACB settings with either Emulator or VITAR | When using the VETAR connected
to the HDMI port of the root MACB the settings
for all the MACB in the system 0x3. |
   |
516
|
Fri Dec 16 14:02:12 2022 |
NH | AIDA System off for christmas break | The AIDA NIM crate, pis and workstation have
been powered off for the Christmas break
and will not be accessible |
|
515
|
Thu Oct 6 16:51:49 2022 |
NH | Oscilloscope analysis | Investigating AIDA noise with a TA041 differential
probe and oscilloscope
|
24x |
514
|
Wed Sep 14 19:07:07 2022 |
PJCS | INFO : Three Merger Statistics explained | There are three Merger statistics that
can be used to better understand how the
data flow through the Merger system is proceeding. |
|
513
|
Thu Sep 8 12:37:18 2022 |
NH | Proxy Port Changed | The proxy in Firefox, Yum and AnyDesk has
been changed as the old wasn't working
|
|
512
|
Thu Sep 8 12:31:25 2022 |
NH | Retrying AIDA DataAcq v10 | Startup AIDA with ribbon cable connected to
aida03 and aida07 for noise
|
  |
511
|
Tue Aug 30 13:48:32 2022 |
NH | AIDA Single Switch Configuration | The second switch was moved back to CARME
so AIDA has been configured back to using
a single switch
|
|
510
|
Tue Jul 5 08:53:20 2022 |
TD | To Do |
In no particular order
|
|
509
|
Wed Jun 29 10:48:26 2022 |
NH, OH | AIDA Dismounted | All detectors removed from single and triple
AIDA snouts
|
|
508
|
Tue Jun 28 10:11:35 2022 |
OH, NH | MIDAS Data Aq V10 | 11:11 Rebooted FEEs and changed aidacommon
in /MIDAS/linux-ppc_4xx/startup to point
to the new V10 DataAq that Patrick produced
|
  |
507
|
Tue Jun 28 09:37:19 2022 |
NH | Tues 28 June 08:00- | Experiment over
10:37 - Stop DAQ & Tape
|
|
506
|
Mon Jun 27 23:11:47 2022 |
TD | Tuesday 28 June 00:00-08:00 | 00:07 Zero stats & all histograms
ASIC settings 2021Apr29-13-16-00
|
30x |
505
|
Mon Jun 27 17:03:09 2022 |
MA | Monday 27th June 16:00-00:00 | 16:00 Took over the shift from OH no beam
yet.
18:00 Still no beam yet. |
6x |
504
|
Mon Jun 27 06:45:22 2022 |
OH | Monday 27th June 08:00-16:00 | 07:45 Spoke to David and the beam has been
gone since about 05:30
Reason for the loss of beam is a vacuum |
   |
503
|
Sun Jun 26 23:04:30 2022 |
Marc | new shift - Monday 27 June 0:00 to 8:00 | 0:00 Experiment continues to run smoothly.
AIDA DAQ has been rather stable. Leakage
current has gone up for the past two days |
12x |
502
|
Sun Jun 26 08:51:20 2022 |
OH, NH | Sunday 26 June 08:00-24:00 | 09:51 Taken over from Tom following the night
shift
Experiment is still running smoothly
|
27x |
501
|
Sat Jun 25 22:48:57 2022 |
TD | Sunday 26 June 00:00-08:00 | 23.43 Check ASIC control
Zero stats & all histograms
|
30x |
500
|
Sat Jun 25 17:06:42 2022 |
BA, AA | Saturday 25 June 2022 16:00-00:00 | Took over the night shift from Magda
18:00 : attachments 1-3 |
9x |
499
|
Sat Jun 25 06:55:07 2022 |
MS | Saturday 25 June 2022 8:00-16:00 | Took over the night shift from Tom.
7:00 The beam is back.
8:00 : attachments 1-3 |
15x |
498
|
Fri Jun 24 22:36:45 2022 |
TD | Saturday 25 June 00:00-08:00 |
23:36 Check ASIC contorol
Zero stats & all histograms
|
31x |
497
|
Fri Jun 24 15:04:06 2022 |
Marc | Friday 24th June - evening shift | 16:05 - Last checked was at 15:30. (see
previous entry. All running smoothly.
Next wide check will be in about |
11x |
496
|
Fri Jun 24 06:56:06 2022 |
OH | Friday 24 June 008:00-16:00 | 07:56 Took over from AM. No issues reported
overnight
Stats ok - attachment 1
|
16x |
495
|
Fri Jun 24 00:34:55 2022 |
AM | Friday 24 June 00:00-08:00 | 01:30 Attachments 1-5,
white rabbit and fpga timestamp failures,
otherwise all good |
20x |
494
|
Thu Jun 23 15:10:17 2022 |
Philippos Papadakis | Thursday 23 June 16:00-24:00 | 16:20
System wide checks.
Clock status OK
|
15x |
493
|
Thu Jun 23 06:56:02 2022 |
OH | Thursday 23 June 08:00-16:00 | 07:56 Taken over from AM
Current White Rabbit Error status:
Base Current Difference
|
11x |
492
|
Wed Jun 22 23:18:45 2022 |
AM, OH, TD, MA | Thursday 23 June | 00:00 TD Noticed on the last stats uploaded
that AIDA08 had stopped sending signals.
Attempted to recover restarting merger |
19x |
491
|
Wed Jun 22 15:09:37 2022 |
MA | Wednesday 22 June 16:00-00:00 | 16:00 Take over from Magda mrning shift
Beam was off of sometime for tunning
to Mercurey setting. |
12x |
490
|
Wed Jun 22 10:09:12 2022 |
PJCS | INFO: FEE64 supply Voltages | Study of the FEE64 power supply distribution
has yielded the following :-
The most sensitive regulator, as |
|
489
|
Wed Jun 22 07:12:46 2022 |
MS | Wednesday 22 June 8:00-16:00 | Took over the night shift from Tom.
8:10 : attachments 1-3
10:00 : attachments 4-6 |
16x |
488
|
Tue Jun 21 23:29:16 2022 |
TD | Wednesday 22 June 00:00-08:00 |
00.28 all histograms & stats zero'd
baseline counters
|
33x |
487
|
Tue Jun 21 10:19:05 2022 |
OH ML | Tuesday 21st June 08:00-00:00 | 11:19 Stats ok 0xa on all FEE - attachment
1
Current pulser settings - attachment |
12x |
486
|
Mon Jun 20 09:26:23 2022 |
OH, TD | Monday 20 June | 10:26 MACB HDMI for AIDA03 and AIDA04 reseated.
Merger and DAQ restarted - All happily
sending data
|
11x |
485
|
Sun Jun 19 19:55:20 2022 |
NH | aida03 and aida04 | aida03 and aida04 seem to stop sending data
quickly
|
 |
484
|
Sun Jun 19 12:36:10 2022 |
OH | Sunday 19th June | Yesterday the sum/inverter was found to be
a considerable source of noise in the system.
|
    |
483
|
Sun Jun 19 00:22:24 2022 |
TD | Sunday 19 June |
01.20
|
14x |
482
|
Sat Jun 18 09:46:57 2022 |
OH, NH | Friday 17th June | Summary of the day:
Small improvements were made throughout the
day but by far the biggest change to the |
23x |
481
|
Tue Jun 14 12:15:33 2022 |
NH | S450 Archived | All MIDAS files for S450 (/TapeData/S450)
have been copied to lustre and archived |
|
480
|
Mon Jun 13 16:38:37 2022 |
NH | Mon 13 Jun | Remove Caen HV Supply
Open up, C23 was grounded (HV grounded in |
|
479
|
Fri Jun 10 10:32:46 2022 |
NH | Fri 10 Jun | Probe PSU output with 'scope again and 10X
probe
|
8x |
478
|
Wed Jun 8 09:55:28 2022 |
NH | Weds 8 May | 10.30 : Turn off AIDA FEE64s
Prepare a ground cable for snout |
 |
477
|
Wed Jun 1 08:00:18 2022 |
TD | 2xDSSSDs spectra layouts |
Spectra layouts for 2x DSSSDs can be found
in directory ~/LayOut/2xDSSSDs |
|
476
|
Wed Jun 1 07:35:13 2022 |
TD | Wednesday 1 June |
Attachment 1 - DSSSD bias & leakage current
- OK
|
15x |
475
|
Tue May 31 17:09:35 2022 |
TD | Tuesday 31 May | 2x MSL type BB18(DS)-1000 installed
|
20x |
474
|
Tue May 31 09:05:00 2022 |
NH, TD | Tuesday 31 May | DSSD stack will be as follows
1 3208-7 1019 125 |
|
473
|
Wed May 25 14:28:26 2022 |
NH | Wed 25 May | The n+n FEEs were all removed to adjust the
platform back to the single configuration
|
|
472
|
Tue May 24 08:35:00 2022 |
NH | Tue 24 May | 09:00 - Remove bias of DSSDs and pwoer of
FEEs
|
|
471
|
Mon May 23 16:41:45 2022 |
TD | Monday 23 May | 17.40 DAQ stopped
Attachment 1 - DSSSD bias & leakage current
|
  |
470
|
Sun May 22 14:12:55 2022 |
TD | Sunday 22 May | 15.11 DAQ stopped
Attachment 1 - DSSSD bias & leakage current
|
  |
469
|
Sat May 21 15:30:45 2022 |
TD | Saturday 21 May | 16.31 DAQ stopped
Attachment 1 - DSSSD bias & leakage current
|
  |
468
|
Fri May 20 16:22:01 2022 |
TD | Friday 20 May | 17.20 DAQ stopped
Attachment 1 - DSSSD bias & leakage current
|
  |
467
|
Thu May 19 17:00:46 2022 |
TD | Thursday 19 May |
18.01 DAQ stopped
|
   |
466
|
Wed May 18 15:29:50 2022 |
TD | Wednesday 18 May | file no storage mode, no MBS data transfer
ADC control register 0x0
|
8x |
465
|
Tue May 17 16:53:20 2022 |
TD | Tuesday 17 May 16.00-00.00 |
17.35 zero all histograms
zero statistics
|
24x |
464
|
Tue May 17 06:58:51 2022 |
OH | Tuesday 17th May | 08:00 Still have beam for the time being.
Currently on file R4_2049
|
6x |
463
|
Mon May 16 23:02:54 2022 |
AM | Tuesday 17th May 00:00-08:00 | 01:00: Attachments
1-3, System checks good
03:00: |
12x |
462
|
Mon May 16 17:23:27 2022 |
TD | NewMerger stats | |
   |
461
|
Mon May 16 15:12:50 2022 |
RDP, MA | Monday 16th May 16:00-00:00 | 16:00 Took over the shift from Magda.
17:30 Updates: Attached (1-3)
19:30 Updates: Attached (4-6) |
12x |
460
|
Mon May 16 14:07:06 2022 |
TD | Lost activity monitor | |
|
459
|
Mon May 16 07:03:16 2022 |
MS, OH | Monday 16th May 08:00-16:00 | 08:00 Took over the night shift from Tom
(Attachments 1-3)
10:00: Attachement 4-6 |
17x |
458
|
Sun May 15 23:00:08 2022 |
& TD | Monday 16 May 00:00-08:00 |
00.01 DAQ continues
file S450/R4_1319
|
25x |
457
|
Sun May 15 15:51:49 2022 |
ML | Sunday 15th May 16:00-0:00 | Status at 16:45 (CET)
The expreiment continues to run
smoothly. |
12x |
456
|
Sun May 15 07:40:52 2022 |
OH | Sunday 15th May 08:00-16:00 | 08:40 FEE Temperatures ok - attachment 1
FEE statistics - Attachment 2
Bias and leakage currents ok - attachment |
10x |
455
|
Sat May 14 23:02:33 2022 |
TD | Sunday 15 May 00:00-08:00 |
Pb beam c. 1e+9/spill
|
48x |
454
|
Sat May 14 21:34:11 2022 |
BA, AA | Saturday 14 May | FEE64 module aida09 global clocks failed,
6
Clock status test result: Passed |
  |
453
|
Sat May 14 19:39:13 2022 |
BA, AA | Saturday 14 May | FEE64 module aida09 global clocks failed,
6
Clock status test result: Passed |
  |
452
|
Sat May 14 17:34:40 2022 |
BA, AA | Saturday 14 May | FEE64 module aida09 global clocks failed,
6
Clock status test result: Passed |
  |
451
|
Sat May 14 15:35:24 2022 |
BA, AA | Saturday 14 May | |
  |
450
|
Sat May 14 13:01:58 2022 |
MS, OH | Saturday 14 May | 14:00 |
  |
449
|
Sat May 14 10:59:52 2022 |
MS, OH | Saturday 14 May | 12:00 |
  |
448
|
Sat May 14 09:00:21 2022 |
MS, OH | Saturday 14 May | 10:00 |
  |
447
|
Sat May 14 06:57:21 2022 |
MS, OH | Saturday 14 May | 08:00 Took over the shift from Philippos
In system wide checks aida09 fails
ASIC clock check but it is bit 6 which is |
|
446
|
Sat May 14 03:14:00 2022 |
PP | Shift Checks | 4:13
AIDA stats OK
Leakage current OK |
8x |
445
|
Sat May 14 01:52:03 2022 |
ML | 2h-Shift Checks | AIDA stats ok
Leakage current ok
Temperatires ok |
    |
444
|
Fri May 13 23:37:59 2022 |
ML | 2h-Shift Checks |
Took over remotely from CB about an
hour ago
At 0h40 (CET):
Stats still
ok - attach 1
Leakage still OK - Attach |
   |
443
|
Fri May 13 15:00:23 2022 |
CB | 13 March - 16:00-24:00 shift | 16:00 Took over remotely from OH, TD who remain
in GSI
|
10x |
442
|
Fri May 13 07:36:18 2022 |
OH, TD | Friday 13 May | 08:36 DAQ Still running smoothly
Currently on run R1_328
At current data rates we have just |
35x |
441
|
Fri May 13 06:56:14 2022 |
OH, BA, MA | Low Rates |
|
|
440
|
Fri May 13 06:47:33 2022 |
ML | 2h-Shift Checks | It has been a quiet night. AIDA has been
stable.
Stats appears stable appear |
  |
439
|
Fri May 13 04:45:23 2022 |
ML | 2h-Shift Checks | AIDA is running ok. Note that, in the last
hour or so, the rate of implant in DSSD1
is back to higher rate ~10 Hzto 15 Hz as |
  |
438
|
Fri May 13 02:56:36 2022 |
ML | 2h-Shift Checks | All seemnormal on AIDA side.
Stats as before
Temperatures stable |
  |
437
|
Fri May 13 02:28:36 2022 |
ML | beam dropped | We see less counts in AIDA implant DSSD1
(~1Hz) than at the beginning of the night
shift (~10Hz). |
|
436
|
Fri May 13 00:50:16 2022 |
ML | 2h-Shift Checks | All seems ok (apart for my virgin
network connection usually quite reliable
but not so tonight !?!). |
  |
435
|
Thu May 12 22:49:39 2022 |
BA, MA | Low Rates | The rate is very low in all aida, not sure
. |
|
434
|
Thu May 12 21:43:49 2022 |
BA, MA | Rates |
Statistic check (screenshot attached).
Temperatures OK (screenshot attached). |
  |
433
|
Thu May 12 16:30:47 2022 |
NH, OH, TD | Notes for mounting a second switch | Update DHCPD as expected
Make sure new ethernet device is trusted
in firewall
|
|
432
|
Thu May 12 10:10:51 2022 |
TD, OH | AIDA data throughput |
Attachment 1 - top aida06
|
15x |
431
|
Thu May 12 07:09:43 2022 |
TD | Thursday 12 May 08:00-24:00 |
08.10 AIDA DAQ status
|
27x |
430
|
Thu May 12 00:37:52 2022 |
OH, AM | Thursday 12 May | 01:37 Over the past hour the degrader settings
have been checked by impinging 209Pb onto
the detectors.
|
13x |
429
|
Wed May 11 19:08:41 2022 |
CB, PP | Wednesday 11th May - evening shift | 20:19 no beam yet.
Statistic check (screenshot attached).
Temperatures OK (screenshot attached). |
   |
428
|
Wed May 11 07:04:13 2022 |
OH, NH, TD | Wednesday 11th May | 08:00 Overnight at some point it appears that
aida14 has stopped producing any ASIC data
It is currently sending data pause |
9x |
427
|
Tue May 10 07:41:28 2022 |
NH TD | Tuesday 10 May | 8:15am S4 CONTROLLED ACCESS to remove beam
plug and check bPlas/AIDA response to beam
|
29x |
426
|
Mon May 9 08:36:23 2022 |
TD, NH | Monday 9 May |
09.32 Re-start AIDA
|
31x |
425
|
Sun May 8 17:43:57 2022 |
NH | AIDA FEE64 Layout | AIDA FEE64 layout with respect to the beam. |
|
424
|
Sun May 8 15:10:45 2022 |
TD | Sunday 8 May |
DSSSD bias and FEE64 power OFF during Ge
install
|
6x |
423
|
Sat May 7 13:27:39 2022 |
TD | AIDA configuration |
Adaptor PCB link settings
|
|
422
|
Sat May 7 09:55:53 2022 |
TD | Saturday 7 May |
Attachments 1 & 2 per FEE64 1.8.L spectra
|
15x |
421
|
Fri May 6 09:31:41 2022 |
TD NH | Friday 6 May |
10:30 DAQ appears to have stopped
|
33x |
420
|
Wed May 4 16:50:04 2022 |
NH | AIDA 4th May | bPlas has been mostly cabled in now
Reseat LAN cable (AIDA Switch -> AIDA-3) |
 |
419
|
Tue May 3 14:15:17 2022 |
NH | AIDA Updates 2 May |
PAUSE/RESUME stats are dominated by 2x FEE64s
- aida07 and aida16
|
|
418
|
Mon May 2 13:08:49 2022 |
NH | AIDA Updates 2 May | The snout endcap has been sealed with new
copper tape over the red tape
|
 |
417
|
Mon May 2 09:52:20 2022 |
OH | Shifter checklist 2022 | Every 30 minutes:
- Reload statistics (Web browser tab screen |
|
416
|
Thu Apr 28 18:45:09 2022 |
NH | Alpha Background Run | Runs are in /TapeData/BGAPR22 located on 8TB
disk "SecondDrive" (N.B. 2A experiment is
using ThirdDrive)
|
  |
415
|
Wed Apr 27 15:47:46 2022 |
NH, HA, SA, CJ | AIDA 27 Apr 22 | Remount snout in S4 (photos to come)
Alignment with DEGAS OK but a bit short...
DEGAS platform will be brought 31 mm closer |
7x |
414
|
Fri Apr 8 08:59:42 2022 |
OH, NH | Friday 8th April | 09:59 Noticed that the right hand ribbon cable
for aida07 was offset vertically by 1. i.e.
Only 1 of the 2 rows of pins in. Unbiassed |
11x |
413
|
Thu Apr 7 12:42:14 2022 |
OH, NH, TD | Thursday 7 April |
Re-assembled triple snout re-installed on
the AIDA stand - DSSSD ribbon cables and |
8x |
412
|
Wed Apr 6 10:09:21 2022 |
OH, NH, TD | Wednesday 6 April | DSSSD bias OFF. DAQ STOPped
Attachment 6 FEE64 temps OK
|
6x |
411
|
Tue Apr 5 12:46:52 2022 |
TD | MSL type BB18(DS)-1000 triple - cable lengths |
OH/NH disassembling current AIDA triple snout
and DSSSD stack.
|
|
410
|
Tue Apr 5 08:52:16 2022 |
TD | Tuesday 5 April | DSSSD bias OFF. DAQ STOPped
Attachment 1 FEE64 temps OK
|
6x |
409
|
Mon Apr 4 09:19:49 2022 |
TD | Monday 4 April | AIDA adaptor PCB config - all LKs removed
*except*
|
12x |
408
|
Sun Apr 3 10:56:11 2022 |
TD | Sunday 3 April |
all system wide checks OK *except* aida10
failed ADC calibration
|
35x |
407
|
Sat Apr 2 10:42:49 2022 |
TD | Saturday 2 April | 11.10 Found aida08 not running - merger/TapeServer
stalled
|
21x |
406
|
Wed Mar 9 10:15:56 2022 |
NH TD | Wednesday 9 March - noise tests contd. | 11.15 - S4 is currently closed no access
If access possible (should be) plan is
|
38x |
405
|
Tue Mar 8 10:26:56 2022 |
TD NH | Tuesday 8 March 2022 |
11.25 System wide checks OK except aida06
clock fail status 6 - attachments 1-13
|
40x |
404
|
Mon Mar 7 12:20:26 2022 |
NH, TD | AIDA Grounding Checks Mon 7 Mar | Remove all DSSD ribbon cable braid ground
from FEE64s to check n+n ground loops
|
14x |
403
|
Sat Mar 5 10:43:45 2022 |
OH, NH | Grounding and noise tests 4th-5th March Summary | Tests performed
After
leaving overnight noticed the pulser was
not fully inserted into aida09
Little |
13x |
402
|
Fri Mar 4 08:40:36 2022 |
OH, NH | AIDA Noise and grounding tests 3rd March - Results | Grounding implemented
Each
FEE given direct connection to heavy duty
earth bar via adaptor board lemo ground
Earth |
9x |
401
|
Thu Mar 3 08:45:03 2022 |
OH, NH | AIDA Noise and grounding tests 3rd March | Pulser spectra
Changes
to the nn pulser circuit yesterday have improved
the peak widths on the p+n side
p+n |
8x |
400
|
Wed Mar 2 16:00:01 2022 |
OH, NH | Noise tests - 2nd March | An interesting find with the new sum inverter
is that with the pulser at default settings
the pulser peak on the n+n side is above |
|
399
|
Wed Mar 2 10:05:13 2022 |
OH | DAQ Status 2nd March 2022 | > DAQ was found yesterday with none of the
FEEs reporting any good events on any FEE
- attachment 1
|
|
398
|
Wed Mar 2 09:08:44 2022 |
OH | DAQ Status 2nd March 2022 | DAQ was found yesterday with none of the FEEs
reporting any good events on any FEE - attachment
1
|
    |
397
|
Tue Mar 1 08:29:19 2022 |
OH, NH | Findings from the upgrade to centos7 from SL6 | Required packages not included as default
csh
(For running MIDAS scripts)
xinetd
(Runs the rdate server for the FEEs)
Telnet |
 |
396
|
Wed Feb 16 10:32:52 2022 |
PJCS | AIDA software changes and some suggestions | The Pi console monitor program now outputs
a full date/time to file for each line of
report. |
|
395
|
Fri Feb 11 12:09:24 2022 |
NH | Fri 11 Feb AIDA Work | Use beam break between FRS experiments to
finalise getting AIDA testable for PJCS etc.
|
|
393
|
Tue Feb 1 15:00:28 2022 |
NH | AIDA CentOS 7 Startup | Starting up AIDA on CentOS 7 to confirm things
work (based on some CARME experience)
|
|
392
|
Fri Oct 29 23:00:04 2021 |
NH | Ground cable information | For future refinement of the AIDA ground cables
using copper bus bar or similar:
|
|
391
|
Fri Oct 29 10:49:01 2021 |
OH | Noise tests - 28th October -29th October | Plan for the day was to connect a mechanical
ground to the FEEs connected to DSSD0 1,
2, 3, 4, 9, 10, 11 and 12
|
24x |
390
|
Wed Oct 27 15:12:50 2021 |
OH, NH | Noise tests - 27th October | Began by installing the triple.
Only the upstream detector was connected
to the FEEs the downstream detector cables |
6x |
389
|
Tue Oct 26 16:08:46 2021 |
NH | FEE and Adaptor board groundings | > During investigations into the noise today
by NH and OH a couple of findings were made
on the grounding of the FEEs and Adaptor |
|
388
|
Thu Oct 21 16:18:51 2021 |
OH | FEE and Adaptor board groundings | During investigations into the noise today
by NH and OH a couple of findings were made
on the grounding of the FEEs and Adaptor |
|
387
|
Thu Oct 21 09:33:08 2021 |
OH | Noise tests - 21st Oct 2021 | Pulser was left on overnight with the bias
filters on
First plot is FEE14 not FEE8
|
    |
386
|
Thu Oct 21 09:08:41 2021 |
OH | MIDAS start up messages | |
    |
385
|
Wed Oct 20 13:26:08 2021 |
NH, OH | AIDA Noise + Filtering | Turn on AIDA (SL6) to investigate noise situation
and impact of new RC filters on bias input
|
6x |
384
|
Tue Oct 19 16:17:49 2021 |
OH | AIDA-gsi server to centos 7 | Following advice of CU have updated the server
to centos 7.
|
|
383
|
Mon Oct 18 14:20:32 2021 |
OH | Top bar launchers | launcher1
TclHttpd @8015 for AIDA
/MIDAS/Linux/startup/HTTPD
|
|
382
|
Mon Oct 18 13:03:23 2021 |
OH | DHCPd.conf backup | |
|
381
|
Mon Oct 18 13:01:04 2021 |
OH | Proxy config | Proxy for firefox and anydesk |
 |
380
|
Fri Sep 3 13:39:42 2021 |
OH | Additional noise tests performed at GSI | Contents of an email from NH to OH and TD
about tests performed on an additional bias
supply:
|
   |
379
|
Sun Aug 22 18:40:08 2021 |
TD | AIDA Noise | > >
> > 11:30 BNC PB-5 pulser polarity chnaged
from positive to negative (using LOCAL control)
|
|
378
|
Fri Aug 20 14:33:24 2021 |
TD | AIDA Noise | >
> 11:30 BNC PB-5 pulser polarity chnaged
from positive to negative (using LOCAL control)
|
6x |
377
|
Thu Aug 19 10:47:13 2021 |
TD | MIDAS spectra | |
|
376
|
Thu Aug 19 10:37:50 2021 |
TD | AIDA Noise |
11:30 BNC PB-5 pulser polarity chnaged from
positive to negative (using LOCAL control)
|
|
375
|
Tue Aug 17 12:39:25 2021 |
NH, TD | AIDA Noise | Investigate noise situation of AIDA (no DSSSDs
connected)
|
16x |
374
|
Mon Jul 5 14:36:15 2021 |
NH | Monday 5 July | AIDA Unbiased and powered down
All FEEs were running before stoppage. None
had rebooted.
|
|
373
|
Tue Jun 22 13:01:19 2021 |
NH | AIDA Data on lustre | The following runs are now copied to lustre
If necessary they can be removed from AIDA
HDDs for space.
|
|
372
|
Mon Jun 21 10:21:50 2021 |
NH | Monday 21 June | 11.21 - DAQ continues OK (no storage mode)
11.21 - System checks ok *except*
|
8x |
371
|
Thu Jun 17 11:14:04 2021 |
TD | Thursday 17 June | 12.15 DAQ continues OK - no storage mode
12.15 System checks OK *except*
|
8x |
370
|
Wed Jun 16 18:05:03 2021 |
TD | Wednesday 16 June |
19.04 All system wide checks OK *except*
|
7x |
369
|
Tue Jun 15 12:35:48 2021 |
TD | Tuesday 15 June | 13.35 DAQ continues OK - no storage mode
All system wide checks OK *except*
|
8x |
368
|
Tue Jun 15 12:33:46 2021 |
TD | Monday 14 June | |
|
367
|
Sun Jun 13 17:29:04 2021 |
TD | Sunday 13 June | 18.15 DAQ continues file S496/R32_946
Hot HEC channels multiple FEE64s from
c. 06.00 this morning
|
8x |
366
|
Sat Jun 12 14:53:56 2021 |
TD | aida15 - system console output - 15.33 Sat 12 June reboot | 04/11:03:40|get_ASICBlk : Blk=1 : bytes =
196096 : Blk_Status = 00000088 : Offset =
00100000 : Databuffer : 273090556 : 0 => |
|
365
|
Sat Jun 12 13:56:00 2021 |
TD | Saturday 12 June | 14.11 DAQ continues file S496/R32_697
High data rate from c. 06.00 this morning
ASIC check
|
9x |
363
|
Thu Jun 10 11:32:27 2021 |
TD | Thursday 10 June | 12.26 DAQ continues file S496/R32_428
High data rate from c. 06.00 this morning
ASIC check
|
9x |
362
|
Wed Jun 9 19:17:21 2021 |
TD | FEE64 system console logs | |
|
361
|
Wed Jun 9 11:27:47 2021 |
TD | Wednesday 9 June | 12.30 DAQ continues OK file S496/R32_373
System wide checks OK *except*
|
11x |
360
|
Tue Jun 8 10:51:55 2021 |
TD | Tuesday 8 June | 11.52 DAQ continues file S496/R32_340
High data ratefrom c. 18.00 yesterday
evening to c. 06.00 this morning
|
9x |
359
|
Mon Jun 7 10:33:48 2021 |
TD | Monday 7 June | 11.30 DAQ continues OK - file S496/R32_144
alpha background
|
9x |
358
|
Sun Jun 6 14:15:05 2021 |
TD | Sunday 6 June | 14.15 DAQ continues file S496/R32_117
Hot HEC channels multiple FEE64s from
c. 06.00 this morning
|
9x |
357
|
Sat Jun 5 11:20:15 2021 |
TD | Saturday 5 June |
12.25 DAQ continues OK - file S496/R32_48
alpha background
|
9x |
356
|
Fri Jun 4 09:47:44 2021 |
TD | aida15 - system console output - 00.02 Friday 4 June reboot | 03/23:58:55|kernel BUG at mm/slab.c:2974!
03/23:58:55|Oops: Exception in kernel mode,
sig: 5 [#1]
|
|
355
|
Fri Jun 4 08:40:52 2021 |
TD | Friday 4 June |
09.20 DAQ found stopped
files in /TapeData/S496 indicate increased |
8x |
354
|
Thu Jun 3 10:16:39 2021 |
TD | Thursday 3 June | 11.19 Hot HEC channels multiple FEE64s
ASIC check fixes
|
8x |
353
|
Wed Jun 2 14:53:05 2021 |
OH, TD | Merger/Tapeserver issues after reboot of aida-3 | On 01/06/21 aida-3 was rebooted as the graphical
user interface had frozen.
We were also unable to connected to MIDAS |
6x |
352
|
Sun May 30 08:57:53 2021 |
TD | Sunday 30 May | 09.55 DAQ continues - file S496/R18_588
alpha background
|
9x |
351
|
Sat May 29 13:13:45 2021 |
TD | Saturday 29 May | 14.10 DAQ continues - file S496/R18_539
alpha background
|
9x |
350
|
Fri May 28 10:29:07 2021 |
TD | Friday 28 May | 11.30 DAQ continues - file S496/R18_472
alpha background
|
9x |
349
|
Thu May 27 12:25:53 2021 |
TD | Thursday 27 May | 13.29 DAQ continues - file S496/R18_443
alpha background
|
9x |
348
|
Wed May 26 13:21:59 2021 |
TD | Wednesday 26 May |
14.22 DAQ continues - file S496/R18_369
alpha background
|
9x |
347
|
Tue May 25 11:16:54 2021 |
TD | Tuesday 25 May | 12.16 DAQ continues file S496/R18_324
alpha background
|
7x |
346
|
Mon May 24 09:57:07 2021 |
TD | Monday 24 May |
07.30 Hot HEC channels multiple FEE64s
ASIC check fixes
|
19x |
345
|
Sat May 22 23:40:32 2021 |
TD | Sunday 23 May |
00.40 Unable to connect to FEE64s 1-7
DAQ power cycled OK
|
    |
344
|
Sat May 22 13:40:56 2021 |
TD | Saturday 22 May |
14.37 DAQ continues file S496/R15_14
|
23x |
343
|
Fri May 21 23:19:23 2021 |
TD | Saturday 22 May 00:00-08:00 |
00.16 413Gb remaining /TapeData filesystem
estimate 413/2 x 177s = 10.15h to filesystem |
29x |
342
|
Fri May 21 15:03:48 2021 |
CA | Friday 21st May 16:00 - 00:00 shift | 16:00 CA takes over
DSSD1 n+n sides 0x20 on ASICs 1-3 and
0x64 on ASIC 4
|
17x |
341
|
Fri May 21 07:03:50 2021 |
OH, TD | Friday 21 May 08:00-16:00 | 08:00 OH and TD take over
Current settings p+n sides at 0xc
DSSD1 n+n sides 0x20 on ASICs 1-3 and |
9x |
340
|
Fri May 21 06:09:28 2021 |
ML | system wide checks | System wide checks at 7am CET
All seems fine, nothing significant
to report |
8x |
339
|
Fri May 21 02:17:47 2021 |
ML | system wide checks | Wide check at 3am -
All fine.
Same output as at the beginning |
9x |
338
|
Thu May 20 23:26:11 2021 |
Marc Labiche (ML) | system wide checks | System Wide check at begining of the night
shift:
----- |
9x |
337
|
Thu May 20 14:56:33 2021 |
OH | Thursday 20th May 16:00 - 24:00 | 16:00 p+n threshold 0xc (120keV)
n+n threshold DSSD2 0x20 (250keV) ASICS
1-3, 0x64 ASIC4
|
10x |
336
|
Thu May 20 07:03:02 2021 |
CA | Thursday May 20th 08:00 - 16:00 shift | 08:00 CA takes over
08:04 DAQ continues ok - writing to file |
40x |
335
|
Thu May 20 01:00:30 2021 |
Muneerah | Thu May 20 00:00-08:00 | 02:00 System check
The ucesb has stopped around 00:26,
the team in the lab ask us to restart AIDA! |
13x |
334
|
Wed May 19 20:02:27 2021 |
NH | AIDA Dead Time Analysis | Some plots of AIDA dead time (From PAUSE/RESUME
data items) against time.
|
6x |
333
|
Wed May 19 18:50:01 2021 |
LS | Wednesday 19th May 20:00 to 24:00 | 20.00
Stats okay (Fig 1)
Analysed R14_762 deadtimes are okay (Fig |
21x |
332
|
Wed May 19 14:54:36 2021 |
CB | Wednesday 19 May 16:00 - 20:00 | 15:55 Took over from OH
16:00
|
11x |
331
|
Wed May 19 11:24:55 2021 |
RDP | Monday 19 April 12 noon shift | System checks gave 11 passed/1 failed, except
for ADC timimgs: 10 passed/2 failed.
|
  |
330
|
Wed May 19 09:12:52 2021 |
Muneerah, Betool, LS | 17th May 00:00-12:00 | Auther: Muneerah and Betool
02:00 The dead time have been checked every
30 min from the start of the shift, it seemes |
17x |
329
|
Wed May 19 09:09:27 2021 |
MS, JS, OH | May 16 08:00-24:00 | 16th May 08:00 - 12:00 shift
Author: MS
|
21x |
328
|
Wed May 19 09:06:58 2021 |
CA | May 16th 00:00 - 08:00 | 16th May 2021 - 00:00 - 08:00 Shift
Author: CA
|
16x |
327
|
Wed May 19 06:59:33 2021 |
OH | Wednesday 19th May 08:00 - 16:00 | 08:00 OH Takes over for CA
Thresholds 0xc for p+n and 0x19 for
n+n ASICs 1-3 and 0x64 for n+nASIC4
|
9x |
326
|
Tue May 18 23:11:40 2021 |
CA | Wednesday May 19th 00:00 - 08:00 | 00:00 CA takes over
LS has performed usual checks, see
previous Elog entry - all ok
|
14x |
325
|
Tue May 18 18:47:25 2021 |
LS | Tuesday 18th May 20:00 - 24:00 |
19.50
Stats okay (Fig 1)
|
24x |
324
|
Tue May 18 14:57:43 2021 |
CB | Tuesday 18th May 16:00 - 20:00 | 15:58 Took shift over from OH and PP
Stats OK - attach 1
|
11x |
323
|
Tue May 18 06:50:53 2021 |
OH, PP | Tuesday 18th May 08:00 - 16:00 | 08:00 OH Takes over
Statistics still look good from yesterday. |
6x |
322
|
Mon May 17 23:17:16 2021 |
DSJ | Tuesday 18 May 0.00-8.00 shift | 0.00 - Taking over from Lewis
00.30 - Rates look ok (fig1), Ran
R14_147 through analyser, dead times |
39x |
321
|
Mon May 17 19:05:23 2021 |
LS | Monday 17th May 20.00-24.00 | 20.30
Checked stats, seem okay (Fig 1)
Ran R14_99 through analyser, dead times all |
21x |
320
|
Mon May 17 15:14:35 2021 |
LJW | Monday 17th May 2021 16:00-20:00 | 16:25
Informed at shift handover that
there has been no beam. |
9x |
319
|
Fri May 14 19:01:09 2021 |
JS | Friday 14th May 20:00-00:00 | 20:00 Taking over from CB
20:33 Doing full checks.
|
27x |
318
|
Fri May 14 17:30:56 2021 |
TD | Analysis of S496 files R5_18, R5_32 & R5_45 |
S496/R5_18 - attachment 1
|
  |
317
|
Fri May 14 14:57:32 2021 |
CB + OH, TD, NH | Friday 14th May 16:00-20:00 | 16:00 Shift taken over from CA
16:00
|
33x |
316
|
Fri May 14 07:05:08 2021 |
CA | Friday 14th May 08:00 - 16:00 shift | 08:00 CA takes over
08:28 Problem at UNILAC - no beam
|
24x |
315
|
Thu May 13 23:58:05 2021 |
OH | Friday 14th May 00:00 - 08:00 | 00:00 OH Takes over.
PULSER SETTINGS
---------------------------
|
21x |
314
|
Thu May 13 21:21:25 2021 |
Philippos and Marc | system wide checks | Wide checks at Time 22:22 CET
Clock status test result: Passed
16, Failed 0 |
13x |
313
|
Thu May 13 19:52:36 2021 |
Philippos and Marc | Connection problem | At 20:15, We lost connection with AnyDesk.
Liliana restarted Anydesk on the
AIDA DAQ PC. All seemed fine on AIDA DAQ |
|
312
|
Thu May 13 17:05:42 2021 |
Muneerah, MArc | Thu 13 May 04:00-00:00 | 19:08 System Check
Rates from ucesb below 1 kHz
Atachment 1 Stats |
  |
311
|
Thu May 13 07:10:17 2021 |
CA, TD | May 13th 08:00 - 16:00 shift | 08:10 all system wide checks ok *except*
all FEE64 modules fail ADC calibration
|
34x |
310
|
Thu May 13 00:51:40 2021 |
Muneerah and OH | Thu 13 May 00:00-08:00 | 01:50 Beam on
The leakage current and the voltages been
monitered every half an hour and logged in |
7x |
309
|
Wed May 12 18:53:38 2021 |
JS | Wed 12 20:00-00:00 | 19:51 Shift change
Stats ok
|
    |
308
|
Wed May 12 15:03:10 2021 |
CB, OH | Wed 12 16:00-20:00 | 16:00 Shift change
Statistics OK (attach 1)
|
15x |
307
|
Wed May 12 06:59:55 2021 |
OH, RDP | Wed 12 May 08:00-16:00 | 08:00 While the rates appeared to have quietened
around 7am they are back to their previous
rate now.
|
   |
306
|
Wed May 12 02:07:59 2021 |
MA, BA | Wed 12 May 00:00-08:00 | 03:12 System Check
attachment 1 : Current
attachment 2 : Temperature |
9x |
305
|
Tue May 11 07:21:24 2021 |
OH | Tuesday 11 May | 08:20 System wide checks all ok except
Base Current Difference
aida01 fault 0xc10c : 0xc10e : 2
|
  |
304
|
Mon May 10 11:45:15 2021 |
NH | Confusing behaviour of rates | Testing the behaviour of powering down the
fast ADCs in aida01 (the hot FEE)
|
|
303
|
Mon May 10 08:49:37 2021 |
TD | Monday 10 May | 09.51 All system wide checks OK *except*
|
21x |
302
|
Sun May 9 19:29:23 2021 |
TD | Sunday 9 May | |
14x |
301
|
Sat May 8 09:05:04 2021 |
TD | Saturday 8 May |
09.54 FEE64 temperatures OK - attachment
1
|
27x |
300
|
Fri May 7 19:47:54 2021 |
PJCS | HowTo mitigate excessive temperature in an FEE64 | After tests in the Daresbury T9 system.
Disabling the waveform ADCs in
an FEE64 which is running with the FPGA over |
|
299
|
Fri May 7 13:11:22 2021 |
NH | Friday 7th May | 14:11 - Alpha has been running most of morning
Just saw rates in tape spike to 6 MB/s... |
    |
298
|
Thu May 6 16:59:20 2021 |
TD, OH | Wednesday 6 May |
18.09 NH has installed jumpers LK2-4 for
*all* adaptor PCBs
|
13x |
297
|
Wed May 5 20:10:16 2021 |
OH | Wednesday 5 May Alpha run | 21:10 After all of the works in S4 today the
noise in the system was considerably worse
It was decided that before starting |
12x |
296
|
Wed May 5 16:57:48 2021 |
TD | Tuesday 5 May |
17.55 Grafana DSSSD bias & leakage current
- attachment 1
|
 |
295
|
Tue May 4 21:26:43 2021 |
OH | Alpha run | 22:26 Tape server set up to write to directory
May21
DAQ stopped
|
10x |
293
|
Tue May 4 14:26:50 2021 |
NH | FEE64 Adapter wiring diagram and layout | A figure showing the FEE64 numbers (a previous
diagram had 2/6 and 4/8 on the wrong side
of the DSSD) and the wiring of pulser/HV
|
|
292
|
Tue May 4 14:05:48 2021 |
NH | MACB Cables | During the installation of the 4 new HDMI
cables for aidas13-16 the HDMI cables between
the MACB Root and the MACB leaves were looped |
 |
291
|
Tue May 4 13:53:22 2021 |
OH, TD | Tuesday 4 May |
NH has re-checked alignment of AIDA adaptor
PCBs cf. Elog:290
|
7x |
290
|
Tue May 4 12:11:39 2021 |
OH TD | Bias test of triple | 13:11 System wide checks all ok except *aida02
fails ADC calibration*
Collecting the file size of each FEE64 Options |
    |
289
|
Sat May 1 08:45:19 2021 |
TD | AIDA 24cm x 8cm DSSSD + triPlast assembly for S496 |
Photos provided by Nic Hubbard and Helena
Albers from the assembly of 2x triPlast
|
12x |
288
|
Fri Apr 30 08:38:12 2021 |
TD | Friday 30 April | 09.40 All system wide checks OK *except*
FEE64 module aida07 failed
|
6x |
287
|
Thu Apr 29 16:56:42 2021 |
OH | Config changes for 16FEE | New set of ASIC settings produced 2021Apr29-13-16-00
- Note no longer odd/even for polarity of
signal
|
|
286
|
Wed Apr 28 17:39:35 2021 |
TD | Preparations for setup of aida13-aida16 |
Updated /etc/dhcpd.conf
|
   |
285
|
Wed Apr 28 10:06:11 2021 |
NH | Wednesday 28 April | 11.05 All system wide checks OK *except*
WR
|
    |
284
|
Tue Apr 27 08:26:01 2021 |
TD | Tuesday 27 April - aida09 system console reports 'received undefined information code: 1' and 'code: 9' |
09.25 All system wide checks OK *except*
|
6x |
283
|
Mon Apr 26 09:41:21 2021 |
TD | Monday 26 April | 10.42 All system wide checks OK *except*
Base Current Difference
|
    |
282
|
Sun Apr 25 10:22:16 2021 |
TD | Sunday 25 April | 11.25 All system wide checks OK *except*
Base Current Difference
|
9x |
281
|
Sat Apr 24 15:48:38 2021 |
TD | aida12 crash - 08:33:03 Monday 19 April |
from ttyUSB5 system console log
|
|
280
|
Sat Apr 24 15:28:17 2021 |
TD | aida05 crash - 11:08:07 Tuesday 20 April |
From ttyUSB7 system console log
|
|
279
|
Sat Apr 24 15:13:33 2021 |
TD | aida09 crash - 08:10:52 Tuesday 20 April |
From ttyUSB6 system console log
|
|
278
|
Sat Apr 24 13:33:44 2021 |
TD | aida01 restart - 22.07UTC+2 Tuesday 20 April |
From aida01 /var/log/messages
|
|
277
|
Sat Apr 24 13:25:23 2021 |
NH, TD | AIDA FEE64 crash log for S460 |
https://docs.google.com/spreadsheets/d/1CyL5L-WLsh9fsS7CN989GC3ktqiPQOkG8YIDJrrGxYA/edit#gid=0 |
|
276
|
Sat Apr 24 10:21:06 2021 |
NH | Saturday 24.04.2021 | I have restarted some more of the consoles
that disappeared since the aidas-gsi issue
last night
|
 |
275
|
Sat Apr 24 05:17:51 2021 |
ML | End of beam | Beam has now stopped (end of expriment)
and background measurements starts. |
|
274
|
Sat Apr 24 04:50:01 2021 |
ML | Spectra | Some screenshot of spectra attached |
8x |
273
|
Sat Apr 24 04:07:23 2021 |
ML | ucesb scalers | ucesb is back - see attachment |
|
272
|
Sat Apr 24 03:32:11 2021 |
ML | quick checks | Quick system checks after crash
Temperature ok.
Still cannot connect remotely to |
|
271
|
Sat Apr 24 03:16:19 2021 |
TD | aida-gsi window manager/login session crashed |
Looks like the aida-gsi window manager/login
session crashed.
|
|
270
|
Sat Apr 24 03:13:05 2021 |
ML | AIDA DAQ Reset | AIDA PC crashe completely at ~3:30am CET.
Tom was called and he Restarted
the AIDA system at ~4am with Run 52. |
|
269
|
Sat Apr 24 01:44:09 2021 |
ML | system wide checks | system wide checks:
|
7x |
268
|
Fri Apr 23 20:19:52 2021 |
TD | Friday 23 April 20.00-00.00 |
21.19 system wide checks
|
7x |
267
|
Fri Apr 23 15:13:28 2021 |
DSJ | 23 April 16.00-20.00 shift | 16:00 taking over from Muneerah
16:15 - temp, ucesb, database,
stats etc. look ok. |
6x |
266
|
Fri Apr 23 13:13:46 2021 |
MA, TD | Friday 23 April 12.00-16.00 |
14.15 System wide checks
|
6x |
265
|
Fri Apr 23 12:12:08 2021 |
TD | aida02 restart - 03.13UTC+2 Friday 23 April |
Overnight aida02 crashed and restarted itself.
A power cycle of all FEE64s was required |
|
264
|
Fri Apr 23 07:09:57 2021 |
LJW | Friday 23rd April 08:00-12:00 | 08:30
System Checks
|
9x |
263
|
Thu Apr 22 23:17:46 2021 |
DJ | Friday 23rd April 00:00-08:00 | Took over from TD ... All is well.
00:45 System Wide Checks
Clock: 12 passed, 0 failed. |
27x |
262
|
Thu Apr 22 15:18:20 2021 |
TD | Thursday 22 April 16.00-00.00 |
16.20 zero all histograms, statistics & merger
statistics
|
36x |
261
|
Thu Apr 22 11:03:57 2021 |
JS | Thursday 22nd April: 12:00-16:00 | 12:00 Taking over from Liam, all is well.
12:10 Beam dropped out for a minute, implantation |
12x |
260
|
Thu Apr 22 07:06:37 2021 |
LPG | Thursday 22nd April: 08:00-12:00 | 08:01 - Taking over from Corrigan...
ucesb ok!
|
8x |
259
|
Wed Apr 21 23:05:39 2021 |
CA | 22nd April 00:00 - 08:00 | 00:07 stats/ucesb/DB checks all ok
00:30 all system wide checks ok *except* |
16x |
258
|
Wed Apr 21 15:01:14 2021 |
LJW, MA | Wednesday 21st April 16:00-20:00 | 16:30
All System checks okay, except for: |
13x |
257
|
Wed Apr 21 11:27:55 2021 |
OH | Wed 21st April 12:00-16:00 | 12:30 MBS has crashed
This time AIDA is still running fine.
|
6x |
256
|
Wed Apr 21 06:56:42 2021 |
MS | Wednesday 21 April 08:00-12:00 2021 | 8:00
|
10x |
255
|
Tue Apr 20 23:08:22 2021 |
CA | Wed Apr 21 00:00-08:00 | 00:00 - CA takes over
00:15 - stats/database/ucesb ok - attachment |
23x |
254
|
Tue Apr 20 19:00:00 2021 |
JS | Tuesday 20th 20:00-00:00 | 20:00
Taking over from Liam, all systems good.
|
12x |
253
|
Tue Apr 20 14:56:38 2021 |
LPG | Tuesday 20th 16:00-20:00 | 16:00
New shift, ready to go! Beam stopped just |
15x |
252
|
Tue Apr 20 10:54:41 2021 |
CB | April 20 12:00 - 16:00 shift | 11:55 CA got AIDA back online
12:00 Stats OK (attach 1)
|
9x |
251
|
Tue Apr 20 07:06:20 2021 |
CA | April 20th 08:00 - 12:00 | 08:09 DAQ crashes at start of shift (!)
again happens at same time as beam |
    |
250
|
Mon Apr 19 23:01:02 2021 |
OH | 20th April 00:00-08:00 | 23:52 Noticed AIDA07 has dropped out of the
merger and is producing no statistics
Powercycling to recover
|
17x |
249
|
Mon Apr 19 19:19:18 2021 |
CB | Monday 19 April 20:00-24:00 shift | 20:19
Statistics OK (attach 1)
No faults found
|
11x |
248
|
Mon Apr 19 16:58:29 2021 |
LJW | Monday 19th April 16:00-20:00 Shift | System Checks @ ~18:10 :
|
6x |
246
|
Mon Apr 19 13:27:08 2021 |
RDP | Monday 19th April 12 noon shift |
12.30 There appear to be fewer HEC events
in aida09 cf. aida11 and upstream DSSSDs
|
  |
244
|
Mon Apr 19 06:46:15 2021 |
MS | Monday 19 April 07:47-12:00 2021 | 07:47
|
9x |
243
|
Mon Apr 19 01:08:40 2021 |
MA | Monday 19th April 00:00-08:00 | 03:09 General check
Rates, temptature, voltages are
OK attached 1, 2, 3, 4 |
8x |
242
|
Sun Apr 18 21:43:01 2021 |
ML-OH | system wide checks | New System wide check 10pm CET
Clock status test result: Passed
12, Failed 0 |
20x |
241
|
Sun Apr 18 19:05:33 2021 |
ML-OH | All spectra reset | All spectra reset at 8pm CET.
|
|
240
|
Sun Apr 18 17:27:44 2021 |
ML-OH | system wide checks | system wide checks at 6pm CET
##################################################
Difference noted from previous |
17x |
239
|
Sun Apr 18 11:18:21 2021 |
TD | Sunday 18 April 12.00-16.00 | 12.18 DAQ continues file NULL/R30_490
ASIC settings 2019Dec19-16.19.51
|
25x |
238
|
Sun Apr 18 07:30:20 2021 |
DJ, TD | Sunday 18th April 08:00-12:00 | ---- 09-26
FEE64 module aida06 failed
FEE64 module aida07 failed |
7x |
237
|
Sun Apr 18 00:57:36 2021 |
BA, MA | Sanday 18 April 00.00-08.00 | 02:01 Beam has stopped at 01:37 and returend
at 01:43 for few min and then stopped again
and not knowen how it will take until it |
12x |
236
|
Sat Apr 17 19:39:56 2021 |
DJ, TD | Saturday 17 April 20.00-00.00 |
19.37 per FEE64 rate spectra - attachments
1 & 2
|
13x |
235
|
Sat Apr 17 15:29:27 2021 |
MS | Saturday 17 April 16:00-20:00 2021 | 16:21
|
6x |
234
|
Sat Apr 17 11:27:38 2021 |
JS, TD | Saturday 17th April 12:00-16:00 | Base Current Difference
aida05 fault 0x1a52 : 0x1a53 : 1
aida06 fault 0x4f3e : 0x4f3f : 1
|
10x |
233
|
Sat Apr 17 07:28:02 2021 |
LPG, TD | Saturday 08:00 - 12:00 | 08:30 CEST
HV and leakage currents: elog:233/1
|
9x |
232
|
Fri Apr 16 23:13:57 2021 |
CA, MA, BA | Saturday 16th 00:00 - 08:00 | 00:05 DESPEC tuning beam, changing spill length
00:21 DAQ continues ok
|
14x |
231
|
Fri Apr 16 15:52:08 2021 |
DSJ | 16 April 16.00 shift | Clock status test result: Passed 12, Failed
0
|
12x |
230
|
Fri Apr 16 11:42:45 2021 |
OH | Shifter checklist | Every 30 minutes:
- Reload statistics (Web browser tab screen |
|
229
|
Fri Apr 16 07:19:00 2021 |
OH, MA | Friday 16th April 08:00-16:00 | 08:19 System wide checks all ok
Baselines reset on WR and FPGA checks
for the start of the day
|
14x |
228
|
Fri Apr 16 05:13:09 2021 |
ML | system wide checks | system wide checks:
Done around 4am.
################################# |
16x |
227
|
Fri Apr 16 02:15:12 2021 |
NH | Implants/Punchthrough | Rate with beam < 1000 Hz still
Not Uranium but some fragments possibly produced |
|
225
|
Fri Apr 16 00:47:36 2021 |
OH-ML | AIDA DAQ Reset | 1:18 (CET) No rate in AIDA 07.
Oscar did a AIDA DAQ reset and
a quick wide check. |
|
224
|
Thu Apr 15 22:38:18 2021 |
OH | Thursday 15th April - 16th April | 23:38 They are testing the finger detector
It is a lot of U they are putting into
it which is passing through to AIDA ~3500Hz
|
|
223
|
Thu Apr 15 13:28:07 2021 |
TD | Thursday 15 April |
13.28 System wide checks
|
19x |
222
|
Thu Apr 15 12:23:22 2021 |
NH | Fast Discriminator Masks | It is possible to get good rates with the
fast discriminators with a few hot channels
mask in aida03 and aida10
|
|
221
|
Wed Apr 14 15:14:55 2021 |
PJCS | Note Well | Please note that for the forseeable future
the system function in the menu on the System
Wide Checks browser page labelled as |
|
220
|
Wed Apr 14 12:14:47 2021 |
PJCS | Corrections and changes | 14/4/21 @12:00 UK time
Corrected the failure of FADC re-calibration
All modules. The problem was a missing > |
|
219
|
Wed Apr 14 09:55:12 2021 |
TD | Tuesday 13 April |
Two further comments
|
|
218
|
Tue Apr 13 16:49:41 2021 |
TD | Tuesday 13 April |
Rename Options directory
|
8x |
217
|
Tue Apr 13 14:24:48 2021 |
TD | S460 information | aida-gsi Anydesk 897170655
DESPEC remote working - https://sf.gsi.de/d/a4fb2134e06a450ca777/
|
|
216
|
Mon Apr 12 18:36:31 2021 |
NH | [HowTo] AIDA & MBS | Some instructions on AIDA and MBS and troubleshooting
are available here
https://sf.gsi.de/lib/c9ba0b4c-26bc-43d0-9b49-213fc594610f/file/AIDA%20and%20MBS%20Tips%20and%20Troubleshooting.pdf |
|
215
|
Mon Apr 12 16:49:40 2021 |
TD | Monday 22 April - pulser walkthrough & alpha background | 16.54 Pulser walkthrough
ASIC settings 2019Dec19-16.19.51
|
6x |
214
|
Sat Apr 10 12:15:54 2021 |
OH, TD | Saturday 10 April | 12:37 DAQ found crashed. Had crashed at some
time following 7:50am following previous
statistics update
|
16x |
213
|
Fri Apr 9 16:45:29 2021 |
NH, HA. GA, GZ, OH, TD | AIDA Noise | AIDA snout was reconnected and we saw originally
a very good reduction in noise from S452
(fig 1)
|
6x |
212
|
Fri Apr 9 16:40:36 2021 |
NH | nnrpi1 Update | nnrpi1 is now back in S4 with a new SD card
and new raspberry pi.
The freezing originally continued (it was |
|
211
|
Wed Mar 31 15:46:09 2021 |
PJCS | HowTo : Synchronize the ASIC clocks. | To Synchronize the FEE64 ASIC clocks to
rise at the same timestamp time use the System
function Synchronise the ASIC clocks |
|
210
|
Wed Mar 31 15:40:42 2021 |
PJCS | Check Options files function added | There is a new operation available in the
System Functions menu in the System Wide
Checks page. ( The page needs to be reset |
|
209
|
Wed Mar 31 12:46:10 2021 |
PJCS | HowTo : Calibrate the LMK3200 clock devices | Here is a document showing how and where
to calibrate the LMK3200 devices that lock
to the system clock and generate the clocks |
|
208
|
Tue Mar 30 13:21:50 2021 |
NH | nnrpi1 (tty logs) | We had issues with all TTYs reading out on
nnrpi1, I have moved them from an apparently
faulty USB hub to a new one.
|
|
207
|
Thu Mar 25 18:04:51 2021 |
TD | MSL type BB18(DS) 24cm x 8cm cabling & bias configuration | |
|
206
|
Wed Mar 24 10:31:03 2021 |
TD | AIDA to do (or to understand) list |
In no particular order
|
|
205
|
Fri Mar 19 17:16:45 2021 |
TD | Friday 19 March |
Screenshots of successive displays of Options
for FEE64s aida01, 02, 03 and 04. |
8x |
204
|
Thu Mar 18 11:53:10 2021 |
TD | Thursday 18 March |
12.40 The snout w/bPlast plus 3xDSSSDs was
removed yesterday
|
7x |
203
|
Wed Mar 17 10:31:09 2021 |
TD | Wednesday 17 March | 11.28 DAQ continues OK - no storage
ASIC settings 2019Dec19-16.19.51
|
11x |
202
|
Tue Mar 16 09:26:47 2021 |
TD | Tuesday 16 March | 10.25 DAQ continues OK - no storage
ASIC settings 2019Dec19-16.19.51
|
6x |
201
|
Mon Mar 15 11:47:51 2021 |
TD | Monday 15 March | 12.48 DAQ continues OK - no storage
ASIC settings 2019Dec19-16.19.51
|
8x |
200
|
Sun Mar 14 17:35:05 2021 |
TD, OH | Sunday 14 March - Pulser Walkthrough | 18.01 Pulser walkthrough at end of run S452
file S452/R57_0 ... R57_28
|
   |
199
|
Sun Mar 14 08:16:00 2021 |
OH PJCS | Sunday 14th March 2021 | > 08:07 Realised the correlation scalers hadn't
been enabled again. Have enabled them now.
R56_151
|
|
198
|
Sun Mar 14 07:04:30 2021 |
OH | Sunday 14th March 2021 | 08:07 Realised the correlation scalers hadn't
been enabled again. Have enabled them now.
R56_151
|
20x |
197
|
Sat Mar 13 23:08:08 2021 |
CA | 14th March 00:00-08:00 | 00:08 DAQ continues ok - writing to file R46_221
ASIC settings 2019Dec19-16.19.51
|
22x |
196
|
Sat Mar 13 17:00:49 2021 |
TD | Saturday 13 March 18.00-00.00 UTC+1 | 18.00 DAQ continues OK - file R46_64
ASIC settings 2019Dec19-16.19.51
|
9x |
195
|
Sat Mar 13 07:08:52 2021 |
CA, TD, LS | March 13th 08:00 - 17:00 | ASIC settings 2019Dec19-16.19.51
DSSSD#1 slow comparator 0xa
DSSSD#2 slow comparator 0xa
|
20x |
194
|
Fri Mar 12 23:56:00 2021 |
OH | Saturday March 13th 00:00-08:00 | 24:00 Joined in with TD to debug the dropout
of AIDA05 - See https://elog.ph.ed.ac.uk/DESPEC/193
After both powercycles and telnet reboots |
11x |
193
|
Fri Mar 12 20:12:10 2021 |
TD | Friday 12 March | 21.12 DAQ continues OK - file R33_617
ASIC settings 2019Dec19-16.19.51
|
    |
192
|
Fri Mar 12 10:09:19 2021 |
LS, CA | Friday 12th March 11.00- | 11.00(Germany) System wide checks okay except:
FEE64 module aida09 global clocks failed, |
17x |
191
|
Fri Mar 12 08:32:46 2021 |
TD | Friday 12 March 08.00- | 08.31 DAQ continues OK - file R33_263
ASIC settings 2019Dec19-16.19.51
|
    |
190
|
Thu Mar 11 23:07:23 2021 |
CA | March 12th 00:00 - 08:00 shift | 00:10 DAQ continues OK - file R30_372
ASIC settings 2019Dec19-16.19.51
|
15x |
189
|
Thu Mar 11 07:09:25 2021 |
OH, LS | Thursday 11th of March | 08:00 System wide checks all ok *exctept aida07
and 10 on ADC calibration
N.B. wrong date given to the screenshots |
29x |
188
|
Thu Mar 11 00:38:33 2021 |
TD | Thusrday 11 March 00.00-08.00 | 01.38 DAQ continues OK - file R22_388
ASIC settings 2019Dec19-16.19.51
|
11x |
187
|
Wed Mar 10 07:16:02 2021 |
OH, LS | Wednesday 10th March 08:00-24:00 | 08:16 System wide checks all ok
Statistics - attachment 1
Temperatures - attachment 2
|
39x |
186
|
Tue Mar 9 23:12:17 2021 |
CA | March 10th 00:00 - 08:00 | 00:13 beam back, but some spills 'missing'
DESPEC keeping beam as is, but may |
20x |
185
|
Tue Mar 9 20:04:20 2021 |
TD | Tuesday 9 March | 21.04 DAQ continues OK - file R20_379
ASIC settings 2019Dec19-16.19.51
|
7x |
184
|
Tue Mar 9 10:23:45 2021 |
TD | Analysis of R14 (w/ NewMerger with min info code 4 & 5 data items) |
Per https://elog.ph.ed.ac.uk/DESPEC/183 R14
data files have been analysed
|
 |
183
|
Tue Mar 9 06:58:14 2021 |
CA, LS | March 9th 08:00-11:00 shift | 08:00 DAQ continues ok - writing to file R13_120
Merger ok - 4.6M data items/s
|
23x |
182
|
Mon Mar 8 22:51:47 2021 |
OH | Tuesday 9th March | 18:00 ASIC settings 2019Dec19-16.19.51
DSSSD#1 slow comparator 0xa
DSSSD#2 slow comparator 0xa
|
10x |
181
|
Mon Mar 8 17:02:38 2021 |
CA | Monday 8th March 18:00 - 00:00 | 18:00 ASIC settings 2019Dec19-16.19.51
DSSSD#1 slow comparator 0xa
DSSSD#2 slow comparator 0xa
|
12x |
180
|
Mon Mar 8 09:45:20 2021 |
CA, TD | Analysis R7_20 (new version of NewMerger with min info code 4 & 5 data items) |
An analysis of file R6_70 can be found at
https://elog.ph.ed.ac.uk/DESPEC/177 attachment |
 |
179
|
Mon Mar 8 07:02:17 2021 |
CA, LS | Monday 8th March 08:00 - 17:00 | 08:00 ASIC settings 2019Dec19-16.19.51
DSSSD#1 slow comparator 0xa
DSSSD#2 slow comparator 0xa
|
24x |
178
|
Sun Mar 7 23:13:50 2021 |
OH | Monday 8th March | 00:00 ASIC settings 2019Dec19-16.19.51
DSSSD#1 slow comparator 0xa
DSSSD#2 slow comparator 0xa
|
13x |
177
|
Sun Mar 7 07:23:26 2021 |
CA, LS | March 7th | 08:00 ASIC settings 2019Dec19-16.19.51
DSSSD#1 slow comparator 0xa
DSSSD#2 slow comparator 0xa
|
14x |
176
|
Sat Mar 6 23:37:22 2021 |
TD | Sunday 7 March 00.00-08:00 |
00.30 UNILAC problem - no beam
|
18x |
175
|
Sat Mar 6 06:52:06 2021 |
OH, LS | Saturday 6th March | 07:50 Pulser settings - attachment 1
07:52 System wide checks all ok
|
30x |
174
|
Fri Mar 5 23:45:41 2021 |
TD | S452 information |
aida-gsi Anydesk 897170655
|
 |
173
|
Fri Mar 5 23:21:59 2021 |
CA, OH | March 6th 2021 00:00 - 08:00 shift | 00:23 206Hg setting - No beam in AIDA yet
AIDA not writing to file
|
18x |
172
|
Fri Mar 5 22:53:05 2021 |
OH | Thursday 5th of March | There has been no beam for much of the day.
It was lost at 13:00 and not regained until
22:00 German time.
|
|
171
|
Thu Mar 4 11:16:13 2021 |
OH | Further cases of possible database issues | During testing for the S452 experiment yesterday
we were running through the steps of a restart.
|
 |
170
|
Thu Mar 4 11:03:04 2021 |
NH | S452 Operating Notes | Please note the following "quirks" currently
present in S452 that are important
|
|
169
|
Thu Mar 4 10:13:48 2021 |
OH | AnyDesk connection instuction | |
|
168
|
Wed Mar 3 15:02:19 2021 |
CA | MBS command line startup | /MIDAS/Linux/startup/datarelaymbs |
|
167
|
Sat Feb 27 12:47:57 2021 |
NH | AIDA Rates/Situation | Updated figures as of 27 Feb 2021 - S4 is
now closed for experiments but may open on
Wednesday.
|
10x |
166
|
Mon Feb 22 18:18:47 2021 |
NH | AIDA Rates/Situation | The ground cable (LEMO00) in aida07 was disconnected,
it has been reinserted and the waveforms
now look good.
|
8x |
165
|
Fri Feb 12 17:31:04 2021 |
NH | Background run | A background run has been started to check
the AIDA channels.
It is in /TapeData/BGFEB21
|
|
164
|
Sat Feb 6 10:45:55 2021 |
NH | AIDA Status Feb 2021 | > 01.02.2021 - Four new sets of PSU, HDMI
& Ethernet cables were ran from the DESPEC
rack to the AIDA frame, in preparation for |
   |
163
|
Mon Feb 1 15:19:07 2021 |
NH | AIDA Status Feb 2021 | 01.02.2021 - Four new sets of PSU, HDMI &
Ethernet cables were ran from the DESPEC
rack to the AIDA frame, in preparation for |
|
162
|
Tue Nov 24 13:10:42 2020 |
NH | Waveforms with frame grounded | AIDA Frame is now grounded to platform (will
add photo later)
|
   |
161
|
Tue Oct 6 15:48:06 2020 |
NH | nnrpi kernel update | FYI:
The SD card on nnrpi2 was full again causing |
|
160
|
Tue Jul 7 08:50:27 2020 |
NH | nnrpi2 does not boot | MIDAS was not working properly on nnrpi2 -
power cycled. Now system does not boot properly,
it gets stuck trying to login claiming "PAM |
|
159
|
Tue Jul 7 08:04:57 2020 |
NH | Power failure in Messhuette | Last week there was a power failure in Messhuette
(FRS control room) - AIDA-3 and MBS were
affected.
|
|
157
|
Wed Mar 18 18:02:37 2020 |
PJCS | New Firmware loaded in all FEE64s | Loaded new frimware into all 12 FEE64s
in the system. IWR_Dec19_4.bin
Booted ok after power-cycle. |
|
156
|
Mon Mar 16 07:25:24 2020 |
NH | End of Experiment | 08:25 Beam Stopped, end of S480
Tape Server stopped at R12_2097
Started again in no storage mode |
|
155
|
Mon Mar 16 05:08:12 2020 |
TA, AM, ES | Regular entry |
02:14 all system wide checks okay
FEE temperatures okay
leakage
currents okay and recorded to spreadsheet |
6x |
154
|
Mon Mar 16 01:07:29 2020 |
TA, AM, ES | Regular entry |
02:14 all system wide checks okay
FEE temperatures okay
leakage
currents okay and recorded to spreadsheet |
13x |
153
|
Sun Mar 15 06:45:00 2020 |
CA, LS, CG | Sunday 15th March 08:00 - 24:00 | ASIC settings 2019Oct31-13.24.23
slow comparator 0xa
|
35x |
152
|
Sat Mar 14 23:09:28 2020 |
MB, SJ, MR | Sunday 15th March 0:00-8:00 | AIDA ELog filled out by DESPEC night shift
members Matthew Brunet (MB), Shaheen Jazrawi
(SJ), and Matthias Rudiger (MR). |
16x |
151
|
Sat Mar 14 07:13:59 2020 |
DK, TD, LS, CG | Saturday 14th March 08:00 - 24:00 | 8:15
All system wide checks passed
Leak currents are continuing to creep |
41x |
150
|
Fri Mar 13 23:06:39 2020 |
CA, OH | Saturday 14th March 00:00 - 08:00 | ASIC settings 2019Oct31-13.24.23
slow comparator 0xa
|
19x |
149
|
Fri Mar 13 15:48:11 2020 |
Friday 13th 16:00 - 24:00 | OH, TD | ASIC settings 2019Oct31-13.24.23
slow comparator 0xa
|
11x |
148
|
Fri Mar 13 08:01:53 2020 |
DK | Friday 13th 8:00 to 16:00 shift | 8:00 - See previous shift's elog entry. All
good.
|
8x |
147
|
Thu Mar 12 23:44:48 2020 |
DK | PID Plots | Labelled PID plots. Note that the A/Q calibration
is shifted by +0.02 units, and Z calibration
is offset by about +5.
|
  |
146
|
Thu Mar 12 23:04:24 2020 |
LS, DK | Friday 13th 00:00 to 08:00 shift | Shift file starting R9_639
system wide checks okay
|
20x |
145
|
Thu Mar 12 15:25:08 2020 |
Thursday, March 12 16:00 to midnight shift | CA, DK, LS | 16:30
All system wide checks passed
Temps shown in attachment #1
|
12x |
144
|
Thu Mar 12 08:04:24 2020 |
OH, CA, TD | Thursday 12th March 08:00-16:00 | ASIC settings 2019Oct31-13.24.23
slow comparator 0xa
|
14x |
143
|
Wed Mar 11 23:44:27 2020 |
OH | Thursday 12th March 00:00-08:000 | ASIC settings 2019Oct31-13.24.23
slow comparator 0xa
|
10x |
142
|
Wed Mar 11 17:21:33 2020 |
CA, TD, OH | Wednesday 11 March 16.00-00.00 | 18.25 DAQ currently running No Storage mode
ASIC settings 2019Oct31-13.24.23
|
23x |
141
|
Wed Mar 11 16:29:17 2020 |
CA, LS, TD | slow comparator threshold comparison | 17.29 good event statistics for slow comparator
threshold at 100 keV (0xa) - attachment 1
|
   |
140
|
Tue Mar 10 23:17:31 2020 |
DK, LS, PW, TD, CA | Wednesday 11 March 2020 | 00:15 Shift change near AIDA R6_475
We continue with spill cycles around 0.2 |
43x |
139
|
Tue Mar 10 12:36:44 2020 |
TD | [How To] Monitor AIDA | All AIDA operations are controlled from the
Dell T410 Workstation aidas1 using web browsers.
|
|
138
|
Tue Mar 10 10:16:56 2020 |
OH | AIDA Implant Range Scan | A range scan was performed using AIDA.
See the GSI elog: https://elog.gsi.de/despec/S480/19 |
|
137
|
Tue Mar 10 08:48:07 2020 |
TD, OH, NH, LS, CA, DK, PJW, RP | Tuesday 10 March 08.00-16.00 |
c 09.00 aida03 drops Histogram enable at
DAQ GO
|
41x |
136
|
Mon Mar 9 13:50:08 2020 |
LS, CA, DK | 09/03/2020 system checks | 14:52 Attachment 1 shows low energy spectra.
AIDA06 shows strange baseline, possible double-hit?
|
13x |
135
|
Mon Mar 9 12:47:18 2020 |
NH | Report: Unusual AIDA Timestamp Error | Noted using my AIDA event builder code, a
timewarp was reported and I noticed the ADC
event had a different timestamp LSB to the |
|
134
|
Sat Feb 29 10:56:38 2020 |
NH | Alpha Hit Pattern | Circa 700 GB of data
A few channels in DSSD 1 missing? Otherwise |
|
133
|
Tue Feb 25 09:55:59 2020 |
NH | HV Scope Traces | Looking at the HV bias in a scope to see if
it is the source of noise especially on the
p+n side
|
10x |
131
|
Fri Feb 21 12:16:06 2020 |
TD, NH | Report - high - unusual FEE64 crash during startup | |
|
130
|
Fri Feb 21 09:39:38 2020 |
TD, NH | Friday 21 February |
Attachment 1 - layout 2 - stat spectra
Attachment 2 - detector biases & leakage |
7x |
129
|
Fri Feb 21 09:21:10 2020 |
NH | (Early) Alpha Analysis | Overnight alpha data
======================================
|
    |
128
|
Thu Feb 20 16:35:55 2020 |
TD, NH | Thu 20 February 2020 | DSSD#3 (most downstream) was replaced
Noted that old DSSD wafer was not sitting |
12x |
127
|
Wed Feb 19 10:58:29 2020 |
NH, PJCS | Report: aida09 Kernel Panic & Lost WR | > aida09 crashed over the weekend and automatically
rebooted.
> After the reboot the WR timestamp sent |
|
126
|
Mon Feb 17 09:48:34 2020 |
NH | Report: aida09 Kernel Panic & Lost WR | aida09 crashed over the weekend and automatically
rebooted.
After the reboot the WR timestamp sent to |
    |
125
|
Fri Feb 14 12:50:29 2020 |
NH | AIDA Noise 14.02.2020 | Electricians changed the grounding for the
DESPEC platform to a large pillar that goes
deep into soil
|
6x |
124
|
Wed Jan 29 19:09:51 2020 |
NH | aida10 troubleshooting | > aida10 has no ASIC events when running without
pulser
>
|
|
122
|
Wed Jan 29 12:22:13 2020 |
NH | aida10 troubleshooting | aida10 has no ASIC events when running without
pulser
|
  |
121
|
Mon Jan 20 14:01:23 2020 |
NH | AIDA Noise 20.01.2019 | > Checking noise
>
> DESPEC crate entirely offline except:
|
6x |
120
|
Wed Jan 15 14:53:41 2020 |
NH | AIDA Noise 14.01.2019 | Checking noise
DESPEC crate entirely offline except:
|
6x |
119
|
Fri Dec 20 12:04:03 2019 |
NH, TD | Friday 20 December 2019 | Fig 1: DSSD Leakage currents ok
Fig 2: FEE Temps ok
|
7x |
118
|
Fri Dec 20 10:19:29 2019 |
NH | Alpha Analysis (Pre Run) | ======================================
|
 |
116
|
Thu Dec 19 14:45:42 2019 |
TD | FEE64 ASIC settings - different diode link settings | [npg@aidas-gsi 2019Oct31-13.24.23]$ pwd
/MIDAS/DB/EXPERIMENTS/AIDA/2019Oct31-13.24.23
[npg@aidas-gsi 2019Oct31-13.24.23]$ /bin/csh
|
|
115
|
Thu Dec 19 14:26:09 2019 |
TD | BNC PB-5 - Manual | |
|
114
|
Thu Dec 19 14:07:38 2019 |
TD | Report - low - aida08 ADC spectrum size 32k cf. 64k channels expcted using layout | Note that aida08 spectrum 1.8.L size is (shown
as) 32k channels cf. 64k channels expected
using layout - attachment 1 - 5
|
6x |
113
|
Thu Dec 19 13:09:32 2019 |
TD | MIDAS app launcher command lines |
App Path
|
|
112
|
Thu Dec 19 13:06:02 2019 |
TD, NH | AIDA@DESPEC To Do list | FAIR phase 0 DESPEC experiments are scheduled
March-May 2020 see
|
|
111
|
Thu Dec 19 10:11:59 2019 |
TD | Report - medium - FEE64 panics during boot |
Some of the FEE64s aida01 .. aida12 panic
during boot
|
|
110
|
Thu Dec 19 09:22:52 2019 |
TD | Thusrday 19 December | 10.15 Found DAQ running at c. 1M data items/s
|
21x |
109
|
Wed Dec 18 17:39:27 2019 |
TD | Wednesday 18 December |
18.40 Detector biases & leakage currents
OK - attachment 1
|
23x |
108
|
Wed Dec 18 02:44:27 2019 |
NH | 17.12.19, 18.12.19 | Beam: 238U (850 MeV/u)
Fragment: 168Re
|
7x |
107
|
Mon Dec 16 14:48:02 2019 |
NH TD | AIDA 16.12.19 | DSSD Snout was reassembled to install the
second bPlast (downstream)
(fig 1)
|
    |
106
|
Wed Dec 11 10:58:52 2019 |
NH | Initital Dec. Analysis | Fig 1: 2D hit pattern of each AIDA DSSD for
MBS run 24 (40Ar primary beam, centred)
|
11x |
105
|
Sat Dec 7 02:39:02 2019 |
DK, NH et al. | 34Si Fragments PID | See attached
Tentative PID (DK) as attachment #3
|
   |
104
|
Fri Dec 6 20:34:56 2019 |
CA, DK, NH | Dec 6 Night Shift | 21.35 AIDA DAQ currently running, TapeService
on, not writing to storage
|
15x |
103
|
Thu Dec 5 22:00:11 2019 |
CA, DK, NH | Dec 5 Night Shift | 15:30
TD found the merger was running at ~43
Mb/s
|
30x |
102
|
Wed Nov 27 16:33:50 2019 |
NH | New Bias & Waveforms | HV core (-160 V) now attached to bottom
3 FEEs (with grounded kapton)
DSSDs bias OK, leakages unchanged |
 |
101
|
Mon Nov 25 13:25:51 2019 |
NH | Waveforms | > AIDA Waveforms immediately after beam at
21.11.19 8am or so.
|
 |
100
|
Fri Nov 22 08:55:31 2019 |
NH | Waveforms | AIDA Waveforms immediately after beam at 21.11.19
8am or so. |
   |
99
|
Wed Nov 20 20:36:14 2019 |
CA, CB, DK, NH | 20/21 November Night Shift | 21:35 (NH) - A few merger timestamp errors
were appearing - Decided to powercycle all
FEE64s, all checks OK.
|
17x |
98
|
Tue Nov 19 21:22:06 2019 |
CA, CB, DK, NH | 19/20 November 2019 - overnight shift | 19 November
Waiting for 40Ar beam to be delivered for |
15x |
97
|
Wed Nov 13 15:38:20 2019 |
NH | [HowTo] MBS for AIDA (Nov 19) | For MBS integration AIDA forwards its data
to an MBS Foreign Data Receiver (FDR), currently
the PC assigned for this task is x86l-94.
|
|
96
|
Wed Nov 13 09:10:37 2019 |
CA, NH, OH | 13th November 2019 | 10:10 moved HV cables from 11 and 12 to 10
and 9 - all detectors biased in same configuration
|
   |
95
|
Wed Nov 13 01:54:00 2019 |
TD | Aida10 Unusual Pulser Peak | Double peaking normally indicates periodic
baseline variations with period comparable
to shaping time |
|
94
|
Tue Nov 12 14:10:18 2019 |
NH, OH, CA | Aida10 Unusual Pulser Peak | aida10 (no alpha rate recorded) has a strange
pulser peak - looks like double peak?
Pulser terminator checked and OK... |
  |
93
|
Tue Nov 12 13:59:07 2019 |
NH, OH, CA | Awfully noisy waveforms | Rates in all FEEs much higher than before
- waveform shows very noisy in all systems.
- attachment 1
|
8x |
92
|
Tue Nov 12 13:22:19 2019 |
OH, NH, CA | AIDA09 Replacement | ASIC 1 on FEE9 was not producing signals.
A new HDMI cable was installed but still |
|
91
|
Mon Nov 11 10:34:19 2019 |
NH | Alpha Analysis (11.11.19) | Analysis of approx 20GB of data since Thursday
night (not continuous - estimate of hours
incominG) |
9x |
90
|
Thu Nov 7 10:22:26 2019 |
Nic, Patrick | Reply: WR Timestamps | > > > All 12 FEEs have valid WR Timestamps
> > > Had to powercycle aida09 once as before
raw readout was displaying upper 12 bits |
|
89
|
Tue Nov 5 11:15:41 2019 |
NH PJCS | Report: aida10 database corruption | If ypou now check the actual values in
the aida10 Options screen are the correct
? |
|
88
|
Tue Nov 5 10:32:21 2019 |
NH PJCS | Report: aida10 database corruption | Hi, see the attached files in the original
comment. I think the BACKUPCorrupt is quite
old now. |
|
87
|
Tue Nov 5 10:02:45 2019 |
NH PJCS | Report: aida10 database corruption | This looks like the Options not the ASIC
data. That will be in the CONTENTS file of
the FEE64 named directory within the dated |
|
86
|
Mon Nov 4 13:17:12 2019 |
NH | Report: aida10 database corruption | Often aida10 database seems to get corrupted
(ASIC values are all 0xad)
Unsure why it's only aida10, might |
 |
85
|
Mon Nov 4 08:48:18 2019 |
NH | Alpha Status 04.11.2019 | Alpha run has been running over the weekend.
Stats looking positive with notable exception
that aida10 is not sending much data. Will |
    |
84
|
Fri Nov 1 18:09:03 2019 |
CA, TD, NH | Summary - 29.10-1.11.19 | > >
> >
> > - evidence of c. 1MHz extrinsic noise |
|
83
|
Fri Nov 1 15:45:04 2019 |
CA, TD, NH | Summary - 29.10-1.11.19 | >
>
> - evidence of c. 1MHz extrinsic noise for |
|
82
|
Fri Nov 1 15:24:40 2019 |
CA, TD, NH | Summary - 29.10-1.11.19 |
3x MSL type BB18(DS)-1000 installed
|
|
81
|
Fri Nov 1 14:17:15 2019 |
Nic, Patrick | Reply: WR Timestamps | > > All 12 FEEs have valid WR Timestamps
> > Had to powercycle aida09 once as before
raw readout was displaying upper 12 bits |
|
80
|
Fri Nov 1 10:46:13 2019 |
CA, TD, NH | Friday 1st November 2019 | 11.46 - attachments 1,2,3: bias/leakage currents,
good event statistics and fee temperatures
(respectively)
|
6x |
79
|
Fri Nov 1 09:30:17 2019 |
NH | Database error stopping run? | Strange error about Database options appearing
during run stop.
Does not seem to affect stopping/starting/data? |
|
78
|
Thu Oct 31 19:02:42 2019 |
Patrick | WR Timestamps | > All 12 FEEs have valid WR Timestamps
> Had to powercycle aida09 once as before
raw readout was displaying upper 12 bits |
|
77
|
Thu Oct 31 17:13:54 2019 |
NH | Alpha Run | Beginning Alpha run
Thresholds set to 100 (LEC/Slow) 255 (LEC/Fast) |
|
76
|
Thu Oct 31 16:43:03 2019 |
NH CA TD | Merger & MBS Performance | Testing merger with 3 DSSDs connected, with
thresholds of 10 (slow comparator), 2 (HEC)
and 255 (fast discriminator)
|
   |
75
|
Thu Oct 31 15:24:35 2019 |
TD | waveform spectra issues? | |
12x |
74
|
Thu Oct 31 14:27:06 2019 |
NH | WR Timestamps | All 12 FEEs have valid WR Timestamps
Had to powercycle aida09 once as before raw
readout was displaying upper 12 bits of WR |
|
73
|
Thu Oct 31 14:23:37 2019 |
TD | aida09 asic 1 no data readout? | |
   |
72
|
Thu Oct 31 10:33:34 2019 |
TD | [How To] Start TclHttpd server on Raspberry Pi | 1) Login to Raspberry Pi
ssh pi@nnrpi2
|
|
71
|
Thu Oct 31 09:36:37 2019 |
TD | RIKEN LayOut directory | |
|
70
|
Thu Oct 31 09:00:58 2019 |
CA, TD, NH | Thursday 31 October | 10.05 DSSSD # 1-3 detector biases & leakage
currents OK - see attachment 1
Ambient temperature +21.2 deg C, d.p. |
14x |
69
|
Wed Oct 30 11:51:08 2019 |
CA, TD, NH | Detector biasing | 12.51 - detector bias and leakage currents
- detector 3 has no current
|
  |
68
|
Wed Oct 30 08:06:49 2019 |
CA, TD, NH | DSSSD stack | Ribbon cables fitted with kapton PCB couplers
-> 3 isolated + 1 non-isolated for each cable
length/DSSSD
|
6x |
67
|
Mon Oct 28 17:02:11 2019 |
TD | AIDA @ DESPEC cable length estimates | AIDA snout length 37cm (from end of AIDA snout
support assembly)
|
|
66
|
Mon Sep 16 12:24:46 2019 |
NH | AIDA Monitoring | https://docs.google.com/spreadsheets/d/1fZBybHodSvacUdKpDM_VKznqTrh49CKnqKVQVMEDWGs/edit?usp=sharing
A spreadsheet of ambient conditions
and water temperature in S4 for AIDA. |
|
65
|
Fri Aug 16 15:09:05 2019 |
NH | Aida Waves (All FEEs) | Included waves for all 12 FEEs
FEES 1, 5, 9 all show this HF noise
on it, and are all located at the top of |
12x |
64
|
Thu Aug 15 14:09:39 2019 |
NH | AIDA Waveforms | Noise seemed less strong today, unsure
of cause. I had re-taped snout as tape was
not sticking well. |
10x |
63
|
Wed Aug 14 14:40:18 2019 |
NH | High Frequency Noise in AIDA | Figures of intense high frequency noise
pickup in AIDA.
Will investigate source in S4. |
 |
61
|
Tue Aug 13 13:55:45 2019 |
NH | New Merger/MBS Test Runs | AIDA Finally working again so can test
VP's added WAVE histogramming mode.
Fig 1. Waveforms for FEE12 ASIC |
6x |
60
|
Tue Aug 13 09:17:02 2019 |
NH | AIDA 12/08/2019 | The other 6 FEEs were powered on one at
a time to check no issues. None found.
All 12 FEEs plugged back into the |
|
59
|
Fri Aug 9 12:33:14 2019 |
NH | Aida Test 9/8/19 - FEES 1-6 | The water issue has been resolved - merely
incorrect temperature guages (replaced).
Water is 20 C as required. |
|
58
|
Mon Aug 5 12:57:48 2019 |
NH | S4 Conditions | Temp: 26 C
Humidity: 50%
Dew Point: ~15 C |
|
57
|
Wed Jul 10 10:46:05 2019 |
NH | AIDA/Water 10.07.2019 | S4 Conditions:
25.2 C, 35.3% RH, Td = 8.8 C
Water temperature: 20 C |
|
56
|
Mon Jul 8 16:02:52 2019 |
NH, PJCS | Aida 28/6/2019 - WATER OFF | It occurs to to wonder if there is now
water inside the modules. If the metal of
the cooling plates is damp on the outside |
|
55
|
Mon Jul 1 09:42:16 2019 |
NH | AIDA - 01/07/2019 | Water is still off.
Checked dewpoint sensor still flashing
red light (indicating an error). Pipes warmed. |
|
54
|
Fri Jun 28 15:05:49 2019 |
NH | Aida 28/6/2019 - WATER OFF | The humidity sensor hasn't reported
humdities under 90% yet. In order to try
and help dry the pipes and turn AIDA the |
|
53
|
Tue Jun 4 09:19:10 2019 |
NH | AIDA Interlock | Currently AIDA's interlock is off (no
lights) and hence AIDA cannot be powered
on. Under investigation - power supply has |
|
52
|
Mon May 20 13:51:45 2019 |
NH | HowTo Verify WR Times | The latest version of MIDAS has a new page
to check the WR timestamp of each FEE.
In AIDA Hardware control click: |
  |
51
|
Mon May 20 13:33:06 2019 |
NH | New Merger/MBS Test Runs | New merger has been worked on by VP which
fixes the timewarp issues and MBS integration.
Currently no WAVE capture is supported, |
8x |
50
|
Fri Apr 12 15:15:33 2019 |
NH | Report - FEE Kernel Panics (Update on 48) | Update on issue #48 - the "confusing
state" is that the FEE has restarted
and hence is undefined again. |
|
49
|
Thu Apr 11 15:28:10 2019 |
CA, OH, CB, TD | Offline analysis of 290319 files R13 and R16 | X projection of high energy channel ADC hits
from R13 (attachment 1)
|
6x |
48
|
Wed Apr 10 14:53:50 2019 |
NH | Report - FEE stops sending data | it seems a FEE somtimes enters a confusing
state and stops sending data
the current merger requires all FEEs |
6x |
47
|
Wed Apr 10 12:33:58 2019 |
NH VP | AIDA Firmware Update | All AIDA modules have been updated to a
new firmware 0xd330701 which should fix the
merger issues. |
 |
46
|
Thu Apr 4 14:07:45 2019 |
NH | Pulser Configuration | Pulser settings during April run:
Rate 2Hz
Delay 250 NS |
|
45
|
Wed Apr 3 15:24:06 2019 |
CA, TD | report - low inquire save state | On Save/Restore Module Settings Tab - "inquire
save state" returns an error. See attachment
1. |
|
44
|
Wed Apr 3 13:29:20 2019 |
CA, CB, TD | report - high DAQ stall | aida10 runs very fast for a period, then stops.
Have to restart Merger and TapeServer to |
    |
43
|
Wed Apr 3 09:28:07 2019 |
CA, CB, TD | Wednesday April 3rd 2018 | Arrived at 09:54 - sequenced all FEE64 on
FEE64 not mounted properly - something happened |
11x |
42
|
Tue Apr 2 09:42:20 2019 |
NH | MBS Server setup | MBS is currently configured and accepting
data:
x86l-94 : AIDA Foreign data receiver |
|
41
|
Tue Apr 2 07:39:44 2019 |
CB, TD, CA | Tuesday 2nd April 2019 | 08:15 beam tuning in progress
08.30 System wide checks
|
36x |
40
|
Mon Apr 1 10:20:20 2019 |
CA, NH, TD | MBS Relay startup | To start the MBS Data Relay, click on the
top rightmost of the MIDAS icons (MBS relay,
Data Relay => MBS) in
|
 |
39
|
Mon Apr 1 10:14:09 2019 |
CA, NH, TD | Friday 29th March 2019 | 16:30 good event statistics ok, with 9,10,11,12
running faster (detector FEE64s) (attachment
1)
|
  |
38
|
Mon Apr 1 08:50:35 2019 |
CA, TD | Monday April 1st 2019 - BEAM START | 09.28 System wide checks - aida06 global clock
failure (check clock status)
|
18x |
37
|
Tue Mar 26 14:59:30 2019 |
NH | 26.03.2019 | Fig1: FEE temperatures from 26.03.2019
all OK
AIDA has been on today and merger |
 |
36
|
Tue Mar 26 14:58:09 2019 |
NH, VP | How to reset merger | If the merger doesn't show "xfer
Links => Merger" at the top, it is
probably the merger setting has been corrupted |
|
35
|
Mon Mar 25 14:44:33 2019 |
NH | Merger issues | Plastic electronics have now been installed.
Water is not leaking still.
FEE64s powered on at 15:00 to test |
    |
34
|
Fri Mar 22 15:44:20 2019 |
NH | End of 22.03.2019 | End of the day - Detector unbiased and
FEE64s turned off for the weekend as they
won't be monitored. |
|
33
|
Fri Mar 22 14:31:24 2019 |
NH | A few more setup notes I forgot | Extra FEE information:
HV braid (0 V) goes into FEE10
(beam left) with LK1 jumper |
   |
32
|
Fri Mar 22 13:47:32 2019 |
NH | Resolution Checks | Pulser widths of AIDA modules, resolution
is fairly typical. AIDA09 still noisier than
other DSSD channels. |
  |
31
|
Fri Mar 22 13:32:55 2019 |
NH | AIDA Startup @ 22.03.2019 | FEEs have been on this morning, temperatures
attached
Leakage current attached |
 |
30
|
Thu Mar 21 15:45:48 2019 |
NH, AS, SA, MP | Snout Mounting & DSSD Bias | The AIDA snout (with single DSSD and plastic)
has been mounted in S4 cave again. DSSD was
connected to the outer 4 FEEs again. |
|
29
|
Thu Mar 21 14:10:10 2019 |
NH | FEE Temperatures | All 12 FEEs powered up, and temperatures
are OK.
Passed ASIC checks and system wide |
|
28
|
Thu Mar 21 13:59:08 2019 |
NH | AIDA Water | The AIDA water system has been repaired
and not leaking.
The interlock did not light immediately, |
|
27
|
Wed Mar 20 15:35:10 2019 |
NH | AIDA Setup @ GSI | Currently
- 1 DSSD and Plastic is in the
AIDA snout, which is not mounted |
|
26
|
Sun Feb 3 16:01:49 2019 |
TD | Updated analyses GSIJan19/R1_0 - R10_0 |
Summary statistics for WR test data files
R1_0, R2_0, R3_0, R4_0, R7_0 and R10_0 (attachments |
7x |
25
|
Fri Jan 25 12:33:03 2019 |
NH | aida-gsi Access | If you have a GSI account it should be
possible to connect to aida-gsi remotely
via the following |
|
24
|
Fri Jan 25 12:26:17 2019 |
NH | Raspberry Pi Startup & Information | Two raspberry pis:
nnrpi1 - FEE64 system consoles
nnrpi2 - AC Mains Relay & CAEN |
|
23
|
Fri Jan 25 09:54:12 2019 |
CA, TD, NH, VP | 25th January 2019 | 09.30: AIDA still running, not writing to
file
Still running ok
|
8x |
22
|
Thu Jan 24 17:16:32 2019 |
CA, TD, NH, VP | MACB time switch settings | AIDA@DESPEC MACB time switch settings before
and after WR are as follows.
|
 |
21
|
Thu Jan 24 14:53:26 2019 |
TD | Analyses GSIJan19/R2_0 - R4_0 |
Attachment 1 - summary analysis R2_0
|
   |
20
|
Thu Jan 24 13:29:40 2019 |
TD | Analysis of GSIJan19/R1_0 |
Analysis of file GSIJan19/R1_0 (white rabbit
data)
|
   |
19
|
Thu Jan 24 10:26:09 2019 |
VP, TD, CA, NH | Start up Merger | 1) Click on item "Merger MK2" in top panel
This kills any existing Merger existence
and loads a new copy.
|
|
18
|
Thu Jan 24 10:21:26 2019 |
CA, TD, NH, VP | January 24th 2019 | 10.30 AIDA setup and DAQ start complete
writing to file TapeData/NULL/R1
|
7x |
17
|
Wed Jan 23 17:03:08 2019 |
CA, TD, NH, VP | January 23rd 2019 | Post White Rabbit update
Attachment 1 - 1.8.W pulser waveform spectra |
6x |
16
|
Wed Jan 23 09:25:55 2019 |
CA, TD, NH, VP | MACB and setup pics 23.01.19 | Pictures of leaking cooling pipe
MACB switch settings
|
11x |
15
|
Wed Jan 23 08:12:01 2019 |
CA, TD, NH, VP | January 22nd 2019 | Attachment 1 - 1.8.W waveform spectra with
pulser signal
|
   |
14
|
Mon Dec 10 11:09:53 2018 |
CA, TD | Scintillator pictures/system checks for 07.12.18 | 11.12 System wide checks ok (attachment 1)
FEE temperatures ok (attachment 2)
bias/leakage currents ok (attachment |
14x |
13
|
Sat Dec 8 15:22:52 2018 |
TD | Raspberry Pi startup | nnrpi1 - FEE64 system consoles
nnrpi2 - USB-controlled ac mains relay, CAEN |
|
12
|
Thu Dec 6 20:45:47 2018 |
TD | Analysis of R2_0 |
Analysis of data file R2_0
|
 |
11
|
Thu Dec 6 15:31:40 2018 |
TD, CA, NH | December 6th 2018 | 15.32 1.8.W spectra for aida01, aida09, aida10,
aida11, aida12 (attachments 1, 2, 3)
2 and 3 expanded in Y and X direction |
11x |
10
|
Sat Dec 1 16:43:47 2018 |
TD | Analysis of WRTest28Nov18.dat |
Attachment 1 - analyser.f
|
  |
9
|
Sat Oct 13 17:26:53 2018 |
TD | Analysis of 3 October 2018 T4 tests | |
7x |
8
|
Thu Sep 20 09:39:17 2018 |
TD | AIDA downstream window |
The AIDA downstream window ('endcap' of snout)
consists of 2x 50um aluminized mylar sheets |
|
7
|
Thu Sep 20 09:16:31 2018 |
TD | Images of AIDA setup for DESPEC | Attachment 1 - detail of MSL type BB18(DS)-1000
p+n junction side strips, bias ring, multiple
guard rings, filed
|
9x |
6
|
Thu Sep 20 08:57:48 2018 |
OH | Thursday 20th September | 09:57 Alpha run stoppped R2_0 -> R2_2
Thresholds dropped from 0x64 to 0xa
|
   |
5
|
Wed Sep 19 12:40:41 2018 |
OH, TD | Single DSSD Installation and Tests | On the 18th of September we installed DSSD
3208-14
|
21x |
4
|
Thu Sep 13 17:21:59 2018 |
TD | AIDA upstream window |
The upstream window of the AIDA detector
assembly consists of 2x 50um aluminized mylar |
|
3
|
Thu Sep 6 15:42:53 2018 |
OH, TD | FEE64 system console logs |
Kernel panics during boot cab be observed
in files ttyUSB1 and ttyUSB5
|
|
2
|
Thu Sep 6 15:25:19 2018 |
OH, TD | AIDA@DESPEC Setup Photos | |
|
1
|
Thu Sep 6 15:11:17 2018 |
OH, TD | Tuesday 4 September - Thursday 6 September | Attachment 1 - FEE64 temperatures illustrating
(we assume) a faulty PSU temperature sensor
for aida07
|
19x |